Excel Inst en
Excel Inst en
December 2002
Manufactured by SOREDEX
P.O. BOX 250
FIN-00031 SOREDEX, FINLAND
Tel. +358 10 394 8412
Fax. + 358 9 701 5261
Contents
1 Introduction ..................................................................................... 1
General ............................................................................................. 1
Using this manual and the separate appendices................................. 2
Associated documents ...................................................................... 3
Warnings and precautions ................................................................. 4
Preinstallation requirements ............................................................... 5
General .......................................................................................................... 5
Film version .................................................................................................... 6
Imaging plate versions .................................................................................... 6
Digital Versions ............................................................................................... 6
Unit description .................................................................................. 7
Panoramic units .............................................................................................. 7
Cephalometric units (Excel Ceph only) ........................................................... 8
Unit dimensions and space requirements ........................................... 9
Installation tools and hardware ......................................................... 11
Standard tools ............................................................................................... 11
Special tools .................................................................................................. 11
Hardware ...................................................................................................... 11
Overview of the installation and alignment procedure ....................... 12
2 Installing the unit .......................................................................... 13
Unpacking the unit ........................................................................... 13
Attaching the wall bracket to the wall ................................................. 13
Attaching the column to the wall bracket ........................................... 14
Installing the rotating unit .................................................................. 16
Preparing the cassette carriage drive mechanism ............................ 21
Checking the position of the follower ............................................................ 23
Removing the counterweight pin ....................................................... 24
Levelling the unit .............................................................................. 25
Checking and connecting the power supply ...................................... 26
1 Introduction
General
This manual and the two separate appendices de-
scribe how to install and set up the different versions
of the Cranex dental panoramic x-ray series.
Associated documents
“Appendix M - Checking and aligning the Cranex
Excel D (X-Mind Pano D) and Cranex BaseX D (X-
Mind Pano 2.5 D)”
For units with serial numbers ending with the letters
DM.
• When installing a dental x-ray unit always observe local and national safety,
radiation control and electrical regulations.
• Do not install a dental x-ray unit in an environment where there are corrosive or
explosive vapours.
• When checking the alignment of the radiation beam take the necessary steps to
protect yourself from the radiation.
• The aperture plate in the collimator is made of lead (Pb) which is a toxic mate-
rial. Do not touch it with your bare hands.
Preinstallation requirements
General
The room in which the unit is to be installed and the
position from where the user will take exposures
must be correctly shielded from the radiation that is
generated when the unit is operated. Since radiation
safety requirements vary from country to country
and state to state it is the responsibility of the in-
staller to ensure that all safety regulations are met.
Make sure that the floor in the room where the unit is
to be used can support the weight of the unit.
Film version
Make sure that the darkroom and processing equip-
ment are correctly set up and in working order.
Digital Versions
Make sure that the PC to which the unit is to be
connected is correctly set up and that the imaging
handling software being used is installed and operat-
ing.
For information on how to do this refer to appendix
8200910DM, if you are installing a unit with a serial
number ending in DM, or appendix 8200910DS if you
are installing a unit with a serial number ending in
DS.
Unit description
Panoramic units
Standard tools
- Electric drill
- Drill for hardware being used to attach the unit to
the wall
- 17 and 19 mm AF wrenches (spanners)
- Allen keys (Hexagon socket wrenches) to fit allen
screws from M1.5 to M8
- Flat blade and Philips screwdrivers
- Spirit level
- Pliers and wire cutters
- Digital multimeter (e.g. Fluke)
- Torx screwdriver (supplied)
Special tools
Panoramic units:
- Ball and pin phantom (part number 6900029)
- Beam bar (part number 6900027)
- Fluorescent screen (PAN) (part number 6900028)
- Radiation shield (PAN) (part number 6902210)
- Calibration filter (DM digital units only) 6904270
Hardware
Suitable mounting hardware for attaching the column
to the wall. Mounting screws or bolts must be at
least 10 mm (3/8”) diameter and each bolt must
withstand a pull-out force of 160 kg (350 lb).
WARNING
The wall and each fastening bolt must be able to
support a pull out force of at least 160 kg (350
lb).
4. Slide the rotating unit into the upper shelf until the
two M4 holes in the X-Y slide assembly can be
seen through the rectangular aperture in the
upper shelf. Rotate the focal trough positioning
knob on the side of the lower shelf until the two
holes in the connection bar line up with the two
holes in the X-Y slide assembly. Attach the con-
nection bar to the X-Y slide assembly with the
two M4 screws and washers.
DM units ONLY
For units with serial numbers ending with DM.
Place the hub (supplied separately) in the upper
shelf and connect the power supply cable and
two ethernet cables to the hub. The ethernet
cables can be connected to any of the hub ports.
The other cables are connected as described
previously.
DS units ONLY
For units with serial numbers ending with DS.
Connect the two interface cables together and
then the two power supply cables. The other
cables are connected as described previously.
2. CAUTION
DO NOT use sharp tools when performing the
next step as they could damage the rubber
surface.
Remove the protective covering from the cas-
sette roller.
CAUTION
If there is resistance when you turn the rotating unit
STOP, DO NOT use force. The profile follower may
be in the wrong position. To check, refer to the next
step.
See steps:
- “Routing the cables through the cephalometric
arm” on page 52
IMPORTANT NOTE
If the mains resistance is significant, the input
voltage will drop below the required level when
an exposure is taken. In this case, select a
terminal one step lower to ensure proper opera-
tion.
CORRECT
If the beam is in the correct position push the TA
switch down to the “off” position and then check
the position of the focal trough. See “Checking
the position of the focal trough” on page 32.
NOT CORRECT
IF the beam is NOT in the correct position, it is
too high or low, too far to one side or not verti-
cal, then the position of the primary slit must be
adjusted.
See the next section, “Adjusting the position of
the panoramic beam”.
CORRECT
If the light beam is on the middle of the balls, it
indicates that the focal trough is correctly posi-
tioned.
The length and symmetry of the image must now
be checked. See “Checking the image” on page
37
NOT CORRECT
If the light beam does NOT touch the middle of
the balls, the position of the focal trough must be
adjusted. See the next section, “Adjusting the
position of the focal trough”.
CORRECT
If the ball/pin image is correct the unit is correctly
aligned and the panoramic head support can now
be installed. See “Installing the panoramic head
support” on page 44.
NOT CORRECT
If the ball/pin image is not correct the image
must be adjusted.
If the image is not symmetrical, its symmetry
must be adjusted. To do this see “Adjusting the
symmetry of the image” on page 39.
Panoramic units
1. Remove the screws that hold the front panel in
position, and support the front panel so that you
can access the Pulse Logic Unit (PLU)
CAUTION
Use an insulated tool to support the front panel.
Switch the TA switch up to the “TEST” position.
Cephalometric units
If you are installing a cephalometric unit also check
and adjust the cephalometric kV/mA levels.
CORRECT
If the rings are correctly positioned, the align-
ment of the x-ray field must be checked.
See section “Checking the position of the X-ray
field” on page 59
NOT CORRECT
If the rings are not concentric to within 1 mm, the
earposts must be realigned.
Refer to the next section, “Realigning the
earposts”.
CORRECT
If the exposed area is correctly positioned, a
cephalometric exposure time test must be car-
ried out. See “Cephalometric exposure time test”
on page 62.
NOT CORRECT
If the area is not in the correct, too far to one
side, too high or low or tilted, the x-ray field
must be repositioned. Refer to the next section,
“Repositioning the x-ray field”.
NOTE
If you cannot align the x-ray field enough by
adjusting the collimator, you will have to adjust
the cephalometric tube. Refer to the Excel Serv-
ice Manual for information on how to do this.
Time
0.4 s ±15%
0.8 s ±15%
1.6 s ±15%
3.2 s ±15%
6 Technical specifications
Classification
Complies with IEC class I type B, IPX0
Complies with IEC 601-1, IEC 601-2-7 and EN 55011 standards
Group 1, class B
Complies with DHHS Radiation Performance Standard, 21CFR Subchapter J.
Description
Dental panoramic and panoramic/cephalometric x-ray units with a high frequency
switching mode x-ray generators. The panoramic version takes panoramic expo-
sures. The panoramic/cephalometric version takes panoramic and cephalometric
exposures and uses a dedicated x-ray source to take cephalometric exposures.
Generator
TUBE
- OPX/105, DE 100/15 or equivalent
FOCAL SPOT
- 0.5 mm IEC 336
TARGET ANGLE
- 5º
TARGET MATERIAL
- Tungsten
OPERATING TUBE POTENTIAL
- Panoramic 63 - 81 kV (±5 kV) at preselected settings
OPERATING TUBE CURRENT
- 6 mA and 10 mA (±1 mA) at preselector,
4.5 mA and 6 mA or 7.5 mA and 10 mA in compensation mode
MAXIMUM TUBE CURRENT
- 11 mA
MAXIMUM OUTPUT POWER
- 945 W nominal
FILTRATION
- minimum filtration 2.5 mm Al
BEAM QUALITY
- HVL over 2.5 mm Al at 81 kV
OUTER SHELL TEMPERATURE
- +50ºC (122ºF) maximum
DUTY CYCLE
- Panoramic: 1:15, 81 kV/10mA
- Cephalometric: 1:20, 81 kV/10mA
Power requirements
INPUT VOLTAGE
- 175-250 VAC, 50/60 Hz, adjustable, single phase, grounded socket
MAXIMUM LINE CURRENT
- 7 A at 81 kV/10mA
MAXIMUM LINE RESISTANCE
- 1 ohm
MAXIMUM LINE FUSING
- 10 A slow (main fuse 8A slow in device)
LINE SAFETY SWITCH (when required)
- Approved type, min. 10 A 250 VAC
EARTH LEAKAGE CIRCUIT BREAKER (when required)
- Approved type, min. 16 A 250 VAC, breaker activation leakage current in
accordance with local regulations.
Mechanical parameters
PANORAMIC
- SID 520 mm (±10 mm)
- Magnification factor 1.3
CEPHALOMETRIC
- SID 1635 - 1700 mm ±20 mm
- SID, USA only, fixed 1700 mm ±20 mm
- SOD (Source object distance) 1500
- Magnification factor 1.09 - 1.13 (Non US unit, set during installation)
- Magnification factor, USA only, fixed 1.13
WEIGHT
- Panoramic unit 150 kg (330 lb.)
- Pan Ceph unit 195 kg (430 lb.)
DIMENSIONS
- Panoramic unit (HxWxD) 2240 x 1200 x 970 mm (88" x 47" x 38")
- Pan Ceph unit (HxWxD) 2240 x 1200 x 1850 mm (88" x 47" x 74")
VERTICAL HEIGHT OF CHIN REST
- 850-1700 mm (33"-67")
Timer
PANORAMIC EXPOSURE TIMES:
50 Hz 60 Hz
Normal 19 s (±15%) 16 s (±15%)
Child 17 s (±15%) 14 s (±15%)
Sides 11 s (±15%) 9 s (±15%)
Middle 10 s (±15%) 8 s (±15%)
TMJ 3.3 + 3.3 s (±15%) 2.8 + 2.8 s (±15%)
Max 240 mAs Max 202 mAs
BACK-UP TIME
- 23.5 s (±1.5s)
CEPHALOMETRIC
- 81 kV, 1800 mAs/h (81 kV, 10 mA, duty cycle 1:20)
(One 3.2 s exposure per 1 minute cool-down period)
Measurement bases
- The kV is measured by monitoring the current flowing through a 450 Mohm,
1% feedback resistor connected between the tube anode and ground.
- Both parameters are measured using a Soredex kV/mA meter (pt. no.
4800177) or with a digital multimeter (DMM) according to a specified
procedure. The exposure time is measured using a radiation probe positioned in
the primary beam.
Cephalometric collimator
18 x 24 A or 24 x 18 A for lateral projection,
18 x 24 S for P.A. and A.P. projections
24 x 30 aperture optional
OPX/105
DE 100/15 ö