EUA Cepheid Xpert Ifu Lab
EUA Cepheid Xpert Ifu Lab
EUA Cepheid Xpert Ifu Lab
XPRSARS-COV2-10
Cepheid®, the Cepheid logo, GeneXpert® and Xpert® are trademarks of Cepheid.
AccuPlex™ is a trademark of SeraCare Life Sciences.
Windows® is a trademark of Microsoft Corporation.
Cepheid
904 Caribbean Drive
Sunnyvale, CA 94089 USA
Phone: +1 408 541 4191
Fax: +1 408 541 4192
3 Intended Use
The Xpert Xpress SARS-CoV-2 test is a rapid, real-time RT-PCR test intended for the
qualitative detection of nucleic acid from the SARS-CoV-2 in upper respiratory specimens
(i.e., nasopharyngeal, oropharyngeal, nasal, or mid-turbinate swab or nasal wash/ aspirate)
collected from individuals suspected of COVID-19 by their healthcare provider.
Testing of nasopharyngeal, oropharyngeal, nasal, or mid-turbinate swab or nasal
wash/aspirate specimens using the Xpert Xpress SARS-CoV-2 test run on the GeneXpert
Dx and GeneXpert Infinity systems is limited to laboratories certified under the Clinical
Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. § 263a, that meet
requirements to perform high or moderate complexity tests.
Testing of nasopharyngeal, nasal, or mid-turbinate swab specimens using the Xpert Xpress
SARS-CoV-2 test run on the GeneXpert Xpress System (Tablet and Hub Configurations)
is limited to laboratories certified under CLIA that meet requirements to perform high,
moderate, or waived complexity tests. Testing of these specimens is authorized for use at
the Point of Care (POC), i.e., in patient care settings operating under a CLIA Certificate of
Waiver, Certificate of Compliance, or Certificate of Accreditation.
Results are for the detection of SARS-CoV-2 RNA. The SARS-CoV-2 RNA is generally
detectable in upper respiratory specimens during the acute phase of infection. Positive
results are indicative of active infection with SARS-CoV-2; clinical correlation with
patient history and other diagnostic information is necessary to determine patient infection
status. Positive results do not rule out bacterial infection or co-infection with other
viruses. The agent detected may not be the definite cause of disease. Laboratories within
the United States and its territories are required to report all results to the appropriate
public health authorities.
Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole
basis for treatment or other patient management decisions. Negative results must be
combined with clinical observations, patient history, and epidemiological information.
Testing with the Xpert Xpress SARS-CoV-2 test is intended for use by trained operators
who are proficient in performing tests using either GeneXpert Dx, GeneXpert Infinity
and/or GeneXpert Xpress systems. The Xpert Xpress SARS-CoV-2 test is only for use
under the Food and Drug Administration’s Emergency Use Authorization.
automated sample processing, and real-time RT-PCR for detection of viral RNA.
6 Reagents and Instruments
6.1 Materials Provided
The Xpert Xpress SARS-CoV-2 kit contains sufficient reagents to process 10 specimens
or quality control samples. The kit contains the following:
Note The bovine serum albumin (BSA) in the beads within this product was produced
and manufactured exclusively from bovine plasma sourced in the United States.
No ruminant protein or other animal protein was fed to the animals; the animals
passed ante- and post-mortem testing. During processing, there was no mixing of
the material with other animal materials.
10.3 Assay/Reagent
• Do not open the Xpert Xpress SARS-CoV-2 cartridge lid except when adding
specimen.
• Do not use a cartridge that has been dropped after removing it from the packaging.
• Do not shake the cartridge. Shaking or dropping the cartridge after opening the
cartridge lid may yield non-determinate results.
• Do not place the sample ID label on the cartridge lid or on the barcode label on the
cartridge.
• Do not use a cartridge with a damaged barcode label.
• Do not use a cartridge that has a damaged reaction tube.
• Each single-use Xpert Xpress SARS-CoV-2 cartridge is used to process one test. Do
not reuse processed cartridges.
• Each single-use disposable pipette is used to transfer one specimen. Do not reuse
disposable pipettes.
• Do not use a cartridge if it appears wet or if the lid seal appears to have been broken.
• Wear clean lab coats and gloves. Change gloves between the handling of each
specimen.
• In the event of a spill of specimens or controls, wear gloves and absorb the spill with
paper towels. Then, thoroughly clean the contaminated area with a 10% freshly
prepared household chlorine bleach. Allow a minimum of two minutes of contact
time. Ensure the work area is dry before using 70% denatured ethanol to remove
bleach residue. Allow surface to dry completely before proceeding. Or, follow your
institution’s standard procedures for a contamination or spill event. For equipment,
follow the manufacturer’s recommendations for decontamination of equipment.
• Biological specimens, transfer devices, and used cartridges should be considered
capable of transmitting infectious agents requiring standard precautions. Follow your
institution’s environmental waste procedures for proper disposal of used cartridges and
unused reagents. These materials may exhibit characteristics of chemical hazardous
waste requiring specific disposal. If country or regional regulations do not provide
clear direction on proper disposal, biological specimens and used cartridges should be
disposed per WHO [World Health Organization] medical waste handling and disposal
guidelines.
11 Chemical Hazards5,6
• Signal Word: Warning
• UN GHS Hazard Statements
• Harmful if swallowed.
• May be harmful in contact with skin.
• Causes eye irritation.
2. Repeat on the other nostril with the same swab, using external pressure on the outside
of the other nostril (see Figure 3). To avoid specimen contamination, do not touch the
swab tip to anything other than the inside of the nostril.
3. Remove and place the swab into the tube containing 3 mL of viral transport medium or
3 mL of saline. Break swab at the indicated break line and cap the specimen collection
tube tightly.
10
13 Procedure
13.1 Preparing the Cartridge
Important: Start the test within 30 minutes of adding the sample to the cartridge.
1. Remove a cartridge from the package.
2. Check the specimen transport tube is closed.
3. Mix specimen by rapidly inverting the specimen transport tube 5 times. Open cap on the
specimen transport tube.
4. Open the cartridge lid.
5. Remove the transfer pipette from the wrapper.
6. Squeeze the top bulb of the transfer pipette completely and then place the pipette
tip in the specimen transport tube (see Figure 5).
7. Release the top bulb of the pipette to fill the pipette before removing from the
tube. After filling pipette, excess sample will be seen in the overflow reservoir
bulb of the pipette (see Figure 5). Check that the pipette does not contain bubbles.
8. To transfer the sample to the cartridge, squeeze the top bulb of the transfer pipette
completely again to empty the contents of the pipette (300 µL) into the large
opening (Sample Chamber) in the cartridge shown in Figure 6. Dispose of the
used pipette.
11
Sample Chamber
(Large Opening)
Note Take care to dispense the entire volume of liquid into the Sample Chamber. False
negative results may occur if insufficient sample is added to the cartridge.
12
13
Note Do not turn off or unplug the instruments while a test is in progress. Turning
off or unplugging the GeneXpert instrument or computer will stop the test.
15 Quality Control
15.1 Internal Controls
Each cartridge includes a Sample Processing Control (SPC) and Probe Check Control
(PCC).
Sample Processing Control (SPC) – Ensures that the sample was processed correctly.
The SPC verifies that sample processing is adequate. Additionally, this control detects
sample-associated inhibition of the real-time PCR assay, ensures that the PCR reaction
conditions (temperature and time) are appropriate for the amplification reaction, and that
the PCR reagents are functional. The SPC should be positive in a negative sample and can
be negative or positive in a positive sample. The SPC passes if it meets the validated
acceptance criteria.
Probe Check Control (PCC) – Before the start of the PCR reaction, the GeneXpert
System measures the fluorescence signal from the probes to monitor bead rehydration,
reaction tube filling, probe integrity, and dye stability. The PCC passes if it meets the
validated acceptance criteria.
15.2 External Controls
External controls should be used in accordance with local, state, and federal accrediting
organizations as applicable.
16 Interpretation of Results
The results are interpreted automatically by the GeneXpert System and are clearly shown
in the View Results window. The Xpert Xpress SARS-CoV-2 test provides test results
based on the detection of two gene targets according to the algorithms shown in Table 1.
14
See Table 2 to interpret test result statements for the Xpert Xpress SARS-CoV-2
test.
Result Interpretation
The 2019 novel coronavirus (SARS-CoV-2) target nucleic acids are
detected.
• The SARS-CoV-2 signal for the N2 nucleic acid target or signals
SARS-CoV-2
for both nucleic acid targets (N2 and E) have a Ct within the valid
POSITIVE
range and endpoint above the minimum setting
• SPC: NA; SPC is ignored because coronavirus target amplification
occurred
• Probe Check: PASS; all probe check results pass
The 2019 novel coronavirus (SARS-CoV-2) nucleic acids may be
present.
Sample should be retested according to the Retest Procedure in
Section 17.2. For samples with a repeated presumptive positive
SARS-CoV-2 result, additional confirmatory testing may be conducted, if it is
PRESUMPTIVE necessary to differentiate between SARS-CoV-2 and SARS-CoV-1 or
POS other Sarbecovirus currently unknown to infect humans, for
epidemiological purposes or clinical management.
• The SARS-CoV-2 signal for only the E nucleic acid target has a
Ct within the valid range and endpoint above the minimum setting
• SPC: NA; SPC is ignored because a target amplification has
occurred.
• Probe Check: PASS; all probe check results pass
The 2019 novel coronavirus (SARS-CoV-2) target nucleic acids are not
detected.
SARS-CoV-2 • The SARS-CoV-2 signals for two nucleic acid targets (N2 and E)
NEGATIVE do not have a Ct within the valid range and endpoint above the
minimum setting
• SPC: PASS; SPC has a Ct within the valid range and endpoint
above the minimum setting
• Probe Check: PASS; all probe check results pass
15
Result Interpretation
SPC does not meet acceptance criteria. Presence or absence of the
2019 novel coronavirus (SARS-CoV-2) nucleic acids cannot be
INVALID determined. Repeat test according to the Retest Procedure in Section
17.2.
• SPC: FAIL; SPC and SARS-CoV-2 signals do not have a Ct
within valid range and endpoint below minimum setting
• Probe Check – PASS; all probe check results pass
Presence or absence of the 2019 novel coronavirus (SARS-CoV-2)
nucleic acids cannot be determined. Repeat test according to the Retest
Procedure in Section 17.2.
• SARS-CoV-2: NO RESULT
ERROR • SPC: NO RESULT
• Probe Check: FAIL1 ; all or one of the probe check results fail
1 If the probe check passes, the error is caused by the maximum
The Xpert Xpress SARS-CoV-2 test includes an Early Assay Termination (EAT) function which will
provide earlier time to results in high titer specimens if the signal from the target nucleic acid reaches a
predetermined threshold before the full 45 PCR cycles have been completed. When SARS-CoV-2 titers
are high enough to initiate the EAT function, the SPC amplification curve may not be seen and its
results may not be reported.
16
17 Retests
17.1 Reasons to Repeat the Assay
If any of the test results mentioned below occur, repeat the test once according to
instructions in Section 17.2, Retest Procedure.
• A PRESUMPTIVE POS result indicates the 2019 novel coronavirus (SARS-CoV-2)
nucleic acids may be present. Only one of the SARS-CoV-2 nucleic acid target was
detected (E gene) while the other SARS-CoV-2 nucleic acid target (N2 gene) was not
detected.
• An INVALID result indicates that the control SPC failed. The sample was not
properly processed, PCR is inhibited, or the sample was not properly collected.
• An ERROR result could be due to, but not limited to, Probe Check Control
failure, system component failure, no sample added, or the maximum pressure
limits were exceeded.
• A NO RESULT indicates that insufficient data were collected. For example,
cartridge failed integrity test, the operator stopped a test that was in progress, or
a power failure occurred.
If an External Control fails to perform as expected, repeat external control test
and/or contact Cepheid for assistance.
17
18 Limitations
• The performance of this test was established based on the evaluation of a limited
number of clinical specimens. Clinical performance has not been established with all
circulating variants but is anticipated to be reflective of the prevalent variants in
circulation at the time and location of the clinical evaluation. Performance at the time
of testing may vary depending on the variants circulating, including newly emerging
strains of SARS-CoV-2 and their prevalence, which change over time.
• Performance of the Xpert Xpress SARS-CoV-2 test has only been established in
nasopharyngeal swab specimens. Use of the Xpert Xpress SARS-CoV-2 test with
other specimen types has not been assessed and performance characteristics are
unknown.
• Oropharyngeal, nasal swabs and mid-turbinate swabs (self-collected under supervision
of or collected by a healthcare provider) as well as nasal wash/aspirate are considered
acceptable specimen types for use with the Xpert Xpress SARS-CoV-2 test but
performance with these specimen types has not been established.
• A false negative result may occur if a specimen is improperly collected, transported or
handled. False negative results may also occur if inadequate numbers of organisms are
present in the specimen.
• As with any molecular test, mutations within the target regions of Xpert Xpress SARS-
CoV-2 could affect primer and/or probe binding resulting in failure to detect the
presence of virus.
• This test cannot rule out diseases caused by other bacterial or viral pathogens.
• The performance of this device has not been assessed in a population vaccinated against
COVID-19.
18
Testing of nasopharyngeal, nasal, or mid-turbinate swab specimens using the Xpert Xpress SARS-
CoV-2 test run on the GeneXpert Xpress System (Tablet and Hub Configurations) is limited to
laboratories certified under CLIA that meet requirements to perform high, moderate, or waived
complexity tests. Testing of these specimens is authorized for use at the Point of Care (POC), i.e., in
patient care settings operating under a CLIA Certificate of Waiver, Certificate of Compliance, or
Certificate of Accreditation.”
20 Performance Characteristics
20.1 Clinical Evaluation
The performance of the Xpert Xpress SARS-CoV-2 test was evaluated using archived
clinical nasopharyngeal (NP) swab specimens in viral transport medium. A total of 45
SARS-CoV-2 positive and 45 SARS-CoV-2 negative NP swab specimens were tested with
Xpert Xpress SARS-CoV-2 in a randomized and blinded fashion.
All the 45 SARS-CoV-2 positive specimens and 30 of the 45 SARS-CoV-2 negative
specimens were collected during COVID-19 pandemic in the US and had previously been
characterized as positive or negative for SARS-CoV-2 by an EUA RT-PCR test. Fifteen of
the 45 SARS-CoV-2 negative NP swab specimens were collected before December 2019
and are expected to be negative for SARS-CoV-2.
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) were
determined by comparing the results of the Xpert Xpress SARS-CoV-2 test relative to the
expected results. Results of these 90 archived clinical NP swab specimens are shown in
Table 3. The PPA was 97.8% (95% CI: 88.4% - 99.6%) and the NPA was 95.6% (95% CI:
85.2% - 98.8%).
Expected Results
Total 45 45 90
20
21 Analytical Performance
21.1 Analytical Sensitivity (Limit of Detection) – Live SARS-CoV-2 Virus
Studies were performed to determine the analytical limit of detection (LoD) of the Xpert
Xpress SARS-CoV-2. The LoD of Xpert Xpress SARS-CoV-2 was established using one
lot of reagent and limiting dilutions of live SARS-CoV-2 virus (USA_WA1/2020 strain)
prepared in viral transport medium and NP swab clinical matrix. The concentration level
with observed hit rates greater than or equal to 95% in the LoD determination study were
0.0050 and 0.0200 PFU/mL for the N2 target and E target, respectively (Table 4).
Verification of the estimated LoD claim was performed on one reagent lot in replicates of
20 prepared in pooled NP swab clinical matrix. The LoD is the lowest concentration
(reported as PFU/mL) of live SARS-CoV-2 virus samples that can be reproducibly
distinguished from negative samples ≥ 95% of the time with 95% confidence. The claimed
LoD is 0.0200 PFU/mL (Table 4).
21
For the N2 target, 211 matching sequences were excluded due to ambiguity codes, which
reduced the total to 109,995 sequences. Xpert Xpress SARS-CoV-2 had 97.29% match to
the sequences with the exception of 2,919 sequences that had a single mismatch and sixty-
three sequences with two or more mismatches. The 63 sequences with additional
mismatches were:
• One sequence contained two mismatches in the probe binding region and a third
mismatch between oligos;
• Two sequences contained one mismatch in the forward primer region and a second
mismatch in the reverse primer region;
• Six sequences contained one mismatch in the forward primer region and a second
mismatch between oligos;
• Twenty-seven sequences contained one mismatch in the probe binding region and a
second mismatch in the reverse primer region; and
• Twenty-seven sequences contained one mismatch in the forward primer region and a
second mismatch in the probe binding region.
None of these mismatches are predicted to have a negative impact on the performance of
the assay.
22
23
22 References
1. Centers for Disease Control and Prevention. https://www.cdc.gov/coronavirus/2019-
ncov/index.html. Accessed February 9, 2020.
2. bioRxiv. (https://www.biorxiv.org/content/10.1101/2020.02.07.937862v1). Accessed
March 3, 2020.
3. Centers for Disease Control and Prevention. Biosafety in Microbiological and
Biomedical laboratories (refer to latest edition).
http://www.cdc.gov/biosafety/publications/
4. Clinical and Laboratory Standards Institute. Protection of Laboratory Workers from
Occupationally Acquired Infections; Approved Guideline. Document M29 (refer to
latest edition).
5. REGULATION (EC) No 1272/2008 OF THE EUROPEAN PARLIAMENT AND OF
THE COUNCIL of 16 December 2008 on the classification labeling and packaging of
substances and mixtures amending and repealing, List of Precautionary Statements,
Directives 67/548/EEC and 1999/45/EC (amending Regulation (EC) No 1907/2007).
6. Occupational Safety and Health Standards, Hazard Communication, Toxic and Hazard
Substances (March 26, 2012) (29 C.F.R., pt. 1910, subpt. Z).
24
24 Technical Assistance
Before contacting Cepheid Technical Support, collect the following information:
• Product name
• Lot number
• Serial number of the instrument
• Error messages (if any)
• Software version and, if applicable, Computer Service Tag number
25
25 Table of Symbols
Symbol Meaning
Catalog number
Do not re-use
Batch code
Caution
Manufacturer
Country of manufacture
Control
Expiration date
Temperature limitation
Biological risks
Cepheid
904 Caribbean Drive
Sunnyvale, CA 94089 USA
Phone: +1 408 541 4191
Fax: +1 408 541 4192
For use under Emergency Use Authorization (EUA) Only
26