Sp-Uv1000 en
Sp-Uv1000 en
Sp-Uv1000 en
SP-UV1000
USER MANUAL
Contents
Safety ............................................................. 1
Unpacking .......................................................... 1
Installation ....................................................... 2
1. Environment Required.................................................................................................. 2
2. Install Spectrophotometer...................................................................................... 2
Overview ........................................................... 2
Symbols ............................................................ 3
1. Appearance ......................................................................................................................... 4
2. Keypad ................................................................................................................................... 5
Functions .......................................................... 6
Measuring .......................................................... 8
1. Photometry ......................................................................................................................... 8
2. Quantitation..................................................................................................................... 9
3. Utility................................................................................................................................. 9
Troubleshooting ................................................... 15
I
Warranty .......................................................... 20
II
Safety
Please follow the guidelines below, and read this manual in its entirety
to ensure safe operation of the unit.
Package Contents
Description Quantity
Spectrophotometer 1PC
10mm Glass Cuvette 4PCS
10mm QuartzCuvette 2PCS
Power Cord 1PC
User’s Manual 1PC
Dust Cover 1PC
Unpacking
Open the package, according to carefully check the packaging packing list
items, if found inside the packaging are missing or damaged items please
1
contactus and authorized contractual partners.
Installation
1. Environment Required
To ensure the best performance, the following conditions are required:
2. Install Spectrophotometer
Placement
Place the instrument on the stable table carefully.
Overview
SP-UV1000 Spectrophotometer is an electrical measure instrument which is
widely used in the laboratories.
Caution, Danger!
Caution, Hot!
Ground
Fuse
Main Specifications
Optical System Single beam
Wavelength Range 200—1000nm
Wavelength Accuracy ±2nm
Wavelength Repeatability 0.8nm
Photometric Range -0.3—3A, 0—200%T
Photometric Accuracy ±0.5%T
Photometric Repeatability 0.3%T
Spectral Bandwidth 4nm
Stray Light 0.3%T@220nm&360nm
Stability ±0.002A/h@500nm
Work Mode Photometry, Quantitation
Interface USB, Parallel(printer)
Power Requirement AC 110/220V, 50/60Hz
3
Dimensions 490x360x210
Weight 14kg
Work Environment 15—35℃,15—70% relative humidity
StoreEnvironment -10—50℃, 15—70% relative humidity
3 1
Back View
8 11
6
5 7 9 10
4
2. Keypad
3. Description of Keys
5
Function Key:Functions according to the screen
Functions
Photometry
Display results as Abs, %T or Energy.
Quantitation
Use a Standard Sample to establish Standard Curve.
Getting Started
The following chart describes the basic operation of the instrument.
Warm up 20 minutes,
Any key to skip
100.0%T
Basic Quantitative
6
Important Guidelines
Reagents and dilution buffers can cause cauterization and other damage
to health.
Samples (nucleic acids, proteins, bacteria cultures) can be infectious
and cause serious damage to health.
During sample preparation, measuring procedures and maintenance and
cleaning work, observe all local laboratory safety precautions (e.g.
wear protective clothing and gloves, use of disinfectant) regarding
the handling of sample material.
Dispose of measuring solutions and cleaning and disinfectant materials
in accordance with the relevant local laboratory regulations.
General Operating
Select Application
Set Wavelength
G O TO ʎ
Test interface, press key to set wavelength, , to modify
WL: 500.0nm
Please enter WL.:
500.0 nm
OK
Set Parameters
SET
Press enter into setup interface, , to select items or
P R IN T
Test Interface, press the key , then press key , to select
7
“Clear Data, not Print ”, (left)to delete.
Blank
ZE R O
Put the Reference in the light path, press to do blank.
Measure Samples
P R IN T
Test Interface, press the key , , select “Print, clear
After got the Standard Curve by Marked, press , input the file
Measuring
1. Photometry
Step 1. StartPhotometry
G O TO ʎ
Press to set wavelength,press , to input wavelength
Step 3. Blank
ZE R O
Put the Reference in the light path and press to do blank.
Step 4. Measuresamples
Put the sample in the light path, and then the result displays on the
2. Quantitation
Step 1. Start Quantitation
9
2 methods to establish Standard Curve:
press (left).
(left) to confirm.
OK Cancel OK Cancel
11
Create Curve C=KxA+B WL.
Load Curve C=1.000xA+0.2 500.0 <
C=0.02xA+0.32 470.0
Clear Curve
OK Cancel OK Cancel
Step 4. Blank
ZE R O
Put the Reference in the light path, press to do blank.
500.0nm 0.000A
No. WL Abs Conc.
1 500.0 0.039 0.078
2 500.0 0.042 0.084
3 500.0 0.041 0.082
3. Utility
SET
Main Interface, press to go into utility setting.
12
WL: 500.0nm 0.000A Test Mode
D2 Lamp On
100.0%T W Lamp On
Basic Quantitative OK Cancel
Test Mode
13
Test Mode √W Lamp On
D2 Lamp On On √
W Lamp On Off
OK Cancel OK Cancel
Note: During the course, open the lid of the compartment is prohibited.
Dark Current
Reset WL
Get Dark Current!
Version
OK Cancel
Reset Wavelength
Dark Current
Reset WL
Calibrating WL!
Version
OK Cancel
About Version
14
Dark Current
Reset WL Software: V2.4.2
Version Hardware: V1.2
OK Cancel
Troubleshooting
Review the information in the table below to troubleshoot operating
problems.
15
Repair and Maintenance
1. Daily Maintain
Check the Compartment
After measurement, the cuvettes with sample solutions should be taken out
of the compartment in time. Or the volatilization of the solution would
make the mirror go moldy. Users must pay more attention to the corrosive
sample and liquid easy to volatilize. Any solution remains in the
compartment should be wiped off immediately.
Surface Clean
The cover of the instrument is with paint. Please use wet towel to wipe
off the drips on the surface immediately. Organic solution is forbidden
to be used to clean the cover. Please wipe off the dirt on the cover timely.
After every test or after a solution change, the cuvettes should be cleaned
carefully, or the remains on the surface would cause measuring error.
Danger! Be sure to switch off the power and unplug the socket
before replacement!
16
Step 4. Replace a new fuse
Pick out the spare fuse (3.15A/250V) and replace it to the working
position.
Replace Lamps
Hot! Wait 20 minutes before open the lamp chamber after power
off to avoid scald!
17
Step 4. Open the cover of the light chamber
Unscrew the 2 screws on the light chamber cover and remove it.
Light
chamber
D2
Lamp
1
2
18
filament in the same direction as the old one face.
W
Lamp Filament
Right Position
Switch
Mirror
2
Step 7. Finish
Reset the cover of the light chamber and fix the screws. Reset the
cover of the instrument and fix the screws. Recover the Pole in the
19
compartment, then the course finished.
Warranty
Wewarrant that this product will be free from defects in material and
workmanship for a period of one (1) year from date of purchase. If a defect
is present, we will, at its option, repair, replace, or refund the purchase
price of this product at no charge to you, provided it is returned during
the warranty period. This warranty does not apply if the product has been
damaged by accident, abuse, misuse, or misapplication, or from ordinary
wear and tear.
For your protection, items being returned must be insured against possible
damage or loss. This warranty shall be limited to the replacement of
defective products. IT IS EXPRESSLY AGREED THAT THIS WARRANTY WILL BE IN
LIEU OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE WARRANTY OF
MERCHANTABILITY.
Equipment Disposal
This equipment is marked with the crossed out wheeled bin symbol to indicate
that this equipment must not be disposed of with unsorted waste.
For more information about where you can drop off your waste of equipment,
please contact your local dealer from whom you originally purchased this
20
equipment.
By doing so, you will help to conserve natural and environmental resources
and you will ensure that your equipment is recycled in a manner that
protects human health.
Thank you!