GFG g450 Manual
GFG g450 Manual
GFG g450 Manual
Microtector II – G450
1 to 4-Gas Detector
Congratulations!
This manual will help you to operate the detector quickly and safely.
-2-
Content
INTRODUCTION 5
Design 6
OPERATIONAL HINTS 6
Switching On and Off 6
Additional Messages during Detector Start 7
Detection mode 8
Battery Capacity and Battery Alarm 8
Alarms 8
Reset of Alarms 9
STEL, TWA, Peak, Minimum values 9
Flip-Flop Display, Zoom Display 9
Peak – Display peak values 9
Turn On /Off Lights 10
Display Illumination 10
Storing Measurement Data with the Data Logger 10
Influence of Oxygen and Interfering Gases 10
Special Notes for LEL Monitoring 10
-3-
Charging of battery pack 20
ANNEX 21
Cleaning 21
-4-
Introduction
For your safety
According to § 3 of the law about technical working media and consumer products, device and production
safety law (GPSG), this manual points out the proper use of the product and serves to prevent dangers. It
must be read and adhered to by all persons who use, service, maintain and check this product.
This device can do the job designed to do only, if it is used, serviced, maintained and checked according to
the instructions given by GfG Gesellschaft für Gerätebau.
The warranties made by GfG with respect to the product are voided, if the product is not used, serviced,
maintained and checked in accordance with GfG’s instructions.
The above does not alter statements regarding warranties and liabilities in GfG’s general conditions of sale
and delivery. Repairs must only be done by skilled personnel resp. by trained persons. Modifications and
changes of the product require GfG’s permission. Unauthorized modification of the product result in the
exclusion of any liability for possible damage. Make sure that only genuine GfG accessories are used with the
product. Repairs require the use of spare parts released by GfG.
The temperature class of the detector depends on the supply module used. When using the „NiMH-II“
accumulator, temperature class T4 is valid for ambient temperatures of –20°C to +50°C, while temperature
class T3 is valid when using the „NiMH“ accumulator. Both supply modules are identified by a black
enclosure with an inside label showing the type and temperature class. When using the Alkaline batteries
(grey housing), temperature class T4 is valid for ambient temperatures from -20°C to +45°C resp.
temperature class T3 for ambient temperatures of -20°C to +55°C.
For the use in explosion endangered areas with a measurement function for the explosion protection there is
a supplement for the G450 to the above mentioned EC-Type Examination Certificate of DEKRA EXAM GmbH
according to guideline 94/9/EG. Basis of the test were the standards DIN EN 60079-29-1 „Gas detection
instruments – requirements to the operational behavior of instruments for the measurement of combustible
gases” and DIN EN 50271 “Electronic instruments for the detection and measurement of combustible gases,
toxic gases or oxygen – requirements and testing for warning instruments, that use software and/or digital
technology”.
Furthermore the G450 was examined on its measurement ability by DEKRA EXAM GmbH on the basis of the
standards DIN EN 50104 “Electronic instruments for the detection and measurement of oxygen –
requirements to the operational behavior and testing method” and DIN EN 45544-1/-2 “Electronic
instruments for the direct detection and direct measurement of the concentration of toxic gases and vapours
part 1: common requirements and testing methods” and part 2: requirements to the operational behavior of
instruments for the measurement of concentration in threshold ranges”. This is approved by the relevant
Type Examination Certificate with the number PFG 09 G 001.
The tests of the measuring function contain followed sensors and detection ranges:
EC-Type Examination Certificate MK221-0, MK221-1 for 0..100% LEL CH4, C3H8, C6H14 (CC)
BVS 06 ATEX E 017 X (4. supplement)
Type Examination Certificate MK369-0 for 5..500 ppm CO (EC)
PFG 09 G 001 MK427-0 for 0..25 %-Vol. O2 (EC)
MK429-0 for 0,2..100 ppm H2S (EC)
The functions being marked with [#] were not subject of the test of the measurement function.
-5-
Special conditions for safe use
In explosion endangered areas the G450 must be used properly. This means that the detector must be
carried at your body and must not be laid down unattended, to prevent an electrostatic charge of the clip. In
case readings in gas-free environments show a permanent zero-point deviation, a zero-point adjustment is
necessary. Especially after a heavy impact stress the zero-points of the sensors have to be checked and
optionally re-adjusted. In case the CC sensor shows “over-range” after a stress impact, the alarm has to be
reset in fresh air and the zero-point has also to be re-adjusted. If the G450 is operated continuously for
more than one day, the instrument should be turned off and on again every 24 hours latest. Within the
system option menu the deactivation of the latching alarm is not allowed for the use as a function tested
measurement instrument.
Design
Carrying strap
Display
Keys
Alarm-LEDs
Battery pack
Contacts
(back)
for periphery
Operational Hints
Switching On and Off
-6-
After switching on the G450 starts a self-test and displays information about the firmware version, the built-
in sensors with detection ranges and alarm thresholds and the date of the next inspection. During the self
test the visual and audible alarms are triggered like gas alarms.
Alarm thresholds and calibration data are displayed for all sensors connected. Only as an example it is only
CO which is being described here. Depending on the status of the sensors, the instrument may provide
additional messages, which may have to be confirmed. Please refer to “Additional messages during detector
start” for further information.
Once the self test is completed, the instrument is ready to use after about one minute. By hitting the middle
key readings and messages can be reset.
When a docking station is used for instrument check, the G450 may include intervals for bump test and
calibration of sensors. The dates for the next bump test or for the next calibration are calculated
automatically on the basis of the last check. Depending on what becomes necessary next, the date for the
next bump test or for the next calibration will be indicated, when the detector is started. Should the
relevant date be exceeded, the G450 indicates this as „overdue“. This message must be acknowledged by
key.
-7-
Detection mode
The G450 is ready for operation, if all measurement values, the unit, the
gas, the battery capacity and the time are displayed.
The detector checks whether the preset thresholds for the individual gases
are exceeded or deviated (O2).
When more than two measurement values are displayed simultaneously, either the gas type or the unit is
shown. By hitting the right key (ZOOM) measurement values can be displayed individually with gas type and
unit.
Alarms
Should the measured gas concentration exceed a pre-set threshold, the detector immediately gives an
audible and visual alarm. The display indicates which gas has caused the alarm. An extremely loud acoustic
alarm (103 dB(A) at 30 cm) and bright flashing alarm LEDs provide reliable warning for dangerous gas
concentrations. In case of a gas alarm the colour of the whole display turns into orange or red depending on
the alarm status. The G460 provides up to three alarm modes. The LO-alarm AL1 can be reset, while the HI-
alarms AL2 and AL3 are latching (default). There are three alarm levels for oxygen and combustible gases
(e.g. CH4), and two thresholds for toxic gases (CO, H2S). For toxic gases the G460 provides additional
alarms for exceeding of short term exposure level (STEL) and time weighted average (TWA). For further
information see “Alarm Thresholds – Standard Setpoints” and “Alarms – Adjusting the Alarm Thresholds”.
The alarm can also be triggered in combination with a vibration alarm, if the instrument provides a relevant
“battery pack with integrated vibrator”.
Number
Kind of alarm Sensors Description
of Alarms
oxygen 3 An instantaneous alarm is activated immediately, if the
Instantaneous combust. gases 3 gas concentration exceeds resp. falls below a pre-set
value (AL) toxic gases 2 threshold. The alarm thresholds are adjustable.
The short term exposure level (STEL) is the average
Short term concentration over a period of 15 minutes. The STEL alarm
exposure level toxic gases 1
is not latching. It resets automatically as soon as the
(STEL)
concentration has fallen below the threshold.
The time weighted average (TWA) refers to an 8 hours
Time weighted shift and calculates the average concentration. The TWA
average toxic gases 1
alarm cannot be reset. It is only de-activated, if the
(TWA)
detector is switched off.
The alarms are prioritized as follows: Power fault, overrange, AL3, TWA > AL2, STEL > AL1, underrange >
temperature fault.
-8-
Reset of Alarms
The latching (default) alarms 2 and 3 can be reset by pressing the RESET key, if the gas concentration has
fallen below or exceeded (O2) the pre-set thresholds. Alarm 1 is not latching and resets automatically, when
the alarm condition does not exist any longer. If the detection range of the CC sensor (e.g. CH4) is
exceeded, the display additionally reads „OVER RANGE“ instead of the value, for gas concentrations above
110 % LEL. In this case the sensor is deactivated to avoid damage. The alarms and the message “OVER
RANGE” remain. This alarm can only be reset by pushing the RESET key. Then the display asks:
Only if you made sure that the sensor is not exposed to combustible
gas but to fresh air only, you may answer this question with YES .
In this case the sensor turns on again and indicates gas concentrations
after a short warm-up time!
For further details please refer to „Special Notes for LEL Monitoring“.
During peak mode (activation by left key PEAK) peak values can be
monitored and displayed. The display shows an animated symbol in the left
bottom corner. Within zoom display the peak value will be displayed in the
top right corner instead of the actual gas concentration.
Pressing RESET during peak mode, the peak memory will be reset to the current gas concentration.
Pressing RESET during zoom display, the peak memory and the peak value memory will be reset to the
current gas concentration. By pressing PEAK again, the peak mode is deactivated.
-9-
Turn On /Off Lights
The G450 is optionally available with a rechargeable battery pack with lights. The lights can be switched on
by keeping the left key pressed for approx. 3 seconds, and turned off by pressing this key shortly. The lights
are useful e.g. when the device is linked to a cord and let down into a sewer system. Using the lights can
prevent the device from being dipped into water.
Display Illumination
The display illumination is turned on for approx. 10 seconds whenever you hit any key. It turns off
automatically after that time. Should the battery or accumulator be almost exhausted, the display
illumination cannot be activated any longer.
Service Mode
Press the middle key (RESET) for approx. 5 seconds to activate the service mode. In the service mode the
G450 can be adjusted by changing of program parameters. Certain menu points require the access code
„0011“ to prevent accidental change of important functions by unauthorized persons. During the service
mode all alarms are deactivated. The main menu is the first menu point in the service mode.
-10-
Main Menu
The menu points of the main menu are:
Menu control: The functions of the keys are explained in bold letters in the bottom line of the display.
After confirming the running number by pressing the left key “EDIT“, the location entry will follow:
-11-
Parameter Full shows the occupancy of the data logger.
= Scroll down to next parameter
ERASE = Deletes data. A security check is effected
“Delete data?” Confirm with the right key YES,
resp. deny with the left key NO.
EXIT = Back to main menu
Interval:
The data recording interval can be selected by and between
1 second and 60 minutes.
The recorded data can be read and transmitted to a PC by means of the drop-in charger or the smart
charger cap and an optional USB adapter cable.
AutoCal – AutoCal-Adjustment
Within the menu point AutoCal, several sensors can be calibrated simultaneously with fresh air (ZERO) or
test gas (CAL). Normally all sensors except of the CO2 sensor can be set with fresh air without any further
adjustment. For adjustments with test gas the sensors have to be activated according to the test gas /
mixture being used (s. chapter “AutoCal Air …” and “AutoCal Gas …”. The menu point AutoCal can be
selected within the main menu, but s also activated automatically, when the “Smart Cap” or the “Smart
Charger Cap” is mounted.
The following functions can be chosen:
ZERO = AutoCal with fresh air
CAL = AutoCal with test gas
EXIT = Back to main menu
-12-
AutoCal adjustment with test gas mixture not successful (e.g. due to wrong test gas concentration)
An AutoCal adjustment with fresh air is only successful, if the measured value does not differ by more than
±10% full scale from the nominal value 0.0 resp. not more than ±5.2 %-Vol. O2 from the nominal value
20.9 %-Vol. O2. A successful AutoCal adjustment with test gas is only completed, if the measured value
does not differ by more than 25% from the nominal “CalGas” value (see sensor menu “Calibration”). In case
of higher deviations the related sensor is marked with “Fault” in the subsequent AutoCal-report. In this case
an adjustment with “ZERO” resp. “CAL” or in the docking station is necessary. The adjustment with gas-free
fresh air can be done in diffusion mode. Zero gas (gas-free air) and test gas can be supplied with a volume
flow of 0.5 to 0.6 l/min by means of the “Smart Cap” or the “Smart Charger Cap”.
= scroll down
CHANGE = Change selected parameter
EXIT = Back to main menu
Service menu
Enter the service menu by selecting ”Service”. Within the service menu the G450 can be adjusted by
changing program parameters.
The menu points are only accessible with the code „0011“. The code prevents important functions being
changed by mistake or by unauthorised persons. In service mode no alarms can be released.
-13-
After entering code “0011” the display reads:
From here you reach the system menu (see section „system menu“), to
perform general adjustments. Within the menu point System the
individual sensors can be zeroed or calibrated. Information can be called
up or alarm thresholds can be adjusted.
Select Sensors for adjustment of sensor specific functions. With DETECT
you leave the service menu and return to detection mode
Normally zero gas is 0.0, so that this concentration does not need to be changed. In special cases, however,
you may push the key GAS to slightly increase the zero gas concentration; please refer to the following
picture. After entering GAS the display shows
-14-
ABORT = Aborting the adjustment and switching to the CH4-menu
A constant measurement value being captured after a stabilization time of 10 seconds the adjustment will be
executed and acknowledged with “OK”. Using CC- and O2-sensors the stabilization time is longer but
restricted to a maximum of 3 minutes.
(*1): The zero point adjustment of the oxygen sensor has factory-made settings of 100%Vol nitrogen. For
the detection of usual alarm thresholds of ≥17%Vol O2 a readjustment done by the operator is not
necessary. In this case an adjustment of the sensitivity is sufficient.
For adjusting the sensitivity the sensor menu point „Calibration“ has to be selected.
After entering ”GAS“ you can define the test gas concentration in a range of 10 to 105% of the
measurement end value:
A constant measurement value being captured after a stabilization time of 25 seconds the adjustment will be
executed and acknowledged with “OK”. The stabilization time is restricted to a maximum of 3 minutes.
(*2): The sensitivity adjustment of sensors that measure certain combustible gases within LEL-range, e.g.
n-Hexane, n-Nonane or other si,milar “heavy” vapours, is not uncomplicated. Apart from the availability of
such a test gas it is necessary to notice that the gas supply accompanies with a long stabilization time of
-15-
several minutes. Alternatively the sensitivity adjustment can be performed with a comparable gas (e.g.
Propane). The CC-sensor MK221-0 can be e.g. with a comparable gas of 0,85%Vol C3H8 (Propane) adjusted
to 65%LEL n-Hexane. The cross sensitivities for those sensors are described in chapter “Sensor
specifications”.
= Scroll down
SELECT = Select menu point
EXIT = Back to sensor menu
After selecting the alarm thresholds (e.g.: Alarm 1) the value can be entered:
The selected alarm threshold is flashing, the value can be changed now:
Except of %LEL detection ranges all thresholds can be freely adjusted within the entire detection range or
complete deactivated (0 resp. “----“). For %LEL detection ranges thresholds are limited to a maximum of
60%LEL.
Calibration data - Date & status of the last calibration and zeroing
Within the sensor menu point „CalDates“ the date of the last sensitivity
calibration and if calibration was successful (√)
or not () can be displayed.
In this menu point you can set the unit for CH4 to %LEL or %Vol.
The volume concentrations in brackets correspond to full scale deflection.
This allows to set the detection range to the country-specific LEL value.
When unit and gas type being changed the instrument has to be restarted after exiting the service menu.
This has to be done before running a bump test or AutoCal adjustment with a docking station.
-16-
System menu – General settings
Selecting „System“ in the system menu, following reading is displayed:
- Bump test (status, date of last and next bump test,
interval)
- Calibration (status, date of last and next calibration,
interval)
- Inspection (date of next inspection)
- Time (date + time)
- System options (selection of menu language, vibration
alarm on/off, latching alarm on/off,
autostore on/off)
- Sensor selection (activation resp. de-activation of individual
sensors)
- AutoCal – air (Adjustment with fresh air)
- AutoCal – gas (Adjustment with test gas)
- Information (info about detector type, software version,
serial number and battery type)
-17-
Inspection - Date of next inspection
To remind you of the date for the next maintenance resp. inspection, you can enter a date. When it expires,
the G450 automatically triggers an alarm. When the entered date is expired, the G450 reports a reminder
every time it is switched on.
Within the service menu, “Inspection“ has to be selected.
Here the parameter to be changed can be selected (day, month and year).
EXIT = Back to system menu
SELECT = Selects the blinking parameter
>> = moves to next parameter
= Scroll down
CHANGE = Change language resp. vibration alarm
BACK = Back to service menu
All options can be changed. „Language“ allows to chose German, English (UK), English (US) and French.
Under „Vibrator“ (only if battery pack with vibrator is available) you can turn the vibration alarm on or off.
„AL-Latching“ determines whether gas alarms 2 and 3 can only be reset by pressing the RESET key or
whether these alarms reset automatically when the gas concentration has fallen below the thresholds. The
deactivation of self-latching alarms is not allowed for the use of a function tested instrument „Autostore“
selects whether leaving the service mode saves all changes automatically or whether saving the changes
must be confirmed by keystroke.
On = Sensor active
Off = Sensor inactive
If the indicator is in brackets, this means that the sensor is not available;
it is indicated, however, how an additional sensor would react.
= Scroll down to next sensor
On/Off = Activation / Deactivation of sensor
EXIT = Back to service menu
Adjustment of which sensors are to be adjusted with fresh air. Generally all
sensors will be adjusted and show “ON“.
= Scroll down to next sensor
On/Off = Calibration / non-calibration of sensor in program
EXIT = Back to service menu
-19-
Charging of battery pack
Caution: The detector must not be charged in hazardous locations.
The detector must not be charged when turned on (detection mode).
The charge contacts must be kept clean. (s. chapter “Cleaning”)
-20-
4., 5., 6.: The instrument being used occasionally the battery pack should be charged after each use, due to
parts of the sensor electronics have to be provided with energy also in power off mode. In case of instrument not
being used for a long time and battery pack completely discharged the instrument has to be charged about 2
days before the next use. Possibly the battery pack will be charged for a short time (e.g. 11min) in the quick
charge mode switching to the trickle charge mode afterwards. A normally discharged battery pack will be
charged in about 4 to 4,5 hours during quick charge mode to 90% of its normal capacity. After further 8 hours
trickle charge the battery pack reaches 100% of its normal capacity. In case of not reaching the normal
operational time with a fully charged battery pack a reason for that might be the “Lazy-Battery-Effect”. That
effect has an impact on the discharging behavior that despite of fully charged battery pack the battery symbol
quite early shows an empty status in which the instrument can be operated for a long time.
Annex
Cleaning
Polluted enclosures can be cleaned with a damp cloth. Do not use solvents and detergents! It is
important to notice that outer charge contacts of the G460 and charge contact pins of the
charging adapter will be kept clean. In case of bad contacts of the charging adapter the NiMH
battery pack will be charged incompletely or not at all.
-21-
Maintenance and inspection
Maintenance and inspection include a regular check and adjustment of sensitivity and zero point. A bump
test or adjustment of the device is necessary as well. Gas monitoring devices can react differently on
environmental conditions. It is important, independent from maintenance duties, to test the device before
putting into operation (s. DIN EN 60079-29-2 chapter 9.2 as well as in Germany see BG-Chemistry,
Guideline T 021 and T 023). This test comprises following checks:
Visual control regarding physical damages
Visual control regarding gas supply entry holes
Charging status of battery / rechargeable battery pack
Display with zero gas and with test gas as well as alarms being triggered
The response behaviour of oxygen sensors can be checked with appropriate test gas (<18 %Vol O2) in
combination with the docking station, calibration cap “Smart Cap” or the charger cap “Smart Charger Cap”.
The simplest way of checking the response behaviour is by exhaled air.
Service - Repair
The DIN EN 60079-29-2 “… Gasmessgeräte – Auswahl, Installation, Einsatz und Wartung von Geräten für
die Messung von brennbaren Gasen und Sauerstoff“, the DIN EN 45544-4 „... Elektrische Geräte für die
direkte Detektion und direkte Konzentrationsmessung toxischer Gase und Dämpfe, Teil 4: Leitfaden für
Auswahl, Installation, Einsatz und Instandhaltung“ as well as relevant national guidelines have to be noticed.
Service comprises in Germany the "Explosionsschutz-Richtlinien", the „BGR 500, chapter 2.33" (formerly:
UVV Gase), the maintenance, inspection and repair of gas monitoring devices. Guidelines T 021 and T 023 of
the BG Chemistry describe the proper measures. The bump test has to be executed before first operation
and at least once a year and comprises:
Status of the zero point
Charging status of the battery
Pump and diffusion inlet
Display with zero gas and standard test gas and adjustment, if necessary
Alarm signal release, e.g. with alarm test gas
Constantly amplified signal with standard test gas
Response time
The check must be done by an expert, and the result must be confirmed in writing. Any repair of the G450
must generally be done according to the manufacturer’s instructions and with genuine spare parts.
Calibration Accessories
For controlling of the response sensitivity the instrument has to be provides with test gas. By using the
“Smart Cap” or the “Smart Charger Cap” the diffusion inlets can be covered, with the result that sensors will
be provided with test gas with a flow of 0,5 – 0,6 l/min. Alternatively the check for relevant test gases can
take place by using the docking station DS400.
Attention:
Test gases, especially toxic gases, can be hazardous. It is necessary not to inhale test gases.
Work places where calibrations with test gas are performed should be ventilated sufficiently
depending on the gas type, concentration and amount. In special cases an exhaust resp. a gas
drain is functional. The safety lecture written on the gas bottles as well as the safety data sheets
of the test gases have to be noticed.
-22-
Test with Docking Station DS400
The bump test required by T021 and T023 as well as the adjustment of the
Microtector II can be done easily and quickly by means of the docking station
DS400.
The bump test is started and effected automatically. The effective time
amounts to approx. 20 seconds. Adjustment is started by just pushing one
button, and is completed within a few minutes. The test result is indicated by
means of a green resp. red LED. Detailed values are shown in the display of
the detector (bump test report, AutoCal-Air report, AutoCal-Gas report). You
do not need a PC; all relevant data are automatically stored on a SD card in
the docking station.
The first bump test of a Microtector II G450 in the docking station can activate
the interval for bump test and adjustment automatically. For additional
information about the functions of the docking station please refer to the
operation manual for the docking station DS400. Before using the docking
station DS400 the operation manual has to be read and followed.
8. “Time set back to …” Clock not set or buffer battery empty Quit message, set clock
or call in GfG-Service
9. “Sensor defect!” Sensor defect or not available Switch device off and on
or call in GfG-Service
10. “Data faulty!” Sensor data faulty Switch device off and on
or call in GfG-Service
11. “No sensors!” No sensor activated in service program Activate available sensor
within the service menu
12. Gas indication “START” Sensor still in start up phase Wait some seconds
(“STRT”)
13. Gas indication “????” Measurement wit CC sensor not If occurring with fresh
possible due to oxygen indication air the oxygen sensor
<10%Vol has to be adjusted or
replaced
14. Gas indication “----“ / “ERROR” No gas indication due to faulty sensor Deactivate sensor within
or sensor data the service program or
call in GfG-Service
15. Gas indication “UNDER” or Massive falling down below detection Perform zero point
“UNDER RANGE” range adjustment
-23-
16. Gas indication “OVER” or Gas concentration too high or high Leave the range of high
“OVER RANGE” cross sensitivity (EC sensors) or gas concentrations and
protect switch activated (CC sensors) acknowledge message
quitting CC-sensor and
switch to fresh air range
17. Gas indication “TEMP” or Sensor operated out of specific Find area of normal
“TEMP ERROR” temperature range or hardware defect temperature or call in
at 0°C<Ta<30°C GfG-Service
18. Gas indication “POWER” or Power supply of sensor is corrupted Call in GfG-Service if
“POWER ERROR” repeated
19. Gas indication “P+T” s. gas indication “TEMP” and “POWER” see above
20. “Remove charger!” Alkaline batteries not rechargeable Demount charger from
the instrument
21. “Remove battery pack!” Instrument can not be shut down due Demount battery pack
to hardware defect from the instrument or
call in GfG-Service
22. “No sensors for AutoCal-Air No sensors for automatic fresh air resp. Enable sensors for the
(Gas) enabled!” test gas adjustment enabled automatic adjustment
within the service menu
23. “Zero failure – measurement Measurement gas possible available Perform zero point
value too high!” (too low) but zero point drift too positive adjustment within
(negative) measurement gas free
environment or call in
GfG-Service
24. “Calibration failure – Wrong test gas concentration or sensor Check test gas and
measurement value too low!” sensitivity too low (high) reference value or call in
(too high) GfG-Service
25. “Zero (calibration) failure – Extreme sensor signal deviation or Repeat process or call in
signal not discoverable!” hardware defect GfG-Service
26. “Storing failure!” Parameter can not be saved when Switch instrument on /
leaving service program off, then repeat settings
within the service menu
or call in GfG-Service
-24-
Accessories and spare parts
Description Part No.
1. Alkaline battery pack (without batteries) 1450200
2. Alkaline battery pack with vibrator (without batteries) 1450202
3. Alkaline battery (pack of 10) 1450204
4. Rechargeable NiMH battery pack 1450206
5. Rechargeable NiMH battery pack with vibrator 1450207
6. Rechargeable NiMH battery pack with lights 1450208
7. Rechargeable NiMH battery pack with vibrator and lights 1450209
8. Smart Charger Cap (charge, calibrate, data transfer) 1450215
9. Plug-in charger 100-240VAC 1450216
10. Charging cable for cars 1450218
11. Drop-in charger G400-DIC1 / Drop-in charger G400-DIC2 [#] 1450219 / 23
12. Drop-in charger G400-DIC1S / Drop-in charger G400-DIC2S (with strap) [#] 1450220 / 24
13. Calibration cap “Smart Cap” (calibration) 1450225
14. Transportation and storing case (plastic) [#] 1450229
15. USB Interface cable for PC 1450232
16. Data logger set 1 with GfG-Interface software for Microtector II 1450233
17. Docking station DS400 with DIC1D / with DIC2D [#] 1450401 /02
18. MK221-0 Sensor for 100% LEL combustible gases and vapours 1450703
19. MK221-1 Sensor for 100% LEL combustible gases (with increased intoxication 1450704
resistance)
20. MK369-0 Carbon monoxide sensor CO 1450701
21. MK380-0 Dual sensor for carbon monoxide CO and hydrogen sulphide H2S [#] 1450706
22. MK380-0 Oxygen sensor O2, (2 years) 1450708
23. MK427-0 Oxygen sensor O2, (3 years) 1450707
24. MK429-0 Hydrogen sulfide H2S 1450705
The spare parts and the accessories should be stored at ambient temperatures of 0...30°C. Storage time
should not be longer than 5 years. Electrochemical sensors should not be stored for more than ½ year.
When you store oxygen sensors be aware of the fact that storage reduces the expected lifetime of the
sensor. When storing spare sensors, make sure that the ambient atmosphere is free from corrosive media
and sensor poisons. For NiMH battery packs a storing time of only one year is valid. Before storing the
battery pack has to be charged completely. In case of storing lasts more than ½ year the battery pack has
to be demounted.
-25-
Sensor type and detection range
Slot Sensor type Detection Gas Resolution T-Band *
(ID) range
EC1 MK 429-0 0 .. 100 (200) ppm H2S Hydrogen sulfide 0.2 ppm ±1.0 ppm
at (*1): T-Band = Tolerance bandat (*2): or one of the following combustible gases and vapours
-26-
Sensor specification
MK221-0 Catalytic combustion sensor for combustible gases and vapours
Detection range: 0,0 .. 100 %LEL
Response time: t50: ≤10 s t90: <20 s for CH4 t50: ≤12 s t90: <30 s
for C3H8
t50: ≤25 s t90: <65 s for C6H14
Pressure (70)80...110 kPa: max. 5(7)% LEL or 10% of display (referred to 100 kPa)
Humidity 0%...95% r.h.: max. 3% LEL or 10% of C3H8 display (referred to 0% r.h. @ 40°C)
or 30% of CH4 display (referred to 0% r.h. @ 40°C)
Temperature-20(-10)...+40(55)°C: max. 5% LEL or 10(15)% of display (referred to 20°C)
Flow velocity 0 .. 6 m/s: max. 1% LEL or 15% of display @ flow velocity ≥ 1.5 m/s
Cross sensitivities [#] at 50%LEL: Gas supply CH4 display C3H8 display n-Hexane display
2,00%Vol H2 ca.65%LEL ca.100% LEL ca.135% LEL (theor.)
2,20%Vol CH4 = 50% LEL ca.75% LEL ca.100% LEL
0,85%Vol C3H8 ca.33% LEL = 50% LEL ca.65% LEL
0,70%Vol C5H12 ca.32% LEL ca.48% LEL ca.63% LEL
0,70%Vol C4H10 ca.31% LEL ca.47% LEL ca.62% LEL
0,50%Vol C6H14 ca.27% LEL ca.38% LEL = 50% LEL
0,55%Vol C7H16 ca.22% LEL ca.32% LEL ca.41% LEL
0,55%Vol C8H18 ca.17% LEL ca.25% LEL ca.32% LEL
May vary from sensor to sensor and depend on the gas concentration and on the age of the sensor.
-27-
MK427-0 Electrochemical sensor for oxygen O2
Detection range: 0 .. 25 %Vol
Response time: t20: ≤8 s t90: <25 s
Pressure (70)80...120(130) kPa: max. 0,4(0,6)%Vol or 2(3)% of the detection range (referred to 100kPa)
Humidity 0...95% r.h.: max. 0,5%Vol or 2,5% of the detection range (referred to 50 r.h. @40°C)
Temperature (-20)-10...+55°C: max. 0,5(0,8)%Vol or 2,5(4,0)% of the detection range (referred to 20°C)
Expected lifetime: 3 years in air
MK429-0 Electrochemical sensor for hydrogen sulfide H2S
Detection range: 0,2 .. 100 ppm (200 ppm [#]) zero point devation < 0,4 ppm
Response time: T50: ≤15 s t90: <30 s
Pressure 70...130 kPa: max. 0,2%Vol or 5% of the display (referred to 100kPa)
Humidity 5%...95% r.h.: max. 0,2%Vol or 2% of the display (referred to 50 r.h. @20°C)
Temperature -20...+40(55)°C: max. 0,2%Vol or 5(16)% of the display (referred to 20°C)
Long term stability per month: max. 0,2%Vol or 2% of the display (referred to laboratory conditions)
Cross sensitivities SO220%; NO2-20%; CO<1%; NO<0,2%; H2<0,1%; (*1)
Expected lifetime: 3 years
(*1) Displayed value with reference to the supplied gas concentration
zu (*1): oder ein anderes der nachfolgend aufgeführten brennbaren Gase und Dämpfe
-28-
Technical Data
Type: G450
Detection principle: Electrochemical (EC): for toxic gases and oxygen
Catalytic combustion (CC): for combustible gases and vapours
(up to 100 %LEL)
Detection range: See section „Sensor type and Detection range“
Response time: See section „Sensor specification“
Expected sensor lifetime: 2...3 years - see section „Sensor specification“
Gas supply: Diffusion with flow velocity of 0 .. 6 m/s or
Pump by means of attachable electrical sampling pump G400-MP1[#]
Display: Illuminated full-graphic LCD, automatical size adjustment for optimal read
out, display of battery capacity, gas concentration as instantaneous and
peak value
Alarm: Depending on gas type 3 or 2 instantaneous and 2 dosimeter alarms,
battery alarm.
visual and audible warning and display indication, colouring of display
depending on alarm status (orange/red)
Buzzer: 103 dB (reduceable to 90 dB)
Zero point and sensitivity Manually or automatically with calibration program by “Smart Cap” or
calibration: “Smart Charger Cap” test gas supply with 0.5..0.6 l/min.
Power supply: 1. NiMH battery module (colour: black), 2500Ah, rechargeable
Im=600mA (max. charging current)
Um=6V DC (max. voltage) or
2. Alkaline battery module (colour: grey), non-rechargeable
with 2x mignon 1.5V Type: DURACELL PROCELL MN1500 LR6 AA
Operational time(1)NiMH-II: approx. 30h (EC+WTCH4); approx. 17h (EC+WT); approx. 130h (EC)
Alkaline: approx. 25h (EC+WTCH4); approx. 14h (EC+WT); approx. 170h (EC)
Climate conditions:
for operation: -20...+55°C | 5...95% r. h. | 700...1300hPa
for storage: -25...+60°C | 5...95% r. h. | 700...1300hPa (recommended 0...+30°C)
Casing:
Material: Rubberized plastic
Dimensions: 75 x 110 x 55 mm (WxHxD)
Weight: 290 g
Protection: IP 67
Approvals and tests:
Labelling and ignition II2G Ex ia d IIC T4 -20°C≤Ta≤+55°C for NiMH-II (black)
protection: Ex ia d IIC T3 -20°C≤Ta≤+55°C for NiMH (black)
Ex ia d IIC T4/T3 -20°C≤Ta≤+45°C/+55°C for Alkaline (grey)
EC-Type Examination BVS 06 ATEX E 017 X (for measuring function and electronic Ex-
Certificate: protection see chapter “application and purpose”)
Examination Certificate: PFG 09 G 001 (for measuring function see chapter “application and
purpose”)
EMC Test: DIN EN 50270 : 2006 Radio shielding: Type class I
Interference resistance: Type class II
(*1): The operational time will be decreased by hitting keys (display illumination and lights) and triggered gas alarms.
-30-
-31-
-32-
-33-
-34-
-35-