Precision RXi System Installation Manual
Precision RXi System Installation Manual
Precision RXi System Installation Manual
Technical
Publication
Direction 5123446-100
Revision 11
Precision RXi System Installation Manual
42011G
GE Medical Systems
Blank page
2. Introduction............................................................................................................ 19
2.1. Applicable standards .............................................................................................................20
2.1.1. Safety .................................................................................................................................20
2.1.2. EMC (IEC 60601-1-2:2001/ EN 60601-1-2:2001 / CEI EN 60601-1-2:2003) .................20
2.1.3. Electromagnetic Compatibility (EMC)..............................................................................21
2.2. Tools and materials needed but not shipped with the product..........................................21
General Medical MERATE SpA can modify / update the present documentation without notice.
Copying and divulgating the present documentation is forbidden without a written General Medical
MERATE SpA authorization.
Revision History
WARNING: THE REMOTE TABLE AND GENERATOR ARE CONFIGURED AND ADJUSTED IN
MANUFACTURE. IN ORDER TO MAKE EASY THE PACKING PROCEDURE SOME
INSTALLATION CODES ARE MODIFIED.
THESE CODES MUST BE RESTORED AT THE EQUIPMENT START UP.
IN ORDER TO RESTORE THESE CODES PLEASE MAKE REFERENCE TO THE PROCEDURE
IN INSTALLATION MANUAL 5123446-100 chap. 6.21 page 190 and chap. 6.22 page 191
1.5. Adjustment
Warning: fill in the following items just in case of a system with an Automatic Collimator
1.5.1 Adjustment
WARNING: IN CASE OF OVERHEAD TUBE SUSPENSION (OTS) ADJUSTMENT WITH
AUTOMATIC COLLIMATION ON WALL STAND
□ MAKE REFERENCE TO
Task and procedures manual 5121726 –100 Book 3 chap. “OTS automatic collimator”
Make Over Head Tube Suspension (OTS) Focal Distance adjustment following procedure in Task and
□ procedures manual 5121726 –100 Book 3 chap “Overhead Tube Suspension(OTS)”
Do the collimator mounted on OverHead Tube Suspension (OTS) ADJUSTMENTS following the
□ procedure in
Task and procedures Manual 5121726 –100 Book 3 chap “Overhead Tube Suspension (OTS) ”
1.7. Covers
1.7.1 Covers
For equipment covers procedure make reference to
□ Installation Manual 5123446-100 chap. 6.20 page 180
2. Introduction
This document has the purpose to collect the main information to allow the installation of the Precision
RXi system.
Warning
THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE
EXPOSURE FACTORS, OPERATING INSTRUCTIONS AND MAINTENANCE
SCHEDULES ARE OBSERVED.
Although this equipment was designed and manufactured according to the most up-to-date safety
Standards, the X-ray beam always represents a danger, should not the operator be properly qualified
and trained.
Excessive exposures to X-rays cause damages to the human body: therefore, all the necessary
precautions must be taken to prevent unauthorized or unskilled personnel from operating this
equipment, thus jeopardizing themselves and other people.
Before executing any operation, the person being qualified and authorized to operate this equipment,
must be informed about the protection and safety measures established by the International Committee
for the Radiological Protection, as well as any other relevant national Standards.
For a proper use of this equipment, it is required that the operator should previously and carefully refer
to the present Operator’s Manual.
A special attention should be paid to the sections “Safety Recommendations”, “Protection Measures”,
“Functioning and Safety Controls”.
Note:
This equipment Is in compliance with the requirements established by the European Directive 93/42
CEE applying to medical devices, thus it is provided with the EC mark stating the code number
identifying the notified body responsible for carrying out the procedures foreseen for the compliance
certification.
The Manufacturer: General Medical Merate via Partigiani, 25 – 24068 Seriate (BG) Italy.
E-mail [email protected] Tel. (+39) 035 4525311
The Manufacturer (GMM) and the Distributor (GE Medical Systems) shall decline all
responsibility all responsibility with regard to the good functioning of the equipment should the
installation and/or the maintenance of it be executed by unauthorized personnel.
2.1.1. Safety
The Precision RXi remote R&F system comply with the following regulatory requirements and design
standards:
2.2. Tools and materials needed but not shipped with the product
Tools needed for installation are:
• Drill machine with tips (millimeters tip will be better)
• Allen keys in millimeters (NOT INCHES)
• Wrench keys (for hexagonal head screws) in millimeters (NOT INCHES)
• Set of screwdrivers (French and crossed head)
• Level tool
• Measuring tape
• Digital voltmeter and oscilloscope
• Scaffold (plateform) and ladder to reach ceiling suspension.
• GenWare® and a laptop computer for setup and calibration if required (for analogical room
version only)
• A calibrated radiation meter with detectors that will allow for R/min and µR type
measurements (or µGy and Gy/min).
• A strobe or reed type tachometer to verify that the anode is rotating up to speed.
• A sufficient selection of patient absorbers to allow AEC and ABS calibration. A suggested
selection is 3/4 inch Al (quantity 2); 1 mm of Cu (quantity 8), Water in containers of 5.0.
10.0, 15.0 cm thickness.
• Test phantoms to verify the imaging system with the generator.
The pertinent national regulations must be fulfilled in each relevant country. Anyway, it is strongly
recommended that the compliance with the above-mentioned prescriptions should be observed, unless
these ones are contradictory to the national laws in force for the safety of the operator, patients and
third parties.
Monitor
Ceiling stand Monitor Wall bucky Table
suspension
230 (220/240) 230 Vac 220/240Vac
Supply voltage Na 100 / 240 Vac
Vac +6% -10% ±10%
Mains frequency 50/60Hz ±0.5 Hz Na 50/60 Hz 50 / 60 Hz 50/60Hz
Max. current Na 1.5 / 0.75 A rms 0.16 A 15 A
300 VA (lamp of
Operating power diaphragm 100 W Na 55W (max.) 40 VA (approx.) 3,5 kVA (approx.)
included)
Operating
10 to 40 °C. 10 to 40 °C. 10 to 40 °C 10 to 40 °C
temperature
Operating
Humidity (non 30% to 75% 30% to 75% 30% to 75% 30% to 75%
condensing
Operating
70 kPa to 110 kPa 50 kPa to 106 kPa 70 kPa to 106 kPa 70 kPa to 106 kPa
pressure
88 x 120 x 111 62 x 222 x 39 or 210 x 165 x 80
Max component 440 cm rails 415 x 340 x 83
cm (W x H x D) 65 cm cm (W x H x D)
dimensions length mm (W x H x D)
extensible (W x H x D) (table on jacks)
Rails and
Max component transversal bridge 75 kg (including
6.2 kg 170 kg / 200 kg 660 kg / 790 kg
Weight 175 kg; two monitors)
stand 140 kg
Analogical
Analogical TV Analogical TV
system B&W UPS
chain 512 chain N33
monitor
Supply voltage 24Vdc ± 20% 110 / 240 Vac 24Vdc ± 20% 80 / 138 Vac
Mains frequency Na 50 Hz Na 50 / 60 Hz
Output voltage:
Max. current 0.83 A 450 / 250 mA 0.83 A
100/110/120 Vac
Operating power 20W 50W (approx.) 20W 2kVA 1.6kW
Operating
10 to 40 °C 0 to 40°C. 10 to 40 °C 0 to 40 °C
temperature
Operating
Humidity (non 30% to 75% 10% to 80% 30% to 75% 95%
condensing
Operating
70 kPa to 106 kPa 67 kPa to 106 kPa 70 kPa to 106 kPa ---
pressure
Max component 30.2 x 36.7 x 6 37.3x14x42.4 cm
22.6x12x4.6 cm 23.5x19.2x6.5 cm
dimensions cm (W x H x D) (W x H x D)
Max component
0.55 kg 4 kg 1.5 kg 30 kg
Weight
April2008 Page 23
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
OPERATING
Ambient temperature range: 10 to 40 °C.
Relative humidity: 30% to 75%
Atmospheric pressure range: 70 kPa to 110 kPa
LCD type: Totoku 18.1 inch (46 cm) monochrome LCD TFT
Power: 110 / 240 Vac; 50/60 Hz; 2.0 / 1.0 A
Power consumption: Normal operation 60W (approx.);
Power management state less than 10 W
OPERATING
Ambient temperature range: 5 to 35 °C.
Relative humidity (non dewing): 30% to 80%
Atmospheric pressure range: 701 hPa to 1013 hPa
April2008 Page 25
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
WS-T and WS
OPERATING
Ambient temperature range: 10 to 40 °C.
Relative humidity: 30% to 75%
Atmospheric pressure range: 700 hPa to 1060 hPa
Complete with grid and pre-set for measure chamber of automatic exposure control.
Complete with grid and pre-set for measure chamber of automatic exposure control
WS
bucky vertical travel 136.5 cm
movement mode/braking manual/mechanical
balancing mode counterweights
minimum distance from floor of the bucky
67.3 cm
upper edge
weight 170 kg
WS-T
bucky vertical travel 136.5 cm
movement mode/braking manual/mechanical
balancing mode counterweights
bucky tilting -20° e +90°
bucky rotation on its axis 90°
minimum distance from floor of the bucky
64.8 cm
upper edge
minimum distance from floor of the bucky in
69.3 cm
horizontal position
weight 200 kg
BUCKY
cassette formats from 13x18 to 35x43 cm
April2008 Page 27
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
BUCKY
grid characteristics 90 lines/inch R=12:1 F= 180 cm
STANDARD ACCESSOIRES
bucky with cassette sensing
measure chamber of automatic exposure control
NORMATIVE REFERENCES
standard compliance EN 60 601-1, 60 601-2-32
classification (IEN/ CEI EN 60 601) class I – type B
3.4. Positioner
Important Note:
Normally the power supply of the remote-controlled table is taken by a 16A, delayed 400Vac nominal,
placed in the mains supply panel board of the exam room, with intervention differential current not
higher than 30 mA in compliance with IEC/CEI Standards.
In case the electrical distribution panel isn’t preset with an ON/OFF switch, the switch must be
accessible to the operator, to make him able to switch ON/OFF the table
In compliance with UL/CSA the equipment must be supplied by two delayed fuse of 16A – 400Vac
with an ON/OFF switch placed in the mains supply panel board of the exam room,
The earth terminal of the equipment must be connected in a visible manner to the earth equipotential
node of the diagnostic system by means of a protection conductor, with the minim section of 10 mm2.
The supply cable from the mains supply panel board to the equipment electrical cabinet must have a
minim section of 2.5 mm2
All the conductors must be made in copper. All the supply connections and earth cables must be in
compliance with the national rules or equivalent.
April2008 Page 29
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
3.4.2. Positioner environmental characteristics
OPERATING
Ambient temperature range: 10 to 40 °C.
Relative humidity: 30% to 75% no condensation.
Altitude: 700 to 1060 hPa
3.4.3. Weights
(1)
The weight of the control panel is given by the weight of the control panel itself plus the weight of the support base in the simple T version.
(2)
The weight of the power cabinet is referred to the complete version.
3.5. Generator
April2008 Page 31
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
Room
Interface
Transformer
Referring to the following figure, move the line voltage tap from the 400 VAC position to the 480
VAC position.
Referring to the following figure, change the line voltage taps to the “480 VAC” position. Because
the primary of this transformer consists of two 240 V windings in series, BOTH leads shown
in figure need to be moved from the 200 V taps to the 240 V taps.
April2008 Page 33
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
Switch the generator mains power OFF, AND WAIT 5 MINUTES FOR THE DC BUS
CAPACITORS TO FULLY DISCHARGE.
STEPS 2 AND 3 APPLY TO MILLENIA GENERATORS ONLY. IN INDICO 100
GENERATORS, THE DUAL SPEED STARTER BOARD IS FULLY ACCESSIBLE WITH
THE APPROPRIATE GENERATOR SIDE PANEL(S) REMOVED.
Loosen the two nuts securing the clamping bracket at the top of the dual speed starter. Slide the
bracket up and gently remove the dual speed starter assembly. The cables connected to the
dual speed starter do not need to be removed in order to change the EPROM.
Carefully rotate the dual speed starter chassis to fully expose the dual speed starter board inside the
unit.
Locate and carefully remove the existing EPROM on the dual speed starter board (U26, refer to
figure 1E-6).
Carefully insert the replacement EPROM into the socket observing the orientation per figure 1E-6.
Minimum Minimum
Minimum
Recommended Generator Recommended Apparent
Mains Recommended Ground Wire
Mains Disconnect Momentary Distribution Mains
Voltage Generator Size
to Generator Line Current Transformer Resistance
Service Rating
15 ft/5 m max) Rating
65 kW 3∅ Generator
400 VAC #6 (15 mm2) 125 A 100 A 85 kVA #4 (25mm2) 0.13 Ω
480 VAC #6 (15 mm2) 105 A 100 A 85 kVA #4 (25mm2) 0.19 Ω
80 kW 3∅ Generator
400 VAC #6 (15 mm2) 155 A 100 A 105 kVA #4 (25mm2) 0.11 Ω
480 VAC #6 (15 mm2) 130 A 100 A 105 kVA #4 (25mm2) 0.15 Ω
All wiring and grounding should be in compliance with the national electrical code or equivalent.
All wiring must be copper.
For all installations, a separate copper ground cable #4 AWG (25 mm2) is required from the
building distribution ground to the ground terminal located inside the main disconnect switch
fuse block.
The disconnect switch shall be located within reach of the operator.
April2008 Page 35
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
OPERATING
Ambient temperature range: 10 to 40 °C.
Relative humidity: 30% to 75%.
Atmospheric pressure range: 500 to 1060 hPa (375 to 795 mm Hg).
3.6. Console
OPERATING
Ambient temperature range: 10 to 40 °C.
Relative humidity: 30% to 75% no condensation
Atmospheric pressure range: 500 hPa to 1060 hPa
April2008 Page 37
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 EQUIPMENTS DATA AND CHARACTERISTICS
OPERATING
Ambient temperature range: 0 to 40 °C.
Relative humidity: 10% to 80% no condensation
Atmospheric pressure range: 670 hPa to 1060 hPa
NOTES: Power must be from a dedicated A.C. line and be free of noise, spikes, surges, and
brownouts, which exceed the nominal voltage by +10%. If these conditions cannot be met, the
optional Power Conditioner is required.
OPERATING
Ambient temperature range: 18 to 30 °C.
Relative humidity: 20% to 80%, no condensation
Atmospheric pressure range: 0 to 3000 m (≈650 to 1020 hPa)
April2008 Page 39
GE MEDICAL SYSTEMS PRECISION RXI SYSTEM INSTALLATION MANUAL
DIRECTION 5123446-100, REVISION 11 CEILING STAND INSTALLATION
READ THIS MANUAL CAREFULLY in order to learn how to operate and service this equipment
correctly and to avoid personal injury or unit damage. This manual and the safety labels on the
equipment may also be available in other languages (for information contact the manufacturer).
The complete documentation which is shipped with the unit, and this manual in particular, must be
carefully preserved for future references or necessities.
MEASURES in this manual are exclusively expressed in the metric system.
THE EQUIPMENT RIGHT AND LEFT SIDES are established by the user, facing the unit (as shown
in this manual cover).
THIS UNIT IS DESIGNED to be exclusively used during the normal X-Ray operations as pointed out
as INTENDED USE. Use in any other way is considered as contrary to the intended use.
The manufacturer declines all responsibility for damage or injury caused by a use which differs from
the intended use; the user is the only responsible party of these consequences. The careful observance
of the functioning, maintenance and repair instructions, which are pointed out by the manufacturer, are
also an essential part of the equipment use.
THE RADIOGRAFIC STAND, object of this manual, MUST BE USED, OVERHAULD AND
REPAIRED EXCLUSIVELY BY PERSONNEL that is qualified and familiar with all its
characteristics and acquainted with the relevant safety rules (accident prevention). Laws regarding
accident prevention and all industrial safety rules must always be observed.
The manufacturer is not responsible for damages or injury caused by any arbitrary modification carried
out on this unit.
4.1.2. Tests
The equipment testing has been performed by the manufacturer with procedures that are to verify that
the equipment, subject of this manual, is without faults and that its working features comply to the
ones foreseen for this type of use.
This kind of test is exclusively carried out by qualified personnel and its result is specified on the
inspection report. The manufacturer keeps, in conformity with law, the only report copy every unit has.
4.1.3. Guarantee
Guarantees are supplied only in accordance with the contract conditions.
Damages which are caused by incorrect equipment use or by non-observance of the rules in this
manual will not be considered under guarantee.
ATTENTION! The guarantee is automatically voided if interventions on the unit are made by non-
authorized personnel.
4.1.5. Labels
On the unit there must be:
• part number label;
• wire ropes replacement label;
• serial number label;
• Warning and/or danger labels
Their shape and placement is shown in the drawings on the following pages.
Do not remove the adhesive labels on the unit: they are an integral part of the instructions and of the
documentation that accompany the equipment
4.1.6. CE Conformity
The equipment subject of this manual, is built craftsman like and in conformity with the EN 60 601
safety regulations for the electromedical equipment. The equipment complies with the European
Directive 93/42 relating the medical devices of class 1 (annex VII).
The CE mark is shown on the adhesive label that identifies the equipment.
The Declaration of Conformity is supplied with the documents that come with the equipment. A copy
of this Declaration of Conformity can be asked for from the manufacturer.
France
Germany
Spain
Hungary
Poland
Portugal
Romania
Russia
Turkey
Holland
Slovakia
it is necessary to replace the safety labels as shown in PET23GE01 drawing (following page) with the
ones included in the product documentation.
The labels are attached on three sheets, that is to say one for each type of safety label: each sheet
includes the different translations for the countries listed above.
Proceed to select the applicable language making reference to the code marked in the following image
and to select the label to be applied based on the spoken language in the installation country.
101604-EN
101614-EN
101594-EN
All units are manufactured following the latest technical standards in safety matters. In any case,
their use can be dangerous, if they are not correctly used by untrained personnel or if they work
differently from the projected aims.
In particular, refer to the Operator Manual to know the safety measures for the equipment operations.
• Instructions and rule which regard this kind of equipment and are in force in a country must
be respected by the operator and the installer.
• The unit must work only in rooms, which are used for medical aims (refer to regulations in
force).
• The unit is not planned to be used in rooms which are explosions risk.
The equipment can be used and subjected to maintenance operations only by authorized and trained
personnel. In particular, they must know the possible dangers coming from a wrong equipment use
and the residuals risk.
In order to avoid major risks to patients and operators and/or possible structural failure, AVOID
ANY MODIFICATIONS on the equipment with the purpose to increase its travels, performances
and/or load capacity.
Refer to the drawings and to the technical features, written in this manual, to be aware of the
performances and the maximum loads that can be supported.
• All operations to do on the unit must be carried out when it does not work. The person who is
responsible of the maintenance must completely disconnect the unit, before the adjustment or
replacement of an equipment part.
• Keep hands, feet, dresses, jewels and hair at safety distance from any moving part in order to
avoid their getting caught.
• Keep all parts in good condition and correctly installed. Repair immediately damages,
replacing the consumed and broken parts.
Do not change any part of the unit; changes which are not authorized can compromise the equipment
functioning and safety. They void the manufacturer’s guarantee as well.
In particular, it is not possible to make inoperative or replace devices and/or circuits which permit
the radiations emission in accordance with the safety laws.
• After the maintenance, it is necessary to check the correct mounting and functioning of all
electrical and mechanical devices before make the unit operative again.
The unit mobile parts movement must be carefully done in order to avoid collision risks and / or
damages to people in the examination room or to the unit functional integrity.
When the motorized vertical movement is used, it is necessary to avoid to create situations of conflict
of the telescope mobile parts against any obstacle and not to insist on the controls when the end of
travels have been reached.
In order to avoid major risks for patient sand/or damages to the equipment, the movements must be
effected using exclusively the proper control devices, make sure to stop the movements before
releasing the grip.
The unit can not be subjected to damaging stresses, such as the violent stop of the mobile parts on the
end of travel bumpers, the excessive torsion of the electrical function cables, the forced movements of
the unit without releasing the brakes, etc.
WARNING! All patients parts which are at risk of radiation exposure, must always be protected by
additional shielding devices according to the radiation safety law.
Contact the manufacturer in order to use alternative mounting systems or for different dimensional
needs.
The breaking of the main rope is the consequence of missed periodical checks and maintenance; the
telescopic stand, being suspended to the ceiling, requires very careful maintenance of the suspended
parts in order to avoid potential dangers for the operators and to assure the correct functioning of
the equipment itself.
Never use the ceiling stand after the safety device intervention.
DO NOT TRY TO FORCE THE RADIOGENIC TUBE TOWARDS THE BOTTOM WHEN THE
MAIN ROPE IS BROKEN.
Immediately contact technical assistance and replace the broken rope.
On the carriage, which is made of a rigidly bolted heavy steel plate, the motor springs for the units
balancing are mounted; it is possible to hold one or more reserve springs as well. Its innovative
conception optimizes all available space and permits to hold up to 10 springs, in contrast to the 7 or 8
which are generally used, guaranteeing a maximum balanceable weight of 75 kg.
The telescope is made of extruded aluminium; it has a light weight and is not susceptible to corrosion;
it guarantees at the same time the required mechanical strength and durability.
Moreover, the light weight makes all aerial structures lighter; these are mechanically less stressed and
requires limited maintenance.
Contact the manufacturer to inquire about the availability of other accessories (not included in the
chart) and/or to ask for specific solutions.
To use the lifting device, place the metal pallet that the telescope is mounted to under the bridge, then
continue with the following procedures:
1) Remove the carriage rear ball bearing supports (93CSC3 – dis. Mi930031-b – Pos. 3);
2) Remove the front ball bearing supports from the carriage, right (93CSC1 - dis. Mi930031-b –
Pos. 2) and left (93CSC2 - dis. Mi930031-b – Pos. 1);
3) Place the lifting bar under the bridge and fasten it by the two screws that are part of the supply
(TCCE M8x12 - dis. MI930030-B – Pos. 7) to the proper holes made in the carriage
shoulders;
4) Remove the closure head of the bridge dis. Mi930031-b – Pos. 4);
5) Insert in the bridge rails the carriage supports previously disassembled, paying attention to the
right insertion direction;
6) Insert the two threaded tie rods (dis. MI930030-B – Pos. 9) in the bridge rails and lock them
in position by the nuts and the screws that are part of the supply (dis. MI930030-B – Pos. 1,
2, 3, 4, 5);
7) Insert the other end of the tie rods (dis. MI930030-B – Pos. 9) inside the proper holes that are
on the lifting bar device (dis. MI930030-B – Pos. 12);
8) Using the two knobs (that are to be screwed from the bottom on the threaded tie rods – dis.
MI930030-B – Pos. 14), lift the carriage up to the bridge height by alternatively acting on the
two knobs;
9) Once the right point is reached, fasten the carriage ball bearing supports to the carriage again
and then proceed with completion of the system installation.
WARNING!!!
Before removing the lifting device, check that the support bearings are correctly inserted, in the
guides fastened to the rails, and that they are properly fastened to the carriage. This safety check is
very important to avoid potential risks to patients and technical staff that may result from carriage
support failure
4.5. Installation
Before using the equipment, the operators should read the par no. 0 about the SAFETY MEASURES.
The equipment must be installed and can be maintained only by authorized and trained personnel.
Specifically, they must be informed about the possible dangers caused by improper equipment use
and the residual risks.
ATTENTION!!!
The installation of hanging units or pars can be very dangerous; only the necessary personnel for the
installation must be in the room during this operation. All other people must be sent out.
• The manufacturer assumes no responsibility in case of faulty unit operation which can be
caused by negligence or elusion of the basic maintenance rules listed below.
• All ordinary or extraordinary operations on the equipment must be registered, documented
and the information contained must be readily findable.
Maintenance and documentation duties must be observed; anything contrary to the case, the
equipment manufacturer responsibilities cease automatically.
In order to avoid major risks for patients and operators and/or dangerous structural limitations,
AVOID ABSOLUTELY modifications on the equipment with the purpose to increase its travels,
performances and/or loads.
Refer to the drawings and to the technical features, written in this manual, to be aware of the
performances and the maximum loads that can be supported.
Before the installation, the unit can be kept in the warehouse for some time with its original
packaging. In any case, although every metal part is carefully covered by protection paint or
oxidation superficial treatment, the equipment is not planned to work or be stocked outdoor. It cannot
withstand the humidity; for this reason, do not keep it outdoor or in extreme environment conditions.
It is absolutely forbidden to pass by or under the load during the lifting operations.
Make sure the cases are well fixed during the transport to avoid blows which could damage their
contents.
Pay attention to the moving hanging loads in order to avoid dangers to people or other things.
The wooden cases are made to be lifted through a lifting device (forklift truck). If they must be lifted
from the top (crane), it is necessary to sling them through suitable ropes, which must twine round the
case.
Refer to the instructions outside the cases for their moving and lifting.
The CASE No. 1, whose dimensions are 460x95x55 cm (gross weight, about 260 kg), contains:
• the longitudinal rails with the closing crossbars and the screws sets for the rails fixing to the
support structures;
• the complete transversal bridge;
• lift device (if available)
• the cables rail (to be placed externally to one of the longitudinal rail);
• the box with the cables hoses; the large carriage cover (in two parts); the tube cover.
The CASE No. 2, whose dimensions are 100x75x140 cm (gross weight, about 230 kg), contains:
• the telescopic stand mounted on a disposable metal structure;
• the collimator (if supplied);
• a cardboard box too, which contains the control board, the accessories and the finishing parts
of the telescopic stand.
The metal structure makes easier not only the stand moving on the ground through a carriage, but also
its lifting and inserting on the sliding rails of the transversal frame. This operation is much easier if the
proper device (which is supplied on demand) is used for the stand lifting.
Index contains of the ceiling suspension case (picture 2 on the following page).
ATTENTION.
The static load which acts on each screw in the vertical direction is equal to 2000 N, while the stress
in the horizontal direction is of about 150 N for each screw.
For mounting methods which differ from the suggested ones, or for heavier suspensions, this value
could be higher or lower.
The supporting surfaces of the longitudinal rails props must be parallel and carefully leveled (± 2
mm on the length). Possible minimum differences can be corrected during the mounting phase,
putting the suitable thickness between the rails and the supporting surfaces.
The more common fixing methods of the rails to the ceiling are listed below.
ATTENTION!!!
The measuring of the ceiling stand support structures is exclusively under the competence of the
designer who is responsible of the rooms preparation. Kind and dimension of the I-beams are shown
in our drawings just as an indication.
4.6. Mounting
4.6.1. Equipment mounting.
In order to avoid major risks to patients and operators and/or possible structural failure, AVOID
ANY MODIFICATIONS on the equipment with the purpose to increase its travels, performances
and/or load capacity.
Refer to the drawings and to the technical features, written in this manual, to be aware of the
performances and the maximum loads that can be supported.
It is important to check the rails parallel to each other and horizontally; If needed, shim where
necessary. Moreover, check, using a meter, that the rails are parallel because an incorrect mount
could compromise the correct sliding of the transversal bridge.
To reduce the exercise backlash of the stand, it is advised to mount the bridge so that its guide
bearing (i.e. the ones with vertical axis) will be where the stand telescope will be mounted.
ATTENTION. To make easier the insertion the transversal bridge guiding bearings, it is better to
remove the RIGHT and LEFT BRACKETS (3.4 and 5.4) from the bridge crossbars (4.4) and to insert
the guiding bearing brackets from the longitudinal rails head (1.4).
When this operation is finished, it is possible to put the bearing brackets back in position.
5) Check distances and alignment of the transversal bridge in respect to the longitudinal rails,
before locking the loosened crossbar.
ATTENTION!!!
Check that the support and sliding bearings (2.5 )are correctly inserted in the guides fastened to the
rails before removing the lifting device. This check is very important in order to avoid dangers that
might come from it for the safety of the patient and technical staff.
It is possible to immediately regulate the bearings backlash or postpone this operation when
mounting is over.
To regulate the bearings tightening, use an Allen wrench and act only on the contrast bearing (the one
placed under the rail) to move the eccentric one on which the bearings are mounted (3.5).
ATTENTION. Check that the contrast bearings are not closed against the guides and the bridge
slides along the rails correctly and freely from frictions. In contrary case, correct the bearings
adjustment (3.5).
Therefore, regulate the distance of the brakes by acting on the fastening screws (4.5) and leaving a
distance of 0,8 – 1 mm between the brake and the steel plate fastened to the rail.
When the operation is over, mount again the covers of the bridge cross bars (8.5) and, in
correspondence of the cover with the opening in the central door, mount the support plate of the
optional micros (11.5). This plate is sent dismounted in the accessories box (17.2), to avoid damages
during the transport.
The centering cams for detent (10.3A) are already complete with the screws and the fastening square
nuts, and must be inserted in the lower channel of a rail of the bridge (in correspondence of the detent
mounted on the carriage of the telescopic stand) and in the lower channel of one of the longitudinal
rails (in correspondence of the detent mounted on the bridge cross bar).
Their position can be adjusted later and modified in any moment (also when mounting is over) thank
to the possibility to make the centering cams sliding along the whole rails and bridge length. The
function of the two centering cams (supplied with the standard supply) is to identify a specific work
position (for example, for the alignment with a table or with a wall bucky) or to spot a parking position
(for example if, in the same room, a remote controlled equipment must work)
If you wish to mount one or more safety micro and the respective action cams, contact the
manufacturer to receive the specific instructions and the availability of the different kits of micro.
Therefore, proceed with mounting the hose support on the bridge (19.3A) by fastening it to the bridge
upper part (as shown in picture 1). The square nuts already inserted inside the rails of the bridge, allow
to mount the support in any position and on the right or left side according to the needs.
Connect then the cable coming from the brakes of the bridge to the terminal board that is in the upper
part of the support (19.3A) together with the connecting cable of the longitudinal brakes (06.3A)
contained in the box. The connecting cable must enter in the hose support upper hole, go through the
piece of rubber sheath that takes up to the carriage and then being connected to the control electronic
board.
The cables connecting operations should be postponed when the mounting is over and following
carefully what written on the electrical schematics standardly supplied and contained in the
Technical Electrical Manual
Then mount the guide for the cables sliding (07.2) to the extern of one of the longitudinal rails placing
it (using the rail lateral groove) toward right or left so to address the cables toward the generator. This
guide can be placed on the front or rear rail (always externally to the rails) according to the installation
needs.
Then insert, inside the cables rail, the 3 sliding hose support (01.3A), after removing the safety screw
at the ends of the cables guide.
The cables and support sheaths mounting will be postponed at the end of the system installation, after
completing the mechanical parts assembling.
* Supplied only if the equipment configuration foresees the assembly of two cable
** Supplied with standard ceilstand.
The cables and support hoses mounting should be delayed to the end of the installation of equipment,
after having completed the assembly of the mechanical parts.
At the last, use the supplied colored bands (contained in the accessories box 20.3 and 21.3) to spot the
directions of the equipment moving (on this subject, see the control push buttons color that are on the
control board).
Especially the green band (14.3) spots the longitudinal direction and must be applied under the front
rail (in a position that the operator can clearly see) inside the present lowering.
The blue band (15.3) spots the lateral direction and must be applied under one of the rails of the bridge
(in a position that the operator can clearly see) inside the lowering.
The yellow band spots the vertical direction and it is already applied on the main element of the
telescopic stand.
Any kind of damage must be immediately communicated to the forwarder and the manufacturer.
2) Mount the cable support assemblies (06.9 e 07.9) on the rear side of the carriage on the
internal part of the bearing support using the bolt TCCE M10x25 already inserted in the
required position (see picture 9).
N.B. The right and left sides are recognizable from the mounting impossibility on the opposite part.
3) on the contrary, postpone the longitudinal brake mounting (03.9) and of the detent (01.9); this
operation must be done only after the stand inserting into the rails. In this way, its accidental
damaging is avoided;
When the stand is on the metal structure, it can be easily moved through a lifting truck and lifted
manually (weight ≈ 130 kg) or through a lifting device which is supplied on request.
4) before the lifting, take out the closing crossbar on transversal rails head (04.7) and the
bumpers (03.7);
5) into the transversal rails (02.7) insert the stand carriage (05.7;
6) once this operation is finished, mount again the bumpers (03.7) and the closing crossbar
(04.7).
ATTENTION
Before removing the lifting device, verify that the supporting and sliding bearings are correctly
inserted into the guides, which are fixed to the rails. This control is very important in order to avoid
injuries to patients or technical personnel
It is possible to immediately regulate the bearings backlash or postpone this operation when
mounting is over.
To regulate the bearings fastening, use an Allen wrench and act on the contrast bearing only (the one
placed under the rail) to move the eccentric one on which the bearings are mounted.
ATTENTION. Check that the bearings are not tightened against the guides and that the carriage
sliding on the bridge happens correctly and free of frictions. In contrary case correct the bearings
regulation.
The balancing springs action keeps completely closed the telescope; for this reason, in order to bring
the arm at the desired height, it is necessary to extend by strength the telescope down and keep it in
position through the locking device, designed to this aim (4.8).
This device consists of a screw M12; once it is screwed in the block fixed to one of the carriage
shoulder (2.8), it enters one of the holes of the balancing springs winding drum (1.8), locking its
rotation.
At this point start the radiogenic unit mounting.
ATTENTION!!!
Follow the above described instructions to lock the rotation of the springs winding drum. Do not use
foreign matters or other systems which could be dangerous for the assembler’s safety.
Picture 9 Carriage
* Suggested tube fastening screws (refer to the technical instructions that come with the X-Ray tube)
ATTENTION. During these operations and to avoid to damage the tube, refer to the technical
instructions that come with the it.
The instructions supplied with this manual are valid only indicatively.
Between the X-Ray tube bracket and the support must be inserted the control board support bracket
(05.11 o 4.12),
On the bracket there are two rows of holes that permit to adapt the position of the control board
support fork to the diameter of the tube bracket that is to be mounted. The holes to use are those which
allow keeping the support fork at the minimum distance from the tube.
When a Varian 130B tube is mounted, it is advisable to use the second and the fifth hole (starting to
count them from the top) that are in the control box support plate (see picture below).
* Suggested tube fastening screws (refer to the technical instructions that come with the X-Ray tube)
ATTENTION
Carefully follow the mounting instructions and make sure that the collimator is correctly assembled.
Uncorrected mounting could be dangerous, it could be cause the collimator fall.
IMPORTANT NOTE: you shall check that the fixation screws used to secure the collimator are the
proper ones as follows: verify that they protrude through the flange and spacers far enough to
engage at least 6 threads (that is to say 6 mm) into the tube port face.
1) Determine the distance from the focal spot to the tube port face.
2) Subtract the resulting distance from 80 mm. and determine how many 1,5 mm spacers
combined with the 18,5 mm thickness of the mounting flange will be required to make up the
difference.
3) The outer face of the collimator-mounting flange must be at 80 mm. from the focal spot.,
where the allowable tolerance is ±1 mm.
4) Select four bolts of suitable thread (TSEI M6) and of such a length that they protrude through
the flange and spacers far enough to engage at least 6 threads into the tube port face.
FOCAL SPOT
18,5
ATTENTION. During these operations and to avoid to damage the collimator, refer to the technical
instructions that come with it.
The instructions supplied with this manual are valid only indicatively
Collimator rotated to left side Centred collimator Collimator rotated to the right side
The collimator has the possibility to rotate on the right and on the left. It is possible to lock the
collimator with a screw in the centre position in case the country regulatory do not allows the ±90°
collimator rotation.
The previous picture (on the right) shows the screw position to lock the collimator rotation feature.
200 cm
180 cm
100 cm 4
90 cm
2 3
parking
position
0 cm
example
Wall 5
Stand 1
CSA007 – Kit micro-switch for table alignment (picture 14) – for manual collimator use only
It is composed of 1 micro-switches mounted on the transversal bridge tie (07.14), 1 cam for the
activation of the micro-switch (inserted in the lower channel of a rail of the bridge) (06.14) and no cam
(3.13) to activate the mechanical detent (it is used one of the detent cam, supplied with the standard
supply, placed in the lower channel of a rail in such a way that, at the position identified by the detent,
corresponds the activation of the lateral micro-switch).
CSA028 – Kit micro-switch for table alignment (picture 15) – for automatic collimator use only
It is composed of 4 micro-switches mounted on the transversal bridge tie, 1 cam for the activation of
the micro-switch (inserted in the lower channel of a rail of the bridge) and no cam to activate the
mechanical detent (it is used one of the detent cam, supplied with the standard supply, placed in the
lower channel of a rail in such a way that, at the position identified by the detent, corresponds the
activation of the lateral micro-switch).
The longitudinal micro-switch kits are composed of the micro-switch plate of support, of the micro-
switches and of the connection cables, already connected and tested at the factory. To mount them, it
is necessary to insert the cable in the milled slot (12.5) placed inside the bridge cross bar (1.5) and
then connect the cables to the terminal board placed on the hose support on the bridge (19.3A). Then,
mount the support plate of the micro-switches in correspondence of the cover of the bridge cross bar
with the opening in the central part. The position and disposition are the ones shown in picture 5.
For the connection modalities and of electrical interfacing, refer to the electrical schematics in the
OTS Section of “System Tasks and Procedures”.
Picture 14– CSA007– Kit to mount the 1 micro-switches on the longitudinal movement
(complete of micro switch and centring cams – furnished with manual-type collimator only).
Picture. 15 – CSA028
From the electric point of view the equipment is composed of the following elements:
• ceiling suspension;
• control board;
• X-Ray tube;
• collimator.
Wait to make control board electrical connections until the final installation phase.
Before starting to mount the cables it is advisable to choose the proper work positions to use: on the
right of the telescope (standard) or on the left (by using the left hose support arm supplied with the
equipment, the support can be detached from the right bracket that is not used and then mounted on
the left one).
It’s also important to make sure in advance that the cable lengths are enough to cover the required
distance.
Assemble the function cables of the X-Ray tube (A.T., rotating anode starter, etc.) and insert them,
together with the cable coming from the control board, into the plastic hoses; to do this, follow the
course shown in the enclosed drawings.
1) First mount the cables support arm on the telescopes right side (16.3A) (or on the left, by
using the second bracket supplied with the equipment 15.3A), fixing it through screws to the
tube number 2 lower end.
2) Then, prepare the hose supports on the carriage (02.3A) and on the bridge (19.3A).
3) Insert the electrical cables into the issued plastic hose.
Before proceeding to the cable connections, check that the previously mounted hose does not limit in
any way the radiogenic unit mobility by bringing it towards the extreme positions of its vertical
travel.
Following the suggested instructions, there should be no problem of hose interference with the stands
foreseen movements.
A second piece of plastic hose (about 7 meters long) is supplied for the cables coming out of the
carriage.
6) Cut a piece, about 1.5 meters long, to cover the distance between the carriage (06.9 or 07.9)
and the hose support fixed to the transversal bridge.
7) Fix one end to the support on the carriage left side (07.9) (or on the right side, if the left
disposition is preferred (06.9)) and insert the cables, which do not stop inside the carriage
(A.T. cables, collimator, transversal brake), paying attention to fix the cable surplus through
the use of cable ties; these cables must be arrange so that there is no interference with the
equipment mechanical parts.
8) Insert the other end into the hose support on the transversal bridge. The cables for the brake
and micro switch controls must then pass through the hole located in the upper side of the
hose support (these cables must then be connected to the terminal board on the support upper
side).
9) Now insert the remaining piece of the white plastic hose and lock it in the support on the
transversal bridge (after having inserted all cables in it), together with the tube coming from
the carriage (19.3A). Arrange well the cables which come from the hose support and tighten
the screws, making a union inside the support.
10) Then mount the last three sliding hose supports (01.3A) and insert them in the cables rail
(07.2) applied to the longitudinal rail (10.2); insert the white plastic hose, leaving about one
meter distance between a clamp and another.
11) Cut the hose at the required length and connect all supply cables.
Connect all cables and electrical devices and check that all correctly works, before mounting the
covers.
For the correct connection sequence refer to the enclosed electrical diagrams.
The suggested arrangement of the cables are shown in the following drawings. Possible other
solutions can be studied and chosen together with the installer and / or manufacturer.
Mount the plastic cover of the X-Ray tube (13.11) and the carriage cover in two parts (09.9 and 11.9 or
13.9).
The closing of the equipment plastic covers must be performed when the assembling is over, after
balancing and connecting all what is necessary, and only after the equipment testing and verification.
The micro-switch longitudinal kits are composed of the micro-switch plate of support, of the micro-
switches and of the connection cables, already connected and tested at the factory. To mount them, it
is necessary to insert the cable in the milled slot (12.5) placed inside the bridge cross bar (1.5) and
then connect the cables to the terminal board placed on the hose support on the bridge (19.3A). Then,
mount the support plate of the micro-switches in correspondence of the cover of the bridge cross bar
with the opening in the central part. The position and disposition are the ones shown in picture 5.
For the connection modalities and electrical interfacing, refers to the electrical schematics in the
OTS Section of “System Tasks and Procedures”
Only after the mechanical and electrical preparation it is possible to remove the safety screw that
avoids the rotation of the drum for the springs winding. Refers to picture 8.
4.6.4.1. Balancing of the radiogenic unit rotating movement on the horizontal axis.
This balancing must be done at the end of the mounting , when it is possible to unlock the permanent
electromagnetic brake of the rotating movement on the horizontal axis. It is important that all cables of
the equipment are correctly positioned as well.
Unlock the brake and rotate the radiogenic unit to verify it is balanced.
If balancing is not correct, find the barycentre by moving the radiogenic unit up and down in its
support by utilizing the regulation possibilities which are offered by numerous holes on the support
(08.9).
In mounting of the control board and of the electrical cables, nearly the whole final weight rests on
the stand; At this time, the vertical movement can be approximately balanced (this must be perfected
in final phases of assembly).
The balancing device uses, as active elements, some motor springs and, as passive elements, some
metal counterweight plates (11.10 and 09.10) that which must be fixed on the proper spaces of the
horizontal support arm that holds the radiogenic unit (08.10).
To allow a correct closing of the cover of the horizontal arm (11.3A) mount the counterweights plates
as shown in picture 10.
In the stand standard configuration, to contrast the telescope and x-ray system weight are used 7
springs, the eighth spring is available for heavier radiogenic units.
4.7. Calibrations
4.7.1. Display calibration.
For instructions on the equipment electronic instruments calibration, refer to the specific technical
instruction that come with the equipment.
Adjustments must be done at the end of the mounting, when X-Ray tube and collimator are installed.
ATTENTION!!!
It is important not to remove the mechanical end of travel (EOF) to avoid the potentiometers
breaking.
This ceiling suspension version, is supplied with both ends of travel (both on the rotation around the
horizontal axis and on the rotation around the vertical axis).
To move the end of travel on the rotation around the vertical axis do as follow:
1) Remove the metallic block fastened to the detent disk by a countersink screw;
2) Uncouple the potentiometer from the gearing;
3) Move the end of travel block on a different hole (there are four different ways of fastening,
offered by four different holes);
4) Check the researched freedom of moving;
5) Couple again the potentiometer to the gearing;
6) Check again that everything works properly;
7) Proceed in calibrating the unit electronically (about this, see the Electrical Manual);
ATTENTION!!!
During the movements of the ends of travel, make sure to uncouple the potentiometer so as to avoid
its damage in case a wrong movement takes the potentiometer beyond the measuring interval.
A procedure like the one here described is also valid for moving the end of travel around the
horizontal axis.
But in this case it is necessary to act on the stop dowel.
1) Slacken the 6 fastening screws (11.20) placed on the lower side of the horizontal arm (04.20);
2) Act on 6 regulation dowels (10.20) placed on the lower side of the horizontal arm (04.20),
next to the fastening screws (11.20).
Especially, lightly unscrew the regulation dowel in correspondence of the point in which you want to
increase the distance between the brake (03.20) and the anchor fastened to the detent disk (01.20).
Follow the opposite procedure if you wish to decrease the air between brake and anchor.
3) Block again the 6 fastening screws (11.20) placed on the lower side of the horizontal arm
(04.20).
Especially, lightly unscrew the regulation dowel in correspondence of the point where you want to
increase the distance between the brake (07.20) and the anchor fastened to the detent disk (05.20).
Follow the opposite procedure if you wish to decrease the air between the brake and the anchor.
For Automatic collimator calibration make reference to the RALCO collimator instruction manual in
“System Tasks and Procedures”.
Necessary tools:
• Tape measure
• Level tool
• Standard metric tool
• Flat ranch 9/16”
• Flat wrench 3/4”
• Pipe wrench 12 sides 3/8”
• Male hexagonal adapter 5/16”
• Hexagonal adapter 1-1/16”
• Measuring tape
• Hexagon socket screw key 2,5mm; 5mm
• Pipe wrench 10mm
• Screwdriver for slotted head screws
• Wrench 7mm; 10mm
• Torque wrenches (5-25Nm)
102
540
978
1107
272 max.883
1195
5.1. Prerequisites
• Stationary rails already installed according to instructions in the section 03.
• Bridge and Mounting Plate already assembled (see Figure 5-1).
Figure 5-1.
Pos. Description
01 LONGITUDINAL RAILS
02 TRANSVERSAL BRIDGE
03 RIGHT BEARING SUPPORT
04 GUIDE BRIDGE TIE
05 LEFT BEARING SUPPORT
06 BUMPERS
07 RAILS HEAD
Figure 5-2.
2) Position the monitor suspension lifting device (B) in order to assemble the suspension as
shown in 5-2 and 5-3.
Figure 5-3.
Figure 5-4.
4) Remove the monitor suspension case top cover and walls and move the remaining case base
with the monitor suspension on, as close as possible to the lifting device, make reference to
figure 5-5
Figure 5-5.
5) Do position the bracket (A 5-6) that will secure the monitor suspension to the lifting device
and check if the 4 head screws are fasten on the bracket (A) between the two monitor
suspension projecting parts (B)
B B
Figure 5-6.
A
Figure 5-7.
7) Take out the monitor suspension case base from the monitor suspension equipment and move
the monitor support harm (C) parallel to the carriage as shown in the 5-8. This step will give
the opportunity to reach monitor suspension balancing on the lifting device
A
C
Figure 5-8.
8) Lift up the monitor/monitors suspension by rotating simultaneously the two devices as shown
in A and B 5-8).
9) Do fix the monitor suspension to the carriage previously fixed with N° 4 screws T.C.E.I.
M8x60 + N° 4 plain washers Ø 8 e + 4 grower washers Ø 8 (A 5-9)
Figure 5-9.
10) Remount the safety washer, the castle nut and the cotter pin removed at step “3.” Figure 5-10
NOTE The castle nut must be tightened by hand until the hole for cotter pin is accessible, then
drive in the cotter pin and bend its extremities around the nut
COTTER PIN
CASTLE NUT
BRIDGE MOUNTING
PLATE SAFETY WASHER
Figure 5-10.
Monitor Monitor
Power Cable Video Cable
Figure 5-11
Figure 5-12.
1) Use the furnished belts to keep the Monitor Holder in a lower position for easier when
working.
2) Mount the monitor to the suspension with screws M4x12 (4 pcs) furnished in the support kit.
Connecting the cables to the monitor please refer to LCD Monitor Documentation.
6. Table installation
See pre installation manual for room preparation information and base plate fixation.
Before fixing the elevation group, you shall make sure that the plate for the fixation of the table to the
floor is positioned according to the specifications detailed in pre-installation manual and safety
regulations.
Possible discrepancies compared with the specifications of pre-installation may determine a bad
functioning of the equipment.
Table
cable
support
slot
Shim
0.5 mm
1.0 mm
1.5 mm
6 shims are provided for each
measure specified in the table. These
Fig. 3: Shims provided as serial standard accessories are to be used for leveling next to the
special screws.
Fig. 4: Base plate elevating with a floor minimum bearing of 1065 kg / square meter
Shim
0,5 mm
1 mm
1,5 mm
6 shims are provided for each
measure specified in the table. These
Fig. 6: Shims supplied as serial standard accessories are to be used for levelling next to the
special screws.
4) Secure the tilting group to the fixation plate by means of the 6 M12x25 hexagonal-head
screws (A/Fig. 10).
A
Fig. 10: non elevating table fixation of the tilting group
2) Remove the four bolts that hold the tilting group to the palette. (Fig. 12)
3) Lower the dolly wheels to raise the tilting group. (Fig. 13)
4) Roll the tilting group off the palette and place on the fixation plate that was previously fixed
or embedded into the floor.
5) Lower the tilting group onto the fixation plate and adjust the height of the wheels by means of
the 4 screws placed on the ends of the jacks (A/Fig. 13).
6) Remove the pair of jacks provided for the transport (Fig. 14) by loosing the 4 bushes fixing
the jacks themselves to the elevation group (B/Fig. 15).
7) Secure the tilting group to the fixation plate by means of the 6 M12x80 hexagonal-head
screws.
Fig. 16
Fig. 17
Fig. 18
Fig. 19 shows how the table appears after being unwrapped from the cellophane sheets.
4) Remove the cables (A/Fig. 19) from the support (B/Fig. 19), remove the support itself by
unloosing the two M8x70 screws with washer and nut (C/Fig. 19).
A C
2) Take off the screw and mount the arm into the support pole (pay attention to the cables
position)
4) Take off the M8x20 hexagonal-head screws fixing the column to the trolley
6) Lift and rotate the column in order to position the column to the floor (Warning: pay attention
to not damage the compressor during this phase !)
Attention
to the
compressor
4) Disconnect the pinion (B/ Fig. 22) of the tilting group from the gear (D/ Fig. 22).
5) Place the rotation shaft (A/ Fig. 22) making sure that the set screw is facing upwards.
6) By means of the 4 screws ensuring the adjustment of the jack height you shall adjust the table
height (B/Fig. 21 so as to obtain a perfect alignment of the central bar guides compared with
the trolley of the base.
7) Insert the central bar in the tilting group paying attention in order to avoid any possible
damage to the bearings or the bar painting (Fig. 23).
8) Slide the bar so as to place it in the centred position compared with the tilting group. Now you
shall verify that the tab of the rotation shaft id directed upwards
9) Connect the pinion (B/ Fig. 22) of the tilting group to the gear (D/ Fig. 22).
10) Secure the pinion (B/ Fig. 22) by means of the special M8x10 dowel (C/ Fig. 22).
D
A
Fig. 22 : Pinion of the tilting Fig. 23 : Fitting of the adjustable height-central bar
11) Again check perfectly center the central bar with respect to the tilting group.
12) Unloose the 8 M8x25 screws with flat washer and grower (A/Fig. 24) and remove the jack
(B/Fig. 24).
13) Remove the trolley (C/Fig. 24) by slipping it from the central bar.
1) Mount the lifting arm to the trolley support pole and tighten the screws
Warning!
It is very important to
fix this screw.
2) Move the screw pulley position to the last lifting arm hole. Pay attention to the lifting device
position, make reference to the below drawings
4) Remove the two bolts from the top of the stand trolley and turn the foot-end rear bolt back
until it is flush with the inside of the casting as shown in (Fig. 27).
Fig. 27
Warnings
to the
compressor
ATTENTION
Insert here the wooden block
Fig. 28
8) Lift the stand and place one end in the stand trolley and the other on the wooden block as
shown in (Fig. 28)
ATTENTION
Insert here the
wooden block
ATTENTION!
Do not block
A these screws!
A
B
Fig. 29: Stand back fixation Fig. 30: Stand front fixation
11) Complete fixing the stand to the trolley using the other the M12x50 screw with flat washer
(A/Fig. 31) and grower on the front side and the two M12x50 screws with flat washer and
grower (A/Fig. 32) on the top rear.
A
A
Fig. 31: Stand fixation (front side) Fig. 32 : Stand fixation (back side)
A
A
ATTENTION:
The base plate for the elevating table allows only one position for fixing the table cable support,
while the base plate for the non-elevating table allows two fixation positions. Please make reference
to Fig. 1: Base plate for not elevating table in order to establish which position shall be used.
1) Remove the covers on the cable support and bolt the cable support to the base plate using the
four M6X16 screws with flat washer and grower (B/Fig. 38).
2) Place the cables applying to the tilting and elevation (A/Fig. 38) within the four fixation holes
of the table cable support (B/Fig. 38).
Fig. 39
4) Fasten the cables to the bottom cross-piece at the point on the cables indicated by the second
piece of black tape.
5) Verify that there is 850 ±50 mm of cable between the table and the cable support.
6) Tie-wrap the cables in place in the cable exit bracket (Fig. 40).
Fig. 40
7) Run the ground wire from the cable group into the table tilting group and connect to the
ground stud (Fig. 41)
8) Replace the front cover of the cable support (Fig. 42)
Fig. 41 Fig. 42
9) Using tie wraps, fasten the cables from the stand to the top cross-piece of the cable support at
the point on the cables indicated by first piece the black tape
10) Fasten the cables to the bottom cross piece at the point on the cables indicated by the second
piece of black tape.
11) The cables should run down the back of the cable support and there should be 1800±50mm of
cable between the column and the cable support. (Fig. 44)
12) Attach the rear cover of the cable support.
Fig. 43
Fig. 44
Fig. 46
1) Before fixing the cabinet to the wall, verify that the wall can support 50kg per bolt.
2) The wall should be drilled at 1978 mm from the floor.
3) Drill 2 φ12mm, 60mm deep holes, 1978 mm from the floor for the support bracket.
4) The distance between the holes should be 400 mm
400 mm
1978 mm
from the Brackets
floor to
the holes
brackets
support
Table cabinet
Front view
3
3
23 5.5 5
34
ATTENTION: the preferred fixation is the fixation to the wall (see Fig. 47)!
227
Ø 13
487
Cabinet dimension
Cabinet width 600 mm
Cabinet depth 340 mm
Cabinet height 1960 mm
Cabinet weight 140 kg ÷ 170 kg
Fig. 50
Useful length of the cables between the control panel and the base of the right side of the power
cabinet: 15 meters.
No direct connection between the control panel and the remote-controlled table.
Should the use of the support plate be foreseen, it will be necessary to fix the support plate onto the
floor by means of two screw-nogs.(bolts)
ATTENTION: Should not the tubes be the standard types, please make reference to the specific kits.
ATTENTION
Carefully follow the mounting instructions and make sure that the collimator is correctly assembled.
Uncorrected mounting could be dangerous, it could be cause the collimator fall
IMPORTANT NOTE: you shall check that the fixation screws used to secure the collimator are the
proper ones as follows: verify that they protrude through the flange and spacers far enough to
engage at least 6 threads (that is to say 6 mm) into the tube port face.
1) Determine the distance from the focal spot to the tube port face (Fig. 53)
2) Subtract the resulting distance from 80 mm. and determine how many 1,5 mm spacers
combined with the 18,5 mm thickness of the mounting flange will be required to make up the
difference.
3) The outer face of the collimator-mounting flange must be at 80 mm. from the focal spot.,
where the allowable tolerance is ±1 mm.
4) Select four bolts of suitable thread (TSEI M6 ) and of such a length that they protrude through
the flange and spacers far enough to engage at least 6 threads into the tube port face.
18,5
FOCAL SPOT
Fig. 53
1
3
4
2
Make sure when putting back the collimator cover that you do not short the thermal device to
the ground through the cover screws
B
C
ATTENTION
Make sure to reassemble correctly the external cover without leaving fissures or small openings
between the parts
Collimator rotated to left side Centered collimator Collimator rotated to the right side
It is possible to lock the collimator with a screw in the center position in case the country regulatory do
not allows the ±90° collimator rotation.
The following picture shows the screw position to lock the collimator rotation feature.
A
B
1S3
display console board 1 Puls
Stop push button
Polycarbonate control
panel for console
Joystick (with
switches)
Console control
panel board
Optical fibers
3Mt11
3Mt14 Motore Cassetto In/Out
Motore Diaframmi a Croce In/Out cassette holder
Crossed diaphragms motor
3Mt15
Motore Griglia
3Mt10 Grid motor
Motore Piano p.p.
Tabletop motor 3Pt17
Potenziometro Piano
Longitud.
Long. Tabletop
3Pt4 potentiometer
Potenziometro Seriografo 3Mt17
SFD potentiometer Motore
Piano Longitud.
3Mt5 Long table top motor
Motore Distanza Focale
Focal distance motor 3Mt4
Motore Seriografo
3Mt3 SFD motor
Motore Stativo 3Mt7,8,9 – 3Pt7,8,9
Stand motor Motori e potenziometri
3Mt3 Collimatore Parallelo
Potenziometro Stativo Perpendicolare e Iride
Stand potentiometer Potentiometer sand motor
Parralel/perpendicular collimator Iris
3Mt1
Motore Ribaltamento 3Mt6
Tilting motor compressor motor
3Pt1
Potenziometro Ribaltamento 3Mt18
Tilting potentiometer footrest rotating
motor
3Pt2
Potenziometro Elevazione
Potentiometer elevation 3Mt16
I.I. motor
3Mt2
Motore Elevazione
Levation motor
6.13. Remote controlled table exam room position and connections lengths
The Fig. 61 at page. 157 shows the layout of the exam room table components, the cables input/output
and maximum cables connections length. This layout is a mere example since several combinations are
possible. We consider the remote controlled table position and its control panel: this is always placed
in front of the remote controlled table itself but it can be moved onto the left the right with reference to
the center of gravity of the equipment according to the normal or reverse usage of the remote
controlled table itself. This table control panel position ensures an easier and more immediate patient
visibility when the remote controlled table is in vertical position +90° or –90° (reverse).
The most frequent exam room equipment combinations are the following
1) Generator cabinet and remote controlled table cabinet: both positioned on the left wall side
behind the remote controlled table
2) Generator cabinet and remote controlled table cabinet both positioned on the right wall side
behind the remote controlled table
3) Generator cabinet on the left wall side and table cabinet on the right wall side behind the
remote controlled table
4) Generator cabinet positioned on the right wall side and table cabinet positioned on the left
wall side behind the remote controlled table.
The connection cables lengths between the generator cabinet and the table cabinet are not important in
the first or second combinations type
Others combinations are possible depending on the exam room dimension and configuration. In .the
exam room can be positioned other devices, for instance generator, digital system, stand, wall bucky,
monitor trolley or monitor suspension, ect., for this components position and their cables length and
position make reference to their proper technical manuals.
Power supply cables and equipotential:
Power supply cables and ground cables are supplied in their maximum length, as far as the supply
cables and ground cables are concerned, we have indicated only the maximum length (not the usable
length) since we do not know the mains supply panel provided in the X-ray room.
The main supply panel must be easily reachable by the operator, but since the equipments inside the
room are live (hot) this panel must be placed next to the cabinets (in order to meet with the Standards
requirements) and must be visible to the Service engineer intervening on the equipments (when the
Service engineer clears one or some of the installed equipments he must be sure that nobody can feed
it/them again: in case the main supply panel is not visible to the engineer it is necessary to protect it by
means of a safety panel (or other protection) equipped with an alarm signal alerting the engineer it is
necessary to lock it and remove the key before starting working on the equipments.
Cables length connection between the cabinet and the console:
The useful cables length between table cabinet and the console is 14 meters.
Cables length connection between tilting table and generator cabinet:
The useful cables length between tilting table and generator cabinet is 4meters and it is determined by
12meters high voltage cables
Cables length connection between table and its cabinet
The useful length between table and its cabinet is 3meters.
Cables length connection between generator cabinet and table cabinet:
This is necessary only in case the two cabinets are positioned far from each other, the cables length
must be of 6 meters
In addition take always into consideration 1meter of margin just in case of long distance.
code
destination
Each wire is identify by a small ring or label that shows the connection.
Make reference to the interconnections tables in the Drawings manual (5142538-100 Drawings).
The Table models are in compliance with the requirements of the 89/336/CEE Directive applying to
the electromagnetic compatibility (please make reference to section 0 ‘General notes’). This is why
all the connections must be executed in respect of the following instructions:
All the cables must be secured to the structure by means of the proper conductive cable fastener.
The conductive cable-fasteners must be secured by means of a picots washer and flat washer,
according to what shown in the scheme of Fig. 62 .
Make sure that the peeling of the cables are properly executed before fixing the cables themselves.
The conductive cable-fasteners used must be the same as those specified in the following
drawings and must not be replaced in any case by cable-fasteners having different
dimensions.
Fixing
screw
Flat
washer
Conductive
cable-fastener
Picots
washer
Type Description
A = Small Conductive cable-fastener FGC-3 (M4)
B = Medium Conductive cable-fastener FGC-5 (M4)
C = Large Conductive cable-fastener FGC-8 (M4)
ATTENTION
The equipment must not work without the conductive fixing cable-fasteners in any case.
Equiponential node:
Make sure to respect the instructions specified in Section 8 “Electrical connections”
Execute the ground connection with the yellow/green wire having a minimum section of 10 mm2.
Connect ground cable cod.55429 to 20GND0.
1) The cabinet arrives disconnected from the table and from the console.
2) The cables to be connected between the cabinet and the console arrive connected only to the
console.
3) The cables to be connected between the cabinet and table arrive connected only to the table.
4) On the table the cables of the motor 3MT3, the potentiometer 3PT3 and the stand are
disconnected and should be reconnected during the installation.
Left side
Cables B
NOTE – For reference only, detection chamber will arrive installed in table.
Secure the chamber with the four fixing blocks previously removed (A/Fig. 65).
A
Fig. 65 : Fixation of the detection chamber
Close the parallel diaphragms and move the grid-holder to the X-ray position.
Place and secure the cable of the detection chamber along the guide by means of the proper clamps
(A/Fig. 66).
Connect the cable of the detection chamber with the extension before the input of the cable-holders
chain (Fig. 67).
The procedure for inserting the chamber in the reverse position is the same as the procedure used
for the standard fitting.
If you insert the detection chamber in the reverse position, the cables path is turned of 180°
compared with the standard position (Fig. 68).
The connection with the extension cable is executed at the beginning of the guide placed on the
right side of the spot film device.
Execute the fixation of the cables with the clamps and check that they do not interfere with the
movement of the parallel diaphragms.
A
Fig. 69 : Fitting of the grid
≈1m
< 0,7 m
≈ 20°
8) Fix the steel rope to the table and to the Image Intensifier.
10) Move the SFD in order to align it with the Image Intensifier.
WARNING !!
DO NOT MOVE
THE SFD
DOWNWARDS !
12) If removed, fit the bar of the intensifier group (A/Fig. 71) by means of four M5x10 screws
(position B) and two M5x10 screws (position C).
C
Fig. 71 : Support bar of the intensifier group
ATTENTION!
For analogical version cameras make reference to the following paragraph:
7.6 NICAL N23 camera connection
7.7 NICAL N33 camera connection
15) Verify that the cables loop allows the ascent and descent movements of the intensifier system.
ATTENTION!
If you unscrew the three alignment screws contemporarily you may create a mechanical contact
between the I.I. sliding guide and its support. In order to avoid this issue it may be useful to insert a
spacer in the position showed in the following figure.
Fig. 73
1
The Image Intensifier system includes the Intensifier tube plus the TV chain.
6.18. Starting up
1) Verify that the positions of the table and the X-ray tube housing are perfectly parallel to the
floor .
2) Verify the correct connection of the cables.
3) Check the earth equipotential node.
4) Verify the supply voltage.
5) Verify the proper functioning of the low ceiling safety (if available).
6) Check and configure the table setting in function of exam room characteristic and system
composition
One wrong setting may caused possible equipment crashing to the ceiling or the walls
ATTENTION
To make easy the equipment disassembling and installation, the equipment is set with the collimator
NOT PRESENT (through the code 019 value 004 the equipment is set with collimator “NOT
PRESENT). Remember to set the code 019 value “005” or “003” (this value is depending on the
equipment collimator type with iris or without iris) after having assembling the equipment collimator.
See paragraph 6.21 page. 190.
The equipment is calibrated and configured in manufacture, do not change the equipment calibration if
it is not required. During equipment installation the service engineer does not have to set any table end
of stroke or modify the configuration related to the hardware. The modification of that configuration
without a proper hardware setting/up-dating can cause equipment damage. Make reference to
“Adjustment manual” to change equipment setting
N. 3 screws M4x8
A
N. 4 screws M4x8 N. 4 screws M4x8
N. 3 screws M4x8
N. 1 screws M4x12
C
E
1) Fit the front cover of the stand (A/Fig. 74) and fix it by means of two M4x8 screws
(A/Fig. 75)
2) Fit the stand’s cover (B/Fig. 74) and fix it by means of five M4x8 screws (A/Fig. 76).
3) Fix the two fixation plates by means of M4x10 screws (B/Fig. 76).
B
Fig. 76 : Stand’s cover
4) Divide the top cover of the stand in two parts (Fig. 77).
5) Fit the top cover by combining the two parts to each other and fixing them with the proper
M3x10 screws.
6) Secure the cover onto the stand (C/Fig. 74) by means of one M4x12 screw on the top side and
two M4x8 screws on the front side.
B B
C A
A C
7) Divide the top cover of the stand in two parts (Fig. 78).
Fig. 79
12) Divide the cover of the X-ray tube in two parts (Fig. 80).
F
B
A
E
Fig. 81
B
A
Fig. 82
Fig. 83
Fig. 84
The tables are configured and tested at the manufacturing plants with the collimator specified in the
customer’s order. Normally such a configuration is not subject to any further adjustments, but should
you need to execute a modification you shall carry out the following procedure:
Sequence of commands to be executed in order to configure the collimator:
Move the 25S3-Sw1.1 onto ON (the control panel display applying to the ”layer height and tilting
angle” will visualize 000).
By means of the P34 (increase) and P35 (decrease) push-buttons you shall select the code
concerned (this is visualised on the “layer height” display).
The value 004 is used only for the initial phase of the installation
in order to avoid the intervention of alarms (determined by the
absence of the collimator) when the equipment is switched on.
You shall absolutely avoid using this value with the equipment in
normal functioning.
Select the P33 (Enter) push-button to memorize the new value or the P57 or P48 (Reset) push-button
if you do not want to confirm it.
• While one of these buttons is pressed, the value visualised on the “Tilting angle” display will
be the value 000. As soon as the button is released the memorized value will be displayed.
Make reference to the adjustment manual 5121726-100 Book 1 in order to set the parameter values.
NOTE:
Information in this section is for reference only. These steps will not be required during installation
A
B
Fig. 87: Left head and right head
Fig. 89 : Fixation of the right bar Fig. 90 : Fixation of the right bar
Fig. 93 :
D B
// 0,80
A C
AC = 2100 ±1 mm
DB = 2100 ±1 mm
3) Verify that the length between the external of the two heads is equal to 2100 ±1 mm.
4) Check if the parallelism between the two headstocks is not more than the tolerance limit of
0,80 millimeters.
5) Screw the heads and the bars by means of the three fixation screws shown in Fig. 942 paying
your attention to the lengths as follows:
6) Fit the cable-holders between the SFD group and the back bar.
7) Insert the cables in the cable-holder.
8) Insert the cables of the SFD and of the dynamo of the SFD transmission in the back bar by
using the specific draft cord.
9) Insert the cables of the SFD and of the dynamo of the SFD transmission in the right head by
using the specific draft cord.
10) Insert the cables of the motor, of the transversal tabletop microswitches and of the SFD’s
transmission potentiometer in the back bar by using the specific draft cord.
11) Insert the cables of the motor and of the transversal tabletop microswitches in the left head by
using the specific draft cord.
12) Insert the SFD’s transmission potentiometer in the central bar by using the draft cord.
13) Fix the cables on both sides of the bar by means of the special threaded rod (Fig. 95).
2
Fig. shows the fixation of the right head; for the fixation of the left head, note that the M10x20 screw is always in the most external position.
14) Fix the cables coming out from the frame by means of the clamps (Fig. 96).
15) Fix the cover of the cable-holder chain by means of the three M4x10 screws (A/Fig. 97).
16) Manually verify the proper sliding of the SFD group.
17) Fit the spot-film-device and fix it to the transmission chain by means of the two M8x40
screws (A/Fig. 98).
Qt Code Description
1 71598 Template for the cassette size calibration
1 44700 Spanner for the frame bearings transversal cassette holder movement adjustment
1 44699 Spanner for the back bearings spot film device carriage adjustment
1 44701 Spanner for the front bearings spot film device carriage adjustment
1 44463 Bush for CD manual motoreductor rotation
1 44464 Bush for manual step by step motor reduction rotation
1 52360 Cable for PC serial input
1 55394 Parallel cable for boards programming
REAR
M10
300 406
36
85
15 30 455
16.1 (409)
13.9 (353)
TOP
3.3 (85)
VIEW
FRONT
LEVELLING
20.1 (511)
FEET
(4 PLACES)
DIMENSIONS ARE IN INCHES (MM)
SPACERS
FILE: IN_MTG.CDR SUPPLIED BY
CUSTOMER
(4 PLACES)
4.5 32.5
1.5
Room interface board;
6 TDI board; GIM
connection; AEC board;
16 32.5 DAP board; Touch screen;
RS232 remote diagnostic
57
Power supply input
3.5 cable (3 phase).
18.5 9
All operations to do on the unit must be carried out when it does not work. The person who is
responsible of the maintenance must completely disconnect the unit, before the adjustment or
replacement of an equipment part.
Keep hands, feet, dresses, jewels and hair at safety distance from any moving part in order to prevent
any injuries from occurring.
Do not change any part of the unit; changes which are not authorized can compromise the equipment
functioning and safety.
Center of
gravity
7.2.1.2. WS-T Wall bucky stand overall dimension and center of gravity
Center of
gravity
Model WS-T
In addition to the functions described above, the model WS-T, the bucky can be tilted from the vertical
position to the horizontal position (+90°) or to -20°. The distance of the wall-front panel ranges from
650 to 720 mm, depending on the fastening type.
The cassette can be loaded up from different sides depending on Bucky type:
1) Bucky for cassette size sensing (automatic): the cassette can be loaded up from the left or right
hand side, when ordering we recommend to order for the correct side
2) Bucky for manual collimation: the cassette can be loaded up from the right/or left hand side and
it can be decided also during installation phase
Note
Installation can be made on the wall or can be freely set up in the room using the freestanding kit.
Exposures of standing and sitting patients from the head to the knee.
Option accessories
Bucky with cassette sensing
Floor-wall fastening
Floor base fastening
Arm support
Measure chamber of automatic exposure control
Area of application
Exposures of standing and sitting patients from the head to the knee.
Important note
Proper use of this product requires that operating personnel be familiar with the operating
instructions. They should be carefully studied prior to starting up the equipment.
7.2.4. Installation
The equipment must be installed and can be maintained only by authorized and trained personnel.
Specifically, they must be informed about the possible dangers caused by improper equipment use and
the residual risks.
ATTENTION!!!
All masonry works for the room preparation are at customers or installers expense.
It is important to know exactly the room features where the unit will be installed in order to make a
correct preparation, according to the required safety rules as well.
7.2.5. Mounting of the wall bucky stand fixation only onto the floor
If the wall bucky stand is the model with a fixation only onto the floor follows the items below:
1) Place the wall bucky stand in a vertical position and move it in its working place
2) Check very carefully if the wall bucky stand is positioned in the place wanted
3) Use a pencil to mark 4 holes and point where to fix the wall bucky stand to the floor
4) By a percussion hammer drill with a widiam bit of D16 make 4 holes 90 mm deeper
5) To fix the wall bucky stand to the floor use 4 small blocks (standard supplied with the
equipment) D16 – M10 and insert the small blocks without the screw inside the 4 holes
already made on the floor.
6) Re-Place the wall bucky stand in the position decided and insert the 4 small block screws to
fix the wall bucky stand (I) Fig. 102).
7) During the installation phase is very important to check by a proper level if the wall bucky
stand is perfectly in vertical position
8) Same adjustment can be made by putting proper shims (standard supplied with the
equipment) between the base plate and the floor ( make reference to the base plate in Fig. 103
bigger base plate)
9) As soon as the installation phase is finished, cover the 4 screws smalls blocks by the 4 black
covering of diameter 30 mm, standard supplied with the equipment, make reference to (L)
Fig. 102.
7.2.6. Wallstand fixing onto the wall and floor (Figure 13)
On the contrary, if the stand has to be directly fixed to wall and floor, after having fixing the stand to
the floor (make reference to Mounting of the wallstand fixation only onto the floor) it is necessary to
proceed in as follow:
1) Mount the wall bracket to the stand (make reference to (X) Fig. 104) by the two proper screws
M5x6 ((H) Fig. 104)
2) Check if the wallstand is in a perfect vertical position
3) Through a pencil mark to wall the two fixing holes positions;
4) Move the stand a little away in order to work more easily and make the two holes;
5) Through a percussion drill with a Widiam drill of D12, make 2 holes 70 millimeters deeper in
the wall for the wall bracket fixation,
6) Use two metallic anchor fishers D12-M6 to fix the wall bracket ((L) Fig. 104)
7) The wallstand fixation to the wall by proper wall bracket gives the opportunity to move the
wall stand a little bit to the right or to the left wall side or up-ward and back-ward to be able
to reach a perfect wallstand vertical position.
L
H
Fig. 104 Wallbucky fixation onto the floor and the wall
ATTENTION!!!
Proceed carefully to prevent the counterweight (which slides inside the column) from injuring
persons or damaging things, if it is not correctly balanced.
Higher cover
4) Move the carriage wall bucky stand to the lowest end of stroke and lock it by its proper
device4
5) Place the higher cover (Fig. 102 (E)) in its proper higher place, move it in a way that it is
going to fix exactly inside the two screw cover slots
6) Tighten the cover with the two screws TBEI M4x25 (Fig. 102 ( F ), and the screw with the
large rounded headed screw M4 X 10 (Fig. 102 ( D ))
Remember to re-position the top cover (Fig. 102 (G )) by the 2 screws T.S.E.I.M5X16 (Fig. 102 ( H))
GND
115VAC
from I.T.
For
PC;
Video
Splitter;
Monitor; Touch
Modem; screen
Foot
switch
Cable
input
Fig. 107
7.3.1. Spot Film Device and Stand Group movement reversed (option)
The Spot Film Device and Stand group joystick is made with a potentiometer of 10V, thanks to it is
possible to stop the movements in the middle of the stroke (5V with a released joystick)
Usually by tipping the joystick forward the spot film device and stand group move towards the head
side (minimum end of stroke and joystick tension output is going down to 0V), by tipping the joystick
backward the spot film device and stand group will move toward the feet side (the maximum end of
stroke and joystick tension output will increase towards a 10V).
Y
Pin 3
Pin 1
1400
660mm 660m
660mm Single
Double
8. Electrical connections
See also cable tables and schematics manual.
Coloured tapes have been added on external cables that are connected during installation to ease
installation work. Colour shall represent the component of destination, that is the part to which cable
shall be connected.
Ceiling
GMM Stand
Positioner
AEC
Film DAP
I.I. AEC
Power AEC Wall
Camera Buky
cabinet
CPU
GIM
RS232 CPI
I.T.
Generator Auxiliary
Monitor
Touch
RS232 Foot
Remote
Diagnostic Switch V.S. RS232
Remote
Modem Diagnostic
Infimed
Digital system Console
Ceiling
GMM Stand
Positioner
AEC
Film DAP
I.I. AEC
Power Wall
Camera Buky
cabinet
I.I.
CPU
Nxx
CPI
Generator Auxiliary
Monitor
Touch
Foot
Switch
Video Outputs
Console
ST3
FM528
ST3
NOTE:
bridge 2-3
Parking
switches
SC2 pins 8 - 9 to cables J1, J2
OTS ST3 1-4
(code 56194A)
ST102
Directions of the main function cable may vary due to installation requirements. Install as required.
Strip terminal [ST102] serves remote braking and position detector switches.
M2
M2 pins 3 - 4 to J2
J2 to ST3 pin 5-6
ST3 pin 5-6 to
Generator Room Interface Board
TB4 – TB8 (code 56400)
8.3.5. Ceiling stand brakes and switches connections (automatic collimator version)
FM457-2C board
Collimator
NOTE:
LED
bridge 2-3
25 S3 SC2 pins 8 – 9
to OTS ST3 1-4
(code 56194A)
Parking switches
cables
Positioner 25 S3 board
ST102
Parcking switch
and FFD switches
Brakes
ST102
FM457-2C
205 cm
180 cm
100 cm
90 cm
0 cm
Detents position for automatic collimation option PBL (Positive Beam Limitation)
FM457-2C to
25 S3 CP13
(code 56427)
M2
M2 pins 3 - 4 to
OTS J2
Tube 2 indicator
M2 pins 3 - 4 to
Generator room interface board
TB8 / TB4 – pin 2, pin 1
(code 56400)
2B
3A
2C
Wall
Bucky
1
1A 3B
3
Collimator Generator
TB3 1-2
CN1
PCB FM 457-2C interface unit in the OTS trolley room interface board: CN1 - TB3 connector.
230 VAC
Connection
TB11-2 – TB10
230 VAC
Connection
TB11-2 – TB10
FM457-2C to
25 S3 CP13
(code 56427)
Equiponential node :
Make sure to respect the instructions for the Electrical Power supply
Execute the ground connection with the yellow/green wire having a minimum section of 10 mm2.
Connect ground cable cod. 55429 to 20GND0.
positioner 25 S3 board 25 S4
(for injector option only)
positioner cabinet
Positioner 25 S3
AMP11 - AMP12 -
AMP13 to generator
TDI J6
(code 56380)
Positioner 25 S3
AMP11 - AMP12 -
AMP13 to generator
TDI J6
(code 56380)
ATTENTION:
in some PC configurations the
COM serial ports could be different!
NOTE: if you choose a different PC COM port, remember to configure properly the GenWare
diagnostic SW.
Anode starter to
generator tank 1
(code 55674)
MAX 400cm
50mm
Photocell Self-Refractive
WARNING!
The 100 foot LVDS cable that connects the Precision tower (computer tower) to the camera should
only be plugged in or unplugged while the system is powered off. If this cable is connected or
disconnected while there is power to the system, it is very likely that there will be permanent damage
to the camera and/or the IAPDB.
Infimed Digital PC
Photo Pick-up camera
Generator AEC interface Board J3
(code 726-795-G1)
+24
VDC
Image Intensifier room interface board
3CM20
AMP5
Connect the cable code 56416 to the rear of the touch screen
1) Make 2 holes in the SFD cover in the position shown in the following pictures
Measure in mm
Cable A
3) tilt the table at 90° and dismount the SFD keyboard protection
Cables B
Cable A
Connection table: cable A, 7 x 25
Wire color / N. Coming from Terminal type destination Terminal type
White 1 3CM19-1 Soft-solder 3AMP19-1 AMP Contact
Brown 2 3CM19-2 Soft-solder 3AMP19-2 AMP Contact
Green 3 3CM19-3 Soft-solder 3AMP19-3 AMP Contact
Yellow 4 3CM19-4 Soft-solder 3AMP19-4 AMP Contact
Gray 5 3CM19-5 Soft-solder 3AMP19-5 AMP Contact
Pink 6 3CM19-6 Soft-solder 3AMP19-6 AMP Contact
Cable B
Connection table: cable B, 7 x 25
Wire color / N. Coming from AMP Contact destination Terminal type
White 1 3AMP19-1 AMP Contact 3S1-P44A-c Soft-solder
Brown 2 3AMP19-2 AMP Contact 3S1-P44A-no Soft-solder
Green 3 3AMP19-3 AMP Contact 3S1-P44C-no Soft-solder
Yellow 4 3AMP19-4 AMP Contact 3S1-P44B-c Soft-solder
Gray 5 3AMP19-5 AMP Contact 3S1-P44B-no Soft-solder
Pink 6 3AMP19-6 AMP Contact 3S1-P44D-no Soft-solder
Bianco = White
Marrone = Brown
Verde = Green
Giallo = Yellow
Grigio = Gray
Rosa = Pink
Cover board
WARNING:
on request of the equipments operator, it is possible to reverse the SFD and column movement by
modifying the cabling of the console on the table harvests. Make reference to the following procedure
in order to do this.
The remote joystick for spot film device and stand group movement is made by two switches that stop
the spot film device and stand group movement as soon as the lever is released
Usually by tipping the cloche lever forward, the spot film device and stand group will move toward the
head side (minimum end of stroke), by tipping the cloche lever backward, the spot film device and
stand group will move toward the feet side (maximum end of stroke)
Or:
1. Open the joystick
2. Invert the connection on the output of the push buttons P44A and P44C (brown wire with -
green wire)
3. Reassemble the covers of the joystick and the spot film device
4. Check that the spot film device and stand group movement is reversed.
5. By tipping the cloche lever forward, the spot film device and stand group will move toward
feet side
6. By tipping the cloche lever backward, the spot film device and stand group will move toward
head side
56112
55966
56176A
55969
55968
ABS contacts
Positioner cabinet
MV1 220 VAC out
monitor
power supply
(code 56111)
Positioner cabinet
video out
(code 56112)
Power supply
(code 56111)
1400
video cable
(code 56112)
660mm 660m
660mm
Ground cable
D38 - cod. 56114
55968
56090
56112 56112
56406
56176A
56402
56403
56401
56113 55969
New camera
interface board
code 56109A
Camera head
Flat cable
Camera board
Sync 56406
Camera board
power supply
Flat cable
TV chain SYNC to
camera board SYNC
(code 56406)
0V24
ABS contacts
ABS
SYNC
J2
Positioner 25 S3
AMP11 - AMP12 -
AMP13 to generator
TDI J6
(code 56380)
Positioner 25 S3 board Generator TDI board
8.7.7. Analogical TV chain N33 power supply and video signals (analogical version only)
When using this camera (1kx1k) with coaxial connection you must insert a filter switching the switch
as shown below:
1400
660mm 660m
660mm
Ground cable
D38 - cod. 56114
Power supply
from I.T.
(code 068-113)
Generator
TDI board J7
ABS and
pulsed fluoro
(code 726-654-G1)
Power supply
to I.T.
(positioner cabinet)
(code 068-113)
P1 – P4 to generator TDI
J4A-J4B-J4C-J4D-J4F
(J4A = 726-619-G1 = P1)
(J4B = 726-619-G2 = P2)
(J4C = 726-619-G3 = P3)
(J4D = 726-619-G4 = P4)
(J4F = 726-619-G6 = P6)
PF ABS TABLE
J6 J10
J7 J11
J1
3
J2 2
1
J3
Generator
TDI 2 F
C
board D
J4
A
B
R S T Neutral
PF ABS TABLE
J6 J10
J7 J11
J1
3
J2 2
1
J3
Generator
TDI 2 F
C
board D
J4
A
B
PFL
TDI board
Infimed Camera
To J2 – channel 1: To J3 – channel 2:
positioner DAP ceiling stand DAP
(code 40358) (code 55993)
positioner DAP
positioner DAP connector
Positioner TUBE 1
X-Ray housing To positioner X-Ray housing
Anode H-Voltage thermal switch (code 55656)
(Tube 1 code 18875)
To positioner X-Ray housing
anode starter (code 55674)
TUBE 2
TUBE 2 GND (code 55783)
NOTE: if you choose a different PC COM port, remember to configure properly the GenWare
diagnostic SW.
25 S3 board
Positioner
Positioner
Injector Cabinet
56442
56443 56420
56444 P1 3S1 AMP1 AMP1 25 S4
board
GND 56453 GND
AMP2
56431
GND
TDI
J10
Generator
J11
J1
3 Connections
56421 between
J2 2 25 S4 AMP2 25 S3 board
and 25 S4 board
J3 1
Injector GROUND
cable connection
(code 56453)
Injector signal
cable connection
(code 56442)
(code 56443)
(code 56444)
To 25 S4 board
(code 56420)
(code 56431 GND)
AMP1
For 25 S3 board
connection
AMP2
25 S4 board frontal view 25 S4 board lateral view
I.T. 110VAC
56120
Video splitter 56120 I.T. 110VAC
LAN
Input Output 1 Output 2
56207 Console Monitor
56209
COM 4 COM5
56226
GIM 56218
Transformer
56210
56184
SVGA video
(code 56209)
Ground cable
D38 - cod. 56114
COM1 - 56216
56240
56234
Modem
LAN
GIM
56184
handswitch J3 to
console positioner
keyboard
(code 56183)
8.12.2. Digital PC
WARNING:
For some PC hardware the COM serial ports could have different position and different COM
number. Please pay attention in order to configure the correct port number for the remote diagnostic
programs (GenWare and GMM Terminal).
WARNING!
The 100 foot LVDS cable that connects the Precision tower (computer tower) to the camera should
only be plugged in or unplugged while the system is powered off. If this cable is connected or
disconnected while there is power to the system, it is very likely that there will be permanent damage
to the camera and/or the IAPDB.
Positioner Positioner
cabinet
Console
Cod. 55510
230 Vac
Cod. 55513
Supplier /
Amplifier
Cod. 56297 Cod. 56296
Microphone Loudspeaker
Remote T O O R + A
230 V A B U U X X T O R + A
Console 2 1 1 2 A B U U
Microphone Loudspeaker
Power supplier / amplifier detail
Power
Supply
screw
230 Vac
Cod. 55513
cable
screw
screw
Cable route
7) Connect all the cable with the amplifier in the positioner cabinet
ATTENTION!
This procedure is applicable only if the PlatinumOne digital system by InfiMed is provided with the
“ClinicalApp” version 2007.12.14.02 (or further versions) software.
ATTENTION!
You shall execute this optional procedure only if especially requested by the final customer and only
on systems meant for countries where an external NTSC video recorder (VCR) is used.
8.15.1. Introduction
Should you need to record image sequences under TV format on the external DVD video recorder
(Sony DVD recorder), you shall eliminate a certain number of lines of the radiological image so as to
avoid any incompatibility problem coming from the video signal formats.
In order to avoid an excessive image degradation, the lines to be eliminated are preferably chosen in
the upper part and lower part of the image itself that is, exactly the areas where the patient name and
other data are usually entered in text modality.
The NTSC standard provides for a lower number of TV lines compared to the PAL system and
consequently with the NTSC TV Record it is not possible to reread this text information correctly.
This is the reason why a patch has been released allowing you to modify the position of this text
(overlay) by positioning it towards the centre of the image.
The following (optional) procedure ensures the installation of a preferred overlay type: either A) or
B):
A) the preset overlay (PAL) type: having text written externally to the image (either higher or
lower);
B) the NTSC type having text written more internally to the image (either less higher or less
lower).
5) In the right side of the window area you shall double-click on the “setup.exe” file to run the
installation program.
6) When the InstallShield “Welcome” screen appears, click on the “Next” key to
continue.
7) When the InstallShield “Start Copy Files” window appears, double click on the “Next”
key to continue.
8) When the prompt for the type of overlay to be installed appears, make sure that the “NTSC
adjusted overlays” option is selected and click on the key to continue.
9) Once the installation is complete, click on the key.
10) You shall now switch off the system and restart it so that the new overlay file can be
downloaded in the IAPDB/IAPDB-MF Board.
6) When the InstallShield “Welcome” screen appears, you shall select the “Repair” option and
click on the “Next” key to continue.
7) When the prompt for the type of overlay to be installed appears, you shall make sure that the
“Default overlays” option is selected and then click on the “Next” key to
continue.
8) Once the installation is complete, you shall click on the “Finish” key.
9) At this point you must switch off the system and restart it so that the new overlay files can be
downloaded in the IAPDB/IAPDB-MF Board.
9. Digital PC configuration
Technical notes to update programs and other installations on the digital PC
WARNING
To install new program and change PC configuration parameter is necessary to be an
“administrator” user. Remember to log in with the user name “serviceapp” (and with the
appropriate password).
If the auto log-in process is enabled hold down the shift key during the start up process to have the
log-in window.
*****
ATTENTION:
Windows auto log-in function should be already disabled.
The following procedure must be done in case it is necessary to update the digital system.
At windows start up, in case the system does not show the log in mask, please perform the following
procedure in order to disable the auto log in function.
WARNING!
If the automatic log-in procedure is activated, it will not be possible to make any remote diagnostic
activities. As a matter of fact, all the diagnostic programs can be started only from “administrator”
users. As a result, a remote user interested in remote diagnostic, must log-out the user “operator” in
order to be able to log-in as a “serviceapp” user. This last procedure step is disabled in case of the
Auto log-in functioning.
7) Just click the button, in order to obtain an empty password for the user “operator”.
In other words, as soon as the log-in mask opens up, the user “operator” will be able to work
with the computer simply by selecting the “OK” button (no password needed).
A remote code execution security issue has been identified in the DHCP Client service that could
allow an attacker to remotely compromise a Windows-based system and gain control over it.
9.2.1.2. Solution
The DHCP Client service is not critical to the functions of the imaging system. It is, however, critical
to the system network connection used for DICOM functionality. In order to address the security
concerns presented by the DHCP Client vulnerability, the DHCP Client service must be disabled. To
maintain DICOM capability, the IP Configuration information must be manually set in the Windows
Network Properties. The following instructions will detail these processes
Warning!
Do disable the DHCP service only if the customer has not decided to use it for his local area
network management.
NOTE:
Disabling the DHCP Client requires administrator-level privileges to the operating system. Steps 1-4
provide admin-level logon instructions; the remaining steps detail DHCP Client service operation.
1) Power up the digital imaging system. As the system boots, hold down the <SHIFT> key to
interrupt auto-logon and allow for administrator logon.
2) Replace the ‘operator’ user name with the admin-level user ‘serviceapp’. The user name is
case-sensitive.
3) Type the correct password in the password field. The password is case-sensitive.
4) Click the OK button or press <ENTER>. The ‘serviceapp’ user account starts the Service
Application automatically.
5) Close the Service Application. The Clinical Application will start automatically.
6) Close the Clinical Application. This will provide un-impeded access to the Windows
desktop.
7) At the Windows desktop, right-click on the My Computer icon and select Manage from the
menu.
8. On the left side of the System Management window, expand the Services and Applications
category and select the Services line.
9. Locate and double-click on the DHCP Client service in the list of installed services.
12. Click on the Apply button to commit the changes, and click the OK button to close the
window.
13. Verify that the Services list in the Computer Management window updates with the new
DHCP Client information.
To perform this step is necessary know the IP address and subnet mask. These information must be
requested to the Network administrator where the equipment will be installed.
WARNING!
IP address and subnet mask must be requested to the Network administrator where the equipment
will be installed.
4) On the command line type ping following by a known accessible IP address (in the example
192.168.0.155).
5) If the network is working the reply is showed in the next figure.
It is very important to enable the access to internet only for the time necessary to perform Windows
Update and to disable the access immediately after the completion of this procedure
DO NOT leave internet access enabled!
1) Double click on “My Network Place”, select the “Local Area Connection”, right click on it
and select “Properties”. In the Properties windows select “Internet Protocol (TCP/IP)” and
click on the “Properties” button in order to reach the following windows (or follow the steps
described in the par. 9.2 page 315)
2) Add the proper value for the parameters “Default gateway” and “Preferred DNS server”.
These parameters are user site dependent.
3) Save the Data by pushing the OK button. Now it will be possible to reach Internet through the
hospital network.
Remember to disable Internet access immediately after the completion of this procedure!
3) Select “I connect through a local area network (LAN)”, and click “Next”.
4) Select “I connect through a local area network (LAN)”, and click “Next”.
5) Click “Next” also for the successive windows (Automatic discovery …).
4) Select the “Custom” button and “Yes” for the following information message.
7) At the end of this process, you shall restart your computer to ensure the updates take effect.
2) In the left frame you will see the following selections (under “Select by Type”).
High Priority (nn)
Software, Optional (nn)
Hardware, Optional (nn)
checkbox
10) Perss “OK” when Windows asks you if you want to restart the computer.
11) After the PC will be restarted, run again all these steps in order to verify if new patches are
found. See the following picture as an example. Repeat these steps until you will not have any
new patch in the “Critical Updates and Service packs” windows.
(Remember: do not install any Windows Operating System Service Pack, “Software,
Optional” patches, or “Hardware Optional” patches.)
Remember to disable Internet access immediately after the completion of this procedure!
ATTENTION:
Genware and GMM_Terminal COM ports should be already configured.
Perform this procedure in order to verify the configuration only.
WARNING:
For some PC hardware the COM serial ports could have different position and different COM
number. Please pay attention in order to configure the correct port number for the remote diagnostic
programs (GenWare and GMM Terminal).
WARNING:
For some PC hardware the COM serial ports could have different position and different COM
number. Please pay attention in order to configure the correct port number for the remote diagnostic
programs (GenWare and GMM Terminal).
ATTENTION:
This is a necessary step in order to enable the remote diagnostic capabilities.
Two different IIP programs configurations are possible: the first one have the access via local area
network, the second one give a modem access opportunity.
A valid gateway IP address is necessary to configure remote access via local area network.
Dial-out Prefix for calling numbers outside the hospital to configure remote access via modem.
Please Contact hospital network Administrator.
5) Type in the command line, “iipadmin config” and push “Enter” button to start the interface
configuration. (See “Voyager_IIP_Installation_Steps_Using_IS.doc” for more details)
6) Please wait until the IIP configuration Stat Screen comes up (illustrated in the figure below)
2) Enter the Gateway IP Address. This is IP Address of the system that can connect to the
OnLine Center. Ask Hospital Network Administrator about it.
3) Make sure that the number entered in each box is only three digits and does not start with
zeroes.
7) Click on the Checkout Now Button and wait for a minute until Reboot Needed window pops
up
8) Click Cancel on the Configure Insite Now popup
9) Click on the Exit Button to exit IIPConfig Window
10) Click Ok button in the popup
11) Wait for a minute for prompt in the iLinQ Shell (Note: Don’t kill the window)
12) Enter Exit at the prompt
13) Reboot the System.
14) When Logon Dialog pops up enter
Username as serviceapp with the correct password
15) Click on the OK Button and Wait for a minute for application to start
17) Click on Cancel if any Modem Installation Wizards popup. (No for any modem installation)
1) Under Device Connection Tab Select the Device Connection Type to be Modem from
available selections
2) Leave the Default Selections unchanged as shown below except if the dial-out prefix is
different. Note Following is Very Important
Dialout prefix is the number you dial (from hospital) before you dial long distance or international
phone number. Consult System Administrator for further information
3) Click on Apply
4) If any Dialog Box pops up asking Location Info enter details and click on OK Button.
5) When Phone and Modem options dialog should popup
6) If Multitech MT 5634ZBA already exists listening on Com1 select that. Remove any other
modem that exists. Select the modem and click on the remove button
22) Click on Cancel if any Modem Installation Wizards popup. (No for any modem installation)
Note: Continue with the installation procedure from where you left
1) Select “My Computer” with the right button of your mouse and select “Properties”.
2) Then select the “Hardware” tab and press on the “Device Manager” button.
3) Double click on the “Modem” device group and by pressing on the right button of your mouse
select each modem defined in this group which is different from the MT5634ZBA modem.
4) For each modem (except MultiTech MT5634ZBA) select the “Disable” option
3a) Should not these windows appear, please select “My Computer” with the right button of your
mouse and select “Properties”.
3b) Then select the “Hardware” tab and press the “Device Manager” button.
3c) Select the device group “Modem” with the right button of your mouse
3d) Select the “Scan for hardware changes” option.
If you do not have the “RXi Software CD”, you can download the Modem driver from internet after
entering the following web site:
http://www.multitech.com/SUPPORT/Families/MultiModemZBA/drivers.asp
WARNING
This is a necessary step in order to enable the DICOM capabilities. Please note that the information
needed to complete this procedure must be requested to the on-site Network Administrator.
Should any error messages appear when you open the Configuration window, you shall just delete
the previous configured devices
DICOM_Store
Before starting the following procedure, you shall verify the possible existence of any previous
Worklist drivers: select the “Name” menu and press the upward arrow several times in order to
make sure you reach the very first driver of the list: if you find the only name being available is
“Empty”, this means there is no Worklist device currently configured and you shall therefore proceed
according to the following steps:
5. Change the "Your Locale (location)" setting to select the appropriate language.
ATTENZIONE
Questa procedura è valida solo se sul computer digitale Infimed Platinumone è installato il software
“ClinicalApp” versione 2007.12.14.02 (o versioni successive).
ATTENZIONE
Eseguire questo passo opzionale solo se espressamente richiesto dal cliente finale e solo per sistemi
destinati a paesi nei quali si utilizza un videoregistratore (VCR) esterno con il sistema NTSC.
9.8.1. Introduzione
Nel caso si vogliano memorizzare delle sequenze di immagini sul videoregistratore DVD esterno
(Sony DVD recorder) in formato televisivo, per problemi di compatibilità dei formati dei segnali video
è necessario eliminare un certo numero di righe dell’immagine radiologica.
Per non degradare troppo l’immagine radiologica, le righe da eliminare vengono scelte
prevalentemente in alto ed in basso, proprio nella zona dove di solito viene sovraimpresso il nome del
paziente ed altre informazioni scritte in modalità testo.
Dato che lo standard NTSC utilizza un numero di righe televisive inferiore rispetto al sistema
PAL ne consegue che nel filmato TV NTSC non è più possibile rileggere correttamente queste
informazioni in formato testo.
Per questo motivo è stata rilasciata una patch che consente di modificare la posizione delle scritte
sovraimpresse sullo schermo (overlay) in modo da spostarle un po’ più verso il centro dell’immagine.
5) Nell’area di destra della finestra, dare doppio click sul file “setup.exe” per avviare il
programma di installazione.
6) Quando il programma InstallShield visualizza la finestra “Welcome”, fare click sul pulsante
“Next” per continuare.
7) quando il programma InstallShield visualizzerà la finestra “Start Copy Files”, fare click sul
pulsante “Next” per continuare.
8) Quando apparirà la richiesta di selezione del tipo di overlay da installare, assicurarsi che sia
selezionata l’opzione “NTSC adjusted overlays”, e fare click sul pulsante per
continuare.
9) Una volta completata l’installazione, fare click sul pulsante .
10) a questo punto è necessario spegnere il computer e riavviarlo in modo che i nuovi file per
l’overlay vengano scaricati nella scheda IAPDB/IAPDB-MF.
ATTENZIONE
Eseguire questo passo solo se è stato erronneamente installato l’overlay versione NTSC e si intende
ritornare ad utilizzare l’overlay predefinito adatto ai sistemi VCR di tipo PAL.
7) Quando apparirà la richiesta di selezione del tipo di overlay da installare, assicurarsi che sia
selezionata l’opzione “Default overlays”, e fare click sul pulsante per continuare.
5) In case you have not your license key (0000-0000-0000-0000) you need to ask for a new one
If you have an internet connection available, connect to the following site in order to obtain your
license key:
http://www.cpii.com/CMP/TouchLicWeb/TSLicenseKey.aspx
or
http://oldstuff.cpii.com/CMP/TouchLicWeb/TSLicenseKey.aspx
Enter the requested information in the Touch Screen License Key Generator screen.
Use the System Code number (example: 4C5E EA54 E43E 45A4) and the Customer Code number
(8662).
Type your e-mail address.
Clicking ? accesses the help function.
Click Submit when finished.
Record the license key when it is displayed. Double check that the license key was recorded correctly,
this is critical
A second way to have the License key is to send the following type of e-mail.
ATTENTION:
The Touch Screen should be already configured.
COM 2
If the program will ask you if you want to save the changes, please press the “Yes” button.
ATTENTION:
The Touch Screen receptors should be already updated.
Only if needed, please set the appropriate picture for the receptors of the “Acquisition Mode” screen.
Press the “Main Menu” button in order to go back to the Main Menu screen
From the “Main menu” select “System Utilities” (remember the password) and from this menu, select
“Receptor Symbols”.
DSA
WARNING
Please note that for systems provided with the DSA, it is important that the DSA option is receptor
# 6 (please make reference to the location in the above picture).
3) Select the APR data file to be restored. The selection will be highlighted.
4) Press Restore.
5) After a brief delay, a pop-up window will be displayed indicating that restore was successful.
6) Press OK to continue.
WARNING:
The Tubes should be already calibrated.
WARNING:
the following procedures produce x-rays. take all safety precautions to protect personnel from X-
radiation.
WARNING:
Deeper information about GenWare program isn’t the aim of the present manual.
CAUTION:
always verify the manufacturer of the tube insert. if the x-ray tube has been rebuilt, the tube insert
and tube housing may be from different manufacturers.
WARNING:
Before starting the auto calibration procedure, close completely the collimator shutters in order to
avoid X-ray beams emission and damage the I.I. device
This utility allows for tube one or tube two to be calibrated for both focal sizes. When selected, the
generator is placed into auto calibration mode. The rad and fluoro settings are disabled at this time.
To start the tube calibration process, press and hold the X-ray exposure button. The exposure values
are shown in the Auto Tube Calibration window. When auto-calibration is completed on the currently
selected focus and tube, the process will stop. Repeat the above steps to calibrate the remaining focal
spot or tube.
Press Close to exit the tube calibration utility.
The tube calibration may be aborted by releasing the X-ray switch. In this case, an error is reported,
however, the process may then be restarted again or the utility may be exited.
By pressing the Tube function , the GenWare program will open the Auto tube calibration
window. Make reference to the following figure:
Double click on the “My Computer” icon. A window like the following will appear.
Double click on the “Local Disk (C:)” and select the option “Folder” from the menu File – New.
See the following picture.
Type a properly name for the new directory. In the example we will use “Config_Backup”.
Save all the backup files decrypted in the following procedures in this folder and do not forget to copy
this directory also on a floppy disk.
For the 12” Touch Screen: connect the serial cable to the J4 Datalink port Touch Screen (COM3).
For the 15” Touch Screen: connect the serial cable to the COM1 Touch Screen (in order to have the
Datalink function) or directly to the J15 generator CPU board.
For positioner: connect the serial cable to the CN3 positioner CPU board.
Follow the instructions described in the paragraphs 11.3 CPI Generator backup, 11.4 APR backup and
11.6 Positioner parameters backup and save all the backup files on a floppy disk.
Please execute the backup of the APR file being used only if this one was modified by the User.
1) Press “System Utilities” in the “Main menu”. A window will be displayed asking you to
enter your password.
2) In “System Utilities” menu, press APR Backup / Restore. A pop-up window will be
displayed showing the available backup files and the factory-default APR files in different
languages.
You must have also a standard serial cable for modem connections (null-modem). This is a description
of the electrical connection for this type of cable.
Cable example:
use a Female DB-9 connector (Touch screen)
and a Female DB-9 connector (laptop PC)
Touch Screen connection (J4 Datalink) Example of a PC serial port COM1 connector
(J4 Datalink serial port COM3)
Next you have to select which serial communications port you would like to use. For a laptop this will
most likely be COM1. What you need to do is select the <Connect Using: COM1> option and press
the OK button. If you receive a message saying unable to open COM1 then close HyperTerminal and
repeat these steps but choose a different communication port.
NOTE: if you have selected all of the available communications ports in the list and all of them give
you an error message when you try to open them, make sure that you are not running any other
program that uses the serial port such as GenWare. If you are not running any serial port software
there is a possibility that your laptop does not have a serial port. If in this case you will require either
another laptop to perform this task or you will need to buy a PCMCIA Serial Port card from a local
computer store.
57600 Baud
No Parity
8 Data Bits
1 Stop Bit
Hardware Flow Control
From the Touch Screen Main Menu, press the SYSTEM UTILITIES button and type the password to
access the System Utilities menu.
Select the file you want to send by highlighting it and press the bottom left button on the touch screen.
From the system utilities menu, press APR Backup / Restore. A pop-up window will display showing
the available backup files and the factory-default APR files in various languages.
Insert
USB stick
Connect the USB memory stick to either USBA or USBB on the rear of the touch screen console and
press .
After a brief auto-detection sequence in which the touch screen looks for the USB key, the USB File
Transfer Utility window will open.
To copy an APR file from the touch screen to a USB memory stick device, select the desired file from
the Touch Screen window and press . Once the file is copied successfully it will appear in the
USB Device window of the USB File Transfer Utility.
When finished, press Close to exit.
Select “Config_Backup” as
destination directory.
Select “Config_Backup” as
destination directory
Select “Config_Backup” as
destination directory
Select “Config_Backup” as
destination directory.
Run “GMM Terminal” (“Start” button, “Programs” “GMM Teleservice”, “GMM Terminal”).
Press the “Connect” button to initiate the connection with the table then press the Setup button and
type the password.
Press the Backup button to perform the backup of the EEPROM data.
The program requests a password to perform this procedure. Type “123” in the password window and
press the “OK” button.
The program shows a window where you can type some information about the Positioner. These data
allow you to understand where the backup comes from.
All these information are mandatory but we recommend to insert properly information in the fields
“Serial N.”, “Customer”, “Date”, and “Signature”.
The others fields “Lot” and “Order” are used for GMM internal purpose.
Press the “OK” button to continue. The program shows the “Save As ...” window.
File Name:
Save As:
The file name is automatically build going by the serial number typed in the “Serial n” box of the
previous window. Is possible to change this file name but we do not recommend to do this.
Select the appropriate folder to save the file and press the “Save” button.
A message shows you if the backup was properly executed (see following picture).
WARNING
If you have the Norton Ghost v.10 installed on your digital PC and you have also a CD with a
different Norton Ghost version (7.5 or 8.3), please do not use the Norton Ghost v.10 any more, but
make sure to use only the CD version.
To verify if the Norton Ghost 10 program is available on the PC, check if the symbol identifying
Norton Ghost is included in the bar of active applications (below on the right side)
or push the “Start” button , select the “Programs” option and verify
if it is present in the list of installed programs.
The number of CDs needed depends from the HD occupation and the type of data stored on it.
In a very approximate way the number of CD-R/RW required to perform a complete Hard Disk backup
can be estimated in one CD for every GByte of HD data (plus one CD).
WARNING:
If you have the Norton Ghost v.10 installed on your digital PC, please follow the procedure at chap.
11.7.1 page 392 .
If you have any other Norton Ghost version, please follow the procedure at chap. 11.7.2 page 394.
WARNING :
In this case is important not use other programmes during backup execution and move the bar to the
left side screen in order to reach maximum execution Ghost speed.
During backup execution one or more messages as for instance the one in the following window, will
be displayed. In this case replace the CD into the recorder player with a new CD to go on making data
saving.
At the backup end, close the following window by pushing the button “x” placed on the right upper
window side
11.7.2. Backup the system with Norton Ghost version 7.5 or 8.3
These versions of Norton Ghost are not installed on the Hard Disk of digital PC but they are supplied
on floppy disk support (v. 7.5) or on CD (v.8.3).
WARNING:
In order to launch the execution program it is necessary that the floppy disk reader (Floppy Driver)
or the CD player (CD Driver) are devices able to execute the procedures of PC automatic boot-up
with a higher priority compared to the one of hard disk (Hard Disk).
In case the system is not configured as explained above, the insertion of disks in the related players
will not result into any effect and the system will keep on executing operative system Windows 2000.
The procedure to set PC boot up is described in paragraph 11.7.3 page 398.
8. Push OK to continue;
9. In the following screen, select the CD drive in the “Look in”:
12. Answer NO to the following question: “Copy a bootable floppy to the CD/DVD disk?”
13. Answer YES to the following question: “Proceed with Drive Backup to CD/DVD?” to any
other WARNING message;
14. If the CD with Norton Ghost has been left in the CD/DVD device, the following message
warning operator to insert a blank CD will appear:
16. When it is required, insert the other CDs/DVDs until back up is completed;
17. Take the last CD out and close the program by selecting in the menu the option “Quit”;
18. Re-boot the system;
19. Put the CDs in a safe place.
Execute the following procedure, only in case the Norton Ghost program Ghost (v.7.5 or 8.3) IS NOT
automatically starting upon PC boot up.
WARNING:
The aspect of BIOS configuration window is linked to the type and version of Motherboard. Indeed,
its aspect could also be different from what displayed in the following screens.
3 Select in the menu the “Boot” and “Boot Device Priority” option and then press “ENTER”.
4 The situation could be similar to the one shown in the picture where Hard Disk (HD) is the
second boot up device and therefore it is endowed with a higher priority compared to the CD
device;
Floppy disk
Hard disk
CD / DVD
5 The priority of CD device shall be increased so that it is higher compared to the one of Hard
Disk (HD);
CD / DVD
7 In the menu displayed in the middle of the window, select the CD/DVD device and press
“Enter”;
8 Now the CD priority should be set as higher compared to the one of Hard Disk (HD);
Floppy disk
CD / DVD
Hard disk
9 Press the function button F10 to exit and save configuration changes;
10 The PC will be restarted and will automatically execute the program that is registered in the
CD inserted in the CD/DVD player.
WARNING:
By using this PC configuration, in the case the CD with Norton Ghost is left inside the CD/DVD
player, the back up program is going to automatically start. This could confuse user or cause some
trouble. In order to avoid such risk, once the back up and restore procedure is terminated, we
recommend to set again the priority of automatic boot up on its original value.
Please find Chinese labels attached to the Chinese operator manual version
In the following paragraphs you will find on the left the standard English label supplied with each unit
and on the right the label with Chinese text: the latter must be stuck on the standard one as it is
indicated by drawings.
T1
C1
T1
WARNING This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed
T1
警告 本X射线装置可能会危及患者和操作员,除非遵守安全曝光系数和操作说明书的要求
1. WARNING
C1 Do not connect items which are not specified parts of the system
2. ATTENTION
Ne accouplez pas equipment que ne sont pas classifiée come parts
du system.
警告
请勿连接不是系统专用部件的项目.
G1
G3
G2
高压
G1 G3
在面板打开或移开的情况下
会被接触到
G2
G1 G3
G2
M1
M1
小心
弹簧处于 高度压缩状态
请勿试图进行拆卸,有关的
完整说明请参阅
13. Appendix
13.1. Dip switches information
13.1.1. Positioner dip switches
The 25S3-Sw2 dip-switch is used to short-circuit the analogical input should not the relevant
transmission be present. Do not execute any adjustments of potentiometers or acquisitions of end-of-
strokes with reference potentiometer present and dip-switch being ON.
Make reference to the topographic drawing code 55308f.1 for the jumpers positions.
JW4
Please note that whenever the 25S4 board is present, you shall cut JW-3 jumper of the I/O board
(25S3)!
Console PC
Video 2
Console 512 TV
Set the DVD recorder “SETUP MENU” – “TV SYSTEM” with the value “PAL”
TB7 TB10
0V (GND) 230 VAC Room Interface
Printer
55541
Touch Screen
56437
COM1
Notes: