CP 150 12-Manual Eletrocardiografo
CP 150 12-Manual Eletrocardiografo
CP 150 12-Manual Eletrocardiografo
electrocardiograph
Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed
physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result
from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use
published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc. CP 150, and CardioPerfect are trademarks of Welch Allyn,
Inc.
Welch Allyn
4341 State Street Road
Skaneateles Falls, NY 13153-0220 U.S.A
www.welchallyn.com
iii
Contents
Powering the electrocardiograph .......................................................... 1
Troubleshooting .................................................................................... 15
Lead-quality problems ........................................................................................ 15
System problems ............................................................................................... 17
iv Contents CP 150 12-lead resting electrocardiograph
1
Press .
2 Powering the electrocardiograph CP 150 12-lead resting electrocardiograph
3
Electrode examples, left to right: arm clamp (reusable), Welsh cup (reusable), tab
electrode (disposable), monitoring electrode (disposable).
• For reusable electrodes: Use electrode paste, gel, or cream to cover an area
the size of each electrode but no larger. Secure the arm and leg clamps. Apply
the Welsh cups (suction electrodes) to the chest.
• For disposable tab electrodes: Place the electrode tab between the “jaws” of
the connector. Keep the tab flat. Verify that the metal portion of the connector
makes contact with the skin side of the electrode tab.
• For all disposable electrodes: Lightly tug on the connector to ensure that the
lead is securely attached. If the electrode comes off, replace it with a new
electrode. If the connector comes off, reconnect it.
7
Item Area
1 Device status
2 Content
3 Navigation
Content area
The Content area includes 3 test selection buttons and a preview selection button:
8 ECG home screen CP 150 12-lead resting electrocardiograph
• Auto ECG
• Rhythm ECG
• Stat ECG
• Electrode Placement
Navigation area
The Navigation area includes the following tabs:
• ECG home: Displays ECG test types and provides shortcuts to several controls.
• Saved tests: Accesses the patient ECG tests.
• Settings: Accesses device configuration settings.
To navigate to a tab, touch the tab in the Navigation area with the corresponding name.
The active tab is highlighted.
9
Rhythm ECG A continuous, real-time printout of rhythm strips with a user-defined lead configuration.
Rhythm ECGs are printouts only. They cannot be saved.
Stat ECG An auto ECG that starts without waiting for you to enter patient data or adjust the
waveforms.
10 About the test types CP 150 12-lead resting electrocardiograph
11
Note The ECG configuration settings can be changed in the Settings tab. The
following settings may appear differently if the default settings have been
modified.
Any flashing dots on the lead-status screen indicate unattached or poorly attached
leads.
When all leads have been attached for several seconds, the ECG preview screen
appears.
Item Button
1 Leads button
3 Speed button
4 Filters button
Quick reference guide Performing an Auto ECG test 13
Troubleshooting
Lead-quality problems
“Artifact” message on the screen
Artifact is signal distortion that makes it difficult to accurately discern the waveform
morphology.
Causes
• The patient was moving.
• The patient was shivering.
• There is electrical interference.
Actions
Wandering baseline
Wandering baseline is an upward and downward fluctuation of the waveforms.
Causes
• Electrodes are dirty, corroded, loose, or positioned on bony areas.
• The electrode gel is insufficient or dried.
• The patient has oily skin or used body lotions.
• Rising and falling of chest during rapid or apprehensive breathing.
Actions
• Clean the patient’s skin with alcohol or acetone.
• Reposition or replace the electrodes.
• Verify that the patient is comfortable, warm, and relaxed.
• If wandering baseline persists, turn the baseline filter on.
16 Troubleshooting CP 150 12-lead resting electrocardiograph
Muscle tremor
Causes
• The patient is uncomfortable, tense, nervous.
• The patient is cold and shivering.
• The exam bed is too narrow or short to comfortably support arms and legs.
• The arm or leg electrode straps are too tight.
Actions
• Verify that the patient is comfortable, warm, and relaxed.
• Check all electrode contacts.
• If interference persists, turn the muscle-tremor filter on. If interference still persists,
the problem is probably electrical in nature. See the suggestions for reducing AC
interference (in a related troubleshooting tip).
AC interference
AC interference superimposes even-peaked, regular voltage on the waveforms.
Causes
• The patient or technician was touching an electrode during recording.
• The patient was touching a metal part of an exam table or bed.
• A lead wire, patient cable, or power cord are broken.
• Electrical devices in the immediate area, or lighting, or wiring concealed in walls or
floors are interfering.
• An electrical outlet is improperly grounded.
• The AC filter is turned off or set incorrectly.
Actions
• Verify that the patient is not touching any metal.
• Verify that the AC power cable is not touching the patient cable.
• Verify that the proper AC filter is selected.
• If interference persists, unplug the electrocardiograph from AC power and run it on
the battery. If this solves the problem, you’ll know that the noise was introduced
through the power line.
• If interference still persists, the noise may be caused by other equipment in the
room or by poorly grounded power lines. Try moving to another room.
Quick reference guide Troubleshooting 17
A dot might be flashing on the lead-status screen. Or one or more leads might appear as
a square wave.
Causes
• Electrode contact might be poor.
• A lead might be loose.
• A lead might be defective.
Actions
• Replace the electrode.
• Verify that the patient’s skin has been properly prepared.
• Verify that electrodes have been properly stored and handled.
• Replace the patient cable.
System problems
The electrocardiograph won't turn on when it is plugged in
Causes
• The AC power connection is faulty.
• An AC fuse is blown.
• There is no AC power.
Actions
• Check the AC power source.
• Check the AC fuses.