VIVIX-S 1717N Specifications
VIVIX-S 1717N Specifications
VIVIX-S 1717N Specifications
Contents
1. Instruction ......................................................................................................................................... 3
2. Products ............................................................................................................................................. 5
3. Performance .................................................................................................................................... 13
4. Defect ............................................................................................................................................... 15
1. Instruction
1.1.1 Target
This document is intended for customers who use the VIVIX-S 1717N detector.
1.1.2 Symbols
This product should be operated under the safety instructions with the warning or caution symbol in this
manual. It is important for you to read and understand the contents to operate the products safely.
This symbol is used to indicate a potentially hazardous situation that may cause death,
personal injury or substantial property damage if the instructions are ignored. Users
should be well acquainted with this symbol and the related contents.
Information
This symbol is used for indicating product related references and supplementary
information. Users are recommended to read the sentences with this notice carefully.
1.1.3 Notations
Bold Types
Words in bold indicate products terms, or the sentences which are needed to transmit clear meaning to the
customers.
1.3 Contact Us
For comments or inquiries regarding this document and relevant products, contact via email below.
Item Contents
Department Customer Support Team at Vieworks
E-mail [email protected]
2. Products
2.1 Detector
2.1.1 Specifications
Item Specifications
FXRD-1717NA (CsI) FXRD-1717NB (Gadox)
Model
FXRD-1717NAW (CsI) FXRD-1717NBW (Gadox)
Image Sensor TFT: a-Si (Amorphous Silicon)
FXRD-1717NA(W): Csl: TI (Thallium doped Caesium Iodide)
X-ray Scintillator type
FXRD-1717NB(W): Gd2O2S:Tb (Gadolinium oxysulfide)
Pixel Pitch 0.14㎜ (140㎛)
Field of View 17” x 17”
Active Area (H x V) 430.08㎜ × 430.08㎜
Active Array 3072 x 3072 pixels
FXRD-1717NA(W): 426.72㎜ x 426.72㎜
Effective Area
FXRD-1717NB(W): 428.4㎜ x 428.4㎜
FXRD-1717NA(W): 3048 x 3048 pixels
Effective Array
FXRD-1717NB(W): 3060 x 3060 Pixels
Grayscale 16 bit
Spatial Resolution Min. 3.5 lp/㎜
Image Acquisition Time (Wired) 1.5 sec.
Image Acquisition Time (Wireless) 3 sec.
Capacity for Image Bakcup Max. 200 images
Recommended Cycle Time 15 sec.
AED (Auto Exposure Detection)
X-ray Synchronization Control
DR Trigger (External line trigger)
DC +24V, Max. 1A
Rated Power Supply Wired: Powered by SCU with a tether interface cable.
Wireless: Powered by a battery pack (3,100 ㎃ h X 2)
Power Consumption Max. 24W
Operating Time 6 hours (when sleep mode is off)
Dimensions (H × W × D) 460㎜ × 460㎜ × 15.5㎜
Weight 4.2kg (Wired) 4.3kg (Wireless / without a battery pack)
Wired: Gigabit Ethernet (1000BASE-T) via PoE (Power over
Image Transfer Ethernet)
Wireless: IEEE802.11n
Data Transmission Rate (Wired) Max. 1Gbps
Data Transmission Rate (Wireless) Max. 300Mbps (MIMO 2X2)
Item Description
Dimensions (H × W × D) 460.0㎜ × 460.0㎜ × 15.5㎜
Curvature of Edges R6.5
The allowed tolerance of thickness of a detector is from -2.0㎜ ~ +0.5㎜. (Under the
ISO4090 regulation).
Item Description
Dimensions (H × W × D) 460.0㎜ × 460.0㎜ × 15.5㎜
Curvature of Edges R6.5
The allowed tolerance of thickness of a detector is from -2.0㎜ ~ +0.5㎜. (Under the
ISO4090 regulation).
2.1.3 Functions
Name Description
Used for tighten the tether interface cable.
A Tether interface connector
Used for wired connection between a detector and SCU.
B Tether interface holder Fix/release holder of the tether interface cable.
Detector status indicator
DATA LED (Blue): Indicates communication and transmission
C Status indicator
ACTIVE LED (Orange): Indicates ready to work
POWER LED (Green): Indicates power On/Off status
D Power button Detector power button
Name Description
Used for tightening the tether interface cable.
A Tether interface connector
Used for wired connection between a detector and SCU.
B Tether interface holder Fix/release holder of the tether interface cable.
C Battery 1 lock/unlock lever Applied for a FXRB-03A battery pack.
D Battery 2 lock/unlock lever Applied for a FXRB-03A battery pack.
E Antenna for wireless LAN Antennas for wireless communication (2EA)
Detector status indicator
DATA LED (Blue): Indicates communication and transmission
F Status indicator
ACTIVE LED (Orange): Indicates ready to work
POWER LED (Green): Indicates power On/Off status
G Power button Detector power button
H AP ON /OFF Button Turns on / off the AP mode.
I Battery Remaining LED Notices the remaining of battery in 5 levels.
Specifications
Item Specifications
Model FXRB-03A
Type Lithium Ion
Rated Power Supply Output: DC +7.6V
Capacity 3,100mAh
Number of Cell 2S1P (2 Series 1 Parallel)
Life Approx. 500 times (Fully charged/Discharged completely, 1 cycle)
Dimension (H × W × D) 160.0㎜ × 61.8㎜ × 5.7㎜
Weight 115g
The battery operation time increases under the sleep mode depending on the operational
condition and environment.
Drawing Sheet
2.3.1 Specifications
Item Specifications
Model FXRC-02A
Simultaneous Charging 2 battery packs
Charging time 2 1/2 hours
Rated power supply DC +24V, 2A Max.
Dimension (H × W × D) 192.0㎜ × 101.0㎜ × 26.0㎜
Weight 0.8 ㎏
2.4.1 Specifications
Item Specifications
Model FXRC-03A
Simultaneous Charging 2 battery packs (FXRB-01A (1ea) / FXRB-03A (1ea))
FXRB-01A: 3 hours
Charging time
FXRB-03A: 2 1/2 hours
Rated power supply DC +24V, 2A Max.
Dimension (H × W × D) 163.6 ㎜ × 190 ㎜ × 34.0 ㎜
Weight 0.5 ㎏
3. Performance
4. Defect
Type Description
Single Defect Isolated defects, adjacent pixels are normal. (Class 0)
Cluster Defect More than consecutive 2 pixels are defected. (Class 1~Class 8)
Defect occur horizontal direction from left ro right, or vertical direction from top
Line Defect
to bottom.
5. Regulatory Information
Item
Wired Model: Class Ⅰ
Type of protection against electrical shock
Wireless Model: Class Ⅰ or Internally Powered
Degree of protection against electrical shock Type B applied parts
IP53 (Degrees of protection against ingress of water and
Degree of protection against ingress of water
dust provided by enclosure.)
Operation mode Continuous operation
NOT suitable for use in the presence of a flammable
Flammable anesthetics anesthetic mixture with air or with oxygen or nitrous
oxide.
USA / Canada
Item
IEC 60601-1:2012 (ed. 3.1) Medical electrical equipment- Part1: General requirements for safety
Medical electrical equipment – Part 1: General requirements for
CAN/CSA-C22.2 No. 60601-1(2008)
safety (adopted amendment 2:1995 to IEC60601-1)
Medical electrical equipment-Part 1-2: Collateral standard:
IEC 60601-1-2: 2014 (ed.4)
Electromagnetic compatibility
IEC 62304:2006 Medical device software-software life cycle processes
ISO 14971:2012 Medical Device- Application of risk management to medical devices
European Union
Item
MDD (Medical Device Directive) 93/42/EEC as amended by 2007/47/EC
Medical devices – Quality Management systems – Requirements for
EN ISO 13485:2016
regulatory purposes
EN 60601-1: 2006(ed.3) Medical electrical equipment- Part1: General requirements for safety
IEC 60601-1: 2012(ed. 3.1) Medical electrical equipment- Part1: General requirements for safety
Medical electrical equipment-Part 1-2: Collateral Standard :
EN 60601-1-2: 2014(ed.4)
Electromagnetic compatibility-Requirements and tests
EN 62304:2006 Medical device software-Software life cycle processes
ISO 14971: 2012 Medical device – Application of risk management to medical devices.
Vieworks.Co., Ltd
(Gwanyang-dong) 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si,
Gyeonggi-do, 14055 Republic of Korea
Telephone: +82-70-7011-6161
Fax: +82-31-386-8631
Homepage: http://www.vieworks.com