Fundamental Principles of Computed Tomography (CT) : For B.S. Radiologic Technology
Fundamental Principles of Computed Tomography (CT) : For B.S. Radiologic Technology
Fundamental Principles of Computed Tomography (CT) : For B.S. Radiologic Technology
FUNDAMENTAL
PRINCIPLES OF COMPUTED
TOMOGRAPHY (CT)
for B.S. Radiologic Technology
All x-ray imaging is based on the absorption of x-rays as they
pass through the different parts of a patient’s body.
1
Although also based on the variable absorption of x-rays by different
tissues, COMPUTED TOMOGRAPHY Imaging provides a different form of
imaging known as “cross-sectional imaging”.
The cross-sectional images produced are used for a variety of diagnostic and
therapeutic purposes.
2
COMPUTED TOMOGRAPHY
Is the process of creating a cross-sectional tomographic plane (slice)
of any part of the body which is then reconstructed by a computer using x-
ray absorption measurements collected at multiple points about the
periphery (external surface) of the part being scanned.
BENEFITS OF CT SCANNER
3
COMPUTED TOMOGRAPHY SCANNING HAS BEEN VARIOUSLY IDENTIFIED
AS:
HISTORICAL PERSPECTIVE
4
The original Axial CT image in 1975 that
shows a coarse 128 x 128 matrix were
physicians fascinated by the ability to see
the soft tissue structures of the brain
including the black ventricles.
In 1971, the first full scale CT unit used to examine the skull and diseases of
the brain called the Head Scanning was installed at Atkinson Morley’s
Hospital, England.
The first CT unit in the U.S. was installed in 1973 at the Mayo Clinic and
Massachusetts General Hospital.
5
PRINCIPLES OF OPERATION
In fact, it is this beam width that typically specifies the slice thickness to be
imaged. The x-ray tube is rigidly linked to an x-ray detector located on the
other side of the subject so that they moved synchronously.
6
Together, the tube and the detector scan across the subject, sweeping the
narrow x-ray beam through the slice
This linear transverse scanning motion of the tube and the detector across
the subject is referred to as a translation. The arrangement is diagramed in
Figure 2.
7
When the “Source-Detector” assembly makes one sweep/translation across
the patient’s body, the internal structures of the body “attenuate” the x-ray
beam according to their density and effective atomic numbers.
8
The computer processing of this data involves the effective
superimposition of each intensity
profile to reconstruct the anatomic
structures. And through the use of
simultaneous equations, a matrix of
values is obtained that represent the
“cross-sectional anatomy”.
9
It uses a single detector element to
capture a beam of x-rays and was designed
to scan the head only.
The original EMI scanner
required 180 translations, and each is
separated by a one degree rotation. These
systems incorporate two detectors and
split the finely collimated x-ray beam so that
two slices could be imaged during each scan.
The scan time for first generation scanners was almost five (5)
minutes to complete one scan and it requires a water-filled bag for patient
positioning and detector normalization during scanning and to eliminate air
interfaces.
The water is
necessary also to moderate
the abrupt change in x-ray
attenuation that would
occur between air and skull
bone.
10
It requires 15 seconds to 5 minutes to gather sufficient data
before it can reconstruct an image. And because of this inherently slow
scanning speed as well as other limitations, these systems were used
exclusively for “Neurologic” work”. The detectors used in the early scanners
were Sodium Iodide Scintillation crystals with photomultiplier tubes.
11
Major Characteristics of the First Generation CT Units:
12
Because of multiple detector array, a single translation results in the same
number of data points as several translations with a first generation CT
scanner.
The Second
generation CT scanner is equipped with a
Linear Detector Array.
• Up to 30 detectors.
13
3. THIRD GENERATION SCANNER
14
The malfunctioning of the detector or detectors results in the
absence of signal thereby having these artifacts on the reconstructed image.
However, a properly formulated “image reconstruction algorithm” or
“software corrected image reconstruction algorithm” minimizes such
artifacts.
15
Another technique, called the “overscan” may also be
available in this unit. Overscanning uses a scan of more than 360 degrees.
This overscan may be displayed as segmented data from 2, 3 or 4 scans as a
single image. These fourth-generation scanners are designed as for the third
generation in which it has a rotate motion only.
With the fourth-generation scanners, the x-ray source rotates,
but the detector assembly does not. Radiation detection is accomplished
through a fixed circular array of detectors which contains as many as 1,000
individual elements.
These units are capable of
one-second scanning time and can
accommodate variable slice thickness
through automatic pre-patient collimation
and can provide the image manipulation
capabilities of earlier scanners. The
fixed circular array of detectors in fourth-
generation CT scanners does not result in a constant beam-path from the
source to all detectors, but it does allow each detector to be calibrated and
its signal normalized during each scan.
16
5. FIFTH-GENERATION SCANNERS
The Fifth generation of scanners is represented by the
technology utilized in a dedicated cardiac unit designed around a rotating
electron beam. This is sometimes referred to as “Electron Beam
Tomography” (EBT) or “Cardiac Cine CT.
17
The x-ray tube is replaced with a huge electron gun that uses
a deflection coil to direct a 30 degrees beam of electrons in a 210 degrees
arc around a 360 degrees anode ring. The fan-beam that is produced is
intercepted by a 360 degrees detector array. Because the electron beam
rotates around the anode ring, there is no moving gantry in this design.
18
6. SIXTH GENERATION (HELICAL/SPIRAL COMPUTED
TOMOGRAPHY)
Helical (Spiral) scanners were made possible by advances in
slip ring connection technology. Slip rings consist of brushes that fit
into grooves to permit the current and voltage to the x-ray tube to be
supplied while the tube is in continuous rotation around the gantry.
This permits scanning of the entire
body in a helical pattern without
stopping the tube.
When the patient
table (couch) is moved slowly during
the x-ray exposure while the tube is
in continuous rotation, data
comprising a continuous helical scan
of the patient is acquired.
The primary advantage of the Helical scanning is a much shorter total scan
time (30-40 sec. for the entire abdomen). This in turn permits the use of
less contrast media.
Another advantage is that for many patients the entire
examination can be completed in one-breath hold, eliminating overlaps and
missed areas due to variations in the amount of air in the lungs between
scan sections. This also reduces the possibility of motion artifacts.
19
7. SEVENTH GENERATION (MULTISECTION or MULTISLICE
COMPUTED TOMOGRAPHY)
Computed Tomography units with multiple detectors are
known by several terms.
20
Section thickness with
single section CT scanners is
determined by the collimator
size, not the width of the
detector.
Because MSCT scanners are
exposing multiple detectors
simultaneously, section thickness
is determined by the width of the
detector, even though the collimation of the beam is wide enough to
expose several other bands of the detectors.
21
RELEVANT NOMENCLATURES
• Algorithm – is a procedure for solving a certain type of mathematical
problem. It is a step by step method of getting a diagnosis.
22
• Data Acquisition System (DAS) –
radiation detection system that measures
amount of radiation passing through
patient.
It refers to a method by which
the patient is systematically scanned by
the x-ray tube and detectors to collect
enough information/data for image
reconstruction
• Field of View (FOV) – is the area of anatomy displayed by the CRT; this can be
adjusted to include entire body section of a specific part of the patient
anatomy being scanned.
• Scan – is the motion of an x-ray tube and radiation detector system required
to collect data for reconstructing CT image.
23
• Scan Diameter – referred to as the zoom or focal plane of a CT scan.
• Scan Field – circular field of measurement covered by the x-ray tube and
detector array.
24
Whereas, in a typical CT study, from 10-20 or more individual
slices are required for a single CT examination. Low tissue density that
would normally be obscured by higher density anatomy on a conventional
radiograph can be clearly visualized in CT.
For this reason, CT is valuable in neurologic work in which
brain is surrounded by the skull. Likewise, in many body examinations, low
tissue density that would be hidden or blend with surrounding anatomy can
be clearly visualized in CT.
It should also be noted that the CT image displays the entire
cross-section of the slice of anatomy that was scanned. Thus, the size and
location of any pathologic condition can be determined with extreme
accuracy within a given CT slice.
25
Spiral CT has emerged as a new and improved diagnostic tool.
It provides improved imaging of anatomy compromised by respiratory
motion.
Spiral CT is particularly good for the chest, abdomen and pelvis. It has also
the ability to perform conventional transverse imaging for regions of the
body where motion is not a problem such as the head, spine and
extremities.
INTERPOLATION ALGORITHM
The ability to reconstruct an image at
any Z-axis position is due to “Interpolation”.
26
During Spiral CT, image data is received continuously. When
an image is reconstructed, the plane of the image does not contain
enough data for reconstruction. Data must be estimated by
“interpolation”.
Data interpolation is performed by a special computer called
“interpolation algorithm”. The first interpolation algorithms used a
360 degree “linear interpolation”.
2. Cubic-Spline Interpolation
27
Given the two
red points, the blue line is the
linear interpolation between
the points
28
The disadvantage of the 180 degree interpolation algorithm is
the increased image noise (reduced visibility of low contrast objects)
compared with 360 degree interpolation algorithm.
However the use of a cubic-spline interpolation algorithm can
produce that is called a “breakup” artifact at high contrast interfaces
such as bone-tissue.
In addition, to improve Sagittal and Coronal reformatted
views, 180 degree interpolation algorithm allows imaging as a “Pitch”
greater than 1.
Pitch is also defined as the ratio of the table speed per rotation and the total
collimation (slice thickness).
29
Pitch is expressed as a ratio; 1:1, 1:5 or 2:1 and a Pitch Ratio of 1:1 will
result in the best image quality.
I Where P = Pitch
P = ------ I = Table increment per 360 degrees
B B = Beam width in millimeter
8mm
= ----------------
5mm
= 1.6mm
PR ≈ 1.6:1
30
Increasing Pitch above 1:1 increases the volume of tissue that
can be imaged in a given time. The ability to image a large volume of tissue
in a single breath-hold is the principal advantage to Spiral Computed
Tomography. This is particularly helpful in CT Angiography, Radiation
Therapy treatment planning and imaging uncooperative patients.
QUESTION:
How much tissue will be imaged if collimation is set to 8mm,
with a scan time of 25 seconds, and a pitch of 1.5:1?
= 300mm ≈ 30 centimeters
What if the gantry rotation time is not 360 degrees in one second?
If the tube heat capacity is limited, slower rotation may be necessary. In
such a situation, the equation would become:
QUESTION:
31
Collimation x Pitch x Scan Time
Tissue Imaged = --------------------------------------------------------
Gantry Rotation
= 80 millimeters
≈ 8 centimeters
SCANNER DESIGN
Spiral CT is made possible by the use of slip ring, which allows the gantry to
rotate continuously without interruption.
32
Take note that conventional CT scanning is performed with a
pause between each gantry rotation. During the pause, the patient couch is
moved and the gantry may be rewound to a starting position.
In Slip Ring gantry system, power and electric signals are
transmitted through a stationary rings within the gantry that make
continuous rotation possible and eliminating the need for electrical cables.
X-RAY TUBE
In conventional CT, the x-ray tube is energized for one
rotation, usually 1 second and every 6-10 seconds that allows the tube to
cool between scans.
Spiral CT places a considerable thermal demand on the x-ray
tube. Because of the continuous rotation and energization of the x-ray tube
for longer times, higher power levels must be sustained.
“High heat capacity” and “High Cooling Rates” are trademarks of x-ray tubes
designed for Spiral CT. Spiral CT x-ray tubes have anode heat storage
capacity of 5 Million Heat Units (HU) or more. Less than 3 million HU is
unacceptable. It has also an anode cooling rate of 1 MHU per
minute.
33
RADIATION DETECTORS
34
Because the crosstalk between adjacent detectors is related
to the angle of the incoming x-ray beam, some scanners are designed with
more narrow detectors at the isocenter of the x-ray beam where the
incoming x-ray photon arrive at near 90 degrees angles.
Additional techniques for decreasing detector crosstalk
include both pre- and post-patient collimation. The post-patient collimator
is a set of lead shields immediately in front of the detectors.
There are also innovative post-acquisition algorithms that
show promise of additional increases in resolution by correlating
information from overlapping sets of binned detectors.
35
CT SYSTEM COMPONENTS
Components of a Conventional x-ray unit are the following:
1. Operating Console/Control
Console
3. X-ray Tube
36
A. GANTRY
The center around which the gantry rotates is called the “axis
of rotation”. The gantry includes a 50-85cm (20”-34”) aperture for the
patient.
Obese patients who exceed the tabletop weight limits must
never be forced into the aperture. Most gantries can be angled up to 30
degrees to permit positioning for partial coronal images. This is especially
desirable in obtaining transverse scans perpendicular to the vertebral
column.
Positioning lights are usually mounted on the gantry as well.
Intense halogen lights and low-power red laser lights are used for
positioning. The body part of interest must be properly centered to the
aperture because the extreme edge of the scanning field produces a
severely degraded image.
There are often three positioning lights for accurate sagittal,
coronal and transverse centering.
37
The Subsystems of the Gantry includes:
A. X-ray Tube
B. Radiation Detector Array
C. High Voltage Generator
D. Patient Support Couch/Patient Couch
E. Mechanical Support
X-RAY TUBE
Unlike the x-ray tubes used for
taking intra-oral radiography, the CT x-ray
source (x-ray tube) must put out a
continuous stream of X-rays of a simple
pulse.
38
Finally, the energy spectrum of the x-ray beam defines how
well the x-rays can penetrate the body, as well as their expected relative
attenuation as they pass through materials of different density.
Higher energy x-rays penetrate more effectively than lower-
energy ones, but are less sensitive to changes in material density and
composition. Lower energy x-rays are more prone to attenuation by soft
tissue, and since soft tissue is often the specific target of the CT scan, the
beam constraints a lot of low frequency x-ray photons.
On the other hand, Large Focal spot is used for high technique
studies of large anatomy.
Example:
39
In addition, 0.5 – 5.0 million heat units’ anodes of layered
alloys, cylindrical anodes and liquid-cooled and air-cooled tube housing
designs have been developed. A CT tube may produce 30 exposures per
examination. Because most CT units are scheduled for 10-20 examinations
per day, a tube may accumulate 10,000 exposures in a single month. It is
not unusual for a CT tube to fail after several months. Only a few last a full
year.
The radiation beam is double collimated; One at the tube exit
and again at the detector entrance. This collimator assists in eliminating
scatter information. Collimation is variable from 1mm to 13mm and is
usually controlled by the
software program. The
dimension of the collimation
width determines the voxel
length or section thicknesses.
DETECTOR ASSEMBLY
An electronic component of CT scanner that measures
remnant radiation exiting the patient and converting the radiation to an
analog signal proportionate to the radiation intensity measured.
Detector assembly is a
sensor that measures the
extent to which the x-ray
signal has been attenuated
by the object.
The individual detectors are arranged in a one-dimensional
line or arc instead of in a 2D array. The detector must be capable of
responding with extreme speed to a signal without lag, must quickly discard
the signal and prepare for the next.
They must also respond consistently and be small in size.
They are usually placed with a source-to-image receptor distance of 44
inches (110cm).
40
CT Detectors should have high capture efficiency, high
absorption efficiency and high conversion efficiency. These three
parameters are called the “Detector Dose Efficiency”.
Capture Efficiency
+Absorption Efficiency
+Conversion Efficiency
DOSE EFFICIENCY
CT detectors should also have high stability, fast response time and wide
dynamic range.
Stability is controlled by how often the detectors must be
recalibrated to meet quality control standards.
Response Time is the speed with which the detector can react
to recognize an incoming photon and recover for the next input.
The Dynamic range is the ratio of the largest signal that can
be measured to the smallest.
Typical modern scanners are capable of dynamic ranges of
1,000,000 to 1.
The early CT scanners employ one detector, but modern CT
scanners uses multiple detectors in an array numbering up to 2,400, and has
Two General Classifications:
1. Scintillation Detector
2. Gas Detector
41
SCINTILLATION DETECTOR
42
The earliest scanners used a “Sodium
Iodide Crystal” which was quickly replaced
by “Bismuth Germanate”. And the current
crystals of choice being used today are the
“Cesium Iodide and Cadmium Tungstate”.
GAS DETECTOR
Gas-filled detectors are also used in CT scanners. They are
constructed of a large metallic chamber
with baffles spaced only at approximately 1
to 1.5 millimeter intervals/apart. The
spacing determines the maximum detector
(and therefore display pixel size)
resolution.
43
In Gas Detector, the entire detector array is hermitically (air-
tight) sealed and filled under pressure with a high atomic number inert gas
such as “Xenon” or Xenon-Krypton” mixture. Gas-filled detectors are less
efficient than Solid-State detectors. The problem can be partially overcome
by the following three ways:
44
The detected energy
comprises a digital
signal that is sent to
the computer.
Although ion gas
detector efficiency
may be only 45%,
these detectors can
be packed extremely close in the detector array, permitting detection
efficiency of slightly less than 50% which is similar to that of Scintillation-PM
detectors.
Xenon detectors are highly
directional, in other words, they must be
set in a fixed position oriented to the x-
ray source. This is why Helical scanners
do not use xenon detectors.
45
TWO BASIC DESIGNS OF RADIATION DETECTORS USED IN TODAY’S
CT SCANNERS
46
COLLIMATION
1. PRE-PATIENT COLLIMATOR
Is mounted on the tube
housing or adjacent to it. It has a contoured
aperture for emitting a generally rectangular
shaped x-ray fan-beam. It limits the area of
the patient that intercepts the useful beam
and thereby determines the slice thickness
and patient dose.
As its name implies, the Pre-
patient collimator is positioned between the
x-ray source and the patient.
Since x-ray photons emitted from the x-ray
tube comes a very wide range, the pre-
patient collimator restricts the x-ray flux (flow) applied to a narrow region as
it is applied to the patient.
For a Single slice CT, it is not only reduces dose to the patient,
it also defines the slice thickness of the imaging plane. For the Multislice
CT, the slice thickness is defined by the detector aperture instead of the
collimator. Because 99% of the x-ray photons emitted from the x-ray tube
are blocked by the pre-patient collimator, x-ray tube efficiency for CT is
poor.
47
Because of geometric
limitations, the x-ray beam, after passing
through the pre-patient collimator, has
two regions:
1. Umbra
2. Penumbra
48
2. POST-PATIENT COLLIMATOR or PRE-DETECTOR COLLIMATOR
49
FILTERS
The x-ray photons emitted
from the x-ray tube exhibit a wide
spectrum; many soft (low energy) x-
rays are present.
50
PATIENT SUPPORT COUCH
The patient
couch is one of the more
important components of
the CT scanner. It supports
the patient comfortably,
and must be constructed of
a low atomic number (Z)
material so that it does not
interfere with x-ray beam
transmission and patient
imaging and to reduce attenuation of the x-ray beam.
The CT table may be
either flat or curved. It is usually
made of carbon graphite fiber to
decrease beam attenuation.
Because the top must
extend beyond the table to move
the patient into the gantry aperture,
it must be capable of supporting the
entire weight of the patient without sagging when fully extended.
Tabletops are rated for maximum weight and it is critical that
the radiographer must make sure that this weight is not exceeded.
Extensive damage to the table may results when attempts are
made to examine very large patients who exceed the tabletop weight limit.
The top is motor-driven to permit the patient to be moved the
exact desired distance between sections. Section intervals may be
controlled automatically by a program initiated at the control console.
The table must also be capable of vertical movement, both for
positioning within the aperture and for ease of patient transfer. Likewise the
patient couch should be capable of automatic indexing so that the operator
does not have to enter the examination room between each scan.
51
COMPUTER
The computer is a unique
subsystem of the CT scanner. It would
not be possible were it not for the ultra-
high speed digital computer. Large
capacity computer is required in order
that depending on image format; as
many as 30,000 equations must be solved simultaneously.
52
Many CT scanners use an “Array Processor” instead of
microprocessor for image reconstruction. This array processor does many
calculations simultaneously. It is significantly faster than the
microprocessor, wherein an image is reconstructed in less than one second.
The CT computer is designed to control data acquisition,
processing, display and storage.
The CT console provides the radiographer with access to the
software program that controls data acquisition, processing and display.
Remote controls may also be linked to the system to permit display and
storage functions.
Control of data storage may be available at the control
console, remote workstations or at the storage units themselves.
A system program is used to start up the CT unit. This
program turns on and performs quality assurance checks on numerous
components in the x-ray equipment and computer hardware systems,
warms up the x-ray tube, etc. It also permits the radiographer to record
various problems that need the attention of the service engineer.
One of its main functions is to perform calibration checks on
the detectors, although this may be an ongoing process during scanning.
A diagnostic program with specific quality assurance tests is
used by service engineers to troubleshoot the system.
The CT console operates from a menu or index directory of operations. The
radiographer simply uses a keyboard, mouse trackball, or other input device
to indicate the desired operation.
At the beginning of each examination, patient information,
such as identification, history and the like is entered or input via a radiology
or hospital data system. This information permits retrieval of the images at
a later date and will also be displayed adjacent to each image.
The data acquisition program controls a variety of operations,
including tube and detector collimation (pixel size), matrix size, gantry
angle, tabletop entrance into the gantry aperture, section increment
movements of the tabletop, x-ray tube voltage and amperage, scan speed,
pitch, detector/resolution, and the direction of detector signals to the
digital image processing section of the computer.
53
EQUIPMENTS & METHODOLOGY
Major areas and equipments required for generating a CT
scan includes:
1. Examination Area
2. Operators Console
3. Computer Room
4. Diagnostic Viewing Console
EXAMINATION ROOM
The major components in the patient area are the patient
couch and gantry. The table and gantry should be positioned so that
the operator can see into the gantry to observe the patient when the
technologist is seated at the operator’s console.
It is often advisable to add mirrors or a TV system to help
observed the patient who is on far side of the gantry from the
operator’s console.
The gantry houses the x-
ray tube and detector system. And
the entire motion of the detector
and the x-ray tube occurs within the
housing of the gantry.
To obtain other than
perpendicular, cross-sectional
projections with respect to the
patient’s table, it is possible to tilt
most gantries before acquiring the data.
54
OPERATOR’S CONSOLE/CONTROL CONSOLE
Operator’s
console is a special area that
contains meters and controls
for selecting radiographic
technique factors and for
proper mechanical movement
of the gantry.
Many CT
scanners are equipped with two
consoles; one for CT radiographer to operate the unit and one for the
radiologist to view the image and manipulate its contrast, size and general
visual appearance.
The operator’s console contains many meters and controls for
selecting proper radiographic technique factors, for the gantry and patient
couch, and for computer commands that allow image reconstruction and
transfer.
The physician’s viewing console accepts the reconstructed
image from the operator’s console and displays for viewing and diagnosis.
A typical operator’s console contains controls and monitors
for the various technique factors. Operation is generally is excess of
100kVp. The usual mA station will be 100mA if the x-ray beam is continuous
and several hundreds mA if it is a pulsed beam. The scan time is often
selectable and varies from 1-5 seconds.
The thickness of the tissue slice to be imaged can also be
adjusted. Nominal thicknesses are 1-10 millimeters, but some units provide
slice thickness as small as 0.5mm for high resolution scanning. Slice
thickness is selected from the console by automatic collimator adjustment.
Controls are also provided for automatic movement and
indexing of the patient couch. This allows the operator to program for
contiguous slices, for intermittent slices or spiral scanning.
The operating console usually has two television monitors.
One is used by the operator to indicate patient data on the scan (hosp. i.d.,
name, patient number, age and gender), and to provide identification for
each scan (scan number, techniques and couch position).
55
The second monitor is used by the operator to view the
resulting image before transferring it to either hard copy or to the
physician’s viewing console.
Once the patient has been positioned on the table and moved
into the gantry, the technologist controls the CT system from the operator’s
console.
Nowadays, many improvements of the operator’s console had
been made to simplify the commands for the technologist so that
understanding and complex computer terminology is minimized. This
simplification allows the technologist to focus most attention on the patient
and the diagnostic results.
Before the technologist can begin to scan, patient data must
be entered into the computer, the proper radiographic parameters must be
selected and the x-ray tube rotor must be initiated.
After the exposure for an individual slice has been completed,
the table is automatically indexed (moved) to the position for the next slice.
Today, most CT systems include the
capability that enables the technologist to
generate a “localizer image” of the patient
that is viewed at the operator’s viewing
console.
56
IMAGE CHARACTERISTICS
To the human observer, the internal structures and functions
of the human body are not generally visible. However, by various
technologies, images can be created through which the medical professional
can look into the body to diagnose abnormal conditions and guide
therapeutic procedures. The medical image is a window to the body. No
image window reveals everything. Different medical imaging methods
reveal different characteristics of the human body.
With each method, the range of image quality and structure
visibility can be considerable, depending on characteristics of the imaging
equipment, skill of the operator, and compromises with factors such as
patient radiation exposure and imaging time.
The figure below is an overview of the medical imaging
process. The five major components are the patient, the imaging system,
the system operator, the image itself, and the observer, the objective is to
make an object or condition within the patient's body visible to the
observer.
57
They can be changeable system components, such as
intensifying screens in radiography, transducers in sonography, or coils in
magnetic resonance imaging (MRI). However, most variables are adjustable
physical quantities associated with the imaging process, such as kilovoltage
in Radiography, gain in Sonography, and echo time (TE) in MRI. The values
selected will determine the quality of the image and the visibility of specific
body features.
The image contained in CT scanning is
unlike that obtained in
conventional
radiography wherein
image is formed directly
on the radiation detector
such as films. With CT
scanners, the x-rays form a stored electronic image that is displayed as a
matrix of intensities.
IMAGE MATRIX
Matrix – is the number of pixels that are
used to display the image on the CRT. The
CT image that is viewed on the CRT is
actually made up of thousands of small
pixels (picture elements), or it is composed
of “matrix of pixels” where each pixel
represent a volume of tissue (voxel).
58
The CT scan format
consists of many cells each assigned
a number and displayed as a density
or brightness level on the video
monitor.
The original image format
consists of an 80 x 80 matrix for a
total of 6,400 individual cells of information.
Each cell of information is “pixel” (picture element), and the
numeric information contained in each pixel is a “CT Number” or
“Hounsfield Unit” for image display purposes.
The “Pixel” is a two-dimensional representation of a corresponding tissue
volume, while the tissue volume is known as a “Voxel” (volume element).
The voxel is determined by multiplying the pixel size by the
thickness of the CT scan slice, or it is an individual pixel with the associated
volume of tissue based on slice thickness.
The diameter of the constructed image is
called the “Field of View”. When
the FOV is increased for a fixed matrix
size; for example a 512x512 matrix, the
size of each pixel is increased
proportionately.
When the matrix size is increased, for example from 512 x 512
to 1024 x 1024, pixel size is smaller.
“CT Pixel Size” is determined by dividing the FOV by the
Matrix size.
Example:
Compute the Pixel size for the following characteristics of a CT
scanner used for CT Brain.
59
B. Field of view of 20cm, 512x512 Matrix.
200mm
= --------------- = 0.4mm/pixel
512 pixels
360mm
= -------------- = 0.7mm/Pixel
512 Pixels
CT NUMBER
Is defined as a relative comparison of the x-ray attenuation of
each voxel of tissue with an equal volume of water.
It is an arbitrary number assigned by computer to indicate
relative density of a given tissue which is expressed in “Hounsfield Unit”
(HU) where the units range from -1000 to +1000 for each pixel.
CT number value is calculated from the attenuation coefficient value for
each voxel and becomes the value for the corresponding pixel in the digital
image.
Water (H2O) is used as the reference and calibration material
for CT numbers. Water has the assigned CT number value of zero. Tissues
or other substances that are more dense than water will have positive (+)
values and those that are less dense will have negative (-) values.
60
A CT Number of -1000 corresponds to air
CT Number of +1000 to a dense
bone
CT Number of “0” indicates water
Muscle is 50 HU
White Matter is 45 HU
Gray Matter is 40 HU
Blood is 20 HU
CSF is 15 HU
Fat is -100
Lung is -200
61
CT image quality characteristic is visibility. That is the visibility
of anatomical structures, various tissues, and signs of pathology. However,
visibility depends on a somewhat complex combination of the five (5) more
specific image characteristics:
1. Contrast Sensitivity
2. Visibility of Detail (as affected by blurring) sometimes called “spatial
resolution”.
3. Visual (Image) Noise
4. Artifacts
5. Spatial or Geometric characteristic of image
62
CONTRAST SENSITIVITY
With CT imaging
the principle source of physical
contrast within a body are the
differences in physical density
among the tissues. An
exception is when an iodine-
based contrast medium is used where it becomes more of an atomic
number (Z) effect.
Compared to other x-ray imaging modalities CT has a very
high contrast sensitivity for "seeing" the soft tissues and differences among
the tissues in the body. Within a
body there will be tissues and
objects with a range of densities
and physical contrast. As
illustrated here things like bones,
bullets, and barium have a very
high physical contrast relative to
the soft tissues. Imaging them is
not a problem. The real challenge is imaging the very low density
differences between and among the soft tissues. That is where CT excels!
Contrast sensitivity determines the range of visibility with respect to
physical contrast. If a procedure has low contrast sensitivity then only
objects with high physical contrast will be visible.
When a procedure, such as CT, has high contrast sensitivity
then tissues with small differences in density will be visualized. If the
contrast sensitivity is low, either because of limitations of the specific
imaging modality or the adjustments of the imaging protocol factors then
tissues that have small differences in density (physical contrast) will not be
visible.
63
Contrast is the degree of difference between two substances
in some parameters that depend on the technique used. The ability to
distinguish material of one composition from another without regard for
size or shape is called “Contrast Resolution”.
In CT scanning, the amount of radiation penetrating the
patient is determined also by the “mass density” of the body part.
Consider the situation of “Fat-Muscle-Bone” structure. In this three
structures, not only are the atomic numbers are different, but the mass
densities also are different. Although these differences are measurable,
they are not imaged well in conventional radiography.
The CT image is able to amplify these differences between the
atomic numbers of the structure and the mass density. The amplified
differences in subject contrast results in a “High-Contrast Resolution”. The
contrast resolution provided by CT scanners is considerably better than that
available in conventional radiography principally because of the scatter
radiation rejection of the fan-beam collimator.
The ability to image low-contrast objects with a CT scanner is
limited by the size and uniformity of the object and by the Noise of the CT
system. While the ability of the CT scanner to demonstrate different tissue
density is “Contrast Resolution”.
VISIBILITY OF DETAIL
A characteristic of all imaging methods, including human
vision, is that there is some blurring that occurs within the process. The
effect of this blurring is that it reduces visibility of detail (small objects and
features). When we have blurred vision we can't read the fine print. Each
medical imaging method has inherent sources of blurring that limits visibility
of detail and determines the types of diagnostic procedures it can be used
for. For example, radiography which has relatively low blur and provides
high visibility of detail is used for visualizing small bone fractures.
64
The general relationship
of detail to the blurring
within the imaging
process is illustrated
here. As the blurring
increases more and
more of the small
objects become
invisible.
In the CT imaging
process there are several sources of blurring that collectively limit visibility
of detail.
There are adjustable protocol factors associated with each of
these sources of blur as we will soon see. Now that brings up and
interesting question, If we can adjust the blurring why not set it to a low
value and get very high visibility of detail? The problem is when we reduce
the blurring it increases another undesirable image characteristic, visual
noise, and can also lead to increased radiation dose to the patient. That is
why we must have optimized imaging protocols that take all of these factors
into account and provide a proper balance.
SPATIAL RESOLUTION is the ability of CT scanner to
demonstrate small objects within the body plane being scanned. In Spatial
resolution, the larger the pixel size and the lower the subject contrast, the
poorer will be the spatial resolution.
The design of the Pre-patient and Pre-detector collimation
will affect also the level of scatter radiation and influencing spatial
resolution by affecting the contrast of the system.
The ability of the CT scanner to reproduce with accuracy a
high-contrast edge is expressed mathematically as the “Edge Response
Function” (ERF). The measured edge response function can be transformed
into another mathematical expression called the “Modulation Transfer
Function” (MTF). The MTF and its graphic representation are most cited to
express the Spatial Resolution of a CT scanner.
65
VISUAL or IMAGE NOISE
All
medical imaging
methods produce
images with some
visual noise. This is
generally an
undesirable
characteristic that
reduces visibility of
certain types of objects and structures as illustrated here.
Specifically, noise reduces the visibility of low-contrast
objects. This is especially significant in CT which is often used to image low-
contrast differences among tissues. Let's take a moment to distinguish
between noise and blurring. Both are characteristics that reduce visibility,
but of different types of objects. Noise reduces visibility of low-contrast
objects, blurring reduces visibility of small objects or detail.
The noise in a CT image can be adjusted with a combination of
protocol factors. So, why don't we adjust it to a low value and have great
visibility? The challenge is that the factors that can be used to adjust and
reduce noise also have an effect on either image detail (blurring) or
radiation dose to the patient. That is why we must have optimized imaging
protocols that take all of these factors into account and provide a proper
balance.
66
The typical CT image is of a slice through the body. During the
image reconstruction phase the slice is divided into a matrix of Voxels. It is
the size of the Voxels that has a major impact on image blurring, noise, and
on radiation dose to the patient.
SYSTEM NOISE
Unlike screen-film radiographs, images acquired with
computed tomography (CT) never look overexposed in the sense of being
too dark or too light. The normalization of CT data to represent a fixed
amount of attenuation relative to that of water ensures that the image
always appears properly exposed. Thus, CT users have not been compelled
to decrease the tube current–time product or the peak kilovoltage for
scanning in small patients, and, as a result, such patients often were
exposed to an excessive radiation dose.
In addition, image quality changes noticeably (specifically, with
regard to the amount of noise or graininess) according to the patient’s body
habitus. For scanning in large patients, the dose must be increased to obtain
diagnostic-quality images. Thus, there has been a tendency to increase the
tube current–time product (patient dose) to avoid excessive noise on
images, particularly for large patients and at thin-section CT, which is more
readily available with the newer generations of scanners.
As the growth in CT utilization increased, particularly in
pediatric patients, and as concern about the population dose from CT began
to be expressed in the scientific literature and lay press, it became clear that
the responsible use of CT required an adjustment of technique factors on
the basis of patient size (attenuation characteristics).
In response to these concerns, the radiology community
(radiologists, medical physicists, and manufacturers) implemented CT dose
management procedures that correspond to the principle of ALARA (as low
as reasonably achievable). The guiding principle in selecting the right dose
for a CT examination is that the specific patient attenuation and the specific
diagnostic task must be taken into account. For large patients, a dose that
is higher than that for small patients is consistent with the ALARA principle.
67
In a CT scanner, if one images a uniform material (e.g., a
water bath) and looks at the CT-numbers for a localized region, one would
find that the CT-numbers are not all the same, but that they vary around
some average or mean value. This variation is called the noise of the system.
Noise is a very important measure of CT scanner performance since the
naturally occurring difference in attenuation coefficient between normal
and pathological tissues is small. CT numbers above or below the average
value is the Noise of the system. Noise of a CT scanner can be measured by
scanning a uniform water phantom.
This should be done for all potential modes (subject size, kVp,
and scan diameters/pixel widths) of clinical use. The noise should be
indicated by the standard deviation (computed from for a sufficient number
of pixels (e.g. 25 or more). Noise should be examined for both central and
peripheral regions of a scan.
System Noise should be evaluated daily by scanning a 20cm diameter water
bath. All scanners have the ability to identify a region of interest (ROI) and
compute the mean and standard deviation of CT Numbers in that ROI.
When the CT radiographer measures Noise, the ROI must
encompass at least 100 pixels. Such noise measurements should include
five determinations --- four on the periphery and one in the center to
monitor the spatial uniformity.
68
Even though the most common display of the results of a CT
scan is a gray level picture which is interpreted for clinical results end
photographed for storage, the fundamental measurement of CT-scanners is
narrow beam transmission resulting in a cross sectional reconstruction of
numbers presumably related to the x-ray linear attenuation coefficient, μ.
The linear attenuation coefficient depends upon (1) physical density
(grams/cm3), (2) atomic composition, and (3) photon energy.
Most CT scanners utilize a polychromatic x-ray source and
provide numbers related to an x-ray linear attenuation coefficient which has
been averaged over the various energies in the photon spectrum being
detected. Because of this, CT numbers relative to water may depend on
the size of the object being scanned and other physical properties of the
object.
Since the number scale in any given CT scanner is arbitrary,
one must determine the contrast scale (that is, the change in linear
attenuation coefficient per CT-number) in order to reduce the CT-number
standard deviation to a machine independent basis.
Linearity
It is also of interest to establish whether CT-numbers vary in a
linear fashion with the linear attenuation coefficients of the material
studied. The verification of linearity is of interest because it establishes the
constancy of contrast scale over the range of CT-numbers of clinical interest.
Obviously, one must have the means of varying the linear x-
ray attenuation by amounts adequate to give a reasonable difference in CT-
numbers. Linearity is the consistent representation of CT Numbers.
69
70
71
ARTIFACTS
An artifact is
generally something that
appears in an image that is not
a direct visualization of an
object or structure in the
body. There are quite a few
possible artifacts coming from
a variety of conditions that
can occur during an imaging
procedure. Some are very
obvious such as streaks and "ghosts" while others are less visible but more
in the form of changes in how certain areas or objects are displayed.
With the advances in CT technology many artifacts are less
common. The most effective approach to learning CT artifacts is through
their observation and analysis while viewing clinical images.
SOURCES OF ARTIFACTS
1. Patient
* moving
* breathing
* metals
* concentrated Gastrografin (contrast media)
* presence of barium sulfate
* orthopedic plates & screws
2. Machine
* ring artifacts
72
73
74
75
76
77
78
79
QUALITY ASSURANCE IN CT SYSTEM
History of CT and Evolution of Spiral Scanners
80
The development of faster scan times has also created some
drawbacks such as: the need for x-ray tubes with higher heat ratings; more
powerful generators to sustain added heat volumes; and, increased image
noise consistent with the rapid reconstruction of images. This "image noise"
can cause an artifact known as the venetian blind artifact. This occurs with
multi-slice scanners and appears as bright and dark bands superimposed on
three dimensional images.
Another drawback from helical scanners is the notable
difference in low contrast resolution. This problem has created the need for
additional test tools and more suitable phantoms for spiral and multi-slice
scanners. Even after a close review; the benefits of spiral scanners definitely
still outweigh the drawbacks.
Although Preventive Maintenance is generally performed by
the vendor, the technologist should be routinely involved in scanning
phantoms and performing other tests to ensure that the scanner is
operating properly. Quality Assurance on CT scanners is subject to all the
misalignment, miscalibration and malfunctioning difficulties.
CT scanners have the additional complexities of multi-
motional gantry, the interactive console and the associated computer. Each
of these subsystems allows more possibilities for drift and instability,
resulting in degradation of image quality. For this reason, a Quality
Assurance Program is essential for each scanner. Such programs include
daily, weekly, monthly and annual measurement and observations in
addition to the on-going preventive maintenance.
81
This photograph of a CT
phantom used to perform
quality assurance testing.
Phantoms are made of Plexiglas,
which has a density of 120
Hounsfield units. The black
holder is also used to perform
fine alignments of the phantom
in the gantry and to the
localizing laser lights. There are several types of phantoms, for example, one
type may be used for head dose calculations, and another for body dose
calculations. Different manufacturers will recommend testing their
equipment with specific phantoms. All phantoms must meet performance
standards set by the Food and Drug Administration (FDA).
82
Acceptance testing includes the following baseline type tests:
*Light alignment
*Slice thickness
*Noise
*Contrast resolution
*Heat Unit linearity
*Uniformity
*Spatial resolution
*Patient dose
83
This chart shows the
various CT numbers
(Hounsfield units)
calculated for various
tissues and substances
based on the density of
water. Notice that at the
extremes is bone (+1000
HU) and air (-1000 HU).
Water has a CT number
of zero, which is used to
test for the function of
the algorithm that
calculates CT numbers.
84
The CT number for water- (average & standard deviation) test
is done to ensure equipment manufacturer specifications for CT number,
field uniformity, and noise. The test for CT number of water is done daily.
Possible causes for the CT number of water to be out of range is
miscalibration of the algorithm generating CT numbers.
This is a type of problem that needs immediate attention of
the biomedical engineer or radiation safety officer. When the CT number for
water and air fail the recommended range it must be immediately corrected
to insure accuracy of the displayed CT image. If the CT number fluxuates
significantly, but remains within the acceptable range, this too should be
brought to the attention of the radiation safety officer.
85
Test 2 - Noise and Field Uniformity
86
Standard deviation describes the difference between the
lowest ROI value and the maximum ROI value. Testing is performed daily
with the CT number for water and also upon acceptance of new equipment.
87
An increase in the standard deviation indicates the image is
becoming "noisier." Possible causes include decrease in tube output,
increased detector, amplifier(s), analog-to-digital converter, or other issues.
88
The
photograph
on the left
is of a
phantom
designed
for multiple
QA testing.
The yellow
arrow
points to the bar pattern in the phantom that is used to measure scanner
high-resolution. The resulting image is seen on the right. Notice the bar
pattern is filled with air giving 100% contrast. The smallest row of bars is
recorded and compared to baseline value in this test. Expected result is that
a complete set of bars or holes in some rows in the range of 0.75 to 1.0 mm
or 0.5% contrast for 5 mm for modern CT scanners.
As we have discussed, spatial resolution is a measure of detail
resolution. Sometimes it is practical to evaluate the spread of information
within the CT system. To do this we look at what is known as the
modulation transfer function (MTF). The MTF is the most common method
of describing spatial resolution in CT, digital radiography, and film-screen
radiography systems. The MTF analysis allows us to compare system
performance on a day-to-day basis, or to compare a system's performance
against another CT system.
Modulation transfer function is expressed in line pairs per
centimeter (lp/cm). When counting a line pair, one line and its adjacent
space are called a line pair. To measure MTF directly a line pair phantom is
imaged and the number of line pairs is counted. If 5 line pairs are counted,
the spatial resolution is reported as 5 lp/cm. If 10, 15, or 20 line pairs are
seen, the spatial resolution is reported as 10 lp/cm, 15 lp/cm or 20 lp/cm
respectively. The number of line pairs seen in a given length is known as the
spatial frequency.
Because CT scanners are not created equally, how well a given
scanner displays an object is also a function of the size of that object and
the spatial resolution of the scanner. An object's size in a given length is also
known as its' spatial frequency.
89
What this means is that how frequently an object fits into a
given space is it spatial frequency. Generally speaking, the smaller an object
is the higher spatial frequency and the more difficult it is to be displayed
accurately.
Likewise, a large object will have low spatial frequency and
will be more accurately displayed. If an object is displayed accurately as it is,
then the MFT is given a value of 1.0. The modulation transfer function scale
is from zero to 1. A MTF value of 0 would mean the image is blank and
contains no information about the object scanned. Scanned objects will
have values between 0 and 1; however, the closer to 1 an object is the
better the MTF of the scanner.
This picture
demonstrates
how object
size is related
to their
spatial
frequency.
Small objects
have low
spatial frequency since more of them fit into a prescribed length. Large
objects have high spatial frequency since they fit fewer times into a given
length.
In practical terms, the information needed during quality
assurance testing or when comparing CT scanners for purchase appear in
graph form called MTF graphs. Modular transfer function is plotted along
the y-axis and object spatial frequency along the x-axis. When comparing
the function of a scanner over time, or when comparing the performance of
different scanners, we look at what is called limiting resolution. Limiting
resolution is the spatial frequency at MTF of 0.1 for any scanner. MTF of 0.1
is referenced because it is the lowest MTF that will result in a visible CT
image. A scanner with a higher spatial frequency will be able to image small
objects.
90
This graph of MTF and
spatial frequency for
three scanners is
shown. At limiting
resolution (0.1 MTF)
scanner "A" have a
spatial frequency of
11.5 lp/cm. Scanner
"B" at a MTF of 0.1
gives a spatial frequency of 17.0, and scanner "C" the spatial frequency is
20.0. Interpretation of the graph implies that scanner "C" is better able to
display small objects than scanners "B or C".
91
CT can resolve small differences in tissue densities because
there must be at least a 10% density difference to be detected with
conventional x-ray imaging. This is why we use a 15% increase or decrease
in the kVp to change the scale of contrast in conventional radiography. CT
on the other hand is able to resolve density differences as little as 0.1%.
Notwithstanding, the size of a low contrast object, its inherent density
(calcium vs. fat), image noise, and viewing window setting will in part
determine its detectability.
To perform this test various types of low-contrast inserts are
available for the CT phantom: therefore, scanning for this test is
manufacturer specific. Our low contrast detectability test phantom image is
defined by the smallest hole size visible for a given contrast level and dose.
The phantom contains a doped polystyrene membrane suspended in water.
The membrane is pierced with holes ranging from 10.0mm, 7.5mm, 5.0mm,
3.0mm and 1.0mm.
The basic of this test is that the number of object visualized
on the phantom image is determined, and the mean value of each visualized
hole and surrounding material is recorded. The smallest holes that should
be visualized is 5 mm in diameter or smaller for 5% contrast objects.
This low
contrast
detectability
phantom
image
displays
various sized
holes used to
determine
low contrast (left). The various sizes are labeled on the right image;
however, it is difficult to see the smallest holes. This test measures the
scanners ability to detect an objects density when it is close to background
density.
92
Low-resolution contrast is determined as the difference in HU
of objects and background. High noise in the image will cause a decrease in
low-contrast resolution. To get an accurate measure of low contrast we
need to know the CT number for the polystyrene membrane. This is
accomplished by taking the CT number for water over an area that does not
include the membrane. A second measurement is taken over an area that
includes the membrane superimposed on water (called water plus the
membrane). The CT number taken for water is subtracted from the water
plus membrane to get the CT number for the membrane.
To perform this test, measure using a box ROI above and
below the membrane in the water section (labeled A and B in the phantom
image below). Take a box ROI in the polystyrene membrane above the holes
(labeled B) and below the holes (labeled C). Subtract "A" from "B" and
subtract "D" from "C". When the measurements are completed and
recorded adjust the Window Width to 20 and the Window Level to the CT
number recorded for water. This will allow for an accurate reading of the
number of holes visible.
93
Test 5 - Slice thickness (sensitivity profile)
94
The quality assurance phantom on the right is used to
perform several tests including the slice thickness test. On the left is the
resulting image used to determine slice thickness. The block pattern
showing line thickness appears along the edges (yellow oval) of the pattern.
Because
several factors affect
viewing of the line
pairs for width
sensitivity profile, it
is recommended on
some scanners that
the window/level
setting be
standardized. This
chart shows the
standard viewing
windows for the GE
Lightspeed scanners for each slice thickness being tested for. A slice
thickness of 10mm is viewed at W/L 250/50, whereas for 3mm the
recommended viewing is 250/-50 W/L. When this test is properly performed
and the collimators working correctly, the number of visible lines should
equal the chosen slice thickness.
95
This image
taken on the
CT phantom
shows 10 one-
millimeter lines
on the corners
of the image.
The slice
thickness is
10mm
according to this image, which is what was set at the console for the slice
thickness test. The window and level setting for this image was set at
250/50 according to the manufacturer�s recommendation. The results of
this test confirm that the collimators that shape the x-ray beam are open to
the appropriate size. One can also vary the slice thickness to test for
linearity of the system. Collimators that shape slice thickness should be
accurate to +/- 1mm of the setting at 10mm.
96
PATIENT PREPARATION and USE OF CONTRAST AGENTS
The patient preparation for most CT examinations is minimal.
In certain cases, a contrast medium is used to enhance designated areas and
to better differentiate disease processes. Although the specified
preparation varies according to the hospital and physician, the following
preparations are frequently used.
• Patient is to arrive 1 ½ hrs prior to exam time, so we can give the oral
preparation in the department
97
The night before the exam before going to bed, all patients should
have 10 mls of Gastrografin mixed in 450 mls (16 ozs.) of clear fluid
(including water).
Patients should have nothing to eat, drink or smoke for two (2) hours
prior to the test, other than the Gastrografin preparation.
CHEST CT PREPARATION:
• Please notify the technologist if you have had a barium study within
the last 3 days.
98
I.V. contrast medium is occasionally needed to enhance the
contrast in CT scans of the brain, body organs and extremities. The contrast
medium used is the same as that employed for Excretory Urography. Oral
Contrast agents (Barium or an Iodinated contrast agents for GI work) are
used for enhancement in certain body examinations.
PATIENT POSITIONING:
Patient Positioning is determined by;
99
One exception is the CT
examination of the spine for
which slices through the disk
space are obtained at various
required angles. Up to 25
degrees of angulation maybe
required for imaging of the L5-
S1 intervertebral spaces.
• For routine Brain scans and most body scans, a one (1) centimeter slice
thickness is generally acceptable.
100
• In some studies such as generalized survey for Lymphoma or metastasis,
one (1) centimeter slices are generally taken every other centimeter.
• The total number of slice depends on the size of the anatomic area to
be scanned.
The brain, being a three-dimensional object, can be cut into three different
planes of orientation. They are:
• Coronal
• Sagittal
• Horizontal
101
TWO KINDS OF BODY POSITIONING EMPLOYED IN CT SCANNING
GANTRY ANGULATION
1. “0” Angle
2. “+” Angle
3. “-” Angle
102
CT SCANNING PROCEDURES
CRANIAL CT
A. STANDARD CT BRAIN (With or Without Contrast)
Indications:
A CT scan of the brain may be performed to assess the brain for tumors and
other lesions, injuries, intracranial bleeding, structural anomalies such as
hydrocephalus, infections, brain function or other conditions, particularly
when another type of examination such as X-rays or physical examination
are not conclusive.
Evaluate the extent of bone and soft tissue damage in patients with
facial trauma, and planning surgical reconstruction.
Diagnose diseases of the temporal bone on the side of the skull,
which may be causing hearing problems.
Determine whether inflammation or other changes are present in
the paranasal sinuses.
Plan radiation therapy for cancer of the brain or other tissues.
103
Guide the passage of a needle used to obtain a tissue sample
(biopsy) from the brain.
Assess aneurysms or arteriovenous malformations through a
technique called CT angiography.
104
PROCEDURES:
• Head is positioned in
normal AP of the skull.
• One centimeter slice
thickness is generally
employed.
• The first slice/line starts
from the OML to the
vertex.
1. POSTERIOR FOSSA
105
Located centrally in the posterior
fossa is the foramen magnum. The
posterior fossa is surrounded by deep
grooves containing the transverse sinuses
and sigmoid sinuses. This location within
the head requires a special use of scan
parameters to observe and categorized
pathologies.
Additionally, special care and
attention has to be taken to artifacts caused mainly by beam
hardening.
106
AXIAL CUTS
107
Patient is in supine position in normal head positioning.
1-2 millimeters slice thickness/increment is employed
Axial scanning along the OML.
Filming program is one film for Bone window and one film for soft
window.
108
PETROUS PYRAMID & INTERNAL
AUDITORY MEATUS
109
CT OF ORBITS
• Normal head positioning
• 2 millimeters slice thickness
• Scanning Technique: Axial/Transverse & Coronal Cuts
• Filming program is Soft & Bone windows
• Inject 80ml of contrast medium.
CT SELLA TURCICA
The
Pituitary gland
is the master
gland of the
body because
it controls
most of the body’s endocrine functions by
means of the hypothalamic-pituitary axis. The Sella Turcica is a depression
hollowed out on the upper surface of the body of the Sphenoid bone
wherein the pituitary gland lies within it.
110
The sellar and parasellar region is an anatomically complex
area where a number of neoplastic, infectious, inflammatory,
developmental and vascular pathologies can occur. Differentiation among
various etiologies may not always be easy, since many of these lesions may
mimic the clinical, endocrinologic and radiologic presentations of pituitary
adenomas. The diagnosis of sellar lesions involves a multidisciplinary effort,
and detailed endocrinologic, ophthamologic and neurologic testing are
essential. CT and, mainly, MRI are the imaging modalities to study and
characterize normal anatomy and the majority of pathologic processes in
this region.
Technique:
111
Computed tomography (CT) was
the first imaging modality to directly
visualize the pituitary gland, hypothalamus,
and optic chiasm. The bony structures in
this region can be well evaluated with CT.
It is more sensitive than either
plain radiographs or magnetic resonance
imaging (MRI) in the detection of
calcifications within soft tissues. However,
intravenous contrast agents frequently are
necessary to improve the image contrast and
to enhance the vasculature, and CT involves
radiation exposure.
112
Computed Tomography
Scan of the Spine
Examinations of the spinal
column with both, Magnetic Resonance
(MR) imaging and Computed Tomography
(CT), often require a precise three-
dimensional positioning, angulation and
labeling of the spinal disks and the
vertebrae. A fully automatic and robust
approach is a prerequisite for an automated
scan alignment as well as for the
segmentation and analysis of spinal disks
and vertebral bodies in Computer Aided
Diagnosis (CAD) applications.
CT scans of the spine can
provide more detailed information about
the vertebrae (bones of the spine) and other
spinal structures and tissues than standard
X-rays of the spine, thus providing more
information related to injuries and/or diseases of the spine.
113
Anatomy of the Spine
The spinal column is made up of 33 vertebrae that are
separated by spongy disks and classified into distinct areas.
114
If you are pregnant or suspect that you may be pregnant, you
should notify your physician. Radiation exposure during pregnancy may lead
to birth defects. If it is necessary for you to have a CT of the spine, special
precautions will be made to minimize the radiation exposure to the fetus.
Nursing mothers should wait 24 hours after contrast material is injected
before resuming breastfeeding.
If contrast dye is used, there is a risk for allergic reaction to
the dye. Patients who are allergic to or sensitive to medications should
notify their physician. Studies show that 85 percent of the population will
not experience an adverse reaction from iodinated contrast; however, you
will need to let your physician know if you have ever had a reaction to any
contrast dye, and/or any kidney problems. A reported seafood allergy is not
considered to be a contraindication for iodinated contrast.
Patients with kidney failure or other kidney problems should
notify their physician. In some cases, the contrast dye can cause kidney
failure. The effects of kidney disease and contrast agents have attracted
increased attention over the last decade, as patients with kidney disease are
more prone to kidney damage after contrast exposure.
Also, patients taking the diabetes medication (Glucophage)
should alert their doctor before having IV contrast, as it may cause a rare
condition called metabolic acidosis. If you take metformin, you will be asked
to stop taking it 24 hours before and for 48 hours after your injection. A
blood test may be required before you can start taking metformin again.
Generally, a CT scan of the bones, joints, and soft tissue follows this process:
• You will be asked to remove any clothing, jewelry, or other objects that
may interfere with the procedure, such as eyeglasses, hairpins,
dentures, and possibly hearing aids.
• If you are to have a procedure done with contrast, an intravenous (IV)
line will be started in the hand or arm for injection of the contrast dye.
For oral contrast, you will be given medication to swallow.
115
You will lie on a scan table that slides into a large, circular opening of
the scanning machine.
Pillows and straps may be used to prevent movement during the
procedure.
The technologist will be in another room where the scanner controls
are located. However, you will be in constant sight of the
technologist through a window.
Speakers inside the scanner will enable the technologist to
communicate with and hear you.
You will have a call button so that you can let the technologist know
if you have any problems during the procedure.
The technologist will be watching you at all times and will be in
constant communication.
• The scanner will begin to rotate around you and X-rays will pass
through the body for short amounts of time. You will hear clicking
sounds, which are normal.
• The X-rays absorbed by the body's tissues will be detected by the
scanner and transmitted to the computer. The computer will
transform the information into an image to be interpreted by the
radiologist.
It will be important for you to remain very still during the procedure.
You may be asked to hold your breath at various times during the
procedure.
If contrast dye is used for your procedure, you will be removed from
the scanner after the first set of scans has been completed. A second
set of scans will be taken after the contrast dye has been
administered.
If contrast dye is used for your procedure, you may feel some effects
when the dye is injected into the IV line. These effects include a
flushing sensation, a salty or metallic taste in the mouth, a brief
headache, or nausea and/or vomiting. These effects usually last for a
few moments.
When the procedure has been completed, you will be removed from
the scanner.
116
If an IV line was inserted for contrast administration, the line will be
removed.
You may be asked to wait for a short period of time while the
radiologist examines the scans to make sure they are clear.
CT of the lumbar is
always in supine
position with the
Scanogram taken in
Lateral view. The
slice thickness is from
2-3mm.
Slices are taken from
the intervertebral spaces perpendicular from the superior end plate of
the body to the inferior end plate from L1 to S1.
Printed in Soft & Bone window.
CT Myelogram
CT scanning involving the
diagnostic tool of Myelography is used to
evaluate both the bony structure of the spine as
well as the nerve structures. In preparation for
this procedure the patient may be asked to
remove any metal, hearing aids, and/or
removable dental work. This procedure requires
the patient to follow a few preps that will be
provided by your ordering physician.
This procedure requires two
parts. During the first portion of this procedure
the patient will have Contrast dye injected into
his or her cerebral spinal fluid which coats the
spinal canal.
117
The contrast dye will coat the
spinal canal, the spinal cord, and the
nerve roots, which will illuminate any
malformations in the spine during the
scan. The patient will then be
transferred to the CT table where they
will be asked to lie on his or her back for
the second portion of this procedure.
Positioning sponges and Velcro may be used to reduce motion and aligning
the spine for the scan.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room
the table will then move through the scanner producing a “scout”
image. There will be a slight pause while the technologist uses the scout
image to set up the exam.
The table will then move through the scanner at least one
more time for the examination. Patients are not permitted to drive
immediately following this procedure and will need to make arrangements
for transportation home.
CT Cervical Spine
CT scanning of
the Cervical Spine is
typically used to evaluate
the spine before and after
surgery, detect various
types of tumors, diseases,
fractures, spinal stenosis,
disc herniations, and or
any malformations. In
preparation for this procedure the patient may be asked to remove any
metal, hearing aids, and/or removable dental work. If the procedure is an
Enhanced Study it calls for contrast media to be used and further
preparations may be needed.
118
The patient typically lies on his or her back and their head is
placed into a cradle. Small positioning sponges or Velcro straps may be
used to reduce motion and keep the head and neck aligned properly for the
scan.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room
the table will then move through the scanner producing a “scout”
image. There will be a slight pause while the technologist uses the scout
image to set up the exam. The table will then move through the scanner at
least one more time for the examination depending on whether or not
contrast media is needed.
119
Note: CT Cervical is excellent in demonstrating the complete bony ring of
C1.
CT Thoracic Spine
CT scanning of the Thoracic or
Dorsal Spine is typically used to evaluate
the spine before and after surgery,
detect various types of tumors, diseases,
fractures, spinal stenosis, disc
herniations, and or any
malformations. In preparation for this
procedure the patient may be asked to
remove any metal, hearing aids, and/or
removable dental work. If the
procedure is an Enhanced Study it calls
for contrast media to be used and further preparations may be needed.
120
The patient typically lies on his or her back with their feet
facing the scanner. Positioning sponges may be placed under the knees for
comfort and to assist in aligning the spine for the scan. If possible the
patient will be asked to raise their arms above their head. The technologist
will center the area being scanned using the laser lights on the scanner.
Shortly after the technologist leaves the room the table will
then move through the scanner producing a “scout” image. There will be a
slight pause while the technologist uses the scout image to set up the
exam. The table will then move through the scanner at least one more time
for the examination depending on whether or not contrast media is
needed. You must inform the technologist, radiology nurse and/or
physician of any allergies you may have before your exam.
While positioning you on the exam table, the technologist will
explain your procedure. If contrast dye is being used, it will be injected
through your IV. During the injection you may experience a warm sensation
all over your body and a metal taste in your mouth. This is normal.
If you experience any itching, sneezing, nasal congestion, scratchy throat or
swelling of your face you should notify the technologist immediately.
Most often, you will be asked to lie flat on your back with your arms
positioned above your head.
The table you are on will slide into the scanner. A portion of
the body is covered by the scanner, and the scanner is open at the back and
the front, allowing the patient to see out.
The technologist will always be able to see and hear you during your exam.
You will be asked to hold very still and at times to hold your breath.
This procedure usually takes approximately 30 minutes.
121
The goal of diagnostic imaging in patients who have sustained
thoracic spinal trauma is to correctly identify spinal fractures, to identify
injuries to the spinal cord and nerve roots, to aid in surgical planning, and to
judge the stability of postoperative fixation. This article highlights the typical
patterns of injury within a classification based on the mechanism of injury
while focusing on the imaging methods that are most useful in clinical
practice.
CT Paranasal Sinuses
CT scanning of the Sinuses is typically used to evaluate the
sinus cavities for fluid, thickened sinus membranes, planning for surgery,
tumors, and to detect any inflammation. In preparation for this procedure
the patient may be asked to remove any metal, hearing aids, and/or
removable dental work. If the procedure is an Enhanced Study it calls for
contrast media to be used and further preps may be needed.
122
The patient typically lies on his or her back and their head is
placed into a cradle. Depending on what is being evaluated during the
procedure the patient may be asked to lie on his or stomach and place their
chin on a positioning device. Small positioning sponges or Velcro straps may
be used to reduce motion and keep the head aligned properly for the scan.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room
the table will then move through the scanner producing a “scout”
image. There will be a slight pause while the technologist uses the scout
image to set up the exam. The table will then move through the scanner at
least one more time for the examination depending on whether or not
contrast media is needed.
CORONAL CUTS:
- 5 millimeters slice thickness
- Reference point is from the back of the sphenoid sinuses up to the
frontal sinuses.
123
CT OF NECK REGION
CT of neck for soft tissue evaluation generally includes
imaging of the following anatomic structures:
Pharynx, Thyroid Glands, Nasopharynx, Parotid glands & Submandibular
glands.
A. CT THYROID GLANDS
A CT Soft Tissue Neck is
typically ordered to evaluate the soft
tissues and organs of the
neck. Capturing images of all anatomy
including the tonsils, adenoids, airways,
thyroid, glands, and the blood vessels:
evaluating for tumors, masses, and or
swelling. In preparation for this
procedure the patient may be asked to
remove any metal, hearing aids, and/or removable dental work.
124
If the procedure is an
Enhanced Study it calls for contrast media to
be used and further preps may be needed.
The patient typically lies on his
or her back and their head is placed into a
cradle. Small positioning sponges or Velcro
straps may be used to reduce motion and
keep the head aligned properly for the
scan. The technologist will center the area being scanned using the laser
lights on the scanner.
Shortly after the technologist leaves the room the table will
then move through the scanner producing a “scout” image. There will be a
slight pause while the technologist uses the scout image to set up the
exam.
The table will then move through the scanner at least one
more time for the examination depending on whether or not contrast
media is needed. The technologist may ask you to withhold from
swallowing during the scan to reduce motion and will inform you when it is
ready to resume normal swallowing activities.
CT Upper Extremity
A CT scan of the Upper Extremities is typically used to
evaluate abnormalities found within the soft tissue and bony anatomy of
that extremity. CT scans of
this nature assess the
muscles, bones, joints, and
soft tissues for arthritis,
fractures, tumors,
bleeding, and
inflammation.
In preparation for this procedure the
patient may be asked to remove any metal and change into a gown. If the
procedure is an Enhanced Study it calls for contrast media to be used and
further preparations may be needed.
125
The patient
typically lies on his or her
back. Small positioning
sponges or Velcro straps
may be used to reduce
motion and keep the
Extremity being imaged
aligned properly for the
scan. The technologist
will center the area being scanned using the laser lights on the
scanner. Shortly after the technologist leaves the room, the table will move
through the scanner producing a “scout” image. There will be a slight pause
while the technologist uses the scout image to set up the exam. The table
will then move through the scanner at least once more depending on
whether or not contrast dye is being used.
CT Lower Extremity
A CT scan of the Lower
Extremities is typically used to evaluate
abnormalities found within the soft tissue and
bony anatomy of that extremity. CT scans of
this nature assess
the muscles,
bones, joints, and
soft tissues for
arthritis,
fractures, tumors, bleeding, and inflammation.
In preparation for this procedure the
patient may be asked to remove any metal and
change into a gown. If the procedure is an
Enhanced Study it calls for contrast media to
be used and further preparations may be needed.
126
The patient typically lies on his
or her back. Small positioning sponges or
Velcro straps may be used to reduce
motion and keep the Extremity being
imaged
aligned
properly for
the scan. The
technologist
will center the area being scanned using the
laser lights on the scanner. Shortly after the
technologist leaves the room, the table will
move through the scanner producing a “scout”
image. There will be a slight pause while the
technologist uses the scout image to set up the exam. The table will then
move through the scanner at least once more depending on whether or not
contrast dye is being used.
CT CHEST
CT scanning of the Chest is typically used to evaluate
abnormalities found on conventional chest X-rays or in diagnosing clinical
symptoms of coughing, chest pain, fever, or trauma. CT scans of the chest
may also evaluate for Pulmonary Embolisms (Blood Clots), tumors, and lung
nodules. In preparation for this procedure the patient may be asked to
remove any metal and change into a gown.
127
The patient typically lies on his or her back with their feet
facing the scanner. Positioning sponges may be placed under the knees for
comfort and to assist in aligning the body for the scan. If possible the
patient will be asked to raise their arms above their head. The technologist
will center the area being scanned using the laser lights on the scanner.
Shortly after the technologist leaves the room, a recorded
voice will request the patient to hold their breath. At this point the table
will then move through the scanner producing a “scout” image, and the
patient will be instructed to breathe normally. There will be a slight pause
while the technologist uses the scout image to set up the exam. The table
will then move through the scanner at least one more time for the
examination depending on whether or not contrast media is needed.
Procedure:
128
CT Abdomen and Pelvis
CT scans of the Abdomen, Pelvis, or
a combination of the two are typically used to
evaluate abnormalities found within the organs of
the abdominopelvic cavity. CT scans of this nature
assess the liver, urinary tract, spleen, intestines,
and vasculature for multiple diseases and or
infections. In preparation for this procedure the
patient may be asked to
remove any metal and
change into a gown.
The patient typically lies on his or her back
with their feet facing the scanner. Positioning
sponges may be placed under the knees for comfort
and to assist in aligning the body for the scan. If
possible the patient will be asked to raise their arms
above their head. The technologist will center the
area being scanned using the laser lights on the
scanner. Shortly after the technologist leaves the room, a recorded voice
will request the patient to hold their breath.
At this point the table will then move through the scanner
producing a “scout” image, and the patient will be instructed to breathe
normally. There will be a slight pause while the technologist uses the scout
image to set up the exam. The table will then move through the scanner at
least one more time for the examination depending on whether or not
contrast media is needed.
129
7. Abdominal aorta
8. Inferior Vena Cava
9. Abdominal lymph nodes
10. Other retroperitoneal structures.
Procedure:
• Scanogram is taken in AP.
• Contrast medium is a must before the
procedure.
• Axial cuts only with 10mm slice
thickness.
• “0” gantry angulation.
• Reference line starting above the
diaphragm dome to the symphysis
pubis.
130
CT Liver
The CT scan of the Abdomen using a Liver Protocol is typically
used to evaluate abnormalities found within the Liver and Hepatic
System. CT scans of this nature assess for liver tumors, bleeding, jaundice,
and liver diseases. In preparation for this procedure the patient may be
asked to remove any metal and change into a gown. This procedure is an
Enhanced Study meaning it calls for contrast media to be used. Further
preps may be needed and should be provided by the ordering physician. If
you are allergic to IV contrast, diabetic, or experiencing lower kidney
functions inform the ordering physician and the technologist.
131
Procedure:
• Without contrast
CT Pancreas
The CT Scan of the Abdomen
using a Pancreatic Protocol is typically used to
evaluate abnormalities found within the
Pancreas. CT scans of this nature assess for
tumors and Pancreatitis, or inflammation of
the pancreas. In preparation for this procedure
the patient may be asked to remove any metal
and change into a gown. This procedure is an
Enhanced Study meaning it calls for contrast
media to be used. Further preparations may be needed and should be
provided by the ordering physician.
If you have allergic to I.V. contrast, diabetic or experiencing lower kidney
functions, inform the physician and the technologist.
The patient may be asked to drink an oral contrast prior to the
scan in order to fill the stomach and small intestines. During the scan the
patient typically lies on his or her back with their feet facing the
canner. Positioning sponges may be placed under the knees for comfort
and to assist in aligning the body for the scan. The technologist will place an
Intravenous (IV) Catheter in the patient’s arm. Note: this procedure is an
Enhanced Study and calls for IV contrast to be used. If possible the patient
will be asked to raise their arms above their head.
132
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room, a
recorded voice will request the patient to hold their breath. At this point
the table will then move through the scanner producing a “scout” image,
and the patient will be instructed to breathe normally. There will be a slight
pause while the technologist uses the scout image to set up the exam. The
table will then move through the scanner several times as the IV Contrast is
injected. This catches the pancreas in multiple stages of enhancement
which aids with the diagnosis.
Procedure:
CT Cardiac Score
Cardiac Scoring in CT is used to
evaluate for a condition called Coronary
Artery Disease. Major risk factors for
Coronary Artery Disease include but are not
limited to diabetes, smoking, high blood
pressure, high blood cholesterol, and
obesity.
133
In preparation for this procedure the patient may be asked to
remove any metal and change into a gown. If you are a smoker or drink
large amounts of caffeine you may be asked to withhold from doing so a
few hours prior to your exam. A complete prep will be provided by your
ordering physician.
The patient typically lies on his or her back with their feet
facing the scanner. Positioning sponges may be placed under the knees for
comfort and to assist in aligning the body for the
scan. If possible the patient will be asked to
raise their arms above their head. A set of
electrodes will be placed on the patient’s chest
connecting them to an ECG (electrocardiograph)
machine which records the electrical activity of
the heart. The technologist will center the area
being scanned using the laser lights on the scanner.
Shortly after the technologist leaves the room, a recorded
voice will request the patient to take a shallow breath in and hold it. At this
point the table will then move through the scanner producing a “scout”
image, and the patient will be instructed to breathe normally. There will be
a slight pause while the technologist uses the scout image to set up the
exam. The table will then move through the scanner at least one more time
for the examination and the request to suspend breathing will be
announced again.
CT Urogram
A CT Urogram is typically used to evaluate abnormalities
found within the Urinary System. CT scans of this nature assess the kidneys,
ureters, and bladder for
blockages, kidney stones,
infections, growths, and other
diseases. In preparation for this
procedure the patient may be
asked to remove any metal and
change into a gown.
134
This procedure is an Enhanced Study meaning it calls for
contrast media to be used. Further preps may be needed and should be
provided by the ordering physician. If you are allergic to IV contrast,
diabetic, or experiencing lower kidney functions inform the ordering
physician and the technologist.
The patient typically lies on his or her back with their feet
facing the scanner. Positioning sponges may be placed under the knees for
comfort and to assist in aligning the body for the scan. The technologist will
place an Intravenous (IV) Catheter in the patient’s arm. Note: this
procedure is an Enhanced Study and calls for IV contrast to be used.
If possible the patient will be asked to raise their arms above
their head. The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room, a
recorded voice will request the patient to hold their breath. At this point
the table will then move through the scanner producing a “scout” image,
and the patient will be instructed to breathe normally.
There will be a slight pause while the technologist uses the
scout image to set up the exam. The table will then move through the
scanner as the IV Contrast is being injected. Once that portion of the scan in
completed the technologist will assist the patient in relaxing his or her arms
at their side. The patient will have to wait a total of ten minutes at this
point in order to allow the IV Contrast to fill the Urinary Tract.
After the ten minute mark, the patient will be asked to raise
their arms for another scan. If the urinary tract is filled with contrast the
scan is complete, if not, the patient will be asked to lay prone (on their
abdomen) for another scan to evaluate the Urinary System.
135
CT Angiography Head / Circle of Willis
CT Angiography of the head is
typically used to evaluate abnormalities
found within the blood vessels that run
through the
brain. CT
scans of this
nature assess
the arteries
and veins for
plaque, atherosclerosis, dissections, and
aneurysms. In preparation for this procedure
the patient may be asked to remove any
metal and change into a gown.
This procedure is an Enhanced Study meaning it calls for
contrast media to be used. Further preps may be needed and should be
provided by the ordering physician. If you are allergic to IV contrast,
diabetic, or experiencing lower kidney functions inform the ordering
physician and the technologist.
The patient typically lies on his or her back and their head is
placed into a cradle. Small positioning sponges or Velcro straps may be
used to reduce motion and keep the head and neck aligned properly for the
scan. The technologist will place an Intravenous (IV) Catheter in the
patient’s arm. Note: this procedure is an Enhanced Study and calls for IV
contrast to be used.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room,
the table will move through the scanner producing a “scout” image.
There will be a slight pause while the technologist uses the scout image to
set up the exam. The table will then move through the scanner at least two
additional times.
136
CT Angiography Neck (Carotids)
137
CT Angiography Chest
CT Angiography of the
chest is typically used to evaluate
abnormalities found within the blood
vessels that run through that portion of
the body. CT scans of this nature assess
the arteries and veins for plaque,
atherosclerosis, dissections, and
aneurysms. In preparation for this
procedure the patient may be asked to
remove any metal and change into a
gown. This procedure is an Enhanced
Study meaning it calls for contrast media to be used. Further preps may be
needed and should be provided by the ordering physician. If you are allergic
to IV contrast, diabetic, or experiencing lower kidney functions inform the
ordering physician and the technologist.
The patient typically lies on his or
her back with their feet facing the
scanner. Positioning sponges may be placed
under the knees for comfort and to assist in
aligning the body for the scan. The
technologist will place an Intravenous (IV)
Catheter in the patient’s arm. Note: this
procedure is an Enhanced Study and calls for
IV contrast to be used. If possible the patient
will be asked to raise their arms above their head.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room, a
recorded voice will request the patient to hold their breath. At this point
the table will then move through the scanner producing a “scout” image,
and the patient will be instructed to breathe normally. There will be a slight
pause while the technologist uses the scout image to set up the exam. The
table will then move through the scanner at least two additional times.
138
CT Angiography Abdomen/Pelvis
CT Angiography of the Abdomen, Pelvis,
or a combination of the two is typically used to
evaluate abnormalities found within the blood
vessels that run through that portion of the
body. CT scans of this nature assess the arteries
and veins for plaque, atherosclerosis, dissections,
and aneurysms. In preparation for this procedure
the patient may be
asked to remove
any metal and change into a gown.
This procedure is an Enhanced
Study meaning it calls for contrast media to
be used. Further preps may be needed and
should be provided by the ordering
physician. If you are allergic to IV contrast,
diabetic, or experiencing lower kidney
functions inform the ordering physician and
the technologist.
The patient typically lies on his or her back with their feet
facing the scanner. Positioning sponges may be placed under the knees for
comfort and to assist in aligning the body for the scan. The technologist will
place an Intravenous (IV) Catheter in the patient’s arm. Note: this
procedure is an Enhanced Study and calls for IV contrast to be used. If
possible the patient will be asked to raise their arms above their head.
The technologist will center the area being scanned using the
laser lights on the scanner. Shortly after the technologist leaves the room, a
recorded voice will request the patient to hold their breath. At this point
the table will then move through the scanner producing a “scout” image,
and the patient will be instructed to breathe normally. There will be a slight
pause while the technologist uses the scout image to set up the exam. The
table will then move through the scanner at least two additional times.
139
CT Angiography Extremities
CT Angiography of the Extremities is
typically used to evaluate abnormalities found within
the blood vessels that run through that portion of the
body.
CT scans of this nature assess the arteries
and veins for plaque, atherosclerosis, dissections, and
aneurysms. In preparation for this procedure the
patient may be asked to remove any metal and change
into a gown.
This procedure is an Enhanced Study
meaning it calls for contrast media to be used. Further
preps may be needed and should be provided by the
ordering physician. If you are allergic to IV contrast, diabetic, or
experiencing lower kidney functions inform the ordering physician and the
technologist.
The patient typically lies on his or her back. Small positioning
sponges or Velcro straps may be used to reduce motion and keep the
Extremity being imaged aligned properly for the scan. The technologist will
place an Intravenous (IV) Catheter in the patient’s arm. Note: this
procedure is an Enhanced Study and calls for IV contrast to be used. The
technologist will center the area being scanned using the laser lights on the
scanner. Shortly after the technologist leaves the room, the table will move
through the scanner producing a “scout” image. There will be a slight pause
while the technologist uses the scout image to set up the exam. The table
will then move through the scanner at least two additional times.
140
CT Heart Angiography
CT Heart Angiography is commonly used to evaluate for a
condition called Coronary Artery Disease. Major risk factors for Coronary
Artery Disease include but are not limited to diabetes, smoking, high blood
pressure, high blood cholesterol, and obesity.
During this procedure the Heart,
Cardiac blood vessels and circulation, and
great vessels are imaged in both high
resolution and 3-Dimensionally as the heart
beats.
In preparation for this procedure
the patient may be asked to remove any
metal and change into a gown. If you are a
smoker or drink large amounts of caffeine you may be asked to withhold
from doing so a few hours prior to your exam. A complete prep will be
provided by your ordering physician.
The patient typically lies on his
or her back with their feet facing the
scanner. Positioning sponges may be placed
under the knees for comfort and to assist in
aligning the body for the scan. The
technologist will place an Intravenous (IV)
Catheter in the patient’s arm. Note: this
procedure is an Enhanced Study and calls for
IV contrast to be used. If you are allergic,
diabetic, or experiencing lower kidney functions inform the ordering
physician and the technologist.
After the IV access is established, a set of electrodes will be
placed on the patient’s chest connecting them to an ECG
(electrocardiograph) machine which records the electrical activity of the
heart. If possible the patient will be asked to raise their arms above their
head. The technologist will center the area being scanned using the laser
lights on the scanner. Shortly after the technologist leaves the room, a
recorded voice will request the patient to take a shallow breath in and hold
it.
141
At this point the table will then move through the scanner
producing a “scout” image, and the patient will be instructed to breathe
normally. There will be a slight pause while the technologist uses the scout
image to set up the exam. A medication that slows or stabilizes the
patient’s heart may be used to increase the diagnostic value of the
images. The table will then move through the scanner at least one more
time for the examination and the request to suspend breathing will be
announced again.
142