0% found this document useful (0 votes)
50 views4 pages

5000F (C20) - Specification

This document provides specifications for the STAR5000F(C20) obstetric monitor made by Shenzhen Comen Medical Instrument Co., Ltd. It lists configurations that include modules for fetal and maternal monitoring. Key features summarized are: - Monitor displays fetal heart rate, TOCO, fetal movement, and other parameters. - Includes options for 3-lead ECG, NIBP, temperature, respiration, SpO2, and EtCO2 monitoring. - Provides data storage, alarms, printing, and can connect to central monitoring systems. - Ultrasound module measures FHR with 1MHz pulses and autocorrelation.

Uploaded by

Ruben Soliz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
50 views4 pages

5000F (C20) - Specification

This document provides specifications for the STAR5000F(C20) obstetric monitor made by Shenzhen Comen Medical Instrument Co., Ltd. It lists configurations that include modules for fetal and maternal monitoring. Key features summarized are: - Monitor displays fetal heart rate, TOCO, fetal movement, and other parameters. - Includes options for 3-lead ECG, NIBP, temperature, respiration, SpO2, and EtCO2 monitoring. - Provides data storage, alarms, printing, and can connect to central monitoring systems. - Ultrasound module measures FHR with 1MHz pulses and autocorrelation.

Uploaded by

Ruben Soliz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Specification: STAR5000F(C20)

SHENZHEN COMEN MEDICAL INSTRUMENT CO.,LTD


Floor7,Block5, 4th Industrial Park of Nanyou, Nanshan District,
Shenzhen City 518052, China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: [email protected]
Web: www.szcomen.com
Obstetric monitor
STAR5000F (C20)

Standard Configuration 1: Fetal and Maternal monitor: C31 module (3-lead


Fetal monitor : FHR, TOCO, FM, AFM, printer ECG,NIBP,TEMP,RESP,Masimo SPO2, EtCO2 interface), FHR,
Standard Configuration 2: TOCO, FM, AFM, printer
Fetal and Maternal monitor: C30 module (4.3inch led Standard Configuration 5:
screen,3-lead ECG,NIBP,TEMP,RESP,Masimo SPO2, EtCO2 Fetal and Maternal monitor: C31 module (3-lead
interface), FHR, TOCO, FM, AFM, printer ECG,NIBP,TEMP,RESP,Comen SPO2, EtCO2 interface), FHR,
Standard Configuration 3: TOCO, FM, AFM, printer
Fetal and Maternal monitor: C30 module (4.3inch led Optional Configuration:
screen,3-lead ECG,NIBP,TEMP,RESP,Comen SPO2, EtCO2 Twins monitoring, Fetal stimulator, ETCO2(ETCO2 is just
interface), FHR, TOCO, FM, AFM, printer available in Fetal and maternal monitor )
Standard Configuration 4: Safety Standards
ISO 13485:2003 approved, CE marking according to
MDD93/42/EEC
Physical Characteristics rate:
Product Size: 400 ㎜×300mm×95 mm Ultrasound: 1.0MHz ±10%
Net Weight: 5kg FHR Range: 50bpm- 210bpm
Display: 12.1’’ Color TFT LCD touch screen Resolution: 1bpm
Resolution: 800X600 Accuracy: ± 2 bpm
Method: folded TOCO
Operation Environment TOCO range: 0-100 Relative(%)
working 0-40℃ Resolution: ±1%
Temperature: Zero Mode: Manual
Humidity: 15-85% Non-linear error: ≤±8%
Power Supply AC100-250V, 50/60Hz Fetal Movement(FM)
Battery Type: Rechargeable lithium-ion battery Mode: Auto; manual
Battery Capacity: 14.8V DC 4400mAh Technique: Pulsed Doppler Ultrasound
Battery Recharging Maximum 6.5 hours for charging; Range: 0-100(%)
Time: Resolution: 1%
Battery working 4 hours for continuous working Fetal Stimulator
period: Method: Vibration
Interface: USB(data input or output, upgrade) Mode: Manual(3 times), continuing
AC power supply socket Voltage: DC 3V, (AA battery)
Parameter cable interface Respiration
RS232 port Method: impedance method
RJ45 port RR measurement Adult: 0-120bpm
Data: Input/output(patient’s case) range:
Network: Central Monitor System Pediatric/Neonate: 0 -150bpm
Date storage Resolution: ±1 rpm
patient case 3000 groups Accuracy: 6-150rpm:±1rpm or 1%, whichever is
Alarm event recall: 300 groups greater
NIBP Recall: 2000 groups 0-6rpm: unspecified
Trend Graph: 120 hours RESP Apnea Alarm: 10s-40s
Trend Table: 120 hours Alarm: Audible and visual alarm; alarm
freezed wave recall 240s for Adult, 60 hours for fetal events recallable
Recorder Sweep Speed: 6.25,12.5,25,50 mm/s
Record width: 150mm Gain Selection: X0.25, X0.5, X1, X2
Recording parper: 150mm*90mm or ECG
150mm*100mm , Z-fold, thermo Lead Type Cardio TecTM 12-leads ECG Analysis, 5
sensitive Lead and 3 lead selectable
Fast print speed: 25mm/s(stored traces) Lead selection: 3-lead: I; II; III
Paper Advance 1/2/3cm/min selectable 5-lead: I; II; III; aVR; aVL
Speed: 12-Lead I; II; III; aVR; aVL;aVF; V1-V6.
Record Data,time, TOCO type,paper Gain Selection: X 0.125, X0.25, X0.5, X1, X2, X4,auto
information : speed,FHR type,bed no.,etc. Sweep Speed: 6.25,12.5, 25, 50mm/s
Printer: Built-in thermal recorder; external Bandwidth : MON Mode:0.5 Hz~40 Hz
laser printer DIA Mode: 0.05Hz~130 Hz
Ultrasound OPE Mode:1 Hz~20 Hz
Technique: Ultrasound Pluse Doppler with ST Mode:0.2Hz~40Hz
autocorrelation
Pulse repetition 1 MHz
Defib. Protection: Withstand 4000VAC/50Hz voltage in Comen SPO2
isolation ,Again electrosurgical Measurement & 1~100%
interference and defibrillation alarm range
Recovery time : 3s Resolution: 1%
ST SEGMENT -2.0mV - +2.0Mv (Automatic) Accuracy: ±2% (70~100%,Ped, non-motion+
detection: ±3% (70-100%, Neo, non-motion);
Arrhythmia Yes 0-69% unspecified
Analysis : Alarm range 0~100%
pacemaker detectable Pulse rate: Range: 20~300bpm
detection: Resolution:1bpm
Alarm: Yes, audible and visual alarm, alarm Accuracy: ±3bpm(non-motion)
events recallable ±5bpm(motion)
NIBP Alarm range:20~250bpm
Method Automatic oscillometric EtCO2(Sidestream)
Work mode: Manual / Automatic/Continuous CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to 20kPa
Measurement Unit: mmHg / Kpa selectable Range: (at 760mmHg)
Measurement Systolic, Diastolic, Mean Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
types: ± 5% of reading (41 – 70 mm Hg)
Range of systolic Adult Mode:40-270mmHg ± 8% of reading (71 –100 mm Hg)
pressure: Pediatric Mode:40-200mmHg ± 10% of reading (101 –150 mm Hg)
Neonate Mode 40-135mmHg awRR range 0-150rpm
Range of diastolic Adult Mode:10-215mmHg RR accuracy: ±2rpm (0-70rpm)
pressure: Pediatric Mode:10-150mmHg ±5rpm (>70rpm)
Neonate Mode 10-100mmHg Response time: <240msec (10% to 90%)
Range of mean Adult Mode:20-235mmHg Delay time: <2s
pressure: Pediatric Mode:20-165mmHg EtCO2(mainstream)
Neonate Mode 20-110mmHg CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to 20kPa
Over-pressure both Hardware and software over Range: (at 760mmHg)
protection: pressure protection Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
Accuracy: Less than ±5mmHg ± 5% of reading (41 – 70 mm Hg)
Resolution 1mmHg ± 8% of reading (71 –100 mm Hg)
PR from NIBP: 40-240bpm ± 10% of reading (101 –150 mm Hg)
Temperature (Dual Channel) awRR range 0-150rpm
Measurement 0-50°C (32 to 122F) awRR precision ±2rpm (0-70rpm)
range: ±5rpm (>70rpm)
Resolution: 0.1°C Delay time: <2s
Accuracy: ±0.1°C (without probe) *Notice: Specifications subject to changes without prior
Channel: 2 channels, T1, T2, TD (Temperature notice. All rights reserved by Comen
Difference)