Model A5000 Series Instruction Manual: Watanabe Electric Industry Co., LTD

This 3 sentence summary provides an overview of the instruction manual for the A5000 series measurement device: The instruction manual outlines the features and proper use of the A5000 series measurement device, which can measure various electrical parameters including voltage, current, resistance, temperature, and frequency. It describes the model codes and input/output units available. The manual also provides instructions on safely mounting the device, including cutting dimensions for the panel and ensuring the panel can securely hold the device.

Uploaded by

Vasile Toth
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
381 views12 pages

Model A5000 Series Instruction Manual: Watanabe Electric Industry Co., LTD

This 3 sentence summary provides an overview of the instruction manual for the A5000 series measurement device: The instruction manual outlines the features and proper use of the A5000 series measurement device, which can measure various electrical parameters including voltage, current, resistance, temperature, and frequency. It describes the model codes and input/output units available. The manual also provides instructions on safely mounting the device, including cutting dimensions for the panel and ensuring the panel can securely hold the device.

Uploaded by

Vasile Toth
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

Instruction Manual for A5000 Series 1/12

A 5 X X X - X X

MODEL A5000 SERIES Input unit


01: DC voltage measuring unit
INSTRUCTION MANUAL (range 11: ±99.99 mV)
02: DC voltage measuring unit
(range 12: ±999.9 mV ; range 13: ±9.999 V)
(range 14:±99.99 V ; range 15: ±600 V)
03: DC current measuring unit
(range 23: ±9.999 mA; range 24: ±99.99 mA)
(range 25: ±999.9 mA)
04: AC voltage measuring unit (average rms)
(range 11: 99.99 mV; range 12: 999.9 mV)
(range 13: 9.999 V)
05: AC voltage measuring unit (average rms)
! Caution (range 14: 99.99 V; range 15: 600 V)
(1) Do not apply a voltage or current exceeding the maximum allowable 06: AC voltage measuring unit (true rms)
value; otherwise, it may damage the equipment. (range 11: 99.99 mV; range 12: 999.9 mV)
(2) Use a power voltage within the operation range; otherwise, it may result (range 13: 9.999 V)
in a fire, electrical shock, or malfunction. 07: AC voltage measuring unit (true rms)
(3) The contents of this manual are subject to change without notice. (range 14: 99.99 V; range 15: 600 V)
(4) Although the contents of this manual have been prepared with extra 08: AC current measuring unit (average rms)
care, if you have any questions, or find errors or missing information, (range 23: 9.999 mA; range 24: 99.99 mA)
contact the sales agent from which you purchased the product or (range 25: 999.9 mA)
Watanabe Electric Industry Co.,Ltd. 09: AC current measuring unit (average rms)
(5) After reading this manual thoroughly, keep it in a convenient place for (range 26: 5 A)
future reference. 10: AC current measuring unit (true rms)
(6) The mark on a label shows the measurement tail range of the input (range 23: 9.999 mA; range 24: 99.99 mA)
specification of 8.1. clause. (range 25: 999.9 mA)
11: AC current measuring unit (true rms)
(range 26: 5 A)
1.Before Using the Product 12: Resistance measuring unit
Thank you for purchasing the A5000 series. This manual should be passed on to 13: Temperature measuring unit (TC)
the person who operates the product. Examine the product for damage caused by 14: Temperature measuring unit (RTD)
transportation or any other defects. If you find any damage or defects, contact the 15: Frequency measuring unit
(inputs: open collector, logic, and magnet)
sales agent from which you purchased the product or Watanabe Electric
16: Frequency measuring unit(input: 50 to 500 Vrms)
Industry Co., Ltd.
17: Strain gauze input unit (load cell)
1.1.Model Codes 18: Process signal measuring unit (4 to 20 mA or 1 to 5 V)

The model lineup of the A5000 series is shown below. Check that the model 2.Mounting the Product
code and specifications of your product match those you specified when
2.1.Dimensions for Cutting Panel
ordering.
Cut the panel for mounting according to the following dimensions.
A 5 X X X - X X
92 +0.8
-0
Power unit
1: 100 to 240 V AC ±10%
2: 9 to 60 V DC
45 +0.6
-0

Display unit

70 mm min.
1: Single display
2: Multi display

Output unit
0: None
1: Comparison
2: Analog
3: RS-232C
4: RS-485
5: Comparison and analog 120 mm min.
6: Comparison, analog, and RS-232C
7: Comparison, analog, and RS-485
2.2.Mounting the Product to the Panel
To mount the A5000 to the panel, remove its fittings and insert it through the
hole in the front of the panel. From the back of the panel, fix the product to the
panel with the fittings.

WATANABE ELECTRIC INDUSTRY CO., LTD. UU-33554v


2021.07
Instruction Manual for A5000 Series 2/12
3.3.3. DC Current Measuring Unit
! Caution
No. Name Description
(1) Mount the product to a panel that is strong enough to hold the product. Positive input terminal for range 23
1 23
If the panel is not strong enough or the product is not fixed tightly, it (±9.999 mA)
Positive input terminal for range 24
may fall down and cause injury. 1 2 3 4 5 2 24
(±99.99 mA)
(2) The A5000 does not have a power switch, and will thus be immediately 3 25
Positive input terminal for range 25
ready for operation upon connecting it to a power supply. (±999.9 mA)
4
(3) If the product is installed inside other equipment, provide sufficient heat LO Negative input terminal
dissipation to ensure that the temperature inside the equipment does 5
not exceed 50℃.

3. Terminal Arrangement 3.3.4. AC Voltage Measuring Unit (Ranges 11 to 13)


Comparison Serial No. Name Description
Analog output
output communication
Positive input terminal for ranges 11
1 11-12
(99.99 mV) and 12(999.9 mV)
1 2 3 2 13 Positive input terminal for range 13 (9.999 V)
12 13 14 15 16 17 18 19 20 22 24 25 26
3 LO Common input terminal

2 2
1 3 3.3.5. AC Voltage Measuring Unit (Ranges 14 and 15)
No. Name Description

1 14 Positive input terminal for range 14 (99.99 V)

1 2 3 2 15 Positive input terminal for range 15 (600 V)

1 3 LO Common input terminal


6 7 8 9 10
1

3.3.6. AC Current Measuring Unit (Ranges 23 to 25)


External Power
Input (the shape depends on the unit.) control connector No. Name Description

3.1.Power 1 23
Positive input terminal for range 23 (9.999
mA)
No. Name Description Positive input terminal for range 24 (99.99
1 2 3 4 5 2 24
mA)
Power terminal without polarity for both Positive input terminal for range 25 (999.9
10 POWER 3 25
DC and AC mA)
10 11
Power terminal without polarity for both
11 POWER 4
DC and AC
LO Negative input terminal
3.2.External Controls 5

No. Name Description


3.3.7. AC Current Measuring Unit (Range 26)
Control for hold function. Enabled when
6 7 8 9 6 HOLD short-circuited or at the same potential as
No. Name Description
COM.
Control for digital zero function. Enabled
7 DZ when short-circuited or at the same 1 HI Input terminal
potential as COM.
Control for peak hold function. Enabled 1 2 2 LO Input terminal
8 PH when short-circuited or at the same Applicable solderless terminals
potential as COM.
9 COM Common for all external control terminals.
min.

min.
mm

mm
5.8

5.8
3.3.Input Signals
3.3.1 DC Voltage Measuring Unit (Range 11)
3.3.8.Resistance Measuring Unit
No. Name Description

1 HI Positive input terminal


No. Name Description
1 2 3 2 NC Do not connect this terminal.
1 HI Input terminal for all ranges
3 LO Negative input terminal
1 2 3 4 5 2 LO Input terminal for all ranges
Constant current for four-wire resistance
3 +S
measurement(positive)
3.3.2.DC Voltage Measuring Unit (Range 12) Constant current for four-wire resistance
4 -S
measurement(negative)
Common terminal (grounding terminal for
No. Name Description 5 COM
input circuit)
Positive input terminal for range 12
1 12 ※Set to the 4-wire system when shipped.
(±999.9 mV)
Positive input terminal for range 13 When changing to the 2-wire system, locate
1 2 3 4 5 2 13 the ST1 socket on the resistance measurement
(±9.999 V)
Positive input terminal for range 14 unit to the “2” positions.
3 14
(±99.99 V)
Positive input terminal for range 15 (±600
4 15 4-wire
V)
5 LO Negative input terminal
2-wire
Instruction Manual for A5000 Series 3/12
3.3.9. Temperature Measuring Unit (TC) 3.5. Analog Output
No. Name Description No. Name Description

1 + Positive terminal for thermocouple 24 25 26 24 COM Common terminal for analog output

1 2 3 2 NC Do not connect this terminal. 25 A-OUT Current output terminal (4 to 20 mA)


Voltage output terminal (1 to 5 V, 0 to 1 V,
3 - Negative terminal for thermocouple 26 V-OUT
and 0 to 10 V)

3.3.10. Temperature Measuring Unit (RTD) 3.6. Serial Communication


20212223
No. Name Description No. Name Description
RS-232C: transmission;
20 RXD(+)
1 A Resistance sensor wire RS-485: Non-reverse output
(2)(3)(4)(5)
RS-232C: reception;
21 TXD(-)
1 2 3 2 B Resistance sensor wire RS-485: Reverse output

Connection of three-wire sensor 22 NC Do not connect this terminal.


3 C Elimination of wire resistance
Modular jack:
When A or B is disconnection, it is displayed as OL, and when RJ-14(6P 4C) 23 SG Common terminal for communications
C is disconnection, it is displayed as ----.
The analog output at the time of a burnout becomes + side
at the time of A or B disconnection, and is set to 0V or 1V,
! Caution
and 4mA at the time of C disconnection. (1) Use 12 to 28 AWG wire for the power, input (except for range 26),
external control, and comparison output connectors.
(2) Tighten the screws for the power, input (except for range 26), external
3.3.11. Frequency Measuring Unit (Open collector, logic, and
magnet) control, and comparison output connectors to a torque of 0.5 to 0.6 Nm.
(3) Use 16 to 28 AWG wire for the analog output connector.
No. Name Description (4) Tighten the screws of analog output connector to a torque of 0.22 to
0.25 Nm.
1 HI Positive input terminal
(5) Each wiring except a power supply is given as under full-length 30m.
2 LO Negative input terminal
If 30m is exceeded, it will become out of the scope of EN/IEC standard.
1 2 3 4 5

3 +15V Power output for sensor (positive) 4. Components and their Functions
4 0V Power output for sensor (negative) The front panel design of the A5000 series of unit meters differs depending on the
Common terminal (grounding terminal for display unit selected. The names and functions of each unit are as shown below.
5 COM
input circuit)
4.1.Multi-display Unit
3.3.12.Frequency Measuring Unit (500 Vrms) (2) (4)

No. Name Description (1)


(5)
1 HI Input terminal

1 2 3 2 NC Do not connect this terminal. (3)

3 LO Input terminal
(6) (7) (8) (9)
3.3.13. Strain Gauge Input Unit (Load cell) No. Name
Main Functions
During measurement During parameter setup
No. Name Description IIndicates information on the
(1) Main display Indicates the measured value.
parameter to be set.
1 +SIG Positive input terminal Indicates the result of judgment and
HI turns on if the measured value > HI
1 2 3 4 5 2 -SIG Negative input terminal judgment value.
Indicates the result of judgment and
Judgment
3 +EXC Power output for sensor (positive) (2) GO turns on if LO judgment value ≦ the
indicators
measured value ≦ HI judgment value.
4 -EXC Power output for sensor (negative) Indicates the result of judgment and
LO turns on if the measured value < LO
Common terminal (grounding terminal for
5 COM judgment value.
input circuit) ME Turns on if “digital zero backup” is on.
Turns on if “peak hold/valley
PH
3.3.14. Process Signal Measuring Unit hold/peak - valley hold” is on.
Function
(3) DZ Turns on if “digital zero” is on.
indicators
No. Name Description Turns on if remote control is being
RE performed through RS-232C or RS-
1 V-IN Positive input terminal for 1 to 5 V range 485 interface.
Indicates the HI side judgment value.
1 2 3 2 A-IN Positive input terminal for 4 to 20 mA range Indicates the item in the
(4) Sub-display 1
maximum/minimum/(maximum-
3 LO Negative input terminal minimum)/input value monitoring mode.
Indicates the LO side judgment value. Indicates the item to be set.
Indicates information on the item in
3.4. Comparison Output (5) Sub-display 2
the maximum/minimum/(maximum-
minimum)/input value monitoring mode.
No. Name Description Pressing the Enter and Mode keys
Returns to the measurement
together changes to the parameter
12 LO-b LO output terminal (b contact) mode.
12 13 14 15 16 17 18 19 setting mode.
Pressing the Enter and Increment keys
13 LO-c Common terminal for LO output together changes to the
maximum/minimum/(maximum-
14 LO-a LO output terminal (a contact) (6) Enter key
minimum)/input value monitoring mode.
Switches from the
15 GO-c Common terminal for GO output maximum/minimum/(maximum-
maximum/minimum/(maximum-
16 GO-a GO output terminal (a contact) minimum)/input value monitoring mode to
the comparative judgment reading mode.
17 HI-b HI output terminal (b contact)

18 HI-c Common terminal for HI output

19 HI-a HI output terminal (a contact)


Instruction Manual for A5000 Series 4/12
Pressing the Mode and Enter keys Pressing the Shift and Mode keys
together changes to the parameter Selects the item to be set. together changes to the shift function
setting mode. setup mode.
Pressing the Mode and Shift keys Holding down the Shift key for about
(7) Mode key together changes to the shift function one second moves to the HI judgment
(8) Shift key
setup mode. value indicator.
Pressing the Mode and Incremental Selects from items in the
keys together turns on/off the “Digital maximum/minimum/(maximum-
zero” indicator. minimum)/input value monitoring mode.
Pressing the Shift and Mode keys (Hold down the key for about one second.)
together changes to the shift function Pressing the Increment and Mode keys Changes the value or content
setup mode. together turns on/off the “Digital of a selected digit.
Selects from items in the zero” indicator. (Increments the value)
maximum/minimum/(maximum- Holding down the Increment key for
(8) Shift key minimum)/input value monitoring mode. about one second moves to the LO
(Hold down the key for about one second.) judgment value indicator.
Pressing the Increment and Mode keys Changes the value or content Pressing the Increment and Enter keys
(9) Increment key
changes to the
together turns on/off the “Digital of a selected digit.
maximum/minimum/(maximum-
zero” indicator. (Increments the value)
Pressing the Increment and Enter keys minimum)/input value monitoring mode.
Resets the
changes to the
maximum/minimum/(maximum-
(9) Increment key maximum/minimum/(maximum-
minimum)/input value monitoring mode.
minimum)/input value monitoring mode. (Hold down the key for about one second.)
Resets the
maximum/minimum/(maximum-
minimum)/input value monitoring mode. 5. Parameter Setup
(Hold down the key for about one second.)
5.1. Differences between Display Units

4.2. Single Display Unit

(2) (1) (3) (1) (3)

Setup contents Parameter name

5.1.1.Multi-display Unit

(6) (7) (8) (9) (6) (7) (8) (9)


Main Functions
No. Name
During measurement During parameter setup
IIndicates information on the
Indicates the measured value.
parameter to be set.
(1) Main display Indicates information on the item in
the maximum/minimum/(maximum-
minimum)/input value monitoring mode. Setup contents and parameter name
Indicates the result of judgment and
HI turns on if the measured value > HI
Note: Pressing the Mode key displays the next parameter.
judgment value.
Indicates the result of judgment and
Judgment
(2) GO turns on if LO judgment value ≦ the 5.1.2.Single Display Unit
indicators
measured value ≦ HI judgment value.
Indicates the result of judgment and
LO turns on if the measured value < LO Note 1: Pressing the mode key with the parameter name shown changes the
judgment value.
Turns on if “peak hold/valley display to the parameter information indication. If there is no key
PH
hold/peak - valley hold” is on.
Turns on if “digital zero” is on. operation for about one second when the parameter name is shown, the
DZ Flashes when linearization data output
values are set. display automatically changes to the parameter information indication
Function
(3) ME Turns on if “digital zero backup” is on.
indicators
Turns on if remote control is being
performed through RS-232C or RS-
(however, this change does not automatically occur for parameters
RE 485 interface.
Flashes when linearization data input
PH/S-HI/FSC, etc., right after COND/COM/MET is indicated).
values are set.
Pressing the Mode and Enter keys Note 2: Pressing the Mode key when the parameter information indication is
Returns to the measurement
together changes to the parameter
mode. shown results in the next parameter being displayed.
setting mode.
Pressing the Enter and Increment keys
together changes to the Note 3: If there is no key operation for about 8 seconds with the parameter
maximum/minimum/(maximum-
(6) Enter key information indication shown, the display returns to the parameter
minimum)/input value monitoring mode.
Switches from the
maximum/minimum/(maximum- name indication.
maximum/minimum/(maximum-
minimum)/input value monitoring mode to
the comparative judgment reading mode.
Pressing the Mode and Enter keys
together changes to the parameter Selects the item to be set.
setting mode.
Pressing the Mode and Shift keys
(7) Mode key together changes to the shift function
setup mode.
Pressing the Mode and Incremental
keys together turns on/off the “Digital
zero” indicator.
Instruction Manual for A5000 Series 5/12
5.2. Moving to the Parameter Setup Mode
Measurement Calibration data

Shift data

Condition data Comparator data Scaling data Linearization data

Pressing the ENTER key saves the data and returns to the measurement mode.

(Data are backed up with EEPROM even when the power is turned off.)

5.3. Data Lists and Default Settings


Default Equipped Input unit number Output unit number
Indication Name value as 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 0 1 2 3 4 5 6 7
Condition data
PVH Peak hold setup PH ○
RANG Measurement range setup *1 × ○ ○ ○ ○ ○ ○ ○ × ○ × ○ ○ ○ ○ ○ × ○
11 15 25 13 15 13 15 25 26 25 26 14 B JPB 14 14 2A
AVG Number of averaging 1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × ○ ○
MAV Number of moving averaging setup OFF ○
S.UD Step wide setup 1 ○
BLNK Indication blank setup OFF ○
UNIT Unit setup C × × × × × × × × × × × × ○ ○ × × × ×
BAUD Baud rate setup 9600 × × × ○ ○ × ○ ○
DATA Data length setup 7 × × × ○ ○ × ○ ○
P.BIT Parity bit setup E × × × ○ ○ × ○ ○
S.BIT Stop bit setup 2 × × × ○ ○ × ○ ○
T- Delimiter setup CR.LF × × × ○ ○ × ○ ○
ADR Equipment ID setup 00 × × × × ○ × × ○
A.OUT Analog output setup OFF × × ○ × × ○ ○ ○
B.UP Digital zero backup setup OFF ○
LINE Linearization setup CLR ○
I.SEL Input selection OC × × × × × × × × × × × × × × ○ × × ×
TR T Tracking zeroing time setup 00 ○ ○ ○ × × × × × × × × × × × × × ○ ○
TR V Tracking zeroing width setup *2 01 ○ ○ ○ × × × × × × × × × × × × × ○ ○
SNSR Sensor power setup 5 × × × × × × × × × × × × × × × × ○ ×
PON Power-on delay setup OFF ○
PRO Protect setup OFF ○
U-NO. Unit number Indication setup ON ○
Comparator data
S-HI HI side judgment value setup 1000 × ○ × × × ○ ○ ○
S-LO LO side judgment value setup 500 × ○ × × × ○ ○ ○
H-HI HI side hysteresis setup 0 × ○ × × × ○ ○ ○
H-LO LO side hysteresis setup 0 × ○ × × × ○ ○ ○
Scaling data
FSC Full scale Indication value setup *1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × × × × ○
9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999
FIN Full scale input value setup *1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × × × × ○
9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999 *3
OFS Offset indication value setup *1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × × × × ○
0 0 0 0 0 0 0 0 0 0 0 0 0
OIN Offset input value setup *1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × × × × ○
0 0 0 0 0 0 0 0 0 0 0 0 *4
PS Pre-scaling value setup 1 × × × × × × × × × × × × × × ○ ○ × ×
PPR Frequency division setup 1 × × × × × × × × × × × × × × ○ ○ × ×
DLHI Digital limiter HI value setup 9999 ○ ○ ○ × × × × × × × × ○ ○ ○ ○ ○ ○ ○
DLLO Digital limiter LO value setup -9999 ○ ○ ○ × × × × × × × × ○ ○ ○ ○ ○ ○ ○
AOHI Analog output HI indication setup 9999 × × ○ × × ○ ○ ○
AOLO Analog output LO indication setup 0 × × ○ × × ○ ○ ○
DEP Decimal point position setup None ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ × × ○ ○ ○ ○
Linearization data *5 ○
Calibration data
ZERO Zero input value *6 0 × × × × × × × × × × × × × × × × ○ ×
SPIN Span input value *6 2000 × × × × × × × × × × × × × × × × ○ ×
SPAN Span indication 9000 × × × × × × × × × × × × × × × × ○ ×
Shift data
SHF Shift data setup 0 ○
*1 Each value in the lower part of a cell in the columns on the right is the default value.
*2 Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
*3 5000 for 1 V range and 2000 for 2 A range
*4 1000 for 1 V range and 400 for 2 A range
*5 Linearization data are not set up for the default values.
*6 This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
Instruction Manual for A5000 Series 6/12
5.4. Information on Each Parameter 5.4.2 Method of Setting Comparator Data

Indication Name Setup options


Default This section explains comparator data and shows a typical example of setting
value
Condition data the HI side judgment value. The same method applies to all other parameters.
PVH Peak hold setup PH (peak hold)/VH (valley hold)/PVH (peak-valley hold) PH
RANG Measurement range setup *1 *1
Number of averaging operations HI side judgment value : 900
AVG 1/2/4/8/10/20/40/80 1
setup HI side hysteresis value : 200
Number of moving average LO side judgment value : 300
MAV OFF/2/4/8/16/32 OFF
operations setup
LO side hysteresis value : 150
S.UD Step width setup 1(1digit)/2(2digit)/5(5digit)/0(10digit) 1
BLNK Indication blank setup OFF/B-3/B-2/B-1/ON OFF
UNIT Unit setup C/F C Indicated value
BAUD Baud rate setup 9600/4800/2400/384(38400)/192(19200) 9600 HI side judgment value
DATA Data length setup 7(7bit)/8(8bit) 7bit 1000
P.BIT Parity bit setup E (even number), O (odd number), N (none) E 900
S.BIT Stop bit setup 2(2bit)/1(1bit) 2 HI side hysteresis
200 range
T- Delimiter setup CR.LF(CR+LF)/CR CRLF
ADR Equipment ID setup 01 to 99 00
A.OUT Analog output setup OFF/0-1(0 to 1V)/0-10(0 to 10V)/1-5(1 to 5V)/4-20(4 to 20mA) OFF
B.UP Digital zero backup setup OFF/ON OFF 500
LINE Linearization setup OFF/ON CLR LO side hysteresis
150 range
I.SEL Input selection OC (open collector)/LGC (logic)/MAG (magnet) O.C 300
TR T Tracking zeroing time setup 00 to 99 00
TR V Tracking zeroing width setup *2 00 to 99 01 LO side judgment value
SNSR Sensor power setup 10(10V)/5(5V) 5
PON Power on delay time setup OFF/ON OFF
PRO Protect setup OFF/1 to 30 OFF
0
U-NO. Unit number indication setup OFF/ON ON
Comparator data
S-HI HI side judgment value setup -9999 to 9999 1000 Judgment GO HI GO LO GO
S-LO LO side judgment value setup -9999 to 9999 500
H-HI HI side hysteresis setup 0 to 999 0
H-LO LO side hysteresis setup 0 to 999 0
Scaling data Multi-display unit Single display unit
FSC Full scale indication value setup -9999 to 9999
FIN Full scale input value setup *In the case of AC input, please use it by "+" setting by all means.
*1
OFS Offset indication value setup It is not displayed normally when I use it by "-" setting.
OIN Offset input value setup
PS Pre-scaling value setup 0.001 to 5.000 1.000
PPR Frequency division setup 1 to 100 1 (1) Press the Mode and Enter keys together during
DLHI Digital limiter HI value setup -9999 to 9999 9999 measurement.
DLLO Digital limiter LO value setup -9999 to 9999 -9999
Analog output HI indication
AOHI -9999 to 9999 9999
setup
Analog output LO indication (2) Press the Shift key a few times to display the
AOLO -9999 to 9999 0
setup comparator data menu.
Decimal point indication position
DEP None/place of 100/place of 101/place of 102/place of 103 None
setup
Linearization data *2 *2
Calibration data
ZERO Zero input value -0.300 to 2.000 0.000 (3) Press the Mode key a few times to display the
parameter to be set.
SPIN Span input value 1.000 to 3.000 2.000
SPAN Span indication 0 to 9999 9000
(4) For a single display unit, press the Mode key to
change to the parameter information indication.
(The display automatically changes to this
indication in about 1 second, except for parameter
S-HI right after COM is indicated.)
5.4.1Method of Setting Condition Data
This section shows a typical example of setting the peak hold parameter.
(5) Press the Shift key (change digit) and press the
&
The same method applies to other parameters. Increment key (change numeric value) to set to 10.
Note:The decimal point in the selected digit
flashes.
Multi-display unit Single display unit
(6) Press the Enter key to return to the
or measurement mode (Pressing the Mode key
changes to the next parameter).

(1) Press the Mode and Enter keys together during


measurement.

Note: The setup conditions are HI side judgment value > LO side judgment
(2) Press the Mode key to change to the peak hold value, HI side judgment value ³ LO side judgment value + LO side
setup mode.
hysteresis, and LO side judgment value £ HI side judgment value-HI
side hysteresis. If these conditions are not satisfied, an error indication
(3) For a single display unit, press the Mode key to
change to the parameter information indication. appears and the display returns to the HI side judgment value setup.
(The display automatically changes to this
indication in about 1 second, except right after
COND is indicated.) 5.4.3 Method of Setting Scaling Data
This section explains comparator data and shows a typical example of setting
(4) Press the Increment key a few times to set to
Valley Hold. the full scale indication parameter. The same method applies to all other
parameters. Indication

(5) Press the Enter key to return to measurement


or mode. (Pressing the Mode key changes to the next Input voltage : 0 to 10V
parameter).
Indication : 0 to 5.000

5000

FSC : 5000
3000 DLHI
F I N : 9999
OFS : 0
O I N : 0
Input
DLHI : 3000
0 10V
DLLO : -2000
DLLO
DEP : The place of 103 is lit. -2000

Note: For the Digital limiter, values larger than the DLHI setpoint are not
indicated even if signals greater that the value set in the DLHI
parameter are input (for DLLO parameter, values smaller than the
DLLO setpoint are not indicated).
Instruction Manual for A5000 Series 7/12

Multi-display unit Single display unit 5.4.4 Method of Setting Calibration Data

5.4.4.1 Actual Load Calibration

(1) Press the Mode and Enter keys together during Actual load calibration means that calibration is carried out by applying
measurement.
actually measured pressure to a sensor such as a load cell connected to the
meter.
Multi-display unit Single display unit
(2) Press the Shift key a few times to change to the
scaling data menu.

(1) Press the Shift, Increment, and Enter keys


(3) Press the Mode key a few times to display the together during measurement.
parameter to be set.

(4) For a single display unit, press the Mode key to (2) Press the Mode key to change to the actual
change to the parameter information indication. load calibration mode.
(The display automatically changes to this
indication in about 1 second, except for parameter (3) Press the Mode key while applying
FSC right after MET is indicated.) pressure that will cause the display to show
zero.
When the Enter key is
pressed, it returns to Err1: When the input at the time of zero calibration
measurement mode. is below -0.3mV/V, it displays.
(5) Press the Shift key (change digit) and press the
& Increment key (change numeric value) to set to 10. Err2: When the input at the time of zero
calibration is more than 1mV/V, it displays.
Note:The decimal point in the selected digit (4) For a single display unit, press the Mode
flashes. When the Enter key is key to change to the parameter information
pressed, it returns to
measurement mode indication.
(6) Press the Enter key to return to the with Zero input value
or measurement mode (Pressing the Mode key only. (5) Press the Shift key (change digit) and
changes to the next parameter). Increment key (change numeric value) to
set 8000.
Note:The decimal point in the selected digit
& flashes.
Note: For the process signal measuring unit, set the full scale input value to Err3: It is the same as that of the time of being an
input at the time of span calibration at the zero
5.000 for the 1 V range and to 20.00 for the 2 A range, and set the offset proofreading time, or when small, it displays.
Err4: When the input at the time of span calibration
input value to 1.000 for the 1 V range and to 4.00 for the 2 A range. is more than 3.3mV/V, it displays.
Err5: When the setup more than the highest
The following explains the frequency measuring unit. (The same method applies decomposition ability is performed, it displays.
(6) Press the Mode key to return to the
to the full scale indication parameter.) measurement mode.

Determining the revolution speed (rpm) using the rotary encoder set to 30 pulses per minute: 5.4.4.2 Equivalent Calibration

(1)Determine the measurement range by calculating the maximum frequency.


Equivalent calibration means that calibration is carried out according to the
The figure below shows an example where the revolution rises to a maximum speed of ratings (specifications) of such a sensor as a load cell. It is not necessary to
about 100 rpm.
connect the sensor or to apply pressure to the sensor.
30 × 100  60 = 50 Multi-display unit Single display unit

Number of pulses per second

Revolution speed per second (1) Press the Shift, Increment, and Enter key
together during measurement.
Number of pulses per revolution at the
rotary encoder
(2) Press the Increment key to select the
(2)Since the number of pulses determined in (1) is 50 per second (50 Hz), set the range to equivalent calibration mode.
range 11(for how to set the range, see the section on setting condition data).

(3)The display unit shows 500 if 50 Hz pulse input is measured under range 11 (when PS=1 (3) Press the Mode key to move to the equivalent
and PPR=1 by default). Therefore, the parameters should be set as PS=2 and PPR=1 so that calibration mode.

the decimal point is positioned in the 101 digit(100.0 is is indicated 50 Hz input).

(4) Press the Shift key to display the zero-input


Note: For the frequency measuring unit, set the relationship between the input setup mode.
Note:For a single display unit, the unit
and indication using the PS and PPR parameters (parameters of FSC, FIN, automatically returns to ZERO indication if
there is no key operation for about 8 seconds.
In such a case, press the Mode key to return
OFS, and OIN are not indicated). to the numerical value indication.
Press the Mode key
while applying & (5) Press the Shift key (change digit) and the
pressure that will Increment key (change numeric value) to set 0.004.
The following explains the scaling of analog output (The same method applies cause the display to Note:The decimal point in the selected digit
show zero.
flashes.
to the full scale indication parameter.)

Output (6) Press the Mode key to change to the span


input value setup mode.
Err1: When the input at the time of zero
20mA calibration is below -0.3mV/V, it displays.
Indication : 0.0 to 500.0
Output : 4 to 20mA Err2: When the input at the time of zero
calibration is more than 1mV/V, it displays.

(7) For a single display unit, press the mode key


to display the parameter information indication.

AOHI : 5000
AOLO : 0 & (8) Press the Shift key (change digit) and the
4mA Increment key (change numeric value) to set
1.002.

Indication
0 5000 (9) Press the Mode key to change to the span
indicating value setup mode.

Note1: For analog output scaling, set the indication value for an output current
(10) For a single display unit, press the Mode key
of 20 mA in the AOHI parameter and set the indication value for an to display the parameter information indication.

output current of 4 mA in the AOLO parameter (for 4-20 mA output). (11) Press the Shift key (change digit) and the
Increment key (change numeric value) to set 2000.
Note:The decimal point in the selected digit
Note2: The analog signal out of the setting range cannot be accurately & flashes.
Err3: It is the same as that of the time of being an
output. input at the time of span calibration at the
zero proofreading time, or when small, it
displays.
Err4: When the input at the time of span calibration
is more than 3.3mV/V, it displays.
Err5: When the setup more than the highest
decomposition ability is performed, it displays.
(12) Press the Mode key to return to the
measurement mode.
Instruction Manual for A5000 Series 8/12

5.4.5 Method of Setting Linearization Data 6.3 Peak Hold Function


The linearization function means a function that changes the slope of straight The Peak Hold function retains one of the maximum (peak hold)/minimum
lines in the relationship between the input and indication by correcting the (valley hold)/maximum - minimum (peak-valley hold) values and provides output
relations at arbitrary points. Linearization data are set using the input value for that value. Selection from these values is made using the condition data.
(indicated value before correction) and the output value (indicated value after The peak hold function is enabled by shortcircuiting the PH and COM terminals
correction) at each arbitrary point.
or setting both terminals to the same voltage level.
Multi-display unit Single display unit
7.Output Function
7.1 Comparison Output Function
(1) Press the Mode key and the Enter key during The A5000 series of unit meters is designed so that the two judgment values HI
measurement.
and LO can be set for the measured (indicated) value to provide the results of
judgment as relay contact output. (This function is effective when the meter is
(2) Press the Shift key a few times to move to the
linearization data menu. equipped with a comparison output unit.) For details on the contact ratings and
other specifications, refer to the section “Output Specifications.”
(3) Press the Mode key to move to the setup for the
number of data to be corrected. 7.2 Analog Output Function
The A5000 series of unit meters can output an analog signal for an indicated
(4) Press the Shift key (change digit) and the
&
Increment key (change numeric value) and then value (when the meter is equipped with an analog output unit). There are four
press the Mode Key to after the number of data to be
corrected has been set. output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection can
Note: The decimal point in the selected digit
flashes. be made using the condition data. In addition, the analog output of the A5000
(5) For a single display unit, press the Mode key to series allows for arbitrary output scaling. This scaling can be achieved by
change to the parameter information indication.
Note: For a single display unit, the unit setting the indication value for an output of the maximum scale value (20 mA for
automatically returns to N-01 indication if there is
RE
no key operation for about 8 seconds. In this 4–20 mA output range) in the AOHI parameter of the scaling data.
case, press the Mode key to return to the numeric
value indication.
7.3 RS-485 Interface Function
(6) Press the Shift key (change digit) and Increment
& key (change numeric value) and then press the Mode
key after the N-01 input value has been set.
The A5000 series can be equipped with an RS-485 interface (when the meter is
DZ
Note: For a single display unit, RE flashes when provided with an RS-485 unit). For details on the RS-485 function, see the
the input value is set and DZ flashes when the
output value is set. separate manual on communication functions.
(7) Press the Shift key (change digit) and Increment
& key (change numeric value) and then press the Mode
key after the N-01 output value has been set. 7.4 RS-232C Interface Function
The A5000 series can be equipped with an RS-232C interface (when the meter is
(8) Press the Mode key to display the setup for output provided with an RS-232C unit). For details on the RS-232C function, see the
values of the data to be corrected.
Note: For a single display unit, the unit is separate manual on communication functions.
automatically returned to N-02 indication if there is
RE
no key operation for about 8 seconds. In this
case, press the Mode key to return to the numeric
8.Specifications and External Dimensions
value indication.
(9) Press the Shift key (change digit) and Increment
8.1 Input Specifications
& key (change numeric value) and then press the Mode
key after the N-02 input value has been set.
8.1.1 DC Voltage Measuring Unit (range 11)
Repeat steps (5) to (9) until all the settings have Maximum
been made. Measurement Highest Input
Range Indication permissible Accuracy
range resolution impedance
input
Offset : ±9999 About
(10) When all the settings have been made, press the 11 ±99.99mV 10μV ±100V ±(0.1% of FS)
Enter key to return to the measurement mode. Full scale : 0 to ±9999 100MΩ

Input circuit : Single ended type


Operating system : ΔΣ conversion
Note: The setup conditions are N-1 < N-2 … N-15 < N-16. Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
6.Control Functions
6.1 Hold Function 8.1.2 DC Voltage Measuring Unit (ranges 12 to 15)
Maximum
The Hold function temporarily retains the indication. The hold function is Range
Measurement
Indication
Highest Input
permissible Accuracy
range resolution impedance
enabled by shortcircuiting the HOLD and COM terminals or setting both input
About
terminals to the same voltage level. As a result the display unit retains the 12 ±999.9mV 100μV
100MΩ
±100V
indication given at that moment. About
13 ±9.999V 1mV ±250V ±(0.1% of FS)
Offset : ±9999 1MΩ
6.2 Digital Zero Function 14 ±99.99V
Full scale : 0 to ±9999
10mV
About
±250V
10MΩ
The Digital Zero function zeros the indication given at an arbitrary timing. About
15 ±600V 100mV ±600V ±(0.15% of FS)
Thereafter, the function shows the amount of change from the point of zeroing. 10MΩ
However, this function serves as an indication resetting function for a frequency Input circuit : Single ended type
measuring unit. Thus, the Digital Zero function can be used to reset the Operating system : ΔΣ conversion
indication when there is no input signal at all. Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Note that, the on/off control of the Digital Zero function can be achieved by
means of terminal control or front panel keys. In the case of terminal control, 8.1.3 DC Current Measuring Unit
the Digital Zero function is turned on by shortcircuiting the DZ and COM Maximum
Measurement Highest Input
terminals or setting both terminals to the same voltage level. The indication at Range Indication permissible Accuracy
range resolution impedance
input
that moment is zeroed. In the case of control with the front panel keys, hold About
down the Mode key and press the Increment key for about 1 second to zero the 23 ±9.999mA 1μA ±100mA
10Ω
±(0.2% of FS)
indication at that moment. Offset : ±9999 About
24 ±99.99mA 10μA ±500mA
Full scale : 0 to ±9999 1Ω
Note: Operation with the control terminals takes priority over operation with About
25 ±999.9mA 100μA ±3A ±(0.3% of FS)
the front panel keys. The Digital Zero function is disabled if the 0.1Ω
control terminals are made to go through the off-on-off sequence with
Input circuit : Single ended type
the function enabled by means of the front panel keys. Operating system : ΔΣ conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Instruction Manual for A5000 Series 9/12
8.1.4 AC Voltage Measuring Unit (average value detection: ranges 8.1.9 AC Current Measuring Unit (average value detection: range
11 to 13) 26)
Maximum Maximum
Measurement Highest Input Measurement Highest Input
Range Indication permissible Accuracy Range Indication permissible Accuracy
range resolution impedance range resolution impedance
input input
11 99.99mV 10μV Offset : ±9999 ±(0.5% of FS
26 5A 1mA (CT) 8A
±100V Full scale : 0 to ±9999 +10digit)
Offset : ±9999 1MΩ ±(0.2% of FS
12 999.9mV 100μV
Full scale : 0 to ±9999 or more +10digit)
Input circuit : CT isoration type
13 9.999V 1mV ±250V Operating system : ΔΣ conversion
Maximum sampling rate : 12.5 times per second
Input circuit : Single ended type Frequency range : 50 Hz or 60Hz
Operating system : ΔΣ conversion Response speed : About 1 second
Maximum sampling rate : 12.5 times per second Crest factor : 4:1 at full scale
Frequency range : 40 Hz to 1 kHz Dead zone : 0 to 99 digits
Response speed : About 1 second
Dead zone : 0 to 99 digits 8.1.10 AC Current Measuring Unit (true rms value: ranges 23 to
25)
8.1.5 AC Voltage Measuring Unit (average value detection: ranges Maximum
Measurement Highest Input
14 and 15) Range
range
Indication
resolution impedance
permissible Accuracy
input
Maximum
Measurement Highest Input About
Range Indication permissible Accuracy 23 9.999mA 1μA 100mA
range resolution impedance 10Ω
input Offset : ±9999 About ±(0.5% of FS
±(0.2% of FS 24 99.99mA 10μA 500mA
14 99.99V 10mV 250V Full scale : 0 to ±9999 1Ω +20digit)
Offset : ±9999 1MΩ +10digit) About
Full scale : 0 to ±9999 or more ±(0.3% of FS 25 999.9mA 100μA 3A
15 600V 100mV 600V 0.1Ω
+10digit)
Input circuit : Single ended type
Operating system : ΔΣ conversion
Input circuit : Single ended type Maximum sampling rate : 12.5 times per second
Operating system : ΔΣ conversion Frequency range : 40 Hz to 1 kHz
Maximum sampling rate : 12.5 times per second Response speed : About 1 second
Frequency range : 40 Hz to 1 kHz Crest factor : 4:1 at full scale
Response speed : About 1 second Dead zone : 0 to 99 digits
Dead zone : 0 to 99 digits *The accuracy applied to a sine wave that equals or exceeds 5% of the measurement range.
*"0" may not be displayed infrequently under the influence of the DC cut
8.1.6 AC Voltage Measuring Unit (true rms value: ranges 11 to 13) condenser of the input part when turning on the power with no signal, but it is not
Maximum trouble.
Measurement Highest Input
Range Indication permissible Accuracy
range resolution impedance It normally operates if the signal in measurement range is given.
input
11 99.99mV 10μV
Offset : ±9999 1MΩ
±100V
±(0.2% of FS
8.1.11 AC Current Measuring Unit (true rms value: range 26)
12 999.9mV 100μV
Full scale : 0 to ±9999 or more +20digit) Maximum
Measurement Highest Input
Range Indication permissible Accuracy
13 9.999V 1mV ±250V range resolution impedance
input
Input circuit : Single ended type Offset : ±9999 ±(0.5% of FS
Operating system : ΔΣ conversion 26 5A 1mA (CT) 8A
Full scale : 0 to ±9999 +20digit)
Maximum sampling rate : 12.5 times per second Input circuit : CT isoration type
Frequency range : 40 Hz to 1 kHz Operating system : ΔΣ conversion
Response speed : About 1 second Maximum sampling rate : 12.5 times per second
Crest factor : 4:1 at full scale Frequency range : 50 Hz or 60Hz
Dead zone : 0 to 99 digits Response speed : About 1 second
*The accuracy applied to a sine wave that equals or exceeds 5% of the measurement range.
*"0" may not be displayed infrequently under the influence of the DC cut Crest factor : 4:1 at full scale
Dead zone : 0 to 99 digits
condenser of the input part when turning on the power with no signal, but it is *The accuracy applied to a sine wave that equals or exceeds 5% of the measurement range.
not trouble. *"0" may not be displayed infrequently under the influence of the DC cut
It normally operates if the signal in measurement range is given. condenser of the input part when turning on the power with no signal, but it is not
trouble.
8.1.7 AC Voltage Measuring Unit (true rms value: ranges 14 and It normally operates if the signal in measurement range is given.
15)
Maximum 8.1.12 Resistance Measuring Unit
Measurement Highest Input
Range Indication permissible Accuracy Measurement Highest Circuit
range resolution impedance Range Indication Accuracy
input range resolution current
±(0.2% of FS
14 99.99V 10mV 250V About
Offset : ±9999 1MΩ +20digit) 11 99.99Ω 10mΩ
Full scale : 0 to ±9999 or more ±(0.3% of FS 5mA
15 600V 100mV 600V About
+20digit) 12 999.9Ω 100mΩ
Offset : ±9999 500μA
Input circuit : Single ended type Full scale : 0 to ±9999 About
±(0.2% of FS)
Operating system : ΔΣ conversion 13 9.999kΩ 1Ω
50μA
Maximum sampling rate : 12.5 times per second About
Frequency range : 40 Hz to 1 kHz 14 99.99kΩ 10Ω
5μA
Response speed : About 1 second
Crest factor : 4:1 at full scale Input circuit : Single ended type
Dead zone : 0 to 99 digits Operating system : ΔΣ conversion
*The accuracy applied to a sine wave that equals or exceeds 5% of the measurement range. Maximum sampling rate : 12.5 times per second
Measuring system : Two-wire system or four-wire system
*"0" may not be displayed infrequently under the influence of the DC cut (internal socket change-over)
condenser of the input part when turning on the power with no signal, but it is Open-circuit voltage : About 5 V
not trouble. 8.1.13 Temperature Measuring Unit (TC)
It normally operates if the signal in measurement range is given. Input Highest
Range Indication Accuracy
sensor resolution
-50.0 to 199.9℃ 0.1℃
8.1.8 AC Current Measuring Unit (average value detection: ranges KA K
(-58.0 to 391.8°F) (0.1°F)
±(0.5% of FS)

23 to 25) KB K
-50 to 1200℃
(-58 to 2192°F)
±(0.2% of FS)
Maximum -50 to 1000℃
Measurement Highest Input J J
(-58 to 1832°F)
Range Indication permissible Accuracy
range resolution impedance -50 to 400℃
input T T ±(0.6% of FS)
(-58 to 752°F)
About 1℃
23 9.999mA 1μA 100mA 0 to 1700℃
10Ω S S
(32 to 3092°F)
(1°F)
Offset : ±9999 About ±(0.5% of FS ±(0.4% of FS)
24 99.99mA 10μA 500mA -10 to 1700℃
Full scale : 0 to ±9999 1Ω +10digit) R R
(14 to 3092°F)
About 100 to 1800℃
±(0.4% of FS)
25 999.9mA 100μA 3A B B Note : The accuracy of range B is applicable
0.1Ω (212 to 3272°F)
to temperatures of 500℃ or more.
Input circuit : Single ended type Input circuit : Single ended type
Operating system : ΔΣ conversion Operating system : ΔΣ conversion
Maximum sampling rate : 12.5 times per second Maximum sampling rate : 6.25 times per second
Frequency range : 40 Hz to 1 kHz Cold junction compensation error : ±2℃(at 10 through 40℃)
Response speed : About 1 second Internal resistance of sensor : 50Ω or less
Crest factor : 4:1 at full scale Linearizer : Digital linearizer
Dead zone : 0 to 99 digits Burnout alarm : It blinks by ---- display.
Instruction Manual for A5000 Series 10/12
8.1.14 Temperature Measuring Unit (RTD) 8.2 Common Specifications
Input Highest Display : 7-segment LED display (character height : 14.2 mm on main
Range Indication Accuracy
sensor resolution
display and 8 mm on sub-display)
PA PT100Ω
-100.0 to 199.9℃ 0.1℃ Polarity indication : Automatically indicated when the calculated result is
±(0.15% of FS)
(-148.0 to 391.8°F) (0.1°F)
JPA JPt100Ω negative.
PB PT100Ω Indication range : -9999 to 9999
-100 to 600℃ 1℃
±(0.3% of FS)
(-148 to 1112°F) (1°F) Over-range alarm : OL or -OL for input signals outside the indicayion range
JPB JPt100Ω
Decimal point : Can be set at an arbitrary digit.
Input circuit : Single ended type
Operating system : ΔΣ conversion Zero indication : Leading aero suppression
Maximum sampling rate : 12.5 times per second External control : HOLD , PH , DZ (reset for frequency measuring unit)
Current through RTD : About 1 mA
External resistance : 10Ω or less per wire Operating temperature and humidity range : 0 to 50℃ ,
Linearizer : Digital linearizer
Burnout alarm : It blinks by ---- display. 35 to 83%RH(non-condensing)
Storage temperature and humidity range : -10 to 70℃ , 60%RH or less
8.1.15 Frequency Measuring Unit (open collector, logic, and Power supply : 100 to 240V AC ±10% for AC power supply unit
magnet)
9 to 60 V DC for DC power supply unit
Measurement Highest Renewal time Power consumption : 7VA max. (AC power supply)
Range Indication Accuracy
range resolution of a display
7W max. (DC power supply)
11 0.1 to 200Hz 0.1Hz 10s External dimensions : 96 mm (W) × 48 mm (H) × 146.5 mm (D)
12 1 to 2000Hz 1Hz 1s Note : Depth (D) denotes the maximum value.
Pre-scale : 0.001 to 5
±(0.2% of FS)
Frequency division : 1 to 100 Weight : 450g
13 0.01 to 20kHz 10Hz 100ms
Withstand voltage : 2000V AC for 1 min. between power terminals and input
14 0.1 to 200kHz 100Hz 100ms
terminal, and between power terminals and each output
terminal (AC power supply)
Maximum
Input type Input voltage lebel
permissible input Withstand voltage : 500V DC for 1 min. between power terminals and input
Open collector LO : 1V or less (5V : 4.7kΩ pull up) terminal, and between power terminals and each output
Logic LO : 1V or less,HI : 2.5 to 15V 15V terminal (DC power supply)

Magnet 0.3 to 30Vp-p


Withstand voltage : 500V DC for 1 min. between input terminal and each output
terminal, and between analog output terminal and
Duty ratio : 50% communication terminals
8.1.16 Frequency Measuring Unit (500 Vrms) 2000 V AC for 1min between ccase and each output terminal
(common to both AC and DC supply)
Measurement Highest Renewal time
Range Indication Accuracy Insulation resistance : 100 M Ω between the abbove terminals when 500 V DC is
range resolution of a display

11 0.1 to 200Hz 0.1Hz 10s


applied
Conformity standard : EN61326-1 EMI : Class A EMS : Industrial locations
12 1 to 2000Hz 1Hz 1s
Pre-scale : 0.001 to 5 EN61010-1 (Transducer is applied to standard by the
±(0.2% of FS)
Frequency division : 1 to 100
13 0.01 to 20kHz 10Hz 100ms
input and output signal line length less than 30m.)
14 0.1 to 200kHz 100Hz 10ms EN IEC 63000
It applies only to the product with which CE mark is
Maximum indicated on the label.
Input type Input voltage lebel
permissible input
Grounding environment : Category Ⅱ, Pollution degree 2
Voltage 50 to 500Vrms 500Vrms
Altitude : 2000m max
Duty ratio : 50% Fuse : K19372 1.0A (DC power supply)

8.1.17 Strain Gage Unit


8.3 Output Specifications
Sensor Zero adjusting Span adjusting Measurement Highest
Accuracy
power range range range resolution 8.3.1 Output for Comparison
5V 0.5μV/digit Judgment
Conditions for comparison
-0.3 to +1mV/V 1 to 3mV/V 0 to 3mV/V ±(0.1% of FS +2digit) result
10V 1μV/digit
Indicated value > Upper limit judgment value HI

Input circuit : Single ended type Lower limit judgment value ≦ Indicated value ≦ Upper limit judgment value GO
Operating system : ΔΣ conversion
Maximum sampling rate : 12.5 times per second Lower limit judgment value > Indicated value LO
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Control system : Micro computer operating system
8.1.18 Process Signal Measuring Unit Judgment value setup range : -9999 to +9999
Hysteresis : Can be set in the range of 1 to 999 digits for each judgment value
Maximum
Range
Measurement
Indication
Input
permissible Accuracy Operating speed : Depends on the sampling rate.
range impedance
input Output method : Relay contact output (Make and break contacts for HI and LO
About
1V 1 to 5V ±100V and make contacts for GO)
Offset : ±9999 100MΩ
±(0.2% of FS) Output rating : 240 V AC, 8 A (resistive load) and 30 V DC, 8 A (resistive load)
Full scale : 0 to ±9999 About
2A 4 to 20mA ±100mA
10Ω
Mechanical life : 20,000,000 times or more
Input circuit : Single ended type Electric life : 100,000 times or more (Resistance load)
Operating system : ΔΣ conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Instruction Manual for A5000 Series 11/12
8.3.2 Analog Output 8.4 External Dimensions

Output type Load resistance Accuracy Ripple

0 to 1V 10kΩ or more

0 to 10V 10kΩ or more ±50mVp-p


±(0.5% of FS)
1 to 5V 10kΩ or more

4 to 20mA 550Ω or less ±25mVp-p


Note: The ripple ratings for the 4-20 mA output are when the load resistance of
250Ω and the output current of 20 mA are applied.

Conversion system : PWM conversion

Resoluton : Equivalent to 13 bits

Scaling : Digital scaling

Response speed : About 0.5 second

8.3.3 Communicating Function

RS-232C RS-485

Synchronization system Start and stop synchronization

Communication system Full duplex Two wire halt duplex


( (Polling selecting system)
Communication rate 38400bps/ 19200bps/ 9600bps/ 4800bps/ 2400bps

Start bit 1bit

Data length 7bit/8bit


Error detection Even parity/ odd parity/ non-parity

BCC (block, check, and character check sum)

Stop bit 1bit/2bit

Character code ASCII code

Communication
No procedure
control procedure 9. Warranty and After-service
Signal name used TXD ,RXD ,SG Non-inversion(+) and inversion(-) 9.1 Warranty
1 1 Up to 31 meters The warranty period shall be one year from the date of delivery. Any failure
Number of
connectable units that arises during this period and the cause thereof is judged to be obviously
attributable to Watanabe Electric Industry Co., Ltd. shall be remedied at no cost.
Line length 15m Up to 500m (total) In EN/IEC conformity, it is under 30m.
Delimiter CR+LF/CR 9.2 After-service
This product is manufactured, tested, inspected, and then shipped under stringent
quality control. Should the product fail, however, contact (or send the product
to) your vendor or Watanabe Electric Industry directly. (It is advisable that you
send a memo describing the failure in as much detail as possible along with the
product returned.)
Instruction Manual for A5000 Series 12/12
Setting table

Default Default
Indication Name value Indication Name value

Condition data Scaling data

PVH Peak hold setup PH FSC Full scale Indication value setup *1

RANG Measurement range setup *1 FIN Full scale input value setup *1

AVG Number of averaging 1 OFS Offset indication value setup *1

MAV Number of moving averaging setup OFF OIN Offset input value setup *1

S.UD Step wide setup 1 PS Pre-scaling value setup 1

BLNK Indication blank setup OFF PPR Frequency division setup 1

UNIT Unit setup C DLHI Digital limiter HI value setup 9999

BAUD Baud rate setup 9600 DLLO Digital limiter LO value setup -9999

DATA Data length setup 7 AOHI Analog output HI indication setup 9999

P.BIT Parity bit setup E AOLO Analog output LO indication setup 0

S.BIT Stop bit setup 2 DEP Decimal point position setup None

T- Delimiter setup CR.LF Linearization data *5

ADR Equipment ID setup 00 Calibration data

A.OUT Analog output setup OFF ZERO Zero input value *6 0

B.UP Digital zero backup setup OFF SPIN Span input value *6 2000

LINE Linearization setup CLR SPAN Span indication 9000

I.SEL Input selection OC Shift data

TR T Tracking zeroing time setup 00 SHF Shift data setup 0

TR V Tracking zeroing width setup *2 01

SNSR Sensor power setup 5

PON Power-on delay setup OFF

PRO Protect setup OFF

U-NO. Unit number Indication setup ON

Comparator data

S-HI HI side judgment value setup 1000

S-LO LO side judgment value setup 500

H-HI HI side hysteresis setup 0

H-LO LO side hysteresis setup 0


*1 Each value in the lower part of a cell in the columns on the right is the default value.
*2 Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
*5 Linearization data are not set up for the default values.
*6 This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
*When the power turns on while pressing all the keys (Enter, Mode, Shift and Increment),
and held pressed until LEDs are all lit, then all data initializes to the default settings.

6-16-19,
6-16-19,Jingumae,
Jingumae,Shibuya-ku,
Shibuya-ku,Tokyo
Tokyo150-0001,
150-0001,Japan
Japan
WATANABE ELECTRIC INDUSTRY CO., LTD. Phone:
Phone:(81)3-3400-6140
(81)3-3400-6141
WATANABE ELECTRIC INDUSTRY CO., LTD. Homepage
Homepage http://en.watanabe-electric.co.jp/
http://www.watanabe-electric.co.jp/en/
ASAHI KEIKI DIVISION