Man 4820 3601 LC Vanquish UHPLC Man48203601 EN
Man 4820 3601 LC Vanquish UHPLC Man48203601 EN
Man 4820 3601 LC Vanquish UHPLC Man48203601 EN
Vanquish
Operating Manual
4820.3601-EN Revision 5.0 • August 2022
*4820.3601-EN-5.0*
Copyright © 2022 Thermo Fisher Scientific Inc. All rights reserved.
Trademarks
Acrobat, Adobe, and Adobe Reader are trademarks of Adobe Systems Incorporated.
Microsoft and Windows are trademarks of Microsoft Corporation.
MP35N is a trademark of SPS Technologies.
Torx is a trademark of Acument Intellectual Properties, LLC.
All other trademarks are property of Thermo Fisher Scientific and its subsidiaries.
Disclaimer
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to
use in the product operation. The document is copyright protected; any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization
of Thermo Fisher Scientific Inc.
This manual is provided "as is." The contents of this manual are subject to being changed,
without notice, in future revisions.
Thermo Fisher Scientific Inc. makes no representations that this document is complete,
accurate, or error-free. Thermo Fisher Scientific Inc. assumes no responsibility and will not be
liable for any errors, omissions, damage, or loss that might result from any use of this document,
even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale.
The Terms and Conditions of Sale shall govern all conflicting information between the two
documents.
Manufacturer's address
Dionex Softron GmbH, Part of Thermo Fisher Scientific, Dornierstrasse 4, D-82110 Germering
Contacting Us
Contacting Us
There are several ways to contact us:
Ordering Information
For ordering information or sales support for HPLC products, contact
your local Thermo Fisher Scientific sales organization. For contact
information, go to Contact Us on http://www.thermofisher.com.
Technical Assistance
For technical support for HPLC products, contact your local Thermo
Fisher Scientific support organization. For contact information, go to
Contact Us on http://www.thermofisher.com.
Contents
1 Using this Manual......................................................................... 9
1.1 About this Manual ............................................................................................... 10
1.2 Conventions ......................................................................................................... 11
1.2.1 Conventions for Safety Messages ............................................................ 11
1.2.2 Special Notices ......................................................................................... 11
1.2.3 Typographical Conventions ...................................................................... 12
1.3 Reference Documentation................................................................................... 13
6 Operation................................................................................... 83
6.1 Introduction to this Chapter ................................................................................ 84
6.2 Safety Guidelines for Operation........................................................................... 85
6.3 Power On/Off Control.......................................................................................... 86
6.4 Preparing the System for Operation .................................................................... 87
6.5 Use of Solvents and Additives.............................................................................. 89
12 Appendix.................................................................................. 147
12.1 Compliance Information .................................................................................... 148
12.1.1 Declarations of Conformity .................................................................... 148
12.1.2 WEEE Compliance .................................................................................. 149
12.1.3 FCC Compliance...................................................................................... 149
12.1.4 Manual Release History.......................................................................... 150
1.2 Conventions
This section describes the conventions that are used throughout this
manual.
Make sure that you understand and follow all safety messages
presented in this manual.
NOTICE
Highlights information necessary to prevent damage to the system or
invalid test results.
• Input that you enter by the keyboard or that you select with the
mouse
Viewpoint
If not otherwise stated, the expressions left and right in this manual
refer to the viewpoint of a person that is facing the system from the
front.
• Index entries
Hardware Documentation
Additional hardware documentation includes the following:
Software Documentation
Additional software documentation includes the following:
• Installation Guide
For basic information about device installation and configuration,
refer to the Installation Guide.
• Reference Card
For a concise overview of the most important workflows, refer to
the Reference Card.
TIP The Chromeleon Help and documents are included in the software
shipment.
Third-Party Documentation
Refer also to the user documentation provided by the manufacturers of
third-party components and materials, for example, Safety Data Sheets
(SDSs).
2 Safety
This chapter provides general and specific safety information and
informs about the intended use of the system.
The safety symbols and signal words in this manual include the
following:
CAUTION
Indicates a hazardous situation that, if not avoided, could result in minor
or moderate injury.
WARNING
Indicates a hazardous situation that, if not avoided, could result in
serious injury.
• Always keep the manual near the system for quick reference.
Symbol Description
Laboratory Practice
Thermo Fisher Scientific recommends that the laboratory in which the
Vanquish system is used follow best practices for LC analyses. This
includes among others:
• Establishing shelf life limits and following them for all consumables
used with the system
Installation
Operation
General Operation
All users must know the hazards presented by the system and the
substances they are using. All users should observe the related Safety
Data Sheets (SDSs).
Protective Clothing
To protect you from chemical splashes, harmful liquids, or other
contamination, put on appropriate protective clothing, such as a lab
coat.
Protective Eyewear
To prevent liquids from striking your eyes, put on appropriate protective
eyewear, such as safety glasses with side shields. If there is a risk of
splashing liquids, put on goggles.
Gloves
To protect you from harmful liquids and avoid personal injury during
maintenance or service, put on appropriate protective gloves.
WARNING—Hazardous Substances
Solvents, mobile phases, samples, and reagents might contain toxic,
carcinogenic, mutagenic, infectious, or otherwise harmful substances.
The handling of these substances can pose health and safety risks.
• Be sure that you know the properties of all substances that you are
using. Avoid exposure to harmful substances. If you have any doubt
about a substance, handle the substance as if it is potentially
harmful.
• Wear personal protective equipment as required by the hazard and
follow good laboratory practice.
• Reduce the volume of substances to the minimum volume required
for sample analysis.
• Avoid handling of solvent reservoirs above head height.
• Do not operate the system in a potentially flammable environment.
• Avoid accumulation of harmful substances. Make sure that the
installation site is well ventilated.
• Dispose of hazardous waste in an environmentally safe manner that
is consistent with local regulations. Follow a regulated, approved
waste disposal program.
WARNING—Biohazard
Biohazardous material, for example microorganisms, cell cultures,
tissues, body fluids, and other biological agents can transmit infectious
diseases. To avoid infections with these agents:
• Assume that all biological substances are at least potentially
infectious.
• Wear personal protective equipment as required by the hazard and
follow good laboratory practice.
• Dispose of biohazardous waste in an environmentally safe manner
that is consistent with local regulations. Follow a regulated,
approved waste disposal program.
WARNING—Self-Ignition of Solvents
Solvents with a self-ignition temperature below 150 °C might ignite
when in contact with a hot surface (for example, due to leakage in the
chromatography system).
Avoid the use of these solvents.
WARNING—Hazardous Vapors
Mobile phases and samples might contain volatile or flammable
solvents. The handling of these substances can pose health and safety
risks.
• Avoid accumulation of these substances. Make sure that the
installation site is well ventilated.
• Avoid open flames and sparks.
• Do not operate the system in the presence of flammable gases or
fumes.
CAUTION—Allergic Reaction
Some capillaries in the system are made of MP35N™, a nickel/cobalt-
based alloy. Individuals with sensitivity to nickel/cobalt may show an
allergic reaction from skin contact.
WARNING—Safety Hazard
In case of emergency, disconnect the system modules from the power
line.
• Use only solvents and additives that are compatible with all parts in
the flow path.
VC‑pumps except The pumps are shipped with UHMW-PE piston seals. Using
VC‑P21 and VC‑P33 tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
In rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with high pH,
ammonium hydroxide containing mobile phases and
prolonged exposure.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals.
Consider the following:
• With the piston seals, also replace the seal wash seals
(part no. 6040.0306, containing 2 seals for use as both,
piston seal and seal wash seal).
• Install new pistons. Exchanging the piston seals but
keeping the pistons that were used with the UHMW-PE
piston seals may impair the seal life of the PTFE seals.
• Note that slightly increased abrasion may be observed
with these seals, depending on the application. Consider
testing the permeability of the static mixer regularly at
short intervals.
VC‑P21 and VC‑P33 The pumps are shipped with carbon-fiber filled PTFE piston
pumps seals. Slightly increased abrasion may be observed with
these seals, depending on the application. Consider testing
the permeability of the static mixer regularly at short
intervals.
VF‑pumps The pumps are shipped with UHMW-PE piston seals. Using
tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
In rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with high pH,
ammonium hydroxide containing mobile phases and
prolonged exposure.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals.
Consider the following:
• With the piston seals, also replace the seal wash seals
(part no. 6040.0306, containing 2 seals for use as both,
piston seal and seal wash seal).
• Install new pistons. Exchanging the piston seals but
keeping the pistons that were used with the UHMW-PE
piston seals may impair the seal life of the PTFE seals.
• In addition, note the following:
¨ Slightly increased abrasion may be observed with the
PTFE seals, depending on the application. Consider
testing the permeability of the static mixer regularly at
short intervals.
¨ The pump should not be operated with pressures
higher than 70 MPa.
¨ The pump is no longer biocompatible.
VH‑pumps The pump is shipped with UHMW-PE piston seals. Using
tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
In rare cases, a shortened lifetime of reversed-phase
(UHMW-PE) piston seals has been observed with high pH,
ammonium hydroxide containing mobile phases and
prolonged exposure.
NOTICE
If the system configuration includes a non-standard detector, for
example, a charged aerosol detector or refractive index detector, refer
to the Operating Manual for the detector for specific recommendations
regarding solvents and additives.
3 System Overview
This chapter introduces you to the system and the main components.
For a description of the solvent rack and system base, see further down
in this chapter. For a description of the other modules, refer to the
Operating Manuals for these modules.
TIP
If your system configuration is a non-standard system configuration
including modules other than those listed in the table, for example, a
fluorescence detector or a charged aerosol detector, refer to the
Operating Manual for the module.
The front side of the solvent rack is magnetic. Here, you can attach a
magnetic label holder, for example, to show a user-specific system
name. Labels are included in the system ship kit.
No. Description
3.2 Operation
The system is designed to be operated from a computer configured with
the Chromeleon Chromatography Data System (CDS). The Chromeleon
software provides complete instrument control, data acquisition, and
data management.
TIP The system can be operated also with other data systems, such as
Thermo ScientificTM XcaliburTM. In this case, installation of additional
software is required in addition to the data system software. For details,
contact the Thermo Fisher Scientific sales organization.
4 Unpacking
This chapter provides information for unpacking the system and informs
you about the scope of delivery.
4.1 Unpacking
Damaged Packaging, Defective on Arrival
Inspect the shipping container for signs of external damage and, after
unpacking, inspect the system for any signs of mechanical damage that
might have occurred during shipment.
If you suspect that the system may have been damaged during
shipment, immediately notify the incoming carrier and Thermo Fisher
Scientific about the damage. Shipping insurance will compensate for the
damage only if reported immediately.
3. Carefully remove the solvent rack and liquid reservoirs from the
shipping container.
6. If applicable
Remove any additional packing material.
TIP Keep the shipping container and all packing material. These items
will be needed if the system is transported to a new location or shipped.
• System base
• Solvent rack
• Solvent reservoirs
• Ship Kit
For details about the kit content, see Ship Kit (} page 139).
5 Installation
This chapter specifies the requirements for the installation site and
describes how to set up, install, and configure the system.
TIP
Before turning on the power to a Vanquish system module for the first
time, verify that the chromatography software is installed on the data
system computer. When the power is turned on, the required USB
drivers are automatically found and the Windows™ operating system
can detect the device.
5. Set up the system in the software. See Setting Up the System in the
Software (} page 80).
7. Recommended:
Perform Instrument Installation Qualification.
In the Chromeleon software, a wizard is available to guide you
through the qualification process. On the Chromeleon 7 Console:
Click Tools > Instrument Qualification > Installation Qualification.
Follow the instructions in the Instruments Installation Qualification
Operating Instructions. The manual provides information about the
required materials and detailed instructions.
NOTICE
If the system is operated with another data system, refer to the
documentation for the software that you are using and/or perform the
qualification manually. The Instruments Installation Qualification
Operating Instructions provide information about the parameters to be
adapted and the required settings.
• Make sure that the installation site has enough power outlets for all
devices in the system.
• Operate the system only under appropriate laboratory conditions
• For specifications, see Specifications (} page 133).
5.3.1 Workbench
The Vanquish system is designed to be placed on a workbench. If the
workbench shall carry a complete Vanquish system and possibly other
instruments, the bench must be capable to bear the weight of all devices
and instruments, including solvents.
For dimensions and weight, see Specifications (} page 133).
Workbench Requirements
Part Requirements
Workbench • Sturdy
• A height that ensures convenient access to the
interior of each device in the system.
• The bench top must be dry, clean, and resistant to
chemicals.
• If the system stack is to be stabilized with the
Bench Clamp kit, the workbench top must fulfill
one of the following requirements:
¨ Thickness: Between 15 and 50 mm
or
¨ Material: Soft enough for wood screws
Site for the workbench The workbench must stand in a secure and level
position that is free of vibrations.
Side Requirements
5.3.4 Condensation
Temperature
Temperature fluctuations can affect the performance of the system.
Avoid locations with significant changes in temperature and strong air
drafts. For example, do not place the system in the direct sunlight, near
heating or cooling sources, or under an air duct.
Humidity
The relative humidity of the operating environment is important for the
performance of the system. Operate the system in the specified
humidity range, with no condensation.
Ventilation
Make sure that the installation site is well ventilated at any time to avoid
potential health hazards and safety risks, which may be caused by
handling hazardous substances, volatiles or gases.
Electromagnetic interference
Electromagnetic sources of strong interference may affect the
performance of the system. In close proximity to the system, avoid
operating equipment generating strong electric or magnetic fields and
only operate certified laboratory equipment. Do not connect equipment
that may cause power disturbances to the same power network used to
supply the system.
1. Set up the system stack. For details, see System Stack Setup
(} page 47).
2. Connect the Waste Lines. For details, see Connecting the Waste
Lines (} page 55).
4. Connect the required signal cables. For details, see Signal Cable
Connections (} page 65).
No. Description
1 Solvent Rack
2 Detector
3 Autosampler
4 Pump
5 System Base
6 Column Compartment
The following table shows which stack stabilization kit you need for
which stack height. The heights are measured from the table surface to
the solvent rack railing. Due to production and installation tolerances,
the actual stack height can be up to 2 cm higher than the limit. This
tolerance is acceptable.
Up to 100 cm None
Example: Configuration with pump, autosampler, diode array
detector and charged aerosol detector
Up to 123 cm One of the following kits:
• Bench clamp kit
• Stack stabilizer kit
• IonBench with stack mounting kit
Up to 139 cm Bench clamp kit
139 cm and higher Build up the system in two separate stacks
NOTICE
Moving the Vanquish system stack after installation even over short
distances on the workbench can damage and impair the functionality of
the system base.
• Therefore, install the system stack at the final location.
• If you have to move the system after installation, see Moving the
System Stack (Unlocking/Locking) (} page 52).
Figure 4: Clip (left) and attaching the clips onto the system base (right)
4. Lift the pump by its carrying handles. Place the pump on the system
base approximately 5 cm before the end of the rails. Push the pump
towards the rear until the pump clicks into place.
No. Component
1 Carrying handles
2 Attachment screw (one on each carrying handle)
6. Remove the two carrying handles by sliding them off the rails
toward the rear.
Figure 6: Sliding off the carrying handle from the left rail
7. Repeat the previous three steps for the autosampler and detector by
placing the autosampler on the pump, and the detector on the
autosampler.
9. Measure the stack height from the table surface to the solvent rack
railing and check whether a stack stabilization is needed (see Safety
Guidelines for Setting Up the System Stack (} page 49)). If necessary,
stabilize the stack.
10. Connect the waste lines (see Connecting the Waste Lines
(} page 55)).
TIP
System modules on the system stack have protective caps on the rear of
the housing. Remove them after installing the system stack and before
attaching the column compartment to the system stack.
11. Attach the column compartment to the system stack (see Attaching
the Column Compartment to the System Stack (} page 62)).
12. Loosen or remove any shipping locks. Refer to the Operating Manual
for each module for information on shipping locks that need to be
removed.
3. Insert the combined locking tools into the lock port on the side of
the system base where the column compartment is installed.
If the column compartment is installed on the right side, insert the
combined locking tools into the right lock port as shown in the figure
below.
4. Insert one locking tool into the lock port on the other side of the
system base.
Figure 9: Inserting locking tools into the system base (system base top
view)
5. Unlock the system base by turning the locking tools as shown below.
In the unlocked position, the locking tools cannot be removed from
the system base.
7. Lock the system base (see Locking the System Base (} page 54)).
2. Lock the system base by turning the locking tools as shown below.
3. Remove the locking tools and store them in a safe place, for
example, in the system base drawer.
The Vanquish system has been designed for optimized and simple leak
liquid and waste handling:
• Leak liquid from the solvent rack, detector, autosampler, and pump
flows through pipes from the modules to the system base drain port.
• Seal wash and needle wash liquids from the pump and autosampler,
as well as condensate from the autosampler, are also guided to the
system base drain port.
• Leak liquid from the column compartment flows to the column
compartment drain port.
• Waste from the detector can also be routed through the system
base.
No. Description
• Connect a waste line to the system drain port (see Connecting the
Waste Line to the System Drain Port (Drain Tubing) (} page 58)).
• When the waste line is set up, test the drain system (see Testing the
Drain System (} page 59)).
• Install a detector waste line that connects the flow cell outlet
through the detector waste outlet to the waste. For installation
instructions, see Installing the Detector Waste Line (} page 60).
TIP
Make sure that you guide the detector waste line through the separate
outlet in the system base, which is located next to the drain port in the
system base.
No. Description
• If you have to cut tubing to length, use a tubing cutter. Make sure
that the cut is at a right angle to the length of the tubing and that
the end is free of nicks or burrs.
5.4.3.2 Connecting the Waste Line to the System Drain Port (Drain Tubing)
Parts required
1. Push an elbow connection piece into the system drain port as far as
it goes in to establish a self-sealing connection.
2. Push the drain tubing into the elbow connection piece with your
hand as far as it goes in to establish a self-sealing connection.
3. Route the free end of the drain tubing into a waste container.
4. Only if guiding the drain tubing toward the system's rear: Insert the
drain tubing into the clips and make sure that the drain tubing does
not form a siphon.
TIP
If routing the drain tubing of the column compartment also toward the
system's rear, you can insert the drain tubing of the column
compartment into the clips as well.
1. Position the waste container below the level of the system stack. To
prevent solvent vapors from evaporating back into the system, make
sure that the distance is sufficient.
3. To prevent liquid from flowing back into the system, the entire drain
tubing must remain below the drain port. Avoid the formation of
siphons.
4. To allow the waste liquid flowing off properly and avoid liquid
accumulation in the system base, the end of the waste line must not
be submerged in waste liquid (see figure below):
a) Shorten the tubing if required.
b) Observe the liquid level in the waste container and empty the
waste container as necessary.
2. Verify that the liquid leaves the system at the system drain port and
into the waste container.
TIP The detector waste line connects the flow cell outlet through the
detector waste outlet to the waste.
• For instructions on how to connect the waste line through the
detector waste outlet to the waste, follow the steps below.
• For instructions on how to connect the waste line from the flow cell
outlet, refer to the Operating Manual for your Vanquish detector.
1. Locate the outlet for the detector waste line in the system base (see
figure below).
3. Route the free end of the waste line into a waste container.
Parts required
• 4 T-slot nuts
• 2 crosspieces
• 4 screws
• 4 spacers
Tools required
Screwdriver, Torx T20
1. Align the screw holes of the two T-slot nuts, the spacers and the
crosspiece.
No. Description
1 Spacer
2 Groove
3 Screw
4 T-slot nut
5 Crosspiece
2. With the screws, fix the T-slot nuts and the spacers onto the
crosspiece.
The T-slot nuts must face each other, as shown above.
3. Slide the T-slot nuts of the mounting bar in the rail on the system
housing and push until stop.
The grooves on the mounting bars must face upwards. Make sure,
that the orientation of the crosspiece is as indicated on the picture.
No. Description
1 Rail
2 Mounting bar
See also
2 Connecting the Waste Lines (} page 55)
No. Description
TIP Thermo Fisher Scientific recommends using the USB ports only as
described above. If the USB ports are used for any other purpose,
Thermo Fisher Scientific cannot ensure proper functionality.
The flat, rectangular end of the cable is the "A" connector. The smaller,
hexagonal end is the "B" connector.
TIP
• The USB standard limits the USB cable length to 5 meters. Each USB
device can be separated from the computer or next USB hub by no
more than 5 meters.
• After connecting the USB cables and before turning on the power to
a device for the first time, verify that the chromatography software
is installed on the computer. When the power is turned on, the
required USB drivers are automatically found and the WindowsTM
operating system can detect the device.
No. Description
- USB connection:
1 From pump to detector
2 From autosampler to pump
3 From column compartment to detector
4 From detector to computer
Follow these steps 1. On the autosampler, pump, and column compartment, connect the
"B" connector of a USB cable to the USB port.
2. Connect the "A" connector of the USB cable from the pump to the
USB hub on the UV/VIS detector. Connect the "A" connector of the
USB cable from the other modules to the USB hub on the UV/VIS
detector or pump.
3. Connect the "B" connector of a USB cable to the USB port on the
UV/VIS detector, and connect the "A" connector of the USB cable to
a USB port (2.0 or higher) on the computer.
In order to control the modules that use a USB hub on another module,
the module that provides the hub must be turned on.
NOTICE
• Never use defective communication cables. If you suspect that a
cable is defective, replace the cable.
• To ensure trouble-free operation, use only the cables provided by
Thermo Fisher Scientific for connecting the system.
No. Description
Follow these steps 1. Connect one end of the system interlink cable to the System
Interlink port on the system base, and the other end to a System
Interlink port on the pump.
2. Connect one end of the system interlink cable to the free System
Interlink port on the pump, and the other end to a System Interlink
port on the autosampler.
3. Connect one end of the system interlink cable to the free System
Interlink port on the autosampler, and the other end to a System
Interlink port on the detector.
4. Connect one end of the system interlink cable to the free System
Interlink port on the detector, and the other end to a System
Interlink port on the column compartment.
NOTICE
Condensation in a device can damage the electronics.
• Before connecting the devices to the power line, be sure that no
condensation is present in the devices.
• If you suspect that condensation is present, allow the device to
warm up to room temperature slowly. Wait until the condensation is
completely gone before proceeding.
4. Insert the signal and power cables into the cable clips (see Guiding
Cables through the Cable Clips (} page 72)).
6. Lock the system base against moving (see Locking the System Base
(} page 54)).
1. Set up the flow connections for the pump. For details, refer to the
Operating Manual of the pump.
2. Set up the flow connections for the autosampler. For details, refer to
the Operating Manual of the autosampler.
3. Set up the flow connections for the column compartment except the
column. For details, refer to the Operating Manual of the column
compartment.
¨ Connect the pre-heater from port 2 on the autosampler
injection valve to the column outlet tubing using a Viper union.
The union replaces the column for the flushing procedure.
¨ Do not install the column yet!
¨ Connect a waste tubing to the outlet from the column
compartment, for example, post-column cooler (if applicable) or
the detector inlet capillary (depending on system configuration),
and column switching valve (if applicable) and route it to waste.
¨ Do not install a flow cell yet.
8. Flush the system without flow cell (see Flushing the System
(} page 82)).
10. Flush the system again with flow cell (see Flushing the System
(} page 82)).
See also
2 Connecting the Waste Lines (} page 55)
Each guide can hold up to three tubes or lines. In each module, push the
tube (or line) into the appropriate guide.
Figure 29: Tubing chase with tubing guides (left: view from inside, right:
view from top)
Tubing Brackets
Tubing brackets are available for holding the tubing in place. Slip the
bracket side onto the drain pipe.
• Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the product or additional or
spare capillaries and tubing as recommended by Thermo Fisher
Scientific.
• Install capillaries and fittings only at the positions for which they are
intended.
NOTICE
• Tighten or loosen Viper capillaries only with your fingers. Do not use
tools other than the knurl that comes with the capillary.
• To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries only when the system pressure is down to zero.
No. Description
1 Knurl
2 Capillary
3 Slot
TIP Note the slot in the knurl. You can easily remove the knurl from the
capillaries through this slot if space is limited.
TIP
Depending on the modules used in the system, you may also find other
fitting connections. Refer to the Operating Manual for the respective
module for information on how to handle these types of fittings.
NOTICE—System damage
Shipping locks must be properly loosened or removed before operation
to avoid damage to the system.
2. Lock the system base (see Locking the System Base (} page 54)).
1. Turn on each system module with its main power switch. Verify that
the autosampler is turned on before the pump flow is on and
pressure builds up.
3. Turn on the system power by pressing the lower push button on the
front left of the system base (system power button).
TIP Turn a module off with the main power switch, when instructed to
do so, for example, during maintenance. Pressing the system power
button will not be sufficient to turn off the power to the module
completely.
See also
2 Power On/Off Control (} page 86)
TIP In order to start the Chromeleon Instrument Controller, you must log
on under Windows with local administrator privileges, or as a member
of the Windows Chromeleon Operators user group.
2. Select the new instrument and click Add Module on the Edit menu
to add the first module.
4. On the General configuration page for each module, make sure that
simulation mode is deactivated and click Browse to select the
module address.
The Chromeleon software connects to the module and transfers the
settings from the device firmware to the software. Check and
change the settings on the other configuration pages if necessary.
For information about the settings, click Help or press the F1 key.
5. On the File menu, click Save Installation and then close the
Instrument Configuration Manager.
NOTICE
To avoid damage to other modules in the Vanquish system, verify the
following before you begin:
• The detector is not connected to the flow path.
• No column is installed in the column compartment.
2. Have the pump deliver the solvent used for purging for a short time.
2. Have the pump deliver the solvent for a short time. Observe the
pressure limit and other properties of the column.
NOTICE
• When the charged aerosol detector is connected in the system flow
path, flush the detector only when the detector gas flow is turned
on.
• Flush the detector with flow from the pump.
6 Operation
This chapter provides information for routine operation and for
shutdown.
NOTICE
Pay attention also to the following guidelines:
• To prevent damage resulting from leakage or from running the
pump dry, always set the lower pressure limit for the pump.
• If there is evidence of leakage in the device, turn off the pump flow
and remedy the situation immediately.
• If the pump flow is interrupted, act appropriately to protect the
components in the detector. For details, refer to the Operating
Manual for the detector.
• Always verify that the autosampler is turned on before the pump
flow is on and pressure builds up. If the autosampler is turned off,
for example, after a power failure, stop the pump flow and wait until
the pressure is down to zero before turning on the autosampler or
other modules again.
• When the power is on, the system power button is pressed in. When
the power is off, the system power button stands out.
• If the main power switch on a device is off, you cannot turn on the
device with the system power button.
• To turn off a device completely, you have to turn it off with the main
power switch on the device. Pressing the system power button will
not be sufficient to turn off the power to the device completely.
NOTICE
Flush the system flow path thoroughly before operating the device for
the first time:
• When you install devices or components to the system, always flush
them to waste before connecting them in the system flow path.
Follow the instructions in Flushing the System (} page 82).
• Some components of the system are filled with isopropanol when
the system is shipped from the manufacturing site. When operating
the system for the first time, use solvents that are miscible with
isopropanol. If they are not, use an appropriate intermediate
solvent.
• Verify that any air bubbles are completely flushed out of the system
flow path.
Refer to the Operating Manual for each system module for any steps
that may be required before initial operation.
• Check the liquid level in the solvent reservoirs. Verify that the
amount of solvent is sufficient for the analysis.
• Close the doors of all modules in the Vanquish system, if not already
done.
System Equilibration
System equilibration should include the following operations:
• Purging the pump (all channels, including those not used for the
application)
Solvent Reservoirs
• Before you change solvents, make sure that the new solvent is
miscible with the previous solvent. If the solvents are not miscible,
flocculation may occur.
Mix immiscible solvents with an intermediate solvent to replace
them step by step. For example, use isopropanol.
TIP
The commands and parameters that are available in the Command
window may vary, depending on the user level in the Command window
(Normal, Advanced, or Expert). To change the user level, right-click in
the commands list and select a different user level.
3. In the Select Channels to Monitor dialog box, select the signals that
you want to monitor.
TIP
Monitor baseline data is overwritten each time baseline monitoring is
started. If you want to save the data permanently, define the location in
the Monitor Baseline Save Preferences dialog box. For details, refer to
the Help.
TIP
If no eWorkflows are available, you can create a new eWorkflow as
described in the Help. As an alternative, use the Sequence Wizard to
create the sequence.
• Sequence Wizard
3. In the Work Area, click the instrument name, and then click Launch.
2. The Sequence Wizard guides you through the process. Complete the
steps in the wizard. Specify the number of samples and/or
standards, the instrument method, processing method, and report
template that you want to use. For detailed information about a
wizard page, click the Help icon.
After you finish the wizard, the sequence is displayed in the Data
view of the Console.
Charged aerosol Check that sufficient gas is available to continue gas flowing
detector through the detector. This is to prevent any build-up of
residue from solvents or analytes. Gas must be flowing when
pump flow is delivered to the detector.
UV/VIS detectors The lamp(s) in the detector can remain turned on.
Variable wavelength detector and VH-D10 diode array
detector only: The shutter can be moved to a closed position
for protection of the flow cell.
Fluorescence Turn off temperature control for the flow cell.
detector
• Make sure that the temperature of the column does not exceed
40 °C.
• When resuming operation, let the flow equilibrate and verify that
the operating parameters for the other system modules are set as
required before proceeding.
TIP With a Vanquish Core system that has been modified for using
normal-phase compatible solvents and additives, see the information
about the flushing liquid in Considerations with Normal-Phase
Compatible Solvents and Additives (} page 130).
System remains in the Flush the system, for Flush the system with
laboratory after example with methanol. several volumes of
shutdown 100% acetonitrile should methanol and water (50:50)
not be used. (for example, 1.0 mL/min
for 10 minutes with the
standard system) to prevent
salt buildup in the fluidics. If
the solvents in the device
are not miscible with water,
use an appropriate
intermediate solvent.
System shall be Flush the system with Flush the system first with
transported or shipped isopropanol. several volumes of
after shutdown methanol and water (50:50)
(for example, 1.0 mL/min
for 10 minutes with the
standard system) to prevent
salt buildup in the fluidics. If
the solvents in the device
are not miscible with water,
use an appropriate
intermediate solvent.
Afterward, flush the system
with isopropanol.
3. Turn off the system with the system power button on the system
base.
If one of the modules shall be removed from the system stack, turn
off all system modules with their main power switch. Pressing the
system power button will not be sufficient to turn off the power to
the devices completely.
If the system shall be transported or shipped after shutdown, follow the
instructions in Transporting or Shipping the System (} page 107).
7 Maintenance and
Service
This chapter gives general guidelines on maintenance and transport of
the system.
WARNING—High Voltage
High voltages are present inside the device that could cause an electric
shock.
Do not open the housing or remove protective panels unless specifically
instructed to do so in this manual.
CAUTION—Spraying Solvent
Solvents can spray when under high pressure.
• Stop the pump flow prior to opening the flow path.
• Wait until the system pressure is down to zero.
• When opening the flow path, wear appropriate protective
equipment.
CAUTION—Hot Surfaces
Surfaces inside the system may become hot during operation. Touching
hot parts might cause burns.
Allow hot surfaces to cool down before starting replacement or
maintenance procedures.
CAUTION—Hydrostatic Pressure
Solvent may spill when you open the flow path. This is due to hydrostatic
pressure in the system when the solvent reservoirs are located above
the pump outlet. Before you loosen a connection in the flow path:
• Turn off the pump flow and wait until the system pressure is down
to zero.
• Unscrew the caps of the solvent reservoirs and remove the solvent
lines together with the caps from the reservoirs.
• Empty the solvent lines. Refer to the Operating Manual for the
pump.
• Retighten the reservoir caps.
General Rules
Depot Repair
• If you need to return a system module for depot repair, follow the
instructions in section Transporting or Shipping the System
(} page 107).
See also
2 Consumables and Replacement Parts (} page 143)
NOTICE
Wipe up all liquids spilled onto the system immediately. If surfaces are
exposed for longer periods, these liquids can cause damage.
Decontamination
Decontamination is required, for example, when leakage or spillage has
occurred, or before service or transport of the system. Use a suitable
cleaning detergent or disinfectant to ensure that the treatment renders
the system safe to handle.
Parts required
• Purified water
NOTICE
Observe the following:
• Only use cleaning detergents that will not damage the surfaces of
the system.
• Never use sharp tools or brushes for cleaning any surfaces.
• Do not use sprays for cleaning.
• Prevent cleaning detergent from entering the flow path.
• Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the device.
Liquids can cause a short circuit when getting in contact with the
electronic components.
Preparations
1. Turn off the power to the system and disconnect the power cord
from the power source.
1. Wipe the surfaces clean with a clean, dry, soft, lint-free cloth or
wipe. If necessary, slightly dampen the cloth or wipe with a solution
of lukewarm water and a suitable cleaning detergent.
3. Wipe the cleaned surfaces with purified water to ensure that all
cleaning detergent residues have been removed.
Preparations 1. Turn off all modules with their main power switch.
2. Remove the solvent reservoirs and solvent lines from the solvent
rack.
5. Grasp the system by its sides and slightly lift the column
compartment up and out of the grooves on the mounting bars.
6. Lift the front part of the solvent rack and remove the solvent rack by
pulling it towards the front.
Removing the System 1. Slide the carrying handles that were installed when the system was
Modules shipped on the rails at the right and left of the module from the rear
as shown in the figure below.
The final position of the handles must match the foam spacers used
for transport (15 cm from the rear of the module to the center of
the handle).
No. Description
1 Carrying handles
2 Attachment screw (one on each carrying handle)
3. Lift the front part of the module by using the carrying handles.
Remove the module from the stack towards the front and carefully
place it on a stable surface.
CAUTION—Possible Contamination
Hazardous substances may have contaminated the device during
operation and may cause personal injury to service personnel.
• Decontaminate all parts of the device that you want to return for
repair.
• Fill in and sign the Health and Safety Form. Thermo Fisher Scientific
refuses to accept devices for repair if the Health and Safety Form is
missing, incompletely filled in, or unsigned.
8 Troubleshooting
This chapter is a guide to troubleshooting issues that may arise during
operation of the system.
Status Indicators
The status indicator LED bar (Light Emitting Diodes) on the front side of
each module and the STATUS LED on the keypad inside each module
provide quick visual feedback on the operational status of the module. If
the firmware detects a problem, the status indicators are red.
The data system logs information about all events related to instrument
operation for the current day in an Instrument Audit Trail. The
Instrument Audit Trail is named with the current date, using the format
yyyymmdd. For example, the Instrument Audit Trail for May 15, 2019, is
named 20190515.
The Instrument Audit Trails can be found on the ePanel Set (Audit
ePanel). In addition, Audit Trails for each instrument are available in the
Chromeleon 7 Console Data view, in the folder of the Instrument.
Messages in the Instrument Audit Trail are preceded by an icon. The icon
identifies the seriousness of the problem. For possible causes and
recommended remedial actions, see the Messages section in the
Operating Manual of the module that is in error state.
Locate the tables in the following sections for the type of symptom you
have, find the possible cause, and use the description of the solution to
help you solve your problem quickly.
TIP
For information on troubleshooting issues that may arise during
operation with the Vanquish charged aerosol detector and the Vanquish
fluorescence detector, refer to the respective Operating Manual for the
detector.
Extra column volume too Use short capillary connections with a suitable inner diameter. Use appropriate
large Viper capillaries.
Improper capillary Check fittings for correct placement. Loosen and retighten Viper fittings (refer to
connections the Viper Installation and Operation Guide). Use only the appropriate fittings (see
Connecting Fittings, Capillaries, and Tubing (} page 77)).
Peak broadening
Negative peaks
Improper reference The sample must not absorb in the range of the reference wavelength. Consider
wavelength using a method without reference wavelength.
Wrong polarization of If you are using analog output, check the analog output polarization.
analog output
Spikes
Air trapped in flow cell • Check all fluid connections for tightness, in particular in the intake path.
• Degas the mobile phase.
• Install a restrictor at the flow cell outlet, observing the pressure specification of
the respective flow cell and the general guidelines for the flow cell.
Note: Excessive backpressure will destroy flow cells!
Particles in flow cell • Flush the system (see Flushing the System (} page 82)). If you suspect that a
capillary is contaminated with particles, in particular between the column and the
flow cell, replace the capillary.
• Flush the flow cell with different flows if particles are trapped in flow cell.
• Check if the column is defective.
UV/VIS lamp old, Check if the lamp is properly seated. Replace the lamp (refer to Detector Operating
defective, or not Manual).
properly installed
Electrical interferences Avoid operation of equipment generating strong electric or magnetic fields in close
from other instruments proximity. Do not connect equipment that may cause power disturbances to the
same power network used to supply the system.
Column temperature • Install a restrictor at the flow cell outlet, observing the pressure specification of
significantly above the respective flow cell and the general guidelines for the flow cell.
boiling point of mobile Note: Excessive backpressure will destroy flow cells!
phase • Install a post-column cooler (refer to Column Compartment Operating Manual).
Environmental conditions
Unstable Make sure that the temperature and air humidity are constant. Use column
environmental thermostatting. Avoid draft. Use insulation on the capillary from the column
conditions compartment to the detector.
Capillary problems
Capillary connections Properly install the capillary connections. Tighten the capillary connections. Replace
not installed properly capillary.
or not tight, dead
volumes in capillary
connections
Pump problems
Piston seals in pump Replace the seals (refer to Pump Operating Manual).
leaking
Air trapped in pump Purge the pump (refer to Pump Operating Manual).
head
Baseline fluctuations See "Periodic baseline fluctuations" in Baseline (} page 119).
Pump pulsation or flow See "Pressure pulsation or inconstant pressure" in Pressure (} page 118).
fluctuations
Irreproducible gradient Change the gradient.
Check the solvent line filters for permeability. Replace the filter frits if necessary
(refer to Pump Operating Manual).
Improper wavelength, Choose a detection wavelength that is located near the apex of the spectrum.
e.g., in a UV spectrum
flank
Detector response time Select a longer response time.
too short
Detector response time Peaks are eventually not separated any more. Select a shorter response time.
too long
Autosampler draws air There is not enough amount of sample in the vial, the needle height setting is
from vial incorrect, or there are too many replicates.
Air trapped in Perform a wash cycle. Check that the injection valve is in Inject position and the
autosampler flow path flow components of the autosampler are flushed with solvent.
Draw speed too high Select a lower draw speed.
Gas content of sample Reduce the draw speed. Degas the sample if possible.
too high or saturated
Needle clogged or Replace the needle (refer to Split Sampler Operating Manual).
needle tip deformed
Piston seals in Replace the metering device head (refer to Split Sampler Operating Manual).
autosampler metering
device leaking
Injection valve or other Inspect the autosampler and the connections on the injection valve. Tighten leaking
autosampler connections. Replace leaking parts (refer to Split Sampler Operating Manual).
components leaking
Needle seat worn out Replace the needle seat (refer to Split Sampler Operating Manual).
8.2.4 Flow
No flow
Flow fluctuations
Pump inlet path Check the solvent inlet lines, solvent line filters, and solvent selectors/proportioning
clogged valve, for example, for signs of clogging.
Air trapped in pump • Purge the pump.
inlet path • Inspect the degasser for indications for leakage. Check in the chromatography
data system whether the degasser has reached the operating vacuum
(Degasser Vacuum = OK). Degassing may not work properly if the flow rate is too
high.
• Inspect the capillary and solvent line connections for leakage; tighten loose fitting
connections.
Inlet and/or outlet Clean the inlet check valve or outlet check valve, and replace the check valve if
check valve necessary (refer to Pump Operating Manual).
contaminated or
defective
Piston seals not sealing Replace the piston seals (refer to Pump Operating Manual).
tightly
8.2.5 Pressure
Pressure pulsation or inconstant pressure
Air trapped in the Purge the pump (refer to Pump Operating Manual).
system
Inlet and/or outlet Clean the inlet check valve or outlet check valve, and replace the check valve if
check valve necessary (refer to Pump Operating Manual).
contaminated or
defective
Pump inline filter (or Check the permeability of the inline filter (or static mixer). Replace the inline filter
static mixer) clogged (or static mixer) if necessary (refer to Pump Operating Manual).
Solvent line filter Check the solvent line filters for permeability. Replace the filter frits if necessary
clogged (refer to Pump Operating Manual).
Pump head Check the compression values and take remedial action (refer to Pump Operating
compression values not Manual).
reached
System clogged with Samples and insufficient solvent purity are frequently the source for particles
particles clogging the system. Follow the instructions for resolving clogging in the
autosampler (refer to Split Sampler Operating Manual).
High backpressure
One or more capillaries Check the capillaries in the system systematically from the detector to the pump.
in the system clogged Replace the capillaries as needed.
or damaged by bending
Pump inline filter (or Check the permeability of the inline filter (or static mixer). Replace the inline filter
static mixer) clogged (or static mixer) if necessary (refer to Pump Operating Manual).
Pre-heater clogged Rinse the pre-heater capillary in reverse direction, using an appropriate solvent. If
necessary, replace the pre-heater (refer to Column Compartment Operating
Manual).
Column switching valve Check the valve for indications of blockage. Replace the valve if necessary (refer to
clogged Column Compartment Operating Manual).
Post-column cooler Rinse the post-column cooler capillary in reverse direction using an appropriate
clogged or defective solvent. If necessary, replace the post-column cooler (refer to Column
Compartment Operating Manual).
Flow cell clogged Clean the flow cell. Replace the flow cell if necessary (refer to Detector Operating
Manual).
Contamination in the Flush the system with an appropriate solvent.
system
Low backpressure
Improper mobile phase Check premixed mobile phase. Perform an OQ gradient accuracy test. A solvent
composition selector/proportioning valve may be defective. Contact Technical Support.
Increasing flow rate Check flow rate settings. Perform OQ flow precision test.
Improper mobile phase Check premixed mobile phase. Perform an OQ proportioning test. A solvent
composition selector/proportioning valve may be defective. Contact Technical Support.
Decreasing flow rate Check capillary connections for leaks. Check flow rate settings.
Piston seal leakage Check the pump for piston seal leakage (refer to Pump Operating Manual).
8.2.7 Baseline
High baseline drift
System not sufficiently Flush the system until equilibration. Usually, a volume of 5–10 times the column
equilibrated volume will be sufficient.
Unstable 1. Make sure that the temperature and the humidity are constant. You can
environmental determine temperature fluctuations by recording the temperature channels.
conditions 2. Verify on the detector that the lamp house cover is properly installed and that the
front doors are closed. Avoid draft.
Flow cell contaminated Clean the flow cell. Replace the flow cell if necessary (refer to the Detector
Operating Manual).
UV/VIS lamp not stable Allow the lamp and optics to warm up for at least 60 minutes. If the lamp is old, you
may need to replace the detector lamp (refer to the Detector Operating Manual).
A new lamp may need burn-in time before the first analysis. For details, refer to the
Detector Operating Manual.
Absorption of eluent Absorbing additives may change the absorption spectrum, depending on the
changes when gradient solvent. Consider varying additive concentrations to level the drift.
is run
Pressure fluctuations • Purge the pump (refer to the Pump Operating Manual).
from pump • Perform a basic tightness test (see Basic Tightness Test (} page 124)), or test the
pump for leakage as described in the Pump Operating Manual.
• Clean the check valves (refer to the Pump Operating Manual).
Air trapped in the Purge the system as necessary (refer to the Pump Operating Manual).
system
UV/VIS lamp old or not Check that the lamp is correctly installed. Replace the lamp (refer to the Detector
properly installed Operating Manual).
Improper reference The sample must not absorb in the range of the reference wavelength. Consider
wavelength using a method without reference wavelength.
Detector response time Select a suitable response time.
too short
Improper wavelength Select an appropriate wavelength.
or optical bandwidth Select a higher bandwidth in particular with critical conditions (low absorption, few
light).
Flow cell not properly Check that the flow cell is installed properly (refer to the Detector Operating
installed Manual).
Detector defective Contact Technical Support.
Optical transmission of Check the lamp age and replace the lamp, if necessary. Use the Cell Transmission
the flow cell too low Check / Flow Cell Wash button on the ePanel for the module and follow the
(Vanquish system with instructions. For this procedure, a diagnostic cell for the VH‑D10 is required. If you
VH-D10 detector only) do not have a diagnostic cell, compare the cell transmission (light intensity) values
of the flow cell before and after a wash procedure and check whether the cell
transmission (light intensity) has increased. If it has not increased, perform an
extended wash cycle and check whether the cell transmission (light intensity) has
increased.
Pressure fluctuations • Purge the pump (refer to the Pump Operating Manual).
from pump • Perform a basic tightness test (see Basic Tightness Test (} page 124)), or test the
pump for leakage as described in the Pump Operating Manual.
• Clean the check valves (refer to the Pump Operating Manual).
Air trapped in the Purge the system (refer to the Pump Operating Manual).
system
Improper reference The sample must not absorb in the range of the reference wavelength. Consider
wavelength using a method without reference wavelength.
UV/VIS lamp defective Check that the lamp is correctly installed. Replace the lamp (refer to the Detector
or not properly Operating Manual).
installed
Piston calibration value Verify that the calibration value of the piston (imprinted on the rear of the piston)
mismatch corresponds to the related piston calibration value in the chromatography data
system. Adapt the value in the data system if necessary.
Ambient temperature Reduce the ambient temperature (for example, by ventilating the room).
too high or setpoint
lower than the
specified difference to
the ambient
temperature
Temperature control Check the setting for TempCtrl in the chromatography data system (TempCtrl=On
turned off turns temperature control on).
Ventilation grid Make sure that the ventilation grids of the column compartment are not obstructed
obstructed in any way.
Ambient air enters • Improper sealing of the column chamber may impair the heating and cooling
column chamber performance of the column compartment and lead to a considerable amount of
condensed water.
• When guiding capillaries through the column chamber cover seal, make sure that
the cover seals properly. If the chamber cover seal is damaged, replace the
chamber cover (refer to Column Compartment Operating Manual).
• Also, make sure that the front door is properly closed.
Column compartment Contact Technical Support.
defective
Pre-heater not The performance may be impaired if the thermal contact is imperfect, or if the pre-
installed correctly heater inlet and outlet are exchanged. Reinstall the pre-heater (refer to Column
Compartment Operating Manual).
Pre-heater defective Replace the pre-heater (refer to Column Compartment Operating Manual).
Post-column cooler not Verify that the post-column cooler screws are tightened and the post-column
installed correctly cooler is seated properly.
Ambient temperature Reduce the ambient temperature (for example, by ventilating the room), and/or
or humidity too high ambient humidity.
System power button Turn on the system with the power button on the front left of the system base.
turned off
Module power turned Turn on the module with its main power switch.
off
Module not connected Connect the power cord.
to the mains
Fuse has blown Replace the fuses (refer to Operating Manual of the module).
Error in electronic Contact Technical Support.
system
One or more modules cannot be turned off from the power button on the system base
System power Check the system interlink connections between the system base and the system
connection modules, see Signal Cable Connections (} page 65).
interrupted
No connection • Check the USB connections from the system to the computer, see Signal Cable
between module and Connections (} page 65). If the module uses a USB hub on another module, check
computer that the module that provides the hub is turned on.
• Check the USB port on the computer. USB standard 2.0 is required.
Repeated unexpected Install a USB Isolator. For ordering information, see Optional Accessories
disconnects in the (} page 141).
chromatography data
system
Automatic Testing
You can set up a schedule in Chromeleon for automatic testing.
Automatic testing is possible only for tests and functions that can run
unattended, without further user interaction, for example the Basic
Tightness Test.
Test Failed
To ensure that you run sequences only on fully operational modules, the
Chromeleon software prevents you from starting a sequence if a
diagnostic test has failed. If a test failed, take remedial action and repeat
the test. A user with special control privileges in the Chromeleon
software can override the failed test result. Refer to the Chromeleon
Help.
See also
2 Basic Tightness Test (} page 124)
2 Grating Motor Test (} page 124)
2 Intensity Test (} page 125)
2 Shutter Motor Test (} page 125)
The Basic Tightness Test checks the tightness of the pump and
autosampler.
2. If the test fails, check the Instrument Audit Trail for a message. For
remedial actions refer to the Troubleshooting section in the
Operating Manual for the module for which the test failed.
The Grating Motor Test checks the mechanical stability of the grating
drive.
The Intensity Test is a test to measure the intensity of the detector lamp
and the transmission performance ("intensity") of the optics. The test is
available for the diode array detector and for the variable wavelength
detector.
2. Turn on the detector lamps (one or both) that you want to test.
The Shutting Motor Test checks the mechanical stability of the shutter
motor and filter paddle.
9 Modifying the
System for Specific
Applications
When
To use very non-polar and harsh organic solvents, such as 100%
acetonitrile, n-hexane, tetrahydrofuran, and similar solvents.
Parts required
• Normal phase (NP) kit, including the parts and instructions for
modifying the pump and autosampler and for replacing the detector
waste line.
• Tubing cutter
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
3. Together with the cap, remove the seal wash line from the reservoir.
4. To empty the seal wash line, press down the lever of the seal wash
pump. Release the lever when the line is empty.
2. As applicable:
a) Replace the pump pistons. Refer to the Operating Manual for
the pump.
b) Replace the flow cell with an NP-compatible flow cell. Refer to
the Operating Manual for the detector.
c) Replace the column switching valve with an NP-compatible
valve. Refer to the Operating Manual for the column
compartment.
• Long-term shutdown
• The seal wash liquid is miscible with the solvent that is used in the
pump.
• The seal wash liquid is compatible with the seal wash tubing.
Situation Action
10 Specifications
This chapter provides important system specifications.
Type Specification
Materials in the flow path For materials in the flow path, refer to the Operating Manuals for the
modules in your system.
For solvent and additive information, see Solvent and Additive
Information (} page 25).
Biocompatibility Core system in the standard Horizon/Flex system in the
configuration: No standard configuration: Yes
Control System base power button
Chromeleon 7
The system can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Good Laboratory Practice (GLP) Predictive Performance functions for scheduling maintenance
features procedures based on the actual operating and usage conditions of the
system.
All system parameters logged in the Chromeleon Audit Trail.
Safety features Leak detection and safe leak handling
Range of use Indoor use only
Ambient operating temperature 5 °C - 35 °C
Ambient storage temperature -20 °C - 45 °C
Ambient operating humidity 20% - 80% relative humidity, non-condensing
Ambient storage humidity Maximum 60% relative humidity, non-condensing
Operating altitude Maximum 2000 m above sea level
Pollution degree 2
Emission sound pressure level System with one column compartment, without Charger:
Typically approx. 55 dB(A), max. approx. 75 dB(A)
System dimensions System with one column compartment, without Charger:
(height x width x depth) 82 cm x 55 cm x 62 cm
System weight Vanquish Core Vanquish Flex: Vanquish Horizon:
With one column compartment, Approx. 80 kg Approx. 90 kg Approx. 110 kg
without Charger
Type Specification
11 Accessories,
Consumables and
Replacement Parts
This chapter describes the standard accessories that are shipped with
the system and the accessories that are available as an option. This
chapter also provides information for reordering consumables and
replacement parts.
Item Quantity in
shipment
Item Quantity in
shipment
Normal-Phase Kits
Miscellaneous
Drainage Kit
12 Appendix
This chapter provides additional information about compliance.
RoHS Compliance
This product complies with the RoHS (Restrictions of Hazardous
Substances) directives:
• European RoHS Directive
Directive on the restriction of the use of certain hazardous
substances in electrical and electronic equipment
The CE mark on the device indicates that the product is compliant
with the directive.
Logo Description
The green logo marks items that do not contain the hazardous
substances identified by the regulations.
Index
A connectors ....................................................... 66
consumables.................................................. 143
additives ........................................................... 25
information ................................................. 25 D
use............................................................... 89
algae ................................................................. 89 decontamination ........................................... 105
audit trail........................................................ 112 delivery ............................................................ 37
Diagnostics..................................................... 123
B digital I/O ......................................................... 66
drainage........................................................... 55
buffer use ......................................................... 89
buffers .............................................................. 25 E
concentration.............................................. 25
information ................................................. 25 EAC mark ....................................................... 148
use............................................................... 89 ePanel Set ........................................................ 92
equilibration .................................................... 88
C eWorkflow ....................................................... 94
cable connections............................................. 65 F
capillaries
guide ........................................................... 75 FCC................................................................. 149
install........................................................... 77 flow connections
Viper............................................................ 77 drain system ............................................... 55
CE mark .......................................................... 148 overview ..................................................... 74
charged aerosol detector ........................... 30, 41 waste lines.................................................. 55
chloride concentration ..................................... 27 fluorescence detector................................ 30, 41
Chromeleon................................................ 33, 91 flushing ............................................................ 82
audit trail................................................... 112
instrument configuration ............................ 80 G
Smart Shutdown.......................................... 96
gloves............................................................... 21
Smart Standby............................................. 96
guide hole ........................................................ 75
Smart Startup .............................................. 88
guidelines
system setup ............................................... 80
installation.................................................. 40
Chromeleon 7................................................... 91
maintenance............................................. 101
automated module control ......................... 94
operation .................................................... 85
client............................................................ 91
ePanel Set.................................................... 92
eWorkflow................................................... 94 I
Instrument Configuration Manager ............ 80 installation
Instrument Controller ........................... 80, 91 capillaries and tubing ................................. 75
module control............................................ 92 flow connections ........................................ 73
monitor baseline ......................................... 93 flushing ....................................................... 82
Sequence Wizard ........................................ 94 hardware setup .......................................... 47
cleaning .......................................................... 105 power connections ..................................... 71
condensation.............................................. 45, 71 safety guidelines......................................... 40
signal cable connections............................. 65
site requirements........................................ 43 P
software setup ............................................ 80
system ......................................................... 41 peak shape..................................................... 113
system stack................................................ 50 pH range .......................................................... 27
waste lines................................................... 55 power connections .......................................... 71
Instrument Audit Trail .................................... 112 power considerations ...................................... 44
intended use..................................................... 18 power cord ...................................................... 44
Interlink ...................................................... 32, 66 power on/off ................................................... 86
protective clothing........................................... 21
L
R
leak liquid ......................................................... 55
LED bar ........................................................... 112 regulatory compliance................................... 148
locking the system............................................ 54 replacement parts ......................................... 143
long-term shutdown ........................................ 97 requirements
long-term shutdown (NP)............................... 130 power cord ................................................. 44
system clearance ........................................ 44
M workbench.................................................. 43
RoHS mark ..................................................... 148
maintenance ............................................ 99, 104
cleaning ..................................................... 105 S
decontamination....................................... 105
general rules.............................................. 103 safety glasses ................................................... 21
interval ...................................................... 104 safety guidelines
interval (NP) .............................................. 131 case of emergency...................................... 24
introduction .............................................. 100 general........................................................ 19
maintenance (NP) ..................................... 131 general hazards .......................................... 22
safety guidelines ....................................... 101 installation.................................................. 40
maintenance............................................. 101
operation .................................................... 85
N
power cord ................................................. 44
normal-phase compatible solvents and additives protective equipment................................. 21
.................................................................. 128 qualification of personnel........................... 19
servicing.................................................... 101
O safety symbols ................................................. 16
scope of delivery.............................................. 37
operating conditions ........................................ 45
seal wash system
electromagnetic interference ..................... 46
seal wash liquid (NP) ................................ 130
humidity ...................................................... 45
Sequence Wizard ............................................. 94
mechanical shocks ...................................... 46
service.............................................................. 99
temperature................................................ 45
setup
ventilation ................................................... 46
hardware .................................................... 47
vibration ...................................................... 46
software ..................................................... 80
operation.......................................................... 33
ship kit ........................................................... 139
Chromeleon................................................. 91
shipping ......................................................... 107
interrupt...................................................... 96
short-term shutdown ...................................... 96
power on/off ............................................... 86
shutdown
preparation ................................................. 87
long-term.................................................... 97
safety guidelines ......................................... 85
long-term (NP).......................................... 130
overview (functional) ....................................... 29
short-term .................................................. 96
T
transport ........................................................ 107
troubleshoot system ...................................... 113