Liq Manual 51-5081A-HT
Liq Manual 51-5081A-HT
Liq Manual 51-5081A-HT
PN 51-5081A-HT/rev.J
January 2011
Model 5081-A-HT
HART® Smart Two-Wire Chlorine, Dissolved Oxygen, and
Ozone Transmitter
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards.
Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure
they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into
your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to
this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct
manual, telephone 1-800-654-7768 and the requested manual will be provided. Save this Instruction Manual for future
reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable
local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount.
Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at
risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being
performed by qualified persons, to prevent electrical shock and personal injury.
CAUTION
If a Model 275 Universal Hart® Communicator is used with these transmitters, the software within the Model 275 may require
modification. If a software modification is required, please contact your local Rosemount Analytical Service Group or National
Response Center at 1-800-654-7768.
3.0 WIRING.................................................................................................................... 11
3.1 Wiring Model 499A Oxygen, Chlorine, and Ozone Sensors .................................... 11
3.2 Wiring Model 499ACL-01 (Free Chlorine) Sensors and pH Sensors....................... 12
3.3 Wiring Model Hx438 and Gx448 Sensors................................................................ 14
7.0 PROGRAMMING..................................................................................................... 35
7.1 General .................................................................................................................... 35
7.2 Default Settings........................................................................................................ 35
7.3 Output Ranging........................................................................................................ 37
i
MODEL 5081-A TABLE OF CONTENTS
ii
MODEL 5081-A TABLE OF CONTENTS
LIST OF TABLES
Number Title Page
7-1 Default Settings ....................................................................................................... 36
17-1 Replacement Parts for Model 5081-A Transmitter ................................................... 98
iii
MODEL 5081-A TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
1-1 Transmitter Display During Calibration and Programming ....................................... 4
1-2 Infrared Remote Controller....................................................................................... 4
1-3 HART Communication.............................................................................................. 4
1-4 Model 5081-A Mounting and Dimensional Drawings................................................ 5
2-1 Mounting the Model 5081-A on a Flat Surface ......................................................... 8
2-2 Using the Pipe Mounting Kit to attach the Model 5081-A to a pipe .......................... 9
2-3 Load/Power Supply Requirements ........................................................................... 10
2-4 Power Supply/Current Loop Wiring .......................................................................... 10
3-1 Amperometric sensors with standard cable ............................................................. 11
3-2 Amperometric sensors with optimum EMI/RFI cable or Variopol cable.................... 11
3-3 Free Chlorine sensor with standard cable and 399VP-09 pH sensor without .........
internal preamplifier.................................................................................................. 12
3-4 Free Chlorine sensor with standard cable and 399-14 pH sensor with internal ......
preamplifier............................................................................................................... 12
3-5 Free Chlorine sensor with standard cable and 399-09-62 pH sensor without .........
internal preamplifier.................................................................................................. 13
3-6 Free Chlorine sensor with optimum EMI/RFI cable or Variopol cable and 399VP-09
pH sensor without internal preamplifier .................................................................... 13
3-7 Free Chlorine sensor with optimum EMI/RFI cable or Variopol cable and 399-14 ..
pH sensor with internal preamplifier ......................................................................... 13
3-8 Free Chlorine sensor with optimum EMI/RFI cable or Variopol cable and 399-09-62
pH sensor without internal preamplifier .................................................................... 13
3-9 Hx438 and Gx448 sensors......................................................................................... 14
4-1 FMRC Explosion-Proof Installation .......................................................................... 15
4-2 FM Intrinsically Safe Installation Label ..................................................................... 16
4-3 FM Intrinsically Safe Installation ............................................................................... 17
4-4 CSA Intrinsically Safe Installation Label ................................................................... 19
4-5 CSA Intrinsically Safe Installation ............................................................................. 20
4-6 ATEX Intrinsically Safe Installation Label ................................................................. 22
4-7 ATEX Intrinsically Safe Installation ........................................................................... 23
5-1 Process Display Screen ........................................................................................... 25
5-2 Program Display Screen .......................................................................................... 25
5-3 Infrared Remote Controller and label ....................................................................... 26
5-4 Menu Tree ................................................................................................................ 28
6-1 Connecting the HART Communicator ...................................................................... 29
6-2 Menu Tree (HART) ................................................................................................... 31
9-1 Sensor Current as a Function of Dissolved Oxygen Concentration ......................... 53
10-1 Sensor Current as a Function of Free Chlorine Concentration ................................ 57
10-2 Dual Slope Calibration.............................................................................................. 60
11-1 Determination of Total Chlorine ................................................................................ 62
11-2 Sensor Current as a Function of Total Chlorine Concentration ................................ 62
11-3 Dual Slope Calibration.............................................................................................. 65
12-1 Sensor Current as a Function of Ozone Concentration ........................................... 67
13-1 Calibration Slope and Offset .................................................................................... 70
14-1 Current Output Calibration ....................................................................................... 75
16-1 Simulate dissolved oxygen....................................................................................... 93
16-2 Simulate chlorine and ozone .................................................................................... 93
16-3 Simulate pH.............................................................................................................. 94
16-4 Three-wire RTD Configuration ................................................................................. 94
16-5 Simulating RTD Inputs.............................................................................................. 95
16-6 Checking for a Poisoned Reference Electrode ........................................................ 96
17-1 Exploded View of Model 5081-A Transmitter ........................................................... 97
iv
MODEL 5081-A SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1
MODEL 5081-A SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
3
MODEL 5081-A SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
MILLIMETER
INCH
5
MODEL 5081-A SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
MODEL
5081-A SMART TWO-WIRE MICROPROCESSOR TRANSMITTER
CODE REQUIRED SELECTION
HT Analog 4-20 mA output with superimposed HART digital signal
FF Foundation Fieldbus digital output (Available in May, 2002)
1.11 ACCESSORIES
POWER SUPPLY: Use the Model 515 Power Supply to provide dc loop power to the transmitter. The Model 515
provides two isolated sources at 24Vdc and 200 mA each. For more information refer to product data sheet
71-515.
ALARM MODULE: The Model 230A alarm Module receives the 4-20 mA signal from the 5081-A transmitter and
activates two alarm relays. High/high, low/low, and high/low are available. Hysteresis (deadband) is also
adjustable. For more information, refer to product data sheet 71-230A.
HART COMMUNICATOR: The Model 275 HART communicator allows the user to view measurement values as
well as to program and configure the transmitter. The Model 275 attaches to any wiring terminal across the
output loop. A minimum 250 Ω load must be between the power supply and transmitter. Order the Model
275 communicator from Rosemount Measurement. Call (800) 999-9307.
ACCESSORIES
MODEL/PN DESCRIPTION
515 DC loop power supply (see product data sheet 71-515)
230A Alarm module (see product data sheet 71-230A)
23572-00 Infrared remote controller (required, one controller can operate any 5081 transmitter)
2002577 2-in. pipe mounting kit
9241178 Stainless steel tag, specify marking
6
MODEL 5081-A SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2.1 Unpacking and inspection
2.2 Orienting the display board
2.3 Installation
2.4 Power supply/current loop
2.3 INSTALLATION
2.3.1 General information
1. The transmitter tolerates harsh environments. For best results, install the transmitter in an area where temperature
extremes, vibrations, and electromagnetic and radio frequency interference are minimized or absent.
2. To prevent unintentional exposure of the transmitter circuitry to the plant environment, keep the cover lock in place
over the circuit end cap. See Figure 2-1. To remove the circuit end cap loosen the lock nut until the tab disengages
from the cap. Then unscrew the cover.
3. The transmitter has two ¾-inch conduit openings, one on each side of the housing. See Figure 2-1.
4. Use weathertight cable glands to keep moisture out of the analyzer. If both a chlorine and pH sensor are being used,
install a cable gland with a dual hole seal insert.
5. If conduit is used, plug and seal the connections at the transmitter housing to prevent moisture from getting inside
the transmitter.
NOTE
Moisture allowed to accumulate in the housing can affect the performance
of the transmitter and may void the warranty.
7
MODEL 5081-A SECTION 2.0
INSTALLATION
INCH
MILLIMETER
8
MODEL 5081-A SECTION 2.0
INSTALLATION
MILLIMETER
INCH
FIGURE 2-2. Using the pipe mounting kit (PN 2002577) to attach the Model 5081-A to a pipe.
9
MODEL 5081-A SECTION 2.0
INSTALLATION
SECTION 3.0
SENSOR WIRING
3.1 Wiring Model 499A oxygen, chlorine, and ozone sensors
3.2 Wiring Model 499ACL-01 (free chlorine) and pH sensors
NOTE
The Model 5081-A transmitter leaves the factory configured for use with the Model 499ADO sen-
sor (ppm dissolved oxygen). If a 499ADO sensor is not being used, turn to Section 7.5.3 and
configure the transmitter for the desired measurement (ppb oxygen, oxygen measured using a
steam-sterilizable sensor, free chlorine, total chlorine, or ozone) before wiring the sensor to the
transmitter. Operating the transmitter and sensor for longer than five minutes while the trans-
mitter is improperly configured will greatly increase the stabilization time for the sensor.
Be sure to turn off power to the transmitter before wiring the sensor.
FIGURE 3-1. Amperometric sensors FIGURE 3-2. Amperometric sensors with optimum
with standard cable. EMI/RFI cable or Variopol cable.
11
MODEL 5081-A SECTION 3.0
SENSOR WIRING
12
MODEL 5081-A SECTION 3.0
SENSOR WIRING
FIGURE 3-5. Free chlorine sensor with standard FIGURE 3-6. Free chlorine sensor with optimum
cable and 399-09-62 pH sensor without internal EMI/RFI cable or Variopol cable and 399VP-09 pH
preamplifier. sensor without internal preamplifier.
13
MODEL 5081-A SECTION 3.0
SENSOR WIRING
14
MODEL 5081-A SECTION 4.0
INTRINSICALLY SAFE & EXPLOSION PROOF INSTALLATIONS
SECTION 4.0
INTRINSICALLY SAFE & EXPLOSION PROOF INSTALLATIONS
15
16
FIGURE 4-2. FM Intrinsically Safe Installation Label
FIGURE 4-3. FM Intrinsically Safe Installation (1 of 2)
17
18
FIGURE 4-3. FM Intrinsically Safe Installation (2 of 2)
FIGURE 4-4. CSA Intrinsically Safe Installation Label
19
20
FIGURE 4-5. CSA Intrinsically Safe Installation (1 of 2)
FIGURE 4-5. CSA Intrinsically Safe Installation (2 of 2)
21
22
FIGURE 4-6. ATEX Intrinsically Safe Installation Label
FIGURE 4-7. ATEX Intrinsically Safe Installation (1 of 2)
23
24
FIGURE 4-7. ATEX Intrinsically Safe Installation (2 of 2)
MODEL 5081-A SECTION 5.0
DISPLAY AND OPERATION
SECTION 5.0
DISPLAY AND OPERATION
5.1 Display Screens
5.2 Infrared Remote Controller (IRC) - Key Functions
5.3 Menu Tree
5.4 Diagnostic Messages
5.5 Security
5.6 Using Hold
Concentration of oxygen,
ozone, or chlorine
Temperature in °C or °F
DBMJcsBuF
Submenus, prompts, and
Commands for submenus, EXIT NEXT ENTER diagnostic measurements
prompts, or diagnostics appear here
25
MODEL 5081-A SECTION 5.0
DISPLAY AND OPERATION
26
MODEL 5081-A SECTION 5.0
DISPLAY AND OPERATION
5.5 SECURITY
5.5.1 Purpose. Use the security code to prevent program settings and calibrations from accidentally being changed.
To program a security code, refer to Section 7.5.
PROGRAM
Je 1. If settings are protected with a security code, pressing PROG or CAL on the remote
111
controller causes the Id screen to appear.
EXIT ENTER
2. Use the arrow keys to enter the security code. Press ENTER.
3. If the security code is correct, the first submenu appears. If the code is incorrect, the
process display reappears.
4. To retrieve a forgotton code number, enter 555 at the Id prompt. The present security
code will appear.
27
28
MAIN DISPLAY
(O2 only)
PROMPT
Pvuqvu!!Dvs SELECTION HGM!!1211 uJNF 121
Dpef!!111 eFMuB! 1/13
DISPLAY AND OPERATION
SECTION 5.0
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
SECTION 6.0
OPERATION WITH MODEL 275
CAUTION
4-20 mA + Digital
250
ohm
Model 5081-A
Smart
Control System
Transmitter
Hand Held
Communicator Bridge
(“Configurator”)
Computer
29
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
6.3 Operation
6.3.1 Off-line and On-line Operation
The Model 275 Communicator features off-line and on-line communications. On-line means the communicator is
connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas-
urement data, change program settings, and read diagnostic messages. Off-line means the communicator is not
connected to the transmitter. When the communicator is off line, the operator can still program settings into the
communicator. Later, after the communicator has been connected to a transmitter, the operator can transfer the
programmed settings to the transmitter. Off-line operation permits settings common to several transmitters to be
easily stored in all of them.
PROGRAM
1. Press PROG on the infrared remote controller.
IBsu
EXIT NEXT ENTER
2. Press NEXT until the screen at left appears. Press ENTER.
Beesftt 11
3. The HART menu tree is shown at left. Use the arrow keys to change set-
EXIT NEXT ENTER tings. Press ENTER to store. Press NEXT to move to the next item on the
menu.
QsfBNc 16
EXIT NEXT ENTER
cvsTu Pgg
EXIT NEXT ENTER
Je 1111111
EXIT NEXT ENTER
30
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
--------------------------------------------------------------------------------
5081A 275 Menu Tree
--------------------------------------------------------------------------------
Device setup
Process variables
View Fld Dev Vars
Oxygen *
Temp
Snsr Cur
pH #
pH mV #
GI #
Temp Res
View PV-Analog 1
PV is Oxygen *
PV
PV % rnge
PV AO
View SV
SV is Temp **
SV
View TV
TV is Snsr Cur ***
TV
View 4V
4V is Temp Res ****
4V
View Status
Diag/Service
Test device
Loop test
View Status
Master Reset
Fault History
Hold Mode
Calibration
Zero Main Sensor
Air Calibration
In-process Cal
Dual Range Cal #####
Adjust Temperature
pH 2-Pt Cal #
pH Auto Cal #
Standardize pH #
D/A trim
31
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
Diagnostic Vars
Oxygen
Snsr Cur
Sensitivity
Zero Current
pH Value #
pH mV #
pH Slope #
pH Zero Offset #
GI #
Temp
Temp Res
Noise rejection
Basic setup
Tag
PV Range Values
PV LRV
PV URV
PV
PV % rnge
Device information
Distributor
Model
Dev id
Tag
Date
Write protect
Snsr text
Descriptor
Message
Revision #'s
Universal rev
Fld dev rev
Software rev
Hardware rev
Detailed setup
Sensors
Oxygen *
Oxygen Unit [ppm, ppb, %sat] *, *****
Oxygen Sensor [ADO, TRDO, SSDO1, SSDO2] ##
Salinity ###
Bar Press Unit [inHg, mmHg, bar, kPa, atm] ##
Man Bar Press ##
Sensor SST
Sensor SSS
Sensor Zero Limit
Dual Range Cal [Disable, Enable] ####
32
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
pH #
pH Value
pH Comp [Auto, Manual]
Manual pH
Preamp loc [Sensor, Xmtr]
Autocal [Manual, Standard, DIN 19267, Ingold, Merck]
pH Slope
pH SST
pH SSS
pH Zero Offset Limit
pH Diagnostics
Diagnostics [Off, On]
GFH
GFL
Imped Comp [Off, On]
Temperature
Temp Comp [Auto, Manual]
Man. Temp
Temp unit [ºC, ºF]
Temp Snsr
Signal condition
LRV
URV
AO Damp
% rnge
Xfer fnctn
AO lo end point
AO hi end pt
Output condition
Analog output
AO
AO Alrm typ
Fixed
Fault
Loop test
D/A trim
HART output
PV is Oxygen *
SV is Temp **
TV is Snsr Cur ***
4V is pH ****
Poll addr
Burst option [PV, %range/current, Process vars/crnt]
Burst mode [Off, On]
Num req preams
Num resp preams
33
MODEL 5081-A SECTION 6.0
OPERATION WITH MODEL 275
Device information
Distributor
Model
Dev id
Tag
Date
Write protect
Snsr text
Descriptor
Message
Revision #'s
Universal rev
Fld dev rev
Software rev
Hardware rev
Local Display
AO LOI Units [mA, %]
Xmtr ID
Noise rejection
Load Default Conf.
Review
Sensors
Outputs
Device information
PV
PV AO
PV LRV
PV URV
--------------------------------------------------------------------------------
Notes:
* Can be Oxygen, Free Cl, Ozone, Ttl Cl, or Chlrmn
** Can be *, Temp, pH, GI
*** Can be *, Snsr Cur, Temp, pH, GI
**** Can be *, Snsr Cur, Temp, pH, GI, Temp Res, Not Used
***** Units for Ozone can be ppm or ppb. For any of the chlorines, unit is
always ppm.
34
MODEL 5081-A SECTION 7.0
PROGRAMMING
SECTION 7.0
PROGRAMMING
7.1 General
7.2 Default Settings
7.3 Output Ranging
7.4 Temperature Settings
7.5 Display
7.6 Factory Default
7.7 HART
7.8 Calibration Setup
7.9 Line Frequency
7.10 pH Measurement
7.11 Barometric Pressure
7.1 GENERAL
This section describes how to do the following:
1. assign values to the 4 and 20 mA outputs
2. set the current generated by the transmitter during hold
3. set the current generated by the transmitter when a fault is detected
4. change sensor diagnostic limits
5. enable and disable automatic temperature correction
6. program the type measurement (oxygen, ozone, or chlorine)
7. setup stabilization criteria for calibration
8. enable automatic pH correction for chlorine measurements
9. choose units for barometric pressure (oxygen only)
10. choose limits for diagnostic fault messages
IMPORTANT
Before changing any default settings, configure the transmitter for the measurement you want to
make: oxygen, free chlorine, total chlorine, or ozone. Changing the measurement ALWAYS returns
the transmitter to factory default settings.
35
MODEL 5081-A SECTION 7.0
PROGRAMMING
C. Display dISPLAY
1. Type of measurement tYPE Oxygen, ozone, free chlorine, total chlorine oxygen
2. Units (oxygen only) Unit ppm, ppb, or % ppm
3. Units (ozone only) Unit ppm or ppb ppm
4. Sensor (oxygen only) SEnSor 499ADO, 499ATrDO, Hx438 or Gx338, other biopharm 499ADO
5. Temperature units tEMP °C or °F °C
6. Output current units OutPut mA or % of full scale mA
7. Security code CodE 000 to 999 000
F. HART HArt
1. Address AddrESS 00 to 15 00
2. Preamble PrEAMb 05 to 20 05
3. Burst bUrSt on or off off
4. ID Id 0000000 to 9999999 0000000
7.3.1 Purpose
This section describes how to do the following:
1. assign values to the 4 and 20 mA outputs
2. set the output current generated during hold
3. set the output current generated when a fault is detected
4. control the amount of dampening on the output signal
5. generate a test current.
7.3.2 Definitions
1. CURRENT OUTPUTS. The transmitter provides a continuous 4 - 20 mA output directly proportional to the concentra-
tion of the analyte. The analyte is the substance being determined (oxygen, chlorine, or ozone).
2. HOLD. During calibration and maintenance the transmitter output may be outside the normal operating range. Placing
the transmitter on hold prevents false alarms or the unwanted operation of chemical dosing pumps. The transmitter
output can be programmed to remain at the last value or to generate any current between 3.80 and 22.00 mA. During
hold, the transmitter displays the present concentration and temperature. The word HOLD appears in the display.
3. FAULT. A fault is a system disabling condition. When the transmitter detects a fault, the following happens:
a. The display flashes.
b. The words FAULT and HOLD appear in the main display.
c. A fault or diagnostic message appears in the display.
d. The output signal remains at the present value or goes to the programmed fault value. Permitted values
are between 3.80 and 22.00 mA.
e. If the transmitter is in HOLD when the fault occurs, the output remains at the programmed hold value. To alert the
user that a fault exists, the word FAULT appears in the main display, and the display flashes. A fault or diagnostic
message also appears.
f. If the transmitter is simulating an output current when the fault occurs, the transmitter continues to generate the sim-
ulated current. To alert the user that a fault exists, the word FAULT appears in the display, and the display flashes.
4. DAMPEN. Output dampening smooths out noisy readings. But it also increases the response time of the output. To
estimate the time (in minutes) required for the output to reach 95% of the final reading following a step change, divide
the setting by 20. Thus, a setting of 140 means that, following a step change, the output takes about seven minutes
to reach 95% of final reading. The output dampen setting does not affect the response time of the process display. The
maximum setting is 255.
5. TEST. The transmitter can be programmed to generate a test current.
37
MODEL 5081-A SECTION 7.0
PROGRAMMING
PvuQvu 1. Press PROG on the remote controller. The OutPut submenu appears.
EXIT NEXT ENTER
PROGRAM
5NB 11/11 2. Press ENTER. The screen displays the 4 MA prompt. Use the arrow keys to change
EXIT ENTER
the setting. Press ENTER to save.
PROGRAM
31NB 21/11 3. The screen displays the 20 MA prompt. Use the arrow keys to change the setting.
EXIT ENTER Press ENTER to save.
PROGRAM
IpMe 32/11 4. The screen displays the HoLd prompt. Use the arrow keys to change the setting to
EXIT ENTER the output desired when the transmitter is in hold. The range is 3.80 to 22.00 mA.
Entering 00.00 causes the transmitter to hold the output at the value it was when
placed in hold. The hold setting overrides the fault setting. Press ENTER to save.
PROGRAM
GBVMu 33/11 5. The screen displays the FAULt prompt. Use the arrow keys to change the setting to
EXIT ENTER the output desired when the transmitter detects a fault. The range is 3.80 to 22.00 mA.
Entering 00.00 causes the transmitter to hold the output at the value it was when the
fault occurred. Press ENTER to save.
PROGRAM
eQo 6. The screen displays the dPn prompt. Use the arrow keys to change the setting. The
111
EXIT ENTER
range is 0 to 255. Press ENTER to save.
PROGRAM
uftu 2 3/11 7. The screen displays the tESt prompt. Use the arrow keys to enter the desired test cur-
EXIT ENTER rent. Press ENTER to start the test. Press EXIT to end the test.
38
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.4.2 Definitions
1. AUTOMATIC TEMPERATURE COMPENSATION - OXYGEN, CHLORINE, AND OZONE. The oxygen, chlo-
rine, and ozone sensors used with the 5081-A transmitter are membrane-covered amperometric sensors. The
permeability of the membrane, or the ease with which the analyte passes through the membrane, is a func-
tion of temperature. As temperature increases, permeability increases, and the analyte diffuses more readily
through the membrane. Because sensor current depends on diffusion rate, a temperature increase will cause
the sensor current and transmitter reading to increase even though the concentration of analyte remained con-
stant. A correction equation in the software automatically corrects for changes in membrane permeability
caused by temperature. Temperature is also used in the pH correction applied to free chlorine readings and in
automatic air calibration of oxygen sensors. In automatic temperature correction, the transmitter uses the tem-
perature measured by the sensor for all calculations in which temperature is used.
2. MANUAL TEMPERATURE COMPENSATION - OXYGEN, CHLORINE, AND OZONE. In manual temperature
compensation, the transmitter uses the temperature entered by the user for membrane permeability and pH
corrections and for air calibration calculations. It does not use the actual process temperature. Do NOT use
manual temperature correction unless the measurement and calibration temperatures differ by no more than
about 2°C. Manual temperature correction is useful if the sensor temperature element has failed and a replace-
ment sensor is not available.
3. AUTOMATIC TEMPERATURE COMPENSATION - pH. The transmitter uses a temperature-dependent factor
to convert measured cell voltage to pH. In automatic temperature compensation the transmitter measures the
temperature and automatically calculates the correct conversion factor. Temperature is also used in automat-
ic buffer calibration. For maximum accuracy, use automatic temperature correction.
4. MANUAL TEMPERATURE COMPENSATION - pH. In manual temperature compensation, the transmitter con-
verts measured voltage to pH using the temperature entered by the user. It does not use the actual process
temperature. Do NOT use manual temperature compensation unless the process temperature varies no more
than about 2°C or the pH is between 6 and 8. Manual temperature compensation is useful if the sensor tem-
perature element has failed and a replacement sensor is not available.
39
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.4.3 Procedure
PROGRAM 1. Press PROG on the remote controller.
uFNQ 2. Press NEXT until the tEMP submenu appears. Press ENTER.
EXIT NEXT ENTER
PROGRAM
3. The screen displays the tAUtO (automatic temperature compensation) prompt.
uBVuP PO Press é or ê to toggle between On and OFF. Press ENTER to save.
EXIT ENTER
PROGRAM
4. If you disable tAuto, the tMAN prompt appears. Use the arrow keys to change the
uNBo 136/1
temperature to the desired value. To enter a negative number, press è or ç until no
EXIT ENTER
digit is flashing. Then press é or ê to display the negative sign. The temperature
entered in this step will be used in all measurements (oxygen, chlorine, ozone,
or pH), no matter what the process temperature is. Press ENTER to save.
40
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.5 DISPLAY
7.5.1 Purpose
This section describes how to do the following:
1. Configure the transmitter to measure oxygen, free chlorine, total chlorine, or ozone
2. Choose concentration units
3. Set the temperature units to °C or °F
4. Set the output to current or percent of full scale
5. Enter a security code.
7.5.2 Definitions
1. MEASUREMENT. The transmitter can be configured to measure dissolved oxygen (ppm and ppb level), free chlorine,
total chlorine, or ozone.
2. FREE CHLORINE. Free chlorine is the product of adding sodium hypochlorite (bleach), calcium hypochlorite (bleaching
powder), or chlorine gas to fresh water. Free chlorine is the sum of hypochlorous acid (HOCl) and hypochlorite ion (OCl-)
3. TOTAL CHLORINE. Total chlorine is the sum of free and combined chlorine. Combined chlorine generally refers to
chlorine oxidants in which chlorine is combined with ammonia or organic amines. Monochloramine, used to disinfect
drinking water, is an example of combined chlorine. The term total chlorine also refers to other chlorine oxidants such
as chlorine dioxide. To measure total chlorine, the sample must first be treated with a mixture of acetic acid and potas-
sium iodide. Total chlorine reacts with iodide to produce an equivalent amount of iodine, which the sensor measures.
4. OUTPUT CURRENT. The transmitter generates a 4-20 mA output signal directly proportional to the concentration of oxy-
gen, chlorine, or ozone in the sample. The output signal can be displayed as current (in mA) or as percent of full scale.
5. SECURITY CODE. The security code unlocks the transmitter and allows access to all menus.
7.5.3 Procedure
PROGRAM
6. For best results make the following settings based on the sensor being used.
Sensor Units
499ADO ppm or %
499ATrDO ppb
Gx448 ppm or %
Hx438 ppm or %
PROGRAM
Voju QQc 7. If you chose O3 in step 3, the screen at left appears. Press é or ê to toggle between
EXIT ENTER
ppm and ppb. Press ENTER to save.
42
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.6.2 Procedure
eJTQMBZ
EXIT NEXT ENTER
2. Press NEXT until the dEFAULt appears in the display. Press ENTER.
PROGRAM
3. Use é or ê to toggle between nO and yES. With yES showing, press ENTER to return
uZQF 13 to factory default settings.
EXIT ENTER
7.7 HART
For more information, see Section 6.3.2.
43
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.8.2 Definitions
1. STABILIZATION CRITERION. The transmitter can be programmed not to accept calibration data until the reading has
remained within a specified concentration range for a specified period of time. For example, a stability criterion of
0.05 ppm for 10 seconds means that calibration data will not be accepted until the reading changes less than 0.05 ppm
over a 10-second period. The transmitter calculates the concentration using the present calibration data, or in the case
of a first time calibration, the default sensitivity.
2. SENSOR ZERO LIMIT. Even in the complete absence of the substance being determined, all amperometric sensors
generate a small current called the zero or residual current. The transmitter compensates for the residual current by
subtracting it from the measured current before converting the result to a concentration value. The zero current varies
from sensor to sensor. The transmitter can be programmed not to accept a zero current until the value has fallen below
a reasonable limit.
3. SALINITY (DISSOLVED OXYGEN ONLY). The solubility of oxygen in water depends on the concentration of dissolved
salts in the water. Increasing the concentration decreases the solubility. If the salt concentration is greater than about
1000 ppm, the accuracy of the measurement can be improved by applying a salinity correction. Enter the salinity as
parts per thousand (‰). One percent is ten part per thousand.
4. DUAL SLOPE CALIBRATION (FREE AND TOTAL CHLORINE ONLY). Free and total chlorine sensors from
Rosemount Analytical (Model 499ACL-01 and 499ACL-02) become non-linear at high concentrations of chlorine. Dual
slope calibration allows the analyzer to correct for the non-linearity of the sensor. For more information see Section
10.4 or 11.4.
7.8.3 Procedure
1. Press PROG on the remote controller.
PROGRAM
Dbm!tfuVq 2. Press NEXT until the CAL SEtUP submenu appears. Press ENTER.
EXIT NEXT ENTER
PROGRAM
tqbo!Dbm 3. The screen displays the SPAn CAL prompt. To set the stabilization criteria, press
ENTER.
EXIT NEXT ENTER
PROGRAM
4. The screen displays the StABiLiSE prompt. Press ENTER.
tubcjmjtf
EXIT NEXT ENTER
PROGRAM 5. Set the stabilization time between 0 and 99 seconds. The default value is 10 sec-
uJNF 21 onds. Press ENTER to save.
EXIT ENTER
44
MODEL 5081-A SECTION 7.0
PROGRAMMING
PROGRAM 6. Set the stabilization range to between 0.01 and 9.99 ppm. The default values are
shown in the table. Press ENTER to save.
efmub 1/16
EXIT ENTER Oxygen 0.05 ppm or 1%
Free chlorine 0.05 ppm
Total chlorine 0.05 ppm
Ozone 0.01 ppm
PROGRAM 7. The display returns to the StABiLiSE prompt. Press NEXT. The next screen depends
on the measurement being made. For free or total chlorine see step 8. For oxygen,
Tubcjmjtf see step 9. For ozone see step 10.
EXIT NEXT ENTER
PROGRAM 8. If the measurement is free or total chlorine, the SLOPE prompt appears. Use é or ê
to toggle between SnGL (single) or duAL (dual) slope. Press ENTER. Go to step 10.
tmpqf TohM
EXIT ENTER
NOTE
For the vast majority of applications, single slope calibration is accept-
able. Dual slope calibration is useful in fewer than 5 % of applications.
PROGRAM 9. If the measurement is oxygen, the SALnty (salinity) prompt appears. Use the arrow
keys to enter the salinity of the water. Press ENTER. Go to step 10.
tbmouz 1/11
EXIT ENTER
PROGRAM 10. The display returns to the SPAn CAL screen. Press NEXT.
tqbo!Dbm
EXIT NEXT ENTER
PROGRAM 12. Enter the desired zero limit. The units are the same as the units programmed in Sec-
tion 7.5. Default limits are given in the table.
mjnju 11/11
EXIT ENTER Oxygen (ppm) 0.05 ppm
Oxygen (ppb) 2.0 ppb
Oxygen (% saturation) 1%
Free chlorine 0.05 ppm
Total chlorine 0.05 ppm
Ozone 0.01 ppm or 10 ppb
45
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.9.2 Procedure
1. Press PROG on the remote controller.
PROGRAM
2. Press NEXT until the LinE FrEq submenu appears. Press ENTER.
mjof!gsf:
EXIT NEXT ENTER
PROGRAM 3. Use é or ê to toggle the line frequency between 50 and 60 Hz. Press ENTER to
mjof save.
71
EXIT ENTER
4. Press RESET to return to the main display.
46
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.10 pH MEASUREMENT
NOTE
The pH measurement submenu appears only if the transmitter has been configured to measure
free chlorine. pH is not available with any other meassurement.
7.10.1 Purpose
This section describes how to do the following:
1. Enable and disable automatic pH correction for free chlorine measurements
2. Set a pH value for manual pH correction
3. Enable and disable pH sensor diagnostics
4. Set upper and lower limits for glass impedance diagnostics
5. Enable and disable automatic pH calibration
6. Set stability criteria for automatic pH buffer calibration.
7.10.2 Definitions
1. AUTOMATIC pH CORRECTION. Free chlorine is the sum of hypochlorous acid (HOCl) and hypochlorite ion (OCl-).
The relative amount of each depends on pH. As pH increases, the concentration of HOCl decreases and the concen-
tration of OCl- increases. Because the sensor responds only to HOCl, a pH correction is necessary to properly con-
vert the sensor current into a free chlorine reading. The transmitter uses both automatic and manual pH correction. In
automatic pH correction the transmitter continuously monitors the pH of the sample and corrects the free chlorine read-
ing for changes in pH. In manual pH correction, the user enters the pH of the sample. Generally, if the pH changes
more than about 0.2 units over short periods of time, automatic pH correction is best. If the pH is relatively steady or
subject only to seasonal changes, manual pH correction is adequate.
2. REFERENCE OFFSET. The transmitter reading can be changed to match the reading of a second pH meter. If the dif-
ference (converted to millivolts) between the transmitter reading and the desired value exceeds the programmed limit,
the transmitter will not accept the new reading. To estimate the millivolt difference, multiply the pH difference by 60.
3. pH SENSOR DIAGNOSTICS. The transmitter continuously monitors the pH sensor for faults. A fault means that the
sensor has failed or is possibly nearing failure. The only pH sensor diagnostic available in the 5081-A is glass imped-
ance.
4. GLASS IMPEDANCE. The transmitter monitors the condition of the pH-sensitive glass membrane in the sensor by
continuously measuring the impedance across the membrane. Typical impedance is 100 to 500 MΩ. A low impedance
(<10 MΩ) means the glass membrane has cracked and the sensor must be replaced. An extremely high impedance
(>1000MΩ) implies that the sensor is aging and may soon need replacement. High impedance might also mean that
the glass membrane is no longer immersed in the process liquid.
5. AUTOMATIC pH CALIBRATION. The transmitter features both automatic and manual pH calibration. In automatic cal-
ibration, screen prompts direct the user through a two-point buffer calibration. The transmitter recognizes the buffers
and uses temperature-corrected values in the calibration. The table in Section 13.1 lists the standard buffers the trans-
mitter recognizes. The transmitter also recognizes several technical buffers: Merck, Ingold, and DIN 19267. During
automatic calibration, the transmitter does not accept data until programmed stability limits have been met.
6. MANUAL pH CALIBRATION. If automatic pH calibration is deactivated, the user must perform a manual calibration. In
manual calibration the user judges when readings are stable and manually enters the buffer values. Because manu-
al calibration greatly increases the chance of making an error, the use of automatic calibration is strongly rec-
ommended.
47
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.10.3 Procedure
PROGRAM
4. The screen displays the dIAgnOStIC submenu header. Prompts under this header
eJbhoptuJD allow the user to change the reference offset and pH sensor diagnostic limits.
EXIT NEXT ENTER
The default settings are: reference offset 60 mV
pH sensor diagnostics off
To keep the default settings, press NEXT. Go to step 11.
To change the reference offset or to enable or make changes to the glass diagnostic
PROGRAM settings, press ENTER. Go to step 5.
sPggtfu 171 5. The rOFFSEt (reference offset) prompt appears. Use the arrow keys to change the
EXIT ENTER offset to the desired value in mV. Press ENTER to save.
PROGRAM
6. The dIAg (diagnostics) prompt appears. Press é or ê to toggle between OFF (dis-
eJbh Pgg able) or On (enable). Press ENTER to save.
EXIT ENTER
PROGRAM
7. The IMPtC (glass impedance temperature correction) prompt appears. Press é or ê
JnquD Po to toggle between OFF (disable) or On (enable). Because glass impedance is a
EXIT ENTER
strong function of temperature, correcting glass impedance for temperature effects is
strongly recommended. Press ENTER to save.
PROGRAM
8. The GFH (glass fault high) prompt appears. Use the arrow keys to change the set-
HgI 2111 ting to the desired value. The default setting is 1000 MΩ. Entering 0000 disables the
EXIT ENTER feature. Press ENTER to save.
When the glass electrode impedance exceeds the limit, the transmitter displays the
GLASSFAIL diagnostic message and sets a fault condition.
PROGRAM 9. The GFL (glass fault low) prompt appears. Use the arrow keys to change the setting
HgM to the desired value. The default setting is 10 MΩ. Entering 0000 disables the feature.
1121
Press ENTER to save.
EXIT ENTER
When the glass electrode impedance falls below the limit, the transmitter displays the
GLASSFAIL diagnostic message and sets a fault condition.
48
MODEL 5081-A SECTION 7.0
PROGRAMMING
PROGRAM
eJbhoptuJD 10. Once diagnostic limits have been set, the display returns to the dIAgnOStIC sub-
menu header. Press NEXT.
EXIT NEXT ENTER
PROGRAM 11. The PH CAL submenu header appears. Prompts under this header allow the user to
enable or disable automatic buffer calibration, select the buffers to be used, and set
QI!Dbm
stabilization criteria for pH calibration. The default settings are:
EXIT NEXT ENTER
Automatic buffer calibration On
Buffers Standard (see Section 7.10.2)
Stabilization <0.02 pH in 10 seconds
cbVuP Po 12. The bAUtO (automatic buffer calibration) prompt appears. Press é or ê to toggle
EXIT ENTER between OFF (disable) or On (enable). Press ENTER to save. Keeping automatic
buffer calibration enabled is strongly recommended.
PROGRAM
cvggfs Tue 13. The buFFEr prompt appears. Press é or ê to scroll through the list of available
EXIT ENTER
buffers. See Section 13.1 for a list of the buffer values.
Std Standard buffers
ErC Merck buffers
InG Ingold buffers
din DIN 19267 buffers
tubcjmjtf 14. The StAbiLiSE (stabilize) prompt appears. To change stabilization criteria, press
EXIT ENTER ENTER. To leave stabilization criteria at the default values, press EXIT.
PROGRAM
15. Set the stabilization time between 0 and 99 seconds. The default value is 10 seconds.
ujnf 2/1 Press ENTER to save.
EXIT ENTER
PROGRAM
16. Set the stabilization range to between 0.02 and 0.50 pH. Press ENTER to save.
efmub 1/13
EXIT ENTER
17. Press RESET to return to the main display.
49
MODEL 5081-A SECTION 7.0
PROGRAMMING
7.11.1 Purpose
This section describes how to do the following
1. Set the units for barometric pressure
2. Enter a pressure other than the calibration pressure for percent saturation measurements.
7.11.2 Definitions
1. BAROMETRIC PRESSURE. Because the current generated by an amperometric oxygen sensor is directly propor-
tional to the partial pressure of oxygen, the sensor is generally calibrated by exposing it to water saturated air. See
Section 9.1 for more information. To calculate the equivalent concentration of oxygen in water in ppm, the transmitter
must know the temperature and barometric pressure. This submenu lets the user specify the units for barometric pres-
sure.
2. PERCENT SATURATION PRESSURE. Oxygen is sometimes measured in units of percent saturation. Percent satu-
ration is the concentration of oxygen divided by the maximum amount of oxygen the water can hold (the saturation
concentration) at the temperature and pressure of the measurement. Generally, the pressure during the measurement
is assumed to be the same as the pressure when the sensor was calibrated. If the measurement and calibration pres-
sures differ, the measurement pressure can be entered as a separate variable.
7.11.3 Procedure
1. Press PROG on the remote controller.
PROGRAM
cbs!qsftt
2. Press NEXT until the bAr PRESS submenu appears. Press ENTER.
EXIT NEXT ENTER
PROGRAM
3. The Unit prompt appears. Press é or ê to scroll through the list of units:
Voju! ooIH
EXIT ENTER nnHG mm Hg
1000PA kPa
Atn atm
bAr bar
InHG in Hg
Press ENTER to save.
PROGRAM
&!tbu!q 4. If % saturation units were selected in Section 7.5, the % SAt P (saturation pressure)
EXIT NEXT prompt appears. Press NEXT.
PROGRAM
5. Use the arrow keys to enter the desired pressure. The transmitter will use this pres-
q!nbo 871 sure to calculate percent saturation. Press ENTER.
EXIT ENTER
SECTION 8.0
CALIBRATION — TEMPERATURE
8.1 INTRODUCTION
All four amperometric sensors (oxygen, ozone, free chlorine, and total chlorine) are membrane-covered sensors. As
the sensor operates, the analyte (the substance to be determined) diffuses through the membrane and is consumed
at an electrode immediately behind the membrane. The reaction produces a current that depends on the rate at
which the analyte diffuses through the membrane. The diffusion rate, in turn, depends on the concentration of the
analyte and how easily it passes through the membrane (the membrane permeability). Because the membrane per-
meability is a function of temperature, the sensor current will change if the temperature changes. To correct for
changes in sensor current caused by temperature, the transmitter automatically applies a membrane permeability
correction. Although the membrane permeability is different for each sensor, the change is about 3%/°C at 25°C, so
a 1°C error in temperature produces about a 3% error in the reading.
Temperature plays an additional role in oxygen measurements. Oxygen sensors are calibrated by exposing them
to water-saturated air, which, from the point of view of the sensor, is equivalent to water saturated with atmospheric
oxygen (see Section 9.1 for more information). During calibration, the transmitter calculates the solubility of atmos-
pheric oxygen in water using the following steps. First, the transmitter measures the temperature. From the tem-
perature, the transmitter calculates the vapor pressure of water and, using the barometric pressure, calculates the
partial pressure of atmospheric oxygen. Once the transmitter knows the partial pressure, it calculates the equilib-
rium solubility of oxygen in water using a temperature-dependent factor called the Bunsen coefficient. Overall, a
1°C error in the temperature measurement produces about a 2% error in the solubility calculated during calibra-
tion and about the same error in subsequent measurements.
Temperature is also important in the pH measurement required to correct free chlorine readings.
1. The transmitter uses a temperature dependent factor to convert measured cell voltage to pH. Normally, a slight
inaccuracy in the temperature reading is unimportant unless the pH reading is significantly different from 7.00.
Even then, the error is small. For example, at pH 12 and 25°C, a 1°C error produces a pH error less than ±0.02.
2. During auto calibration, the transmitter recognizes the buffer being used and calculates the actual pH of the
buffer at the measured temperature. Because the pH of most buffers changes only slightly with temperature,
reasonable errors in temperature do not produce large errors in the buffer pH. For example, a 1°C error caus-
es at most an error of ±0.03 in the calculated buffer pH.
Without calibration the accuracy of the temperature measurement is about ±0.4°C. Calibrate the transmitter if
1. ±0.4°C accuracy is not acceptable
2. the temperature measurement is suspected of being in error. Calibrate temperature by making the transmitter
reading match the temperature measured with a standard thermometer.
51
MODEL 5081-A SECTION 8.0
CALIBRATION — TEMPERATURE
8.2. PROCEDURE
1. Place the sensor and a calibrated reference thermometer in a container of water at ambient temperature. Be
sure the temperature element in the sensor is completely submerged by keeping the sensor tip at least three
inches below the water level. Stir continuously. Allow at least 20 minutes for the standard thermometer, sen-
sor, and water to reach constant temperature.
CALIBRATE
52
MODEL 5081-A SECTION 9.0
CALIBRATION — OXYGEN
SECTION 9.0
CALIBRATION — OXYGEN
9.1 INTRODUCTION
As Figure 9-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolved
oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero stan-
dard) and to a solution containing a known amount of oxygen (full-scale standard).
The zero standard is necessary because oxygen sensors, even when no oxygen is present in the sample, gener-
ate a small current called the residual current. The analyzer compensates for the residual current by subtracting it
from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing
before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The
recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used.
The Model 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low resid-
ual current and does not normally require zeroing. The residual current in the 499A TrDO sensor is equivalent
to less than 0.5 ppb oxygen.
The purpose of the full-scale standard is to establish the slope of the calibration curve. Because the solubility of
atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural
choice for a full-scale standard is air-saturated water. However, air-saturated water is difficult to prepare and use,
so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air-sat-
urated water are identical. The equivalence comes about because the sensor really measures the chemical poten-
tial of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the
sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water
and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical
potential of oxygen in the two phases (air and water) is the same.
Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a
membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical poten-
tial between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxy-
gen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sen-
sor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate
and the sensor current.
When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current.
Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the
same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is
convenient to use water-saturated air for calibration.
Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air and keys in the
barometric pressure. The transmitter monitors the sensor current. When the current is stable, the transmitter stores
the current and measures the temperature. From the temperature, the transmitter calculates the saturation vapor
pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure from the baromet-
ric pressure. Using the fact that dry air always contains
20.95% oxygen, the transmitter calculates the partial pres-
sure of oxygen. Once the transmitter knows the partial pres-
sure of oxygen, it uses the Bunsen coefficient to calculate the
equilibrium solubility of atmospheric oxygen in water at the
prevailing temperature. At 25°C and 760 mm Hg, the equilib-
rium solubility is 8.24 ppm.
Often it is too difficult or messy to remove the sensor from the
process liquid for calibration. In this case, the sensor can be
calibrated against a measurement made with a portable lab-
oratory instrument. The laboratory instrument typically uses a
membrane-covered amperometric sensor that has been cali-
brated against water-saturated air.
FIGURE 9-1. Sensor Current as a Function
of Dissolved Oxygen Concentration
53
MODEL 5081-A SECTION 9.0
CALIBRATION — OXYGEN
A new sensor or a sensor in which the electrolyte solution has been replaced may require several hours (occa-
sionally as long as overnight) to reach a minimum current. DO NOT START THE ZERO ROUTINE UNTIL THE
SENSOR HAS BEEN IN ZERO SOLUTION FOR AT LEAST TWO HOURS.
CALIBRATE 4. The screen shows the value (in units selected in Section 7.5.3) below
which the reading must be before the zero current will be accepted.
1!bu!!1/16
Assume the units are ppm. The screen shows 0.02. Therefore, the read-
EXIT ENTER
ing must be below 0.02 ppm before the zero will be accepted. For a
499ADO sensor 0.02 ppm corresponds to about 50 nA. To change the
zero limit value, see Section 7.8.3. Press ENTER.
NOTE
The number shown in the main display may change. During the zero
step, the previous zero current is suppressed, and the concentration
shown in the main display is calculated assuming the residual cur-
rent is zero. Once the transmitter accepts the new zero current, it is
used in all subsequent measurements.
CALIBRATE 5. The tiME dELAy message appears and remains until the zero current is
ujnf!efmbz below the concentration limit shown in the previous screen. If the current
is already below the limit, tiME dELAy will not appear. To bypass the
EXIT ENTER
time delay, press ENTER.
CALIBRATE 6. O donE shows that the zero step is complete. Press EXIT.
1!epof
EXIT
7. Press RESET to return to the main display.
54
MODEL 5081-A SECTION 9.0
CALIBRATION — OXYGEN
Tfotps!Dbm 5. Press NEXT. The SEnSor CAL submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
CALIBRATE
ooIH 7. The screen shows the units selected for barometric pressure. Press
NEXT.
EXIT NEXT
CALIBRATE
8. Use the arrow keys to enter the barometric pressure. Press ENTER.
Qsftt! 871/1 NOTE
EXIT ENTER
Be sure to enter the actual barometric pressure. Weather
forecasters and airports usually report barometric pressure
corrected to sea level; they do not report the actual baro-
metric pressure. To estimate barometric pressure from alti-
tude, see Appendix A.
CALIBRATE
9. The tiME dELAy message appears and remains until the sensor read-
ujnf!efmbz ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT ENTER delay, press ENTER.
CALIBRATE
Dbm!ePof 10. This screen appears when the calibration is complete. The concentration
shown in the main display is the solubility of atmospheric oxygen in
EXIT
water at ambient temperature and barometric pressure. Press EXIT.
11. To return to the main display, press RESET.
12. During calibration, the transmitter calculates the sensitivity (nA/ppm) of
the sensor. To check the sensitivity, go to the main display. Press DIAG.
Press NEXT until the SenSitvtY (sensitivity) prompt appears. Press
ENTER to display the sensitivity in nA/ppm. Typical values at 25°C are
given in the table.
Sensor nA/ppm
499ADO 1,800 - 3,100
499ATrDO 3,600 - 6,100
Hx438 and Gx448 4.8 - 9.8
55
MODEL 5081-A SECTION 9.0
CALIBRATION — OXYGEN
CALIBRATE
Tfotps!Dbm 3. Press NEXT. The SEnSor CAL submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE 4. Press NEXT. The Air CAL prompt appears. Press NEXT.
B js!Dbm
EXIT NEXT ENTER
CALIBRATE
5. The In ProCESS prompt appears. Press ENTER.
Jo!QspDftt
EXIT ENTER
CALIBRATE
ujnf!efmbz 6. The tiME dELAy message appears and remains until the sensor read-
ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT NEXT
delay, press ENTER.
CALIBRATE
7. The GrAb SPL (grab sample) message appears. Press ENTER.
Hsbc!tqm
EXIT ENTER
CALIBRATE
8. Use the arrow keys to change the flashing display to the value indicated
by the standard instrument. Press ENTER to save.
Dbm 4/31
EXIT ENTER
56
MODEL 5081-A SECTION 10.0
CALIBRATION - FREE CHLORINE
SECTION 10.0
CALIBRATION — FREE CHLORINE
10.1 INTRODUCTION
As Figure 10-1 shows, a free chlorine sensor generates a current directly proportional to the concentration of free
chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero stan-
dard) and to a solution containing a known amount of chlorine (full-scale standard).
The zero standard is necessary because chlorine sensors, even when no chlorine is in the sample, generate a
small current called the residual current. The transmitter compensates for the residual current by subtracting it
from the measured current before converting the result to a chlorine value. New sensors require zeroing before
being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the
following makes a good zero standard:
• Deionized water containing about 500 ppm sodium chloride. Dissolve 0.5 grams (1/8 teaspoonful) of table
salt in 1 liter of water. DO NOT USE DEIONIZED WATER ALONE FOR ZEROING THE SENSOR. THE
CONDUCTIVITY OF THE ZERO WATER MUST BE GREATER THAN 50 μS/cm.
• Tap water known to contain no chlorine. Expose tap water to bright sunlight for at least 24 hours.
The purpose of the full-scale standard is to establish the slope of the calibration curve. Because stable chlorine
standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liq-
uid. Several manufacturers offer portable test kits for this purpose. Observe the following precautions when tak-
ing and testing the grab sample.
• Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does
not alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the
sensor.
• Chlorine solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sen-
sor when the chlorine concentration is at the upper end of the normal operating range.
Free chlorine measurements made with the 499ACL-01 sensor also require a pH correction. Free chlorine is the sum of
hypochlorous acid (HOCl) and hyprochlorite ion (OCl-). The relative amount of each depends on the pH. As pH increas-
es, the concentration of HOCl decreases and the concentration of OCl- increases. Because the sensor responds only
to HOCl, a pH correction is necessary to properly convert the sensor current into a free chlorine reading.
The transmitter uses both automatic and manual pH correction. In automatic pH correction, the transmitter con-
tinuously monitors the pH of the solution and corrects the free chlorine reading for changes in pH. In manual pH
correction, the transmitter uses a fixed pH value entered by the user to make the correction. Generally, if the pH
changes more than about 0.2 units over short periods of time, automatic pH correction is best. If the pH is rela-
tively steady or subject only to seasonal changes, manual pH correction is adequate.
During calibration, the transmitter must know the pH of the sample. If the transmitter is using automatic pH cor-
rection, the pH sensor (properly calibrated) must be in the process liquid before starting the calibration. If the
transmitter is using manual pH correction, be sure to enter the pH value before starting the calibration.
The Model 499ACL-01 free chlo-
rine sensor loses sensitivity at high
concentrations of chlorine. The
5081-A transmitter has a dual
slope feature that allows the user
to compensate for the non-lineari-
ty of the sensor. However, for the
vast majority of applications, dual
slope calibration is unnecessary.
57
MODEL 5081-A SECTION 10.0
CALIBRATION - FREE CHLORINE
CALIBRATE
CALIBRATE
1!bu! 1/13 4. The screen shows the value (in units ppm) below which the reading
EXIT ENTER
must be before the zero current will be accepted. The screen shows
0.02. Therefore, the reading must be below 0.02 ppm before the zero
will be accepted. For a typical 499ACL-01 sensor, 0.02 ppm corre-
sponds to about 7 nA. To change the zero limit value, see Section 7.8.3.
Press ENTER.
NOTE
The number shown in the main display may change. During the
zero step, the previous zero current is suppressed, and the con-
centration shown in the main display is calculated assuming the
residual current is zero. Once the transmitter accepts the new zero
current, it is used in all subsequent measurements.
CALIBRATE
ujnf!efmbz 5. The tiME dELAY message appears and remains until the zero current
is below the concentration limit shown in the previous screen. If the cur-
EXIT ENTER
rent is already below the limit, tiME dELAy will not appear. To bypass
the time delay, press ENTER.
CALIBRATE
1!ePof 6. O donE shows that the zero step is complete. Press EXIT.
EXIT
7. Press RESET to return to the main display.
58
MODEL 5081-A SECTION 10.0
CALIBRATION - FREE CHLORINE
CALIBRATE
CALIBRATE
ujnf!efmbz 5. Press ENTER. The tiME dELAy message appears and remains until
EXIT NEXT
the sensor reading meets the stability criteria set in Section 7.6. To
bypass the time delay, press ENTER.
NOTE
As soon as the stability criteria are met (or ENTER is
pressed to bypass the time delay), the transmitter stores
the sensor current. Therefore, if the chlorine level in the
process liquid drifts while the sample is being tested, there
is no need to compensate for the change when entering
test results in step 7.
CALIBRATE
Hsbc!tqm 6. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of free
EXIT ENTER chlorine in the sample. Press ENTER.
CALIBRATE
Dbm!! 4/31 7. Use the arrow keys to change the flashing display to the concentration
EXIT ENTER
of chlorine determined in the grab sample. Press ENTER to save.
59
MODEL 5081-A SECTION 10.0
CALIBRATION - FREE CHLORINE
CALIBRATE
CALIBRATE
6. The CAL Pt 1 prompt appears. Adjust the chlorine concentration until it
Dbm!qu2 is near the upper end of the linear range of the sensor. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
ujnf!efmbz 7. The tiME dELAy message appears and remains until the sensor read-
ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT NEXT
delay, press ENTER.
NOTE
As soon as the stability criteria are met (or ENTER is pressed
to by-pass the time delay), the transmitter stores the sensor
current. Therefore, if the chlorine level in the process liquid
drifts while the sample is being tested, there is no need to
compensate for the change when entering test results.
CALIBRATE
Hsbc!tqm 8. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of free
EXIT ENTER
chlorine in the sample. Press ENTER.
60
MODEL 5081-A SECTION 10.0
CALIBRATION - FREE CHLORINE
CALIBRATE
Qu2!! 4/11 9. The Pt1 prompt appears. Use the arrow keys to change the flashing
display to the concentration of chlorine determined in the grab sample.
EXIT ENTER
Press ENTER to save.
CALIBRATE
Dbm!qu3 10. The CAL Pt 2 prompt appears. Adjust the concentration of chlorine until
it is near the top end of the range, i.e., near concentration C2 shown in
EXIT NEXT ENTER
Figure 10-2. Press ENTER.
CALIBRATE
ujnf!efmbz 11. The tiME dELAy message appears and remains until the sensor read-
ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT NEXT
delay, press ENTER.
CALIBRATE
Hsbc!tqm 12. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of free
EXIT ENTER
chlorine in the sample. Press ENTER.
CALIBRATE
Qu3!! 7/11 13. The Pt2 prompt appears. Use the arrow keys to change the flashing dis-
EXIT ENTER
play to the concentration of chlorine determined in the grab sample.
Press ENTER to save.
61
MODEL 5081-A SECTION 11.0
CALIBRATION - TOTAL CHLORINE
SECTION 11.0
CALIBRATION — TOTAL CHLORINE
9.1 INTRODUCTION
Total chlorine is the sum of free and combined chlorine. The continuous determination of total chlorine requires two
steps. See Figure 11-1. First, the sample flows into a conditioning system (SCS 921) where a pump continuously
adds acetic acid and potassium iodide to the sample. The acid lowers the pH, which allows total chlorine in the
sample to quantitatively oxidize the iodide in the reagent to iodine. In the second step, the treated sample flows to
the sensor. The sensor is a membrane-covered amperometric sensor, whose output is proportional to the con-
centration of iodine. Because the concentration of iodine is proportional to the concentration of total chlorine, the
analyzer can be calibrated to read total chlorine.
Figure 11-2 shows a typical calibration curve for a total chlorine sensor. Because the sensor really measures
iodine, calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a solu-
tion containing a known amount of iodine (full-scale standard).
The zero standard is necessary because the sensor, even when no iodine is present, generates a small current
called the residual current. The transmitter compensates for the residual current by subtracting it from the meas-
ured current before converting the result to a total chlorine value. New sensors require zeroing before being
placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero stan-
dard is sample without reagent added.
The purpose of the full-scale standard is to
establish the slope of the calibration curve.
Because stable total chlorine standards do not
exist, the sensor must be calibrated against
a test run on a grab sample of the process
liquid. Several manufacturers offer portable
test kits for this purpose. Observe the following
precautions when taking and testing the grab
sample.
• Take the grab sample from a point as close
as possible to the inlet of the SCS921 sam-
ple conditioning system. Be sure that taking
the sample does not alter the flow through
the SCS921. Sample flow must remain FIGURE 11-1. Determination of Total Chlorine
between 80 and 100 mL/min.
• Chlorine solutions are unstable. Run the
test immediately after taking the sample.
Try to calibrate the sensor when the chlo-
rine concentration is at the upper end of
the normal operating range.
The Model 499ACL-02 (total chlorine) sensor
loses sensitivity at high concentrations of chlo-
rine. The 5081-A transmitter has a dual slope
feature that allows the user to compensate for
the non-linearity of the sensor. However, for
the vast majority of applications, dual slope
calibration is unnecessary.
62
MODEL 5081-A SECTION 11.0
CALIBRATION - TOTAL CHLORINE
CALIBRATE
CALIBRATE
1!bu! 1/13 6. The screen shows the value (in units ppm) below which the reading
EXIT ENTER
must be before the zero current will be accepted. The screen shows
0.02. Therefore, the reading must be below 0.02 ppm before the zero
will be accepted. For a typical 499ACL-02 sensor, 0.02 ppm corre-
sponds to about 20 nA. To change the zero limit value, see Section
7.8.3. Press ENTER.
NOTE
The number shown in the main display may change. During the
zero step, the previous zero current is suppressed, and the con-
centration shown in the main display is calculated assuming the
residual current is zero. Once the transmitter accepts the new zero
current, it is used in all subsequent measurements.
CALIBRATE
ujnf!efmbz 7. The tiME dELAY message appears and remains until the zero current
is below the concentration limit shown in the previous screen. If the cur-
EXIT ENTER
rent is already below the limit, tiME dELAy will not appear. To bypass
the time delay, press ENTER.
CALIBRATE
1!ePof 8. O donE shows that the zero step is complete. Press EXIT.
EXIT
9. Press RESET to return to the main display.
63
MODEL 5081-A SECTION 11.0
CALIBRATION - TOTAL CHLORINE
CALIBRATE
CALIBRATE
ujnf!efmbz 5. Press NEXT. The tiME dELAy message appears and remains until the
EXIT NEXT
sensor reading meets the stability criteria set in Section 7.8. To bypass
the time delay, press ENTER.
NOTE
As soon as the stability criteria are met (or ENTER is
pressed to bypass the time delay), the transmitter stores
the sensor current. Therefore, if the chlorine level in the
process liquid drifts while the sample is being tested, there
is no need to compensate for the change when entering
test results in step 7.
CALIBRATE
Hsbc!tqm 6. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of total
EXIT ENTER chlorine in the sample. Press ENTER.
CALIBRATE
Dbm!! 4/31 7. Use the arrow keys to change the flashing display to the concentration
EXIT ENTER
of chlorine determined in the grab sample. Press ENTER to save.
64
MODEL 5081-A SECTION 11.0
CALIBRATION - TOTAL CHLORINE
CALIBRATE
CALIBRATE
6. The CAL Pt 1 prompt appears. Adjust the chlorine concentration until it
Dbm!qu2 is near the upper end of the linear range of the sensor. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
ujnf!efmbz 7. The tiME dELAy message appears and remains until the sensor read-
ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT NEXT
delay, press ENTER.
NOTE
As soon as the stability criteria are met (or ENTER is pressed
to by-pass the time delay), the transmitter stores the sensor
current. Therefore, if the chlorine level in the process liquid
drifts while the sample is being tested, there is no need to
compensate for the change when entering test results.
CALIBRATE
Hsbc!tqm 8. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of total
EXIT ENTER
chlorine in the sample. Press ENTER.
65
MODEL 5081-A SECTION 11.0
CALIBRATION - TOTAL CHLORINE
CALIBRATE
Qu2!! 4/11 9. The Pt1 prompt appears. Use the arrow keys to change the flashing
display to the concentration of chlorine determined in the grab sample.
EXIT ENTER
Press ENTER to save.
CALIBRATE
Dbm!qu3 10. The CAL Pt 2 prompt appears. Adjust the concentration of chlorine until
it is near the top end of the range, i.e., near concentration C2 shown in
EXIT NEXT ENTER
Figure 11-3. Press ENTER.
CALIBRATE
ujnf!efmbz 11. The tiME dELAy message appears and remains until the sensor read-
ing meets the stability criteria set in Section 7.8. To bypass the time
EXIT NEXT
delay, press ENTER.
CALIBRATE
Hsbc!tqm 12. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of total
EXIT ENTER
chlorine in the sample. Press ENTER.
CALIBRATE
Qu3!! 7/11 13. The Pt2 prompt appears. Use the arrow keys to change the flashing dis-
EXIT ENTER
play to the concentration of chlorine determined in the grab sample.
Press ENTER to save.
66
MODEL 5081-A SECTION 12.0
CALIBRATION - OZONE
SECTION 12.0
CALIBRATION — OZONE
12.1 INTRODUCTION
As Figure 12-1 shows, an ozone sensor generates a current directly proportional to the concentration of ozone in
the sample. Calibrating the sensor requires exposing it to a solution containing no ozone (zero standard) and to a
solution containing a known amount of ozone (full-scale standard).
The zero standard is necessary because ozone sensors, even when no ozone is in the sample, generate a small
current called the residual current. The transmitter compensates for the residual current by subtracting it from the
measured current before converting the result to an ozone value. New sensors require zeroing before being
placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. Either of the fol-
lowing makes a good zero standard:
• Deionized water.
• Tap water known to contain no ozone. Expose tap water to ozone-free air for several hours.
The purpose of the full-scale standard is to establish the slope of the calibration curve. Because stable ozone stan-
dards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid.
Several manufacturers offer portable test kits for this purpose. Observe the following precautions when taking and
testing the grab sample.
• Take the grab sample from a point as close to the sensor as possible. Be sure that taking the sample does not
alter the flow of the sample to the sensor. It is best to install the sample tap just downstream from the sensor.
• Ozone solutions are unstable. Run the test immediately after taking the sample. Try to calibrate the sensor
when the ozone concentration is at the upper end of the normal operating range.
67
MODEL 5081-A SECTION 12.0
CALIBRATION - OZONE
CALIBRATE
CALIBRATE
1!bu! 1/13 4. The screen shows the value (in units ppm) below which the reading
EXIT ENTER
must be before the zero current will be accepted. The screen shows
0.02. Therefore, the reading must be below 0.02 ppm before the zero
will be accepted. For a typical ozone sensor, 0.02 ppm corresponds to
about 7 nA. To change the zero limit value, see Section 7.8.3. Press
ENTER.
NOTE
The number shown in the main display may change. During the
zero step, the previous zero current is suppressed, and the con-
centration shown in the main display is calculated assuming the
residual current is zero. Once the transmitter accepts the new zero
current, it is used in all subsequent measurements.
CALIBRATE
ujnf!efmbz 5. The tiME dELAY message appears and remains until the zero current
is below the concentration limit shown in the previous screen. If the cur-
EXIT ENTER
rent is already below the limit, tiME dELAy will not appear. To bypass
the time delay, press ENTER.
CALIBRATE
1!ePof 6. O donE shows that the zero step is complete. Press EXIT.
EXIT
7. Press RESET to return to the main display.
68
MODEL 5081-A SECTION 12.0
CALIBRATION - OZONE
CALIBRATE
CALIBRATE
ujnf!efmbz 5. Press NEXT. The tiME dELAy message appears and remains until the
EXIT NEXT
sensor reading meets the stability criteria set in Section 7.6. To bypass
the time delay, press ENTER.
NOTE
As soon as the stability criteria are met (or ENTER is
pressed to bypass the time delay), the transmitter stores
the sensor current. Therefore, if the chlorine level in the
process liquid drifts while the sample is being tested, there
is no need to compensate for the change when entering
test results in step 7.
CALIBRATE
Hsbc!tqm 6. The GrAb SPL (grab sample) prompt appears. Take a sample of the
process liquid and immediately determine the concentration of ozone in
EXIT ENTER
the sample. Press ENTER.
CALIBRATE
Dbm!! 4/31 7. Use the arrow keys to change the flashing display to the concentration
EXIT ENTER of ozone determined in the grab sample. Press ENTER to save.
69
MODEL 5081-A SECTION 13.0
CALIBRATION - pH
SECTION 13.0
CALIBRATION — pH
13.1 INTRODUCTION
A new pH sensor must be calibrated before use. Regular recalibration is also necessary.
A pH measurement cell (pH sensor and the solution to be measured) can be pictured as a battery with an extreme-
ly high internal resistance. The voltage of the battery depends on the pH of the solution. The pH meter, which is
basically a voltmeter with a very high input impedance, measures the cell voltage and calculates pH using a con-
version factor. The value of the voltage-to-pH conversion factor depends on the sensitivity of the pH sensing ele-
ment (and the temperature). The sensing element is a thin, glass membrane at the end of the sensor. As the glass
membrane ages, the sensitivity drops. Regular recalibration corrects for the loss of sensitivity. pH calibration stan-
dards, also called buffers, are readily available.
Two-point calibration is standard. Both automatic calibration and manual calibration are available. Auto calibration
avoids common pitfalls and reduces errors. Its use is recommended.
In automatic calibration the transmitter recognizes the buffer and uses temperature-corrected pH values in the cal-
ibration. The table below lists the standard buffers the controller recognizes. The transmitter also recognizes sev-
eral technical buffers: Merck, Ingold, and DIN 19267. Temperature-pH data stored in the controller are valid
between at least 0 and 60°C.
70
MODEL 5081-A SECTION 13.0
CALIBRATION - pH
CALIBRATE
QI!Dbm 5. Press NEXT until the PH CAL submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
CALIBRATE
DBM!cG2 7. The CAL bF1 prompt appears. Rinse the sensor and place it in the first
EXIT NEXT ENTER buffer. Be sure the glass bulb and reference junction are completely
submerged. Swirl the sensor. The main display will show the pH of the
buffer based on the previous calibration. Press ENTER.
CALIBRATE
cg2 8. bF1 flashes until the pH reading meets the stability criteria programmed
EXIT ENTER in Section 7.10.
CALIBRATE
cg2! 5/12 9. Once the reading is stable, the display changes to look like the figure
at left. The flashing number is the nominal pH, that is, the pH of the
EXIT ENTER
buffer at 25°C. If the flashing number does not match the nominal pH,
press é or ê until the correct pH appears. Press ENTER to save.
CALIBRATE
DBM!cG3 10. The CAL bF2 prompt appears. Remove the sensor from the first buffer.
EXIT NEXT ENTER Rinse the sensor and place it in the second buffer. Be sure the glass
bulb and reference junction are completely submerged. Swirl the sen-
sor. The display will show the pH of the buffer based on the previous
calibration. Press ENTER.
CALIBRATE
cG3 11. bF2 flashes until the pH reading meets the stability criteria programmed
EXIT NEXT ENTER in Section 7.10.
CALIBRATE
cg3! 21/11 12. Once the reading is stable, the display changes to look like the figure
at left. The flashing number is the nominal pH, that is, the pH of the
EXIT ENTER
buffer at 25°C. If the flashing number does not match the nominal pH,
press é or ê until the correct pH appears. Press ENTER to save.
71
MODEL 5081-A SECTION 13.0
CALIBRATION - pH
CALIBRATE
QI!DBM 5. Press NEXT until the PH CAL prompt appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
CALIBRATE
DBM!cg2 7. The CAL bF1 prompt appears. Rinse the sensor and the thermometer
and place them in the first buffer. Be sure the glass bulb and reference
EXIT NEXT ENTER
junction are completely submerged. Swirl the sensor. The main display
will show the pH of the buffer based on the previous calibration. Press
ENTER.
CALIBRATE
cg2! 15/12 8. Wait until the pH reading in the main display is constant. Use the arrow
EXIT ENTER keys to change the flashing display to the value of the buffer at the
measurement temperature. Press ENTER.
CALIBRATE
DBM!cg3 9. The CAL bF2 prompt appears. Rinse the sensor and the thermometer
and place them in the second buffer. Be sure the glass bulb and refer-
EXIT NEXT ENTER
ence junction are completely submerged. Swirl the sensor. The main
display will show the pH of the buffer based on the previous calibration.
Press ENTER.
CALIBRATE
cg3!! 21/11 10. Wait until the pH reading in the main display is constant. Use the arrow
EXIT ENTER
keys to change the flashing display to the value of the buffer at the
measurement temperature. Press ENTER.
72
MODEL 5081-A SECTION 13.0
CALIBRATION - pH
13.4 STANDARDIZATION
1. The pH measured by the transmitter can be changed to match the reading from a second or reference instru-
ment. The process of making the two readings agree is called standardization, or one-point calibration.
2. During standardization, the difference between the two pH values is converted to the equivalent voltage. The
voltage, called the reference offset, is added to all subsequent measured cell voltages before they are con-
verted to pH. If a sensor that has been calibrated with buffers is then standardized and placed back in a buffer,
the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset.
3. Install the sensor in the process liquid. Once readings are stable, measure the pH of the liquid using a refer-
ence instrument. Normally, it is acceptable to test a grab sample. Because the pH of the process liquid may
change if the temperature changes, measure the pH immediately after taking the grab sample. For poorly
buffered samples, it is best to determine the pH of a continuously flowing sample from a point as close as pos-
sible to the process sensor.
QI!DBM 5. Press NEXT until the PH CAL submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
Tue!QI 6. Press NEXT until the Std PH submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
7. Be sure the process pH and temperature are stable. Measure the pH of
Tue 18/11 the process liquid using the reference instrument. Use the arrow keys
EXIT ENTER to change the flashing display to match the reading from the reference
meter. Press ENTER to save.
73
MODEL 5081-A SECTION 13.0
CALIBRATION - pH
298
slope at 25°C = (slope at t°C)
t°C + 273
Changing the slope overrides the slope determined from the previous buffer calibration.
CALIBRATE
CALIBRATE
4. Press NEXT until PH SLOPE appears. Press ENTER.
qI!tmPqf
EXIT NEXT ENTER
CALIBRATE
5. The SLOPE prompt appears. Use the arrow keys to change the flash-
tmPqf!! 6:/27 ing display to the desired slope. Press ENTER to save.
EXIT ENTER
74
MODEL 5081-A SECTION 14.0
CALIBRATION — CURRENT OUTPUT
SECTION 14.0
CALIBRATION — CURRENT OUTPUT
14.1 GENERAL
Although the transmitter outputs are calibrated at the factory, they can be trimmed in the field to match the read-
ing from a standard current meter. Both the 4 mA and the 20 mA outputs can be trimmed. During output calibra-
tion the transmitter is in Hold. The output current will go to the value programmed in Section 7.3.
14.2 PROCEDURE
1. Wire an accurate milliammeter as shown in Figure 14-1.
PVuqVu!DBM 3. Press NEXT until the OUtPUt CAL submenu appears. Press ENTER.
EXIT NEXT ENTER
CALIBRATE
Dvs 5/111 4. Use the arrow keys to change the display to match the reading from the
EXIT NEXT ENTER
milliammeter. Press ENTER.
CALIBRATE
Dvs 31/11 5. Use the arrow keys to change the display to match the reading from the
EXIT NEXT ENTER milliammeter. Press ENTER. Press RESET to return to the main dis-
play.
75
MODEL 5081-A SECTION 15.0
DIAGNOSTICS
SECTION 15.0
DIAGNOSTICS
15.1 GENERAL
The 5081-A transmitter can display diagnostic information that is useful in troubleshooting. The diagnostics avail-
able depend on the measurement being made. To read diagnostic information, go to the main display and press
DIAG on the infrared remote controller. Press NEXT until the mnemonic for the desired information appears. Refer
to the appropriate section below for more information.
TYPE O3 or tCL Transmitter is measuring ozone (or total chlorine). Press NEXT to view diagnostics.
SEnSor Cur Press ENTER to display raw current from sensor (note units).
SEnSitvtY Press ENTER to display sensitivity. Sensitivity is calculated during calibration. It is the
measured current divided by concentration.
O CurrEnt Press ENTER to display the zero current measured during calibration (note units).
5081-A-Ht This is the model number. Press ENTER to display the software revision (SFtr) level.
Press NEXT to show the hardware revision (HArdr) level.
FAULtS Press ENTER to scroll through existing fault messages.
76
MODEL 5081-A SECTION 15.0
DIAGNOSTICS
TYPE FCL Transmitter is measuring free chlorine. Press NEXT to view diagnostics.
SEnSor Cur Press ENTER to display raw current from sensor (note units).
SEnSitvtY Press ENTER to display sensitivity. Sensitivity is calculated during calibration. It is the measured
current divided by concentration.
O CurrEnt Press ENTER to display the zero current measured during calibration (note units).
PH Press ENTER to view pH diagnostics. Press NEXT to skip pH diagnostics.
InPut Current pH sensor input voltage in millivolts.
SLOPE Sensor slope in millivolts per unit pH. Slope is calculated during buffer calibration. See Figure 13.1.
OFFSt Sensor voltage in millivolts in pH 7 buffer.
GIMP Glass impedance in MΩ.
5081-A-Ht This is the model number. Press ENTER to display the software revision (SFtr) level. Press
NEXT to show the hardware revision (HArdr) level.
FAULtS Press ENTER to scroll through existing fault messages.
77
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
SECTION 16.0
TROUBLESHOOTING
16.1 WARNING AND FAULT MESSAGES
16.2 TROUBLESHOOTING WHEN A WARNING OR FAULT MESSAGE IS SHOWING
16.3 TEMPERATURE MEASUREMENT AND CALIBRATION PROBLEMS
16.4 OXYGEN MEASUREMENT AND CALIBRATION PROBLEMS
16.5 FREE CHLORINE MEASUREMENT AND CALIBRATION PROBLEMS
16.6 TOTAL CHLORINE MEASUREMENT AND CALIBRATION PROBLEMS
16.7 OZONE MEASUREMENT AND CALIBRATION PROBLEMS
16.8 pH MEASUREMENT AND CALIBRATION PROBLEMS
16.9 SIMULATING INPUT CURRENTS - DISSOLVED OXYGEN
16.10 SIMULATING INPUT CURRENTS - CHLORINE AND OZONE
16.11 SIMULATING INPUTS - pH
16.12 SIMULATING TEMPERATURE
16.13 MEASURING REFERENCE VOLTAGE
78
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
79
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.2.6 bAd rtd, TEMP HI, TEMP LO, and rtd OPEn
These messages usually mean that the RTD (or thermistor in the case of the Hx438 and GX448 sensors) is open or short-
ed or there is an open or short in the connecting wiring.
1. Verify all wiring connections, including wiring in a junction box if one is being used.
2. Disconnect the RTD IN, RTD SENSE, and RTD RETURN leads or the thermistor leads at the transmitter. Be sure to
note the color of the wire and where it was attached. Measure the resistance between the RTD IN and RETURN leads.
For a thermistor, measure the resistance between the two leads. The resistance should be close to the value in the
table in Section 16.12. If the temperature element is open or shorted, replace the sensor. In the meantime, use man-
ual temperature compensation.
3. For oxygen measurements using the HX438, the Gx448, or other steam-sterilizable sensor using a 22kNTC, the TEMP
HI error will appear if the controller was not properly configured. See Section 7.4.3 or 7.4.4.
80
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.2.8 pH In
pH In means the voltage from the pH measuring cell is too large.
1. Verify all wiring connections, including connections in a junction box.
2. Check that the pH sensor is completely submerged in the process liquid.
3. Check the pH sensor for cleanliness. If the sensor looks fouled or dirty, clean it. Refer to the sensor instruction manu-
al for cleaning procedures.
4. Replace the sensor.
81
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
82
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.2.14 AdC
AdC means the analog to digital converter has failed.
1. Verify that sensor wiring is correct and connections are tight. Be sure to check connections at the junction box if one
is being used. See Section 3.0.
2. Disconnect sensor(s) and simulate temperature and sensor input.
To simulate See Section
Dissolved oxygen 16.9
Ozone or chlorine 16.10
pH 16.11
Temperature 16.12
3. If the transmitter does not respond to simulated signals, the analog PCB has probably failed. Call the factory for assis-
tance.
83
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.4.1 Zero current is substantially greater than the value in Section 9.2.
1. Is the sensor properly wired to the transmitter? See Section 3.0.
2. Is the membrane completely covered with zero solution and are air bubbles not trapped against the membrane? Swirl
and tap the sensor to release air bubbles.
3. Is the zero solution fresh and properly made? Zero the sensor in a solution of 5% sodium sulfite in water. Prepare the
solution immediately before use. It has a shelf life of only a few days.
4. If the sensor is being zeroed with nitrogen gas, verify that the nitrogen is oxygen-free and the flow is adequate to pre-
vent back-diffusion of air into the chamber.
5. The major contributor to the zero current is dissolved oxygen in the electrolyte solution inside the sensor. A long zero-
ing period usually means that an air bubble is trapped in the electrolyte. To ensure the 499ADO or 499A TrDO sensor
contains no air bubbles, carefully follow the procedure in the sensor manual for filling the sensor. If the electrolyte solu-
tion has just been replaced, allow several hours for the zero current to stabilize. On rare occasions, the sensor may
require as long as overnight to zero.
6. Check the membrane for damage and replace the membrane if necessary
84
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.4.3 Sensor current during air calibration is substantially different from the value in Section 9.3.
1. Is the sensor properly wired to the transmitter? See Section 3.0. Verify that all connections are tight.
2. Is the membrane dry? The membrane must be dry during air calibration. A droplet of water on the membrane during
air calibration will lower the sensor current and cause an inaccurate calibration.
3. If the sensor current in air is very low and the sensor is new, either the electrolyte flow has stopped or the membrane
is torn or loose. For instructions on how to restart electrolyte flow see Section 16.4.2 or refer to the sensor instruction
manual. To replace a torn membrane, refer to the sensor instruction manual.
4. Is the temperature low? Sensor current is a strong function of temperature. The sensor current decreases about 3%
for every °C drop in temperature.
5. Is the membrane fouled or coated? A dirty membrane inhibits diffusion of oxygen through the membrane, reducing the
sensor current. Clean the membrane by rinsing it with a stream of water from a wash bottle or by gently wiping the
membrane with a soft tissue. If cleaning the membrane does not improve the sensor response, replace the membrane
and electrolyte solution. If necessary, polish the cathode. See the sensor instruction sheet for more information.
16.4.4 Process and standard instrument readings during in-process calibration are substantially different.
This error warning appears if the current process reading and the reading it is being changed to, ie, the reading from the
standard instrument, are appreciably different.
1. Is the standard instrument properly zeroed and calibrated?
2. Are the standard and process sensor measuring the same sample? Place the sensors as close together as possible.
3. Is the process sensor working properly? Check the response of the process sensor in air and in sodium sulfite solu-
tion.
85
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
86
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.5.3 Sensor current during calibration is substantially less than 250 nA/ppm at 25°C and pH 7.
1. Is the temperature low or is the pH high? Sensor current is a strong function of pH and temperature. The sensor cur-
rent decreases about 3% for every °C drop in temperature. Sensor current also decreases as pH increases. Above
pH 7, a 0.1 unit increase in pH lowers the current about 5%.
2. Sensor current depends on the rate of sample flow past the sensor tip. If the flow is too low, chlorine readings will be
low. Refer to the sensor instruction sheet for recommended sample flows.
3. Low current can be caused by lack of electrolyte flow to the cathode and membrane. See step 4 in Section 16.5.2.
4. Is the membrane fouled or coated? A dirty membrane inhibits diffusion of free chlorine through the membrane, reduc-
ing the sensor current and increasing the response time. Clean the membrane by rinsing it with a stream of water from
a wash bottle. DO NOT use a membrane or tissue to wipe the membrane.
5. If cleaning the membrane does not improve the sensor response, replace the membrane and electrolyte solution. If
necessary, polish the cathode. See the sensor instruction sheet for details.
87
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
88
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
16.7.3 Sensor current during calibration is substantially less than 350 nA/ppm at 25°C.
1. Sensor current is a strong function of temperature. The sensor current decreases about 3% for every °C drop in tem-
perature.
2. Sensor current depends on the rate of sample flow past the sensor tip. If the flow is too low, ozone readings will be
low. Refer to the sensor instruction sheet for recommended sample flows.
3. Low current can be caused by lack of electrolyte flow to the cathode and membrane. See step 3 in Section 16.7.2.
4. Is the membrane fouled or coated? A dirty membrane inhibits diffusion of ozone through the membrane, reducing the
sensor current and increasing the response time. Clean the membrane by rinsing it with a stream of water from a wash
bottle or gently wipe the membrane with a soft tissue.
If cleaning the membrane does not improve the sensor response, replace the membrane and electrolyte solution. If
necessary, polish the cathode. See the sensor instruction sheet for details.
89
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
90
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
91
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
92
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
voltage (mV)
current (µA) =
resistance (kΩ) FIGURE 16-1. Simulate dissolved oxygen.
C. Note the sensor current. It should be close to the value in the table. The actual value depends on the voltage of the
battery. To view the sensor current, go to the main display and press DIAG. Then, press NEXT. SEnSor Cur will appear
in the display. Press ENTER. The display will show the sensor current. Note the units: uA is microamps: nA is
nanoamps.
D. Change the decade box resistance and verify that the correct
current is shown. Calculate the current from the equation:
The voltage of a fresh 1.5 volt battery is about 1.6 volt (1600 mV).
94
MODEL 5081-A SECTION 16.0
TROUBLESHOOTING
95
MODEL 5081-A SECTION 16.0
MAINTENANCE
96
MODEL 5081-A SECTION 17.0
MAINTENANCE
SECTION 17.0
MAINTENANCE
17.1 OVERVIEW
This section gives general procedures for routine maintenance of the 5081-A transmitter. The transmitter needs
almost no routine maintenance.
97
MODEL 5081-A SECTION 17.0
MAINTENANCE
Location in Shipping
Figure 17-1 PN Description Weight
1 23992-00 PCB stack consisting of the CPU (part 3) and analog (part 4) boards, 1 lb/0.5 kg
display board is not included, CPU and analog boards are factory-
calibrated as a unit and cannot be ordered separately
2 23638-01 LCD display PCB 1 lb/0.5 kg
5 33337-02 Terminal block 1 lb/0.5 kg
6 23593-01 Enclosure cover, front with glass window 3 lb/1.5 kg
7 33360-00 Enclosure, center housing 4 lb/1.5 kg
8 33362-00 Enclosure cover, rear 3 lb/1.0 kg
9 6560135 Desiccant in bag, one each 1 lb/0.5 kg
10 9550187 O-ring (2-252), one, front and rear covers each require an O-ring 1 lb/0.5 kg
12 note Screw, 8-32 x 0.5 inch, for attaching terminal block to center housing *
13 note Screw, 8-32 x 1.75 inch, for attaching circuit board stack to center *
housing
14 33342-00 Cover lock 1 lb/0.5 kg
15 33343-00 Locking bracket nut 1 lb/0.5 kg
16 note Screw, 10-24 x 0.38 inch, for attaching cover lock and locking bracket *
nut to center housing
NOTE: For information only. Screws cannot be purchased from Rosemount Analytical.
* Weights are rounded up to the nearest whole pound or 0.5 kg.
98
MODEL 5081-A SECTION 18.0
RETURN OF MATERIAL
SECTION 18.0
RETURN OF MATERIAL
99
MODEL 5081-A-HT APPENDIX A
APPENDIX A
BAROMETRIC PRESSURE AS A FUNCTION OF ALTITUDE
The table shows how barometric pressure changes with altitude. Pressure values do not take into account
humidity and weather fronts.
100
MODEL 5081-A-HT APPENDIX B
APPENDIX B
Model 4000 Percent Oxygen Sensor for Measurement in Gas
APPLICATIONS
The Model 4000 sensor is used to monitor percent and calculates the equilibrium solubility of atmospher-
oxygen concentrations in gaseous streams. The most ic oxygen at the prevailing temperature and pressure.
common use is to monitor oxygen headspace concen- Replacing the membrane requires no special tools or
tration for nitrogen blanketing applications. The sen- fixtures. To replenish the electrolyte solution, unscrew
sors are also used in other applications where oxygen the fill plug, add the reagent, and replace the plug.
levels are controlled and monitored.
TYPES OF SENSORS:
FEATURES Rechargeable with a fast response flow assembly —
The Model 4000 Series Percent Oxygen sensor is a allows minimum volume gas flow that permits mount-
membrane-covered amperometric sensor. The mem- ing sensor in a flowing gas stream. Sample is supplied
brane consists of a gas permeable Teflon ®1 mem- at slightly above atmospheric pressure, flows through
brane stretched tightly over a gold cathode. A silver the assembly, and discharges to atmospheric pres-
anode and an electrolyte solution complete the inter- sure. Internal volume is low to minimize sensor
nal circuit. The sensor body is constructed of response time. Refer to Figure 1 for mounting instruc-
Ryton®2, which can withstand exposure to hydrocar- tions.
bons and other corrosive chemicals.
Rechargeable In-line flow — In line pressure com-
Gas permeates through the membrane, and the oxy- pensated flow assembly permits mounting the sensor
gen in the sample is reduced at the cathode. A volt- in a variable pressure gas stream at pressures up to
age is applied across the cathode and anode, 50 psig. This may or may not include a gland on the
generating an electrical current that is directly propor- sensor body. Refer to Figures 2, 3, and 4 for mounting
tional to the oxygen concentration in the sample. instructions.
Since the rate of oxygen diffusion through the mem-
brane is temperature dependent, the sensor response
must be corrected for permeability caused by temper- *30%
RYTON is resistant to 30% sulfuric acid, 85% phosphoric acid,
sodium hydroxide, gasoline, aliphatic alcohols, esters, ethers,
ature. A Pt 100 RTD in the sensor accurately meas- and ketones, as well as to aromatic amines. It is not particularly
ures temperature, and the analyzer automatically suited for service in strong oxidizing agents, aliphatic amines, chlori-
performs the correction. nated hydrocarbons, or aromatic nitrites, aldehydes, and nitro com-
pounds.
The sensor is easy to maintain. For calibration simply
expose the sensor to ambient air and press the air
1
calibration button. The analyzer measures the baro- Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
2
metric pressure using an on-board pressure sensor Ryton is a registered trademark of Chevron Phillips Chemical Company LP.
101
MODEL 4000 SPECIFICATIONS
SPECIFICATIONS
Range: 0-25% Oxygen
Linearity: For constant sample temperature after correction for sensor zero offset: ±1% of full scale
Repeatability: ± 0.1% of range
Stability:
Zero drift ± 0.25% O2 per week @ 25°C;
Span drift ± 0.25% O2 per week @ 25°C
Response Time: 90% in 20 seconds for a step change, using an equilibrated sensor at 25°C
Sample Pressure: 0 to 50 PSIG
Sample Temperature: 32 to 110°F
Humidity: up to 95% non-condensing
Wetted materials: Ryton®1, Teflon®2
1
Ryton is a registered trademark of Chevron Phillips Chemical Company LP.
2
Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
102
MODEL 4000 INSTALLATION
CALIBRATION
Air calibration is recommended. Using a certified span gas is an option, but since the concentration of oxygen in
ambient air is close to 21% at sea level, this is the best solution. Refer to the instrument instruction manual for
details on how to access the calibration menu.
TROUBLESHOOTING
The most frequent fault is a progressive development of insensitivity of the sensor. If sensor calibration is slug-
gish, then the sensor should be recharged with new electrolyte and the membrane may need to be replaced.
Listed below are some common troubleshooting techniques
103
MODEL 4000 WIRING
104
MODEL 4000 PROCEDURE
105
106
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira-
tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90
days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man-
ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por-
tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit-
able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli-
gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per-
sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri-
od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT-
NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
The right people,
the right answers,
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
right now.
http://www.raihome.com