DI-710 Series

Low-Cost, Portable, USB/


Ethernet Data Logger
System

16 Single-ended, 8
Differential Analog Inputs

Stand-alone Data Logger


Option Allows Data to be
Saved to SD

Multiple Hardware Options


Front of DI-710
The DATAQ Instruments DI-710 Series
of products is a family of instruments Features
for general purpose and stand-alone data
logger data acquisition applications. Stand-alone Data Logger File Protection
Options include interface type, input Operation When powered down unexpectedly, the
voltage range, and PC-connected or stand- Use a Secure Digital Card to record and DI-710 Stand-alone model retains all data
alone data logger operation. Interface store data—up to 2GB. A FIFO memory saved to its memory card.
configuration allows the DI-710 to record
options are USB or Ethernet. Gain ranges Easy to Connect & Use
have selectable factors per channel of continuously using a circular buffer
Installs in seconds. Simply connect to your
1, 2, 4, and 8, or 1, 10, 100, and 1000. approach. A push button allows manual
computer's USB port or to an Ethernet
Instruments with the stand-alone data start/stop control over the recording
port. Connect power, then connect your
logger option feature a built-in multimedia process. A multi color LED shows
signals to the provided screw terminal
socket that accepts standard Secure Digital instrument status (Record, Standby, Busy,
blocks (16 ports each). Stand-alone data
(SD) memories to which acquired data Error).
loggers just require a Secure Digital Card
may be stored without a connected PC. Wide Signal Measurement Range and power.
SD memories are the same commonly Suitable for use with all types of
available mass storage devices used
Includes Software
transducers, the 16-channel single-ended,
Be up and running minutes out of the
with digital cameras and MP3 players. 8-channel differential DI-710 features a
box with WINDAQ software. WINDAQ
Memories ranging in size from 16 MB to per-channel measurement range of ±10 V
Recording and Playback software is
2 GB are supported. Instruments without over four gain ranges. This allows you to
included free with the purchase of every
this option must remain tethered to a simultaneously measure a wide range of
DI-710 instrument. Record at rates up
PC’s USB or Ethernet port during data signals with ease.
to 1000 Hz using WINDAQ Acquisition
acquisition and use the PC’s own program
and memory to store acquired data.
Flexible Programmability software. Purchase an option Unlock
Channel-by-channel software selection Code to record data as high as 4,800 Hz
All DI-710 Series products feature 14-bit of gain and single-ended/differential throughput to pc. Use Playback software
measurement resolution, sixteen analog operation. (WWB) to review, measure, and analyze
input channels that may be configured for your data during or after a recording
single ended or differential operation per High Throughput Rate
session.
channel, and an 8-bit digital bi-directional Supports sample throughput rates up to
4800 samples/sec to PC (depending on DATAQ Instruments Hardware Manager
port. PC-connected instruments stream Software allows you to effectively manage
data to the PC from as low as 0.048 Hz host computer speed) or up to 14,400
samples/sec to memory card (stand-alone and run multiple units installed to your PC,
up to as high as 4,800 Hz throughput rate. your network, or even over the Internet. It
Stand-alone data loggers store to their SD data loggers).
includes configuration software for stand-
memory from as low as 0.0017 Hz up to High Resolution alone data loggers allowing a complete
as high as 14,400 samples per second. All 14-bit resolution analog to digital data acquisition configuration to be
DI-710 instruments are supplied with two conversion provides a responsive designed and downloaded from any local
removable, 16-position screw terminal instrument capable of registering changes or remote PC. Upload software allows you
access connectors. as small as one part in 8.192 (±0.012% of to read data stored to an SD card over the
the full scale measurement range). DI-710’s Ethernet interface.
DATAQ Instruments, Inc. • 241 Springside Drive • Akron, Ohio 44333 • Tel: 330-668-1444 • Email: [email protected] • www.dataq.com
DI-710 Block Diagram

330-668-1444 2 www.dataq.com
Ways to transfer SD data files

Rear Panel

Power Jack “Mode” LED


May be powered by the Tri-color LED
provided AC adaptor, or indicates instrument
from any 9-36 VDC source. status: Standby, Re-
Consumes 2 watts. cording, Error.

Interface “Control” Pushbutton Removable Storage Slot


May be configured with an Allows manual start/stop local con- Accommodates standard and readily available multi-media
Ethernet or USB interface. trol over the recording process and memory cards for mass storage. These are the same memories
instrument configuration. used by consumer electronic devices like MP3 players and
digital cameras. Accepts memory sizes from 16 MB to 2 GB.
www.dataq.com 3 330-668-1444
Deployment Methods for Ethernet Models
Six Deployment Methods
Ethernet Models add a new dimension to Data Acquisition applications allowing data access over a LAN to a PC on the other side of
your facility or over the internet to a PC on the other side of the world. For more information view our application note on the internet
at: http://www.dataq.com/applicat/articles/data_logger_ethernet02.htm.

Deployment #1 - One PC directly connected to one DI-710 with a crossover cable.

Deployment #2 - One or more PCs, one or more DI-710s on a Dedicated Network.

Deployment #3 - One or more PCs, one or more DI-710s on a LAN.

330-668-1444 4 www.dataq.com
Deployment Methods for Ethernet Models
Deployment #4 - One or more PCs, one or more DI-710s on a Wireless Network.

Deployment #5 - One or more PCs, one or more DI-710s remotely to different LAN subnets.

Deployment #6 - One or more PCs, one or more DI-710s remotely over the Internet.

View the Ethernet Resource Page for application notes and demos on Ethernet Data Acquisition
with DI-71X products. Go to http://www.dataq.com/applicat/ethernet.htm for more information.
www.dataq.com 5 330-668-1444
DI-710 Specifications
Analog Inputs Controls (Stand-alone models)
Number of Channels: 16 Single push-button: Manual control Record and Standby
Channel Configuration: 16 single-ended; 8 differential; program- Calibration
mable per channel Calibration cycle: One year
Measurement range, Accuracy, and Resolution Digital I/O
Bits: 8 bidirectional bits
Gain Range Accuracy* Resolution
Configuration: Each bit is programmable as Input or Output
PGL Models: 1 ±10V ±.05%FSR ±50μV ±1.22mV
Output voltage levels: Min. “1” 3V @ 2.5mA sourcing; Max. “0” 0.4V @
10 ±1V ±.05%FSR ±50μV ±122µV 2.5mA sinking
100 ±100mV ±.05%FSR ±50μV ±12.2µV Output current: Max. source, -2.5 mA; Max. sink, 2.5mA
1000 ±10mV ±.05%FSR ±50μV ±1.22µV Input voltage levels: Min. required “1” 2V; Max allowed “0” 0.8V

PGH Models 1 ±10V ±.05%FSR ±50μV ±1.22mV Ethernet Interface


Type: 10/100Base-T
2 ±5V ±.05%FSR ±50μV ±610µV
Connector: RJ-45
4 ±2.5V ±.05%FSR ±50μV ±305µV Protocol: TCP/IP
8 ±1.25V ±.05%FSR ±50μV ±153µV Server Type: DHCP
*Test Conditions: Single channel, 100S/s, Averaging mode. Removable Memory (Stand-alone models)
Input Impedance, single-ended: 1MΩ Type: Standard SD (not SDHD or SDxC). Minimum
Input impedance, differential: 1MΩ each input to common speed of 13x. For sample rates above 2kHz SD
speed should be 133x or higher.
Input bias current: 10µA for a 10V input, single channel
Capacity: 16 MB to 2 GB
Input offset voltage: Auto-zero
Input offset current: 2nA (single channel) Real Time Clock (Stand-alone models)
Max. normal mode voltage: 30V DC or peak AC Type: Date, hour, minute, second
Max. common mode voltage: 30V DC or peak AC Resolution: 1 second
Common mode rejection: 80db, gain=1, 1KΩ unbalance Accuracy: 20 ppm
Channel-to-channel crosstalk Indicators
rejection: -75db @ 100Ω unbalance Stand-alone models: Three-color LED indicating Record, Standby, and
Gain temperature coefficient: 50 ppm/°C Error conditions
Offset temperature coefficient: 0.25µV/°C Standard models: Power LED
Digital filtering: Standard: Conditional over-sampling Transfer Rate to PC
Stand-alone: None
Real Time: up to 4,800 samples per second
A/D Characteristics From Memory Card: up to 2,400 samples per second (Ethernet only)
Type: Successive approximation
General
Resolution: 14-bit
Panel indicators: Mode LED
Monotonicity: ±2 LSB
Panel-accessible Power: 5 VDC @ 10 mA max
Conversion Time: 69µs
Panel Controls: Control push button (Stand-alone models)
Scanning Characteristics Panel Slots: Accepts SD-type flash memory
Max. throughput sample rate:* Standard: 4,800 Hz (requires Unlock Code)
Input connectors: Two, removable sixteen position terminal blocks
Stand-alone: 14,400 Hz**
Operating Environment: 0°C to 70°C
*When acquiring more than one channel at a gain of 100 max throughput is 7200 Hz;
When acquiring more than one channel at a gain of 1000 max throughput is 900 Hz. Enclosure: Aluminum base with steel wrap-around. Aluminum
**Dependent on SD card used. Low speed SD cards can sample up to 2,000 end-panels with plastic bezels.
Hz; High speed cards can sample up to 14,400 Hz. Some high speed cards Dimensions: 57/16 "D × 41/8 "W × 11/2 "H (13.81D × 10.48W × 3.81H cm.)
cannot sample as high at 14,400 Hz but their capability can only be determined Weight: 14 oz.
by trial and error (Model 101014-2G has been tested and approved).
Power Requirements: USB (stand-alone): 9 to 36 VDC, 2 watts max
Min. throughput sample rate: Standard: 0.0034 Hz USB (non stand-alone): 5V @ < 2 watts max
Stand-alone: 0.0017 Hz (supplied by USB port)
Max. scan list size: 17 entries Ethernet: 9 to 36 VDC, 2.5 watts max
Sample buffer size: 2kb

330-668-1444 6 www.dataq.com
Ordering Guide
Description Order No. Description Order No.
DI-710-UH USB Instrument DI-710-EH Ethernet Instrument
Low cost, portable, USB data logger with programmable DI-710-UH Low cost, portable, Ethernet data logger with programma- DI-710-EH
gain ranges of 1, 2, 4, and 8. ble gain ranges of 1, 2, 4, and 8. Includes Power Supply.

DI-710-UL USB Instrument DI-710-EL Ethernet Instrument


Same as DI-710-UH but with programmable gain ranges DI-710-UL Same as DI-710-EH but with programmable gain ranges DI-710-EL
of 1, 10, 100, and 1000. of 1, 10, 100, and 1000. Includes Power Supply.

DI-710-UHS USB Stand-alone Instrument DI-710-EHS Ethernet Stand-alone Instrument


Same as DI-710-UH but with stand-alone capability. DI-710-UHS Same as DI-710-EH but with stand-alone capability. DI-710-EHS
Includes Power Supply. Includes Power Supply.

DI-710-ULS USB Stand-alone Instrument DI-710-ELS Ethernet Stand-alone Instrument


Same as DI-710-UL but with stand-alone capability. DI-710-ULS Same as DI-710-EL but with stand-alone capability. DI-710-ELS
Includes Power Supply. Includes Power Supply.

SD Card Reader WINDAQ/HS-710


WINDAQ/
Reads Secure Digital (SD) and MultiMedia Card (MMC) 101014-CR High-Speed Unlock Code for WINDAQ Acquisition soft-
HS-710
Memories. For use with Stand-alone models. ware allowing higher sample rates (4.8kHz).

SD Card Power Connector


Low Speed 2GB SD Card (up to 2 kHz sample rate) 101014-2GS Power Connector (for customers who want to provide 100952
High Speed 2GB SD Card (up to 14.4 kHz sample rate) 101014-2G their own power).

DATAQ Instruments, Inc.


241 Springside Drive
Akron, Ohio 44333
Phone: 330-668-1444
Fax: 330-666-5434

Data Acquisition Product Links


(click on text to jump to page)
Data Acquisition | Data Logger

DATAQ, the DATAQ logo and WinDaq are registered trademarks of DATAQ Instruments, Inc. All rights reserved. Copyright © 2014 DATAQ Instruments, Inc.
The information on this data sheet is subject to change without notice. DI-8B modules are designed and manufactured by Dataforth Corporation