Manual DAQ970A-DAQ973A - Eng - Tds

Download as pdf or txt
Download as pdf or txt
You are on page 1of 42

Keysight DAQ970A/DAQ973A

Data Acquisition System


Sample More Signals, Faster and Dynamically

Find us at www.keysight.com Page 1


Key Features
 3-slot mainframes with built-in 6 ½ digit DMM
 Basic 0.003% DCV accuracy
 9 switch, RF and control plug-in modules, including a new 4-channel simultaneous sampling digitizer
 Up to 450 channel/s scan rate
 Up to 120 channels per system
 Up to one million points scanning memory
 Measures and converts 12 different input signals:
Temperature with thermocouple, RTDs and thermistor; dc/ac volts; 2- and 4-wire resistance;
frequency and period; dc/ac current and capacitance

 Large 4.3” color display for ease of setup and viewing data
 LAN and USB for easy connectivity to your PC (DAQ973A comes with additional GPIB)
 Code compatible with the 34970A/34972A
 USB flash drive support to copy/log data in standalone applications
 BenchVue DAQ with enhanced time/frequency domain measurement included

Measurements you can trust


We took the measurement engine from our high-performance benchtop DMM and embedded it inside a
3-slot mainframe. You get the benefit of proven Keysight measurement performance, universal inputs
with built-in signal conditioning, and modular flexibility, all in a low-cost, compact data acquisition
package. The DAQ970A/DAQ973A features 6½ digits (22 bits) of resolution, 0.003% basic dcV accuracy,
and ultra-low reading noise. Combine that with scan rates of up to 450 channels/sec, and you’ve got the
speed and accuracy you need to get the job done right the first time.

Built-in signal conditioning to get your job done faster


Whether you need to measure temperature, ac/dc volts, resistance, frequency, or current, the
DAQ970A/DAQ973A can handle it. The internal auto-ranging DMM directly measures 12 different
functions, eliminating the need for expensive external signal conditioning. And our unique design allows
complete per channel configurability for maximum flexibility and quick, easy set up. It’s like having an
independent, high-performance DMM behind each channel.

Large 4.3” color display for ease of setup, configure and viewing data
The DAQ970A/DAQ973A 4.3” graphical color display and soft keys are designed to be intuitive for easy
configuration and measurement displays in multiple formats. You can view the measurements in numeric
digit, bar graph, trend chart and histogram, giving you a quick overview of your measurement.

Find us at www.keysight.com Page 2


Standard connectivity to the PC
The DAQ970A has built-in Gigabit LAN and USB 2.0. With the standard LAN connection, you also get the
added benefit of a graphical Web interface for easy configuration of measurements and monitoring of
results using a standard Web browser. In addition to LAN and USB, the DAQ973A comes with a built-in
GPIB interface.

Convenient data storage with USB flash drive


The DAQ970A/DAQ973A also feature a built-in USB memory port so that you can use a USB flash drive
to collect data without being connected to a PC. Data can be logged directly to the USB flash drive,
extending your instrument’s memory, or copied from internal memory for transfer to a computer in another
location.

Virtual front panel control via Web interface


The built-in graphical Web interface provides easy access and control of the instrument using a Web
browser such as Chrome and Internet Explorer. Using this interface, you can configure measurements,
define and execute scan lists, or monitor measurement results from anywhere on the network. Simply
enter the instrument host name or IP address into the browser URL and gain access to the instrument’s
front panel capabilities.

 Specify per channel measurement configuration


 Define and execute switch scans
 Monitor measurement readings
 View and save data
 View error queue
 View instrument information like module configuration, relay counts, firmware revisions, and more
Additionally, since the Web interface is built into the instrument, you can access it on any operating
system that supports a Web browser without having to install any special software. Password protection
and LAN lock out are also provided to limit access. The graphical Web interface makes it easy to
configure measurements, set up and execute scans or troubleshoot your designs from anywhere on the
network.

Custom configurations that grow with you


Three module slots and nine switch/control modules allow you to customize the DAQ970A/DAQ973A to
meet your unique requirements. Buy only what you need—and add more modules later as your
application grows.

Find us at www.keysight.com Page 3


Front Panel at a Glance

Label Description
1 USB port
2 On/Standby switch with LED indicator
3 Display
4 Softkeys
5 Measurement operation menu (to control the initiation of the measurements)
6 Measurement configuration menu (to set parameter for measurements)
7 Knob
8 Cursor navigation keypad

Find us at www.keysight.com Page 4


Rear Panel at a Glance

Label Description
1 Slot identifier (100, 200, 300)
2 GPIB interface (DAQ973A only)
3 AC mains input
4 LAN interface
External trigger input, alarm outputs, channel advance input, and channel
5
close output
6 USB interface
7 Chassis ground screw

Find us at www.keysight.com Page 5


The Keysight DAQ970A/DAQ973A offers Unequaled Versatility for Your Data Acquisition
Applications
In the past, you had to make a choice. On the one hand, you could choose the simple operation and low
cost of a data logger. On the other hand, you had the flexibility and higher performance of a modular data
acquisition system. The Keysight DAQ970A/DAQ973A Data Acquisition System gives you the best of
both worlds: a simple user interface with low per-channel cost, modular flexibility, standard connectivity
and impressive measurement performance. Whether you’re an R&D engineer working on characterizing
your latest design, or a manufacturing engineer building a test system or troubleshooting a process, the
DAQ970A/DAQ973A Data Acquisition System offers the best combination of price and measurement
performance.

It’s a data logger


Configured with a 20-channel relay multiplexer, the DAQ970A/DAQ973A becomes a powerful, low-cost
data logger for simple characterization applications. What’s more, the DAQ970A/DAQ973A with its LAN,
USB or GPIB interfaces is ideal for easy set up and control for data logging applications in remote
locations.

It’s a data acquisition front end


The DAQ970A/DAQ973A is an automated test system with excellent measurement performance—it’s got
the accuracy, resolution, and speed you need.

An easy-to-use data logger for monitoring and characterization applications


Data loggers are used to monitor multiple signals (temperature, voltage, etc.) over extended periods of
time to identify irregularities. Example applications include environmental chamber monitoring,
component inspection, benchtop testing, process troubleshooting, and temperature profiling. The
Keysight DAQ970A/DAQ973A is easy to use for a multitude of data logging and monitoring applications,
either stand-alone or with a computer. Its flexible, modular design makes it scalable from 4 to 120
channels, and lets you add actuator, digital I/O, and analog output channels for simple control. Its small
size, ruggedized features and USB memory port on the DAQ970A/DAQ973A make it perfect for portable
applications. Use USB, LAN or GPIB interfaces for simple connection to the PC and support of remote
applications. The DAQ970A/DAQ973A can be set up at remote locations and accessed through the Web
interface on a network connection or through the USB port by copying instrument configurations and
measurement data results to a USB flash drive.

Better measurement with fewer hassles


Tired of putting up with the mediocre measurement performance you get with most data loggers or plug-in
data acquisition boards? The DAQ970A/DAQ973A offers 6½ digits of resolution and 0.003% basic 1-year
dcV accuracy.

Find us at www.keysight.com Page 6


And the DAQ970A/DAQ973A measures and converts 12 different input signals:

 Temperature with thermocouples, RTDs, and thermistors


 dc and ac volts
 2- and 4-wire resistance
 Frequency and period
 Capacitance and diode
 Strain

What’s more, each channel is independently configurable. This means you can configure channel 1 for
dcV, channel 2 for a K-type thermocouple, and channels 3 and 13 for a 4-wire RTD measurement, all on
the same module, all in a single scan. For custom linear conversions, use the Mx+B scaling function on
any channel. You can even display a custom engineering label like RPM or PSI to identify your
measurement units.

Versatile alarms
Alarms are available on a per-channel basis as well. Enter a high limit, a low limit, or both. The
DAQ970A/DAQ973A compares each reading to its limits and flags any out-of-range measurements. You
can assign one of four TTL alarm outputs to any input channel to trigger external alarm lights, sirens, or
send a TTL pulse to your control system, all without a PC connected.

Scanning made simple


The DAQ970A/DAQ973A automatically builds a scan list that includes all configured inputs (even digital
inputs from the Keysight multifunction module) in ascending order by channel number. You can pace
scans by setting the DAQ970A/DAQ973A internal timer for automatic scanning at a specific interval, by
manually pressing a front-panel button, or by sending a software command or external TTL trigger pulse.

Monitor any input


A special display mode monitors a selected input channel, continuously updating the display with new
readings - even during a scan. Or, when using the DAQ970A/DAQ973A with built-in LAN, the channels
can be monitored over the network using the graphical Web interface. It’s great for keeping an eye on a
key input, or for troubleshooting your system before a test.

Non-volatile memory and USB flash drive adds convenient, portability


All readings are automatically time-stamped and stored in a nonvolatile 1 million reading memory -
enough memory to hold more than a week’s worth of data (20 channels scanned every five minutes). The
nonvolatile memory holds your data even after power is removed, so you can use the
DAQ970A/DAQ973A to collect data at a remote location for later uploading to a PC. Or if you need even
more memory, the DAQ970A/DAQ973A USB port can be used to log data directly to a USB flash drive or
to copy the data from the reading memory without being connected to a computer.

A powerful, flexible data acquisition system for automated test


The DAQ970A/DAQ973A gives you the resolution, accuracy, repeatability, and speed you’ve come to
expect from a Keysight data acquisition system. It provides the measurement muscle you need, along
with signal routing and control capability, in a flexible, modular format that can grow and change to match
your varied applications.

Find us at www.keysight.com Page 7


Powerful measurements
The internal 6½ digit DMM brings the power and performance of a world-class stand-alone DMM to the
DAQ970A/DAQ973A, but at a fraction of the cost and in a fraction of the space. It’s as accurate as the
best bench DMM available: 0.003% basic 1-year dcV accuracy, 0.05% basic 1-year acV accuracy, and
0.004% basic 1-year resistance accuracy. Our patented A-D technology offers incredible linearity (2 ppm
of reading +1 ppm of range) along with 22 bits of real resolution. And since it is an integrating A/D, it
provides excellent noise rejection as well - a nice change from noisy and sampling A/Ds. No more
averaging lots of samples just to see the real data you wanted. And if you need high scan rates, the
DAQ970A/DAQ973A is capable of delivering fully converted measurements at speeds up to 450 ch/s.
The input section of the DMM is optically isolated and shielded from the DAQ970A/DAQ973A earth-
referenced circuitry and computer interface, offering up to 300 V of input isolation. This is important for
reducing ground loop and common mode voltage errors associated with long wiring runs and floating
measurement sources.

Flexible functionality
The DMM is installed inside the chassis rather than in one of the slots, leaving all three mainframe slots
free for switch and control modules. You can choose from eight different modules to get the precise
functionality you need now—while giving you flexibility for future expansion.

The internal DMM gives you the flexibility to measure 12 types of inputs easily and inexpensively. The
built-in signal conditioning and conversion routines turn raw inputs directly into real information. Each
measurement channel is independently configurable, so you can set different measurement functions,
scale factors and alarm limits, even on adjacent channels. Advanced measurement features such as
offset compensation, variable integration time, and delay are also selectable on a per-channel basis.

Software drivers
Your months of test system software development time need not go to waste. Software drivers that
support (C, C#, Visual Basic, Visual Studio), LAN, USB and GPIB on the DAQ970A/DAQ973A and SCPI
programming language make integration even easier.

The functionality you need


We put a lot of thought into defining and designing the modules for the DAQ970A/DAQ973A in order to
cover a broad spectrum of switching and signal routing requirements with fewer modules. The result?
Simplified ordering and easier configuration. And while we were at it, we improved performance and
density. The DAQ970A/DAQ973A modules can switch from microvolts to 300 volts, dc to 2 GHz, and with
densities as high as 120 single-ended channels or 96 matrix cross-points per frame. Plus, simple control
capabilities like analog outputs, open collector digital outputs, and isolated Form-C relays for controlling
higher-powered devices are available.

Easy scanning
The DAQ970A can easily scan with external instruments. It builds a scan list that includes all enabled low
frequency multiplexer inputs. Scans are controlled with the external channel advance input, or with the
front panel Step key.

Find us at www.keysight.com Page 8


Connect to the company network
With the LAN interface, the instrument is easily connected to the company network to collect
measurement data to a central database, remotely access the instrument’s set up, or monitor
measurement data from anywhere on the network.

Keysight quality
We know you can’t afford instrument downtime due to hardware failures and unscheduled maintenance.
That’s why our engineers designed reliability into the DAQ970A/DAQ973A: A rugged enclosure, state-of-
the-art surface mount construction throughout, reduced parts counts, and rigorous and thorough testing
on all aspects of the product.

Take the guesswork out of relay maintenance


The DAQ970A/DAQ973A uses our proprietary relay maintenance system to help you to predict relay end-
of-life and avoid costly production line downtime. It automatically counts every individual switch closure
and stores it in nonvolatile memory on each module. You can query the total number of cycles on any
individual channel so you can schedule maintenance and avoid erratic end-of-life failures.

Find us at www.keysight.com Page 9


Customize your Keysight DAQ970A with plug-in modules
A complete selection of plug-in modules gives you high quality measurement, switching, and control
capabilities to choose from. Modules include both low-frequency and RF multiplexers, a matrix switch, a
general-purpose switch, a multifunction module and a digitizer module that includes digital input/output,
analog output, and totalizer capabilities. You can mix and match modules to get just the functionality you
need right now—then change or add more channels later as your application grows. Modules for the
DAQ970A/DAQ973A are designed to make your testing easier, faster, and more reliable. Here’s how:

Higher throughput
Our unique architecture incorporates a high-performance microprocessor on each module, offloading the
mainframe processor and minimizing backplane communications for faster throughput.

More channels in less space


Surface mount construction and a highly integrated design minimize the space required for relay drive
and interface circuitry. High density on-module connectors save both board and connector space
normally required by a terminal block. We use the latest technology to squeeze the most out of the
remaining board space, giving you up to 40 single-ended channels in roughly the same space used by
many data acquisition system terminal blocks.

Convenient connections
On-module screw-terminal connectors make wiring more convenient. Built-in strain relief cable routing
and cable tie points keep your wiring secure and safe from accidental tugs and pulls. An internal analog
bus routes signals from any of the low frequency multiplexers directly to the internal DMM, without the
need for external connections.

Find us at www.keysight.com Page 10


Keysight DAQ970A/DAQ973A modules-at-a-glance selection guide
Model description Type Speed Max Max Bandwidth Thermal Comments
(ch/sec) volts amps offset
DAQM900A 2-wire solid state 450 120 V 20 mA 10 MHz < 4 µV Built-in cold junction
20 ch Multiplexer (4-wire selectable) reference

DAQM901A 2-wire armature 80 300 V 1A 10 MHz 0 µV Built-in cold junction


20 ch Multiplexer (4-wire selectable) reference
+ 2 current channels 2 additional current
channels
(22 total)

DAQM902A 2-wire reed 250 300 V 50 mA 10 MHz < 4 µV Built-in cold junction
16 ch Multiplexer (4-wire selectable) reference

DAQM903A SPDT/form C 120 300 V 1A 10 MHz < 1 µV


20 ch Actuator/GP
Switch

DAQM904A 2-wire armature 120 300 V 1A 10 MHz < 1 µV


4 x 8 Matrix

DAQM905A Common low 60 42 V 0.7 A 2 GHz < 4 µV 1 GHz bandwidth through


Dual 4 ch RF Mux (unterminated) BNC-to-SMB adapter cable
50 Ω

DAQM907A Two 8-bit digital I/O 42 V 400 Open drain


Multifunction Module ports mA

26-bit event counter 42 V 100 kHz Selectable input threshold

Two 16-bit analog ±12V 10 mA dc Max 40 mA total output per


outputs frame

DAQM908A 1-wire armature 100 300 V 1A 10 MHz No four-wire measurements


40 ch Single-Ended (common low)
Mux

DAQM909A Simultaneous 800 kSa/s1 36 Vpk 24-bit resolution, differential


4 ch Digitizer sampling inputs

1. Sa mpling rate.

Find us at www.keysight.com Page 11


Simplify your data gathering and analysis with Keysight BenchVue DAQ software
(now included with DAQ970A/DAQ973A purchase)
Do you want PC-based data logging capability, but don’t want to spend hours programming? BenchVue
DAQ software is the answer. This Windows-based application is designed to make it a snap to use your
PC for gathering and analyzing measurements. Use it to set up your test, acquire and archive
measurement data, and perform real-time display and analysis of the incoming measurements.

A familiar spreadsheet environment makes it easy to configure and control your tests. And a rich set of
colorful graphics provides many options for analyzing and displaying your data—all with point-and-click
ease. Set up multiple graphics using strip charts, histograms, bar and scatter charts, individual channel
results, and more. And of course, you can use BenchVue DAQ to easily move data to other applications
for further analysis, or for inclusion in your presentations and reports.

When used with DAQM909A digitizer, you can access these advanced measurement capabilities:

 Frequency domain measurements: FFT, THD, THD+N, SINAD


 Frequency domain chart sample size up to 1M samples
 Time domain oscilloscope-like measurements
 Time domain chart sample size up to 50M samples

Find us at www.keysight.com Page 12


Data logging feature checklist
 From 1 to 120 channels of analog input
 Measurements include dc volts, ac volts, thermocouple, thermistor and RTD temperature
measurements, 2- and 4-wire Ohms, dc current, ac current, frequency, and period and capacitance
 6½ digits (22 bits) of resolution
 50,000 readings/s reading speed
 1 million reading nonvolatile memory including time-stamp
 Scaling and alarms available on each channel
 Full-featured front panel for stand-alone configuration, troubleshooting, and data viewing
 Nonvolatile storage for five complete instrument states
 Built-in LAN or USB interfaces to support remote data logging applications
 BenchVue DAQ software for configuration and data analysis included

ATE feature checklist


 3-slot card-cage with 6½ digit (22 bit) internal DMM 0.003% basic 1-year dcV accuracy; 0.05% acV
accuracy
 Up to 120 single-ended measurements or 96 matrix cross-points in a 3½” high, half-rack instrument
 Nine switch and control modules include low- frequency and RF multiplexers, 4-ch digitizer, matrix
and actuation switches, digital input and output, analog output, and event recording
 Scan rates up to 450 ch/s
 BenchVue software and graphical web interface to speed up test development and monitor tests
remotely
 Relay maintenance feature for system maintenance

Get better measurements with built-in signal conditioning


The Keysight DAQ970A/DAQ973A architecture offers advantages over other data acquisition solutions
which rely on external or plug-in signal conditioning modules for handling functions other than dcV:

 Minimizes external wiring and the resultant potential for noise and errors to enter your system
 Reduces hidden costs and overall system cost by avoiding unnecessary cables, breakout boxes and
signal conditioning elements
 Simplifies your configuration - for faster, easier setup - with fewer connections and components
 Takes the guesswork out of error analysis.
 Measurement accuracies are specified to include all system-related errors
 Improves reliability, with fewer interconnects and fewer parts that can fail

Find us at www.keysight.com Page 13


Spec interpretation guide
The following pages list the technical specifications for the Keysight DAQ970A/DAQ973A Data
Acquisition System and its modules. The explanations and examples below are helpful in understanding
how to interpret these specifications:

 Measurement accuracy is specified as percent of reading plus percent of range, where reading is the
actual measured value and range is the name of the scale (1 V, 10 V, etc.) - not the full-scale value
(1.2 V, 12 V, etc.).
 DMM measurement accuracies include all switching errors. Switching errors are also listed separately
in the module specifications section. Temperature measurement accuracies include ITS-90
conversion errors. The thermocouple accuracies include the reference junction error as well.
 Accuracies are listed as either 24-hour, 90-day, or 1-year specifications. This refers to the length of
time since the instrument’s last calibration. Use the specification that matches your calibration cycle.
The 24-hour specifications are useful for determining short-term relative performance.

Example 1: Basic dcV accuracy


Calculate the accuracy of the following measurement: 9 V dc input 10 V dc range 1-year accuracy
specifications Normal operating temperature (18 - 28°C)

From the following page, the 1-year accuracy is: 0.003% of reading + 0.0004% or range

Which translates into: (0.003/100 x 9 V)+ (0.0004/100 x 10 V) = 310 µV

For a total accuracy of: 310 μV/9 V = 0.0034%

Example 2: Extreme operating temperature


When the DAQ970A is used outside of its 18 - 28°C temperature range, there are additional temperature
drift errors to consider. Assume the same conditions in Example 1, but at a 35°C operating temperature.

The basic accuracy is again: 0.003% of reading + 0.0004% of range = 310 µV

Now, multiply the 10 V temperature coefficient from the DC voltage specifications table by the number of
degrees outside of operating range for additional error (non ACAL):

(0.0005% reading + 0.0001% range) /°C x (35 - 28°C)

= (0.0005% reading + 0.0001% range) /°C x 7°C

= 0.0035% reading + 0.0007% range = 385 µV

Total error is then: 310 µV + 385 µV = 695 µV or 0.0077%

Find us at www.keysight.com Page 14


Example 3: Thermocouple measurement accuracy
Calculating the total thermocouple reading error is easy with the DAQ970A/DAQ973A - just add the listed
measurement accuracy to the accuracy of your transducer. Switching, conversion, and reference junction
errors are already included in the measurement specification.

For this example, assume a J-type thermocouple input reading 150°C.

From the temperature characteristics table, total error is: Thermocouple probe accuracy + 1.0°C

The probe vendor specifies accuracy of 1.1°C or 0.4%, whichever is greater.

Total error is then: 1.0ºC + 1.1ºC = 2.1ºC total, or 1.4%

Example 4: acV accuracy


The acV function measures the true RMS value of the input waveform, regardless of waveshape. Listed
accuracies assume a sinewave input. To adjust accuracies for non-sinusoids, use the listed crest factor
adder.

For this example, assume a ± 1 V square wave input with 50% duty cycle and a 1 kHz frequency.

Accuracy for 1 V, 1 kHz sinusoid is: 0.05% reading + 0.02% range = 0.7 mV or 0.07%

Find us at www.keysight.com Page 15


DAQ970A/DAQ973A Accuracy Specifications ± (% of reading + % of range)
DC Voltage and Resistance. Automated Calibration (ACAL) capable.
Non ACAL With ACAL
24 Hour4 90 Days 1 Year 2 Year
Range3 Temperature Temperature
Tcal ± 1 ºC Tcal ± 5 ºC Tcal ± 5 ºC Tcal ± 5 ºC
Coefficient/ºC5 Coefficient/ºC6

Accuracy ± (% of reading + % of range)1

DC voltage2

100 mV 0.0030 + 0.0050 0.0040 + 0.0060 0.0050 + 0.0060 0.0065 + 0.0060 0.0005 + 0.0005 0.0002 + 0.0005

1V 0.0015 + 0.0006 0.0025 + 0.0006 0.0035 + 0.0006 0.0050 + 0.0006 0.0005 + 0.0001 0.0002 + 0.0001

10 V 0.0010 + 0.0003 0.0021 + 0.0004 0.0030 + 0.0004 0.0045 + 0.0004 0.0005 + 0.0001 0.0002 + 0.0001

100 V 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0055 + 0.0006 0.0005 + 0.0001 0.0002 + 0.0001

300V 0.0020 + 0.0020 0.0035 + 0.0020 0.0040 + 0.0020 0.0055 + 0.0020 0.0005 + 0.0001 0.0002 + 0.0001
2,7
Resistance

100 Ω 0.0030 + 0.0050 0.0050 + 0.0060 0.0060 + 0.0060 0.0070 + 0.0060 0.0006 + 0.0005 0.0002 + 0.0005

1 kΩ 0.0020 + 0.0007 0.0030 + 0.0007 0.0040 + 0.0007 0.0050 + 0.0007 0.0006 + 0.0001 0.0002 + 0.0001

10 kΩ 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001

100 kΩ 0.0020 + 0.0005 0.0030 + 0.0005 0.0040 + 0.0005 0.0050 + 0.0005 0.0006 + 0.0001 0.0002 + 0.0001

1 MΩ 0.0020 + 0.0005 0.0060 + 0.0005 0.0070 + 0.0005 0.0080 + 0.0005 0.0010 + 0.0002 0.0002 + 0.0002

10 MΩ 0.010 + 0.001 0.020 + 0.001 0.025 + 0.001 0.030 + 0.001 0.0030 + 0.0004 0.0030 + 0.0004

100 MΩ 0.100 + 0.001 0.250 + 0.001 0.350 + 0.001 0.400 + 0.001 0.1000 + 0.0001 0.0100 + 0.0001

1000 MΩ 2.000 + 0.001 2.500 + 0.001 3.500 + 0.001 4.000 + 0.001 1.0000 + 0.0001 0.1000 + 0.0001

1. Specifications are for 60-minute war m- up, integration setting of 10 or 100 NPLC, auto-zero on, AC slo w filter. ACAL run within the last 2 days
& 901 module connectivity.
2. These specifications include offset characteristics contributed by the 901 Module & backpl ane (which adds DCV + 2 uV, 4- wire oh ms + 2 mΩ )
3. 20% over range on all ranges, except 300 VDC, 300 VAC, 1 ADC, 1 AAC, an d di ode test have 0%.
4. Relative to calibrati on standards. 24 hrs spec only applicabl e when calibrated wi th the sa me DAQM901 module that is being used to verify
spec.
5. Add this for each °C outside the last ACAL ± 5 °C.
6. Add this for each °C outside the last external standards cali bration ± 5 °C.
7. Specifications are for 4-wire oh ms fun ction or 2-wire ohms using math null for offset. W ithout math null, add 2 Ω additional error in 2-wire
oh ms function. The 100M and 1G oh m ranges are 2-wire only. See the manua l for low power oh ms specification and measure ment currents.

Find us at www.keysight.com Page 16


DC Current and Other DC Functions
24 Hour3 90 Days 1 Year 2 Year Temperature
Range2
Tcal ± 1 ºC Tcal ± 5 ºC Tcal ± 5 ºC Tcal ± 5 ºC Coefficient/ºC5

Accuracy ± (% of reading + % of range)1

Characteristics – typical : DC current

1 µA 0.017 + 0.005 0.040 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0010

10 µA 0.015 + 0.002 0.040 + 0.002 0.050 + 0.002 0.060 + 0.002 0.0015 + 0.0006

100 µA 0.012 + 0.001 0.040 + 0.001 0.050 + 0.001 0.060 + 0.001 0.0015 + 0.0004

Specifications: DC current

1 mA 0.007 + 0.003 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0015 + 0.0005

10 mA 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.060 + 0.020 0.0020 + 0.0020

100 mA 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.060 + 0.005 0.0020 + 0.0005

1A 0.050 + 0.006 0.070 + 0.010 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010

Characteristics – Typical: Diode Test4

5V 0.002 + 0.010 0.008 + 0.010 0.010 + 0.010 0.012 + 0.020 0.0010 + 0.0010

Specifications: AC voltage
True RMS AC Voltage6,7

100 mV, 1 V, 10 V, and 100 V ranges

3 Hz – 5 Hz 0.50 + 0.02 0.50 + 0.02 0.50 + 0.02 0.50 + 0.02 0.010 + 0.003

5 Hz – 10 Hz 0.10 + 0.02 0.10 + 0.02 0.10 + 0.02 0.11 + 0.02 0.008 + 0.003

10 Hz – 20 kHz 0.02 + 0.02 0.04 + 0.02 0.05 + 0.02 0.06 + 0.02 0.007 + 0.003

20 kHz – 50 kHz 0.05 + 0.03 0.06 + 0.03 0.07 + 0.03 0.08 + 0.03 0.010 + 0.005

50 kHz – 100 kHz 0.15 + 0.05 0.15 + 0.05 0.15 + 0.05 0.15 + 0.05 0.060 + 0.008

100 kHz – 300 kHz 1.00 + 0.1 1.00 + 0.1 1.00 + 0.1 1.00 + 0.1 0.200 + 0.020

300 V range

3 Hz – 5 Hz 0.50 + 0.05 0.50 + 0.06 0.50 + 0.06 0.50 + 0.06 0.010 + 0.008

5 Hz – 10 Hz 0.10 + 0.05 0.10 + 0.06 0.10 + 0.06 0.11 + 0.06 0.010 + 0.008

10 Hz – 20 kHz 0.02 + 0.05 0.04 + 0.06 0.05 + 0.06 0.06 + 0.06 0.010 + 0.008

20 kHz – 50 kHz 0.05 + 0.09 0.06 + 0.09 0.07 + 0.09 0.08 + 0.09 0.010 + 0.0012

50 kHz – 100 kHz 0.15 + 0.15 0.15 + 0.15 0.15 + 0.15 0.15 + 0.15 0.060 + 0.020

100 kHz – 300 kHz 1.00 + 0.3 1.00 + 0.3 1.00 + 0.3 1.00 + 0.3 0.200 + 0.050

Specifications: True RMS AC Current7’8

100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges

3 Hz – 5 kHz 0.07 + 0.04 0.09 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006

5 kHz – 10 kHz (typical) 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.030 + 0.006
(typical)
1. Specifications are for 60-minute war m-up, i ntegration setting of 10 or 100 NPLC, auto-zero on, AC slo w filter. ACAL run within the last 2 days &
901 module connectivity.
2. 20% over range on all ranges, except 300 VDC, 300 VAC, 1 ADC, 1 AAC, and diode test have 0%.
3. Relative to calibration standards. 24 hrs spec only applicable when calibrated with the sa me D AQM901 module th at is being used to verify spec.
4. Specifications are for the voltage measured at the input terminals. The 1 mA te st current is typical. Vari ation in the current source wil l create
so me variation in the voltage drop across a di ode juncti on.
5. Add this for each °C outside Tcal ± 5 °C.
6. Specifications are for sinewave input >0.3% of range and > 1 mVr ms. For the 300 Vr ms range inputs >1 % of rang e. All VAC ranges limited to 8 x
10 ^7 Volt–Hz.
7. Low-frequency perfor mance: three filter settings are available: 3 Hz, 20 Hz, 200 Hz. Frequencies greater than these filter settings are specified
with no additional errors.
8. Specifications are for sinewave input > 1% of range and > 10 µA AC.

Find us at www.keysight.com Page 17


Characteristics: Capacitance (Typical)
Capacitance1

1.0000 nF 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.05 + 0.05

10.000 nF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01

100.00 nF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01

1.0000 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01

10.000 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01

100.00 µF 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01

Specifications: Frequency
Frequency2,3

100 mV, 1V, 10V, 100 V, and 300 V ranges3

3 Hz – 10 Hz 0.070 0.070 0.070 0.070 0.0002

10 Hz – 100 Hz 0.030 0.030 0.030 0.030 0.0002

100 Hz – 1 kHz 0.003 0.006 0.007 0.010 0.0002

1 kHz – 300 kHz 0.002 0.005 0.007 0.009 0.0002

Square wave 0.001 0.004 0.006 0.008 0.0002

Additional frequency errors (% of reading)2

Aperture (resolution/range) 1 second 0.1 second 0.01 second 0.001 second

3 Hz – 40 Hz 0 0.100 0.160 0.160

40 Hz – 100 Hz 0 0.030 0.160 0.160

100 Hz – 1 kHz 0 0.020 0.200 0.200

1 kHz – 300 kHz 0 0.004 0.030 0.240

Square wave4 0 0.000 0.000 0.003

1. Specifications are for using Math Null zeroing. High dissipation factor capacitors may sho w different results than a single frequency
measure ment. Film capacitors usually have lower dissipation factors than other dielectrics.
2. Specifications are for sine wave input unless stated otherwise.
3. Input > 100 mV. For 10 mV to 100 mV i nputs, multiply % of reading error x10. Amplitude 10%–120% of range e xcept 30%–100% for the 300
ACV range. Specifications are for 1-second gate time (7-digits).
4. Square wave input specified for 10 Hz – 300 KHz for 1 second aperture. For s horter apertures the mini mu m frequenc y requires > 2 cycles.

Find us at www.keysight.com Page 18


Characteristics: Temperature (Typical)
Thermocouple DAQ970A/DAQ973A Probe1 Ref junction
temperature ranges & DMM accuracy

E -150 ºC to 1000 ºC 1.0 ºC

J -150 ºC to 1200 ºC 1.0 ºC

K -100 ºC to 1200 ºC 0.9 ºC

N -100 ºC to 1300 ºC 1.0 ºC

R 300 ºC to 1760 ºC 0.5 ºC

T -100 ºC to 400 ºC 0.9 ºC

Thermocouple DAQ970A/DAQ973A Probe1 Ref junction


extended & DMM accuracy
temperature ranges

E -200 ºC to -150 ºC 1.4 ºC

J -210 ºC to -150 ºC 1.6 ºC

K -200 ºC to -100 ºC 1.7 ºC

N -200 ºC to -100 ºC 2 ºC

R -50 ºC to 300 ºC 2.9 ºC

T -200 ºC to -100 ºC 1.7 ºC

RTD Probe DMM2

PT100 (DIN/IEC 751) [3] 0.05 ºC

Thermistor

2.2 K, 5 K, 10 K [3] 0.1 ºC

Other Measurement Functionality

Strain

Direct strain – measured as 2 or 4 wire ohms (customer enters gauge parameters)

Bridge strain – voltage measurement (1/4, ½ & full bridge, Bending & Poisson configurations, requires external power supply & bridge completions)

1. Thermocouple probe accuraci es would co me fro m ther mo couple suppli er and should be selected based on temp ran ges required.
2. DMM accuracy a ssu mes measure ment with DAQM901 A. For module Oh ms me asure ments - see respective s witch mo dules.
3. Accuracy specificati on depends on Mfg specificati ons of device.

Find us at www.keysight.com Page 19


Single Channel Reading Rates to I/O or Internal Memory (Nominal)
DAQ970A DAQ970A DAQ973A DAQ973A
Int Memory (1M) LAN, USB Int Memory (1M) LAN, USB, GPIB

Single ch ASCII rdgs – DCV, Ohms 50,000/s 50,000/s 50,000/s 50,000/s

Single ch ASCII rdgs – ACV, ACI 250/s 250/s 250/s 250/s

Single ch ASCII rdgs – Freq, period 800/s 800/s 800/s 800/s

Single ch (chg Scale) 100/s 100/s 100/s 100/s

Single ch (chg Func) 50/s 50/s 50/s 50/s

Scanning Measurement Rates to I/O or Internal Memory (Nominal)


DAQ970A DAQ970A DAQ973A DAQ973A
Int Memory (1M) LAN, USB Int Memory (1M) LAN, USN, GPIB

Scanning DCV or Ohms channels

DAQM900A (INIT, MEAS, FETch) 450 450 500 500

DAQM901A 80 80 80 80

DAQM902A 250 250 250 250

DAQM902A (INIT, FETch) 240 240 240 240

DAQM902A (MEAS) 240 240 240 240

DAQM902A (scale, alarm) 220 220 220 220

DAQM902A (Alt channel) 80 80 80 80

DAQM908A 80 80 80 80

Scanning ACV channels

DAQM900A 90 90 90 90

DAQM901A 50 50 50 50

DAQM902A 90 90 90 90

DAQM908A 50 50 50 50

Scanning temperature – thermistor or TC channels

DAQM900A 150 150 150 150

DAQM901A 50 50 50 50

DAQM902A 150 150 150 150

Scanning digital in TOTALIZER channels

DAQM907A (Digital IN) 275 275 275 275

DAQM907A (Totalizer) 240 240 240 240

Find us at www.keysight.com Page 20


Data out of Memory (FETch of 50K reading) (Nominal)
DAQ970A/DAQ973A DAQ970A/DAQ973A
Single Channel DAQ973A over GPIB
over USB over LAN

Readings 55,000/s 120,000/s 2,500/s

Readings (with timestamp) 35,000/s 60,000/s 1,500/s

Readings (format option ON) 25,000/s 50,000/s 1,000/s

Noise performance for DC voltage, DC current, and resistance (Measured)


Integration time Digits1 RMS Noise adder (% of range + fixed base)2

DC Volt Ohms DC current3

100 PLC/1.67 s (2 s) 6.5 0 0 0

10 PLC/167 ms (200 ms) 6.5 0 0 0

1 PLC/16.7 ms (20 ms) 6.5 0.0001 + 0.5 µV 0.0001 + 0.5 mΩ 0.0006 + 0.02 nA

0.2 PLC/3 ms (3 ms) 6.5 0.0005 + 3 µV 0.0010 + 10 mΩ 0.0050 + 5 nA

0.06 PLC/1 ms (1 ms) 6 0.0020 + 3 µV 0.0020 + 10 mΩ 0.0070 + 10 nA

0.02 PLC/400 µs (300 µs) 6 0.0020 + 3 µV 0.0020 + 10 mΩ 0.0070 + 10 nA

1. For DCV on the 10-volt range with zer o volts input and auto zero on.
2. RMS noise adder measured with zero volts input and auto zero on.
3. The follow DCI ranges have the se additi onal multi pliers: The 10 mA by 5 x, th e 100 mA b y 2 x.

DC and AC Current Burden Voltage at Full Scale (Measured)


DC current range Burden voltage

1 µA < 0.0011 V

10 µA < 0.011 V

100 µA < 0.11 V

1 mA < 0.11 V

10 mA < 0.027 V

100 mA < 0.27

1A < 0.7

A/D linearity (Measured)


0.0001% of reading + 0.0001% of range

Find us at www.keysight.com Page 21


Measurement Characteristics
DC voltage

Measurement method Keysight patented continuously integrating multi-slope IV A/D converter

A/D Linearity 0.0001% of reading + 0.0001% of range

Input resistance

0.1 V, 1V, 10V range Selectable 10 MΩ or >10 GΩ

100 V, 300 V range 10 MΩ ± 1%

Input bias current < 30 pA at 25 ºC

Input protection 300 V on all ranges

True RMS AC voltage

Measurement type AC-coupled True RMS. Measures the AC component of the input.

Measurement method Digital sampling with anti-alias filter

Maximum input 250 DCV, 300 Vrms

Input impedance 1 MΩ ± 1%, in parallel with < 100 pF

Input protection 300 Vrms

DC current

Input protection 1.6A 250V fuse on DAQM901A

True RMS AC current

AC measurement method Digital sampling with anti-alias filter

Input protection 1.6A 250V fuse on DAQM901A

AC crest factor and peak input

10:1 maximum crest factor, (3:1 at full-scale). Measurement bandwidth limited to 300 kHz for signal plus
Crest factor
harmonics.

Peak input 300% of range or maximum input

Will select higher range if peak input overload is detected during auto range. Overload is reported in
Overload ranging
manual ranging.

Resistance

Measurement method Will select 4-wire or 2-wire ohms. Current source referenced to LO input.

Maximum lead resistance (4-wire ohms) 10% of range per lead for 100 Ω and 1 kΩ ranges. 1 kΩ on all other ranges

Input protection 300 V

Temperature

Thermocouple

Conversion ITS-90 software compensation

Reference junction type Internal, Fixed, and External

Open thermocouple check Selectable per channel. Open > 5 kΩ.

RTD PT100 and PT1000 Transducer Type

Thermistor 44004, 44006, 44007 Series

Find us at www.keysight.com Page 22


Measurement Characteristics Continued
Measurement noise rejection

60 Hz (50 Hz) for 1 kΩ LO lead unbalance (± 500 V peak maximum)


- DCV CMRR: 140 dB
- ACV CMRR: 70 dB

Integration time Normal mode rejection1

≥ to 1 PLC 60 dB2

< 1 PLC 0 dB

External trigger latency <1.4ms (with display ON and monitor ON <100ms)

External trigger jitter <16us (with display ON and monitor ON <18ms)

Frequency and period

Measurement method Reciprocal-counting technique. Measurement is AC-coupled using AC measurement functions.

Voltage ranges 100 mVrms full scale to 300 Vrms. Auto or manual ranging.

Gate time 1 ms, 10 ms, 100 ms, or 1 s

All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals.
Measurement considerations
Shielding inputs from external noise pickup is critical for minimizing measurement errors.

1. For power-line frequency ± 0.1%


2. For power-line frequency ± 1%, the NMR is 40 dB. For ± 3%, use 30 d B.

Find us at www.keysight.com Page 23


General Characteristics
Line power

Power supply (Mains supply voltage fluctuation


100 – 240 Vac
is not allowed to exceed ± 10%)

Power line frequency 50/60/400 Hz

Power consumption 45 watts

Environment

Operating temperature Full accuracy at 0 to 55 ºC

Storage temperature -40 to +70 ºC

Full accuracy to 80% RH, at 40 ºC non-condensing


Operating humidity
(Humidity degrades linearly to 50% RH at 55 ºC non-condensing)

Operating altitude Up to 3,000 m

Mechanical

Rack dimensions (W x H x D): 212.6 mm x 88.5 mm x 348.3 mm

Bench dimensions (W x H x D): 261.0 mm x 103.7 mm x 378.0 mm

Weight DAQ970A/DAQ973A: 4.2 kg

Regulatory

EN 61010-1:2010 (3rd Edition)


ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
Safety
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current revisions
Pollution Degree 2

IEC 61326
EN 61326
CISPR
EMC
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current revisions

Acoustic noise - Nominal 35 dBA

Triggering conditions

External Delay, Jitter, Minimum Pulse Width, Maximum rate

Voltmeter complete Polarity, Pulse Width

Computer interfaces LXI (rev 1.4) USB, LAN, GPIB (DAQ973A only)

Language SCPI-1999, IEEE-488, 34970A/34972A compatible

USB 2.0 – high speed mass storage (MSC) class device


Front panel USB host port
Capability: Import/export instrument configuration files, save volatile readings and screen captures

Software BenchVue DAQ (Included with DAQ970A/DAQ973A)

Find us at www.keysight.com Page 24


Switch Modules Characteristics
The Keysight DAQ970A/DAQ973A accuracy specifications already include the switching offset and
reference junction errors shown below. These errors are listed separately for determining system error
with external measurement devices. Up to three modules, in any combination, can be inserted into a
single mainframe. The DAQ970A/DAQ973A’s internal DMM connections are accessible only through the
DAQM900A, DAQM901A, DAQM902A, and DAQM908A low-frequency multiplexers. On-module screw
terminals accept wire sizes from 16-gage to 22-gage. Twenty-gage wire is recommended for high channel
count applications. The DAQM905A RF Multiplexer uses SMB connectors. A standard set of (10) BNC-to-
SMB adapter cables is provided with each RF module for convenient BNC connections.

Multiplexer Multiplexer Multiplexer Actuator Matrix RF multiplexer Multifunction

DAQM900A DAQM901A DAQM902A DAQM903A DAQM904A DAQM905A DAQM908A

Number of channels 20 20 + 2 16 20 (SPDT) 4 x 8 (2 wire) Dual 1x4 (50Ω) 40 (1 wire)

Scanning speed 450 ch/s 80 ch/s 250 ch/s 80 ch/s

Open/close speed 120 ch/s 120 ch/s 120 ch/s 120 ch/s 60 ch/s 70 ch/s

Internal DMM measurement functions supported

Voltage AC/DC Yes3,4 Yes Yes No No No Yes

Current AC/DC No Yes No No No No No

Frequency/period Yes Yes Yes No No No Yes

Ohms – 2 wire Yes2 Yes Yes5 No No No Yes

Ohms – 4 wire Yes2 Yes Yes5 No No No No

Thermocouple Yes Yes Yes No No No No

RTD 2-wire No Yes Yes No No No Yes

RTD 4-wire No Yes Yes No No No No

Thermistor No Yes Yes No No No Yes

Capacitance No Yes Yes No No No Yes

Characteristics – typical: Input

Voltage (dc, ac rms) 120 V 300 V 300 V 300 V 300 V 42 V 300 V

Current (dc, ac rms) 0.02 A 1A 50 mA 1A 1A 0.7 A 1A

Power (W, VA) 2.4 W 50 W 2W 50 W 50 W 50 W 50 W

Characteristics – typical: DC

Offset voltage
< 4 µV 0 µV6 < 4 µV < 1 µV < 1 µV < 4 µV < 1 µV
(incremental to DMM specs)

Offset resistance (4-wire)


< 4 mΩ 0 mΩ6 < 4Ω
(incremental to DMM specs)

Initial closed channel R < 50 Ω < 1.0 Ω < 1.0 Ω < 0.2 Ω < 1.0 Ω < 0.5 Ω < 1.0 Ω

Isolation Ch-Ch, Ch-Earth > 10 GΩ > 10 GΩ > 10 GΩ > 10 GΩ > 10 GΩ > 1 GΩ > 10 GΩ

1. DMM accuracy a ssu mes measure ment with DAQM901 A. For 900 module Ω measure ments - see note 2.
2. The series resistance of the DAQ M900 A limits the use of the 100 Ω and 1 kΩ resistance range. The precision of 2-wire Ω is limited by this on
resistance. Keysight reco mmend s the use of 4-wire Ω for these 2 ranges. Ma xi mu m resistance range of DAQ M900 A is 1 MΩ. T he 10 MΩ and
above ranges are not specified on this module.
3. For AC Volts - The i nput impedance of this module decreases with frequency. A source i mpedance of 5 Ω or less will maintain specifications
over the full frequency range. A source impedance of 50 oh ms or less will maintain specifications to 5 kHz.
4. For DC Volts - Addi tional settling time may be needed for short integration time and higher source impedance.
5. Resistance range ≥ 1 MΩ is Typical.
6. Absolute Offset is <2µV, Incre mental Offset to 901 = 0µV. 2µV included in DCV’s % of range specs. O ffset for this module already included in
speci fications above.

Find us at www.keysight.com Page 25


Switch Modules Characteristics Continued
Multiplexer Multiplexer Multiplexer Actuator Matrix RF multiplexer Multifunction

DAQM900A DAQM901A DAQM902A DAQM903A DAQM904A DAQM905A DAQM908A

Characteristics – typical: AC

Ch-Ch crosstalk (at 1 MHz) NA - 50 dB -55 dB - 60 dB - 50 dB - 110 dB - 25 dB

Capacitance (Hi-Lo) NA < 50 pF < 65 pF 10 pF < 50 pF < 20 pF < 50 pF

Capacitance (Lo-Earth) NA < 150 pF < 135 pF < 80 pF < 80 pF NA < 80 pF

Volt-Hertz limit 10˄8 10˄8 10˄8 10˄8 10˄8 10˄10 10˄8

Characteristics – typical: Other

T/C CRJ accuracy1 0.8 ºC 0.8 ºC 0.8 ºC

Unlimited
Switch life (no load) within FET 100 M 100 M 100 M 100 M 5M 100 M
Bank2

Unlimited
Switch life (rated load)3 within FET 100 K 100 K 100 K 100 K 100 K 100 K
Bank2

Operating temperature All cards: 0 to 55 ºC

Storage temperature All cards: -40 ºC to 70 ºC

Humidity (non-condensing) All cards: 40 ºC to 80% RH

Characteristics – nominal: DAQM905A

AC 10 MHz 100 MHz 500 MHz 1 GHz 1.5 GHz 2 GHz

Insertion loss (db) -0.1 -0.4 -0.6 -1 -1.2 -3

SWR 1.02 1.05 1.2 1.2 1.3 1.4

1. Errors included in DMM measure ment accuracy specifications


2. This module has an Ar mature backplane & 2 W /4 W relays with 100 M operational life (FET life withi n banks is unlimited). Limiting backplane &
2 W /4 W relay cycling is recommended .
3. Applies to resistive loads only.

Find us at www.keysight.com Page 26


Multiplexer selection
Choose between the broad functionality of the DAQM901A, the high-speed scanning of the DAQM902A,
the DAQM900A solid-state module, or the single-ended density of the DAQM908A. These four modules
are the only way to connect to the DAQ970A/DAQ973A internal DMM. They can be used to scan with
external instruments as well. All multiplexer modules employ break-before-make scanning, ensuring only
one closed channel (or channel pair) at a time.

DAQM900A
20-Channel General Purpose Multiplexer

 450 ch/s scanning


 Two- and four-wire scanning
 Built-in thermocouple reference junction
 120 V switching

The DAQM900A module is divided into two banks of 10 two-wire channels each. All 20 channels switch
both HI and LO inputs, thus providing fully isolated inputs to the internal DMM or an external instrument.
During 4-wire resistance measurements, channels from Bank A are automatically paired with channels
from Bank B to provide the source and sense connections. The module has a built-in thermocouple
reference junction to minimize errors due to thermal gradients when measuring thermocouples.

Find us at www.keysight.com Page 27


DAQM901A
20-Channel General Purpose Multiplexer

 80 ch/s scanning
 Two- and four-wire scanning
 Built-in thermocouple reference junction
 300 V switching

The Keysight DAQM901A is the most versatile multiplexer for general purpose scanning. It combines
dense, multifunction switching with 80-channel/second scan rates to address a broad spectrum of data
acquisition applications.

Two- and four-wire channels can be mixed on the same module. Two additional fused inputs (22
channels total) route up to 1 A of current to the internal DMM, allowing ac and dc current measurements
without the need for external shunt resistors.

Find us at www.keysight.com Page 28


DAQM902A
16-Channel High-Speed Multiplexer

 Scanning up to 250 ch/s


 Two- and four-wire scanning
 Built-in thermocouple reference junction

The Keysight DAQM902A employs reed relays to achieve scan rates up to 250 channels per second. Use
this module for high-throughput automated test applications as well as high-speed data logging and
monitoring tasks.

Sixteen two-wire inputs switch up to 300 V. Two and four-wire channels may be mixed on the same
module. User provided shunt resistors are required for current measurements.

Find us at www.keysight.com Page 29


DAQM903A
20-Channel Actuator/General Purpose Switch

 SPDT (Form C) latching relays


 300 V, 1 A actuation and control

This general-purpose switch module has 20 independent single-pole, double-throw (SPDT) relays. Use it
to cycle power to products under test, control indicator and status lights, and to actuate external power
relays and solenoids. Combine it with matrix and multiplexer modules to build custom switch systems. Its
300 V, 1 A contacts can handle up to 50 W, enough for many powerline switching applications.

Find us at www.keysight.com Page 30


DAQM904A
4x8 Two-wire Matrix Switch

 32 two-wire cross-points
 300 V, 1 A switching

The Keysight DAQM904A gives you the most flexible connection path between your device under test
and your test equipment, allowing different instruments to be connected to multiple points on your DUT at
the same time.

Rows or columns may be connected between multiple modules to build 8x8, 4x16 or larger matrices, with
up to 96 cross-points in a single frame.

Find us at www.keysight.com Page 31


DAQM905A
Dual 4-Channel RF Multiplexer 50 Ω

 2 GHz bandwidth
 BNC to SMB adapter cables included

The Keysight DAQM905A RF multiplexers offer broadband switching capabilities for high frequency and
pulsed signals. Use them to route test signals between your device under test and your signal generator,
oscilloscope, spectrum analyzer, or other instrumentation.

The RF multiplexer are arranged as two independent 1x4 multiplexers, each with a common shield and a
switched center conductor. Connections can be made directly to SMB inputs with 2 GHz usable
bandwidth, or to the BNC-to-SMB adapters provided with 1 GHz bandwidth. Multiple banks may be
cascaded together for applications requiring even larger topologies—create a 16:1 multiplexer in a single
frame.

50 Ω MUX typical AC performance graphs

Find us at www.keysight.com Page 32


DAQM907A
Multifunction Module

 16 bits of digital input and output


 100 kHz totalizer input
 Two ± 12 V analog outputs or ± 24 mA outputs

The Keysight DAQM907A allows great flexibility for a variety of sense and control applications. It
combines two 8-bit ports of digital input and output, a 100 kHz gated totalizer, and two ± 12 V or ± 24 mA
analog outputs—all on a single earth-referenced module. The digital inputs and totalizer input may be
included in a scan. Alarm limits for the digital and event counter inputs are evaluated continuously,
capturing and logging alarm conditions even between scans.

On top of that, it has two additional channels (Channel 6 and 7) that provide the ability to sense output
current when sourcing voltage, or sense output voltage when sourcing current.

Find us at www.keysight.com Page 33


DAQM907A Multifunction Module Characteristics
Measurement Functionality

Digital I/O (Port 1, 2) 16 bits, Input or Output, non-isolated

Totalizer 100 kHz Input

Analog outputs 2 channels, ± 12 V or ± 24 mA

Characteristics – typical: [1, 2, 3]

DAC (output & readback) Accuracy

Voltage ± 0.027% of output + 4.4 mV

Current ± 0.115% of output + 4.4 µA

Characteristics – nominal

DIO Port 1, 2 8-bit, input or output, non-isolated

Vin(L) < 0.8 V (TTL)

Vin(H) > 2.0 V (TTL)

Vout(L) < 0.8 V @ lout = -400 mA

Vout(H) > 2.4 V @ lout = 1 mA

Vin(H) max < 42 V with external open drain pull-up

Alarming Maskable pattern match or state change

Speed 4 ms (max) alarm sampling

Latency 5 ms (typical) to DAQ970A alarm output

Read/write speed 95/s

Max count 2˄26 - 1

Totalize input 100 kHz (max) Rising or falling edge, programmable, min H or L time 5µS

Signal level 1 Vp-p (min) 42 Vpk (max)

Threshold 0 V or TTL, jumper selectable

Gate Input TTL-Hi, TTL-Lo, or none

Count reset Manual or Read + Reset

Read speed 85/s

DAC 1, 2 ± 12 V or ± 24 ma non-isolated

Resolution 100 µV, 0.2 µA

Voltage mode current 15 mA max per channel

Current mode compliance 12 V

Settling time 1 ms to 0.01% of output

1. ACAL required withi n the last 30 days to achieve performance ac curacies.


2. ACAL assu me s ± 2 ºC with no additional TC adders.
3. Mainfra me DMM – calibrated through the DAQ M901A module.

Find us at www.keysight.com Page 34


Digital input/output
Use the digital outputs with an external power supply to control microwave switches and attenuators,
solenoids, power relays, indicators, and more. Use the digital inputs to sense limit switch and digital bus
status. There are no complex handshake modes; reads and writes are initiated either from the front panel
or the bus.

Port 1, 2 8-bit, input or output, non-isolated

Vin(L) < 0.8 V (TTL)

Vin(H) > 2.0 V (TTL)

Vout(L) < 0.8 V @ Iout = –400 mA

Vout(H) > 2.4 V @ Iout = 1 mA

Vin(H) max < 42 V with external open drain pull-up

Alarming Maskable pattern match or state change

Speed 4 ms (max) alarm sampling

Latency 5 ms (typical) to DAQ970A alarm output

Read/write speed 95/s

Totalize input
Count events from devices like photo interrupters, limit switches, and Hall-effect sensors.

It keeps an updated total which can be read via the front panel or programmatically at any time. With 26
bits of resolution, it can count events at full speed for nearly 11 minutes without an overflow.

Max count 226 - 1

Totalize input 100 kHz (max) Rising or falling edge, programmable

Signal level 1 Vp-p (min) 42 Vpk (max)

Threshold 0 V or TTL, jumper selectable

Gate Input TTL-Hi, TTL-Lo, or none

Count reset Manual or Read + Reset

Read speed 85/s

Find us at www.keysight.com Page 35


Analog output
Use the two electronically calibrated analog outputs to source bias voltages to your device under test, to
control your analog programmable power supplies, or use the outputs as setpoints for your control
systems. The outputs are programmed directly in volts, either from the front panel or from the bus.

On top of that, it has two additional channels (Channel 6 and 7) that provide the ability to sense output
current when sourcing voltage, or sense output voltage when sourcing current.

DAC 1, 2 ± 12 V or ± 24 mA analog outputs

Resolution 100 uV, 0.2 uA

IOUT 15 mA max per channel

Settling time 1 ms to 0.01% of output

Accuracy ±(% of output + mV)

1 year ± 5 °C Voltage: ± 0.027% output + 4.4 mV

Current: ± 0.115% of output + 4.4 uA

Find us at www.keysight.com Page 36


DAQM908A
40-Channel Single-Ended Multiplexer

 80 ch/s scanning
 Single-wire switching for common low applications
 Built-in thermocouple reference junction

Use the Keysight DAQM908A for the greatest density in common low applications, such as battery test,
component characterization, and benchtop testing. Each module switches 40 one-wire inputs. All two-wire
internal measurements except current are supported. The module low connection is isolated from earth
and can float up to 300 V.

Find us at www.keysight.com Page 37


DAQM909A
4-Channel 24-bit Digitizer

 Simultaneous sampling
 Up to 800 kSa/s at 24-bit
 Time/frequency domain measurement with Benchvue DAQ

The DAQM909A is a four-channel digitizer module capable of providing four simultaneously-sampled


channels with sampling rate as high as 800k samples per second with a maximum resolution of 24 bits.
The input channels of the DAQM909A can be configured as a differential input or a single-ended input
and each of the channels can provide up to 4mA of constant current to power up an external IEPE
transducer.

When three DAQM909A modules are inserted within the same instrument, all 12 channels can be
synchronized to provide simultaneous sampling.

Find us at www.keysight.com Page 38


DAQM909A Continued
Measurement Functionality

DC Volt

ADC per channel 800 kSa/s @ 24-bit

Range (Hi, Lo Input to Earth)1 0.3 Vpk, 1 Vpk, 3 Vpk, 18 Vpk

Range (Differential input) 0.6 Vpk, 2 Vpk, 6 Vpk, 36 Vpk

Range (Differential input – sinewave Vrms)2 0.42 Vrms, 1.4 Vrms. 4.2 Vrms, 25 Vrms

Residual noise (20 Hz – 20 kHz) 7 µVrms, 16 µVrms, 52 µVrms, 175 µVrms

IEPE current source OFF or ON, ON (2mA – 4mA)

IEPE currents source compliance voltage 20 V

Trigger count 1 to 85k

Sample count 1 to 16M

Pre-trigger count 0 to sample count

Trigger arm event (System) IMM, Scan-sweep, Alarm-#

Trigger event (channel) Software, IMM, Level/Slope

Trigger event (same module) External/Slope, Chan-#

Trigger event (module-module) Any Chan-#

External trigger input (one per module) TTL compatible

External trigger delay 0 – 1000s, 1 µs steps

External trigger latency 1-sample

External Trigger-In terminals Screw terminal (≥ 20-gage wire)

Measurement input terminals Screw terminal (≥ 20-gage wire)

Frequency or Time domain anti-alias filters Yes

Time-stamped measurements Yes (Acoustic Start)

Acquisition memory/channel 16M to 48M Samples

Log multiple digitized records (volatile) Yes

Automatic file copy to flash drive Yes

1. The DAQM909 is ali as-protected for out-of-band inputs to 6MHz with a mplitudes < 10dB belo w the range’s ma xi mu m.
2. Balanced differential input, amplitude = range ma x -1d B (and < 12Vr ms), 50k Sa/s, frequency mode, 20kHz analysis bandwidth, DC-coupled.

Find us at www.keysight.com Page 39


DAQM909A Continued
Characteristics: Typical

Accuracy (DC)1 0.017% of readings + 0.004% of range (ACAL)

2
THD (1 kHz) - 103 dB

SFDR (1 kHz)2 - 105 dB

2,3
THD (20 Hz - 20 kHz) - 83 dB

SFDR (20 Hz - 20 kHz)2 - 85 dB

Characteristics: Nominal

Flatness (dc - 20 kHz) 0.1 dB

Input impedance (Hi or Lo to Earth) 1 MΩ | | 400 pF

Single Ended (SE) Lo-Earth 150 Ω

Coupling AC, DC

A/D converter (per channel) 24-bit Delta-Sigma

Sample rate

Time domain 800 k/2˄n n=0,1,2,… 9

Frequency domain 100 k/2˄n n=0,1,2,… 6

Acquisition bandwidth

Time domain (-3 dB) 0.2*Fs or 125 kHz max

Frequency domain (-0.1 dB) 0.4*Fs or 30 kHz max

Analog bandwidth (-3 dB) 125 kHz (measured)

Channel to channel gain match (1 kHz) 0.003 dB

4
Channel to channel phase match (1 kHz) 0.045º

Channel to Channel cross-talk (1 kHz)4 -100 dB

Common mode input voltage

Range (0.3, 1, 3 Vpk) ±8V

Range (18 Vpk range) ± 18 V

Common mode rejection 60 dB (dc – 1 kHz, DIFF in)

Overvoltage protection ± 42 V

1. AC accuracy = DC a ccuracy + Flatness


2. Balanced differential input, amplitude = range ma x -1d B (and < 12Vr ms), 50k Sa/s, frequency mode, 20kHz analysis bandwidth, DC-coupled.
3. THD analysis BW is up to 20kHz. All har monics above 20kHz are e xcl uded. W orst case THD is usually at 6.67kHz.
4. Perfor mance within the sa me module.

Find us at www.keysight.com Page 40


Rack mounting

To rack mount a single instrument, or the adapter kit (part number: DAQ190A)

To rack mount two instrument side-by-side, order the lock-link kit (part number: DAQ194A) and flange kit
(part number: DAQ191A)

Find us at www.keysight.com Page 41


Ordering Information

Mainframe
DAQ970A Data Acquisition System Mainframe with USB and LAN

DAQ973A Data Acquisition System Mainframe with USB, LAN and GPIB

Option Z54 ANSI Z540 compliant calibration

Modules
DAQM900A 20-Channel solid-state multiplexer

DAQM901A 20-Channel armature multiplexer

DAQM902A 16-Channel reed multiplexer

DAQM903A 20-channel actuator/general purpose switch

DAQM904A 4 x 8 Two-wire matrix switch

DAQM905A Dual 4-Channel RF multiplexer, 50 Ω

DAQM907A Multifunction module

DAQM908A 40-Channel single-ended multiplexer

DAQM909A 4-Channel simultaneous sampling digitizer

Accessories
11062A Kelvin clip set

34307A 10-pack of J-type thermocouples

34308A 5-pack of 10 kΩ thermistors

34905-60001 Kit of 10 SMB-to-BNC adapter cables, 50 Ω

Learn more at: www.keysight.com


For more information on Keysight Technologies’ products, applications or services,
please contact your local Keysight office. The complete list is available at:
www.keysight.com/find/contactus

Find us at www.keysight.com Page 42


This information is subject to change without notice. © Keysight Technologies, 2019, Published in USA, November 8, 2019, 5992-3168EN

You might also like