Eco Titrator
Eco Titrator
Eco Titrator
Product manual
8.1008.8001EN / 2021-08-27
Metrohm AG
Ionenstrasse
CH-9100 Herisau
Switzerland
+41 71 353 85 85
[email protected]
www.metrohm.com
Eco Titrator
Product manual
8.1008.8001EN /
2021-08-27
Technical Communication
Metrohm AG
CH-9100 Herisau
Disclaimer
Deficiencies arising from circumstances that are not the responsibility of
Metrohm, such as improper storage or improper use, etc., are expressly
excluded from the warranty. Unauthorized modifications to the product
(e.g. conversions or attachments) exclude any liability on the part of the
manufacturer for resulting damage and its consequences. Instructions and
notes in the Metrohm product documentation must be strictly followed.
Otherwise, Metrohm's liability is excluded.
■■■■■■■■■■■■■■■■■■■■■■ Table of contents
Table of contents
1 Overview 1
1.1 Product description .............................................................. 1
1.2 Product versions ................................................................... 2
1.3 Symbols and conventions .................................................... 3
1.4 Additional information ......................................................... 3
1.5 Accessories ............................................................................ 3
2 Safety 5
2.1 Intended use ......................................................................... 5
2.2 Responsibility of the operator ............................................. 5
2.3 Requirements for operating personnel ............................... 6
2.4 Safety instructions ................................................................ 6
2.4.1 Danger from electrical potential ............................................... 6
2.4.2 Danger from biological and chemical hazardous substances ..... 6
2.4.3 Danger from highly flammable substances ............................... 7
2.4.4 Danger from leaking liquids ..................................................... 7
2.4.5 Danger during transport of the product ................................... 8
2.5 Design of warning messages ............................................... 8
2.6 Meaning of warning signs ................................................... 9
3 Functional description 11
3.1 System overview ................................................................. 11
3.1.1 Signals ................................................................................... 11
3.1.2 Interfaces and connectors ...................................................... 11
3.2 Eco Titrator – Overview ..................................................... 13
3.2.1 Dosing unit ............................................................................ 17
3.2.2 Bottle unit ............................................................................. 20
3.3 Function ............................................................................... 20
3.3.1 Magnetic stirrer ..................................................................... 20
3.3.2 Dosing unit ............................................................................ 20
3.4 Indicators and controls ...................................................... 21
3.5 Interfaces and connectors ................................................. 22
3.6 Remote interface ................................................................ 23
3.7 Remote control ................................................................... 25
3.8 Arithmetic algorithms ........................................................ 27
■■■■■■■■ III
Table of contents ■■■■■■■■■■■■■■■■■■■■■■
5 Installation 31
5.1 Setup location ..................................................................... 31
5.2 Preparing the assembly ...................................................... 31
5.3 Switching the instrument on and off ................................ 33
5.4 Initial assembly of the cylinder unit .................................. 34
5.5 Mounting the support rod ................................................. 36
5.6 Preparing the bottle unit ................................................... 37
5.7 Mounting the tubing connections .................................... 38
5.8 Mounting the electrode ..................................................... 39
6 Start-up 40
6.1 Switching the instrument on and off ................................ 40
6.2 Setting the language, date and time ................................ 41
6.3 Setting the dialog type ...................................................... 42
IV ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Table of contents
8 Maintenance 167
8.1 Maintenance ..................................................................... 167
8.2 Cleaning the product surface .......................................... 167
8.2.1 Performing maintenance on the cylinder unit ....................... 168
8.2.2 Mounting the cylinder unit .................................................. 172
8.3 Displaying system data .................................................... 173
8.4 Resetting the system ........................................................ 174
9 Troubleshooting 176
10 Disposal 177
11 Technical specifications 178
11.1 Ambient conditions .......................................................... 178
11.2 Energy supply ................................................................... 178
11.3 Dimensions ........................................................................ 179
11.4 Housing ............................................................................. 179
11.5 Connectors specifications ................................................ 180
11.6 Display specifications ....................................................... 181
11.7 Operation specifications .................................................. 181
11.8 Measurement specifications ............................................ 182
11.9 Stirrer specifications ......................................................... 183
11.10 Liquid handling specifications ......................................... 183
■■■■■■■■ V
■■■■■■■■■■■■■■■■■■■■■■ Overview
1 Overview
Titration modes
The following titration modes are supported:
■ DET
■ MET
■ SET
Calibration mode
The CAL calibration mode is used to calibrate the electrode.
Connectors
The instrument is equipped with the following connectors:
■ USB 1 and USB 2
■ Ethernet
■ Remote
■ 24 VDC Power OUT
■ 24 VDC Power IN
■ Ind
■ Ref
■ Temp
■ Pol
■■■■■■■■ 1
Product versions ■■■■■■■■■■■■■■■■■■■■■■
The required numbers for the customer service can be found on the type
plate (see example):
NOTICE
2 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Overview
1.5 Accessories
Up-to-date information on the scope of delivery and on optional accesso-
ries can be found on the Metrohm website. Download this information as
follows:
1 Go to https://www.metrohm.com.
2 Enter the article number of the product (e.g. 2.1001.0010) into the
search field.
The search result is displayed.
4 On the Included parts tab, click the link to download the PDF.
■■■■■■■■ 3
Accessories ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
4 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Safety
2 Safety
■■■■■■■■ 5
Requirements for operating personnel ■■■■■■■■■■■■■■■■■■■■■■
6 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Safety
■■■■■■■■ 7
Design of warning messages ■■■■■■■■■■■■■■■■■■■■■■
DANGER
8 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Safety
WARNING
CAUTION
■■■■■■■■ 9
Meaning of warning signs ■■■■■■■■■■■■■■■■■■■■■■
10 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
3 Functional description
The status indicator uses colors to display the operating status of the
instrument.
Table 4 Status indicator
Signal Color Meaning
Green Ready for operation
Orange In operation
Yellow Waiting
■■■■■■■■ 11
System overview ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
12 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
■■■■■■■■ 13
Eco Titrator – Overview ■■■■■■■■■■■■■■■■■■■■■■
14 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
■■■■■■■■ 15
Eco Titrator – Overview ■■■■■■■■■■■■■■■■■■■■■■
16 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
3 Flat stopcock
■■■■■■■■ 17
Eco Titrator – Overview ■■■■■■■■■■■■■■■■■■■■■■
18 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
1 Connector for the tubing connection to 2 Connector for the tubing connection to
the bottle the cylinder unit
■■■■■■■■ 19
Function ■■■■■■■■■■■■■■■■■■■■■■
3.3 Function
3.3.1 Magnetic stirrer
The magnetic stirrer ensures that the sample is well mixed. The stirring
rate can be adjusted depending on the amount and viscosity of the sam-
ple.
A clamping ring and an electrode holder can be fastened to the support
rod of the magnetic stirrer.
20 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
cylinder of the cylinder unit and is responsible for accurate dosing of the
solution.
The flat stopcock switches between filling and emptying the dosing cylin-
der of the cylinder unit.
Once the cylinder unit is put into place, the dosing drive and the flat stop-
cock handle the following functions:
■ Raising and lowering the piston:
Solution is aspirated while the piston is being lowered. The dosing cyl-
inder fills up.
Solution is dosed while the piston is being raised. The dosing cylinder
empties.
■ Rotating the flat stopcock:
The position of the flat stopcock determines which connectors the sol-
ution flows through.
Figure 10 Indicators
5 Start
■■■■■■■■ 21
Interfaces and connectors ■■■■■■■■■■■■■■■■■■■■■■
1 USB 2 Ethernet
Connect USB flash drive, printer, balance, Remote control via local network
etc.
3 Remote 4 Power IN
Connect analog remote control / Eco Dosi- Connect power supply unit
mat
5 Pol 6 Temp
Connect a polarizable electrode Connect a temperature sensor (Pt1000 or
NTC) or an electrode with integrated tem-
perature sensor
7 Ref 8 Ind
Connect a reference electrode Connect a non-polarizable electrode (e.g.
ion-selective, pH, redox or other electrode)
NOTICE
22 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
9 6
6 9
Figure 13 Pin assignment of remote socket and remote plug
The above figure of the pin assignment applies to all Metrohm instru-
ments with 9-pin D-Sub remote connector.
Table 6 Inputs and outputs of the remote interface
Pin no. Assignment Function
1 Output 0 Ready/EOD
2 Output 1 Activate/Dosimat
3 Output 2 Titration/Determination
4 Output 3 Cond. OK
5 Output 4 Error
6 0 volt (GND)
7 +5 Volt
8 Input 0 Start
9 Input 1 Stop
Inputs
+5 V approx. 5 kΩ Pull-up
tp
tP > 100 ms
active = low, inactive = high
Outputs
■■■■■■■■ 23
Remote interface ■■■■■■■■■■■■■■■■■■■■■■
Open Collector
tp
tp > 200 ms
active = low, inactive = high
IC = 20 mA, VCEO = 40 V
+5 V: maximum load = 20 mA
Activ./Dos. OUTPUT 1
Titr./Determ. OUTPUT 2
Cond. OK OUTPUT 3
Error OUTPUT 4
Start INPUT 0
Stop INPUT 1
Ready/EOD OUTPUT 0
Activ./Dos. OUTPUT 1
Titr./Determ. OUTPUT 2
Cond. OK OUTPUT 3
Error OUTPUT 4
Start INPUT 0
Stop INPUT 1
24 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
EOD EOD
Ready/EOD OUTPUT 0
Activ./Dos. OUTPUT 1
Titr./Determ. OUTPUT 2
Cond. OK OUTPUT 3
Error OUTPUT 4
Start INPUT 0
Stop INPUT 1
Activ./Dos. OUTPUT 1
Titr./Determ. OUTPUT 2
Cond. OK OUTPUT 3
Error OUTPUT 4
Start INPUT 0
Stop INPUT 1
Transfer protocol
The data communication is synchronous. On each command there is an
acknowledgement by the instrument.
A command must be sent to the instrument with the control characters
CRLF as terminator. The acknowledgements of the instrument will also be
transferred with CRLF as terminator.
The instrument does not send spontaneous messages.
■■■■■■■■ 25
Remote control ■■■■■■■■■■■■■■■■■■■■■■
26 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
The values of the variables are only available after the end of a determina-
tion (in the status 'ready').
Acknowledge- Comment
ment of the
instrument
OK Command executed
E1 Method not found
E2 Invalid variable
E3 Invalid command
Rounding-off process
Measured values and results are rounded to the defined number of deci-
mal places (commercial rounding, in accordance with the US Pharmaco-
peia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4,
then it will be rounded off; if this digit is 5, 6, 7, 8 or 9, then it will be
■■■■■■■■ 27
Arithmetic algorithms ■■■■■■■■■■■■■■■■■■■■■■
Statistics
The arithmetic mean value and the absolute and relative standard devia-
tions of results are calculated:
You can statistically evaluate a maximum of five results (1 ≤ k ≤ 5) calcula-
ted in a determination. A statistical series can contain a maximum of 20
determinations (1 ≤ n ≤ 20).
The following convention applies to the subsequent formulas:
1 ≤ n ≤ 20 and 1 ≤ k ≤ 5.
Mean value:
Explanations
The individual values are incorporated in the statistics with full accuracy.
15 significant places are yielded when the 64 bit numerical format is
applied for the floating-point number in decimal presentation.
The accuracy can be controlled by the selection of the prefix of the unit
(milli, micro) and the number of decimal places.
Example
The result displayed, 1234.56789158763 mg/L , has 15 significant pla-
ces. It should be rounded off to three decimal places according to the
above rounding-off process:
■ 1234.568 mg/L.
28 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Functional description
NOTICE
■■■■■■■■ 29
Delivery ■■■■■■■■■■■■■■■■■■■■■■
4.1 Delivery
Inspect the delivery immediately upon receipt:
■ Check the delivery against the delivery note to ensure completeness.
■ Check the product for damage.
■ If the delivery is incomplete or damaged, contact your regional Met-
rohm representative.
4.2 Packaging
The product and accessories are supplied in protective special packaging.
Keep this packaging to ensure safe transportation of the product. If a
transport locking device is present, keep this as well for future reuse.
30 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Installation
5 Installation
WARNING
■■■■■■■■ 31
Preparing the assembly ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
The flat side of the plug of the power supply unit must face
downwards.
Connect the power supply unit to the Power IN connector. Note the
orientation (see figure).
The instrument can now be switched on and off.
32 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Installation
Prerequisite
The power cord is connected.
The instrument is switched off with the key. In order to prevent acci-
dental switch-off, the key must be held for an extended time.
NOTICE
To power down the product, you must unplug the power cord
from the energy supply.
■■■■■■■■ 33
Initial assembly of the cylinder unit ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
If the instrument has already been set up, the cylinder unit is disas-
sembled and assembled with the Manual control ▶ Exchange cyl-
inder unit function.
Prerequisite
The instrument is switched on. The push rod is in the lowest position.
Procedure
1 If the instrument is switched on for the first time, the installation wiz-
ard is started automatically.
The installation wizard can be opened manually with the Manual
control ▶ Installation wizard function.
34 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Installation
NOTICE
■■■■■■■■ 35
Mounting the support rod ■■■■■■■■■■■■■■■■■■■■■■
Procedure
NOTICE
The clamping ring is used as the lower stop for the electrode
holder. The clamping ring prevents the electrode holder with the
mounted electrode from being lowered too far.
Procedure
1 Push the clamping ring over the support rod with the indent facing
upward as far as it will go.
4 To fix the electrode holder in place, release the locking lever at the
desired height.
36 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Installation
5 Secure the adsorber tube in place with the clip for SGJ 14/15.
■■■■■■■■ 37
Mounting the tubing connections ■■■■■■■■■■■■■■■■■■■■■■
2 Screw the prepared bottle cap onto the bottle and tighten it by
hand.
NOTICE
Tighten all the tubing connections by hand. Use no other aids. The
thread of the screw nipples and the tubing openings must not be
deformed.
38 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Installation
Prerequisite:
■ The cylinder unit, electrode holder and bottle unit are set up.
1 Screw the 6.1805.090 tubing securely to the cylinder unit and to the
flat stopcock.
2 Screw the 6.1805.090 tubing securely to the bottle unit and to the
flat stopcock.
5 Insert the buret tip into the guide sleeve on the electrode holder.
3 Buret tip
■■■■■■■■ 39
Switching the instrument on and off ■■■■■■■■■■■■■■■■■■■■■■
6 Start-up
Prerequisite
The power cord is connected.
The instrument is switched off with the key. In order to prevent acci-
dental switch-off, the key must be held for an extended time.
NOTICE
To power down the product, you must unplug the power cord
from the energy supply.
40 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Start-up
Prerequisite:
■ The instrument is switched on.
■■■■■■■■ 41
Setting the dialog type ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
If the instrument is switched off, the dialog type that was set remains
activated.
2 Expand the [Dialog type] input field. Select the Routine dialog
type.
42 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Start-up
6 Expand the [Dialog type] input field. Select the Expert dialog type.
All the user settings are available.
■■■■■■■■ 43
Display elements and controls ■■■■■■■■■■■■■■■■■■■■■■
Figure 22 User guidance – Status display, touch screen and control bar
3 Control bar
The status display uses flashing patterns to display the operating status
of the instrument.
The touch screen is used to enter data and to control runs. Results and
evaluations are also displayed on the touch screen.
The control bar is used to switch the instrument on and off, to control
the stirring rate during a determination and to start and stop determina-
tions.
44 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
Prerequisite
The power cord is connected.
The instrument is switched off with the key. In order to prevent acci-
dental switch-off, the key must be held for an extended time.
■■■■■■■■ 45
User interface ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
To power down the product, you must unplug the power cord
from the energy supply.
Work areas
Once the instrument is ready for operation, the work areas can be selec-
ted. Work areas that cannot be selected are grayed out.
NOTICE
If necessary, click the button so that the piston moves to the basic
position, the status display lights up and the status indicator is green.
46 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Sample data
Access to the sample data: sample size, unit, ID1 and ID2
Parameters
Access to the parameters, grouped into the following categories:
■ Start conditions
■ Titration parameters
■ Stop conditions
■ Evaluation
■ Calculation
■ Statistics
■ Reports
Results
Access to the calculated results and the stop criterion.
Live status
Access to the graphic display of the ongoing determination.
Menu path
In the menu path, clicking one of the elements of the menu path can be
used to call up the respective menu.
■■■■■■■■ 47
User interface ■■■■■■■■■■■■■■■■■■■■■■
■ Input fields
The respective keyboard is called up by clicking the button of the input
field.
■ Keyboards
Input of text, numbers or characters. An appropriate keyboard is called
up, depending on the type of input.
NOTICE
Help texts
Help texts (in English) are available for the input fields. The respective
help text is called up by long pressing (for at least 3 seconds) on an
input field. Default values and possible inputs are displayed.
Status indicator
The status indicator on the touch screen uses colors to display the operat-
ing status of the instrument.
Icon
The following functions can be called up with the different icons -
depending on the context:
■ Save
■ Delete
■ Export
■ etc.
NOTICE
48 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Brightness
Input range 1 to 10
Default value = 7
3 Increase brightness
Keyboards
Different keyboard types are available.
■■■■■■■■ 49
User interface ■■■■■■■■■■■■■■■■■■■■■■
9 Switch keyboard
50 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
The formula editor allows for the entry of formulas. The formula editor is
equipped with an automatic syntax check. This is triggered as soon as a
formula is applied. The generally valid rules of priority apply for the calcu-
lation operations.
Variable Description
C00 Sample size
EP# Volume of endpoint EP# (# = 1–9)
CI# Sample identification (# = 1–2)
R# Result (# = 1–5)
FP# Volume of fixed point FP# (# = 1–2)
CV0# Common variable (# = 1–5)
SMN# Mean value of result R# (# = 1–5)
TITER Titer of selected solution
CONC Concentration of selected solution
Var List of additional variables
Templates List of predefined calculation formulas
"#" stands for a sequential number that you must enter manually. Exam-
ple: If you apply the variable EP# in the formula, only EP is entered. In
order to define the EP to be used, the corresponding number has to be
added manually after "EP". Example: EP5
Variables
Clicking on [Var] displays a list with additional variables. You can enter
these variables either directly into the formula or also by selecting them
from the list and applying them with [OK].
Variable Description
MIM Initial measured value, i.e. measured value prior to the
processing of the start conditions
MSM Start measured value, i.e. measured value after the pro-
cessing of the start conditions
MCV End volume, i.e. total dosed volume at the end of the
titration
ET# Temperature at endpoint EP# (# = 1–9)
■■■■■■■■ 51
User interface ■■■■■■■■■■■■■■■■■■■■■■
Variable Description
EM# Measured value of endpoint EP# (# = 1–9)
ED# Time at endpoint EP# (# = 1–9)
MSV Start volume
MEN Electrode zero point pH(0)
MSL Electrode slope
DD Duration of the entire determination
MST Start temperature
MCT End temperature
FT# Temperature at fixed point FP# (# = 1–2)
FM# Measured value of fixed point FP# (# = 1–2)
FD# Time at fixed point FP# (# = 1–2)
For information on the meaning of the placeholder Molw, see the follow-
ing note.
Calculation templates
Pressing [Templates] displays a list with calculation templates. You can
apply these templates directly with .
NOTICE
Some templates contain the placeholder Molw, which stands for the
molar mass of the sample. You must replace this placeholder with the
correct value in the calculation formula.
Template Description
Content % Content in %
Unit of the sample size = g
Content Content in mmol/L
mmol/L
Unit of the sample size = mL
Content Content in mol/L
mol/L
Unit of the sample size = mL
Content g/L Content in g/L
Unit of the sample size = mL
52 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Template Description
Content Content in ppm
ppm
Unit of the sample size = g
Titer Titer calculation
Unit of the sample size = g
Blank mean Blank value as mean value of single results
value
Blank single Blank value as single value
value
■■■■■■■■ 53
Manual control ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
54 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
■■■■■■■■ 55
Manual control ■■■■■■■■■■■■■■■■■■■■■■
4 Use the button to stop the process and fill the cylinder.
Measuring manually
56 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
CAUTION
CAUTION
■■■■■■■■ 57
Manual control ■■■■■■■■■■■■■■■■■■■■■■
3 ■ Make sure that the tubing from the bottle cap is removed.
■ [Continue]
The message Exchanging cylinder unit... appears and the piston is
lowered down to the height at which the cylinder unit can be disas-
sembled.
58 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
■ To mount the new cylinder unit, the piston rod should protrude by
approx. 6 mm from the dosing cylinder. If necessary, pull the pis-
ton carefully out of the dosing cylinder with the 6.1546.040 pis-
ton tool.
While doing so, make sure that the sealing lips and the piston in
the dosing cylinder are not damaged.
■ Couple the piston rod with the push rod.
When doing so, the hook profile of the piston rod must be care-
fully aligned in the hook profile of the push rod.
■ Carefully push the cylinder unit downwards. The piston is pushed
into the dosing cylinder.
Screw the light protection of the cylinder unit securely into the
thread of the housing.
7 ■ Make sure that the cylinder unit has been mounted correctly.
■ [Continue]
8 [Continue]
NOTICE
Make sure that the value for the cylinder volume in the System
menu is the same as the volume of the mounted cylinder unit.
9
NOTICE
The cylinder unit has been replaced and is ready for use.
■■■■■■■■ 59
Manual control ■■■■■■■■■■■■■■■■■■■■■■
60 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Stirring rate
The stirring rate can be adjusted in 15 steps. The default value is 8.
Prerequisite:
■ The magnetic stirrer controls are opened: Start page ▶ Manual con-
trol ▶ Stirrer
■ The stirrer is switched on.
Click on the button repeatedly until the desired stirring rate has
been reached.
Each click reduces the stirring rate by one step. The current stirring
rate is displayed.
Click on the button repeatedly until the desired stirring rate has
been reached.
Each click increases the stirring rate by one step. The current stirring
rate is displayed.
■■■■■■■■ 61
Methods ■■■■■■■■■■■■■■■■■■■■■■
7.5 Methods
Definition
A method determines how determinations are carried out. The titration
mode, measured quantity and further parameters are defined in a
method.
Methods are saved under a freely selectable method name. A method
name consists of a maximum of 12 characters.
62 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Method list
The [Method] button on the start page shows a list of all the saved meth-
ods. Methods can be created, exported and deleted here.
Titration mode
Each method is based on a titration mode. The following titration modes
are available:
■ MET – Monotonic equivalence point titration
3 measured quantities are available: pH, U or Ipol
■ DET – Dynamic equivalence point titration
3 measured quantities are available: pH, U or Ipol
■ SET – Endpoint titration
3 measured quantities are available: pH, U or Ipol
■ CAL – Calibration of pH electrodes
Measured quantity: pH
■■■■■■■■ 63
Methods ■■■■■■■■■■■■■■■■■■■■■■
Procedure
Procedure
1 Load the method that you want to change in the method selection
bar on the Start page.
2
Open the Parameters work area:
If you modify method parameters, then you can save these as your own
method. A maximum of 120 methods can be saved.
64 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Procedure
The method that you want to save is loaded in the method selection
bar. The method is marked [New] or [Modified].
3 Enter the desired name with the keyboard. Finish with [OK].
The name that was entered appears in the method selection bar. The
method is now saved in the method list.
Procedure
■■■■■■■■ 65
Methods ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
■ Click on .
The new method appears in the method selection bar. The method is
marked with [New]. Example: MET_pH [New]
66 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Deleting a method
Procedure
Exporting a method
Procedure
■■■■■■■■ 67
Methods ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
If a method with the same name already exists on the USB flash
drive, then the following warning appears: Store method:
Method name already exists. Do you want to overwrite
the name?.
■ [Yes]: The method on the USB flash drive will be overwritten.
■ [No]: The method will not be exported.
Importing a method
Procedure
68 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
Input options
There are 2 options for entering sample data:
■ Directly in the Sample data work area.
■ Automatic request immediately after the start of the determination.
The data for the sample can be entered in the Sample data work area,
even if the determination is ongoing.
■■■■■■■■ 69
Sample data ■■■■■■■■■■■■■■■■■■■■■■
ID1
The sample identification ID1 can be used in calculations as the variable
CI1.
Input: max. 10 characters
Default value: empty
ID2
The sample identification ID2 can be used in calculations as the variable
CI2.
Input: max. 10 characters
Default value: empty
Sample size
The value of the sample size can be used in calculations as the variable
C00.
Input range –999,999,999 to 9999999999
Default value 1.0
Unit
Unit of the sample size.
Selection:
■ g
■ mg
■ µg
■ mL
■ µL
■ Pieces
■ User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.
Default value: g
70 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
The automatic sample data request can be controlled for each method
separately. The following parameters are available under [Start condi-
tions] in the Parameters work area:
■ Request sample ID
■ Request sample size
■ Request sample unit
■ If the Hold at request parameter is activated, the run will be paused
and must be continued with after the sample data has been
entered.
■ If Hold at request is deactivated, then the titration will be started in
the background. This dialog will be displayed until the entering of the
sample data is confirmed with or until the sample data is transfer-
red from the balance, even if the titration is already completed. This
ensures that the sample data is available for calculations.
2 Click on [ON].
■■■■■■■■ 71
Sample data ■■■■■■■■■■■■■■■■■■■■■■
3
Go to the Sample data work area:
The sample table appears.
1
Go to the Sample data work area:
72 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Editing samples
1
Go to the Sample data work area:
3 Edit sample:
1
Go to the Sample data work area:
3 Delete sample:
■■■■■■■■ 73
Sample data ■■■■■■■■■■■■■■■■■■■■■■
1
Go to the Sample data work area:
1
Go to the Sample data work area:
74 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
■■■■■■■■ 75
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
■ Ethernet settings
■ Service
■ About
■ Change password
■ COM port settings
User name
A user name can be entered here for the report. This parameter will only
be printed if a user has been defined.
Input: max. 12 characters
Default value: empty
76 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Instrument name
An instrument name can be entered here for the report. This parameter
will only be printed if a designation has been defined.
Input: max. 10 characters
Default value: empty
Language
Set the dialog language.
Dialog type
The user dialog can be limited for routine operations. The resetting of the
dialog will take effect as soon as you exit the start page.
■ Dialog type Expert (default value)
In the dialog type Expert, all the user settings are available.
■ Dialog type Routine
In the dialog type Routine, the availability of the settings is restricted.
The [System] and [Methods] menus as well as the Parameters work
area can only be opened with a password. Methods can be loaded,
however, on the start page.
NOTICE
The Routine setting will only take effect once the [System] menu
has been exited.
To switch back to expert mode, open the [System] menu and enter
the password:
■ Password for firmware version 57.1008.0010 or higher:
METROHM9100
■ Password for firmware version 57.1008.0009 or lower: MSH9101
Then activate the Expert dialog type.
Selection:
■ Expert
■ Routine
Default value: Expert
Time
Current time. Only numbers that make sense can be entered.
Format: hh:mm:ss
■■■■■■■■ 77
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
Date
Current date. Only numbers that make sense can be entered.
Format: YYYY:MM:DD
PREP warning
If the PREP warning is activated, the recommendation to execute the
Prepare buret (PREP) function appears:
■ After the instrument is switched on.
■ Each time a buret unit is attached.
All tubing and the cylinder are rinsed with this function.
Switch:
■ OFF
■ ON
Default value: ON
Beep
If Beep is activated, a short acoustic signal sounds in the following cases:
■ When a key is pressed.
■ At the end of the determination.
■ When the system remains conditioned without interruption for 10 sec-
onds.
Switch:
■ OFF
■ ON
Default value: ON
Temperature sensor
The instrument supports the use of two different temperature measure-
ment techniques:
78 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
R (25 °C)
This parameter is visible only when Temperature sensor = NTC.
Nominal resistance of the NTC sensor at 25 °C.
Input range 1,000 to 99,999 Ω
Default value 30,000 Ω
B value
This parameter is visible only when Temperature sensor = NTC.
Material constant of the NTC sensor. B values of NTC sensors are fre-
quently based on different reference temperatures (usually 25 °C and 50–
100 °C).
Input range 1,000 to 9,999 K
Default value 4,100 K
■■■■■■■■ 79
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
Sensor data
Name
The designation of the sensor is used for unambiguous identification.
Input: max. 24 characters
Default value: empty
Type
The sensor type is displayed.
80 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Selection:
■ pH electrode
■ Metal electrode
■ Temperature sensor
■ Other sensor
Slope
This parameter is only visible with pH electrodes.
Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0% is used as the slope.
Input range –999.9 to 999.9 %
Default value 100 %
pH(0)
This parameter is only visible with pH electrodes.
pH value of the pH electrode at 0 mV. Apart from the slope, pH(0) is the
second characteristic of the calibration curve.
Input range –20.000 to 20.000
Default value 7.000
Calibration temperature
This parameter is only visible with pH electrodes.
Temperature at which the last calibration was carried out.
Input range –20.0 to 150.0 °C
Default value 25.0 °C
Calibration date
This parameter is only visible with pH electrodes.
Date of the last calibration.
Monitoring
This parameter is only visible with pH electrodes.
Activate and deactivate the calibration monitoring.
Switch:
■ OFF
■ ON
Default value: OFF
Time interval
This parameter is visible only when Monitoring = ON.
■■■■■■■■ 81
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
When you start a method, you will be notified if this time interval (in days)
has already elapsed. You can then select whether or not you would still
like to start the method.
Input range 1 to 999 d
Default value 999 d
Solution data
The solution data for New ( ) and Edit ( ) are the same, therefore
only pictures for System ▶ Solutions ▶ New are shown:
82 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Name
The designation of the solution is used for unique identification.
Input: max. 24 characters
Default value: empty
Titer
Titer of the solution.
Input range –999,999,999 to 9999999999
Default value 1.000
Titer unit
Unit of the titer.
■■■■■■■■ 83
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ µmol/mL
■ mmol/L
■ mol/L
■ g/L
■ mg/L
■ mg/mL
■ µg/L
■ ppm
■ %
■ mEq/L
■ empty
■ User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. An empty entry can be generated this way as well.
Default value: empty
Concentration
Concentration of the solution.
Input range –999,999,999 to 9999999999
Default value 1.000
Concentration unit
Unit of the concentration.
Selection:
■ mol/L
■ %
■ g/L
■ mEq/L
■ mg/L
■ mg/mL
■ mmol/L
■ ppm
■ µg/L
■ µmol/mL
■ User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. An empty entry can be generated this way as well.
Default value: mol/L
84 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Monitoring
Activate and deactivate titer monitoring.
Switch:
■ OFF
■ ON
Default value: OFF
Time interval
This parameter is visible only when Monitoring = On.
When you start a method, you will be notified if this time interval (in days)
has already elapsed. You can then select whether or not you would still
like to start the method.
Input range 1 to 999 d
Default value 999 d
■■■■■■■■ 85
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
86 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
If the parameter Statistics is set to [ON], then the mean value of the
results will be assigned to the respective common variable.
PC/LIMS report
Specification of the storage location for the PC/LIMS report. The PC/LIMS
report is a machine-readable report with all of the important data for a
determination. It can be saved as follows:
■ as a TXT file on a USB flash drive.
■ to a LIMS via the Ethernet interface and an RS-232 instrument server.
Selection:
■ USB flash drive
■ Ethernet/RS-232
Default value: USB flash drive
USB flash The report will be saved as a TXT file on the USB flash
drive drive in the folder pc_lims_report.
Ethernet/ The report is sent via an RS-232 instrument server. The
RS-232 interface parameters are set on the RS-232 instrument
server (see Application Bulletin AB-435).
Printer
If a printer is connected, then the printer type needs to be defined here in
order for the reports to be printed out correctly.
■■■■■■■■ 87
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
The printers that have the designation ESC-POS are so-called POS printers
(point-of-sale printers), i.e. they print on continuous paper.
Selection:
■ PDF (Save on USB flash drive)
■ Custom (ESC-POS)
■ PostScript
Default value: Custom (ESC-POS)
Commercially available A4 printers that communicate via PostScript can be
connected directly via USB.
Keyboard layout
Layout of the on-screen keyboard.
Selection:
■ English US
Balance
Selection:
■ Sartorius
For balances with RS-232 interface: Use the 6.2148.050 USB/RS-232 Con-
verter.
Configure the serial interface: System ▶ COM port settings
The parameters set for the RS-232 interface on the balance must match
those on the Eco Titrator.
88 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Importing a method
Procedure
■■■■■■■■ 89
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
Procedure
90 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Creating a backup
Procedure
Restoring
Procedure
■■■■■■■■ 91
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
Display test
The [Display test] button offers settings for brightness, various test
images and a calibration program for the screen:
92 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Keyboard test
■ Start the test: [Keyboard test]
■ Press the five keys of the control bar one after the other:
Logs
■ Show error log: [Logs]
■ Save error log to a USB flash drive:
Mode
This network configuration can be done manually or automatically.
Selection:
■ Static
The network configuration is done manually. The input fields IP
address, Subnet mask and Gateway are used for this.
■ DHCP
The network configuration is assigned automatically via a server.
Default value: DHCP
2 Enter the current password and then the new password twice.
■■■■■■■■ 93
System – Configuration ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
NOTICE
Baud rate
Transfer rate in characters per second.
Selection:
■ 1,200
■ 2,400
■ 4,800
■ 9,600
■ 19,200
■ 38,400
■ 57,600
■ 115,200
Default value: 9,600
Data bits
Number of data bits.
94 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Selection:
■ 7
■ 8
Default value: 8
Stop bits
Number of stop bits.
Selection:
■ 1
■ 2
Default value: 1
Parity
Type of parity testing.
Selection:
■ Even
■ None
■ Odd
Default value: None
Handshake
Type of the data transfer protocol.
Selection:
■ Hardware
■ Software
■ none
Default value: Hardware
NOTICE
■■■■■■■■ 95
Carrying out a pH calibration ■■■■■■■■■■■■■■■■■■■■■■
3 Measuring buffer 1
■ Immerse the pH electrode in the buffer 1.
■ Press the key.
■ Enter the current temperature. If a temperature sensor is connec-
ted, then the temperature will be measured automatically.
The measurement is started. The on-screen display changes to the
Live status work area:
96 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
■■■■■■■■ 97
Carrying out the determination ■■■■■■■■■■■■■■■■■■■■■■
5 Live modifications
Make live modifications if necessary:
■ Editing the sample data of the running determination
■ Editing the method parameters of the running determination
■ Modifying the stirring rate
98 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
The sample data can be entered or modified in the Samples work area
while a determination is running. The sample data entered at the end of
the titration in the Samples work area is always used in calculations.
Click on .
The Samples work area appears. The determination continues to
run in the background.
Click on .
The Live status work area appears again.
NOTICE
■■■■■■■■ 99
Results ■■■■■■■■■■■■■■■■■■■■■■
dosed, then these modifications will not be taken into account until the
next determination.
Click on .
The Parameters work area appears. The determination continues to
run in the background.
Click on .
The Live status work area appears again.
1 The stirring rate of the magnetic stirrer can be changed with the con-
trol bar while a determination is carried out.
7.10 Results
100 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Curve
Recalculating
■■■■■■■■ 101
Results ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
All the results of the determination that was carried out last are recalcula-
ted with the Recalc function. This is necessary if, for example, the calcula-
tion, the titer or the sample size has been modified.
Statistics
By clicking on the key, the statistical overview of a determination ser-
ies is displayed.
NOTICE
The mean value Mean, the absolute standard deviation s abs and the rel-
ative standard deviation s rel are displayed in the overview. For the mean
value, the number of individual results from which it has been calculated is
displayed in parentheses. In this example, it is 3.
The Statistics row shows how many determinations have already been car-
ried out and how many determinations are to be carried out in total. 3 out
of 4 determinations were carried out in this example.
102 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Details
Shows further data of the determination series.
The result and the sample size of each determination are shown.
A determination can be removed from the Statistics in the col-
umn On/Off. That row is then marked with . All the results
from the highlighted determination are removed from the statis-
tics. The statistics are automatically recalculated.
Reset
Deletes all statistics data.
The statistics data is deleted automatically in the following
cases:
■ Once all the determinations of the determination series have
been carried out and a new determination is started.
■ Once a new method is being loaded.
Increase
Adds an additional sample to a determination series, e.g.
because a determination was faulty and had to be removed
from the statistics. The second number in the Statistics line will
be increased automatically by one.
■■■■■■■■ 103
Printing reports ■■■■■■■■■■■■■■■■■■■■■■
■ Report as in method
The reports that are defined in the method will be printed out.
■ PC/LIMS
Machine-readable report with all of the data for a determination. This
report can be saved as a TXT file to a connected USB flash drive.
Preparing to print
104 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
7.12 Parameters
Titrations
Dynamic equivalence Dynamic equivalence point titration is a titration mode for all standard
point titration (DET) titrations. The reagent is added in variable volume steps. The volume steps
vary as a function of the slope of the curve. An attempt is made to reach
constant measured value changes with each dosing. The optimal volume
for dosing is determined from the measured value changes of the previous
dosings. Measured value acceptance is drift-controlled (equilibrium titra-
tion) or after a waiting time. Equivalence points are evaluated automati-
cally.
U/mV
V/mL
Figure 44 Reagent dosing for DET
Monotonic equiva- Monotonic equivalence point titration is a titration mode for titrations
lence point titration with relatively high signal fluctuations or suddenly occurring potential
(MET) jumps and for slow titrations or slow-response electrodes. The reagent is
added in constant volume steps. Measured value acceptance is drift-con-
trolled (equilibrium titration) or after a waiting time. Equivalence points are
evaluated automatically.
U/mV
V/mL
Figure 45 Reagent dosing for MET
Endpoint titration Endpoint titration is a titration mode for rapid routine determinations to a
(SET) preset endpoint (e.g. titrations in accordance with special standards) and
titrations for which reagent overflow must be avoided. The titration termi-
nation at the endpoint takes place either drift-controlled or after a waiting
■■■■■■■■ 105
Parameters ■■■■■■■■■■■■■■■■■■■■■■
time. The volume dosed until the endpoint is used for calculating the con-
tent of the sample.
U/mV
Endpoint
Control range
V/mL
t/s
106 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Request sample ID
Selection of the sample identification that is queried at the start of the
determination.
Selection:
■ ID1
■ ID2
■ ID1&ID2
■ Off
Default value: Off
Start volume
Volume that is dosed prior to the start of the titration.
■■■■■■■■ 107
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Dosing rate
Rate at which the start volume is dosed.
Input range 0.02 to Max. mL/min
Additional selection: Max. = maximum dosing rate.
Default value: Max.
NOTICE
Pause
Waiting time, e.g. for the stabilization of the measured value after the
start, for dissolving solid substances or a reaction time after the dosing of
a start volume.
Input range 0 to 999,999 s
Default value 0s
Activation pulse
If it is activated, an activation pulse is sent to a remote line that starts a
connected Dosimat. We recommend to define a Start delay time for
this.
108 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Switch:
■ OFF
■ ON
Default value: ON
Stirrer
If this parameter is activated, then the stirrer is switched on at the start of
the determination.
■■■■■■■■ 109
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Switch:
■ OFF
■ ON
Default value: ON
Stirring rate
Setting the stirring rate. The stirring direction is always clockwise.
Conversion: Value x 120 ± 5 rpm = stirring rate in rpm
e.g.: 8 × 120 ± 5 rpm = 960 ± 40 rpm
Input range 1 to 15
Default value 8
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted, then the temperature will be measured continuously. For determina-
tions in pH mode, the value is used for temperature compensation (elec-
trode slope is adjusted accordingly).
Input range –20.0 to 150.0 °C
Default value 25.0 °C
Sensor
Open the selection list and select a sensor.
The selection depends on the measuring mode. Sensors are defined and
listed under System ▶ Sensors, e.g.:
Solution
Open the selection list and select a solution.
Solutions are defined and listed under System ▶ Solutions, e.g.:
110 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Titration rate
3 predefined sets of parameters and 1 set of parameters that can be
defined manually are available for setting the titration rate.
Selection:
■ Slow: For titrations in which the finest details are to be visible. This
can, however, also lead to an increase in noise, which may result in
unwanted equivalence points.
■ Optimal: For all standard titrations. The parameters have been opti-
mized for the most frequent applications.
■ Fast: For fast and less critical titrations.
■ User: The individual titration parameters can be modified.
Default value: Optimal
■■■■■■■■ 111
Parameters ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
Min. Increment
This parameter is only active if the titration rate is set to User.
This smallest permitted volume increment is added at the start of the titra-
tion and with steep curves in the region of the equivalence point. Very
small values should only be used if a low titrant consumption is expected.
Otherwise unwanted equivalence points may be evaluated.
Input range 0.05 to 999.90 µL
Default value 10.0 µL
Max. Increment
This parameter is only active if the titration rate is set to User.
A maximum volume increment should be selected in the following cases:
■ If titrant consumption is very low until the equivalence point is reached.
■ If a start volume is dosed until shortly before the equivalence point is
reached.
■ If the change of direction in the jumping range is very abrupt, because
otherwise it is easily possible for an excessively large volume to be
dosed in the region of the equivalence point.
The value should not be set to less than 1/100 of the cylinder volume.
112 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
It is not advisable to select similar volumes for the minimum and the
maximum increment. Monotonic equivalence point titration (MET) is
appropriate for these applications.
Signal drift
This parameter is only active if the titration rate is set to User.
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. This type of titration is
often referred to as equilibrium titration.
■■■■■■■■ 113
Parameters ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
Dosing rate
This parameter is only active if the titration rate is set to User.
Rate at which the volume increments are dosed.
Input range 0.01 to Max. mL/min
Additional selection: Max. = maximum dosing rate.
Default value: Max.
NOTICE
114 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Ipol
This parameter is only active with Ipol determinations.
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement.
Selection:
■ 1 µA
■ 20 µA
■ 50 µA
■ 100 µA
Default value: 1 µA
Electrode test
This parameter is only active with Ipol determinations.
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started.
Switch:
■ OFF
■ ON
Default value: OFF
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration.
Adjust this volume to the size of the titration vessel in order to prevent the
contents from overflowing.
Input range 0.00000 to 9,999.99 mL
Default value 100.000 mL
Additional selection: Off
■■■■■■■■ 115
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Stop EP
The titration is canceled when the specified number of equivalence points
has been found.
Input range 1 to 9
Default value 9
Additional selection: Off
Volume after EP
The entered volume will be dosed when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.
Input range 0.01000 to 9,999.99 mL
Additional selection: Off
Default value: Off
Stop time
The titration is canceled when the specified time has elapsed since the
start of the titration.
Input range 0 to 999,999 s
Additional selection: Off
Default value: Off
Filling rate
Rate at which the dosing cylinder is filled after the titration.
116 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
The maximum filling rate depends on the cylinder volume (see table).
If volatile solvents/solutions or solutions with a high viscosity are
used, the filling rate must be reduced accordingly so that the dosing
unit is not overloaded.
7.12.1.4 Evaluation
Parameters ▶ Evaluation
The parameters for the evaluation of the titration curve are defined under
[Evaluation].
Window
If ON is selected, a measured value range (window) can be defined. Only
equivalence points that are within this window are recognized.
Only one window can be defined.
Switch:
■ OFF
■ ON
Default value: OFF
■■■■■■■■ 117
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ All: All equivalence points will be recognized.
■ Greatest: Only the equivalence point with the greatest ERC value, i.e.
the steepest jump, will be recognized.
■ Last: Only the last equivalence point will be recognized.
■ Off: No evaluation takes place.
Default value: All
118 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
EP criterion
The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.
Input range 0 to 200
Default value 5
■■■■■■■■ 119
Parameters ■■■■■■■■■■■■■■■■■■■■■■
pH
VE V/mL
Figure 49 Tubbs method for determining the equivalence point
The figure shows that the evaluation still requires measured values from
the measuring point list even after the equivalence point.
For the recognition of the EPs found, the set EP criterion is compared to
the ERC (Equivalence point Recognition Criterion) found. The ERC is the
first derivative of the titration curve combined with a mathematical func-
tion that is more sensitive for flat jumps than for steeper ones. EPs whose
ERC is smaller than the defined EP criterion will not be recognized.
The ERC is displayed under Results for each discovered and recognized
EP.
If you adjust the EP criterion retroactively in order to recognize more or
fewer EPs, then you can initiate the reevaluation under Results.
7.12.1.5 Calculation
Parameters ▶ Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations.
For each calculation, you can define whether the result is to be saved as a
titer or as a common variable.
The five possible calculations are shown in a list:
120 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Editing a calculation
2 Click on .
■■■■■■■■ 121
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Result name
The result name is the text that will be shown in the result view and in the
report.
Input: max. 12 characters
Default value: empty
Decimal places
Number of decimal places used to display the result.
Input range 0 to 5
Default value 2
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Switch:
■ OFF
■ ON
Default value: OFF
Save as titer
The result can be saved as titer for the selected solution (the solution must
be selected). If statistics has been switched on, then the current mean
value of the determination series will be saved.
122 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Switch:
■ OFF
■ ON
Default value: OFF
R1=… – R5=…
Shows the calculation formula. The formula editor is opened for the defi-
nition.
Result unit
The result unit is displayed and saved along with the result.
Selection:
■ %
■ mol/L
■ mmol/L
■ g/L
■ mg/L
■ mg/mL
■ ppm
■ g
■ mg
■ mL
■ mg/piece
■ °C
■ µL
■ mL/min
■ User-defined
Default value: %
7.12.1.6 Statistics
Parameters ▶ Statistics
The statistics calculation of a multiple determination is activated under
[Statistics] and definition is made as to how many determinations the
series contains.
Statistics
If it is set to ON, statistics calculations will be carried out for all of the
defined results.
■■■■■■■■ 123
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Switch:
■ OFF
■ ON
Default value: OFF
Number of samples
The number of determinations that are carried out for the statistics calcu-
lations.
Input range 2 to 20
Default value 3
7.12.1.7 Reports
Parameters ▶ Reports
The reports that will be printed out automatically or saved as a PDF report
after a determination are defined under [Reports].
Results
The result report contains the calculated results, equivalence points, end-
points, sample data, etc.
Switch:
■ OFF
■ ON
Default value: OFF
Curve
The report is shown as a curve.
Switch:
■ OFF
■ ON
Default value: OFF
Calculation/statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes checking with an external program
possible.
If the switch under Parameters ▶ Statistics is set to ON, the following
data will be printed out as well:
■ Result and sample size of the individual determinations
■ Mean value as well as absolute and relative standard deviation
124 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
Parameters
All of the parameters of the current method are shown in the parameter
report.
Switch:
■ OFF
■ ON
Default value: OFF
NOTICE
The printer for the report data above is defined under Start
page ▶ System ▶ External devices ▶ Printer.
■ If a connected printer is selected, then the report contains the
data defined by the switches above.
■ If [PDF] is selected as printer and at least one switch is set to ON,
then the report contains the complete data except for the measur-
ing point list. The measuring points are only printed if the switch
measuring point list is set to ON.
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the important
data for a determination. This report can be saved as a TXT file on a USB
flash drive or sent to a LIMS via an RS-232 interface. The output location is
defined under Start page ▶ System ▶ External devices ▶ PC/LIMS
report.
The file name of the TXT file has the following structure:
■■■■■■■■ 125
Parameters ■■■■■■■■■■■■■■■■■■■■■■
PC_LIMS_Report_ID1_YYYYMMDD-hhmmss.txt
Switch:
■ OFF
■ ON
Default value: OFF
126 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Request sample ID
Selection of the sample identification that is queried at the start of the
determination.
Selection:
■ ID1
■ ID2
■ ID1&ID2
■ Off
Default value: Off
Start volume
Volume that is dosed prior to the start of the titration.
Input range 0.00000 to 9,999.99 mL
Default value 0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range 0.02 to Max. mL/min
Additional selection: Max. = maximum dosing rate.
Default value: Max.
NOTICE
■■■■■■■■ 127
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Pause
Waiting time, e.g. for the stabilization of the measured value after the
start, for dissolving solid substances or a reaction time after the dosing of
a start volume.
Input range 0 to 999,999 s
Default value 0s
Activation pulse
If it is activated, an activation pulse is sent to a remote line that starts a
connected Dosimat. We recommend to define a Start delay time for
this.
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
128 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Switch:
■ OFF
■ ON
Default value: OFF
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Switch:
■ OFF
■ ON
Default value: ON
Titration rate
3 predefined sets of parameters and 1 set of parameters that can be
defined manually are available for setting the titration rate.
Selection:
■ Slow: For titrations in which the finest details are to be visible. This
can, however, also lead to an increase in noise, which may result in
unwanted equivalence points.
■ Optimal: For all standard titrations. The parameters have been opti-
mized for the most frequent applications.
■ Fast: For fast and less critical titrations.
■ User: The individual titration parameters can be modified.
Default value: Optimal
■■■■■■■■ 129
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Min. waiting 0s 0s 0s
time
Max. waiting 38 s 26 s 21 s
time
NOTICE
Volume increment
This parameter is only active if the titration rate is set to User.
Volume dosed at each dosing step.
High accuracy requires using the correct volume increment. A good guide-
line is 1/20 of the expected endpoint volume. The volume increment
should be closer to 1/100 of the endpoint volume for steep jumps and
closer to 1/10 for flat jumps.
Small volume increments are used for determining blank values or with
very asymmetrical curves. The accuracy of the evaluation cannot be
increased by using smaller increments as the measured value changes
between two measuring points are then of the same order of magnitude
as the noise.
Input range 0.00005 to 999.900 mL
Default value 0.10000 mL
Dosing rate
This parameter is only active if the titration rate is set to User.
Rate at which the volume increments are dosed.
Input range 0.01 to Max. mL/min
Additional selection: Max. = maximum dosing rate.
Default value: Max.
130 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
Signal drift
This parameter is only active if the titration rate is set to User.
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. This type of titration is
often referred to as equilibrium titration.
NOTICE
■■■■■■■■ 131
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Stirrer
If this parameter is activated, then the stirrer is switched on at the start of
the determination.
Switch:
■ OFF
■ ON
Default value: ON
Stirring rate
Setting the stirring rate. The stirring direction is always clockwise.
Conversion: Value x 120 ± 5 rpm = stirring rate in rpm
e.g.: 8 × 120 ± 5 rpm = 960 ± 40 rpm
Input range 1 to 15
Default value 8
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted, then the temperature will be measured continuously. For determina-
tions in pH mode, the value is used for temperature compensation (elec-
trode slope is adjusted accordingly).
Input range –20.0 to 150.0 °C
Default value 25.0 °C
132 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Sensor
Open the selection list and select a sensor.
The selection depends on the measuring mode. Sensors are defined and
listed under System ▶ Sensors, e.g.:
Solution
Open the selection list and select a solution.
Solutions are defined and listed under System ▶ Solutions, e.g.:
Ipol
This parameter is only active with Ipol determinations.
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement.
Selection:
■ 1 µA
■ 20 µA
■ 50 µA
■ 100 µA
Default value: 1 µA
■■■■■■■■ 133
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Electrode test
This parameter is only active with Ipol determinations.
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started.
Switch:
■ OFF
■ ON
Default value: OFF
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration.
Adjust this volume to the size of the titration vessel in order to prevent the
contents from overflowing.
Input range 0.00000 to 9,999.99 mL
Default value 100.000 mL
Additional selection: Off
134 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Stop EP
The titration is canceled when the specified number of equivalence points
has been found.
Input range 1 to 9
Default value 9
Additional selection: Off
Volume after EP
The entered volume will be dosed when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.
Input range 0.01000 to 9,999.99 mL
Additional selection: Off
Default value: Off
Stop time
The titration is canceled when the specified time has elapsed since the
start of the titration.
Input range 0 to 999,999 s
Additional selection: Off
Default value: Off
Filling rate
Rate at which the dosing cylinder is filled after the titration.
Input range 0.01 to Max. mL/min
Additional selection: Max. = maximum filling rate.
Default value: Max.
NOTICE
The maximum filling rate depends on the cylinder volume (see table).
If volatile solvents/solutions or solutions with a high viscosity are
used, the filling rate must be reduced accordingly so that the dosing
unit is not overloaded.
■■■■■■■■ 135
Parameters ■■■■■■■■■■■■■■■■■■■■■■
7.12.2.4 Evaluation
Parameters ▶ Evaluation
The parameters for the evaluation of the titration curve are defined under
[Evaluation].
Window
If ON is selected, a measured value range (window) can be defined. Only
equivalence points that are within this window are recognized.
Only one window can be defined.
Switch:
■ OFF
■ ON
Default value: OFF
136 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Selection:
■ First: Only the first equivalence point will be recognized.
■ Greatest: Only the equivalence point with the greatest ERC value, i.e.
the steepest jump, will be recognized.
■ Last: Only the last equivalence point will be recognized.
Default value: First
■■■■■■■■ 137
Parameters ■■■■■■■■■■■■■■■■■■■■■■
: EP volume
138 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
For the recognition of the EPs found, the set EP criterion is compared to
the ERC (Equivalence point Recognition Criterion) found. The ERC is the
sum of the measured value changes before and after the jump:
7.12.2.5 Calculation
Parameters ▶ Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations.
For each calculation, you can define whether the result is to be saved as a
titer or as a common variable.
The five possible calculations are shown in a list:
■■■■■■■■ 139
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Editing a calculation
2 Click on .
Result name
The result name is the text that will be shown in the result view and in the
report.
Input: max. 12 characters
Default value: empty
Decimal places
Number of decimal places used to display the result.
140 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Input range 0 to 5
Default value 2
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Switch:
■ OFF
■ ON
Default value: OFF
Save as titer
The result can be saved as titer for the selected solution (the solution must
be selected). If statistics has been switched on, then the current mean
value of the determination series will be saved.
Switch:
■ OFF
■ ON
Default value: OFF
R1=… - R5=…
Shows the calculation formula. The formula editor is opened for the defi-
nition.
Result unit
The result unit is displayed and saved along with the result.
■■■■■■■■ 141
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ %
■ mol/L
■ mmol/L
■ g/L
■ mg/L
■ mg/mL
■ ppm
■ g
■ mg
■ mL
■ mg/piece
■ °C
■ µL
■ mL/min
■ User-defined
Default value: %
7.12.2.6 Statistics
Parameters ▶ Statistics
The statistics calculation of a multiple determination is activated under
[Statistics] and definition is made as to how many determinations the
series contains.
Statistics
If it is set to ON, statistics calculations will be carried out for all of the
defined results.
Switch:
■ OFF
■ ON
Default value: OFF
Number of samples
The number of determinations that are carried out for the statistics calcu-
lations.
Input range 2 to 20
Default value 3
142 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
7.12.2.7 Reports
Parameters ▶ Reports
The reports that will be printed out automatically or saved as a PDF report
after a determination are defined under [Reports].
Results
The result report contains the calculated results, equivalence points, end-
points, sample data, etc.
Switch:
■ OFF
■ ON
Default value: OFF
Curve
The report is shown as a curve.
Switch:
■ OFF
■ ON
Default value: OFF
Calculation/statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes checking with an external program
possible.
If the switch under Parameters ▶ Statistics is set to ON, the following
data will be printed out as well:
■ Result and sample size of the individual determinations
■ Mean value as well as absolute and relative standard deviation
Switch:
■ OFF
■ ON
Default value: OFF
■■■■■■■■ 143
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Switch:
■ OFF
■ ON
Default value: OFF
Parameters
All of the parameters of the current method are shown in the parameter
report.
Switch:
■ OFF
■ ON
Default value: OFF
NOTICE
The printer for the report data above is defined under Start
page ▶ System ▶ External devices ▶ Printer.
■ If a connected printer is selected, then the report contains the
data defined by the switches above.
■ If [PDF] is selected as printer and at least one switch is set to ON,
then the report contains the complete data except for the measur-
ing point list. The measuring points are only printed if the switch
measuring point list is set to ON.
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the important
data for a determination. This report can be saved as a TXT file on a USB
flash drive or sent to a LIMS via an RS-232 interface. The output location is
defined under Start page ▶ System ▶ External devices ▶ PC/LIMS
report.
The file name of the TXT file has the following structure:
PC_LIMS_Report_ID1_YYYYMMDD-hhmmss.txt
Switch:
■ OFF
■ ON
Default value: OFF
144 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Request sample ID
Selection of the sample identification that is queried at the start of the
determination.
■■■■■■■■ 145
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ ID1
■ ID2
■ ID1&ID2
■ Off
Default value: Off
Start volume
Volume that is dosed prior to the start of the titration.
Input range 0.00000 to 9,999.99 mL
Default value 0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range 0.02 to Max. mL/min
Additional selection: Max. = maximum dosing rate.
Default value: Max.
NOTICE
146 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Pause
Waiting time, e.g. for the stabilization of the measured value after the
start, for dissolving solid substances or a reaction time after the dosing of
a start volume.
Input range 0 to 999,999 s
Default value 0s
Activation pulse
If it is activated, an activation pulse is sent to a remote line that starts a
connected Dosimat. We recommend to define a Start delay time for
this.
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
■■■■■■■■ 147
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Switch:
■ OFF
■ ON
Default value: ON
Stirrer
If this parameter is activated, then the stirrer is switched on at the start of
the determination.
Switch:
■ OFF
■ ON
Default value: ON
Stirring rate
Setting the stirring rate. The stirring direction is always clockwise.
Conversion: Value x 120 ± 5 rpm = stirring rate in rpm
e.g.: 8 × 120 ± 5 rpm = 960 ± 40 rpm
Input range 1 to 15
Default value 8
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted, then the temperature will be measured as follows:
■ For measured quantity Ipol: Temperature measurement in the begin-
ning and at the end of the titration.
■ For measured quantities U and pH: Continuous temperature measure-
ment.
For determinations in pH mode, the value is used for temperature com-
pensation (electrode slope is adjusted accordingly).
148 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Sensor
Open the selection list and select a sensor.
The selection depends on the measuring mode. Sensors are defined and
listed under System ▶ Sensors, e.g.:
Solution
Open the selection list and select a solution.
Solutions are defined and listed under System ▶ Solutions, e.g.:
Titration direction
We, at Metrohm, recommend that you specify whenever possible whether
the change of the measured value is positive or negative.
If two endpoints have been set, then the titration direction will be defined
automatically. In this case, the setting will be ignored.
■■■■■■■■ 149
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ +: Positive measured value change, i.e. in the direction of a higher pH
value, greater voltage or greater current.
■ -: Negative measured value change, i.e. in the direction of a lower pH
value, lesser voltage or lesser current.
■ Auto: The titration direction is determined automatically from the ini-
tial measured value and the set endpoint.
Default value: Auto
Extraction time
Minimum duration of the titration. The titration will not be canceled dur-
ing the extraction time, even if the endpoint has already been reached.
The titration is, however, canceled if a stop condition is fulfilled in this
time. The entry of an extraction time may be advisable, for instance, for
the titration of sparingly soluble samples.
Input range 0 to 999,999 s
Default value 0s
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 to 999,999.0 s
Default value 2.0 s
Ipol
This parameter is only active with Ipol determinations.
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement.
Selection:
■ 1 µA
■ 20 µA
■ 50 µA
■ 100 µA
Default value: 1 µA
Electrode test
This parameter is only active with Ipol determinations.
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started.
150 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Switch:
■ OFF
■ ON
Default value: OFF
Titration rate
3 predefined sets of parameters and 1 set of parameters that can be
defined manually are available for setting the titration rate.
Selection:
■ Slow: For titrations with a long reaction time or a steep increase/
decrease at the endpoint. The titration time can be very long.
■ Optimal: For all standard titrations. The parameters have been opti-
mized for the most frequent applications.
■ Fast: For titrations that show a flat curve progression at the endpoint.
The titration is carried out very fast.
■ User: The individual titration parameters can be modified.
Default value: Optimal
■■■■■■■■ 151
Parameters ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
Minimum rate
This parameter is only active if the titration rate is set to User.
Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration.
This parameter has a vital influence on the titration rate and thus also on
the accuracy. The smaller the selected minimum rate, the slower the titra-
tion.
Input range 0.01 to 9,999.00 µL/min
Default value 25.00 µL/min
Maximum rate
This parameter is only active if the titration rate is set to User.
Rate at which dosing is carried out outside of the control range.
Input range 0.01 to Max. mL/min
Default value 10.00 mL/min
Additional selection: Max. = maximum dosing rate.
152 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
■■■■■■■■ 153
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Stop criterion
The titration is canceled when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion was defined, the titra-
tion is completed according to the stop conditions.
Selection:
■ Drift: The titration is canceled once the stop drift has been reached.
■ Time: The titration is canceled once the Delay time has been
reached.
■ Off: The titration will not be canceled until the stop conditions are
fulfilled.
Default value: All
Stop drift
This parameter is only active if the stop criterion is set to Drift.
The titration is canceled as soon as the endpoint and the stop drift have
been reached.
Input range 1 to 999 µL/min
Default value 20 µL/min
Delay time
This parameter is only active if the stop criterion is set to Time.
When the endpoint has been reached, the Delay time is allowed to
elapse after the last dosing and the titration is then stopped.
Input range 0 to 999 s
Default value 10 s
NOTICE
The parameters and their selection and input ranges are identical to
the Control parameters EP1.
154 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration.
Adjust this volume to the size of the titration vessel in order to prevent the
contents from overflowing.
Input range 0.00000 to 9,999.99 mL
Default value 100.000 mL
Additional selection: Off
Stop time
The titration is canceled when the specified time has elapsed since the
start of the titration.
Input range 0 to 999,999 s
Additional selection: Off
Default value: Off
Filling rate
Rate at which the dosing cylinder is filled after the titration.
Input range 0.01 to Max. mL/min
Additional selection: Max. = maximum filling rate.
Default value: Max.
NOTICE
The maximum filling rate depends on the cylinder volume (see table).
If volatile solvents/solutions or solutions with a high viscosity are
used, the filling rate must be reduced accordingly so that the dosing
unit is not overloaded.
■■■■■■■■ 155
Parameters ■■■■■■■■■■■■■■■■■■■■■■
7.12.3.6 Calculation
Parameters ▶ Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations.
For each calculation, you can define whether the result is to be saved as a
titer or as a common variable.
The five possible calculations are shown in a list:
Editing a calculation
2 Click on .
156 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Result name
The result name is the text that will be shown in the result view and in the
report.
Input: max. 12 characters
Default value: empty
Decimal places
Number of decimal places used to display the result.
Input range 0 to 5
Default value 2
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
■■■■■■■■ 157
Parameters ■■■■■■■■■■■■■■■■■■■■■■
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Switch:
■ OFF
■ ON
Default value: OFF
Save as titer
The result can be saved as titer for the selected solution (the solution must
be selected). If statistics has been switched on, then the current mean
value of the determination series will be saved.
Switch:
■ OFF
■ ON
Default value: OFF
R1=… - R5=…
Shows the calculation formula. The formula editor is opened for the defi-
nition.
Result unit
The result unit is displayed and saved along with the result.
Selection:
■ %
■ mol/L
■ mmol/L
■ g/L
■ mg/L
■ mg/mL
■ ppm
■ g
■ mg
■ mL
■ mg/piece
■ °C
■ µL
■ mL/min
■ User-defined
Default value: %
158 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
7.12.3.7 Statistics
Parameters ▶ Statistics
The statistics calculation of a multiple determination is activated under
[Statistics] and definition is made as to how many determinations the
series contains.
Statistics
If the switch is set to ON, statistics calculations will be carried out for all of
the defined results.
Switch:
■ OFF
■ ON
Default value: OFF
Number of samples
The number of determinations that are carried out for the statistics calcu-
lations.
Input range 2 to 20
Default value 3
7.12.3.8 Reports
Parameters ▶ Reports
The reports that will be printed out automatically or saved as a PDF report
after a determination are defined under [Reports].
Results
The result report contains the calculated results, equivalence points, end-
points, sample data, etc.
Switch:
■ OFF
■ ON
Default value: OFF
Curve
The report is shown as a curve.
■■■■■■■■ 159
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Switch:
■ OFF
■ ON
Default value: OFF
Calculation/statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes checking with an external program
possible.
If the switch under Parameters ▶ Statistics is set to ON, the following
data will be printed out as well:
■ Result and sample size of the individual determinations
■ Mean value as well as absolute and relative standard deviation
Switch:
■ OFF
■ ON
Default value: OFF
Switch:
■ OFF
■ ON
Default value: OFF
Parameters
All of the parameters of the current method are shown in the parameter
report.
Switch:
■ OFF
■ ON
Default value: OFF
160 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
NOTICE
The printer for the report data above is defined under Start
page ▶ System ▶ External devices ▶ Printer.
■ If a connected printer is selected, then the report contains the
data defined by the switches above.
■ If [PDF] is selected as printer and at least one switch is set to ON,
then the report contains the complete data except for the measur-
ing point list. The measuring points are only printed if the switch
measuring point list is set to ON.
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the important
data for a determination. This report can be saved as a TXT file on a USB
flash drive or sent to a LIMS via an RS-232 interface. The output location is
defined under Start page ▶ System ▶ External devices ▶ PC/LIMS
report.
The file name of the TXT file has the following structure:
PC_LIMS_Report_ID1_YYYYMMDD-hhmmss.txt
Switch:
■ OFF
■ ON
Default value: OFF
■■■■■■■■ 161
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute.
Input range 0.1 to 999.0 mV/min
Default value 2.0 mV/min
Additional selection: Off: Measured value acceptance will take place after
the maximum waiting time has elapsed. This can be useful when the elec-
trode is slow to respond.
Temperature
Calibration temperature. If a temperature sensor is connected, the calibra-
tion temperature is automatically applied. The calibration temperature can
also be entered manually.
The calibration temperature allows for the temperature compensation dur-
ing a determination. If a temperature sensor is connected, this happens
automatically. If there is no temperature sensor, the temperature can be
entered manually as a method parameter. The calibration temperature is
necessary for selecting the corresponding pH value of the buffer.
Input range –20.0 to 150.0 °C
Default value 25.0 °C
162 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Sensor
Open the selection list and select a sensor.
Only sensors of the pH electrode type can be selected.
Stirrer
If this parameter is activated, then the stirrer is switched on at the start of
the determination.
Switch:
■ OFF
■ ON
Default value: ON
Stirring rate
Setting the stirring rate. The stirring direction is always clockwise.
Conversion: Value × 120±5 rpm = stirring rate in rpm
e.g.: 8 × 120±5 rpm = 960±40 rpm
Input range 1 to 15
Default value 8
7.12.4.2 Buffers
Parameters ▶ Buffers
The statistics calculation of a multiple determination is activated under
[Buffers] and definition is made as to how many determinations the ser-
ies contains.
Buffer type
Selection of a predefined buffer series or definition of special buffers. In
the case of predefined buffer series, the instrument automatically recogni-
zes which buffer is involved.
■■■■■■■■ 163
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Selection:
■ Baker
■ Beckmann
■ DIN
■ Fisher
■ Fluka Basel
■ Hamilton
■ Merck CertiPUR
■ Merck Titrisol
■ Metrohm
■ Mettler
■ NIST
■ Precisa
■ Radiometer
■ Special
Default value: Metrohm
NOTICE
Merck CertiPUR:
Reference temperature = 25 °C.
When using Merck CertiPUR buffers with a reference temperature of
20 °C, the buffer type Merck Titrisol must be selected.
NOTICE
Special:
Up to 5 calibration buffers can be defined in the method. The buffers
are not recognized automatically here. Enter the buffers exactly in the
specified sequence.
Number of buffers
Number of buffers that are used for calibration.
If calibration is accomplished with more than two buffers, then certain
buffers can be used repeatedly in order to give them more statistical
weight. The first two buffers must, however, always be different from one
another.
164 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Operation and control
Selection:
■ 1
■ 2
■ 3
■ 4
■ 5
Default value: 2
Buffer 1 pH
This parameter is only available if the buffer type is set to Special.
Input range –20.000 to 20.000
Default value 7.000
Buffer 2 pH
This parameter is only available if the buffer type is set to Special.
Input range –20.000 to 20.000
Additional selection: Off
Default value: 4.000
Buffer 3 pH
This parameter is only available if the buffer type is set to Special.
Input range –20.000 to 20.000
Additional selection: Off
Default value: Off
Buffer 4 pH
See buffer 3 pH
Buffer 5 pH
See buffer 3 pH
7.12.4.3 Reports
Parameters ▶ Reports
The reports that will be printed out automatically or saved as a PDF report
after a determination are defined under [Reports].
Results
The result report contains the specifications for the calibration (slope,
pH(0), etc.).
■■■■■■■■ 165
Parameters ■■■■■■■■■■■■■■■■■■■■■■
Switch:
■ OFF
■ ON
Default value: OFF
Parameters
All of the parameters of the current method are shown in the parameter
report.
Switch:
■ OFF
■ ON
Default value: OFF
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the important
data for a determination. This report can be saved as a TXT file on a USB
flash drive or sent to a LIMS via an RS-232 interface. The output location is
defined under System ▶ External devices.
The file name of the TXT file has the following structure:
PC_LIMS_Report_ID1_YYYYMMDD-hhmmss.txt
Switch:
■ OFF
■ ON
Default value: OFF
166 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance
8 Maintenance
8.1 Maintenance
Regularly perform maintenance work on the product to prevent malfunc-
tions and to ensure a long service life.
■ Metrohm recommends having the products maintained by specialist
personnel of Metrohm AG as part of an annual service. Shorter mainte-
nance intervals may be necessary if you frequently work with caustic
and corrosive chemicals.
■ Only perform maintenance work that is described in this instruction.
Contact your regional Metrohm service representative for further main-
tenance work and repairs. The regional Metrohm service representative
offers every form of technical advice for maintenance and service of all
Metrohm products.
■ Only use spare parts that meet the technical requirements of the man-
ufacturer. Original spare parts always meet these requirements.
WARNING
■■■■■■■■ 167
Cleaning the product surface ■■■■■■■■■■■■■■■■■■■■■■
WARNING
Electrical potential
Contact with electrical potential can cause serious injuries or death.
■ Operate the product only if it is in perfect condition. The housing
must also be intact.
■ Only use the product with the covers fitted.
■ Protect live components (e.g. power supply unit, power cord, con-
nection sockets) against moisture.
■ Always have maintenance work and repairs on electrical compo-
nents carried out by a regional Metrohm service representative.
Prerequisite:
■ The product is switched off and disconnected from the energy supply.
Required accessories:
■ Cleaning cloth (soft, lint-free)
■ Water or ethanol
CAUTION
168 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance
CAUTION
Maintenance steps
■ Emptying the cylinder unit as far as possible and removing it
■ Taking the cylinder unit apart
■ Cleaning the cylinder unit
■ Assembling the cylinder unit
Once the maintenance steps have been carried out professionally, the cyl-
inder unit can be mounted again.
■■■■■■■■ 169
Cleaning the product surface ■■■■■■■■■■■■■■■■■■■■■■
NOTICE
Push the dosing cylinder out of the light protection from above.
Lift the piston carefully out of the dosing cylinder. Use the
6.1546.040 piston tool to accomplish this.
The individual parts can now be cleaned and checked.
1 Clean the individual parts of the cylinder unit with deionized water.
3 Check the individual parts of the cylinder unit (dosing cylinder, piston,
sealing lips and piston rod) for the following defects:
■ Are rough areas or scratches visible on the dosing cylinder?
170 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance
Required accessories:
■ 6.2803.010 paraffin grease
NOTICE
2 Carefully slide the piston far enough into the dosing cylinder that the
piston rod still protrudes out of it by approximately 6 mm.
3 Push the dosing cylinder far enough into the light protection so that
its flange is securely up against the mounting ring (gray plastic ring).
The cylinder unit can now be mounted.
■■■■■■■■ 171
Cleaning the product surface ■■■■■■■■■■■■■■■■■■■■■■
Prerequisite
The instrument is switched on.
The [Exchange cylinder unit] process was executed up to the point
when the cylinder unit can be removed from the instrument.
The maintenance has been carried out professionally or a new cylinder
unit is ready to be mounted.
The instrument is carrying out the [Exchange cylinder unit] procedure
and the push rod of the dosing drive is at the height at which the cylinder
unit can be mounted. The following message is displayed:
Procedure
1 To mount the cylinder unit, the piston rod must be pulled out of the
dosing cylinder by approx. 6 mm. If necessary, pull the piston care-
fully out of the dosing cylinder with the 6.1546.040 piston tool.
While doing so, make sure that the sealing lips and the piston in the
dosing cylinder are not damaged.
172 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance
Make sure that the cylinder unit has been mounted correctly.
[Continue]
6 [Continue]
NOTICE
Make sure that the value for the cylinder volume in the System
menu is the same as the volume of the mounted cylinder unit.
7
NOTICE
■■■■■■■■ 173
Resetting the system ■■■■■■■■■■■■■■■■■■■■■■
CAUTION
If the system is reset, all user data (methods, solutions, etc.) will be
deleted. The instrument will be reset to factory settings. The pass-
word for the Expert dialog type is: METROHM9100
We recommend creating a backup of the system at regular intervals
in order to avoid data losses.
The program version does not change when resetting the system.
Prerequisite:
■ The instrument is switched off.
174 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Maintenance
NOTICE
■■■■■■■■ 175
■■■■■■■■■■■■■■■■■■■■■■
9 Troubleshooting
Messages on malfunctions and errors are displayed in the control software
or in the embedded software (e.g. on the display of an instrument) and
contain the following information:
■ Descriptions of causes of malfunctions (e.g. jammed drive)
■ Descriptions of problems with the control (e.g. missing or invalid
parameter)
■ Information on how to solve the problem
System components with status display elements also indicate malfunc-
tions and errors with a red flashing LED.
Troubleshooting on the product is often only possible with the control
software or the embedded software (e.g. initializing, moving to a defined
position).
See also
176 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Disposal
10 Disposal
Properly dispose of chemicals and of the product to reduce negative
effects on the environment and public health. Local authorities, waste dis-
posal companies or dealers provide more detailed information on disposal.
Observe the WEEE EU directive (WEEE = Waste Electrical and Electronic
Equipment) for the proper disposal of waste electronic equipment within
the European Union.
■■■■■■■■ 177
Ambient conditions ■■■■■■■■■■■■■■■■■■■■■■
11 Technical specifications
Storage +5 to +45 °C
Instrument
Input
Nominal voltage 24 VDC
Power consumption max. 20 W
Output
Nominal voltage 24 VDC
Power output max. 45 W
USB connector
Nominal voltage 5V
178 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Technical specifications
Current at the power supply unit 500 mA max. output current per
channel
Protection
Internal fuse 1.5 A
11.3 Dimensions
Measurements
Width 286 mm
Height
without cylinder unit 220 mm
with cylinder unit 358 mm
with support rod 508 mm
Depth 286 mm
11.4 Housing
Materials
Cover PP 20% filled with talc
Back panel 1.4301 stainless steel
Base PP 20% filled with talc
Front foils PET EBA 180, anti-glare
IP degree of protection IP 40
■■■■■■■■ 179
Connectors specifications ■■■■■■■■■■■■■■■■■■■■■■
Power OUT
Socket round plug 4-pin
Remote
Socket D-Sub 9-pin
Ethernet
Type CAT 6
Socket RJ-45
Cable type min. FFTP shielded
Cable length max. 10 m from Metrohm acces-
sories
USB
Type 2.0
Socket type A
Cable type shielded
Cable length max. 5 m from Metrohm acces-
sories
Measuring inputs
Ind
Socket type F measuring input for
potentiometric electro-
des
Temp
180 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Technical specifications
Keys 5 keys
■■■■■■■■ 181
Measurement specifications ■■■■■■■■■■■■■■■■■■■■■■
Temperature
Pt1000
Measuring range –150 to +250 °C
Measuring resolution 0.1 °C
Measuring accuracy ±0.4 °C in the measuring range
–20.0 °C to +150.0 °C
NTC 30 kOhm
Measuring range –5 to +250 °C
Measuring resolution 0.1 °C
Measuring accuracy ±0.6 °C in the measuring range
+10.0 °C to +40.0 °C
Polarizer
Ipol DC
Polarization current 1, 20, 50, 100 µA can be selected
Measuring range 0–3,500 mV
Measuring resolution 0.1 mV
Reference conditions
Instrument status min. 30 minutes in operation
Adjusting interval annual
182 ■■■■■■■■
■■■■■■■■■■■■■■■■■■■■■■ Technical specifications
Dosing drive
Dosing resolution 20,000 steps per cylinder vol-
ume
Dosing accuracy according to ISO/DIN 8655-3
Tubing
Tubing nipple outer thread M6
Inner diameter 2 mm
Material FEP fluorinated ethylene
propylene
■■■■■■■■ 183