Manual DAQ970A-DAQ973A - Eng - Tds
Manual DAQ970A-DAQ973A - Eng - Tds
Manual DAQ970A-DAQ973A - Eng - Tds
Large 4.3” color display for ease of setup and viewing data
LAN and USB for easy connectivity to your PC (DAQ973A comes with additional GPIB)
Code compatible with the 34970A/34972A
USB flash drive support to copy/log data in standalone applications
BenchVue DAQ with enhanced time/frequency domain measurement included
Large 4.3” color display for ease of setup, configure and viewing data
The DAQ970A/DAQ973A 4.3” graphical color display and soft keys are designed to be intuitive for easy
configuration and measurement displays in multiple formats. You can view the measurements in numeric
digit, bar graph, trend chart and histogram, giving you a quick overview of your measurement.
Label Description
1 USB port
2 On/Standby switch with LED indicator
3 Display
4 Softkeys
5 Measurement operation menu (to control the initiation of the measurements)
6 Measurement configuration menu (to set parameter for measurements)
7 Knob
8 Cursor navigation keypad
Label Description
1 Slot identifier (100, 200, 300)
2 GPIB interface (DAQ973A only)
3 AC mains input
4 LAN interface
External trigger input, alarm outputs, channel advance input, and channel
5
close output
6 USB interface
7 Chassis ground screw
What’s more, each channel is independently configurable. This means you can configure channel 1 for
dcV, channel 2 for a K-type thermocouple, and channels 3 and 13 for a 4-wire RTD measurement, all on
the same module, all in a single scan. For custom linear conversions, use the Mx+B scaling function on
any channel. You can even display a custom engineering label like RPM or PSI to identify your
measurement units.
Versatile alarms
Alarms are available on a per-channel basis as well. Enter a high limit, a low limit, or both. The
DAQ970A/DAQ973A compares each reading to its limits and flags any out-of-range measurements. You
can assign one of four TTL alarm outputs to any input channel to trigger external alarm lights, sirens, or
send a TTL pulse to your control system, all without a PC connected.
Flexible functionality
The DMM is installed inside the chassis rather than in one of the slots, leaving all three mainframe slots
free for switch and control modules. You can choose from eight different modules to get the precise
functionality you need now—while giving you flexibility for future expansion.
The internal DMM gives you the flexibility to measure 12 types of inputs easily and inexpensively. The
built-in signal conditioning and conversion routines turn raw inputs directly into real information. Each
measurement channel is independently configurable, so you can set different measurement functions,
scale factors and alarm limits, even on adjacent channels. Advanced measurement features such as
offset compensation, variable integration time, and delay are also selectable on a per-channel basis.
Software drivers
Your months of test system software development time need not go to waste. Software drivers that
support (C, C#, Visual Basic, Visual Studio), LAN, USB and GPIB on the DAQ970A/DAQ973A and SCPI
programming language make integration even easier.
Easy scanning
The DAQ970A can easily scan with external instruments. It builds a scan list that includes all enabled low
frequency multiplexer inputs. Scans are controlled with the external channel advance input, or with the
front panel Step key.
Keysight quality
We know you can’t afford instrument downtime due to hardware failures and unscheduled maintenance.
That’s why our engineers designed reliability into the DAQ970A/DAQ973A: A rugged enclosure, state-of-
the-art surface mount construction throughout, reduced parts counts, and rigorous and thorough testing
on all aspects of the product.
Higher throughput
Our unique architecture incorporates a high-performance microprocessor on each module, offloading the
mainframe processor and minimizing backplane communications for faster throughput.
Convenient connections
On-module screw-terminal connectors make wiring more convenient. Built-in strain relief cable routing
and cable tie points keep your wiring secure and safe from accidental tugs and pulls. An internal analog
bus routes signals from any of the low frequency multiplexers directly to the internal DMM, without the
need for external connections.
DAQM902A 2-wire reed 250 300 V 50 mA 10 MHz < 4 µV Built-in cold junction
16 ch Multiplexer (4-wire selectable) reference
1. Sa mpling rate.
A familiar spreadsheet environment makes it easy to configure and control your tests. And a rich set of
colorful graphics provides many options for analyzing and displaying your data—all with point-and-click
ease. Set up multiple graphics using strip charts, histograms, bar and scatter charts, individual channel
results, and more. And of course, you can use BenchVue DAQ to easily move data to other applications
for further analysis, or for inclusion in your presentations and reports.
When used with DAQM909A digitizer, you can access these advanced measurement capabilities:
Minimizes external wiring and the resultant potential for noise and errors to enter your system
Reduces hidden costs and overall system cost by avoiding unnecessary cables, breakout boxes and
signal conditioning elements
Simplifies your configuration - for faster, easier setup - with fewer connections and components
Takes the guesswork out of error analysis.
Measurement accuracies are specified to include all system-related errors
Improves reliability, with fewer interconnects and fewer parts that can fail
Measurement accuracy is specified as percent of reading plus percent of range, where reading is the
actual measured value and range is the name of the scale (1 V, 10 V, etc.) - not the full-scale value
(1.2 V, 12 V, etc.).
DMM measurement accuracies include all switching errors. Switching errors are also listed separately
in the module specifications section. Temperature measurement accuracies include ITS-90
conversion errors. The thermocouple accuracies include the reference junction error as well.
Accuracies are listed as either 24-hour, 90-day, or 1-year specifications. This refers to the length of
time since the instrument’s last calibration. Use the specification that matches your calibration cycle.
The 24-hour specifications are useful for determining short-term relative performance.
From the following page, the 1-year accuracy is: 0.003% of reading + 0.0004% or range
Now, multiply the 10 V temperature coefficient from the DC voltage specifications table by the number of
degrees outside of operating range for additional error (non ACAL):
From the temperature characteristics table, total error is: Thermocouple probe accuracy + 1.0°C
For this example, assume a ± 1 V square wave input with 50% duty cycle and a 1 kHz frequency.
Accuracy for 1 V, 1 kHz sinusoid is: 0.05% reading + 0.02% range = 0.7 mV or 0.07%
DC voltage2
100 mV 0.0030 + 0.0050 0.0040 + 0.0060 0.0050 + 0.0060 0.0065 + 0.0060 0.0005 + 0.0005 0.0002 + 0.0005
1V 0.0015 + 0.0006 0.0025 + 0.0006 0.0035 + 0.0006 0.0050 + 0.0006 0.0005 + 0.0001 0.0002 + 0.0001
10 V 0.0010 + 0.0003 0.0021 + 0.0004 0.0030 + 0.0004 0.0045 + 0.0004 0.0005 + 0.0001 0.0002 + 0.0001
100 V 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0055 + 0.0006 0.0005 + 0.0001 0.0002 + 0.0001
300V 0.0020 + 0.0020 0.0035 + 0.0020 0.0040 + 0.0020 0.0055 + 0.0020 0.0005 + 0.0001 0.0002 + 0.0001
2,7
Resistance
100 Ω 0.0030 + 0.0050 0.0050 + 0.0060 0.0060 + 0.0060 0.0070 + 0.0060 0.0006 + 0.0005 0.0002 + 0.0005
1 kΩ 0.0020 + 0.0007 0.0030 + 0.0007 0.0040 + 0.0007 0.0050 + 0.0007 0.0006 + 0.0001 0.0002 + 0.0001
10 kΩ 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001
100 kΩ 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001
1 MΩ 0.0020 + 0.0005 0.0060 + 0.0005 0.0070 + 0.0005 0.0080 + 0.0005 0.0010 + 0.0002 0.0002 + 0.0002
10 MΩ 0.010 + 0.001 0.020 + 0.001 0.025 + 0.001 0.030 + 0.001 0.0030 + 0.0004 0.0030 + 0.0004
100 MΩ 0.100 + 0.001 0.250 + 0.001 0.350 + 0.001 0.400 + 0.001 0.1000 + 0.0001 0.0100 + 0.0001
1000 MΩ 2.000 + 0.001 2.500 + 0.001 3.500 + 0.001 4.000 + 0.001 1.0000 + 0.0001 0.1000 + 0.0001
1. Specifications are for 60-minute war m- up, integration setting of 10 or 100 NPLC, auto-zero on, AC slo w filter. ACAL run within the last 2 days
& 901 module connectivity.
2. These specifications include offset characteristics contributed by the 901 Module & backpl ane (which adds DCV + 2 uV, 4- wire oh ms + 2 mΩ )
3. 20% over range on all ranges, except 300 VDC, 300 VAC, 1 ADC, 1 AAC, an d di ode test have 0%.
4. Relative to calibrati on standards. 24 hrs spec only applicabl e when calibrated wi th the sa me DAQM901 module that is being used to verify
spec.
5. Add this for each °C outside the last ACAL ± 5 °C.
6. Add this for each °C outside the last external standards cali bration ± 5 °C.
7. Specifications are for 4-wire oh ms fun ction or 2-wire ohms using math null for offset. W ithout math null, add 2 Ω additional error in 2-wire
oh ms function. The 100M and 1G oh m ranges are 2-wire only. See the manua l for low power oh ms specification and measure ment currents.
1 µA 0.017 + 0.005 0.040 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0010
10 µA 0.015 + 0.002 0.040 + 0.002 0.050 + 0.002 0.060 + 0.002 0.0015 + 0.0006
100 µA 0.012 + 0.001 0.040 + 0.001 0.050 + 0.001 0.060 + 0.001 0.0015 + 0.0004
Specifications: DC current
1 mA 0.007 + 0.003 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0015 + 0.0005
10 mA 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.060 + 0.020 0.0020 + 0.0020
100 mA 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0005
1A 0.050 + 0.006 0.070 + 0.010 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010
5V 0.002 + 0.010 0.008 + 0.010 0.010 + 0.010 0.012 + 0.020 0.0010 + 0.0010
Specifications: AC voltage
True RMS AC Voltage6,7
3 Hz – 5 Hz 0.50 + 0.02 0.50 + 0.02 0.50 + 0.02 0.50 + 0.02 0.010 + 0.003
5 Hz – 10 Hz 0.10 + 0.02 0.10 + 0.02 0.10 + 0.02 0.11 + 0.02 0.008 + 0.003
10 Hz – 20 kHz 0.02 + 0.02 0.04 + 0.02 0.05 + 0.02 0.06 + 0.02 0.007 + 0.003
20 kHz – 50 kHz 0.05 + 0.03 0.06 + 0.03 0.07 + 0.03 0.08 + 0.03 0.010 + 0.005
50 kHz – 100 kHz 0.15 + 0.05 0.15 + 0.05 0.15 + 0.05 0.15 + 0.05 0.060 + 0.008
100 kHz – 300 kHz 1.00 + 0.1 1.00 + 0.1 1.00 + 0.1 1.00 + 0.1 0.200 + 0.020
300 V range
3 Hz – 5 Hz 0.50 + 0.05 0.50 + 0.06 0.50 + 0.06 0.50 + 0.06 0.010 + 0.008
5 Hz – 10 Hz 0.10 + 0.05 0.10 + 0.06 0.10 + 0.06 0.11 + 0.06 0.010 + 0.008
10 Hz – 20 kHz 0.02 + 0.05 0.04 + 0.06 0.05 + 0.06 0.06 + 0.06 0.010 + 0.008
20 kHz – 50 kHz 0.05 + 0.09 0.06 + 0.09 0.07 + 0.09 0.08 + 0.09 0.010 + 0.0012
50 kHz – 100 kHz 0.15 + 0.15 0.15 + 0.15 0.15 + 0.15 0.15 + 0.15 0.060 + 0.020
100 kHz – 300 kHz 1.00 + 0.3 1.00 + 0.3 1.00 + 0.3 1.00 + 0.3 0.200 + 0.050
3 Hz – 5 kHz 0.07 + 0.04 0.09 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
5 kHz – 10 kHz (typical) 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.006
(typical)
1. Specifications are for 60-minute war m-up, i ntegration setting of 10 or 100 NPLC, auto-zero on, AC slo w filter. ACAL run within the last 2 days &
901 module connectivity.
2. 20% over range on all ranges, except 300 VDC, 300 VAC, 1 ADC, 1 AAC, and diode test have 0%.
3. Relative to calibration standards. 24 hrs spec only applicable when calibrated with the sa me D AQM901 module th at is being used to verify spec.
4. Specifications are for the voltage measured at the input terminals. The 1 mA te st current is typical. Vari ation in the current source wil l create
so me variation in the voltage drop across a di ode juncti on.
5. Add this for each °C outside Tcal ± 5 °C.
6. Specifications are for sinewave input >0.3% of range and > 1 mVr ms. For the 300 Vr ms range inputs >1 % of rang e. All VAC ranges limited to 8 x
10 ^7 Volt–Hz.
7. Low-frequency perfor mance: three filter settings are available: 3 Hz, 20 Hz, 200 Hz. Frequencies greater than these filter settings are specified
with no additional errors.
8. Specifications are for sinewave input > 1% of range and > 10 µA AC.
1.0000 nF 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.05 + 0.05
10.000 nF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
100.00 nF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
1.0000 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
10.000 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
100.00 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
Specifications: Frequency
Frequency2,3
1. Specifications are for using Math Null zeroing. High dissipation factor capacitors may sho w different results than a single frequency
measure ment. Film capacitors usually have lower dissipation factors than other dielectrics.
2. Specifications are for sine wave input unless stated otherwise.
3. Input > 100 mV. For 10 mV to 100 mV i nputs, multiply % of reading error x10. Amplitude 10%–120% of range e xcept 30%–100% for the 300
ACV range. Specifications are for 1-second gate time (7-digits).
4. Square wave input specified for 10 Hz – 300 KHz for 1 second aperture. For s horter apertures the mini mu m frequenc y requires > 2 cycles.
N -200 ºC to -100 ºC 2 ºC
Thermistor
Strain
Bridge strain – voltage measurement (1/4, ½ & full bridge, Bending & Poisson configurations, requires external power supply & bridge completions)
1. Thermocouple probe accuraci es would co me fro m ther mo couple suppli er and should be selected based on temp ran ges required.
2. DMM accuracy a ssu mes measure ment with DAQM901 A. For module Oh ms me asure ments - see respective s witch mo dules.
3. Accuracy specificati on depends on Mfg specificati ons of device.
DAQM901A 80 80 80 80
DAQM908A 80 80 80 80
DAQM900A 90 90 90 90
DAQM901A 50 50 50 50
DAQM902A 90 90 90 90
DAQM908A 50 50 50 50
DAQM901A 50 50 50 50
1 PLC/16.7 ms (20 ms) 6.5 0.0001 + 0.5 µV 0.0001 + 0.5 mΩ 0.0006 + 0.02 nA
1. For DCV on the 10-volt range with zer o volts input and auto zero on.
2. RMS noise adder measured with zero volts input and auto zero on.
3. The follow DCI ranges have the se additi onal multi pliers: The 10 mA by 5 x, th e 100 mA b y 2 x.
1 µA < 0.0011 V
10 µA < 0.011 V
1 mA < 0.11 V
10 mA < 0.027 V
1A < 0.7
Input resistance
Measurement type AC-coupled True RMS. Measures the AC component of the input.
DC current
10:1 maximum crest factor, (3:1 at full-scale). Measurement bandwidth limited to 300 kHz for signal plus
Crest factor
harmonics.
Will select higher range if peak input overload is detected during auto range. Overload is reported in
Overload ranging
manual ranging.
Resistance
Measurement method Will select 4-wire or 2-wire ohms. Current source referenced to LO input.
Maximum lead resistance (4-wire ohms) 10% of range per lead for 100 Ω and 1 kΩ ranges. 1 kΩ on all other ranges
Temperature
Thermocouple
≥ to 1 PLC 60 dB2
< 1 PLC 0 dB
Voltage ranges 100 mVrms full scale to 300 Vrms. Auto or manual ranging.
All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals.
Measurement considerations
Shielding inputs from external noise pickup is critical for minimizing measurement errors.
Environment
Mechanical
Regulatory
IEC 61326
EN 61326
CISPR
EMC
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current revisions
Triggering conditions
Computer interfaces LXI (rev 1.4) USB, LAN, GPIB (DAQ973A only)
Open/close speed 120 ch/s 120 ch/s 120 ch/s 120 ch/s 60 ch/s 70 ch/s
Characteristics – typical: DC
Offset voltage
< 4 µV 0 µV6 < 4 µV < 1 µV < 1 µV < 4 µV < 1 µV
(incremental to DMM specs)
Initial closed channel R < 50 Ω < 1.0 Ω < 1.0 Ω < 0.2 Ω < 1.0 Ω < 0.5 Ω < 1.0 Ω
Isolation Ch-Ch, Ch-Earth > 10 GΩ > 10 GΩ > 10 GΩ > 10 GΩ > 10 GΩ > 1 GΩ > 10 GΩ
1. DMM accuracy a ssu mes measure ment with DAQM901 A. For 900 module Ω measure ments - see note 2.
2. The series resistance of the DAQ M900 A limits the use of the 100 Ω and 1 kΩ resistance range. The precision of 2-wire Ω is limited by this on
resistance. Keysight reco mmend s the use of 4-wire Ω for these 2 ranges. Ma xi mu m resistance range of DAQ M900 A is 1 MΩ. T he 10 MΩ and
above ranges are not specified on this module.
3. For AC Volts - The i nput impedance of this module decreases with frequency. A source i mpedance of 5 Ω or less will maintain specifications
over the full frequency range. A source impedance of 50 oh ms or less will maintain specifications to 5 kHz.
4. For DC Volts - Addi tional settling time may be needed for short integration time and higher source impedance.
5. Resistance range ≥ 1 MΩ is Typical.
6. Absolute Offset is <2µV, Incre mental Offset to 901 = 0µV. 2µV included in DCV’s % of range specs. O ffset for this module already included in
speci fications above.
Characteristics – typical: AC
Unlimited
Switch life (no load) within FET 100 M 100 M 100 M 100 M 5M 100 M
Bank2
Unlimited
Switch life (rated load)3 within FET 100 K 100 K 100 K 100 K 100 K 100 K
Bank2
DAQM900A
20-Channel General Purpose Multiplexer
The DAQM900A module is divided into two banks of 10 two-wire channels each. All 20 channels switch
both HI and LO inputs, thus providing fully isolated inputs to the internal DMM or an external instrument.
During 4-wire resistance measurements, channels from Bank A are automatically paired with channels
from Bank B to provide the source and sense connections. The module has a built-in thermocouple
reference junction to minimize errors due to thermal gradients when measuring thermocouples.
80 ch/s scanning
Two- and four-wire scanning
Built-in thermocouple reference junction
300 V switching
The Keysight DAQM901A is the most versatile multiplexer for general purpose scanning. It combines
dense, multifunction switching with 80-channel/second scan rates to address a broad spectrum of data
acquisition applications.
Two- and four-wire channels can be mixed on the same module. Two additional fused inputs (22
channels total) route up to 1 A of current to the internal DMM, allowing ac and dc current measurements
without the need for external shunt resistors.
The Keysight DAQM902A employs reed relays to achieve scan rates up to 250 channels per second. Use
this module for high-throughput automated test applications as well as high-speed data logging and
monitoring tasks.
Sixteen two-wire inputs switch up to 300 V. Two and four-wire channels may be mixed on the same
module. User provided shunt resistors are required for current measurements.
This general-purpose switch module has 20 independent single-pole, double-throw (SPDT) relays. Use it
to cycle power to products under test, control indicator and status lights, and to actuate external power
relays and solenoids. Combine it with matrix and multiplexer modules to build custom switch systems. Its
300 V, 1 A contacts can handle up to 50 W, enough for many powerline switching applications.
32 two-wire cross-points
300 V, 1 A switching
The Keysight DAQM904A gives you the most flexible connection path between your device under test
and your test equipment, allowing different instruments to be connected to multiple points on your DUT at
the same time.
Rows or columns may be connected between multiple modules to build 8x8, 4x16 or larger matrices, with
up to 96 cross-points in a single frame.
2 GHz bandwidth
BNC to SMB adapter cables included
The Keysight DAQM905A RF multiplexers offer broadband switching capabilities for high frequency and
pulsed signals. Use them to route test signals between your device under test and your signal generator,
oscilloscope, spectrum analyzer, or other instrumentation.
The RF multiplexer are arranged as two independent 1x4 multiplexers, each with a common shield and a
switched center conductor. Connections can be made directly to SMB inputs with 2 GHz usable
bandwidth, or to the BNC-to-SMB adapters provided with 1 GHz bandwidth. Multiple banks may be
cascaded together for applications requiring even larger topologies—create a 16:1 multiplexer in a single
frame.
The Keysight DAQM907A allows great flexibility for a variety of sense and control applications. It
combines two 8-bit ports of digital input and output, a 100 kHz gated totalizer, and two ± 12 V or ± 24 mA
analog outputs—all on a single earth-referenced module. The digital inputs and totalizer input may be
included in a scan. Alarm limits for the digital and event counter inputs are evaluated continuously,
capturing and logging alarm conditions even between scans.
On top of that, it has two additional channels (Channel 6 and 7) that provide the ability to sense output
current when sourcing voltage, or sense output voltage when sourcing current.
Characteristics – nominal
Totalize input 100 kHz (max) Rising or falling edge, programmable, min H or L time 5µS
DAC 1, 2 ± 12 V or ± 24 ma non-isolated
Totalize input
Count events from devices like photo interrupters, limit switches, and Hall-effect sensors.
It keeps an updated total which can be read via the front panel or programmatically at any time. With 26
bits of resolution, it can count events at full speed for nearly 11 minutes without an overflow.
On top of that, it has two additional channels (Channel 6 and 7) that provide the ability to sense output
current when sourcing voltage, or sense output voltage when sourcing current.
80 ch/s scanning
Single-wire switching for common low applications
Built-in thermocouple reference junction
Use the Keysight DAQM908A for the greatest density in common low applications, such as battery test,
component characterization, and benchtop testing. Each module switches 40 one-wire inputs. All two-wire
internal measurements except current are supported. The module low connection is isolated from earth
and can float up to 300 V.
Simultaneous sampling
Up to 800 kSa/s at 24-bit
Time/frequency domain measurement with Benchvue DAQ
When three DAQM909A modules are inserted within the same instrument, all 12 channels can be
synchronized to provide simultaneous sampling.
DC Volt
Range (Differential input – sinewave Vrms)2 0.42 Vrms, 1.4 Vrms. 4.2 Vrms, 25 Vrms
1. The DAQM909 is ali as-protected for out-of-band inputs to 6MHz with a mplitudes < 10dB belo w the range’s ma xi mu m.
2. Balanced differential input, amplitude = range ma x -1d B (and < 12Vr ms), 50k Sa/s, frequency mode, 20kHz analysis bandwidth, DC-coupled.
2
THD (1 kHz) - 103 dB
2,3
THD (20 Hz - 20 kHz) - 83 dB
Characteristics: Nominal
Coupling AC, DC
Sample rate
Acquisition bandwidth
4
Channel to channel phase match (1 kHz) 0.045º
Overvoltage protection ± 42 V
To rack mount a single instrument, or the adapter kit (part number: DAQ190A)
To rack mount two instrument side-by-side, order the lock-link kit (part number: DAQ194A) and flange kit
(part number: DAQ191A)
Mainframe
DAQ970A Data Acquisition System Mainframe with USB and LAN
DAQ973A Data Acquisition System Mainframe with USB, LAN and GPIB
Modules
DAQM900A 20-Channel solid-state multiplexer
Accessories
11062A Kelvin clip set