Operational Steps DM4P
Operational Steps DM4P
The Software opens as below with initialization of stepping-stage for X & Y home-positions –
The below confirmation message appears and it needs to confirm by pressing YES option.
(Here basically two files are created with the same name entered by the operator but their
extensions are different; one is with .lg1 extension and the 2 nd one is with .lg2 extension. The
.lg2 extension file contains the experiment o/p data).
• Go to SLIDE→ CONFIGURE to enter the no. of counts and scan-pattern. The below screen
opens -
Enter target counts in “Number of Steps” area and select ‘Elliptical’ Scan-pattern for Round-
shaped pallet or ‘Rectangular’ pattern for Square pallet. Close the window by pressing OK
button.
• Define scan-area from the pallet by clicking SLIDE→ AREA OF INTEREST. The below screen
comes-
USER
The stepper-stage (petrog) can be moved to reach the desired sample-position by pressing
the soft-buttons around the Displayed slide-area –
Once AOI is created correctly by defining both the reference points, a red coloured area
will be formed. The X & Y-step size will be displayed as below-
This step size depends on two factors -> ‘AOI size’ & ‘no. of counts’ entered. As AOI is mostly
fixed due to fixed sample diameter in a particular lab, the step-size can be adjusted by
entering no. of counts more than 500 (e.g., if 500 counts give step size of 750 micron,
around 1000 counts can be entered to obtain a step-size of 500 micron.) Close the AOI-
selection window by confirming OK.
• The distribution of counting-points can be visualised from SLIDE→ SHOW POINTS
• To see the report, open a new excel file and from ‘File→ Open’ option browse for the
particular .lg2 file and advance through the steps to finish (the COMMA option should be
checked while moving through the steps.)