Ka01164pen 0517
Ka01164pen 0517
Ka01164pen 0517
17
71353700
Products Solutions Services
Serial number
2. www.endress.com/deviceviewer Endress+Hauser
Operations App
3.
A0023555
2 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Table of contents
Table of contents
1 Document information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Document function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Terms and abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5 Turn down calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.1 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.2 Influence of the installation position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.3 Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 Mounting the profiled seal for the universal process adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.5 Mounting instructions for oxygen applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.6 Post-installation check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.1 Connecting the measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.2 Connection conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.3 Connection data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.4 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7 Operation options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.1 Plug-on display PHX20 (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Endress+Hauser 3
Document information Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
1 Document information
1.1 Document function
The Brief Operating Instructions contain all the essential information from incoming
acceptance to initial commissioning.
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in personal
injury.
Symbol Meaning
Open-ended wrench
A0011222
4 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Document information
Permitted Tip
Procedures, processes or actions that Indicates additional information.
are permitted.
A Reference to page
Symbol Meaning
1. , 2. , 3. … Series of steps
A, B, C, ... Views
1.3 Documentation
The document types listed are available:
In the Download Area of the Endress+Hauser Internet site: www.endress.com →
Download
The nameplate provides information on the Safety Instructions (XA) that are relevant for
the device.
6 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Document information
p
0
LRL LRV URV URL MWP OPL
A0029505
1 OPL The OPL (over pressure limit = sensor overload limit) for the measuring device depends on
the lowest-rated element, with regard to pressure, of the selected components, i.e. the
process connection has to be taken into consideration in addition to the measuring cell.
Also observe pressure-temperature dependency. For the relevant standards and additional
notes, see the "Pressure specifications" section of the Operating Instructions .
The OPL may only be applied for a limited period of time.
2 MWP The MWP (maximum working pressure) for the sensors depends on the lowest-rated
element, with regard to pressure, of the selected components, i.e. the process connection
has to be taken into consideration in addition to the measuring cell. Also observe
pressure-temperature dependency. For the relevant standards and additional notes, see
the "Pressure specifications" section of the Operating Instructions .
The MWP may be applied at the device for an unlimited period.
The MWP can also be found on the nameplate.
Endress+Hauser 7
Document information Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
p - Pressure
1=2 3
LRL LRV URV URL
A0029545
1 Calibrated/adjusted span
2 Zero point-based span
3 URL sensor
Example
URL
TD =
|URV - LRV|
8 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Basic safety instructions
3 Product description
See Operating Instructions.
10 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Incoming acceptance and product identification
4.2.1 Nameplate
1
2 Cerabar
Made in Germany,
D-79689 Maulburg
3 Ord. cd.:
4 Ser. no.:
5 Ext. ord. cd.:
Date:
TAG:
A0024456
1 Manufacturer's address
2 Device name
3 Order number
4 Serial number
5 Extended order number
Endress+Hauser 11
Installation Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
5 Installation
5.1 Installation conditions
• No moisture may enter the housing when installing or operating the device, or when
establishing the electrical connection.
• For metal M12 plugs: Do not remove the protection cap (only for IP69 and Ex ec version) of
the M12 plug terminal until shortly before the electrical connection.
• Do not clean or touch process isolating diaphragms with hard and/or pointed objects.
• Do not remove the protector on the process isolating diaphragm until just before
installation.
• Always firmly tighten the cable entry.
• Point the cable and connector downwards where possible to prevent moisture from entering
(e.g. rain or condensation water).
• Protect the housing from impact
• The following instruction applies for devices with a gauge pressure sensor and M12 plug or
valve plug:
NOTICE
If a heated device is cooled during the cleaning process (e.g. by cold water), a vacuum
develops for a short time, whereby moisture can penetrate the sensor through the
pressure compensation element (1).
Device could be destroyed!
‣ In this case, mount the device with the pressure compensation element (1) pointing
diagonally downwards where possible or to the side.
12 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Installation
1 1 1
A0022252
A B C
A0024708
Type Process isolating diaphragm Process isolating diaphragm Process isolating diaphragm
axis is horizontal (A) pointing upwards (B) pointing downwards (C)
PMP11 Calibration position, no effect Up to +4 mbar (+0.058 psi) Up to –4 mbar (–0.058 psi)
PMP21
PMP23
PMC11, PMC21 Calibration position, no effect Up to +0.3 mbar (+0.0044 psi) Up to –0.3 mbar (–0.0044 psi)
< 1 bar (15 psi)
PMC11, PMC21 Calibration position, no effect Up to +3 mbar (+0.0435 psi) Up to –3 mbar (–0.0435 psi)
≥1 bar (15 psi)
Endress+Hauser 13
Installation Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
5.4 Mounting the profiled seal for the universal process adapter
For details on mounting, see KA00096F/00/A3.
Are the measuring point identification and labeling correct (visual inspection)?
Does the pressure compensation element point downwards at an angle or to the side?
To prevent the penetration of moisture: are the connecting cables/plugs pointing downwards?
14 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Electrical connection
6 Electrical connection
6.1 Connecting the measuring unit
6.1.1 Terminal assignment
LWARNING
Risk of injury from the uncontrolled activation of processes!
‣ Switch off the supply voltage before connecting the device.
‣ Make sure that downstream processes are not started unintentionally.
LWARNING
Supply voltage might be connected!
Risk of explosion!
‣ Make sure that no supply voltage is applied when connecting.
‣ Switch off the supply voltage before connecting the device.
LWARNING
Limitation of electrical safety due to incorrect connection!
‣ In accordance with IEC/EN61010 a separate circuit breaker must be provided for the
device .
‣ The device must be operated with a 500 mA fine-wire fuse (slow-blow).
‣ When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or
Installation or Control Drawings.
‣ All explosion protection data are given in separate documentation which is available upon
request. The Ex documentation is supplied as standard with all devices approved for use in
explosion hazardous areas.
‣ Protective circuits against reverse polarity are integrated.
Connect the device in the following order:
1. Check whether the supply voltage matches the supply voltage indicated on the
nameplate.
2. Connect device in accordance with the following diagram.
Switch on supply voltage.
For devices with a cable connection: do not close reference air hose (see (a) in the following
drawings)! Protect reference air hose against penetration by water/condensate.
Endress+Hauser 15
Electrical connection Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
4 to 20 mA output
Device M12 plug Valve plug Cable
1 brown = L+
2 blue = L-
3 green/yellow = ground connection
(a) reference air hose
0 to 10 V output
Device M12 plug Valve plug Cable
PMC11 0.5A 1 -
PMP11 L+
2 1 3
4
3
2 L–
L–
A0017576
L+
0.5A
A0022822
4 to 20 mA output PMC11 10 to 30 V DC
PMP11
PMC21
PMP21
PMP23
0 to 10 V output PMC11 12 to 30 V DC
PMP11
16 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Electrical connection
2 PMC11 ≤ 26 mA > 21 mA
PMP11
PMC21
PMP21
PMP23
3 PMC11 < 12 mA 11 V
PMP11
R L max
[W ]
1068
10 20 30 UB
[V]
1
A0029452
1 Power supply 10 to 30 V DC
2 RLmax maximum load resistance
UB Supply voltage
Endress+Hauser 17
Operation options Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
Are all the cable glands installed, firmly tightened and leak-tight?
7 Operation options
7.1 Plug-on display PHX20 (optional)
No display or other operation facility is required to operate the device. However, devices with
a valve plug can be fitted with the optional local display PHX20.
A 1-line liquid crystal display (LCD) is used. The local display shows measured values, fault
messages and information messages. The device display can be turned in 90° steps. Depending
on the orientation of the device, it is therefore easy to read the measured values.
7.1.2 Installation
1 2 3 4
A0022208
1. Place seals between the sensor/plug-on display and the plug-on display/connector.
2. Fit the plug-on display (2) between the connector (3) and connector socket (1) of the
sensor.
3. Replace the securing screw (4) with the longer screw included in the delivery.
18 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Operation options
4. An adhesive label (included in the delivery) with information on the technical unit can
be fixed beneath the LED display.
Pin assignment
LWARNING
Is the supply voltage switched off?
Risk of electric shock!
‣ Switch off the supply voltage before connecting the device.
• PIN 1: L+ (supply voltage UB)
• PIN 2: L- (0 V)
• PIN 3: Not assigned
Supply voltage
The supply voltage (mostly 24 V DC) must be greater than the sum of the voltage drop Us at
the sensor, the voltage drop of 5 V at the display and other voltage loss Ua (such as additional
evaluation and power loss).
The following therefore applies: Ub = Us + 5 V + Ua
Post-connection check
If supply voltage is present, is the device ready for operation and do values appear on the display module?
7.1.5 Commissioning
NOTICE
Risk of injury from the uncontrolled activation of processes!
‣ Ensure that no uncontrolled processes are activated in the system.
Configuring the menu items
To configure the menu items, loosen the 4 Phillips screws (1) on the display and remove the
cover.
Endress+Hauser 19
Operation options Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
1 1
A B
1 1
A0022209
20 Endress+Hauser
*71353700*
71353700