General Specifications
GS 01C25B01-01E
The high performance differential pressure transmitter EJX110A features single crystal silicon resonant sensor and is suitable to measure liquid, gas, or steam flow as well as liquid level, density and pressure. EJX110A outputs a 4 to 20 mA DC signal corresponding to the measured differential pressure. Its highly accurate and stable sensor can also measure the static pressure which can be shown on the integral indicator or remotely monitored via BRAIN or HART communications. Other key features include quick response, remote set-up using communications, diagnostics and optional status output for pressure high/low alarm. The multi-sensing technology provides the advanced diagnostic function to detect such abnormalities as an impulse line blockage or heat trace breakage. FOUNDATION Fieldbus protocol type is also available. All EJX series models in their standard configuration, with the exception of the Fieldbus type, are certified by TV as complying with SIL 2 for safety requirement.
0.04% of Span
(0.005 0.0049 URL/span)% of Span 70 kPa (280 inH2O) 500 kPa (2000 inH2O)
T08E.EPS
STANDARD SPECIFICATIONS
Refer to GS 01C25T02-01E for Fieldbus communication type marked with .
V 0.04% of Span (0.005+0.00125 URL/span)% of Span 500 kPa (2000 inH2O) 14 MPa (2000 psi)
T09E.EPS
Reference accuracy
mmH2O(/D4)
X X
span span
kPa
inH2O(/D1)
mbar(/D3) 1 to 100
X
URL (upper range limit)
Measurement span Reference accuracy X span X span X URL (upper range limit)
Span 0.4 to 40 0.1 to 10 L Range 10 to 10 40 to 40 Span 0.5 to 100 2 to 400 M Range 100 to 100 400 to 400 2.5 to 500 Span 10 to 2000 H Range 500 to 500 2000 to 2000 Span 0.07 to 14 MPa 10 to 2000 psi V Range 0.5 to 14 MPa 71 to 2000 psi
10 to 1000 100 to 100 1000 to 1000 50 to 10000 5 to 1000 1000 to 1000 10000 to 10000 25 to 5000 0.025 to 5 kgf/cm2 5000 to 5000 5 to 5 kgf/cm2 0.7 to 140 bar 0.7 to 140 kgf/cm2 5 to 140 bar 5 to 140 kgf/cm2
T01E.EPS
0.04% of Span
(0.015 0.005 URL/span)% of Span 2 kPa (8 inH2O) 10 kPa (40 inH2O)
T03E.EPS
PERFORMANCE SPECIFICATIONS
Zero-based calibrated span, linear output, wetted parts material code S and silicone oil, unless otherwise mentioned. For Fieldbus communication type, use calibrated range instead of span in the following specifications. Specification Conformance EJX series ensures specification conformance to at least 3 . Reference Accuracy of Calibrated Span (includes terminal-based linearity, hysteresis, and repeatability)
Measurement span Reference accuracy X URL (upper range limit) X X span span M
Square Root Output Accuracy The square root accuracy is a percent of flow span.
Output 50% or Greater 50% to Dropout point Accuracy Same as reference accuracy Reference accuracy 50 Square root output (%)
T03E.EPS
0.04% of Span
(0.005 0.0035 URL/span)% of Span 10 kPa (40 inH2O) 100 kPa (400 inH2O)
T02E.EPS
Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Phone: 81-422-52-5690 Fax.: 81-422-52-2018
<<Contents>> <<Index>>
2
Vibration Effects Amplifier housing code 1: Less than 0.1% of URL when tested per the requirements of IEC60770-1 field or pipeline with high vibration level (10-60 Hz, 0.21 mm peak to peak displacement/60-2000 Hz 3 g) Amplifier housing code 2: Less than 0.1% of URL when tested per the requirements of IEC60770-1 field with general application or pipeline with low vibration level (10-60 Hz 0.15mm peak to peak displacement /60-500 Hz 2g) Mounting Position Effects Rotation in diaphragm plane has no effect. Tilting up to 90 degree will cause zero shift up to 0.4 kPa (1.6 inH2O) which can be corrected by the zero adjustment. Response Time (Differential pressure) 90 msec for Wetted Parts material code S. 150 msec for Wetted Parts Material Code H, M, T, A, B and D. When amplifier damping is set to zero and including dead time of 45 msec (nominal) Static Pressure Signal Range and Accuracy (For monitoring via communication or on indicator. Includes terminal-based linearity, hysteresis, and repeatability) Range Upper Range Value and Lower Range Value of the statice pressure can be set in the range between 0 and Maximum Working Pressure(MWP). The upper range value must be greater than the lower range value. Minimum setting span is 0.5 MPa(73 psi). Measuring either the pressure of high pressure side or low pressure side is user-selectable. Accuracy Absolute Pressure 1MPa or higher : 0.2% of span Less than 1 MPa: 0.2% (1 MPa/span) of span Gauge Pressure Reference Gauge pressure reference is 1013 hPa (1 atm)
Note : Gauge pressure variable is based on the above fixed reference and thus subject to be affected by the change of atomospheric pressure.
E1: Reference Accuracy of Calibrated Span E2: Ambient Temperature Effects per 28 C change E3: Static Span Effects per 6.9 MPa change
Total Accuracy (M capsule) 0.12% of Span @1:1 Rangedown 0.25% of Span @ 5:1 Rangedown
Total accuracy is a comprehensive measure of transmitter total performance, covering all major factors in actual installation, that cause errors in measurement. As a standard measure, YOKOGAWA uses this to evaluate transmitter performance.
Total Accuracy
E1: Reference Accuracy of Calibrated Span E2: Ambient Temperature Effects per 28 C change E3: Static Span Effects per 6.9 MPa change E4: Static Zero Effects per 6.9 MPa change E5: Overpressure Effects upto overpressure 25MPa
Not only a day-to-day changes in temperature can affect the measurement and lead to unnoticed errors; fluctuaion of line pressure, incorrect operation of three/five valve manifold leading to over-pressure events, and other phenomena can have the similar result. Total Accuracy factors in such changes and errors and provides much comprehensive and practical determination of how a transmitter will perform under actual plant operation.
Static Pressure Effects per 6.9 MPa (1000 psi) Change Span Effects L, M, H and V capsules 0.075% of span Effect on Zero
Capsule M H, V L Effect 0.02% URL 0.028% URL 0.05% URL
FUNCTIONAL SPECIFICATIONS
Output Two wire 4 to 20 mA DC output with digital communications, linear or square root programmable. BRAIN or HART FSK protocol are superimposed on the 4 to 20 mA signal. Output range: 3.6 mA to 21.6 mA Output limits conforming to NAMUR NE43 can be pre-set by option code C2 or C3. Failure Alarm (Output signal code D and E) Analog output status at CPU failure and hardware error; Up-scale: 110%, 21.6 mA DC or more (standard) Down-scale: 5%, 3.2 mA DC or less Analog output status at process abnormality (Option code /DG6); The result of process abnormality detected by the advanced diagnostic function can be reflected to an analog alert status. The following three setting modes are available.
Overpressure Effects Overpressure condition: up to maximum working pressure M, H and V capsules 0.03% of URL Stability (All normal operating condition, including overpressure effects) M, H and V capsules 0.1% of URL per 10 years Power Supply Effects (Output signal code D and E) 0.005 % per Volt (from 21.6 to 32 V DC, 350 )
GS 01C25B01-01E
<<Contents>> <<Index>>
Mode Burnout Standard /C1 Option Code /C2 /C3 110 %, 21.6mA or more -2.5%, 3.6mA or less -1.25%, 3.8mA or less 103.1%, 20.5mA or more
T11E.EPS
3
Fall back Off
Signal Characterizer (Output signal code D and E) User-configurable 10-segment signal characterizer for 4 to 20 mA output. Status Output (optional, output signal code D and E) One transistor contact output (sink type) to output the status of user configurable high/low alarm for differential pressure/static pressure. Contact rating : 10.5 to 30 V DC, 120 mA DC max. Refer to Terminal Configuration and Wiring Example for Analog Output and Status Output. SIL Certification EJX series transmitters except Fieldbus communication type are certified by TV in compliance with the following standards; IEC 61508: 2000; Part1 to Part 7 Functional Safety of Electrical/electronic/programmable electronic related systems; SIL 2 capability for single transmitter use, SIL 3 capability for dual transmitter use.
Holds to a specified value within the output range from 3.6mA to 21.6mA
Damping Time Constant (1st order) Amplifier damping time constant is adjustable from 0.00 to 100.00 sec by software and added to response time.
Note: For BRAIN protocol type, when amplifier software damping is set to less than 0.5 sec, communication may occasionally be unavailble during the operation, especially while output changes dynamically. The default setting of damping ensures stable communication.
Update Period Differential pressure: 45 msec Static pressure: 360 msec Zero Adjustment Limits Zero can be fully elevated or suppressed, within the lower and upper range limits of the capsule. External Zero Adjustment External zero is continuously adjustable with 0.01% incremental resolution of span. Re-range can be done locally using the digital indicator with rangesetting switch. Integral Indicator (LCD display, optional) 5-digit numerical display, 6-digit unit display and bar graph. The indicator is configurable to display one or up to four of the following variables periodically.; Measured differential pressure, differential pressure in %, scaled differential pressure, measured static pressure. See also Factory Setting. Burst Pressure Limits 69 MPa (10,000 psi) Self Diagnostics CPU failure, hardware failure, configuration error, and over-range error for differential pressure, static pressure and capsule temperature. User-configurable process high/low alarm for differential pressure and static pressure is also available, and its status can be output when optional status output is specified. Advanced Diagnostics (optional) Applicable for Output signal code E and F. Impulse line blockage detection The impulse line condition can be calculated and detected by extracting the fluctuation component from the differential pressure and static pressure signals. The EJX110A detects the impulse line abnormality particularly which side of impulse line is plugged. Heat trace monitoring The change of the flange temperature calculated by using the two temperature sensors built in the EJX enables to detect the heat trace breakage or the abnormal temperature due to the failure.
0 (32)
40 (104)
80 (176)
120 (248)
F01E.EPS
Process temperature C ( F)
GS 01C25B01-01E
<<Contents>> <<Index>> Supply & Load Requirements (Output signal code D and E. Optional features or approval codes may affect electrical requirements.) With 24 V DC supply, up to a 550 load can be used. See graph below.
4
European Pressure Equipment Directive 97/23/EC Sound Engineering Practice (for all capsules) With option code /PE3 (for M, H and V capsules and wetted parts material code S.) Category III, Module H, Type of Equipment: Pressure Accessory-Vessel, Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2
600 R= External load resistance 250 R( ) E-10.5 0.0244 Digital Communication range BRAIN and HART
PHYSICAL SPECIFICATIONS
Wetted Parts Materials Diaphragm, Cover Flange, Process Connector, Capsule Gasket, and Vent/Drain Plug Refer to MODEL AND SUFFIX CODES. Process Connector Gasket PTFE Teflon Fluorinated rubber for option code N2 and N3 Non-wetted Parts Materials Bolting B7 carbon steel, 316L SST or 660 SST Housing Low copper cast aluminum alloy with polyurethane, mint-green paint (Munsell 5.6BG 3.3/2.9 or its equivalent), or ASTM CF-8M Stainless Steel Degrees of Protection IP67, NEMA4X Cover O-rings Buna-N, fluoro-rubber (optional) Name plate and tag 304 SST, 316 SST (optional) Fill Fluid Silicone, fluorinated oil (optional) Weight [Installation code 7, 8 and 9] 2.8 kg (6.2 lb) for wetted parts material code S without integral indicator, mounting bracket, and process connector. Add 1.5 kg (3.3lb) for Amplifier housing code 2. Connections Refer to MODEL AND SUFFIX CODES. Process connection of cover flange: IEC61518
42
F02E.EPS
Figure 2. Relationship Between Power Supply Voltage and External Load Resistance
Supply Voltage 10.5 to 42 V DC for general use and flameproof type. 10.5 to 32 V DC for lightning protector (option code / A. ) 10.5 to 30 V DC for intrinsically safe, type n, or nonincendive. Minimum voltage limited at 16.6 V DC for digital communications, BRAIN and HART Load (Output signal code D and E) 0 to 1290 for operation 250 to 600 for digital communication Communication Requirements (Approval codes may affect electrical requirements.) BRAIN Communication Distance Up to 2 km (1.25 miles) when using CEV polyethylene-insulated PVC-sheathed cables. Communication distance varies depending on type of cable used. Load Capacitance 0.22 F or less Load Inductance 3.3 mH or less Input Impedance of communicating device 10 k or more at 2.4 kHz. , EMC Conformity Standards EN61326-1 Class A, Table2 (For use in industrial locations) EN61326-2-3
<<Contents>> <<Index>>
-D -E -F Measurement L span (capsule) M H V Wetted parts material *2 Process connections 0 1 2 3 4 5 Bolts and nuts material J G C Installation -7 -8 -9 -B -U Amplifier housing 1 2 Electrical connection 0 2 4 5 7 9 A C D D E N B D G J K M N
Integral indicator
Mounting bracket
Optional Codes
The marks indicate the most typical selection for each specification. *1: Not applicable for output signal code F. *2: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid can contaminate the users process fluids. Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material. *3: Not applicable for electrical connection code 0, 5, 7 and 9. *4: Material of a blind plug is aluminum alloy or 304 SST. *5: Only applicable for Wetted parts material code S.
GS 01C25B01-01E
<<Contents>> <<Index>>
Table. Wetted Parts Materials
Wetted parts material code S H M T A D B Cover flange and process connector ASTM CF-8M ASTM CF-8M
*1 *1
6
Capsule Hastelloy C-276 *2 (Diaphragm) F316L SST (Others) Hastelloy C-276 Monel Tantalum Hastelloy C-276 *2 Tantalum Monel
*2
Capsule gasket Teflon-coated 316L SST PTFE Teflon PTFE Teflon PTFE Teflon PTFE Teflon PTFE Teflon PTFE Teflon
Drain/Vent plug 316 SST 316 SST 316 SST 316 SST Hastelloy C-276 *2 Hastelloy C-276 *2 Monel
M04E.EPS
# # #
Monel equivalent
*4
*1: Cast version of 316 SST. Equivalent to SCS14A. *2: Hastelloy C-276 or ASTM N10276. *3: Indicated material is equivalent to ASTM CW-12MW. *4: Indicated material is equivalent to ASTM M35-2. The #marks indicate the construction materials conform to NACE material recommendations per MR0175/ISO15156. Please refer to the latest standards for details. Selected materials also conform to NACE MR0103.
Code
FF1
FS1
FU1 *1
CENELEC ATEX (KEMA) Flameproof Approval Applicable Standard: EN 60079-0, EN 60079-1, EN 61241-0, EN 61241-1 Certificate: KEMA 07ATEX0109 II 2G, 2D Exd IIC T4, T5, T6 Ex tD A21 IP6X T85, T100, T120 Degree of protection : IP66 and IP67 Amb. Temp. (Tamb) for gas-proof : T4 ; 50 to 75C (58 to 167F), T5; 50 to 80 C (58 to 176 F), T6; 50 to 75C (58 to 167F) *3 Max. process Temp.(Tp): T4; 120 C (248 F), T5; 100 C (212 F), T6; 85 C (185 F) Max. surface Temp. for dust-proof : T85C (Tamb: 40 to 40 C, Tp:85 C), T100C (Tamb: 40 to 60 C, Tp:100 C), T120C (Tamb: 40 to 80 C, Tp:120 C) *3 CENELEC ATEX CENELEC ATEX (KEMA) Intrinsically safe Approval *1*2 Applicable Standard: EN 50014, EN 50020, EN 50284, EN 50281-1-1 Certificate: KEMA 03ATEX1544X II 1G, 1D EEx ia IIC T4 Degree of protection : IP66 and IP67 Amb. Temp.(Tamb) for gas-proof: 50 to 60 C (58 to 140 F) *3 Maximum Process Temp.(Tp) for gas-proof :120 C Electrical data : Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH Max. surface Temp. for dust-proof : T85C (Tamb: 40 to 60 C, Tp:80 C), T100C (Tamb: 40 to 60 C, Tp:100 C), T120C (Tamb: 40 to 60 C, Tp:120 C) *3 Combined KF21, KS2 and Type n *1*2 Type n Applicable Standard: EN 60079-15 Referential Standard: IEC 60079-0, IEC 60079-11 II 3G Ex nL IIC T4, Amb. Temp.: 50 to 60 C (50 to 140 F) *3 Ui=30 V DC, Ci=10 nF, Li=0 mH *1: *2: *3: Applicable for Electrical connection code 2, 4, 7, 9, C and D. Not applicable for option code /AL. Lower limit of ambient temperature is 15C (5F) when /HE is specified.
KF21
KS2
KU21
T05E.EPS
GS 01C25B01-01E
<<Contents>> <<Index>>
Item *1 Description CSA Explosionproof Approval Certificate: 2014354 Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.0.5, C22.2 No.25, C22.2 No.30, C22.2 No.94, C22.2 No.60079-0, C22.2 No.60079-1, C22.2 No.61010-1-04 Explosion-proof for Class I, Groups B, C and D. Dustignition-proof for Class II/III, Groups E, F and G. When installed in Division 2, SEAL NOT REQUIRED Enclosure: TYPE 4X, Temp. Code: T6...T4 Ex d IIC T6...T4 Enclosure: IP66 and IP67 Max.Process Temp.: T4;120 C(248 F), T5;100 C(212 F), T6; 85 C(185 F) Amb.Temp.:50 to 75 C(58 to 167 F) for T4, 50 to 80 C(58 to 176 F) for T5, 50 to 75 C(58 to 167 F) for T6 *3 Process Sealing Certification Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01 No additional sealing required Primary seal failure annunciation: at the zero adjustment screw CSA Intrinsically safe Approval *1*2 Certificate: 1606623 [For CSA C22.2] Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.25, C22.2 No.94, C22.2 No.154, C22.2 No.213, C22.2 No.1010.1 Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, Groups E, F & G, Class III, Division 1, Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2, Groups E, F & G, Class III, Division 1 Enclosure: Type 4X, Temp. Code: T4 Amb. Temp.:50 to 60 C(58 to 140 F) *3 Electrical Parameters: [Intrinsically Safe] Vmax=30V, Imax=200mA, Pmax=0.9W, Ci=10nF, Li=0 [Nonincendive] Vmax=30V, Ci=10nF, Li=0 [For CSA E60079] Applicable Standard: CAN/CSA E60079-0, CAN/CSA E60079-11, CAN/CSA E60079-15, IEC 60529:2001-02 Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66 and IP67 Amb. Temp.:50 to 60 C(58 to 140 F) *3, Max. Process Temp.: 120 C(248 F) Electrical Parameters: [Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0 [Ex nL] Ui=30V, Ci=10nF, Li=0 Process Sealing Certification Dual Seal Certified by CSA to the requirement of ANSI/ISA 12.27.01 No additional sealing required Primary seal failure annunciation: at the zero adjustment screw Combined CF1 and CS1 *1*2 IECEx Flameproof Approval *1 Applicable Standard: IEC 60079-0:2004, IEC60079-1:2003 Certificate: IECEx CSA 07.0008 Flameproof for Zone 1, Ex d IIC T6...T4 Enclosure: IP66 and IP67 Max.Process Temp.: T4;120 C(248 F), T5;100 C(212 F), T6; 85 C(185 F) Amb.Temp.:50 to 75 C(58 to 167 F) for T4, 50 to 80 C(58 to 176 F) for T5, 50 to 75 C(58 to 167 F) for T6 *3 IECEx Intrinsically safe, type n and Flameproof Approval *1*2 Intrinsically safe and type n Applicable Standard: IEC 60079-0:2000, IEC 60079-11:1999, IEC 60079-15:2001 Certificate: IECEx CSA 05.0005 Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66 and IP67 Amb. Temp.:50 to 60 C(58 to 140 F) *3, Max. Process Temp.: 120 C(248 F) Electrical Parameters: [Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0 [Ex nL] Ui=30V,Ci=10nF, Li=0 Flameproof Applicable Standard: IEC 60079-0:2004, IEC60079-1:2003 Certificate: IECEx CSA 07.0008 Flameproof for Zone 1, Ex d IIC T6...T4 Enclosure: IP66 and IP67 Max.Process Temp.: T4;120 C(248 F), T5;100 C(212 F), T6; 85 C(185 F) Amb.Temp.:50 to 75 C(58 to 167 F) for T4, 50 to 80 C(58 to 176 F) for T5, 50 to 75 C(58 to 167 F) for T6 *3
7
Code
CF1
CS1
CU1
SF2
IECEx Scheme
SU2
Combination Combination of KU21, FU1 and CU1 *1*2 of approval *1: *2: *3: Applicable for Electrical connection code 2, 4, 7, 9, C and D. Not applicable for option code /AL. Lower limit of ambient temperature is 15C (5F) when /HE is specified.
V1U
T05Eb.EPS
GS 01C25B01-01E
<<Contents>> <<Index>>
OPTIONAL SPECIFICATIONS
Item Color change Coating change 316 SST exterior parts Fluoro-rubber O-ring Lightning protector Amplifier cover only*9 Description Code P PR X2 HC HE A
Painting
Amplifier cover and terminal cover, Munsell 7.5 R4/14 Anti-corrosion coating*1*9
316 SST name plate, tag plate and zero-adjustment screw*10 All O-rings of amplifier housing. Lower limit of ambient temperature: 15C (5F) Transmitter power supply voltage: 10.5 to 32 V DC ( 10.5 to 30 V DC for intrinsically safe type.) Allowable current: Max. 6000 A ( 1 40 s ), Repeating 1000 A ( 1 40 s ) 100 times Applicable Standards: IEC 61000-4-4, IEC 61000-4-5 Transistor output (sink type) Contact rating : 10.5 to 30 V DC, 120 mA DC(max ) Degrease cleansing treatment Low level : 0 to 2 V DC
Status output*2
AL K1 K2 K5 K6 K3 D1
Oil-prohibited use*3
Degrease cleansing treatment and fluorinated oilfilled capsule. Operating temperature 20 to 80 C( 4 to 176 F) Degrease cleansing and dehydrating treatment Degrease cleansing and dehydrating treatment with fluorinated oilfilled capsule. Operating temperature 20 to 80 C( 4 to 176 F) Fluorinated oil filled in capsule Operating temperature 20 to 80 C( P calibration (psi unit) 4 to 176 F) (See Table for Span and Range Limits.)
Calibration units*4
D3 D4 U1 GS A1 C1 C2 C3 N1 N2 N3 N4 CA CB
T06E.EPS
Total length: 119 mm (standard: 34 mm); Total length when combining with option code K1, K2, K5, and K6: 130 mm. Material: 316 SST Gold-plated 316L SST capsule gasket. Without drain and vent plugs. Surface of isolating diaphragms are gold plated, effective for hydrogen permeation. Overpressure effects for M, H and V capsules: 0.06% of URL Failure alarm down-scale : Output status at CPU failure and hardware error is 5%, 3.2mA DC or less. NAMUR NE43 Compliant Output signal limits : 3.8 mA to 20.5 mA Failure alarm down-scale : Output status at CPU failure and hardware error is 5%, 3.2 mA DC or less. Failure alarm up-scale : Output status at CPU failure and hardware error is 110%, 21.6 mA or more.
Body option*7
Side L H
N1 and Process connection, based on IEC61518 with female thread on both sides of cover flange, with blind kidney flanges on back. N2, and Material certificate for cover flange, diaphragm, capsule body, and blind kidney flange 304 SST tag plate wired onto transmitter (316 SST when /HC is specified) Data configuration for HART communication type
Software damping, Descriptor, Message
*1: *2 : *3: *4: *5: *6: *7: *8: *9: *10: *11: *12:
Not applicable with color change option. Check terminals cannot be used when this option code is specified. Not applicable for output signal code F. Applicable for Wetted parts material code S, M, H and T. The unit of MWP (Max. working pressure) on the name plate of a housing is the same unit as specified by option codes D1, D3, and D4. Applicable for vertical impulse piping type (Installation code 7) and Wetted parts material code S, H, M and T. Applicable for output signal codes D and E. The hardware error indicates faulty amplifier or capsule. Applicable for wetted parts material code S, M, H and T; process connection codes 3, 4, and 5; installation code 9; and mounting bracket code N. Process connection faces on the other side of zero adjustment screw. Also see Ordering Information. Not applicable for amplifier housing code 2. 316 or 316L SST. The specification is included in amplifier housing code 2. Applicable for wetted parts material code S; process connection code 0 and 5; and installation code 8 and 9. Not applicable for option code U1, N2, N3 and M11. No PTFE is used for wetted parts. Applicable for wetted parts material code S.
GS 01C25B01-01E
<<Contents>> <<Index>>
Item Advanced diagnostics*9 Multi-sensing process monitoring Impulse line blockage detection *10 Heat trace monitoring PED 97/23/EC Category III, Module H, Type of Equipment: Pressure Accessory-Vessel, Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2 Lower limit of ambient and process temperature: 29 C Cover flange *3 Cover flange, Process connector Pressure test/ Leak test certificate*5
*1: *2: *3: *4: *5: *6: *7: *8: *9: *10:
*4
9
Description Code DG6
PE3
Material certificate*2
Applicable for measurement span code M, H and V and wetted parts material code S. If compliance with category III is needed, specify this option code. Material traceability certification, per EN 10204 3.1B. Applicable for process connections codes 0 and 5. Applicable for process connections codes 1, 2, 3, and 4. The unit on the certificate is always Pa unit regardless of selection of option code D1, D3 or D4. Applicable for capsule code L. Also applicable for capsule M, H and V when combined with Wetted Parts Material code H, M, T, A, D and B. Applicable for capsule codes M, H and V when combined with Wetted Parts Material code S. Pure nitrogen gas is used for oil-prohibited use (option codes K1, K2, K5, and K6). Applicable only for output signal code -E. The change of pressure fluctuation is monitored and then detects the impulse line blockage. See TI 01C25A31-01E for detailed technical information required for using this function.
GS 01C25B01-01E
<<Contents>> <<Index>>
10
DIMENSIONS
Vertical Impulse Piping Type (INSTALLATION CODE '7') Wetted Parts Material code : S
Unit : mm (approx.inch)
52 (2.05)
54 (2.13)
97 (3.82)
6(0.24)
110(4.33) 12 (0.47) 39 (1.54) 2-inch pipe (O.D. 60.5 mm) Vent/Drain plugs
223(8.78)
97(3.82)
95(3.74)
67(2.64)
41 (1.61)
54 (2.13)
Horizontal Impulse Piping Type (INSTALLATION CODE '9') (For CODE '8', refer to the notes below.) Wetted Parts Material code : S
External indicator Conduit connection (optional) 54 (2.13) 89(3.50) 67 (2.64) 95(3.74) Conduit connection Integral indicator (optional) Shrouding bolt (for flame-proof type) 175(6.89) 78(3.07) Zero adjustment 110(4.33) 12 (0.47) 39 (1.54) 70 (2.76) 129(5.08) 117(4.61)*3
GS 01C25B01-01E
145(5.71)
6 (0.24)
47 (1.85)
* 1: * 2: * 3: * 4:
When installation code 8 is selected, high and low pressure side on above figure are reversed. (i.e. High pressure side is on the right side.) When option code K1, K2, K5 or K6 is selected, add 15mm(0.59 inch) to the value in the figure. When option code K1, K2, K5 or K6 is selected, add 30mm(1.18 inch) to the value in the figure. Not available when option code GS is selected.
70(2.76)
<<Contents>> <<Index>>
11
DIMENSIONS
Vertical Impulse Piping Type (INSTALLATION CODE '7') Wetted Parts Material code : H, M, T, A, B and D
Unit : mm (approx.inch)
6(0.24)
54 (2.13)
97 (3.82)
110(4.33) 12 (0.47) 39 (1.54) 2-inch pipe (O.D. 60.5 mm) Vent/Drain plugs
52 (2.05)
234(9.21)
102(4.02)
72(2.83)
Conduit connection Zero adjustment Ground terminal High pressure side 54 (2.13) Low pressure side
Horizontal Impulse Piping Type (INSTALLATION CODE '9') (For CODE '8', refer to the notes below.) Wetted Parts Material code : H, M, T, A, B and D
External indicator Conduit connection (optional) 54 (2.13) Conduit connection 94(3.70) 72(2.83) Integral indicator (optional) Shrouding bolt (for flame-proof type) 95(3.75) Zero adjustment 110(4.33) 9*4 (0.35) 39 (1.54)
78(3.07)
159(6.26)
194(7.64)
143(5.63)
Vent plug Drain plug
May 29, 2009-00
6 (0.24)
124(4.88)
Drain plug 46 (1.81) 125(4.92) Process connection Process connector (optional) 54 (2.13)
47 (1.85)
*1: When Installation code 8 is selected, high and low pressure side on above figure are reversed. (i.e. High pressure side is on the right side.) *2: When Option code K1, K2, K5, or K6 is selected, add 15 mm(0.59 inch) to the value in the figure. *3: When Option code K1, K2, K5, or K6 is selected, add 30 mm(1.18 inch) to the value in the figure. *4: 15 mm(0.59 inch) for right side high pressure.
128*3 (5.04)
GS 01C25B01-01E
70 (2.76)
70 (2.76) 148(5.83)*2
46 (1.81)
78(3.07)
95 (3.74)
12
95(3.74)
78(3.07)
145(5.71)
175(6.89)
Ground terminal
Electrical connection for code 5, 9, A, and D. Drain plug Vent plug Drain plug 41 (1.61) 115(4.53)
129(5.08)
Vent plug Drain plug
May 29, 2009-00
6 (0.24)
Shrouding bolt (for flame-proof type) Process connection Process connector (optional)
*1: When Option code K1, K2, K5, or K6 is selected, add 30 mm(1.18 inch) to the value.
6 (0.24)
54 (2.13)
95(3.74) 79 (3.11)
78(3.07)
70 (2.76)
Electrical connection for code 5 and 9. External indicator Conduit connection (optional)
70 (2.76)
54 (2.13) 117(4.61)
Low pressure side
219(8.62)
Ground terminal
Vent plug
Process connection
129(5.08)
GS 01C25B01-01E
109(4.29)
<<Contents>> <<Index>>
13
Terminal Wiring
Check meter connection hook*1*2 SUPPLY CHECK or ALARM + + + Power supply and output terminal External indicator (ammeter) terminal*1*2 or Status contact output terminal*2 (when /AL is specified) Ground terminal
Terminal Configuration
Communication terminals (BT200 etc.) connection hook
SUPPLY + SUPPLY
*1: When using an external indicator or check meter, the internal resistance must be 10 or less. A check meter or indicator cannot be connected when /AL option is specified. *2: Not available for fieldbus communication type.
F05E.EPS
Connection
Analog output
EJX electrical terminal
SUPPLY
Description
Distributor
+ +
+
250
24V DC
CHECK
Shielded cable
+ +
24V DC
Distributor
250
ALARM
External power supply 30V DC, 120mA max Use two-wire separately shielded cables.
Magnetic valve
AC power supply
F06E.EPS
GS 01C25B01-01E
<<Contents>> <<Index>>
14
< Factory Setting >
Tag number Software damping *1 Output mode As specified in order 2.00 sec or as specified in order Linear unless otherwise specified in order
Calibration range lower range value As specified in order Calibration range As specified in order upper range value Selected from mmH2O, mmH2O(68F), Calibration range mmAq*2, mmWG*2, mmHg, Pa, hPa*2, kPa, unit MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O, inH2O(68F), inHg, ftH2O, ftH2O(68F) or psi. (Only one unit can be specified.) Display setting Designated differential pressure value specified in order. (% or user scaled value.) Display mode 'Linear' or 'Square root' is also as specified in order. '0 to 25 MPa' for M and H capsule with wetted parts material S, and '0 to 16 MPa' for L capsule with wetted parts material S and all capsules with wetted parts material other than S, absolute value. Measuring high pressure side.
T07E.EPS
4. Display scale and units (for transmitters equipped with the integral indicator only) Specify either 0 to 100 % or Range and Unit for engineering units scale: Scale range can be specified with range limit specifications up to 5 digits (excluding any decimal point) for low or high range limits within the range of -32000 to 32000. Unit display consists of 6-digit, therefore, if the specified scaling unit excluding / is longer than 6-characters , the first 6 characters will be displayed on the unit display. 5. Tag Number (if required) For BRAIN communication type, specify upto 16 letters. The specified letters will be written in the amplifier memory and engraved on the tag plate. For HART communication type, specify software tag (up to 8 letters) to be written on the amplifier memory and Tag number(up to 16 letters) to be engraved on the tag plate seperately. 6. Other factory configurations (if required) Specifying option code CA or CB will allow further configuration at factory. Following are configurable items and setting range. [/CA : For HART communication type] 1) Descriptor(up to 16 characters) 2) Message (up to 30 characters) 3) Software damping in second (0.00 to 100.00) [/CB : For BRAIN communication type] 1) Software damping in second (0.00 to 100.00)
*1: To specify these items at factory, option code CA or CB is required. *2: Not available for HART protocol type.
All Rights Reserved. Copyright 2004, Yokogawa Electric Corporation Subject to change without notice.
GS 01C25B01-01E