BS-360E&BS-370E&BS-360S - Operation Manual - V7.0 - EN
BS-360E&BS-370E&BS-360S - Operation Manual - V7.0 - EN
BS-360E&BS-370E&BS-360S - Operation Manual - V7.0 - EN
Chemistry Analyzer
Operator’s Manual
© 2017-2022 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator's Manual, the issue date is 2022-12.
, , , , BeneView, WATO,
BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other
countries. All other trademarks that appear in this manual are used only for informational or
editorial purposes. They are the property of their respective owners.
All information contained in this manual is believed to be correct. Mindray shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only
if:
all installation operations, expansions, changes, modifications and repairs of this product are
conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and local
requirements; and
the product is used in accordance with the instructions for use.
Warning
It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.
Note
This equipment must be operated by skilled/trained clinical professionals.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the use of parts or accessories not approved by
Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
EC - Representative
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Address: Eiffestraβe 80, 20537 Hamburg, Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726
Preface
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Please read this manual thoroughly before using
the product. Observance of this manual is a prerequisite for proper performance and correct
operation, and it ensures patient and operator safety. All graphics including screens and
printouts in this manual are for illustration purpose only and must not be used for any other
purposes. The screens and printouts on the actual product should prevail.
Intended audience
This manual is intended for medical laboratory professionals to do the following:
Learn about the system hardware and software.
Perform daily operating tasks.
Maintain and troubleshoot the system.
Related documents
The following documents are provided for searching information about the instrument:
Operator's Manual
Contains instrument compositions, operating instructions, maintenance and troubleshooting
methods. At the beginning of this manual is a table of contents, which provides references of all
chapters for searching desired information. At the end of this manual is a glossary and index,
which provide term definitions and index to key words.
This manual is based on the maximum configuration and therefore some contents may not
apply to your product. It you have any questions, please contact us.
Operation Card
Provides daily operating procedure for quickly guiding you through basic operations of the
instrument. It includes pre-startup checks and startup, operations before test, routine test, daily
performance and powering off.
Maintenance Card
Provides regular and irregular maintenance of the instrument to help you maintain it so that it
can work normally.
For detailed maintenance instructions, see the Operator's Manual.
Online help
Contains detailed descriptions of the software screens and parameters. It also covers the
Operator's Manual, which enables you to retrieve information related to the software screens
and operation tasks.
Conventions
Graphical symbols, formats and abbreviations are used to get better visual effects and
readability. To help you understand this manual correctly, this section provides statements of
pictures, terms and applicable models used in this manual.
Symbols and formats
The following symbols and formats are used:
Online help
The operating software provides a context-sensitive online help, which can help you better
understand the screen parameters and perform correct operations. The online help is related to
software screens, and it can display information related to menu page, maintenance item,
maintenance command, and event log.
You can open the online help window in the following ways:
Alt+F1: press this shortcut key combination on any screen.
: click this icon on the top-right of any screen.
: click this button to the left of a maintenance item, a maintenance command, or an event
log.
Safety information
This chapter provides you with safety symbols used in this manual and their meanings,
summarizes the safety hazards and operating precautions that should be considered seriously
when the instrument is being operated, and lists the labels and silkscreens that have been
applied to the instrument and their indications.
Safety symbols
Safety symbols are used in this manual in order to remind you of the instructions necessary to
operate the product safely and in accordance with its function and intended use. A safety symbol
and text constitutes a warning as shown in the table below:
Symbol Meaning
Caution
Biological risks
Summary of hazards
This section lists hazards of the instrument itself. The hazards of specific operation are included
in the warning information of each operation task.
Observe the following safety precautions when using the product. Ignoring any of them may
lead to personal injury or equipment damage.
WARNING
If the product is used in a manner not specified by our company, the protection provided by the product may be
impaired.
Electric shock hazards
WARNING
When the MAIN POWER is turned on, users other than the servicing personnel authorized by our company
must not open the rear cover or side cover.
Spillage of reagent or sample on the product may cause equipment failure and even electric shock. Do not
place sample and reagent on the product. In case of spillage, switch off the power immediately, remove the
spillage and contact our Customer Service Department or your local distributor.
Moving Parts Hazards
WARNING
Do not touch such moving parts as sample/reagent carousel, reaction carousel, probe, mixer, and cuvette
wash station, when the system is in operation.
Exercise caution while using the ISE module Prevent your hair, legs or other parts of your body from being
hurt by the driving parts.
Do not put your fingers or hands into any open part when the system is in operation.
Photometer lamp hazards
WARNING
Eye injury could occur from light emission from the photometer lamp. Do not stare into the lamp when the
system is in operation.
If you want to replace the photometer lamp, first switch off the MAIN POWER and then wait at least 5
minutes for the lamp to cool down before touching it. Do not touch the lamp before it cools down, or you
may get burned.
Sample, calibrator and control hazards
BIOHAZARD
Inappropriately handling samples, controls and calibrators may lead to biohazardous infection. Do not touch
samples, controls, calibrators, mixtures, or waste with your bare hands. Wear gloves and lab coat and, if
necessary, goggles.
In case your skin contacts the sample, control or calibrator, follow the standard laboratory safety procedure
and consult a doctor.
The serum samples remaining in the electrodes may contain a great number of viruses. Wear gloves to
prevent infection while operating around the electrodes.
WARNING
Reagents,diluted wash solution and concentrated wash solution are corrosive to human skins. Exercise caution
when using reagents and concentrated wash solution. In case your skin or clothes contact them, wash them off
with soap and clean water. If reagents or wash solution spills into your eyes, rinse them with much water and
consult an oculist.
Waste hazards
BIOHAZARD
Some substances contained in reagent, control, calibrator, concentrated wash solution, and waste are
subject to regulations of contamination and disposal. Dispose of the waste in accordance with your local or
national rule for biohazard waste disposal and consult the manufacturer or distributor of the reagents for
details.
Wear gloves and lab coat and, if necessary, goggles.
System disposal hazards
WARNING
Materials of the analyzer are subject to contamination regulations. Dispose of the waste analyzer in accordance
with your local or national rule for waste disposal.
Fire and explosion hazards
WARNING
Ethanol is flammable substance. Please exercise caution while using ethanol around the instrument in order to
prevent fire and explosion.
Removal of analyzer from use for repair or disposal
WARNING
When the analyzer is not in use, for example, in repair, transportation or disposal process, please clean and
sterilize the parts (sample probe, reagent probe, etc.) or surfaces that may cause biohazdards and remind
the person who handles the device of the related hazards.
Cleaning and Decontamination
CAUTION
Appropriate decontamination should be carried out in accordance with laboratory safety regulations if
reagent, sample or other liquids are spilled onto the equipment. In case of large-amount liquid ingression,
please contact our customer service department or the local distributor.
No decontamination or cleaning agents can be used which could cause a HAZARD as a result of a reaction
with parts of the equipment or with material contained in it. Strong acid or alkaline solutions are forbidden to
clean the equipment.
If there is any doubt about the compatibility of the decontamination or cleaning agents with parts of the
equipment or with material contained in it, please contact our customer service department or the local
distributor.
CAUTION
Recommended detergent: water and 75% ethanol.
Prohibited detergent: materials that may corrode metals, for example, 3% hydrogen peroxide.
The user shall perform regular cleaning to the cover of the analyzer. Use the specified materials to clean the
equipment only. For any damage to the instrument or other accidents caused by using materials other than
specified, Mindray will not provide any warranty.
Mindray does not claim the validity of the listed chemicals in infection control. For effective control of infection,
please consult the Infection Prevention Department of the hospital or the epidemic professionals.
Disinfection may damage the system to some extent. It is recommended to perform disinfection only when
necessary according to your laboratory protocol.
Do not use any cleaning agents which could cause a HAZARD as a result of a reaction with parts of the
equipment or with material contained in it.
If you accidentally spill hazardous material (for example, samples and reagents) on the instrument, clean and
disinfect the instrument. Recommended detergents and disinfectants include water and 75% ethanol. Do not
use materials that may corrode metals (for example, 3% hydrogen peroxide). Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures when handling them and the
contacted areas in the laboratory.
Software and Cybersecurity
WARNING
Data should be transmitted in a closed network or virtual isolated network environment. The laboratory is
responsible for the security of the virtual isolated network environment.
Make sure that the network authorization information (such as user information and password) is secure and
not obtained by unauthorized persons.
Please use Microsoft firewall and kill the virus regularly.
Notification of Adverse Events
NOTE
As a health care provider, you may report the occurrence of certain events to SHENZHEN MINDRAY
BIO-MEDICAL ELECTRONICS CO., LTD., and possibly to the competent authority of the Member state in
which the user and / or patient is established.
These events, include device-related death and serious injury or illness. In addition, as part of our Quality
Assurance Program, SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. requests to be notified
of device failures or malfunctions. This information is required to ensure that SHENZHEN MINDRAY
BIO-MEDICAL ELECTRONICS CO., LTD. provides only the highest quality products.
Summary of precautions
This section lists precautions to be understood during instrument operation. The precautions of
specific operation are included in the warning information of each operation task.
To use the product safely and efficiently, pay attention to the following operating precautions.
Intended use
WARNING
The instrument is an automated chemistry analyzer for in vitro diagnostic use in clinical laboratories and
designed for in vitro quantitative determination of clinical chemistries in serum, plasma, urine and cerebrospinal
fluid samples.
Please consult us before you use the instrument for other purposes.
When drawing a clinical conclusion, please also refer to patients' clinical symptoms and other test results.
Environment precautions
CAUTION
Please install and operate the system in an environment specified by this manual. Installing and operating the
system in other environment may lead to unreliable results and even equipment damage.
To relocate the system, please contact our Customer Service Department or your local distributor.
Installation precautions
WARNING
The product is a permanently connected equipment, and it is switched on and off via a switch or breaker.
Before installing the system, ensure that the building in which the equipment will be located has been equipped
with a switch or breaker that complies with IEC 61010-1+AMD1, is in close proximity to the equipment and
within easy reach of you, and is marked as the disconnecting device for the equipment.
NOTE
The safety of any system incorporating the equipment is the responsibility of the assembler of the system.
Electromagnetic Noise Precautions
WARNING
requirements described in this part of IEC 61326.
This equipment is not intended for use in residential environments and may not provide adequate protection to
radio reception in such environments.
This equipment is designed for use in a PROFESSIONAL HEALTHCARE FACILITY ENVIRONMENT. It is likely
to perform incorrectly if used in a HOME HEALTHCARE ENVIRONMENT. If it is suspected that performance is
affected by electromagnetic interference, correct operation may be restored by increasing the distance between
the equipment and the source of the interference.
The electromagnetic environment should be evaluated prior to operation of the device.
Do not use this device in proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional
RF sources), as these can interfere with proper operation.
NOTE
It is the manufacturer’s responsibility to provide equipment electromagnetic compatibility information to the
customer or user.
It is the user’s responsibility to ensure that a compatible electromagnetic environment for the equipment can be
maintained in order that the device will perform as intended.
The calculation formula to dete rmine the separation distance between an IVD MEDICAL EQUIPMENT and a
mobile phone is given by d = 6/E ∙√P, where d is the minimum separation distance in metres, P is the
maximum power in watts, and E is the immunity test level in V/m.
Operating precautions
CAUTION
Take the clinical symptoms or other test results of the patient into considerations when making diagnosis
based on the measuring results produced by the system.
Operate the system strictly as instructed by this manual. Inappropriate use of the system may lead to
unreliable test results or even equipment damage or personal injury.
When using the system for the first time, run calibrations and QC tests to make sure the system is in proper
state.
Be sure to run QC tests every time when you use the system, otherwise the result may be unreliable.
Start the operating software again when the analyzing unit is powered off.
Do not uncover the reagent carousel when the system is in operation. Keep the reagent carousel cover
closed.
The RS-232 port on the analyzing unit is used for connection with the operation unit only. Do not use it for
other connections. Use the cables provided by our company or your local distributor for the connection.
The operation unit is a personal computer with the operating software installed. Installing other software or
hardware on the computer may interfere with the system operation. Do not run other software when the
system is working.
Computer virus may destroy the operating software or test data. Do not use the computer for other
purposes or connect it to the Internet. If the computer is infected by virus, please install anti-virus software
to check for and clear virus.
Do not touch the display, mouse or keyboard with wet hands or hands with chemicals.
Do not place the MAIN POWER to ON again within 10 seconds after placing it to OFF; otherwise the
system may enter the protection status. If it does so, place the MAIN POWER to OFF and place it to ON
again.
Any serious incident that has occurred in relation to the device shall be reported to the manufacturer and
the competent authority of the country in which the user and/or the patient is established.
Chemistry parameter configuration precautions
CAUTION
To define such parameters as sample volume, reagent volume and wavelength, follow the instructions in this
manual and the instructions of reagents.
CAUTION
To prevent ISE electrodes from being damaged due to water scarcity, if the system, when equipped with an ISE
module will be powered off for a long time, perform the electrode storage maintenance.
Sample precautions
CAUTION
Use samples that are completely free of insoluble substances like fibrin or suspended matter; otherwise the
sample probe may be clogged.
Medicines, anticoagulants or preservative in the samples may lead to unreliable results.
Hemolysis, icterus or lipemia in the samples may lead to unreliable test results; running a serum index test
therefore, is recommended.
Store the samples properly. Improper storage may change the compositions of samples and lead to
unreliable results.
Sample volatilization may lead to unreliable results. Do not leave the sample open for a long period.
Prepare sufficient sample volume before analysis.
Load samples to correct positions on the sample carousel before the analysis begins; otherwise reliable
results may not be obtained.
Reagent, calibrator and control precautions
CAUTION
Use proper reagents, calibrators and controls on the system.
Select appropriate reagents according to the performance characteristics of the system. Consult the
reagent suppliers, our company or our authorized distributor for details, if you are not sure about your
reagent choice.
Store and use the reagents, calibrators and controls strictly as instructed by the suppliers; otherwise,
reliable results or best performance of the system may not be obtained. Improper storage of reagents,
calibrators and controls may lead to unreliable results and bad performance of the system even in validity
period.
Perform calibration after changing the reagents, otherwise reliable results may not be obtained.
Contamination caused by carryover among reagents may lead to unreliable test results. Consult the reagent
suppliers for details.
ISE calibration precautions
BIOHAZARD
The calibrators contain preservatives. In case your skin contacts calibrators, wash them off with soap and water.
In case the calibrators spill into your eyes, rinse them with water and consult an oculist. If you swallow them by
mistake, see a doctor.
CAUTION
Use the calibrators specified by our company. Use of other reagents or calibrators may result in unreliable
results, or damage the Hydropneumatic system, or even shorten the electrodes life span.
Prior to using the calibrators, check if they are within the expiration date.
Place them correctly; otherwise, it may cause unreliable results, or leak, or module damage.
ISE cleaning solution biohazards
BIOHAZARD
The ISE cleaning solution is sodium hypochlorite. Use the ISE cleaning solution carefully to prevent it from
contacting your skins or eyes. If your skins or eyes contact the ISE cleaning solution, rinse them off with fresh
water and consult a doctor.
Data archiving precautions
NOTE
The system automatically stores the data to the built-in hard disk. Data loss, however, is still possible due to
mis-deletion or physical damage of the hard disk. You are recommended to regularly archive the data to such
medium as CDs.
To avoid the data loss caused by unexpected power failure, UPS (uninterrupted power supply) is
recommended.
External equipment precautions
WARNING
For operating instructions and precautions of the computer and printer, please refer to their operation manuals.
External equipment connected to the analogue and digital interfaces must be authorized and complied with
relevant safety and EMC standards (e.g., IEC 60950 Safety of Information Technology Equipment Standard and
CISPR 22 EMC of Information Technology Equipment Standard (CLASS B)). Any person, who connects
additional equipment to the signal input or output ports and configures an IVD system, is responsible for
ensuring that the system works normally and complies with the safety and EMC requirements. If you have any
questions, consult the technical services department of your local representative.
Tube and liquid container precautions
WARNING
When the tube or the part that contain liquid become aged or damaged, please stop its use immediately and
contact our customer service department or your local distributor to check and replace it.
NOTE
Check the safe state of the equipment after repair. Make sure the equipment is safe and then offer it to the
customer.
For the label marked with , please consult the related documentations in order to find out
the nature of the potential HAZARDS and any actions which have to be taken to avoid them.
Check the labels regularly for cleanliness and integrity. If any of the labels becomes vague or
peels off, contact our Customer Service Department or your local distributor for replacement.
The general meaning assigned to geometric shapes, safety colors and contrast colors for safety
signs are as follows.
Date of Manufacture
Manufacturer
CE marking
Symbol Meaning
The following definition of the WEEE label applies to
EU member states only: The use of this symbol
indicates that this product should not be treated as
household waste. By ensuring that this product is
disposed of correctly, you will help prevent bringing
potential negative consequences to the environment
and human health. For more detailed information with
regard to returning and recycling this product, please
consult the distributor from whom you purchased the
product.
In Vitro diagnostic medical device
Biological risks
Caution
“ON” (Power)
“OFF” (Power)
Serial interface
Computer network
Alternating current
Warning labels
Biohazard warning
This label indicating the risk of biohazardous infection is located in the following positions:
Probe
Waste outlet
Waste tank
ISE module
This symbol and text located on the left side panel of the analyzer. Please turn off the main
power before opening the small door.
Chemical hazard
This label is applied on the diluted wash solution tank. Please take protective measures to
prevent chemical hazard.
Cabinet
This label is applied at the middle of the cabinet back. If you want to relocate your analyzer,
contact our Customer Service Department or your local distributor. When relocating it, do not
carry it with the cabinet. Only when the cabinet is supported by the anchors, can the analyzer
be placed on the cabinet.
4 Calibration························································································································· 4-1
4.1 ISE calibration ................................................................................................................................................................... 4-2
4.1.1 Calibration setup ........................................................................................................................................... 4-2
4.1.2 Calibration status and alarm ..................................................................................................................... 4-2
4.1.3 Results recall.................................................................................................................................................... 4-3
4.2 Biochemistry calibration .............................................................................................................................................. 4-5
4.2.1 Calibration setup ........................................................................................................................................... 4-5
4.2.2 Calibration status and alarm ...................................................................................................................4-11
4.2.3 Reagent blank................................................................................................................................................4-12
4.2.4 Recalling calibration results ....................................................................................................................4-15
5 QC ··································································································································· 5-1
5.1 Overview ............................................................................................................................................................................. 5-2
5.1.1 QC procedure .................................................................................................................................................. 5-2
5.1.2 QC result flags ................................................................................................................................................. 5-2
5.1.3 Control status................................................................................................................................................... 5-2
5.2 QC setup............................................................................................................................................................................... 5-3
5.2.1 Defining/Editing a control ......................................................................................................................... 5-3
5.2.2 Setting up control concentrations........................................................................................................... 5-4
5.2.3 Setting up QC rules ........................................................................................................................................ 5-4
5.2.4 Auto QC .............................................................................................................................................................. 5-5
5.2.5 Deleting a control ........................................................................................................................................... 5-6
5.3 Recalling control results ................................................................................................................................................ 5-6
5.3.1 Result > History screen ............................................................................................................................... 5-6
5.3.2 Levey-Jennings screen ................................................................................................................................ 5-8
5.3.3 Recalling cumulative sum chart............................................................................................................... 5-9
5.3.4 Recalling Twin-Plot chart .........................................................................................................................5-10
5.3.5 QC > Results screen .....................................................................................................................................5-11
5.3.6 Recalling QC Summary..............................................................................................................................5-14
6 Program ···························································································································· 6-1
6.1 Sample management ...................................................................................................................................................... 6-2
6.2 Sample programming and processing .................................................................................................................... 6-3
6.2.1 Processing samples with LIS .................................................................................................................... 6-3
6.2.2 Processing bar-coded samples ................................................................................................................. 6-4
6.2.3 Batch programming ..................................................................................................................................... 6-5
6.2.4 Adding samples ............................................................................................................................................... 6-6
6.2.5 Adding/Modifying chemistries ................................................................................................................ 6-7
6.2.6 Rerunning samples ....................................................................................................................................... 6-7
6.2.7 Sample blank..................................................................................................................................................6-13
6.2.8 Whole Blood Test .........................................................................................................................................6-14
6.3 Serum Index .....................................................................................................................................................................6-15
6.3.1 Introduction ...................................................................................................................................................6-15
6.3.2 Theory of Serum Index ..............................................................................................................................6-15
6.3.3 Serum Index Setup ......................................................................................................................................6-16
6.3.4 Auto Serum Index ........................................................................................................................................6-17
6.3.5 Running SI Chemistry ................................................................................................................................6-17
6.4 Extended functions........................................................................................................................................................6-17
1 System description
This chapter describes the system from the installation, hardware, software and specifications
perspectives, including:
Installation requirements and methods of the instrument
Hardware components
Introduction of software screens
Technical specifications
CAUTION
Install the instrument in a place meeting the requirements presented in this section; otherwise, it will not perform
as intended.
Installation environment
The following environment requirements must be satisfied:
The system is for indoor use only.
The bearing platform should be level (with gradient less than 1/200).
The bearing platform should be able to support at least 150Kg weight.
The bearing ground should be able to support at least 200Kg weight.
The installation site should be well ventilated.
The installation site should be free of dust.
The installation side should not be in direct sun.
The installation site should be kept away from a heat or draft source.
The installation site should be free of corrosive gas and flammable gas.
The bearing platform should be free of vibration.
The installation site should be kept away from large noise and power supply interference.
Keep the system away from brush-type motors and electrical contact device that is frequently
switched on and off.
Do not use such devices as mobile phones and radio transmitter near the system.
The system should be installed in a place with altitude height -400-2000 m.
Power supply
The following power supply requirements must be satisfied:
Connect the system to a power supply meeting the requirements specified in this manual.
For more information on power supply, see 1.4.2Power supply requirements on page 1-40.
Use the three-wire power cord provided with the system, which has good grounding
performance.
Connect the system to a properly-grounded power socket.
Configure the grounding voltage correctly.
WARNING
Make sure the power socket is grounded correctly. Improper grounding may lead to electric shock or equipment
damage. Check if the power sockets outputs voltage meeting the specified requirements and has a proper fuse
installed.
CAUTION
Operating the system in an environment other than the specified may lead to unreliable test results. If the
temperature or relative humidity does not meet the above-mentioned requirements, use air-conditioning
equipment.
CAUTION
The water supply must meet the requirements; otherwise insufficiently purified water may result in misleading test
results.
The high-/low-concentration waste produced during cuvette wash are drained separately. The
high-concentration waste is discharged to the provided 10 L waste tank, and the low-concentration
waste is discharged to the provided 15 L waste tank or to a drain outlet.
BIOHAZARD
Dispose of the waste liquid according to the local regulations.
After installing the instrument, connect it with the fluidic components as instructed in the figure
below.
BIOHAZARD
Wear gloves and lab coat, if necessary, goggles.
CAUTION
When connecting the tubes, exercise caution to avoid folding or pressing them.
FL02 FL04
FL03
FL01
DI water tank Diluted wash solution tank Low-conc. waste tank High-conc. waste tank
Wall
Min. 500
Maximum of 2500
Min. 500
Operation Unit
660
Analyzing Unit
Front
860
Min. 500
Item Description
Hard disk defragment At least 500 GB for hard disk. Install the operating system in the C
drive and the operating software of the instrument in the D drive.
Make sure that the C drive is over 30 GB and the D drive over 100
GB, and the disk file system is of NTFS format. Deselect the two
options at the bottom of the disk properties window: "Compress
drive to save disk space" and "Allow Indexing Service to index this
disk for fast file searching".
Operating system The operating system installed on the computer must be an
activated or free version Microsoft Windows 10 (64 bit).
Application software Except for the operating system, other application software must
not be installed or reserved on the computer. If an anti-virus
application has been installed, then remove the automatic
scheduled scanning and add the operating software and BSLOG to
the trust list.
Screen saver and system Turn off the screen saver and BS Special Power Policy power
standby scheme, and then disable the hibernation option.
Screen display Set the screen resolution as 1280*1024 pixels.
properties
Sound card and speaker The system must be configured with sound card and speaker.
WARNING
The system should be installed only by technicians of or authorized by our company.
The system should be installed by technicians of or authorized by our company. Before the
technicians arrive, prepare a proper site to install the system.
Before installation
When you receive the package, check it carefully. If you find any signs of mishandling or damage,
file a claim immediately with our Customer Service Department or your local distributor.
After opening the package, check the delivered goods against the packing list, and then visually
check the system appearance. If you find anything missing or damaged, alert our Customer Service
Department or your local distributor immediately.
System relocation
If you want to relocate your system, contact our Customer Service Department or your local
distributor.
(8)
(7)
(6)
Analyzing unit
The analyzer, determines various clinical chemistries in samples and generates test results. It is
composed of the following components:
Sample/Reagent handling system
Mixer assembly
Reaction system
Cuvette wash station
Photometric system
ISE unit
Cabinet
Operation unit
A computer with the operating software installed to perform test requisition, measurement,
reaction process monitoring, result calculation, and input, storing and query of test data.
Output unit
A printer for printing out test results and other data.
Accessories and consumables
Includes cuvette, lamp, concentrated wash solution (CD80), and other accessories and
consumables required by test.
(4)
(5)
(1) ISE sample injection port (2) Aspirate port on inner ring
(3) Aspirate port on middle ring (4) Aspirate port on outer ring
(5) Sample/Reagent carousel
Carousel positions
The carousel holds both sample and reagent, and it includes the outer ring, middle ring and inner
ring. The three rings provides the following positions:
Outer ring: No.1-50, can hold 50 samples; supporting bar code scanning.
Middle ring: can hold 50 20mL or 40mL reagent bottles or 50 sample tubes with adapter;
supporting bar code scanning.
Inner ring: 50 positions are provided. When 40mL reagent bottles are placed in the middle ring,
the inner ring cannot hold the 20mL reagent bottles. When 20mL reagent bottles or sample
containers are placed on the middle ring, 20mL reagent bottles can be placed on the inner ring.
The following fixed positions are provided for special reagents:
W for physiological saline(No.50)
D for probe wash solution(No. 49)
CAUTION
Every day before analysis, remove the plugs on the sample/reagent carousel in order to prevent mechanical reset
failure and bending the reagent probe. Restore the plugs after finishing tests of the day.
Ensure that the sample/reagent carousel is closed while the system is running tests. Opening the carousel cover
during test could result in probe collision or other failures.
NOTE
The refrigeration unit is powered independently from the analyzing unit, and it is operational once the MAIN
POWER is put to the ON position.
Installing/Removing the sample/reagent carousel
WARNING
Before installing or removing the sample/reagent carousel, make sure that the analyzer is in standby status or is
shut down, and the sample/reagent carousel has stopped.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
(2)
(1)
(3)
CAUTION
Make sure the carousel cover is closed; otherwise the refrigeration performance could be degraded and the probe
could be damaged.
Before system operation, make sure that the carousel cover is closed properly; otherwise the probe could be
damaged.
The sample/reagent compartment and the carousel could be contaminated during measurement. If sample or
reagent spills in the compartment or on the carousel, switch off the analyzing unit power, and wipe them with cloth
soaked with water or disinfector.
Sample containers
Sample containers are used to hold sample.
Sample tubes varying in specification requires different minimum sample volumes. Each sample
tube must contain the minimum amount of sample; otherwise, correct aspirating cannot be
ensured. The minimum sample volume is the sum of the sample volume for analysis (sum of
defined sample volume for chemistry and the excessive aspiration 5μl) and the dead volume of the
sample container.
Table 1.2 Specification of sample containers
Sample Container Specification Dead Volume
Microtube Φ14×25 mm, 0.5 120μL
mL(Not used for ISE
test)
Φ14×25 mm, 2 mL 150μL
(Not used for ISE test)
Φ12×37 mm, 2 120μL
mL(Not used for ISE
test)
Primary tube or plastic Φ12×68.5 mm 8mm higher than the
tube Φ12×99 mm unacceptable sample
level.
Φ12.7×75 mm
Φ12.7×100 mm
Φ13×75 mm
Φ13×95 mm
Φ13×100 mm
For the tests of the whole blood(centrifuged),onlyΦ12×68.5 mm, Φ12×99 mm, Φ12.7×75 mm,
Φ12.7×100 mm, Φ13×75 mm, Φ13×95 mm, Φ13×100 mm anticoagulation tubes can be used. The
sample height in the tube should be no higher than 55mm and the blood cell level should be no
lower than 10mm. Microcups are not allowed. To ensure the clinical performance and avoid the
system alarm, EDTA anticoagulation tubes are recommended.
Reagent bottles
20mL and 40mL reagent bottles are used.
Loading/Unloading sample tube
WARNING
Before installing or removing a sample tube, make sure that the sample/reagent carousel and the probe have
stopped.
Do not use sample tubes other than the specified ones.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
To load a sample tube, insert it into the tube holder until the tube bottom contacts the groove of the
tube rack.
To unload a sample tube, grab it and pull it upwards to remove from the tube holder.
Loading/Unloading reagent bottle
WARNING
Before installing or removing a reagent bottle, make sure that the sample/reagent carousel and the probe have
stopped.
Do not use reagent bottles other than the specified.
Some reagents may hurt human skins. Exercise caution when using the reagents. In case your skin or clothes
contact them, wash them off with clean water. In case the reagents spill into your eyes, rinse them with much
water and consult an oculist.
To load a reagent bottle, insert it into the bottle holder until the bottle bottom contacts the groove of
the holder.
To remove a reagent bottle, grab it and pull it upwards to remove from the bottle holder.
(1)
The table below lists the specifications of the bar code reader:
Table 1.3 Specifications of the bar code reader
Symbology Codabar, ITF, Code128, Code39, UPC/EAN, and Code93
Minimum bar code Outer ring:0.19 mm ~ 0.50 mm
density Middle ring:0.25 mm ~ 0.50 mm
Length Sample bar code: 3-27 digits
Reagent bar code: 13-30 digits
Format and content User-defined
Maximum width 55mm
Minimum height Sample bar code: 10 mm
Reagent bar code: 12 mm
Maximum inclination ±5°
angle
Print quality No less than Class C according to the ANSI MH10.8M Print Quality
Specification.
Width and narrowness Sample bar code: (2.5-3.0):1
Reagent bar code: 2.5:1
Print paper Coated paper or matte paper. Printing bar code on common paper
may result in vague bar code or degraded bar code label. You are
not suggested to print bar code on common print paper.
Characters Meaningful characters, such as numbers (0-9) and upper-case
letters (A-Z). You are recommended to print the check digit in
order to check that a bar code is read accurately.
Dispenser assembly
The dispenser assembly located at the upper-right corner of the sample/reagent carousel is
composed of the probe, probe arm, probe rotor, syringe, wash well, and related fluidic path. It
aspirates the specified amount of sample and reagent from a sample tube and a reagent bottle and
then dispenses them into a cuvette for reaction.
Figure 1.9 Dispenser assembly
(1)
(2)
(4)
(3)
WARNING
When the system is in operation, do not place any part of your body or any obstacle in the route where the sample
probe arm moves; otherwise, personal injury or equipment damage may be caused.
Probe
One probe is available to add sample and reagent with the following volume range:
Biochemistry: 2-35 μL, with increment of 0.1 μL.
ISE test: 70 μL for serum and plasma, and 140 μL for diluted urine.
Reagent: 10-200 μL, with increment of 0.5 μL.
Besides adding sample and reagent, the probe has the following functions:
Vertical obstruct detection: Detects obstacles in the vertical direction. When the probe collides
with an obstacle in the vertical direction, the auto guard system is started to prevent the probe
from being damaged.
Level detection and tracking: Detects the liquid level in sample tube and reagent bottle and
determines the depth of lowering down into the liquid based on the specified aspirate volume.
(1)
(2)
(4)
(3)
When stirring is finished, the mixer moves automatically to the wash well for cleaning.
Reaction carousel
The reaction carousel is a single-ring turntable, which can hold 8 cuvette segments. Each cuvette
segment includes 10 cuvettes.
The reaction carousel is heated by means of air bath to provide a constant environment at
37±0.3°C with fluctuation of ±0.1°C.
Figure 1.11 Reaction carousel
(1)
(2)
(4)
(3)
Reaction cuvette
Plastic reaction cuvette segments are used. Each segment includes 10 cuvettes (5 mm × 5 mm). The
light path length of the cuvette is 5 mm, and the internal dimension is 5 mm (length) × 5 mm
(depth) × 29.5 mm (height).
The cuvettes used on the analyzer can be washed automatically and should be replaced every 3
months.
(3)
(1) (2)
The cuvette wash station performs 8-phase wash on cuvettes by using diluted CD80 wash solution
and deionized water, so that the cuvettes can be contamination free and dry during the test
process.
After wash, the liquid waste is discharged in two flows: high-concentration waste and
low-concentration waste. The system provides level detection of high-concentration waste. When
the high-concentration waste exceeds the specified volume, the system gives an alarm to remind
you to empty the high-concentration waste tank.
For drainage requirements of high-/low-concentration waste, see Water supply and drainage on
page 1-13.
WARNING
The ISE unit must be operated by skilled/trained doctors, nurses or clinical professionals.
Exercise caution while using the ISE unit Prevent your hair, legs or other parts of your body from being hurt by the
driving parts.
The ISE (Ion Selective Electrode) unit consists of the ISE module, the pump module and the reagent
module, used in combination with Na, K, Cl, and reference electrodes to measure the concentration
of Na+, K+ and Cl- ions in serum, plasma and diluted urine.
If not specifically noted, "ISE module" represents the ISE unit.
(3)
(1)
(2)
In ISE test, 70 μL serum or plasma, or 140 μL diluted urine (diluted at the ratio of 1:10) is
required.
The following components are comprised in the ISE unit:
Reagent module: includes the calibrator A, calibrator B, waste container, and chip for measuring
reagent volume.
Pump module: includes the calibrator A pump, calibrator B pump and waste pump.
1.2.8 Cabinet
The cabinet is used to hold the analyzer and can at least bear 150 kg weight.
(1)
(2)
(3)
(4)
(1)
(1) Printer
CAUTION
Use the accessories, power cords and consumables manufactured or recommended by our company in order to
achieve the promised system performance and safety. If needed, contact our customer service department or your
local distributor.
The Accessories and consumables list is subject to change; if needed, contact our customer service
department or your local distributor.
Table 1.4 Accessories and consumables
No Part Name Remark
1 Needle .0.25+/-0.01mm*125mm round head Accessory
2 Valve Washer,10-32,18011Telfon washer Accessory
3 20ml reagent bottle brown Accessory
4 40ml reagent bottle brown Accessory
5 20ml reagent bottle Accessory
6 Reagent bottle label Accessory
7 BS200 white cap of reagent bottle Accessory
8 BS200 red cap of reagent bottle Accessory
9 Accessory kit bar code Accessory
10 Parameter list Accessory
11 Cross screwdriver 102*100 Accessory
12 Serial port cable Accessory
13 Cuvette (Surface processed) Accessory
14 10L water tank Accessory
15 Water tank Accessory
16 Filter Accessory
(5) (3)
(4)
: Start icon. Select it to display the Start Conditions window, on which you are allowed to
start new analysis or resume testing.
: Pause icon. Select it to stop dispensing of sample and reagent. Then you are allowed to
load new samples or reagents on the sample/reagent carousel. To resume the test, select .
: Emergency stop icon. Select it to stop all tests and other actions. To restore the system
into Standby status, execute the Home command.
: STAT icon. Select it to display the STAT Sample Program window, on which you are
enabled to program emergency samples quickly.
: Online help icon. Select it to display the online help of the current window, where you can
find description of parameters and operations.
Function window
It displays the page or window related to the selected function button or shortcut icon. It is used to
perform system operations.
: used to program patient samples and control samples, and view sample carousel status.
: used to recall test results of patient samples and controls and view the result statistics
and test statistics.
: used to set reagents, define/edit calibrators, request calibrations, recall calibration results,
and view reagent carousel status.
: used to execute instrument commands, set up chemistry and system parameters, perform
system maintenance, and view component status.
Page
Window
A window has visible boundaries, which consists of title bar, content area and buttons.
The figure below is an example of window:
Figure 1.18 Example of window
Window
Tab
Click a tab to access the working page that it indexes.
The figure below is an example of tab:
Figure 1.19 Example of tab
Tabs
Drop-down list
Drop-down list
Combo box
A combo box includes a field and a drop-down list, in which you can manually input characters or
select an option.
The figure below is an example of combo box:
Figure 1.21 Example of combo box
Combo box
Edit box
An edit box is a field in which you can input characters manually.
The figure below is an example of edit box:
Figure 1.22 Example of edit box
Edit box
Radio button
A radio button is a circle with text beside. It indicates a set of choices from which only one can be
selected.
The figure below is an example of radio button:
Figure 1.23 Example of radio button
Radio button
Check box
A check box is a square box with text beside. It indicates a set of choices from which one or more
can be selected.
The figure below is an example of check box:
Check box
Button
A button is used to open a window or to execute a defined function.
The figure below is an example of button:
Figure 1.25 Example of button
Button
Scroll bar
A scroll bar is used to display the hidden contents when they are too many to be shown on one
screen. A vertical scroll bar moves the screen up and down, and a horizontal scroll bar moves the
screen left and right.
The figure below is an example of scroll bar:
Figure 1.26 Example of scroll bar
List
A list holds multiple chemistries or panels, or contains texts or charts in the form of table.
The figure below is an example of list:
List
Slider
A slider is used to select a scale continuously. Click and hold the slider and drag it to the desired
scale.
The figure below is an example of slider:
Figure 1.28 Example of slider
Slider
Program Sample
Quality Control
Status
Current By Sample
Result
History By Chemistry
Statistics
Reagent Reagent/Calibration
Biochemistry Calibration
ISE Calibration
Reagent Carousel Status
Cal Setup
Levey-Jennings
QC Cumulative sum
Twin-Plot
Results
Summary
QC Setup
Commands
Utility
Chemistries
System Setup Status Summary
Maintenance Count
Status Temperature
Hydro
Power
Alarm Error Log
Edit Log
Exit
Select the icon on the upper right corner to display the help topic related to the current
screen.
Figure 1.30 Accessing the online help from the main screen
Select the icon in front of each maintenance command or regular maintenance item to
display the relevant operating instructions.
Figure 1.31 Accessing the online help from the Maintenance window
Select the icon in front of each error log to display the corresponding topic.
Figure 1.32 Accessing the online help from the Error Log screen
Select the icon on a warning message window to display the corresponding descriptions
and solutions.
Select the icon on an error message window to display the corresponding descriptions
and solutions.
Press the shortcut key combination Alt+F1 to display the topic related to the current page or
window.
1 Select the icon on the upper right corner of the main screen, or press the shortcut key
combination Alt+F1.
2 Select the following tabs to view relevant information:
Contents: to navigate through all topics of the online help.
Index: to view topics related to the input keywords.
Search: to view topics containing the input keywords.
Favourites: to view your favorite topics.
3 Read the help topics. Move the scroll bar on the right side of the help window to view more
information.
4 Select to close the help window.
Photometric system
Table 1.10 Specifications of the photometric system
Light source 12V/20W tungsten-halogen lamp, 2000 hours life span
Colorimetric component Reaction cuvette
Light-splitting mode Holographic concave flat-field gratings
Measuring wavelength 12 wavelengths: 340nm, 380nm, 412nm, 450nm,
505nm, 546nm, 570nm,605 nm,660 nm,700 nm,740
nm and 800nm
Absorbance measurement 0 - 3.5 A
range
Measuring period 10 seconds
Reaction volume 100 μL - 360 μL
Average water consumption
≤ 6.5 L/H
CAUTION
Use the fuse specified in this manual.
NOTE: “25/30 cycles" means "25 cycles for 50 Hz test" or "30 cycles for 60 Hz test”.
Performance criterion:
The equipment shall continue to operate as intended during and after the test.
The equipment shall continue to operate as intended after the test.
LOSS OF FUNCTION is allowed, provided the function is self-recoverable or can be restored
by the operation of the controls.
1.4.12 Contraindication
None.
This chapter describes a typical daily operating procedure of the instrument. For instructions of
more operations, see other chapters in this book.
All operations described in this chapter are based on complete configuration of the instrument.
If you do not have certain optional module, please neglect the relevant steps or operate in
another way provided.
To understand the detailed information of software screens, see the Online Help.
Routine test
1. Sample processing
2. Checking results
3. Checking test status and test control
2.2.2 Startup
The instrument can be started manually or automatically. For manual startup, you need to
switch on the power supply manually; for auto startup, you should set up the weekday and time
for the instrument to start up automatically.
Manual startup
You need to switch on the power supply and log on the operating software. When the
instrument is powered on, the operating software runs automatically and requires you to enter
the username and password.
To switch on the power supply
1 Turn on the main power switch on the rear panel.
Place the switch to the position to turn it on. Place the switch to the position to
turn it off.
2 Turn on the analyzing unit power switch on the left panel.
Figure 2.2 Analyzing unit power switch
(1)
Place the switch to the position to turn it on. Place the switch to the position to
turn it off.
3 Turn on the printer.
4 Turn on the monitor and computer of the operation unit.
To start the operating software
NOTE
If virtual reagent carousel is used, please make sure that the loaded one is the NO.1 reagent carousel before
starting the operating software each time.
1 When the operation unit (computer) is turned on, the operating software will run
automatically.
2 Enter the username and password in the Login window, and then select OK.
3 Select Quick Mode if analyzer initialization procedure is not needed.
NOTE
The default username and password for administrator is Admin. Please note that the password is case
sensitive. You are recommended to change the password when logging on the system for the first time in
order to prevent others from abusing the privileges of the administrator.
If an operator forgets his password, he may ask the administrator to log on the system and delete the
username and then redefine a username; or he may contact our customer service department or your
local distributor. If the administrator forgets his password, contact our customer service department or
your local distributor.
CAUTION
To ensure accurate test results, do not start measurement until the system status turns to Standby and
the system has been turned on for about 20 minutes, so that the light source and reaction temperature
gets steady.
Auto startup
You should specify the weekday and time for auto startup. When the set time is reached, the
system will start up automatically if it is off.
To set up auto startup timer
1 Select Utility > System Setup, and click Instrument F1.
2 Select Sleep/Awake.
3 Select Auto Startup Setup.
Figure 2.3 Auto Startup Setup window
NOTE
After setting up the auto startup time, ensure that the operation unit and the analyzer are connected to power
supply; otherwise, they cannot be started up automatically.
6 Select the help button in front of the new alarm messages to view relevant descriptions and
solutions.
7 Take actions according to the recommended solutions.
3 View the reagent status. When a reagent is insufficient or exhausted, the corresponding
chemistry name and chemistries left will be indicated as follows:
Yellow: indicates that the reagent is insufficient or expired, and the analysis will
continue. Refill or replace the reagent.
Red: indicates that the reagent is exhausted or at least one reagent type is not loaded,
and the analysis is stopped. Refill or replace the reagent.
4 View the calibration status. When the calibration is succeeded or failed, the Cal Status
column of the chemistry shows the calibration status in corresponding color.
Yellow: indicates that the calibration factors of the chemistry have been calculated, or
extended, edited or overridden.
Red: indicates that the calibration of the chemistry fails or expired, or the chemistry
needs to be calibrated.
5 Check the calibration time left. If it will be expired, perform calibration immediately.
For more information about calibration, refer to 2.3.2Calibrationon page 2-16.
Safety information
WARNING
The probe tip is sharp and may cause puncture wounds. To prevent injury, exercise caution when working
around the probes.
BIOHAZARD
Wear gloves and lab coat, if necessary, goggles.
Do not touch the reagent directly with your body; otherwise, skin wound or inflammation may be caused.
NOTE
Before loading biochemistry reagent, ensure that there are no air bubbles inside the reagent bottle so as to
avoid inaccurate test results.
If a chemistry is set with sample pretreatment, ensure that the chemistry reagent and the pretreatment reagent
are loaded to the same reagent carousel; otherwise, the chemistry cannot be run.
Manual load
When loading reagents manually, you need to enter the reagent information, which is the only
information source of the loaded reagents. If loaded reagents are bar-coded, the reagent
information cannot be edited; otherwise, all reagent information except for position, chemistry
and reagent type can be edited.
Manually loaded reagents have the letter “M” (Manual) appearing near them.
Figure 2.6 Flag for manually loaded reagents
(1)
(1)
CAUTION
If the system is running tests, click and wait until the system status becomes Pause before
removing the reagent carousel cover. Otherwise, probe collision or other error may occur.
2 Load reagents to the set positions according to the reagent load list, and then uncap the
reagent bottles.
3 Restore the sample/reagent carousel cover.
4 Select End Load F2.
5 Select Inventory F3 to check the volume of the loaded reagents and refresh the number of
tests left on the screen.
For more information on reagent inventory check, see 3.2.7 Checking and auto refreshing
reagent inventory on page 3-6.
Auto load
Auto load is to load bar-coded reagents to the reagent carousel, which are identified by bar code
scanning. The closed reagents can be loaded only through bar code scanning.
To load bar-coded reagents
1 Remove the sample/reagent carousel cover.
CAUTION
If the system is running tests, click and wait until the system status becomes Pause before
removing the reagent carousel cover. Otherwise, probe collision or other error may occur.
2 Place the reagents in idle positions of the reagent carousel and then uncap the reagent
bottles.
3 Restore the sample/reagent carousel cover.
4 Select End Load F2.
The system scans all reagent positions and read the reagent information from the bar code.
NOTE
When installing ISE reagent pack, do not twist, press and squeeze the pipes of the ISE Module otherwise
the ISE pipes may be clogged.
CAUTION
Use the concentrated wash solution specified by our company. Using other wash solutions may cause
inaccurate test result.
NOTE
Before loading wash solution, ensure that there are no air bubbles inside the reagent bottle to avoid
affecting washing effects.
8 Restore the sample/reagent carousel cover.
9 Click End Load F2.
1 Select Reagent > Reagent/Calibration, and click Load F1 to display the Load Reagent
window.
2 Or select Reagent > Reagent Carousel Status, click the position on the reagent carousel
graph to load the pretreatment reagent, and select Load F1.
NOTE
Pretreatment reagent can be only loaded on one reagent carousel. Make sure that it is on the same
carousel as the chemistry reagent set with sample pretreatment; otherwise, the chemistry cannot be run.
3 Enter the following information:
Bar code
Chemistry
Bottle type
Lot number
Serial number
Reagent type(R0)
Expiration date
4 Select Load F3, and select Exit F5 to close the window.
5 Remove the sample/reagent carousel cover.
6 Place the pretreatment reagent in the set position of the sample/reagent carousel.
7 Restore the sample/reagent carousel cover.
8 Select End Load F2.
2.3.2 Calibration
Calibration is performed to obtain calibration factors for calculate sample test results. The
calibration test procedure is as shown below:
Requesting calibration
Request ISE test
Request biochemistries
Preparing calibrators
Starting analysis
Yes
Performing QC test
Requesting calibration
Calibration request includes ISE test and biochemistries.
CAUTION
After changing electrodes or other consumables, perform a calibration. You are recommended to perform
calibration at least once every day to ensure accurate results.
2 Select ISE.
3 Click Calibrate F5.
To request biochemistry calibration
1 Select Reagent > Reagent/Calibration.
2 Select a reagent carousel from the Reagent Carousel drop-down list.
3 Select the down-arrow button on the right side of the screen to display the biochemistry
reagent/calibration screen.
Figure 2.10 Biochemistry Reagent/Calibration screen
Preparing calibrators
BIOHAZARD
Inappropriate handling of calibrators may lead to biohazardous infection. Do not touch the calibrators directly
with your hands. Wear gloves and lab coat, if necessary, goggles. In case your skin contacts the calibrators,
follow standard laboratory safety procedure and consult a doctor.
CAUTION
Do not use expired calibrators; otherwise, unreliable test results may be caused.
To prepare calibrators
1 Select Reagent > Reagent/Calibration.
2 Select a reagent carousel from the Reagent Carousel drop-down list.
3 Select the down-arrow button on the right side of the screen to display the biochemistry
reagent/calibration screen.
4 Select Load List F4.
The calibrator list shows all requested chemistries as well as calibrators, positions,
concentration, lot number and expiration date.
5 Select Print F7, and select Close F8.
6 Load calibrators to the sample carousel according to the calibrator list.
Starting analysis
After requesting calibrations and load calibrators to the sample carousel, you can start the
calibration test.
To start calibration test
1 Select on upper right corner of the main screen. The Start Conditions window is
displayed.
Figure 2.11 Start Conditions window
2.3.3 QC
QC results are tools used to monitor the system performance. To check if the system is running
normally and steadily, you are recommended to run control samples every day.
The QC test procedure is as shown below:
Figure 2.12 QC test procedure
Programming controls
Preparing controls
Starting analysis
Checking QC results
Yes
Programming controls
Before routine test, biochemistries, ISE test, calculations, and panels should be run with control
samples. Prior to programming controls, make sure that the QC parameters have been set
correctly; otherwise, the chemistries cannot be requested.
To program controls
1 Select Program > Quality Control
Figure 2.13 Quality Control screen
Preparing controls
BIOHAZARD
Inappropriate handling of control samples may lead to biohazardous infection. Do not touch the control samples
directly with your hands. Wear gloves and lab coat, if necessary, goggles. In case your skin contacts the control
samples, follow standard laboratory safety procedure and consult a doctor.
CAUTION
Do not use expired control samples; otherwise, unreliable test results may be caused.
To prepare controls
1 Select Program > Sample.
2 Select List F5.
The sample list shows all programmed patient samples, control samples and chemistries.
3 Select Print F7.
Samples and controls are printed out separately.
4 Select Exit F8.
5 Load control samples to the sample carousel according to the printed list.
Starting analysis
After programming and load the control samples, you can start the QC test.
To start QC test
1 Select on upper right corner of the main screen. The Start Conditions window is
displayed.
2 Select a sample carousel to which the control samples are loaded.
3 Select a reagent carousel to which the reagents are loaded.
4 Select OK to start analysis.
Checking QC results
After the QC test is complete, check if the test results are within the normal range and the data
points on the QC chart are normal. If you see any abnormity, troubleshoot the error immediately.
To check QC results
1 Select Result > Current, and click the By Sample option button.
2 Check for result flags in the result list.
If you see result flags, troubleshoot the error according to 11.4.1 Data alarms and
corrective actions on page 11-9.
3 Select QC > Levey-Jennings or Twin-Plot, and check if the data points on the charts are
normal.
4 After taking actions, you can start the routine test.
BIOHAZARD
Inappropriate handling of samples may lead to biohazardous infection. Do not touch the samples directly with
your hands. Wear gloves and lab coat, if necessary, goggles. In case your skin contacts the samples, follow
standard laboratory safety procedure and consult a doctor.
CAUTION
Do not use expired samples; otherwise, unreliable test results may be caused.
NOTE
Before loading sample, ensure that there are no air bubbles inside the sample cup so as to avoid inaccurate
test results.
2 Input the sample information, including: sample ID, carousel No. and position, STAT
property, sample type, comment, and patient ID.
3 Select chemistries and panels to be run.
4 To input patient information, click Demog F1.
5 To set number of replicates and dilution factors, click Options F2.
6 Click Save F8.
7 To program more samples, repeat steps 2-6.
1 Select on upper right corner of the main screen. The STAT Sample Program window
is displayed.
Figure 2.15 STAT Sample Program window
2 Input the sample information, including: sample ID, carousel No. and position, sample type,
and sample cup.
3 Confirm the default chemistries.
4 To select more chemistries, perform the following steps:
a. Click Chems F3.
b. Select chemistries and panels to be run for the samples.
c. Click Save F7.
5 To input patient information, click Demog F1.
6 To set number of replicates and dilution factors, click Options F2.
7 Click Save F7.
8 To program more samples, repeat steps 2-7.
9 Click Close F8 to close the window.
To prepare samples
1 Select Program > Sample.
2 Select List F5.
The sample list shows all programmed samples, controls and chemistries.
3 Select Print F7.
Samples and controls are printed out separately.
4 Select Exit F8.
5 Load samples to the sample carousel according to the printed list.
To start sample analysis
1 Select on upper right corner of the main screen. The Start Conditions window is
displayed.
2 Select a sample carousel to which the samples are loaded.
3 Select a reagent carousel to which the reagents are loaded.
4 Select a patient sample range: All or Partial. When you select Partial, you should specify a
sample position range for analysis.
5 Select OK.
Adding samples
You can add routine sample and STAT sample at any time.
For details about adding samples, see 6.2.4 Adding samples on page 6-6.
Adding chemistries
You can add chemistries to samples of any status. Whether to change the program information
will be determined based upon the sample status.
For details about adding chemistries, see 6.2.5 Adding/Modifying chemistries on page 6-7.
Rerunning samples
The system supports manual rerun and auto rerun. Manual rerun can be performed through
the List window and the Current or History screen. Auto rerun is based on the set critical
range of the ISE test and the rerun conditions of biochemistries. When the conditions are met,
the relevant chemistries will be rerun automatically.
For details about rerunning samples, see 6.2.6 Rerunning samples on page 6-7.
2 View the status of calibrators, controls and samples on the sample carousel graph.
Refer to the explanations of various sample statuses on the lower-right corner of the
screen.
3 To view the detailed information of certain sample, select the sample position on the sample
carousel graph.
The detailed information of the selected sample position is displayed on the right side of the
screen.
3 Check the reagent volume status according to the explanation in the middle of the carousel
graph. If a reagent is insufficient or exhausted, replace it immediately.
For instructions of loading reagent in Running status, see 3.2.3Loading biochemistry
reagents in Running status on page 3-4.
4 To view the detailed information of certain reagent, select the reagent position on the
reagent carousel graph.
The detailed information of the selected reagent position is displayed on the right side of
the screen.
5 Select the following buttons to perform respective operations:
Switching carousels
Switching carousels means changing sample carousel and reagent carousel during
measurement, so that the samples and reagents on them can be tested.
Switching sample carousel
The system supports 10 virtual sample carousels, on all of which samples can be programmed in
order to improve the test efficiency.
After samples on multiple sample carousels are programmed, if those on the current sample
carousel are about to finish sample dispensing during test process, the screen shows the
countdown for sample loading and the Start Conditions window pops up. Select desired sample
carousel, load samples to it, and then select OK to resume the test.
Switching reagent carousel
The system supports 2 virtual reagent carousels, on both of which biochemistry reagents, wash
solution and physiological saline can be loaded. However, reagents of the same chemistry must
be loaded on the same carousel, and only the chemistries on the same carousel can be tested in
every batch of tests.
To run chemistries on the other reagent carousel, click to display the Start Conditions
window. Select the other reagent carousel, load reagents to it, and then select OK to resume the
test.
Pause
"Pause" means to stop addition of sample and reagent temporarily during test process, so that
you can load/unload sample and reagent on the carousel. After you click , when the started
tests finish sample/reagent dispensing, the system enters the Pause status. Then you can start
loading/unloading sample and reagent.
Emergency stop
Emergency stop will terminate all measurements on the instrument, and all tests that are not
finished yet will be invalidated. Do not use emergent stop unless it is really needed, for example,
system failure. Emergency stop can be performed in any system status.
Select the icon on upper right corner of the screen, and then select OK. All unfinished
actions of the system are cancelled, all pumps and valves are turned off, and the system enters
the Stopped status.
To restore system failure, select Utility > Commands, and then select Home. To resume the
analysis, select the icon.
Daily maintenance
Powering off
Operations after powering off
NOTE
When the analyzing unit power is switched off, the refrigeration system is still running. If you are going to
store the system for over 7 days, switch off the main power.
1 Remove the sample/reagent carousel cover, and then remove the calibrators, controls and
patient samples.
2 Check the analyzer panel for stains and wipe them off with clean gauze if any.
3 Check the high-concentration waste tank and the low-concentration waste tank. Clear them
if necessary.
This chapter describes reagent and calibration operations related to ISE and biochemistry tests.
All set biochemistry reagents are displayed. The reagent name is indicated by different colors
according to the reagent volume and loading status.
Yellow: indicates that the reagent is insufficient or expired.
Red: indicates that the reagent is exhausted or at least one reagent type is not loaded.
The following procedure is only applicable to unloading the reagents without bar code; for those
reagents with barcode, when the reagents are taken away from the reagent carousel, they are
unloaded automatically.
To unload biochemistry reagents
1 Select Reagent > Reagent/Calibration, and select the up and down arrow buttons to
display the biochemical regent/calibration screen.
2 Or select Reagent > Reagent Carousel Status.
3 Select the desired reagent.
4 Select Load F1, and then Unload F4.
5 Remove the sample/reagent carousel cover, take out the reagent , and restore the cover.
6 Select Exit F5 to close the window.
7 Select End Load F2.
3 To display a header on the reagent/calibration screen, select the check box on the left.
4 To cancel displaying a header, deselect the corresponding check box.
5 Select Up and Down to adjust the display order of the reagent information.
6 Select Save to save the settings.
7 Select Exit to close the window.
All reagents of selected chemistry: check the inventory of all reagent types of the
selected chemistry.
4 Select Check.
The reagent carousel graph refreshes the reagent status automatically.
The Reagent/Calibration screen refreshes the Tests Left of the selected chemistry,
Chems Left, and the Volume of the selected wash solution.
Canceling reagent inventory check
To cancel reagent inventory check, select Close on the Check window, and then select No
Invent. F3 on the Reagent/Calibration screen, or on the Reagent Carousel Status screen.
Auto refreshing reagent inventory
1 Select Utility > System Setup.
2 Select Instrument F1, and then select Reagent/Calibration Setup.
3 Select the option Auto Refresh Reagent Inventory, which is unselected by default.
4 Click Save.
5 Click Exit to close the window.
This chapter provides calibration setup, calibration status, and calibration result recall of ISE
test and biochemistry.
5 Select OK.
Defining a calibrator
The system allows the definition of up to 99 calibrators. You are allowed to add, edit and delete
calibrators only when the system status is not Running.
To define a calibrator
1 Select Reagent > Cal Setup.
NOTE
Calibrators of a chemistry must be placed and analyzed on the same sample carousel.
5 Select Save to save your input information.
6 To define more calibrators, click New and then repeat step 3 to 7.
7 Select Close to exit the window.
8 To edit a calibrator, select it, click Edit F2, and then change the settings as the steps above.
Importing a calibrator
Calibrator parameters such as calibrator name, lot number, expiration date, concentration of
each chemistry and dilution parameter can be imported.
To import a calibrator
1 Select Reagent > Cal Setup.
2 Select Define F1.
3 Select Import and insert USB drive.
4 Select the path of .cif file.
Only .cif file can be imported; each .cif file stores information for one calibrator and each
time only one calibrator can be imported. When the system reads the calibrator
information, the following window is displayed:
5 Select the chemistries you want to import. All chemistries are selected by default.
6 Enter the lot number, select OK, and then select Close.
When one calibrator correponds to two or more sets of reagent paramters, enter the lot
number of the reagent.
7 Select Save and then Close.
You are allowed to edit or delete the calibrator dilution factors when the system is not running
any tests.
To set up calibrator dilution factors
1 Select Reagent > Cal Setup.
2 Choose the desired calibrator and chemistry.
3 Select Dilute F5.
Figure 4.6 Calibrator Dilution Setup window
4 Set up the unit, concentration, aspirated volume, neat sample volume, and diluent volume.
The input of aspirated volume and neat sample volume must be an integer multiple of
0.1 within 2 μL - 35 μL. The aspirated volume is required.
The input of diluent volume must be an integer multiple of 0.5 within 100 μL - 200 μL.
This field can be left blank.
If the neat sample volume and diluent volume are defined, ensure that the sum of the
two volumes is within 125 μL - 235 μL.
5 Select Save.
6 To edit the dilution factors, select the number button on the left, click Edit, and change the
settings.
7 To delete the dilution factors, select the number button on the left and click Delete.
8 Select Close to exit the window.
Auto calibration
Based on the auto calibration conditions, the system can determine chemistries that need to be
calibrated and remind you through calibration status and color indication.
Setting up auto calibration
1 Select Reagent > Cal Setup.
2 Select Rules F4.
3 Choose a chemistry from the Chem drop-down list.
4 Choose auto calibration conditions:
Bottle changed
Lot changed
Unavailable for closed chemistries, of which calibration will be run automatically when
reagent lot number is changed.
Calibration time
NOTE
If the Manage Reagents by Lot option on the System Setup screen is enabled, Bottle Changed
and Lot Changed will not appear. When a different reagent lot is used, the system will request and run
calibration automatically.
5 Select Save F7.
Auto calibration reminding
When the auto calibration conditions are satisfied, the system will remind you through the
calibration status, prompt message and color indication.
If you choose the Bottle Changed option, the system will display a message indicating
calibration is required when you use a different bottle of reagents.
If you choose the Lot Changed option, the system will display a message indicating
calibration is required when you use reagents of a different lot.
If you choose the Cal Time option, the system will remind you in 30 minutes before the
calibration is timed out and display the chemistry name and calibration status with yellow.
Removing auto calibration
1 Select Reagent > Cal Setup.
2 Select Rules F4.
3 Choose a chemistry from the Chem drop-down list.
4 Deselect all auto calibration conditions.
5 Select Save F7.
Deleting a calibrator
You are allowed to remove the calibrators other than WATER. When a calibrator is deleted, all
calibration settings and its position are cleared, and it cannot be used for programming. The
stored test results of the calibrator can be recalled according to the chemistry name. Only
calibrators that are not requested or run can be deleted.
To delete a calibrator
1 Select Reagent > Cal Setup.
2 Choose a calibrator you want to remove.
3 Select Delete F6.
4 Select OK. The selected calibrator is deleted.
The response value current displayed is the updated reagent blank response.
4 Select the reaction data table to view the reagent blank reaction data.
Calibrator response
K factor (for linear calibrations only)
5 Select the calibration time range.
6 Select Search F1.
The graphical trend of the selected chemistry within the specific period is displayed.
7 Select the Tabular Trend tab to view the trend data.
Figure 4.11 Tabular Trend tab page
4 Select a point on the curve. Relevant measuring period and absorbance are displayed on
the right of the window.
5 Select a filter condition from the following options:
None: observe reaction curve and data in the default mode.
Chemistry: observe reaction curve of the results for the selected test.
Calibrator: observe reaction curve of the results for the selected calibrator.
6 Choose the Reaction Data tab to view the reaction data.
The window shows the calibrators and reagents used in calibration, and reagents for
reagent blank test.
2 Select Close to exit the window.
5 Select OK.
Calibration override
The Calibration Override option allows the system to override a failed calibration and calculate
results based on the failed calibration factors. Calibration override is only applied to failed
calibrations. Results that are obtained based on failed calibration factors will be flagged with
"OVE".
CAUTION
Before overriding a calibration, make sure that the calibration factors are within the acceptance limits of your
laboratory. The magnitude of the error should be totally under the control of your laboratory. Use of overridden
calibration factors may lead to unreliable results and influence the doctor’s diagnosis. Think twice before
overriding a failed calibration.
To override a calibration
1 Select Reagent > Reagent/Calibration.
2 Choose a chemistry you want to override.
3 Select Cal Options F8.
4 Select Calibration Override from the Calibration Options window.
5 Select OK. The failed calibration factors of the selected chemistry can be used for result
calculation.
Removing Cal Overridden status
Recalibrate the chemistry to remove its Cal Overridden status.
Reject
If the current calibration fails but sample analysis needs to be performed immediately, you may
use the Reject function to reject the current calibration factors, and use the latest valid ones for
calculating sample results, which will be flagged with "CALJ". Calibration factors of status other
than Requested and Cal Required can be rejected. Rejected calibration factors cannot be
rejected again.
Rejecting a calibration
1 Select Reagent > Reagent/Calibration.
2 Choose a chemistry you want to reject.
3 Select Cal Options F8.
4 Select Reject from the Calibration Options window.
5 Select OK. Calibration factors of the selected chemistry are rejected.
Removing Reject status
Recalibrate the chemistry to remove its Reject status.
5.1 Overview
QC test is performed on samples provided with known concentration range of various analytes
by authority divisions or reagent suppliers. By comparing with the given range, the test results
obtained on this instrument can be used to judge if the instrument is in normal status and the
sample results are reliable.
To ensure the system performance, run control samples every time after you perform a
calibration, or change the reagent lot, or maintain and troubleshoot the instrument.
5.1.1 QC procedure
After you define a chemistry, control, and QC rules there is no need to edit them frequently, and
you are only required to run control samples every day to make sure that the system works well.
Run control samples according to the following procedure:
Figure 5.1 QC procedure
Program Running
Load control Recall QC Print real-time
control control
Daily operations samples results QC results
samples samples
5.2 QC setup
Perform QC settings in the following order:
Define a control
Set up control concentrations
Set up QC rules
Auto QC
You can delete controls while the system is in non-test status.
3 Set up the control name, No., lot number, expiration date, and sample type.
4 Assign positions for the control.
You are allowed to assign one position of each sample carousel for the control.
5 Select OK to save your input information.
6 To define more controls, select New and repeat step 3 to 5.
7 Select Exit to exit the window.
5.2.4 Auto QC
The system provides the auto quality control function. The conditions for auto quality control
include:
Number of samples: indicates the number of patient samples. After the given number of
samples is finished, the system will run the selected control(s) automatically.
When calibrated: The system will automatically run the chemistry for the selected control(s)
every time when the chemistry is calibrated. Auto QC is not applicable to non-measurement
calibrations, such as recalculation and editing.
When the control samples automatically run are selected, all chemistries configured for the
control samples will be run.
To set up and run auto QC
1 Select Utility > System Setup.
2 Select Instrument F1.
3 Choose 9 QC Evaluation.
Figure 5.4 QC Parameters window
When Calibrated: select the checkbox to allow the system to run controls when a
chemistry is calibrated.
6 Choose controls to be run automatically.
One or more controls can be selected.
7 Select OK.
During test, the system will insert QC runs automatically once the conditions are met.
To remove auto QC status
To remove an auto QC status, clear the auto QC settings on the QC Parameters window.
3 Select Search F1. The twin-plot chart area displays the recent 10 results of control X and
control Y for the chemistry.
Figure 5.9 Twin-Plot screen
4 Select a point on the curve. Relevant measuring period and absorbance are displayed on
the right of the window.
5 Select a filter condition from the following options:
None: observe reaction curve and data in the default mode.
Chemistry: observe reaction curve of the results for the selected test.
Control: observe reaction curve of the results for the selected control.
6 Select the Reaction Data tab to view the reaction data.
Figure 5.12 Control reaction data
Next F6: to view the reaction curve and data of the next QC test.
Print F7: to print the current reaction curve or data.
8 Select Close F8 to close the window.
To add QC comments
1 Search for desired QC results on the Results screen.
2 Choose a QC result in the result list.
3 Select Comment F5.
4 Type in comments for the selected QC result.
Up to 100 characters can be entered.
5 Select OK.
To archive QC data
1 Search for desired QC results on the Results screen.
2 Select Archive F6.
3 Select OK.
CAUTION
Prepare the sample according to the procedure recommended by the tube manufacturer. For collection and
preparation of samples, please see the reagent Instructions for Use. Use clean tubes, microcups and other
disposable materials specified by the manufacturer. Do not reuse disposables.
When using vacuum collection tube for sample collection, make sure that the cap of the vacuum collection tube
is clean.
Sample container types
The sample carousel supports blood collecting tube, centrifugal tube, plastic tube and Microtube,
which are available in the following specifications:
Microtube: Φ14×25 mm, 0.5 mL (Beckman); Φ14×25 mm, 2 mL (Beckman); Φ12×37 mm, 2
mL (Hitachi).
Blood collecting tube or plastic tube: Φ12×68.5 mm, Φ12×99 mm, Φ12.7×75 mm,
Φ12.7×100 mm, Φ13×75 mm, Φ13×95 mm, Φ13×100 mm.
For the tests of the whole blood(centrifuged),onlyΦ12×68.5 mm, Φ12×99 mm, Φ12.7×75 mm,
Φ12.7×100 mm, Φ13×75 mm, Φ13×95 mm, Φ13×100 mm anticoagulation tubes can be used.
The sample height in the tube should be no higher than 55mm and the blood cell level should be
no lower than 10mm. Microcups are not allowed. To ensure the clinical performance and avoid
the system alarm, EDTA anticoagulation tubes are recommended.
Sample volume
The amount of sample required for a common measurement is 2-35 μL, with an increment of
0.1 μL. Analysis with insufficient samples may lead to inaccurate results.
If a sample is exhausted during the analysis, the system will automatically invalidate all
incomplete chemistry of the sample. Before running samples, make sure that they are sufficient
in volume for analysis.
Loading samples
BIOHAZARD
Wear gloves and lab coat, if necessary, goggles.
To load samples
1 Check if the sample inside the sample tube is sufficient for analysis and the bar code label is
applied correctly.
2 Check the system status.
Unloading samples
BIOHAZARD
Wear gloves and lab coat, if necessary, goggles.
To unload samples
1 Check if the sample carousel and the sample probe have stopped moving.
Latest samples: to download samples that are programmed on the current day but have
not been downloaded.
Samples with the following IDs: to download samples with the specified program date
and ID. Enter the sample IDs or ID range to download.
Sample with the following bar code: to download the sample with the specified bar code.
Enter the bar code of the desired sample.
3 Select OK.
4 Confirm the sample information and selected chemistries/panels on the Sample List
screen.
5 Load the samples to idle positions of the sample carousel.
6 Select the icon, set the test conditions, select the Sample Crsl Bar Code check box,
and then click OK to start analysis.
To download non-bar-coded samples
1 After downloading samples from LIS, select List F5.
2 Select Unpositioned F2,and select Assign.
3 Select the date the desired samples are programmed.
4 Type in the single sample ID or ID range in the ID field.
5 Choose a sample carousel on which you will place the sample.
6 Enter the sample position.
The options include all available positions of the selected sample carousel.
To assign position for single sample, input the position number in the first edit box.
To assign positions for multiple samples, enter the start position number in the first edit
box, and then the end position number in the second edit box. The system will assign
positions for the samples ascending according to the sample ID.
7 Select OK.
8 Load the samples to the assigned positions on the sample carousel.
9 Select the icon, set the test conditions, and then click OK to start analysis.
If your system is not connected with a LIS host, you can program bar-coded samples with the
default panel or program them manually one by one or by batch. This section describes two
methods of manually programming samples without LIS.
For sample analysis with default panel, see 7.7.5Setting up and running default panel on page
7-27.
Before processing bar-coded samples, check if the following conditions are satisfied:
The sample bar code reader has been configured.
The Sample Crsl Bar Code check box on the Sample Bar Code window is selected.
The system status is Standby or Pause.
To process bar-coded samples without LIS -- Method 1
1 Program samples manually according to "2.4.1Programming and processing samples"
(Page 2-23).
7 Select the icon, set the test conditions, and then click OK to start analysis.
10 Select the icon, set the test conditions, and then click OK to start analysis.
BIOHAZARD
Inappropriate handling of samples may lead to biohazardous infection. Do not touch the samples directly with
your hands. Wear gloves and lab coat, if necessary, goggles. In case your skin contacts the samples, follow
standard laboratory safety procedure and consult a doctor.
CAUTION
Do not use expired samples; otherwise, unreliable test results may be caused.
3 Select the icon, set the test conditions, and then click OK to start analysis.
If the system is in Standby status, select the icon, set the test conditions, and then
click OK to start analysis.
3 Type in the ID or bar code of the sample you desire to rerun or enter the barcode of the
sample.
4 Click Select.
Figure 6.4 Rerun Samples window
Separate single samples with comma, e.g. 5, 7, 9; and connect multiple continuous samples
with a dash, e.g. 1-3.
4 Select Batch.
Figure 6.5 Rerun Batch window
8 Load samples to the assigned positions, and select to start the analysis.
If the sample is on the current sample carousel, it is analyzed automatically.
Otherwise, you should specify the sample carousel and position to start the analysis.
To rerun samples by chemistry
1 Select Result > Current Results or History Results, and choose the By Chemistry option.
2 Search for desired sample results.
3 Choose the chemistry and samples you desire to rerun.
4 Select Rerun F5.
5 To run sample blank for all samples, select the Sample Blank check box.
6 Modify the following information for single sample:
Sample volume (2-35)
Predilution factor (4-100)
Off-line dilution factor (2-9999)
Sample blank
Pretreatment
7 Select OK.
8 Load samples to the original positions, and select to start the analysis.
If the samples are on the current sample carousel, they will be analyzed automatically.
Otherwise, you should specify the sample carousel and position to start the analysis.
5 The latest rerun result is the default one. To change the default result, choose a result, and
then select Set Defaults.
The Default column of the result shows Y, which stands for Yes.
6 Select Exit to exit the window.
For the test of whole blood (centrifuged) sample, the parameters can be set up on the chemistry
parameter screen or on the chemistry options window and rerun window. On the chemistry
parameter screen, once sample pretreatment is selected, all the tests of this chemistry will
perform pretreatment. You can enable or disable pretreatment on the chemistry options window
or rerun window so that manually pretreated sample can be tested.
Preparing sample
Put the centrifuged whole blood sample (2000rpm,5min) into the sample positions.
NOTE
For the tests with blood cell pretreatment, please prepare the centrifuged whole blood sample. For the
centrifuged whole blood sample,onlyΦ12×68.5 mm, Φ12×99 mm, Φ12.7×75 mm, Φ12.7×100 mm, Φ13×75 mm,
Φ13×95 mm, Φ13×100 mm anticoagulation tubes can be used. The sample height in the tube should be no
higher than 55mm and the blood cell level should be no lower than 10mm. Microcups are not allowed. To
ensure the clinical performance and avoid the system alarm, EDTA anticoagulation tubes are recommended.
If manual pretreatment is required due to abnormal test results, please deselect the pretreatment option on the
chemistry options window and the rerun window.
Sample cup
Replicates
Off-line dilution
Predilution
Sample blank
Pretreatment
11 Select Save F8.
12 Click .
(1)
(2)
(3)
The figure above shows the absorption spectrum of interferents in serum samples. (1) refers to
lipemia, (2) refers to hemolysis, and (3) refers to icterus.
The three interferents are selective to wavelength and have complex absorption spectrums.
They cannot be removed completely by means of double-wavelength measurements. The serum
index option can be used to analyze the interferents contained in samples, helping clinical
professionals to evaluate the test results.
Six wavelengths are chosen to determine the serum index. The equations of serum index are as
follows:
Lipemia: primary wavelength of 660, secondary wavelength of 700.
AL A660 A700 , lipemia index: L 1 / C AL
Hemolysis: primary wavelength of 570, secondary wavelength of 605.
AH A570 A605 , hemolysis index: H 1/ A ( AH B AL )
Icterus: primary wavelength of 450, secondary wavelength of 505.
AI A450 A505 , icterus index: I 1/ D [ AI E ( AH B AL ) F AL ]
Where,
B and F: determined by the absorption spectrum of lipemia.
E: determined by the absorption spectrum of hemolysis.
C: determined by single lipemia.
A: determined by single hemolysis.
D: determined by single icterus.
4 Type in the print name of lipemia in the Print Name of the Lipemia area. Up to 15
characters can be entered.
The lipemia index will appear as the print name on patient reports and as “SI” on other
reports.
5 Repeat step 4 to define print names for hemolysis and icterus.
6 Select Save F7.
7 Select Close F8.
1 Select Utility-Chemistries.
2 Choose the SI chemistry.
3 Select Define F1.
4 Mark the Use Qualitative Result checkbox in the Lipemia area.
The Range and Flag fields below are activated for editing.
5 Type in the detection range in the first edit box of the Range field, and then enter a flag in
the first edit box of the Flag field.
For instance, type in “10” in the first edit box of the Range field in the Lipemia area, and
then enter “+” in the Flag field of the same row. If the lipemia volume (L1) contained in a
sample is lower than 10, the “+” sign will be added to the result in the patient report. Type
in “20” in the second edit box below the Range icon and “+-” in the second edit box below
the Flag icon. If the lipemia volume (L2) is greater than 10 and lower than 20, the result
will be flagged with the “+-” sign. The cycle continues. If the result is greater than L5, the
six flag will appear on the patient report.
6 Repeat step 4-5 to define ranges and flags for hemolysis and icterus.
7 Select Save F7.
8 Select Close F8.
Sample list
The sample list shows all patient and controls samples that have been programmed but not
analyzed yet. On the Sample List screen, you can search samples, assign position for
unpositioned samples, download program information from LIS, rerun tests, and print the
sample list.
For assigning position for unpositioned samples, see 6.4.3 Viewing unpositioned samples on
page 6-20.
For downloading program information from LIS, see 6.2.1 Processing samples with LIS on page
6-3.
For manual rerun, see Manual rerun on List window on page 6-7.
Chemistry list
The chemistry list shows the summary of chemistries that are requested on the current day or
requested before but not finished yet. On the Chemistry List screen, you can download
program information from LIS and rerun tests.
For downloading program information from LIS, see 6.2.1 Processing samples with LIS on page
6-3.
For manual rerun, see Manual rerun on List window on page 6-7.
7 Select the icon, set the test conditions, and then click OK to start analysis.
4 Find desired sample information and mark the corresponding Customize checkbox.
Click the checkbox again to deselect it.
5 Select Save.
6 Select Exit to close the window.
4 Select the desired information and the default value and then click Add.
5 Select the desired information and click Delete to delete it from the demographics list.
6 Select Up, Down, Home and End button to adjust the displayed order of patient
demographics.
7 Select OK to save the settings.
8 Select Exit to close the window.
Result optimizing will not affect storage, transmission and archiving of results. Only users who
have the permissions of system setup are allowed to optimize result display.
To optimize result display
1 Select Utility > System Setup.
2 Select Instrument F1.
3 Select Optimize Result Display.
Figure 6.21 Optimize Result Display window
4 Find desired chemistry, and mark the corresponding Low and High checkboxes.
5 To optimize result display of all chemistries, select Select All.
6 To cancel all settings, select Clear.
7 Select Save.
8 Select Cancel to close the window.
The sample type includes R, E and C. R stands for routine sample, E for STAT sample, and
C for control.
The Host column indicates the transmission status of the sample. Y means that the
sample has been sent to the LIS host, and N means the opposite.
The Print column indicates the print status of the sample. Y means that the sample has
been printed, and N means the opposite.
When certain test of a control sample or patient sample triggers a data alarm, the sample
will appear in yellow. The pretreated chemistries will be highlighted in purple.
2 Choose a result recall mode:
By sample
By chemistry
3 When recalling results by sample, choose a sample in the left list. The right list displays all
results of the sample.
4 When recalling results by chemistry, choose a chemistry in the left list. The right list
displays all results of the chemistry.
5 Choose the following buttons as needed:
Search F1: to inquire sample results.
Options F2: to delete, edit and print samples, recall rerun results, customize result
display options, recalculate results, compensate results, archive results, and observe
result trend.
Demog F3: to view patient demographics of the sample.
Reac Curve F4: to view the reaction curve and data of the selected test.
Rerun F5: to rerun a finished sample.
Review F6 to review the sample result.
Print F7: to print sample results.
Host F8: to transmit the selected sample results to the LIS host.
To recall current results
1 Select Result > Current.
2 Select Search F1.
Figure 6.23 Recall results window
The sample type includes R, E and C. R stands for routine sample, E for STAT sample, and
C for control.
The Hostcolumn indicates the transmission status of the sample. Y means that the
sample has been sent to the LIS host, and N means the opposite.
The Print column indicates the print status of the sample. Y means that the sample has
been printed, and N means the opposite.
2 Choose a result recall mode:
By sample
By chemistry
3 Select Search F1 to search for desired results.
4 When recalling results by sample, choose a sample in the left list. The right list displays all
results of the sample.
5 When recalling results by chemistry, choose a chemistry in the left list. The right list
displays all results of the chemistry.
6 Choose the following buttons as needed:
Options F2: to delete, edit and print samples, recall rerun results, customize result
display options, compensate results, archive results, and observe result trend.
Demog F3: to view patient demographics of the sample.
Reac Curve F4: to view the reaction curve and data of the selected test.
Rerun F5: to rerun a finished sample.
Review F6 to review the sample results.
Print F7: to print sample results.
Host F8: to transmit the selected sample results to the LIS host.
To recall history results
1 Select Result > History Results.
2 Select Search F1.
3 Select the program date range you want to recall. Select the start date in the first box and
the end date in the second box.
4 Enter one or more search conditions.
5 Select OK. The samples matching the condition are displayed on the screen.
6 Select a function button to perform relevant operations.
5 Select a point on the curve. Relevant measuring period and absorbance are displayed on
the right of the window.
6 Select a filter condition from the following options:
None: observe reaction curve and data in the default mode.
Chemistry: observe reaction curve of the results for the selected test.
Sample: observe reaction curve of the results for the selected sample.
7 Choose the Reaction Data tab to view the reaction data.
Figure 6.28 Sample reaction data
Prev F4: to view the reaction curve and data of the previous test.
Next F5: to view the reaction curve and data of the next test.
Print F7: to print the current reaction curve or data.
9 Select Close F8 to close the window.
Viewing reagent information
On the reaction curve window, you are allowed to view the reagents in sample measurement,
the calibrators and reagents used in calibration, and reagents for reagent blank test.
To view reagent information
1 Select Reagent F1 on the Reaction Curve window.
Figure 6.29 Reagent window
The window shows the calibration date and time; sample measurement date and time;
calibrators, reagents for reagent blank test; and reagents for sample analysis.
2 Select Close to exit the window.
Adjusting display range
The maximum absorbance display range of reaction curve can be adjusted automatically or
manually. The adjustment is only applicable to the currently-displayed curve, which will restore
the default display when opened next time.
To adjust display range
1 Select Adjust F3 on the Reaction Curve window.
CAUTION
Edit Results function gives doctors with freedom to modify the results, and therefore, must be used with
cautions. Only users that have sufficient permissions are allowed to edit results.
To edit results
1 Select Result > Current or History.
2 Choose a result recall mode:
By sample
By chemistry
3 Select Search F1 to search for desired results.
4 Choose a sample or chemistry in the sample list which includes the off-system chemistries
as well.
5 Select Options F2, and select Edit Results.
The screen shows the samples or chemistry and all measured results.
6 Choose a chemistry to edit, and then input result in the Final Result column.
By sample
By chemistry
3 Select Options F2, and select Customize Result Display.
b. Choose desired header names in the Result List Setup area. Use the Up and Down
buttons to adjust the display order of the header names.
6 Select Save to save the settings and close the window.
To compensate results
1 Select Result > Current or History.
2 Choose the By Chemistry option.
3 Choose the chemistry that you want to compensate in the left list.
4 Select Options F2, and select Compensate Results.
All results of the chemistry are displayed in the list at the bottom.
Figure 6.41 Compensate window
6 Move the cursor to certain point on the graphic trend. The actual result, final result,
completion time, reagent lot number, serial number, and calibration time are displayed on
the right of the window.
7 To show all results of repeated analysis or rerun tests, select the Include Replicate Results
checkbox.
8 To observe result trend of other sample tests, select Prev F1 or Next F2.
9 Select Exit F8 to close the window.
NOTE
It may take a long time to archive a large amount of results. You are recommended not to archive results
over one week each time.
2 Select Options F2.
3 Select Archive.
4 Select OK.
By Sample: To view all requested samples and the quantity of its requested chemistries.
By Test: To view test requisitions and reagent volume for the chemistries.
3 Select or enter the start date and end date in the Date field. the start date cannot be later
than the end date.
4 Select Search F1.
All samples or tests requested during the period are displayed in the middle list of the Tests
screen.
5 Select Print to print out the currently displayed statistic information of measurements.
7 Select Print F7 to print out the statistic graph and statistic data.
This chapter describes the setup methods of closed-/open-reagent chemistries and special
chemistries, as well as the extended chemistry functions.
CAUTION
While importing chemistries, do not switch off the analyzing unit main power or exit the operating software.
If an imported chemistry is no longer needed, it can be deleted with the Delete F2 button on the
Chemistries screen.
All chemistries contained in the default parameter form are displayed in the Available
Chemistries list.
4 Use the following buttons to import desired chemistries:
Add All>>: add all chemistries in the Available Chemistries list to the Imported
Chemistries list.
Add ->: add the selected chemistries in the Available Chemistries list to the Imported
Chemistries list.
<-Remove: remove the selected chemistries from the Imported Chemistries list.
<<Remove All: remove all chemistries from the Imported Chemistries list.
5 Select Import.
All imported chemistries are enabled by default and can be used for measurement. If the
result unit is changed, the corresponding chemistry must be recalibrated.
6 Select Exit.
Twin chemistries
Pretreatment parameters
7 Select Exit.
Defining a chemistry
Up to 200 chemistries can be defined.
To define a chemistry
1 Select Utility > Chemistries.
2 Choose a blank frame in the chemistry list, and select Define F1.
Figure 7.3 Chemistry Type window
Chem
Chemistry name is the only identity of a chemistry and must not be duplicate. A chemistry name
can be composed of up to 10 characters.
No.
No. is a unique number for chemistry. It can be left blank but must not be duplicate. Chemistry
number is composed of numbers, and it ranges from 1-400 for open-reagent chemistries.
Sample type
Sample type refers to the samples to which the chemistry is applicable. The options include
serum, plasma, urine, CSF and other. The options available in the Sample Type drop-down list
are those supported by the chemistry, and the default is the default sample type.
The system allows definition of chemistry parameters for more than one sample type, including
the processing parameters and error detection limits. During definition of chemistries, the
parameters should be firstly defined for serum sample, and then other sample types. Such
chemistries will be calibrated with serum sample parameters by default.
Chemistry
Chemistry is the complete form of chemistry name. It can be composed of up to 50 characters.
The input is not case sensitive. The Chemistry field can be left blank or duplicate.
A chemistry is only represented by its print name on patient reports and appears on other
reports in the form of short name.
Print Name
Print name is displayed on patient reports representing a chemistry. It can be composed of up to
15 characters. The print name can be edited and duplicate. When this field is left blank, the
short form of the chemistry name will appear on reports. A chemistry is represented by its
short name on all reports other than patient reports.
Reaction Type
Reaction type is a measurement theory based on which chemistries are run for samples and
then calculated. The system supports three reaction types, which are Endpoint, Fixed-time and
Kinetic.
Table 7.1 Reaction types
Reaction Type Description
Endpoint Qualitative analysis is performed based on the absorption spectrum and
absorbed light intensity of the reactant when the reaction becomes
equilibrious.
Fixed-time For this reaction type, the reaction velocity is directly proportional to the
substrate concentration. As the substrate is consumed continuously, the
reaction velocity is decreasing gradually, and so is the absorbance change
rate. It will take a long time for such reaction to become equilibrium, and
the reaction can get steady only after a delay.
Kinetic Kinetic, also called continuous monitoring method, is used to continuously
measure the multiple change points of a reactant or substrate’s
concentration which varies with the enzymatic reaction, thus calculating
the initial velocity of the enzymatic reaction and then the enzyme activity.
This reaction type is mainly used for measurement of enzyme activity.
Reaction Direction
Reaction direction refers to the change trend of absorbance during the reaction process, and
includes two options:
Positive: indicates increasing absorbance with time.
Negative: indicates decreasing absorbance with time.
Primary Wavelength
The primary wavelength is chosen based on the light absorption features of the reactant and
used to measure the absorbed light intensity.
Table 7.3 Input range of incubation time, blank time and reaction time for fixed-time and Kinetic analysis
Fixed-time and Incubation Blank time Reaction time K
Kinetic time
When the blank absorbance is read before the reaction begins,
Single-reagent 0 -13≤N<P≤-1 1≤L≤<M≤59 K1
Double-reagent 0<F≤34 -F≤N<P≤-1 1≤L≤<M≤ K2
34,F+M≤59
When the blank absorbance is not subtracted,
Single-reagent 0 N and P are 1≤L≤<M≤59 0
blank.
Double-reagent 0<F≤34 N and P are 1≤L≤<M≤ 0
blank. 34,F+M≤59
The blank time and reaction time are almost the same for both fixed-time and Kinetic analysis,
except that M-L≥2 is required for Kinetic analysis, that is, the reaction time should include at
least 3 measuring points.
Sample Volume, Standard, Aspirated, Diluent, Increased, and Decreased
Sample volume is the standard sample amount, which should be dispensed in a normal test. It
ranges from 2 μL to 35μL with an increment of 0.1 μL. The default is 2 μL. A maximum of one
decimal is allowed.
Aspirated volume refers to the amount of sample used for dilution at the specified ratio. It
ranges from 2 μL to 35 μL with an increment of 0.1 μL. The default is blank. A maximum of one
decimal is allowed.
Diluent volume refers to the amount of diluent used for sample dilution. It ranges from 100 μL
to 200 μL with an increment of 0.5 μL. The default is blank. A maximum of one decimal is
allowed.
NOTE
If aspirated volume for dilution and diluent volume are defined, ensure the total sum of them is within 125μL -
235 μL; otherwise, the settings cannot be saved.
The diluent volume for standard, increased and decreased analysis can be defined in the same way.
Decreased sample volume indicates the sample amount required for a decrement test. It ranges
from 2 μL to 35μL with an increment of 0.1 μL. The default is blank. A maximum of one decimal
is allowed.
Increased sample volume indicates the sample amount required for an increment test. It ranges
from 2 μL to 35μL with an increment of 0.1 μL. The default is blank. A maximum of one decimal
is allowed.
Sample Blank
Sample blank is similar to sample analysis except for use of equivalent amount of physiological
saline Sample blank is used for removal of non-chromogenesis reaction, such as influence of
sample interference (Hemolysis, icterus and lipemia) on absorbance readings. Sample blank is
only effective for single-reagent endpoint chemistries.
Mark the Sample Blank checkbox with a tick. The chemistry will be sample blanked before the
reaction begins, and the Sample Blank checkbox on the Options and Rerun windows will be
selected automatically and cannot be modified.
Auto Rerun
The Auto Rerun option is used to rerun the chemistries when the auto rerun conditions are
satisfied.
Mark the Auto Rerun checkbox means enabling the auto rerun option.
For more information about auto rerun, see 6.2.6Rerunning samples on page 6-7.
Reagent Volume
Reagent volume specifies the reagent amount, which should be dispensed for measurement.
R1: 100 μL to 200 μL with an increment of 0.5 μL.
R2: 10 μL to 200 μL with an increment of 0.5 μL.
NOTE
The combined volume of all reagents and sample must be within 100 μL and 360 μL.
Linearity Range
The linearity range indicates the measurable range of the system, during which the test result is
linear to the response R. Determine the linearity range according to the reagent package insert.
The linearity range for standard, increased and decreased sample volume test should be set
separately. The input should be no more than 12 digits, and the default is blank.
The system compares the calculated sample concentration with the linearity range When the
high limit is exceeded, the > sign will appear near the result; when the low limit is exceeded, the
< sign will appear.
The default is blank, which means not performing this check.
Linearity Limit
Linearity limit is only applicable to Kinetic analysis, in which the absorbance change is linear to
the reaction time. If the reagent undergoes substrate depletion or the photometer fluctuates, or
the reaction mixture is not stirred evenly, the test results may be unreliable. Therefore, the
linearity of the measuring period is calculated and then compared with the set linearity limit.
If the reaction data within the linearity range does not satisfy the linearity limit, the system will
flag the test result with "LIN" on the patient report.
The linearity limit can be any number between 0 and 1 with a maximum of 2 decimals. The
default is blank, which means not performing this check.
Substrate Depletion
The Substrate Depletion option is only applicable to Kinetic and fixed-time analysis. It can be
obtained through the following formula:
Substrate depletion limit = Input substrate depletion limit + K(L1-Lb)
Where,
L1: refers to the absorbance of primary wavelength measured at the first measuring point
when sample is dispensed and stirred in sample analysis.
Lb: refers to the absorbance of primary wavelength measured at the first measuring point
when sample is dispensed and stirred in a reagent blank test or calibration with
0-concentration calibrator.
K: correction factor of liquid volume
Results will not be adjusted when L1-Lb≤0 or the measurement is not a reagent blank or
0-concentration calibration. Substrate depletion is not applicable for calibrations.
We deem that substrate depletion occurs if the primary wavelength absorbance of the first
measuring point is greater than the substrate depletion limit in ascending reactions or lower
than the substrate depletion limit in descending reactions. When substrate depletion occurs, the
system will flag the test result with "BOE" in the patient report.
The substrate depletion limit can be any number within -35,000~35,000. The default is blank,
which means not performing this check.
R1 Blank Absorbance Range
The R1 Blank Abs indicates the allowable range of the maximum absorbance in the previous
period prior to sample dispensing. The input range must be within -35,000~35,000, and the
low limit lower than the high limit.
If the maximum absorbance in the previous period prior to sample dispensing is beyond the set
range, the system will flag the test result with "RBK".
The default is -35,000~35,000; the field can be left blank.
Mixed Blank Absorbance Range
The Mixed Blank Abs indicates the allowable range of the absorbance measured at the end point
of a zero-concentration calibrator reaction or a reagent blank reaction. The input range must be
within -35,000~35,000, and the low limit lower than the high limit.
If the absorbance measured at the reaction end point is beyond the set range, the system will
flag the test result with "MBK".
The default is -35,000~35,000; the field can be left blank.
Blank Response
The Blank Response specifies the allowable range of the response in a zero-concentration
calibrator analysis or a reagent blank test. The input range can be any number within
-35,000~35,000, and the low limit lower than the high limit.
If the response is beyond the set range, the system will flag the test result with "BLK".
The default is -35,000~35,000; the field can be left blank.
On-board Stability
It refers to the number of days that the reagent can be kept valid since uncapped at the first
time.
The input range must be within 1-999 days. The default is blank.
Twin Chemistry
Twin Chemistry is associated with the current chemistry, and the two chemistries are run with
the same reagent. Results of two twin chemistries are calculated in the same test.
The chemistry whose result will be firstly calculated should be defined prior to the associated
chemistry. Volume of the shared reagent and sample volume must be the same for the two
chemistries. Only the two chemistries that have had no reagents loaded can be configured as
twins.
For more information about twin chemistries, see 7.5Twin chemistr on page 7-23.
A q 4 A q3
q 4 q3
PC1 PC 2
A q 2 A q1
q 2 q1
Formula 2:
3 Choose a chemistry.
4 Double click the Slope field and then input the slope.
5 Double click the Offset field and then input the offset.
6 Select Save to save your input information.
7 To restore the factory settings of slope and offset, select Restore Defaults.
8 Select Close the exit the window.
The system provides auto rerun of ISE test. When ISE test result is beyond the set critical range,
the ISE test will be rerun automatically.
Prior to defining the reference/critical range, ensure that you have sufficient permissions and
the system status is not Running.
7 Select Save F7. The reference/critical range are displayed in the middle list.
8 Select Prev F4 or Next F5 to set up reference/critical range for more chemistries.
9 Select Exit F8 to close the window.
3 Choose the chemistry name, sample type, gender and age range.
4 Choose a reference/critical range you want to remove.
5 Select Delete F2, and then select OK.
6 To clear all ranges of the chemistry, select Del All F3, and then select OK.
7 Select Exit F8 to close the window.
Unit
The unit of K, Na and Cl is mmol/L which can be viewed but cannot be edited.
Decimal
The decimal of the result can be viewed but cannot be edited.
Measurement range
The measurement range can be viewed but cannot be edited.
For instance, type in "10" in the first edit box of the Range field under Na, and then enter
"+" in the Flag field of the same row. If the Na concentration (L1) contained in a sample is
less than or equal to 10, the "+" sign will be added to the result in the patient report. Type
in "20" in the second edit box below the Range icon and "+-" in the second edit box below
the Flag icon. If the Na concentration (L2) is greater than 10 and less than or equal to 20,
the result will be flagged with the "+-" sign. The cycle continues. If the result is greater than
L5, the six flag will appear on the patient report.
5 Repeat steps 5-6 to flag the qualitative result for K and Cl.
6 Click OK to save the setup.
7 Select Exit to close the window.
5 Select OK.
The chemistry list on the request screens are refreshed automatically.
Requesting calibration
Twin chemistries can be requested for calibration in the same way as normal chemistries.
When either of the twin chemistries is requested, the other twin will be requested automatically,
and finally both chemistries will be calibrated. You are allowed to recall the calibration results,
calibration curves and reaction curves of the two chemistries.
7.7 Panels
A couple of chemistries combined together for certain clinical purposes can constitute a panel,
such as liver function, kidney function, etc. Panels can help fast programming of samples.
Panels can be composed of biochemistries and ISE chemistries. The system allows a maximum
of 100 panels to be defined. Only users with sufficient permissions are allowed to define, modify
and delete panels.
ID number
Print name
Attribute and reference value
Result unit and decimal place
5 To set up result flags for quantitative result, perform the following steps:
a. Select the Use Qualitative Result check box.
b. Enter the qualitative range and flag.
For instance, type in "10" in the first edit box of the Range field, and then enter "+" in the
Flag field of the same row. If the chemistry result (L1) contained in a sample is less than or
equal to 10, the "+" sign will be added to the result in the patient report. Type in "20" in the
second edit box below the Range icon and "+-" in the second edit box below the Flag icon.
If the chemistry result (L2) is greater than 10 and less than or equal to 20, the result will
be flagged with the "+-" sign. The cycle continues. If the result is greater than L5, the six flag
will appear on the patient report.
6 Select Save F7.
7 Select Exit F8.
7.11 Reflex
The Reflex option allows related chemistries to be requested and run automatically when the
deciding biochemistry's result is within specified range. Each biochemistry may embrace
multiple reflex conditions, and each condition may contain a maximum of 20 related chemistries.
Reflex conditions and related chemistries are open for observation, but only users with
corresponding permission are allowed to set, modify or delete reflex relation.
This chapter provides descriptions of system commands, system setup, instrument setup, print
setup, bar code setup, LIS setup, and user setup.
8.1.1 Home
The Home command is used to initialize the biochemistry system and the ISE module and to
recover them from failures, making all components return to the home positions. When the
Home command is executed, the system status becomes Standby
To home the system
1 Select Utility > Commands.
2 Select Home.
When the Start Analysis When Temperature is Steady checkbox is selected, the system
will check before analysis begins if the reaction temperature is normal. If the temperature is
normal, you are allowed to select to start analysis; otherwise, a message will appear
indicating analysis is forbidden in current condition.
When the Start Analysis When Temperature is Steady checkbox is not selected, the system
will still check before analysis begins if the reaction temperature is normal and within
37±2.0°C. If the temperature is normal, you are allowed to select to start analysis;
otherwise, the system will remind you that the results may be influenced if you continue to
start analysis. You may continue or abort the analysis.
Alarm for each reagent bottle
Each chemistry can have more than one bottle of reagent loaded. You can set up alarms for the
case that each reagent is used up.
Select the Alarm for each reagent bottle option. When the reagent of a bottle is exhausted, the
system will give an alarm. If the option is not selected, the system will not give an alarm.
Manage reagents by lot
This option is used to monitor the calibration status and time of each reagent lot, supports
reagent lot calibration, and displays calibration results of each reagent lot.
When this option is enabled, special attentions should be paid for the following operations:
Loading reagents: You are required to input the lot number when loading reagents manually.
The lot number of bar-coded reagents cannot be left blank; otherwise, reagent load will fail.
Viewing calibration status and requesting calibration: You can view calibration status and
time of each reagent lot, and request calibration accordingly.
Recalling calibration results: You can recall calibration results of each reagent lot on the
Biochemistry Calibration screen.
Auto calibration: Auto calibration by reagent bottle or lot is forbidden. When a different
reagent lot is used, the system will request and run calibration automatically. Reagent lots
with valid calibration factors will not be calibrated again when used for measurement.
Special wash before test
Select this option to execute a special wash with wash solution for the probe and mixer before a
batch of tests. If it is unselected, no special wash will be performed before a batch of tests.
Pause immediately
Select this option to pause the analyzer immediately after one test period once the Pause button
on the right corner of the main screen is clicked.
Result display settings
This option is used to set up flags and color for results less than or greater than the reference
range, as well as color for results less than or greater than the critical range.
Click the relevant color setup button, choose desired color, and then select OK. The system will
display flags in the Flag column of the Current Results and History Results screens and on
patient reports if the test result is less than or greater than the reference range. The flags can be
composed of numbers, letters and symbols for no more than 10 digits. The default flags for
reference range are "^" and "v". If a result is greater than the high limit, "^" will appear near the
result; if a result is less than the low limit, "v" will appear near the result.
If test results are beyond the critical range, they will appear in the set color.
8.3.1 Sleep/Awake
Sleep/Awake feature includes the Auto Sleep Setup, Auto Startup Setup and Auto Awake Setup
options.
The Auto Sleep Setup option is used to set up the time interval of auto sleep time of the system.
After the sleep time interval is set up, a countdown will begin from the moment that the system
status becomes Standby. When the time interval is elapsed, the system will begin sleeping.
Except for the auto sleep setting, the system can be woken up by means of the wake up
command.
The Auto Startup Setup and Auto Awake Setup options allow you to define date and time of
starting up or waking up the system. When the time is reached, the system will be started up or
woken automatically no matter if it is off or sleeping.
NOTE
If auto sleep is not enabled, some components, such as lamp, may get aged quickly and degraded in
performance. You are recommended to enable this option.
6 Select Save.
When the interval is elapsed, the system will starts to sleep and the system status becomes
Sleep.
7 Select Exit.
NOTE
After setting up the auto startup time, ensure that the operation unit and the analyzer are connected to power
supply; otherwise, they cannot be started up automatically.
1 Select Utility - System Setup.
5 Choose the weekday for auto startup, and then set up the specific time.
Any time within a week (from Monday to Sunday) can be defined for the system to start up
automatically.
6 Select Save.
When the date and time is reached, the system will be started up automatically if it is off.
7 Select Close.
NOTE
After setting up the auto awake time, ensure that the operation unit and the analyzing unit are connected to
power supply; otherwise, they cannot be woken up automatically.
1 Select Utility - System Setup.
2 Select Instrument F1.
3 Select 1 Sleep/Awake.
4 Select 3 Auto Awake Setup.
5 Choose the weekday for auto startup, and then set up the specific time.
Any time within a week (from Monday to Sunday) can be defined for the system to wake up
automatically.
6 Select Save.
When the date and time is reached, the system will be woken up automatically if it is
sleeping.
7 Select Exit.
3 View the version number of the operating software, model and number of open channels.
4 To view the version numbers of the smart module software, select Details.
Figure 8.8 Smart module software version window
When adjusted, the date and time will influence the time left of reagents and calibration, shelf
life of samples, and run length of two-control evaluation. The date and time cannot be edited
when the system status is Running. Modification of the date and time will not affect samples on
the Current Results screen or QC evaluation and Twin-Plot chart
To set up the system date and time
1 Select Utility > System Setup, and select Instrument F1.
2 Select Date/Time.
Figure 8.9 Date/Time window
3 Choose a voice tone from the drop-down list, and then select the corresponding Test
button to test the voice effect until the proper one is found.
4 Select Save to save the settings.
After customizing sample positions, the positions on the sample carousel and reagent carousel
are indicated as follows:
The set positions of the inner ring on the Program > Status screen are indicated in white,
which means available for sample.
The positions indicated with two crosses"××"on the sample carousel graph can be only used
for reagents.
The positions indicated with a red cross " " on the Reagent > Reagent Carousel Status
screen means unavailable for reagents.
Code128
Code39
UPC/EAN
Code93
Code 128, Code 93 and UPC/EAN requires a check digit by default, and other symbologies
are not compulsive. The Code 128 is selected by default and cannot be modified.
CAUTION
You are recommended to enable the check function for all symbologies in order to prevent misreading of
bar code.
4 Set up sample bar code applications.
Enable/Disable sample carousel bar code
When this option is selected, the system scans the entire sample carousel to locate samples
at the beginning of test.
Enable or disable auto numbering of bar-coded samples
When this option is enabled, the system will automatically number the bar-coded samples
during bar code scanning. The start number will be the next available one since the last
sample is programmed. The default start number for every day is 1.
Extract sample information
° When this option is selected, the system will automatically extract the sample
information according to the barcode.
° Only when LIS communication mode is unidirectional, can the option be enabled.
Define STAT sample positions on sample carousel
Input the start and end positions within the range of 1~50. If the sample positions on the
middle ring are customized, these positions can be set as STAT positions. The set positions
will be indicated by E (Emergent) on the sample carousel status screen. Samples placed in
the specified STAT positions will be taken automatically as emergent samples.
The Sample Crsl Bar Code and Auto Number Scanned Samples options are selected by
default.
5 Select OK to save the setup.
6 Select Format
7 Define the bar code digits.
The system can scan a sample bar code of fixed length or within 3-27 digits. The Interleaved
2 of 5 only supports bar code of even number length and the digits of the barcode must be
defined.
To use a fixed-length bar code,
° Mark the Fixed Digits checkbox of relevant symbology.
° Type in the number of digits in the edit box to the right of the Fixed Digits field.
To use a sample bar code within 3-27 digits, you have no need to define the fixed digits.
8 Select OK to save the settings.
To perform reagent bar code setup
1 Select Utility > System Setup, and select Bar Code F4.
2 Choose Reagent Bar Code.
3 Select or deselect Analyze barcode of open reagent.
If Reagent barcode system is configured, the option Analyze barcode of open reagent
is not selected by default
When Analyze barcode of open reagent is selected, once the barcode of the open
reagent is identified, its information is analyzed according to its setup.
While loading the reagent manually, you can enter the barcode of the open reagent on
the condition that Analyze barcode of open reagent is not selected.
4 Choose a bar code symbology and set up the check digit status.
The following symbologies are provided:
Codabar
Interleaved 2 of 5
Code128
Code39
UPC/EAN
Code93
Code 128, Code 93 and UPC/EAN requires a check digit by default, and other symbologies
are not compulsive. The Code 128 is selected by default and cannot be modified.
CAUTION
You are recommended to enable the check function for all symbologies in order to prevent misreading of
bar code.
5 Define the total length of reagent bar code.
Type in the total length of the reagent bar code in the T field. The input range is 13-30
digits. The Interleaved 2 of 5 only supports bar code of even number length.
Type in the start digit of the reagent bar code in the S field.
Type in the end digit of the reagent bar code in the E field.
6 Determine reagent bar code compositions.
Type in the number of digits for reagent information in the Digits field.
Type in the start digit of the reagent information in the S field.
Type in the end digit of the reagent information in the E field.
Table 8.1 Reagent bar code compositions
Reagent Information Number of Digits
Chemistry number 0-4 digits
Chemistry name 0-10 digits
Reagent type 1 digit ("1" stands for R1 and "2" stands for R2)
Serial number 0-5 digits
Bottle type 1-3 digits
Lot number 0-18 digits
Expiration date 0, 4, 6 or 8 digits (4 digits: yymm; 6 digits:
yyyymm; 8 digits: yyyymmdd)
7 Select OK.
Sample program
LIS Chemistry Analyzer
Result transfer
The host communication parameters, such as transmission mode, IP address and port, should be
set up prior to use of the LIS host. To download sample program information from or sent
results to the host, you need to set up the chemistry code used for identification of chemistries
on both the LIS host and the analyzer, which, otherwise, cannot identify the chemistries
simultaneously.
Parameter Description
Port Enter the interface number of the LIS host.
Serial If you choose Serial as the transport mode, set up the following parameters:
communication Serial port: The default is COM1.
parameters Data bits: 7 or 8. The default is 8.
Stop bits: 1 or 2. The default is 1.
Parity: None, Odd, or Even. The default is None.
Baud rate: 300, 1200, 2400, 4800, 9600, or 19200. The default is 9600.
Protocol Choose a protocol for connection between the analyzer and the LIS host from the
Protocol drop-down list. The options include HL7 and ASTM 1394.
Mode Choose a data transmission mode for the analyzer and LIS host. The available
options are Unidirectional and Bidirectional.
Unidirectional: You are only allowed to send results and patient demographics to
the host rather than downloading sample programs from it.
Bidirectional: You are allowed to send results and patient demographics to the
host and downloading sample programs from it.
Timeout Enter the time out limit for querying the LIS host. The input range is 30s-60s, and
the default is 30s.
If the time out limit is exceeded when you attempt to download sample programs
from, or send results to, or connect the analyzer with the LIS host, the system will
give an alarm indicating communication timed out.
Auto Connect to LIS When the checkbox is selected, the system will connect to the LIS host
automatically when started up.
Retry after When the checkbox is selected, the system will try to reconnect the LIS host for
Disconnection every set interval once the connection is interrupted.
Interval Input the time interval for which the system will try to reconnect the LIS host for
every set interval once the connection is interrupted. The default is 30 seconds.
Send Complete When the checkbox is selected, the system will automatically send results to the LIS
Samples host after a sample changes from In Progress to Complete. This function is only
applicable to samples analyzed on the current day rather than those analyzed
before.
Send Incomplete When the checkbox is selected, the system will automatically send results to the LIS
Samples host after a sample changes from In Progress to Incomplete. This function is only
applicable to samples analyzed on the current day rather than those analyzed
before.
Advanced options Select Advanced. The Advanced window appears, providing the following options:
Send Programmed Samples: When the checkbox is selected, the system will
automatically send the program information to the LIS host once a single or
batch routine and STAT samples are programmed.
Rerun Finished Chemistries When Downloaded: When the checkbox is
selected, chemistries that have been finished will be rerun if downloaded again.
If this option is not selected, they will be neglected.
Send Actual Results and Rerun Results: When the checkbox is selected, all
actual results and rerun results of each chemistry will be sent to the LIS. If this
option is not selected, only the default result will be sent.
Bypass Results Beyond Linearity Range: When the checkbox is selected, those
results that are beyond the linearity range will not be sent to the LIS. If this
option is not selected, they will be sent.
Ignore Alarms for Unknown Chemistries: When the checkbox is selected, the
system will not give an alarm if the samples downloaded from the LIS host
Parameter Description
contain unknown chemistries without identification code. If this option is not
selected, an alarm will be given indicating sample programming failure.
4 Select Save to save your input information.
5 Select Connect to connect the analyzer with the LIS host.
NOTE
The default username and password for administrator is Admin. Please note that the password is case sensitive.
You are recommended to change the password when logging on the system for the first time in order to prevent
others from abusing the privileges of the administrator.
If an operator forgets his password, he may ask the administrator to log on the system and delete the username
and then redefine a username; or he may contact our customer service department or your local distributor. If
the administrator forgets his password, contact our customer service department or your local distributor.
The Template Modifying Software is affiliated with the Operating Software and used to create or
edit print templates, which illustrate the contents and format of patient reports.
The Template Modifying Software can be started separately or together with the Operating
Software. To start the Template Modifying Software, select the Edit button on the Print page of
the operating software.
The following sections introduce the Template Modifying Software by menus and toolbars.
Option Description
Option Description
If the template has more than one page, and are available.
: Go to the first page.
: Go to the previous page.
: Go to the specified page.
: Go to the next page.
: Go to the last page.
: Select to expand the template view among 25%, 50%, 75% and 100%.
The default is 100%.
: Select to shrink the template view.
: Print out the template. It is equivalent to the Print option in the File
menu.
: Select to exit the preview window and return to the template.
Exit Select this option to close the Template Modifying Software.
You can use the shortcut key Alt+F4 instead.
If the template is changed, the following dialog box pops up.
NOTE
The control(s) you have cut or copied can only be pasted on the current Template Modifying Software rather
than another one or other software.
NOTE
Selecting a control while holding the Ctrl key copies the control.
Line Select this option to insert a line in the editing area. The mouse pointer changes
into a +. Click once in the editing area and drag the mouse to draw a line.
Rectangle Select this option to insert a rectangle in the editing area. The mouse pointer
changes into a +. Click once in the editing area and drag the mouse to draw a
rectangle.
Option Description
Label Select this option to insert a label in the editing area. The mouse pointer
changes into a +. Click once in the editing area and drag the mouse to draw a
label.
Label is a type of text control and the contents on a label will not change when
printed.
Text Select this option to insert a text control in the editing area. The mouse pointer
changes into a +. Click once in the editing area and drag the mouse to create a
text.
The contents in a text control will be replaced by the actual test data when
printed.
Title Select this option to insert a title in the editing area. The mouse pointer changes
into a +. Click once in the editing area and drag the mouse to create a title.
Title is a type of text control. The “%s” will be replaced by a hospital name
when printed. Please note “%s” is added by user and not produced
automatically.
Image Select this option to insert an image in the editing area. The mouse pointer
changes into a +. Click once in the editing area and drag the mouse to create an
image.
The image on the template is for illustration only and will be replaced by
corresponding curve graph when printed.
Option Description
Center H Align one or multiple controls to the horizontal center of current template.
Center V Align one or multiple controls to the vertical center of current template.
Even Space Arrange three or more controls with same space horizontally.
H
Even Space Arrange three or more controls with same space vertically.
V
Same Width Adjust the specified controls to the same width as the lastly-selected control.
Same Height Adjust the specified controls to the same height as the lastly-selected control.
Same Size Adjust the specified controls to the same width and height as the
lastly-selected control.
9.1.7 Set(S)
The Set menu only includes an option, Print ID. See the figure below.
Figure 9.6 Set menu
Select Print ID. The Set Print ID dialog box is displayed. You can enable or disable the print
fields and view the corresponding ID of each field.
Figure 9.7 Set Print ID window
Option Description
Chinese Switch the screen language to Chinese. Not available.
English Switch the screen language to English. Not available.
9.1.10 Page
When no control is selected, the property window shows the properties of the current template,
such as paper, print type, etc.
Figure 9.10 Page property area
The following table shows the correspondence between the shortcut buttons and menu options.
Table 9.9 Common tools
Button Menu Option Button Menu Option
New File/New Cut Edit/Cut
Save File/Save (not enabled) Copy Edit/Copy
Save As File/Save As Paste Edit/Paste
Import File/Import Delete Edit/Delete
Export File/Export Zoom View/25%-200%
Delete File/Delete Property View/Property Window
Print File/Print Rpt List View/Report Window
Preview File/Preview Print ID Set/Print ID
The following table shows the correspondence between the shortcut buttons and menu options.
Insert/Line Format/Bottom
Insert/Rectangle Format/Center H
Insert/Label Format/Center V
9.4.1 Line
When a line control is selected, the property window shows the properties of the line.
Figure 9.13 Line property area
NOTE
The control coordinate originates from the upper-left corner of the editing area, from
which the X axis (positive) is extended horizontally to the right and the Y axis (positive)
vertically to the bottom. The unit is mm.
Start Y Set the Y-coordinate value of the start point.
End X Set the X-coordinate value of the end point.
End Y Set the Y-coordinate value of the end point.
Line Width Set the width of the line. The unit is mm.
Group No. A group gathers multiple controls that will be used frequently on the
template. e.g. a line of controls constitutes a group. The group No. is 0 if not
defined.
Line Color Set the color of the line.
Print Enable or disable printing the line on actual reports.
9.4.2 Rectangle
When a rectangle control is selected, the property window shows the properties of the rectangle.
Figure 9.14 Rectangle property area
9.4.3 Label
When a label control is selected, the property window shows the properties of the label.
Figure 9.15 Label property area
9.4.4 Text
When a text control is selected, the property window shows the properties of the text.
Figure 9.16 Text property area
9.4.5 Title
When a title control is selected, the property window shows the properties of the title.
9.4.6 Image
When an image control is selected, the property window shows the properties of the image.
Figure 9.18 Image property area
(1)
(2)
This chapter provides you with maintenance of the instrument, including frequently-used
maintenance commands and scheduled maintenance procedures. The purpose, time, system
status, precautions and steps of each maintenance procedure are described here.
10.1 Overview
Maintenance of the system should be performed regularly by trained personnel to ensure
reliable performance and reduce unnecessary service calls. Even you are only an operator, it is
important for you to read this chapter. Your thorough understanding will help you obtain the
best performance of the system.
In the case of maintenance that is beyond your capability or not covered in this chapter, contact
our customer service department or your local distributor.
Maintenance frequency
The maintenance frequencies stated in this manual are based on 720 tests/day, and
720*25=18,000 tests/month.
WARNING
Maintain the system strictly as instructed by this manual. Inappropriate maintenance may lead to unreliable
results, equipment damage or personal injury.
Do not touch the components other than those specified in this chapter.
Performing unauthorized maintenance procedures can damage the instrument and cause personal injury, or
invalidate the applicable warranty provisions in the service contract.
Do not spill water or reagent on mechanical or electrical components of the system.
Shut down and turn off the system and disconnect the power plugs before cleaning. Take necessary
measures to prevent water ingression, otherwise equipment damage or personal injury may be caused.
Replacement of the photometer lamp should be done when the system power has been switched off for at
least 5 minutes.
If the system fails and needs servicing, contact our Customer Service Department or your local distributor.
The system may need to be stopped or transported during servicing, which will probably cause biohazards,
electric shock hazards and moving part hazards. Exercise caution when prepare the system for servicing.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
CAUTION
To wipe off dust from the system surface, use a soft, clean and moistened cloth soaked with soap water
rather than organic solvents such as ethanol. After cleaning, wipe the surface dry with dry cloth.
Replacement of major parts as photometer lamp, probe and mixer must be followed by re-calibration.
After performing maintenance, make verification to ensure that the system runs normally.
If the system is to be stored for a long time (over 1 week) or transported, contact our customer service
department or your local distributor to perform necessary maintenance in order to ensure the system’s
optimal performance in following use.
Accessories
Cross screwdriver (φ4.7×100): for install and removing the probe
Round-head needle, 0.25+/-0.01mm×125mm: for unclogging the probe
Spare parts
Table 10.1 Spare parts
Part name Related maintenance Comment
Lamp Replacing lamp Regularly-replaced part.
Replace it when it serves for over
2000 hours or in every 6 months.
Syringe Replacing syringe Replace it when it serves for over
100,000 times.
Reaction cuvette Replacing cuvette Consumable. Replace it when needed.
Probe Replacing probe Regularly-replaced part.
Replace it when needed.
ISE electrode Replacing ISE electrode Regularly-replaced part.
Replace it when needed.
Electrode case Storing ISE electrode Use it when store the ISE electrodes.
ISE cleaning solution Cleaning electrode tubes Consumable. Use it when cleaning the
electrode tubes every day.
Tools to be prepared by user
Before maintenance, prepare the following tools:
Table 10.2 Tools to be prepared by user
Item Applicable maintenance
Clean gauze Checking syringe; cleaning rotors, probe/mixer
exterior and cuvette wash station
Cotton swabs Cleaning wash well and sample/reagent
compartment
Suction cleaner Cleaning fans and dust screens
Hair brush Cleaning dust screen
Tweezers Removing/Installing probe and syringe washer
Thread syringe Unclogging the probe
Tube brush or ultrasound cleaner Cleaning filter core
Beaker Unclogging and cleaning the probe
Ethanol Cleaning the photometer lens, probe, mixer and
wash station
NaClO (0.5% sodium hypochlorite Cleaning wash wells
solution)
Fiber-free gloves Cleaning and replacing reaction cuvettes
Large water container Cleaning deionized water tank
Screen and keyboard wash solution Cleaning screen and keyboard
Sample tube Cleaning ISE electrodes
10.1.3 Concepts
The Maint window includes the following tab pages:
CAUTION
"Cleaning ISE waste tube" is not provided on the software. To avoid clogged waste tube after long-term use,
when you see alarm “No fluid in tube” or “Air in calibrator A” or “Air in calibrator B”, and check that the waste
tube is clogged, perform this maintenance item.
Warning
The probe and mixer are sharp and vulnerable. To prevent injury and equipment damage, exercise caution
when working around the probe and mixer. Keep away from the probe and mixer to avoid collision with them.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
How to do
1 Open the protective shield of the analyzer.
2 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
3 Select Clean Probes/Mixers/Wash Wells.
4 Check the exterior of the probe and mixer for stains. If stains exist, perform the "Cleaning
probe exterior" and "Cleaning the mixer" procedures.
5 Select Continue to clean the probe interior.
6 Check the liquid flow of the probe.
If the liquid flow is sprayed out or does not come out vertically, the probe may be
clogged. Perform the "Special wash probe" procedure, and then check it again.
If the abnormity remains, perform the "Cleaning probe interior" procedure.
If the abnormity still remains, perform the "Replacing the probe" procedure, or contact
a service engineer.
7 Select Second Wash. The probe interior wash can be performed again.
8 Select Continue.
9 Observe the water flow of the probe/mixer wash wells, and check if the water reaches to
about 5 mm of the probe/mixer from the tip.
10 If the water flow is abnormal, contact a service engineer.
11 Select Continue and Done.
12 Restore the protective shield.
When to do
You are recommended to do this maintenance procedure every day before starting the analysis.
System status
Make sure that the system is powered off, or the system status is Incubation or Standby.
How to do
1 Check that the diluted wash solution tank has sufficient wash solution.
2 If the wash solution is insufficient, fill more.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Dispose of the waste in accordance with your local or national guidelines for biohazard waste disposal.
How to do
1 Check the high-/low-concentration waste tubes, and ensure that they are not leaking or
bent.
2 Check if the high-/low-concentration waste tanks are full. If yes, empty them.
When to do
You are recommended to do this maintenance procedure every day before starting the analysis.
System status
Make sure that the system is powered off, or the system status is Incubation or Standby.
Precautions
CAUTION
You are recommended to replace the probe wash solution every day in order to prevent probe clogging and
cross contamination.
While the system is running tests, do not try to fill probe wash solution until the system status becomes
Standby.
How to do
1 Check the volume of the probe wash solution on the sample/reagent carousel position 49#.
2 If necessary, fill more or replace the wash solution. To ensure wash effects, you are
recommended to replace it.
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
The cleaning solution may hurt your eyes and skins. Exercise caution while using the cleaning solution. If your
eyes contact the cleaning solution, rinse them off with fresh water and consult a doctor.
CAUTION
Please use consumables recommended by our company. Use of other consumables may degrade the system
performance.
NOTE
After performing this procedure, recalibrate the ISE electrodes prior to starting analysis.
How to do
1 Select Utility > Maintenance > Maintenance > ISE Maintenance.
2 Choose Clean Electrode Tubes. The maintenance guide window shows.
3 Open the upper protective shield of the analyzer.
4 Fill a reagent bottle with at least 2.5 mL ISE cleaning solution, and then load it to ISE
cleaning solution position (Position 48#) on the sample/reagent carousel.
5 Select Continue. The system starts cleaning the ISE electrode tubes.
6 Select Done.
After finishing the maintenance, run ISE calibration.
For more information on ISE calibration, see Requesting calibration on page 2-17.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
How to do
Warning
The probe tip is sharp and may cause puncture wounds. To prevent injury, exercise caution when working
around the probes. If the probe is bent or damaged, replace it immediately; otherwise, unreliable results may be
obtained.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Switch off the analyzing unit power.
2 Pull the probe arm to the highest point and then rotate it to move the probe to a position
convenient for cleaning.
3 Use gauze soaked with ethanol to gently wipe the probe exterior. Clean the probe tip until
it becomes clear without stain.
4 Use gauze moistened with deionized water to clear the ethanol on the probe.
CAUTION
Do not pull the probe horizontally to prevent probe damage.
5 After finishing the cleaning, close the shielding cover, and turn on the analyzing unit power
switch.
6 Select Utility > Commands > Home to reset the system.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
CAUTION
Exercise caution while working around the mixer. If it is bent or damaged, replace it immediately; otherwise,
unreliable results may be obtained.
How to do
1 Switch off the analyzing unit power.
2 Pull the mixer arm to the highest point and then rotate it to move the mixer to a position
convenient for cleaning.
3 Use gauze soaked with ethanol to gently wipe the mixer exterior until it becomes clear
without stain.
4 Use gauze moistened with deionized water to clear the ethanol on the mixer.
CAUTION
Do not pull the mixer horizontally to prevent damage.
5 After finishing the cleaning, close the shielding cover, and turn on the analyzing unit power
switch.
6 Select Utility > Commands > Home to reset the system.
Precautions
NOTE
When a cuvette is deemed dirty, clean or replace it immediately, and then perform the cuvette check again.
How to do
1 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
2 Choose Cuvette Check.
3 Make sure that the lamp has been turned on for over 5 minutes. Select Continue and then
select Start.
When finishing the check, the system refreshes the cuvette status based on the check
results and highlights the dirty cuvettes in yellow.
4 Record the cuvettes highlighted in yellow and replace them.
5 Select Result. The Cuvette Check Results window appears and shows the latest check
results at all wavelengths.
6 To view the test result and current status of certain cuvette, click the cuvette No. in the
result list. The Cuvette Status window pops up.
7 Choose the following buttons as needed:
|<: to view the first cuvette.
<: to view the previous cuvette.
>: to view the next cuvette.
>|: to view the last cuvette.
Print: to print the results currently displayed on the screen.
Exit: to close the Cuvette Status window.
8 Select Exit to close the Cuvette Check window.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
The wash solution may hurt your eyes and skins. Exercise caution while using the wash solution. If your eyes
contact the wash solution, rinse them off with fresh water and consult a doctor.
CAUTION
Please use consumables recommended by our company. Use of other consumables may degrade the system
performance.
How to do
6 Select Done.
When to do
You are recommended to perform this procedure every 3 months.
Materials required
Water container
System status
Make sure that the system status is Standby
How to do
1 Remove the quick connector from the outlet of the diluted wash solution tank.
2 Clean the diluted wash solution tank repeatedly with deionized water.
3 Connect the outlet tube of the diluted wash solution tank.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Switch off the analyzing unit power.
2 Rotate the probe and mixer to keep them away from the wash wells.
3 Use clean cotton swabs moistened with NaClO to clean the wash wells.
4 After finishing the cleaning, turn on the analyzing unit power switch.
5 Select Utility > Commands > Home or execute the Clean Probes/Mixers/Wash Wells
command, and check if the wash wells have a normal water flow.
Purpose
To clean the cuvette wash station in order to avoid waste buildup and cross contamination.
When to do
This procedure should be performed on monthly basis.
Materials required
Gauze, ethanol, deionized water, waste container (large beaker)
System status
Make sure that the system status is Standby
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Dispose of the used gauze in accordance with your local or national guidelines for biohazard waste
disposal.
How to do
1 Remove the cuvette wash station and use ethanol-moistened gauze to wipe the wash
probes and wipe block.
2 Use gauze moistened with deionized water to clear the ethanol on the wash probes.
3 Restore the wash station.
4 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
5 Choose Prime Wash Station The maintenance guide window shows. Select Continue.
6 Enter the wash cycle (1~100). Select Continue.
7 When the cleaning and priming are finished, select Done.
8 Select Utility > Commands, and then select Home to put the instrument into the Standby
status.
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
System status
Make sure that the status of the ISE module is Standby or Stopped.
How to do
1 Select Utility > Maintenance > Maintenance > ISE Maintenance, and choose the Clean
Sample Injection Port option.
2 Select Continue.
3 Use clean cotton swab soaked with ethanol to wipe the sample injection port (interior of
the sample injection cup of the ISE module ) until it is clean.
4 Use a clean cotton swab soaked with DI water to wipe the interior and periphery of the
sample injection port.
5 Select Done.
6 Select Purge A and Purge B to prime the ISE module, each for 3 times.
7 Restore the cover of the ISE module.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Select Utility > Maintenance > Maintenance > ISE Maintenance.
2 Choose Pump Calibration.
3 Fill the sample tube with at least 1000 μL deionized water and then place it in No.40
position of the sample carousel.
4 Select Start.
When the calibration is complete, the calibration results are displayed on the screen.
5 Select Done.
6 To view the detailed results, select Pump Calibration Result. The detailed results are
displayed in the data list.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Select Utility > Maintenance > Maintenance > ISE Maintenance.
2 Choose Air Bubble Detector Calibration.
3 Select Start.
When the calibration is complete, the calibration results are displayed on the screen.
4 Select Done.
5 To view the detailed results, select Air Bubble Detector Calibration Result. The detailed
results are displayed in the data list.
Precautions
NOTE
Before checking the lamp, perform the Cuvette Check procedure and replace or clean the dirty cuvettes;
otherwise, the photometer check results are unreliable.
To ensure the photometer's measurement performance, replace the lamp in the case of weak light intensity.
How to do
1 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
2 Choose Photometer Check.
3 Select Continue and then select Start.
4 If an alarm occurs during the check, operate as follows:
If the alarm indicates the lamp is off, check if the lamp has been turn on. If not, execute
the Home command; if yes, contact our customer service department or your local
distributor.
If the alarm indicates light intensity too strong, contact our customer service
department or your local distributor.
If the alarm indicates light intensity weak, replace the lamp.
For more information on replacing lamp, refer to 10.8.1 Replacing lamp on page 10-27.
5 When the test is finished, check the test results.
On the left of the screen shows the absorbance at each wavelength in the current
photometer check; on the right of the screen shows that of the previous photometer
check.
6 Check the results of the previous and current photometer check to understand the lamp
status.
7 Choose the following buttons as needed:
Print: to print the photometer check results currently available on the screen.
Exit: to close the window.
8 Select Done to close the Photometer Check window.
System status
Make sure that the system status is Standby or Stopped.
Precautions
CAUTION
Too hot lamp may burn you. Do not replace the lamp until it gets cool.
Please use consumables recommended by our company. Use of other consumables may degrade the
system performance.
Do not touch the light entrance on the lamp housing or the lens in front of the lamp. In case the light
entrance is dirty, use cotton swabs moistened with absolute ethanol to clean it.
How to do
1 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
2 Choose Replace Lamp. The maintenance guide window pops up. Select Continue.
3 Make sure that the lamp has cooled down for 5 minutes, and then select Continue.
4 Open the small door at the right panel.
5 Loosen the nuts on the cable terminals, and then remove the O-ring connectors from the
terminals.
6 Loosen the retaining screw on the left side of the lamp.
7 Remove the lamp from the lamp housing.
CAUTION
Do not hold the lamp by its bulb to prevent contamination and damage.
8 Install the new lamp, and the retaining screw, O-ring connectors, cable terminal nuts and
lamp cover plate in the reversed order.
9 Select Done.
Perform the Photometer Check procedure to ensure the system power is normal.
For more information, refer to 10.7.6 Photometer check (page 10-26).
System status
Make sure that the system status is not Running.
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Dispose of the used gauze in accordance with your local or national guidelines for biohazard waste
disposal.
CAUTION
Do not spill liquid on the analyzer. Liquid ingression may cause equipment damage.
How to do
1 Use clean gauze moistened with ethanol to clean the analyzer panels and carousel covers.
2 Use wash solution to clean the screen and keyboard.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Dispose of the used gauze in accordance with your local or national guidelines for biohazard waste
disposal.
CAUTION
Do not spill water or ethanol into the sample compartment to prevent equipment damage.
How to do
1 Remove the sample/reagent carousel cover and the carousel, and then store them
properly.
2 Use clean gauze soaked with deionized water or ethanol to clean the interior of the
compartment. If necessary, use gauze moistened with neutral wash solution.
3 Use clean gauze soaked with deionized water or ethanol to clean the carousel, and then use
cotton swabs dipped with ethanol to clean the sample and reagent positions.
4 Install the sample/reagent carousel and the carousel cover.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
To remove the probe
1 Recall the maintenance logs and check if the probe has been removed and reinstalled for 3
times. If it has, replace the washer with a new one.
a. Prepare a new washer.
b. Moisten it with deionized water.
Store the washer properly to avoid being lost.
2 Switch off the analyzing unit power.
3 Loosen the screws on the arm cover and remove the cover from the arm base.
4 Press the circuit board with one hand and unplug the tube connector with the other hand.
5 Use a small screwdriver to remove the retaining screw from the probe and take out the
spring.
6 While holding the connector on the probe with one hand, unscrew the tube connector
counterclockwise with the other hand until the tube connector is disconnected. Remove
the tube from the probe.
Exercise caution to prevent the washer from dropping out. If the washer drops out, store it
in a clear place for later installation. To replace the washer, take it out from the tube
connector.
7 To replace the washer, remove it.
8 Remove the sample probe.
When to do
This procedure should be performed when there are dirty substances or dust on the rotors.
Materials required
Clean gauze
System status
Make sure that the system status is Standby.
Precautions
Warning
The probe and mixer tip are sharp and may cause puncture wounds. To prevent injury, exercise caution when
working around the probes and mixers.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Dispose of the used gauze in accordance with your local or national guidelines for biohazard waste
disposal.
How to do
1 Switch off the analyzing unit power.
2 Pull the probe/mixer arm to the highest point, and then rotate the arm to move the
probe/mixer to a position convenient to operate.
3 Wipe the rotor with clean gauze.
4 After finishing the cleaning, turn on the analyzing unit power switch.
5 Select Utility > Commands > Home to reset the probe and mixer.
Precautions
Warning
The probe tip is sharp and may cause puncture wounds. To prevent injury, exercise caution when working
around the probes.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
To remove the old probe
Remove the old probe.
For more information on removing probe, see 10.9.3 Cleaning probe interior on page 10-30.
Precautions
Warning
The mixer tip is sharp and may cause puncture wounds. To prevent injury, exercise caution when working
around the mixer.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
CAUTION
When trying to pull out the mixer, concentrate your force in the direction of the axis on the mixer arm Biased
force may damage the mixer and/or the axis.
Ensure the mixer is all the way pushed to the end.
How to do
Replacing the mixer includes the following steps:
Removing the mixer
Installing and checking the mixer
To remove the mixer
1 Switch off the analyzing unit power.
2 Gently pull the mixer to its highest point and rotate it to a position convenient to operate.
3 Pinch the mixer by the knurled part with one hand and unscrew (counterclockwise) the
retaining nut with the other hand until the mixer loosened.
Figure 10.2 Mixer
Knurled part
Mixer
Mixer
2 Pinch the mixer by the flat part and align the hole of the nut to the axis on the mixer and
push the nut onto the mixer until it reaches the end of the mixer.
3 Screw the lock nut clockwise to tighten the mixer.
4 Visually check whether the mixer is vertical to the bar arm.
5 If it is not, remove the mixer and reinstall it.
6 Pull the mixer arm to its highest point and rotate it back to a position above its wash well.
7 Turn on the analyzing unit power switch.
8 Select Utility > Commands, and execute the Home command.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Switch off the analyzing unit power, and open the upper shield of the analyzer.
2 Loosen counterclockwise the four retaining screws on top of the syringe, and then remove
the screws and the fixing blocks.
3 Loosen counterclockwise the retaining screw at the bottom of the syringe and then remove
it.
4 Hold the T piece with one hand and the syringe connector with the other hand. Loosen the
syringe counterclockwise and then remove the washer.
5 Soak the syringe connector in the deionized water beaker, pull the plunger head to aspirate
half syringe of deionized water, and then push the plunger head to remove the air. Repeat
this pull-push operation until the air bubbles are removed from the syringe. Fill the syringe
with half cylinder of deionized water to prevent new bubbles.
6 Put the washer in the T piece. Hold the T piece with one hand and the syringe connector
with the other hand, and then screw the T piece clockwise.
7 Install the syringe on the bracket.
8 Install the fixing blocks and 4 retaining screws while having the retaining screws not
tightened.
9 Align the plunger head to the retaining screw at the bottom of the syringe, and then tighten
clockwise the retaining screw.
10 Pinch the plunger guide cap to adjust the syringe height to make the syringe head over the
upper fixing block for 7.5 scales.
11 Tighten the four retaining screws on the fixing blocks.
12 Turn on the analyzing unit power switch.
13 Perform the Home maintenance procedure. Check the new syringe for leak and bubbles,
and if there is, perform the Check Sample/Reagent Syringe procedure.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
How to do
1 Prepare a new syringe assembly and washer, put the plunger head in the deionized water
beaker to remove air from the syringe, and then moisten the washer in the deionized
water.
2 Switch off the analyzing unit power.
3 Open the upper shield of the analyzer. You will see one syringe. They are, from left to right,
sample syringe, reagent syringe 1 and reagent syringe 2.
4 Loosen counterclockwise the four retaining screws on top of the syringe, and then remove
the screws and the fixing blocks.
5 Loosen counterclockwise the retaining screw at the bottom of the syringe and then remove
it.
6 Hold the T piece with one hand and the syringe connector with the other hand. Loosen the
syringe counterclockwise and then remove the washer.
7 Soak the new syringe connector in the deionized water beaker, pull the plunger head to
aspirate half syringe of deionized water, and then push the plunger head to remove the air.
8 If there is no washer inside the T piece, put the new washer in the T piece. Hold the T piece
with one hand and the syringe connector with the other hand, and then screw the T piece
clockwise.
9 Install the syringe on the bracket.
10 Install the fixing blocks and 4 retaining screws while having the retaining screws not
tightened.
11 Align the plunger head to the retaining screw at the bottom of the syringe, and then tighten
clockwise the retaining screw.
12 Pinch the plunger guide cap to adjust the syringe height to make the syringe head over the
upper fixing block for 7.5 scales.
13 Tighten the four retaining screws on the fixing blocks.
14 Turn on the analyzing unit power switch.
15 Perform the Home maintenance procedure. Check if the new syringe has leak. If it does,
perform the Check Sample/Reagent Syringe procedure to check the syringe.
Precautions
Warning
While installing the reaction cuvettes, exercise caution to avoid scratching them. Do not touch the optical
surface of the reaction cuvettes. If the optical surface is polluted, the obtained absorbance may be inaccurate.
While installing the reaction cuvettes, make sure that the optical surface is confronted with the outside of the
reaction carousel.
Wear gloves free of fibre and powder to avoid polluting the optical surface of the reaction cuvettes.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
CAUTION
Please use consumables recommended by our company. Use of other consumables may degrade the system
performance.
While installing the reaction cuvettes, exercise caution to avoid scratching them. Do not touch the optical
surface of the reaction cuvettes. If the optical surface is polluted, the obtained absorbance may be inaccurate.
While installing the reaction cuvettes, make sure that the optical surface is confronted with the outside of the
reaction carousel.
Wear gloves free of fibre and powder to avoid polluting the optical surface of the reaction cuvettes.
How to do
1 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
2 Choose Replace Cuvette.
3 Open the cuvette replacement cover.
4 Select Continue.
5 Type in the position number of the cuvette you want to replace.
6 Find the specified cuvette position, grab the cuvette segment with your thumb and
forefinger, and then take it out.
7 Install a new cuvette segment on the positioning pins, and then press it with the retaining
screw.
8 Close the cuvette replacement cover.
9 Select Done.
How to do
1 Place more than 20 mL concentrated wash solution in position D of the reagent carousel.
2 Select Utility > Maintenance > Maintenance > Biochemistry Maintenance.
3 Choose Special Wash Probe.
4 Set replicates and wash volume, and then select Continue.
6 When the cleaning is finished, select Done.
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
CAUTION
Exercise caution not to spray water or ethanol or other liquids on the glass of the bar code scanning window.
How to do
1 Remove the sample/reagent carousel cover and the carousel.
2 Use clean gauze to clean the bar code scanning window inside the compartment.
3 If necessary, use gauze soaked with ethanol or deionized water. Make sure that there is no
trace or dust left on the glass.
4 Install the carousel and the carousel cover.
Other electrodes:
° When 10,000 ISE tests are performed.
° When the ISE electrodes are used for 6 months since installation.
° When calibration fails or quality control is abnormal as the result of degraded electrode
performance.
Materials required
Reference electrode, ISE electrode
System status
Make sure that the status of the ISE module is Standby or Stopped
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles.
NOTE
After performing this procedure, recalibrate the ISE electrodes prior to starting analysis.
How to do
To remove the electrodes
1 Select Utility > Maintenance > Maintenance > ISE Maintenance.
2 Choose Replace Electrode.
3 Select desired electrodes, and enter the lot number and expiration date.
4 Select Add and then select OK.
5 Select Continue.
6 Open the ISE side door and remove the cover of the shielding box.
7 Open the electrode case, take out the electrode, remove the tapes around its inside, and then
use clean tissue to wipe it.
NOTE
Take out the insert from the reference electrode, and ensure no crystallized salt exists in and around it. If
needed, clean the electrode with warm water.
Make sure the red ball of the reference electrode floats on the internal fluid. Make sure the O rings of all
electrodes remain intact.
8 Remove all electrodes from the ISE module.
To install the new electrodes
1 To install a new reference electrode:
a. Place the reference electrode at the bottom of the ISE module and make the rear part of
the electrode contact closely with the internal wall of the ISE module.
b. Loosen the compressor and ensure the electrodes are fixed tightly.
2 Install other electrodes in the order of Cl, K, Na, and spacer from bottom to top.
3 If the O ring is lost, install a new one.
There are two more O rings in each electrode case.
4 Check if the electrode positions are correct:
The Na, K and Cl electrodes are of the same size and shape. Ensure that the electrodes
are inserted in the correct order.
If one of the electrodes cannot be easily pushed into the housing, check the electrode
first and then repeat the installation process.
Check if the 5 electrodes are relatively on the same straight line; otherwise, liquid
cannot flow through the electrode tubes smoothly.
5 Select Continue. The system primes the tubes with calibrators A and B.
6 Select Done.
7 Restore the cover of the shielding box and close the side door of the ISE module.
8 Run ISE calibration.
NOTE
The new electrode can be calibrated successfully only after certain time period.
9 If the calibration fails, perform the following operations:
Run ISE calibration for multiple times so that the electrodes can get steady quickly.
Or, drip little serum sample in the electrode channel, and leave it for 10-30 minutes, and
then run calibration again.
Purpose
To store the electrodes separately to prevent them from being damaged due to lack of water
while the analyzer is powered off.
Materials required
Electrode cases and tapes
When to do
Perform this procedure when the analyzer is going to be powered off for over 3 days. If it will
be powered off for no more than 3 days, prime the ISE electrodes to protect them from being
damaged.
System status
Make sure that the status of the ISE module is Standby or Stopped
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
NOTE
Make sure that the storage temperature is below 40℃.
How to do
1 Remove the reagent pack according to 10.9.13 Removing reagent pack on page 10-41.
2 Open the ISE side door and remove the cover of the shielding box.
3 Remove all electrodes from the ISE module.
4 To save the reference electrode:
a. Put back the insert to the cell of the reference electrode and prevent the crystallized salt
from clogging the cell.
b. Store the electrode in an electrode case at the room temperature in a sun-shielding place.
5 To save the Na, K and Cl electrodes:
a. Take out a little calibrator A from the reagent pack, inject it into the cell of the electrode
and seal it with tape. Make sure proper amount of calibrator is injected into the cell of the
electrode.
b. Store the capped electrodes in an electrode case at the room temperature in a
sun-shielding place.
6 Restore the cover of the shielding box and close the side door of the ISE module.
Materials required
Unclogging tool for the ISE waste tube, bleaching agent (Dilution Ratio 1:1) or 50% sodium
hypochlorite, and deionized water
System status
Make sure that the system is not running test.
Precautions
BIOHAZARD
Wear gloves and lab coat, and if necessary, goggles during the maintenance process.
Note
Excessive bleach and DI water flushed into the ISE reagent pack waste bag may cause waste bag over
expansion and clog the Cal A & Cal B reagent flow.
Do not spill liquid on the analyzer. Liquid ingression may cause equipment damage.
How to do
1 Ensure the analyzer is in Standby status. Open the ISE cover on the left panel.
Figure 10.4 Open the ISE cover
(1)
Figure 10.5 Removing the right angle adapter and waste pump
(2)
(1)
(3)
(2)
(1)
(1) and (2) Unclogging tool (3) Tube fitting for right angle adaptor
4 Press the wand release button to remove the wand from the current in use ISE reagent
pack and keep it in a save place. Engage the wand to an old used-up reagent pack.
5 Inject bleach into the ISE waste tube and soak the tube for 5 minutes. Discharge the waste
into the reagent pack.
6 If the bleach cannot be injected into the ISE pack, remove the wand and push down to open
the waste valve manually with a sharp object, and then inject again.
If bleach can go through this time, the waste bag was clogged and cannot be used.
If bleach still cannot be injected, replacing the ISE wand is recommended.
(1)
The filter on the DI water tube and the diluted wash solution tube should not be used for over
one year. When replacement is needed, contact our customer service department or your local
distributor.
This chapter describes how to view and edit error logs and edit logs and how to locate failure
and determine relevant corrective actions. Read this chapter thoroughly to achieve the best
performance of the instrument.
Help
Every error log is provided with online help information. Select the icon in front of an error
log. The descriptions, possible causes and solutions of the error are displayed.
Every error log contains the event ID, date/time, error description (by processing method),
event class (by subsystem) and symptom.
Choose the following buttons as needed:
Search F1: to search for error logs by date, event ID, symptom, or event class.
Refresh F2: to refresh the error logs based on the current search conditions.
Delete F3: to remove specified error logs on the screen.
Print F7: to print all error logs currently displayed on the screen.
Every edit log contains the serial number, date/time, operator, event type and description.
Choose the following buttons as needed:
Search F1: to search for edit logs based on the occurring date.
Refresh F2: to refresh the edit logs based on the current search conditions.
Delete F3: to remove specified edit logs on the screen.
Print F7: to print all edit logs currently displayed on the screen.
Refresh F2: to refresh the logs based on the current search conditions.
Delete F3: to remove specified logs on the screen.
Print F7: to print all logs currently displayed on the screen.
Flag
Flag is also called data alarm. When calibration error or failure, or sample result error occurs due
to the sample, reagent or system failure, a flag will appear near the corresponding calibration
result or sample results.
Error log
All alarms are recorded in the error logs. By recalling the error logs you are enabled to master
the current status of the system and troubleshoot errors.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Or The sample probe collides
with other object.
Sample syringe 4. Collision error:
movement error. The collision remains.
Position: 5. Moving vertically is not
Error: allowed in current position:
The sample probe moves
vertically in an unknown
position.
Sample probe horizontal
movement error
1. Sensor status error:
The sample probe assembly is
probably forced to move
horizontally.
2. Failed to find the zero
position:
The sample probe assembly is
obstructed when rotating.
3. Collision occurs during
horizontal movement:
The sample probe assembly is
obstructed when rotating.
4. Moving horizontally is not
allowed in current position:
The sample probe assembly is
probably forced to move
vertically.
Sample syringe movement
error.
1. Sensor status error:
The syringe assembly is
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
probably forced to move.
2. Failed to find the zero
position:
The syringe assembly is
probably jammed.
A01007 Sample probe unit Warning Sample probe collides / 1. Collision occurs during 1. Collision occurs during
with an obstacle when aspirating: aspirating:
aspirating The sample probe collides Remove the obstacle, and then
Sample position: with other object. recover failure by performing the
Sample ID/bar code: Home maintenance procedure.
Specific position:
A01021 Sample probe unit Warning The probe collides with / Clog detection board Recover the failure. If this message
an obstacle when communication error. appears for 3 times, contact our
aspirating. customer service department or
your local distributor.
A01024 Sample probe unit Warning Insufficient sample / There is no sample or 1. Check if the sample is sufficient,
or insufficient sample on the and then try again.
Sample probe level designated position. 2. If the error remains, contact our
detection failed. customer service department or
your local distributor.
A01027 Sample probe unit Warning Sample is insufficient or There is no sample or 1. Check if the sample is sufficient,
contains air bubbles insufficient sample on the and then try again.
Position: designated position. 2. If the error remains, contact our
Sample ID/bar code: customer service department or
or your local distributor.
Probe level detection
failed
Position:
Sample ID/bar code:
A01028 Sample probe unit Error Sample probe fails to / There is no deionized water, 1. Check if the water supply is
detect liquid level during or the deionized water is not
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
cleaning supplied normally. normal.
2. Recover the failure for 3 times. If
the error remains, contact our
Customer Service Department or
your local distributor.
A01029 Sample probe unit Error Sample is insufficient or / 1. The sample contains clots, 1. Check that the sample is
contains fibrins and clots or is too thick. preprocessed correctly; or check if
or Probe clog detection 2. The sample probe is the sample contains foreign
failed. clogged. matters such as clot. If it does,
Position: change the sample.
Sample ID/bar code: 2. Clean the sample probe with
diluted wash solution. If the
problem remains, remove the
sample probe and unclog it, and
then continue with the
measurement.
A01030 Sample probe unit Error Probe is clogged during / The sample probe is clogged. 1. Clean the sample probe with
cleaning. diluted wash solution. Remove the
sample probe and unclog it.
2. If the problem remains, contact
the manufacturer.
A01033 Sample probe unit Warning Sample probe fails to / There is no reagent or 1. Check if R1 volume is sufficient
detect liquid level on insufficient reagent in the and the reagent bottle is free of air
reaction carousel when reaction cuvette. bubbles, and then try again.
dispensing. 2. If the problem remains, contact
Cuvette No.: the manufacturer.
Sample ID/bar code:
Chemistry:
or
Sample probe level
detection failed.
Cuvette No.:
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Sample ID/bar code:
Chemistry:
A01036 Probe unit Error Probe level detection Level detection board Recover the failure. If this message
board communication communication error appears for 3 times, contact our
error customer service department or
your local distributor.
A01039 Probe unit Error Instruction execution / Instrument instructions Switch off the analyzing unit
error cannot be executed. power and switch on it again.
Recover failure by performing the
Home maintenance procedure. If
this message appears for 3 times,
contact our customer service
department or your local
distributor.
A02007 Probe unit Warning Probe R2 collides with / 1. Collision occurs during 1. Collision occurs during
an obstacle when aspirating: aspirating:
aspirating The probe R2 collides with Remove the obstacle and then
Reagent position: other object. recover the failure.
Specific position:
A02023 Probe unit Warning Insufficient reagent / There is no reagent or 1. Check if the reagent is sufficient,
Or insufficient reagent on the and then try again.
Probe R2 level detection designated position. 2. If the error remains, contact our
failed. customer service department or
your local distributor.
A02025 Probe unit Warning Probe dispenses / 1. The probe aspirates 1. Check if the reagent satisfies the
insufficient reagent nothing. requirement and is sufficient in
volume, and then try again.
2. Recover the failure. If this
message appears for 3 times,
contact our customer service
department or your local
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
distributor.
A02027 Probe unit Warning Water residues exist in / DI water residual exists in Recover the failure. If this message
the cuvette or Probe cuvette. appears for 3 times, contact our
level detection failed customer service department or
your local distributor.
A02032 Probe unit Warning Reagent is insufficient or 1. Air bubbles exist in the 1. Check if the reagent bottle
contains air bubbles reagent bottle. contains air bubbles, and then try
Chemistry: 2. The reagent bottle does not again.
Position: meet the requirements. 2. Check if the reagent bottle meets
the requirements.
3. If the error remains, contact our
customer service department or
your local distributor.
A05006 Mixer unit Error Mixer vertical movement / 1. Sensor status error. The Recover failure by performing the
error assembly is probably forced Home maintenance procedure. If
Mixer horizontal to move vertically. this message appears for 3 times,
movement error 2. Failed to find the zero contact our customer service
position. The mixer assembly department or your local
is probably jammed distributor.
3. Vertical movement is not
allowed in current horizontal
position.
The reagent mixer moves
vertically in an unknown
position.
1. Sensor status error. The
assembly is probably forced
to move vertically.
2. Failed to find the zero
position The mixer assembly
is obstructed when rotating
3. Horizontal movement is
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
not allowed in current vertical
position.
A05007 Mixer unit Error Instruction execution / Instrument instructions Switch off the analyzing unit
error cannot be executed. power and switch on it again.
Recover failure by performing the
Home maintenance procedure. If
this message appears for 3 times,
contact our customer service
department or your local
distributor.
A06006 Reaction carousel Error Reaction carousel / Reaction carousel movement Switch off the analyzing unit
unit movement error error power and switch on it again.
Error: 1. Failed to find the home Recover failure by performing the
position Home maintenance procedure. If
The reaction carousel is this message appears for 3 times,
obstructed or blocked. contact our customer service
department or your local
2. The coder missed steps
distributor.
The reaction carousel is
obstructed or blocked.
3. The reaction carousel
missed steps when moving to
the home position.
The reaction carousel is
obstructed or blocked.
A06008 Reaction carousel Error Instruction execution / Instrument instructions Switch off the analyzing unit
unit error cannot be executed. power and switch on it again.
Recover failure by performing the
Home maintenance procedure. If
this message appears for 3 times,
contact our customer service
department or your local
distributor.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
A07006 Sample/reagent Error Sample carousel / Sample carousel movement Recover the failure. If this message
carousel unit movement error error appears for 3 times, contact our
Error: 1. Failed to find the home customer service department or
position your local distributor.
The sample carousel is
obstructed or blocked.
2. The coder missed steps
The sample carousel is
obstructed or blocked.
3. The sample carousel
missed steps when moving to
the home position.
The sample carousel is
obstructed or blocked.
A07009 Sample/reagent Error Sample bar code reader / The sample bar coder reader Recover the failure. If the error still
carousel unit error goes wrong due to system remains, contact our Customer
failure. Service Department or your local
distributor.
A07010 Sample/reagent Warning Sample bar code error / Sample bar coder reader does Try again. If your attempt fails,
carousel unit Position: not work normally due to contact our customer service
communication error. department or your local
distributor.
A07011 Sample/reagent Error Sample bar code sending / Sample bar coder sending Recover the failure or reboot the
carousel unit buffer is full buffer is full due to analyzing unit.
communication error.
A09011 Sample/reagent Error Reagent bar code reader / The reagent bar coder reader Recover the failure. If the error still
carouse unit does not work normally goes wrong due to system remains, contact our Customer
failure. Service Department or your local
distributor.
A09012 Sample/reagent Warning Reagent bar code error / Reagent bar coder sending Try again. If your attempt fails,
carousel unit Position: buffer is full due to contact our customer service
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
communication error. department or your local
distributor.
A09014 Sample/reagent Error Reagent bar code / Reagent bar coder reader does Recover the failure or reboot the
carousel unit sending buffer is full not work normally due to analyzing unit.
Position: communication error.
A11005 Wash station Error Wash station movement / Wash station movement error Switch off the analyzing unit
error 1. Sensor status error power and switch on it again.
Error: The wash station assembly is Recover failure by performing the
probably forced to move. Home maintenance procedure. If
this message appears for 3 times,
2. Failed to find the home
contact our customer service
position
department or your local
The wash station assembly is distributor.
obstructed by other object.
3. The wash station collides
with an obstacle when
moving.
The wash station collides with
other object, or the wash
probes then collide with the
reaction carousel.
A11013 Wash station Error Water tank is empty / 1. The water tank is empty. 1. Check if the water level inside
2. The low-level floater of the the water tank is low.
water tank goes wrong. 2. Check if the error is accidental.
3. If the error is not accidental,
contact our customer service
department or your local
distributor.
A11015 Wash station Error Insufficient diluted wash / 1. The low-level floater of the 1. Check the floater of the diluted
solution diluted wash solution tank wash solution tank.
goes wrong. 2. Fill the diluted wash solution
2. The diluted wash solution tank.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
tank is empty. 3. Check if the error is accidental.
4. If the error is not accidental,
contact our customer service
department or your local
distributor.
A11020 Wash station Error High concentration / 1. The high concentration 1. Check the high-concentration
waste tank is full waste tank is full waste tank. If it is full, replace the
2. The floater of the high waste tank, close the full tank and
concentration waste tank goes dispose of the waste properly.
wrong. 2. Check if the error is accidental.
3. If the error is not accidental,
contact our customer service
department or your local
distributor.
A11034 Wash station Error Cuvette wash syringe / Sensor status error. The Switch off the analyzing unit
movement error. syringe assembly is probably power and switch on it again.
forced to move.2.Failed to find Recover failure by performing the
the mechanical zero position. Home maintenance procedure. If
The syringe assembly is this message appears for 3 times,
probably jammed. contact our customer service
department or your local
distributor.
A11038 Wash station Error Instruction execution / Instruction execution error Switch off the analyzing unit
error power and switch on it again.
Recover failure by performing the
Home maintenance procedure. If
this message appears for 3 times,
contact our customer service
department or your local
distributor.
A12005 Temperature unit Warning Reaction carousel T1 1. The ambient temperature 1. Check if the error is accidental.
temperature is out of is out of range. 2. If the error is not accidental,
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
range 2. The temperature sensor contact our customer service
goes wrong. (component department or your local
error and cable error) distributor.
3. The temperature
protection switch goes
wrong. (component error and
cable error)
4. The heater goes wrong.
(component error and cable
error)
5. Temperature control fan
error.
6 PCB error
7. Parameters are lost.
8. Electromagnetic
interference exists.
A12006 Temperature unit Warning Temperature of wash / 1. The ambient temperature 1. Check the temperature of the
solution for cleaning is out of range. deionized water for cleaning the
cuvettes is out of range 2. The temperature sensor whole unit.
Temperature: goes wrong. (component 2. Check if the water supply is
error and cable error) normal and has the temperature
3. The temperature between 15°C-30°C.
protection switch goes 3. Check if the error is accidental.
wrong. (component error and 4. If the error is not accidental,
cable error) contact our customer service
4. The heater goes wrong. department or your local
(component error and cable distributor.
error)
5. PCB error
6. Parameters are lost.
7. Electromagnetic
interference exists.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
A12007 Temperature unit Warning Temperature of / 1. The ambient temperature 1. Check the temperature of the
deionized water for is out of range. deionized water for cleaning the
cleaning cuvettes is out 2. The temperature sensor whole unit.
of range goes wrong. (component 2. Check if the water supply is
error and cable error) normal and has the temperature
3. The temperature between 15°C-30°C.
protection switch goes 3. Check if the error is accidental.
wrong. (component error and 4. If the error is not accidental,
cable error) contact our customer service
4. The heater goes wrong. department or your local
(component error and cable distributor.
error)
5. PCB error
6. Parameters are lost.
7. Electromagnetic
interference exists.
A12010 Temperature unit Warning Reagent preheating / 1. The ambient temperature 1. Check the temperature of the
temperature is out of is out of range. deionized water for cleaning the
range. 2. The temperature sensor whole unit.
goes wrong. (component 2. Check if the water supply is
error and cable error) normal and has the temperature
3. The temperature between 15°C-30°C.
protection switch goes 3. Check if the error is accidental.
wrong. (component error and 4. If the error is not accidental,
cable error) contact our customer service
4. The heater goes wrong. department or your local
(component error and cable distributor.
error)
5. PCB error
6. Parameters are lost.
7. Electromagnetic
interference exists.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
A14007 Temperature unit Warning Reagent refrigeration / 1. Reagent refrigeration 1. Check the reagent refrigeration
temperature is out of temperature is out of range. temperature sensor.
range. TDISP 2. The temperature sensor 2. Check if the error is accidental.
Temperature: goes wrong. (component 3. If the error is not accidental,
error and cable error) contact our customer service
3. Radiator goes wrong. department or your local
(component error and cable distributor.
error)
4. PCB error
5. Parameters are lost.
6. Electromagnetic
interference exists.
A14008 Temperature unit Warning Reagent refrigeration / 1. Reagent refrigeration 1. Check the reagent refrigeration
temperature is out of temperature is out of range. temperature sensor.
range. TDISP 2. The temperature sensor 2. Check if the error is accidental.
Temperature: ? goes wrong. (component 3. If the error is not accidental,
error and cable error) contact our customer service
3. Radiator goes wrong. department or your local
(component error and cable distributor.
error)
4. PCB error
5. Parameters are lost.
6. Electromagnetic
interference exists.
A21001 Probe Interior Error Probe interior wash / 1. Sensor status error. The Switch off the analyzing unit
Wash Unit syringe movement error. syringe assembly is probably power and switch on it again.
Error: forced to move.2.Failed to find Recover failure by performing the
the mechanical zero position. Home maintenance procedure. If
The syringe assembly is this message appears for 3 times,
probably jammed. contact our customer service
department or your local
distributor.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
A22001 ISE unit Error Slope out of range, SLEX 1. Electrode installation 1. Reinstall the electrode.
electrode: incorrect. 2. Replace the calibrator.
2. Calibrator expired. 3. Replace the problematic electrode
3. Electrode degenerated. and rerun.
4. Bubbles in reference 4. Remove the electrode and clap on
electrode. it to eliminate bubbles. Reinstall
5. Reference electrode has the electrode and run calibration.
been used for a long time. 5. Replace reference electrode and
6. Electrodes interfered. rerun.
7. Module or tubing 6. Troubleshoot the electrodes by
temperature above 32°C. replacing them in different groups.
7. Monitor temperature, if too high,
relocate equipment.
A22002 ISE unit Error Air in sample. Position: SMPA 1. Sample is insufficient or 1.&2. Increase the sample volume.
contain much bubbles after At least 90μl sample should be
dispensing. prepared.
2. No or insufficient sample 3. Electrode is not installed
has been dispensed into the correctly. Reinstall it.
sample injection port. 4. Check the waste tube, and if
3. Liquid leakage due to that necessary, replace it.
the electrodes are not
properly installed.
4. The waste pump tube is
aging or broken.
A22004 ISE unit Error ISE unit cannot be / 1. ISE power supply failure. 1. Replace the 24V power supply
connected. 2. ISE communication cable board.
failure. 2. Replace the ISE communication
3. Communication interface cable.
or pins failure. 3. Replace the interface or pins.
4. ISE main control board 4. Replace the ISE main control
failure. board.
A22005 ISE unit Error ISE unit response error / 1. ISE communication cable 1. Replace the ISE communication
failure. cable.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
2. Communication interface 2. Replace the interface or pins.
or pins failure. 3. Replace the ISE main control
3. ISE main control board board.
failure. 4. Upgrade the operating software
4. Software failure. or reinstall it.
A22006 ISE unit Error Purge A and B failed. / 1. Leaks exist due to 1. Reinstall the electrode and check
improperly-installed electrode for O ring.
or missing O ring. 2. Use warm water to clean and
2. Sample injection port or unclog the sample injection port
electrode inside is clogged. with fresh water and unclog the
3. Calibrator is exhausted. electrode tube. Check the reference
4. Prime combinations are electrode for crystallized salt.
not enough. 3. Replace the reagent pack.
5. Pump tube is aging, 4. Increase the prime cycle.
blocked, or broken. 5. Replace the pump tube.
6. Calibrator cannot be 6. Unclog the reagent pack tube
dispensed normally due to with warm water.
clogged reagent pack tube.
A22007 ISE unit Warning ISE reagent is going to / Calibrator is exhausted. Replace the reagent pack with a
be exhausted. new one.
A22008 ISE unit Error Voltage overflow, VOUT 1. Electrode or reagent pack 1. Replace the problematic electrode
electrode: failed. and reagent pack.
2. Electrode is unsteady. 2. New electrode will become
3. New reagent pack is steady after 15 minutes since
unsteady. installed.
4. Reference electrode has 3. Run a couple of calibrations after
been used for over 6 months. installing new reagent pack.
5. ISE main control board 4. Replace the reference electrode.
failure. 5. Replace the ISE main control
board.
A22009 ISE unit Error Electrode slope drift. VDRF/ 1. Electrode or reagent pack 1. Replace the problematic electrode
(during calibration) SLDR failed. and reagent pack.
Or 2. Electrode is unsteady. 2. New electrode will become
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Electrode voltage drift. 3. New reagent pack is steady after 15 minutes since
(during sample analysis) unsteady. installed.
Electrode: 4. Reference electrode has 3. Run a couple of calibrations after
been used for over 6 months. installing new reagent pack.
5. ISE main control board 4. Replace the reference electrode.
failure. 5. Replace the ISE main control
6. Ambient temperature board.
fluctuates drastically. 6. Control the ambient temperature
to make the fluctuation within
+/-4℃.
A22010 ISE unit Error Voltage noise, electrode: NOIS 1. Electrode failure. 1. Replace the electrode.
2. Environment interference. 2. Relocate the instrument.
3. ISE main control board 3. Replace the ISE main control
failure. board.
4. Salt buildup around 4. Clean the tubes and electrodes.
electrodes or tubes due to
fluidic leaks.
A22011 ISE unit Error Air in calibrator B PUGB 1. Calibrator B is exhausted. 1. Replace the reagent pack with a
2. Bubbles exist in calibrator new one.
tube B. 2. Perform purge B to remove
3. Pump tube B is aging, bubbles.
blocked, or broken. 3.&4. Replace the pump tube.
4. Waste pump tube B is 5. Clean the sample injection port
aging, blocked, or broken. and reinstall electrodes.
5. Sample injection port and 6. Replace the air bubble detector.
fluidic path are blocked or
leaking.
6. Air bubble detector fails.
A22012 ISE unit Error Air in calibrator A PUGA 1. Calibrator A is exhausted. 1. Replace the reagent pack with a
2. Bubbles exist in calibrator new one.
tube A. 2. Perform purge A to remove
3. Pump tube B is aging, bubbles.
blocked, or broken. 3.&4. Replace the pump tube.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
4. Waste pump tube B is 5. Clean the sample injection port
aging, blocked, or broken. and reinstall electrodes.
5. Sample injection port and 6. Replace the air bubble detector.
fluidic path are blocked or
leaking.
6. Air bubble detector fails.
A22013 ISE unit Error ISE pump calibrating / 1. Pump tube is aging. 1. Replace the pump tube.
failed! 2. Sample probe 2. Replace the sample probe.
aspiration/dispensing failure.
A22014 ISE unit Error Air bubble detector / 1. Air bubble detector board is Replace the bubble detector.
failure eroded due to the leaks at the
joint of sample injection port
and bubble detector.
2. Air bubble detector fails.
A22015 ISE unit Error Reading reagent pack / 1. Reagent pack is not 1. Install reagent pack.
chip error installed. 2. Replace the wand.
2. Reagent pack wand fails.
A22016 ISE unit Error Reagent pack chip / 1. Reagent pack is not 1. Install reagent pack.
writing error. Unload the installed. 3. Replace the wand.
reagent pack and load it 3. Reagent pack wand fails.
again.
A22017 ISE unit Error Air in ISE cleaning / 1. ISE cleaning solution is 1. Place sufficient ISE cleaning
solution insufficient. solution.
2. Waste pump tube B is 2. Replace the pump tube.
aging, blocked, or broken. 3. Clean the sample injection port
3. Sample injection port and and reinstall electrodes.
fluidic path are blocked or 4. Replace the air bubble detector.
leaking.
4. Air bubble detector fails.
A22018 ISE unit Error No fluid in tubing CALF 1. Waste pump tube B is 1. Place sufficient ISE cleaning
aging, blocked, or broken. solution.
2. Sample injection port and 2. Replace the pump tube.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
fluidic path are blocked or 3. Clean the sample injection port
leaking. and reinstall electrodes.
3. Air bubble detector fails. 4. Replace the air bubble detector.
A22019 ISE unit Error Saving calibration result DEP 1. ISE communication cable 1. Replace the ISE communication
error failure. cable.
2. Communication interface 2. Replace the interface or pins.
or pins failure. 3. Replace the ISE main control
3. ISE main control board board.
failure. 4. Upgrade the operating software
4. Software failure. or reinstall it.
A22021 ISE unit Error Command format or RESP 1. ISE communication cable 1. Replace the ISE communication
execution error failure. cable.
2. Communication interface 2. Replace the interface or pins.
or pins failure. 3. Replace the ISE main control
3. ISE main control board board.
failure. 5. Upgrade the operating software
4. Software failure. or reinstall it.
A22022 ISE unit Error No fluid in tubing / 1. Waste pump tube B is 1. Place sufficient ISE cleaning
aging, blocked, or broken. solution.
2. Sample injection port and 2. Replace the pump tube.
fluidic path are blocked or 3. Clean the sample injection port
leaking. and reinstall electrodes.
3. Air bubble detector fails. 4. Replace the air bubble detector.
A22023 ISE unit Error No reagent module has / 1. Reagent pack is not 1. Install reagent pack.
been loaded. installed. 2. Replace the wand.
2. Reagent pack wand fails.
A22024 ISE unit Error ISE response check code RESP 1. The communication wire 1.Replace the communication wire
error between ISE and the 2.Change the interface or the pin.
middle-layer unit goes wrong. 3.Change the main control board.
2. Communication interface
4.Upgrade the software or reinstall
or pin error.
the software.
3.Main control board does not
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
function.
4.Software error
A22027 / Error Fluidic prime failed. / Fluidic is not primed. 1. Turn off the analyzing unit
power and reswitch it on.
2. If the error occurs for
continuous three times, please
contact our customer service or
your local distributor.
A22036 / Error Initializing sample bar / Sample bar code reader failed 1. Recover failure by performing
code reader failed. due to system error. the Home maintenance procedure.
2. If this error remains
contact our customer service
department or your local
distributor.
A22038 / Error Scanning reagent bar / Scanning reagent bar code 1. Turn off the analyzing unit
code failed. failed. power and reswitch it on.
2. If the error occurs for
continuous three times, please
contact our customer service or
your local distributor.
A22039 / Error Unmatched software / 1.Version inquiry instruction 1. Turn off the analyzing unit
version. failed. power and reswitch it on.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Sensor disconnected or department or your local
temperature below 0℃. distributor.
Warning No.3:
Heater is heating for a long
time
Warning No.4:
Heater control is locked.
A22043 Temperature Warning Temperature of cuvette / Warning No. 1: 1. Turn off the analyzing unit
control unit wash solution is out of Sensor shortcut or power and reswitch it on.
range: %s\r\n temperature above 75℃. 2. If the error is not accidental,
Temperature: %s Warning No. 2: contact our customer service
Sensor disconnected or department or your local
distributor.
temperature below 0℃.
Warning No.3:
Heater is heating for a long
time
Warning No.4:
Heater control is locked.
C00007 Operating system Error CPU performance low / The CPU is too busy. Reboot the computer and
operating software. If this message
appears for 3 times, contact our
customer service department or
your local distributor.
C00011 Operating system Error The last abnormal exit / The operating software is Restart the operating software, and
may cause carryover not abnormal, or the instrument execute the Special Wash
handled. Execute the power is cut off unexpectedly. maintenance command before
Special Wash starting analysis.
maintenance command
before starting analysis
to ensure accurate
results.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
C00012 Operating system Warning Sound card failure / No sound card is installed. Reinstall the sound card or the
Sound card failure. Incorrect sound card driver.
sound card driver.
C00013 Operating system Error The cuvette status may / The operating software is not Take out and check the standby
not be refreshed due to exited normally due to some cuvettes. If they are used, replace
the last abnormal exit of reasons.(BS-230) them. When you are uncertain if
the system. To ensure they are used, replace them.
the correct test results,
please check the cuvettes
and replace them.
C01001 Instrument Error Equipment cannot be / The serial cable is not Check the serial port connection.
connection connected connected; or the analyzing Replug the cable. Check if the
unit power is switched off. analyzing unit is powered on. Start
the initialization again. Restart the
computer and analyzing unit. If
three continuous attempts are
failed, contact our customer service
department or your local
distributor.
C02001 Database Error Database initialing failed / The database file is damaged Reboot the computer and
or lost. analyzing unit. If three continuous
attempts are failed, contact our
customer service department or
your local distributor.
C02002 Database Error Database upgrade failed / The database file is damaged Reboot the computer and
or lost. analyzing unit. If three continuous
attempts are failed, contact our
customer service department or
your local distributor.
C02004 Database Warning Database backup failed / The database file is damaged Reboot the computer and
or lost. analyzing unit. If three continuous
attempts are failed, contact our
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
customer service department or
your local distributor.
C02005 Database Warning Reading/Writing / The database does not work Reboot the computer and
database failed normally. analyzing unit. If three continuous
attempts are failed, contact our
customer service department or
your local distributor.
C03001 Result calculation Warning Result cannot be RCE Absorbance data for Rerun the test. If the error
calculated calculation is incomplete, or remains, contact our customer
Sample ID/bar code: the dividend is 0. service department or your local
Position: distributor.
Chemistry:
C03002 Result calculation Warning Absorbance out of range ABS 1. Probe R1 dispenses 1. Observe the reaction curve. If
Sample ID/bar code: insufficient reagent, or air the absorbance of R1 is too high,
Position: bubbles exist in the reagent. check the reagent for air bubbles
2. The reagent is placed in an and the syringe for leaking.
Chemistry:
incorrect position or is 2. Check if the reagent has been
abnormal. placed in the correct position.
3. The sample concentration 3. Rerun the test after dilution.
is too high, resulting in great 4. Contact our customer service
response. department or your local
4. The absorbance data used distributor.
for calculation is incomplete
(due to photoelectric data
loss), or the error of division
by zero occurs.
C03003 Result calculation Warning R1 blank absorbance out RBK The reagent goes wrong; the Check if the reagent is sufficient
of range cuvette is not clear; the without air bubbles and the
reaction cuvette is chemistry parameters are
overflowed; or insufficient reasonable. If yes, replace the
reagent is dispensed. reagent and then rerun the test.
Check if the cuvette is normal. If
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
the error remains, contact our
customer service department or
your local distributor.
C03004 Result calculation Warning Substrate depletion BOE The sample concentration is Check the reaction curve and the
Sample ID/bar code: too high, and substrate substrate depletion limit. Rerun the
Position: depletion occurs during test with diluted sample.
fixed-time measurements.
Chemistry:
C03005 Result calculation Warning Result cannot be ENC The sample concentration is Check the reaction curve and the
calculated too high, and substrate substrate depletion limit. Rerun the
Sample ID/bar code: depletion occurs within the test with diluted sample.
Position: lag time of rate check
measurements.
Chemistry:
C03006 Result calculation Warning Linearity limit out of LIN The measuring points for Check the reaction curve and the
range result calculation are substrate depletion limit. Rerun the
Sample ID/bar code: nonlinear, because the sample test with diluted sample.
Position: concentration is too high, or
the substrate depletion limit is
Chemistry:
not specified or unreasonable.
C03007 Result calculation Warning Prozone check error PRO Antibody excess occurs due to Check the reaction curve and the
Sample ID/bar code: too high sample prozone check parameters. Rerun
Position: concentration. the test with diluted sample.
Chemistry:
C03008 Result calculation Warning Sample concentration is RRN The sample concentration Rerun the test with diluted sample.
higher than that of the exceeds the high limit of the
highest-level calibrator calibrator concentration.
Sample ID/bar code:
Position:
Chemistry:
C03009 Result calculation Warning Mixed blank absorbance MBK The reagent goes wrong; the Check if the reagent is sufficient
out of range cuvette is not clear; the without air bubbles and the
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Chemistry: reaction cuvette is chemistry parameters are
overflowed; or insufficient reasonable. Check if the cuvette is
reagent is dispensed. normal. Replace the reagent and
then rerun the test. If the error
remains, contact our customer
service department or your local
distributor.
C03010 Result calculation Warning Blank response out of BLK The reagent goes wrong; Check if the reagent is sufficient
range insufficient reagent is without air bubbles and the
Chemistry: dispensed; the cuvette chemistry parameters are
contains air bubbles; the light reasonable. Check if the cuvette is
drifts; or the cuvette is normal. Replace the reagent and
overflowed. (BS-240) then rerun the test. If the error
The reagent goes wrong; remains, contact our customer
insufficient reagent is service department or your local
dispensed; the cuvette distributor.
contains air bubbles; the light
drifts;(BS-230)
C03011 Result calculation Warning Calibration repeatability DUP The difference between the Check if the acceptance limit is
exceeds limit. Chem: maximum and minimum reasonable, troubleshoot the error,
response of the calibrator and then recalibrate.
exceeds the specified limit.
C03012 Result calculation Warning Calibration sensitivity SEN The difference of final Check if the acceptance limit is
exceeds limit. Chem: response of the maximum reasonable and the reagent and
and minimum concentration calibrator are normal, and then
calibrators exceeds the recalibrate.
specified limit.
C03013 Result calculation Warning Calibration curve SD CSD The calculated standard Check if the acceptance limit is
exceeds limit, Chem: deviation of the calibration reasonable and the reagent and
curve exceeds the specified calibrator are normal, and then
limit. recalibrate.
C03014 Result calculation Warning Calibration DET The calculated determination Check if the acceptance limit is
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
determination coefficient of the calibration reasonable and the reagent and
coefficient exceeds limit, curve exceeds the specified calibrator are normal, and then
Chem: limit. recalibrate.
C03015 Result calculation Warning Calibration slope FAC The slope difference is Check if the acceptance limit is
difference out of range. applicable to linear reasonable and the reagent and
Chem: calibration only and refers to calibrator are normal, and then
the K factor (slope) difference recalibrate.
between two adjacent
calibrations. It exceeds the
specified limit.
C03016 Result calculation Warning Calibration curve is not MON The calibration data and Check if the calibrator is defined
monotonic, Chem: calibration curve are not and placed correctly, and then
monotonic. recalibrate.
C03017 Result calculation Warning Calibration curve is not COV For nonlinear calibration, a Check that the reagent and
convergent, Chem: satisfying base cannot be calibrator are normal, and then
calculated and no calibration recalibrate. If the error remains,
curve is drawn. contact our customer service
department or your local
distributor.
C03018 Result calculation Warning Chemistry: 1-2s The QC result is between ±2 No actions are required.
Control: 1-2s warning and ±3 standard deviations
from the assigned mean
concentration.
C03019 Result calculation Warning Chemistry: 1-3s The QC result is greater than Check if the reagent is qualified
Control: 1-3s out of ±3 standard deviations from and control is normal. If the error
control the assigned mean remains, contact our customer
concentration. service department or your local
distributor.
C03020 Result calculation Warning Chemistry: 2-2s Results of two controls or two Check if the reagent is qualified
Control: 2-2s out of results of one control within a and control is normal. If the error
control run are simultaneously remains, contact our customer
greater than +2 or -2 standard service department or your local
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
deviations from the assigned distributor.
mean.
C03021 Result calculation Warning Chemistry: R-4s One result of a run is greater Check if the reagent is qualified
Control: R-4s out of than +2 standard deviations and control is normal. If the error
control from the assigned mean and remains, contact our customer
the other greater than -2SDs. service department or your local
distributor.
C03022 Result calculation Warning Chemistry: 4-1s Results of two runs in Check if the reagent is qualified
Control: 4-1s out of two-control evaluation or four and control is normal. If the error
control continuous results of a control remains, contact our customer
are greater than +1 or -1 service department or your local
standard deviation from the distributor.
assigned mean concentration.
C03023 Result calculation Warning Chemistry: 10-x Results of five runs in Check if the reagent is qualified
Control: 10-x out of two-control evaluation or ten and control is normal. If the error
control continuous results of a control remains, contact our customer
that are being compared are service department or your local
on the same side. distributor.
C03024 Result calculation Error Biochemistry test period / 1. Software error Rerun the test. Reboot the
time out. Cannot 2. Operating system error operating software, analyzing unit
continue and computer. If the error remains,
contact our customer service
department or your local
distributor.
C03026 Result calculation Warning Photoelectric data is lost / Communication error. If the error persists, contact our
customer service department or
your local distributor.
C03027 Result calculation Warning Chemistry: 2.7s Multiple QC data and Check if the reagent is qualified
Control: 1.0-2.7 out of threshold values or and control is normal. If the error
control cumulative sum exceed remains, contact our customer
±2.7SD. service department or your local
distributor.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
C03028 Result calculation Warning Chemistry: 3.0s Multiple QC data and Check if the reagent is qualified
Control: 1.0-3.0 out of threshold values or and control is normal. If the error
control cumulative sum exceed remains, contact our customer
±3.0SD. service department or your local
distributor.
C03029 Result calculation Warning Chemistry: 5.1s Multiple QC data and Check if the reagent is qualified
Control: 0.5-5.1 out of threshold values or and control is normal. If the error
control cumulative sum exceed remains, contact our customer
±5.1SD. service department or your local
distributor.
C03030 Result calculation Error Photoelectric TD 1. Software error 1. Rerun the operating software.
measurement period is 2. Reboot the operation unit.
out of range 3. If the error remains, contact our
Sample ID/bar code: customer service department or
Position: your local distributor.
Chemistry:
C03031 Result calculation Error Multiple consecutive / 1. Software error 1. Rerun the operating software.
photoelectric 2. Reboot the operation unit.
measurements are time 3. If the error remains, contact our
out customer service department or
Sample ID/bar code: your local distributor.
Position:
Chemistry:
C04001 Sample bar code Warning Duplicate sample bar / Duplicate bar code is used. Replace the duplicate sample bar
code. code label.
Sample ID/bar code:
Position 1:
Position 2:
C04002 Sample bar code Warning Bar code has no / The sample of the bar code Program the sample of the bar
corresponding has not been programmed. code.
programming.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Sample ID/bar code:
Position:
C04006 Sample bar code Warning Sample is expired / The sample is loaded after its The sample is expired. Replace the
Sample ID/bar code: shelf life is exceeded. sample and program it again.
Position: Reject the expired sample. If the
sample shelf life is too short,
change it to a reasonable one.
C04008 Sample bar code Warning Sample bar code too long. / The bar code length is greater Redefine the bar code with no more
Position: than the maximum value of than 27 digits.
27 digits.
C04009 Sample bar code Warning Sample bar code is less / The sample bar code is too Reprint the bar code and ensure it
than 3 digits. short, less than the minimum is no less than 3 digits.
position: range of 3 digits.
C04012 Sample bar code Warning Sample bar code analysis / Barcode information does not Reset the barcode format or
error conform with the barcode reprint the barcode and scan it.
Sample bar code: format
Position:
C05001 Reagent bar code Warning Duplicate reagent bar / Incorrect reagent or reagent Reprint the reagent bar code, or
code bar code is being used, or an replace the reagent bottle with an
Reagent: invalid reagent bar code is invalid bar code.
Position 1: being used. Bar code is
aligned with reagents, and
Position 2:
cannot be used again for new
reagent when a reagent is
exhausted.
C05002 Reagent bar code Warning Reagent bar code data / Incorrect reagent bar code is Print the new reagent bar code
error. being used, or reagent bar with correct settings and check the
Position: code is not configured bar code against the settings.
reasonably. The reagent bar Replace the reagent bottle, or
code contains incomplete or contact the reagent supplier.
incorrect reagent
information, such as
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
expiration date, reagent
volume, etc.
C05003 Reagent bar code Warning Reagent bar code / Incorrect reagent bar code is Check the reagent bar code
analysis error being used, or reagent bar settings, or reprint the reagent bar
Position: code settings are incorrect. code against the settings. Contact
The system fails to extract the reagent supplier.
reagent information from the
bar code.
C05006 Reagent bar code Error Wash solution position / Reagent rather than wash Reposition the reagent, or remove
on reagent carousel is solution is placed in the fixed it from the fixed reagent position.
occupied by another wash solution position (D) on
reagent reagent carousel.
Position:
C05008 Reagent bar code Error Physiological saline / Reagent rather than Reposition the reagent, or remove
position on reagent physiological saline is placed it from the fixed wash solution
carousel is occupied by in the fixed physiological position.
another reagent saline position (W) on
Position: reagent carousel.
C05010 Reagent bar code Error The pretreatment / Reagent rather than Reposition the reagent, or remove
reagent position is pretreatment reagent is it from the pretreatment reagent
occupied by other placed in the pretreatment position.
reagent. Position: reagent position on reagent
carousel.
C05011 Reagent bar code Error The ISE cleaning solution / Reagent rather than ISE Reposition the reagent, or remove
position is occupied by cleaning solution is placed in it from the ISE cleaning solution
another reagent. the pretreatment reagent position.
position on reagent carousel.
C06001 Host Error LIS initialization error / Host file is damaged or does Reinstall the operating software.
communication not exist.
C06002 Host Error LIS communication / Host parameters error Re-set or modify the host
communication parameter error communication parameters.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
C06003 Host Error LIS communication / Communication error If the error occurs accidentally,
communication error send or receive the instruction
again. If the error still remains,
contact our customer service
department or your local
distributor.
C06004 Host Error LIS host cannot be / Abnormal network Check Disconnections and network
communication connected connection or the LIS host is cable. Check if LIS host and LIS
not started. station can start normally.
C06005 Host Warning Sending sample results / Communication error If the error occurs accidentally,
communication failed. send or receive the instruction
Sample ID/bar code: again. If the error still remains,
Position: contact our customer service
department or your local
distributor.
C06006 Host Warning Sending sample / Communication error If the error occurs accidentally,
communication information failed. send or receive the instruction
Sample ID/bar code: again. If the error still remains,
Position: contact our customer service
department or your local
distributor.
C06007 Host Warning Inquiring sample / LIS host failure. If the error occurs accidentally,
communication information failed. neglect it. If the error occurs
Sample ID/bar code: frequently, contact the
Position: manufacturer of LIS or contact our
customer service department or
your local distributor.
C06008 Host Warning Downloading sample / Incorrect channel settings or Check and re-set the chemistry
communication failed. insufficient or redundant correspondence between the
Sample ID/bar code: chemistries on the LIS host. operating software and the LIS
Position: host.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
C07003 Light source Error Light intensity is too / 1. The lamp is not installed 1. Check if the lamp is installed
weak correctly. correctly.
2. The cuvette is 2. Perform the diluted wash
contaminated. procedure and then the lamp check
3. The lamp is aging. procedure.
4. The wash station dispenses 3. Replace the lamp.
liquid incorrectly. 4. Check if the wash station
5. The photoelectric collection dispenses liquid with correct
board goes wrong. volume to reaction cuvettes.
5. If your attempt fails, contact our
customer service department or
your local distributor.
C07004 Light source Warning Cuvette blank out of / 1. The cuvette is 1. Open the reaction carousel and
range contaminated. check if the lamp is turned on. If it
Cuvette No.: 2. The lamp is aging. is not, rerun the operating
3. The lamp is not installed software.
correctly. 2. Check if the lamp is installed
4. The wash station dispenses correctly.
liquid incorrectly. 3. Perform the diluted wash
5. The photoelectric collection procedure and then the cuvette
board goes wrong. check procedure.
4. Replace or clean the failed
cuvette.
5. Replace the lamp.
6. Check if the wash station
dispenses liquid with correct
volume to reaction cuvettes.
7. If your attempt fails, contact our
customer service department or
your local distributor.
C07005 Light source Error Lamp is not turned on / 1. The lamp is damaged. 1. Open the reaction carousel and
2. The lamp cable is not check if the lamp is turned on. If it
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
connected properly. is not, rerun the operating
3. The power board of the software.
lamp is not connected 2. Check if the lamp cable is
properly. tightened.
4. The power supply of the 3. Replace the lamp.
analyzing unit is 4. If your attempt fails, contact our
disconnected. customer service department or
5. The photoelectric collection your local distributor.
board goes wrong.
C07006 Light source Error Light intensity is too / 1. A cuvette position has no 1. Check if all cuvette positions
strong cuvette installed. have cuvettes installed.
2. The circuit gain is too high 2. Contact our customer service
and beyond the measurement department or your local
range. distributor to adjust the gain.
C07007 Light source Error Dark current is too high / 1. The photoelectric collection If three continuous attempts are
Channel: board goes wrong. failed, contact our customer service
AD: 2. The photoelectric collection department or your local
board or its shielding box is distributor.
not properly earthed.
C07009 Light source Error Water blank out of range L! 1. The cuvette wash station is 1. Check if the cuvette is
(10X) overflowing. overflowing.
2. The lamp has been replaced 2. Check if the Replace Lamp
incorrectly. command is executed during lamp
3. Cuvette check is not replacement.
performed after maintenance. 3. Check if the Cuvette Check
4. The cable connectors are command is executed after
not tightened. maintenance.
5. The retaining screw is not 4. Check if the cleaning liquid inside
tightened. the cuvette is no less than half of
6. The wash station dispenses the cuvette.
insufficient fluid. 5. Check if the cable connectors
and retaining screw of the lamp
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
7. The lamp is aged. have been tightened.
6. Check if the reaction curve
fluctuates irregularly. If yes, replace
the lamp.
7. If the error remains, contact our
customer service department.
C07012 Other error of Warning Storage device error. / No U disk is inserted. No file is Check if a U disk is inserted or full.
operation unit Cannot import data found in the U disk, or file Check if the storage device is
error, or file is damaged. The U damaged.
disk is locked or damaged.
C07013 Other error of Warning Storage device error. / No U disk is inserted. Check if a U disk is inserted or full.
operation unit Cannot export data Insufficient disk space. The U Check if the storage device is
disk is locked or damaged. damaged.
C07014 Other error of Warning Reagent exhausted / All reagents of the reagent Refill or replace the reagent.
operation unit Chemistry: type for the chemistry are less
Position: than the minimum limit. All
reagents of the type are too
little to be detected.
C07016 Other error of Warning Insufficient wash / Insufficient wash solution on Refill the wash solution on the
operation unit solution the reagent carousel. reagent carousel.
Position:
C07017 Other error of Warning Wash solution is / The wash solution on the Refill the wash solution on the
operation unit exhausted reagent carousel is exhausted. reagent carousel.
Position:
C07022 Other error of Warning Less than X tests are left / All reagents of the reagent Refill or replace the reagent.
operation unit in biochemistry reagent. type for the chemistry are less
Chemistry: than the minimum limit. All
reagents of the type are too
little to be detected.
C07023 Other error of Warning Chemistry: %s, 30 / The calibration factors will be Recalibrate the chemistries.
operation unit minutes left for next expired.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
calibration.
C07027 Other error of Warning Calibrator %s has been / The calibrator is expired. Replace the calibrator.
operation unit expired
C07028 Other error of Warning Chemistry: %s, lot / The reagent is expired. Replace the reagent.
operation unit No.: %s, position: %s,
has been expired
C07029 Other error of Warning Chemistry: %s, lot / The on-board stability time of Replace the reagent.
operation unit No.: %s, the reagent pack is too long.
position: %s, has
exceeded the on-board
stability time
C07034 Other error of Warning Insufficient physiological / Insufficient physiological Refill the physiological saline on
operation unit saline saline. the reagent carousel.
Position:
C07035 Other error of Warning Physiological saline is / Physiological saline is Refill the physiological saline on
operation unit exhausted exhausted. the reagent carousel.
Position:
C07036 Other Warning Chemistry: %s. / The calibration factors have Recalibrate the chemistry.
Calibration factors are been expired.
expired
C07037 Other Warning Reagent bottle number / Serial number of the reagent Recalibrate the chemistry.
of %s chemistry is is changed.
changed. Please
recalibrate
C07038 Other Warning Reagent lot number / Lot number of the reagent is Recalibrate the chemistry.
of %s chemistry is changed.
changed. Please
recalibrate
C07039 Other Warning Calibration factors of %s / The calibration factors are Recalibrate the chemistry.
chemistry are expired. expired.
Event ID Component Event Error Message and Flag Probable Causes Corrective Actions
class Event Log
Recalibrate
C07040 Other Warning Reagent exhausted / 1. The reagent is running out. Refill or replace the reagent.
Chemistry: 2. The reagent is too little to
be detected.
C07041 Other Error ISE reagent is less / ISE reagent inventory is Check the inventory. If the reagent
than %s below the alarm limit is insufficient, load the reagent.
C07042 Other Warning %s, lot number: %s, / One or more special reagents Replace them with new reagents.
position: %s, has been have been expired.
expired
C07043 Other Warning Pretreatment reagent is / The pretreatment reagent is Add more pretreatment reagent in
exhausted. running out. specified position of the reagent
Position: carousel.
C07044 Other Warning Pretreatment reagent is / The pretreatment reagent is Add more pretreatment reagent in
insufficient. insufficient. specified position of the reagent
Position: carousel.
C07045 Other Warning ISE cleaning solution is / ISE cleaning solution is Add the ISE cleaning solution.
exhausted. exhausted.
C07046 Other Warning ISE cleaning solution is / ISE cleaning solution is Add the ISE cleaning solution.
insufficient. insufficient.
This chapter gives brief introduction of the operation theories of the instrument, which include:
Principles of biochemistry measurement
Calibration math model and calculation of factors
QC determination methods
Prozone check
Principles of ISE measurement
12.1 Overview
The system is a fully automated computer-controlled clinical chemistry analyzer allowing
random selection of chemistries. It is capable of running a variety of chemistries based on the
operation theories and measurement principles.
The system performs measurement and generates the test results in the following procedure:
Figure 12.1 Measurement workflow
AD value
Absorbance
Response
Calibration factors
QC conclusion
The system measures the light intensity through photoelectric conversion, linear amplification
and AD conversion, and then calculates the reaction mixture’s absorbance and the absorbance
change rate, that is, the response, based on which the calibration factors are obtained. The
system performance is evaluated according to the test results of the control samples. If the
system is working normally, you may start the analysis of patient samples and the system will
calculate the sample results with the calibration factors.
Calculation of K factor
The system provides four K factors for result calculation, which are expressed through the
following equations:
VR1
k1
VR1 VS
VR1 VS
k2
VR1 VS VR 2
Where, VR1 and VR2 are the volumes of R1 and R2; V s is the actual sample volume dispensed for
reaction.
Calculation of response
The response in endpoint measurements is calculated as follows:
R Ai k Ab
k is the calculation factor and varies with the chemistry parameters.
S S 0 e kt
Where,
S0: the initial substrate concentration
e: base of the natural log
k: velocity constant
The change of substrate concentration Δ[S] over a fixed time interval, t1 to t 2 , is related to [S0]
by the following equation:
[ S ]
[ S 0] kt1 kt 2
e e
That is, the change in substrate concentration is directly proportional to its initial concentration
within a fixed time interval (Tl-Tm). This is the common feature of rate measurements. Within
this interval, the absorbance change is directly proportional to the analytes concentration. The
fixed-time reaction is also called, rate reaction, first-order Kinetic reaction and two-point Kinetic
reaction.
It is available in single-interval and double-interval according to the input mode of measuring
points. In the double-interval reaction, the sample blank, which is the absorbance change at two
points within the incubation time, is subtracted from the reaction absorbance.
The fixed-time measurements allow the check of substrate depletion at the two measuring points.
When detecting substrate depletion, the system will flag the test result with "BOE" and give an
alarm.
AM AL A AN
R 60*( k P )
tM t L tP tN
In fact, it is impossible for the substrate concentration to be absolutely high, and the reaction
will be no longer a zero-order reaction when the substrate is consumed to certain degree.
Therefore, the reaction type only stands within certain reaction period. In addition, the reaction
can become steady only after a period of time, because the reaction is complicated at the
beginning and there are miscellaneous reactions due to complex serum compositions.
In Kinetic reaction, the concentration or activity is obtained according to the absorbance change
among specified measuring points.
Determination of linearity
range
Enter substrate
No
depletion limit?
Yes
Number (N)
of measuring points without No
No
substrate depletion between L
and M′ is greater than
or equal to 3.
Yes
The number (N) of measuring points within the substrate depletion limit is monitored for
different operations:
If N≥3, the linearity range includes all measuring points from the reaction start point to the
substrate depletion limit; If substrate is depleted, the result will be flagged with “BOE”.
If N=2, the system will give the flag "NLN"and “BOE” while using two measuring points for
calculating the response.
If N=0 or 1, when Enzyme Linear Extension option is enabled. For Enzyme Linear Extension,
the system will give the flag”EXP”,”NLN” and “BOE”.
Calculation of response
Absorbance change rate⊿ALM' within the reaction time
The response ⊿ALM' within L-M' is calculated with the least square method.
M'
(T T ) ( A A)
i i
A LM' 60 * i L M'
(T T )
iL
i
2
Where,
L: start point of the linearity rangeM': end point of the linearity range
Ai: absorbance measured at measuring point i
A : average absorbance within L-M'
Ti: actual measuring time (second) at measuring point i
T : average measuring time within L-M
If there are less than two measuring points without substrate depletion within the reaction time,
the system will calculate the absorbance change rate by extending the enzyme linearity rang.
Absorbance change rate ⊿ANP within the blank time
The absorbance change rate ⊿ANP within the blank time is calculated with the same equation
as ⊿ALM'.
If N=P=0, the absorbance change rate within the blank time is 0.
Calculation of response
The response in Kinetic measurements is calculated as follows:
R ALM' K A NP
k is the calculation factor and varies with the chemistry parameters.
Af Ab
Linearity= 100 Linearity Limit
Au ,v
Where, A f , Ab and Au ,v are the absorbance change rates in the front part, back part
and at all measuring points of the reaction. These three values are calculated based on the
number of measuring points within the linearity range.
When N>8, A f is the absorbance change rate of the first 6 measuring points, Ab of
the last 6 measuring points, and Au ,v of all measuring points.
The system will compare the calculated linearity with that defined for the chemistry, and will
flag the test result with “LIN” and given an alarm if the configured linearity is exceeded.
Lag time
Reaction Time
Absorbance
Substrate depleted
In high-activity enzyme measurements, the substrate may be depleted quickly and the reaction
curve will appear obviously nonlinear (as a smooth curve). If the measurement is performed
based on the general procedure, the system will flag the test result with "NLN" (no linearity
interval), reminding the user to rerun the test after diluting the sample. This will more or less
bring troubles to the user.
Extending enzyme linearity range:
Suppose the reaction start time is t1 and the reaction time is tL-tM, then t1-tL is the lag time.
If the number (N) of valid measuring points within tL-tM is less than 2 and too few to calculate
the response, the sample response can be obtained by extending the enzyme linearity range.
Calculation of ⊿Amax:
The linearity range t1-tL' without substrate depletion is found within the lag time t1-tL.
If the number (N) of valid measuring points within tL-tM is less than 2, the system will not
calculate the response but flag the test result with "ENC" (no calculation interval) and give an
alarm;
or the system calculates the reaction rate ⊿A=60*(Ai+1-Ai)/(ti+1-ti), i=1, 2…L' with the lag
time t1-tL'. The maximum ⊿A is taken as the response of the sample. Therefore, the enzyme
linearity range is extended via the lag time. The results calculated by extending the enzyme
linearity range will be flagged with "EXP" and "NLN".
Calculation formula: C K ( R R 0 )
C 2 C1 C1
The formula contains two factors, K and R0, where K , and R0 R1 .
R2 R1 K
The calibration math model requires two calibrators. C1 and C2 are the concentrations of
calibrator 1 and 2; R1 and R2 are the responses of calibrator 1 and 2.
Multi-point linear calibration
Calculation formula:
C K (R R 0 )
The formula contains two factors, K and R0. The calibration math model requires n(n≥3)
calibrators. Ci is the concentration of calibrator i. Ri is the response of calibrator i. K and R 0 can
be calculated with the least square method:
n n n
CiRi ( Ci)( Ri ) / n
K i 1
n
i 1
n
i 1
Ri
i 1
2
( Ri ) 2 / n
i 1
n
( Ci) / n
R0 ( Ri ) / n i 1
i 1 K
1
Calculation formula: R R0 K
1 exp[ (a b ln C )]
The formula contains four factors, which are R0, K, a and b.
The calibration math model requires at least four calibrators. The four factors can be calculated
with the L-M method.
This calibration type is applied to the chemistries which have a calibration curve with the
response reversely proportional to the concentration.
Logit–Log 5P
1
R R0 K
Calculation formula: 1 exp[ (a b ln C cC )]
The formula contains five factors, which are R0, K, a, b and c. The calibration math model
requires at least five calibrators, and calculates the five factors with the L-M method.
This math model has the same application with the Logit-Log 4P except for a higher fitting.
Exponential 5P
R R0 R R0 2 R R0 3
ln C a b( ) c( ) d( )
Calculation formula: 100 100 100
The formula contains five factors, which are R0, a, b, c and d. The calibration math model
requires at least five calibrators. The response (R) of the first calibrator (with internal
converting concentration of 0) is R0, which is given.
R R0
x
Suppose, y ln C and 100 .
Then, y a bx cx dx
2 3
can be calculated with the least square method for polynomial
expressions.
Parabola
R aC 2 bC R0
Calculation formula:
The formula contains three factors, which are a, b and R0. The calibration math model requires
at least three calibrators. The three factors can be calculated with the least square method.
Spline
R R0i ai (C Ci ) bi (C Ci ) 2 ci (C Ci ) 3
Calculation formula:
The calibration math model requires 2-9 calibrators. Suppose the number of calibrators is n,
R a b ci
then the calculation formula contains 4(n-1) factors, which are 0 i , i , i , and . Due to the
subsection fitting, this math model has be best fit curves than other math models.
Logit–Log 3P
1
R R0 K
Calculation formula: 1 aC
The calibration math model requires 3-10 calibrators. Use L-M method to calculate R0, K and a.
Line
Calculation formula:
C K i ( R R 0i )
The calibration math model requires 2-10 calibrators. Suppose the number of calibrators is n,
12.4 QC evaluation
The system provides the Westgard rules for evaluating QC results of the chemistries, and give
alarms and flags when the obtained QC results are beyond the reference range. Since every
chemistry may have one or more control samples, the QC results can be evaluated with different
rules accordingly. Those controls that are not included in any lots will be evaluated as single
controls.
(1) An asterisk "*" indicates a random error, which requires no special action but must not be
ignored.
(2) A "#" symbol indicates a systematic error, which requires special consideration.
Control data
No
>2S In-control
Yes No
Yes
12S Warning
No
No No No
13S 22S 41S 10X
Yes Yes Yes Yes
Out of control
(1) An asterisk "*" indicates a random error, which requires no special action but must not be
ignored.
(2) A "#" symbol indicates a systematic error, which requires special consideration.
The random errors in two-control evaluation correspond to those in single-control evaluation as
follows:
22SA\22SW corresponding to 22s.
41SA\41SW corresponding to 41s.
10XA\10XW corresponding to 10x.
The procedure of two-control evaluation is shown in the figure below:
Figure 12.6 Two-control evaluation workflow
Measured values of X
and Y controls
No
12S In control
Yes
No
No No No No No No No
13S 22SA R4S 22SW 41SA 41SW 10XA 10XW
Yes Yes Yes Yes Yes Yes Yes Yes
When Cumulative Sum QC is performed, first the difference between the QC test result and the
threshold (k) and the accumulative sum are calculated and then the software judges whether the
QC test result is out of control by calculating if the accumulative sum exceeds the control
limit(h).
Concentration C
In the reaction of antigen and antibody, the amount of generated insoluble compound is closely
related to the proportion of antigen and antibody. The maximum amount of compound will be
generated at a proper proportion of antigen and antibody, at this point least light is passed and
the greatest absorbance is obtained. For other proportions, the amount of insoluble compound
will decrease with more light passed and lower absorbance calculated. Therefore, samples with
quite different concentrations may generate the equivalent amount of insoluble
antigen/antibody compound, and can have the same test results without a Prozone check. The
Prozone check, therefore, is necessary for antigen-antibody reactions.
V1 is the threshold of │Aq2- Aq1│. Y means the condition │Aq2- Aq1│< V1 is satisfied; N
means that│Aq2- Aq1│≥V1 is satisfied.
V2 is the threshold of │Aq4- Aq3│. As for Y/N, Y means the condition │Aq4- Aq3│< V2 is
satisfied; N means that│Aq4- Aq3│≥ V2 is satisfied.
PC1 and PC2 are the low limit and high limit of the threshold of formula 1. Y means the value of
PC is within the range of formula 1; N means out of the range of formula 1.
V3 is the threshold of formula 2. Y means the value of formula 2 is within the defined range; N
means out of the defined range.
Select A/ O from the drop down list. The default is A. A means there is And logic between
formula 1 and formula 2; O means there is Or logic between formula 1 and formula 2.
For single reagent chemistry:-13≤q1, q2, q3, q4, q5, q6≤M. M is the end point of the defined
reaction time.
For double reagent chemistry:-34≤q1, q2, q3, q4, q5, q6≤M. M is the end point of the defined
reaction time.
q1≠q2 and q3≠q4 but q5=q6 is allowed.
The input range for PC1, PC2, V1, V2 and V3 is [-35,000, 35,000]. And PC1<PC2.
Formula 1:
A q 4 A q3
q 4 q3
PC1 PC 2
A q 2 A q1
q 2 q1
Formula 2:
When q5≠q6,Aq6- Aq5≥V3.
When q5=q6, Aq5≥V3.
Where,
Aq5 is the absorbance of q5 measuring point.Aq6 is the absorbance of q6 measuring point.
Single point calibration: 80 μL calibrator is dispensed into the ISE module to perform single
point calibration.
Electronic Interface
Description Serial interface
Interface Purpose The host receives and executes instructions from the PC
through this interface, and returns the execution results to
the PC through this interface.
Glossary
Absorbance
The difference between the amount of light entering a solution (incident light) and the amount
of light passing through the solution (transmitted light) without being absorbed, to determine
the concentration of the substance in the solution.
Analyzing unit
The analyzing unit, the analyzer, determines various clinical chemistries in samples and displays
the test results. It consists of the sample/reagent handling system, reaction system, cuvette
wash station, photometric system, and mixer assembly Auto rerun
When a result is beyond the defined range or satisfies the defined conditions, the chemistry will
be run again.
Bar code reader
It scans the bar code label on sample tube and reagent bottle to identify the sample and reagent.
Batch program
Batch program is to program a group of samples with identical programming information, with
the exception of the sample ID.
Blank time
Blank time refers to the period between dispensing of the second reactant (reagent or sample)
in reversed order and of the last reactant (reagent or sample).
Bottle type
Volume of the reagent bottle.
Calibration curve
A calibration curve reflects the mathematical relation between calibrator concentration and
response. It is drawn based on the obtained response and the multiple values between the
minimum and maximum concentrations of the calibrator.
Calibration factor
Calibration factor is obtained based on the equation of calibrator concentration (known) and
response (calibration math model).
Calibration math model
Calibration math model is used to calculate calibration factors and create calibration curves. It
includes single-point K factor, two-point linear, multi-point linear, Logit-Log4P, Logit-Log5P,
Exponential5P, Polynomial5P, Parabola and Spline.
Calibration trend
Calibration trend summarizes a chemistry’s calibrations during a period of time and reflect the
trends of the calibrations.
Carryover
Fixed-time
In fixed-time measurements, namely, rate measurements, the reaction velocity (v) is directly
proportional to the substrate concentration [S] within a specific period, that is, v=k[S].
Flag
Flag is a manufacturer-defined symbol, which appears on patient reports or result list when a
result is beyond the user-defined reference range or exceeds the defined limits.
High-concentration waste
High-concentration waste is produced during the phase 1 cuvette wash. It can be drained to the
provided high-concentration waste tank and then disposed of according to your local or national
regulations.
History results
Stored results are those programmed and analyzed before the current day.
Increased
Increased indicates the sample volume required for analysis and can be defined on the
Define/Edit Chemistries window.
Initialization
Initialization is a series of operations automatically performed by the system during the startup
procedure. It includes parameters check, reset, testing, cleaning and priming.
Inventory check
Used to check the remaining volume of the biochemistry reagents, sample probe wash solution
and reagent probe wash solution and refresh the tests left and wash solution volume on the
Reagent/Calibration screen.
ISE
ISE is the abbreviation of Ion Selective Electrode. It consists of the ISE module, pump module and
reagent module, and is used to measure the concentration of Na, K and Cl ions in serum, plasma
and diluted urine.
K factor
C K ( R R0 )
K factor is manually input for single-point linear calibration formula and
used to calculate results.
Lamp
Lamp is located on the photometer assembly and used to measure the absorbance of mixture in
a reaction cuvette. It should be replaced regularly.
Linearity
Degree of linearity for a reaction curve or calibration curve. Reaction curve linearity is available
in fixed-time measurements while calibration curve linearity specifies the allowable
concentration range for result calculation.
LIS
LIS stands for Laboratory Information System. It is a host computer and communicates with
chemistry analyzers through the internet interface.
L-J chart
A Levey-Jennings (L-J) chart, drawn based on the QC date (X) and test results (Y), shows the QC
result trend of a chemistry during the specified period. The graphical trends of up to 3 controls
can be displayed on one L-J chart and distinguished with different colors.
Lot number
Lot number is assigned to controls, calibrators or wash solutions of the same lot for identifying
manufacture date, quality, expiration date and other related information.
Low-concentration waste
Low-concentration waste is produced during phase 2-4 cuvette wash and probe/mixer cleaning.
It can be drained to the provided low-concentration waste tank or the sewer of your laboratory.
Mask/Unmask chemistries
Used when a chemistry needs to be disabled temporarily due to abnormal result or reagent
exhaustion. The masked chemistry will have a symbol appearing on its upper-left corner,
and will still be displayed on the Sample, Quality Control and Reagent/Calibration screens
but not run for sample analysis. Masked chemistries cannot be requested until they are
unmasked.
Mixer
The system provides one mixer for stirring the mixture inside a reaction cuvette when sample
and R2 are respectively dispensed.
Multi-sample report
Containing the results of multiple samples, and can be printed out on the Current Results and
History Results screens.
Off-line dilution
Prior to analysis, samples are diluted manually based on specific ratio.
Offset
Offset is a value added or subtracted to compensate a result. It is often used along with the slope
in the equation y=kx+b, in which k is the slope and b is the offset.
Off-system chemistry
All the chemistries that are not run by the analyzer are referred to as the off-system
chemistries.
Online help
Online help provides you with help information about the screens. If you do not understand a
parameter or an operation on a screen, you can go to the online help for relevant information.
Access the online help from the following screens:
Select the icon on the upper right corner to display the help topic related to the current
screen.
Select the button in front of each maintenance instruction or item to display the
relevant operating instructions.
Select the button in front of each error log to display the corresponding topic.
Click the button on a warning message window to display the corresponding
descriptions and solutions.
Press the shortcut combination key Alt+F1 to display the topics related to the current screen
or window.
Open-reagent chemistry
Open-reagent chemistry, an opposite of the closed-reagent chemistry can be measured by using
the reagents provided by other manufacturers. It can be user-defined, edited and deleted.
Operation unit
The operation unit, a computer configured with the operating software, controls the analyzing
unit to finish tests and produce test results.
Output unit
A printer used to print out test results and other data.
Panel
Consists of a couple of chemistries combined together for certain clinical purposes, such as liver
function, kidney function, etc. Panels can help fast programming of samples.
Patient demographics
Patient demographics contain information related to the patient and sample, such as patient
name, age, gender, collection date/time, etc.
Physiological saline
0.9% sodium chloride solution, used for reagent blank and sample dilution.
Predilution
Prior to analysis, samples are diluted automatically based on the defined dilution factor.
Primary wavelength
The primary wavelength is chosen based on the light absorption features of the reactant and
used to measure the absorbed light intensity.
Prime
Prime is an action to replace the reagents in tubing of the ISE module. A prime is required to
replace the reagents in tubing with new ones during the startup procedure or when a reagent is
changed.
Print name
Print name appears on a patient report representing a chemistry, and if left blank, will be
replaced by the short name of the chemistry.
Probe
The probe aspirates the specified amount of sample and reagent and then dispenses it into a
cuvette for reaction and analysis.
Probe ash solution
CD80 alkaline concentrated wash solution. It is placed in position D of the sample/reagent
carousel, and used for special cleaning the probe, in order to prevent cross contamination.
Prozone check
Prozone check is intended to checking samples with quite different concentrations, which may
generate the equivalent amount of insoluble antigen/antibody compound and can have the
same test results. The rate check method is supported.
Pull-down list
A control of the software screen or window. Select the down-triangle button on the right of a
pull-down list to show multiple options.
QC panel
Used for analysis of control samples.
QC rule
A set of rules to evaluate if the QC results are under control and the analyzing system is stable.
Examples of QC rule are 1-2s, 1-3s, etc.
QC summary
Contains the mean values and standard deviations of controls analyzed within the specified
period, as well as the set mean and SD value. The obtained results are compared with the set
values to judge if the system is working normally.
Qualitative analysis
Qualitative analysis is used to analyze every sample for the detection of lipemia, hemolysis and
icterus and calculate the numeric values of the index. If the volume of the interferents contained
in a sample is beyond the set range, a flag will be added to the patient report.
Random error
An alarm of quality control monitoring. A random error may occur when the lowest and highest
values of QC results respectively exceed -2SD/-3SD and +2SD/+3SD.
Reaction carousel
Reaction carousel is a turntable, and used to hold reaction cuvettes and transmit each of them to
the photometric position for signal detecting and absorbance calculation.
Reaction curve
A reaction curve reflects the relationship of the absorbance measured at the primary
wavelength, secondary wavelength and primary-secondary wavelength. It is drawn based on the
absorbance of the sample-reagent mixture measured within the reaction period. The system
provides 4 types of reaction curves: calibration reaction curve, QC reaction curve, sample blank
reaction curve, and sample reaction curve.
Reaction cuvette
Reaction cuvette is a carrier in which reagents and samples react with each other and then
carried to the photoelectric position for signal detecting and response calculation.
Reaction direction
Reaction direction refers to the change trend of absorbance during the reaction process. It
includes positive and negative.
Reaction time
For endpoint analysis, the reaction time refers to the time span from the start point of the
reaction to the end point; for fixed-time and Kinetic analysis, it refers to the period from
reaction equilibrium to the end of monitoring.
Reagent blank
In the reagent blank test, the reagents react with the physiological saline and the blank
absorbance is calculated to correct the calibration factors. Only the reagents that are in
Calibrated status can be requested for reagent blank.
Reagent carryover
Cross contamination between the reagent probe and the mixers. When the number of tests
between the contaminating chemistry and the contaminated is less than or equal to the defined
number (N), and no concentrated wash is inserted between the two chemistries, it indicates
that the reagents underlie the risk of carryover.
Reagent inventory alarm limit
Alarm limit of reagents and wash solutions. When the reagent inventory is lower than the alarm
limits during or before the analysis, the system will give an alarm and display the reagent or
wash solution name in yellow on the Reagent/Calibration screen.
Reference range
Reference range is a user-defined range consisting of low limit and high limit. When a result is
beyond the reference range, a flag will appear near the result.
Release
Used to clear the specified sample position or all positions on the current sample carousel.
When a sample is released, its results and programming information can be still recalled. The
released position can be used for programming of new samples.
Replicates
Number of times to run a test, to ensure accurate results.
Result statistics
Result statistics option can summarize the total chemistries and the distribution trend of its
results and provide the test data and graph.
Sample blank
Sample blank is similar to sample analysis except for use of equivalent amount of physiological
saline Sample blank is used for removal of non-chromogenesis reaction, such as influence of
sample interference (Hemolysis, icterus and lipemia) on absorbance readings.
Sample/Reagent carousel
The sample/reagent carousel is located on left side of the analyzer panel. It holds sample tubes
and reagent bottles and carries each of them to the aspirate position for aspirating.
Sample comments
Remarks for some special samples, such as, ** sample has hemolysis; ** sample needs to be
analyzed immediately, etc.
Sample log
Contains the controls and patient samples that are not complete within the recent 24 hours due
to certain reasons. Based on the sample log you are allowed to rerun the samples or take other
actions for the controls and samples.
Sample panel
Used for analysis of patient samples.
Sample type
Type of sample. The sample type options include serum, plasma, urine, CSF and other.
Page
Page is a part of the software interface. It is rectangular and contains various controls, such as
edit box, function button, etc.
Secondary wavelength
The secondary wavelength is used to remove the interference in primary wavelength values and
eliminate the influence of noise, such as light flash and drift, and scratches on cuvettes, etc. It
cannot be the same as the primary wavelength.
Serial number
Sequence number of the reagent bottle.
Slope
Multiplied with the test result to make it consistent with that obtained on other instruments. It
is often used along with the offset in the equation y=kx+b, in which k is the slope and b is the
offset.
Special calculation
Special calculation is derived from calculation of certain chemistries and has specific clinical
purposes, such as A/G, TBil-DBil, etc.
Special wash
Special wash is to clean the probe, mixer and reaction cuvettes by using the probe wash solution,
with the aim of eliminating carryover and preventing waste from leaving in the waste tubes.
Standard deviation (SD)
Standard deviation is the mean of deviations from the mean value. It is an index to judge the
measurement accuracy under specific conditions. In this manual, SD refers to the standard
deviation of control concentration.
Standby
Standby is one of the system statuses. When the system status is Standby, it indicates that all
tests are finished and all actions of the system have stopped.
STAT
STAT means emergent, including common STAT and quick STAT program. STAT sample program
allows emergent samples to be programmed and analyzed with high priority. Common STAT
program is used in daytime to run emergent samples with higher priority than routine samples.
Quick STAT program is mainly used in nighttime and weekends to program emergent samples
quickly with higher priority than routine and common STAT samples.
Symbology
Symbology is a set of rules for encoding and decoding information contained in a bar code label.
The system provides a couple of symbologies, such as Codabar, ITF, code128, code39, UPC/EAN,
and Code93.
Systematic error
An alarm of quality control monitoring. A systematic error may occur when both the lowest
value and highest value of a QC result are on the same side.
Test statistics
On the Tests screen, you can view test requests and reagent application for each chemistry
during a period, and sample requests and the quantity of its chemistries.
Transmit
Transmit is an action sending specified sample results or QC results to the LIS host.
Twin chemistries
Twin chemistries are run with the same reagents and calculated through the same test. For two
twin chemistries, the sample volume, volume of shared reagent, calibration replicates, and auto
calibration conditions should be the same. When either of the two chemistries is requested for
calibration, quality control or sample analysis, the other chemistry will be automatically
requested, and finally results of both chemistries will be calculated.
Twin-Plot chart
A twin-plot chart, drawn based on the results of control X and control Y in the same run, is used
to detect systematic errors and random errors. It shows the recent 10 QC results of a chemistry
and excludes those that have been deleted.
Two-control evaluation
In two-control evaluation, two results are obtained: Xn and Yn, which are used to define a point
on the Twin-plot chart. In this way, a complete twin-plot chart is drawn based on all the QC
results and used for detecting systematic errors and random errors.
Unpositioned samples
Samples without positions assigned or with positions not assigned successfully, including those:
downloaded from the LIS host and not positioned yet.
that are in Incomplete status when their positions are assigned for new samples.
that are incomplete when their positions are released.
Wash solution
All wash solution used by the instrument is CD80 alkaline concentrated wash solution. It is used
to clean the probe, mixer and reaction cuvettes.
Westgard rule
Westgard rule is used for monitoring of quality control. In the Westgard rule, single rules such as
12S, 13S, 22S and 41S are combined to evaluate results of single or multiple controls.
Index
High-concentration waste, 10-15
A Host, 11-53
host communication, 11-53
absorbance, 12-2
Absorbance, 11-45
auto serum index, 6-17 I
Installation requirements, 1-11
B ISE module, 8-5, 10-21
C L
calibration curve, 12-8, 12-9 Light source, 11-54, 11-55
calibration math model, 12-9 Linear, 12-8
Calibration math model, 12-9 linearity range, 12-7
Carryover, 7-29 Linearity range, 12-7
CAUTION, 8, 6-2 LIS, 6-17, 6-28, 7-27, 8-21, 11-53, 11-54, 9
Cleaning the dust screen, 10-3 Lot number, 8-20
Cleaning the filter core, 10-3 Low limit, 8-5
Cleaning the wash wells, 10-3
clearing samples, 6-17
Control, 5-12
M
Critical range, 6-11 Measuring point, 12-6
Current results, 6-27 Mouse, 1-36
D P
default panel, 6-17, 7-27 Panels, 7-26, 7-27
Default panel, 6-17, 7-27 physiological saline, 11-52, 11-57
dust screens, 10-3 Primary tube, 1-20
Dust screens, 10-3 print name, 6-16
Print name, 6-16
E
R
Endpoint, 12-2
endpoint measurements, 12-2 random error, 6
Endpoint measurements, 12-2 Random error, 6
Reaction carousel, 11-30
Reaction curve, 12-7, 12-13
F reagent blank, 4-12
reagent volume, 6-16, 11-52
Filter core, 10-3 Reagent volume, 6-16
fixed-time measurements, 12-4 response, 12-8, 12-9
Fixed-time measurements, 12-4
H S
Sample carousel, 11-31
High limit, 8-5 Sample carousel outer ring, 11-31
high-concentration waste, 10-15 sample list, 6-37
Bibliography
1. Burtis C.A., Ashwood E.R. and Bruns D.E. Tietz Fundamentals of Clinical Chemistry. 6th Ed.
Saunders/Elseviser, St. Louis, Missouri, 2008, 63-262.
2. Burtis C.A., Ashwood E.R. and Bruns D.E. Tietz Textbook of Clinical Chemistry and Molecular
Diagnostics. 5th Ed., Saunders/Elseviser, St. Louis, Missouri, 2012, 7-59.
3. Gauglitz G. and Vo-Dinh T. Handbook of Spectroscopy. WILEY-VCH Verlag GmbH & Co. KGaA,
Weinheim, 2003, 37-162.
4. Gottschalk P.G. and Dunn J.R. The five-parameter logistic: A characterization and comparison
with the four-parameter logistic. Analytical Biochemistry, 2005; 343: 54–65.
5. Levey S. and Jennings E.R. The use of control charts in the clinical laboratory. Am J Clin Pathol,
1950; 20: 1059-1066.
6. Madsen K., Nielsen H.B. and Tingleff O. Methods for Non-Linear Least Squares Problems.
2nd Ed., Informatics and Mathematical Modelling, Technical University of Denmark, Lyngby,
Denmark, 2004.
7. Westgard J.O. and Barry P.L. Cost-Effective Quality Control: Managing the quality and
productivity of analytical processes. AACC Press, Washington DC, 1986.
8. Westgard J.O., Barry P.L., Hunt M.R. and Groth T. A multi-rule Shewhart chart for quality
control in clinical chemistry. Clin Chem 1981; 27: 493-501.