Unipulse 400

The document describes the UniPulse 400, a portable defibrillator analyzer. It can analyze monophasic, biphasic, and pulsed waveforms, and has a pacer function. Key features include a simple interface, large color screen to view waveforms, memory storage, and compatibility with many defibrillator types. It provides comprehensive testing of defibrillators with accurate measurements of values like energy, voltage, current, and more.

Uploaded by

Eliel Velasquez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
49 views6 pages

Unipulse 400

The document describes the UniPulse 400, a portable defibrillator analyzer. It can analyze monophasic, biphasic, and pulsed waveforms, and has a pacer function. Key features include a simple interface, large color screen to view waveforms, memory storage, and compatibility with many defibrillator types. It provides comprehensive testing of defibrillators with accurate measurements of values like energy, voltage, current, and more.

Uploaded by

Eliel Velasquez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

rigelmedical.

com

UniPulse 400
An easy to use defibrillator
analyzer with pacer

A portable comprehensive Why buy the UniPulse 400? Key Features:


defibrillator analyzer capable of
> Dedicated fast keys and simple to use > Light handheld, battery operation
analyzing all monophasic, biphasic,
operating system enables biomedical /
standard, and pulsating waveforms… > Pacer functionality
clinical engineers to select the relevant
The UniPulse 400 is designed to test all test function in seconds > Compatible with monophasic,
defibrillators accurately and efficiently. biphasic and pulsed waveforms
> A large vibrant colour screen
An unrivalled functionality and our that displays both test results and > Reliable and non-inductive test
pacer function make the UniPulse 400 waveforms in amazing detail load of 50Ω
the ultimate defibrillator analyzer on the > Patient simulator with ECG
> Large internal memory and on-board
market. high-level output
results storage will reduce test time and
help work towards a modern paperless
system

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400
rigelmedical.com

What’s in the box? Technical Specification:


> AC Power Supply Defib - Energy Measurement
Load Resistance 50Ω ± 1% non-inductive
> USB cable
Range 0 - 199.9 Joules
> ECG Snap Adapters (pack of 10) Accuracy ± 1% of reading ± 0.1 Joule
Range (High) 200 - 600 Joules
> Calibration Certificate
Accuracy ± 1% of reading ± 1 Joule
> UniPulse 400 Analyser Resolution 0.1 Joules
Voltage 0 - 5000 Volts
> UniPulse 400 Carry Case
Current 0 - 100 Amps
> UniPulse 400 Quick Start Guide Sampling Rate 250 kHz sampling frequency
Maximum pulse width (capture time) 5us – 120ms
Optional Accessories:
Charge Time
> Bluetooth barcode scanner Measurement is initiated by pressing the function key and simultaneously
(339A923) charging the defibrillator. Charge time is recorded after the defibrillator has
> Test ’n’ Tag Elite 2 Bluetooth printer discharged into the UniPulse 400
(339A989) Clock accuracy 0.10s to 100s +/-0.05s
> Thermal label rolls - pack of 5 AED Pulse Mode Waveforms
(339A949) Normal Sinus Rhythm (NSR) 20 - 300 BPM, Amplitude 1.00mV
> Paddle adaptor box (386A950) (±2%)
Asystole (ASYS)
> Paddle adaptor box lead - Red Ventricular Fibrillation - Coarse (VFBC) 240 BPM, Amplitude 1.00mV (±2%)
(386A952) Ventricular Fibrillation - Fine (VFBF) 240 BPM, Amplitude 1.00mV (±2%)
> Paddle adaptor box lead – Black Monomorphic Ventricular Tachycardia 210 BPM, Amplitude 1.00mV (±2%)
(386A953) (MVT)
Atrial Fibrillation (AFB) 20 - 300 BPM,
> Applied part adaptor (pack of 10) Amplitude 1.00mV (±2%)
(404A951)
ECG Simulator
> AC power supply (386A011) ECG simulation with high-level output.
> Battery pack 9.6V/2400mAh - NiMH ECG Waveforms - Sinus
(386A008) Normal Sinus Rhythm (NSR) 20 - 300 BPM,
> Replacement carry case (410A950) Amplitude 0.50 - 5.00mV (±2%)
ST Elevation (STE) 20 - 300 BPM,
> Med-eBase software (383A910) Amplitude 0.50 - 5.00mV (±2%)
ST Depression (STD) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Myocardial Infarction (MI) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Tall T (TT) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400
rigelmedical.com

Service and Warranty Technical Specification (Continued):

The UniPulse 400 comes with a ECG Waveforms – Ventricular Arrhythmias


free upgraded 24-month warranty, Premature Ventricular Contraction 80 BPM,
subject to terms and conditions, - Intermittent (PVCI) Amplitude 0.50 - 5.00mV (±2%)
available at Bigeminy (BIG) 80 BPM,
Amplitude 0.50 - 5.00mV (±2%)
rigelmedical.com/register Trigeminy (TRIG) 80 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Ventricular Flutter (VFLT) 240 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Ventricular Fibrillation - Coarse (VFBC) 240 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Ventricular Fibrillation - Fine (VFBF) 240 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Monomorphic Ventricular Tachycardia 210 BPM,
(MVT) Amplitude 0.50 - 5.00mV (±2%)
Right-focal Premature Ventricular 80 BPM,
Contraction (RFPVC) Amplitude 0.50 - 5.00mV (±2%)
ECG Waveforms – Atrial Conduction Arrhythmias
First Degree AV Bock (FAVB) 80 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Second Degree AV Block 80 BPM,
- Mobitz I (SAVB_MI) Amplitude 0.50 - 5.00mV (±2%)
Second Degree AV Block 80 BPM,
- Mobitz II (SAVB_MII) Amplitude 0.50 - 5.00mV (±2%)
Third Degree AV Block (TAVB) 50 BPM,
Amplitude 0.50 - 5.00mV (±2%)
ECG Waveforms – Atrial Arrhythmias
Sinus Arrhythmia (SAR) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Missing Beat (MB) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Atrial Flutter (AFLT) 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Atrial Fibrillation (AFB) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Paroxysmal Atrial Tachycardia (PAT) 180 BPM,
Amplitude 0.50 - 5.00mV (±2%)
Premature Junctional Contraction (PJC) 20 - 300 BPM,
Amplitude 0.50 - 5.00mV (±2%)

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400
rigelmedical.com

Technical Specification (Continued):


ECG Pacer Waveforms
Synchronous Atrial (AAI) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV,
Pulse width 0.1 – 2.0ms
Asynchronous Atrial (AOO) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV,
Pulse width 0.1 – 2.0ms
Pacer (PCR) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV,
Pulse width 0.1 – 2.0ms
Ventricular Pacer (VVI) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV,
Pulse width 0.1 – 2.0ms
Atrial & Ventricular Pacer (DDD) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV,
Pulse width 0.1 – 2.0ms
R-Wave Detection (RWD) 20 - 300 BPM,
Pulse amplitude 0.50 – 5.00mV
ECG Performance Waveforms
Sine (SINE) 0.1 - 300Hz, 1.00 – 10.00mV
Square (SQ) 0.1 - 300Hz, 1.00 – 10.00mV
Triangle (TRI) 0.1 - 300Hz, 1.00 – 10.00mV
Sawtooth (SAW) 0.1 - 300Hz, 1.00 – 10.00mV
Inverse Sawtooth (INVSAW) 0.1 - 300Hz, 1.00 – 10.00mV
Pulse (PULSE) 0.1 - 300Hz, 0.50 – 5.00mV
ECG Noise Selection
Amplitude 0 – 10.00mV
Frequency 50 - 60Hz
ECG Waveform Output
Low Level 5-lead ECG and on Paddles
Hi Level Output Jack
ECG Accuracy
Rate ± 1% bpm
Amplitude ± 2% (LA-LL), ± 10% (Paddles)
Lead II: 1 – 10 mV
(in steps of 0.5 mV). Other leads are
proportional to Lead II by the
following percentages:
Lead I: 60 %
Lead II: 100 %
Lead III: 40 %

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400
rigelmedical.com

Technical Specification (Continued):


ECG Accuracy (Continued)
V1: 63 % [ Reference LA ]
V2: 71 % [ Reference LA ]
V3: 68 % [ Reference LA ]
V4: 80 % [ Reference LA ]
V5: 55 % [ Reference LA ]
V6: 49 % [ Reference LA ]
Pacer Input
Fix Load 50Ω
Accuracy 1 %, non-inductive
Over voltage protection 5000V
Variable Load 50 to 1600Ω in 50Ω steps
Accuracy 1 %, non-inductive
Pulse Rate 5.0 to 800 ppm
Accuracy +/- (0.5% rdg + 0.1ppm)
Heart rate selection 20 – 300 bpm
Under & overdrive 85% (20 bpm min) and
115% (300 bpm max)
Wave form selection NSR, VFibC, VFibF, MVT, AFib,
Missing Beat, R-Wave detection
Pulse Current Amplitude 5.00 – 200mA
Accuracy +/- (1% rdg +0.02mA)
Current Measurements Average (RMS),
Leading edge, Trailing edge,
Peak (the highest during the pulse)
Pulse Width 1.00 - 100ms
Accuracy +/- (0.5% rdg + 0.01ms)
Pulse Energy 1µJ – 2.00J
Accuracy +/- 4% + 10µJ
Pacer Manufacturer Algorithms
CU Medical, GE, HP, Laerdal, Mindray, Philips, PhysioControl, Schiller,
WelchAllyn, Zoll
Pacer Refractory Periods
Refractory Period test 15 – 500mS (Paced and sensed)
Accuracy +/- 1ms
Pacer Sensitivity Test
Wave form, R Wave Polarity Normal and Reversed,
selectable
Dynamic Sensitivity 0.05mV to 5.00mV in 50µV steps
Pacer Interference Test (Immunity):
Heart rate 20-300 bpm
Frequency 50 or 60 Hz
Noise level in mV 0-15.0mV

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400
rigelmedical.com

General Specifications:
Dimensions 220mm x 150mm x 90mm
(L x W x H)
Weight 1.5kg
Operation 9.6V/2400mAh
Nickel Metal Hydride battery pack
Battery charge time 2.5 hours
Battery capacity (fully charged) 12 hours
Mains Supply 110/230V AC; 48 to 66Hz,
35VA power supply
Storage Environment -15°C to +60°C
Operating Conditions 0°C to +40°C
Environmental Protection IP 40
Communication USB
Display LCD colour graphic display ¼” VGA
Memory 100 test results including graphs
Impact rating 5J

Rev 2

Passionate about patient safety.


T: +44 (0) 191 587 8730 E: [email protected]

See the UniPulse 400 for yourself, request your FREE live
online demonstration today at rigelmedical.com/unipulse400