Cobas e 411 Reference Data 4.15
Cobas e 411 Reference Data 4.15
Version: 2.0
page 1 of 12
r cobas e 411
Refer.DAT installation procedure
Content:
1 Prerequisites................................................................................................................. 4
2 Download of Reference Data .......................................................................................... 4
3 Preparation of Reference Data CD or DVD or USB flash drive .......................................... 5
3.1 USB flash drive............................................................................................................... 5
3.2 CD/DVD......................................................................................................................... 7
4 Reference file update ..................................................................................................... 9
5 Verification Steps ......................................................................................................... 11
page 3 of 12
r cobas e 411
Refer.DAT installation procedure
1 Prerequisites
page 4 of 12
r cobas e 411
Refer.DAT installation procedure
● Ensure that a USB flash drive is inserted in your Roche computer. Format the
USB flash drive before use to make sure it’s free of malicious software.
● Open the folder Refer_update_4.xx.zip on your computer and double click
Refer update 4.xx.iso.
page 5 of 12
r cobas e 411
Refer.DAT installation procedure
● Select and right click on the cobas e 411 folder that opens, then select Send
to your USB flash drive location. (If you do not see the Send to option in the
menu, ensure that the e 411 folder has the check box selected, then right click
on the folder)
● Check to ensure the files in the image below were copied to your USB flash
drive.
page 6 of 12
r cobas e 411
Refer.DAT installation procedure
3.2 CD/DVD
page 7 of 12
r cobas e 411
Refer.DAT installation procedure
● Select and right click on the cobas e 411 folder, then Select Send to and your
DVD drive location. (If you do not see the Send to option in the menu, ensure
that the e 411 folder has the check box selected, then right click on the folder)
● Check to ensure the files in the image below were copied to your CD/DVD.
page 8 of 12
r cobas e 411
Refer.DAT installation procedure
The Reference files are specific for each laboratory. For this reason, the unit settings of the
fourth unit must be checked before the update is performed.
Hint: Please note the settings in Alarm Setting screen and set then again after reading the
reference data.
page 9 of 12
r cobas e 411
Refer.DAT installation procedure
Insert the reference update CD/DVD into the DVD-RAM drive or the USB flash drive with the
reference files in the USB Port named “Service”.
page 10 of 12
r cobas e 411
Refer.DAT installation procedure
Remove the CD/DVD from the DVD-RAM drive or the USB flash drive from the USB port.
Please set the 3rd Unit Setting and the Alarm Setting like the first steps of this chapter.
5 Verification Steps
● Select the Help icon on the lower left corner of the cobas e 411 screen.
● Select System Information. You may need to expand the “Publication Infor-
mation” folder by clicking on the + sign, to see the System Information.
● Verify the correct version of Reference Data has been installed.
page 11 of 12
r cobas e 411
Refer.DAT installation procedure
page 12 of 12