Matachana Type Autoclave
Matachana Type Autoclave
Matachana Type Autoclave
2
Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Safety advice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4. Installation and start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
5. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6. Running a sterilization cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8. Troubleshooting, alarms and messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9. Recycling and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
ANNEXES
1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
2 Maintenance of dental instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
3 Sterilization load preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4 Water quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5 Example of a cycle data report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6 Accessories and spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7. W&H Installation check-list for sterilizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3
1. Introduction
For your safety and the safety of your patients
The purpose of this manual is to provide you with information about the MS-22 sterilizer to ensure:
• proper installation and set-up;
• optimal use;
• safe and reliable operation;
• compliance with regular maintenance and servicing requirements.
Directive 97/23/CEE (PED – Pressure Equipment Directive) for every sterilization chamber designed and manufactured in
conformity to the ANNEX 1 and to the procedure described in the form D1 Annex III.
Directive 2002/96/CEE (RAEE) for disposal of parts coming from electrical or electronic products.
See the Declaration of Conformity and the Warranty Card in the enclosed documents.
5
2. Unpacking
If the sterilizer comes from a cold location, wait until all external and internal surfaces
are free from moisture before switching it ON.
The sterilizer must be removed from the box and transported by two people.
Weight: 40.5 kg
Check the external conditions of the box and the sterilizer. In case of any damage,
immediately contact your dealer or the shipping agent that has carried out the
transport.
However, we recommend to keep the original packaging should you ever have to ship
or transport the sterilizer.
6
Contents of the package
Reversible rack
Funnel
Wall spacer
Tray holder
Fast guide
Declaration of conformity
Documentation CD
Drain tube Warranty card
Works tests report
7
3. Safety advice
8
4. Installation and start-up
Placement
Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials.
Do not place the sterilizer so that it is difficult to open the service door and operate on the controls in it.
Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
Place the sterilizer in a well ventilated room.
If installed in a cabinet, this shall be provided with an opening of at least 200x150 mm on the rear side.
The sterilizer must not be operated in presence of explosive atmospheres.
9
Electrical connection
Electrical connection
The electrical power supply to the sterilizer must fulfil all applicable standards in the country of use,
and must comply with the data label on the back of the sterilizer.
Use only the cord set provided by the manufacturer.
Connect the cord set to the socket provided in the back of the sterilizer.
Do not use cable extensions nor multiple sockets/adapters
Connect the mains cable to a wall socket with the following characteristics:
• Single - phase 200 - 240 V, 50/60 Hz - 10A, on a dedicated circuit;
• Protected by a suitable circuit breaker and a residual-current device. All protection devices must be
10
Water tanks
Use only high quality distilled or demineralized water (see ANNEX 4).
Do not add any chemical / additive to the water.
11
Chamber furniture
The chamber furniture consists of the trays, the tray rack and the steam diffuser plate.
Chamber rack
Insert the rack into the sterilizer chamber, align it at the center/bottom of the chamber and push it gently
into position until it clicks.
The chamber rack is reversible and can accommodate 5 trays horizontally or 3 cassettes vertically.
If inserted in a 90° degree rotated position, the rack holds 3 trays or 3 cassettes horizontally.
12
Controls, commands, connections
Front view Service door
Chamber door
USB port
Display
Control panel
Identification
label
Service door
Door pin
Door seal
Dust filter Sterilization chamber Used water drain (grey) Clean water drain (blue)
13
Controls and commands
Rear view Detail of the hydraulic connections (optional)
Accessory The water supply system must deliver demineralized water meeting the
power socket requirements as listed in ANNEX 4.
Do not add any chemical / additive to the water.
Mains plug socket The manufacturer’s warranty is void if the sterilizer was used with water
containing either chemical additives, or contaminant levels exceeding
Mains cable guide those listed in ANNEX 4.
Condenser grid The maintenance of the external water filling system must be done in
exact accordance with the information of the instructions for use given
with the relevant system.
14
Controls and commands
“SLEEP” mode
If the sterilizer is not used for 12 hours, (the time interval can be changed, see Chapter 5 - Programming) it
will automatically switch to “SLEEP” mode.
In “SLEEP” mode the display remains dark and the sterilizer chamber is no longer heated to save energy.
Exit from “SLEEP” mode through any of the following actions:
- Press any button on the control panel;
- Open or (if it is open)close the chamber door.
15
Display and icons
First line
Shows the title/purpose of the current page and invites you to take an action (e.g. select a cycle).
16
Icons
If one or more icons of the display are preceded by the cursor, please take the actions as outlined below:
If an icon is preceded by the cursor, this means that an information message is present in the MESSAGES menu.
Follow the instructions provided in Chapter 8 to read the relevant messages.
General warning
One or more messages require your attention, or an action is required (e.g. maintenance).
Tank warning
The clean water tank needs to be filled, or the used water tank has to be drained, or a message about the water
quality is present.
Printer
An external device (printer, PC, etc.) is not working properly, or is OFF, or is disconnected from the sterilizer.
Door locked
The door is locked. During a sterilization cycle this does not indicate any anomaly.
17
Control buttons
UP button
Moves to the upper item in the list.
Increases a number or a parameter.
BACK button
Aborts the action/function.
Moves to the previous screen
without confirming/making any
changes nor saving any parameters.
OK button (confirmation button)
Confirms the active option.
Confirms a number or a date. DOWN button
Saves a configuration or a parameter. Moves to the lower item in the list.
Reduces/decreases a number or a parameter.
18
5. Programming
Initial setup
Before using the sterilizer please program important parameters such as date, time, language, display backlight and contrast.
This is done by means of the SETUP functions.
Menu
Confirm (OK) the MENU option to access all sub-menus and navigate through
sub-menus with the UP and DOWN buttons.
See the following pages for a complete description of the options available and
how to use them.
Setup
Scroll (DOWN) to SETUP and confirm (OK) to access and set the main parameters
of the sterilizer.
See the following pages for a complete description of the parameters you can set
and how to do it.
19
Table 1: list of the MENU options
(*) available/effective only if an endorsed device (printer, logger, PC, water supply, etc.) is connected and enabled in the SETUP menu.
20
Table 1: list of the MENU options (continued)
21
Table 2: Detail of the SETUP options
MENU SUB-MENU WHAT IT DOES AND HOW TO SET IT
Sets the language. The active language is displayed: press OK and scroll other available languages with UP or DOWN. When the new language is
displayed press OK to confirm, or BACK to exit without saving.
Sets the date and time display formats. Press OK to access the function and then scroll with UP and DOWN until the
preferred format is displayed. Press OK to confirm. Press BACK to exit without saving.
Sets the time and date which will be used for the cycle report and for the delayed cycle start option.
By pressing OK the cursor is positioned on the date. Change the month, year and the day with UP or DOWN. By pressing OK, the
changes are saved and the cursor moves to the time setting. The procedure for setting the time is the same.
During the procedure, you can press BACK to return to the SETUP menu without saving.
Sets the operator or dental clinic name which will be used for the cycle report.
There are 18 characters (capital letters and numbers) plus space, the dash and the point. You can store only one name. By pressing OK the saved
(continues on next page)
name is displayed, or a series of dashes if no name is saved. Press UP and DOWN to change characters. Press OK to save a character and the cursor
will move to the next character. To return to the previous character, press BACK. To go to the next character without changing it, just press OK without
pressing either UP or DOWN. To go to the last character hold OK for two seconds. Press BACK on the first character to exit without saving. Press OK on
the last character to save the name as displayed.
Sets the time before the sterilizer will enter “SLEEP” mode.
In “SLEEP” mode the sterilizer consumes less energy. It is advised to set a short “SLEEP” mode time in order to save energy. See Chapter 4
“CONTROLS AND COMMANDS” for a description of “SLEEP” mode. Press OK to view the current time. Press UP or DOWN to increase or decrease the time
by increments of 10 minutes from 10 minutes to 12 hours. Press OK to save the time. Press BACK to exit without saving.
Increases or decreases the sound volume.
Press OK to view the current setting. To decrease or increase the volume press UP or DOWN: a sound will be emitted as an example. Press OK to save
the new setting. Press BACK to exit without saving.
Sets the display contrast.
Press OK to view the current setting. Press UP to increase or DOWN to decrease the contrast. Press OK to save the new setting, or BACK to exit without
saving.
Sets the device that is Serial port not in use.
connected to the serial port.
Serial port used for cycle report printer.
See note (*) for setting
instructions. Serial port used for an external PC/LOGGER (see ANNEX 6 – Accessories).
22
Table 2: Detail of the SETUP options (continued)
Sets the HOT SURFACES warning A warning appears while the door is open and the chamber hot.
(**) See note(*) for instructions No warning appears.
Sets the PC/LOGGER warning A warning appears if the PC/LOGGER is not detected or if data saving fails.
(**) See note(*) for instructions No warning appears.
Sets the unit for pressure
See note (*) for instructions.
(**) Sets the unit for temperature
(*) Enables automatic saving of the cycle report on the USB device
(*) Enables automatic printing of the cycle report.
(*) Enables saving of a complete data set of parameters each second (takes more memory space)
(**) Enables printing the plateau temperature at fixed time steps (use the next option to set the step).
(**) Sets the time step and enables printing the plateau temperature at the set time interval.
Note (*) The current setting is displayed: press OK to enable changes and then UP/DOWN to scroll the available options. Press OK to set a new option, or BACK to exit without saving.
Note (**) This option is available for advanced users only. See the next page for instructions about how to log in as an advanced user.
23
How to log on as an advanced user
Some programmable options of the MS sterilizer can be changed only after logging in as an advanced user.
This is to prevent accidental changes or unexpected operation of the sterilizer.
Hiding a cycle program, making it inaccessible to users, is an example of option that can be accessed by advanced users only.
Use the UP, DOWN and OK buttons to browse the menu, choosing the following options
in sequence: MENU – SERVICE – CURRENT LEVEL.
The current level is now displayed. If you want to change it, press OK.
The screen as shown to the left will appear: you can now type in the advanced user password
(000123) using the UP, DOWN, BACK and OK buttons as follows:
- UP/DOWN to increase/decrease the current number (indicated by the cursor );
- OK to save the number and move to the next one; ^
- BACK to move to the previous number.
Press OK on the LAST number to confirm the password.
Press BACK on the FIRST number to abort the procedure.
After making the desired changes in the advanced user level, return to the user level by setting all numbers to zero, or switch OFF the
sterilizer and then ON again.
24
6. Running a sterilization cycle
Place the sterilization load in the sterilizer chamber and close the door.
See ANNEXES 2 and 3 on how to properly prepare and place the load.
Switch the sterilizer ON by pressing the mains switch behind the service door.
The start screen will show the default sterilization program, preceded by the cursor.
(The default cycle program can be changed by the advanced user; see Chapter 5 – Programming).
25
Running a sterilization cycle
After selecting the cycle:
- the first line of the display shows the selected cycle.
See the following pages for a description of each cycle program (temperatures, times, maximum load weights).
26
The available sterilization cycles
The sterilizer is not suitable for the sterilization of liquids, textiles and hollow instruments .
27
The available sterilization cycles
28
The available sterilization cycles
29
The sterilization cycle profile
All available sterilization cycles feature the same basic pressure profile as shown in the graph below.
The duration of the sterilization phase (or plateau time) and the sterilization temperatures differ between the various cycles.
Pressure
LEGEND
Pre-heating (this is not considered a
PHE
part of the cycle)
De-aeration (purging of air from the
PG1
sterilization chamber)
30
The “Delayed start” options
The evaporation time may be increased according to the sterilization protocol in force and to the type of load processed.
The evaporation time may be increased only after acceding as an ADVANCED USER.
When changing the evaporation time, ensure that the load is always dry at the end of a sterilization cycle in order to avoid
wicking of moisture and, potentially, microorganisms from hands, gloves or environmental surfaces.
After selecting a cycle, press UP or DOWN until the SETUP option appears and confirm (OK).
Scroll the sub-menu options by pressing UP or DOWN (the current value is displayed) and
follow the instructions in the table below.
32
Cycle in progress
33
Cycle in progress
INFO screen and menu options
While a cycle is in progress, you can view the main cycle parameters in real time.
On the “cycle in progress” screen, press UP or DOWN until the INFO option appears. Other menu items are also available at this stage.
Screen title
Cycle time
Phase time
Heating element temperature Legend of the parameters
Steam pressure displayed when scrolling the
Temperature in the chamber INFO screen.
Heating element power output
Theoretical temperature
Additional chamber sensor temperature
Mains voltage
Mains frequency
Total water injected
Water conductivity
34
Manual stop
Once a cycle is aborted, a reset phase commences to safely release any steam pressure from the
chamber. This may take several minutes. Do not switch off the sterilizer! Wait until the reset
phase is completed.
At this stage you can access some menu items by pressing UP or DOWN.
When selecting the INFO option (see picture) you can view the sterilizer parameters in real time (see previous page).
35
Manual stop
The message “LOAD NOT STERILE” means that the load is not sterile. Do not use
items on patients!
The message “STERILE CONDITIONS ACHIEVED - CYCLE INTERRUPTED” means that
the load might be wet. Wet items are for immediate use only!
Press OK to unlock the door as requested in the second line of the screen.
(a waiting message appears wile the door is unlocking)
Open the chamber door and remove the load, or repeat the sterilization
cycle.
36
End of a sterilization cycle
When a cycle is successfully finished, the “CYCLE COMPLETED” message appears on
the screen and the “Unlock door” option is preceded by the cursor.
At this stage you can press DOWN or UP until the INFO option appears; confirm INFO to access cycle parameters
for mechanical sterilization monitoring (see previous pages).
This is only possible prior to unlocking the chamber door.
Confirm (OK) to unlock the door (the cursor near the “padlock” icon disappears).
Wait the door to unlock, then open the chamber door.
WARNING! When you open the door, pay attention to possible steam
sprays: Risk of burns!
Do not touch the chamber, the inner porthole and the internal
fittings as long as they are hot.
37
7. Maintenance
Before carrying out any maintenance on the sterilizer, switch the unit OFF and remove the mains cable.
Wear personal protective equipment (gloves, goggles, etc.) during normal use of the sterilizer and when carrying out
cleaning and maintenance tasks.
Before accessing the chamber and the connected parts, be sure that the sterilizer is cold.
Follow the instructions in this chapter when carrying out any maintenance on the sterilizer.
Maintenance program
The maintenance program is outlined in the table on the next page.
It includes the replacement of certain wearing parts (consumables) which is imperative to ensure the safe and faultless operation of
the sterilizer.
Maintenance counters
The sterilizer keeps track of the age of consumables by keeping memory of the number of cycles
executed since the last replacement.
When one counter reaches the maximum, a replacement message appears on the screen and the
consumable needs to be replaced; replace the consumable.
38
Maintenance program
39
Monthly or 50-cycle maintenance
Clean the chamber with a damp sponge and a mild detergent solution paying
attention not to bend or damage the temperature probe inside the sterilizer
chamber. Rinse with water.
Clean the steam diffuser plate, the trays and the tray rack with a damp sponge
and a mild detergent solution. Rinse with water.
Reposition all pieces of the chamber accessories properly.
Ensure that the steam diffuser plate is correctly placed and engaged, as this is
essential for the sterilization process.
The trays, the tray holder and the steam diffuser plate may also be cleaned in a
washer disinfector.
40
Monthly or 50-cycle maintenance
1–2: Remove the filter cap at the back of the chamber (bottom/center)
by turning it counter-clockwise.
4–5–6: Insert the filter in the cap, attach the filter cap and lock it by
turning clockwise.
41
3 month or 400-cycle maintenance
42
6 month or 800-cycle maintenance
When cleaning the tanks, be careful not to touch the water level sensors. If moved from their original position, the operation
of the sterilizer could be impaired.
43
1 year or 800-cycle maintenance
Pull out the used door seal by hand (easy if seal and fingers are dry).
Carefully clean the seal seat and the chamber face side with a cotton swab moistened
with isopropyl alcohol.
First: up and down Moisten the new seal with water. This will make placement much easier!
Insert the new seal in the sequence as illustrated in the pictures to the left.
Complete the operation by evenly inserting the seal on the entire circumference;
ensure the seal does not stick out (no bumps or deformations)!
Regular service is imperative to ensure continuous and effective operation of the sterilizer.
It is recommended to carry out a general service every 4000 cycles or five years by an authorized service technician.
The service includes replacement of consumables and other important internal components, a check of the entire unit
with special care for the safety systems, and cleaning of areas and components that cannot be accessed by the user.
45
Resetting the maintenance counters
Use the UP, DOWN and OK buttons to browse the menu, choosing the following options in
sequence: MENU – DEVICE INFO – SERVICE COUNTERS
A confirmation request appears: scroll the answer to YES by pressing UP or DOWN and
then confirm with OK.
46
8. Troubleshooting, alarms and messages
If the cursor appears to the left of one or more icons, there is related information pending.
All messages can be viewed by means of the MESSAGES sub-menu.
Use the UP, DOWN and OK buttons to browse the menu, choosing the following options in
sequence: MENU – MESSAGES.
If there is more than one message pending, you can scroll within messages with UP or
DOWN.
The icon that is preceded by the cursor is related to the pending message.
The cursor that precedes the icon disappears as soon as the relevant message has been read and the condition that
gave rise to the message has been fixed.
The cursors that precede the message icons are not visible while a cycle is in progress.
47
Messages
The water level inside the clean water tank is below the minimum. Fill the clean water tank.
The water level inside the used water tank is at maximum level. Drain used water tank.
Don’t touch the chamber or the load with bare hands: high temperature, risk of burns!
The bacteriological filter needs to be replaced.
The dust filter needs to be replaced.
The door seal needs to be replaced.
The 4000 cycle overhaul needs to be performed. Call for service.
The 20000 cycle overhaul needs to be performed. Call for service.
The CPU board battery needs to be replaced. Call for service.
PC/Logger not detected (disconnected or not powered).
Cycle report printer configured but not detected (disconnected or not powered).
File saving error (check presence and connection of the USB drive).
Error in the USB communication (between the main board and the USB4 board)
Error in saving the HTML cycle report
Error in saving the SCL cycle report
Bad USB device
NOTE: for any message not listed in this table, call service.
48
Alarm stop
In case certain important sterilization parameters are not met, the sterilizer will generate an alarm code
and abort the cycle automatically.
The sterilizer enters into a reset phase; a wait message and an alarm code are displayed on the screen.
At this stage select and confirm “Info” to view the sterilizer parameters (see Chapter 6 of this manual).
Do not switch off the sterilizer: It will take some time (several minutes) to reset the system and reach safe conditions in the sterilizer
chamber before it is possible to open the sterilizer door and remove the load.
Alarm end
When the reset phase is over, you will be asked to press BACK to get to the
“Unlock door” option.
49
Alarms
NOTE: for any alarm not listed in this table, call technical service.
50
Alarms
NOTE: for any alarm not listed in this table, call technical service.
51
Resetting the thermal overload
52
Troubleshooting
53
Troubleshooting
Before sending the sterilizer for technical service, remove the mains cable, empty both water tanks and use the original or
appropriate packaging.
54
9. Recycling and disposal
MS sterilizers are mainly built from fiber-reinforced polymers, metals and electronic components.
In case of disposal:
- separate the various components according to the materials they are made of;
- drop the sterilizer with a company that specializes on the recycling of related products;
- do not abandon the sterilizer in unsecured places;
- always refer to current/applicable laws and rules in the country of use.
55
ANNEX 1. Technical data
56
ANNEX 2. Maintenance of dental instruments
External disinfection
This procedure reduces the risk of infection during cleaning and maintenance of the instrument.
Wear protective gloves during disinfection.
Refer to the instructions of the instrument manufacturer.
Avoid using abrasive disinfectants, pH must be between value 2.5 and 9; no chlorine based disinfectants.
Residual disinfectants on instruments can cause extensive damage to your instrumentation during sterilization (oxidation, alteration
of technical characteristics of seals, rubbers, fiber optics, etc.).
External cleaning
This procedure involves the removal of residues (blood, dentine, etc.) that adhere to the surface of the instruments.
Wear protective gloves during cleaning.
Refer to the instructions of the instrument manufacturer.
Use a soft, damp brush and take care not to scratch the surface of the light ports.
57
ANNEX 3. Sterilization load preparation
58
Loading the chamber
Tubes
Rinse, drain and dry tubes after washing.
Place tubes on a tray allowing the ends to remain open. Do not bend tubes.
Paper side up
Wrapped/bagged items
Place the bags on trays allowing adequate space in-between bags.
Ensure that packs do not touch the sterilizer chamber walls.
Place sterilization bags with the paper side facing up.
Never place the load or the trays directly into the chamber without the chamber rack as
Space between bags this could affect the steam and temperature distribution. The load must always be
supported by the chamber rack.
Before initiating a sterilization cycle, always check that the steam diffuser plate is properly
positioned.
An improper positioning of the steam diffuser plate could result in bad steam
quality and could impair the sterilization process, with risk of non sterile load and
cross infection.
Sterility at the end of the cycle is not guaranteed if the steam diffuser plate was
not correctly placed.
59
ANNEX 4. Water quality
The sterilizer uses distilled or demineralized water to generate steam for the sterilization process.
The table below lists the maximum content of minerals and the specifications for the water used for steam sterilization (see EN13060 ANNEX C).
Aluminium tray
Printer model S'Print part n. F523205x
part n. 19721108
Tray holder
part n. F523001x
USB pen drive
part n. V000004x Funnel
part n. F540903x
Door seal
part n. F460504x
Permanent drain kit Wall spacer
part n. X051052x part n. F190107x
Safety bracket kit
Drain tube kit with fittings part n. X051019x
part n. A812110x
Drain tube
part n. S230900x
External water
filtration system
USB Osmo
S’Print
drive
64
Accessories and spare parts
190x43x300/375 *
Standard chamber rack for 3 cassettes (*)
190x43x300/375 *
part n. F523009x
190x43x300/375 *
205x35x300/375 *
Standard chamber rack for 3 USA size cassettes (*)
210x35x300/375 *
part n. F523021x
205x35x300/375 *
190x32x300/375 *
190x70x300/375 *
Standard chamber rack for 2 implant cassettes (*)
190x70x300/375 * part n. F523017x
(*) All racks shown in this page, if rotated
90°, accept 5 standard aluminium trays.
65
Consumables
66
ANNEX 7. W&H Installation check-list for sterilizers
1. Was the head of the clinic/practice present during all the in-service? Yes No
2. Is the packing of the sterilizer undamaged? Yes No
3. When unpacked, is the sterilizer undamaged? Yes No
4. Are all the contents of the package available (sterilizer ship-with)? Yes No
5. Are all the ordered accessories available with the sterilizer? Yes No
6. Have you removed all the protection covers from the sterilizer and from all the ship-with? Yes No
7 Were all sections of the instruction for use of the sterilizer covered during the in-service? Yes No
8 Is the allocated countertop for the sterilizer, levelled and flat? Yes No
9 Are the recommended ventilation conditions, in the allocated area for the sterilizer, complied with? Yes No
10 Are the required minimum clearances complied with? Yes No
11 Have you explained the water quality required for the use of the sterilizer? Yes No
12 Have you shown to the head of the clinic/practice the procedure for filling and draining the main and used water tanks? Yes No
13 Have you shown to the head of the clinic/practice how to program the sterilizer? Yes No
14 Have you shown to the head of the clinic/practice the cycle options? Yes No
15 Have you shown to the head of the clinic/practice what the Messages and Alarms mean? Yes No
16 Have you shown to the head of the clinic/practice how to manually abort a cycle? Yes No
17 Have you shown to the head of the clinic/practice the maintenance program and procedures? Yes No
18 Have you shown to the head of the clinic/practice how to use all of the accessories? Yes No
19 Have you shown to the head of the clinic/practice the advantages of having a USB connection for a pen drive? (if available) Yes No
Have you suggested to the head of the clinic/practice to periodically backup the data, stored on the USB pen drive, on another
20 Yes No
safe support? (if available)
67
W&H Installation check-list for sterilizers (continued)
Have you shown to the head of the clinic/practice the advantages of having a Ethernet connection (remote data saving)? (if
21 Yes No
available)
22 Have you run a sterilization cycle (134°C) with the tray rack and trays inserted? Yes No
23 Have you explained to the head of the clinic/practice the correct load type for each available sterilization program? Yes No
24 Have you shown to the head of the clinic/practice how to prepare and place the load in the sterilizer chamber? Yes No
25 Have you explained to the head of the clinic/practice to use only original parts and accessories on the sterilizer? Yes No
26 Have you shown and explained to the head of the clinic/practice the SAFETY ADVICE section? Yes No
Purchased From
Installed by
Dr./Clinic Name
Address
Phone
W&H UK LIMITED, 6 Stroud Wood Business Centre, Park Street, St.Albans, Herts AL2 2NJ
t +44 1727 874990 f +44 1727 874628 [email protected]
A-DEC AUSTRALIA CO.INC., Unit 8, 5-9 Ricketty Street, Mascot NSW 2020,
t +61 2 83324000 f +61 2 83324099 [email protected]
Ivoclar Vivadent NZ, P.O.Box 5243, Wellesley Street, 12 Omega Street, Auckland,
t +64 9 914 9999 f +64 9 914 9958 [email protected]
Manufacturer Iso 13485,
93/42 EEC – Annex II
W&H Sterilization S.r.l.
Italy, I-24060 Brusaporto (Bg), via Bolgara, 2