C300B Three Phase Power Calibrator Data Sheet en

The C300B is a 3-phase power calibrator and tester used to calibrate and test a wide range of power engineering devices with high accuracy levels of 0.02% or 0.05%. It can generate voltages up to 560V and currents up to 120A to automatically test devices like electricity meters, power quality analyzers, and protective relays. Test results include calculating measurement errors and standard deviations. The C300B also enables manual calibration and has high output burdens to test older analog instruments.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
48 views6 pages

C300B Three Phase Power Calibrator Data Sheet en

The C300B is a 3-phase power calibrator and tester used to calibrate and test a wide range of power engineering devices with high accuracy levels of 0.02% or 0.05%. It can generate voltages up to 560V and currents up to 120A to automatically test devices like electricity meters, power quality analyzers, and protective relays. Test results include calculating measurement errors and standard deviations. The C300B also enables manual calibration and has high output burdens to test older analog instruments.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Three Phase Calmet Calibration

Power Calibrator Calibration


and Tester of
Power Engineering Data Sheet
Devices

C300B Power Calibrator and Tester


x 3-phase voltage source up to 560V
x 3-phase current source up to 120A and 1-phase up to 360A
x Single pair of current sockets for each current outputs
x Accuracy class 0.02% or 0.05% to calibrate digital instruments
x Single product in a single case without auxiliary amplifiers
x High burden of outputs to drive older analogue instruments
x Manual mode and automatic test procedures

The C300B Calibrator and Tester is used for calibration and testing a wide range of measuring instruments used
in power engineering and enables testing:

AC voltmeters, AC ammeters, frequency meters, phase angle


meters, power factor meters, wattmeters, VARmeters,
VAmeters, clamps meters and much more,

power quality analysers, recorders and flickermeters


IEC 61000-4-30 class A for EN 50160 compatibility or individual
requirements of user.

The C300B enables automatic testing with calculating of errors and standard deviation:

electricity meters EN 50470 with accuracy relative to an internal reference of the C300B (or
relative to an external reference meter) including:
measure the basic error and influence of frequency, voltage, self-heating, distortion,…, checking
the starting current and no-load run,

electrical measuring transducers for converting a.c. electrical quantities EN 60688


(voltage, current, active power, reactive power, frequency, phase angle, power factor),

current instrument transformers EN 60044 including measure the accuracy of current and
phase angle as well as burden measurement,

current clamps with AC & DC voltage and current output including measure the accuracy of
current and phase angle,

protective relays EN 60255 including checking of tripping time and tripping level
characteristics of ANSI#21 Distance relays, ANSI#27/59 Under/over voltage relays, ANSI#32
Directional power relays, ANSI#50/51 Time overcurrent relays, ANSI#81 Frequency relays and
more protective relays.

www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 1/6
The C300B as a power calibrator of sinusoidal signals Calmet .
____________
Calibration

The accuracy of the basic parameters of the C300B Calibrator -


voltage, current, power and energy in a wide voltage 21 ... 560V
and current 0.05 ... 120A ranges, is expressed as a percentage of
the setting value, without the component as a percentage of the
range value. This form of errors specification is very useful when
checking instruments, such as electricity meters.
Energy errors diagrams of the C300B Calibrator class 0.02 as a
function of current settings for balanced and unbalanced loads at
230V/50Hz and power factor cosM=1 and 0.5L, achieved in
automatic test procedure using Calpro300TS software and
reference meter Radian RD31.

Specification for the sinusoidal signals


Uncertainty 1)
Parameter Range Settings span Resolution Maximum load
class 0.02 class 0.05
70V 0.5000...70.0000V 0.0001V 560mA@70V
140V 1.000...140.000V 0.001V 2)4) 2)4) 280mA@140V
Voltage U r0.02% r0.05%
280V 2.000...280.000V 0.001V 140mA@280V
560V 5.000...560.000V 0.001V 70mA@560V
Voltage short term [1h] stability r0.005% 2) r0.010% 2)
Voltage long term [1year] stability r0.01% 2) r0.02% 2)
Voltage temperature drift per 1qC r0.0005% 2) r0.0010% 2)
0.5A 0.001000...0.500000A 0.000001A [email protected]
6A 0.05000...6.00000A 0.00001A 8.5V@6A
2)5) 2)5)
Current I 20A 0.2000...20.0000A 0.0001A r0.02% r0.05% 3.3V@20A
7)
0.95V@60A
120A 1.000...120.000A 0.001A 7)
0.70V@120A
2) 2)
Current short term [1h] stability r0.005% r0.010%
Current long term [1year] stability r0.01% 2) r0.02% 2)
2)
Current temperature drift per 1qC r0.0005% r0.0010% 2)
Frequency f 40.000...500.000Hz 0.001Hz r0.005%
Phase shift M 0.00...r360.00q 0.01q r0.05q 2) r0.10q 2)
Active power P 0...3x67200.0W 0.00001-1W r0.02% 2)3) r0.05% 2)3)
Reactive power Q 0...3x67200.0var 0.00001-1var r0.02% 2)3) r0.05% 2)3)
2)
Apparent power S 0...3x67200.0VA 0.00001-1VA r0.02% r0.05% 2)
Power short term [1h] stability r0.005% 2)8) r0.010% 2)8)
Power long term [1year] stability r0.01% 2)8) r0.02% 2)8)
Power temperature drift per 1qC r0.0005% 2) r0.0010% 2)
Time 6) 1...36000s 1s r0.01% r0.001s
Energy calculated from settings of power and time r0.02% 2)3) r0.05% 2)3)
1)
absolute extended uncertainty under confidence level of 95% covers reference uncertainty of standards, stability in 12 months,
influence quantities (ambient temperature in range +20…+26qC, humidity and power supply voltage according to the table 2.3,
load acc. to table 2.1, frequency in range 45…65Hz) and nonlinearity. For frequency band below 45Hz and above 65Hz – linear
rise up typically to double value for frequency 40Hz and 500Hz
2)
of setting value, for settings from 10% of current range and 30% of voltage range
3)
uncertainty of power P(Q) under cosM(sinM)=1, for cosM(sinM)z1 linear rise up to 0.15% (class 0.02) or 0.30% (class 0.05) for
cosM(sinM)=0.5
4)
for voltage below 30% of range uncertainty 0.006% of range (class 0.02) or 0.015% of range (class 0.05)
5)
for current below 10% of range uncertainty 0.002% of range (class 0.02) or 0.005% of range (class 0.05)
6)
for energy dosage
7)
0.85V@60A and 0.50V@120A using the AKD300 current cables length 1m
8)
stability of power S and P(Q) under cosM(sinM)=1, for cosM(sinM)z1 linear rise up to 0.04% for cosM(sinM)=0.5

General parameters
Weight 28kg
Width (with handle) x height (with feet and handle) x depth (with feet) (480x200x560)mm
Power supply 90V…264V / 47...63Hz / 900VA

www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 2/6
The C300B as a power quality calibrator of nonsinusoidal and time variable signals Calmet .
____________
Calibration

Power Quality function enables generation of nonsinusoidal voltage and


current waveforms with harmonics, interharmonics and subharmonics as well
as simulation of voltage, current, phase shift and frequency variations as a
function of time (dips, interruptions, swells, flicker, fluctuations and ramp
signals).
PQ functions meet all accuracy requirements of power quality testing to the
EN 61000-4 series of standards.

Harmonics
Harmonics are voltages and currents with a frequency, that is an
integral multiple of the fundamental frequency. Harmonic testing is
defined in EN 61000-4-7 and EN 61000-4-13.
The C300B can generate multi-harmonic distortion with independent
superposition of harmonic components of current and voltage, with
levels 0…100% and phase angle 0…360q of the first harmonic.

Interharmonics
Interharmonics are voltages with a frequency, that is a non-integral
multiple of the fundamental frequency. For example, in 50Hz supply
system, 150Hz is a harmonic (the third) but 175Hz is an
interharmonic. Interharmonic testing is defined in EN 61000-4-7 and
EN 61000-4-13.
The C300B can generate interharmonics at a user-definable
frequency up to 3200Hz, amplitude and phase angle.
C300B output signals recorded by digital oscilloscope
Dips, Interruptions, Swells and Shocs Single voltage dip Periodic voltage swells
Dips are a temporary reduction of the voltage below
nominal (Interruptions below 1% nominal) and Swells are
a temporary increase of the voltage above nominal up to
200% nominal for a time from 10ms up to several minutes.
Shocs (Inrush current) are a temporary increase of the
current above nominal when first turned on of an electrical Single voltage interruption Single current shoc
device. Dips, Interruptions and Swells testing are defined in
EN 61000-4-11 and EN 61000-4-34.
The C300B can generate a user-definable fast and slow
changes of voltage and current independently.

Flicker
Performance test curve
Voltage fluctuations

Flicker is a specific measurement, which sets out to measure the human 10,000
[%]

sensitivity of a flickering light caused by supply voltage fluctuation around of a 230V 50Hz rectangular modulation

120V 60Hz rectangular modulation

nominal value. Flickermeter testing is defined in EN 61000-4-15.


The C300B generates voltage changes for performance testing and displays the 1,000

results in Pst / Plt severity including combined frequency/voltage changes,


harmonic/interharmonic distortion and phase jumps.

0,100
0,1 1,0 10,0 100,0 1000,0 10000,0
Changes per minute [1/min]

Specification for the nonsinusoidal and time-varying signals


Uncertainty 1)
Parameter Settings span Resolution Conditions
class 0.02 class 0.05
magnitude 0...100% output value 0.01% r0.02% 2)
Harmonics up to 64th or 3200Hz
phase 0…360q 0.01q r0.5q 2)
Interharmonics in voltage 0...30% output value 0.01% r0.2% 3) for 16…9000Hz
magnitude 0...100% nominal value 6 digits r0.05% 4)
Dips
duration 0.02…999s 0.001s 0.001s
magnitude 0...200% nominal value 6 digits r0.05% 4)
Swells
duration 0.02…999s 0.001s 0.001s
Pst 0…40 0.00001 r1% acc. IEC61000-4-15
0.1...4000 changes/min
Flicker modulation 7 digits or 4 digits
or 0.000833…33.33Hz
duration 1s…999h 1s
1)
absolute extended uncertainty under confidence level of 95% covers reference uncertainty of standards, stability in 12 months, for 45-65Hz
2)
0,02% of output value and 0,5q for frequency range of harmonics 80-120Hz with linear rise up to 0.2% of output value and 4q for 3200Hz
3)
0,2% of output value for frequency range of interharmonics 16-120Hz with linear rise up to 2% of output value for 9000Hz
4)
with uncertainty of voltage and current for calibrator class 0.05 in setting span of voltage and current

www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 3/6
The C300B as a tester of power engineering devices Calmet .
____________
Calibration

Automatic test system

Calibrator C300B
Test System function for testing of electricity meters,
Impulse measurement of industrial transducers, current clamps,
Input #2
current transformers and protection relays in fully automatic
Reference Electricity Meter way under sinusoidal and nonsinusoidal conditions, for
Three Phase
Calibrator example the C300B can be used to type test from 0.1% to
Device Under Test
2% energy meters acc. to EN 50470.
Impulse Electricity Meter
Input #1
Current output provides currents in full current range from
Measurement Transducer, 1mA up to 120A through compact banana sockets without
Current Clamps, needed to manual switch current terminals during test.
Multimeter Current Transformer
Input

Protection Relay
Start/Stop
Input

Parameters of Inputs and Outputs for automatic tests functions


1) Number of
Input / output Range Uncertainty Conditions
inputs/outputs
Impulse Input frequency range 3)
for counting pulses from electricity 0.000001Hz…200kHz
0...2V/4...30V 0.001% @ tt1s 2
meter, photo scanning head or max test time
reference meter 1193[h]/f[kHz]
DC Voltage 0...r14.0000V 0.02% + 0.5mV
DC Current 0... r24.0000mA 0.02% + 1PA
AC Voltage 0...10.0000V 0.05% + 0.5mV
Multimeter Input 0...16.0000mA 0.05% + 1.6PA 1
AC Current 0...200.000mA 0.05% + 10PA in 45…65Hz range
0...6.0000A 0.05% + 300PA
Phase shift 0…360.00q ref. to I1 0.1q 2)
Start/Stop Input range of input voltage
0.001...100.000s 0.001s 3
for time measurement 15...250V DC/AC
Binary Output of Operate/Standby time of state change output load
2
for signalling state of Calibrator 0.001s 250VDC/0.5A/10VA
4) open collector
Impulse Output for C300B testing 0.0001Hz…210kHz 0.000…0.009% 1
28V/100mA
1)
absolute extended uncertainty under confidence level of 95% including stability in 12 months
2)
from 5% of current and voltage range
3)
fmax=120kHz for both inputs simultaneously
4)
0.003%+0.006%/t[s] for ft2Hz and 0.003%x—f[Hz] for f<2Hz, where t is test time and f is output frequency

Examples of the C300B applications


Electricity meter testing with accuracy Electricity meter testing with accuracy
referenced to an internal reference of the C300B referenced to an external reference meter

Meter Under Test Meter Under Test Reference Meter

www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 4/6
The Calpro 300 PC software package for Windows Calmet .
____________
Calibration

Calpro 300 PC software features:


x using a modern concept, which allows the operator to create own test Advantages of Calpro 300 PC soft:
procedures - this is very important because new requirements for
x user-friendly operation,
new meter generations can be realized easily without changing the
complete software, x database for devices and test
procedures,
x the automated mode - direct execution of the complete test
procedure automatically and requires no more additional handling by x fully-automatic test procedures,
operator unless it will not be defined in the test procedure, x continuous monitoring of the test,
x the manual mode - direct execution of single test step. It offers an x tables and graphics for
ideal solution for tests and evaluation of entire specifications for presentation of results,
devices under test without generating the complete test procedure, x operator interface available in
x computer database of customers, devices, measurement procedures several languages,
as well as edition of results, diagrams, tables of results and reports, x automatic measurements report
x traditional manual settings the value of all parameters of output generation.
signals.

Versions of Calpro 300 PC software:


Calpro 300 Basic PC software version enables traditional manual setting:
x the value of U+I+M+f+P+Q+S in symmetric and asymmetric circuit of
connection,
x the wave shape of voltage and current with using harmonics, interharmonics
and shape function.

Calpro 300 PQ Power Quality PC software version enables


generating sinusoidal and nonsinusoidal voltage and current, which
value is changed in time for testing meters, recorders and power
quality analyzers with the following functions:
x Slow Ramp for generate voltage and current which value is
changed relative slow to the time,
x Fast Ramp for generate voltage and current which value is
changed relative fast to the time,
x Flicker for generate voltage fluctuation (Flicker) levels expressed
in Plt and Pst coefficients.

Calpro 300 TS Test System PC software version enables using a modern concept, which allows the operator
to create own test procedures with using automated / manual mode for automatic testing the following devices:
x electricity meters, x protection relays (Quick function for quick relay's
testing, Trigger Time function for tripping time
testing and Trigger Level function for tripping level
testing),

x current clamps,
x current transformers,
x measurement transducers,

Calpro 300 Basic + TS and Calpro 300 Basic + PQ have the following functions:
x Type for entering data to testing devices database, x Result for visualization, edition and storing
x Procedure for entering data to measuring procedures measurements results in form of tables and
database, diagrams, easy Report generation, printing and
exporting data to MS Excel,

x Client function for collecting data in clients database


and Admin function for using clients database during
x Auto Test for performing automatic test of device, reports edition.

www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 5/6
The C300B Calibrator's set and applications Calmet .
____________
Calibration

Current clamps testing with using the C300B Calibrator, Electricity meter testing with
ZW100/10A coil and AD300 sockets adapter using the TB10 three phase fully
automated 0.02 (0.05) accuracy
class test station, consisting of
the C300B Calibrator, ER20 rack,
EH20 quick connection device,
CF102 photo head and AKD300
cables

C300B Calibrator's equipment


All completed C300B Calibrator’s set consists
Optionally for C300B Calibrator are available:
of:
x C300B calibrator case class 0.02 or 0.05, x computer Laptop,
x power cord, x Calpro 300TS PC Soft for automatic test of electric equipment,
x Calpro 300 Soft – Basic Version, x Calpro 300PQ PC Soft for Power Quality measurement devices testing,
x USB / RS232 adapter, x C300LabView-LabView Driver for C300B Calibrator,
x fuse T4A, 250V, 5x20 (2units), x AKD300 current cables up to
x set of safety voltage cables (6units), 120A (6units) with set (18units) of
x set of safety current cables up to 20A (6units), replaceable terminals,
x set of accessories for safety cables (12units banana
plug +12units Cu), x KAS300 transportation case for
x AD300 sockets adapter, portable work,
x C091A T3475-001 plug Amphenol for Calibrator inputs,
x operation manual of calibrator and software (2units),
x warranty card,
x calibration certificate.

x CF102 photo head with holder for


inductive meter and meter with
LED,

x MPX8 Eight Inputs Multiplexer


with MPX8 PC Soft for
simultaneously testing up to eight
electricity meters,

x ER20 rack for hanging of meter


under test,

x EH20 quick connection device,

x ZW100/10A coil 100 turns / 10A,

x ZW10/20A coil 10 turns / 20A,

x RS232 – Bluetooth adapter for wireless connection.

Calmet Ltd
Kukulcza 18, 65-472 Zielona Gora, Poland
Phone +48 68 324 04 56 Fax +48 68 324 04 57
E-mail: [email protected] Web access: http://www.calmet.com.pl
www.calmet.com.pl Three Phase Power Calibrator and Tester of Power Engineering Devices C300B Data sheet EN 2015-03 6/6