ISCO 6712 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 242
At a glance
Powered by AI
The document discusses installation and operation guidelines for portable samplers. It emphasizes reading the manual before use and contacting technical support in case of issues.

The document mentions confined spaces as potential hazards and outlines hazard severity levels to identify hazardous conditions.

The document provides a one year limited factory warranty for covered products against failure due to faulty parts or workmanship. It specifies what is and isn't covered under warranty.

Edited by Foxit Reader Copyright(C) by Foxit Software Company,2005-2008 For Evaluation Only.

6712 Portable Samplers


Installation and Operation Guide

Part #69-9003-588 of Assembly #60-9004-334 Copyright 2001. All rights reserved, Teledyne Isco, Inc. Revision X, August 11, 2008

Foreword
This instruction manual is designed to help you gain a thorough understanding of the operation of the equipment. Teledyne Isco recommends that you read this manual completely before placing the equipment in service. Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction. If the problem persists, call or e-mail the Teledyne Isco Technical Service Department for assistance. Simple difficulties can often be diagnosed over the phone. If it is necessary to return the equipment to the factory for service, please follow the shipping instructions provided by the Customer Service Department, including the use of the Return Authorization Number specified. Be sure to include a note describing the malfunction. This will aid in the prompt repair and return of the equipment. Teledyne Isco welcomes suggestions that would improve the information presented in this manual or enhance the operation of the equipment itself. Teledyne Isco is continually improving its products and reserves the right to change product specifications, replacement parts, schematics, and instructions without notice. Contact Information
Customer Service

Phone:

(800) 228-4373 (402) 464-0231

(USA, Canada, Mexico) (Outside North America)

Fax: Email:
Technical Service

(402) 465-3022 [email protected] (800) 775-2965 (800) 228-4373 (Analytical) (Samplers and Flow Meters)

Phone:

Email:

[email protected]

Return equipment to: 4700 Superior Street, Lincoln, NE 68504-1398


Other Correspondence

Mail to: Email: Web site:

P.O. Box 82531, Lincoln, NE 68501-2531 [email protected] www.isco.com

Revised September 15, 2005

6712 Portable Sampler Safety


6712 Portable Sampler Safety

General Warnings

Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed in the following general warnings:

WARNING
This instrument has not been certified for use in hazardous locations as defined by the National Electrical Code.

WARNING
Avoid hazardous practices! If you use this instrument in any way not specified in this manual, the protection provided by the instrument may be impaired; this will increase your risk of injury.

AVERTISSEMENT
viter les usages prilleux! Si vous utilisez cet instrument dune manire autre que celles qui sont specifies dans ce manuel, la protection fournie de linstrument peut tre affaiblie; cela augmentera votre risque de blessure.

This product is often installed in confined spaces. Some examples of confined spaces are manholes, pipelines, digesters, and storage tanks. These spaces may become hazardous environments that can prove fatal for those unprepared. These spaces are governed by OSHA 1910.146 and require a permit before entering. Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts, These three levels are described in the following sample alerts.

CAUTION
Cautions identify a potential hazard, which if not avoided, may result in minor or moderate injury. This category can also warn you of unsafe practices, or conditions that may cause property damage.

WARNING
Warnings identify a potentially hazardous condition, which if not avoided, could result in death or serious injury.

iii

6712 Portable Sampler Safety

DANGER
DANGER limited to the most extreme situations to identify an imminent hazard, which if not avoided, will result in death or serious injury.
Hazard Symbols The equipment and this manual use symbols used to warn of hazards. The symbols are explained below.

Hazard Symbols
Warnings and Cautions
The exclamation point within the triangle is a warning sign alerting you of important instructions in the instruments technical reference manual.

The lightning flash and arrowhead within the triangle is a warning sign alerting you of dangerous voltage inside the product.

Pinch point. These symbols warn you that your fingers or hands will be seriously injured if you place them between the moving parts of the mechanism near these symbols.

Symboles de scurit
Ce symbole signale lexistence dinstructions importantes relatives au produit dans ce manuel.

Ce symbole signale la prsence dun danger dlectocution.

Risque de pincement. Ces symboles vous avertit que les mains ou les doigts seront blesss srieusement si vous les mettez entre les lments en mouvement du mcanisme prs de ces symboles

Warnungen und Vorsichtshinweise


Das Ausrufezeichen in Dreieck ist ein Warnzeichen, das Sie darauf aufmerksam macht, da wichtige Anleitungen zu diesem Handbuch gehren. Der gepfeilte Blitz im Dreieck ist ein Warnzeichen, das Sei vor gefhrlichen Spannungen im Inneren des Produkts warnt.

Vorsicht Quetschgefahr! Dieses Symbol warnt vor einer unmittelbar drohenden Verletzungsgefahr fr Finger und Hnde, wenn diese zwischen die beweglichen Teile des gekennzeichneten Gertes geraten.

iv

6712 Portable Sampler


Table of Contents

Section 1 Introduction
1.1 1.2 1.3 1.4 1.5 About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . About 700 Series Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SDI-12 Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Memory to Store Monitoring Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pump Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1-2 1-2 1-2 1-3

Section 2 Installation/Preparation
2.1 Preparing the Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 2.2 Installing the Distributor Arm and Discharge Tube. . . . . . . . . . . . . . . . . . . . . . . . 2-1 2.3 Installing Bottle Kits for Multiplex Sampling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 2.3.1 Aligning Bottle Kits in the Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 2.3.2 Aligning Bottle Kits in the Compact Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2.4 Installing Bottle Kits for Composite Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2.4.1 Optional Jumbo Bottle Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2.5 Installing a Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9 2.5.1 Connecting an External 12-Volt DC Source . . . . . . . . . . . . . . . . . . . . . . . . . 2-9 2.6 Suction Line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 2.6.1 Attaching Suction Line to Pump Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 2.7 Strainers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 2.7.1 Alternative to Strainers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 2.8 How Does the Sampler Work? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12 2.8.1 A Typical Sampling Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12 2.8.2 Measuring Sample Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9 Installing the Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9.1 Icing the Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9.2 Tips for Routing Suction Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9.3 Intake Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9.4 Positioning a Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 2.9.5 Using the Suspension Harness or ProHanger . . . . . . . . . . . . . . . . . . . . . . 2-14 2.10 Connecting External Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14 2.10.1 Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14 2.10.2 Rain Gauge Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15 2.11 Locking the Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16 2.12 Servicing the Sampler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16 2.13 Recovering the Sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17 2.14 Collecting Sample Bottles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17 2.14.1 Compact Sampler Bottle Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17 2.14.2 Standard Sampler Bottle Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18

Section 3 Programming Introduction


3.1 Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1.1 Selecting a Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 Using Menus and Entering Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2.1 Selecting Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2.2 Entering Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 3-3 3-3 3-3 3-3

6712 Portable Sampler Table of Contents


3.3 Quick View Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.3.1 Paging Through Quick View Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.3.2 Changing Settings in a Quick View Screen . . . . . . . . . . . . . . . . . . . . . . . . . 3-4 3.4 Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 3.4.1 Setting the Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6 3.4.2 Menu Screens: Clock Start Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6 3.5 Menu Screens: Site Descriptions and Program Names . . . . . . . . . . . . . . . . . . . . . 3-7 3.6 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.6.1 Information Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.6.2 Help Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9 3.6.3 Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10 3.6.4 Operation Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11 3.7 Menu Trees . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

Section 4 Standard Programming


Switching Between Standard and Extended Modes . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Language Selection, Units of Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2 Pacing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 4.4.1 Trigger Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 4.5 Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.5.1 Sequential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.5.2 Bottles Per Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.5.3 Samples Per Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.5.4 Composite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.5.5 Continuous Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 4.6 Start Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 4.6.1 How Do Start Times Work? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 4.6.2 Sampler Enable/Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9 4.7 Running Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10 4.7.1 Run Time Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10 4.7.2 Module Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11 4.8 Interrupting a Running Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12 4.8.1 Stop Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12 4.8.2 Resume Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.3 View Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.4 Grab Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.5 Pump Tube Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.6 Calibrate Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.7 Cal/Adj Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.8 Adjust Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13 4.8.9 Adjust Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 4.8.10 Power Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 4.9 Other Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 4.10 Manual Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14 4.11 Grab Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15 4.12 Calibrate Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16 4.12.1 Calibration Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16 4.13 Operating the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17 4.14 Moving the Distributor Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18 4.15 Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19 4.15.1 Collecting Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19 4.15.2 Viewing the Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19 4.15.3 Configuring Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20 4.15.4 Sampling Results Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21 4.15.5 Combined Results Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21 4.15.6 Summary Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21 4.1 4.2 4.3 4.4

vi

6712 Portable Sampler Table of Contents


4.16 System IDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29 4.17 Programming for 700 Series Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29

Section 5 Extended Programming


5.1 Extended and Standard Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 5.2 One-Part and Two-Part Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 5.3 Storage for Extended Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 5.3.1 Selecting a Stored Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 5.4 Programming for 700 series Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.5 Programming SDI-12 Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.6 Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4 5.6.1 Storm Water Runoff Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 5.7 Suction Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12 5.8 Rinses and Retries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12 5.9 Pacing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12 5.9.1 Event Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12 5.9.2 Event Paced Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13 5.9.3 Nonuniform Time Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15 5.9.4 Nonuniform Clock Time Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15 5.9.5 Nonuniform Interval Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15 5.9.6 Random Interval Pacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15 5.10 Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.1 Sequential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.2 Bottles Per Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.3 Samples Per Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.4 Composite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.5 Multiple Bottle Compositing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.6 Time Switching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16 5.10.7 Continuous Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17 5.11 Flow Proportional Sample Volumes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19 5.11.1 Calculating Variable Sample Volume Settings . . . . . . . . . . . . . . . . . . . . 5-20 5.11.2 Calculations when Using an External Flow Meter . . . . . . . . . . . . . . . . . 5-21 5.11.3 Calculations when Using a 700 Series Flow Module . . . . . . . . . . . . . . . . 5-22 5.12 Sampler Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5.12.1 Sampler Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24 5.13 Sampler Enable Responses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25 5.13.1 Once Enabled Stay Enabled . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25 5.13.2 Repeatable Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25 5.13.3 Sample at Enable or Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25 5.13.4 Delay To Start Of Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26 5.13.5 Dry Period Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26 5.13.6 Resetting the Sample Interval at Enable . . . . . . . . . . . . . . . . . . . . . . . . . 5-26 5.14 Pauses/Resumes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27 5.15 Running Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27 5.15.1 Run Time Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27 5.15.2 Interrupting a Running Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27 5.16 Sampling Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28 5.17 Other Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28 5.18 Software Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28 5.18.1 Liquid Detector Enable/Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28 5.18.2 Programming Style . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29 5.18.3 Measurement Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29 5.18.4 Dual Sampler Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29 5.18.5 Pre-sample and Post-sample Purge Counts . . . . . . . . . . . . . . . . . . . . . . . 5-31 5.18.6 Serial Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32 5.18.7 Program Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34 5.19 Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36

vii

6712 Portable Sampler Table of Contents


5.19.1 SDI-12 Sonde Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.2 Rain Gauge Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.3 I/O Pin Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.4 Dialout Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.5 Pager Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.6 Digital Modem Text Messaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.7 Refrigerator Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.19.8 Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.20 Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.21 Pressurized Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.22 Command Driven Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.22.1 Command Driven Sampler Responses . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36 5-36 5-37 5-39 5-40 5-40 5-41 5-41 5-42 5-44 5-44 5-45

Section 6 SDI-12 Sondes


6.1 6.2 6.3 6.4 6.5 6.6 SDI-12 Data Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Scanning For Sondes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Teledyne Isco Ready Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Other SDI-12 Sondes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sonde Setup - Storing Parameter Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sonde Calibration and Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.6.1 The Calibration Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.6.2 The Validation Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.7 Disconnecting Sondes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1 6-2 6-3 6-3 6-4 6-5 6-6 6-6 6-7

Section 7 Remote Operation


7.1 Computer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.1 Menu Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.2 External Program Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1.3 Remote Control of Sampler Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2 Telephone Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1 7-2 7-4 7-8 7-9

Section 8 Maintenance
8.1 Maintenance Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1 8.2 Cleaning Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1 8.2.1 Controller, Top Cover, Center Section, Retaining Ring, and Tub . . . . . . . . 8-1 8.2.2 Sample Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1 8.2.3 Suction Line, Pump Tube, and Discharge Tube . . . . . . . . . . . . . . . . . . . . . . 8-2 8.2.4 Strainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.3 Cleaning Protocols for Priority Pollutants. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.3.1 Teledyne Isco Glass Sample Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.3.2 Vinyl Suction Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.3.3 Teflon Suction Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.3.4 Teledyne Isco Pump Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2 8.4 Maintenance Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3 8.4.1 Set Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3 8.4.2 Pump Tube Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4 8.4.3 Internal Battery Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6 8.4.4 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6 8.5 Replacing the Pump Tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8 8.5.1 Checklist For Replacing Pump Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 8.6 Opening the Controller Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11 8.7 Replacing the Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11 8.8 Replacing the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13 8.9 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14 8.10 Pump Tube Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15

viii

6712 Portable Sampler Table of Contents

Appendix A Menu Flowcharts Appendix B Material Safety Data Sheets Appendix C General Safety Procedures
C.1 Practical Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 C.1.1 Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 C.1.2 Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2 C.1.3 Adverse Atmospheres . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2 C.1.4 Entering Manholes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2 C.1.5 Traffic Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3 C.1.6 Falling Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3 C.1.7 Removing the Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3 C.1.8 Other Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3 C.1.9 Emergencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4 C.1.10 Field Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4 C.2 Lethal Atmospheres in Sewers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4 C.3 Hazardous Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6

Appendix D Replacement Parts List


D.1 Replacement Parts Diagrams and Listings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1

Appendix E Accessories List


E.1 Order Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1 E.2 Controller, Center Sections, Top Covers, and Tubs . . . . . . . . . . . . . . . . . . . . . . . . E-1 E.3 Compact and Mini Sampler Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1 E.4 Compact Sampler Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2 E.5 Standard Sampler Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2 E.6 Bottle Carriers, Retaining Rings and Racks, Bottle Inserts, Discharge Tubes . . E-3 E.7 Bulk Sets of Bottles with Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3 E.8 General Accessories: Compact and Standard Samplers. . . . . . . . . . . . . . . . . . . . . E-4 E.9 Pump Tubes, Suction Line, Strainers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4 E.10 Data Collection Devices and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5 E.11 Power Sources and Chargers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5 E.12 Modules, Rain Gauges, & Interfacing Instruments . . . . . . . . . . . . . . . . . . . . . . . E-6 E.13 SDI-12 Data Acquisition Connect Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6

Lis t of Illus trations


2-1 Installing the Distributor Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 2-2 Placing the 24-Bottle Kit in a Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 2-3 Placing the 12-Bottle Kit in a Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 2-4 Placing the 8-Bottle Kit in a Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 2-5 Placing the 4-Bottle Kit in a Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2-6 Placing the carrier in the compact tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 2-7 Installing the Composite Tube Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 2-8 Composite Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8 2-9 Attaching the suction line to the pump tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 2-10 Identifying the Sampler Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12 3-1 Menu Tree for Standard Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12 4-1 Sample Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7 4-2 Start Time Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10

ix

6712 Portable Sampler Table of Contents


4-3 Running a Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11 4-4 Taking a Grab Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15 4-5 Report: Program Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24 4-6 Report: Sampling Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25 4-7 Report: Combined Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26 4-8 Report: Combined Results (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27 4-9 Report: Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28 5-1 Sample Distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18 5-2 Factors Affecting Flow-Weighted Variable-Volume Samples . . . . . . . . . . . . . . . 5-20 5-3 Variable-Volume Scenario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21 5-4 Start Time Diagram for Sampler Enable Settings . . . . . . . . . . . . . . . . . . . . . . . . 5-23 5-5 Event Mark Timing Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31 5-6 Password Protected Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-35 6-1 SDI-12 Sonde Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 7-1 Press and hold ?? to connect to the sampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2 7-2 Example list of remote menu commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2 8-1 Removing and Replacing the Pump Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10 8-2 Placing the Pump Tube in the Liquid Detector . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10 8-3 Opening the Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12 8-4 Main Circuit Board Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13 A-1 6712 Menu Tree for Standard Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 A-2 Standard Programming: Programming Screens . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 A-3 Standard and Extended Programming: Start Times . . . . . . . . . . . . . . . . . . . . . . . A-3 A-4 Standard Programming: Quick View; Programming Screens . . . . . . . . . . . . . . . . A-4 A-5 Standard Programming: Quick View; Start Times . . . . . . . . . . . . . . . . . . . . . . . . A-5 A-6 View Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6 A-7 Maintenance Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7 A-8 Manual Functions Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8 A-9 Extended Programming: Programming Screens . . . . . . . . . . . . . . . . . . . . . . . . . . A-9 A-10 Extended Programming: Equipment Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10 A-11 Extended Programming: Pacing and Distribution . . . . . . . . . . . . . . . . . . . . . . A-11 A-12 Extended Programming: Event Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12 A-13 Extended Programming: Programmed Sampler Enable . . . . . . . . . . . . . . . . . A-13 A-14 Extended Programming: Quick View; Programming Screens . . . . . . . . . . . . . A-14 A-15 Extended Programming: Quick View; Equipment Set-Up . . . . . . . . . . . . . . . . A-15 A-16 Extended Programming: Quick View; Pacing and Distribution . . . . . . . . . . . A-16 A-17 Extended Programming: Quick View; Programmed Sampler Enable . . . . . . . A-17 A-18 Extended Programming: Normal View; Software Options . . . . . . . . . . . . . . . . A-18 A-19 Extended Programming: Quick View; Software Options . . . . . . . . . . . . . . . . . A-19 A-20 Extended Programming: Normal View; Hardware . . . . . . . . . . . . . . . . . . . . . . A-20 A-21 Extended Programming: Quick View; Hardware . . . . . . . . . . . . . . . . . . . . . . . A-21 A-22 Analog and Dialout Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-22 A-23 Analog Output and Dialout Quick View Screens . . . . . . . . . . . . . . . . . . . . . . . A-22 A-24 SDI-12 Sonde Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-23

Lis t of Tables
1-1 6712 Standard and Compact Sampler Features . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 1-2 6712 Standard and Compact Sampler Construction Materials . . . . . . . . . . . . . . . 1-7 1-3 Technical Specifications for the 6712C and Standard Samplers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8 1-4 Connector Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16 2-1 Distributor Arm Adjustments and Discharge Tube Lengths for Bottle Kits . . . . 2-2 2-2 Discharge Tubes for Composite Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8 2-3 Selecting the Right Strainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 2-4 Connecting Teledyne Isco Instruments to the Sampler . . . . . . . . . . . . . . . . . . . . 2-16 3-1 About the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

6712 Portable Sampler Table of Contents


4-1 4-2 4-3 4-4 4-5 5-1 5-2 5-3 5-4 5-5 7-1 7-2 7-3 7-4 8-1 8-2 8-3 C-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle . . . . . . . 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample . . . . . . . . . . 4-3 Reports: Program Events, Source Codes, and Error Codes . . . . . . . . . . . . . . . . . 4-22 Programming Example With 730 Module Installed . . . . . . . . . . . . . . . . . . . . . . . 4-30 Programming Example With 750 Module Installed . . . . . . . . . . . . . . . . . . . . . . . 4-33 Selecting a Stored Extended Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Storm Water Runoff Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6 Event Paced Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13 Sampler Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24 Serial Data Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33 6712 Remote Menu Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 Serial Data Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6 Remote Control of Sampler Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8 6712 Remote Phone Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9 Setting the Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4 Resetting the Pump Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5 Checking and Replacing the Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6 Hazardous Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6

xi

6712 Portable Sampler Table of Contents

xii

6712 Portable Sampler


S ec tion 1 Introduc tion

The 6712 Series Samplers are an integral part of Teledyne Iscos comprehensive sampling system. The system includes the 6712 Compact Sampler, the 6712 Standard Sampler and the 6712FR and 6712SR Refrigerated Samplers. The 6712 portable samplers are easily moved from site to site. Constructed from durable, corrosion-resistant materials, the samplers withstand the hostile environments of industrial and municipal monitoring sites. The 6712 Compact differs from the 6712 Standard Sampler only in size and sampling capacity. Both samplers use the same controller and sample-delivery system. Both instruments accept a variety of composite and multiple bottle kits. Because the 6712 Compact and Standard Samplers share many features, this manual includes programming, operation, and maintenance information for both instruments. Except where specifically noted, the contents of this manual apply to both units. This manual contains the information you will need to install, program, operate, and maintain a 6712 portable sampler. There are seven sections and five appendices: Section 1, Introduction lists the features and specifications of the 6712 portable samplers. Section 2, Installation/Preparation steps you through installing the sampler and preparing it for use. Section 3, Getting Started describes features that are common to both standard and extended programming modes. Section 4, Standard Programming shows how to program your sampler using standard programming mode. Section 5, Extended Programming shows how to program your sampler using extended programming mode. Section 6, SDI-12 Sondes provides information on setting up sondes to work with the 6712. Section 7, Remote Operation explains how the 6712 controller can be operated remotely from external devices. Section 8, Maintenance covers the basic maintenance required to keep your sampler in top operating condition.

6712 Controller

1.1 About this Manual

1-1

6712 Portable Sampler Section 1 Introduction The appendices are: Appendix A, Menu Flowcharts Appendix B, Material Safety Data Sheets Appendix C, General Safety Procedures Appendix D, Replacement Parts Appendix E, Accessories List

1.2 About 700 Series Modules

The bay on the controllers side accepts any of Teledyne Iscos 700 Series Modules. The 700 Series includes: 701 pH Parameter Module for monitoring pH and temperature. 710 Ultrasonic Module for monitoring a flow streams level and flow rate with an ultrasonic level sensor. 720 Submerged Probe Module for monitoring a flow streams level and flow rate with a submerged probe. 730 Bubbler Flow Module for monitoring a flow streams level and flow rate with a bubbler system. 750 Area Velocity Module for monitoring a flow streams level, velocity, and flow rate. 780 4-20mA Input Module for interfacing to non-Teledyne Isco devices with 4 to 20 milliampere output signals. The modules are optional accessories. The samplers do not require them for operation. However, the modules offer a number of advantages: they are an economical way to combine flow-rate or parameter monitoring with sampling, and you can program the samplers and modules as a single unit. Finally, the samplers store the readings in memory.

1.3 SDI-12 Sondes

As an option, the sampler accepts up to 16 parameters from up to ten sensors with SDI-12 addresses from 0 - 9 (refer to SDI-12 Sonde Readings, page 1-11, for specific parameters). The data parameters used by the sensing device (sonde) and recording device (sampler) must match. Certain vendors sondes have enhanced commands that facilitate plug and play setup. These Teledyne Isco Ready sondes can tell the 6712 what values they have, their order, and units of measure. Other sondes can be used, but require manual setup to identify proper data types for each data value reported.

1.4 Memory to Store Monitoring Data

The samplers contain enough memory to store five sampling programs, sampling data, Teledyne Isco 700 Series Module readings, and SDI-12 parameter readings. You can view the readings on your samplers display. You can also retrieve the readings and reports so the information can be processed on a personal computer. Readings and reports may be collected with a computer running Teledyne Iscos Flowlink software. Flowlink can download the information through a direct connection, a modem connection (when the

1-2

6712 Portable Sampler Section 1 Introduction sampler is equipped with the optional 4200T dial-out modem or connected to the optional digital cellular modem), or from a 581 Rapid Transfer Device (RTD). The RTD is a quick and simple way to transfer the data from the field to your computer. Reports are easily collected with Teledyne Iscos Samplink software.

1.5 Pump Requirements

The samplers peristaltic pump is extremely powerful. This pump requires a high capacity power source such as Teledyne Iscos model 913 or 923 power pack. Older model power packs do not provide the needed power. A list of recommended power packs and batteries can be found in Installing a Power Source on page 2-9. The pump also requires a pump tube made specifically for the 6712 and 6700 Series samplers. These pump tubes are easily recognized by their blue alignment collars. The 6712 pump tubing is the same as that for Teledyne Iscos 6700 Series samplers, but different from Teledyne Iscos earlier model samplers, such as the 3700 Series. Other types of pump tubing will not work in the 6712. Refer to Replacing the Pump Tube on page 8-8.

CAUTION
This sampler has a high performance pump. As with all such pumps, it relies upon liquid to cool working components. If the sampler is programmed to pump in the absence of liquid in excess of 5 minutes, excessive heat buildup may damage the paddles, rollers, and housing. Ensure that the liquid inlet is completely immersed.

Table 1-1 6712 Standard and Compact Sampler Features


General Features
Top Cover Protects pump, power source, and controller. Lockable latches. Contains a rechargeable desiccant to prevent moisture damage to the electronics, pump, and distributor systems. Control panel sloped 15 degrees for easy reading. Keys labeled with large, vivid icons. 80-character display (4 lines by 20 columns). Display has selectable backlight: Always on or always off. Timed, switching off when keypad is inactive for 60 seconds. Memory for program and data storage. Flash memory for easy software upgrades. NEMA 4X and 6 (IP67) ratings. Secured to tub with lockable latches. Large handles.

Controller

Center Section

1-3

6712 Portable Sampler Section 1 Introduction

Table 1-1 6712 Standard and Compact Sampler Features (Continued)


Adjustable Distributor Arm Discharge Tube and Support Spring A single distributor arm adjusts quickly to fit all bottle kits. Easily removed for composite sampling. Routes sample liquid from pump tube, through distributor arm to sample bottle.

Composite Tube Guide for Com- Keeps discharge tube in place over composite bottles. posite Sampling Tub Large, comfortable handles. Both compact and standard samplers accept a variety of bottle kits. Compact Tubs Only: Removable bottle-carrier insert. Standard Tubs Only: Tub drain. Optional five gallon bottle base: fits the standard sampler 581 Rapid Transfer Device 1640 Liquid Level Actuator 674 Rain Gauge 2100 Series Flow Modules 700 Series Modules 4100 Series Flow Loggers SDI-12 Sondes 4200 Series Flow Meters Refrigerator Temperature Sensor Flowlink, Samplink

Compatible Teledyne Isco Products

Real-Time Displays

As the sampler runs a sampling program, it displays the programs status. The status display may include such information as the time of the next sample, the number of the next bottle, or whether the sampler is disabled or stopped. If the sampler encounters an error while running the program, it displays a message alerting you to the problem. Program the modules from the samplers control panel. The modules program settings become part of the sampling program. Calibrate and program SDI-12 sondes from the samplers control panel. The program settings become part of the sampling program. See Section 6. 512 kilobytes of battery-backed RAM (Random Access Memory), to store: Five sampling programs. A sampling report from the most recently run program. It records as many as 1000 sampling events; events can be the program start time, enable time, sample event information, etc. 700 Series module readings. The readings can be: level, flow rate, velocity, pH, temperature, or the data collected by the 4-20 mA module. Rain gauge and refrigerator temperature readings. SDI-12 sonde readings. The Program Settings report, listing current program settings. The Sampling Results report, listing the events occurring during the program. The Combined Results report, combining sampling events with readings from a rain gauge, module, or SDI -12 sonde. The Module Summary report, summarizing flow rate or parameter readings. The Rainfall Summary report, listing a summary of rainfall readings. ASCII data output from the interrogator port.

Programming for Modules Setup for SDI-12 Sondes Memory for Stored Programs and Readings

Five Reports Available

Serial Data Output

1-4

6712 Portable Sampler Section 1 Introduction

Table 1-1 6712 Standard and Compact Sampler Features (Continued)


Units of Measure External Sampler Enable A variety of metric and English units of measure for length, flow rate, flow volume, and temperature. Teledyne Isco flow meters and flow loggers have a programmable sampler enable feature that lets them send an electronic signal to a 6712 that enables (starts) or disables (stops) a running sampling program. Dual Sampler Mode operates two samplers. Operate sampler functions using RS-232 communications. Pump Tube Warning. The 6712 displays a warning to inspect the tube. Internal Battery Warning. From the Maintenance screen, the 6712 displays a warning when it is time to replace the internal battery. The internal battery preserves stored data when the 6712 is without external power. The 6712 controller can be ordered with a factory-installed 2400 baud dialout modem. With the modem you can: connect to the sampler and download data using Flowlink software. program the sampler to call a contact list when an alarm condition exists. use the Remote Commands to control the samplers operation from a remote location. When programming the sampler, press the [?] (Help) key for a brief help note. All help topics appear in the index. Standard programming lets you set up typical sampling programs quickly. Extended programming includes all features available in standard programming plus additional features.

Dual Sampler Mode Command Driven Operation Warning Messages

Optional Dialout 4200T Modem

On-Line Help Notes Two Programming Levels

Sample Delivery System Features


Peristaltic Pump Benefits: Liquid moves continuously under pumped flow. The pump has no metering chambers or gravity fed internal tubing to trap sediment or residual liquid. Sample liquid contacts only the strainer, suction line, tube coupling, pump tube, bulkhead fitting, and sample bottles. Non wetted Detection: Sample liquid never touches the detector. The sampling cycle always includes a pre-sample purge and post-sample purge that clears the suction line of residual liquid. Simply disconnect the pump tube from the bulkhead fitting on the center section, and place the pump tube over your sample container.
3 3

LD90 Liquid Detector Two Line Purges in Sampling Cycle Easy Grab Samples Vinyl and PTFE Suction Lines

/8-inch ID (Inside Diameter) vinyl line. /8-inch ID PTFE lined with polyethylene jacket. Standard Programming Features

Pacing

Uniform Time Pacing: Sampling at regular time intervals. Flow Pacing: Sampling at regular flow-volume intervals. Composite: Samples deposited in a single large bottle. Sequential: Only one sample placed in each bottle. Samples Per Bottle: Multiple samples placed in each bottle. Bottles Per Sample: One sample deposited in multiple bottles.

Distribution Methods

1-5

6712 Portable Sampler Section 1 Introduction

Table 1-1 6712 Standard and Compact Sampler Features (Continued)


Three Flexible Start-Time Settings Start Immediately: Starts the sampling program immediately. Delayed Start: Starts the sampling program after a user-definable delay of 1 to 999 minutes. Clock Time: Starts the sampling program at a user-definable time on one or more days of the week. Continuous Sampling: When sample bottles are regularly replaced, the sample distribution can restart with the first bottle set after the last bottle set is filled, without interrupting the running program.

Option for a Continuous Running Program

Extended Programming Features


Pacing Uniform Time Pacing: Sampling at regular time intervals. Nonuniform Time Pacing: Sampling at irregular time intervals. Random Interval Pacing: Sampling at unique random time intervals generated by the controller. Flow Pacing: Sampling at regular flow-volume intervals. Event Pacing: Sampling each time a user-definable event occurs. Allows for sample sizes to be based on flow. (This option is available only with uniform time pacing.) Composite: Samples deposited in a single large bottle. Sequential: Only one sample placed in each bottle. Samples Per Bottle: Multiple samples placed in each bottle Bottles Per Sample: One sample deposited in multiple bottles Multiple Bottle Compositing: A combination of samples per bottle and bottles per sample distribution methods. Time Switched Bottles or Bottle sets: Control the sample distribution using clock times. A 6712 sampler can be programmed to enable or disable a running sampling program when readings received from a connected rain gauge, module, or SDI-12 Sonde meet certain conditions. Create intermittent sampling schedules. Two-part programming lets you set up a sampling program that divides the bottles into two groups, filling each group according to separate pacing, distribution, sampler enable, and pause and resume settings. This is ideal for storm-water run-off sampling.

Flow Proportional Sample Volumes Distribution Methods

Programmable Sampler Enable

Pauses and Resumes Two-Part Programming

Auto Suction Head or Fixed Suc- The suction head, is the vertical distance from the flow stream to the liquid detection Head tor. Extended programming has two settings. Auto-Suction Head: The head is automatically determined. Fixed Suction Head: A user-definable measurement for the head. Suction Line Rinses Sampling Retries Program setting for the number of times (0 to 3) that the 6712 rinses the suction line before drawing a sample. Program setting for the number of times (0 to 3) that the 6712 attempts to sample if it fails to deliver the entire sample volume.

1-6

6712 Portable Sampler Section 1 Introduction

Table 1-1 6712 Standard and Compact Sampler Features (Continued)


Three Flexible Start Time Settings Run Immediately: Starts the sampling program immediately. Delayed Start: Starts the sampling program after a user-definable delay of 1 to 999 minutes. Clock Time: Starts the sampling program at a user-definable time on one or more days of the week. Continuous Sampling: When sample bottles are regularly replaced, the sample distribution can restart with the first bottle set after the last bottle set is filled, without interrupting the running program.

Option for a Continuous Running Program

Table 1-2 6712 Standard and Compact Sampler Construction Materials


Item
Top Cover Center Section Tubs Bottle Inserts Plastic Retaining Rings Bottle Carriers Bottle Rack Controller Case Distributor Arm Distributor Arm Nut Pump Rollers Control Panel Connector Labels Distributor Shaft Pump Shaft Pump Paddles Pump Band Bulkhead Fitting Latches Handles Metal Retaining Rings Power-Source Draw Catches: Amphenol Connectors Retaining Ring Draw Cords (inside standard tub) Retaining Ring Tie-Down (inside standard tub) Iridited aluminum EPDM (Ethylene Propylene Diene Monomer) rubber Cadmium Plated Aluminum Polypropylene jacketed rubber with stainless steel clip Stainless steel Polyester Delrin Powder coated steel Noryl

Material
ABS (Acrylonitrile Butadiene Styrene) plastic

1-7

6712 Portable Sampler Section 1 Introduction

Table 1-2 6712 Standard and Compact Sampler Construction Materials (Continued)
Drain Plug (standard tub only) Webbing for Compact Bottle Carriers Polypropylene. CPVC with polypropylene retaining strap

Table 1-3 Technical Specifications for the 6712C and Standard Samplers
General Notes: 1. All weights may vary 1 lb (1/2 kg). 2. All dimensions may vary 1/4 inch (0.64 cm). 3. Sample delivery specifications valid for the following conditions and ranges, unless otherwise stated: 75F (24C) ambient, sample liquidtap water at 50 to 80F (10 to 27C), sample volumes from 50 ml to 1000 ml, suction line lengths of 25 ft (7.6 m) or less, suction heads from 1 to 20 ft (0.3 to 6.1 m), atmospheric pressure between 29.92 and 31.89 inHg (760 and 810 mmHg), and a power source of a Teledyne Isco High Capacity Power Pack at 12.5 volts DC no load output.

6712 Controller, Pump, and Tubing: Mechanical Specifications


Weight of Controller: Weight of Nickel Cadmium Battery: Weight of Lead Acid Battery: Weight of Power Pack: Controller only: 13.0 lbs (5.9 kg) With pump tube: 13.2 lbs (6.0 kg) 4.3 lbs (2.0 kg) 5.7 lbs (2.6 kg) 6.0 lbs (2.7 kg)

Weight of Battery Backed Power 6.6 lbs (3.0 kg) Pack: Controller Dimensions: Length:10.3 in (26.0 cm) Width:12.5 in (31.7 cm) Height:10 in (25.4 cm) Operational: 32F to 120F (0C to 49C) Storage: 0F to 140F (-18C to 60C) 2,000 Meters II 2 95% RH Maximum NEMA 4X and 6 Pump: IP17 Enclosure: IP67

Temperature Ranges: Maximum Altitude: Installation Category: Pollution Degree: Humidity: Enclosure: Typical Delivered Volume Accuracy: Typical Repeatability:

(The ability to deliver the programmed sample volume.) 10 ml or 10% of programmed value, whichever is greater. (The ability to repeat the delivered volume for a set of samples collected under the same conditions.) 5 ml or 5% of the average of the maximum and minimum sample volume in a sample set, whichever is greater, at lifts up to 25 feet.

1-8

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
Typical Pump Flow Rate and Line Transport Velocity: Suction Head 3 feet 5 feet 10 feet 15 feet 20 feet 23 feet 25 feet Flow Rate 3.9 liters per minute 3.9 liters per minute 3.7 liters per minute 3.6 liters per minute 3.3 liters per minute 3.1 liters per minute 2.8 liters per minute Line Transport Velocity 3.0 ft/s (0.91 m/s) 3.0 ft/s (0.90 m/s) 2.9 ft/s (0.87 m/s) 2.7 ft/s (0.83 m/s) 2.5 ft/s (0.77 m/s) 2.3 ft/s (0.71 m/s) 2.2 ft/s (0.66 m/s)

Maximum Suction Head:

(The maximum suction head is the vertical height at which a sample can be taken.) 28 ft (8.5 m) at 30 inHg (762 mmHg) Tube life may vary depending on the abrasiveness of the sample liquid. Recommended maximum 1,000,000 pump counts. This equates to 912 standard samples. A standard sample is 200 ml at 5 ft (1.5 m) suction head using a 10 ft (3 m) vinyl suction line.

Pump Tube Life:

6712 Controller: Electrical Specifications


Controller Internal Battery: Real Time Clock Accuracy: Program Memory: Sampler Power Requirements: 5 years minimum (maintains internal logic, program settings & stored data) 1 minute per month Nonvolatile programmable Flash. Can be field updated via interrogator connector. Nominal: 12 volts DC by battery or AC power converter Operational: 11 to 13 Volts DC Absolute Maximum: 14.4 Volts DC

Power Pack Requirements: 120VAC 12VAC, 60 Hz., 1.0 Amp. Use only Teledyne Isco-made Note: This is the only version certified by UL. The line cord (mains power packs certified by UL. (P/N connect cable) is the Disconnect Device. 60-1684-088) Power Pack Requirements: (Use only Teledyne Isco-made CE certified power packs. (P/N 60-3004-190) 230VAC 23VAC, 2.0 Amp, 50/60 Hz. Note: The line cord (mains connect cable) is the Disconnect Device.

Power Consumptiona
6712 Controller in Standby Average Current of Accessories Controller Display Backlight 701 pH Module
b

10 mA

228 mA 11 mA 11 mA 11 mA 24 mA 31 mA 12 mA 10 mA 11 mA

710 Ultrasonic Moduleb 720 Submerged Probe Moduleb 730 Bubbler at 2 ft of liquid level at 10 ft of liquid level 750 Area Velocity Moduleb 780 4-20 mA Moduleb YSI 600 with pH, DO, Conductivity, and Temperatureb Moduleb

1-9

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
Programmable Analog 4-20 mA Output Optionc CDMA cellular telephone modem standby current current while communicating GSM cellular telephone modem standby current current while communicating
a. Current ratings when input power is 12.5 VDC and ambient temperature is 77 F (25 C) b. Average current when programmed to take readings at 15 minute intervals. c. Rating for three outputs.

124 mA 35 mA 150 mA 62 mA 230 mA

Sampler Only Operating Current:

Approximately 30 mA based on 200 ml sample every hour, 10 ft (3 m) suction line, and 5 ft (1.5 m) suction head.

Teledyne Isco External Battery Capacity:Number of Complete Standard Sampling Routines Nickel-Cadmium Battery Sampler Only: Sampler with pH Module: Sampler with Ultrasonic Module: Sampler with Submerged Probe Module: Sampler with Bubbler Module (1-2 ft head): Sampler with Area Velocity Module: Sampler with 4-20 mA Module: 6 6 6 6 3 4 6 Lead-Acid Battery 9 9 9 9 5 6 9

A Standard Sampling Routine is a 200 ml sample taken every hour in 24 bottle mode, with a 10 foot (3 m) vinyl suction line at 5 feet (1.5 m) of head. The module is set to take a reading every 15 minutes. Flow Meter Signal Requirement: 5 to 15 volt DC pulse or isolated contact closure of at least 25 millisecond duration. (4 - 20 mA or pulse duration signal may be converted with optional interface unit.)

1-10

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
6712 Controller: Software Specifications
Suction Line: Programmable 3 to 99 ft (1 to 30 m) lengths of: 3/8" (0.95 cm) I.D. Vinyl 3/8" (0.95 cm) I.D. PTFE-lined with polyethylene jacket From 1 minute to 99 hours and 59 minutes in 1 minute increments between consecutive samples. Nonuniform times in minute intervals or clock time Random time intervals between consecutive samples From 1 to 9,999 flow pulses in single-pulse intervals Flow paced in volume with attachable flow module Inches or millimeters pH 8 bits representing pH 16 bits representing temp. 0.1 pH storage resolution 0.1 C storage resolution 0.0001 m storage resolution 0.0001 m storage resolution 0.0001 m storage resolution 0.0001 m storage resolution 0.001 ft/s storage resolution 0.1% storage resolution

Sample Frequency:

Rainfall Reading Units: Module Readings:

Ultrasonic 16 bits representing level Submerged Probe 16 bits representing level Bubbler 16 bits representing level Area Velocity 16 bits representing level 16 bits representing velocity 420 mA 16 bits representing percent 16 bits representing level 32 bits representing flow SDI-12 Sonde Readings: Parameter Temperature Conductivity Specific Conductance Total Dissolved Solids Salinity Dissolved Oxygen pH ORP Level Ammonium-Nitrogen Ammonia-Nitrogen Nitrate-Nitrogen Turbidity Chloride Chlorophyll Other

Range -40 - 100 C 0 - 100 mS/cm 0 - 100 mS/cm 0 - 90 g/l 0 - 70 ppt 0 - 20 mg/l 0 - 14 -999 - 999 mv 0 - 6.5279 m 0 - 200 mgN/l 0 - 200 mgN/l 0 - 200mgN/l 0 - 5000 NTU 0 - 1000 mg/l 0-400 ug/l -99999.0-99999.0

Resolution 0.1 C 0.01 mS/cm 0.01 mS/cm 0.01 g/l 0.1 ppt 0.1 mg/l 0.1 0.1 mv 0.0001 m 0.1 mgN/l 0.1 mgN/l 0.1 mgN/l 0.1 NTU 0.1 mg/l 0.1 ug/l 0.0001

Storage Bytes 2 2 2 2 2 1 1 2 2 2 2 2 2 2 2 4

1-11

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
Module Reading Conversions: Level and 420 mA readings converted to flow rate units: gallons per second gallons per minute million gallons per day cubic feet per second cubic meters per second cubic meters per hour cubic meters per day liters per second Velocity units: feet per second Totalized flow units: gallons cubic feet liters meters per second million gallons cubic meters

Devices supported in flow conversion: Weirs: V-notch; 22.5, 30, 45, 60, 90, and 120 degrees Rectangular; with and without end contractions Cipoletti. Flumes: Parshall; 1", 2", 3", 6", 9", 1.5', 2', 3', 4', 5', 6', 8', 10', 12' Palmer-Bowlus; 4", 6", 8", 12", 15", 18", 21", 24", 27", 30", 48" Trapezoidal; LG 60 V, 2" 45 WSC, 12" 45 SRCRC H; 0.5', 0.75', 1', 1.5', 2', 2.5', 3', 4.5' Equation: Q = a x Hb+c x Hd where: Q = flow H = head a,b,c, and d = entered values Manning Formula: Round, U-channel, Rectangular, Trapezoidal Area Velocity: Round, U-channel, Rectangular, Trapezoidal Data Points: 50 level-flow rate points, 50 level-area points Refrigerator Temperature Sensor Optional sensor and 5 ft (1.55 m) cable to record temperature readings. Intended for use with fiberglass refrigerated (FR) samplers. Range: -40.0 C to 100 C Accuracy specification: 0.3 C over a range of 0.0 to 10.0 C. Reading data is 2 bytes, with a storage resolution of 0.1 C.

6712 Standard Portable Sampler: Physical Specifications


Dimensions: Sampler Tub Configurations: Height: 27 inches (68.6 cm) Diameter: 19.96 inches (50.7 cm) 24 wedge polypropylene bottles, 1,000 ml 24 round glass bottles, 350 ml 12 round polypropylene bottles, 1,000 ml 12 round glass bottles, 950 ml 8 round polyethylene bottles, 2.0 liters 8 round glass bottles, 1.8 liters 4 round polyethylene bottles, 3.7 liters (1 gal) 4 round glass bottles, 3.7 liters (1 gal) 1 round polyethylene bottle, 9.4 liters (2.5 gal) 1 round glass bottle, 9.4 liters (2.5 gal)

Dry Weight:

32 lbs (15 kg)

This includes the top cover, center section, tub, controller, pump tube, distributor arm, distributor arm nut, and discharge tube for the 24 bottle configuration.

1-12

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
Dry Weight With: 24 plastic bottles, hold down, no caps 36 lbs 24 glass bottles, expander ring, no caps 55 lbs 12 plastic bottles, insert, expander ring, no caps 37 lbs 12 glass bottles, insert, expander ring, no caps 45 lbs 8 plastic bottles, insert, hold down, no caps 36 lbs 8 glass bottles, insert, hold down, no caps 46 lbs 4 plastic bottles, insert, hold down, no caps 36 lbs 4 glass bottles, insert, hold down, no caps 45 lbs 1 plastic bottle and composite tube guide, no cap, arm or nut 34 lbs 1 glass bottle and composite tube guide, no cap, arm or nut 38 lbs 24 plastic bottles 24 glass bottles 12 plastic bottles 12 glass bottles 8 plastic bottles 8 glass bottles 4 plastic bottles 4 glass bottles 1 plastic bottle 1 glass bottle 101 lbs(46 kg) 104 lbs(48 kg) 89 lbs (41 kg) 95 lbs (43 kg) 84 lbs (38 kg) 90 lbs (41 kg) 98 lbs (45 kg) 104 lbs(48 kg) 90 lbs (41 kg) 94 lbs (43 kg) (17 kg) (25 kg) (17 kg) (21 kg) (17 kg) (21 kg) (17 kg) (21 kg) (16 kg) (18 kg)

Filled Bottle Weight:

This includes everything in the Dry Weight, Dry Weight With, and Ice Capacity specifications, and the bottles are filled to the capacities listed in Sampler Tub Configurations. This weight does not include a power source. Ice Capacity: 24 plastic bottles 24 glass bottles 12 plastic bottles 12 glass bottles 8 plastic bottles 8 glass bottles 4 plastic bottles 4 glass bottles 1 plastic bottle 1 glass bottle 12 lbs 30 lbs 16 lbs 16 lbs 12 lbs 12 lbs 28 lbs 25 lbs 35 lbs 35 lbs (5.4 kg) (13.6 kg) (7.3 kg) (7.3 kg) (5.4 kg) (5.4 kg) (12.7 kg) (11.3 kg) (15.9 kg) (15.9 kg)

3/4" 3/4" 1/4" approximate ice cube size. Ice capacity will vary with the size of the cubes.

Cooling Capacity Configuration: 24 plastic bottles 24 glass bottles 12 plastic bottles 12 glass bottles 8 plastic bottles 8 glass bottles 4 plastic bottles 4 glass bottles 1 plastic bottle 1 glass bottle Sample Size 500 ml Sample 175 ml Sample 500 ml Sample 475 ml Sample 1,000 ml Sample 900 ml Sample 1,850 ml Sample 1,850 ml Sample 12400 ml Samples 12400 ml Samples Cooling Water 61F 39F 36F 36F 50F 37F 34F 37F 37F 36F (16C) (4C) (2C) (2C) (10C) (3C) (1C) (3C) (3C) (2C) Last Sample Last Sample Temp 63F 41F 54F 54F 57F 54F 34F 37F 37F 36F (17C) (5C) (12C) (12C) (14C) (12C) (1C) (3C) (3C) (2C) Sample 24 Sample 24 Sample 12 Sample 12 Sample 8 Sample 8 Sample 4 Sample 4 Composite Sample Composite Sample

The cooling capacity is the temperature of the sample and the cooling water 24 hours after the last sample is taken. The amount of ice specified for the ice capacity was used. The temperature of the sample liquid is 65F (18C). The ambient temperature is 75F (24C). Average Thermal Resistance Factor: R-10

The average R-value for a complete sampler. This includes a tub, center section, top cover, controller, pump tube, distributor arm, and discharge tube.

1-13

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
6712 Standard Sampler with Optional 5 Gallon Bottle: Physical Specifications
Dimensions Base Alone: Assembled Sampler: Sampler Tub Configuration: Dry Weight: Height: 15.4 inches (39.1 cm) Diameter: 19.9 inches (50.6 cm) Height: 31.5 inches (80.0 cm) Diameter: 19.9 inches (50.6 cm) 1 round glass bottle, 5 gallons (18.9 liters) 52 lbs (24 kg)

This includes the top cover, center section, bottle base, 5 gallon glass bottle, controller, pump tube, and discharge tube. Ice Capacity: 21 lbs (9.5 kg)
3/4" 3/4" 1/4" approximate ice cube size. Ice capacity will vary with the size of the cubes.

Filled Bottle Weight:

115 lbs (52 kg)

This includes everything in the Dry Weight and the Ice Capacity specifications, and the bottle is filled to the capacities listed in Sampler Tub Configurations. This weight does not include a power source. Cooling Capacity: Sample Size 12800 ml Samples Cooling Water 36F (2C) Composite Sample Temperature 36F (2C)

The cooling capacity is the temperature of the sample and the cooling water 24 hours after the last sample is taken. The amount of ice used is specified in ice capacity. The sampler is set up to take one sample every hour for twelve hours filling the bottle half full. The temperature of the incoming sample liquid is 65F (18C). The ambient temperature is 75F (24C). Average Thermal Resistance Factor: R-10

This is the average R-value for the complete sampler. This includes a bottle base, center section, top cover, controller, pump tube, discharge tube, and composite tube guide.

6712 Compact Portable Sampler: Physical Specifications


Dimensions: Sampler Tub Configurations: Height: 27.6 inches (70.1 cm) Diameter: 17.75 inches (45.1 cm) 24 wedge polyethylene bottles, 500 ml 12 round polyethylene bottles, 500 ml 12 round glass bottles, 375 ml 1 round polyethylene bottle, 9.4 liters (2.5 gallon) 1 round glass bottle, 9.4 liters (2.5 gallon)

Dry Weight:

31 lbs (14 kg)

This includes the top cover, center section, tub, controller, pump tube, distributor arm, distributor arm nut, and discharge tube for the 24 bottle configuration. Dry Weight With: 24 plastic bottles, insert, hold down, no caps 35 lbs 12 plastic bottles, insert, expander ring, no caps 35 lbs 12 glass bottles, insert, expander ring, no caps 43 lbs 1 plastic bottle and composite tube guide, no cap, arm or nut 33 lbs 1 glass bottle and composite tube guide, no cap, arm or nut 37 lbs (16 kg) (16 kg) (20 kg) (15 kg) (17 kg)

1-14

6712 Portable Sampler Section 1 Introduction

Table 1-3 Technical Specifications for the 6712C and Standard Samplers (Continued)
Filled Bottle Weight: 24 plastic bottles 12 plastic bottles 12 glass bottles 1 plastic bottle 1 glass bottle 70 lbs 62 lbs 66 lbs 66 lbs 73 lbs (32 kg) (28 kg) (30 kg) (30 kg) (33 kg)

This includes everything in the Dry Weight, Dry Weight With, and Ice Capacity specifications, and the bottles are filled to the capacities listed in Sampler Tub Configurations. This weight does not include a power source. Ice Capacity: 24 plastic bottles 12 plastic bottles 12 glass bottles 1 plastic bottle 1 glass bottle 8 lbs 13 lbs 13 lbs 12 lbs 15 lbs (3.6 kg) (5.9 kg) (5.9 kg) (5.4 kg) (6.8 kg)

/4" x 3/4" x 1/4" approximate ice cube size. Ice capacity may vary with the size of the cubes.
Sample Size 250 ml Sample 250 ml Sample 175 ml Sample 12400 ml Samples 12400 ml Samples Cooling Water 41F 34F 34F 43F 39F (5C) (1C) (1C) (6C) (4C) Last Sample Last Sample Temp 55F 41F 43F 43F 39F (13C) (5C) (6C) (6C) (4C)

Cooling Capacity: Configuration: 24 plastic bottles 12 plastic bottles 12 glass bottles 1 plastic bottle 1 glass bottle Sample 24 Sample 12 Sample 12 Composite Sample Composite Sample

The cooling capacity is the temperature of the sample and the cooling water 24 hour after the last sample is taken. The amount of ice specified for the ice capacity was used. The temperature of the sample liquid is 65F (18C). The ambient temperature is 75F (24C). Average Thermal Resistance Factor: R-11

The average R-value for a complete sampler. This includes a tub, center section, top cover, controller, pump tube, distributor arm, and discharge tube.

1-15

6712 Portable Sampler Section 1 Introduction

Edited by Foxit Reader Copyright(C) by Foxit Software Company,2005-2008 For Evaluation Only.

Table 1-4 Connector Specifications


Connector Icon Pin Location Diagram Pin Function
A Ground B +12 volts DC
A

POWER SOURCE

2-PIN MALE, SEALED

A +12 volts DC

RAIN GAUGE

9-PIN FEMALE, SEALED

B Ground C Programmable I/O1 pin D Rain gauge input

E A F B G I C D H

E Not Connected F SDI-12 Data G Not Connected H Programmable I/O2 pin I Programmable I/O3 pin

INTERROGATOR

6-PIN FEMALE, SEALED


B A C D F E

A Switched +12 volts DC B Ground C Connection Sense D Transmit E Receive F NC

FLOW METER

6-PIN MALE, SEALED


B C

A +12 volts DC B Ground C Flow Pulses In D Bottle Number Out Dual Sampler Out E Event Mark Out F Inhibit In

8-PIN MALE, SEALED MODULE


2 8 7 1 4 5

1 NC 2 +12 volts DC 3 Ground 4 Module Sense 5 Receive

(IN MODULE BAY)

6 Transmit 7 Clock 8 Control

1-16

6712 Portable Sampler


S ec tion 2 Ins tallation/Preparation

2.1 Preparing the Sampler

When sampling with a single bottle, the composite tube guide is installed and the discharge tube is threaded through the guide. When sampling with multiple bottles, both the compact and standard 6712 Samplers use a distributor arm. It is adjustable so that you can position the discharge tube over the bottles of each kit. Each bottle kit also uses a discharge tube of a specific length. Table 2-1 lists the correct discharge tube lengths for each bottle kit. After installing the distributor arm and discharge tube, run the distributor to each bottle position to ensure proper installation. If using bulk tubing to cut your own tubes, cut the ends of the tube square, not at an angle, and cut the length accurately.

2.2 Installing the Distributor Arm and Discharge Tube

The two-piece distributor arm has four positions, marked on the underside with the letters A, B, C, and D. Adjust the arm to the correct position for the bottle kit being used. Table 2-1 shows the correct position for each kit. Thread the discharge tube through the spring and arm so that the tube protrudes from the arm 1/16 to 1/8 inch. If too much tube protrudes from the arm it can catch on bottles and jam the distributor arm. Attach the arm to the distributor shaft. Secure the arm with the black plastic nut. Slide the free end of the discharge tube onto the bulkhead fitting.

Distributor Arm

Composite Tube Guide

2-1

6712 Portable Sampler Section 2 Installation/Preparation

Step 1
The two-piece arm has four positions, marked on the underside with the letters A, B, C, and D. Adjust the arm to the correct position for the bottle kit. Table 2-1 shows you the correct position for each kit. Thread the discharge through the spring and arm so that the tube protrudes from the arm 1/16 to 1/8 inch.

Step 2
Attach the arm to the distributor shaft. The arm and shaft are keyed so that you can align them correctly. Secure the arm to the shaft with the black plastic nut.

Step 3
Slide the end of the discharge tube onto the bulkhead fitting.

Figure 2-1 Installing the Distributor Arm

Table 2-1 Distributor Arm Adjustments and Discharge Tube Lengths for Bottle Kits
Compact 24-Bottle Kit

500-milliliter polyethylene bottle


Distributor Arm Position: B Discharge Tube: 60-9003-257 Discharge Tube Length: 14 /4 inches (36.2 cm)
1

Compact 12-Bottle Kit

375-milliliter glass bottles 500-milliliter polyethylene bottles


Distributor Arm Position: B Discharge Tube: 60-9003-257 Discharge Tube Length: 14 /4 inches (36.2 cm)
1

2-2

6712 Portable Sampler Section 2 Installation/Preparation

Table 2-1 Distributor Arm Adjustments and Discharge Tube Lengths for Bottle Kits
Standard 24-Bottle Kit

350-milliliter glass bottles 1,000-milliliter polypropylene bottles


Distributor Arm Position: D Discharge Tube: 60-9003-260 Discharge Tube Length: 17 /4 inches (43.8 cm)
1

Standard 12-Bottle Kit

950-milliliter glass bottles 1,000-milliliter polypropylene bottles


Distributor Arm Position: C Discharge Tube: 60-9003-262 Discharge Tube Length: 16 /2 inches (41.9 cm)
1

Standard 8-Bottle Kit

1.8-liter glass bottles 2.0-liter polyethylene bottles


Distributor Arm Position: B Discharge Tube: 60-9003-264 Discharge Tube Length: 15 /4 inches (40.0 cm)
3

Standard 4-Bottle Kit

3.7-liter (1-gallon) glass bottles 3.7-liter (1-gallon) polyethylene bottles


Distributor Arm Position: A Discharge Tube: 60-9003-266 Discharge Tube Length: 15 inches (38.1 cm)

2.3 Installing Bottle Kits for Multiplex Sampling

The samplers are shipped from the factory with sample bottles in place. When using the sampler for the first time, you will only need to remove the bottle lids. When installing cleaned bottles or a new kit, keep these guidelines in mind: Never mix glass and plastic bottles together. Install all bottles to insure that they remain in place. Align the bottles correctly. If the bottles are misaligned, the sampler may miss the bottle mouth, or the sampler will deposit samples in the wrong bottle.

2-3

6712 Portable Sampler Section 2 Installation/Preparation Each time you change from one size bottle to another, you must adjust the length of the distributor arm and install a discharge tube of the correct length. Each time you install a bottle kit, check the program settings for the number of bottles and bottle volume. You must adjust the distributor arm and use a different discharge tube for each bottle kit. See Installing Distributor Arm and Discharge Tube, and Table 2-1 for more information. 2.3.1 Aligning Bottle Kits in the Standard Tub Figures 2-2 through 2-5 show each kit being installed in the standard tub. Refer to Table 2-1 to adjust the distributor arm to the correct length and use the correct length of discharge tube.

1. Install the bottles then secure the plastic retaining ring with the three draw cords. (For glass bottles use the stainless steel retaining ring). 2. Adjust the distributor arm to the correct position, insert the discharge tube, and attach the distributor arm.

Figure 2-2 Placing the 24-Bottle Kit in a Standard Tub

2-4

6712 Portable Sampler Section 2 Installation/Preparation

1. Slide the 12-bottle insert into the tub, aligning the label with the corresponding numbers on the outside of the tub. 2. Place the bottles in the insert. Secure the bottles against the wall with the stainless steel retaining ring. 3. Adjust the distributor arm to the correct position, insert the discharge tube, and attach the distributor arm.

Figure 2-3 Placing the 12-Bottle Kit in a Standard Tub

1. Position the 8-bottle insert in the tub, aligning the label with the corresponding numbers on the outside of the tub. 2. Place the bottles in the insert then secure the plastic bottle insert with the three draw cords. 3. Adjust the distributor arm to the correct position, insert the discharge tube, and attach the distributor arm.

Figure 2-4 Placing the 8-Bottle Kit in a Standard Tub

2-5

6712 Portable Sampler Section 2 Installation/Preparation

1. Place the rack in the tub, aligning the label with the corresponding numbers molded on the outside of the tub. 2. Place bottles in the rack and secure the hold down over the bottles with the three draw cords. 3. Adjust the distributor arm to the correct position, insert the discharge tube, and attach the distributor arm.

Figure 2-5 Placing the 4-Bottle Kit in a Standard Tub

Compact Sampler Tub Alignment Notches

Aligning the carrier

Figure 2-6 Placing the carrier in the compact tub

2-6

6712 Portable Sampler Section 2 Installation/Preparation 2.3.2 Aligning Bottle Kits in the Compact Tub To align the bottles: 1. Place the bottles in the carrier, and secure them with the appropriate retaining ring. 2. Place the carrier in the tub, aligning it as shown in Figure 2-6. The carrier will drop into the notches when properly aligned.

2.4 Installing Bottle Kits for Composite Sampling

You can install a single 21/2 gallon bottle in either the standard or compact sampler tub for composite sampling. Single bottle sampling uses a composite tube guide instead of a distributor arm to position the discharge tube over the bottle. Install the composite tube guide and discharge tube as shown in Figure 2-7. Discharge tube part numbers and lengths are listed in Table 2-2. The compact and standard sampler bottle bases accept a 2 1/2 gallon glass or plastic bottle. Place the bottle in the recess formed in the bottle base.

2.4.1 Optional Jumbo Bottle Base

You can install an optional bottle base on the standard sampler to collect larger composite volumes. The optional base replaces the standard tub and holds a 4 gallon (15 liter) polyethylene bottle, 5 gallon (19 liter) glass bottle or 5.5 gallon (20.8 liter) polyethylene bottle. Prepare the sampler as shown in Table 2-2 and Figure 2-7, and place the bottle in the recess formed in the bottle base.

Step 1
Both the compact and standard samplers use the same composite tube guide to hold the discharge tube in place over the composite bottle. Attach the composite tube guide by firmly sliding it onto the distributor-shaft housing. It will snap in place. The guide has a slot that fits over a key on the housing to align it.

Step 2
Thread the discharge tube through the guide as shown. Attach the tube to the bulkhead fitting.

Removal
To remove the composite tube guide, rotate the metal clip to one side while pulling the guide away from the center section.

Figure 2-7 Installing the Composite Tube Guide

2-7

6712 Portable Sampler Section 2 Installation/Preparation

Table 2-2 Discharge Tubes for Composite Sampling


Configuration
Compact 1 Bottle
1 1

Discharge Tube
60-9003-258

Discharge Tube Length


10 inches (25.4 cm)

9.4-liter (2 /2 gallon) glass bottle 9.4-liter (2 /2 gallon) polyethylene bottle


Standard 1 Bottle 60-9003-271 12 inches (30.5 cm)

9.4-liter (2 /2 gallon) glass bottle 9.4-liter (2 /2 gallon) polyethylene bottle


Jumbo Bottle Option 60-9003-271 12 inches (30.5 cm)
1

15 liter (4 gallon) polyethylene bottle 19 liter (5 gallon) glass bottle 20.8 liter (5.5 gallon) polyethylene bottle
Standard samplers only. Requires the Jumbo Bottle Base

5 Gallon Glass Composite Bottle

2.5 Gallon Plastic Composite Bottle

2.5 Gallon Glass Composite Bottle

Figure 2-8 Composite Bottles

2-8

6712 Portable Sampler Section 2 Installation/Preparation

2.5 Installing a Power Source

The 6712 Sampler requires 12-volt DC power. Power sources available from Teledyne Isco are listed below. 946 Lead Acid Battery 934 Nickel Cadmium Battery 913 High Capacity Power Pack (120 volts AC, 50/60 Hz) 923 High Capacity Power Pack (240 volts AC, 50/60 Hz) 914 Battery Backed Power Pack (120 volts AC, 50/60 Hz) 924 Battery Backed Power Pack (240 volts AC, 50/60 Hz) (the 924 is not certified for use in Europe) 948 45 Amp Hr Battery
Power Source Connector Icon Connect the power source to the sampler connector that is labeled with this icon.

To install the smaller batteries or power packs on the sampler: 1. Place the power source in the samplers battery recess. 2. Secure the power source with the black rubber draw-catches. 3. Attach the power-source cable to the power-source connector on the controller.

CAUTION
Use only Teledyne Isco-made power packs. If using a power pack with the 6712, Teledyne Isco recommends you use the new 913 or 923 power pack. Older model power packs may not provide sufficient power. Refer to the Teledyne Isco Power Products Guide shipped with your sampler for more information on Teledyne Iscos power sources. If you use a 120 or 240 volt AC-connected power pack, the AC line cord is the "Disconnect Device." Always install the sampler with the line cord and plug visible and readily accessible, so that power can be disconnected easily. 2.5.1 Connecting an External 12-Volt DC Source You can also power the sampler with an external 12-volt DC source, such as an automotive battery. Connect the sampler to the battery with the External 12-Volt DC Power Cable.

CAUTION
Be sure to attach the positive and negative clips to the correct battery terminals. If polarity is reversed, the sampler may be permanently damaged.

2-9

6712 Portable Sampler Section 2 Installation/Preparation

2.6 Suction Line

The suction line is the tubing from the sampling point to the pump intake. The 6712 uses a 3/8-inch ID suction line of lengths 3 to 99 feet. Teledyne Isco offers vinyl or PTFE suction lines. The PTFE tubing has a polyethylene jacket to protect it from kinks and abrasions. Guidelines for Measuring and Cutting the Suction Line: Cut the line to the desired length but use the shortest length feasible for the installation. Cut the line in 1 foot increments. For instance, 4 feet, not 31/2. If using metric units of measure, cut the line in increments of 0.1 meter. Do not include the length of the strainer in the measurement. The vinyl line contains a very low parts-per-million level of phenols. If phenol content affects your sample analysis, use the PTFE suction line. When installing the sampler, be sure the vertical distance between the liquid level and the pump is as small as possible.

2.6.1 Attaching Suction Line to Pump Tube

Attach the vinyl suction line to the pump tube with the 3/8-inch tubing coupler. First, screw the threaded end into the suction line u n t i l t h e f l a t s u r f a ce i s f l u s h a g a i n s t t h e s u c t i o n l i n e (Figure 2-9). Then, push the other end of the coupler into the end of the pump tube until the other flat surface is flush against the tubing.

1.

2.

Figure 2-9 Attaching the suction line to the pump tubing

2-10

6712 Portable Sampler Section 2 Installation/Preparation Once the coupler is attached to the pump tube, removal is difficult, and may require cutting the tube.

2.7 Strainers

Teledyne Isco offers three styles of strainer that help prevent solids from clogging the suction line: 3/8-inch standard weighted polypropylene strainer for routine and priority pollutant sampling 3/8-inch stainless steel low flow strainer 3/8-inch CPVC-coated strainer for acidic liquid sources To select the right strainer for your application, see Table 2-3. To install the optional stainless strainer in PTFE suction line, heat the end of the suction line to make it more pliable, then carefully screw the strainers threaded connector into the suction line.

2.7.1 Alternative to Strainers

When sampling from high velocity streams with heavy suspended solids, some field investigations suggest that more representative samples are obtained without the strainer. Consider attaching a short piece of thin walled aluminum tubing to the end of the suction line; anchor the tubing so that the inlet opens upstream. The aluminum tubings thin wall will not disturb the flow stream, and most sample analyses disregard aluminum ions. Under most conditions, the pre-sample purge removes any debris over the tubing entrance.

Table 2-3 Selecting the Right Strainer


Strainer 3/8 Stainless Steel Ends, Polypropylene Center
1

Dimensions
10.5

Application Maximum Depth


Routine applications. Vinyl Line: 22 feet (6.7m).

.400 Diameter

1/4 Diameter

PTFE Line: 15 feet (4.5m).

3/8 Stainless Steel Low Flow

1 .400 diameter

Low flow applications


.562 diameter
11.5

Vinyl Line: 22 feet (6.7m). Teflon Line: 15 feet (4.5m).

3/8 CPVC

.400 diameter 1- 5/16 Diameter

Highly acidic liquids.

Vinyl Line: 4 feet (1.2m).

Note
The suction lines will float when filled with air during the purge cycles and when depths exceed the listed ratings. Secure the suction lines when sampling at depths exceeding the maximum ratings. All strainer dimensions are in inches. PTFE suction line is compatible only with the 3/8 stainless steel strainer.

2-11

6712 Portable Sampler Section 2 Installation/Preparation

2.8 How Does the Sampler Work?

When the sampler takes a sample, it draws liquid through the strainer and suction line to the pump tube. The liquid flows through the pump tube and past the liquid detector, which senses the liquid. From the detector, the liquid follows the pump tube through the pump to the bulkhead fitting and then through the discharge tube to the sample bottle. 1. The sampler moves the distributor arm over the bottle that is to receive the sample. 2. The pump reverses for the pre-sample purge. 3. The pump direction changes, filling the suction line. 4. When the detector senses liquid, the sampler begins measuring the sample. 5. After depositing the sample, the pump again reverses for the post-sample purge. Sampling cycles vary somewhat according to program settings for distribution. The sampler can move the distributor arm clockwise and counterclockwise, making a number of distribution methods possible. (In this manual, the words sample event refer to a full sampling cycle for any distribution.) Other variations include extended-program settings for suction-line rinses and sampling retries. A sampler running a program with line rinses completes the pre-sample purge and then rinses the line. The suction line is purged after each rinse. When programmed for sampling retries, the sampler will attempt to pump a sample again if it fails on previous attempts.

2.8.1 A Typical Sampling Cycle

a. Controller. b. Peristaltic pump. c. Liquid Detector. d. Strainer e. Suction line f. Stainless Steel Coupling g. A pump tube routed from the liquid detectors intake port t h r o u g h t h e p u m p, o u t t h e detectors discharge port, and over the side of the controller to a bulkhead fitting at the top of the center section. h. Bulkhead fitting. i. Discharge tube running from the bulkhead fitting on the inside of the center section through the spring and distributor arm. j. Distributor arm and spring.

f e g c b

h a i

j d

Figure 2-10 Identifying the Sampler Components

2-12

6712 Portable Sampler Section 2 Installation/Preparation 2.8.2 Measuring Sample Volume Sample volume refers to the amount of liquid delivered into a bottle. The volume is determined by the programmed value. The volume is dependent on the volume per revolution of the pump, which is dependent on the suction head. The 6712 delivers the sample by counting the pump revolutions and automatically compensating for the suction head. The following paragraphs briefly describe initial sampler installation.

2.9 Installing the Sampler

CAUTION
Tests indicate that the pump produces sound levels in excess of 85db at one meter. Prolonged exposure to this sound could result in hearing loss and requires the use of protective ear plugs. 2.9.1 Icing the Samples Ice is placed in the center of the bottle kit or around the composite bottle. The ice capacity varies according to the bottle kit. To increase cooling for the multiple-bottle kits use an ice-water bath. Place as much ice as possible in the center of the bottle kit and slowly add cold water. Continually add ice, as the water will melt some ice. Always use the retaining rings or hold downs. They hold the bottles in position and keep the bottles from floating. 2.9.2 Tips for Routing Suction Line Route the line so that it runs continuously downhill from the sampler to the liquid source. This helps drain the line during pre-sample and post-sample purges. When the sampler is used in below freezing temperatures, there is a risk of the suction line being frozen. A suitably warm sampling source can usually prevent this, provided there are no loops in the suction line. Some situations may require more protective measures, such as insulation of the suction line, or heat tape. Thoroughly draining the suction line minimizes the possibility of frozen liquid clogging the line. For representative samples, place the intake in the main current of the flow stream, not in an eddy or at the edge of flow. Placing an intake at the bottom may produce samples with excess heavy solids and no floating materials, while placement at the top may produce the opposite conditions. Place the sampler on a relatively flat, horizontal surface. Placing the sampler on a steep incline may cause the sample to miss the bottle mouth. When installing the sampler, be sure the vertical distance between the level of the liquid and the pump is as small as possible.

2.9.3 Intake Placement

2.9.4 Positioning a Sampler

2-13

6712 Portable Sampler Section 2 Installation/Preparation

CAUTION
When using the 6712 sampler with a 750 or 710 module, transmitters such as cell phones or walkie talkies must not be operated within 3 meters of the sampler. If water level readings are fluctuating due to TV/radio station transmitter towers in the area, the sampler must be relocated.

WARNING
The 6712 samplers have not been approved for use in hazardous locations as defined by the National Electrical Code.

WARNING
If you must enter a manhole or other dangerous location to install the sampler, observe standard safety precautions. Refer to Appendix C for a discussion of safety considerations. 2.9.5 Using the Suspension Harness or ProHanger Teledyne Isco has a stainless steel suspension harness to lower or raise the sampler through a manhole or suspend the sampler during operation. You can suspend the harness from any suitable support. Prohanger is an X-shaped bracket that fits inside the lip of a manhole and lets you suspend the sampler using a cable harness. The 6712 sampler can be used in conjunction with several types of external instruments. To connect the sampler to a Teledyne Isco open-channel flow meter, 4100 series flow logger, 2100 series flow module, or 1640 Liquid Level Actuator, attach the sampler-to-flow meter cable to the Flow Meter Connector. To connect the sampler to a closed-pipe flow meter, attach a Teledyne Isco 4-20 mA Input Interface device to the samplers Flow Meter connector. When using the sampler with an external instrument, remember these guidelines: Both the sampler and the flow instrument must be on and running a program. If using a flow meter, logger, or module for flow pacing, the sampler and the flow instrument must both run programs with flow pacing settings. If using a flow meter or flow logger for trigger pacing, the sampler must run a program with flow pacing settings, and the flow instrument must run a program with trigger pacing settings. When the sampler runs an event-paced program, it disregards pacing pulses from a flow meter or flow logger. However, the sampler continues to monitor for enable signals.

2.10 Connecting External Instruments


2.10.1 Flow Meter

Flow Meter Connector Icon

2-14

6712 Portable Sampler Section 2 Installation/Preparation 2.10.2 Rain Gauge Connector Connect the rain gauge, SDI-12 Sonde, or Refrigerator Temperature Sensor connect cables to the nine-pin Rain Gauge Connector. When connecting a rain gauge and an SDI-12 Sonde, or rain gauge and a Refrigerator Temperature Sensor, use the appropriate Y-connect cable. See the Accessories appendix at the end of this manual.

WARNING
Rain Gauge Connector Icon

Most Teledyne Isco flow meters are not certified for use in hazardous locations as defined by the National Electrical Code. Contact your Teledyne Isco representative for your equipments status. Never operate equipment in hazardous locations that has not been certified as safe for such use. Refer to Appendix C in the back of this manual for more safety information

Note
Select Extended Programming to use the SDI-12 sonde. (See Switching Between Standard and Extended Modes on page 4-1.)

Note
Update the Hardware Setup whenever you add or remove an SDI-12 sonde. (See Section 6.) In Extended Programming, rain gauge and SDI-12 sonde readings are only available if they are selected in the Hardware Setup.

2-15

6712 Portable Sampler Section 2 Installation/Preparation

Table 2-4 Connecting Teledyne Isco Instruments to the Sampler


Connector Icon Power Source Connect These Instruments:
12 volt power source

674 Rain Gauge Programmable I/O Pins (pins C, H, and I)

Rain Gauge

SDI-12 Sonde Refrigerator Temperature Sensor 581 RTD (Rapid Transfer Device) IBM PC or compatible computer running FLOWLINK or SAMPLINK

Interrogator

External Modem External Data Logger/Controller 1640 Liquid Level Actuator 3000 Series Flow Transmitters 2100 Series Flow Modules 4100 Series Flow Loggers

Flow Meter

4200 Series Flow Meters Master/Slave Sampler Pulse Duration Input Interface 4-20 mA Input Interface Closed-Pipe Flow Meters (using 4-20 mA Input Interface)

2.11 Locking the Sampler

Teledyne Isco offers stainless steel cables to lock the sampler (standard size use P/N 68-6700-134; compact use 68-6700-111). To lock the sampler, place the hook through the center section latch. Repeat this for the other two latches, and then pull the ends of the cables together over the top of the sampler. Lock the loops with the padlock. You can also lock each latch individually, using three padlocks, one for each of the latches.

2.12 Servicing the Sampler

Servicing the sampler includes: Recovering the sampler from the monitoring site Collecting the filled sample bottles and replacing them with clean bottles Replacing a battery, if required Collecting the data recorded by the sampler during the program.

2-16

6712 Portable Sampler Section 2 Installation/Preparation Samplers are usually serviced after they complete a sampling program. When working up a service schedule, you will need to estimate the program completion time. You should also be familiar with the programs settings for the start time or if the sampler is programmed for continuous sampling. Continuous sampling is controlled by the distribution setting entered at the RUN CONTINUOUSLY? screen. A YES response directs the sampler to reset the distribution sequence and continue the countdown to the next sample event. Appropriate service intervals will prevent overfilled bottles.

2.13 Recovering the Sampler

When recovering the sampler, keep it level to prevent spilling the collected samples. If the batterys charge is questionable, replace it with a fully-charged battery. The 6712 has a fuel gauge that gives an indication of power usage. The controller keeps track of how much power has been consumed since the last time it lost power. The current power consumption, as well as the previous power consumption, can be accessed by pressing the STOP key while in the main menu. Keep in mind that the sampler has no idea as to how much charge was originally stored in the battery (or even if a battery is being used), and therefore has no idea as to the remaining capacity of its power source.

2.14 Collecting Sample Bottles

If you return the entire sampler back to the office, place caps on the bottles and drain any water from the carrier or tub. If you leave the sampler at the site and only need to collect the bottles, follow these guidelines for bottle collection. To collect the bottles from a compact sampler: 1. Open the latches, and lift the center section from the tub. Place caps on the bottles. 2. Remove the bottles by lifting the bottle carrier from the tub. Attach a label to the carrier detailing the location, date, etc.

2.14.1 Compact Sampler Bottle Collection

Removing Bottles From the Compact Sampler

3. Pour any ice water out of the tub.

2-17

6712 Portable Sampler Section 2 Installation/Preparation 4. Slide a new bottle carrier filled with clean bottles into the tub. 5. Pour more ice into the carrier or tub.

2.14.2 Standard Sampler Bottle Collection

To collect the bottles from a standard sampler: 1. Open the latches, and lift the center section from the tub. Place caps on the bottles. 2. Drain any ice water from the standard tub by opening tubs drain plug. Remember to replace the plugs cap when done. 3. Attach a label to the tub or on each bottle detailing the location, time, date, etc. 4. Replace the standard tub and/or bottles with a clean set. 5. Pour more ice into the tub.

Drain Plug Location on the Tub

2-18

6712 Portable Sampler


S ec tion 3 Programming Introduc tion

Before programming the 6712, you should become familiar with its keys and how to use the screens. The 6712 Sampler has two sets of programming modes. The first mode, Standard Programming, lets you set up typical sampling programs easily and efficiently. The second mode, Extended Programming, lets you create sophisticated programs for complex sampling applications. All of the sampling features available in standard programming mode are available in extended programming mode. However, extended programming provides many additional features. Standard and extended programming features are explained in separate sections. This section focuses on some basic features that are common to both programming modes, such as using the menus and setting the clock.

CAUTION
This sampler has a high performance pump. As with all such pumps, it relies upon liquid to cool working components. If the sampler is programmed to pump in the absence of liquid in excess of 5 minutes, excessive heat buildup may damage the paddles, rollers, and housing. Ensure that the liquid inlet is completely immersed.

3.1 Initial Operation

Turn the sampler on by pressing the On/Off key. It is labeled with this icon: . The start-up screen appears first.
6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key.

It remains on the display for about eight seconds or until you press a key. It tells you the type of programming screens the sampler is currently using standard or extended and gives you a tip about the on-line help. The main menu screen, shown below, appears next.

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

(If a module is attached, a different screen may appear first. See the module manual for more information.)

3-1

6712 Portable Sampler Section 3 Programming Introduction The sampler will continue to use the program mode you selected, even if the sampler is turned off and on again, until you manually switch to the other programming mode.
With the display backlight activated, the screen will glow for several seconds following your last keystroke. If you press a key and the screen text does not change, it is probably because the keystroke turned the backlight on, instead of performing its function. Simply press the same key again.

Table 3-1 About the Keypad

Key

Name

Function

On-Off

Turns sampler on or off.

Stop Enter Help Down-Right Arrow Up-Left Arrow Numbers Decimal Point Pump Reverse Pump Forward

Stops the pump or distributor. Pauses a running sampling program. In programming screens, returns to a previous screen. Accepts a menu choice or number entry and goes to next screen. In programming screens, displays a brief help message. Selects the menu option right or below the current choice. Selects the menu option left or above the current choice. Types a number. Types a decimal point. Press when at the main menu to run the pump in reverse. Press when at the main menu to run the pump forward.

3-2

6712 Portable Sampler Section 3 Programming Introduction The sampler is programmed at the factory so that the standard programming screens appear when you first turn the sampler on. You can switch between programming modes by typing in a numerical command at the main menu: At the main menu type 6712.2 to show the extended programming screens. At the main menu type 6712.1 to show the standard programming screens. 3.1.1 Selecting a Language You have the option of having screens display in English or another language. The default is English. At the main menu, type 6712.8:
ENGLISH SPANISH

Use the arrow keys to move between the choices (languages other than those shown above could appear on your unit). When the language choice you want is blinking, press (Enter). The (Enter) key

always accepts the blinking option.

3.2 Using Menus and Entering Numbers


The Main Menu
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

A menu is a list of options. The main menu has four options: RUN PROGRAM VIEW REPORT OTHER FUNCTIONS The options identify the operations that can be performed from the menu. For example, to run the sampling program, select RUN; to modify the program, select PROGRAM. In menu screens, one menu option always blinks. Press when the blinking option is the one you want; it will accept your choice and go to the next screen. Select a different option by pressing an arrow key until the option you want blinks. Then press .

3.2.1 Selecting Menu Options

3.2.2 Entering Numbers

Number-entry screens prompt you to enter a value. An example is shown to the left. To enter a number: Press the number keys to type the number. Then, press . As soon as you press , the sampler saves your number and moves to the next screen. In some screens, you can use the . (Decimal Point) key in a number. Some screens display the range of acceptable numbers between parentheses. If an entry is too low or high, the controller beeps and erases the entry. Type a new number and continue.

SUCTION LINE LENGTH IS __ ft (3-99)

Type the length of the suction line. Acceptable values are shown between the parentheses. Press to accept the number.

3-3

6712 Portable Sampler Section 3 Programming Introduction

3.3 Quick View Screens

Quick view screens are a special type of menu screen. They show the current program settings and let you move quickly through the program. You must change the programming style to QUICK VIEW/CHANGE to see the quick view screens, as shown below:

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

Select OTHER FUNCTIONS.

Select PROGRAMMING STYLE.

PROGRAMMING STYLE NORMAL QUICK VIEW/CHANGE

Select QUICK VIEW/CHANGE.

3.3.1 Paging Through Quick View Screens

The arrows in the corners of each quick view screen are menu options that let you move from one quick view screen to another. Select the up arrow ( ) to go to the previous screen. Select the down arrow ( ) to go to the next screen. Press Stop to return to the main menu. Because the forward arrow is always blinking when the screen appears, you can simply press to go to the next quick view screen, making it easy to page through the screens.

3.3.2 Changing Settings in a Quick View Screen

Although the quick view screens offer you a quick way to see the program settings, they also provide you with a way to change settings. Using quick view screens to change settings is sometimes a faster way to change a program because you can go quickly to the setting or settings that needs updating. To change the program settings in a quick view screen: Press an arrow until the setting blinks. Press . The 6712 then displays the screen used to change the setting. The Quick View screens below shows you how to change the sample volume. Begin by paging through the quick view screens until you see the screen containing the sample volume setting 200 ml SAMPLES. When you change a setting, the sampler stores the new settings and returns to the updated quick view screen. You can repeat this process until all settings shown in the quick view screens have been changed. When you are done, select the screens forward arrow to move to the next quick view screen, or press Stop to return to the main menu.

3-4

6712 Portable Sampler Section 3 Programming Introduction

Quick View Screens in Standard Programming


RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

200 ml SAMPLES

To see the Quick View screens, select PROGRAM.

SITE DESCRIPTION: SITE 29

Press the arrow key until 200 ml SAMPLES blinks. Then press .

SAMPLE VOLUME: 250 ml (10-1000)

Use the keypads arrow key to select for the next quick view screen. When is blinking, press . Use the keypads arrow key to select for the previous quick view screen. When is blinking, press . Use the keypads arrow key to select the program setting. When the setting is blinking, press to display the screen used to change the setting. Press Stop for the main menu. All quick view screens work the same way.

Type the new sample volume and press .

250 ml SAMPLES

3.4 Clock and Calendar

You may have to set the internal clock or calendar. The samplers are shipped with their clocks set to U.S. Central Standard time. If your sampler operates in a different time zone, you will have to reset the clock. The example to the left shows you how to find the time and date screen. When setting the clock and calendar, use a 24-hour clock for times and the day-month-year format for dates. The illustration at the left shows blanks for each position that accepts an entry. The positions are called fields. When this screen appears on your sampler, the fields will contain the samplers current settings for the time and date. To move from one field to another without changing the setting, press the arrow keys. Use this technique when you want to change only one or two settings. Change the setting by typing a new number. Press to accept the new setting. For example, to enter 2:00 p.m. (2:00 p.m is 14:00 on a 24-hour clock), type 14. Press . Next, type 0 (zero) for the minutes, and press . To enter a date, such as January 22, 2001, type: 22 01 01 .

Entering Times and Dates


ENTER TIME AND DATE: HH:MM DD-MON-YY __:__ __-__-__

Move from one field to another by pressing the arrow keys. Change a setting in a field by typing a number. Press to accept the new setting.

3-5

6712 Portable Sampler Section 3 Programming Introduction 3.4.1 Setting the Clock and Calendar
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

Select OTHER FUNCTIONS.

Select MAINTENANCE.

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS ENTER TIME AND DATE: HH:MM DD-MON-YY 14:00 22-JAN-01

Select SET CLOCK.

Enter the time and date.

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Press Stop to return to the main menu.

3.4.2 Menu Screens: Clock Start Times

There are four start time settings you can use in a sampling program. This section discusses only one of these settings, the clock start time. A clock start time lets you set up a sampling program that starts at the same time on one or more days of the week. For example, you may want to begin sampling at 6:00 a.m. on Monday through Friday (refer to the screen below). To select a day or days from the menu: 1. Press an arrow key until the cursor is on the correct day. Then, press . 2. Repeat these steps until each day you want is blinking. 3. Press an arrow until DONE blinks. Press . To remove days from the list: 1. Press an arrow key until the cursor is on the day you want to remove. Press to deselect it. 2. When finished, select DONE and press

3-6

6712 Portable Sampler Section 3 Programming Introduction

NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL

Using the Clock Start Time Menu

Select CLOCK TIME.


FIRST SAMPLE AT: 06:00

Type the time of day and press .


SELECT DAYS: SU MO TU WE TH FR SA DONE

Press an arrow key until the cursor is on the correct day. Then, press . To remove a day, move the cursor to it and press to deselect it. Repeat these steps until each day you want is blinking. Press an arrow until DONE blinks. Press .

3.5 Menu Screens: Site Descriptions and Program Names

The sampler has two text-entry menus, one for site description and another for an extended-program name. They work in the same way. A site description is commonly a number, address, or other short note that helps identify the monitoring site. Extended programming has a similar screen that lets you enter program names for stored programs. The example on the next page shows how to change the site description from FACTORY01 to SITE 29. Line 1 contains text between two quotation marks. Lines 2 and 3 contain the menu options numbers, letters, and punctuation marks used to spell out the description. In line 3, the space between the ampersand (&) and the double quotes (") is a space character. Line 4 contains two additional options: BACK-UP, an option that moves the cursor on the text line to the left one character. DONE, an option that tells the sampler to save the text. The cursor is the blinking rectangle that identifies the current field. When the screen first appears, the cursor blinks on the first character of the text in line 1. The matching character in line 2 or 3 also blinks. When you press , the cursor on line 1 moves to the right one position, and, at the same time, the sampler blinks on the matching character on lines 2 or 3.

3-7

6712 Portable Sampler Section 3 Programming Introduction To change any character in the text line: 1. Press the left or right arrow until the replacement character on lines 2 or 3 blinks. Changing the Site Name
SITE: FACTORY01 ABCDEFGHIJKLMNOPQRST UVWXYZ-& 0123456789 BACK-UP DONE

2. Press . The replacement character will appear on the first line, and the cursor will move to the next character. 3. Repeat steps 1 and 2 until the text in the top line is complete. 4. Finish the entry by selecting DONE. Then, press . Some additional tips for using the text-entry menus:
SITE: SITE 29_ ABCDEFGHIJKLMNOPQRST UVWXYZ-&_"0123456789 BACK-UP DONE

SITE: SACTORY01 ABCDEFGHIJKLMNOPQRST UVWXYZ-& 0123456789 BACK-UP DONE

Press the Down-Right arrow until the S on line 2 blinks. Then, press .
SITE: SICTORY051 ABCDEFGHIJKLMNOPQRST UVWXYZ-& 0123456789 BACK-UP DONE

Press 29 on the keypad, then Up-Left until space blinks. Press .


SITE: SITE 29 ABCDEFGHIJKLMNOPQRST UVWXYZ-&_"0123456789 BACK-UP DONE

Press the Up-Left arrow until the I blinks. Press . Continue pressing arrow keys to select T, E, and space. Press after selecting each character.

Press the Down-Right arrow until DONE blinks. Press .

Hold an arrow key down for faster movement. If your description uses all ten spaces in the text line, the sampler automatically moves to the next screen as soon as you add the tenth character. You can enter numbers by pressing the number keys. Erase characters by replacing them with a space character. Select the BACK-UP option to move the cursor to the left. To leave the menu before completing the entry, press Stop. The sampler will display this menu:
SAVE CHANGES? YES NO

Select YES to save any changes made. Select NO to leave the original description (or program name) unchanged.

3-8

6712 Portable Sampler Section 3 Programming Introduction

3.6 Messages

The sampler has four sets of messages that supplement the programming screens. Information messages tell you about programming status. Help Notes are brief comments associated with programming screens describing the setting or offering assistance. Warning messages tell you something is different tha Teledyne Iscos standard. Operation messages report the samplers status as it runs a program. These are brief messages about the programming status, such as ROM PASSED TEST or the status message that displays when you first start up the sampler. Information messages are displayed for four seconds, unless they are terminated by pressing any key on the keypad. To see a Help note: 1. Press the ? (Help) key. 2. When the note requires several screens, the word more appears in the lower-right corner. Press for the next note. 3. Move back and forth between screens by pressing the arrow keys. 4. Press Stop at any time to return to the programming screen. Most help notes refer to a related section in this manual. Screens that require extensive explanation have notes that contain only a reference to the manual. Sections mentioned in the notes appear in the index and the table of contents.

3.6.1 Information Messages

3.6.2 Help Notes

3-9

6712 Portable Sampler Section 3 Programming Introduction

SUCTION LINE LENGTH IS 7 ft (3-99)

Press ? (Help).

Using Help Notes

Enter the length of suction line only, Although this example shows the excluding strainer and pump tube. <more> help notes that are associated Press or Down-Right to see the next screen. Press Stop to return to the programming screen. See Installing Suction Line in your manual. <more> Press or Down-Right to see the next screen. Press Up-Left to see the previous screen. Press Stop to return to the programming screen. Press the red STOP key to exit HELP. Press Up-Right to see the previous notes. Press Stop to return to the programming screen.

with the suction-line length screen, you can use the same technique to see help notes for other screens.

3.6.3 Warning Messages

Warning messages appear when the sampler determines something is out of the ordinary. For example, the following warning appears when you type a number that is larger than the standard bottle volume:

BOTTLE VOLUME IS 3500 ml (300-30000)

WARNING! STANDARD BOTTLE VOLUME EXCEEDED!

The 6712 does accept nonstandard volumes because you might use a nonstandard container occasionally. For a list of recommended volumes for Teledyne Isco bottles press the ? (Help) key at the Bottle Volume Is screen, or refer to the Technical Specifications.

3-10

6712 Portable Sampler Section 3 Programming Introduction The 6712 uses the bottle and sample-volume settings to determine the maximum number of samples that can be deposited without overfilling the bottles. Entering a volume that exceeds the standard volume may cause the sampler to overfill the bottle. 3.6.4 Operation Messages An operation message reports the samplers status as it runs a program. An example is shown below.

PROGRAM WILL START AT 06:00 TU 20-FEB 05:56:22 TU 20-FEB

3.7 Menu Trees

Selecting a menu option will take you to a number-entry screen or another menu screen. The screens are organized in a branching structure that resembles a tree. Refer to Figure 3-1 for a simple chart of the menu tree for standard programming. The standard programming structure is different much simpler than the extended programming structure. Charts that show you most of the samplers screens appear in Appendix A, Flowcharts. The entire structure changes a little when a module is attached because the sampler adds a set of screens for a module as soon as it detects the modules presence. The manuals for each module contain menu trees for its programming screens.

3-11

6712 Portable Sampler Section 3 Programming Introduction

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

See Running Programs on page 4-10

See the screen charts in Appendix A.

VIEW DATA PRINT DATA SYSTEM IDs CONFIGURE REPORTS

See Reports on page 4-19.

MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

SET CLOCK SET PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

See Figure A-8, Manual Functions Screens.

PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE

Figure 3-1 Menu Tree for Standard Programming

3-12

6712 Portable Sampler


S ec tion 4 S tandard Programming

The 6712 Sampler has two sets of programming screens. The first set, called standard programming screens, lets you set up typical sampling programs easily and efficiently. The second set, extended programming screens, lets you create sophisticated programs for complex sampling applications. All of the sampling features available within the standard programming screens are available in extended programming. However, extended programming provides several additional features which are discussed in the next section. Menu flowcharts for both sets of programming screens are in Appendix A.

4.1 Switching Between Standard and Extended Modes

Factory default settings cause the sampler to begin in standard programming mode the first time you turn it on. The start-up screen tells you which programming mode the sampler is currently using. You can switch between programming modes by entering a numerical command at the main menu: At the main menu, type 6712.1 to enter standard programming mode. At the main menu, type 6712.2 to enter extended programming mode.
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS Main Menu

4.2 Language Selection, Units of Length

You have the option of displaying screens in English or Spanish (or possibly another language). The default is English. At the main menu, type 6712.8 to display:
ENGLISH SPANISH

Use the arrow key to move between the two choices. When the desired choice is blinking, press . If you select the English option, you will advance to a second screen that asks you to select your units of length:
SELECT UNITS FOR LENGTH: ft m

4-1

6712 Portable Sampler Section 4 Standard Programming Use the arrow key to move between the two choices, feet or meters. When the desired choice is blinking, press . The default unit of length for the Spanish option is meters.

4.3 Programming Examples

The following table is an example of time-paced sampling. Further details about some of the menu items are provided in later sections.

Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle
1
6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS SITE DESCRIPTION: FACTORY051 CHANGE? YES NO

Turn the sampler on by pressing the On/Off key. Press . This screen disappears on its own after 8 seconds. The option PROGRAM will be blinking. Press .

The option NO will be blinking. Press . For the purposes of this example, it isnt necessary to change the description.

NUMBER OF BOTTLES: 1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key until the correct number blinks. Press . For this example, select 24.

BOTTLE VOLUME IS 1000 ml (300-30000)

Type the volume for the bottles in your kit. For this example, 1000 is correct, so simply press .

SUCTION LINE LENGTH IS 10 ft (3-99)

Type the length of the suction line, then press . If you change the length, the sampler will display a message, PLEASE WAIT!. . . GENERATING PUMP TABLES. Because this example requires samples every 15 minutes, select TIME PACED by pressing an arrow until the option TIME PACED blinks. Then, press . Type 0 for hours and press . Type 15 for minutes and press . Tip: Move back and forth between hours and minutes by pressing an arrow key.

TIME PACED FLOW PACED

TIME BETWEEN SAMPLE EVENTS 0 HOURS, 15 MINUTES

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOTTLE

Because this program requires one sample in each bottle, select SEQUENTIAL by pressing an arrow until the option SEQUENTIAL blinks. Then, press . For this example, select NO by pressing an arrow until the option blinks. Then, press . Selecting YES allows the program to run indefinitely by repeating the sample distribution. Continuous sampling assumes that filled bottles are replaced with empty bottles at regular service intervals.

10

RUN CONTINUOUSLY? YES NO

4-2

6712 Portable Sampler Section 4 Standard Programming

Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle (Continued)
11
SAMPLE VOLUME: 200 ml (10-1000)

Type the volume of the sample you want deposited in each bottle. Then, press .

12

NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL FIRST SAMPLE AFTER A 5 MINUTE DELAY (1-999) PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

For this example, select DELAYED START by pressing an arrow until the option blinks. Then, press .

13

Type the delay period you want between the time you run the program and the time the sampler takes the first sample. Then, press .

14

Run the program immediately by selecting YES. Select NO if you want to run the program later by selecting RUN from the main menu. Press after making your choice. In this example, NO is selected. Run the program by selecting RUN and pressing .

15

The following table is an example of flow-paced sampling. Program type: Standard Site description: SITE 29 Bottle kit: 24, 1000 milliliter bottles Suction line length: 7 feet Pacing: Flow pacing, every two pulses Distribution: 2 bottles per sample Sample Volume: 250 milliliters Start time: Clock Time, 6:00 a.m. on Monday, Wednesday, Friday Module: No module installed

Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample
1
6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS SITE DESCRIPTION: FACTORY051 CHANGE? YES NO

Select PROGRAM.

Select NO.

4-3

6712 Portable Sampler Section 4 Standard Programming

Edited by Foxit Reader Copyright(C) by Foxit Software Company,2005-2008 For Evaluation Only.

Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample (Continued)
4
NUMBER OF BOTTLES: 1 2 4 8 12 24

Select 24.

BOTTLE VOLUME IS 1000 ml (300-30000)

Enter 1000.

SUCTION LINE LENGTH IS 10 ft (3-99)

Enter 7.

TIME PACED FLOW PACED

Select FLOW PACED.

FLOW BETWEEN SAMPLE EVENTS: 2 PULSES (1-9999)

Enter the number of pulses between sample events.

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOTTLE

Select BOTTLES/SAMPLE.

10

RUN CONTINUOUSLY? YES NO

For this example, select NO by pressing an arrow until the option blinks. Then, press . Selecting YES allows the program to run indefinitely by repeating the sample distribution. Continuous sampling assumes that filled bottles are replaced with empty bottles at regular service intervals. Enter 2.

11

1 BOTTLES PER SAMPLE EVENT (1-24)

12

NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL START FLOW COUNT AT: 00:00

Select CLOCK TIME.

13

Enter 6, then enter 0. This is the time at which the sampler will start the pacing countdown. Select MO, WE, and FR. Use the arrow keys and to select the days. When the correct days are blinking, select DONE and press . Select NO.

14

SELECT DAYS: SU MO TU WE TH FR SA DONE

15

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO

4-4

Edited by Foxit Reader Copyright(C) by Foxit Software Company,2005-2008 For Evaluation Only.
6712 Portable Sampler Section 4 Standard Programming

4.4 Pacing
Standard Time Pacing and Flow Pacing Screens for Samplers without Flow Modules
TIME PACED FLOW PACED

Sample pacing is the rate at which the sampler takes samples. Depending on the type of pacing you select, the rate is controlled by the samplers internal clock or by inputs received from connected instruments. Standard programming provides time pacing and flow pacing (refer to Step 7 in the previous two programming examples). In time-paced sampling, the interval between samples is a constant time interval. When you program the sampler for time pacing, the sampler prompts you to enter the time between sample events in hours and minutes. Time paced programs always take a sample at the start time. Flow paced sampling requires a flow meter, flow logger, or a module. A flow meter or flow logger paces a sampler by sending an electronic signal to the sampler after measuring a specified volume of liquid. Because each pulse represents a volume interval, flow pacing rates are proportional to the volume of water flowing through the channel. When you program the sampler for flow pacing and are using a flow meter or flow logger, the sampler prompts you to enter the interval between sample events in pulses. The sampler initiates a sample event when the set number of pulses is received. Programs that are flow paced do not take a sample at the start time. After setting the program start time, you will be prompted to enter a maximum run time of zero to 999 hours. If you would like the program to run indefinitely, enter 0 (zero) for maximum run time.

When programming, you will be asked to select time or flow paced sampling. Depending upon your choice, you will see one of the following screens:
TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES

FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999)

Flow Pacing Screen for Samplers with Flow Modules


FLOW BETWEEN SAMPLE EVENTS: ____ Mgal (0.001 -99999)

The flow pacing screens change when you attach a flow module (see example to the left). Because the sampler is more closely integrated with the modules, the standard flow pacing screen prompts you for the flow volume between sample events instead of pulses between events. The flow volume units displayed are what you had previously programmed. Teledyne Isco 4200 Series Flow Meters and 4100 Series Flow Loggers send pulses for trigger pacing. The flow meter or flow logger sends pulses at two different time intervals when trigger pacing a sampler. Trigger pacing, for example, lets you sample less frequently when the trigger condition a condition based on level, flow, or rainfall remains below a set point. The intervals are determined by the flow instruments program settings for trigger pacing. When conditions change and readings pass the set point, triggering the new rate, you can sample more frequently. Or, when the trigger condition is a clock setting, the instrument changes the pacing interval according to the time of day. Using a clock trigger condition, for instance, the flow meter or flow logger can pace the sampler slowly at night and more frequently during the day. (For more information about trigger pacing, see your flow meter or FLOWLINK manual.)

4.4.1 Trigger Pacing

4-5

6712 Portable Sampler Section 4 Standard Programming The flow meter or flow logger sends the same pulse for both flow and trigger pacing, and the sampler cannot distinguish between them. Therefore, to program a sampler for trigger pacing, simply program it for flow pacing.

4.5 Distribution

Distribution describes how the sampler is to deposit samples. A sample is the volume of liquid deposited in a bottle. A sample event includes the full sampling cycle and may deposit a sample into more than one bottle. In standard programming, you can program the sampler for these different distribution methods: Sequential Bottles Per Sample Samples Per Bottle Composite

4.5.1 Sequential

In sequential distribution, the sampler deposits one sample in each bottle. A sequential sample represents a snapshot of the flow stream at a point in time. In bottles per sample distribution, the sampler deposits a sample in each of a set of bottles. A bottle set includes at least two bottles but may include all bottles. Use bottles per sample when the volume to be collected is larger than the amount one bottle can hold or when you need identical samples. In samples per bottle distribution, the sampler deposits samples from several sample events in a single bottle before moving to the next bottle. Use samples per bottle distribution to collect a series of small composite samples. For single bottle configurations, samples per bottle distribution is known as composite sampling. A composite sample represents an average of the flow streams characteristics during the sampling period. Sample programs can be run indefinitely by selecting YES at the RUN CONTINUOUSLY? screen. Continuous sampling resets the distribution when the distribution sequence is complete. That is, when the last bottle/set is reached, the next bottle/set is the first bottle/set. All pacing modes except RANDOM are supported. When running a program continuously, the bottles must be serviced at regular intervals to prevent overfilling the bottles. The sampler assumes that the next bottle/set is empty and ready to receive samples. The figure on the next page steps through the programming screens for sequential, bottles per sample, and samples per bottle distribution.

4.5.2 Bottles Per Sample

4.5.3 Samples Per Bottle

4.5.4 Composite

4.5.5 Continuous Sampling

4-6

Edited by Foxit Reader Copyright(C) by Foxit Software Company,2005-2008 For Evaluation Only.
6712 Portable Sampler Section 4 Standard Programming

Sequential

STANDARD PROGRAMMING

DISTRIBUTION: SEQUENTIAL

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOT TLE

RUN CONTINUOUSLY? YES NO

Each bottle receives one sample from one sample event.

Bottles-per-Sample

STANDARD PROGRAMMING

DISTRIBUTION: 2 BOTTLES/SAMPLE

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOT TLE

2 BOTTLES PER SAMPLES EVENT (1-24)

RUN CONTINUOUSLY? YES NO

Multiple bottles receive a sample from one sample event.

Samples-per-Bottle

2 1

4 3

6 5

8 7

STANDARD PROGRAMMING

DISTRIBUTION: 2 SAMPLES/BOT TLE

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOT TLE

2 SAMPLES PER BOTTLE (1-50)

RUN CONTINUOUSLY? YES NO

Each bottle receives a sample from multiple sample events.

Figure 4-1 Sample Distribution

4-7

6712 Portable Sampler Section 4 Standard Programming

4.6 Start Times

It is important to understand the difference between the time at which you run a program and the programs start time. Running a program simply means selecting RUN from the samplers main menu. The start time is the time at which the sampler begins the programs first sample interval countdown. The start time is controlled by your selections from the start time screens. Each program contains start-time settings that tell the sampler when to begin the program. When programming the sampler, you can select one of four start time options: NO DELAY TO START, DELAYED START, CLOCK TIME, or WAIT FOR PHONE CALL. Select NO DELAY TO START when you want the sampler to start as soon as you select RUN. Select DELAYED START when you want the sampler to delay from 1 to 999 minutes before starting the program. Select CLOCK TIME when you want the sampler to begin the program at a specific time on at least one day of the week. Select WAIT FOR PHONE CALL when you want the sampler to begin the program after receiving a remote command to start (optional modem required).

4.6.1 How Do Start Times Work?

As soon as you select RUN from the main menu, the sampler checks the programs start time settings. With a DELAYED START setting, the sampler starts a countdown to the start time. The period between the time you select RUN and the start time youve specified in the program is called the delay to start time. When you run a program with CLOCK TIME settings, the sampler also starts a countdown to start time until the day of week and clock times are met. The sampler may not collect a sample at the start time. The sampler must be enabled, and the pacing selected must call for a sample at start. In any event, the sampler checks the pacing settings and begins the pacing-interval countdown.

4-8

6712 Portable Sampler Section 4 Standard Programming

Delayed Start Screen


1 NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL 1

Clock Start Time Menu


NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL

Select DELAYED START.


FIRST SAMPLE AFTER A ___MINUTE DELAY (1-999) Enter the number of minutes to delay the countdown to the first sample. START FLOW COUNT AFTER A __MINUTE DELAY (1-999) The sampler uses this screen for flow paced programs. 2

Select CLOCK TIME.


FIRST SAMPLE AT 06:00

Enter the time of day.


SELECT DAYS: SU MO TU WE TH FR SA DONE

Press an arrow key until the cursor is on the correct day. Press. Repeat until each day you want is blinking. Press an arrow until DONE blinks. Press .

4.6.2 Sampler Enable/Disable

Teledyne Isco flow meters and flow loggers have a programmable sampler-enable feature that lets them enable (start) or disable (stop) a program according to certain monitored conditions. For example these conditions can be level, flow rate, pH, temperature, percent, rainfall, I/O; or a combination of two conditions. The 1640 Liquid Level Actuator can also be used, as well as any other instrument that can ground the input to pin F. Grounding the input disables the sampler. The sampler disregards disable signals during the delay to start time. However, when the sampler detects a disable signal at the start time, it suspends the program until it is enabled. The diagram in Figure 4-2 shows the samplers responses.

4-9

6712 Portable Sampler Section 4 Standard Programming

Sampler enabled at start time A sample event at the start time occurs for time-paced programs.

Sampler disabled at start time by signal from pin F A sample is not taken at start time if the sampler is disabled. No samples are taken while disabled.

Figure 4-2 Start Time Diagram

4.7 Running Programs

To run a program, select RUN from the main menu. You may also select YES at the RUN THIS PROGRAM NOW? screen at the end of the programming screens. While running a sampling program the sampler displays a variety of messages that report the programs status. For example, if the sampler has not yet reached the programmed start time, it displays the scheduled start time as well as the current time. Once the program reaches the start time, you can determine the time of the next sample, the next bottle to receive a sample, sample distribution, and other information. Other messages appear while the sampler runs through a sampling cycle and takes a sample. When the sampler needs to report multiple messages, it alternates them, displaying each for one to three seconds.

4.7.1 Run Time Screens

4-10

6712 Portable Sampler Section 4 Standard Programming

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO

If RUN had been previously Stopped and no programming changes were made. Multiple Bottles ENTER START BOTTLE: __ (1MAX)

OR
For the first run using a 24-bottle base. PLEASE WAIT! ...TESTING DISTRIBUTOR SYSTEM

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

If applicable FATAL ERROR NO DISTRIBUTOR ARM!

If applicable PROGRAM WILL START AT 06:00 MO 17OCT Single Bottle 05:52:22 MO 17OCT OPERATION ERROR DISTRIBUTOR JAMMED

PURGING SUCTION LINE

PUMPING 250 ml RUN DISPLAYS

RINSING SUCTION LINE

SAMPLE 2 IN 00:14:29

Figure 4-3 Running a Program 4.7.2 Module Readings Samplers with an attached module display the modules readings: pH and temperature for the pH module; level, percentage, or flow rate for the 4-20 mA module; level, flow rate, and flow volume readings for the flow modules; level, velocity, flow rate, and positive flow volume readings for the area velocity module. Some module screens will alternate with the sampler screens.

BOTTLE 2 IN 00:14:39 150 gpm 0.82 ft 00001438200 gal

The Module readings are updated: Every second when displaying the reading during programming. Every minute otherwise.

4-11

6712 Portable Sampler Section 4 Standard Programming An * (asterisk) appears next to the reading if the module was unable to take a reading. If an asterisk appears, the reading displayed is the last available reading. Flow rates will be reported as zero flow if the error persists for more than five minutes (for a level or velocity reading that flow is calculated from).
BOTTLE 2 IN 00:12:00 ***** gpm 5.73 ft 00001457600 gal

If an entire reading is filled with asterisks, the value exceeds the number of characters that can be displayed. The sampler stores the actual reading in memory, so it may be collected later. For those programs that have delayed or scheduled start times, the readings will be displayed while waiting for the start time. Keep in mind that any totalizer values will be reset at the start time. Module and sonde readings are also displayed after a program is complete, until a key is pressed. It is not necessary to run a program to view module or sonde readings. Real time data can be viewed on the display by selecting VIEW REPORT from the main menu.

4.8 Interrupting a Running Program

You can interrupt a sampling program by pressing the Stop key while the sampler is waiting for the next sample event. Pressing Stop places the sampler into Manual Paused operation and records a manual pause in the sample event log. In this state, the sampler allows you to access several functions. You can modify the program, reset alarms, take a grab sample, etc., without adversely affecting the running program. Modifying the program is limited to adjusting the pacing intervals, enable conditions and sample volume. The basic program structure, such as the type of pacing, can not be changed. While in the manual paused state, the sampling program continues to operate as normal, with the exception of taking samples. If a sample was to be taken, it is skipped. The sampler records sample skipped in the sample event log and continues to operate as normal. The manual paused state displays a scrolling menu with several options (see example below). Use the Arrow keys to scroll through the manual paused options and the key to make a selection. Return to the running program by selecting resume program or pressing the Stop key. If you do not make a selection, the sampler automatically resumes the original sampling program five minutes after the last key was pressed. STOP PROGRAM
RESUME PROGRAM VIEW DATA GRAB SAMPLE Manual Paused Screen

4.8.1 Stop Program

This option terminates the running program and records program stopped in the sample event log. The program cannot be resumed.

4-12

6712 Portable Sampler Section 4 Standard Programming 4.8.2 Resume Program Selecting this option will cause the sampler to exit the manual paused state and return to normal program operation. A manual resume event will be logged. This option allows you to view the data recorded by the sampler. See Reports on page 4-19 for more information. When selected, you will be asked to enter the volume to be pumped. A sample will then be taken as if it is to be placed into a container outside the sampler base. Do this by removing the pump tubing from the bulkhead fitting. See Grab Samples on page 4-15. A grab sample event will be logged. If the pump tube is changed, this option should be selected to reset the pump counts on the counter. This will remove the WARNING: REPLACE PUMP TUBING message that may appear while the program is running. A pump tube replaced message is logged. See Replacing The Pump Tube in Section 7, Maintenance. When selected, you will be asked to enter the volume to be pumped. A sample will then be taken as if it is to be placed into a container outside the sampler base. Do this by removing the pump tubing from the bulkhead fitting. After the sample has been pumped, you will then be asked to enter the amount actually delivered. Once you have entered the delivered volume and accepted the calibration, a volume delivered event will be logged. 4.8.7 Cal/Adj Parameters When the sampler is configured for operation with a module, this option becomes available. Selecting this option will display the appropriate level adjustment screens and/or the calibration screens. If the level is adjusted, an adjusted level event is logged. If a parameter is calibrated, an appropriate event is logged. Calibrating a parameter probe will temporarily turn off the partition data storage and the sample enable/disable functions. These functions are disabled during the calibration and for five minutes after the program is resumed. During this time, parameter data normally collected at the data storage interval will be logged as a 252 error message. 4.8.8 Adjust Pacing This option is available when the running program is paced by Time, Flow Pulses or Flow Volume. If you select this option, you are asked to enter a new pacing interval. The sampler logs an interval changed event. The original pacing interval continues to count down until you return the sampler to the running program (select resume program). If the new pacing interval is less than the original count remaining, the new interval is used.

4.8.3 View Data

4.8.4 Grab Sample

4.8.5 Pump Tube Alarm

4.8.6 Calibrate Volume

4-13

6712 Portable Sampler Section 4 Standard Programming If the original count remaining is less than the new pacing interval, the original count will continue to count down to the next sample event. Subsequent samples are then paced by the new interval. 4.8.9 Adjust Volume Select this option to change the sample volume within the limits of the currently programmed sample distribution. If you change a sample volume, a volume changed event is logged. The 6712 has a fuel gauge that gives an indication of power usage. The controller keeps track of how much power has been consumed since the last time it lost power. If the controller is powered by an external battery, this fuel gauge can help you estimate the condition of the battery. Keep in mind that the sampler has no idea as to how much charge was originally stored in the battery (or even if a battery is being used), and therefore has no idea as to the remaining capacity of its power source.

4.8.10 Power Used

4.9 Other Functions


Other Functions Menu
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

In Standard Programming, the Other Functions menu screen includes options for: Maintenance (see Section 8). Manual Functions (see below). Programming Style (see Section 3).

Select OTHER FUNCTIONS.


MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

4.10 Manual Functions

The manual functions programming screens let you: Take grab samples. Calibrate sample volumes. Operate the pump manually. Move the distributor arm.

4-14

6712 Portable Sampler Section 4 Standard Programming

4.11 Grab Samples


Taking a Grab Sample
GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR

Grab samples let you take a single sample on demand, collecting the sample in an external container (Figure 4-4). Keep in mind: When the sampler delivers a grab sample, it runs through a complete sampling cycle, using the current settings for volume and for line rinses or retries. If it is a two-part program, the sampler uses the settings for part A. To take a grab sample while the sampler is running a program, you must interrupt the program by pressing the Stop key. Restart the program by running it again. To take a grab sample: 1. Disconnect the pump tube from the bulkhead fitting. 2. Place the end of the tube over a collection container. 3. Follow the steps in Taking a Grab Sample at left. 4. Reconnect the pump tube to the bulkhead fitting.

Select GRAB SAMPLE


SAMPLE VOLUME: ___ ml (10-9990)

Enter the desired volume


GRAB SAMPLE

PRESS WHEN READY!

PURGING SUCTION LINE

PUMPING 200 ml

PURGING SUCTION LINE

GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR

Press STOP to return to main menu.

Figure 4-4 Taking a Grab Sample

4-15

6712 Portable Sampler Section 4 Standard Programming

4.12 Calibrate Volume

The sampler delivers accurate sample volumes without calibration. If you find that sample volumes vary significantly from the programmed values, first check the suction line for proper installation. Be sure it slopes continuously downhill to the liquid source and drains completely after each sampling cycle. Then, compare the actual length of the suction line to the suction line length settings in the program to see that they match. Also check the pump tube for excessive wear and replace it if necessary. You may want to calibrate when: A new pump tube is installed. Run the pump for five minutes before calibrating. The sample source is above the sampler. Sampling from pressurized lines (15 PSI Maximum).

4.12.1 Calibration Tips

For the best results, calibrate the sampler after its installed on site. Use a graduated cylinder for volume measurement. The sampler clears the previous calibration setting when you reinitialize the sampler. You cannot calibrate while the sampler is running a program. When the sampler delivers the sample, it runs through a complete sampling cycle, using the current settings for volume line rinses and retries. If it is a two part program, the sampler uses the volume setting for part A. To calibrate: 1. Disconnect the pump tube from the bulkhead fitting. 2. Place the end of the tube over a collection container. 3. Follow the steps in Calibrating Sample Volumes in the margin. 4. Reconnect the pump tube to the bulkhead fitting.

4-16

6712 Portable Sampler Section 4 Standard Programming Calibrating Sample Volumes 1


RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

VOLUME DELIVERED: ____ml

Select OTHER FUNCTIONS

Measure the volume delivered and enter the amount here.

MAINTENANCE MANUAL FUNCTIONS MODULE HARDWARE

10

200 ml ARE YOU SURE? YES NO This screen appears when the amount you entered in step 9 was more than twice, or less than half, the programmed volume. Select YES when the volume delivered matches the volume displayed on this screen. Select NO to enter the volume delivered.

Select MANUAL FUNCTIONS

GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR

Select CALIBRATE VOLUME

SAMPLE VOLUME: ___ ml (100-9990)

Enter the desired volume


CALIBRATE VOLUME

PRESS WHEN READY!

PURGING SUCTION LINE

PUMPING 200 ml

PURGING SUCTION LINE

4.13 Operating the Pump

You can manually operate the samplers pump in the forward or reverse direction. You can do this to test the pump, draw a large sample, or purge the suction line. When you run the pump manually, the sampler reports the number of pump counts during the pumps operation. Use this feature to obtain an estimate of the pump counts required to purge a nonstandard suction line. You can also manually run the pump from the main menu. Press the 1 key and then to run the pump in reverse. Press the 3 key and then to run the pump forward.

4-17

6712 Portable Sampler Section 4 Standard Programming Operating the Pump


GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR

CAUTION
Tests indicate that the pump produces sound levels in excess of 85db at one meter. PROLONGED EXPOSURE TO THIS SOUND LEVEL COULD RESULT IN HEARING LOSS AND REQUIRES THE USE OF PROTECTIVE EAR PLUGS.

SELECT DIRECTION PUMP FORWARD PUMP REVERSE

PURGING. . . PRESS THE RED STOP KEY WHEN DONE

PURGED _____COUNTS

4.14 Moving the Distributor Arm

The sampler lets you reposition the distributor arm. Use this feature to verify a bottle location when installing a bottle kit. You cannot move the distributor arm while the sampler is running a program.

CAUTION
DO NOT rotate the distributor arm manually; this will severely damage the distributor drive gears.

Moving the Distributor Arm


GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR

Select MOVE DISTRIBUTOR


GO TO BOTTLE ____ (1-MAX)

Enter the bottle number to which you want to move the distributor
NOW AT BOTTLE ____

4-18

6712 Portable Sampler Section 4 Standard Programming

4.15 Reports

The 6712 records a variety of data while running a sampling program. It uses the data to produce four reports: The Program Settings report, listing the program settings. The Sampling Results report, listing the program settings, time of samples and other program events. The Combined Results report, combining the sample event times with rainfall or module data. The Summary report, listing daily summaries of data collected. The Summary report may be rainfall and/or module data. The sampler stores the data for each report in memory where they remain until you select RUN. Selecting RUN clears the memory so that it can store the data from the next program.

Interrogator

4.15.1 Collecting Reports

There are a number of ways to collect the reports: View reports (program settings, sampling results, and abbreviated summary reports) on the samplers display. Collect the reports (except for summary reports) with a Teledyne Isco 581 RTD and use a computer running flowlink or samplink to transfer the reports from the RTD to a file on the computer. Collect the reports (except for summary reports) with an IBM PC or compatible computer running flowlink.

581 RTD

Collect the reports with and IBM PC or compatible computer running samplink. Because the display area is small, the data on the reports will be displayed a little differently from the data that is transferred to and viewed on your computer. For the Sampling Report, the sampler displays each program event, one at a time. For the Module and Rainfall Data reports, it displays daily summaries, instead of the full reports available with the rtd, Flowlink, or Samplink. If you select Sampling Report or Rainfall (step 3 in Viewing Data, shown in the margin), the sampler begins displaying the report data. The sampler advances automatically through the report items, displaying each item briefly. While the sampler advances automatically through the displays: Stop the automatic displays by pressing Stop once. Then, use the arrow keys to move manually through the report. Return to the main menu by pressing Stop twice. At the end of the report, the sampler leaves the last item displayed until you press: the arrow keys to move forward or backward manually through the report. Stop to return to the main menu.

4.15.2 Viewing the Data

4-19

6712 Portable Sampler Section 4 Standard Programming to start the automatic displays again. Viewing Data 1
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

Select VIEW REPORT.


VIEW DATA SYSTEM IDs CONFIGURE REPORTS

If you select Module Data in step 3, the sampler displays the screen shown in step 4. Select the Daily Summary option and the sampler displays the stored data just as the Sampling Report was described above. Selecting the Current Readings option causes the sampler to display real-time readings. When there are multiple screens, they will be displayed alternately. You can quickly advance through the screens by pressing an arrow key or . Pressing Stop exits the Current Readings mode.

Select VIEW DATA.


SELECT DATA TO VIEW: SAMPLING REPORT MODULE DATA RAINFALL DATA

Select a report or data type.


VIEW: DAILY SUMMARY CURRENT READINGS

Select DAILY SUMMARY to view the stored data, or CURRENT READINGS to view real-time data.

4.15.3 Configuring Reports Configuring Reports 1


RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

The Configure Reports option lets you specify which of the Sampling Reports will be included in any output request. Each sampler is shipped with the option Flowlink Report selected. The Samplink Report and Flowlink Report options allow you to tailor the output to the Teledyne Isco software used to collect the reports. The Samplink Report option provides the Sampling Reports, but disables the output of the detailed partition data. Flowlink Report provides the Sampling Reports and the detailed partition data. Selecting the Flowlink Report option enables the output of the detailed partition data so Flowlink can add the data to the sites database file. When you are familiar with the available Sampling Reports, you may want to specify an output containing only the specific data you require. The Custom Report option will allow selection of detailed data. Data collection time and storage space may be optimized using this option.

Select VIEW REPORT.


VIEW DATA SYSTEM IDs CONFIGURE REPORTS

Select CONFIGURE REPORTS


SAMPLINK REPORT FLOWLINK REPORT CUSTOM REPORT ALL REPORTS

Select the option that matches your data collection software, or tailor the output according to the reports needed.

4-20

6712 Portable Sampler Section 4 Standard Programming 4.15.4 Sampling Results Report As the sampler runs a program, it records the programs events. Program events include such items as sample events, program enables and disables, power losses, and so on. Table 4-3 lists each event with a short description. You can view the report by following the steps in Viewing the Data on page 4-19 or print the log as the Sampling Results report. In the text file version of the report, sample-event entries include the time of the event as well as the sample and bottle numbers. The sample column contains the sample number deposited into the bottle. This information reflects the distribution selected for the program. The letters in the Source columns are codes for the cause of the event. The letters in the Error column are codes for the causes of missed samples. The source and error codes appear with explanations at the end of the report. A list of codes appears in Table 4-3. The last column in the Sampling Results report records the pump counts for the sample event. Large variations in pump counts from event to event indicate fluctuating heads, a relocated suction line, or a temporarily clogged strainer.

4.15.5 Combined Results Report

At the time of each sample event, the sampler records the readings from each connected module, rain gauge, or SDI-12 sonde. The Combined Results Report presents the event time, sample and bottle number, and the respective readings from the device. For flow modules, the readings are level and flow rate. For the pH module, the readings are temperature and pH. For the rain gauge and SDI-12 Sonde, the readings vary according to the selections enabled during the hardware setup in extended programming. When recording four or more sonde parameters, the Combined Results report will exceed the standard 40 character width. To read the additional columns, the data must be collected with Flowlink 4 for Windows software.

4.15.6 Summary Report

This report summarizes the rain gauge or module data. The types of readings are the same as those available to the Combined reports. The sampler compiles the data to present a daily total, the days average, maximum, and minimum reading, and a listing of hourly averages. The sampler then graphs the days readings.

4-21

6712 Portable Sampler Section 4 Standard Programming

Table 4-3 Reports: Program Events, Source Codes, and Error Codes
* Indicates events and codes that occur in Standard Programming

Program Events
*Program Run *Program Started *Program Enabled (Disabled) A/B Enabled (Disabled) Sample *Sample Switch Time *Program Done A/B Done *Manual Pause *Program Stopped *Grab Sample Pump Tube Replaced Calibrate Volume Calibrate SDI Parameter *Adjusted Pacing *Volume Changed Adjust Enable Reset Rain History Adjust Suction Head *Manual Resume Program Paused (Resumed) A/B Paused (Resumed) *Power Failed *Power Restored Alarm Occurred Alarm Acknowledged *Missed Password

(listed in the order that they most likely would occur)

The sampler ran the sampling program. The sampler started the sampling program. The sampler was enabled or disabled according to program settings or an external instrument. Program part A or B was enabled or disabled according to the programs settings. A sample was taken as programmed. (See Source Codes and Error Codes in this table.) A sample was taken at programmed switch time. The sampler finished the sampling program. The sampler finished program part A or B. Stop was pressed during the program to enter the Manual Paused state. You selected Stop from the manual paused options, terminating the program. A grab sample was taken while in the Manual Paused state. The pump tube alarm was reset while in the Manual Paused state. The volume was calibrated. An SDI-12 sonde or module parameter was calibrated. The pacing interval was changed while in the Manual Paused state. Sample volume was changed while in the Manual Paused state. Sampler enable thresholds were changed while in the Manual Paused state. The rain history was reset. The suction head was changed while in the Manual Paused state. The resume program option was selected from the Manual Paused state. The program reached a programmed pause or resume time. Program part A or B reached a programmed pause or resume time. The power source was disconnected. The power source was reconnected. An alarm occurred. An alarm was acknowledged. The password was missed.

4-22

6712 Portable Sampler Section 4 Standard Programming

Table 4-3 Reports: Program Events, Source Codes, and Error Codes (Continued)
(The Source and Error Codes Listed below refer to why a sample was taken and what happened while it was being taken.)

Source Codes
*C *D *E Ev *F *G M PH R *S Sw TO *T Calibrate sample Disable Enable Event Flow Grab sample Command driven sample Phone command Resume Start Switch time Time Override time Sample volume delivered during calibration. Sample event at disable time. Sample event at enable time. Event-paced sample event. Flow-paced sample event. Sample volume delivered as a grab sample. A command driven sample. Phone command sample. Sample event at resume time. Sample event at start time. Sample event at switch time. Sample event at bottle-switch time. Time-paced sample event.

Error Codes
*DJ Distributor Jammed *IP *L sample in progress pump latch open The distributor jammed. Sample in progress when report was printing. Unable to take sample because the pump band was opened. There is no distributor arm or the stop is damaged. The sampler was unable to detect liquid. Not enough flow for the minimum sample volume entered for samples with volumes proportional to flow. After the sampler detected liquid and while the sample was being taken, the liquid detector stopped detecting liquid. The composite bottle is full. The sampler was turned off with the On-Off key during the sample event. The sampler was directed to take a sample that would not fit into the bottle. Occurs for flow paced, time override programs only. The power supply failed during the sample event. Note: Power failures during five consecutive sample events results in a LOW BATTERY shut-down. The pump jammed. Sample not taken. You pressed the Stop key during the sample event.

*ND no distributor arm *NL No Liquid Detected LF Not enough flow

*NM No More Liquid M *O Ov *P Bottle Full Sampler Shut Off Probable Overflow Power Failed

*PJ Pump Jammed *Sk Sample Skipped *US User Stopped

Figures 4-5, 4-6, 4-7, 4-8, and 4-9 show examples of reports configured to show program settings, sampling results, combined results, and a summary report. 4-23

6712 Portable Sampler Section 4 Standard Programming

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 1.02 *********** PROGRAM SETTINGS *********** ---------SITE DESCRIPTION: "FACTORY051" ---------UNITS SELECTED: FLOW RATE: cfs FLOW VOLUME: Mgal ---------BUBBLER MODULE: WEIR 90 V-NOTCH ---------24, 1000 ml BTLS 10 ft SUCTION LINE ---------PACING: TIME, EVERY 0 HOURS, 15 MINUTES ---------DISTRIBUTION: SEQUENTIAL

---------200 ml SAMPLES

---------5 MINUTE DELAY TO FIRST SAMPLE ---------------------------------------Figure 4-5 Report: Program Settings

4-24

6712 Portable Sampler Section 4 Standard Programming

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 1.02 *********** SAMPLING RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml COUNT TO SAMPLE BOTTLE TIME SOURCE ERROR LIQUID ------- ------ ----- ------15:03 PGM ENABLED 1,1 1 15:03 S 250 1,1 2 15:18 T 247 1,1 3 15:33 T 247 1,1 4 15:48 T 249 1,1 5 16:03 T 247 1,1 6 16:18 T 247 1,1 7 16:33 T 247 1,1 8 16:48 T 248 1,1 9 17:03 T 237 1,1 10 17:18 T 236 1,1 11 17:33 T 237 1,1 12 17:48 T 241 1,1 13 18:03 T 238 1,1 14 18:18 T 236 1,1 15 18:33 T 237 1,1 16 18:48 T 236 1,1 17 19:03 T 242 1,1 18 19:18 T 237 1,1 19 19:33 T 235 1,1 20 19:48 T 238 1,1 21 20:03 T 237 1,1 22 20:18 T 237 1,1 23 20:33 T 236 1,1 24 20:48 T 230 20:48 PGM DONE 19-DEC

SOURCE S ==> START SOURCE T ==> TIME ----------------------------------------

Figure 4-6 Report: Sampling Results

4-25

6712 Portable Sampler Section 4 Standard Programming

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 1.02 BUBBLER MODULE: 2215419583 Hardware: A00000 Software: 1.07 *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml FLOW TOTAL LEVEL RATE FLOW SAMPLE BOTTLE TIME ft cfs Mgal ------- ------ ---- ----- ----- ------------1,1 1 15:03 0.009 2.495 000000.000 1,1 2 15:18 0.012 2.576 000000.017 1,1 3 15:33 0.012 2.666 000000.035 1,1 4 15:48 0.012 2.773 000000.053 1,1 5 16:03 0.012 2.773 000000.072 1,1 6 16:18 0.012 2.798 000000.091 1,1 7 16:33 0.012 2.798 000000.109 1,1 8 16:48 0.012 2.798 000000.128 1,1 9 17:03 0.012 2.751 000000.147 1,1 10 17:18 0.012 2.773 000000.166 1,1 11 17:33 0.009 2.798 000000.185 1,1 12 17:48 0.012 3.728 000000.207 1,1 13 18:03 0.012 3.728 000000.232 1,1 14 18:18 0.012 3.728 000000.257 1,1 15 18:33 0.009 3.728 000000.282 1,1 16 18:48 0.012 5.273 000000.317 1,1 17 19:03 0.012 5.239 000000.353 1,1 18 19:18 0.012 6.651 000000.397 1,1 19 19:33 0.009 6.651 000000.442 1,1 20 19:48 0.009 6.651 000000.486 1,1 21 20:03 0.012 6.651 000000.531 1,1 22 20:18 0.012 6.651 000000.576 1,1 23 20:33 0.009 3.404 000000.612 1,1 24 20:48 0.012 3.425 000000.634 -----------------------------------------------

Figure 4-7 Report: Combined Results

4-26

6712 Portable Sampler Section 4 Standard Programming

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 1.02 *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml FR-TEMP SAMPLE BOTTLE TIME C ------- ------ ------NO FR-TEMPERATURE ----------------------------------------

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 0.01.0127 *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml SAMPLE BOTTLE TIME ------- ------ ----------- ------ ---NO RAIN GAUGE ----------------------------------------

SAMPLER ID# 3687447734 06:32 19-DEC-02 Hardware: A0 Software: 0.01.0127 SDI-12 DATA *********** COMBINED RESULTS *********** SITE: FACTORY051 Program Started at 15:03 WE 18-DEC-02 Nominal Sample Volume = 200 ml

NO SDI-12 SONDE ----------------------------------------

Figure 4-8 Report: Combined Results (continued)

4-27

6712 Portable Sampler Section 4 Standard Programming

SAMPLER ID# 3687447734 16:41 20-DEC-02 Hardware: A0 Software: 1.02 BUBBLER MODULE: 2215419583 Hardware: A00000 Software: 1.07 SITE: FACTORY051 Summary Report for 19-DEC-02 (TH) Day's Flow: 000000.054 Mgal Average Flow Rate: 0.4437 cfs 07:20 Minimum Flow Rate: 0.0237 cfs 13:25 Maximum Flow Rate: 7.0207 cfs Hourly Average Flow Rate: 00:00-01:00: NO DATA 01:00-02:00: NO DATA 02:00-03:00: NO DATA 03:00-04:00: NO DATA 04:00-05:00: NO DATA 05:00-06:00: NO DATA 06:00-07:00: NO DATA 07:00-08:00: NO DATA 08:00-09:00: 2.519 cfs 09:00-10:00: 2.747 cfs 10:00-11:00: 2.791 cfs 11:00-12:00: 3.456 cfs 12:00-13:00: 5.385 cfs 13:00-14:00: 6.651 cfs 14:00-15:00: 2.591 cfs 15:00-16:00: 1.593 cfs 16:00-17:00: NO DATA 17:00-18:00: NO DATA 18:00-19:00: NO DATA 19:00-20:00: NO DATA 20:00-21:00: NO DATA 21:00-22:00: NO DATA 22:00-23:00: NO DATA 23:00-24:00: NO DATA *---+---+---+---+---+---+* + + I I I I I # I I # I 6.00 + # + I # I I ## I I ## I I ## I 4.00 + ## + I ### I I ### I I ##### I I ####### I 2.00 + ####### + I ####### I I ######## I I ######## I I ######## I 0.00 + ######## + *---+---+---+---+---+---+* Hour Ending: 08: 16: 24: Units are 'cfs' 8.00

Figure 4-9 Report: Summary

4-28

6712 Portable Sampler Section 4 Standard Programming

4.16 System IDs


System IDs
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

In addition to viewing report data and configuring reports, the View Report option on the main menu lets you view system identification information. The sampler reports the sampler model number, ID (identification number), and software version number as system identification notes. The sampler also reports the same information for an attached module. The model and ID numbers remain unchanged through the life of the sampler. The ID numbers appear on the samplers reports and in files created by flowlink and samplink when they interrogate the sampler. The software packages use the ID number to make sure that data from different samplers is not combined in the same file. The software version number changes when your samplers software is updated. The system will also list any software options that you have activated on your unit. These software options are discussed in Section 5.

Select VIEW REPORT.


VIEW DATA SYSTEM IDs CONFIGURE REPORTS

Select SYSTEM IDs.


SAMPLER 6712 ID __________ HARDWARE: __ SOFTWARE: __

Displayed briefly

[MODULE TYPE] [MODEL #] ID: __________ HARDWARE: __ SOFTWARE: __

Displayed briefly
SOFTWARE OPTIONS [Options List]

Press STOP to return to the main menu.

4.17 Programming for 700 Series Modules

The bay on the controllers side accepts any of Teledyne Iscos 700 Series Modules. These modules are optional accessories that are not required for operation of the 6712. However, the modules offer an economical way to combine flow-rate or parameter monitoring with sampling. The programming of these modules is integrated with the programming of the sampler, so attachment of a module will cause different screens to be displayed, depending on the type of module and the selections you make while programming. Two modules that are commonly used are the 730 Bubbler Module, which monitors a flow streams level and flow rate, and the 750 Area Velocity Module for monitoring a flow streams level, velocity, and flow rate. The following sets of screens give you an example of what screens you might see while programming the sampler and module. Refer to Table 4-1 (which does not have a module attached) at the beginning of this section to compare the sequence of screen displays.

4-29

6712 Portable Sampler Section 4 Standard Programming

CAUTION
Most Teledyne Isco accessories and interfacing equipment are not certified for us in hazardous locations as defined by the National Electrical Code. Check applicable guidelines, codes, and regulations of federal, state, city, and county agencies. Refer to Appendix C in the back of this manual for more safety information.

Table 4-4 Programming Example With 730 Module Installed


MODULE INSERTED--DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

With the sampler off, insert the module in the bay on the controller. Turn the sampler on by pressing the On/Off key. Press . Download stored data, if you want to keep it. Pressing DONE erases it and advances to the next screen.

6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key.

This screen will now appear for 8 seconds.

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

The option PROGRAM will be blinking. Press .

SITE DESCRIPTION: FACTORY051 CHANGE? YES NO

The option NO will be blinking. Press . For the purposes of this example, it isnt necessary to change the description.

SELECT UNITS FOR FLOW RATE: cfs gps gpm Mgd lps m3s m3h m3d

Select the desired units for the flow rate. For this example, choose cfs. Press .

SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit

Select the desired units for the flow volume. For this example, choose Mgal. Press .

PROGRAM MODULE? YES NO

Select YES to program the module.

4-30

6712 Portable Sampler Section 4 Standard Programming

Table 4-4 Programming Example With 730 Module Installed (Continued)


8
MODE OF OPERATION: FLOWMETER LEVEL ONLY

If you choose Flowmeter, proceed to screen #9 below. If you choose LEVEL ONLY, proceed to screen #12.

WEIR FLUME DATA POINTS EQUATION MANNING FLOW-INSERT

When you select Flowmeter, you will have several options for performing the flow conversion. For this example, choose Weir. Press .

10

V-NOTCH RECTANGULAR CIPOLLETTI

For this example, choose V-Notch. Press .

11

V-NOTCH WEIR ANGLE: 120 90 60 45 30 22.5

For this example, choose a Weir angle of 90. Press .

12

CURRENT LEVEL IS __.__ ft ADJUST LEVEL TO __.__ ft

The current level will be displayed. Enter an adjustment if needed. If the adjustment is different by more than 0.50 ft, you will be prompted ARE YOU SURE? YES NO Choose NO to re-enter; select YES if correct. Press .

13

NUMBER OF BOTTLES: 1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key until the correct number blinks. Press . For this example, select 24.

14

BOTTLE VOLUME IS 1000 ml (300-30000)

Type the volume for the bottles in your kit. For this example, 1000 is correct, so simply press .

15

SUCTION LINE LENGTH IS 10 ft (3-99)

Type the length of the suction line, then press . If you change the length, the sampler will display a message, PLEASE WAIT!. . . GENERATING PUMP TABLES.

16

TIME PACED FLOW PACED

Select FLOW PACED. If the module is being used as a flowmeter, you will see screen #17a. If the module is being used for level only, you will see screen #17b.

a 17

FLOW BETWEEN SAMPLE EVENTS: 0.075 Mgal (0.001 - 99999)

If the module is being used as a flowmeter, choose the amount of flow between sample events.

4-31

6712 Portable Sampler Section 4 Standard Programming

Table 4-4 Programming Example With 730 Module Installed (Continued)


b 17 FLOW BETWEEN SAMPLE EVENTS: 2 PULSES (1-9999)

If the module is being used for level only, enter the number of pulses between sample events.

18

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOTTLE

Select BOTTLES/SAMPLE.

19

2 BOTTLES PER SAMPLE EVENT (1-24)

Enter 2.

20

RUN CONTINUOUSLY? YES NO

Select NO.

21

SAMPLE VOLUME: 250 ml (10-1000)

Enter 250.

22

NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL

Select CLOCK TIME.

23

START FLOW COUNT AT: 06:00

Enter 6, then enter 0. This is the time at which the sampler will start the pacing countdown.

SELECT DAYS: SU MO TU WE TH FR SA

24

DONE

Select MO, WE, and FR. Use the arrow keys and to select MO, WE, and FR. When the correct days are blinking, select DONE and press .

25

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO

Select NO.

4-32

6712 Portable Sampler Section 4 Standard Programming

Table 4-5 Programming Example With 750 Module Installed


MODULE INSERTED--DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

With the sampler off, insert the module in the bay on the controller. Turn the sampler on by pressing the On/Off key. Press . Download stored data, if you want to keep it. Pressing DONE erases it and advances to the next screen. This screen will now appear for 8 seconds.

6712 SAMPLER STANDARD PROGRAMMING For HELP at any screen press ? key. RUN PROGRAM VIEW REPORT OTHER FUNCTIONS SITE DESCRIPTION: FACTORY051 CHANGE? YES NO SELECT UNITS FOR FLOW RATE: cfs gps gpm Mgd lps m3s m3h m3d

The option PROGRAM will be blinking. Press .

The option NO will be blinking. Press . For the purposes of this example, it isnt necessary to change the description.

Select the desired units for the flow rate. For this example, choose cfs. Press .

SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit

Select the desired units for the flow volume. For this example, choose Mgal. Press .

SELECT UNITS FOR VELOCITY: fps mps

Select the desired units for the flow velocity. For this example, choose fps. Press .

PROGRAM MODULE? YES NO

Select YES to program the module.

MODE OF OPERATION: FLOWMETER LEVEL AND VELOCITY

The 750 can have two different modes of operation: Flowmeter or Level and Velocity. For this example, choose Flowmeter. Press . When you select Flowmeter, you will have two options for performing the flow conversion: Area * Velocity or Level to Flow Rate. For this example, choose Area * Velocity. Press . For this example, choose Channel Shape. Press .

10

FLOW CALCULATION: AREA * VELOCITY LEVEL TO FLOW RATE

11

LEVEL TO AREA: CHANNEL SHAPE DATA POINTS ROUND PIPE U-CHANNEL RECTANGULAR CHANNEL TRAPEZOIDAL CHANNEL

For the channel shape, select Round Pipe. Press .

12

4-33

6712 Portable Sampler Section 4 Standard Programming

Table 4-5 Programming Example With 750 Module Installed (Continued)


ROUND PIPE

Enter the diameter of the round pipe. Press .

13

DIAMETER = 2.00 ft

14

CURRENT LEVEL IS __.__ ft ADJUST LEVEL TO __.__ ft

The current level will be displayed. Enter an adjustment if needed. If the adjustment is different by more than 0.50 ft, you will be prompted ARE YOU SURE? YES NO Choose NO to re-enter; select YES if correct. Press . Enter the amount of zero level offset. Press .

15

ZERO LEVEL OFFSET: _.___ ft

16

MAXIMUM DEPTH FOR VELOCITY MEASUREMENT 2 3 4

(This screen is applicable for standard-sized probes only.) Select the maximum depth for the velocity measurement. Press .

17

NUMBER OF BOTTLES: 1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key until the correct number blinks. Press . From this point forward, the screens will be the same as in Table 4-2.

4-34

6712 Portable Sampler


S ec tion 5 Extended Programming

The 6712 Sampler has two sets of programming screens. The first set, called standard programming screens, lets you set up typical sampling programs easily and efficiently. The second set, extended programming screens, lets you create sophisticated programs for complex sampling applications. All of the sampling features available with standard programming screens, discussed in Section 4, are available with extended programming screens. However, the extended screens provide several additional features and some screen display differences which are discussed in this section. The menu flowcharts for both sets of programming screens are in Appendix A, Menu Flowcharts.

Note
This section assumes that you have read, and are familiar with, the instructions in Section 4, Standard Programming!

5.1 Extended and Standard Mode

The sampler is shipped with factory default settings in standard programming mode when you first turn the sampler on. The start-up screen tells you which programming mode the sampler is currently using. You can switch between programming modes by entering a numerical command at the main menu: At the main menu type 6712.2 to enter extended programming mode. At the main menu type 6712.1 to enter standard programming mode.
RUN EXTENDED 1 PROGRAM VIEW REPORT OTHER FUNCTIONS Main Menu for Extended Mode

5.2 One-Part and Two-Part Programs

Extended programming lets you set up a one-part program or a two-part program. One-part programs let you fill all bottles of the tub with one set of pacing, distribution, and enable settings. Two-part programs add an additional set of extended pacing, distribution, and enable screens to the sampling program. Each set of screens is called a program part. The program parts are simply called part A and part B. Both parts share the program settings for suction-line length, suction head, and rinses and retries. They also share one start-time setting.

5-1

6712 Portable Sampler Section 5 Extended Programming Each part has its own group of bottles. Because settings for each part are independent of each other, the sampler, when running a two-part program, fills each bottle group as if they were being filled by two different programs. Two-part programs finish when both parts are done. Two-part programming is ideal for sampling storm events. Many storm-water run-off monitoring permits require a first-flush grab sample within the first 30 minutes of the storm event and flow-paced samples during the remainder of the event. With a two-part program, you can set up part A for the first-flush sample and part B for the remainder. Table 5-2 shows a storm-water sampling program.

5.3 Storage for Extended Programs

The sampler stores five sampling programs: one standard and four extended. Program storage eliminates the need to reprogram the sampler with frequently used settings. This feature is especially useful for complex extended programs such as two-part programs or programs with many nonuniform-time settings. Teledyne Isco ships the sampler to you with factory programs that you can modify to suit your own needs. All four extended programs are identical to each other. The stored programs will also save your 700 Series Module program settings and any sampler enable conditions using rain gauge data or SDI-12 sonde parameters. These program settings remain as long as you do not change the type of module, or change the rain gauge or SDI-12 Hardware Setup.

5.3.1 Selecting a Stored Program

When the sampler is in extended programming mode, it always has one current and three noncurrent programs, each occupying a permanent location in memory. The first time you access the extended mode, the current program is EXTENDED 1, one of the factory extended programs. The name of the current extended program appears in quotations on the first line of the main menu. The factory programs are named EXTENDED 1, EXTENDED 2, EXTENDED 3, and EXTENDED 4." You can rename them with more descriptive names, making them easier to identify. The program name is one of the extended programs settings. After selecting a stored program, review the settings with the quick view screens. Each sampling program stores all the settings required by the sampler to run a program. These settings include the bottle-kit information (number of bottles and the bottle volume), the suction-line length, the site description, start times, and so on. If you use the same program at two sites each requiring different bottle configurations, suction line lengths, or other details always check these settings before running a program. You may need to make minor changes to the program to make sure the settings match your equipment.

5-2

6712 Portable Sampler Section 5 Extended Programming The current program is the program you see when you select PROGRAM from the main menu and page through the programming screens. It is the program the sampler uses when you select RUN from the main menu. You cannot delete programs, but you can modify them as often as necessary. To run a program with different settings without losing the settings in your current program, select a stored program and modify its settings. For example, assume you have been running the current program, named EXTENDED 1, at site A for several weeks. This program contains all the settings you need for site A, and you dont want to lose those settings. Nevertheless, you need another program because you want to move the sampler to site B, so you select EXTENDED 2 and modify it for site B. The settings in EXTENDED 1 are preserved automatically. When you return the sampler to site A, you wont need to reprogram the sampler. Merely select EXTENDED 1 again, double-check the program settings, and run the program. To select an extended program, follow the steps in Table 5-1.

Table 5-1 Selecting a Stored Extended Program


1
6712 SAMPLER, EXTENDED PROGRAMMING For HELP at any screen, press ? key.

Press the On/Off key to turn the sampler on. Press to clear the start-up message. If the words STANDARD PROGRAMMING appear on the second line of the start-up message, the sampler is using the standard programming screens. To see the extended screens, type 6712.2 at the main menu shown in step 2. Select PROGRAM from the main menu.

RUN EXTENDED 1 PROGRAM VIEW REPORT OTHER FUNCTIONS PROGRAM NAME: "EXTENDED 1" SITE DESCRIPTION: "______"

Select PROGRAM NAME: "EXTENDED 1" from the quick view menu.

SELECT NEW PROGRAM CHANGE PROGRAM NAME

Select SELECT NEW PROGRAM to change the program. You can go through QUICK VIEW screens quickly by pressing at each screen. You can also change settings in quick view screens; see Changing Settings in a Quick View Screen on page 3-4. Select the name of the extended program you want to use. Each programs name appears on the same line. If you inadvertently give two programs identical names, you can still identify them by their location on the screen. To run the program immediately, select YES. To run the program later, select NO.

EXTENDED EXTENDED EXTENDED EXTENDED

1 2 3 4

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO RUN EXTENDED 2 PROGRAM VIEW REPORT OTHER FUNCTIONS

5-3

6712 Portable Sampler Section 5 Extended Programming

5.4 Programming for 700 series Modules

The bay on the controllers side accepts any of Teledyne Iscos 700 Series Modules. These optional accessories are not required to operate the 6712. However, the modules offer an economical way to combine flow-rate or parameter monitoring with sampling. The programming of these modules is integrated with the programming of the sampler, so attachment of a module will cause different screens to be displayed, depending on the type of module and the selections you make while programming. Two modules that are commonly used are the 730 Bubbler Module, which monitors a flow streams level and flow rate, and the 750 Area Velocity Module for monitoring a flow streams level, velocity, and flow rate. Section 4 contains two examples of programming for modules. Module programming is the same in Extended as it was in Standard programming.

5.5 Programming SDI-12 Sondes

As an option, the 6712 accepts up to 16 parameters from up to ten sensors with SDI-12 addresses from 0 - 9. The data parameters used by the sonde and sampler must match. Certain vendors sondes, such as Hydrolab and YSI, have enhanced commands that facilitate plug and play setup. These Teledyne Isco Ready sondes can tell the 6712 what values they have, their order, and units of measure. Other sondes can be used, but require manual setup to identify proper data types for each data value reported. For more information about programming SDI-12 sondes, refer to Section 6.

5.6 Programming Examples

In addition to the features available with standard programming, and discussed in Section 4, extended programming offers: Nonuniform time pacing Random interval pacing Event pacing Multiple bottle compositing Time switched bottles or bottle sets Programmable sampler enable Pauses and resumes Two-part programming Automatic or user-defined suction head Suction line rinses Sampling retries This manual provides several programming examples. Some, demonstrating simple tasks, appear in the margins. Others, demonstrating more complex tasks, appear as examples. Table 5-2 on the following pages steps you through a storm water runoff sampling using extended programming.

5-4

6712 Portable Sampler Section 5 Extended Programming 5.6.1 Storm Water Runoff Sampling Storm water runoff sampling is an ideal application for two-part programming. Typical monitoring permits require that a series of timed samples be taken during the initial portion of the storm event, followed by flow paced samples during the remainder of the event. This example assumes that a Rain Gauge and a Bubbler Flow Module are connected to the sampler.

Note
To enable the Bubbler Module programming screens, insert the module before turning on the sampler Using a 24-bottle kit, it divides the bottles into two groups, with six bottles assigned to part A and 18 to part B. The sampler will begin taking samples after being enabled according to programmable enable controls; that is, when the rain gauge detects 0.15 inches of rainfall in 30 minutes, and the module detects a flow stream level of more than 0.25 feet. Because both program parts use the same enable settings, the sampler will enable both parts at the same time. A list of settings for the program follow: Program type: Extended. Program name: STORM. Site description: SITE 54. Units: Length Ft, Flow Rate gpm, Flow Volume gal Mode of Operation: Flowmeter, Flow-Insert, Round, 12" Bottle kit: 24, 1000-milliliter bottles. Suction-line length: 5 feet. Suction head: Auto suction head. Line rinses: None. Sampling retries: None. Part A and B bottle assignments: Bottles 1 - 6 to part A, bottles 7 - 24 to part B. Part A pacing: Time pacing, sampling every 5 minutes. Part A distribution: Sequential, 1 sample per bottle. Part A sample volume: 800 milliliters. Part A enable: At least 0.15 inches of rainfall in 30 minutes and a flow stream level of more than 0.25 feet; once enabled, stay enabled, sample taken when the sampler is enabled. Part A pause and resumes: None. Part B pacing: Flow pacing, sampling every 2,500 gallons. Part B distribution: Sequential, 1 sample per bottle. Part B sample volume: 500 milliliters. Part B enable: At least 0.15 inches of rainfall in 30 minutes and a flow stream level of more than 0.25 feet;

5-5

6712 Portable Sampler Section 5 Extended Programming once enabled, stay enabled, no sample taken when the sampler is enabled; first sample taken 20 minutes after the sampler is enabled. Part B pause and resumes: None. Start time: Delayed start of 1 minute. 730 Bubbler Module: Installed

Table 5-2 Storm Water Runoff Sampling


6712 SAMPLER EXTENDED PROGRAMMING For HELP at any screen press ? key. RUN EXTENDED 1 PROGRAM VIEW REPORT OTHER FUNCTIONS PROGRAM NAME: EXTENDED 1 CHANGE? YES NO

This screen will appear briefly.

Select PROGRAM. Note: If the start-up screen displayed STANDARD PROGRAMMING, type 6712.2 at this menu before selecting PROGRAM. Select YES. Note: If the quick view screen shows up, press STOP then select OTHER FUNCTIONS, SOFTWARE OPTIONS, QUICKVIEW/CHANGE, NORMAL. Then press STOP and go back to step 2. Select CHANGE PROGRAM NAME.

SELECT NEW PROGRAM CHANGE PROGRAM NAME

NAME: STORM ABCDEFGHIJKLMNOPQRST UVWXYZ-& 0123456789 BACK-UP DONE SITE: DESCRIPTION FACTORY051 CHANGE? YES NO SITE: SITE 54 ABCDEFGHIJKLMNOPQRST UVWXYZ-& 0123456789 BACK-UP DONE

Enter the word STORM for the program name.

Select YES.

Enter SITE 54 for the site description.

SELECT UNITS FOR LENGTH: ft m SELECT UNITS FOR FLOW RATE cfs gps gpm Mgd lps m3s m3h m3d SELECT UNITS FOR FLOW VOLUME cf gal Mgal m3 lit

Select ft.

Select gpm.

Select gal.

10

5-6

6712 Portable Sampler Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)


Select YES

11

PROGRAM MODULE? YES NO

12

MODE OF OPERATION FLOWMETER LEVEL ONLY

Select FLOWMETER

13

WEIR FLUME DATA POINTS EQUATION MANNING FLOW-INSERT WEIR/ORIFICE TYPE V-NOTCH ROUND

Select FLOW-INSERT.

Select ROUND.

14

15

FLOW INSERT SIZE: 6" 8" 10" 12"

Select 12.

16

NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE CURRENT LEVEL IS ___.___ ft.

Select DONE. Note: Partition memory status message will be displayed briefly.

Press ENTER.

17

ADJUST LEVEL TO ___.___ ft. DATA STORAGE INTERVAL IN MINUTES 1 2 5 10 15 30

18

Select 5. (If the data storage interval is different from the current program, you will be informed that the storage interval has changed, and that the stored data will be lost. Download stored data (if desired) and select DONE. Select 24.

19

NUMBER OF BOTTLES: 1 2 4 8 12 24

20

BOTTLE VOLUME IS 1000 ml (300-30000)

Enter 1000.

21

SUCTION LINE LENGTH IS 5 ft (3-99)

Enter 5. The 6712 calculates new pump tables.

22

AUTO SUCTION HEAD ENTER HEAD

Select AUTO SUCTION HEAD.

5-7

6712 Portable Sampler Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)


Enter 0.

23

0 RINSE CYCLES (0-3)

24

RETRY UP TO 0 TIMES WHEN SAMPLING (0-3)

Enter 0.

25

ONE-PART PROGRAM TWO-PART PROGRAM

Select TWO-PART PROGRAM.

26

24 BOTTLES AVAILABLE ASSIGN BOTTLES 1 THRU 6 TO PART A (1-23) UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME TIME BETWEEN SAMPLE EVENTS: 0 HOURS, 5 MINUTES

Enter 6. Screen will say "Beginning Part A" before proceeding with screen 27.

Select UNIFORM TIME PACED.

27

Enter 0 for hours, 5 for minutes.

28

29

1 BOTTLES PER SAMPLE EVENT (1- 6)

Enter 1.

30

SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME

Select NUMBER OF SAMPLES.

31

SWITCH BOTTLES EVERY 1 SAMPLES (1-50)

Enter 1.

32

RUN CONTINUOUSLY? YES NO

Select NO.

33

DO YOU WANT SAMPLE VOLUMES DEPENDENT ON FLOW? YES NO

Select NO.

34

SAMPLE VOLUME 200 ml (10-1000)

Enter 800.

5-8

6712 Portable Sampler Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)


ENABLE: RAIN LEVEL FLOW NONE

Select RAIN. Note: All valid enable options will appear here, so your screen may differ.

35

ENABLE: RAIN

Select AND.

36

AND

OR

DONE

37

ENABLE: RAIN AND LEVEL FLOW NONE

Select LEVEL.

38

RAIN SET POINT: 0.15 INCHES PER __HOURS, __MINUTES ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

Enter 0.15 for inches, then enter 30 minutes.

Select ABOVE SET POINT.

39

40

RESET RAIN HISTORY? YES NO

Select YES.

41

LEVEL CONDITION: SET POINT RANGE RATE OF CHANGE LEVEL SET POINT 0.25 ft (0.001-30.000) ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

Select SET POINT.

Enter 0.25.

42

Select ABOVE SET POINT.

43

44

ONCE ENABLED, STAY ENABLED? YES NO

Select YES.

45

SAMPLE AT ENABLE? YES NO

Select YES.

46

PAUSE RESUME 1. HH:MM DD HH:MM DD 2. HH:MM DD HH:MM DD CLEAR DONE UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME

Select DONE. Screen will say End of Part A then Beginning Part B.

Select FLOW PACED.

47

5-9

6712 Portable Sampler Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)


PACED BY: FLOW PULSES FLOW MODULE VOLUME

Select FLOW MODULE VOLUME.

48

49

FLOW BETWEEN SAMPLE EVENTS: 2500 gal (0.001-99999)

Enter 2500.

50

SAMPLE AT START? YES NO

Select YES. This will cause the first sample to be taken 20 minutes (the delay to start of sampling specified in Step 67) after the enable for Part B.

51

1 BOTTLES PER SAMPLE EVENT (1- 18)

Enter 1.

52

SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME

Select NUMBER OF SAMPLES.

53

SWITCH BOTTLES EVERY 1 SAMPLES (1- 50)

Enter 1.

54

RUN CONTINUOUSLY? YES NO

Select NO.

55

SAMPLE VOLUME: 200 ml (10-1000)

Enter 500.

56

ENABLE: RAIN LEVEL FLOW A DONE NONE

Select RAIN.

ENABLE: RAIN

Select AND.

57

AND

OR

DONE

58

ENABLE: RAIN AND LEVEL FLOW A DONE DONE

Select LEVEL.

59

RAIN SET POINT: 0.15 INCHES PER __HOURS, __MINUTES ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

Enter 0.15 for inches, then enter 30 minutes.

Select ABOVE SET POINT.

60

5-10

6712 Portable Sampler Section 5 Extended Programming

Table 5-2 Storm Water Runoff Sampling (Continued)


Select YES.

61

RESET RAIN HISTORY? YES NO

62

LEVEL CONDITION: SET POINT RANGE RATE OF CHANGE LEVEL SET POINT 0.25 ft (0.100-30.000) ONCE ENABLED, STAY ENABLED? YES NO

Select SET POINT.

Enter 0.25.

63

Select YES.

64

65

SAMPLE AT ENABLE? YES NO

Select NO.

66

START OF SAMPLING AFTER A 15 MINUTE DELAY (1-999) PAUSE RESUME 1. HH:MM DD HH:MM DD 2. HH:MM DD HH:MM DD CLEAR DONE NO DELAY TO START DELAYED START CLOCK TIME WAIT FOR PHONE CALL START STORM AFTER A 72 HOUR DRY PERIOD? YES NO START STORM AFTER A 1 MINUTE DELAY (1-999) PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO

Enter 20.

Select DONE. Screen will say End of Part B.

67

Select DELAYED START.

68

69

Select NO. (Option for samplers using a 0.01 inch tip rain gauge. See Dry Period Option, described on page 5-26.)

Enter 1.

70

Select NO.

71

The above example presented several screens that are not available in standard programming. Descriptions of features available in extended programming follow.

5-11

6712 Portable Sampler Section 5 Extended Programming

5.7 Suction Head


Extended Screen for Suction Head Settings
AUTO SUCTION HEAD ENTER HEAD

Suction head is the vertical distance between the surface of the liquid source and the samplers pump. At each sample event, the sampler determines the suction head automatically. In extended programming, you can manually enter the suction head if you want. Select ENTER HEAD only when the head remains stable and you can measure it accurately.

If you choose AUTO SUCTION HEAD, the head will be determined for you. If you select ENTER HEAD, the screen below will be displayed so you can enter the suction head.
SUCTION HEAD OF ___ ft (min - max)

5.8 Rinses and Retries


Screens for Line Rinses and Sampling Retries
__RINSE CYCLES (0-3)

Rinses and retries are extended programming features. You can program the sampler to rinse the suction line automatically. During a line rinse cycle, the sampler draws liquid through the line to the liquid detector. As soon as it detects liquid, the sampler reverses the pump to purge the line. The sampling retries feature lets you set the number of times, from 0 to 3, that the sampler tries to detect liquid in the line before skipping the sample.

RETRY UP TO __ TIMES WHEN SAMPLING (0-3)

5.9 Pacing
Extended Pacing Screen
UNIFORM TIME PACE FLOW PACED EVENT PACED NONUNIFORM TIME

Sample pacing is the rate at which the sampler takes samples. Depending on the type of pacing you select, the rate is controlled by the samplers internal clock or by inputs received from connected instruments. Standard programming provides time pacing and flow pacing, which are described in Section 4. Extended programming provides additional pacing types, so its pacing screen (see example in margin) has four items listed instead of two. Uniform time pacing is identical to time pacing in standard programming. Flow paced is the same for both types of programming. Event paced and nonuniform time pacing are described below.

5.9.1 Event Pacing

To program the sampler for event pacing, select event paced. Refer to Table 5-3 for an example of event paced sampling that uses three event conditions. Up to fourteen event conditions can be defined (for two-part programs, up to fourteen event conditions can be defined for each part). An event will occur when any of the programmed event

5-12

6712 Portable Sampler Section 5 Extended Programming conditions becomes true. The default distribution setting for event-paced sampling is sequential (each time an event occurs, the sampler takes one sample, placing it in one bottle). However, any distribution setting may be used. If you have programmed multiple conditions, and more than one condition comes true at the same time (within your measurement interval), one sample will be taken and all conditions that caused the event will be noted on reports. Table 5-3 uses rate of change for the event condition (see step 7). This feature enables the event condition under specific circumstances. In the example below, a rate of change is set so that if the level rises more than one foot in a two hour time period, the event condition would be enabled and a sample taken. In determining if the rise occurs in this example, the sampler looks back at the previous two hour time period to see if there is a rise of more than one foot between any valley (low point) or peak (high point) within that time period. Once the condition becomes true (a rise of more than one foot is noted), the event condition is enabled. It will not be enabled again unless the condition becomes false, and then true again, within the two hour time window. In event pacing, up to three different rate of change event conditions can be programmed (for two-part programs, a total of six rate of change event conditions can be defined, three for each part). 5.9.2 Event Paced Sampling This example uses a sampler with an attached 730 Bubbler Flow Module, Rain Gauge, and an SDI-12 sonde at address 5 with four parameters.

Table 5-3 Event Paced Sampling


1
UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME EVENT01: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE

Select EVENT PACED.

Select the parameters that will trigger the event. Use the arrow keys to move through the parameters. To define the first event for this example, select LEVEL.

LEVEL CONDITION: SET POINT RANGE RATE OF CHANGE LEVEL SET POINT: 1.000 ft ( 0.001 - 30.00)

Choose SET POINT.

Enter a set point of 1 foot.

5-13

6712 Portable Sampler Section 5 Extended Programming

Table 5-3 Event Paced Sampling (Continued)


5
ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

Select ABOVE SET POINT.

EVENT02: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE LEVEL CONDITION: SET POINT RANGE RATE OF CHANGE CONDITION IS TRUE WHEN LEVEL RISES FALLS

Select LEVEL again.

Choose RATE OF CHANGE.

Choose RISES. This will enable the event condition if a rise occurs within the time frame and above the amount specified in screen #9. Enter 1 ft and press . Then enter 2 hours.

LEVEL RISES 1.000 ft __HOURS, __MINUTES

10

EVENT03: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE pH5 CONDITION: SET POINT RANGE RATE OF CHANGE pHx RANGE: LOWER: 4.00 UPPER: 10.00 ( 0.00 - 14.00) ENABLED WHEN: INSIDE RANGE OUTSIDE RANGE

Select pH.

Select RANGE.

11

Enter 4 for the lower range and 10 for the upper range.

12

Choose INSIDE RANGE.

13

14

EVENT04: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE PACING: EVENT PACED 3 EVENT CONDITIONS

You will not be entering a fourth event condition, so use the arrow keys to scroll down, and then select DONE.

Proceed with the rest of the programming steps.

15

5-14

6712 Portable Sampler Section 5 Extended Programming 5.9.3 Nonuniform Time Pacing The extended time pacing options are uniform time and nonuniform time. Uniform time pacing is identical to time pacing in standard programming. Nonuniform time pacing uses an irregular interval between sample events; each interval is individually programmable. There are three types of nonuniform time pacing: Nonuniform clock time pacing. Nonuniform interval pacing. Random interval pacing. 5.9.4 Nonuniform Clock Time Pacing For clock time pacing, enter specific times for each sample event (see screen 2). You can enter as many as 99 clock times, but the interval between times cannot be greater than 24 hours. (Remember to use a 24-hour clock when entering times.) The sampler takes a sample at the start time. For nonuniform interval pacing, enter the number of sample events spaced at intervals of minutes: 12 samples at 5 minute intervals, 6 samples at 10 minute intervals, and so on. The sampler accepts as many as 99 nonuniform interval entries. The sampler takes a sample at start time.

5.9.5 Nonuniform Interval Pacing

5.9.6 Random Interval Pacing


Time Pacing Screens
NONUNIFORM TIME: CLOCK TIMES INTERVALS IN MINUTES RANDOM INTERVALS

CLOCK TIMES
2 TAKE SAMPLES AT: 1. START TIME 2. HH:MM 3. HH:MM

To program the sampler for random interval pacing, enter the length of time you want to sample. The run time is the only random programming setting youll need to enter because the sampler derives the number of sample events required for the program from the distribution settings. (You can combine random pacing with any distribution.) Each time you run the program, the sampler generates a new set of random intervals. This makes each sample event unpredictable from run to run. In both nonuniform clock time pacing and nonuniform interval pacing, the sampler takes a sample at the start time. For random pacing, however, it takes the first sample at the end of the first interval, not at the start time.

INTERVALS IN MINUTES
3 FIRST SAMPLE AT START TIME, THEN . . .

QUANTITY AT INTERVAL 1. __ AT ___ MIN 2. __ AT ___ MIN 3. __ AT ___ MIN

RANDOM INTERVALS
5 PROGRAM RUN TIME: __ HOURS, __ MINUTES

5-15

6712 Portable Sampler Section 5 Extended Programming

5.10 Distribution

Distribution describes how the sampler is to deposit samples. A sample is the volume of liquid deposited in a bottle. A sample event includes the full sampling cycle and may deposit a sample into more than one bottle. You can program the sampler for five distribution methods: Sequential Bottles per sample Samples per bottle Composite Multiple Bottle Compositing

5.10.1 Sequential

In sequential distribution, the sampler deposits one sample in each bottle. A sequential sample represents a snapshot of the flow stream at a point in time. In bottles per sample distribution, the sampler deposits a sample in each of a set of bottles. A bottle set includes at least two bottles but may include all bottles. Use bottles per sample when the volume to be collected is larger than the amount one bottle can hold or when you need identical samples. In samples per bottle distribution, the sampler deposits samples from several sample events in a single bottle before moving to the next bottle. Use samples per bottle distribution to collect a series of small composite samples. For single bottle configurations, samples per bottle distribution is known as composite sampling. A composite sample represents an average of the flow streams characteristics during the sampling period. Multiple bottle compositing is a combination of bottles-per-sample and samples-per-bottle. At each sample event, the sampler deposits a sample into a set of bottles. It moves to a new bottle set only after each bottle of the first set contains the programmed number of samples. Use multiple-bottle compositing to collect identical sets of composite samples or a composite sample that is larger than the capacity of a single bottle. Multiple bottle compositing is available only in extended programming. In extended programming, the sampler offers an additional distribution feature, time switching. This feature determines when the sampler fills the next bottle or bottle set, and may be used with time-paced or flow-paced sample programs. Time switching distribution occurs at regular intervals (as programmed in screen 3 in the margin). The switch times are relative to the programmed FIRST SWITCH TIME (screen 4). Switch times occur regardless of the state of the sampler. Be aware that the sampler may leave empty bottles if it becomes disabled. Pause and resumes are an exception; the sampler will not switch bottles during a pause if the current bottle is empty.

5.10.2 Bottles Per Sample

5.10.3 Samples Per Bottle

5.10.4 Composite

5.10.5 Multiple Bottle Compositing

5.10.6 Time Switching

5-16

6712 Portable Sampler Section 5 Extended Programming Time Switched Distribution


1 2 BOTTLES PER SAMPLE EVENT (1-max)

If the program settings instruct the sampler to take a sample at the start time, the sampler also draws a sample at the switch time. When this occurs, the pacing will also be reset at the switch time. Time switching is normally used with flow-paced programs that do not take a sample at start. These programs will place a sample in the current bottle if it is empty at the switch time, and then move to the next bottle/set. Pacing intervals are reset at every switch time. When you use the time switching feature, the volume in each bottle or bottle set may vary. Sample volumes that would exceed the bottle capacity are not taken and a PROBABLE OVERFILL message is logged.

Enter the number of bottles that should receive samples at each sample event.
SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME

Select TIME.

SWITCH BOTTLES EVERY __ HOURS, __ MINUTES

Enter the time between bottle switches.


4 FIRST SWITCH TIME AT: __:__

Enter the first bottle switch time.

5.10.7 Continuous Sampling

Sample programs can be run indefinitely by selecting YES at the RUN CONTINUOUSLY? screen. Continuous sampling resets the distribution when the distribution sequence is complete. That is, when the last bottle/set is reached, the next bottle/set is the first bottle/set. All pacing modes except RANDOM are supported. When running a program continuously, the bottles must be serviced at regular intervals to prevent overfilling the bottles. The sampler assumes that the next bottle/set is empty and ready to receive samples. If you are using this feature with two-part programming, keep in mind that the parts will most likely reset at different intervals.

5-17

6712 Portable Sampler Section 5 Extended Programming

Sequential

EXTENDED PROGRAMMING

Each bottle receives one sample from one sample event.


DISTRIBUTION: SEQUENTIAL 1 BOTTLES PER SAMPLE EVENT (1-24) S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME SWITCH BOTTLES EVERY 1 SAMPLES (1-50) RUN CONTINUOUSLY? YES NO

Bottles-per-Sample

EXTENDED PROGRAMMING

Multiple bottles receive a sample from one sample event.


DISTRIBUTION: 2 BOTTLES/SAMPLE 2 BOTTLES PER SAMPLES EVENT (1-24) S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME SWITCH SETS EVERY 1 SAMPLES (1-50) RUN CONTINUOUSLY? YES NO

Samples-per-Bottle

2 1

4 3

6 5

8 7

EXTENDED PROGRAMMING

Each bottle receives a sample from multiple sample events.


1 BOTTLES PER SAMPLE EVENT (1-24) S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME SWITCH BOTTLES EVERY ___ SAMPLES (1-50) RUN CONTINUOUSLY? YES NO

DISTRIBUTION: 2 SAMPLES/BOT TLE

Multiple Bottle Compositing


3 2 1 3 2 1 6 5 4 6 5 4 8 7 8 7
EXTENDED PROGRAMMING ONLY
S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME

DISTRIBUTION: 2 BOTTLES/SAMPLE 3 SAMPLES/BOT TLE

2 BOTTLES PER SAMPLE EVENT (1-24)

SWITCH BOTTLES EVERY 3 SAMPLES (1-50)

RUN CONTINUOUSLY? YES NO

Time Switched (Flow-paced, No sample at start)


Start Time Time Override

2 1 TO 6 5 4
EXTENDED PROGRAMMING ONLY
DISTRIBUTION: 1 BOTTLES/SAMPLE S W I T C H AT 1 3 : 0 0 THEN EVERY 1:00 S W I T C H B O T T L E S ON: NUMBER OF SAMPLES TIME

8 7

1 BOTTLES PER SAMPLE EVENT (1-24)

SWITCH BOTTLES EVERY 1 HOURS, 0 MINUTES

FIRST SWITCH TIME AT : 1 3 : 0 0

Figure 5-1 Sample Distribution

5-18

6712 Portable Sampler Section 5 Extended Programming

5.11 Flow Proportional Sample Volumes

The 6712 can collect flow proportional sample volumes. For some sampling protocols, these flow-weighted, variable-volume samples can be more representative of the stream. This is because stream characteristics often change as the flow fluctuates. Variable-volume sampling is possible when the sampler is programmed for Uniform Time pacing. To enable this feature, select Y E S a t t h e d i s p l a y a s k i n g D O YO U WA N T S A M P L E VOLUMES DEPENDENT ON FLOW?

DO YOU WANT SAMPLE VOLUMES DEPENDENT ON FLOW? YES NO

After answering YES, the sampler displays screens to program the flow proportional volumes. If the sampler has a 700 Series Flow Module inserted, you will be asked which flow measurement should be used to determine the volumes.
FLOW PULSES FLOW MODULE VOLUME

To base the volumes on an external flow measurement device attached to the Flow Meter connector, select FLOW PULSES. To base the volumes on the 700 Series Flow Module readings, select FLOW MODULE VOLUME. Next, the sampler asks for the SAMPLE VOLUME to collect at each time interval. This volume is expressed as 10 ml for every X number of pulses, or if using a flow module, for every volume unit.
SAMPLE VOLUME: 10 ml FOR EVERY ___ PULSES

Refer to section 5.11.1, Calculating Variable Sample Volume Settings, for a more detailed discussion on this setting. The final screen used to program variable-volume sampling is the MINIMUM SAMPLE VOLUME. If the number of flow pulses or flow module volume is not sufficient enough to collect this minimum volume, the sample is skipped and the flow count carries over to the next sample time. This setting ranges from 10 ml to a maximum determined by the bottle volume.
MINIMUM SAMPLE VOLUME ___ ml (10 - maximum)

5-19

6712 Portable Sampler Section 5 Extended Programming 5.11.1 Calculating Variable Sample Volume Settings The entry for the SAMPLE VOLUME display can be calculated based on expected flow rates and collection volume requirements. The number of pulses or flow module volume that you enter should ensure that the minimum volume of liquid required for analysis is collected without exceeding the bottle volume capacity of the sampler. You may find it helpful to determine a range of values by calculating both possibilities. The number you enter would be selected from this range. This range is illustrated in Figure 5-2.
Total Bottle Capacity

Minimum Required Sample Volume

w Flo ted c w pe Fl o Ex st ted p ec ate x e Gr st E L ea

Time

Figure 5-2 Factors Affecting Flow-Weighted Variable-Volume Samples In other words, at the greatest expected flow, this SAMPLE VOLUME value should be set to prevent filling the bottle before the end of the program. An early full-bottle condition would result in a composite sample that is not representative of the flow stream. At the lowest expected flow, the SAMPLE VOLUME value should be set to ensure that enough sample is collected. If too little sample liquid is collected, there may not be enough for analysis although it is a representative composite sample. For example, consider an NPDES sampling program used to collect a flow-proportional composite sample for BOD, TSS, and NH3-N analysis. The laboratory requires at least 1,000 ml for BOD, 100 ml for TSS, and 400 ml for NH3-N tests. The minimum composite volume needed for analysis is then 1,500 ml, but to allow for duplicates and spillage, this volume is doubled to 3,000 ml. The samples are to be collected over an eight hour period at 15 minute intervals into a 9,400 ml composite bottle. Flow rates at the site range from 1.5 to 2.0 mgd, the equivalent volume of 500,000 and 666,667 gallons respectively over the eight hour period. This flow is measured by an external flow meter that sends a flow pulse every 100 gallons. Figure 5-3 depicts this scenario.

5-20

Flow Volume

6712 Portable Sampler Section 5 Extended Programming

9400 ml

666,667 gal

3000 ml
c pe te low dF
w Flo

500,000 gal

Gr

s at e

x tE

L ea

st E

d cte xp e

4 Hours

Figure 5-3 Variable-Volume Scenario 5.11.2 Calculations when Using an External Flow Meter The highest possible SAMPLE VOLUME value is determined by the minimum volume required for analysis and the least expected flow rate over the sampling duration. To calculate the number of pulses at this high end of the range, you should know: the least expected flow volume that will pass the flow measurement point the flow volume interval between each flow pulse from the external flow meter the minimum composite sample volume required for analysis. Allow for duplicates and spillage. These factors would be applied to the following equation.
(Least Total Flow Vol) / (Flow Pulse Vol) 10 ml for every __ pulses = (Minimum Required Vol) / (10 ml per unit)

SAMPLE VOLUME: 10 ml FOR EVERY ___ PULSES

From the scenario we can apply the factors for the equation.
(500,000 gallons) / (100 gallons per pulse) 10 ml for every 16.7 pulses = (3,000 ml / 10 ml)

Because the SAMPLE VOLUME entry must be in whole numbers, 16.7 is rounded down to 16. It is important to note that this number is inversely proportional to the amount of volume collected. Rounding up would result in less volume collected. The lowest possible SAMPLE VOLUME value is dependent upon the highest expected flow and the available bottle volume. To calculate the number of pulses or volume at this low end, you should know: the greatest expected flow volume that will pass the flow measurement point the flow volume interval between each flow pulse from an external flow meter the available bottle volume for sample collection.

5-21

6712 Portable Sampler Section 5 Extended Programming These factors would be applied to the following equation.
(Greatest Total Flow Vol) / (Flow Pulse Vol) 10 ml for every __ pulses = (Total Bottle Vol) / (10 ml per unit)

From the scenario we can apply the factors to the equation.


(666,667 gallons / 100 gallons) 10 ml for every 7.0 pulses = (9,400 ml / 10 ml)

At high flow rates, a number below 7 would overfill the bottle. At low flow rates, a number greater than 16 would collect too little sample. Therefore, a value should be selected between 7 and 16 pulses. 5.11.3 Calculations when Using a 700 Series Flow Module The same principles discussed in Section 5.11.2 apply. However, the equations differ slightly because the flow module directly measures the streams flow volumes. The equation for the high end of the range is:
(Least Total Flow Vol) 10 ml for every __ units = (Minimum Required Vol) / (10 ml per unit)

Using the same scenario the calculations would be:


SAMPLE VOLUME: 10 ml FOR EVERY ___ Gallons
500,000 gallons 10 ml for every 1667 gallons = (3,000 ml) / (10 ml per gallon)

The equation for the low end of the range is:


(Greatest Total Flow Vol) 10 ml for every __ units = (Total Bottle Vol) / (10 ml per unit)

The scenario would be applied as:


666,667 gallons 10 ml for every 709 gallons = (9400 ml) / (10 ml per unit)

Therefore, a value should be selected between 709 and 1667 gallons for the given scenario.

5-22

6712 Portable Sampler Section 5 Extended Programming

5.12 Sampler Enable

Certain external instruments can enable (start) or disable (stop) a sampler by sending a signal to pin F of the samplers flow meter connector. Teledyne Isco flow meters, flow loggers, and the Model 1640 Liquid Level Actuator have a programmable sampler-enable feature that lets them enable or disable the sampler. You can define the condition (such as level, flow rate, pH, temperature, percent, rainfall, and I/O) or combination of conditions that need to be true in order to enable the sampler. Refer to Figure A-12 in Appendix A for details on programming sampler enables. When the sampler detects that it is disabled at start time, it suspends the program until it is enabled. Once enabled, the sampler collects a sample only when programmed to take a sample at the enable time. The diagram in Figure 5-4 shows the samplers response when enabled or disabled at the start time.

Sampler enabled at start time A sample event at the start time always occurs for time-paced programs (except for random). For flow-paced programs, it occurs when the program requires a sample at start time.

Sampler disabled at start time by signal from pin F A sample event at enable time occurs only when the program requires a sample at enable.

Figure 5-4 Start Time Diagram for Sampler Enable Settings

5-23

6712 Portable Sampler Section 5 Extended Programming 5.12.1 Sampler Enable This example uses a sampler with an attached 730 Bubbler Flow Module, Rain Gauge, and an SDI-12 sonde at address 5 with four parameters.

Table 5-4 Sampler Enable


ENABLE:

NONE PROGRAMMED

In this example, you will be enabling the sampler if it rains .5 in 2 hours and the level goes above 1 foot.

ENABLE: RAIN LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 NONE ENABLE: RAIN

Select RAIN.

Choose AND.

AND

OR DONE

ENABLE: RAIN AND LEVEL FLOW TEMP5 pH5 SP_CO5 DO5 DONE RAIN SET POINT: .5 INCHES PER __HOURS, __MINUTES

Select LEVEL.

Enter .5 for the inches of rain. Press . Enter 2 HOURS. Select ABOVE SET POINT.

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

RESET RAIN HISTORY? YES NO

Select YES.

LEVEL CONDITION SET POINT RANGE RATE OF CHANGE LEVEL SET POINT: +1 ft ( 0.001 - 30.00) ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

Select SET POINT.

Enter 1 foot.

Select ABOVE SET POINT.

10

11

ENABLE: RAIN >0.50 / 2:00 AND LEVEL >1.000 ft

Your enable conditions will be summarized on the display. If correct, press to continue. If necessary, make changes to the enable conditions.

5-24

6712 Portable Sampler Section 5 Extended Programming

5.13 Sampler Enable Responses

In addition to programming enable conditions, extended programming lets you control the samplers response to its enable state. You can: Stay enabled after the first enable. Set up a repeatable enable. Take a sample at enable or disable. Delay the start of sampling after the enable. Reset the sampling interval countdown each time the sampler is enabled. Control the sampling interval countdown while disabled. Enable Part B of a two-part program when Part A is done.

5.13.1 Once Enabled Stay Enabled

For certain monitoring programs, you may want the sampler to continue to sample, even though the conditions that enabled the sampler no longer exist. When you use the ONCE ENABLED, STAY ENABLED feature, after becoming enabled the sampler takes samples until it reaches the end of its program.
ONCE ENABLED, STAY ENABLED? YES NO

Note
The STAY ENABLED feature works only when the sampler becomes enabled following program start. If you need this feature with enable prior to program start, contact Teledyne Isco for details. 5.13.2 Repeatable Enable After enabling the sampler, the enabling instrument (or instruments) continues to monitor the conditions and disables the sampler when the conditions are no longer satisfied. If re-enabled, the sampler resumes the sampling program.

Note
Repeatable enable is called sampler latching in flow meters, flow loggers, and FLOWLINK. Latching a sampler produces the same results as selecting YES; the difference is that the external instrument controls the latching. 5.13.3 Sample at Enable or Disable You have the option of directing the sampler to take a sample as soon as it receives the enable and/or disable signal. If you program the sampler to stay enabled and choose to take a sample at enable, the sample interval will be reset at the enable time. Samples at disable do not affect the sample intervals. Standard programs do not take a sample at the enable time.
SAMPLE AT ENABLE? YES NO

5-25

6712 Portable Sampler Section 5 Extended Programming 5.13.4 Delay To Start Of Sampling If you have selected ONCE ENABLED STAY ENABLED and have chosen to not sample at the enable, you have the opportunity to suspend the start of sampling. This choice is useful for storm programs that require a flow-paced sample taken after the initial grab sample.
START SAMPLING AFTER A ___ MINUTE DELAY (1-999)

5.13.5 Dry Period Option

For samplers used with a 0.01 inch tip rain gauge, you can cause the sampler to wait for a dry period of under 0.10 inches of rain over a 72-hour period before running the program. At RUN time, the sampler asks if you want to restart the 72-hour period. Select this choice if you want the sampler to disregard any dry period that took place prior to the program start. The sampler will wait for up to one year for the dry period to occur.

START EXTENDED 1 AFTER A 72 HOUR DRY PERIOD? YES NO

5.13.6 Resetting the Sample Interval at Enable You may want the pacing countdown to start over each time the sampler becomes enabled. To accomplish this, select YES when asked if you want to reset the sample interval at enable. The next sample event will occur at the end of a full interval. Furthermore, if the sampler is enabled several times during the program, it resets the countdown each time. Use the reset-interval feature carefully. For example, avoid resetting the interval when you sample with uniform-time pacing and want to take samples at specific times. The effects of resetting the interval can be seen in this example: A sampler has been programmed to sample every 15 minutes and to reset the interval when enabled. It receives an enable signal at 10:03 and takes a sample. After resetting the interval, the sampler takes subsequent samples at 10:18 (not 10:15), 10:33 (not 10:30), and so on. If you select NO when asked if you want to reset the sample interval at enable, you will be asked if you want to continue the countdown while disabled. You can have the sampler continue the countdown, repeating it as often as necessary while the sampler is disabled, or you may suspend the countdown while disabled, resuming the countdown as the sampler is enabled. The standard program allows the countdown to continue while disabled.

RESET SAMPLE INTERVAL AT ENABLE? YES NO

If you answer NO to the above screen, you will see:


CONTINUE COUNTDOWN WHILE DISABLED? YES NO

5-26

6712 Portable Sampler Section 5 Extended Programming

5.14 Pauses/Resumes
Pause and Resume Screens
Flowcharts containing the pause and resume screen appear in Appendix A.
PAUSE RESUME 1. 17:00 MO 08:00 TU 2. 17:00 TU 08:00 WE CLEAR DONE

Pause and resume settings, available in extended programming, create an intermittent sampling schedule. A program with pauses and resumes begins sampling at its programmed start time, continuing until the first pause time and day of the week. It then suspends sampling until the first resume time, when it begins sampling again. If the start time falls within a pause period, the sampler will not begin sampling until the next resume time. Pauses and resumes repeat weekly. You can enter up to nine pairs of pause and resume times. The pause and resume times shown as examples to the left suspend the program between 5:00 p.m. Monday and 8:00 a.m. on Tuesday and between 5:00 p.m. Tuesday and 8:00 a.m. on Wednesday. If the start time is Monday at 8:00 a.m., the sampler takes samples between 8:00 a.m. and 5:00 p.m. Monday; between 8:00 a.m. and 5:00 p.m. Tuesday; and, continues after 8:00 a.m. Wednesday. To remove pause and resume settings from a program, select CLEAR. A sampler running a pause/resume program resets the pacing interval at each resume time. You can program the sampler to take a sample at resume times (see screen at left).

SAMPLE AT RESUME? YES NO

5.15 Running Programs

To run an extended program, select run from the main menu. You may also select YES at the RUN THIS PROGRAM NOW? screen at the end of the programming screens. Extended programs scheduled to start at a clock time will not start until the programmed clock time on or after the first valid date. Note that if you have recently calibrated a parameter probe, the sampler may delay the start and display PLEASE WAIT until valid readings are expected from the probe. The delay accounts for the time typically required to install the probe, and lasts for five minutes after you complete the calibration.

5.15.1 Run Time Screens

Run time screens, which report the programs status while running a program, are explained in Section 4. Interrupting a running program is also explained in Section 4. The scrolling menu that appears during the manual paused state can have two extra options when using extended programming: Adjust Enable This option is available for extended sampling programs that have programmed enable conditions. Selecting this option allows you to change the set-points or ranges for the enable conditions. The sampler displays all of the programming screens that define the thresholds of the programmed enables. You cannot change the type of enables used or the way they are combined. If rain is an enabling condition, you also have the option of resetting the rain history.

5.15.2 Interrupting a Running Program

5-27

6712 Portable Sampler Section 5 Extended Programming Adjust Suction Head If you have chosen to manually enter the suction head, and are in the manual paused state, the scrolling menu will have an option to adjust the suction head. If you make changes to the suction head value, the sampler logs a changed suction head event.

5.16 Sampling Reports

The 6712 records a variety of data while running a sampling program. Sample reports can be found in Section 4. Section 4, Table 4-3 contains a list of source codes and program events that may be recorded when using either standard or extended programming.

5.17 Other Functions


Other Functions Menu
RUN PROGRAM NAME PROGRAM VIEW REPORT OTHER FUNCTIONS

In extended programming, the Other Functions menu screen includes options for: Maintenance (see Section 7) Manual Functions (see Section 4) Software Options Hardware

Select OTHER FUNCTIONS.


MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE

5.18 Software Options

The software option screens allow you to set up a number of different software options:

Liquid Detector Enable Normal/Quick View Screens Measurement Interval Dual Sampler Mode Full Bottle Detection

Display Backlighting Event Marks Pump Purge Counts Serial Output Remote Access Interrogator Connect Power

5.18.1 Liquid Detector Enable/Disable

Under some conditions (pressurized lines, sampling foam), a more repeatable sample volume may be delivered by disabling the liquid detector. Selecting NO will turn the Liquid Detector off. If the Liquid Detector is off: you must enter the SUCTION HEAD no RINSES or RETRIES will be available you will not be able to CALIBRATE SAMPLE volumes the RESULTS report will read a COUNT TO LIQUID of 0 A complete purge will occur between samples when using multiple bottles-per-sample distribution.

USE LIQUID DETECTOR? YES NO

5-28

6712 Portable Sampler Section 5 Extended Programming 5.18.2 Programming Style The sampler has two different styles of programming screens: normal and quick view. These are both explained in Section 3. When you are in extended programming, the default style is Quick View, which allows you to move more quickly through the screens. If you prefer a step-by-step display for the screens, switch to the Normal setting. 5.18.3 Measurement Interval If using a 700 Series module or SDI-12 Sonde parameter for enabling, the sampler will request a Measurement Interval. The Measurement Interval allows the sampler to compare the parameter to the enabling conditions at a rate faster than the Data Storage interval. If the Measurement Interval is equal to or greater than the Data Storage interval, the parameter will be checked at the rate of Data Storage interval. Normally, there are two considerations when selecting the Measurement Interval: power consumption and duration of the enabling event. This setting for battery powered samplers will require some degree of balance between the need to conserve power yet detect the enabling event before it has passed. 5.18.4 Dual Sampler Mode Dual Sampler Mode Dual Sampler Mode lets you operate two samplers as a pair. The primary sampler disables the secondary one until the primary sampler completes its sampling program. The samplers then reverse roles with the second sampler becoming the primary one. Setting Up Dual Samplers 1. Turn on the dual sampler mode in both samplers. 2. Install the samplers. 3. Connect the samplers with the Dual Sampler Mode Interconnect Cable by attaching the cable to the samplers flow-meter connectors. If your installation includes a flow meter or flow logger, use the Dual Sampler Mode to Flow Meter Cable instead. 4. Decide which sampler is to be the primary sampler; the primary sampler should be the sampler with the program you want to run first. 5. Turn the primary sampler on and run its sampling program by selecting RUN. After at least 5 seconds, turn the secondary sampler on and select RUN from its main menu. How Does It Work? The primary sampler sends a disable signal out its flow meter connector. The secondary sampler receiving the signal is held disabled until the primary sampler completes its program. When using a flow meter or flow logger a Dual Sampler Mode to Flow Meter cable must be used. This cable is constructed so that the pacing instrument receives only the event marks. Therefore, a connected flow meter will not receive a bottle number signal, nor will it be able to control the sampler inhibit. 5-29

MEASUREMENT INTERVAL 5 15 30 SECONDS 1 2 5 MINUTES STORAGE INTERVAL

DUAL SAMPLER MODE? YES NO

Select YES to put the sampler in the dual sampler mode. When you no longer need dual samplers, return to this screen to turn the feature off.

6712 Portable Sampler Section 5 Extended Programming Bottle Full Detect


BOTTLE FULL DETECT? YES NO

Bottle Full Detection The sampler can detect a full bottle condition when collecting samples in a single bottle (composite sampling). When the sampler detects a full bottle, it stops the sampling routine. A BOTTLE FULL condition will also stop a sampling routine that is set to RUN CONTINUOUSLY. The Bottle Full Detect screen allows you to enable or disable this feature. Selecting YES enables the sampler to detect a full bottle and stop the sampling routine. Selecting NO disables this feature. The Bottle Full Detect setting is part of the sampling program. Because it is a program setting, full bottle detection can be enabled or disabled for each of the four extended programs. If the Bottle Full Detect setting is critical to your sampling program, always check this Software setting after you select a different Extended Program. This feature relies on a post-purge after each sample. When using the Bottle Full Detect, make sure that post-sample purges greater than 100 are used. Bottle Full Detect is always disabled in the Standard Programming mode.

Adjusting the Display Backlighting


BACKLIGHT ALWAYS OFF TIMED BACKLIGHT CONTINUOUS BACKLIGHT

Display Backlighting The display has a backlight that is adjustable so that the backlight is always off, timed or always on. In the timed backlight mode the sampler turns the backlight off after detecting no key strokes within 60 seconds. Any keystroke (except ON-OFF) turns the backlighting on again. In this mode the backlight must be on before the sampler accepts a menu choice or number entry. If you press a key and a screen does not respond, it is because the keystroke turned the backlight on. It is recommended that you use either BACKLIGHT ALWAYS OFF or TIMED BACKLIGHT when using a battery.

Event Marks
EVENT MARK SENT FOR: EVERY SAMPLE COMPLETE SAMPLES

Event Marks The sampler generates an event mark at each sample event, sending it through the flow meter connector to a connected instrument. The event mark may also control an external device; a connected solenoid relay, for example. When using the event mark for a data logger or to control external devices, you may need to adjust the mark by changing the factory settings. The event mark is adjustable through the software setup screens. You can set up the sampler to generate an event mark for EVERY SAMPLE or for COMPLETE SAMPLES. If you select COMPLETE SAMPLES, the sampler sends a three-second pulse only after it successfully collects a sample. The sampler will not send an event mark if an error occurs, such as No Liquid Detected. If you select EVERY SAMPLE, the sampler will send an event mark every time a sample is initiated. The event mark signal can be a: 3-second pulse at the beginning of the pre-sample purge.

5-30

6712 Portable Sampler Section 5 Extended Programming 3-second pulse at the beginning of forward pumping only, marking the time the sampler deposited the sample. Variable duration pulse generated during the pump cycle, from the beginning of the pre-sample purge to the end of the postsample purge. Variable duration pulse generated only during forward pumping. Figure 5-5 shows event mark timing diagrams. Bottle Number Concurrent with the event mark, the sampler also sends a bottle number signal. The bottle number signal is a series of pulses. The number of pulses sent represents the bottle currently accepting a sample (see Figure 5-5).
Presample Purge Rinse Cycle Pump Volume 1 Pump Volume 2 Pump Volume 3 Postsample Purge

Pulse at Beginning of Presample Purge


Event Mark Bottle Number

Pulse at Beginning of Forward Pumping


Event Mark Bottle Number

Variable Duration Pulse During Entire Pump

Event Mark

Bottle Number

Variable Duration Pulse During Forward Pumping

Event Mark Bottle Number

This diagram shows the pulses sent for a bottles-per-sample distribution scheme

Figure 5-5 Event Mark Timing Diagram 5.18.5 Pre-sample and Post-sample Purge Counts
PRE-SAMPLE PURGE: ____COUNTS (10-9999)

POST-SAMPLE PURGE: DEPENDENT ON HEAD FIXED COUNT

The pre-sample purge is when the pump runs in reverse clearing the strainer of any debris before taking a sample. The pre-sample purge is set at 200 pump counts. After the sample is taken a post-sample purge occurs to clear the liquid from the suction line. The duration of the post-sample purge is determined by the length of the suction line to maximize the battery life. The preand post-sample purge can also be manually set.

POST-SAMPLE PURGE: ____COUNTS (10-9999)

5-31

6712 Portable Sampler Section 5 Extended Programming 5.18.6 Serial Output Programming the serial output
PERIODIC SERIAL OUTPUT? YES NO

Serial ASCII data can be output through the controllers interrogator port. The output frequency is user-selectable: every 15 seconds, every minute, every 5 minutes, every 15 minutes, or by command. To receive serial data as a periodic output: 1. Select YES at screen asking if you want periodic serial output.

PORT SETTINGS: 9600N81 4800N81 2400N81 1200N81

2. Select the appropriate baud rate setting in screen . Note that at all baud rates the data is sent with no parity, eight data bits, and one stop bit. If you have difficulty with the connection, try a slower baud rate. 3. Select the desired output interval in screen. 4. Connect to the 6712 controllers interrogator connector using a standard interrogator cable. To receive serial data by command: 1. Select NO at the screen asking for periodic serial output . 2. Connect your computer to the 6712 controllers interrogator connector with a standard interrogator cable, P/N 60-2544-040 (25 pin) or 60-2544-044 (9 pin). 3. Configure your communications software for a baud rate no higher than 19,200 kb, no parity, 8 data bits, and 1 stop bit. 4. From the computers keyboard, type ? repeatedly until the 6712 controller determines the baud rate. Once the baud rate is known, the 6712 returns an identification banner. This banner indicates that the connection is working properly. 5. Type DATA<CR> each time the data string is desired. The serial data is in comma-separated-value format. Only the values that the sampler is set up to measure will be output in the data string. The string is terminated with a <CR><LF>. To read the data in the string, refer to Table 5-5. This serial data can also be accessed from a remote location using the optional dialout modem (see Section 7).

SERIAL OUTPUT EVERY: 15 SEC 1 MIN 5 MIN 15 MIN

Note
The internal modem is disabled when an interrogator cable is connected to the samplers Interrogator port. It cannot receive incoming calls, and the alarm dialout is rendered inoperative, while this cable is connected. Disconnect the interrogator cable in order to use the 4200T modem.

5-32

6712 Portable Sampler Section 5 Extended Programming

Table 5-5 Serial Data Codes


Identifier Parameter
B?1 CR CS2 DE FL ID LE MO PE PH PR RA RTE SS TE TI VE VO VSI VSP COa3 DOa PHa SAa SPa TDa TEa ORa LEa MMa MAa NIa TBa Bottle number and time Current days rain Checksum Description Flow Unit specific identifier Level Model Percent pH Previous days rain Rain Refrigerator Temperature Sampler enable state Temperature Current time Velocity Volume Velocity signal strength Velocity spectrum strength Conductivity Dissolved oxygen pH Salinity Specific Conductance Total Dissolved Solids Temperature Oxidation reduction potential Level Ammonium-nitrogen Ammonia-nitrogen Nitrate-Nitrogen Turbidity

Units
days since 1-Jan 1900 tips (rolls over every 256 tips) unsigned long 6712 Sampler cubic meters per second 10 digit unsigned long meters 6712 percent of full scale none tips (rolls over every 256 tips) tips (rolls over every 256 tips) degrees celsius logical degrees celsius days since 1-Jan-1900 meters per second cubic meters percent percent milliseimens per centimeter milligrams per liter none parts per thousand milliseimens per centimeter grams per liter degrees Celsius millivolts meters milligrams nitrogen per liter milligrams nitrogen per liter milligrams nitrogen per liter nephelometric turbidity units

5-33

6712 Portable Sampler Section 5 Extended Programming

Table 5-5 Serial Data Codes (Continued)


Identifier Parameter
CLa CPa FLa VOa xxDa4 Chloride Chlorophyll Flow rate Flow volume Other data not listed above

Units
milligrams per liter micrograms per liter cubic meters per second cubic meters

The three most recent samples will have entries as B?. The ? will be the first bottle number in the set. If the sample is a grab sample, it will be shown as B0. An a in the identifier indicates an SDI-12 address. XX matches the data label for the SDI-12 parameter. The checksum does not include the checksum, carriage return, and linefeed. The string is terminated with <CR><LF>.

5.18.7 Program Lock

The specialized software option Program Lock (P/N 60-9003-532) may be purchased from Teledyne Isco. Once activated, the purchased option is always available, even if the software is upgraded at a later time. The option must be installed at the factory. Consult the factory for further details. To view which software options have been activated, select VIEW REPORT>SYSTEM IDs. A Software Options screen will list a code for each active option. The code for Program Lock is 1108. The Program Lock adds password security to the 6712 sampling programs. With this option activated, a user-selected password (up to five digits) must be entered before any program changes or run time modifications can be made. Run time modifications include accessing the Manual Pause menu options and shutting off the sampler. You will still be able to run the current program, view reports, and interrogate the sampler without the knowledge of the password. If three consecutive attempts to enter the password fail, a five minute time-out will follow in which most keys are deactivated. The user-selected password defaults to 6712 and may be changed in the Software options that are available in the Extended Program mode (see example at left). While the password may only be changed in the Extended Program mode, the program lock works for both Standard and Extended programs. Should you forget your password, a key code can be obtained from Teledyne Iscos Customer Support department for a nominal fee.

PROGRAM LOCK DISABLED

ENABLE PROGRAM LOCK? YES NO

CHANGE PASSWORD TO ----------

REENTER NEW PASSWORD ---------

5-34

6712 Portable Sampler Section 5 Extended Programming

From Standby
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS ENTER PASSWORD: ____ Programming Screens

VIEW DATA SYSTEM IDs CONFIGURE REPORTS

ENTER PASSWORD: ____

SAMPLINK REPORT FLOWLINK REPORT CUSTOM REPORT ALL REPORTS

ENTER PASSWORD: ____

MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

While Running a Program


SAMPLE 3 IN 00:03:26

Stop Key
ENTER PASSWORD: ____

STOP PROGRAM RESUME PROGRAM VIEW DATA GRAB SAMPLE

ENTER PASSWORD: ____

On/Off Key

Figure 5-6 Password Protected Functions

5-35

6712 Portable Sampler Section 5 Extended Programming

5.19 Hardware Setup


The hardware setup screens make adjustments to the samplers hardware-based features:
RUN PROGRAM NAME PROGRAM VIEW REPORT OTHER FUNCTIONS

Select Other Functions


MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE

Select Hardware

SDI-12 Sonde Setup Rain Gauge Setup I/O Pin Programming Dialout Alarms 5.19.1 SDI-12 Sonde Setup

Pager Numbers Text Messaging Refrigerator Temperature Analog Outputs

The 6712 supports SDI-12 sondes. For more information on setting up sondes, refer to Section 6.

5.19.2 Rain Gauge Setup

With standard programming, the sampler does not record rain gauge readings. However, with extended programming, recording the rain gauge readings and the rain enable option can be turned on or off. Enter the hardware setup and follow the screens to the left in the margin. When asked if you have a rain gauge connected, select YES to record the rain gauge readings and add RAIN as a sampler enable option. Select the appropriate tip for the rain gauge. The sampler will then create a rain memory partition and update the available sampler enable conditions. Select NO to disable the readings and the sampler enable option. When you select NO, the sampler removes the rain memory partition, leaving the memory space available for other reading types.

DO YOU HAVE A RAIN GAUGE CONNECTED? YES NO

Select YES.
RAIN GAUGE: 0.01 inch TIP 0.1 mm TIP

The rain gauges are set up to tip on either .01 inch or .1 mm. Select the desired tip.

5-36

6712 Portable Sampler Section 5 Extended Programming 5.19.3 I/O Pin Programming Pins C, H, and I on the Rain Gauge Connector are user-configured as I/O1, I/O2, and I/O3 (Input/Output pins) respectively. Each pin can be either an input for sampler enabling, or an output for alarms.

Note
The state of the I/O lines is not defined when the sampler is OFF. The enable input requires a pulse signal of 5 to 15 volts. Teledyne Isco offers optional interface modules for converting from a 4-20 mA analog or pulse duration signal. The alarm output can be set for any RUNNING PROGRAM or RUN ERROR state. In addition, the CONDITIONS output can report alarm conditions using the parameters the 6712 is configured to measure. You can set thresholds for one or two conditions and the CONDITIONS output can then trigger alarms when the measured parameters exceed the threshold. Alarm function requires an alarm interface module, factory-installed talking modem, or external digital cellular modem (see Section 5.19.4 Dialout Alarms). Teledyne Isco offers a module designed to provide one, two, or three contact closures. The alarm module can interface the sampler with many types of non-Isco devices. For more information about the alarm module, contact your Teledyne Isco sales representative.

5-37

6712 Portable Sampler Section 5 Extended Programming

Example: I/O Pin Programming


Although the screens displayed may vary depending on your site setup, the programming sequence will be the same or similar to the following example. 1. Select OTHER FUNCTIONS to enable the dialout alarm.
RUN EXTENDED 1 PROGRAM VIEW REPORT OTHER FUNCTIONS

Note: All choices listed are alarm output settings, except for I/O ENABLE. This is the input function to enable the sampler via an external device. 4. If you select CONDITIONS, the sampler will display:
I/O1: RAIN NONE

2. Select hardware and use the down arrow to scroll to the I/O screen.
MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE I/O1 = NONE I/O2 = NONE I/O3 = NONE

Depending on what equipment is connected to your sampler, you may see other parameters in addition to RAIN, such as LEVEL, FLOW, VELOCITY, etc. Under CONDITIONS, you can program limits for a condition to trigger the alarm when the measured parameter exceeds its limit. You can use one or two conditions to trigger one alarm, using AND or OR operators. 5. When you select a parameter with the arrow keys and Enter, you can then program the limits for alarm triggering. For example, if you select rain, the sampler will display:
"RAIN" SET POINT: 1.00 INCHES PER 1 HOURS, 0 MINUTES

3. Select the first pin to program.

SET I/O1:

From this screen, your choices are viewed by scrolling with the arrow buttons:
PGM RUNNING PGM ENABLED

I/O IS HIGH WHEN: ABOVE SET POINT BELOW SET POINT

A ENABLED B ENABLED
PROGRAM DONE

I/O IS HIGH means Alarm is triggered. Select ABOVE or BELOW, then select whether or not to reset rainfall history when the alarm is triggered:
RESET RAIN HISTORY? YES NO

A DONE B DONE
RUN ERROR FLOW-THRU CONDITIONS NONE I/O ENABLE

Continue with programming I/O2 and I/O3 in the same manner, or if additional alarms are not needed, select NONE.

5-38

6712 Portable Sampler Section 5 Extended Programming 5.19.4 Dialout Alarms Dialout alarms require the sampler to be equipped with the factory-installed 2400 baud talking modem, or the external digital cellular CDMA or GSM modem. The dialout alarm allows the sampler to notify a contact list when an alarm condition exists (is true). At the transition to the true state, any programmed I/O output can cause the sampler to warn of the alarm condition. As many as three telephone numbers can be entered. The sampler tries each telephone number in succession until the alarm is acknowledged. For each alarm condition, the sampler will attempt each telephone number entered a maximum of five times, at user specified intervals. Alarms are set up through the I/O programming menu, shown in Section 5.19.3. I/O1, I/O2, and I/O3 represent the three alarms.

Example: Dialout Alarms


If the controller is equipped with a dialout modem, it can be programmed to notify a telephone or pager contact list when an alarm condition is true. Before proceeding the I/O conditions must be defined. Refer to section 5.19.4. These steps program a dialout alarm when the condition defined for I/O1 is true. 1. Select YES to enable the dialout alarm.
DIALOUT WHEN IO1 BECOMES TRUE? YES NO

If the phone number requires extended characters, or if you need to designate a number as a pager, press the decimal key. The controller displays:
SELECT CODE PAGER # *

Use an arrow key to select a code and press Enter to insert it in the number on the DIALOUT NUMBERS screen. 3. Enter the delay time between each attempt. A delay will allow time between attempts so you can return a call to the controller to acknowledge the alarm condition.
__ MINUTE DELAY BETWEEN CALLS (0-99)

The controller asks the same question for IO2 and IO3. Select NO.
DIALOUT WHEN IO2 BECOMES TRUE? NO YES

DIALOUT WHEN IO3 BECOMES TRUE? YES NO

Note: Numbers may be removed from the list by advancing the cursor to the very end of the number with the right arrow key, then pressing the left arrow key to erase the number.

2. Using the number entry keys, enter as many as 3 telephone numbers to the contact list.
DIALOUT NUMBERS: 1. 9,55555550001 2. 9,55555550002 3. P5550003

5-39

6712 Portable Sampler Section 5 Extended Programming When the call is answered, the samplers modem announces the alarm condition (analog talking modem) or sends a digital text message (external cellular modem), and the samplers three-digit ID number. To acknowledge the call, press the (star) key, and the three-digit ID number on your phone. The sampler will reply with an acknowledgment when the keyed entry is correct.

A record of alarms and acknowledgements is available in the Sampling Report. When the sampler detects the alarm condition, it records the type of alarm and the time it occurred. Acknowledgements are a separate Sampling Report entry, recording the time and telephone number that responded to the call. As mentioned previously, as many as three telephone numbers may be entered. The modem can dial up to 25 digits. Enter numbers using the numeric keys on the sampler keypad. When entering telephone numbers longer than 17 digits, the display will scroll. Extended characters, the Comma (,), Pound (#), and Star (*), also may be included in the 25 digits. To enter an extended character, press the Decimal Point key on the sampler keypad. The sampler displays a screen used to select and enter the extended characters. Each comma entered will cause the modem to pause for two seconds before proceeding with the remaining digits. Phone numbers that are no longer needed can be removed. Advance the cursor to the end of the telephone number and then press the Left Arrow key to erase the preceding character. If the cursor is not at the end of the telephone number, the Left Arrow key will only shift the cursor without erasing any characters. Dialout telephone and pager numbers are included in the Program Settings Report.

Note
Connecting either the serial output or the standard interrogator cable to the sampler disables the internal dial-out modem. The interrogator or serial output cable must be disconnected before you can use the modem. 5.19.5 Pager Numbers The extended character screen contains a pager option that designates the phone number as a pager. When programmed as a pager number, an identification string will be sent seven times after the pager number is dialed, with a three second delay between strings. An example identification string would be 828-1, meaning Alarm condition #1 exists at sampler 828. If more than one alarm condition exists, the string would read similar to 828-1-3. 5.19.6 Digital Modem Text Messaging The sampler can send alarms in the form of text messages containing the same information as that found in voice messages. To use this option, you must have: a cellular phone with text messaging capabilities

5-40

6712 Portable Sampler Section 5 Extended Programming an external digital modem recommended: Teledyne Iscos CDMA cellular modem part # 60-5314-489 or GSM cellular modem part # 60-5314-563

Note
If an internal voice modem is installed, the sampler will use that modem, and text messaging will be disabled. When you enter each of the (up to three) phone numbers for the sampler to call, if text messaging is active, a request for the TAP (Telocator Alphanumeric Protocol) server number and communications parameters will be made. As with voice messaging, the sampler will dial each telephone number up to five times per alarm condition, until the message is acknowledged. 5.19.7 Refrigerator Temperature Teledyne Isco offers a temperature sensor intended for use with the 6712FR Refrigerated Sampler. The optional sensor connects to the Rain Gauge port and uses I/O3 to transmit its temperature readings to the 6712.

Note
If alarms are being used, the temperature sensor is simply plugged into the panel mount on the one- or two-alarm output module. Because the temperature sensor uses I/O3, it cannot be used with a three-alarm output module. To configure an attached temperature sensor, select FR-TEMP a s t h e i n p u t f o r I / O 3 . T h e 6 7 1 2 w i l l d i s p l ay t h e N E W HARDWARE SETUP screen. Select DONE and the 6712 reconfigures its memory partitions to record the temperature readings. Temperature readings are updated approximately every 2.5 seconds. The readings are stored as FR-TEMP readings at the data storage interval. You can retrieve a history of temperature readings through the 6712s reports. The temperature data also can be used as a sampler enable condition. 5.19.8 Analog Output Teledyne Isco offers three analog outputs as a factory installed option. The option is installed inside the controller and includes a six-pin connector to interface with external devices. The analog outputs can be programmed for either 0-to-20 mA or 4-to-20 mA operation. With the exception of rain, any parameter that the 6712 sampler is currently measuring can be used to control any (or all) of the analog output channels. Programming the outputs requires parameter levels at each end of the analog range and then assumes a linear relationship between those points. MANUAL CONTROL of each output is also available.

5-41

6712 Portable Sampler Section 5 Extended Programming

Note
The internal analog output can not be installed in a controller that has the internal voice modem also installed. Contact the factory for information regarding external cellular modem options. Flowcharts for analog output screens appear in Appendix A.

5.20 Memory

The 6712 Sampler contains 512 kilobytes of battery backed RAM (Random Access Memory) that stores: Five sampling programs The sampling results from the most recent program The rainfall data collected from rain gauges The data from a refrigerator temperature sensor The data collected from a 700 Series module Parameter data collected from an SDI-12 Sonde Calculated flow information from the current flow module The sampling results memory stores up to approximately 1004 events (the actual number is dependent on the number of SDI-12 parameters). The program run and start time information is logged as four events. These four entries are held for the duration of the program. The remaining 1000 are for logging sample events, program events, and errors from the currently running program. If the running program exceeds 1000 events, the data will roll over. That is, the sampler will replace the oldest stored event with the newest event. When using an SDI-12 sonde, the sampler logs additional entries at each sample event. Module, sonde, refrigerator temperature, and rain gauge readings are stored in memory partitions. The storage capacity depends on the reading types, the number of bytes per reading, and the data storage interval. The sampler clears the sampling results memory when the site description changes. It is important to collect the sampling results before changing the site description the data can not be retrieved afterwards. Keep in mind that the description may change when selecting an extended program. The partition data is cleared when the sampler automatically creates and sizes the memory partitions. This occurs when you: Select a new program with a different storage interval Change the Data Storage Interval in the current program Change the programming mode Install a new 700 Series Module Change 700 Series Module mode of operation Add or remove a rain gauge with the Hardware Setup

5-42

6712 Portable Sampler Section 5 Extended Programming Add or remove a refrigerator temperature sensor as I/O3 Add or remove a SDI-12 sonde parameter with the Hardware Setup Reinitialize the sampler The stored data is lost as the sampler reformats its memory. As a precaution, before clearing the memory, the sampler display will show the following message:

INTERVAL CHANGED -DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

If you want to keep the stored data, download it before proceeding. When DONE is selected, the sampler clears the stored data and reformats the memory partitions. The sampler then displays the new capacity of the partition memory.

PARTITION MEMORY WILL LAST 13 DAYS

If the duration is too short, try changing the data storage interval or eliminating unnecessary reading types in the Hardware or Module Setup. For example, a sampler configured with a 730 Bubbler Module and a rain gauge storing data at a one-minute interval, creates level and rainfall partitions that hold 100 days of readings. Changing the storage interval to two minutes extends the storage capacity to 201 days. The sampler does not have to be running a program to store readings. When the sampler is on, it stores readings at the selected data storage interval. Once the partition is filled, the readings roll over. When the sampler is running a program, the partitions become triggered by the programmed start time or the first enable time, whichever is later. A triggered partition continues to log readings at the selected interval, but will not replace any readings taken after the trigger event. The sampler fills the partition and then stops recording to preserve the data. The first reading preserved in a full partition is one hour before the trigger event. A full partition can be reset by starting another program or changing storage parameters. If you need continuous readings for analysis, be sure to collect the readings before the partition fills or rolls over to avoid gaps in the data.

5-43

6712 Portable Sampler Section 5 Extended Programming

5.21 Pressurized Lines

The sampler can obtain samples from pressurized lines, as long as the line pressure remains below 15 PSI (pounds per square inch). Pressures greater than 15 PSI may prevent the sampler from purging the suction line; moreover, extreme pressures can force liquid past the pump, even when the pump is not running. For more consistent sample volumes from lines under constant pressure, calibrate the samples. For information about sampling with line pressures greater than 15 PSI, contact Teledyne Isco.

5.22 Command Driven Operation

The 6712 controller can be used as a slave to an external controller. The external controller is responsible for determining when to take a sample, how much volume to pump, and where to put the sample. The external controller directly interfaces to the sampler via an RS-232 communications port at 2,400-19,200 baud, 8 data bits, no parity, and 1 stop bit. A comma-separated-value protocol is used by the external controller to make requests, and by the sampler to report results. Use a standard computer connect cable (see Appendix E) to connect the samplers interrogator port and the external controller. For information on serial data codes and their meanings, refer to Table 5-5 Serial Data Codes in this section. The 6712 Sampler will respond to three commands: Turn Sampler On, GET STATUS, and TAKE SAMPLE. All commands must include the instruction set. The checksum value is optional. The available commands are: 1. Turn Sampler On: STS,2,CS,582<CR> 582 is the byte-by-byte sum of the ASCII values in the string STS, 1,CS. <CR> is a carriage return signifying the end of the command. If the sampler is OFF, this command will turn it ON before returning status info. 2. Get Status: STS,1,CS,581<CR> STS,1 instructs the sampler to return the current status. CS,581 is the optional checksum value. The byte-by-byte sum of the ASCII values in the string STS,1,CS is 581. <CR> is a carriage return signifying the end of the command. 3. Take Sample: BTL,2,SVO,100,CS,1039<CR> The TAKE SAMPLE command is only valid if the sampler is WAITING TO SAMPLE. BTL,2 instructs the sampler to place the sample in bottle 2. SVO,100 instructs the sampler to take a 100 ml sample.

5-44

6712 Portable Sampler Section 5 Extended Programming CS,1039 is the optional checksum value. The byte-by-byte sum of the ASCII values in the string BTL,2,SVO,100,CS, is 1039. <CR> is a carriage return signifying the end of the command. 5.22.1 Command Driven Sampler Responses The 6712 Samplers response string is the same for each of the three commands. A sample response string is shown below: MO,6712,ID,2424741493,TI,35523.50000,STS,1,STI,35523.4187 5,BTL,2,SVO,100,SOR,0,CS,xxxx<CR> Note that the samplers response is a series of alphabetical and numerical pairs. The alphabetical characters are headings that describe the following numerical value or code. A complete description of each pair is listed below. MO,6712, is the model number of the sampler. ID,2424741493, is the samplers unique identification number. TI,35523.50000, is the samplers current time. STS,1, is the samplers current status. The possible states are: 1 = WAITING TO SAMPLE. 4 = POWER FAILED (for short time after power is restored). 5 = PUMP JAMMED (must be resolved before continuing). 6 = DISTRIBUTOR JAMMED (must be resolved before continuing). 9 = SAMPLER OFF. 12 = SAMPLE IN PROGRESS. 20 = INVALID COMMAND. identifier code is not supported. bottle requested is not in current configuration sample volume requested is outside its range (10 - 9990 ml) day (Set_Time) must be 5 digits and more recent than 1977 21 = CHECKSUM MISMATCH. 22 = INVALID BOTTLE. (bottle requested is not in the current configuration) 23 = VOLUME OUT OF RANGE. ( the sample volume requested is outside its range (10-9990 ml) STI,35523.41875, is the most recent sample time. BTL,2, is the bottle receiving the most recent sample. SVO,100, is the most recent samples volume. SOR,0, is the results of attempting the most recent sample. The possible results are: 0 = SAMPLE OK. 1 = NO LIQUID FOUND. 2 = LIQUID LOST (not enough liquid to satisfy requirements.) 5-45

6712 Portable Sampler Section 5 Extended Programming 3 = USER STOPPED (using the Stop Key). 4 = POWER FAILED. 5 = PUMP JAMMED. 6 = DISTRIBUTOR JAMMED. 8 = PUMP LATCH OPEN. 9 = SAMPLER SHUT OFF (while sampling). 11 = NO DISTRIBUTOR. 12 = SAMPLE IN PROGRESS. CS,xxxx is the byte-by-byte sum of the entire string MO,...,CS, For commands that are not understood (INVALID COMMAND or CHECKSUM MISMATCH) or when no samples have been taken, the most recent bottle information is left off the response string.

5-46

6712 Portable Sampler


S ec tion 6 S DI-12 S ondes

In order to provide you with the flexibility to use the water parameter sensors that best meet your needs, Teledyne Isco has implemented support for SDI-12 protocol. Support of this standard allows you to use SDI-12 compatible devices in conjunction with your 6712 sampler in Extended Programming mode.

6.1 SDI-12 Data Parameters

The 6712 Sampler can accept up to 16 parameters from up to ten sensors with SDI-12 addresses from 0 - 9 (refer to chart in Figure 6-1). This data can be used in different ways. Data can be logged at user selectable time intervals for later download and processing by Teledyne Iscos Flowlink program. You can set up sampling programs that will be enabled to run only if conditions based on parameters monitored by the connected SDI-12 devices are met. The sampler can also be programmed to take samples each time certain parameter limits are exceeded. The data parameters used by the sensing device (sonde) and recording device (sampler) must match. Otherwise, the equipment could record the data values for one parameter as anothers, such as reporting a pH value as being temperature. Certain vendors sondes (such as Hydrolab and YSI) have enhanced commands that facilitate plug and play setup. These Teledyne Isco Ready sondes can tell the 6712 what values they have, their order, and units of measure. Other SDI-12 compatible sondes can be used, but require manual setup to identify proper data types for each data value reported. This section shows you how to set up both Teledyne Isco Ready and other SDI-12 sondes for use with the 6712.

WARNING
SDI-12 Sondes have not been approved for use in hazardous locations as defined by the National Electrical Code.

CAUTION
Before you install any sonde, you must take the proper safety precautions. Refer to Appendix C in the back of this manual. Additional safety considerations may be required. Check applicable guidelines, codes, and regulations of federal, state, city, and county agencies.

6-1

6712 Portable Sampler Section 6 SDI-12 Sondes

SDI-12 Sonde Data Units:

Parameter Temperature Conductivity Specific Conductance Total Dissolved Solids Salinity Dissolved Oxygen pH ORP Level Ammonium-Nitrogen Ammonia-Nitrogen Nitrate-Nitrogen Turbidity Chloride Chlorophyll Other

Range -40 - 100 C 0 - 100 mS/cm 0 - 100 mS/cm 0 - 90 g/l 0 - 70 ppt 0 - 20 mg/l 0 - 14 -999 - 999 mv 0 - 6.5279 m 0 - 200 mgN/l 0 - 200 mgN/l 0 - 200mgN/l 0 - 5000 NTU 0 - 1000 mg/l 0-400 ug/l -99999.0-99999.0

Resolution 0.1 C 0.01 mS/cm 0.01 mS/cm 0.01 g/l 0.1 ppt 0.1 mg/l 0.1 0.1 mv 0.0001 m 0.1 mgN/l 0.1 mgN/l 0.1 mgN/l 0.1 NTU 0.1 mg/l 0.1 ug/l 0.0001

Storage Bytes 2 2 2 2 2 1 1 2 2 2 2 2 2 2 2 4

Figure 6-1 SDI-12 Sonde Parameters

6.2 Scanning For Sondes

You must be in Extended Programming to use a sonde with the 6712. The sampler can automatically scan for the presence of sondes if you enable that option: From the main menu, select OTHER FUNCTIONS. Select HARDWARE. As you press to move through the hardware screens, you will see one that looks like the following:

NO SDI-12 SONDE AUTO SDI-12 SCAN OFF

Select NO SDI-12 SONDE to scan for sondes. If one is found, you can set it up, and the AUTO SDI-12 SCAN option will be switched to yes. Select the AUTO SDI-12 SCAN option to manually turn on (or off) automatic scanning for sondes when the sampler is turned on. If a sonde has already been connected and set up, the hardware screen will instead display a screen similar to the following:

SDI-12 DATA: TEMP0 pH0 SD_CO0 SAL0 05DATA0 DO0 ORP0 LEVEL0 09DATA0

6-2

6712 Portable Sampler Section 6 SDI-12 Sondes Press the arrow keys until SDI-12 DATA is blinking, then press . You will see the screen below:

NO SETUP SDI-12 SONDE KNOWN SONDES SCAN SDI-12 SONDES

Select SCAN SDI-12 SONDES and press . The system will scan to see what sondes are connected. If it detects a different sonde, it will indicate so, and advance to the setup screens.

6.3 Teledyne Isco Ready Sondes


Connecting a Teledyne Isco Ready Sonde
1 SELECT ADDRESS OF SONDE TO SETUP: __ __ DONE

If the 6712 scans for sondes and finds one that is Teledyne Isco Ready, it will automatically recognize the sondes data types and units. You can then use Hardware options to setup the definitions for data storage and calibrate the sonde. Even with a Teledyne Isco Ready sonde, the 6712 may find a parameter that is not on its list of supported parameters (Figure 6-1). It can still work with such a parameter. Such a parameter will be identified as something like 05DATA1 and you will have to mentally associate the appropriate data type with that label. However, once such data is brought into Flowlink, data sets can be manually changed so that reports show the appropriate labels.

Select the address you want to set up.


Company Model Optional Information SONDE AT ADDRESS __ HAS __ DATA VALUES

Wait a few seconds for the above display to disappear, or press .


You will then be asked if you want to store the parameter data. Refer to SONDE SETUP - STORING PARAMETER DATA on page 6-4.

6.4 Other SDI-12 Sondes

If you have connected an SDI-12 sonde that is not Teledyne Isco Ready, you will need to define the sondes data values. When you select the address of the sonde you want to set up (see screens in the margin), the company model and other information will be displayed along with the number of data values the sonde has. When you advance to the next screen (see #3 below), the data types will be listed as OTHER because you have not defined them yet. Move between the data types using the arrow keys. Select one to change by pressing . A list of possible parameters will be displayed (see #4 below). Use the arrow keys to move through the options. When the one you want is blinking, press . Repeat the process until you have defined all of the data parameters for your sonde. Use the arrow key to move to DONE and press . You will then be asked if you want to store the parameter data.

6-3

6712 Portable Sampler Section 6 SDI-12 Sondes

Connecting Other SDI-12 Sondes


3 SELECT ADDRESS OF SONDE TO SETUP: __ __ DONE

Make your selection and press .


<-DATA FROM SONDE __ 1 = TEMPERATURE(oC) 2 = OTHER 3 = OTHER --->

Select the address of the sonde you want to set up.


Company Model Optional Information SONDE AT ADDRESS __ HAS __ DATA VALUES

Repeat the process until you have defined all the data. Press DONE when finished.
<--DATA FROM SONDE __ 4 = pH 5 = LEVEL DONE

You can wait a few seconds, or press to advance to the next screen:
DATA FROM SONDE __ 1 = OTHER 2 = OTHER 3 = OTHER --->

Press an arrow key until the cursor is on the data you want to identify. Then, press . A scrolling screen listing possible parameters will appear.
<-SONDE ___ , DATA 1: pH TEMPERATURE (oC) TEMPERATURE (oF) --->

You will then be asked if you want to store the parameter data. Refer to SONDE SETUP - STORING PARAMETER DATA in the following paragraphs.

6.5 Sonde Setup - Storing Parameter Data

Whether you are using a Teledyne Isco Ready sonde or other SDI-12 compatible sonde, you will need to tell the 6712 if you want to store all of the parameter data or data from selected parameters. Data is stored in memory partitions. A fixed amount of memory is available to store data. If you are using some parameters to simply trigger an event or enable condition, it is not necessary to store the data. If all of the parameters data will fit in memory, you will be asked if you want to store all of the parameter data. Answering NO will still allow you to selectively store parameter data; otherwise, all data will be stored. You will be asked to store parameter data during the setup process. You can also manually change the setup using options in the Hardware menu. To access the setup option in the Hardware menu, select Other Function from the main menu, then choose Hardware. Advance through the screens until you see one that displays SDI-12 DATA.

6-4

6712 Portable Sampler Section 6 SDI-12 Sondes Select this screen to display the following:

NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES

Select SETUP KNOWN SONDES and then indicate the address of the sonde you want to setup. A screen containing the sonde model and number of data values will appear briefly. Then you will be asked to verify, parameter by parameter, what data you want to store. The screens will look similar to the one below:

SONDE 5, STORE TEMPERATURE DATA? YES NO LABEL IS: TEMP5

Indicate if you want to store the data for that particular parameter and then advance through the rest of the parameter screens. Once you have defined your data storage options, select DONE. You will be prompted that you have entered a new hardware setup and asked if you want to erase the stored data. Indicate YES or NO. A partition memory status message will be displayed and (if a parameter exists that can be calibrated) you will then advance to the calibration screen (refer to SONDE CALIBRATION AND VALIDATION on this page). If you do not want to calibrate or validate, select DONE to return to the Hardware Options screens.

6.6 Sonde Calibration and Validation

It is recommended that calibration of your sonde be done in the lab, according to manufacturers recommendations. However, for your convenience, the 6712 allows for field calibration of pH, Dissolved Oxygen (DO), and Oxidation Reduction Potential (ORP). Field calibrations are stored in the sampler, NOT the sonde, and are used to modify readings from the sonde until a new sonde is connected. When a sonde is moved to another sampler, it will retain calibration constants set in the lab, but will not retain any that were set in the field. Although field calibration is limited to three parameters, you can validate any of the available parameters. Validation allows you to immerse the sonde in a known solution, note the readings, and see if the sonde is functioning properly. The validation process does not modify any readings like calibration does; it merely alerts you to inaccurate ones. No data will be stored, nor will enable conditions be updated, while calibrations or validations are being performed.

6-5

6712 Portable Sampler Section 6 SDI-12 Sondes 6.6.1 The Calibration Screens To access the calibration screens from the main menu, select Other Functions, Hardware. Advance through the screens until you see one that displays SDI-12 DATA. Select this screen to display the following:

NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES

Select SETUP KNOWN SONDES and then select DONE. You will see a screen similar to the following that will list the parameters you can calibrate.

SELECT PARAMETER TO CALIBRATE: pH5 DO5 ORP5 DONE

The sondes address will appear after the parameter (such as pH0 and pH5). If you have multiple sondes attached, use the arrow key to select the correct one to calibrate. Depending upon the parameter you selected, a series of screens will be displayed to step you through the calibration process. pH - You will be given the choice of three different pH solutions: 4.00, 7.00, and 10.00. After making your selection, rinse the probe and place in the pH buffer. If the reading is off by more than 0.5 pH units, you will asked to verify if the reading is correct. DO - For dissolved oxygen, you can choose between two methods: DO Standard and Saturation. The saturation method requires a temperature reading between 0.0 and 47.0 C. You will also need to enter your altitude and barometric pressure. ORP - You will asked to enter the value of your calibration standard. When calibration is complete, select DONE to proceed to the validation screen. If you do not want to validate any parameters, select DONE again to return to the Hardware options screens. 6.6.2 The Validation Screen The validation screen appears after you select DONE on the calibration screen. It will be similar to the screen shown below.

SELECT PARAMETER TO VALIDATE: <-SAL5 05DATA5 DO5 ORP5 LEVEL5 -->

6-6

6712 Portable Sampler Section 6 SDI-12 Sondes Use the arrow keys to select the desired par am eter to validate. Any parameter can be validated. You will be prompted to place the probe in your validation standard. When stable, press . The reading will be displayed at the bottom of the screen. When you are done validating, select DONE to return to the Hardware options screens.

6.7 Disconnecting Sondes

When you have disconnected all sondes, and have interrogated the sampler for all its data, you can indicate that there is no SDI-12 sonde. From the main menu, select Other Functions, Hardware. Advance through the screens until you see one that displays SDI-12 DATA. Select this screen to display the following:

NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES

Select NO SDI-12 SONDE. You will advised that you have a new hardware setup:

NEW HARDWARE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

If you have collected all the data, select DONE. A partition memory message will be displayed briefly, and you will be returned to the Hardware options screens. If you do not have a sonde attached, you may want to turn off automatic scanning for sondes. This will eliminate the time taken to scan for sondes whenever the sampler is turned on.

6-7

6712 Portable Sampler Section 6 SDI-12 Sondes

6-8

6712 Portable Sampler


S ec tion 7 Remote Operation

The 6712 controller can be operated remotely from external devices. External controlling devices may be a computer via RS-232 serial communications or a telephone via touch-tone comm a n d s. R e m o t e t e l e p h o n e o p e r a t i o n r e q u i r e s t h e factory-installed 2400 baud talking modem.

Note
The internal modem cannot be installed in a controller with the analog output option also installed.

7.1 Computer Operation

The 6712 controller can be operated from a remote location using a computer. There are two different levels of computer control: Menu control is used to start and stop the current sampler program (or programs if in Extended programming mode) and to access other menu commands. This level of remote operation relies on the sampler program for much of its capabilities. External program control ignores the samplers program and current state. Using external control, you can turn on the sampler, take a sample, and request sampler status data. The pump, distributor, and serial data output simply respond to the external commands. To operate the sampler remotely on either level, you must first establish RS-232 serial communication between the computer and the 6712 controller. To do so: 1. Connect your computer to the 6712 controllers interrogator port with a standard interrogator cable, P/N 60-2544-044 (for 9-pin serial ports). 2. On your computer, open a terminal emulator program such as HyperTerminal, and configure it for: baud rate of 19,200 or less no parity 8 data bits 1 stop bit VT100 terminal emulation 3. Using your communications software, connect to the sampler. Press and hold the question mark (?) on the computer keyboard until the sampler returns its banner string. A caret (>) prompt will appear on the screen when the sampler is ready to accept commands.

7-1

6712 Portable Sampler Section 7 Remote Operation

Figure 7-1 Press and hold ?? to connect to the sampler 7.1.1 Menu Control Once you have remotely connected to the sampler, menu commands can be entered from your computer. These menu commands are listed in Table 7-1. To access the menu commands, type MENU at the > prompt and press ENTER. The screen will return a list of the available options. Note that the current state of the controller will determine the available menu options. Then, type the desired command(s), confirming with ENTER.

Figure 7-2 Example list of remote menu commands

7-2

6712 Portable Sampler Section 7 Remote Operation

Table 7-1 6712 Remote Menu Commands


Menu Command Description
0 or START 1 or RUN1 This command will immediately start any program set to WAIT FOR PHONE CALL and currently waiting for the remote start command (see Section 4.6). This command will load Program 1 as the samplers current Extended program, reset partition if the storage interval has been changed, and run the program. RUN1 is valid when the sampler is in standby, or when it is waiting for a start command. RUN1 will also run the current program while in Standard programming. These commands will load the selected program number as the samplers current Extended program, reset partition data if the storage interval has changed, and run the program. They are valid when the sampler is in standby, or when it is waiting for a start command. The commands are available for Extended programming only. This command is similar in functionality to a pin F low signal (a signal typically used to enable or disable the sampler from a flow meter). At run time, the sampler starts with the sampler in the enabled state. DISABLE is valid while a program is running. If successful, a display of REMOTE ENABLE IS FALSE will be returned. A Remote Disable is the highest priority disable. If the sampler is disabled with this command, no condition other than menu command 6, ENABLE, will allow the program to become enabled again. 6 or ENABLE This command is used to re-enable a disabled sampler. ENABLE is valid while a program is running. If successful, a display of REMOTE ENABLE IS TRUE will be returned.

2 or RUN2 3 or RUN3 4 or RUN4 5 or DISABLE

7 or TAKE_SAMPLE This command causes the sampler currently running a program to take a sample. The sample will count as one of the programs samples and is placed in the current bottle (Part As current bottle when using two-part programming). TAKE_SAMPLE is valid while the program is running, after the start time, but not while in the MANUAL PAUSE screens. If successful, a response similar to PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a response similar to NO SAMPLE, PROGRAM WAITING START will be returned. ST or STATUS SD or SCREEN_DUMP This command causes the sampler to return its status information and current parameter data. This command causes the sampler to return the information currently displayed on the samplers screen. The remote screen follows the samplers display as the information is updated. This mode can be used to observe from a remote location while the sampler is programmed on-site. This command causes a sampler currently running a program to enter the MANUAL PAUSE menu. While you are connected in this mode, the sampler keypad becomes disabled, and your computers keypad will be redirected to simulate the sampler keypad. The following keys will be activated: <Enter>, decimal point, and number keys. <Esc>, S, or s = STOP. <Backspace>, L or l, and U or u = Left/Up arrows. R or r, and D or d = Right/Down arrows.

P or PAUSE

Sampling reports

You can retrieve the sampling reports at any time using the same serial connection. Type Q to exit the menu. Then at the > prompt, type REPORT and press Enter. The sampler controller will send reports as configured.

7-3

6712 Portable Sampler Section 7 Remote Operation 7.1.2 External Program Control Under external program control, the sampler operates according to received commands. An external program can be as simple as manually entered commands to collect a sample when needed, or as complex as a user-developed process control program (for SCADA or other systems) that directs sample collection according a systems needs. The 6712 controller will respond to four commands: Turn on the sampler Take a sample Send status Send data Note that commands and responses are comma-separated pairs. The first set of alpha-characters is an identifier; the second set of numeric characters indicates the value. The 6712 controller will not receive and process the command unless it is followed by a carriage return. Turn on the sampler If the sampler is OFF, the sampler responds by placing itself in the ON state and returning a status string to report the new status. (See Send Status for a description of this string.) The command to turn the sampler ON and return status is: STS,2<CR> Take a Sample This directs the sampler to take a sample. The command specifies how much volume to take, and which bottle to place the sample in. The sampler responds by taking a sample and returning a status string that reports the new status. An example of a command to take a sample is shown below: BTL,2,SVO,100<CR> This example instructs the sampler to place the distributor arm over bottle 2 (BTL,2) and take a 100 ml sample (SVO,100). While most settings of the sampler program (start times, enables, distribution schemes, etc.) are ignored when responding to external commands, the sampler does rely on two program settings that configure the sampler. These settings are: Number of Bottlesis used to determine the distributor arm position and the acceptable range for the BTL values. Suction Line LengthUsed with liquid detection to deliver correct sample volumes according to the SVO value.

7-4

6712 Portable Sampler Section 7 Remote Operation Send Status This command causes the sampler to return a data string that identifies the controller and lists its current status. The command to send current status is: STS,1<CR> The status is received as pairs of comma-separated identifiers and values. A complete description of each pair is listed below. MO,6712, is the model number of the sampler. ID,2424741493, is the samplers unique ID number. TI,35523.50000, is the samplers current time, shown in a date-time format based on the number of days since 00:00:00 1-Jan-1900, and the time shown as a fraction. The possible states returned by the STS,1 command are: 1 = WAITING TO SAMPLE. 4 = POWER FAILED (for short time after power is restored). 5 = PUMP JAMMED (must be resolved before continuing). 6 = DISTRIBUTOR JAMMED (must be resolved before continuing). 9 = SAMPLER OFF. 12 = SAMPLE IN PROGRESS. 20 = INVALID COMMAND. Possible causes may be: identifier code is not supported. bottle requested is not in current configuration sample volume requested is outside its range (10 - 9990 ml) day (Set_Time) must be 5 digits and more recent than 1977 21 = CHECKSUM MISMATCH. (see Optional checksum on page 7-8) 22 = INVALID BOTTLE. (bottle requested is not in the current configuration) 23 = VOLUME OUT OF RANGE. (the sample volume requested is outside its range (10-9990 ml) STI,35523.41875, is the most recent sample time. BTL,2, is the bottle that received the most recent sample. SVO,100, is the most recent samples volume. SOR,0, is the results of attempting the most recent sample. The possible results are: 0 = SAMPLE OK 1 = NO LIQUID FOUND 2 = LIQUID LOST (not enough liquid) 3 = USER STOPPED (using the Stop Key) 4 = POWER FAILED 5 = PUMP JAMMED

7-5

6712 Portable Sampler Section 7 Remote Operation 6 = DISTRIBUTOR JAMMED 8 = PUMP LATCH OPEN 9 = SAMPLER SHUT OFF (while sampling) 11 = NO DISTRIBUTOR 12 = SAMPLE IN PROGRESS CS,xxxx is the byte-by-byte sum of the entire string MO,...,CS, (see Optional checksum on page 77-8)

Note
For commands that are not understood (INVALID COMMAND or CHECKSUM MISMATCH) or when no samples have been taken, the most recent bottle information is left off the response string. Send Data This command causes the sampler to return the Send Status appended with data from any module, rain gauge, or SDI-12 device. Refer to Table 7-2 to interpret the returned string.

Note
This returns the same data as the programmable software option Serial Output, as described in Section 5.18.6 of this manual. The command to send data is: DATA<CR>

Table 7-2 Serial Data Codes


Identifier Parameter
B?1 CR CS3 DE FL ID LE MO PE PH PR RA RTE SS TE Bottle number and time Current days rain Checksum Description Flow Unit specific identifier Level Model Percent pH Previous days rain Rain Refrigerator Temperature Sampler enable state Temperature

Units
days since 1-Jan-1900 tips2 unsigned long 6712 Sampler cubic meters per second 10 digit unsigned long meters 6712 percent of full scale none tips2 tips2 degrees celsius logical degrees celsius

7-6

6712 Portable Sampler Section 7 Remote Operation

Table 7-2 Serial Data Codes (Continued)


Identifier Parameter
TI VE VO VSI VSP CO_
4

Units
days since 1-Jan-1900 meters per second cubic meters percent percent milliseimens per centimeter milligrams per liter none parts per thousand milliseimens per centimeter grams per liter degrees Celsius millivolts meters milligrams nitrogen per liter milligrams nitrogen per liter milligrams nitrogen per liter nephelometric turbidity units milligrams per liter micrograms per liter cubic meters per second cubic meters

Current time Velocity Volume Velocity signal strength Velocity spectrum strength Conductivity Dissolved oxygen pH Salinity Specific Conductance Total Dissolved Solids Temperature Oxidation reduction potential Level Ammonium-nitrogen Ammonia-nitrogen Nitrate-Nitrogen Turbidity Chloride Chlorophyll Flow rate Flow volume Other data not listed above

DO_ 4 PH_ SA_ SP_


4 4 4

TD_ 4 TE_ OR_ LE_


4 4 4

MM_ 4 MA_ NI_


4 4 4

TB_

CL_ 4 CP_ FL_ VO_


4 4 4

xxD_ 4, 5
1

The three most recent samples will have entries as B?. The ? will be the first bottle number in the set. If the sample is a grab sample, it will be shown as B0. Rainfall tips roll over every 256 counts. The checksum does not include the checksum, carriage return, and linefeed. The string is terminated with <CR><LF>. A _ in the identifier indicates an SDI-12 address. XX matches the data label for the SDI-12 parameter.

2 3

4 5

7-7

6712 Portable Sampler Section 7 Remote Operation Optional checksum Optionally, commands to the sampler can be validated using a checksum. Checksum validation can avoid problems caused by faulty program control or communication lines. When using the checksum, add the CS identifier and a comma at the end of the command, and the checksum value of the string up to the last comma. For example, the command with checksum validation to turn the sampler on would appear as: STS,2,CS,582<CR> The numeric value of each character is its ASCII equivalent expressed in decimal format. The checksum of 582 was calculated as shown below.

83 84 83 44 50 44 67 83 44 582 STS,2,CS,582
7.1.3 Remote Control of Sampler Keypad To control the sampler keypad remotely, type CONTROL at the > prompt and press ENTER. The sampler display appears on your computer monitor as you step through the programming screens. While in this mode, the computer keys will be redirected to simulate the samplers keypad, and the sampler keypad itself will be disabled to avoid any conflict. The active keys and their corresponding functions are given in Table 7-3.

Table 7-3 Remote Control of Sampler Keypad


Computer
<Esc>, S, s L,l,U,u, <Backspace> R, r, D, d O, o <Enter>, arrows, decimal, numbers

Sampler
STOP Left / Up Right / Down ON Same as sampler

NOTE: You will not be able to turn the sampler off remotely.

7-8

6712 Portable Sampler Section 7 Remote Operation

7.2 Telephone Commands

6712 controllers equipped with an internal voice modem can be operated using telephone commands. This factory-installed option includes a 12 ft (3.6 m) modem cable. After connecting to the sampler from a touch-tone telephone, command numbers may be entered from the telephones keypad, followed by the # (pound) button. The commands are listed in Table 7-4. To use telephone commands: 1. Dial the samplers telephone number. When the sampler connects, it answers with Teledyne Isco Sampler Site [number]. If the sampler does not receive a touch-tone command in five seconds, it assumes that a computer modem has called and initiates the modem connect sequence. If this occurs you must hang up and try again. 2. Enter the command. For example, to run program 1, press 1 # . 3. If the command is valid for its current state of operation, the sampler responds with [command number], please acknowledge. If the command is not valid the sampler responds with the command number followed by three quick beeps. Enter a valid command. 4. Acknowledge the valid command by pressing the number. For the example, press 1 . 5. If the number is incorrect, the sampler responds with your entry followed by three quick beeps. You must re-enter the command (step 2). 6. After the command is confirmed, the sampler replies with [command number] acknowledge and executes the command. To enter additional commands, repeat steps 2 through 5. The sampler will hang up after ten seconds of quiet time.

Table 7-4 6712 Remote Phone Commands


Phone Command Name
Start Load and Run Program 1

Description
This command will immediately start any program set to WAIT FOR PHONE CALL and currently waiting for the remote start command (see Section 4.6). This command will load Program 1 as the samplers current Extended program, reset partition if the storage interval has been changed, and run the program. RUN1 is valid when the sampler is in standby, or when it is waiting for a start command. RUN1 will also run the current program while in Standard programming.

1 #

2 3 4

# # #

Load and Run These commands will load the selected program number as the samplers curProgram 2, 3, or 4 rent Extended program, reset partition data if the storage interval has changed, and run the program. They are valid when the sampler is in standby, or when it is waiting for a start command. The commands are available for Extended programming only.

7-9

6712 Portable Sampler Section 7 Remote Operation

Table 7-4 6712 Remote Phone Commands (Continued)


Phone Command Name
Disable

Description
This command is similar in functionality to a pin F low signal (a signal typically used to enable or disable the sampler from a flow meter). At run time, the sampler starts with the phone enable in the enabled state. PHONE DISABLE is valid while a program is running. If successful, a reply of REMOTE ENABLE IS FALSE will be returned. A Remote Disable is the highest priority disable. If the sample is disabled with this command, no condition other than command 6#, ENABLE, will allow the program to become enabled again.

6 7

# #

Enable

This command is used to re-enable a phone-disabled sampler. PHONE ENABLE is valid while a program is running. If successful, a reply of REMOTE ENABLE IS TRUE will be returned. This command causes the sampler currently running a program to take a sample. The sample will count as one of the programs samples and is placed in the current bottle (Part As current bottle when using two-part programming). TAKE_SAMPLE is valid while the program is running, after the start time, but not while in the MANUAL PAUSE screens. If successful, a response similar to PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a response similar to NO SAMPLE, PROGRAM WAITING START will be returned.

Grab Sample

7-10

6712 Portable Sampler


S ec tion 8 Maintenanc e

This section describes how to maintain and troubleshoot your 6712 sampler. If you think your unit requires repair, or if you have questions concerning its operation or maintenance, contact Teledyne Iscos Technical Service Department: Phone: (800) 228-4373 (402) 464-0231 FAX: (402) 465-3022 E-mail: [email protected]

8.1 Maintenance Checklist

Inspect the pump tube for wear. Replace it if necessary. Clean the pump tubing housing. Change the suction line if necessary. Clean the bottles, suction line, strainer and pump tube. Check the humidity indicator. Check the controllers internal battery status and replace the battery every five years. If the keypad label has bubbles under it, the air inside the controller has expanded. To release this pressure, unscrew the flow meter cable or connector cap on the back of the controller. Push down on the label to force the air out, then re-tighten the cable connector or cap.

8.2 Cleaning Guidelines


8.2.1 Controller, Top Cover, Center Section, Retaining Ring, and Tub Clean the top cover, center section, retaining ring, and tub with warm soapy water or by spraying them with a hose. Avoid using a high-pressure hose to clean the controller, especially around the control panel. Extreme pressures may force water past the control-panel seal. When cleaning the controller and top cover, cap the connectors at the back of the controller tightly. Keep a power source connected to the power-source connector, or replace the protective cap (shipped with the sampler) over the power source connector, to protect the pins from moisture damage. Make sure the two pump drain holes (located on the bottom right-hand side of the pump, beneath the pump band) are open and free of debris or buildup. 8.2.2 Sample Bottles The sample bottles have a wide mouth to facilitate cleaning. Wash them with a brush and soapy water, or use a dishwasher. You can autoclave the glass bottles.

8-1

6712 Portable Sampler Section 8 Maintenance 8.2.3 Suction Line, Pump Tube, and Discharge Tube Clean the suction line, pump tube, and discharge tube by placing the end of the suction line in a cleaning solution and pumping it through the delivery system. Rinse with clean water. If these items are severely contaminated, discard them. For critical sampling applications, refer to the following sections. Clean the strainer with a brush and soapy water. Clean sampling equipment is essential for valid laboratory analysis. Teledyne Isco recommends that you develop cleaning protocols in consultation with a laboratory analyst when designing the monitoring program. For example, to clean sample bottles, suction line, and pump tubes, Lair (1974) suggests these protocols used by USEPA Region IV Surveillance and Analysis field personnel engaged in NPDES compliance monitoring. The protocols are based on U.S. Environmental Protection Agency Publications EPA-600/4-77-039 (Sampling of Water and Wastewater by Dr. Phillip E. Shelley). 1. One spectro-grade acetone rinse. 2. Dishwasher cycle (wash and tap water rinse, no detergent). 3. Acid wash with at least 20 percent hydrochloric acid. 4. Dishwasher cycle (wash and tap water rinse, no detergent). 5. Replace in covered Teledyne Isco tubs. 8.3.2 Vinyl Suction Line Use new suction line for each new sampling setup, therefore, no cleaning is required. Use PTFE suction line when sampling for organic compounds. 8.3.3 Teflon Suction Line 1. Rinse twice with spectro-grade acetone. 2. Rinse thoroughly with hot tap water using a brush, if possible, to remove particulate matter and surface film. 3. Rinse thoroughly three times with tap water. 4. Acid wash with at least 20 percent hydrochloric acid. 5. Rinse thoroughly three times with tap water. 6. Rinse thoroughly three times with distilled water. 7. Rinse thoroughly with petroleum ether and dry by pulling air through the line. 8. Dry overnight in a warm oven (use an oven temperature of lower than 150 F), if possible. 9. Cap ends with aluminum foil. 8.3.4 Teledyne Isco Pump Tube 1. Rinse by pumping hot tap water through the tube for at least 2 minutes. 2. Acid wash the tube by pumping at least a 20 percent solution of hydrochloric acid through the tube for at least 2 minutes.

8.2.4 Strainer

8.3 Cleaning Protocols for Priority Pollutants

8.3.1 Teledyne Isco Glass Sample Bottles

8-2

6712 Portable Sampler Section 8 Maintenance 3. Rinse by pumping hot tap water through the tube for at least 2 minutes. 4. Rinse by pumping distilled water through the tube for at least 2 minutes.

Note
For critical sampling applications, consider replacing the suction line, pump tube, and discharge tube. Replacement eliminates the possibility of any cross contamination from previous sampling sites.

8.4 Maintenance Screens

When you select OTHER FUNCTIONS from the main menu, and then select Maintenance, the following screen will be displayed:

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Using these options, you can set the samplers time and date, set the number of pump counts for the pump tube warning, reset the battery warning for the internal battery, and run a set of diagnostic tests for troubleshooting purposes. 8.4.1 Set Clock Entering Times and Dates
ENTER TIME AND DATE: HH:MM DD-MON-YY __:__ __-__-__

Teledyne Isco ships samplers from the factory with the clocks set to the correct Central Standard Time. Reset the clock when installing the sampler in a different time zone, for daylight savings time, or when needed. When setting the clock and calendar, use a 24-hour clock for times and the day-month-year format for dates. The illustration at the left shows blanks for each position that accepts an entry. The positions are called fields. When this screen appears on your sampler, the fields will contain the samplers current settings for the time and date. To move from one field to another without changing the setting, press the arrow keys. Use this technique when you want to change only one or two settings. Change the setting by typing a new number. Press to accept the new setting. For example, to enter 2:00 p.m. (2:00 p.m is 14:00 on a 24-hour clock), type 14. Press . Next, type 0 (zero) for the minutes, and press . To enter a date, such as January 22, 2001, type: 22 01 01 .

Move from one field to another by pressing the arrow keys. Change a setting in a field by typing a number. Press to accept the new setting.

8-3

6712 Portable Sampler Section 8 Maintenance

Table 8-1 Setting the Clock and Calendar


1
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

Select OTHER FUNCTIONS.

Select MAINTENANCE.

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS ENTER TIME AND DATE: HH:MM DD-MON-YY 14:00 22-JAN-01

Select SET CLOCK.

Enter the time and date.

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Press STOP to return to the main menu.

8.4.2 Pump Tube Alarm

The sampler tracks the pump counts in both the forward and reverse cycles with a resettable counter. When the counter reaches the default count of 1,000,000 or the count you have e ntere d, the sa mp le r d is p lay s a m ess ag e, WA RN IN G ! REPLACE PUMP TUBE. The message appears each time you turn the sampler on or run a program until you reset the counter. After replacing the pump tube (see page 8-8), reset the count to zero so the sampler can begin tallying the pump counts for the new tube. Replacing the pump tube does not reset the counter. Experience may suggest a significantly different pump tube life. You can change the alarm count to represent the pump tube life for your application. The example below shows how to set the alarm count and reset the pump counts.

8-4

6712 Portable Sampler Section 8 Maintenance

Table 8-2 Resetting the Pump Counter


1
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

Select OTHER FUNCTIONS.

Select MAINTENANCE.

Select PUMP TUBE ALARM.

1024000 PUMP COUNTS, WARNING AT 1000000

The sampler displays this screen briefly to give you the current pump-count information. The first line shows you the pump counts tallied since the counter was last reset. The second line shows the current alarm setting. To reset the counter to zero, select YES. Always reset the counter after replacing a pump tube. Select NO when merely checking the current count. If necessary, change the pump-count alarm setting by typing the first two digits of the new setting. The sampler accepts entries between 1 and 99. For example, to increase the count to 1,500,000, enter 15. Press STOP to return to the main menu.

1024000 PUMP COUNTS, RESET PUMP COUNTER? YES NO

WARNING AT 1000000 PUMP COUNTS (1 - 99)00000)

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

8-5

6712 Portable Sampler Section 8 Maintenance 8.4.3 Internal Battery Warning The internal lithium battery preserves the samplers memory when power is removed. For an explanation of how to replace the lithium battery, go to Section 8.8.

Table 8-3 Checking and Replacing the Internal Battery


1
RUN PROGRAM VIEW REPORT OTHER FUNCTIONS MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE HARDWARE SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS WARNING! REPLACE INTERNAL BATTERY BEFORE 17-OCT-97 DID YOU REPLACE THE INTERNAL BATTERY? YES NO

Select OTHER FUNCTIONS.

Select MAINTENANCE.

Select INTERNAL BATTERY.

The sampler displays this screen briefly to show you the expiration date.

After replacing the battery, select YES. Select NO only if checking the expiration date. Resetting the alarm directs the sampler to calculate a new expiration date, five years in the future. If you reset the alarm without replacing the battery, the sampler will not be able to alert you to the batterys impending expiration. Press STOP to return to the main menu.

SET CLOCK PUMP TUBE ALARM INTERNAL BATTERY DIAGNOSTICS

8.4.4 Diagnostics

The sampler has several self-diagnostic tests that check the memory (both the RAM and ROM), pump, and the distributor arm position. It also contains screens that let you reset the samplers program settings and memory to factory settings. You do not need to run the diagnostic tests on a routine basis. The diagnostic tests are a troubleshooting tool; Teledyne Iscos Customer Service department staff may ask you to run these tests when they work with you in diagnosing problems with your sampler. Access the diagnostic screens by selecting Other Functions from the main menu, then Maintenance, and Diagnostics. The unit will automatically test the memory and display; you can choose whether it tests the pump and distributor, and whether it reinitializes the unit to its original factory settings.

8-6

6712 Portable Sampler Section 8 Maintenance

PLEASE WAIT! . . . RAM PASSED TEST PURGING . . . PLEASE WAIT! . . . ROM PASSED TEST

If RAM or ROM fails the test, the sampler will report that RAM (or ROM) FAILED TEST. Contact Teledyne Iscos Customer Service for assistance.
ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST

PURGING . . . ON/OFF RATIO = _.__

TEST DISTRIBUTOR YES NO

FINDING ARM FLEXURE

TEST PUMP? YES NO

NOW AT BOTTLE __

PUMPING . . .

RE-INITIALIZE? YES NO

PUMPING . . . ON/OFF RATIO = _.__

Memory Test Screens 1 and 2 at left show you the diagnostic tests for the units memory. If either screen shows you a message saying that the memory failed the test, contact Teledyne Iscos Customer Service. Display Test Screen 3 shows you the display test. You should see the test pattern illustrated in the screen. If the pattern does not match the illustration, the display or its circuitry require repair. Pump Test Screen 4 lets you start the pump test by selecting YES. As the sampler runs the test, it first runs the pump forward for a short period. Next, it displays an ON/OFF ratio number. After displaying the forward pumping ratio, the sampler reverses the pump to obtain a purge ratio. The acceptable range is between 0.80 and 1.25. If either ratio is outside the acceptable range, the pump requires repair. You do not need to pump liquid while running the test, and you can run the test with or without a tube in the pump.

8-7

6712 Portable Sampler Section 8 Maintenance

CAUTION
Tests indicate that the pump produces sound levels in excess of 85db at one meter. Prolonged exposure to this sound level could result in hearing loss and requires the use of protective ear plugs. Distributor Test The distributor test is provided for factory personnel to verify the distributors position as it rotates through the 24 positions. (screens 9 through 11). The distributor arm flexes slightly when it contacts the stop, and the sampler measures this flexure. The sampler uses this measurement to position the arm accurately over each bottle. Reinitialize Controller The sampler allows you to reset all the program settings to the settings shipped with the sampler (screen 12). Reinitializing also clears the memory that stores the samplers event log and module data. Use this option cautiously.

8.5 Replacing the Pump Tube

Replace the pump tube only with Teledyne Iscos 6712 pump tubing. Other pump tubes will not work. The 6712 pump tubing is easily recognized by the blue alignment collars. Improper pump tubes include those made for Teledyne Iscos other samplers (3700, GLS, etc.) and tubing from non-Teledyne Isco vendors. Also note that the discharge tube is not the same as the pump tube. You could experience several problems if you install the wrong pump tubing: The sampler will not pump the liquid. Pump jams Inaccurate sample volumes Faulty liquid detection Inspect the pump tube periodically. Replace the tube when it cracks or appears worn. Inspect the tube frequently when the sample liquid contains a high percentage of suspended or abrasive solids.

Note
The importance of regular tubing replacement cannot be overstated. The key is to replace the tube before failure, not after. When a pump tube ruptures, grit and other abrasive debris can be driven into the pump shaft seal. Over time, this abrasive material will degrade the pump seal, jeopardizing the NEMA 4x 6 rating of the controller. Failure to maintain the pump tube may result in permanent damage to the sampler. Check the condition of the pump tube regularly and if the tube shows signs of fatigue or wear, replace it immediately. A properly maintained sampler will provide the years of reliable service that is expected of a Teledyne Isco Sampler.

8-8

6712 Portable Sampler Section 8 Maintenance 8.5.1 Checklist For Replacing Pump Tube 1. Disconnect power from the sampler.

WARNING
The pumps safety lock prevents the sampler from running the pump when the pump band is open. DO NOT tamper with the safety mechanism. The pump is extremely powerful. The pump rollers can injure you severely if the sampler activates the pump while you are working on it or inside it. Disconnect power from the sampler before replacing the pump tube. 2. Loosen the liquid detectors cover by unscrewing the black knob on top of the detector. Unlatch the pump band. (The band is the rounded metal band that holds the tube in the pump.) 3. Pull the tube away from the bulkhead fitting. Pull it from the pump and detector. 4. Clean the pump rollers and the inside of the pump band. 5. Thread the new tubing through the pump so that the tube follows its natural curve around the pump rollers. You may need to move the pump rollers to make the installation easier. 6. The blue collars on the tube help align the tube in the detector and pump. Align the tube by placing the collars in the grooves inside the liquid detector. 7. Close the detectors cover and tighten the black knob. Close the pump strap and latch it. 8. Reset the pump-tube counter. (See example in table 8-2.) 9. Take a dry grab sample to test the tube installation.

Note
To extend the pump tube life: Always use Teledyne Isco pump tubes. Install the tube properly, aligning the blue collars correctly in the liquid detectors grooves. Follow the natural curve of the tube when wrapping the tube around the pump rollers. Minimize the line rinses and sampling retries in the sampling programs Use the shortest possible suction line. Conditions that shorten tube life: Improper installation. Abrasive materials suspended in sample liquid. Frequent line rinses. Long purge cycles, such as those used with long suction lines.

8-9

6712 Portable Sampler Section 8 Maintenance

Figure 8-1 Removing and Replacing the Pump Tube

Figure 8-2 Placing the Pump Tube in the Liquid Detector

8-10

6712 Portable Sampler Section 8 Maintenance

8.6 Opening the Controller Case

You will need to open the controller case to gain access to: The desiccant bag that dries the controllers interior The battery that provides backup power to the controller

CAUTION
The circuit boards can be damaged from a discharge of static electricity. Always use an antistatic mat when opening the controller to protect the circuit boards. To open the case: 1. Pull the discharge tube and pump tube away from the bulkhead fitting. 2. Remove the distributor arm by unscrewing the nut that attaches the arm to the distributor shaft. DO NOT ROTATE THE DISTRIBUTOR ARM MANUALLY; THIS WILL SEVERELY DAMAGE THE DISTRIBUTOR DRIVE GEARS. 3. Remove the controller from the center section. 4. Open the case by removing the 11 screws that attach the case bottom to the case top.

Note
The case top and bottom fit together with a sealed tongue-and-groove joint. Any damage nicks or cuts to the tongue, groove, or sealing gasket prevent the case from sealing completely when you close it. Use extreme care when opening the case to avoid damaging the joining surfaces. Refer to Figure 8-3 for the location of the desiccant box. To gain access to the main circuit board and battery, remove the two distributor connectors and the EMI shield.

8.7 Replacing the Desiccant

A humidity indicator, labeled INTERNAL CASE HUMIDITY, is located to the right of the keypad. It indicates the amount of moisture present inside the control box. The paper indicator is blue when the control box is dry. The control box is a completely sealed unit. You do not need to open the control box during normal operation. The 6712 Samplers use a desiccant bag inside the controller case to prevent moisture damage to its electronic components. Thus, the humidity indicator should remain blue under normal conditions. If moisture does accumulate, the numbered areas on the indicator will turn light pink or white, starting with the area numbered 20. This indicates that the relative humidity inside the control box exceeds 20%. As more moisture accumulates, the areas numbered 30 and 40 will turn light pink or white, indicating relative humidities of 30% and 40% respectively.

8-11

6712 Portable Sampler Section 8 Maintenance If the 30% area of the humidity indicator turns light pink or white, open the control unit, inspect it for leaks, and replace the desiccant bag.

CAUTION
Do not attempt to reactivate the used desiccant. Reactivation may cause desiccant to leak out of the bag, causing mechanical damage inside the controller. Follow the instructions in Opening the Controller Case in Section 8.6 to open the controller. Remove the bag from the cardboard box (part #69-9003-518) at the front of the case and replace it with a fresh bag of desiccant (part #099-0002-00).

Distributor Connectors

Desiccant Box

Figure 8-3 Opening the Controller

8-12

6712 Portable Sampler Section 8 Maintenance

Keypad Connector Module Connector

Battery

Liquid Detector Connector

Connector Board Connector

Pump Count Connector

Figure 8-4 Main Circuit Board Connections

8.8 Replacing the Internal Battery


Battery Status Display
WARNING! REPLACE INTERNAL BATTERY BEFORE 17-OCT-06

A lithium battery housed inside the controller maintains power to the samplers memory when the controller is disconnected from a power source. If it discharges completely, the sampler will lose all program settings and all data stored in memory when the external power source is disconnected. The battery requires replacement every five years. The sampler keeps track of the time that has passed since the battery was installed. To view this information, select INTERNAL BATTERY from the Maintenance screen (see example at left). After 5 seconds, the sampelr asks if you have replaced the internal battery (see exapmple at left). Select YES to reset the countdown to 59 months from the current date. While replacing the battery, it is likely that the stored readings will be lost and your program will revert to the default program. Before proceeding, retrieve the stored data and record your program settings.

DID YOU REPLACE THE INTERNAL BATTERY? YES NO

Note
Teledyne Isco recommends you replace the battery every five years, as soon as the battery warning appears. To ensure that the replacement meets Teledyne Iscos specifications, use only Teledyne Isco replacement batteries. (See Appendix D, Replacement Parts.)

8-13

6712 Portable Sampler Section 8 Maintenance To replace the internal battery: 1. Open the controller case. Separate the case bottom and top by removing the two distributor connectors. 2. Lift the desiccant box from the case. 3. Remove the EMI shield and grounding straps by loosening the five 1/4-inch stop nuts. 4. Disconnect the module, liquid detector, and pump count connectors from the main circuit board. 5. Remove the three screws and the 1/4-inch stud attaching the main circuit board to the case. 6. Detach the main circuit board from the keypad connector by pulling gently upward on the edges of the board next to the connector. Pull the main board up until the keypad connector is almost even with the edge of the case. Next, detach the main circuit board from the connector board. 7. De-solder the lithium battery, using care to prevent damage to the circuit board. Insert the replacement battery and solder it in place.

Note
Poor soldering techniques can easily damage a printed circuit board. The battery should be replaced by a skilled technician.

WARNING
Danger of explosion if lithium battery is incorrectly replaced. Replace with the same type, ensuring correct polarity. Do not dispose of used lithium battery in fire. 8. Reconnect the main circuit board to the connector board before reconnecting it to the keypad. Replace the 1/4-inch stud and the three screws, and reconnect the module, liquid detector, and pump count connectors. 9. Replace the EMI shield and grounding straps. Reconnect the two distributor connectors. 10. Ensure that all connections and hardware are secure. Replace the desiccant and close the case. 11. Reset the battery alarm.

8.9 Error Messages

If the sampler detects an error that prevents it from taking a sample or continuing the program, it displays an error message.
OPERATION ERROR: PUMP JAMMED

8-14

6712 Portable Sampler Section 8 Maintenance There are messages for the following conditions: Error Messages
ERRORS HAVE OCCURED DURING PROGRAM

PUMP JAMMED DISTRIBUTOR JAMMED PROBABLE OVERFLOW POWER FAILED LOW BATTERY USER STOPPED NO DISTRIBUTOR ARM

NO LIQUID DETECTED NO MORE LIQUID PUMP LATCH OPEN SAMPLER SHUT OFF BOTTLE FULL SAMPLE IN PROGRESS

Since the sampler may encounter more than one error during a program, it does not display the messages continuously. Instead, it records each error and the time it occurred in the Sampling Results and Combined Results report. It alerts you to the recorded errors by displaying the message, ERRORS HAVE OCCURRED DURING PROGRAM. If the sampler detects a low battery condition, it stops the running program and displays the LOW BATTERY error message. A low battery condition exists when five consecutive sample events have resulted in a POWER FAILED error message. Fatal Error The Fatal Error message will only show up if you are using a multiple bottle configuration and the distributor system fails. Failing the test can indicate any of the following conditions: The distributor arm is not attached to the sampler. The distributor-arm stop inside the center section is missing or broken. The distributor drive mechanism inside the controller is malfunctioning.

FATAL ERROR NO DISTRIBUTOR ARM

8.10 Pump Tube Warning


WARNING: REPLACE PUMP TUBE

Occasionally, the sampler displays a warning message to alert you to a potential problem. The warning, REPLACE PUMP TUBE, reminds you to check the condition of the pump tube. Instructions regarding the pump tube can be found elsewhere in this section.

8-15

6712 Portable Sampler Section 8 Maintenance

8-16

6712 Portable Sampler


Appendix A Menu Flowcharts

See Running Programs on page -10.


RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

Stop Key

POWER USED: CURRENT: _._ amp-hr PREVIOUS: _._ amp-hr Note A

See the charts on pages A-2 through A-19.

Note A

See View Report on page A-6.

MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

See Maintenance Screens on page A-7.

See Manual Functions Screens on page A-8.

PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE

Note A: Menu access can be controlled with the optional Program Lock. See Section 5.

Figure A-1 6712 Menu Tree for Standard Programming

A-1

6712 Portable Sampler Appendix A Menu Flowcharts

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

SITE DESCRIPTION: " FAC TO RY " CHANGE? YES NO


NO

YES

SITE: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE


DONE

If STOP key pressed.

SAVE CHANGES? YES NO

Programming screens shown are for a 730 Bubbler Flow Module. Screens for other modules may vary.
If applicable SELECT UNITS FOR F L O W R AT E : cfs gps gpm Mgd lps m3s m3h m3d SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit PROGRAM MODULE? YES NO
YES

Programming screens for a 700 Series module. Refer to Modules manual.

(if 1, 2, 4, or 8 bottles) BOTTLE VOLUME IS _____ lit (0.3-100.0) NUMBER OF BOTTLES: 1 2 4 8 12 24 BOTTLE VOLUME IS _____ ml (300-30000) (if 12 or 24 bottles) SUCTION LINE LENGTH IS __ ft (3-99)

NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE If applicable

T I M E PA C E D F L O W PA C E D

TIME BETWEEN SAMPLE EVENTS __ HOURS, __ MINUTES

FLOW BETWEEN SAMPLE EVENTS 0.075 Mgal (0.001-99999)

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOTTLE ___ BOTTLES PER SAMPLE EVENT (1-max)

Multi-bottle

___ SAMPLES PER BOTTLE (1-max) 1 bottle DO YOU WANT TO RUN CONTINUOUSLY? YES NO Yes No TA K E _ _ _ S A M P L E S (1-999)

R U N C O N T I N U O U S LY ? YES NO

SAMPLE VOLUME: ____ ml (10-max)

See Figure A-3: Standard and Extended Programming: Start Times.

Figure A-2 Standard Programming: Programming Screens

A-2

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-2.

See Notes B & C.


N O D E L AY T O S TA R T D E L AY E D S TA R T CLOCK TIME WAIT FOR PHONE CALL FIRST SAMPLE AFTER A _ _ _ M I N U T E D E L AY (1-999)

See Note D.
F I R S T S A M P L E AT : HH:MM

S E L E C T DAY S : SU MO TU WE TH FR SA DONE

AT L E A S T O N E D AY MUST BE SELECTED! DONE

F I R S T VA L I D DAY IS __-___

(Extended Programming Only)

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO


NO

YES

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

See Run Time Screens on page A-4.

See Note E.

Notes

B. Appears for standard flow-paced programs.

S TA R T F L O W C O U N T AFTER A _ _ _ M I N U T E D E L AY (1-999)

C. Appears for extended programs.

S TA R T " E X T E N D E D 1 " AFTER A _ _ _ M I N U T E D E L AY (1-999)

D. Appears for standard flow-paced programs.

S TA R T F L O W C O U N T AT HH:MM

E. Main menu for extended programming.

RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS

Figure A-3 Standard and Extended Programming: Start Times

A-3

6712 Portable Sampler Appendix A Menu Flowcharts

SITE DESCRIPTION: " FAC TO RY "

SITE: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE

If STOP key pressed.

SAVE CHANGES? YES NO

If applicable UNITS SELECTED F L O W R AT E : c f s FLOW VOLUME: Mgal

SELECT UNITS FOR F L O W R AT E : cfs gps gpm Mgd lps m3s m3h m3d

Units Selected screens are for a 730 Bubbler Flow Module. Screens for other modules may vary.

SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit

If applicable BUBBLER MODULE WEIR 90 V- N OT C H

Programming screens for a 730 Series Module. Refer to Modules manual.

If applicable NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

__, _____ml BOTTLES __ ft SUCTION LINE

NUMBER OF BOTTLES: 1 2 4 8 12 24

BOTTLE VOLUME IS _____ ml (300-30000)

SUCTION LINE LENGTH IS __ ft (3-99) TIME BETWEEN SAMPLE EVENTS __ HOURS, __ MINUTES PA C I N G : TIME, EVERY __ HOURS, __ MINUTES T I M E PA C E D F L O W PA C E D FLOW BETWEEN SAMPLE EVENTS 0.075 Mgal (0.001 - 99999)

Multi-Bottle

SEQUENTIAL BOTTLES/SAMPLE SAMPLES/BOTTLE __ BOTTLES PER SAMPLE EVENT (1-max)

R U N C O N T I N U O U S LY ? YES NO

DISTRIBUTION: BOTTLES/SAMPLE R U N C O N T I N U O U S LY

__ SAMPLES PER BOTTLE (1-max)

1 Bottle

DO YOU WANT TO RUN CONTINUOUSLY? YES NO

No

TA K E _ _ _ S A M P L E S (1-max)

MUST BE SELECTED! SAMPLE VOLUME: ___ ml (10-max)

200 ml SAMPLES

Continued in Figure A-5.

Figure A-4 Standard Programming: Quick View; Programming Screens

A-4

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-4.

FIRST SAMPLE AFTER A _ _ _ M I N U T E D E L AY (1-999)

N O D E L AY T O S TA R T

N O D E L AY T O S TA R T D E L AY E D S TA R T CLOCK TIME WAIT FOR PHONE CALL

F I R S T S A M P L E AT : HH:MM

If applicable

S E L E C T DAY ( S ) : SU MO TU WE TH FR SA DONE

AT LEAST ONE DAY MUST BE SELECTED! DONE

PROGRAMMING COMPLETE RUN THIS PROGRAM NOW? YES NO


NO

FIRST SAMPLE: HH:MM MO TH


YES

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

See Run Time Screens on page 4-10.

Figure A-5 Standard Programming: Quick View; Start Times

A-5

6712 Portable Sampler Appendix A Menu Flowcharts

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

No valid options N O D ATA TO D I S P L AY

SELECT DATA TO VIEW: SAMPLING REPORT MODULE FR-TEMP RAIN SDI-12

VIEW: D A I LY S U M M A R Y CURRENT READINGS

pH MODULE VIEW: p H D ATA T E M P E R AT U R E D ATA

Valid Options Will Be Shown; Screens Will Vary Depending on Devices Attached

V I E W D ATA SYSTEM IDs CONFIGURE REPORT SAMPLER 6712 ID:1234567890 HARDWARE: B0 SOFTWARE: 1.00

Note B SOFTWARE OPTIONS: NONE

When applicable ULTRASONIC MODULE: 0710 ID:1234567890 HARDWARE: A1 SOFTWARE: 1.00

When applicable SDI-12 SONDE SOFTWARE:

Note A

SAMPLINK REPORT FLOWLINK REPORT CUSTOM REPORT ALL REPORTS

REPORTS FOR SAMPLER: PROGRAM SETTINGS S A M P L I N G R E S U LT S BOTH NONE

When applicable REPORTS FOR MODULE: D A I LY S U M M A R Y C O M B I N E D R E S U LT S BOTH NONE

When applicable REPORTS FOR FR-TEMP: D A I LY S U M M A R Y C O M B I N E D R E S U LT S BOTH NONE

Note A: Menu access can be controlled with the optional Program Lock. See Note 8. Note B: 1108 - Program Lock

When applicable REPORTS FOR RAIN: D A I LY S U M M A R Y C O M B I N E D R E S U LT S BOTH NONE

When applicable D O YO U WA N T C O M B I N E D R E S U LT S FOR SDI-12 DATA? YES NO

I N C L U D E D E TA I L D ATA FOR FLOWLINK AND TEXTLINK PROGRAMS? YES NO

Figure A-6 View Report

A-6

6712 Portable Sampler Appendix A Menu Flowcharts

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

E N T E R T I M E A N D D AT E : HH:MM DD-MON-YY __:__ __-__-__

________ PUMP COUNTS, W A R N I N G AT _ _ _ _ _ _ _ .

See Note A.

SET CLOCK PUMP TUBE ALARM I N T E R N A L B AT T E R Y DIAGNOSTICS REPLACE INTERNAL B AT T E R Y B E F O R E __-__-__

______ PUMP COUNTS, RESET PUMP COUNTER? YES NO

W A R N I N G AT _ _ 0 0 0 0 0 PUMP COUNTS (1-99)00000

PLEASE WAIT! . . . TESTING 'RAM'

D I D YO U R E P L AC E T H E I N T E R N A L B AT T E R Y ? YES NO

' R A M ' PA S S E D T E S T

TEST PUMP? YES NO


YES

NO

PLEASE WAIT! . . .TESTING 'ROM' PUMPING . . .

TEST DISTRIBUTOR? YES NO

NO

YES

' R O M ' PA S S E D T E S T

FINDING ARM FLEXURE

PUMPING . . . O N / O F F R AT I O = _ . _ _ ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST ABCDEFGHIJKLMNOPQRST PURGING . . .

N O W AT B O T T L E _ _

DISTRIBUTOR SYSTEM ARM FLEXURE=__ PURGING . . . O N / O F F R AT I O = _ . _ _ RE-INITIALIZE? YES NO

Note
A. The standard programming maintenance screen is shown. Extended programming has SOFTWARE and HARDWARE options.

YES

D E FA U LT S E T T I N G S ? . . . A R E YO U S U R E ? YES NO

Figure A-7 Maintenance Screens

A-7

6712 Portable Sampler Appendix A Menu Flowcharts

RUN PROGRAM VIEW REPORT OTHER FUNCTIONS

MAINTENANCE MANUAL FUNCTIONS PROGRAMMING STYLE

GRAB SAMPLE SAMPLE VOLUME: ____ml (10-9990) P R E S S W H E N R E A DY !

if not known GRAB SAMPLE C A L I B R AT E V O L U M E O P E R AT E P U M P MOVE DISTRIBUTOR COMPACT PORTABLE STANDARD PORTABLE REFRIGERATOR SAMPLE VOLUME: ____ml (100-9990)

C A L I B R AT E V O L U M E P R E S S W H E N R E A DY !

GO TO BOTTLE __ (1-max)

VOLUME DELIVERED: _____ ml

N O W AT B O T T L E _ _

_____ ml! A R E YO U S U R E ? YES NO

SELECT DIRECTION: PUMP FORWARD PUMP REVERSE

PUMPING... PRESS THE RED 'STOP' KEY WHEN DONE!

PUMPED _____ COUNTS

PURGING... PRESS THE RED 'STOP' KEY WHEN DONE!

PURGED _____ COUNTS

Figure A-8 Manual Functions Screens

A-8

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS

"EXTENDED "EXTENDED "EXTENDED "EXTENDED

1" 2" 3" 4"


If STOP key pressed.

PROGRAM NAME: "ISCO 6712 " CHANGE? YES NO


NO

YES

SELECT NEW PROGRAM CHANGE PROGRAM NAME

NAME: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE

SAVE CHANGES? YES NO

SITE DESCRIPTION: " FAC TO RY " CHANGE? YES NO


NO

YES

SITE: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE If


STOP key pressed.

SAVE CHANGES? YES NO

SELECT UNITS FOR LENGTH: ft m

Units Selected screens are for a 730 Bubbler Flow Module. Screens for other modules may vary.

SELECT UNITS FOR FLOW RATE: cfs gps gpm Mgd lps m3s m3h m3d

SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit

If applicable PROGRAM MODULE? YES NO


YES

Programming screens for a 700 Series Module. Refer to the Modules manual.

NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

See Figure A-10.

See Figure A-11.

See Figure A-12.

See Figure A-6.

Figure A-9 Extended Programming: Programming Screens

A-9

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-9.

NUMBER OF BOTTLES: 1 2 4 8 12 24

BOTTLE VOLUME IS _____ ml (300-30000)

SUCTION LINE LENGTH IS __ ft (3-99)

AUTO SUCTION HEAD ENTER HEAD

SUCTION HEAD OF __ ft (min-max)

_ RINSE CYCLES (0-3)

RETRY UP TO _ TIMES WHEN SAMPLING (0-3)

O N E - PA R T P R O G R A M T W O - PA R T P R O G R A M

_ _ B OT T L E S AVA I L A B L E ASSIGN BOTTLES 1 THRU __ TO PA R T ' A ' ( 1 - m a x )

See Figure A-11.

Figure A-10 Extended Programming: Equipment Setup

A-10

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-10.


TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES

U N I F O R M T I M E PA C E D F L O W PA C E D E V E N T PA C E D NONUNIFORM TIME

PA C E D B Y : FLOW PULSES FLOW MODULE VOLUME FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999)

NONUNIFORM TIME: CLOCK TIMES I N T E RVA L S I N M I N U T E S R A N D O M I N T E RVA L S

TA K E S A M P L E S AT : 1 . S TA R T T I M E 2. HH:MM 3. HH:MM S A M P L E AT S TA R T ? YES NO

See Figure A-12: Event Conditions

PROGRAM RUN TIME: __ HOURS, __ MINUTES

F I R S T S A M P L E AT S TA R T T I M E , THEN ... D O YO U WA N T TO R U N C O N T I N U O U S LY ? YES NO Q U A N T I T Y AT I N T E R VA L 1 . _ _ AT _ _ _ M I N 2 . _ _ AT _ _ _ M I N 3 . _ _ AT _ _ _ M I N

__ BOTTLES PER SAMPLE EVENT (1-max)

SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME

SWITCH BOTTLES EVERY ___ SAMPLES (1-max)

Multi-bottle SWITCH BOTTLES EVERY __ HOURS, __ MINUTES FIRST SWITCH TIME AT _ _ : _ _

1 bottle R U N C O N T I N U O U S LY ? YES NO No TA K E _ _ _ S A M P L E S (1-999) R U N C O N T I N U O U S LY ? YES NO

Yes

Uniform Time Paced Only

D O YO U WA N T SAMPLE VOLUMES DEPENDENT ON FLOW? YES NO No

Yes

FLOW PULSES FLOW MODULE VOLUME

SAMPLE VOLUME: 10 ML FOR EVERY ___ PULSES

All other pacing

SAMPLE VOLUME: ___ ml (10-max)

See Figure A-13.

Figure A-11 Extended Programming: Pacing and Distribution

A-11

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-12.

All valid event options will be shown.


EVENT: RAIN LEVEL FLOW pHx TEMPx CONDx TDSx DOx NONE NO EVENT PARAMETERS CURRENTLY AVAILABLE

I/O

RAIN

OTHER OPTIONS

"RAIN" SET POINT: __.__ INCHES PER __HOURS, __MINUTES

"LEVEL" CONDITION: SET POINT RANGE RATE OF CHANGE

"LEVEL" SET POINT: _.__ft (0.001-30.000)

Repeat until conditions are set or until DONE

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

CONDITION IS TRUE WHEN "LEVEL" RISES FALLS

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

RESET RAIN HISTORY? YES NO

"LEVEL" RANGE: LOWER:_.__ft UPPER:_.__ft (0.001-30.000)

ENABLED WHEN: I/O ENABLE IS HIGH I/O ENABLE IS LOW

"LEVEL" RISES _._ ft. __HOURS, __MINUTES

ENABLED WHEN: INSIDE RANGE OUTSIDE RANGE

"LEVEL" FALLS _._ ft. __HOURS, __MINUTES

Figure A-12 Extended Programming: Event Conditions

A-12

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-11.

Continued from Figure A-21.


'A' Done and I/O are not options for I/O Conditions

See Note A.
NONE ENABLE: I/O RAIN LEVEL FLOW TEMPx CONDx TDSx DOx NONE All except NONE ENABLE: RAIN AND OR DONE AND or OR

See Note A.
ENABLE: RAIN AND I/O LEVEL FLOW TEMPx CONDx TDSx DOx DONE

DONE 'A' DONE RAIN "RAIN" SET POINT: __.__ INCHES PER __HOURS, __MINUTES

I/O ENABLED WHEN: I/O ENABLE IS HIGH I/O ENABLE IS LOW

All except I/O, Rain, and 'A' Done "LEVEL" CONDITION: SET POINT RANGE RATE OF CHANGE

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

"LEVEL" RANGE: LOWER:_.__ft UPPER:_.__ft (0.001-30.000)

CONDITION IS TRUE WHEN "LEVEL" RISES FALLS

"LEVEL" SET POINT: _.__ft (0.001-30.000)

RESET RAIN HISTORY? YES NO

ENABLED WHEN: INSIDE RANGE OUTSIDE RANGE

"LEVEL" RISES _.__ft __HOURS, __MINUTES

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

If applicable Repeat R E P E AT FOR SECOND TYPE

If applicable CHECK ENABLE EVERY 0.5 1 2 5 10 15 30 MINUTES

No ONCE ENABLED, S TAY E N A B L E D ? YES NO

S A M P L E AT D I S A B L E ? YES NO

S A M P L E AT E N A B L E ? YES NO

See Note B
Ye s S A M P L E AT E N A B L E ? YES NO Ye s RESET SAMPLE I N T E R VA L AT E N A B L E ? YES NO Ye s CONTINUE COUNTDOWN WHILE DISABLED? YES NO No

No

S TA R T O F S A M P L I N G AFTER A _ _ _ M I N U T E D E L AY (0-999)

If applicable PA U S E RESUME 1.HH:MM DD HH:MM DD 2.HH:MM DD HH:MM DD CLEAR DONE PA U S E & R E S U M E S MUST BE ENTERED I N PA I R S ! S A M P L E AT R E S U M E ? YES NO

A. All valid enable options will be shown. Enable options may include:

See Figure A-5.

Notes

LEVEL, FLOW, pH, TEMP, PERCENT, VEL, RAIN, I/O, 'A'DONE FR-TEMP, TEMPx, CONDx, SP_COx, TDSx, SALx, DOx, pHx, ORPx, LEVELx, NH4x, NH3x, NO3x, TURBx, Clx, FLOWx

B. YES = Latched or Stay Enabled NO = Repeatable Enable

Figure A-13 Extended Programming: Programmed Sampler Enable

A-13

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS

"EXTENDED "EXTENDED "EXTENDED "EXTENDED

1" 2" 3" 4"

SELECT NEW PROGRAM CHANGE PROGRAM NAME PROGRAM NAME: "EXTENDED 1" SITE DESCRIPTIONS: " FAC TO RY " NAME: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE

NAME: __________ ABCDEFGHIJKLMNOPQRST UVWXYZ -&"0123456789 BACK-UP DONE

SAVE CHANGES? YES NO

SAVE CHANGES? YES NO

UNITS SELECTED: LENGTH: ft

SELECT UNITS FOR FLOW RATE: cfs gps gpm Mgd lps m3s m3h m3d UNITS SELECTED: FLOWRATE: cfs FLOW VOLUME: Mgal SELECT UNITS FOR FLOW VOLUME: cf gal Mgal m3 lit If applicable

If applicable BUBBLER MODULE: WEIR 90o V-NOTCH

Programming screens for a 700 Series Modules. Refer to the Modules manual.

NEW MODULE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

May ask for level, offset, etc. Series of screens will vary, depending upon module.
If applicable __ MINUTE D ATA I N T E R VA L DATA STORAGE INTERVAL IN MINUTES: 1 2 5 10 15 30 INTERVAL CHANGED-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE

See Figure A-15.

See Figure A-16

See Figure A-17.

See Figure A-5.

Figure A-14 Extended Programming: Quick View; Programming Screens

A-14

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-14.


NUMBER OF BOTTLES: 1 2 4 8 12 24

__,____ ml BOTTLES __ ft SUCTION LINE AUTO SUCTION HEAD _ RINSES, _RETRIES

SUCTION LINE LENGTH IS __ ft (3-99) BOTTLE VOLUME IS _____ ml (300-30000)

When Liquid Detector is on


_ RINSE CYCLES (0-3) AUTO SUCTION HEAD ENTER HEAD SUCTION HEAD OF IS __ ft (min-max)

RETRY UP TO _ TIMES WHEN SAMPLING (0-3)

O N E PA R T P R O G R A M

O N E - PA R T P R O G R A M T W O - PA R T P R O G R A M

_ _ B OT T L E S AVA I L A B L E ASSIGN BOTTLES 1 THRU __ TO PA R T ' A ' ( 1 - m a x )

See Figure A-16.

Figure A-15 Extended Programming: Quick View; Equipment Set-Up

A-15

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-15.

TIME BETWEEN SAMPLE EVENTS: __ HOURS, __ MINUTES

PA C I N G : TIME, EVERY __ HOURS, __ MINUTES

U N I F O R M T I M E PA C E D F L O W PA C E D E V E N T PA C E D NONUNIFORM TIME

PA C E D B Y : FLOW PULSES FLOW MODULE VOLUME FLOW BETWEEN SAMPLE EVENTS: ____ PULSES (1-9999)

NONUNIFORM TIME: CLOCK TIMES I N T E RVA L S I N M I N U T E S R A N D O M I N T E RVA L S

TA K E S A M P L E S AT : 1 . S TA R T T I M E 2. HH:MM 3. HH:MM S A M P L E AT S TA R T ? YES NO

PROGRAM RUN TIME: __ HOURS, __ MINUTES

F I R S T S A M P L E AT S TA R T T I M E , THEN ...

Q U A N T I T Y AT I N T E R VA L 1 . _ _ AT _ _ _ M I N 2 . _ _ AT _ _ _ M I N 3 . _ _ AT _ _ _ M I N

__ BOTTLES PER SAMPLE EVENT (1-max)

SWITCH BOTTLES ON: NUMBER OF SAMPLES TIME

SWITCH BOTTLES EVERY ___ SAMPLES (1-max)

Multi-bottle DISTRIBUTION: __ BOTTLES/SAMPLE ___ SAMPLES/BOTTLE R U N C O N T I N U O U S LY

SWITCH BOTTLES EVERY __ HOURS, __ MINUTES

FIRST SWITCH TIME AT _ _ : _ _

1 bottle DO YOU WANT TO RUN CONTINUOUSLY? YES NO No TA K E _ _ _ S A M P L E S (1-999) DO YOU WANT TO RUN CONTINUOUSLY? YES NO

Uniform Time Paced Only VOLUME ____ ml SAMPLES D O YO U WA N T SAMPLE VOLUMES DEPENDENT ON FLOW? YES NO Yes SAMPLE VOLUME: 10 ML FOR EVERY ___ PULSES (1-9999) MINIMUM SAMPLE VOLUME: ___ ml (10 - max)

No

SAMPLE VOLUME: ___ ml (10-max)

See FigureA-17.

Figure A-16 Extended Programming: Quick View; Pacing and Distribution

A-16

6712 Portable Sampler Appendix A Menu Flowcharts

Continued from Figure A-16.

Continued from Figure A-21.


'A' Done and I/O are not options for I/O Conditions

See Note A.
'A' ENABLE: RAIN>0.11"/1:00 AND 0.500<LEVEL<1.500 ENABLE: I/O RAIN LEVEL FLOW pH TEMP COND D.O. NONE All except NONE ENABLE: RAIN AND OR DONE AND or OR

See Note A.
ENABLE: RAIN AND I/O LEVEL FLOW pH TEMP COND D. O . NONE

DONE I/O 'A' DONE RAIN "RAIN" SET POINT: __.__ INCHES PER __HOURS, __MINUTES ENABLED WHEN: I/O ENABLE IS HIGH I/O ENABLE IS LOW All except I/O, Rain, and 'A' Done

"LEVEL" CONDITION: SET POINT RANGE RATE OF CHANGE

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

"LEVEL" RANGE: LOWER:_.__ft UPPER:_.__ft (0.001-30.000)

CONDITION IS TRUE WHEN "LEVEL" RISES FALLS

"LEVEL" SET POINT: _.__ft (0.001-30.000)

RESET RAIN HISTORY? YES NO

ENABLED WHEN: INSIDE RANGE OUTSIDE RANGE

"LEVEL" RISES: _.__ft __HOURS, __MINUTES

ENABLED WHEN: ABOVE SET POINT BELOW SET POINT

ONCE ENABLED, S TAY E N A B L E D ? YES NO 'A' ENABLE: R E P E ATA B L E E N A B L E S A M P L E AT D I S A B L E N O S A M P L E AT E N A B L E S A M P L E AT E N A B L E ? YES NO

See Note B.
Repeatable Enable S A M P L E AT D I S A B L E ? YES NO

If applicable R E P E AT FOR SECOND TYPE

If applicable CHECK ENABLE EVERY 0.5 1 2 5 10 15 30 MINUTES

Latched Enable + No Sample at Enable

'A' ENABLE: _ _ M I N U T E D E L AY TO S TA R T O F S A M P L I N G

S TA R T O F S A M P L I N G AFTER A _ _ _ M I N U T E D E L AY (0-999)

Repeatable Enable

'A' ENABLE: COUNTDOWN CONTINUES WHILE DISABLED

RESET SAMPLE I N T E R VA L AT E N A B L E ? YES NO

NO

CONTINUE COUNTDOWN WHILE DISABLED? YES NO

'A' ENABLE: _ PA U S E & R E S U M E S N O S A M P L E AT R E S U M E

PA U S E RESUME 1.HH:MM DD HH:MM DD 2.HH:MM DD HH:MM DD CLEAR DONE

S A M P L E AT R E S U M E ? YES NO

A. All valid enable options will be shown. Enable options may include:

Notes
See Figure A-5.

LEVEL, FLOW, pH, TEMP, PERCENT, VEL, RAIN, I/O, 'A'DONE FR-TEMP, TEMPx, CONDx, SP_COx, TDSx, SALx, DOx, pHx, ORPx, LEVELx, NH4x, NH3x, NO3x, TURBx, Clx, FLOWx

B. YES = Latched or Stay Enabled NO = Repeatable Enable

Figure A-17 Extended Programming: Quick View; Programmed Sampler Enable

A-17

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS

M E A S U R E M E N T I N T E RVA L 5 15 30 SECONDS 1 2 5 MINUTES S TO R AG E I N T E RVA L

MAINTENANCE MANUAL FUNCTIONS S O F T WA R E O P T I O N S HARDWARE

DUAL SAMPLER MODE? YES NO

USE LIQUID DETECTOR? YES NO

BOTTLE FULL DETECT? YES NO

PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE

B A C K L I G H T A LWAY S O F F TIMED BACKLIGHT CONTINUOUS BACKLIGHT

EVENT MARK SENT FOR: EVERY SAMPLE COMPLETE SAMPLES

E V E N T M A R K D U R AT I O N : 3 SECONDS WHILE PUMPING

AT T H E B E G I N N I N G O F : INITIAL PURGE FORWARD PUMPING

DURING: ENTIRE PUMP CYCLE FORWARD PUMPING PRE-SAMPLE PURGE: __ COUNTS (10-9999)

P O S T- S A M P L E P U R G E : DEPENDENT ON HEAD FIXED COUNT

P O S T- S A M P L E P U R G E : ___ COUNTS (10-9999)

PERIODIC SERIAL OUTPUT? YES NO

YES

PORT SETTINGS: 9600n81 4800n81 2400n81 1200n81

SERIAL OUTPUT EVERY: 15 SEC 1 MIN 5 MIN 5 MIN

If applicable P O W E R A LWAY S O N POWER ON/OFF TIMES A L A R M D I A L O U T S O N LY POWER ON OFF 1. _:_ _:_ 2. _:_ _:_ CLEAR DONE

POWER ON/OFF TIMES C A N N O T S PA N M O R E T H A N O N E DAY !

POWER ON/OFF TIMES MUST BE ENTERED I N PA I R S !

Figure A-18 Extended Programming: Normal View; Software Options

A-18

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS

MAINTENANCE MANUAL FUNCTIONS S O F T WA R E O P T I O N S HARDWARE USE LIQUID DETECTOR? YES NO LIQUID DETECT ON QUICK VIEW/CHANGE PROGRAMMING STYLE: NORMAL QUICK VIEW/CHANGE

TA K E M E A S U R E M E N T S EVERY __ MINUTES

M E A S U R E M E N T I N T E RVA L 5 15 30 SECONDS 1 2 5 MINUTES S TO R AG E I N T E RVA L

DUAL SAMPLER MODE? YES NO DUAL SAMPLER ON BTL FULL DECTECT OFF TIMED BACKLIGHT B A C K L I G H T A LWAY S O F F TIMED BACKLIGHT CONTINUOUS BACKLIGHT AT T H E B E G I N N I N G O F : INITIAL PURGE FORWARD PUMPING BOTTLE FULL DETECT? YES NO

EVENT MARK SENT WHILE PUMPING FORWARD

E V E N T M A R K D U R AT I O N : 3 SECONDS WHILE PUMPING

E V E N T M A R K D U R AT I O N : 3 SECONDS WHILE PUMPING

DURING: ENTIRE PUMP CYCLE FORWARD PUMPING

PUMP COUNTS FOR EACH PURGE CYCLE: __ PRE-SAMPLE _ _ P O S T- S A M P L E

PRE-SAMPLE PURGE: __ COUNTS (10-9999)

P O S T- S A M P L E P U R G E : DEPENDENT ON HEAD FIXED COUNT

P O S T- S A M P L E P U R G E : ___ COUNTS (10-9999)

SERIAL OUTPUT EVERY __ MINUTES

PERIODIC SERIAL OUTPUT? YES NO

YES

PORT SETTINGS: 9600n81 4800n81 2400n81 1200n81

SERIAL OUTPUT EVERY: 15 SEC 1 MIN 5 MIN 5 MIN

If applicable P O W E R A LWAY S O N POWER ON/OFF TIMES A L A R M D I A L O U T S O N LY POWER ON OFF 1. _:_ _:_ 2. _:_ _:_ CLEAR DONE

POWER ON/OFF TIMES C A N N O T S PA N M O R E T H A N O N E DAY !

I N T E R R O G AT O R CONNECTOR P O W E R A LWAY S O N

POWER ON/OFF TIMES MUST BE ENTERED I N PA I R S !

Figure A-19 Extended Programming: Quick View; Software Options

A-19

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "EXTENDED 1" PROGRAM VIEW REPORT OTHER FUNCTIONS

MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE

D O YO U H AV E A RAIN GAUGE ATTACHED? YES NO NO YES RAIN GAUGE 0.01 inch TIP 0.1mm TIP

If applicable NEW HARDWARE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE PLEASE WAIT! DONE PA R T I T I O N M E M O R Y W I L L L A S T _ _ DAY S

SCAN SDI-12 SONDES AT TURN-ON TIME? YES NO

YES

If scan finds sonde, see Figure A-24.

See Note A.
<<-> -> -> -> -> -> -> -> -> -> SET I/Ox: NONE I/O ENABLE FR-TEMP PGM RUNNING PGM ENABLED 'A' ENABLED 'B' ENABLED PROGRAM DONE 'A' DONE 'B' DONE RUN ERROR FLOW THRU CONDITIONS NEW HARDWARE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE RUN FLOW THRU PUMP ___SECONDS BEFORE TAKING READINGS (0-120)

See Figure A-13.

See Figure A-22.

Notes
A. <- denotes input signal. -> denotes output signal.

Figure A-20 Extended Programming: Normal View; Hardware

A-20

6712 Portable Sampler Appendix A Menu Flowcharts

RUN "ISCO 6712 " PROGRAM VIEW REPORT OTHER FUNCTIONS

MAINTENANCE MANUAL FUNCTIONS SOFTWARE OPTIONS HARDWARE

0.01 inch TIP RAIN GAUGE

D O YO U H AV E A R A I N G A U G E AT TA C H E D ? YES NO NO

YES

RAIN GAUGE 0.01 inch TIP 0.1mm TIP

If applicable NEW HARDWARE SETUP-D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE

See Figure A-24.


NO SDI-12 SONDE AUTO SDI-12 SCAN ON SCAN SDI-12 SONDES AT T U R N - O N T I M E ? YES NO

I/O1=I/O ENABLE I/O2=PGM RUNNING I/O3=PROGRAM DONE

See Note A.
SET I/Ox: NONE I/O ENABLE FR-TEMP PGM RUNNING PGM ENABLED 'A' ENABLED 'B' ENABLED PROGRAM DONE 'A' DONE 'B' DONE RUN ERROR FLOW THRU CONDITIONS NEW HARDWARE SETUP-DOWNLOAD DATA NOW OR LOSE ALL DATA! DONE RUN FLOW THRU PUMP ___SECONDS BEFORE TAKING READINGS (0-120)

<<-> -> -> -> -> -> -> -> -> ->

See Figure A-17.

See Figure A-23.

Notes

A. <- denotes input signal. -> denotes output signal.

Figure A-21 Extended Programming: Quick View; Hardware

A-21

6712 Portable Sampler Appendix A Menu Flowcharts


Notes
A. Factory installed option. Requires B1 hardware or newer. B. Requires factory-installed Dialout Modem and B2 hardware or newer.

See Figure A-20.

If applicable See Note A ANALOG OUTPUT HARDWARE HAS NOT B E E N I N S TA L L E D ! SETUP MANUAL CONTROL ANALOG OUTPUT _: LEVEL VELOCITY FLOW NONE

SET ANALOG OUTPUT _ TO __.__ mA DONE

ANALOG OUTPUT _: 0-TO-20 mA 4-TO-20 mA

ANALOG OUTPUT _: 0 m A AT _ . _ _ f p s 2 0 m A AT _ . _ _ f p s (-99999 - 99999)

Repeat for each channel

If applicable See Note B

I/Ox=1,2 DIALOUT NUMBERS: I/Ox=3 1. 2. 3.

When DECIMAL POINT key is pressed SELECT CODE: # PAGER

DIALOUT WHEN I/Ox BECOMES TRUE? YES NO

__ MINUTE DELAY BETWEEN CALLS (0-99)

Figure A-22 Analog and Dialout Screens

See Figure A-21.

Note
A. Factory installed option. Requires B1 hardware or newer.
If applicable See Note ANALOG OUTPUT HARDWARE HAS NOT B E E N I N S TA L L E D ! SETUP MANUAL CONTROL ANALOG OUTPUT _: LEVEL VELOCITY FLOW NONE

SET ANALOG OUTPUT _ TO __.__ mA _ ANALOG OUTPUTS DONE

ANALOG OUTPUT _: 0-TO-20 mA 4-TO-20 mA

ANALOG OUTPUT _: 0 m A AT _ . _ _ f p s 2 0 m A AT _ . _ _ f p s (-99999 - 99999)

Repeat for each channel

When DECIMAL POINT key is pressed DIALOUT EACH TIME I/Ox BECOMES TRUE DIALOUT WHEN I/Ox BECOMES TRUE? YES NO I/Ox =3 1. 2. 3. DIALOUT NUMBERS: , SELECT CODE: # PAGER

I/Ox=1,2

__ MINUTE DELAY BETWEEN CALLS (0-99)

Figure A-23 Analog Output and Dialout Quick View Screens

A-22

6712 Portable Sampler Appendix A Menu Flowcharts

NEW HARDWARE SETUP D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES

PLEASE WAIT! LOOKING FOR SDI-12 SONDES

NO SDI-12 SONDE AUTO SDI-12 SCAN ON

_ _ P R E V I O U S LY CONFIGURED SONDES FOUND

__ SONDE(S) FOUND

If Sondes Found

If No Sondes Found

SCAN SDI-12 SONDES AT T U R N O N T I M E ? YES NO

NEW HARDWARE SETUP D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE

PA R T I T I O N M E M O R Y W I L L L A S T _ _ DAY S

Done

SELECT ADDRESS OF SONDE TO SETUP: __ __ DONE

Company Model Optional Information S O N D E AT A D D R E S S _ H A S _ D ATA VA L U E S

Non-Isco Ready Sondes

D ATA F R O M S O N D E _ 1 = OTHER 2 = OTHER 3 = OTHER Repeat Until Done

S O N D E _ , D ATA x x pH DO LEVEL

D ATA F R O M S O N D E _ 4 = OTHER 5 = OTHER DONE

Enough partitions to store this sonde's data.

Not enough partitions to store this sonde's data.

SONDE _, STORE A L L PA R A M E T E R D ATA ? YES NO

SONDE _, STORE T E M P E R AT U R E D ATA ? YES NO LABEL IS: TEMP_

Repeat for all data types.

NEW HARDWARE SETUP D O W N L O A D D ATA N O W O R L O S E A L L D ATA ! DONE

Proceed O N calibration/validation P A R T I T I to M E M O R Y WIL LA T _ D Hardware screens. screensL andSrest_ of A Y S

Figure A-24 SDI-12 Sonde Screens

A-23

6712 Portable Sampler Appendix A Menu Flowcharts

A-24

6712 Portable Sampler


Appendix B Material S afety Data S heets

This appendix provides Material Safety Data sheets for the internal desiccant used in the 6712 sampler. Specific questions regarding the use and handling of these products should be directed to the manufacturer listed in the MSDS.

B-1

6712 Portable Sampler Appendix B Material Safety Data Sheets

101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002

MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT Packaged Desiccant SECTION I -- PRODUCT IDENTIFICATION Trade Name and Synonyms: Chemical Family: Formula: Silica Gel, Synthetic Amorphous Silica, Silicon, Dioxide Synthetic Amorphous Silica SiO2.x H2O

SECTION II -- HAZARDOUS INGREDIENTS Components in the Solid Mixture CAS No % ACGIH/TLV (PPM) OSHA-(PEL) 63231-67-4 >99 PEL - 20 (RESPIRABLE), LIMIT NONE, TLV 5 HAZARD IRRITANT

COMPONENT Amorphous Silica

Synthetic amorphous silica is not to be confused with crystalline silica such as quartz, cristobalite or tridymite or with diatomaceous earth or other naturally occurring forms of amorphous silica that frequently contain crystalline forms. This product is in granular form and packed in bags for use as a desiccant. Therefore, no exposure to the product is anticipated under normal use of this product. Avoid inhaling desiccant dust.

SECTION III -- PHYSICAL DATA Appearance and Odor: Melting Point: Solubility in Water: Bulk Density: Percent Volatile by Weight @ 1750 Deg F: White granules; odorless. >1600 Deg C; >2900 Deg F Insoluble. >40 lbs./cu. ft. <10%.

B-2

6712 Portable Sampler Appendix B Material Safety Data Sheets

101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002

MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT Packaged Desiccant SECTION IV -- FIRE EXPLOSION DATA Fire and Explosion Hazard - Negligible fire and explosion hazard when exposed to heat or flame by reaction with incompatible substances. Flash Point - Nonflammable. Firefighting Media - Dry chemical, water spray, or foam. For larger fires, use water spray fog or foam. Firefighting - Nonflammable solids, liquids, or gases: Cool containers that are exposed to flames with water from the side until well after fire is out. For massive fire in enclosed area, use unmanned hose holder or monitor nozzles; if this is impossible, withdraw from area and let fire burn. Withdraw immediately in case of rising sound from venting safety device or any discoloration of the tank due to fire.

SECTION V -- HEALTH HAZARD DATA Health hazards may arise from inhalation, ingestion, and/or contact with the skin and/or eyes. Ingestion may result in damage to throat and esophagus and/or gastrointestinal disorders. Inhalation may cause burning to the upper respiratory tract and/or temporary or permanent lung damage. Prolonged or repeated contact with the skin, in absence of proper hygiene, may cause dryness, irritation, and/or dermatitis. Contact with eye tissue may result in irritation, burns, or conjunctivitis. First Aid (Inhalation) - Remove to fresh air immediately. If breathing has stopped, give artificial respiration. Keep affected person warm and at rest. Get medical attention immediately. First Aid (Ingestion) - If large amounts have been ingested, give emetics to cause vomiting. Stomach siphon may be applied as well. Milk and fatty acids should be avoided. Get medical attention immediately. First Aid (Eyes) - Wash eyes immediately and carefully for 30 minutes with running water.

B-3

6712 Portable Sampler Appendix B Material Safety Data Sheets

101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002

MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT Packaged Desiccant NOTE TO PHYSICIAN: This product is a desiccant and generates heat as it adsorbs water. The used product can contain material of hazardous nature. Identify that material and treat accordingly.

SECTION VI -- REACTIVITY DATA Reactivity - Silica gel is stable under normal temperatures and pressures in sealed containers. Moisture can cause a rise in temperature which may result in a burn.

SECTION VII --SPILL OR LEAK PROCEDURES Notify safety personnel of spills or leaks. Clean-up personnel need protection against inhalation of dusts or fumes. Eye protection is required. Vacuuming and/or wet methods of cleanup are preferred. Place in appropriate containers for disposal, keeping airborne particulates at a minimum.

SECTION VIII -- SPECIAL PROTECTION INFORMATION Respiratory Protection - Provide a NIOSH/MSHA jointly approved respirator in the absence of proper environmental control. Contact your safety equipment supplier for proper mask type. Ventilation - Provide general and/or local exhaust ventilation to keep exposures below the TLV. Ventilation used must be designed to prevent spots of dust accumulation or recycling of dusts. Protective Clothing - Wear protective clothing, including long sleeves and gloves, to prevent repeated or prolonged skin contact. Eye Protection - Chemical splash goggles designed in compliance with OSHA regulations are recommended. Consult your safety equipment supplier.

B-4

6712 Portable Sampler Appendix B Material Safety Data Sheets

101 Christine Drive Belen, New Mexico 87002 Phone: (505) 864-6691 Fax: (505) 861-2355 ISO 9002

MATERIAL SAFETY DATA SHEET -- September 28, 1998 SORB-IT Packaged Desiccant SECTION IX -- SPECIAL PRECAUTIONS Avoid breathing dust and prolonged contact with skin. Silica gel dust causes eye irritation and breathing dust may be harmful. * No Information Available HMIS (Hazardous Materials Identification System) for this product is as follows: Health Hazard Flammability Reactivity Personal Protection 0 0 0 HMIS assigns choice of personal protective equipment to the customer, as the raw material supplier is unfamiliar with the condition of use.

The information contained herein is based upon data considered true and accurate. However, United Desiccants makes no warranties expressed or implied, as to the accuracy or adequacy of the information contained herein or the results to be obtained from the use thereof. This information is offered solely for the user's consideration, investigation and verification. Since the use and conditions of use of this information and the material described herein are not within the control of United Desiccants, United Desiccants assumes no responsibility for injury to the user or third persons. The material described herein is sold only pursuant to United Desiccants' Terms and Conditions of Sale, including those limiting warranties and remedies contained therein. It is the responsibility of the user to determine whether any use of the data and information is in accordance with applicable federal, state or local laws and regulations.

B-5

6712 Portable Sampler Appendix B Material Safety Data Sheets

Material Safety Data Sheet


Manufacturer : Address:

Indicating Silica Gel


Identity (Trade Name as Used on Label)

MULTISORB TECHNOLOGIES, INC.


(formerly Multiform Desiccants, Inc.)

MSDS Number* : CAS Number* : Date Prepared: Prepared By* :

M75

325 Harlem Road Buffalo, NY 14224 Phone Number (For Information): 716/824-8900 Emergency Phone 716/824-8900
Number:

July 6, 2000 G.E. McKedy

Section 1 - Material Identification and Information


Components - Chemical Name & Common Names (Hazardous Components 1% or greater; Carcinogens 0.1% or greater) %* OSHA PEL ACGIH TLV
3

OTHER LIMITS RECOMMENDE D


3

Silica Gel SiO2 Cobalt Chloride

98.0 >2.0

6mg/m (total dust) 3 0.05mg/m (TWA cobalt metal dust & fume)

10mg/m (total dust) 3 .05mg/m (Cobalt, TWA)

Non-Hazardous Ingredients TOTAL

100

Section 2 - Physical/Chemical Characteristics


Boiling N/A Point Vapor Pressure N/A (mm Hg and Temperature Vapor N/A Density (Air =1) Solubility Insoluble, but will adsorb moisture. in Water Appearance Purple crystals, no odor. and Odor Specific Gravity (H20 = 1) Melting Point Evaporation Rate (__________=1) Water Reactive

2.1 N/A N/A

Not reactive, but will adsorb moisture.

Section 3 - Fire and Explosion Hazard Data


Flash Point and Auto-Ignition Flammability Limits in LEL UEL N/A N/A N/A Methods Used Temperature Air % by Volume Extinguisher Dry chemical, carbon dioxide and foam can be used. Media Special Fire Water will generate heat due to the silica gel which will adsorb water and liberate heat. Fighting Procedures Unusual Fire and When exposed to water, the silica gel can get hot enough to reach the boiling point of water. Flooding with Explosion Hazards water will reduce the temperature to safe limits.

Section 4 - Reactivity Hazard Data


Conditions STABILITY Moisture and high humidity environments. To Avoid Stable Unstable Incompatibility Water. (Materials to Avoid) Hazardous Carbon dioxide, carbon monoxide, water Decomposition Products Conditions HAZARDOUS POLYMERIZATION None. To Avoid May Occur

*Optional

Indicating Silica Gel

B-6

6712 Portable Sampler Appendix B Material Safety Data Sheets

Page 2

Section 5 - Health Hazard Data


PRIMARY ROUTES OF ENTRY HEALTH HAZARDS Inhalation Ingestion NTP CARCINOGEN LISTED IN Skin Absorption Not Hazardous IARC Monograph Acute May cause eye, skin and mucous membrane irritation. Chronic Prolonged inhalation may cause lung damage. OSHA Not Listed

Signs and Symptoms Drying and irritation. of Exposure Medical Conditions Asthma. Generally Aggravated by Exposure EMERGENCY FIRST AID PROCEDURES - Seek medical assistance for further treatment, observation and support if necessary. Eye Contact Flush with water for at least 15 minutes. Skin Wash affected area with soap and water. Contact Inhalation Remove affected person to fresh air. Ingestion Drink at least 2 glasses of water.

Section 6 - Control and Protective Measures


Respiratory Protection Use NIOSH approved dust mask or respirator. (Specify Type) Protective Eye Protection Light cotton gloves. Safety glasses. Gloves VENTILATION Local Exhaust Mechanical (General) TO BE USED Other (Specify) Other Protective None. Clothing and Equipment Hygienic Work Avoid raising dust. Avoid contact with skin, eyes and clothing. Practices

Special

Section 7 - Precautions for Safe Handling and Use/Leak Procedures


Steps to be Taken if Material Sweep or vacuum up and place the spilled material in a waste disposal container. Avoid raising dust. Is Spilled Or Released Waste Disposal Dispose in an approved landfill according to federal, state and local regulations. Methods Precautions to be Cover promptly to avoid blowing dust. Wash after handling. Taken In Handling and Storage Other Precautions and/or Special Keep in sealed containers away from moisture. The silica gel will readily adsorb moisture. Hazards

*Optional

Indicating Silica Gel

B-7

6712 Portable Sampler Appendix B Material Safety Data Sheets

B-8

6712 Portable Sampler


Appendix C General S afety Proc edures

In field installations of Teledyne Isco wastewater samplers and associated equipment, the safety of the personnel involved should be the foremost consideration. The following sections provide safety procedures for working in and around manholes and sewers. the first section offers general safety advice. The second section deals with the special problem of hazardous gases found in sewers.

WARNING
The 6712 has not been approved for use in hazardous locations as defined by the National Electrical Code.

CAUTION
Before any sampler is installed, the proper safety precautions must be taken. The following discussions of safety procedures are only general guidelines. Each situation in which you install a flow meter varies. You must take into account the individual circumstances you are in. Additional safety considerations, other than those discussed here, may be required.

C.1 Practical Safety Precautions

The following procedures are those used by Black & Veatch, a respected consulting firm, and are published here by permission. Field personnel must keep safety uppermost in their minds at all times. When working above ground, rules of common sense and safety prevail. However, when entering manholes, strict safety procedures must be observed. Failure to do so could jeopardize not only your own life, but also the lives of other crew members.

C.1.1 Hazards

There are many hazards connected with entering manholes. Some of the most common hazards are: Adverse Atmosphere The manhole may contain flammable or poisonous gases or the atmosphere may be deficient in oxygen. orced ventilation may be necessary. Deteriorated Rungs Manhole steps may be corroded and not strong enough to support a man. It may be difficult to inspect the rungs because of poor lighting. Traffic Whenever manholes are located in the traveled way, barricades and warning devices are essential to direct traffic away from an open manhole. Falling Objects Items placed near the manhole opening may fall and injure a worker in the manhole.

C-1

6712 Portable Sampler Appendix C General Safety Procedures Sharp Edges Sharp edges of items in or near a manhole may cause cuts or bruises. Lifting Injuries Unless proper tools are used to remove manhole covers, back injuries or injuries to hands or feet may result. C.1.2 Planning Advance planning should include arrangements for test equipment, tools, ventilating equipment, protective clothing, traffic warning devices, ladders, safety harness, and adequate number of personnel. Hasty actions may result in serious injuries. Time spent in the manhole should be kept to a minimum. [Refer to Table C-1, Hazardous Gases, at the end of this appendix.] Before workers enter a manhole, tests should be made for explosive atmosphere, presence of hydrogen sulfide, and oxygen deficiency. Combustible or toxic vapors may be heavier than air, so the tests on the atmosphere must be run at least 3/4 of the way down the manhole. Whenever adverse atmosphere is encountered, forced ventilation must be used to create safe conditions. After the ventilating equipment has been operated for a few minutes, the atmosphere in the manhole should be retested before anyone enters the manhole. When explosive conditions are encountered, the ventilating blower should be placed upwind to prevent igniting any gas that is emerging from the opening. When a gasoline engine blower is used, it must be located so that exhaust fumes cannot enter the manhole. If testing equipment is not available, the manhole should be assumed to contain an unsafe atmosphere and forced ventilation must be provided. It should never be assumed that a manhole is safe just because there is no odor or the manhole has been entered previously. C.1.4 Entering Manholes Since the top of the manhole is usually flush with the surrounding surface, there may not be anything for the person who is entering the manhole to grab on to steady himself. Persons who are entering manholes should not be permitted to carry anything in their hands as they enter the manhole, to ensure that their hands will be free to hold on or grab if they slip. A good method for entering a manhole is to sit on the surface facing the manhole steps or ladder, with the feet in the hole and the arms straddling the opening for support. As the body slides forward and downward, the feet can engage a rung, and the back can rest against the opposite side of the opening. If there is any doubt about the soundness of the manhole steps, a portable ladder should be used. A person should never enter a manhole unless he is wearing personal safety equipment, including a safety harness and a hard hat. Two persons should be stationed at the surface continuously while anyone is working inside a manhole, to lift him out if he is

C.1.3 Adverse Atmospheres

C-2

6712 Portable Sampler Appendix C General Safety Procedures overcome or injured. One man cannot lift an unconscious man out of a manhole. The persons stationed at the surface should also function as guards to keep people and vehicles away from the manhole opening. To avoid a serious injury, a person should not be lifted out of a manhole by his arm unless it is a dire emergency. When more than one person must enter a manhole, the first person should reach the bottom and step off the ladder before the next one starts down. When two men climb at the same time, the upper one can cause the lower one to fall by slipping or stepping on his fingers. C.1.5 Traffic Protection In addition to traffic cones, markers, warning signs, and barricades, a vehicle or a heavy piece of equipment should be placed between the working area and oncoming traffic. Flashing warning signals should be used to alert drivers and pedestrians. Orange safety vests should be worn by personnel stationed at the surface when the manhole is located in a vehicular traffic area. All loose items should be kept away from the manhole opening. This applies to hand tools as well as stones, gravel and other objects. Manhole covers should be removed with a properly designed hook. Use of a pick ax, screwdriver, or small pry bar may result in injury. A suitable tool can be made from 3/4-inch round or hex stock. Two inches of one end should be bent at a right angle and the other end should be formed into a D-handle wide enough to accommodate both hands. Even with this tool, care must be exercised to prevent the cover from being dropped on the toes. The 2-inch projection should be inserted into one of the holes in the cover, the handle grasped with both hands, and the cover lifted by straightening the legs which have been slightly bent at the knees. Other precautions which should be taken when entering a manhole are: Wear a hard hat. Wear coveralls or removable outer garment that can be readily removed when the work is completed. Wear boots or nonsparking safety shoes. Wear rubberized or waterproof gloves. Wear a safety harness with a stout rope attached. Do not smoke. Avoid touching yourself above the collar until you have cleaned your hands.

C.1.6 Falling Objects

C.1.7 Removing the Covers

C.1.8 Other Precautions

C-3

6712 Portable Sampler Appendix C General Safety Procedures C.1.9 Emergencies Every member of the crew should be instructed on procedures to be followed in cases of an emergency. It is the duty of each crew chief to have a list of emergency phone numbers, including the nearest hospital and ambulance service, police precinct, fire station, and rescue or general emergency number. The following equipment will be available for use: Blowers Harnesses Emergency flashers Rain slickers Gas detectors Waders Gloves First aid kits Pick axes Mirrors Safety vests Traffic cones Coveralls Manhole irons Flashlights Ropes Gas masks Breathing apparatus Hard Hats

C.1.10 Field Equipment

C.2 Lethal Atmospheres in Sewers

The following is an article written by Dr. Richard D. Pomeroy, and published in the October 1980 issue of Deeds & Data of the WPCF. Dr. Pomeroy is particularly well known for his studies, over a period of nearly 50 years, in the field of the control of hydrogen sulfide and other odors in sewers and treatment plants. He has personally worked in a great many functioning sewers. In the earlier years he did so, he admits, with little knowledge of the grave hazards to which he exposed himself. It is gratifying that the subject of hazards to people working in sewers is receiving much more attention than in past years, and good safety procedures are prescribed in various publications on this subject. It is essential that people know and use correct procedures. It is less important to know just what the hazardous components of sewer atmospheres are, as safety precautions should in general be broadly applicable, but there should be a reasonable understanding of this subject. It is disturbing to see statements in print that do not reflect true conditions. One of the most common errors is the assumption that people have died from a lack of oxygen. The human body is able to function very well with substantially reduced oxygen concentrations. No one worries about going to Santa Fe, New Mexico, (elev. 2,100 meters), where the partial pressure of oxygen is equal to 16.2% (a normal atmosphere is about 21%) oxygen. When first going there, a person may experience a little shortness of breath following exercise. People in good health are not afraid to drive over the high passes in the Rocky Mountains. At Loveland Pass, oxygen pressure is 13.2% of a normal atmosphere. At the top of Mt. Whitney, oxygen is equal to 12.2%t. Many hikers go there, and to higher peaks as well. After adequate acclimation, they may climb to the top of Mt. Everest, where oxygen is equal to only 6.7%.

C-4

6712 Portable Sampler Appendix C General Safety Procedures The lowest oxygen concentrations that I have observed in a sewer atmosphere was 13 percent. It was in a sealed chamber, near sea level, upstream from an inverted siphon on a metropolitan trunk. A man would be foolish to enter the chamber. Without ventilation, he might die, but not from lack of oxygen. It seems unlikely that anyone has ever died in a sewer from suffocation, that is, a lack of oxygen. Deaths have often been attributed to asphyxiation. This is a word which, according to the dictionary, is used to mean death from an atmosphere that does not support life. The word has sometimes been misinterpreted as meaning suffocation, which is only one kind of asphyxiation. In nearly all cases of death in sewers, the real killer is hydrogen sulfide. It is important that this fact be recognized. Many cities diligently test for explosive gases, which is very important, and they may measure the oxygen concentration which usually is unimportant, but they rarely measure H2S. Death has occurred where it is unlikely that there was any measurable reduction in the oxygen concentration. Waste water containing 2 mg per liter of dissolved sulfide, and at a pH of 7.0, can produce, in a chamber with high turbulence, a concentration of 300 PPM H2S, in the air. This is considered to be a lethal concentration. Many people have died from H2S, not only in sewers and industries, but also from swamps and from hot springs. In one resort area, at least five persons died from H2S poisoning before the people were ready to admit that H2S is not a therapeutic agent. Hardly a year passes in the U.S. without a sewer fatality from H2S as well as deaths elsewhere in the world. The presence of H2S in a sewer atmosphere is easily determined. A bellows-and-ampoule type of tester is very satisfactory for the purpose, even though it is only crudely quantitative. When using a tester of this type, do not bring the air to the ampoule by way of a tube, as this may change the H 2 S concentration. Hang the ampoule in the air to be tested, with a suction tube to the bulb or bellows. Lead acetate paper is very useful as a qualitative indicator. It cannot be used to estimate the amount of sulfide, but it will quickly turn black in an atmosphere containing only a tenth of a lethal concentration. Electrodes or other similar electrical indicating devices for H2S in air have been marketed. Some of them are known to be unreliable, and we know of none that have proved dependable. Do not use one unless you check it at frequent intervals against air containing known H2S concentrations. A supposed safety device that is unreliable is worse than none at all. Remember that the nose fails, too, when it comes to sensing dangerous concentrations of H2S. Various other toxic gases have been mentioned in some publications. It is unlikely that any person has been asphyxiated in a sewer by any of those other gases, except possibly chlorine. The vapor of gasoline and other hydrocarbons is sometimes present in

C-5

6712 Portable Sampler Appendix C General Safety Procedures amounts that could cause discomfort and illness, but under that condition, the explosion hazard would be far more serious. The explosimeter tests, as well as the sense of smell, would warn of the danger. Pipelines in chemical plants might contain any number of harmful vapors. They, too, are sensed by smell and explosimeter tests if they get into the public sewer. Such occurrences are rare. The attempt to instill a sense of urgency about real hazards is diluted if a man is told to give attention to a long list of things that in fact are irrelevant. Be very careful to avoid high H 2 S concentrations, flammable atmospheres, and hazards of physical injuries. Remember that much H2S may be released by the stirring up of sludge in the bottom of a structure. Obey your senses in respect to irritating gases, such as chlorine (unconsciousness comes suddenly from breathing too much). Be cautious about strange odors. Do not determine percent oxygen in the air. There is a danger that the result will influence a man's thinking about the seriousness of the real hazards. Most important, use ample ventilation, and do not enter a potentially hazardous structure except in a good safety harness with two men at the top who can lift you out.

C.3 Hazardous Gases

The following table contains information on the properties of hazardous gases.

Table C-1 Hazardous Gases


Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 0.60 Physiological Effect Max Safe 60 Min. Exposure ppm 300 to 500 Max. Safe 8 Hour Exposure ppm 85 Explosive Range (% by vol. in air) Limits lower/upper 16 25 Likely Location of Highest Concentration Near top. Concentrates in closed upper spaces Most Common Sources Simplest and Cheapest Safe Method of Testing Detectable odor at low concentrations

Ammonia

NH3

Irritant and poisonous. Colorless with characteristic odor.

Causes throat and eye irritation at 0.05%, coughing at 0.17%. Short exposure at 0.5% to 1% fatal. Slight symptoms after several hours exposure at 0.16% to 0.32%. 2% rapidly fatal. Very poisonous, irritating, vomiting, convulsions, psychic disturbance.

Sewers, chemical feed rooms.

Benzene

C6H6

Irritant, colorless anesthetic

2.77

3,000 to 5,000

25

1.3

7.1

At bottom.

Industrial wastes, varnish, solvents.

Combustible gas indicator

Carbon Bisulfide

CS2

Nearly odorless when pure, colorless, anesthetic. Poisonous.

2.64

15

1.3

44.0

At bottom

An insecticide

Combustible gas indicator

C-6

6712 Portable Sampler Appendix C General Safety Procedures

Table C-1 Hazardous Gases (Continued)


Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 1.53 Physiological Effect Max Safe 60 Min. Exposure ppm 40,000 to 60,000 Max. Safe 8 Hour Exposure ppm 5,000 Explosive Range (% by vol. in air) Limits lower/upper Likely Location of Highest Concentration At bottom; when heated may stratify at points above bottom. Most Common Sources Simplest and Cheapest Safe Method of Testing Oxygen deficiency indicator

Carbon Dioxide

CO2

Asphyxiant, Colorless, odorless. When breathed in large quantities, may cause acid taste. Non-flammable. Not generally present in dangerous amounts unless an oxygen deficiency exists. Chemical asphyxiant. Colorless, odorless, tasteless. Flammable. Poisonous.

Cannot be endured at 10% more than a few minutes, even if subject is at rest and oxygen content is normal. Acts on respiratory nerves.

Products of combustion, sewer gas, sludge. Also issues from carbonaceous strata.

Carbon Monoxide

CO

0.97

Combines with hemoglobin of blood. Unconsciousness in 30 min. at 0.2% to 0.25%. Fatal in 4 hours at 0.1%. Headache in few hours at 0.02%.

400

50

12.5 74.0

Near top, especially if present with illuminating gas.

Manufactured gas, flue gas, products of combustion, motor exhausts. Fires of almost any kind. Industrial wastes, solvent, cleaning

CO ampoules.

Carbon Tetra-Chl oride

CCl4

Heavy, ethereal odor.

5.3

Intestinal upset, loss of consciousness, possible renal damage, respiratory failure. Irritates respiratory tract. Kills most animals in a very short time at 0.1%.

1,000 to 1,500

100

At bottom.

Detectable odor at low concentrations.

Chlorine

Cl2

Irritant. Yellow-green color. Choking odor detectable in very low concentrations. Non-flammable. Colorless, pungent suffocating odor.

2.49

At bottom.

Chlorine cylinder and feed line leaks.

Detectable odor at low concentrations.

Formaldehyde

CH2O

1.07

Irritating to the nose.

10

7.0

73.0

Near bottom.

Incomplete combustion of organics. Common air pollutant, fungicide. Service stations, garages, storage tanks, houses.

Detectable odor.

Gasoline

C5H12 to C9H20

Volatile solvent. Colorless. Odor noticeable at 0.03%. Flammable.

3.0 to 4.0

Anesthetic effects when inhaled. Rapidly fatal at 2.4%. Dangerous for short exposure at 1.1 to 2.2%. Acts mechanically to deprive tissues of oxygen. Does not support life.

4,000 to 7,000

1,000

1.3

6.0

At bottom.

1. Combustible gas indicator. 2. Oxygen deficiency indicator.**

Hydrogen

H2

Simple asphyxiant. Colorless, odorless, tasteless. Flammable

0.07

4.0

74.0

At top.

Manufactured gas, sludge digestion tank gas, electrolysis of water. Rarely from rock strata. Insecticide and rodenticide.

Combustible gas indicator.

Hydrogen Cyanide

HCN

Faint odor of bitter almonds. Colorless gas

0.93

Slight symptoms appear upon exposure to 0.002% to 0.004%. 0.3% rapidly fatal.

10

6.0

40.0

Near top.

Detector tube

C-7

6712 Portable Sampler Appendix C General Safety Procedures

Table C-1 Hazardous Gases (Continued)


Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 Specific Gravity or Vapor Density Air = 1 1.19 Physiological Effect Max Safe 60 Min. Exposure ppm Max Safe 60 Min. Exposure ppm 200 to 300 Max. Safe 8 Hour Exposure ppm Max. Safe 8 Hour Exposure ppm 20 Explosive Range (% by vol. in air) Limits lower/upper Explosive Range (% by vol. in air.) Limits lower/upper 4.3 45.0 Likely Location of Highest Concentration Likely Location of Highest Concentration Near bottom, but may be above bottom if air is heated and highly humid. Most Common Sources Simplest and Cheapest Safe Method of Testing Simplest and Cheapest Safe Method of Testing

Gas

Chemical Formula

Common Properties

Physiological Effect*

Most Common Sources

Hydrogen Sulfide

H2S

Irritant and poisonous volatile compound. Rotten egg odor in small concentrations. Exposure for 2 to 15 min. at 0.01% impairs sense of smell. Odor not evident at high concentrations. Colorless. Flammable. Simple asphyxiant. Colorless, odorless, tasteless, flammable.

Impairs sense of smell, rapidly as concentration increases. Death in few minutes at 0.2%. Exposure to 0.07 to 0.1% rapidly causes acute poisoning. Paralyzes respiratory center. Acts mechanically to deprive tissues of oxygen. Does not support life.

Coal gas, petroleum, sewer gas. Fumes from blasting under some conditions. Sludge gas.

1. H2S Ampoule. 2. 5% by weight lead acetate solution.

Methane

CH4

0.55

Probably no limit, provided oxygen percent-age is sufficient for life.

5.0

15.0

At top, increasing to certain depth.

Natural gas, sludge gas, manufactured gas, sewer gas. Strata of sedimentary origin. In swamps or marshes. Sewer gas. sludge gas. Also issues from some rock strata.

1. Combustible gas indicator 2. Oxygen deficiency indicator.

Nitrogen

N2

Simple asphyxiant. Colorless, tasteless. Non-flammable. Principal constituent of air. (about 79%). Colorless Colorless, sweet odor. Reddish-brown. Irritating odor. Deadly poison Colorless, odorless, tasteless. Supports combustion.

0.97

Physiologically inert.

Near top, but may be found near bottom.

Oxygen deficiency indicator.

Nitrogen Oxides

NO N2O NO2

1.04

60 to 150 ppm cause irritation and coughing. Asphyxiant.

50

10

Near bottom.

1.53 1.58 100 ppm dangerous. 200 ppm fatal. 1.11 Normal air contains 20.8% of O2. Man can tolerate down to 12%. Minimum safe 8 hour exposure, 14 to 16%. Below 10%, dangerous to life. Below 5 to 7% probably fatal. Max. naturally occurring level is 0.04 ppm. 0.05 ppm causes irritation of eyes and nose. 1 to 10 ppm causes headache, nausea; can cause coma. Symptoms similar to radiation damage. Variable at different levels.

Industrial wastes. Common air pollutant.

NO2 detector tube.

Oxygen

O2

Oxygen depletion from poor ventilation and absorption, or chemical consumption of oxygen. Where ozone is used for disinfection.

Oxygen deficiency indicator.

Ozone

O3

Irritant and poisonous. Strong electrical odor. Strong oxidizer. Colorless. At 1 ppm, strong sulfur-like odor.

1.66

0.08

0.04

Near bottom.

Detectable odor at 0.015 ppm.

C-8

6712 Portable Sampler Appendix C General Safety Procedures

Table C-1 Hazardous Gases (Continued)


Gas Chemical Formula Common Properties Specific Gravity or Vapor Density Air =1 Variable Physiological Effect Max Safe 60 Min. Exposure ppm Max. Safe 8 Hour Exposure ppm Explosive Range (% by vol. in air) Limits lower/upper 5.3 19.3 Likely Location of Highest Concentration Near top of structure. Most Common Sources Simplest and Cheapest Safe Method of Testing See components.

Sludge Gas

***

Mostly a simple asphyxiant. May be practically odorless, tasteless. Colorless, pungent odor. Suffocating, corrosive, poisonous, non-flammable. Colorless, benzene-like odor.

Will not support life.

No data. Would vary widely with composition.

From digestion of sludge.

Sulfur Dioxide

SO2

2.26

Inflammation of the eyes. 400 to 500 ppm immediately fatal.

50 to 100

10

At bottom, can combine with water to form sulfurous acid.

Industrial waste, combustion, common air pollutant. Solvent.

Detectable taste and odor at low concentration.

Toluene

C5H12 to C9H20 C10H16

3.14

At 200-500 ppm, headache, nausea, bad taste, lassitude. Eye irritation. Headache, dizziness, nausea, irritation of the kidneys.

200

100

1.27 7.0

At bottom.

Combustible gas indicator.

Turpentine

Colorless, Characteristic odor.

4.84

100

At bottom.

Solvent, used in paint.

1. Detectable odor at low concentration s. 2.Combustible gas indicator. Combustible gas indicator.

Xylene

C8H10

Colorless, flammable

3.66

Narcotic in high concentrations. less toxic than benzene.

100

1.1

7.0

At bottom.

Solvent

* Percentages shown represent volume of gas in air. ** For concentration over 0.3%. ***Mostly methane and carbon dioxide with small amounts of hydrogen, nitrogen, hydrogen sulfide, and oxygen; occasionally traces of carbon monoxide.

C-9

6712 Portable Sampler Appendix C General Safety Procedures

C-10

6712 Portable Sampler


Appendix D Replac ement Parts Lis t

D.1 Replacement Parts Diagrams and Listings

Replacement parts are called out in illustrations in this section. Reference the call-outs in the accompanying tables to determine the part number for the item. Replacement parts can be purchased by contacting Teledyne Iscos Customer Service Department. Teledyne Isco, Inc. Customer Service Department P.O. Box 82531 Lincoln, NE 68501 USA Phone: (800) 228-4373 (402) 464-0231 FAX:(402) 465-3022 E-mail:[email protected]

D-1

6712 Portable Sampler Appendix D Replacement Parts List

Controller Mounting Screws 8-32 x 3/4 (4) #8 Lockwashers (4) #8 Flatwashers (4)

D-2

6712 Portable Sampler Appendix D Replacement Parts List

Controller Mounting Screws 8-32 x 3/4 (4) #8 Lockwashers (4) #8 Flatwashers (4)

D-3

6712 Portable Sampler Appendix D Replacement Parts List

Locknut, 10-32

Screw, 10-32 x 3/4

D-4

6712 Portable Sampler Appendix D Replacement Parts List

D-5

6712 Portable Sampler Appendix D Replacement Parts List

D-6

6712 Portable Sampler Appendix D Replacement Parts List

D-7

6712 Portable Sampler Appendix D Replacement Parts List

D-8

6712 Portable Sampler


Appendix E Ac c es s ories Lis t

E.1 Order Information

Prices available on request. Additional items appear in Appendix D, Replacement Parts List. Many other items are available. To order any item, contact your sales representative or the factory. Note that the part name listed on your order acknowledgment and invoice may be different from the item name listed here. When examining these documents, use the part number for reference. Teledyne Isco, Inc. P.O. Box 82531 Lincoln, NE 68501 Telephone Within U.S.A. Toll free: (800) 228-4373 Outside U.S.A:(402) 464-0231 Fax: (402) 465-3022 [email protected]

E.2 Controller, Center Sections, Top Covers, and Tubs


6712 Sampler Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6710-074 (Includes 2 Pump Tubes) 6712 Portable Sampler Instruction Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-334 6712 Portable Sampler Pocket Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-589 Compact Center Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-245 Compact Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3704-129 Mini-Tub (for use with Compact Tub and multiple bottle carrier only) . . . . . . . . . . . . . . . . . . 60-3704-181 Compact Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-248 Standard Center Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-246 Standard Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-247 Standard Tub . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-052

E.3 Compact and Mini Sampler Bottle Kits


24-Bottle Kit (500-milliliter Polyethylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68-6700-022 (Includes Bottle Carrier, Bottles, Retaining Ring and 2 Discharge Tubes.) 12-Bottle Kit (500-milliliter Polyethylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68-6700-023 (Includes Bottle Carrier, Retaining Ring, Bottles, and 2 Discharge Tubes.) 12-Bottle Kit (375-milliliter Glass Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-024 (Includes Bottle Carrier, Retaining Ring, Bottles, and 2 Discharge Tubes.)

E-1

6712 Portable Sampler Appendix E Accessories List

E.4 Compact Sampler Bottle Kits


1-Bottle Kit (9.4-liter Polyethylene) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-025 (Includes Bottle and 2 Discharge Tubes.) 1-Bottle Kit (9.4-liter Glass Bottle). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-026 (Includes Bottle and 2 Discharge Tubes.)

E.5 Standard Sampler Bottle Kits


24-Bottle Kit (1,000-milliliter Polypropylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-006 (Includes Retaining Ring, Bottles, and 2 Discharge Tubes.) 24-Bottle Kit (350-milliliter Glass Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-007 (Includes Retaining Ring, Bottles, and 2 Discharge Tubes.) 12-Bottle Kit (1,000-milliliter Polypropylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-008 (Includes Insert, Bottles, Retaining Ring, and 2 Discharge Tubes.) 12-Bottle Kit (950-milliliter Glass Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-009 (Includes Insert, Retaining Ring Bottles, and 2 Discharge Tubes.) 8-Bottle Kit (2.0-liter Polyethylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-010 (Includes Insert, Bottles, Retaining Ring, and 2 Discharge Tubes.) 8-Bottle Kit (1.8-liter Glass Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-011 (Includes Insert, Bottles, Retaining Ring, and 2 Discharge Tubes.) 4-Bottle Kit (3.7-liter Polyethylene Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-012 (Includes Bottles, Rack, and 2 Discharge Tubes.) 4-Bottle Kit (3.7-liter Glass Bottles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-013 (Includes Bottles, Rack, and 2 Discharge Tubes.) 1-Bottle Kit (9.4-liter Polyethylene Bottle). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-014 (Includes Bottle and 2 Discharge Tubes.) 1-Bottle Kit (9.4-liter Glass Bottle). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-015 (Includes Bottle and 2 Discharge Tubes.) 4 (15 Liter) Gallon Bottle Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-080 (Includes Jumbo Base, Polyethylene Bottle, Cap, Two Discharge Tubes, and Composite Tube Guide.) 5 Gallon (19 Liter) Bottle Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-078 (Includes Jumbo Base, Glass Bottle, Cap, PTFE Cap Liner, Two Discharge Tubes, and Composite Tube Guide.) 5.5 Gallon (20.8 Liter) Bottle Configuration (Includes Jumbo Base, Polyethelene Bottle, Cap, Two Discharge Tubes, and Composite Tube Guide.) ProPak Starter Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-112 (Includes 24 holders with caps, 100 liners, retaining ring, instruction video, instruction sheet.) ProPak Configuration Kit for 6700 Full-size Samplers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-105 (Includes 25 holders, 1,000 liners, retaining ring, instruction video, instruction sheet.) ProPak holders with caps, and standard liners (24) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-087 ProPak disposable liners, 1-liter polyethylene (100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-113 ProPak disposable liners, 1-liter polyethylene (500) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-114 ProPak disposable liners, 1-liter polyethylene (1000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-096 ProPak labels (1000) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-100 ProPak retaining ring for 24 holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-475

E-2

6712 Portable Sampler Appendix E Accessories List

E.6 Bottle Carriers, Retaining Rings and Racks, Bottle Inserts, Discharge Tubes
Bottle Carrier for Compact 12-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-131 Bottle Carrier for Compact 24-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-9004-132 Retaining Ring for Compact 12-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-2904-062 Retaining Ring for Compact 24-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-235 Retaining Ring for Standard 24-Bottle Kit (1000-milliliter Polypropylene Bottles) . . . . . . . . 60-9003-242 Retaining Ring for Standard 24-Bottle Kit (350 milliliter Glass Bottles) . . . . . . . . . . . . . . . . 60-2703-011 Retaining Ring for Standard 12-Bottle Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3704-112 Bottle Insert for Standard 12-Bottle Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-147 Retaining Ring for Standard 8-Bottle Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-149 Bottle Insert for Standard 8-Bottle Kits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-146 Retaining Rack for Standard 4-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-351 Discharge Tube for Compact 12- Bottle Kit and 24-Bottle Kit . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-257 Discharge Tube for Compact 1 Bottle Kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-258 Discharge Tube for Standard 24-Bottle Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-260 Discharge Tube for Standard 12-Bottle Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-262 Discharge Tube for Standard 8-Bottle Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-9003-264 Discharge Tube for Standard 4-Bottle Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-9003-266 Discharge Tube for Standard 1-Bottle Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-271 Bulk Discharge Tube (10 foot Length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-6700-046 Bulk Discharge Tube (50 foot Length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-6700-047

E.7 Bulk Sets of Bottles with Lids


Set of 24, 500-milliliter Polyethylene Bottles with Polyethylene Foam Lined Lids . . . . . . . . . 68-2900-002 Set of 100, 500-milliliter Polyethylene Bottles with Polyethylene Foam Lined Lids . . . . . . . . 68-6700-027 Set of 500, 500-milliliter Polyethylene Bottles with Polyethylene Foam Lined Lids . . . . . . . . 68-6700-028 Set of 12, 500-milliliter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . 68-2900-014 Set of 100, 500-milliliter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . 68-6700-029 Set of 500, 500-milliliter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . 68-6700-030 Set of 12, 375-milliliter Glass Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-2900-015 Set of 100, 375-milliliter Glass Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68-6700-031 Set of 24, 1,000-milliliter Polypropylene Bottles with Polyethylene Foam Lined Lids . . . . . .68-2100-007 Set of 100, 1,000-milliliter Polypropylene Bottles with Polyethylene Foam Lined Lids . . . . . 68-3700-046 Set of 500, 1,000-milliliter Polypropylene Bottles with Polyethylene Foam Lined Lids . . . . . 68-3700-047 Set of 24, 350-milliliter Glass Bottles with PTFE Lined Lids. . . . . . . . . . . . . . . . . . . . . . . . . . 68-2100-001 Set of 100, 350-milliliter Glass Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . 68-3700-048 Set of 12, 1,000-milliliter Polypropylene Bottles with Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3700-035 Set of 100, 1,000-milliliter Polypropylene Bottles with Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-016 Set of 500, 1,000-milliliter Polypropylene Bottles with Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-017 Set of 12, 950-milliliter Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3700-023 Set of 100, 950-milliliter Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-018 Set of 8, 2.0-liter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-2740-026 Set of 100, 2.0-liter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3720-028

E-3

6712 Portable Sampler Appendix E Accessories List


Set of 8, 1.8-liter Glass Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-2740-027 Set of 100, 1.8-liter Glass Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3720-030 3.7-liter Glass Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-147 3.7-liter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-2910-003 Set of 4, 3.7-liter Glass Bottles with PTFE Lined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-146 Set of 4, 3.7-liter Polyethylene Bottles with Unlined Lids . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-020 9.4-liter (21/2-gallon) Glass Bottle with PTFE Lined Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-2700-005 9.4-liter (21/2-gallon) Polyethylene Bottle with Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .299-0013-04 18.9-liter (5-gallon) Glass Bottle with Cap and Teflon Liner (for the 5 Gallon Bottle Base) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-079

E.8 General Accessories: Compact and Standard Samplers


Composite Tube Guide (Recommended for all 1-Bottle Kits) . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-051 Distributor Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-050 Equipment Suspension Platform (For Suspending Samplers In Manhole or Other Locations). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-151 Locking Assy, compact samplers (includes locking cables and padlock) . . . . . . . . . . . . . . . . . . 68-6700-111 Locking Assy, standard samplers (includes locking cables and padlock) . . . . . . . . . . . . . . . . . 68-6700-134 Suspension Harness for Compact and Standard Samplers. . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1394-014 1000-milliliter Plastic Graduated Cylinder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 299-0020-00

E.9 Pump Tubes, Suction Line, Strainers


Pump Tube (Qty 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-062 Pump Tube (Qty 10). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-044 Pump Tube (Qty 25). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-045
3 3

/8-inch Standard weighted polypropylene strainer with 10 foot Suction Line . . . . . . . . . . . . 60-9004-378 /8-inch Standard weighted polypropylene strainer with 25 foot Suction Line . . . . . . . . . . . . 60-9004-379

PTFE Suction Line (10 foot length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1683-146 PTFE Suction Line (25 foot length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-2703-114 Bulk Vinyl Suction Line (100 foot length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68-1680-058 Bulk Vinyl Suction Line (500 foot length) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68-1680-059
3

/8-inch Vinyl Suction Line Accessory Kit (includes tube coupling, /8-inch Standard weighted polypropylene strainer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-367 /8-inch Stainless Steel Low Flow Strainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69-2903-138

clamp, and instructions) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3700-007


3

CPVC Plastic-coated Weighted Strainer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3704-066


3 3/8-inch Suction Line Tubing Coupler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69-4703-106

E-4

6712 Portable Sampler Appendix E Accessories List

E.10 Data Collection Devices and Cables


581 Rapid Transfer Device (RTD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-027 Power Cable 581 Rapid Transfer Device (RTD) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-077 Flowlink for Windows Software New License. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . call factory Flowlink for Windows Software Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . call factory SAMPLINK Software (Includes Instruction Manual) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3774-013 Computer Connect Cable (25-pin). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-2544-040 Computer Connect Cable (9-pin). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-2544-044 Connector to Attach Non-Teledyne Isco Flow Meter to Customer Supplied Cable . . . . . . . . . 68-1680-060 Serial Output Connect Cable (for use with the periodic serial output feature) . . . . . . . . . . . . 60-9004-263 External 12-volt DC Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1394-023 Flow-Meter to Two-Samplers Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-018 Dual Sampler Mode Interconnect Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3704-065 Dual Sampler Mode to Flow-Meter Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-3704-080 Non-Teledyne Isco Flow-Meter to Sampler Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1394-077 Sampler to Flow Meter /1640 Liquid Level Actuator Y Cable . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-019 Sampler to Flow Meter Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-107 2100 Series Sampler Interface Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-2004-260 Solar Panel Y Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-098 Solar Panel/Interrogator Extension Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-2544-028 Solar Panel Cable (25 foot length). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-097 Rain Gauge/Refrigerator Temperature Sensor Y cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-294 4200T Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . factory installed (Contact your sales representative or Teledyne Isco Customer Service.)

Note
The modem is disabled when an interrogator cable is connected to the samplers Interrogator port. It cannot receive incoming calls, and the alarm dialout is rendered inoperative, while this cable is connected. Disconnect the interrogator cable in order to use the 4200T modem.
CDMA Digital Cellular Modem. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5314-489 GSM Digital Cellular Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5314-563 (Requires the purchase of a SIM card from your preferred provider.)

E.11 Power Sources and Chargers


913 High Capacity Power Pack (120-volt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-1684-088 914 Battery-Backed Power Pack (120-volt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-130 921 Nickel-Cadmium Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1684-040 923 High Capacity Power Pack (240-volt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60-1684-093 924 Battery-Backed Power Pack (240-volt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-160 947 Lead-Acid Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-106 948 45-Amp-Hour Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-3000-948 Portable 12-volt DC, 6-Amp Battery Charger for 948 Battery . . . . . . . . . . . . . . . . . . . . . . . . . 341-0118-12 961 Battery Charger (120-volt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3004-059

E-5

6712 Portable Sampler Appendix E Accessories List


965 Five-Station Battery Charger, 120/240 V, with 120 V cord. . . . . . . . . . . . . . . . . . . . . . . . . 68-3000-965 965 Five-Station Battery Charger, 120/240 V, with 240 V cord. . . . . . . . . . . . . . . . . . . . . . . . . 68-3000-966 Solar Panel Battery Charger, 5-watt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5314-478 Solar Panel Battery Charger, 40-watt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5314-347 Solar Panel Battery Charger, 90-watt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-5314-399 Power Products Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9003-092

E.12 Modules, Rain Gauges, & Interfacing Instruments

(Note: Additional accessories for modules and interfacing instruments appear in the accessories list at the back of the manual for each module or instrument.)

701 pH Module with Double Junction pH Sensor with Temperature Sensor. . . . . . . . . . . . . . 68-6700-052 710 Ultrasonic Flow Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-049 720 Submerged Probe Module, 10 ft range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-068 730 Bubbler Flow Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-050 780 Analog Interface Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-040 675 Rain Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3284-013 1640 Liquid Level Actuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-1644-000 4-20 mA Sampler Input Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-3704-075 Refrigerator Temperature Sensor with 5 ft (1.55 m) cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68-6700-136 Internal three channel 4-20 mA programmable output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-266 (Includes multi-analog output cable, 10 ft.)

E.13 SDI-12 Data Acquisition Connect Cables

Note: Y-connect cables can be combined with other Y-connect cables or SDI-12 adapter cables if multiple SDI-12 devices are to be connected.

Rain Gauge Y-Connect Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-294 (This cable allows connection of 2 of the following: rain gauge, SDI-12 adapter cable, SDI-12 Y-connect cable, I/O adapter, temperature sensor.) YSI SDI-12 Adapter Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-337 YSI SDI-12 Y-Connect Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-338 (This cable allows connection of a YSI SDI-12 sonde and a Teledyne Isco Rain Gauge.) Hydrolab Quanta SDI-12 Adapter Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-339 Hydrolab Quanta SDI-12 Y-Connect Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-340 (This cable allows connection of a Hydrolab Quanta SDI-12 sonde and a Teledyne Isco Rain Gauge.) SDI-12 Adapter Cable (6 foot length, no connector). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-341 (This is for use with a customer supplied connector to an SDI-12 device.) SDI-12 Y-Connect Cable (6 foot length, no connector) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60-9004-342 (This is for use with a customer supplied connector to an SDI-12 device and a Teledyne Isco Rain Gauge.)

E-6

6712 Portable Sampler


Index

A
Accessories bulk discharge tubing, E-4 bulk sets of bottles, E-3 cables, E-5 composite tube guide, E-4 data collection accessories, E-5 discharge tube, E-3 distributor arm, E-4 equipment suspension platform, E-4 graduated cylinder, E-4 power sources and battery chargers, E-5 pump tube, E-4 refrigerator temperature sensor, E-6 strainers, E-4 suction line, E-4 suspension harness, E-4 acknowledge dialout alarm, 5-40 alarm acknowledgement, 5-40 alarms, 5-39 Alkaline battery battery warning, 8-13 opening controller case, 8-11 replacing, 8-13 Analog output, 5-41 analog output, 5-41

setting clock and calendar, 3-6, 8-4 Clock start-time menu, 3-6 Combined results report see Sampling reports command responses, 5-45 commands, external, 5-44 Compatible Isco products, 1-4 Composite tube guide accessories list, E-4 Computer control, 7-1 checksum validation, 7-8 external program control, 7-4 menu control, 7-2 remote keypad, 7-8 sampling reports, 7-3 Configuring reports, 4-20 Construction compact and standard sampler, 1-7 Continuous sampling, 2-17, 4-6, 5-17 Count Switched Distribution Sampler Enable, 5-23

D
Data collection accessories accessories list, E-5 data storage, 5-42 Desiccant opening controller case, 8-11 Diagnostic tests distributor test, 8-8 Diagnostics, 8-6 Dialout alarms, 5-39 dialout alarms, 5-39 Discharge tube accessories list, E-3 Display Backlighting, 5-30 Distribution, 4-6, 5-16 bottles-per-sample distribution, 4-6, 5-16 continuous sampling, 4-6, 5-17 count-switched distribution, 5-16 event pacing, 5-12 multiple-bottle compositing, 5-16 random-interval pacing, 5-15 Samples per bottle, 4-6, 5-16 sequential distribution, 4-6, 5-16 time-switched distribution, 5-16 Distributor arm accessories list, E-4 Dual sampler mode, 5-29

B
Backlighting, 5-30 Batteries, 2-9 Bottle kits collecting filled bottles, 2-17 installing, 2-3 Bottle Number Signal, 5-31 Bottles-per-sample distribution, 4-6, 5-16

C
Cables External 12-Volt DC Power Connect Cable, 2-9 Master/Slave-to-Flow-Meter Cable, 5-29 Calendar, 3-5 setting clock and calendar, 3-6, 8-4 Calibration during a manual pause, 4-13 sample volumes, 4-14 cleaning protocols, 8-2 Clock

Index-1

6712 Portable Sampler Index

E
Enable see Sampler enable Equipment suspension platform accessories list, E-4 Error messages, 8-14 Event marks, 5-30 3-second pulse, 5-31 master/slave sampling, 5-29 timing diagram, 5-31 variable-duration pulse, 5-31 Event pacing, 5-12 Extended programming extended programming features, 1-6?? one-part and two-part programs, 5-1 showing extended screens, 4-1, 5-1 External 12-Volt DC Power Connect Cable, 2-9

M
Maintenance, 8-3 alkaline battery, 8-13 checklist, 8-8 cleaning guidelines, 8-1 cleaning protocols, 8-2 diagnostics, 8-6 maintenance programming screens, 8-3 opening controller case, 8-11 pump counts for pump tube warning, 8-8 pump tube, 8-8 pump tube alarm, 8-8 reinitializing, 8-8 Manual functions, 4-14 calibrate volume, 4-14 grab samples, 4-14 moving distributor, 4-18 running pump manually, 4-17 Manual paused operation, 4-12?? Master/slave sampling, 5-29 event marks, 5-29 sampler enable, 5-29 Master/Slave-to-Flow-Meter Cable, 5-29 Materials used in sampler construction, 1-7 Memory, 1-4 capacity, 1-2, 5-43 creating partitions, 5-42 partition sizes, 5-42 roll-over, 5-43 stored data will be lost!, 5-1 triggered, 5-43 memory, 5-42 Menu screens, 3-3 Modem, 7-9 Module summary report see Sampling reports Modules 700 Series Modules, 1-2 displayed readings, 4-11 Multiple-bottle compositing, 5-16

F
Five gallon bottle base, 2-7 Flow pacing, 4-5 Flow proportional sample volumes, 5-19

G
Grab samples, 4-13, 4-15 Graduated cylinder accessories list, E-4

H
hardware analog output, 5-41 cellular modems, 5-40 dialout alarms, 5-39 I/O pin programming, 5-37 pager numbers, 5-40 rain gauge, 5-36 refrigerator temperature, 5-41 sonde, 5-36 Hardware set-up master/slave sampling, 5-29 hardware setup, 5-36

N I
I/O pin programming, 5-37 Installing suction line, 2-13 Interrupting interrupting a running program, 4-12 manual paused operation, 4-12 Nonuniform time pacing, 5-15

O
One-part programs, 5-1 Operating running sampling programs, 4-10, 5-27 Other Functions see Hardware setup, 4-14, 5-28 see Maintenance, 4-14, 5-28 see Manual functions, 4-14, 5-28 see Programming Style, 4-14, 5-28

L
Line rinses, 5-12 Liquid detector enable/disable, 5-28 Locking the sampler, 2-16

P
Pacing, 4-5, 5-12

Index-2

6712 Portable Sampler Index


event pacing, 5-12 flow pacing, 4-5 nonuniform clock-time pacing, 5-15 nonuniform time pacing, 5-15 random-interval pacing, 5-15 time pacing, 4-5, 5-12 trigger pacing, 4-5 Partitions creating, 5-42 Password Protected Functions, 5-35 Pause and resume, 5-27 Periodic serial output, 5-39 Power sources, 2-9 accessories list, E-5 Presample purge event marks, 5-31 pressurized lines, 5-44 Program name menu how to use, 3-7 Program settings report see Sampling reports Program storage see Stored programs Programming changing program name, 3-8 changing site description, 3-8 clock and calendar, 3-5 extended programming features, 1-6?? menu screens, 3-3 program name menu, 3-7 running programs, 4-10, 5-27 site description menu, 3-7 text entry, 3-7 Pump running pump manually, 4-17 Pump Counts pump tube warning, 8-8 Pump counts purging a nonstandard suction line, 5-31 Pump tube accessories list, E-4 replacing, 8-8 warning, 8-4 pump tube warning, 8-15 computer commands, 7-1 telephone commands, 7-9 Replacing internal alkaline battery, 8-13 Reports, 4-19 Resume, 5-27 Retries, 5-12 Rinses, 5-12 RS-232 communications, 7-1 Running programs, 4-10, 5-27

S
Sample volume calibrating, 4-14 flow proportional, 5-19 measuring, 2-13 variable volume sampling, 5-19 Sampler enable, 4-9, 5-23 "dry period" option, 5-26 master/slave sampling, 5-29 repeatable enable, 5-25 sample at enable or disable, 5-25 start delay, 5-26 start times, 4-8 stay enabled, 5-25 Samples per bottle, 4-6, 5-16 Sampling programs running programs, 4-10, 5-27 Sampling reports Collecting reports, 4-19, 5-28 configuring, 4-20 sampling results report, 4-21 Viewing, 4-19 Sampling results report see Sampling reports Sampling retries, 5-12 Sequential distribution, 4-6, 5-16 Serial output, 5-39, 7-6 Servicing, 2-16 collecting filled bottles, 2-17 recovering the sampler, 2-17 service schedule, 2-17 Site description menu how to use, 3-7 sonde setup, 5-36 Start times clock start-time menu, 3-6 event pacing, 5-12 first valid day, 4-8, 5-27 sampler enable, 4-8 Stored data, 5-1 Stored programs, 5-2 effects of reinitializing controller to factory settings, 8-8 selecting a program, 5-2 Stormwater monitoring Two-part programs, 5-2 Strainers accessories list, E-4

Q
Quick View Screens, 3-4 see Program Style

R
rain gauge setup, 5-36 Random-interval pacing, 5-15 Recovering the sampler, 2-17 refrigerator temperature, 5-41 Refrigerator temperature sensor, 5-41, E-6 Reinitialize controller effects on stored programs, 8-8 Remote access

Index-3

6712 Portable Sampler Index


selecting, 2-11 Suction head auto suction head, 5-12 automatic determination, 5-12 manual (ENTER HEAD) settings, 5-12 variations in pump counts for fluctuating heads, 4-21 Suction line accessories list, E-4 attaching to pump tube, 2-10 installing and routing, 2-13 measuring and cutting, 2-10 Suspension harness accessories list, E-4 System IDs, 4-19

T
Telephone commands, 7-9 Text entry numbers, 3-3 program names, 3-7 site descriptions, 3-7 text messaging, 5-40 Time pacing, 4-5, 5-12 Times and dates entering times and dates, 3-5 Time-switched distribution, 5-16 pause and resume, 5-27 Trigger pacing, 4-5 Tube coupling attaching suction line, 2-10 Two-part programs, 5-1 storm-water monitoring, 5-2

V
Variable volume sampling, 5-19 Variable-duration pulse, 5-31 event marks, 5-31 Voice modem, 7-9

W
warning messages, 8-15

Y
YSI 600 about the sonde, 1-2 displayed readings, 4-11

Index-4

Compliance Statements

Name and amount of Hazardous Substances or Elements in the product Hazardous Substances or Elements Component Name (Pb) Circuit Boards Display Wiring Internal Cables Line Cord DC Motor Keypad Connectors X X O O O X O O (Hg) O O O O O O O O (Cd) O O O O O O O X (Cr(VI)) O O O O O O O O (PBB) O O X X X X X O (PBDE) O O O O O O O O

Name and amount of Hazardous Substances or Elements in the product O: ST/ O: Represent the concentration of the hazardous substance in this components any homogeneous pieces is lower than the ST/ standard limitation. X ST/ ( X ) X: Represent the concentration of the hazardous substance in this components at least one homogeneous piece is higher than the ST/ standard limitation. (Manufacturer may give technical reasons to the Xmarks) The Environmentally Friendly Use Period (EFUP) was determined through experience. (207 2007 ) A B The date of Manufacture is in code within the serial number. The first three numbers are the year of manufacture (207 is year 2007) followed by a letter for the month. "A" is January, "B" is February and so on.

Hazmat Table 6712/3700

60-9003-655Rev.

DECLARATION OF CONFORMITY
Application of Council Directive: Manufacturer's Name: Manufacturer's Address: Equipment Type/Environment: Trade Name/Model No: Year of Issue: Standards to which Conformity is Declared: 89/336/EEC The EMC Directive 73/23/EEC The Low Voltage Directive Teledyne Isco, Inc. 4700 Superior, Lincoln, Nebraska 68504 USA Mailing Address: P.O. Box 82531, Lincoln, NE 68501 Laboratory Equipment for Light Industrial/Commercial Environments 6712 Sampler 2001 EN 61326-1998 EMC Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use EN 61010-1 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use Severity Applied

Standard EN61000-4-2 EN61000-4-3 EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-11 CISPR11/ EN 55011 EN61000-3-2, 3-3

Description Electrostatic Discharge Radiated RF Immunity Electrical Fast Transient Surge on AC Lines Conducted RF on AC lines

fo rm ity

Performance Criteria B B A B B B B

Voltage Dips/Short Interruptions RF Emissions Harmonic, Flicker

C E

We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and Standards as of March 6, 2001.

William Foster USA Representative

Williaml Foster Director of Engineering Teledyne Isco, Inc. 4700 Superior Street Lincoln, Nebraska 68504 Phone: (402) 464-0231 Fax: (402) 464-4543
60-9002-068 Rev. A

D ec la ra t

Level 2 - 2kV on ac lines

2kV common mode, 1kV differential mode 150 kHz to 80 MHz, 3V rms, 80% modulated

0.5 cycle, each polarity/100% Group 1, Class A Industrial, Scientific, and Medical Equipment

io n

80 MHz to 1000MHz 80% AM at 1kHz Level 1 10V/m

of C

Level 2 - 4kV contact discharge Level 3 - 8kV air discharge

on

DECLARATION OF CONFORMITY
Application of Council Directive: Manufacturer's Name: Manufacturer's Address: Equipment Type/Environment: Trade Name/Model No: Year of Issue: Standards to which Conformity is Declared: 89/336/EEC The EMC Directive 73/23/EEC The Low Voltage Directive Teledyne Isco, Inc. 4700 Superior, Lincoln, Nebraska 68504 USA Mailing Address: P.O. Box 82531, Lincoln, NE 68501 Laboratory Equipment for Light Industrial/Commercial Environments 4200T Modem 2001 EN 55024-1998 EMC Requirements for Information Technology Equipment EN 60950 Safety Requirements for Information Technology Equipment FCC Part 68

EN61000-4-2 EN61000-4-3 EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-11 EN 55022 EN61000-3-2, 3-3

Electrostatic Discharge Radiated RF Immunity Electrical Fast Transient Surge on AC Lines Conducted RF on AC lines

Level 2 - 4kV contact discharge Level 3 - 8kV air discharge 80 MHz to 1000MHz 80% AM at 1kHz Level 1 10V/m

of C on fo r

Standard

Description

Severity Applied

ity

Performance Criteria B B A B B B B

Voltage Dips/Short Interruptions RF Emissions Harmonic, Flicker

C E

We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and Standards as of July 5, 2001.

William Foster USA Representative

William Foster Director of Engineering Teledyne Isco, Inc. 4700 Superior Street Lincoln, Nebraska 68504 Phone: (402) 464-0231 Fax: (402) 464-4543
60-3212-049 Rev. A

D ec la ra ti o n

Level 2 - 2kV on ac lines

2kV common mode, 1kV differential mode 150 kHz to 80 MHz, 3V rms, 80% modulated

0.5 cycle, each polarity/100% Group 1, Class A, Information Technology Equipment

Warranty

Teledyne Isco One Year Limited Factory Service Warranty *


Teledyne Isco warrants covered products against failure due to faulty parts or workmanship for a period of one year (365 days) from their shipping date, or from the date of installation by an authorized Teledyne Isco Service Engineer, as may be appropriate. During the warranty period, repairs, replacements, and labor shall be provided at no charge. Teledyne Iscos liability is strictly limited to repair and/or replacement, at Teledyne Iscos sole discretion. Failure of expendable items (e.g., charts, ribbon, tubing, lamps, glassware, seals, filters, fittings, and wetted parts of valves), or from normal wear, accident, misuse, corrosion, or lack of proper maintenance, is not covered. Teledyne Isco assumes no liability for any consequential damages. This warranty does not cover loss, damage, or defects resulting from transportation between the customers facility and the repair facility. Teledyne Isco specifically disclaims any warranty of merchantability or fitness for a particular purpose. This warranty applies only to products sold under the Teledyne Isco trademark and is made in lieu of any other warranty, written or expressed. No items may be returned for warranty service without a return authorization number issued from Teledyne Isco. The warrantor is Teledyne Isco, Inc. 4700 Superior, Lincoln, NE 68504, U.S.A.

* This warranty applies to the USA and countries where Teledyne Isco Inc. does not have an authorized dealer. Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their Teledyne Isco dealer for warranty service.

In the event of instrument problems, always contact the Teledyne Isco Service Department, as problems can often be diagnosed and corrected without requiring an on-site visit. In the U.S.A., contact Teledyne Isco Service at the numbers listed below. International customers should contact their local Teledyne Isco agent or Teledyne Isco International Customer Service.

Return Authorization A return authorization number must be issued prior to shipping. Following authorization, Teledyne Isco will pay for surface transportation (excluding packing/crating) both ways for 30 days from the beginning of the warranty period. After 30 days, expense for warranty shipments will be the responsibility of the customer. Shipping Address: Teledyne Isco, Inc. - Attention Repair Service 4700 Superior Street Lincoln NE 68504 USA Teledyne Isco, Inc. PO Box 82531 Lincoln NE 68501 USA Repair service: (800)775-2965 (lab instruments) (800)228-4373 (samplers & flow meters) Sales & General Information (800)228-4373 (USA & Canada) (402) 465-3001 [email protected] Web site: www.isco.com

Mailing address:

Phone:

Fax: Email:
February 1, 2006 P/N 60-1002-040 Rev C

You might also like