Exploring TotalChrom Software For GC 1684694541
Exploring TotalChrom Software For GC 1684694541
Exploring TotalChrom Software For GC 1684694541
1. Launch the
“TotalChrom”
Software
2. Click on “Run”
3. Select “Take
Control…”
4. Select “Clarus 580”
5. Click “OK”
Running GC
• The GC touch screen panel should
show that the GC is in external control.
• Once in external control:
• Build Method
• Build Sequence
Build Method
Build Method
1. Click on “Method”
button
Build Method
2. Select “Create
new method”
3. Click on “OK”
Build Method
4. Type in a description
5. Click “Next”
Click “Next”
e x t”
k “N
Clic
Click
Click “Next”
“Finish”
Build Method
1. Click on “Sequence”
button
Build Sequence
1. Click on “Setup”
button
Running Sequence
1. Select a sequence
(new window will
open / not shown)
2. Click OK
Running Sequence
• The GC will take some
time to upload the
sequence information,
equilibrate, and do pre-
run operation
• You will know that the
sequence upload
properly because the
Instrument-Hands On
and Instrument-Modify
will no longer be grayed
out
• Once the everything is
ready the lettering here
should be in green.
• I/F will read Ready
3. Click on Run
4. Start Run
Chromlab TotalChrom Operational Notes
TotalChrom™ Tips
Introduction
T
he information and tips in this document are intended to provide some
helpful ways for the user of the TotalChrom™ chromatography software to
generate and optimize their methods, reports and sequences and in review
of their data.
Pictured below is the TotalChrom Navigator screen. These system functions are
accessed by a single click on the screen buttons. These functions are labeled with
blue group names and can also be accessed from the drop-down menus of the
same name.
For example the Build group for the function Method, Graphic Edit, Sequence
and Report Format are also available in the Build drop-down menu.
2
Chromlab TotalChrom Operational Notes
Tip #1- Create separate folders for your TotalChrom data, methods & reports and
sequences on the C:\ drive. This will allow for shorter directory paths for easier
viewing in the TC sequence screen and in report headers.
This structure will also prevent possible deletion of system files that are stored in
the PenExe folder. It is NOT a good idea to use the TotalChrom PenExe program
folder for this reason.
Tip #2 - Always save your Method and Report Format files together in the
same folder so that the TotalChrom software can always find your
Report Format files.
C:\TotalChrom Data\2008
C:\TotalChrom Data\2008\June
C:\GC 1 – QC Lab
C:\Your Analysis or Product
Tip #3 - The more you segment your data into separate folders, the easier
it will be to retrieve your data for review and reprocessing or to
move you data for archival purposes.
3
Chromlab TotalChrom Operational Notes
Note: You may need to use Save As to save your first file entered
into a newly created project directory before your files will be
automatically saved in your new project directory instead of in the
TotalChrom default paths.
4
Chromlab TotalChrom Operational Notes
5
Chromlab TotalChrom Operational Notes
6
Chromlab TotalChrom Operational Notes
Use Windows™ Explorer to create this folder. These backup files will come
in very useful if a method file is accidentally changed, lost or is corrupted.
C:\TotalChrom Backup
Step 2. When you are satisfied that your method, report format
and sequence files are finished copy them to your
backup directory. You can use the Windows Save As
function as follows:
Tip #7 - It is also highly recommended that you backup all important files
off-line in case of hard drive failure. Use removable storage
media such as a flash drive, CD or floppy disk.
7
Chromlab TotalChrom Operational Notes
Important Note: It is recommended that you do not change the paths for
methods and report formats or sequences once they are
setup.
8
Chromlab TotalChrom Operational Notes
As an example, you can add the year to your Raw and Result file data
paths. Example: C:\2008 TC Data
9
Chromlab TotalChrom Operational Notes
Chapter 2
Tip #9 - Use the Review function in the Method Editor File drop-down menu.
This will step you through each page of the method. Don’t forget to
click on each of the tabs to review all of the method parameters.
10
Chromlab TotalChrom Operational Notes
11
Chromlab TotalChrom Operational Notes
If the Sampling Rate is set too slow, the peak data at the crest of the
peaks of interest will not be properly collected resulting in accurate
and low area counts.
Example of Sampling Rate set too low. Note that the data at the crest of the
peak was not collected!
Important Note: If the Sampling Rate is too low (slow) the peak
crest data points may be missed causing
inaccurate peak area (low) measurements.
On the other hand, setting the Sampling Rate too high (fast) and collecting
too many data points will result in .Raw data files that are overly large and
can result in peak splitting.
Tip #12 – Use the Graphic Method Editor to view (and count) the actual raw
data points in your data file. See Chapter 3, page 31.
12
Chromlab TotalChrom Operational Notes
Setting the Sampling Rate too high or over sampling can result in your method
being over sensitive to detector noise that may result in peak splitting during
integration.
Example of the Sampling Rate set too high (fast). Note that this peak is
split and will be identified as two separate peaks in the report.
Important Note: If the Sampling Rate is too high (fast) your data
file size will be overly large and peak splitting
may occur.
13
Chromlab TotalChrom Operational Notes
14
Chromlab TotalChrom Operational Notes
b. Use Range 1 for most capillary column installations, use Range 20 for
packed columns.
15
Chromlab TotalChrom Operational Notes
b. Again select the SPL1 (or SPL 2) function and set the Time (min) to
2.00 minutes and the Value back to a standby split flow rate of 15 – 20
mL/min and click on the Add button.
Method Editor - Instrument Control – Instrument Timed Events – Gas Saving mode.
16
Chromlab TotalChrom Operational Notes
Method Editor – Instrument – Control Options – Split Control – Gas saver mode.
Important Note: This split flow saving mode may cause baseline upsets if
you are using a flow sensitive detector such as an ECD or TCD detector.
17
Chromlab TotalChrom Operational Notes
a. Higher numbers for the Noise threshold will move the peak start and
end points in toward the crest of the peak.
b. Lower numbers will move the peak start and end points away from the
crest of the peak.
18
Chromlab TotalChrom Operational Notes
19
Chromlab TotalChrom Operational Notes
d. Select the Enable Peak Detection (+P) function with the Event browse
button. Set the Time to after the crest of the solvent peak and click on
the Add button.
20
Chromlab TotalChrom Operational Notes
c. Set the Scaling Type to Absolute scaling, and set the Full scale
(mV) for your desired plot display. This will fix the vertical axis of the
chromatogram replot so that you can visually compare chromatograms.
Method Editor – Instrument – Control Options – Split Control – Gas saver mode.
21
Chromlab TotalChrom Operational Notes
22
Chromlab TotalChrom Operational Notes
b. Turn off the Relative window to remove this parameter from the
component ID calculations to simplify the ID window calculation.
Tip #22 – Check the “Find tallest peak in window” function so that the
ID software will find the tallest peak in the window as apposed
to the default which is the peak closest to the middle of the
window.
23
Chromlab TotalChrom Operational Notes
Tip #23 – To prevent results calculated by using a curve fit from giving
negative values, choose Force for the Origin Treatment.
24
Chromlab TotalChrom Operational Notes
Tip #24 - It is a good practice to use the same file name for the Report file
as was used for the Method file and to store it in the same
directory as the Method file.
Note: If TotalChrom can not find the method Report file it will use the Default
Report.
b. Click on the Other menu in the Method Editor window and choose the
Report Format Editor function.
c. The Report Format displayed for a new method is the Default Report.
25
Chromlab TotalChrom Operational Notes
c. Click on the Quick paths browse button and select the TotalChrom
Examples directory path: C:\PenExe\TcWS\Ver6.3\Examples.
26
Chromlab TotalChrom Operational Notes
27
Chromlab TotalChrom Operational Notes
28
Chromlab TotalChrom Operational Notes
29
Chromlab TotalChrom Operational Notes
30
Chromlab TotalChrom Operational Notes
Chapter 3
Chapter 3 – Graphic Method Editor Tips
Tip #31 - One way to access the Graphic Editor is from the Method Editor
screen. Using this technique, you will not need to re-select your
method file name in the Graphic Editor.
Click on the Other drop-down menu in the Method Editor and
select Graph Editor.
31
Chromlab TotalChrom Operational Notes
32
Chromlab TotalChrom Operational Notes
b. Using the mouse, rubber box the narrowest peak of interest at its
baseline. DO NOT include peak pre/post tail area.
33
Chromlab TotalChrom Operational Notes
34
Chromlab TotalChrom Operational Notes
35
Chromlab TotalChrom Operational Notes
1. Manual Calibration.
a. Click on the Calibration drop-down menu and select
Calibrate.
Calibration – Calibrate
Important Note: Check to be sure that the data file currently being
viewed in the Graphic Method Editor is a calibration
standard run before using the Calibrate function.
36
Chromlab TotalChrom Operational Notes
Chapter 4
37
Chromlab TotalChrom Operational Notes
c. Enter the Starting Row, Ending Row, Directory and Edit the Base
file name as necessary to give your data files unique names.
38
Chromlab TotalChrom Operational Notes
39
Chromlab TotalChrom Operational Notes
40
Chromlab TotalChrom Operational Notes
41
Chromlab TotalChrom Operational Notes
Chapter 5
1. Manual Integration.
Use manual integration to re-draw the baseline under a
peak of interest.
a. Click the Process drop-down menu.
b. Select Manual Integration.
c. Place the + cursor on the baseline, in front of the peak of interest, click
and drag the new baseline to a point after the peak.
e. Review your new peak area in either Display/Peak Report or with
File/Print Preview Report.
42
Chromlab TotalChrom Operational Notes
43
Chromlab TotalChrom Operational Notes
1. Batch Reprocess
Batch Reprocess is usually performed after method
integration parameters have been optimized in the method
either manually or by using the Graphic Method Editor.
a. In the TC Navigator screen click on Batch Reprocess.
b. Click on the Three Vial Icon or click on File/Sequence.
c. Select your Sequence file using the browse button.
d. Choose your Actual data path with the browse button.
e. Check the Overwrite existing result files option.
Note: Select None for the Batch printer to reprocess the data files only,
without reprinting the results.
44
Chromlab TotalChrom Operational Notes
Chapter 5 – Reprocess
2. Batch Reprocess – continued
Tip #45 - If you reuse the same sequence over and over again as a
template to collect data for numerous runs it may be easier to
find your data using a sequence index file. A sequence index file
is created each and every time a sequence is run.
45
Chromlab TotalChrom Operational Notes
Chapter 5 – Reprocess
2. Batch Reprocess – continued
46
Chromlab TotalChrom Operational Notes
Chapter 6
Chapter 6 – Chromatograms
1. Chromatograms - Overlay
Tip #46 - Using Chromatograms – Overlay you can select two or more
chromatogram files and overlay them for visual inspection.
Chromatograms – Overlay
47
Chromlab TotalChrom Operational Notes
Chapter 6 – Chromatograms
1. Chromatograms – Auto Size
Tip #47 - With Chromatograms – Auto Size you can expand all
selected chromatograms displayed to see the chromatogram
details more easily.
48
Chromlab TotalChrom Operational Notes
Chapter 7
49
Chromlab TotalChrom Operational Notes
50
Chromlab TotalChrom Operational Notes
51
Chromlab TotalChrom Operational Notes
52
Chromlab TotalChrom Operational Notes
53
Chromlab TotalChrom Operational Notes
54
Chromlab TotalChrom Operational Notes
55