IR-HA - Manual (INE-567-P0CE) 2017 - 9 - 15 (002) - 1 PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 48

INST.No.

INE-567-P0CE

IR-HA Series
PORTABLE RADIATION
THERMOMETER
Model: IR-HAI
IR-HAS
IR-HAQ

Always keep this instruction with the unit.

Please be sure to deliver these instructions


with the unit to the end user.
INST.No.INE-567-P0CE

PREFACE/ Request and notices


Thank you for purchasing IR-HASeries Portable Radiation Thermometer.
Please read this instruction manual for using this thermometer correctly, safely and also
preventing troubles in advance.

 Request to designers, instrument controllers, and sale agents


Make sure to deliver this instruction manual to the operator of this thermometer.

 Request to the operator of the thermometer


Follow this instruction manual and use the thermometer correctly.
This instruction manual is necessary for maintenance, too.
Keep this manual with due care until this thermometer is discarded.
If you have unclear points or need technical assistance, please contact your sales agent of CHINO
Corporation.

NOTE
1. The descriptions of this manual are subject to change without notice.
2. If a question has arisen or if an omission was found in this manual, please contact your nearest CHINO’s sales
agent or your dealer.
3. Copy right of this instruction manual belongs to CHINO Corporation. All or part of this instruction manual shall
not be released to the third party.

Trademark
・SD Memory Card is the trademark of Panasonic Corporation, SanDisk Corporation in USA, and TOSHIBA
CORPORATION.

-C1-
INST.No.INE-567-P0CE

Important explanation !
To use this instrument correctly and safely, make sure to observe following safety precautions.

1 Precondition for use


★For key operation, make sure to push the button until it makes "blip" sound.(Except for a few operation)
●The thermometer is designed as a handheld type. Use a tripod or universal head for long term or fixed
mounting measurement.
●The working temperature range of this product is 0 to 50°C. (No dew condensation)
●Do not use the thermometer in dusty places, etc. Remove the dust after using it.
(Refer to "10.1 Cleaning of Objective Lens" for objective lens cleaning and "10.2 Cleaning of External LCD
and Eyepiece Cover" for external LCD and eyepiece cover cleaning.)
●Be careful not to give vibration or impact to the thermometer
●For preventing the consumption of the batteries, make sure to remove the batteries when it is not used.

2 Storage
●Do not store the thermometer in hot and humid places. Make sure to store the thermometer with the lens
cap. It is recommend to stores the thermometer in room temperature with desiccant (silica gel etc.).
●Do not leave the thermometer in extreme high ambient temperature such as beside a rear window or inside
of a trunk of cars. The thermometer may have trouble
●When the thermometer is not used for 2 weeks or more, take out the batteries from it.
Otherwise, the thermometer may be damaged by liquid leakage of the batteries.
●For failures of the thermometer, don’t overhaul it by yourself, and contact your sales agent of CHINO
Corporation.

3 Disposal
●When you discard this product, please obey the regulation of each local government.

4 Symbols in this instruction manual


The symbols shown below are used depending on important degrees for using the thermometer safely and
avoiding unexpected situations.
Important
Symbols Contents
degree
! This symbol is indicated with a title for an explanation with Warning
1
Indicates important information that must be observed to avoid blindness (or
2 Warning other dangers that may result in serious personal injury or death,) or damage to
this product.
Indicates important information that must be observed to avoid the risk of personal
3 Caution
injury or malfunctions of this product.
Indicates supplementary information that the operator is recommended to
4 Remark
understand.

5 Reference Indicates supplementary information or a reference to an operation.

-C2-
INST.No.INE-567-P0CE

Warning and attention for the security !


◆To use this product, make sure to observe following matters and use it correctly.
In addition, keep this instruction manual carefully in the place that you can reach at anytime.
shows an act of the prohibition.

! Warning (May cause death or serious injury)

! Don’t operate this thermometer in places where combustible or volatile gas exists.
It is extremely dangerous to use the thermometer in such environment.

! Don’t use the thermometer if it is broken, smoking or abnormal order is detected.


These may cause fire.


Disassembling or modifying this thermometer may not only cause failure but also be
dangerous to you. Disassembling or modifying this thermometer is prohibited.

! If it is broken, smoking or abnormal order is detected, turn off the power supply
immediately and contact your sales agent of CHINO Corporation.

! Caution (May cause injury or physical damage)

Avoid using in the places where; temperature changes widely, humidity is high, close to


heavy electric circuit, inductive interference is large, static electricity or magnetic field
exists, and mechanical vibration and impact exist. Also avoid using under the atmosphere
where dust and particles exist, corrosive gas exists and electric noise or static electricity
exists and easily to interrupt.


To protect eyes and detecting element, never to see the sun through the finder of the
thermometer.

For the measurement of an object exceeding 1500°C, make sure to turn the beam

! attenuation filter switch "ON (attenuation side)" for protecting your eyes.
When you feel glare on the measurement of an objects lower than 1500°C, turn the beam
attenuation filter switch "ON (attenuation side)".
To use the thermometer safely, strictly observe the contents described in this instruction manual.
! If the contents of this instruction manual are not complied, damage to the thermometer,
functional decline or damage to a system may occur.

-C3-
INST.No.INE-567-P0CE

CONTENTS
1. Introduction ................................................................ 1 8. Memory Input Mode (Display/Saving Data)........... 25
1.1 General ................................................................... 1 8.1 Data Display ........................................................ 25
1.2 Configuration ......................................................... 1 8.2 Data Save ............................................................. 26
2. Model and Attachments ............................................. 2 8.2.1 Manual Memory Mode Saving ......................... 26
2.1 Model ...................................................................... 2 8.2.2 Interval Memory Mode Saving ........................ 27
2.2 Attachments ........................................................... 2 8.3 Data File and Folder............................................ 28
3. Names and Functions of Component parts .............. 3 8.4 File Saving Format .............................................. 29
3.1 Names and Functions of Component parts.......... 3 9. Zero/Span Adjustment Mode ................................... 30
3.2 External Display Marker and 9.1 Initialize Zero/Span Adjustment......................... 31
Viewfinder Inside ........................ 4 9.2 Perform Zero/Span Adjustment .......................... 31
3.3 Operation Key ........................................................ 5 9.3 InGaAs Zero Adjustment..................................... 32
4. Preparation for Measurement ................................... 6 9.4 InGaAs Span Adjustment.................................... 32
4.1 Loading batteries ................................................... 6 9.5 Si Zero Adjustment .............................................. 33
4.2 About SD Card ....................................................... 7 9.6 Si Span Adjustment ............................................. 33
4.3 Inserting SD Card.................................................. 7 9.7 2-Color Zero Adjustment ..................................... 34
4.4 Date/Time Setting at Initial Start-up .................. 8 9.8 2-Color Span Adjustment .................................... 34
4.5 Cautions on Measurement .................................... 9 10. Maintenance and Inspection ................................. 35
4.6 Measuring Distance and Measuring Diameter.. 10 10.1 Cleaning of Objective Lens................................ 35
4.7 Targeting .............................................................. 10 10.2 Cleaning of External LCD and
5. Emissivity (ratio) Setting ........................................ 11 Eyepiece Cover ................... 35
5.1 Emissivity (ratio) Setting .....................................11 10.3 Self-Diagnostic Function ................................... 35
5.2 Auto Emmisivity Setting by Thermocouple ....... 12 10.4 Overflow/Underflow Display ............................. 35
6. Setting Mode............................................................. 13 11. Accessories .............................................................. 36
6.1 Lower Limit Temperature Alarm Setting .......... 14 11.1 Thermocouple ..................................................... 36
6.2 Higher Limit Temperature Alarm ...................... 15 11.2 AC Power Adoptor IR-VHRA............................. 36
6.3 Signal Modulation Selection ............................... 16 11.3 Tripod IR-ZBMT ................................................. 36
6.4 Modulation Degree Setting ................................. 16 11.4 Universal head Model: IR-VMS ........................ 36
6.4.1 Modulation Time Constant .............................. 16 11.5 SD card RZ-SMC□□............................................ 36
6.4.2 Attenuation Rate Setting ................................. 17 12. Specifications .......................................................... 37
6.5 Temperature Unit Selection ................................ 17 12.1 Specifications ..................................................... 37
6.6 Thermocouple Measurement Selection .............. 17 12.2 External Dimensions ......................................... 38
6.7 2-color tyep/Single-color wide range type 13. Emissivity Table..................................................... 39
Selection ........... 18 13.1 Emissivity Table ................................................ 39
6.8 Memory Mode Selection ...................................... 19 13.1.1 Emissivity (λ=0.65μm).................................... 39
6.9 Memory Interval Setting..................................... 19 13.1.2 Emissivity (λ=0.9μm)...................................... 40
6.10 Data Save Folder Selection ............................... 20 13.1.3 Emissivity (λ=1.55μm).................................... 40
6.11 Data Save File Selection.................................... 20 14. Start-up Option ...................................................... 41
6.12 Date/Time Setting.............................................. 21 14.1 Start-up Option .................................................. 41
6.13 SD Card Data Initialization.............................. 21 14.2 Table of Display.................................................. 41
7. Measuring Mode ....................................................... 22 15. List of Parameters .................................................. 42
7.1 Standard Measuring Mode.................................. 23
7.2 Continuous Measuring Mode .............................. 24
-C4-
INST.No.INE-567-P0CE

1. Introduction
1.1 General
IR-HA series is small and light weight portable radiation thermometer equipped with well-lighted viewfinder.
Direct viewfinder realizes measurement of an article with small diameter from a distance.
Digital indication is available in the viewfinder, so measured value can be seen while checking an object.
From device selection IR-HAQ which is high function type equipped with both "2-color type+single-color wide
range type", single-color, medium temperature type IR-HAI, and single-color, high temperature type IR-HAS,
user can choose the device which meet their application.

SD card (*)
1.2 Configuration IR-HA□ RZ-SMC□□

[Surface temperature sensor]

Thermocouple (*)
YC599-99K (C510-01K type)

K ribbon thermocouple
Personal
Computer

Pressure type thermocouple (*)


YC599-99K (C510-02K type)

(*): Sold separately

Tripod (*) Universal head (*) AC power adapter (*)


IR-ZBMT IR-VMS IR-VHRA

-1-
INST.No.INE-567-P0CE

2. Model and Attachments


2.1 Model

IR-HA□N□
Thermometer types
I : Single-color type for medium temperature (300 to 1000°C)
S : For single-color type for high temperature (600 to 2000°C)
Q : High function type (2-color type 600 to 2000°C,
single-color wide range type 400 to 3000°C)

Unit
N : °C fixed (for Japan)
E : °C /°F switchable (for overseas)

2.2 Attachments
Name Quantity
Alkaline AA (UM-3) battery 2
Instruction manual 1

-2-
INST.No.INE-567-P0CE

3. Names and Functions of Component parts


3.1 Names and Functions of Component parts


② ③

⑤ ① ⑯ ⑭




⑦ ⑩
⑨ ⑬


Name Function
①Tripod fixing screw 1/4-20UNC, depth 5mm screw hole for mounting tripod.
②SD card connector cover Cover for SD card slot.
③SD card slot SD card (Sold separately: Max 2GB) insertion slot.
④Lens cap/ Strap Cap for protecting objecting lens. It comes with strap.
⑤Front cover Put it on/take it off at changing batteries.
⑥Battery cover Battery (AA/UM-3) change can be done by taking front cover off then turn
battery cover counterclockwise to take off the cover.
⑦Objective lens Effective diameter is φ20mm.
⑧Operation key Keys to switch each mode, select and set various parameters.
⑨Beam attenuation filter A switch whether or not to turn ON beam attenuation filter (attenuation
switch side).
Refer to "4.5 Cautions on Measurement" and make sure to turn the
switch ON when measuring high temperature object, or when you feel
bright light to protect your eyes.
⑩Connector cover Cover to protect the connector. To attach a connector, peel back the cover
from the bottom and put to the designated connector.
⑪External LCD Displays temperature measured value and parameter.
⑫Backlight sensor Detect the surrounding brightness. When it is dark, it turns ON backlight
on the external LCD for visibly display.
⑬Viewfinder There is an aim mark when you see through the viewfinder. Aline the
center of the circle to the center of the measuring object to measure.
⑭DC power supply jack DC power supply jack to connect AC power adaptor sold separately.
⑮Thermocouple input A connector to connect thermocouple.
connector (K-thermocouple)
⑯Nameplate A nameplate to indicate name of the IR-HA series model, serial number,
and measuring range.
For inquiry, make sure to give us the information on the nameplate.
-3-
INST.No.INE-567-P0CE

3.2 External Display Marker and Viewfinder Inside


[Note] When battery mark is
blinking, saving to the SD card is
disabled.
①Main marker ②Status marker So change the batteries.

Center
circle of the
Tb CONT MEM PEAK Aim mark aim mark
③Measuring unit
Main display
display

F
Sub display

④Sub marker
[External LCD] [Viewfinder inside]

Indication in
Name Marker Function
this manual
①Main marker Tb Not used. "Tb"
" " under main marker "CONT" is lit at continuous
CONT "CONT"
measuring mode.
MEM " " under main marker "MEM" is lit at memory input mode. "MEM"
" " under main marker "PEAK" is lit when "PAEK" is
PEAK selected at signal modulation selection. "PEAK"

"Low
It blinks when it is low battery.
battery"
②Status MEAS It lights at measuring. "MEAS"
marker
HOLD It lights at hold. "HOLD"
AL It lights at lower limit temperature alarm activation. "AL"
AH It lights at higher limit temperature alarm activation. "AH"
③Measurement °C It lights at Celsius temperature display. "°C"
unit display o
o
F It lights at Fahrenheit temperature display. " F"
④Sub marker MAX The highest temperature is displayed on sub display. "MAX"

MIN The lowest temperature is displayed on sub display. "MIN"


Average temperature (600 points moving average value) is
AVE "AVE"
displayed on sub display.
If thermocouple temperature is selected, thermocouple
measured temperature data is displayed on sub display. "oFF"
TC "TC"
is displayed if OFF (no thermocouple measurement) is
selected.
It lights when sub display is displaying emissivity (ratio).
ε
ε (emissivity) is lit for single-color type and εr (emissivity "ε" or "εr"
(εr)
ratio) is lit for 2-color type.
Memory data registration number.
NO "NO"
It lights only at memory input mode.
-4-
INST.No.INE-567-P0CE

3.3 Operation Key

①Measurement switch

Power ON and
measurement
start/stop

③Select key ②Memory key

⑥Registration key

④Up key ⑤Down key

Indication in
Name Function
this manual
Turns ON power supply and starts/stops measurement.
①Measurement
Power supply is automatically turned OFF if no key operation is done for MEASURE
switch
30sec.under "HOLD" status.
It switches normal or continuous measuring mode to memory input
②Memory key MEMORY
mode, or memory input mode to normal or continuous measuring mode.
At measuring, it switches data markers to display on sub display and
③Select key at selecting/setting parameter, it switches selection/setting items of sub SEL
display.

④Up key
At selecting parameter, it selects an item.

At setting parameter, it changes numeric value of registered digit.
⑤Down key 

At selecting parameter, it is used to register selecting item.


⑥Registration At setting parameter, it is used to register set value/changing value.
key
ENT
At manual mode, it is used to save the data at that point.
It is not used at interval memory mode.

-5-
INST.No.INE-567-P0CE

4. Preparation for Measurement


4.1 Loading batteries
1) Put something hard like a coin to the slit of front cover to remove the cover.

Insert to
this slit.

Front cover

Caution Do not remove the front cover with your bare hands.
Removing it by your bare hands may injure your fingers and nails.
Be sure to use something hard such as a coin to remove it.

2) Turn battery cover counterclockwise to open.

Objective lens

Battery cover

3) Load dry cell batteries.

Pay attention
to the polarity.
Alkaline AA (UM-3)
dry cell battery
(Comes with
two batteries)

Caution [Battery Life]


・When the batteries are close to their end of life, low battery indication on
the main marker " " blinks, so change them to new AA (UM-3)
batteries.
[Note for Battery Change]
・Change two batteries at the same time.
・Pay attention to the polarities of the batteries and load tem.

Reference Rechargeable battery and manganese dioxide battery can be used.

-6-
INST.No.INE-567-P0CE

4.2 About SD Card


Caution
・Operation under SD card other than our recommended item is not guaranteed.
・Maximum capacity is 2GB, so 2GB or more can not be used.
・If SD card is formatted (initialized) by PC etc., the card may not work properly. Refer to "6.13 SD
Card Data Initialization" for format and follow the procedure.
・Remove and insert SD card when the power supply is turned OFF.
・Do not turn OFF the power supply of the radiation thermometer while it is accessing SD card.
・If battery voltage is low, it can not access to the SD card. Change the battery.
・If memory function is used when the SD card is not inserted, "Er51" is displayed.
・If the SD card is protected by write protection, "LoCK" is displayed when saving the data.

4.3 Inserting SD Card !


Warning Be sure to remove and insert SD card when the power supply is turned OFF.

◆ Open SD card connector cover and insert SD card.

Operation key

SD card connector cover


SD card slot

When setting SD card,


this side up and insert
the card.

Caution If the card is inserted up side down, it may


cause the failure.
SD card

Reference To remove the SD card, push the SD card lightly and pull out the card when the tip of
the card comes out from the "SD card slot".

-7-
INST.No.INE-567-P0CE

4.4 Date/Time Setting at Initial Start-up


◆ Display to set Date/Time appears by turning ON the power supply by holding down MEASURE key.

Remark If no key operation has done for 30sec., power supply is automatically turned OFF.

[Year setting]
1) Set "Year" by  key or  key.
Register by ENT key. Tb CONT MEM PEAK

Main
Caution If year setting is changed, date setting
display
becomes "1" automatically, so change the
date setting as well.
Sub
display

2) Push SEL key to move to "Month" setting.


Select from 01 to 12 by  key or  key [Month setting]
and register by ENT key. Tb CONT MEM PEAK

Caution
If month setting is changed, date setting
becomes "1" automatically, so change the r
date setting as well.

3) Push SEL key to move to "Date" setting. [Date setting]


Select from 01 to 31* by  key or  key
Tb CONT MEM PEAK
and register by ENT key.

Caution
・If year or month setting is changed, next r

setting becomes "1" automatically.


・Number of days differs by each month.

4) Push SEL key to move to "Hour" setting. [Time setting]


Select from 00 to 23 by  key or  key Tb CONT MEM PEAK
and register by ENT key.

Remark Setting range is 0 to 23.

5) Push SEL key to move to "Minute" setting. [Minute setting]


Select from 01 to 59 by  key or  key Tb CONT MEM PEAK
and register by ENT key.

Remark Setting range is 0 to 59.

-8-
INST.No.INE-567-P0CE

4.5 Cautions on Measurement !


◆ To protect the eyes, make sure to turn "ON (attenuation side)" beam attenuation filter
switch if measuring 1500°C or more.

FILTER
ON

[Enlarged view]

FILTER
ON
To the attenuation side!

Warning Never sight the objective lens of the thermometer to the sunlight directly for protecting your eyes
and a detecting element.

Warning ・For the measurement of an object exceeding 1500°C, make sure to turn the beam attenuation
filter switch "ON (attenuation side)" for protecting your eyes.
・When you feel glare on the measurement of an objects lower than 1500°C, turn the beam
attenuation filter switch "ON (attenuation side)".

Caution ・Light path


Be careful not to introduce water drops, dust particles, smoke, steam, or other foreign
substances into the light path between the object measured and the objective lens of the
thermometer.
・Interference causing high indication
Be careful not to apply the direct sunlight, light of an incandescent lamp, flame or other
thermal radiation to the object measured and the objective lens of the thermometer.

-9-
INST.No.INE-567-P0CE

4.6 Measuring Distance and Measuring Diameter


The relation of measuring distance and measuring diameter is shown below.

φ98
φ33 φ20 φ20

4m 0m
5m
10m

4.7 Targeting
For the accurate temperature measurement, it is necessary to target at an object correctly.
The following figures (1), (2), and (3) show the correct targeting based on the relation of the measuring
distance and the measuring diamete

(1
(1))The
測定 距離 が4 mより
measurement (2The
(2) ) 測measurement
定距 離が4 m (3)測
(3) 定距
The 離が4 mより
measurement
近いと き
distance 4m or less distance
のと き4m 遠いと き4m or more
distance

(1) For the measurement distance 4m or less.


Measurement diameter becomes "larger" than the targeting mark.
(2) For the measurement distance 4m.
The measurement diameter and the inner side of the targeting mark are almost "same".
(3) For the measurement distance 4m or more.
Measurement diameter becomes "smaller" than the targeting mark.

-10-
INST.No.INE-567-P0CE

5. Emissivity (ratio) Setting

Caution Emissivity (ratio) setting can not be done while in memory input mode.
For canceling memory input mode, push MEMORY key to turn off " " under main marker

Remark If no key operation has done for 30sec. while in "HOLD" state, power supply is turned OFF
automatically.

5.1 Emissivity (ratio) Setting


If the emissivity (ration) of the object to be measured is low, displayed temperature indicates lower than the
actual measured temperature, so it is necessary for the emissivity (ratio) to be compensated.
Following procedure shows how to set emissivity (ratio) compensation according to the measuring object for
accurate measurement of the thermometer.

1) Turn ON the power by holding down MEASURE key.


[Emissivity (ratio) setting]
Release MEASURE key to display status marker "HOLD".
Tb CONT MEM PEAK
2) Push SEL key several times while in the "HOLD" status and Status
marker HOLD
display "ε" on the sub marker ("εr" for 2-color type).
3) Push  key or  key to change numeric value on the sub
display. ε
4) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed. Sub marker Sub display

Remark ・Setting range single-color: 1.900 to 0.100


2-color: 1.200 to 0.800 (0.001 step)
・Setting of factory default is 1.000.

Reference
・If the emissivity of the object measured is known, set its value.
・If the emissivity is unknown, measure the temperature of the object by thermocouple then
compare them and change emissivity to display the same value. For emissivity of typical
substance, refer to "13. Emissivity Table".

-11-
INST.No.INE-567-P0CE

5.2 Auto Emmisivity Setting by Thermocouple


The thermometer has a function of contact type thermometer using K thermocouple.
Furthermore, the emissivity of the thermometer can be automatically set by assuming the contact
temperature measured by the thermocouple as a true temperature.
For turning "on" thermocouple measurement (thermocouple measurement enable), execute the setting
according to the section "6.6 Thermocouple Measurement Selection".

1) Turn ON the power by holding down MEASURE key.


Release MEASURE key to display status marker "HOLD". [Auto emissivity selection:
Status
K thermocouple]
2) Peel the connector cover and connect the thermocouple. Tb CONT MEM PEAK
marker

Caution HOLD
Pay attention to the thermocouple position and
connect it properly. C

3) Push SEL key several times while in "HOLD" status and display
sub marker "TC". TC
4) Attach the tip of thermocouple to the object to be measured then
push MEASURE key to take measurements by the thermometer
Sub marker Sub display
and by the thermocouple simultaneously. Temperature
measured by the thermocouple is indicated on the sub display.
*There are two types of measuring mode. Refer to "7.1 Standard
Measuring Mode" and "7.2 Continuous Measuring Mode".
5) Emissivity is set automatically, after the measurement, by
pushing ENT key while in the "HOLD" status,
6) By pushing SEL key several times to display sub marker "ε",
you can check the set emissivity.

Reference
Measuring range of the thermocouple is
300 to 800°C (IR-HAI)
600 to 800°C (IR-HAS)
Single-color: 400 to 800°C, 2-color: 600 to 800°C (IR-HAQ)

Remark Thermocouple input setting of factory default is "oFF"


(thermocouple measurement disable).

Caution
If "on" (thermocouple measurement enable) is set, but the thermocouple is not connected, "oFL" is
displayed on the sub display.

-12-
INST.No.INE-567-P0CE

6. Setting Mode
◆ Set each parameter

1) Turn ON the power by holding down MEASURE key.

2) Start setting mode by holding down SEL key while in the "HOLD" status.

Status flow is shown on the right.


For each setting and selection, refer to the individual
sections. Flow Chart

Remark By holding down SEL key or no key operation


Lower limit temperature alarm
is taken for 1 minute, it returns to the
measuring mode. SEL

Higher limit temperature alarm


SEL PEAk
*Items enclosed by may not be displayed
depending on the setting.
Signal modulation
selection
SEL
dELy

Modulation time constant Attenuation rate

SEL SEL

Temperature unit
SEL

Thermocouple measurement enable/disable

SEL

2-color/Single-color wide range selection

(IR-HAQ) SEL

Memory mode
SEL

Memory interval
SEL

Data save folder selection


SEL

Data save file selection


SEL

Date/Time
SEL

SD card data initialize

SEL

-13-
INST.No.INE-567-P0CE

6.1 Lower Limit Temperature Alarm Setting


This setting is for a judgment of lower limit temperature alarm during measurement.
When the alarm is judged, status marker "AL" (lower temperature alarm) is lit and buzzer rings.
If "oFF" is selected, neither the alarm judgment nor the buzzer are activated.

1) Hold down SEL key while in the "HOLD" status to make it to [Lower limit temperature
the setting mode. alarm setting display]
Tb CONT MEM PEAK
Push SEL key several times to display "AL" on the sub display. Main display
2) Push  key or  key to set OFF or numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Reference ・ Setting temperature range is oFF or following


Sub display
numeric values.
300 to 1000°C (IR-HAI) 600 to 2000°C (IR-HAS)
400 to 3000°C (IR-HAQ)
・Default setting is "oFF".

Caution
<How to reset once set lower limit temperature alarm to "oFF">

1) Hold down SEL key to display "AL" on the sub display.


2) By changing displayed numeric value according to above Reference setting temperature
range with operation key (  key,  key or ENT key) to be below lower limit temperature
value, "oFF" is displayed.
(Lower limit temperature value is IR-HAI: 299°C or lower, IR-HAS: 599°C or lower, IR-HAQ:
399°C or lower)

3) By pushing ENT key in this status, blinking stops and "oFF" is displayed and registration is
completed.

-14-
INST.No.INE-567-P0CE

6.2 Higher Limit Temperature Alarm


This setting is for a judgment of higher limit temperature alarm during measurement.
When the alarm is judged, status marker "AH" (higher temperature alarm) is lit and buzzer rings.
If "oFF" is selected, neither the alarm judgment nor the buzzer are activated.
[Higher limit temperature
1) From setting mode display, push SEL key several times to
alarm setting display]
display "AH" on the sub marker. Tb CONT MEM PEAK

2) Push  key or  key to set OFF or numeric value.


3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Reference ・ Setting temperature range is oFF or following


numeric values.
300 to 1000°C (IR-HAI) 600 to 2000°C (IR-HAS)
400 to 3000°C (IR-HAQ)
・Default setting is "oFF".

Caution
<How to reset once set higher limit temperature alarm to "oFF">

1) Hold down SEL key to display "AL" on the sub display.


2) By changing displayed numeric value according to above Reference setting temperature
range with operation key (  key,  key or ENT key) to be below lower limit temperature
value, "oFF" is displayed.
(Lower limit temperature value is IR-HAI: 299°C or higher, IR-HAS: 599°C or higher, IR-HAQ:
399°C or higher)

3) By pushing ENT key in this status, blinking stops and "oFF" is displayed and registration is
completed.

-15-
INST.No.INE-567-P0CE

6.3 Signal Modulation Selection


This setting is used to extract nothing else than average value and maximum value continuously from the
measurement signal (original signal).
[Signal modulation
1) From setting mode display, push SEL key several times to selection display]
display "modu" on the sub display. Tb CONT MEM PEAK

2) Push  key or  key to select signal modulation.


3) Blinking stops by pushing ENT key and the registration is completed.

It displays the value based on the first order lag of modulation degree set at "6.4.1 Modulation time
dELy
constant setting".
It displays the value based on the original signal if temperature is rising.
PEAk If the temperature is dropping, the value set at attenuation rate setting (refer to the section 6.4.2) is
displayed.

Remark
Setting of factory default is "dELy".

6.4 Modulation Degree Setting


If "dELy" is selected in signal modulation selection, degree of first order lag can be adjusted by setting
modulation time constant.
If "PEAk" in signal modulation selection, signal attenuation degree after tracing peak can be set.

6.4.1 Modulation Time Constant


(This is only valid if "dELy" is selected for signal modulation)
1) From setting mode display, push SEL key several times to
[Modulation degree
display "tAu" on the sub display. setting display]
2) Numeric value on the main display blinks in the order of Tb CONT MEM PEAK
0.0→0.2→0.5→1.0 (sec) when  key is pushed and
0.0→1.0→0.5→0.2 (sec) when  key is pushed.
3) Blinking stops by pushing ENT key and the registration is
completed.

Remark Setting of factory default is "0.0"sec. (Displayed value is based on the


original signal without any modulation).

-16-
INST.No.INE-567-P0CE

6.4.2 Attenuation Rate Setting


(This is only valid if "PEAk" is selected for signal modulation)
[Attenuation rate setting
display]
1) From setting mode display, push SEL key several times to
Tb CONT MEM PEAK
display "dEC" on the sub display.
2) Numeric value on the main display blinks in the order of
0→2→5→10(°C/sec) when  key is pushed and
0→10→5→2(°C/sec) when  key is pushed.
3) Blinking stops by pushing ENT key and the registration is
completed.

Remark Setting of factory default is "0"°C/sec (Hold the highest temperature at


measuring (HOLD)).

6.5 Temperature Unit Selection


Select ºC or ºF for measuring temperature unit. [Temperature Unit selection display]

1) From setting mode display, push SEL key several times to Tb CONT MEM PEAK

display "unit" on the sub display.


2) Push  key or  key to select "C (ºC)" or "F (ºF)".
3) Blinking stops by pushing ENT key and the registration is completed.

Remark Setting of factory default is " C (ºC)".

6.6 Thermocouple Measurement Selection


This setting is used to perform measurement with K thermocouple, sold separately (refer to "11.1
Thermocouple").

1) From setting mode display, push SEL key several times to


[Thermocouple measurement
selection display]
display "tC" on the sub display.
Tb CONT MEM PEAK
2) Push  key or  key to select whether or not to perform
thermocouple measurement.
3) Blinking stops by pushing ENT key and the registration is completed.

Remark Setting of factory default is "oFF" (No


thermocouple measurement).

-17-
INST.No.INE-567-P0CE

6.7 2-color tyep/Single-color wide range type Selection


(*High Function Type IR-HAQ)
This setting is used on high function type IR-HAQ to select to use as 2-color type or single-color wide type.

1) From setting mode display, push SEL key several times to [2-color type/single-color wide
display "CoLr" on the sub display. range type selection display]
Tb CONT MEM PEAK
2) Push  key or  key to select 2-color type or single-color wide
range type. Single-color wide range type is set if "1" is selected
and 2-color type is set if "2" is selected.
3) Blinking stops by pushing ENT key and the registration is completed.

Remark Setting of factory default is "2" (2-color type).

-18-
INST.No.INE-567-P0CE

6.8 Memory Mode Selection


This setting is used to select data saving method, manual memory mode or interval memory in memory input
mode.

mAn Manual memory mode: Measured data is saved when ENT key is pushed.
int Interval memory mode: Measured data is saved by every set interval.

Reference This setting in only valid at memory input mode. (Refer to "8.2.1 Saving in Manual Memory Mode"
or "8.2.2 Saving in Interval Memory Mode".)

1) From setting mode display, push SEL key several times to [Memory mode selection display]
display "mmod" on the sub display. Tb CONT MEM PEAK

2) Push  key or  key to select memory mode.


3) Blinking stops by pushing ENT key and the registration is completed.

Remark Setting of factory default is "mAn" (manual memory mode).

6.9 Memory Interval Setting


This setting is used to set saving interval (sec.) of measured data on interval memory mode.

Reference This setting in valid if interval memory is selected at memory input mode.
(Refer to "8.2.2 Saving in Interval Memory Mode")

1) From setting mode display, push SEL key several times to [Memory interval setting display]
display "int" on the sub display. Tb CONT MEM PEAK

2) Push  key or  key to change numeric value.


3) Blinking stops by pushing ENT key and the blinking digit
moves its position from lower place to the higher place and the
registration is completed.

Remark
・Setting range is 1 to 7200sec.
・Default setting is "60"sec.

-19-
INST.No.INE-567-P0CE

6.10 Data Save Folder Selection


This setting is used to select data saving folder. [Data save folder selection
1) From setting mode display, push SEL key several times to display]
display "FLdr" on the sub display. Tb CONT MEM PEAK

2) Push  key or  key to select saving folder.


3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0000 to 9999.


・Default setting is "0000".

6.11 Data Save File Selection


This setting is used to select data saving file. [Data save file selection display]
Tb CONT MEM PEAK
1) From setting mode display, push SEL key several times to
display "FiLE" on the sub display.
2) Push  key or  key to select saving file.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0000 to 9999.


・Default setting is "0000".

-20-
INST.No.INE-567-P0CE

6.12 Date/Time Setting


*If this is not set at initial start-up, the setting is available to set afterward.
This setting is used to set date/time of this thermometer.
[Date/Time setting display]
1) From setting mode display, push SEL key several times to
Tb CONT MEM PEAK
display "YEAr" on the sub display.
2) Push  key or  key to change 'year'.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place. Registration is r
completed by pushing ENT key once again.
4) Set 'month' "Mo"→'day' "dAy"→'hour' "Hour"→'minute' "Min" as
well.

Remark ・Default setting is 'year': "2016", 'month': "1", 'day': "1", 'hour': "0" and 'minute': "0".

Caution If date/time set is initialized at power ON, it may be a result of low built-in lithium battery.
Contact your nearest sales agent of CHINO Corporation or your dealer.

6.13 SD Card Data Initialization


This setting is used to initialize data saved in the SD card when all those data becomes unnecessary or data
exceeds 9999 and it displays 'over' then saving becomes no longer available.
1) From setting mode display, push SEL key several times to
display "FrMt" on the sub display. [SD card data initialization display]
2) "no" is displayed on the main display at this point.
Tb CONT MEM PEAK
By push  key or  key, "no"←→"yES" indication blinks in
the order.
3) To initialize all saved data, select "yES" (delete) and push ENT
key. It displays "dEL" for a moment and then displays "no".
*If there is no data to delete, it displays "non" for a moment
and then displays "no".

4) Now all the data is initialized.

5) If not initializing, select "no" (not delete) and push ENT key.
"no" stops blinking and displays "no" again.
Tb CONT MEM PEAK
6) To end setting mode, push and hold SEL key for about 2 sec.

Remark Setting of factory default is "no" (Not initialize).

-21-
INST.No.INE-567-P0CE

7. Measuring Mode
◆ There are two types of measuring method.

[Standard measuring mode] ······ This mode is for portable measurement.

[Continuous measuring mode] ··· This mode is for continuous measurement which this thermometer is
fixed and measuring for a long time at same location.
" " under main marker "CONT" is lit.

Status flow is shown below.


For measuring method, refer to the individual sections.
◆ Setting Method
Flow Chart
Start the thermometer by pushing  key
and push and hold MEASURE key while
Standard measuring Continuous
power OFF.
mode measuring mode
* under main marker "CONT" lights.

◆ Canceling Method
Highest temperature
indication Start the thermometer by pushing  key
and push and hold MEASURE key while
SEL
power OFF.
* under main marker "CONT" turns
Average temperature
indication OFF.
SEL

[In the case thermocouple measurement is set


Emissivity
to "on"]
SEL Push SEL key while "HOLD" status to display sub
marker TC.
Thermocouple measurement temperature data is
Thermocouple
temperature indication displayed on the sub display;
SEL ・"oFL" is displayed at 1220°C or more
・"uFL" is displayed at -50°C or less

Lowest temperature *"oFL" is displayed if thermocouple is not


indication connected.
SEL *Accuracy is not guaranteed for 800°C to
1220°C.

Warning Never directly sight the objective lens of the thermometer to the sunlight for protecting your eyes
and a detecting element.

Warning ・For the measurement of an object exceeding 1500°C, make sure to turn the beam attenuation
filter switch "ON (attenuation side)" for protecting your eyes.
・When you feel glare on the measurement of an objects lower than 1500°C, turn the beam
attenuation filter switch "ON (attenuation side)".
(Refer to "4.5 Cautions on Measurement")

Remark If no key operation has done for 30sec. while in "HOLD" state, power supply is turned OFF
automatically.

-22-
INST.No.INE-567-P0CE

Status marker
7.1 Standard Measuring Mode [Standard measuring mode
display]
1) Sight through the viewfinder and match the center circle
Tb CONT MEM PEAK
of the targeting circle to the center of object measured.
MEAS
2) Turn ON the power by holding down MEASURE key.

3) Start measurement by pushing MEASURE key. Status
marker "MEAS" lights and temperature is displayed on the
main display. MAX

Measurement continues while holding down on the key.


Push MEASURE key
4) When MEASURE key is released, status marker "MEAS" Sub marker
turned OFF and status marker "HOLD" lights,
measurement is paused and holds measured value. [Pause]
Tb CONT MEM PEAK
5) By pushing SEL key, the sub marker switches "MAX"
HOLD
(highest temperature)→"AVE" (average temperature)→
"TC" (thermocouple temperature) → "MIN" (lowest C
temperature) and each values can be displayed.

Remark MAX
If the measured value is +21°C or more of measuring
range higher limit, "oFL" is displayed, and -20°C or
less of measuring range lower limit, "uFL" is
displayed.

-23-
INST.No.INE-567-P0CE

7.2 Continuous Measuring Mode


1) Start the thermometer by pushing  key and push and hold
MEASURE key while power OFF. [Continuous measuring display]
2) " " under main marker "CONT" lights and it becomes Tb CONT MEM PEAK
continuous measuring mode.
MEAS
3) Sight through the viewfinder and match the center circle of
C
the targeting circle to the center of object measured.
4) Start measurement by pushing MEASURE key.
At this point, status marker "HOLD" turned OFF and MAX
status marker "MEAS" lights and continuous measuring
starts.
Push MEASURE key
5) When MEASURE key is pushed again, measurement is
paused and holds measured value. At this point, status
marker "HOLD" lights. [Pause]
Tb CONT MEM PEAK
6) By pushing SEL key, the sub marker switches "MAX"
(highest temperature) → "AVE" (average temperature) → HOLD
"TC" (thermocouple temperature) → "MIN" (lowest C
temperature) and each values can be displayed.

[Continuous measuring mode canceling method] MAX

Start the thermometer by pushing  key and push and hold


MEASURE key while power OFF. Continuous measuring mode is
canceled and measurement is switched to standard measuring
mode.
*Check that under main marker "CONT" is turned OFF.

Remark
If the measured value is +21°C or more of
measuring range higher limit, "oFL" is displayed,
and -20°C or less of measuring range lower limit,
"uFL" is displayed.

Caution
Continuous measuring mode consumes battery power quickly so it is recommended to use an
accessory AC adopter (model: IR-VHRA) sold separately for the measurement.

-24-
INST.No.INE-567-P0CE

8. Memory Input Mode (Display/Saving Data)


◆ This is a function of display/saving data.

1) Turn ON the power by holding down MEASURE key


2) Push MEMORY key on standard measuring mode or continuous measuring mode under "HOLD" status to
start memory input mode.
" " under main marker "MEM" is lit.

Status flow is shown on the right.


For each setting and selection, refer to the individual
Flow Chart
sections.

Data number
Remark
SEL
・Push MEMORY key to end memory input mode.
It returns to standard measuring mode or
continuous measuring mode. Thermocouple
temperature indication
・It returns to setting mode by pushing and holding
SEL
SEL key.

・If no key operation has done for 30sec., power


supply is turned OFF automatically. Emissivity

SEL

8.1 Data Display


This setting is used to display data already saved (data number, thermocouple temperature, and emissivity).

1) By push MEMORY key on standard measuring mode or [Data number/thermocouple


continuous measuring mode under "HOLD" status to make it to temperature/emissivity data
the memory input mode. " " under main marker "MEM" is display ]
lit. Tb CONT MEM PEAK
2) Sub marker "NO" is displayed.
3) Push  key or  key to select data number to display.
C
4) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and data is
displayed. NO
5) By pushing SEL key, the main marker switches "TC"
(thermocouple temperature indication)→"ε" (emissivity). Perform
procedure 3) to 4) likewise to display thermocouple temperature
and data of emissivity.

Reference "no" is displayed if there is no registered data.

-25-
INST.No.INE-567-P0CE

8.2 Data Save


◆There are two types of data saving method. *Refer to "6.8 Memory Mode Selection" for memory mode
settings.

[Manual memory mode]············ Measured data is saved at the time when ENT key is pushed.

[Interval memory mode] ··········· Measured data is saved at every set interval time.

Caution ・If you are about to save when the number of the file is exceeded, "OVEr" is displayed on the sub
display for a moment.
・If number of the file is exceeding, " " under main marker "MEN" blinks.

Remark
・Measured data is saved in a file or a folder selected at setting mode.
(Refer to "6.10 Data Save Folder Selection" and "6.11 Data Save File Selection ")

Remark ・By pushing SEL key under the status of data is saved, sub marker "TC" and on the sub display
currently measuring thermocouple temperature are displayed. Furthermore, by pushing SEL
key to display sub marker "ε, emissivity (ratio) is displayed on the sub display.
Emissivity displayed at this point is one set at "5. Emissivity (ratio) Setting".
(If the setting has not done, factory default setting 1.000 is displayed.)
・To change emissivity (ratio) setting, make it to the "HOLD" status and set according to "5.
Emissivity (ratio) Setting".
・Emissivity (ratio) set here is emissivity (ratio) of data number from this time forth.
To change emissivity (ratio) from this time forward, reset it by this method.

8.2.1 Manual Memory Mode Saving "mAn" is displayed only


*If the memory mode setting is not "mAn" (manual memory mode), refer for a moment when
to "6.8 Memory Mode Selection" and change memory mode to "mAn". MEMORY key is pushed.

1) By push MEMORY key on standard measuring mode or c


continuous measuring mode under "HOLD" status to make it to [Manual memory mode display]
the memory input mode. " " under main marker "MEM" is lit. Tb CONT MEM PEAK

2) Start measurement by pushing MEASURE key.


Measuring method is a set measuring mode ("Standard
measuring mode" or "Continuous measuring mode").
3) Follow the procedure below for the saving method.
NO

◆ Saving at standard measuring mode: Temperature data, thermocouple temperature data, and emissivity
set value at that point are saved when ENT key is pushed under the measuring status while MEASURE key
is kept push and hold, or under pause while MEASURE key is released.
When "Str" is displayed on the sub display for a moment and when the data is saved, data registration number
becomes next number.

◆Saving at continuous measuring mode: Temperature data, thermocouple temperature data, and emissivity
set value at that point are saved when ENT key is pushed under the continuous measuring status while
MEASURE key is pushed.
When "Str" is displayed on the sub display for a moment and when the data is saved, data registration number
becomes next number.

-26-
INST.No.INE-567-P0CE

8.2.2 Interval Memory Mode Saving


*If the memory mode setting is not "int" (interval memory mode), refer to
"int" is displayed only for
"6.8 Memory Mode Selection" and change memory mode to "int".
a moment when
MEMORY key is pushed.
1) By push MEMORY key on standard measuring mode or
continuous measuring mode under "HOLD" status to make it to
[Interval memory mode display]
the memory input mode. " " under main marker "MEM" is lit.
Tb CONT MEM PEAK

2) Start measurement by pushing MEASURE key.


Measuring method is a set measuring mode ("Standard
measuring mode" or "Continuous measuring mode").

3) Follow the procedure below for the saving method.


NO

◆ Saving at standard measuring mode: Temperature data, thermocouple temperature data, and emissivity
set value are saved at preset interval under the measuring status when MEASURE key is pushed.
Each time, "Str" is displayed on the sub display for a moment and when the data is saved, data registration
number becomes next number.

◆Saving at continuous measuring mode: Temperature data, thermocouple temperature data, and emissivity
set value are saved at preset interval under the continuous measuring status when MEASURE key is pushed.
When "Str" is displayed on the sub display for a moment and when the data is saved, data registration number
becomes next number.

-27-
INST.No.INE-567-P0CE

8.3 Data File and Folder


[SD card (Maximum2GB)]

Folder
1 file

× Maximum 512 folders

9999 files
9999 data

Remark ・For one file, in the order from 1 to 9999 data can be saved.

Caution
・Measured data is saved in a folder or a file set at setting mode.
・Past data is not available to be overwritten.
・If saving data exceeds 9999 data by continuous measurement etc, a new file will be created.
New file name will be "○○○○-01.csv" and forth until "○○○○-99.csv" saving is available
continuously. When saving data of "○○○○-99.csv" exceeds 9999 data, 'over' is displayed and
continuous saving becomes unavailable. Furthermore, data saved in "○○○○-○○.csv" is not
available to be replayed by the radiation thermometer. To replay the data, use PC etc.
・If saved csv file is edited by PC etc., files may not be recognized by the thermometer.
・Number of the file to be saved in the SD card is limited. So try not to save the files other than
measured data.
・If data can not be saved in the SD card although there is enough free space in the card, it may
be a result of far too many files in the card. Delete some files or move some data to PC.

-28-
INST.No.INE-567-P0CE

8.4 File Saving Format


File Saving Format

Data Radiation Thermocouple


Field Date/Time Status Emissivity CR LR
number temp. temp.
Data string 1 2 3 4 , 2 0 1 7 / 0 1 / 2 3 ? 1 2 : 3 4 : 5 6 , 0 1 3 0 , 1 . 0 0 0 , ? ? 5 0 , ? ? 6 1 ¥r ¥n
Number of
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
character
* ? is for single byte space.

Description
Remarks
Data number ???1 to 9999
Date/Time yyyy/mm/dd?hh:mm:ss
Status □□□□
(1)(2)(3)(4)

(1)Radiation temperature status information


0: All clear
1: Overflow (OFL)
2: Underflow (UFL)
3: Abnormal thermometer
(2)Unit information
0: °C
(3)Thermocouple status information
0: All clear
1: Overflow (OFL)
2: Underflow (UFL)
3: OFF
(4)Spare (fixed to 0)
Emissivity 0.100 to 1.900
Radiation *Save "9999" if radiation thermometer status information
temp. is overflow/underflow.
Thermocouple *Save "9999" if radiation thermometer status information
temp. is overflow/underflow/OFF.

-29-
INST.No.INE-567-P0CE

9. Zero/Span Adjustment Mode


◆ By measuring scale lower limit (zero side) and higher limit (span side) with your black body furnace and
input black body furnace temperature, this function performs zero/span adjustment.

Caution ・Be sure to prepare a black body furnace and a reference radiation
thermometer.
If those instruments can not be provided, do not perform zero/span
adjustment.
・Check that emissivity (ratio) is 1.000 then perform the adjustment.

・Start zero/span adjustment by push and holding  key, ENT key, and MEASURE key at the same time
while power OFF.

Status flow is shown on the right. Flow Chart


For each setting and selection, refer to the individual
sections.
Initialize Zero/Span

Remark SEL
If no key operation has done for 30sec., power
supply is turned OFF automatically.

Perform Zero/Span
Adjustment
SEL

InGaAs Zero Adjustment


(IR-HAI/IR-HAQ) SEL

InGaAs Span Adjustment


(IR-HAI/IR-HAQ) SEL

Si Zero Adjustment
(IR-HAS/IR-HAQ)
SEL

Si Span Adjustment
(IR-HAS/IR-HAQ) SEL

2-Color Zero Adjustment


(IR-HAQ) SEL

2-Color Span Adjustment


(IR-HAQ) SEL

-30-
INST.No.INE-567-P0CE

9.1 Initialize Zero/Span Adjustment


This setting is used to initialize zero/span adjustment values.
[Initialize Zero/Span
Adjustment display]
1) Start zero/span adjustment by push and holding  key, ENT Tb CONT MEM PEAK

key, and MEASURE key at the same time while power OFF.
2) Display "init" on the sub display.
3) Push  key or  key to select "on (initialize)" or "oFF (not
initialize)".
4) Registration is completed by pushing ENT key.

Remark ・ "Init" is display while initializing and key operation is disabled until initialization
completes.
・Setting of factory default is "oFF".

Caution If "on" is selected, all zero/span adjustment values are initialized and return to factory
default.

9.2 Perform Zero/Span Adjustment


Performs zero/span adjustment.
Zero/span adjustment result is reflected to the measured value if
"yES" is set. [Zero/Span adjustment display]
Tb CONT MEM PEAK
1) From zero/span adjustment mode display, push SEL key several
times to display "CAL" on the sub display.
2) Push  key or  key to select "yES (perform)" or "no (not
perform)".
3) Registration is completed by pushing ENT key.

Caution If "on" is selected, zero/span adjustment result is not reflected to the


measured value.

Remark ・Setting of factory default is "no".

-31-
INST.No.INE-567-P0CE

9.3 InGaAs Zero Adjustment (*IR-HAI or IR-HAQ)


Performs InGaAs zero adjustment. [InGaAs zero adjustment
Collimate the thermometer and sight the black body furnace to display]
input the temperature of the furnace. Tb CONT MEM PEAK

1) From zero/span adjustment mode display, push SEL key several


times to display "inGZ" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default for IR-HAI and IR-HAQ is 300 and 400.

Caution At setting, if zero adjustment value (input temperature or measured temperature) exceeds span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

9.4 InGaAs Span Adjustment(*IR-HAI or IR-HAQ)


Performs InGaAs span adjustment. [InGaAs span adjustment
Collimate the thermometer and sight the black body furnace to display]
input the temperature of the furnace. Tb CONT MEM PEAK

1) From zero/span adjustment mode display, push SEL key several


times to display "inGS" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default for IR-HAI and IR-HAQ is 1000 and 600.

Caution At setting, if span adjustment value (input temperature or measured temperature) is below span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

-32-
INST.No.INE-567-P0CE

9.5 Si Zero Adjustment (*IR-HAS ot IR-HAQ)


Performs Si zero adjustment. [Si zero adjustment display]
Collimate the thermometer and sight the black body furnace to
Tb CONT MEM PEAK
input the temperature of the furnace.

1) From zero/span adjustment mode display, push SEL key several


times to display "Si Z" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default for IR-HAS and IR-HAQ is both 600.

Caution At setting, if zero adjustment value (input temperature or measured temperature) exceeds span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

9.6 Si Span Adjustment (*IR-HAS or IR-HAQ)


Performs Si span adjustment.
Collimate the thermometer and sight the black body furnace to
[Si span adjustment display]
Tb CONT MEM PEAK
input the temperature of the furnace.

1) From zero/span adjustment mode display, push SEL key several


times to display "Si S" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default for IR-HAS and IR-HAQ is 2000 and 3000.

Caution At setting, if span adjustment value (input temperature or measured temperature) is below span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

-33-
INST.No.INE-567-P0CE

9.7 2-Color Zero Adjustment (*IR-HAQ)


Performs 2-color adjustment.
Collimate the thermometer and sight the black body furnace to [2-color adjustment display]
input the temperature of the furnace. Tb CONT MEM PEAK

1) From zero/span adjustment mode display, push SEL key several


times to display "2C Z" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default is 600.

Caution At setting, if zero adjustment value (input temperature or measured temperature) exceeds span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

9.8 2-Color Span Adjustment (*IR-HAQ)


Performs 2-color adjustment.
Collimate the thermometer and sight the black body furnace to [2-color adjustment display]
input the temperature of the furnace. Tb CONT MEM PEAK

1) From zero/span adjustment mode display, push SEL key several


times to display "2C S" on the sub display.
2) Push  key or  key to change numeric value.
3) Blinking stops by pushing ENT key and the blinking digit moves
its position from lower place to the higher place and the
registration is completed.

Remark ・Setting range is 0 to 9999.


・Setting of factory default is "2000".

Caution At setting, if span adjustment value (input temperature or measured temperature) is below span
adjustment value (input temperature or measured temperature), setting becomes disabled and
displays 'Er6'.

-34-
INST.No.INE-567-P0CE

10. Maintenance and Inspection

10.1 Cleaning of Objective Lens


Wipe the objective lens periodically with a soft cloth for accurate measurement.

10.2 Cleaning of External LCD and Eyepiece Cover


Wipe the external LCD and eyepiece cover periodically with a soft cloth for clear view.

10.3 Self-Diagnostic Function


This thermometer has self-diagnostic function and displays following indications at abnormality.
Check the contents and take countermeasure.
Error Display
Contents Alarm Countermeasure
(Main Display)

Higher limit over range Status marker Isn't the emissivity set too low?
(The object temperature exceeds AH lights and the Set the correct emissivity by referring
the measuring range of the buzzer turned "5.1 Emissivity (ratio) Setting" and "13.
thermometer) "on". *1 Emissivity Table".
Main Display (Thermometer)

Lower limit under range Status marker AL Isn't the emissivity set too high?
The object temperature is no lights and the Set the correct emissivity by referring
more than the measuring range buzzer turned "5.1 Emissivity (ratio) Setting" and "13.
of the thermometer) "on". *1 Emissivity Table".

Abnormal ambient temperature


(The thermometer is placed in
Only "Er1" is Use the thermometer within the
the environment other than its
displayed. working temperature rage 0 to 50°C.
working temperature: 0°C or
lower, and 50°C or higher)
E2PROM data broken
*2 Re-adjustment is necessary.
(For some reason, data ROM is
Only "Er4" is Send the thermometer to us.
broken and that causes broken
displayed. (Memory data and adjustment data are
memory data, temperature data,
initialized.)
and adjustment data)

10.4 Overflow/Underflow Display


Error Display
Contents Alarm Countermeasure
(Sub Display)
・If "OFL" is displayed in room
Higher limit over range temperature, it means the
Sub Display (TC temperature)

thermocouple is broken. Replace it.


(・ Thermocouple is disconnected
・If "OFL" is displayed while measuring,
or, Only "OFL" is
it means the thermocouple is in the
・(Object) temperature measured displayed.
temperature 1220°C or higher. The
by the thermocouple exceeds
thermocouple may get broken. Stop
1220°C)
measuring by the thermocouple
immediately.
・The object temperature exceeds the
Lower limit under range lower limit measuring range of the
((Object) temperature measured Only "UFL" is thermocouple.
by the thermocouple is below displayed. It may deteriorate the thermocouple.
-50°C) Stop the measuring by the
thermocouple immediately.
*1 This is not output, if alarm value is set to "oFF".
*2 This is displayed at power ON or while displaying memory data.

-35-
INST.No.INE-567-P0CE

11. Accessories
11.1 Thermocouple
The thermometer can be used as a surface thermometer by connecting a K-thermocouple (sold separately).
Connect the thermocouple by inserting its connector to the connector inside the connector cover.
Refer to the figure of "3.1 Names and Functions of Component parts" for the location of thermocouple input
connector.

Name Model Specification


Thermocouple C510-01K K-thermocouple, maximum 500°C, response time about 1.8 sec.
Thermocouple for high C510-02K K-thermocouple, maximum 800°C, response time about 4 sec.
temperature

11.2 AC Power Adoptor (Model: IR-VHRA) !


This is used when using this thermometer by AC power supply.
For connection, insert the plug of the adapter into the DC power jack inside the connector cover.
Refer to the figure of "3.1 Names and Functions of Component parts" for the location of the DC power jack.

Reference ・Specifications: input voltage 100 to 240V AC, output voltage 5V DC

Warning ・Makes sure to use the AC adaptor in the rage of 100 to 240V AC.
・Do not touch the AC adapter or an electrical outlet with wet hands.
It may cause electric shock, fire or damage.
・Do not wet the AC adaptor. It may cause fire.
・Wipe the dust on the AC adaptor. It may cause fire.

Caution ・Make sure to turn OFF the thermometer to connect the AC adaptor.

11.3 Tripod (Model: IR-ZBMT)


This is used for long term measurement or for fixing the thermometer for measurement.
Use tripod fixing screw in the bottom to mount to the universal head.

11.4 Universal head (Model: IR-VMS)


This is used by placing between the tripod and the thermometer for adjustment of its direction and title or for
fixing.

11.5 SD card (Model: RZ-SMC□□)


Operation of SD card other than our recommended product is not guaranteed.
Capacity of the SD card is maximum 2GB and 2GB or more can no be used with this thermometer.

-36-
INST.No.INE-567-P0CE

12. Specifications
12.1 Specifications
High function type
Single-color type for Single-color type for high
Type (2-color type+single-color
medium temperature temperature
wide rang type)
Model IR-HAI IR-HAS IR-HAQ
Measuring System Narrow band radiation thermometer
Detecting Element InGaAs Si Si/InGaAs
Measuring
1.55μm 0.9μm 0.9/1.55μm
Wavelength
Measuring Range 300 to 1000°C 600 to 2000°C 600 to 2000°C (2-color)
400 to 3000°C (Single-color)
Accuracy Rating* ±6°C Below 1000°C: ±6°C Below 1000°C: ±6°C
1000 or higher, below 1500°C 1000 or higher, below 1500°C
: ±0.6% of measured value : ±0.6% of measured value
1500°C or higher 1500 or higher, below 2000°C
: ±1.2% of measured value : ±1.2% of measured value
2000°C or higher
: ±2.4% of measured value
Reproducibility ±1°C
Resolution 1°C
Response Time 0.2s
Emissivity (ratio)
Emissivity correction: 1.900 to 0.100, Emissivity ratio correction: 1.2 to 0.8
Correction
Mathematics Maximum value, Minimum value, Average value
Signal Modulation Delay, Peak
Display System LCD digital 4 digit (External display and viewfinder inside display)
Data Memory Media SD card (Max 2GB)
Memory Data 1 File maximum 9999 data (Number of maximum folder: 512, 1 folder maximum file
number: 9999)
Optical System Fixed focus lens type
Measuring distance 4000mm or less: Measuring diameter Φ20mm
Measuring Distance
Measuring distance 4000mm or more: Measuring diameter Φ(0.013×(Measuring
and Diameter
distance‐4000)+20)mm
Targeting System Direct viewing finder
Lens Diameter Φ20mm
Thermocouple Input K-thermocouple:‐50 to 800°C, Accuracy rating: ±2°C (in 23°C±5°C)
Auto power OFF, LCD backlight, continuous measuring, battery check,
Other Functions
higher/lower limit alarm
Key Switch 6 switches (MEASURE, MEMORY, SEL, ▼, ▲, ENT)
Working
0 to 50°C
Temperature Range
Power Supply Two AA (UM-3) dry cell batteries (rechargeable battery can be used), or AC power
adaptor (sold separately)
Case Material and
Polycarbonate (heat resistant resin is used for a part), white, gray
Color
Weight About 450g (thermometer only)
CE marking EN61326-1 Class B
*At ε=1.0, reference operating condition: 23°C±5°C, relative humidity: 35 to 75%RH

-37-
INST.No.INE-567-P0CE

12.2 External Dimensions

Unit: mm

-38-
INST.No.INE-567-P0CE

13. Emissivity Table


The emissivity are values determined by the material of object, profile of its surface, surface roughness,
oxidized or not, measuring temperature, measuring wavelength and other factors. They are represented by
the thermal radiation ratio "ε" when a black body furnace at the same temperature is measured in the same
wavelength band.

The emissivity "ε" is generally known by a value at the wavelength of 0.65μm when an optical pyrometer is
used. The emissivity changes according to the above factors even in case of the same material. Please use the
following table as a reference.

13.1 Emissivity Table


13.1.1 Emissivity (λ=0.65μm)
Emissivity
Metal Oxide Emissivity
Solid Liquid
Zinc 0.42 ― Alumel* 0.87
Alumel 0.37 ― Chromel* 0.87
Aluminum 0.17 0.12 Constantan* 0.84
Antimony 0.32 ― Ceramics 0.25 to 0.5
Iridium 0.30 ― Cast iron* 0.70
Yttrium 0.35 0.35 55Fe. 37.5Cr. 7.5Al* 0.78
Uranium 0.54 0.34 70Fe. 23Cr. 5Al. 2Co* 0.75
Gold 0.14 0.22 80Ni. 20Cr* 0.90
Silver 0.07 0.07 60Ni. 24Fe. 16Cr* 0.83
Chromium 0.34 0.39 Stainless steel* 0.85
Chromel P 0.35 ― Aluminum oxide 0.22 to 0.4
Cobalt 0.36 0.37 Yttrium oxide 0.60
Constantan 0.35 ― Uranium oxide 0.30
Zirconium 0.32 0.30 Cobalt oxide 0.75
Mercury ― 0.23 Columbium oxide 0.55 to 0.71
Tin 0.18 ― Zirconium oxide 0.18 to 0.43
Carbon 0.8 to 0.9 ― Tin oxide 0.32 to 0.60
Tungsten 0.43 ― Cerium oxide 0.58 to 0.82
Tantalum 0.49 ― Titanium oxide 0.50
Cast iron 0.37 0.40 Iron oxide 0.63 to 0.98
Titanium 0.63 0.65 Copper oxide 0.60 to 0.80
Iron 0.35 0.37 Thorium oxide 0.20 to 0.57
Copper 0.10 0.15 Vanadium oxide 0.70
Thorium 0.54 0.34 Beryllium oxide 0.07 to 0.37
Nickel 0.36 0.37 Magnesium oxide 0.10 to 0.43
80Ni. 20Cr 0.35 ―
60Ni. 24Fe. 16Cr 0.36 ― *Oxidized on surface
Platinum 0.30 0.38
90Pt. 10Rh 0.27 ―
Palladium 0.33 0.38
Vanadium 0.35 0.35
Bismuth 0.29 ―
Beryllium 0.61 0.61
Manganese 0.59 0.59
Molybdenum 0.37 0.40
Rhodium 0.24 0.30

-39-
INST.No.INE-567-P0CE

13.1.2 Emissivity (λ=0.9μm) 13.1.3 Emissivity (λ=1.55μm)


Metal Emissivity Metal Emissivity
Aluminum 0.10 to 0.23 Aluminum 0.09 to 0.40
Gold 0.015 to 0.02 Chrome 0.34 to 0.80
Chrome 0.36 Cobalt 0.28 to 0.65
Cobalt 0.28 to 0.30 Copper 0.05 to 0.80
Iron 0.33 to 0.36 Gold 0.02
Copper 0.03 to 0.06 Steel plate 0.30 to 0.85
Tungsten 0.38 to 0.42 Lead 0.28 to 0.65
Titanium 0.50 to 0.62 Magnesium 0.24 to 0.75
Nickel 0.26 to 0.35 Molybdenum 0.25 to 0.80
Platinum 0.25 to 0.30 Nickel 0.25 to 0.85
Molybdenum 0.28 to 0.36 Palladium 0.23
Platinum 0.22
Rhodium 0.18
Alloy Emissivity
Silver 0.04 to 0.10
Inconel X 0.40 to 0.60
Tantalum 0.20 to 0.80
Inconel 600 0.28
Tin 0.28 to 0.60
Inconel 617 0.29
Titanium 0.50 to 0.80
Inconel 0.85 to 0.93
Tungsten 0.30
Incoloy 800 0.29
Zinc 0.32 to 0.55
Kanthal 0.80 to 0.90
Stainless steel 0.30
Hastelloy X 0.30 Alloy Emissivity
Brass 0.18 to 0.70
Chromel, Alumel 0.30 to 0.80
Semi conductor Emissivity
Constantan, Manganin 0.22 to 0.60
Sillicon 0.69 to 0.71
Incone 0.30 to 0.85
Germanium 0.60
Monel 0.22 to 0.70
Gallium arsenic 0.68
Nickel Chrome 0.28 to 0.85

Ceramics Emissivity
Sillicon carbide 0.80 to 0.83 Ceramics Emissivity
Titanium carbide 0.47 to 0.50 Alumina ceramics 0.30
Silicon nitride 0.89 to 0.90 Red brick 0.80
White brick 0.35
Silicon brick 0.60
Others Emissivity Sillimanite brick 0.60
Carbon pigment 0.90 to 0.95 Ceramics 0.50
Graphite 0.87 to 0.92

Others Emissivity
Asbestos (plate, sheet, fabric) 0.90
Asphal 0.85
Carbo 0.85
Graphite 0.80
Soot 0.95
Cement, Concrete 0.70
Cloth 0.80

-40-
INST.No.INE-567-P0CE

14. Start-up Option


14.1 Start-up Option
The following operation modes are available by the key combinations at the start up.
Key Operation Mode Remarks
Measurement with the same
Standard measuring or
Push MEASURE key only measuring mode as the last
Continuous measuring
start-up
Push  key while pushing MEASURE
Standard measuring
key
Push  key while pushing MEASURE " " under main marker "CONT"
Continuous measuring
key lights
If it is start-up by only MEASURE
Push  key, ENT key, and MEASURE Zero/Span adjustment key at next start-up, it will be
key at the same time mode standard measuring or continuous
measuring.

14.2 Table of Display


The displays indicated on the external LCD are classified as three types.
Mode Outline
Push MEASURE key to perform measurement.
Measuring mode Emissivity setting and automatic calculation of emissivity is available at
holding standard/continuous measuring display.
Displays and sets measurement parameter.
By pushing and holding SEL key in standard/continuous measuring mode or
Setting mode memory input mode, it shifts to setting mode.
In any setting mode, push and hold SEL key or not operating any key for about
one minute, it becomes measuring mode or memory input mode.
Memory number setting, displaying memory data, and saving memory data is
available.
By pushing MEMORY key at standard/continuous measuring mode, it becomes
Memory input mode
memory input mode and lights "MEM".
By pushing MEMORY key again, it returns to standard/continuous measuring
mode.
Performs zero/span adjustment using black body furnace.
Zero/Span adjustment
If no key operation has done for 30sec., power supply is turned OFF
mode
automatically.

-41-
INST.No.INE-567-P0CE

15. List of Parameters


Sub Section
Name Parameter Name Setting range Initial value of this
display manual
Emissivity
(ratio)

Single-color: 1.900 to 0.100


Emissivity (ratio) setting
2-color: 1.200 to 0.800
1.000 5.1

"oFF" or
Lower limit temperature alarm AL 300 to 1000°C (IR-HAI) 6.1
600 to 2000°C (IR-HAS) oFF
Higher limit temperature alarm AH 400 to 3000°C 6.2
(IR-HAQH)
Signal modulation*1 Modu dELy/PEAk dELy 6.3

Modulation time constant tAu 0.0, 0.2, 0.5, 1.0sec. 0.0sec. 6.4.1

Attenuation Rate dEC 0, 2, 5, 10°C/sec. 0°C/sec. 6.4.2

Measuring unit selection unit C, F C 6.5

Thermocouple measurement tC oFF/on oFF 6.6


Setting mode

2-color type/Single-color wide


CoLr 2, 1 2 6.7
range type*2
Memory mode mmod mAn/int mAn 6.8

Memory interval int 1 to 7200sec. 60sec. 6.9

Data saving folder FLdr 0000 to 9999 0000 6.10

Data saving file FiLE 0000 to 9999 0000 6.11


yEAr 2016 to 2099 2016
mo 1 to 12 1
Date/Time dAy 1 to 31 1 6.12
Hour 0 to 23 0
min 0 to 59 0
SD card data initialization Frmt no/yES no 6.13
Initialize zero/span
init oFF/on oFF 9.1
adjustment
Perform zero/span adjustment CAL no/yES no 9.2
Zero

300 (IR-HAI)
InGaAS zero adjustment InGZ 0 to 9999 9.3
400 (IR-HAQH)
Span adjustment mode

1000 (IR-HAI)
InGaAS span adjustment InGS 0 to 9999 9.4
600 (IR-HAQH)
600 (IR-HAS,
Si zero adjustment Si Z 0 to 9999 9.5
IR-HAQH)
2000 (IR-HAS)
Si span adjustment Si S 0 to 9999 9.6
3000 (IR-HAQH)
2-color zero adjustment 2C Z 0 to 9999 600 (IR-HAQH) 9.7

2-color span adjustment 2C S 0 to 9999 2000 (IR-HAQH) 9.8

*1: The parameters of modulation ratio selection differ by the signal modulation mode.
Refer to "6.3 Single Modulation Selection" for details.
*2: 2-color/single-color wide selection is displayed in Model IR-HAQ only

-42-
32-8, KUMANO-CHO, ITABASHI-KU, TOKYO 173-8632

Telephone:+81-3-3956-2171
Facsimile:+81-3-3956-0915
Web site http://www.chino.co.jp/