Matachana Type Autoclave

Download as pdf or txt
Download as pdf or txt
You are on page 1of 72

Posizione logo

Instructions for use

MS-22 EN13060-ST55- ENG - Rev. 3 MS-22


Symbols

Symbols displayed on the product and/or used in this manual:

WARNING! ATTENTION! General explanations,


Risk of injury To prevent damage occurring without risk to persons or objects

HOT SURFACES! HOT STEAM! Call service


Risk of burns Risk of burns

Thermo washer Consult instruction for use Do not dispose of


disinfectable with normal waste

2
Contents

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Safety advice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4. Installation and start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6. Running a sterilization cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8. Troubleshooting, alarms and messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9. Recycling and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

ANNEXES
1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
2 Maintenance of dental instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3 Sterilization load preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4 Water quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5 Example of a cycle data report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6 Accessories and spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7. W&H Installation check-list for sterilizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

3
1. Introduction
For your safety and the safety of your patients
The purpose of this manual is to provide you with information about the MS-22 sterilizer to ensure:
• proper installation and set-up;
• optimal use;
• safe and reliable operation;
• compliance with regular maintenance and servicing requirements.

Please carefully read the safety information in Chapter 3!

Intended use of the product


Small steam sterilizers are widely used for medical purposes, e.g. in general medical practices, dentistry, facilities for personal
hygiene and beauty care and also veterinary practices. They are also used for materials and equipment, which are likely to come into
contact with blood or body fluids, e.g. implements used by beauty therapists, tattooists, body piercers and hairdressers.
The devices is intended for professional use only by trained people.

About this manual


All drawings, images and texts contained in this manual are the property of the manufacturer.
All rights reserved. Even partial duplication of drawings, images or text is prohibited.
The information contained in this document is subject to change without prior notice.

Responsibility of the manufacturer


The manufacturer can only accept responsibility for the safety, reliability and performance of the product when the
product itself is installed, used and serviced in accordance with these instructions for use.
Servicing by unauthorized persons invalidates all claims under warranty and any other claims.
4
Introduction

Qualifications of the users


There are two types of users who may operate the sterilizer:
The Users are the persons who use the sterilizer according to the ADVANCED USER’s instructions. They must be trained in operating
the sterilizer and in its safe use. Training must be regular and evidence of the understanding shall be recorded.
The Advanced user is the head of the clinic/practice, who is legally responsible for the efficiency of the hygiene protocol in place as
well as the sterilization process. He/she is also responsible for the USERS’ training and the correct operation and maintenance of the
equipment.

Conformity to European Standards and Directives


Medical Device Directive 93/42/CEE for devices class IIb, in accordance with the Rule 15 – ANNEX IX of the above Directive.

Directive 97/23/CEE (PED – Pressure Equipment Directive) for every sterilization chamber designed and manufactured in
conformity to the ANNEX 1 and to the procedure described in the form D1 Annex III.

Directive 2002/96/CEE (RAEE) for disposal of parts coming from electrical or electronic products.

European standard EN13060 (small water steam sterilizers).

ANSI/AAMI ST55:2010 Table-top steam sterilizers.

See the Declaration of Conformity and the Warranty Card in the enclosed documents.

5
2. Unpacking

If the sterilizer comes from a cold location, wait until all external and internal surfaces
are free from moisture before switching it ON.
The sterilizer must be removed from the box and transported by two people.
Weight: 40.5 kg

Check the external conditions of the box and the sterilizer. In case of any damage,
immediately contact your dealer or the shipping agent that has carried out the
transport.

The packaging of the product is environmentally friendly and can be disposed of by


industrial recycling companies.

However, we recommend to keep the original packaging should you ever have to ship
or transport the sterilizer.

Open the front door.

All the accessories are in the sterilization chamber.


Remove all items except the trays and the tray rack.

6
Contents of the package

Tray (3 pieces total) Mains cable

Reversible rack

Funnel

USB pen drive

Wall spacer
Tray holder

Fast guide
Declaration of conformity
Documentation CD
Drain tube Warranty card
Works tests report

7
3. Safety advice

•Keep the user manual for further consultation.


•The user is responsible for the proper installation, the correct use and maintenance of the sterilizer in accordance with the instructions listed in
this manual.
•The safety devices of the sterilizer are impaired when the product itself is not installed, used and serviced in accordance with the instructions
provided by the manufacturer..
•The sterilizer has not been designed for the sterilization of liquids, foodstuff or waste.
•The sterilizer must not be used in presence of explosive or flammable gases, vapours, liquids or solids.
•The chamber is automatically heated up to high temperature as soon as the sterilizer is switched on – risk of burns!
•Ensure that the socket the mains cable is connected to is properly grounded.
•The trays and the sterilization load will be hot at the end of each cycle. Use tray or cassette holders to empty the sterilization chamber.
•Do not drink any water coming from the sterilizer.
•Do not exceed the maximum load weight limits as specified in this manual (see Chapter 6).
•Do not remove the name plate or any label from the sterilizer.
•To avoid electrical short circuits, do not pour water or any other liquids over the sterilizer.
•Use only the power cord set provided by the manufacturer. Plug into a wall socket with an effective ground connection.
•Switch off the sterilizer and unplug the mains cable before inspecting, carrying out maintenance or servicing the sterilizer.
•The low-voltage outlet in the rear of the sterilizer is for the connection of specific accessories only: do not connect any device other than those
specifically supplied by the manufacturer.
•All electric devices connected to the sterilizer shall be of Insulation Class II (double insulated) or higher.
•Repairs, maintenance or service must be carried out by service technicians authorized by the manufacturer and using genuine spare parts
only.
•During any repairs, maintenance or transport, ward off unauthorized personnel from the working area.
•In case of transport:
–Completely drain both water tanks (see section “Water Tanks” in Chapter 4).
–Allow the sterilization chamber to cool down.
–Use original or appropriate packaging.

8
4. Installation and start-up

Always wear personal protective equipment.

Placement
Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials.
Do not place the sterilizer so that it is difficult to open the service door and operate on the controls in it.
Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
Place the sterilizer in a well ventilated room.
If installed in a cabinet, this shall be provided with an opening of at least 200x150 mm on the rear side.
The sterilizer must not be operated in presence of explosive atmospheres.

Required minimum clearances


Back side: 50 mm
Right and left sides: 10 mm
Upper side: As required for filling the water tank, 50 mm minimum

9
Electrical connection

Electrical connection
The electrical power supply to the sterilizer must fulfil all applicable standards in the country of use,
and must comply with the data label on the back of the sterilizer.
Use only the cord set provided by the manufacturer.

Connect the cord set to the socket provided in the back of the sterilizer.
Do not use cable extensions nor multiple sockets/adapters

Connect the mains cable to a wall socket with the following characteristics:
• Single - phase 200 - 240 V, 50/60 Hz - 10A, on a dedicated circuit;

• Overvoltage category = II;

• Protected by a suitable circuit breaker and a residual-current device. All protection devices must be

certified according to applicable standards;


• Maximum current absorbed by the sterilizer: 10A;

• A grounded connection is essential.

10
Water tanks

Filling the clean water tank


• Switch the sterilizer ON
• Slide the tank cover to the right to access the clean water tank inlet.
• Remove the cap from the tank inlet;
• Insert the funnel and fill the water tank with app. 3.5 litres of distilled or demineralized water;
• Once the clean water tank is almost full, an audible tone will sound; stop filling;
• Place the cap to close the tank;
• Slide the tank cover back into its original position.

Use only high quality distilled or demineralized water (see ANNEX 4).
Do not add any chemical / additive to the water.

Draining the used and clean water tank


• Open the service door at the front of the sterilizer.
• Put a container (4 litres min.) below the sterilizer and insert into it the free end of the drain tube.
• Insert the drain tube into the right connector (grey) for the used water, or into the left connector
(blue) for the clean water.
• Let the water flow from the tank completely.
• Press the push-button on top of the quick connector to dislodge the drain tube.

11
Chamber furniture

Before touching the chamber furniture, ensure the sterilization chamber is


cold: risk of burns!

The chamber furniture consists of the trays, the tray rack and the steam diffuser plate.

Steam diffuser plate


Ensure that the steam diffuser plate is firmly hooked in its position before starting a
sterilization cycle. An improper positioning of the steam diffuser plate could result in bad
steam quality and could impair the sterilization process, with risk of non sterile load and
cross infection.
Sterility at the end of the cycle is not guaranteed if the steam diffuser plate was not correctly placed.
To hook the steam diffuser plate, slide it into the chamber until it is engaged in the end hooks.
To remove the steam diffuser plate, press it in the center of the end edge (1) and slide it outwards (2).

Chamber rack
Insert the rack into the sterilizer chamber, align it at the center/bottom of the chamber and push it gently
into position until it clicks.
The chamber rack is reversible and can accommodate 5 trays horizontally or 3 cassettes vertically.
If inserted in a 90° degree rotated position, the rack holds 3 trays or 3 cassettes horizontally.

Usable space in the chamber


195 x 195 x 390mm (WxHxD); equal to the volume of 15 litres.

12
Controls, commands, connections
Front view Service door

Water tank cover

Tank filling cover - cap


Bacteriological filter
Mains switch

Chamber door
USB port

Display

Control panel
Identification
label

Service door

Door pin

Door seal

Dust filter Sterilization chamber Used water drain (grey) Clean water drain (blue)

13
Controls and commands
Rear view Detail of the hydraulic connections (optional)

Pressure safety valve cover

Serial communication port

Used water drain

Water supply inlet

Accessory The water supply system must deliver demineralized water meeting the
power socket requirements as listed in ANNEX 4.
Do not add any chemical / additive to the water.
Mains plug socket The manufacturer’s warranty is void if the sterilizer was used with water
containing either chemical additives, or contaminant levels exceeding
Mains cable guide those listed in ANNEX 4.

The water supply system must be fitted with a backflow preventing


device complying to IEC 61770 and to national and local regulations.
Maximum pressure: 8.6 bar, minimum flow 2.5 litres/min, maximum
temperature 35 °C.

Condenser grid The maintenance of the external water filling system must be done in
exact accordance with the information of the instructions for use given
with the relevant system.

14
Controls and commands

Switching ON the sterilizer


Press the mains switch behind the service door to switch ON the sterilizer.
The visual indicator on the mains switch turns green and the START screen (see next page) appears.

“SLEEP” mode
If the sterilizer is not used for 12 hours, (the time interval can be changed, see Chapter 5 - Programming) it
will automatically switch to “SLEEP” mode.
In “SLEEP” mode the display remains dark and the sterilizer chamber is no longer heated to save energy.
Exit from “SLEEP” mode through any of the following actions:
- Press any button on the control panel;
- Open or (if it is open)close the chamber door.

It is also possible to put the sterilizer into “SLEEP” mode manually:


On the START screen, press the BACK button.
A 10 second countdown will commence. At the end of the countdown the sterilizer will enter the
“SLEEP” mode.
The countdown can be stopped at any time by pressing the BACK button.

15
Display and icons

First line
Shows the title/purpose of the current page and invites you to take an action (e.g. select a cycle).

Second line Warning, tank and printer icons


Shows the active option/action (preceded by the cursor >). The cursor (>) preceding an icon means
By pressing OK, the active option/action will be selected/executed. that an information message is to be read.
Pressing UP or DOWN changes the active option/action.

Third and fourth lines “Padlock” icon


Show additional information about the active option/action. The cursor (>) preceding this icon
informs that the door is safely locked.

16
Icons

If one or more icons of the display are preceded by the cursor, please take the actions as outlined below:

If an icon is preceded by the cursor, this means that an information message is present in the MESSAGES menu.
Follow the instructions provided in Chapter 8 to read the relevant messages.

General warning
One or more messages require your attention, or an action is required (e.g. maintenance).

Tank warning
The clean water tank needs to be filled, or the used water tank has to be drained, or a message about the water
quality is present.

Printer
An external device (printer, PC, etc.) is not working properly, or is OFF, or is disconnected from the sterilizer.

Door locked
The door is locked. During a sterilization cycle this does not indicate any anomaly.

17
Control buttons

The control panel shows four buttons:

UP button
Moves to the upper item in the list.
Increases a number or a parameter.

BACK button
Aborts the action/function.
Moves to the previous screen
without confirming/making any
changes nor saving any parameters.
OK button (confirmation button)
Confirms the active option.
Confirms a number or a date. DOWN button
Saves a configuration or a parameter. Moves to the lower item in the list.
Reduces/decreases a number or a parameter.

18
5. Programming

Initial setup
Before using the sterilizer please program important parameters such as date, time, language, display backlight and contrast.
This is done by means of the SETUP functions.

Start screen and menu options


When the sterilizer is switched ON, or when exiting from SLEEP mode, the default
cycle program is displayed, preceded by the cursor.
By pressing DOWN the other available cycle programs are displayed.
Continuously press DOWN until the MENU option appears.

Menu
Confirm (OK) the MENU option to access all sub-menus and navigate through
sub-menus with the UP and DOWN buttons.
See the following pages for a complete description of the options available and
how to use them.

Setup
Scroll (DOWN) to SETUP and confirm (OK) to access and set the main parameters
of the sterilizer.
See the following pages for a complete description of the parameters you can set
and how to do it.

19
Table 1: list of the MENU options

MENU SUB-MENU WHAT IT DOES


- Displays pending messages. Refer to Chapter 8 for a detailed list of messages.
Select a previously recorded cycle. Press OK and then scroll the list of the recorded cycles with
UP /DOWN. Press OK to select the cycle to be viewed or printed.
Displays the selected cycle. Press UP/DOWN to scroll the cycle report.
Prints the selected cycle. Press OK and scroll UP and DOWN to change the number of copies to
(*)
be printed. Once the value is displayed, press OK to print.
(*) Saves a cycle data file on a memory storage device).
Displays all the alarms that have occurred during sterilization cycles.
(continues
on next (*) Prints all the alarms that have occurred of the sterilization cycles stored in memory.
page) Sets important parameters of the sterilizer such as date, time, language, etc.
Confirm (OK) to access all available options. Refer to TABLE 2 for a detailed list of options and related programming.
(*) Enables the automatic water feed Press UP/DOWN to scroll the YES/NO options,
(*) Enables water quality warnings based then press OK to enable/disable the function
(*) on the external/internal sensor (warning), or BACK to exit without saving.
Allows the user to change the current user level. Access to advanced level or service level is
password-protected.
See “How to log in as an advanced user” in the following pages.
Allows the user to enter the activation code in order to enable some optional features.

(*) available/effective only if an endorsed device (printer, logger, PC, water supply, etc.) is connected and enabled in the SETUP menu.

20
Table 1: list of the MENU options (continued)

MENU SUB-MENU WHAT IT DOES


Displays the device brand name; e.g. W&H.
Displays the device model name; e.g. LINA.
Displays the device type; e.g. PRO13-003-22.
Displays the serial number of the sterilizer; e.g. 110009.
Displays the total number of cycles executed by the sterilizer.
Displays the status (number of cycles executed) of consumables. Permits the user to reset
the counter to zero after replacing a consumable. See Chapter 7 (Maintenance) for details.
(continued
Displays the number of cycles executed compared to the 4000 cycle service.
from
Displays the number of cycles executed compared to the 20000 cycle service.
previous
Displays the current software version.
page)
Displays the current system software version.
Displays the current version of the power firmware.
Displays the current version of the USB4 software
Displays the current version of the USB4 system software
Formats the USB device (pen drive) WARNING! All data in will be erased!
Shows the identifier of the hardware key, if connected.
Displays the software version of the PC/logger device, if connected.

21
Table 2: Detail of the SETUP options
MENU SUB-MENU WHAT IT DOES AND HOW TO SET IT
Sets the language. The active language is displayed: press OK and scroll other available languages with UP or DOWN. When the new language is
displayed press OK to confirm, or BACK to exit without saving.
Sets the date and time display formats. Press OK to access the function and then scroll with UP and DOWN until the
preferred format is displayed. Press OK to confirm. Press BACK to exit without saving.
Sets the time and date which will be used for the cycle report and for the delayed cycle start option.
By pressing OK the cursor is positioned on the date. Change the month, year and the day with UP or DOWN. By pressing OK, the
changes are saved and the cursor moves to the time setting. The procedure for setting the time is the same.
During the procedure, you can press BACK to return to the SETUP menu without saving.
Sets the operator or dental clinic name which will be used for the cycle report.
There are 18 characters (capital letters and numbers) plus space, the dash and the point. You can store only one name. By pressing OK the saved
(continues on next page)

name is displayed, or a series of dashes if no name is saved. Press UP and DOWN to change characters. Press OK to save a character and the cursor
will move to the next character. To return to the previous character, press BACK. To go to the next character without changing it, just press OK without
pressing either UP or DOWN. To go to the last character hold OK for two seconds. Press BACK on the first character to exit without saving. Press OK on
the last character to save the name as displayed.
Sets the time before the sterilizer will enter “SLEEP” mode.
In “SLEEP” mode the sterilizer consumes less energy. It is advised to set a short “SLEEP” mode time in order to save energy. See Chapter 4
“CONTROLS AND COMMANDS” for a description of “SLEEP” mode. Press OK to view the current time. Press UP or DOWN to increase or decrease the time
by increments of 10 minutes from 10 minutes to 12 hours. Press OK to save the time. Press BACK to exit without saving.
Increases or decreases the sound volume.
Press OK to view the current setting. To decrease or increase the volume press UP or DOWN: a sound will be emitted as an example. Press OK to save
the new setting. Press BACK to exit without saving.
Sets the display contrast.
Press OK to view the current setting. Press UP to increase or DOWN to decrease the contrast. Press OK to save the new setting, or BACK to exit without
saving.
Sets the device that is Serial port not in use.
connected to the serial port.
Serial port used for cycle report printer.
See note (*) for setting
instructions. Serial port used for an external PC/LOGGER (see ANNEX 6 – Accessories).

22
Table 2: Detail of the SETUP options (continued)

MENU SUB-MENU WHAT IT DOES AND HOW TO SET IT


Sets the printer model
See note (*) for instructions.
Sets the speed of the printer port
Sets the preheating mode Preheats the chamber ONLY if the chamber door is closed.
(**)
See note (*) for instructions Chamber is never preheated.
(continued from previous page)

Sets the HOT SURFACES warning A warning appears while the door is open and the chamber hot.
(**) See note(*) for instructions No warning appears.
Sets the PC/LOGGER warning A warning appears if the PC/LOGGER is not detected or if data saving fails.
(**) See note(*) for instructions No warning appears.
Sets the unit for pressure
See note (*) for instructions.
(**) Sets the unit for temperature
(*) Enables automatic saving of the cycle report on the USB device
(*) Enables automatic printing of the cycle report.
(*) Enables saving of a complete data set of parameters each second (takes more memory space)
(**) Enables printing the plateau temperature at fixed time steps (use the next option to set the step).
(**) Sets the time step and enables printing the plateau temperature at the set time interval.

Note (*) The current setting is displayed: press OK to enable changes and then UP/DOWN to scroll the available options. Press OK to set a new option, or BACK to exit without saving.
Note (**) This option is available for advanced users only. See the next page for instructions about how to log in as an advanced user.

23
How to log on as an advanced user

Some programmable options of the MS sterilizer can be changed only after logging in as an advanced user.
This is to prevent accidental changes or unexpected operation of the sterilizer.
Hiding a cycle program, making it inaccessible to users, is an example of option that can be accessed by advanced users only.

Use the UP, DOWN and OK buttons to browse the menu, choosing the following options
in sequence: MENU – SERVICE – CURRENT LEVEL.
The current level is now displayed. If you want to change it, press OK.

The screen as shown to the left will appear: you can now type in the advanced user password
(000123) using the UP, DOWN, BACK and OK buttons as follows:
- UP/DOWN to increase/decrease the current number (indicated by the cursor );
- OK to save the number and move to the next one; ^
- BACK to move to the previous number.
Press OK on the LAST number to confirm the password.
Press BACK on the FIRST number to abort the procedure.

After making the desired changes in the advanced user level, return to the user level by setting all numbers to zero, or switch OFF the
sterilizer and then ON again.

24
6. Running a sterilization cycle

Place the sterilization load in the sterilizer chamber and close the door.

See ANNEXES 2 and 3 on how to properly prepare and place the load.

Switch the sterilizer ON by pressing the mains switch behind the service door.

The start screen will show the default sterilization program, preceded by the cursor.
(The default cycle program can be changed by the advanced user; see Chapter 5 – Programming).

To select a different cycle program scroll the available options by pressing


UP or DOWN.

Select the desired cycle program by pressing OK.

25
Running a sterilization cycle
After selecting the cycle:
- the first line of the display shows the selected cycle.

- the “START NOW” option appears: press OK to start the


sterilization cycle immediately, otherwise see the next pages for
the “delayed start” options.
- the third and fourth lines show the maximum load weight limits
for the slected cycle.
- if you want to select a different cycle, press BACK to return to
the cycle selection screen (see previous page).

After initiating the cycle:


-the door locks automatically (the cursor appears near the "padlock" icon);
- the sterilization cycle starts;

- the second line shows the approximate residual cycle time;


- the third and fourth lines show the “Cycle-in-progress” information
(see “Cycle in progress” in the following pages of this chapter)

Emergency during the door locking:


While the door is getting locked, the user may reopen it immediately by pressing any button on the panel.
This action is possible during the motor operation and for two seconds more.

See the following pages for a description of each cycle program (temperatures, times, maximum load weights).
26
The available sterilization cycles

In total there are four sterilization cycles available.


To select the appropriate cycle for your load and requirements: see CYCLE SELECTION GUIDE in the table below.
For more details about the cycle parameters: see the CYCLE PROGRAM TABLE in the next page.

CYCLE SELECTION GUIDE


CYCLE NAME DESCRIPTION REF. STANDARD
Very fast processing of unwrapped dental instruments
ANSI/AAMI ST55
Wrapped dental instruments (2 kg).
Cycle for dental instruments, unwrapped (4 kg) or single wrapped (2 kg) , withstanding the
temperature of 134 degrees. EN 13060
(Type S)
Fast processing of a limited load (2 kg) of dental instruments , not bagged.
NOTE The weight in the table above is to be intended as the maximum permitted for the cycle.

For your safety and the safety of your patients


Never exceed the maximum load weight limits as specified in the CYCLE PROGRAM TABLE (see next page) as this
could impair the sterilization process.

The sterilizer is not suitable for the sterilization of liquids, textiles and hollow instruments .

27
The available sterilization cycles

CYCLE PROGRAM TABLE


LOAD Sterilization phase Total cycle time
CYCLE (3) Evaporation
Minutes
PROGRAM Max. Temperature Time(2) time (4)
Type Warm Cold
NAME(1) Weight kg °C minutes minutes
Start Start
Dental load as defined in ISO 17665-3(7).
W&H product families including 2 134 3.5 18'11" 21'30" 2(5)
handpieces and turbines. UNWRAPPED.
Dental load as defined in ISO 17665-3(7).
W&H product families including
2 134 3.5 58' 62' 40(5)
handpieces and turbines.
WRAPPED(6).

NOTES (1) Cycle names could be different depending on country requirements.


(2) Values could be different depending on country requirements.
(3) The total cycle time may vary depending on the type of load, the load weight, and other factors.
(4) The evaporation time can be increased by the SETUP menu (see “Customization of cycle parameters“) if required.
(5) Moisture acceptance criteria according to ANSI/AAMI ST55.
(6) Bagged /wrapped items shall be put in the upper tray only.
(7) With aluminium trays.

28
The available sterilization cycles

CYCLE PROGRAM TABLE


LOAD Sterilization phase Total cycle time
Evaporation
CYCLE Minutes(3)
Max. Temperature Time(2) time (4)
PROGRAM NAME(1) Type Warm Cold
Weight kg °C minutes minutes
Start Start
(5)
Solid, UNWRAPPED. 2(5) 134 3.5 26 2
Solid, UNWRAPPED. 4
134 3.5 47 20
Solid, SINGLE BAGGED. 1(6)

NOTES (1) Cycle names could be different depending on country requirements.


(2) Values could be different depending on country requirements.
(3) The total cycle time may vary depending on the type of load, the load weight, and other factors.
(4) The evaporation time can be increased by the SETUP menu (see “Customization of cycle parameters“) if required.
(5) When using the STANDARD cycle, place the load in the two upper trays only.
(6) Bagged / wrapped items shall be put in the upper tray only.

29
The sterilization cycle profile

All available sterilization cycles feature the same basic pressure profile as shown in the graph below.
The duration of the sterilization phase (or plateau time) and the sterilization temperatures differ between the various cycles.

Pressure
LEGEND
Pre-heating (this is not considered a
PHE
part of the cycle)
De-aeration (purging of air from the
PG1
sterilization chamber)

PPH Rise to the sterilization conditions


Time
PR Process (plateau/sterilization time)
EVAP Low-pressure evaporation
VEN Chamber venting
LEV Pressure leveling

Heating De-aeration Pressure Holding time Moisture Venting and


rise (plateau) evaporation pressure
levelling

30
The “Delayed start” options

After selecting a cycle program, press UP or DOWN to scroll between


the “start now”, “start at...” and “start in...” options.

Select the desired option by pressing OK.

The delayed start option is not available for all cycles.


“Start at...” option
Sets the time and date when the cycle starts.
Press OK: the display shows the last choice. If you accept it press OK, otherwise press UP or DOWN and then OK to select “Set start at...”
to set a new time/date: by pressing UP, DOWN and OK you can change the time/date. Press OK to confirm the change. The cycle will start
at the indicated time. A countdown timer will appear on the display.
Press BACK at any time to abort the procedure.
“Start in...” option
Sets a waiting interval before the cycle starts by increments of 10 minutes, up to 24 hours.
Press OK: the display shows the last interval used. Press UP or DOWN and then OK to select “Set start in...” to set a new interval.
By pressing UP and DOWN you can change the time interval. Press OK to confirm the change. The cycle will start after the programmed
interval. A countdown timer will appear on the display.
Press BACK at any time to abort the procedure.
Stopping the countdown
During the countdown, you can press UP and DOWN to scroll between the two following options:
Press OK to stop the countdown and start the cycle immediately
Press OK to stop the countdown and return to the main menu (a further confirmation will be requested)
31
Customization of cycle parameters

The evaporation time may be increased according to the sterilization protocol in force and to the type of load processed.
The evaporation time may be increased only after acceding as an ADVANCED USER.

When changing the evaporation time, ensure that the load is always dry at the end of a sterilization cycle in order to avoid
wicking of moisture and, potentially, microorganisms from hands, gloves or environmental surfaces.

After selecting a cycle, press UP or DOWN until the SETUP option appears and confirm (OK).
Scroll the sub-menu options by pressing UP or DOWN (the current value is displayed) and
follow the instructions in the table below.

MENU SUB-MENU WHAT IT DOES AND HOW TO SET IT


Sets the cycle as the default cycle, means it will appear first on the start screen. After pressing OK, press OK on YES to set the cycle
as the default cycle (the other cycles will be automatically set to NO); press BACK to exit without saving.
Unhides/hides a cycle. Once a cycle is hidden, it will no longer be visible in the menus and thus it will be impossible to launch it.
After pressing OK you can scroll between YES and NO with the UP and DOWN buttons. Press OK on YES to hide the cycle, press OK on
NO to unhide it, press BACK to exit without change.
After pressing OK, an asterisk will appear near the current value, indicating that it can be
Sets the duration of the evaporation phase. changed by pressing UP or DOWN. After programming the desired value, press OK to confirm or
BACK to exit without saving.

32
Cycle in progress

Information displayed on the screen while a cycle is in progress


First line changing between:
- Name of the current cycle;
- Progress bar (if enabled).
Selectable options (preceded by the cursor to the left)
Countdown
By pressing UP or DOWN, certain options will be available
Approximate residual time until cycle
(e.g., changing the displayed information, aborting the
completion.
cycle, viewing messages, etc.).
Name of the current cycle
phase
(See cycle profile).

Current pressure and temperature


of the sterilizer chamber. Cycle counter Cursor near the "Padlock" icon
Number of the current cycle. Indicates that the door is securely
locked.

33
Cycle in progress
INFO screen and menu options
While a cycle is in progress, you can view the main cycle parameters in real time.
On the “cycle in progress” screen, press UP or DOWN until the INFO option appears. Other menu items are also available at this stage.

Then confirm with OK.

The current parameters of the cycle in progress are displayed.


Press UP or DOWN to view the complete list of values (see table below).
Press BACK to return to the standard “cycle in progress" screen.

Screen title
Cycle time
Phase time
Heating element temperature Legend of the parameters
Steam pressure displayed when scrolling the
Temperature in the chamber INFO screen.
Heating element power output
Theoretical temperature
Additional chamber sensor temperature
Mains voltage
Mains frequency
Total water injected
Water conductivity

34
Manual stop

While a cycle is in progress, you can abort it manually at any time.


Press UP or DOWN until the STOP option appears preceded by the cursor, then proceed as shown below:

Confirm STOP (OK)

Press DOWN until YES appears.

Before the cycle abortion is confirmed, the abortion procedure


Confirm (OK) YES can be interrupted at any time; press BACK several times until
you get to the “cycle in progress” screen and the cycle will go
on as originally programmed.

Once a cycle is aborted, a reset phase commences to safely release any steam pressure from the
chamber. This may take several minutes. Do not switch off the sterilizer! Wait until the reset
phase is completed.

At this stage you can access some menu items by pressing UP or DOWN.
When selecting the INFO option (see picture) you can view the sterilizer parameters in real time (see previous page).

35
Manual stop

When the reset phase is over, press BACK

One of the following messages appears:

The message “LOAD NOT STERILE” means that the load is not sterile. Do not use
items on patients!
The message “STERILE CONDITIONS ACHIEVED - CYCLE INTERRUPTED” means that
the load might be wet. Wet items are for immediate use only!

Press OK to unlock the door as requested in the second line of the screen.
(a waiting message appears wile the door is unlocking)

Open the chamber door and remove the load, or repeat the sterilization
cycle.

36
End of a sterilization cycle
When a cycle is successfully finished, the “CYCLE COMPLETED” message appears on
the screen and the “Unlock door” option is preceded by the cursor.
At this stage you can press DOWN or UP until the INFO option appears; confirm INFO to access cycle parameters
for mechanical sterilization monitoring (see previous pages).
This is only possible prior to unlocking the chamber door.

Confirm (OK) to unlock the door (the cursor near the “padlock” icon disappears).
Wait the door to unlock, then open the chamber door.
WARNING! When you open the door, pay attention to possible steam
sprays: Risk of burns!

If an alarm message appears at the end of the cycle, consult Chapter 8


(Troubleshooting) and, if the problem persists, call for technical service.

Remove the load from the chamber.

WARNING! THE LOAD AND THE STERILIZER ARE HOT!

Always use adequate protections (cassette holder, tray holder


and/or thermal protective gloves) to remove the load!

Do not touch the chamber, the inner porthole and the internal
fittings as long as they are hot.

37
7. Maintenance

Before carrying out any maintenance on the sterilizer, switch the unit OFF and remove the mains cable.
Wear personal protective equipment (gloves, goggles, etc.) during normal use of the sterilizer and when carrying out
cleaning and maintenance tasks.

Before accessing the chamber and the connected parts, be sure that the sterilizer is cold.

Follow the instructions in this chapter when carrying out any maintenance on the sterilizer.

Maintenance program
The maintenance program is outlined in the table on the next page.
It includes the replacement of certain wearing parts (consumables) which is imperative to ensure the safe and faultless operation of
the sterilizer.

Maintenance counters
The sterilizer keeps track of the age of consumables by keeping memory of the number of cycles
executed since the last replacement.
When one counter reaches the maximum, a replacement message appears on the screen and the
consumable needs to be replaced; replace the consumable.

38
Maintenance program

MAINTENANCE PROGRAM TABLE


Frequency (*) # of cycles (*) Operation Consumable Performed by
Clean the door seal and the chamber face side
Clean the chamber, trays and the rack
Monthly 50 Clean the chamber filter
Clean the external sterilizer surfaces
Clean the steam diffuser plate User
Replace the bacteriological filter See ANNEX 5
3 months 400
Replace the dust filter
6 months 800 Clean both water tanks
Yearly 800 Replace the door seal
5 years 4000 General check and service
Service technician
- 20000 General check and service

(*) whichever occurs first

39
Monthly or 50-cycle maintenance

Cleaning the door seal and the chamber face side


Clean the door seal and the outer edge of the chamber with a non-abrasive cloth
moistened with water. If you use a detergent solution, be careful not to get in
contact with the plastic body of the front cover.
Rinse with clean water.
Do not use abrasive products, cutting tools or sharp objects.

Cleaning the chamber and the chamber accessories


Remove the trays from the chamber.
Remove the chamber rack and the steam diffuser plate.

Clean the chamber with a damp sponge and a mild detergent solution paying
attention not to bend or damage the temperature probe inside the sterilizer
chamber. Rinse with water.
Clean the steam diffuser plate, the trays and the tray rack with a damp sponge
and a mild detergent solution. Rinse with water.
Reposition all pieces of the chamber accessories properly.

Ensure that the steam diffuser plate is correctly placed and engaged, as this is
essential for the sterilization process.

The trays, the tray holder and the steam diffuser plate may also be cleaned in a
washer disinfector.
40
Monthly or 50-cycle maintenance

Cleaning the chamber filter


Empty the sterilizer chamber by removing the trays and the rack.

1–2: Remove the filter cap at the back of the chamber (bottom/center)
by turning it counter-clockwise.

3: Remove the cartridge filter and rinse it with tap water.

4–5–6: Insert the filter in the cap, attach the filter cap and lock it by
turning clockwise.

Cleaning the external surfaces of the sterilizer


Clean all external sterilizer covers with a slightly damp cloth moistened with water.
Never use disinfectants, detergents or abrasive products.

41
3 month or 400-cycle maintenance

Replacing the bacteriological filter


Open the service door.
Unscrew the bacteriological filter by hand (counter-clockwise).
Screw on the new bacteriological filter (clockwise) and tighten it snug.

Remember to reset the counter after replacement (see following pages).

Replacing the dust filter


Pull out the dust filter from underneath the sterilizer.
Detach the used filter from the handle.
Attach the new filter to the handle.
Slide the filter back into its original position.

Remember to reset the counter after replacement (see following pages).

42
6 month or 800-cycle maintenance

Cleaning the water tanks


Switch OFF the sterilizer and disconnect the mains cable.
Completely drain both tanks.
Leave the drain tube attached to one of the drain quick connectors.
Turn the 5 screws of the tank cover a ¼ turn with the use of a
screwdriver and lift the cover to gain access to the tanks.
Tap with your fingers on the rubber membrane to remove any
condensate.
Remove the rubber membrane; clean and dry it.
Clean the internal tank surfaces with a soft sponge and a mild detergent
solution, then rinse and dry them. Make sure the drain tube is
connected to the tank you are cleaning (left tank – grey colored
connector; right tank – blue colored connector) to drain the detergent
solution.
Only when both tanks are clean, remove the internal filters (A), clean
them with tap water and put them back into their position. Reposition
the rubber membrane.
Close the cover and tighten the 5 tank cover screws.
Disconnect the drain tube.

Do not use abrasive, strong detergents or disinfectants.


Use a small non-abrasive brush to clean the areas that are difficult to reach.

When cleaning the tanks, be careful not to touch the water level sensors. If moved from their original position, the operation
of the sterilizer could be impaired.
43
1 year or 800-cycle maintenance

Replacing the door seal

Fully open the chamber door.

Pull out the used door seal by hand (easy if seal and fingers are dry).

Carefully clean the seal seat and the chamber face side with a cotton swab moistened
with isopropyl alcohol.

First: up and down Moisten the new seal with water. This will make placement much easier!

Insert the new seal in the sequence as illustrated in the pictures to the left.

Complete the operation by evenly inserting the seal on the entire circumference;
ensure the seal does not stick out (no bumps or deformations)!

Remember to reset the counter after replacement (see following pages).

Then: left and right


44
4000 cycle/5 years general check and service

Regular service is imperative to ensure continuous and effective operation of the sterilizer.

It is recommended to carry out a general service every 4000 cycles or five years by an authorized service technician.

The service includes replacement of consumables and other important internal components, a check of the entire unit
with special care for the safety systems, and cleaning of areas and components that cannot be accessed by the user.

REPLACEMENT PARTS CLEANING CHECKS


Sterilization chamber and external surfaces Pneumatic connections
Chamber filter Electrical connections
Temperature and pressure calibration
Solenoid valves Internal cleaning, with particular care for the
Door locking system
condenser fins and the main board
Pressure safety valve
Steam diffuser plate Safety systems

45
Resetting the maintenance counters

Use the UP, DOWN and OK buttons to browse the menu, choosing the following options in
sequence: MENU – DEVICE INFO – SERVICE COUNTERS

Scroll to the concerned consumable by pressing UP or DOWN.


The consumable status (number of cycles executed and maximum lifespan of the
consumable) is displayed in the third and fourth line of the display.
Press OK to select the concerned consumable.

After selecting, the concerned consumable appears in the first line.

The RESET COUNTER option is displayed: confirm it with OK.

A confirmation request appears: scroll the answer to YES by pressing UP or DOWN and
then confirm with OK.

After being reset, the consumable counter shows zero.

46
8. Troubleshooting, alarms and messages

If the cursor appears to the left of one or more icons, there is related information pending.
All messages can be viewed by means of the MESSAGES sub-menu.

Use the UP, DOWN and OK buttons to browse the menu, choosing the following options in
sequence: MENU – MESSAGES.

If there is more than one message pending, you can scroll within messages with UP or
DOWN.
The icon that is preceded by the cursor is related to the pending message.

The cursor that precedes the icon disappears as soon as the relevant message has been read and the condition that
gave rise to the message has been fixed.

The cursors that precede the message icons are not visible while a cycle is in progress.

47
Messages

ICON MESSAGE DESCRIPTION/ACTION REQUIRED


- The chamber door is locked; no action required.

The water level inside the clean water tank is below the minimum. Fill the clean water tank.
The water level inside the used water tank is at maximum level. Drain used water tank.
Don’t touch the chamber or the load with bare hands: high temperature, risk of burns!
The bacteriological filter needs to be replaced.
The dust filter needs to be replaced.
The door seal needs to be replaced.
The 4000 cycle overhaul needs to be performed. Call for service.
The 20000 cycle overhaul needs to be performed. Call for service.
The CPU board battery needs to be replaced. Call for service.
PC/Logger not detected (disconnected or not powered).
Cycle report printer configured but not detected (disconnected or not powered).
File saving error (check presence and connection of the USB drive).
Error in the USB communication (between the main board and the USB4 board)
Error in saving the HTML cycle report
Error in saving the SCL cycle report
Bad USB device

NOTE: for any message not listed in this table, call service.
48
Alarm stop
In case certain important sterilization parameters are not met, the sterilizer will generate an alarm code
and abort the cycle automatically.
The sterilizer enters into a reset phase; a wait message and an alarm code are displayed on the screen.

At this stage select and confirm “Info” to view the sterilizer parameters (see Chapter 6 of this manual).

Do not switch off the sterilizer: It will take some time (several minutes) to reset the system and reach safe conditions in the sterilizer
chamber before it is possible to open the sterilizer door and remove the load.

Alarm end
When the reset phase is over, you will be asked to press BACK to get to the
“Unlock door” option.

Confirm (OK) to unlock the door.


While the door is unlocking, a waiting message is displayed.
The message “LOAD NOT STERILE” means that the load is not sterile. Do not use
items on patients!
The message “STERILE CONDITIONS ACHIEVED - CYCLE INTERRUPTED” means that
the load might be wet. Wet items are for immediate use only!
Water could be present in the chamber when
Open the chamber door
opening the door: prevent spilling (e.g.,
and remove the load.
place a towel under the chamber door).

49
Alarms

Alarm code DESCRIPTION ACTION


E010 Power failure during a cycle Load cannot be considered sterile. Repeat the cycle.
E02x Internal voltage error Switch the sterilizer OFF and ON. If the problem persists call service.
Switch the sterilizer OFF and ON. If the problem persists call service.
E041 Cycle counter lost
NOTE: Initiating a sterilization cycle is still possible.
Set date and time - Switch the sterilizer OFF and ON. If the problem persists call service.
E042 Internal clock error
NOTE: Initiating a sterilization cycle is still possible.
Disconnect optional accessories from 24VDC rear plug - switch the sterilizer OFF and ON. If the problem
E060 Internal voltage error
persists call service.
E080 Internal overheating Check the dust filter and ensure that the sterilizer fan is not blocked.
E090 Internal voltage error Switch the sterilizer OFF and ON. If the problem persists call service.
E100 Phase timeout Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E101 Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E104 Phase timeout Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E121 Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E130 Overpressure during the sterilization phase
E131 Temperature fluctuation during the steril. phase
E132 Temperature difference too high during PG1 Clean the chamber from residuals of detergents, disinfectants and other chemicals.
E140 Low pressure during the sterilization phase Replace the clean water if it is suspected to be contaminated with chemicals.
E150 Low temperature during the sterilization phase Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
E160 Overtemperature during the sterilization phase Repeat the cycle. If the problem persists call service.
E163 Overpressure detected
E164 Overpressure during PG1

NOTE: for any alarm not listed in this table, call technical service.

50
Alarms

Alarm code DESCRIPTION ACTION


E180-E181 Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E183 Pressure drop timeout Clean the chamber filter. If the problem persists call service.
E184 Overtemperature detected If the problem persists call service.
E215 Fan blocked or faulty electronic control Call service.
E230 Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E231 Overtemperature detected If the problem persists call service.
E232-E233-E234 Internal probe error Switch the sterilizer OFF and ON. If the problem persists call service.
E240 Heating element error Wait for the chamber to cool down. Reset the thermal overload. If the problem persists call service.
Switch the sterilizer OFF. Wait for the chamber to cool down. Switch the sterilizer ON. If the problem
E241 Heating element overheating
persists call service.
E242 Chamber filter blocked Clean the chamber filter. If the problem persists call service.
E310-E320-E380- Check the door seal; clean or replace if necessary. Clean the chamber face side. Clean the chamber
Vacuum timeout
E390 filter. If the problem persists call service.
E510 Door motor: failure after cycle completion Switch the sterilizer OFF and ON. If the problem persists call service.
E520 Door motor: locking timeout If the problem persists call service.
E570 Door motor: unable to detect the door position Switch the sterilizer OFF and ON. If the problem persists call service.
E580 Door motor: door locked check signal lost If the problem persists call service.
E59x Door motor error Switch the sterilizer OFF and ON. If the problem persists call service.
Switch the sterilizer OFF and ON. If the problem persists call service.
E950 Internal memory error
NOTE: Initiating a sterilization cycle is still possible.
E95x-E96x Internal memory error Switch the sterilizer OFF and ON. If the problem persists call service.
E990 Manual stop The cycle has been aborted by the user. Re-process the load.

NOTE: for any alarm not listed in this table, call technical service.

51
Resetting the thermal overload

A safety thermostat is fitted on the sterilizer to prevent overheating


of the electric heater.
If the safety thermostat opens because of too high temperatures,
the alarm E240 or a timeout alarm is generated.
If this happens, proceed as follows:
- Switch the sterilizer OFF and remove the mains cable.
- Wait for the sterilizer to cool down.
- Remove the dust filter.
- Slide your hand underneath the sterilizer where the dust filter
was located and push on the reset button of the thermostat switch
(see pictures to the left).
-A click sound will indicate that the thermostat switch has been
reset.
- Insert the dust filter back into its original position.
-Connect the mains cable and switch the sterilizer ON.
- Wait for the sterilizer to finish the alarm reset phase and follow
the instructions on the display.

If the thermostat opens repeatedly, call technical service.

52
Troubleshooting

PROBLEM POSSIBLE CAUSE SOLUTIONS


The main switch or network circuit breaker is OFF Activate the main switch or network circuit breaker (ON).
The sterilizer remains switched
No voltage at the socket Check the electric circuit.
OFF.
The mains cable is not properly connected Attach the cord set properly.
Water is leaking at the front of Leaks through the chamber door seal Clean or replace the door seal. Clean the chamber face side.
the sterilizer Internal leak. Call technical service.
The cycle commences but there The thermal overload switch is open Reset the thermal overload switch (see “Resetting the thermal overload” in this manual).
is no pressure/temperature rise Electric – electronic fault Call technical service.
Sterilizer not properly levelled Properly level the surface the sterilizer is placed on.
When opening the door at the Comply with the maximum load weight limits for each type of load.
Overloaded chamber
end of the cycle, water spills Always use the chamber rack for trays and cassettes.
from the chamber Chamber filter clogged Remove and clean the chamber filter.
Load incorrectly placed Follow the recommendations as listed in ANNEX 2.
Tap water on instruments when placed in the sterilizer Ensure that instruments are dry before they are placed in the sterilizer.
Use of water of poor quality or water containing
Drain both water tanks. Use water of good quality (see ANNEX 4).
chemical substances
Corrosion or spots on
Organic or chemical residues on the instruments Clean, rinse and dry instruments before placing them in the sterilizer (see ANNEX 2).
instruments
Ensure that instruments of different materials do not touch (aluminum, carbon or stainless
Contact between instruments of different materials
steel, etc.); place them on different trays or cassettes or pouch them (refer to ANNEX 2).
Scale deposits on the chamber Clean the chamber and use water of good quality (refer to ANNEX 4).
Instruments are turning brown or Select a sterilization cycle featuring a lower sterilization temperature. Follow the instructions
Incorrect temperature selected
black. of the instrument manufacturer.
Printer not properly connected or not powered Check the data and the power connection to the printer.
If the printer is connected directly: configure the serial port to “Printer” (see Table 2).
Serial port not configured
The cycle report printer does not If connected via PC/Logger: configure the serial port to “PC/Logger” (see Table 2).
work You are trying to print a stored cycle but the printer is busy to print the data of the cycle in
The cycle is in progress and the automatic report
progress: the requested printout will be queued.
printing is enabled
NOTE: The max. queue is 5 cycles. Longer queues will be ignored.
No cycles are stored in the cycle Power board replaced by service
These service steps cause loss of memory.
history menu Serial number re-entered by service

53
Troubleshooting

PROBLEM POSSIBLE CAUSE SOLUTIONS


When starting a cycle, the Door seal not properly placed; seal sticking out Ensure that the door seal is evenly inserted on the entire circumference.
chamber door locks but re-opens OK button was pressed twice to launch the cycle Try again by pressing OK only once.
immediately. The “Open the Remove any objects interfering with the chamber door. Check the door does not force
Door jammed by external objects or by the load itself
door” message appears. against the load or the chamber furniture.
Water fill system (optional) not installed Install a water fill system.
When the sterilizer is connected Water fill system (optional) not connected Connect the water fill system to the sterilizer (see ANNEX 4 for water quality requirements).
to an automated water supply Water fill system (optional) not configured Enable the water fill system in the “Configuration” sub-menu (see Chapter 5, Table 2).
system: There is no clean water Since water tank filling is attempted only once in-between cycle execution, this event
in the tank, but the automatic When the water fill system attempted to fill the tank, inhibits water loading. Switch the sterilizer OFF and then ON again.
water filling does not start. water was temporarily unavailable Check the external water supply system.
Check for water leaks from the sterilizer.
The sterilizer enters into “Sleep The chamber door has not been opened after the Press any button on the control panel to exit from “Sleep mode”.
mode” immediately after previous cycle had finished and the “Sleep mode
opening the chamber door. delay” has expired
At the end of the cycle the The chamber is in vacuum due to an internal Switch the sterilizer OFF: this will release any internal pressures allowing the chamber door
display reads “Open the door” malfunction to be opened. Call technical service if the problem persists.
but opening the door is Remove the bacteriological filter to get the pressure released. Replace the filter. Note that
The bacteriological filter is blocked
impossible. bacteriological filters need to be replaced every 400 cycles.
The sterilization (PROCESS) The chamber temperature dropped below the minimum
phase of a sterilization cycle was threshold and the software performed a successful Wait for cycle completion. If the problem occurs frequently, call technical service.
longer than expected. recovery

Before sending the sterilizer for technical service, remove the mains cable, empty both water tanks and use the original or
appropriate packaging.

54
9. Recycling and disposal

MS sterilizers are mainly built from fiber-reinforced polymers, metals and electronic components.

In case of disposal:
- separate the various components according to the materials they are made of;
- drop the sterilizer with a company that specializes on the recycling of related products;
- do not abandon the sterilizer in unsecured places;
- always refer to current/applicable laws and rules in the country of use.

The same instructions apply to disposal of all used consumable parts.

55
ANNEX 1. Technical data

TECHNICAL DATA CONFORMITY


Electrical supply: 200 - 240 VAC – 50/60 Hz, single - phase STERILIZER featuring type S sterilization cycles conform with the following standards:
Nominal voltage: 200 - 240 VAC 93/42/CEE Medical Device Directive (MDD)
Max. power: 1750 W 97/23/CEE Pressure Equipment Directive (PED)
Max. current: 8.75 A 2002/96/CEE Waste Electrical and Electronic Equipment (WEEE)
Sterilizer: EN 13060 Small steam sterilizer
Working temperature: from +5°C to +40°C ANSI/AAMI ST55 Table-top steam sterilizers.
Working relative humidity: Max. RH 80% up to 31°C, linearly decreasing to 50% at 40°C Safety requirements for electrical equipment for measurement, control and
Storage temperature /rel. humidity: -20°C to +60°C/0-90% (with empty tanks) IEC 61010-1
laboratory use, general requirements
Max altitude: 4000m asl Safety requirements for electrical equipment for measurement, control and
Min. atmospheric pressure: 0.6 barA IEC 61010-2-040 laboratory use; particular requirements for autoclaves using steam for the
Overall dimensions: W: 450 mm/H: 435 mm/D: 599 mm treatment of medical materials and for laboratory processes.
Min. space required: W: 470 mm/H: 485 mm/D: 650 mm Electrical equipment for measurement, control and laboratory use: EMC
Size of the door movement: W: 360 mm/H: 410 mm/D: 360 mm EN 61326
requirements.
Weight empty: 40.5 kg
Max. weight (fully loaded): 53 kg
Weight per support area 758 kN/m2 PRODUCT LABELS
Max. heat output: 3000 KJoule/hour
Pressure safety valve: 3 bar
Safety thermostat 330°C
Sterilizer chamber:
Total volume: 22 l/Ø 250 mm x D 440 mm
Usable space (for all cycles) 15.5 l/W 195 mm x H 195 mm x D 390 mm
Bacteriological filter: 0.3 µm
Distilled or demineralized water:
Water quality: Fulfilling EN 13060 Ann. C (conductivity < 15μS/cm) Data plate
Average water consumption: 0.7 litres/cycle
Tank volume: 2 x 3.6 litres
External water supply (optional): To be compliant with IEC61770
Pressure: min. 2 bar – max. 8.6 bar
Flow: min. 0.25 – max 0.5 l/min
Communication with other devices: Serial port, USB port
Other Fully micro-processor controlled, process evaluation system
according to EN13060. Manufacturer label
Programmable sleep-mode. Sterilization chamber label

56
ANNEX 2. Maintenance of dental instruments

External disinfection
This procedure reduces the risk of infection during cleaning and maintenance of the instrument.
Wear protective gloves during disinfection.
Refer to the instructions of the instrument manufacturer.
Avoid using abrasive disinfectants, pH must be between value 2.5 and 9; no chlorine based disinfectants.
Residual disinfectants on instruments can cause extensive damage to your instrumentation during sterilization (oxidation, alteration
of technical characteristics of seals, rubbers, fiber optics, etc.).

External cleaning
This procedure involves the removal of residues (blood, dentine, etc.) that adhere to the surface of the instruments.
Wear protective gloves during cleaning.
Refer to the instructions of the instrument manufacturer.
Use a soft, damp brush and take care not to scratch the surface of the light ports.

57
ANNEX 3. Sterilization load preparation

Cleaning the instruments


Clean all instruments thoroughly prior to sterilization.
If possible, clean instruments immediately after use; always follow the instrument manufacturer ‘s instructions.
Remove all traces of disinfectants and detergents.
Rinse and dry carefully all instruments.

The instruments must be carefully rinsed and dried prior to sterilization.


Any residual of chemicals (like cleaning and disinfection products), could affect the purity of the steam and consequently the
whole sterilization process, and could seriously damage the sterilizer.
The manufacturer’s warranty is void in case of damage from chemicals coming from the load or added to it.

Preparing the trays


Do not overload the chamber; adhere to the maximum load weight limits (see cycle program table; the available sterilization cycles).
Always use the chamber rack to allow adequate steam circulation.
Do not overload trays; spread single items on multiple trays.
Place cassettes in the vertical position (if possible) to enhance evaporation.
Place empty containers or non-perforated trays upside down to prevent accumulation of water.
Items made from different materials (stainless steel, carbon steel, aluminum, etc.) must be placed on separate trays.
If the instruments are manufactured from carbon steel, paper should be placed between them and the trays to avoid rusty spots.
Sterilize hinged instruments (e.g., forceps, extraction pliers, etc.) in the open position.

58
Loading the chamber
Tubes
Rinse, drain and dry tubes after washing.
Place tubes on a tray allowing the ends to remain open. Do not bend tubes.
Paper side up
Wrapped/bagged items
Place the bags on trays allowing adequate space in-between bags.
Ensure that packs do not touch the sterilizer chamber walls.
Place sterilization bags with the paper side facing up.

Never place the load or the trays directly into the chamber without the chamber rack as
Space between bags this could affect the steam and temperature distribution. The load must always be
supported by the chamber rack.

Before initiating a sterilization cycle, always check that the steam diffuser plate is properly
positioned.

An improper positioning of the steam diffuser plate could result in bad steam
quality and could impair the sterilization process, with risk of non sterile load and
cross infection.
Sterility at the end of the cycle is not guaranteed if the steam diffuser plate was
not correctly placed.

Before touching, ensure the sterilization chamber is cold: risk of burns!

59
ANNEX 4. Water quality
The sterilizer uses distilled or demineralized water to generate steam for the sterilization process.
The table below lists the maximum content of minerals and the specifications for the water used for steam sterilization (see EN13060 ANNEX C).

FEED WATER SPECIFICATIONS


Contaminants/minerals/qualities Value/Specification
Evaporate residue < 10 mg/l
Silicon oxide, SiO2 < 1 mg/l
Iron < 0,2 mg/l
Cadmium < 0,005 mg/l
Lead < 0,05 mg/l
Heavy metals (excl. iron, cadmium, lead) < 0,1 mg/l
Chloride < 2 mg/l
Phosphate < 0,5 mg/l
Conductivity (at 20°C) < 15 µs/cm
pH value 5-7
Appearance colorless, clean, free from sediment
Hardness < 0,02 mmol/l
Chemical additives No chemicals or additives must be added to the water used for the steam sterilization process, even if
they are specifically claimed for use in steam generators, or for steam production, or as additives for
sterilization, disinfection, cleaning or corrosion protection.
The use of water with a conductivity greater than 15µS/cm may affect the sterilization process and damage the sterilizer.
The use of water with a conductivity greater than 50µS/cm, or not complying with the specifications in the table above, may strongly affect
the sterilization process and seriously damage the sterilizer.
The manufacturer’s warranty is void if the sterilizer was used with water containing contaminant or chemical levels exceeding
those listed in the table above.
60
ANNEX 5. Example of cycle data report

Sterilizer model, serial number, software rev.


Surgery – practice – doctor name
Cycle name
Cycle counter
Programmed sterilization temperature
Programmed sterilization time
Programmed evaporation time
Cycle start date and time
Headers for the table below
Cycle start

De-aeration phase (purging pulses)

End of the de-aeration phase


Phase of pressure rise to sterilization conditions
Sterilization phase (process) start
Min. and max. temperatures
during the sterilization phase (process)
Min. and max. pressures
during the sterilization phase (process)
Process end conditions
Evaporation phase start
Evaporation phase

Evaporation phase end


Pressure levelling phase
Cycle end conditions
Cycle end date and time
Cycle outcome
Tracking code for traceability management
61
ANNEX 6. Accessories and spare parts

Aluminium tray
Printer model S'Print part n. F523205x
part n. 19721108
Tray holder
part n. F523001x
USB pen drive
part n. V000004x Funnel
part n. F540903x

Automatic water feed and drain kit Bacteriological filter


part n. X051110x part n. W322400x

Door seal
part n. F460504x
Permanent drain kit Wall spacer
part n. X051052x part n. F190107x
Safety bracket kit
Drain tube kit with fittings part n. X051019x
part n. A812110x
Drain tube
part n. S230900x

Dust filter Mains cable


part n. F364502x part n. U38010xx
62
Accessories
Cycle report printer (S'Print) - part n. 19721108
S'Print is a compact, reliable and easy-to-use printer that can be connected directly to the serial port located in
the rear of the sterilizer.
S'Print can be easily managed from the sterilizer control panel (See Chapter 5 - Programming) in order to:
- Print cycle data reports (see ANNEX 5 “Example of a cycle data report”) at the end of each cycle either in
automatic or manual print mode;
- Print a report of any cycle stored in the sterilizer memory.
Printouts are very durable and can be stored in the file records for years.

Water feed system - part n. X051110x


Mount this kit in the sterilizer if you want to connect a water filtration system to automatically fill
the clean water tank with demineralized water and drain the used water tank continuously.
The kit needs to be mounted by an authorized service technician, or by the factory upon specific
order request.
Water quality provided by the filtration system has to comply with ANNEX 4. The water supply
pressure must be between 2 and 8,6 bar.

Permanent drain kit - part n. X051052x


This kit is mounted to continuously drain the used water tank, thus manual tank draining is no longer necessary.
The kit needs to be mounted by an authorized service technician, or by the factory upon specific order request.

Drain tube kit with fittings - part n. A812110x


Use this kit to connect the sterilizer permanent drain to a drain pipe.
63
Accessory connection scheme

External water
filtration system

Water feed system

USB Osmo
S’Print
drive

64
Accessories and spare parts

190x43x300/375 *
Standard chamber rack for 3 cassettes (*)
190x43x300/375 *
part n. F523009x
190x43x300/375 *

205x35x300/375 *
Standard chamber rack for 3 USA size cassettes (*)
210x35x300/375 *
part n. F523021x
205x35x300/375 *

190x32x300/375 *

210x32x300/375 * Standard chamber rack for 4 cassettes (*)


210x32x300/375 * part n. F523015x
190x32x300/375 *

190x70x300/375 *
Standard chamber rack for 2 implant cassettes (*)
190x70x300/375 * part n. F523017x
(*) All racks shown in this page, if rotated
90°, accept 5 standard aluminium trays.

65
Consumables

Bacteriological filter - part n. W322400x


Replace every 400 cycles

Door seal - part n. F460504x


Replace every 800 cycles

Dust filter - part n. F364502x


Replace every 400 cycles

400-800 cycle consumable kit - part n. X050315x


This kit consists in a stock of consumables suitable to run 800 cycles.
2x
It includes:
- 1 door seal;
- 2 air filters;
- 2 dust filters with handle.
2x

66
ANNEX 7. W&H Installation check-list for sterilizers
1. Was the head of the clinic/practice present during all the in-service? Yes No
2. Is the packing of the sterilizer undamaged? Yes No
3. When unpacked, is the sterilizer undamaged? Yes No
4. Are all the contents of the package available (sterilizer ship-with)? Yes No
5. Are all the ordered accessories available with the sterilizer? Yes No
6. Have you removed all the protection covers from the sterilizer and from all the ship-with? Yes No
7 Were all sections of the instruction for use of the sterilizer covered during the in-service? Yes No
8 Is the allocated countertop for the sterilizer, levelled and flat? Yes No
9 Are the recommended ventilation conditions, in the allocated area for the sterilizer, complied with? Yes No
10 Are the required minimum clearances complied with? Yes No
11 Have you explained the water quality required for the use of the sterilizer? Yes No
12 Have you shown to the head of the clinic/practice the procedure for filling and draining the main and used water tanks? Yes No
13 Have you shown to the head of the clinic/practice how to program the sterilizer? Yes No
14 Have you shown to the head of the clinic/practice the cycle options? Yes No
15 Have you shown to the head of the clinic/practice what the Messages and Alarms mean? Yes No
16 Have you shown to the head of the clinic/practice how to manually abort a cycle? Yes No
17 Have you shown to the head of the clinic/practice the maintenance program and procedures? Yes No
18 Have you shown to the head of the clinic/practice how to use all of the accessories? Yes No
19 Have you shown to the head of the clinic/practice the advantages of having a USB connection for a pen drive? (if available) Yes No
Have you suggested to the head of the clinic/practice to periodically backup the data, stored on the USB pen drive, on another
20 Yes No
safe support? (if available)
67
W&H Installation check-list for sterilizers (continued)

Have you shown to the head of the clinic/practice the advantages of having a Ethernet connection (remote data saving)? (if
21 Yes No
available)
22 Have you run a sterilization cycle (134°C) with the tray rack and trays inserted? Yes No
23 Have you explained to the head of the clinic/practice the correct load type for each available sterilization program? Yes No
24 Have you shown to the head of the clinic/practice how to prepare and place the load in the sterilizer chamber? Yes No
25 Have you explained to the head of the clinic/practice to use only original parts and accessories on the sterilizer? Yes No
26 Have you shown and explained to the head of the clinic/practice the SAFETY ADVICE section? Yes No

Sterilizer Serial Number Date

Purchased From

Installed by

Dr./Clinic Name

Address

Phone

Receiver’s Signature Installer’s Signature

Make a copy of the Installation Checklist and store it.


68
Notes
Questo testo serve solo per far uscire dritta la pagina PDF
Authorized W&H service partners
Find your nearest W&H service partner at http://wh.com
Simply go to the menu option »Service« for full details. Alternatively please contact:

W&H UK LIMITED, 6 Stroud Wood Business Centre, Park Street, St.Albans, Herts AL2 2NJ
t +44 1727 874990 f +44 1727 874628 [email protected]

A-DEC AUSTRALIA CO.INC., Unit 8, 5-9 Ricketty Street, Mascot NSW 2020,
t +61 2 83324000 f +61 2 83324099 [email protected]

Ivoclar Vivadent NZ, P.O.Box 5243, Wellesley Street, 12 Omega Street, Auckland,
t +64 9 914 9999 f +64 9 914 9958 [email protected]
Manufacturer Iso 13485,
93/42 EEC – Annex II
W&H Sterilization S.r.l.
Italy, I-24060 Brusaporto (Bg), via Bolgara, 2

t +39/035/66 63 000 f +39/035/50 96 988 MS-22 EN13060-ST55- ENG - Rev. 3


wh.com Subject to alterations 11.01.2017

You might also like