Method Validation and Verification
Method Validation and Verification
Verification
2
Outline
3
Context and Definitions
4
Validation versus Verification
Validation
» Establishing the performance specifications of a new diagnostic tool such as a new test, laboratory
developed test or modified method
Verification
» A one-time process to determine performance characteristics of a test before use in patient testing
5
Why Evaluate a Method?
• Document initial performance:
6
Laboratory Regulations
• General and open to some interpretation
7
U.S. Test Categorization
• Determined during FDA pre-market approval
• Waived testing
» Approved for home and point-of-care use
» “Low risk of patient mismanagement if performed incorrectly”
• Non-waived testing
» Moderate Complexity
» High Complexity e.g. LDTs
» Modified Tests
https://wwwn.cdc.gov/clia/resources/testcomplexities.aspx. Accessed February 12, 2018.
8
Regulatory Requirements
9
Waived Tests
Labs have only 3 requirements!
• Pay biennial fee (every 2 years) for CLIA certificate renewal
• Follow manufacturers instructions for use
• Allow the laboratory to be inspected
̶ Generally, for cause (patient complaint)
̶ Random state survey
̶ Periodic inspections not required!
10
Nonwaived Tests
11
Method “Validation” to CLIA
Moderate Complexity High Complexity
» Precision » Precision
» Accuracy » Accuracy
» Reportable Range » Reportable Range
» Reference Range(s)
» Reference Range
» Analytical Sensitivity (LOD)
» Analytical Specificity
• Mnemonic: PARR
» Establish calibration and
control procedures
» Other performance criteria
Halling KC, Schrijver I, Persons DL. “Test Verification and Validation for Molecular Diagnostic Assays. Arch Pathol Lab Med. 2012;136:11-13.
Nichols JH. “Verification of Method Performance for Clinical Laboratories”. Advances in Clinical Chemistry. 2009;47:121-138.
12
Test Modifications
Any change in the intended use or change to an assay that could affect
performance:
• Different sample matrix (urine in a serum assay)
• Promoting different use (screen vs diagnosis)
• Type of analysis (qualitative vs quantitative)
• Incubation times and temperatures
• Sample or reagent dilution
• Using different calibration material or set-point
• Change or eliminating a procedural step
13
Analytical Verification/Validation
• Laboratories are required to perform analytical verification or validation of each nonwaived test,
method, or instrument system before use in patient testing
̶ Regardless of when it was first introduced by the laboratory
̶ Includes instruments of the same make and model and temporary replacement (loaner) instruments
• There is no exception for analytical verification or validation of tests introduced prior to a specific date
• The laboratory must retain records as long as the method is in use and for at least two years after
discontinuation
14
How to Meet the Regulations
• There is no one right way
• Consensus - CLSI protocols
• Literature - do what others have done
• Manufacturer’s recommendations
• Balance between cost and what is reasonable
15
Validation Studies
16
Accuracy
Bias to a “reference” method
̶ Absolute
̶ Relative
17
Accuracy Studies
Method comparison
» Carefully select “reference” method
» Curate high quality samples with a range of analyte concentration
» Analyze >40 specimens by both test and reference method
Best to analyze in duplicate over a period of many days
Data analysis
̶ Scatter plot of data
̶ Calculate regression statistics and estimate bias
̶ Compare results with claims or internal criteria to judge acceptability
CLSI EP-09
18
192 specimens
2 lots of reagent and calibrator
19
Special Considerations
• Medical decision points
• Clinically relevant cutoffs
20
21
22
Precision
• Within-run (Intra-assay)
• Between-run
• Day-to-day (Inter-assay or total)
23
Precision Studies
• Selection of appropriate material
• Verification study
̶ 5 x 5 study design
24
3 controls run twice per day in duplicate x for 20 days
25
Utility of Precision Data
• Future troubleshooting
• Setting QC ranges
26
Reportable Range
Includes:
• Analytical measurement range (AMR)
» Range of values that an instrument can report directly without alteration
or pretreatment of the sample
27
Range for Reporting Patient Samples
• May use the AMR
• May modify AMR to create expanded range (CRR)
» Must document that modifications to the sample and method produce
reliable results
Verify every 6 months
» Must be verified or established before patient testing begins
» Must establish reportable limits (undiluted) and maximum dilution
28
Reportable Range Studies
• Linearity Study
• 5+ concentrations of analyte throughout range
• Spike low sample with known amount of analyte
• Dilute high sample with a blank
• Mix high and low sample to create a curve
• Standard reference materials
• Commercial linearity products
• Two replicates at each level
• Data analysis
o Evaluate linear fit with XY plot
o Calculate slope and intercept
CLSI EP-6
29
Pooled patient serum sample and calibrator A mixed to get 6 sample concentrations
Both instruments demonstrated linearity of the assay consistent with manufacturer’s claims
30
Reference Intervals (RI)
• Labs are not required to establish their own
» Good practice is to verify that RI is appropriate for patient population
• Can use previously established RI or create a new one
» Discretion of the medical director
• Transfer of a RI is acceptable if test subject population and
methodology are the same or comparable
» Verified by testing N ≥ 20 samples
» If ≤ 2 outside limits, then accept
31
Establishing a RI
• Typically the central 95% of the values for the study population
• Considerations
• Exclusion criteria
• Partitioning
• Pre-analytical considerations
• Specimen handling and storage
• Special or unique patient populations
32
Protocol for Full RI Study
• Establish selection criteria for individuals
• Establish a list of interferences or sources of biological variability
• Decide on appropriate number of individuals based on desired
confidence limits (e.g. n=120)
• Collect and analyze specimens
• Evaluate data using histogram to evaluate distribution
CLSI EP28A3C
33
Samples from 20 apparently healthy donors into PST and SST tubes
Samples had PCT concentrations of 0.01 to 0.03 ng/mL confirming the manufacturer’s claims
34
Analytical Sensitivity
• Establishes the analytical sensitivity (lower detection limit) of the
assay
35
Analytical Sensitivity Studies
• Acquire measurements from multiple, independent blank and low-
level samples or pools of samples
» At least four samples of each type
» Can dilute or spike samples to provide low level samples at desired
analyte levels
» Low level sample around assumed LoD
» Obtain a series of replicate results
• Data analysis
» Parametric or nonparametric statistical methods
36
Limit of Blank determination:
Calibrator A (concentration 0 ng/mL) analyzed 10 times on each instrument
Calibrator B (concentration 0.1 ng/mL) analyzed 3 times on each instrument
37
Analytical Specificity (Interferences)
• Refers to the ability of a test or procedure to correctly identify or
quantify an entity in the presence of interfering or cross-reactive
substances
38
Interfering Substances (IFS)
• Interference- a significant difference in test result because of another
component of the sample
39
Identifying Error from IFS
Quantify effects by performing paired difference study:
• Pairs of test samples
• One with potential IFS, the other without
• Measure analyte of interest
• Calculate differences
• May be performed with patient samples
CLSI EP07
40
Summary
• Regulations require performance verification prior to patient
testing
• Precision, accuracy, reportable range and reference interval
must be evaluated, at a minimum, for all nonwaived tests
before patient use
• No “one size fits all” approach to validation/verification
41
References
• Burtis, CA, Ashwood ER. (editors) Tietz Textbook of Clinical
Chemistry, Third Edition. 1999, WB Saunders Co. Philadelphia PA.
• Clinical and Laboratory Standards Institute. Evaluation Protocols.
Various publication dates. Wayne, PA.
• HCFA 42 CFR; Final Rule; Medicare, Medicaid and CLIA Programs;
Regulations implementing Clinical Laboratories Improvement
Amendments of 1988 (CLIA. Fed. Regist. 57;7001-288 (28 February
1992)
• College of American Pathologists Commission on Laboratory
Accreditation. Inspection Checklist. 2016 CAP, Northfield, IL.
• Gruzdys V et al. 2019. Method Verification Shows a Negative Bias
between 2 Procalcitonin Methods at Medical Decision
Concentrations. JALM; 4(1):69-77.
42
© 2020 ARUP LABORATORIES