Measurements & Metrology Basics
Measurements & Metrology Basics
Measurements & Metrology Basics
AND
METROLOGY
(N–SCHEME)
Type setting :
Students’ Media Computer Graphics
Vellore – 632 011.
Preface
N. IYANARAPPAN
([email protected])
1.2 Units
The standards used for measuring each physical quantity is
known as units. The International System of Units (S.l) has divided the
units into three categories.
a) Fundamental units or base units.
b) Supplementary units.
c) Derived units.
3) Second
It is the fundamental unit of time. Second is equal to the
duration of 9192631770 periods of the radiation corresponding to the
transition between the hyper fine levels of the ground state of the
Caesium –133 atom.
4) Kelvin
It is the fundamental unit of temperature. Kelvin is defined as
the fraction 1/273.16 of the thermodynamic temperature of the triple
point of water.
5) Ampere
It is the fundamental unit of electrical current. Ampere is that
constant current which, if maintained in two straight parallel
conductors of infinite length of negligible circular cross section and
placed one metre apart in vacuum would produce a force equal to
2 x 10-7 Newton per unit length between these conductors.
6) Candela
It is the fundamental unit of luminous intensity. Candela is the
luminous intensity in the perpendicular direction of a surface of
1/6,00,000 m2 of a black body at the temperature of freezing platinum
under a pressure of 101.325 kN/m2.
1.2
7) Mole
It is the fundamental unit for amount of substance. Mole is the
amount of substance of a system which contains as many elementary
entities as there are atoms in 0.012 kg of Carbon-12.
b) Supplementary units
1) Radian (rad ) : It is the unit of plane angle. One radian is defined
as the plane angle subtended as the centre of an arc of unit
length at unit radius.
2) Steradian (sr ) : It is the unit of solid angle. One steradian is
defined as the solid angle subtended at the centre by unit area
of a spherical surface at unit radius.
c) Derived units
The derived units are expressed in terms of the fundamental
and supplementary units by defining equations.
1.4
4) To provide input for automatic control systems.
5) To test the functions of elements in a system.
6) To evaluate the performance of a system.
7) To study some basic laws of nature.
8) To ensure interchangeability for promoting mass production.
9) To evaluate the response of the system to a particular point.
10) To check the limitations of theory in practical situations.
11) To establish the validity of design.
1.5
5) Transposition method : This method involves making the
measurement by direct comparison, in which the quantity to be
measured (V ) is initially balanced by a known value ( X ) of the
same quantity. Then X is replaced by the quantity to be
measured and balanced again by another known value (Y ) .
Then, quantity to be measured, V= XY . Example :
Determination of mass by balancing methods and known weights.
6) Coincidence method : This is a differential method of
measurement in which a very minute difference between the
quantity to be measured and the reference is determined by careful
observation of the coincidence of certain lines and signals.
Example : Measurements on Vernier calliper and micrometer.
7) Deflection method : This method involves the indication of the
value of the quantity to be measured directly by deflection of a
pointer on a calibrated scale. Example : Pressure measurement
8) Complementary method : The value of the quantity to be
measured is combined with a known value of the same quantity.
The combination is so adjusted that the sum of these two values
is equal to the predetermined comparison value.
Example : Determination of the volume of a solid by liquid
displacement.
9) Null measurement method : In this method, the difference
between the value of the quantity to be measured and the known
value of the same quantity with which comparison is to be made
is brought to zero.
10) Substitution method : It is a direct comparison method. This
method involves the replacement of the value of the quantity to be
measured with a known value of the same quantity, so that the
effects produced in the indicating device by these two values are
the same. Example : The Borda method of determining mass.
11) Contact method : In this method, the surface to be measured
is touched by the sensor or measuring tip of the instrument.
Example : Measurements using micrometer, Vernier calliper,
and dial indicator.
1.6
12) Contactless method : In this method, there is no direct contact
with the surface to be measured. Example : Optical instruments,
tool maker’s microscope, and profile projector.
1.7
3) Geometric or profile measuring instruments
Profile measurement systems measure and record the profile
of a target by tracing the surface of the target using a stylus. These
instruments are ideal for measuring minute shapes such as screw
threads and thin films in the order of micrometers. In recent years,
profilometer models have been developed that use a laser instead of
a stylus to measure complex shapes.
1.8
1.10 Standardisation
Standardisation is the process of creating, issuing and
implementing standards. Standardisation leads to interchangeability.
It improves economy, efficiency and productivity in a factory and
country. Standardisation is done at various levels.
3) Range
Range is defined as the upper and lower limits an instrument
can measure a value without exceeding the maximum permissible
error.
4) Magnification
In order to measure small difference in dimensions, the
movement of the measuring tip in contact with the work piece must be
magnified. The output signal from a measuring instrument is to be
magnified many times to make it more readable. Magnification may be
achieved by mechanical, electrical, optical, pneumatic principle or a
combination of these.
5) Repeatability
Repeatability is the quality of a measuring instrument to give
the same result when measuring a component using the same method
in the same conditions by the same operator.
6) Resolution
Resolution is the smallest value that an instrument can
measure.
7) Sensitivity
Sensitivity of the instrument denotes the smallest change in
input value to which the instrument responds. It also denotes the
maximum change in an input signal that will not initiate a response on
the output.
8) Hysteresis
It is the difference between the indications of a measuring
instrument when the same value of the measured quantity is reached
by increasing or by decreasing that quantity.
1.11
9) Response time
It is the time taken by an instrument to give an indication after
a sudden change in the measured quantity.
10) Calibration
The process of periodic checking against measuring
instruments and standards of high accuracy is called as calibration.
The following are the requirement of a good calibration system.
1) The measuring equipment in measuring system should be
capable of desired accuracy and precision.
2) Calibration should be carried out using measurement systems
having adequate accuracy, stability and range.
3) All the test and measuring equipment should be securely and
durably labeled.
4) Test and measuring equipment should be calibrated at periodic
intervals with desired accuracy and quality.
5) Records should be maintained for all the test and measuring
equipment included in the calibration system.
Precision
Precision is the closeness of agreement between
independent test results obtained under specified conditions.
Precision refers to a group of measurements rather than a single
measurement. If an instrument is not precise, it will give different
results for same dimension when measured again and again.
1.12
Average Value
Dimension
Dimension
Error Average Value
True Value Error True Value
Frequency Frequency
(a) Precise but not accurate (b) Accurate but not precise
Dimension
Frequency
(c) Accurate & Precise
Accuracy
Accuracy is the closeness of agreement between the
measured value and the true value. The difference between the true
value and the measured value is known as error of measurement.
Accuracy is an important quality of a measuring instrument. As the
exact measurement of a true value is difficult, a set of observations
are made and the mean value is taken as the true value.
1.13
Difference between precision and accuracy
Precision Accuracy
1) Precision is the closeness of Accuracy is the closeness of
agreement between independent agreement between the
test results obtained under same measured value and the true
conditions. value.
2) It indicates the degree of It indicates the degree of
repeatability in the measuring trueness in the measuring
process. process.
3) Precision does not take into Accuracy takes into account
account the accepted value. the accepted value.
4) For high precision, the average For high accuracy, the
value need not be closer to the average value must be closer
true value. to the true value.
5) Good repeatability is sufficient for Good repeatability is not
good precision. sufficient for good accuracy.
6) For high precision, the measured For high accuracy, the
values must be close to the measured values may be
average value. scattered from the average
value.
7) Precision speaks about quality. Accuracy does not speak
about the quality.
8) Precision refers to a group of Accuracy may refer to a single
measurements. measurement.
9) It is concerned with random It is concerned with systematic
errors. errors.
1.15
1.15 Precautions for use of instruments to avoid inaccuracy
The following are the precautions for use of instruments so as
to avoid inaccuracy in measurements.
1) The measuring instrument should be inspected for damage before use.
2) The zero error of the instrument should be checked and adjusted.
3) To eliminate parallax error, the observer should position his eyes
directly above the scale reading.
4) The objects should be gripped gently during measurement.
5) The surface of the objects should be clean and dry.
6) The measuring instrument must be cleaned before and after use.
7) The instrument should not be mixed with other bench tools.
8) It should be kept in in the cases provided, when not in use.
9) Bright surfaces should be protected by a film of oil.
10) Regular preventive maintenance should be done.
11) Recalibration should be done periodically.
12) The manufacturer’s recommendations for maintaining the
instrument in good condition should be followed.
1.16 Reliability
Reliability is the degree of consistency of a measurement. A
test will be reliable when it gives the same repeated result under the
same conditions. For example, a micrometer is a reliable instrument
when it measures the correct dimension each time it is used.
1.17 Error
Error is the difference between the measured value and the
true value. The errors can be expressed either as an absolute error or
a relative error.
1) Absolute error
True absolute error : It is the algebraic difference between the
results of measurement to the true value of the quantity measured.
Apparent absolute error : It is the algebraic difference between
one of the results of measurement to the arithmetic mean of a
series of measurement.
1.16
2) Relative error
It is the ratio of the absolute error and the value of comparison
used for calculation of that absolute error. This value of comparison
may be the true value or the arithmetic mean for series of
measurement.
2) Loading errors
3) Dynamic error
(a) Systematic error
(b) Random error
1.17
1) Static errors
Static errors result from the imperfections or limitations in the
hardware and apparatus of measuring device. The basic sources of
static errors include :
(a) Reading error : Reading error occurs only in readout
device. The factors involved in reading error are parallax, interpolation
and optical resolution.
Parallax error can be eliminated by using a mirror behind the
readout pointer.
Interpolation error can be minimised by using a magnifier over
the scale.
Most of the reading errors can be eliminated by using digital
readout.
2) Loading error
Instrument loading error is the difference between the value
of measurand before and after the measurement. The deformation of
soft component under contact pressure of measuring instrument is an
example of loading error. Measuring system should be selected such
that its sensing element will minimise instrument loading error.
3) Dynamic error
Dynamic error is caused by time variations in the measurand.
It results from the inability of the system to respond faithfully to a time
1.18
varying measurement. This is due to inertia, damping or friction in the
sensing, readout or display system. Dynamic errors are classified as :
(a) Systematic or controllable errors
(b) Random errors
1.19
Comparison of systematic errors and random errors
Review Questions
1) Define metrology.
2) Define unit. State its classification.
3) List out the base units.
4) Explain the two supplementary units.
5) What are derived units? Give examples.
6) What is measurement?
7) Classify the measuring instruments.
8) Define standardisation.
9) List out the international standardisation organisations.
10) What is legal metrology?
11) What is calibration.
12) Differentiate between precision and accuracy.
13) Define reliability.
14) Define error? Differentiate absolute error and relative error.
15) Classify errors in measurements.
16) What is loading error?
1.21
Long Answer Questions
1) Explain the various base units.
2) Explain the needs for inspection.
3) What are the needs for measurement.
4) Explain the important elements in measurement system.
5) Explain the various methods of measurements.
6) Explain the classifications of measuring instruments.
7) Explain the scope of metrology.
8) Explain the functions of BIS.
9) Explain the functions of legal metrology.
10) Explain the important terminology in metrology.
11) Explain the precision and accuracy.
12) Explain the factors affecting the accuracy of measuring system.
13) List out the general rules for accurate measurements.
14) State the precautions in use of instruments to avoid inaccuracy in
measurements.
15) Explain the various sources of errors.
16) Explain the various errors in measurements.
17) Compare systematic errors and random errors.
18) Explain about the selection of measuring instruments.
1.22
Unit – II
LINEAR AND ANGULAR MEASUREMENTS
2.1 Linear measurements
Linear metrology is defined as the science of linear
measurement, for the determination of the distance between two
points in a straight line. Linear measurement is applicable to all
external and internal measurements such as distance, length,
diameter, thickness, etc. The principle of linear measurement is to
compare the dimensions to be measured and aligned with standard
dimensions marked on the measuring instruments.
1 2 3 4 5 6 7 8 29 30
1 2 3 12
2.4 Calipers
2.2
Outside caliper has two-curved legs bent inwards. It is used to
measure outside dimensions like diameter, thickness, etc. with the
help of steel rule. The two legs are joined stiff at the hinge of the legs.
The odd leg caliper is used for scribing parallel lines about a
straight edge of the work piece. It is also used to find the centre of a
cylindrical work.
Squaring
head
Graduated rule
Fig. 2.3 Combination set
2.3
2.6 Feeler gauges
m
5m
0.0 mm
0 .1
2.4
2.7 Fillet and radius gauge
m m
.05
6.0
5m 10
m
2.5
Pitch screw gauge is a very effective and accurate instrument
used to identify or check the pitch of the threads on different threaded
items. It consists of a case made of metal carrying a large number of
blades or threaded strips. They have teeth of different pitches cut on
their edges. The corresponding pitch value is marked on the surfaces.
Locking screw
Rule
Main scale
0 2 3 5 6
0 25
Sliding jaw
Fixed jaw Work piece
Fig.2.7 Vernier caliper
Length, width, depth, outside diameter and inside diameter of
a component can be measured by using vernier caliper. The accuracy
of vernier caliper is 0.02 mm.
Vernier caliper has a main beam called rule. It is graduated in
millimeters. It has a fixed jaw at one end. There is a vernier head with
a sliding jaw. It has a vernier scale marked on it. The sliding jaw slides
over the rule and can be locked in any position by using locking
2.6
screws. A fine adjustment screw is fitted in the vernier head. The
inside face of fixed and sliding jaw are parallel with each other.
Reading
The length of one main scale division is 0.5 mm. The length
of 25 divisions in the vernier scale equal to 12 mm. So length of one
vernier scale division = 12 / 25 = 0.48 mm.
Least count = Length of one main scale division –
Length of one vernier scale division
= 0.50 – 0.48 = 0.02 mm
In some vernier calipers, the length of one main scale division
will be 1 mm. The length of 50 divisions in vernier scale is 49 mm. So
length of one vernier scale division = 49 / 50 = 0.98 mm.
0 5 10 15 20 25
2.7
For example, as shown in the figure,
Main scale reading = 28 mm
Vernier scale coincidence = 16 divisions
Vernier scale reading = 16 x 0.02 = 0.32 mm
Correct reading = 28 + 0.32 = 28.32 mm.
Thump roller
Battery lid
ON switch
OFF switch
Movable jaw
Fixed jaw
Fig.2.9 Digital vernier caliper
Clamp 7
6 Clamping screw
25
Scriber
4
3
0
Vernier scale
Slider
1
Measuring jaw
0
Base
2.11 Micrometer
The micrometer is used to measure length, width, thickness and
diameter of small and medium size components. The outside micrometer
used for measuring external dimensions is shown in the figure.
2.9
It has a ‘C’ shaped frame made of steel. An anvil is fitted at
the left end and a barrel is fitted at the right end of frame. The main
scale is graduated on the barrel. The barrel has a bore with threads of
0.5 mm pitch. A screw of 0.5 mm pitch passes through this bore. The
spindle is fitted in the left end and the thimble is fitted at the right end
of the screw. Thimble is tubular cover. When thimble is rotated, spindle
moves as the screw rotates. Vernier scale is graduated in the beveled
edge at the left end of the thimble. Ratchet is fitted at the right end of
thimble. The ratchet will slip when the pressure on the screw exceeds
a certain limits.
Spindle clamp
Thimble
Spindle Main scale
Anvil 0 5 10 15 15
10
5
Barrel Ratchet
Vernier scale
Frame
Fig.2.11 Micrometer
Reading
30
0 5
25
20
Work piece
Handle
Anvil
Barrel
Extension
Thimble rods
Spindle
2.11
The spindle can be moved to the maximum of 13 mm. So this
micrometer is directly used for measuring the dimensions from 50 mm
to 63 mm. For measuring larger dimensions, the anvil is removed and
suitable extension rod is fitted. The extension rods of 25, 50, 100, 150,
200 and 600 mm are available. The accuracy of this micrometer is
0.01 mm.
Thimble
Barrel
Locking ring
Head
Work piece
Spindle
Pitch diameter
V-shaped Conical
anvil point
Thread
Thread
Wire
Distance between
wires (D)
2.13
This method is used for measuring pitch circle diameter
accurately. In this method, an ordinary micrometer and three wires of
same diameter are used. These wires are made of hardened steel and
finished by lapping. The correct size of wire is selected according to
the size of thread to be measured.
As shown in the figure, two wires are placed in one side of
thread and one wire is placed at the other side of thread. The
micrometer anvil and spindle are adjusted to touch the wires on the
both sides. Now the reading in micrometer (M) is noted.
Then the pitch diameter of thread is calculated by using the
following formula.
= − + +
Where, = Micrometer reading over the wires
= Diameter of wirre
= Included angle of thread
= Pitch of the thread
2.14
Fig.2.17 Optical micrometer
2.15
Fig.2.18 Light wave micrometer
1.002
1.003
1.004
1.005
1.006
1.007
1.008
1.009
1.01
1.02
1.03
1.04
1.05
1.06
1.07
1.08
1.09
1
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
10
2
9
20 30 40 50 60
70 80 90
10 20 40
10 20 40
2.17
Indian standard on slip gauges (IS:2984-1966)
Slip gauges are graded according to their accuracy as Grade 0,
Grade I & Grade II.
2.18
Care of slip gauges
The following care should be taken for slip gauges to maintain
their accuracy for a long period.
1) The gauges should not be left in the wrung condition when not
in use.
2) Before the use of slip gauges, they must be properly cleaned
to remove dust and grease traces.
3) The slip gauges should be kept in their case when not in use.
4) Do not break the wring after use. Slide one gauge over the
other to separate them.
5) Slip gauges should be used in an atmosphere free from dust.
6) The blocks are not to be used again immediately after
separating them.
7) A thin layer of good quality grease should be applied on their
faces before they are kept in their case.
2.19
ANGULAR MEASUREMENTS
2.18 Vernier bevel protractor
Vernier bevel protractor is also known as plain protractor. It is
a simple device mainly used for measuring and laying out angles. It
mainly consists of a head and blade. The blade is about 150mm long.
The head is graduated from 0 to 180o in both directions for easy
reading. It can provide a least count of 1o for smaller protractors and
½ ° for large ones.
The main scale on the dial is divided into four quadrants, each
measuring 90o . Each division on this scale reads 1o . The degrees are
numbered from 0 to 90 on either side of the zeroth division.
2.20
A vernier scale is fitted to the disc. A movable blade is fitted in
the groove of circular dial. The vernier scale has 24 divisions, which
correspond to 46 divisions on the main scale. The divisions on the
vernier scale are numbered from 0 to 60 on either side of the zeroth
division. . An acute angle attachment is provided for the measurement
of acute angles.
Reading
Main scale
10 20
20 30
40
30
50
40
0 15 0 15 30 45
45 3 60
60
Vernier scale
Fig.2.21 Reading of bevel protractor
2.21
Actual reading = Main scale reading +
Vernier scale reading
= 15 + 20 Minutes = 15O 20’
2.22
sine bar. The size of the sine bar is specified by the distance between
the centres of these two rollers. The size varies from 100 mm to 300
mm. The line drawn between the centres of two rollers will be parallel
with the top and bottom surface of the sine bar.
1) Sine centre
Sine centers are used for mounting conical work pieces which
cannot be held on a simple sine bar. Sine center consists of a self-
contained sine bar hinged at one roller and mounted on its own datum
surface. The top surface of the bar is provided with clamps and centers
to hold the conical work pieces. It cannot measures the angle more
than 45 degrees
2.23
Uses of sine bar
1) Measuring unknown angle
Slip gauges
l
e bar
Sin
2.24
3) Checking of unknown angles of heavy component
When components are heavy and difficult to place on the sine
bar, then the sine bar is placed on the component. The height of both
rollers from the horizontal reference surface are measured by a vernier
height gauge. A dial test gauge is used along with the vernier height
gauge for accurate measurement.
Let, ℎ+ = Height of the lower roller, and
ℎ, = Height of the upper roller
Then the angle of inclination of the work piece can be
calculated with the formula,
−
" #$=
%
Limitations of sine bar (Disadvantages of sine bar)
The following are the limitations of sine bar
1) The accuracy of sine bars is limited by measurement of
center distance between the two precision rollers.
2) It cannot be used as a primary standard for angle
measurements.
3) Sine principle is fairly reliable at angles less than 15o, but
becomes inaccurate as the angle increases.
4) Sine bar becomes impractical and inaccurate as the angle
exceeds 45o.
2.22 Clinometer
A clinometer is a special case of application of spirit level for
measuring the inclinations in the vertical plane in relation to the
horizontal plane. The main functional element of a clinometer is the
sensitive vial mounted on a rotatable disc. Vial is a closed glass tube
of accurate size in a spirit level, which is used for storing the liquid. It
is graduated in linear scale and the bubble moves inside it. Vial carries
a graduated ring with its horizontal axis supported in the housing of
the instrument.
2.25
inclined surface, the bubble deviates from the centre. It can be brought
to the centre by rotating the disc. The rotation of the disc can be read
on the scale. It represents the deviation of the surface over which the
clinometer is placed from the horizontal plane.
5 55
0
Minute
Rotatable disc adjustment
screw
Applications
The following two categories of measurement are possible
with clinometer :
a) Measurement of an inclined plane with respect to a horizontal
plane : This is done by placing the instrument on the surface to be
measured and rotating graduated disc to produce zero inclination
on the bubble. The scale value of the disc position will be equal to
the angle of incline.
2.26
surface in turn, and taking the readings with respect to the
horizontal. The difference of both the readings will indicate the
angular value of the relative incline.
2.23 Autocollimator
Principle of auto-collimation
90
°
O
Optical axis
Object and
image Collimating lens Plane
reflector
(a) θ°
Image
O'
2θ°
x
Tilted
O reflector
Object
(b)
2.27
Working of autocollimator
Light source
Target
graticule
Micrometer
Reading on eyepiece
Reflected image
Target wire 0 10 20 of target wire coincide
with setting lines
Scale
Fig.2.28 Reading on eyepiece
2.28
The eyepiece contains an adjustable graticule engraved with
setting lines and a scale. The scale is divided into minutes and half
minutes. The setting lines are adjusted by the micrometer control until
they coincide with the reflected image of the target wires. The scale
is then read to the nearest minute plus the micrometer reading.
Generally the divisions of the micrometer scale represent a
0.5 seconds. Autocollimators can accurately read up to 0.1 seconds
and may be used for distance up to 30 meters.
Prism
Microscope
eyepiece
Light source
Graticule
(glasss screen)
TRansmiitted
image
Reflected
image
Collimating lens
Workpiece
2.29
angles to each other. The image does not fall across a simple datum
line, but across a similar fixed scale at a right angle to the image. Thus
the reading on the reflected scale measures angular deviation in one
plane at 90° to the optical axis.
The reading on the fixed scale gives the deviation in a plane
which is perpendicular to the former plane. This feature enables to
measure angular errors in two planes at the same time. The whole optical
system is enclosed in a tube which is mounted on an adjustable bracket.
Procedure to measure inclination angle
glass screen Eye piece
Reflected image Fixed scale
40 of illuminated scale
40
30 30 30
20
0 10 20 30 40 0 10 30 40 0 10 20 30 40
10 10 10
0 0
2.30
2.25 Comparators
Comparators are devices used to compare the size of the
component with a standard size. The actual size of the component
cannot be measured directly by using comparators. The deviation in
the size of component only can be measured.
Uses / Applications of comparator
1) Comparators are used as laboratory standards
2) They are used as working standards to maintain required
tolerance at all stages of manufacture.
3) Comparators may be used as final inspection gauges where
selective assembly is requires.
4) Comparators are used for checking parts received from
outside sources.
5) Comparators are used for checking newly purchased gauges.
Requirements of a good comparator
1) The scale used in the instrument should be linear and have a
wide range of acceptability for measurement
2) There should not be any backlash and lag between the
movement of the parts.
3) The instrument must be precise and accurate.
4) The indication method should be clear.
5) The indicator should be constant in its return to zero.
6) The design and construction of the comparator should be
strong.
7) The measuring pressure should be uniform for all similar
measuring cycles.
8) The comparator must possess maximum compensation for
temperature effects.
Classification of comparators
All comparators have a magnifying device. According to the
mechanism used for magnification, the comparators are classified as
follows.
1) Mechanical comparator 2) Optical comparator
3) Electrical comparator 4) Pneumatic comparator
2.31
2.26 Mechanical comparator
Rack Gear
Set of
gears
Pointer
2.32
Advantages of mechanical comparator
1) It is cheaper in comparison to the other comparators.
2) It has linear scale, which is easily readable.
3) It is compact and robust in construction.
4) It is easily handled.
5) It does not require any external supply such as electricity, air, etc.
6) It is suitable for ordinary workshops.
First, the slip gauge of required size are built up and placed
between the plunger and the table. Now the point where the beam of
light falls on the screen is noted as zero reading. Then the slip gauges
are removed and the work piece to be checked is placed under the
plunger. If there is any deviation in the size of work piece, the plunger
will move up and down. This makes the lever to tilt the mirror. Now the
beam of light reflected from the mirror falls on the screen at another
point. From this, the deviation in the size of work piece is determined
accurately.
2.33
Light
Scale
Condensing
lens
Projection
lens
Beam of
light
Hinge Lever Mirror
Plunger
Table
Base
Wheatstone Meter
bridge circuit
Stand Plunger
Table
Base
2.35
Advantage of electrical comparator
1) It has very less number of moving parts.
2) The magnification is comparatively high.
3) It can be used for various ranges.
4) It has very light pointer mechanism and hence not sensitive to
vibrations.
5) It has a compact construction.
Work piece
h
constant pressure
Internal
Manometer measuring head
Work piece
Scale
Air escaping
2.36
This comparator has a water tank fitted with a manometer. A
calibrated scale is fitted vertically at the side of the manometer. A dip
tube is immersed in the water tank. Restriction chamber is provided at
the top of the dip tube. The lower end of the dip tube is opened. The
upper end is connected to a flexible pipe through a control orifice. The
top of the manometer tube is connected to the flexible pipe. A
measuring head is connected at the end of flexible pipe.
When the air flows freely through the measuring head, the
water levels in the manometer and in the tank will be same. If there is
some restriction to the flow of air, there will be a pressure difference
between the flexible pipe and the dip tube due to the back pressure.
This pressure difference is shown by the variation of water level in the
manometer.
a) Internal measurement
For checking the size of hole, internal measuring head is
connected to the flexible pipe. It is in cylindrical shape. Two holes are
provided at its sides through which air can escape. First, the
measuring head is introduced into the hole of standard specimen or in
a ring gauge. The level of water in the manometer is marked as zero.
Then the measuring head is introduced in the hole of work piece to be
checked. When the size of hole is smaller, the restriction for air flow
will be more. So the water level in the manometer will fall down from
the zero reading. When the size of hole is larger, the restriction for air
flow will be less. So the water level in manometer will rise above the
zero reading. The deviation can be found out with an accuracy of
0.001 mm by using this method.
2.37
b) External measurement
External measuring head is connected to the flexible pipe to
check the thickness and height of components. Slip gauge blocks are
built up to required size and kept under the measuring head. The zero
reading in the manometer is noted. Then the slip gauges are removed
and the work piece is placed under the measuring head. The gap
between the work piece and measuring head varies according to the
variations in the size of work piece. This variation is indicated by the
water level in the manometer.
2.38
2.30 Comparison of comparators and measuring instruments
2.39
Review Questions
2.41
Unit – III
FORM MEASUREMENT
MEASUREMENT OF SCREW THREADS
3.1 Screw thread
A screw thread is obtained by cutting a continuous helical
groove on a cylindrical surface. The threaded portion engages with a
corresponding threaded hole forming a screwed fastener. It is also
used to transmit power and motion.
3.1
Fig.3.1 Nomenclature of screw thread
3.2
Dedendum : It is the radial distance between the minor diameter
and pitch line for external threads. It is the radial distance between
the major diameter and pitch line for internal threads.
Single-start thread : In case of a single-start thread, the lead is
equal to the pitch. Therefore, the axial distance moved by the screw
equals the pitch of the thread.
Multiple-start thread : In a multiple-start thread, the lead is an
integral multiple of the pitch. Accordingly, a double start will move
by an amount equal to two pitch lengths for one complete revolution
of the screw.
3.2 Errors in thread
Errors in screw threads may occur during its manufacture or
storage. The possible sources of errors in screw thread are :
1) Errors on major diameter
2) Errors on minor diameter
3) Errors on pitch diameter
4) Pitch errors
5) Angle errors
Errors on the major and minor diameters will cause interference
with the mating thread. Due to this error, the root section and wall
thickness will be reduced. The flank contact will be reduced. The
component will become weak in strength.
Errors on the effective diameter will also result in weakening of the
assembly due to interference between the flanks. If it is small, the
threads will be thin on the external screw and thick on the internal
screw.
Pitch and angle errors cause a progressive tightening and
interference on assembly. These errors precisely related to the
effective diameter.
Pitch errors
Errors in pitch are the errors in the effective diameter helix
measured parallel to the axis of the screw thread. The total pitch error
in overall length of the thread is called cumulative pitch error.
3.3
The types of pitch errors are :
1) Progressive error
2) Periodic error
3) Erratic or irregular error
4) Drunken error
3.4
Periodic errors are cyclic. Pitch increases to a maximum then
reduces to minimum. The graph between the cumulative pitch error
and length of thread for this type of error will be in sinusoidal form.
4) Drunken error
It is type of periodic pitch error occurring for one revolution in
drunken threads. In such threads, the thread is not cut to true helix.
The helix will be a curve and not a straight line.
3.5
In this measurement, the micrometer is used as a comparator.
A setting cylinder of size approximately equal to the major diameter of
the thread is taken. Let, 𝑆 is the diameter of the setting cylinder. First,
the micrometer is adjusted on the setting cylinder and the reading is
taken (𝑅1 ) Then the micrometer is adjusted on the major diameter of
the thread and the reading is taken (𝑅2 ).
Now, the major diameter of thread, 𝐷 = 𝑆 ± (𝑅1 − 𝑅2 )
Where, 𝑆 = Diameter of setting cylinder
𝑅1 = Micrometer reading over the setting cylinder
𝑅2 = Micrometer reading over the thread.
Initially, the Instrument is set for the setting cylinder and the
reading is taken. Then the floating head is retracted to engage the tip
of the stylus at the root of the thread under spring pressure. Now, the
new reading is taken.
Let, 𝑆 = Diameter of setting cylinder
𝑅1 = Reading of dial indicator on setting cylinder
𝑅2 = Dial indicator reading on thread
Then, the major diameter internal thread, 𝐷 = 𝑆 ± (𝑅1 − 𝑅2 )
3.7
The threaded work piece is mounted between the centres of
the machine. The correct V-pieces are place on each side of the work
piece and the reading is taken. Then the work piece is replaced by the
setting cylinder and the reading is taken.
Let, 𝑆 = Diameter of setting standard
𝑅1 = Micrometer reading on setting cylinder
𝑅2 = Micrometer reading on thread.
Then, the minor diameter of external thread, 𝐷 = 𝑆 ± (𝑅1 − 𝑅2 )
3.8
3.3.5 Measurement of effective diameter (pitch diameter) of external thread
a) Using thread micrometer
The thread micrometer is
similar to ordinary micrometer except
that it has special contact points to suit
the screw thread form to be measured.
The contact points are selected
depending upon the type and pitch of
the thread to be measured. In this
micrometer, the end of the spindle is Fig.3.7 Measuring effective
pointed to the V-thread form with a diameter using thread
micrometer
corresponding V-recess in the fixed
anvil.
When measuring threads, the angle of the point and the sides
of V-anvil should make contact with the screw thread. Now the reading
of the micrometer gives the effective diameter.
This value should agree with the valued calculated by using
the following relation.
Effective diameter, 𝐸 = 𝐷 − 0.6403 𝑝
Where, 𝐷 = major diameter, 𝑝 = pitch of the thread.
3.9
First the micrometer reading is noted on a setting standard.
The diameter of the setting standard is nearly equal to the effective
diameter of the thread to be checked. Then the setting standard is
replaced by the thread under test and the reading is taken.
Let, 𝑆 = Diameter of setting standard
𝑅1 = Micrometer reading on setting cylinder
𝑅2 = Micrometer reading on thread.
Then, the effective diameter of thread, 𝐸 = 𝑆 ± (𝑅1 ~ 𝑅2 )
This method is used for measuring effective diameter of
counter pitch threads, and during manufacture of threads. The
difficulty with this method is that the micrometer axis may not remain
exactly at right angles to the thread axis.
3.11
3.3.6 Measurement of effective diameter (pitch diameter) of
internal thread
A thread comparator is used to measure the effective
diameter of internal threads. In this case a pair of ball tips engage the
flanks of the threads measure the effective diameter only.
The ball tip on the right is fixed at the end of a measuring jaw
attached to a floating head in the sliding bracket (B). The floating head
has extension in contact with the spindle of the dial indicator. The
movement of floating head towards the indicator is constrained by a
spring. The fixed head (A) carrying the left hand ball tip is adjusted by
a fine screw to set the gauge to the reference standard.
3.12
Setting up reference standard
The reference standard is built up from slip gauges as shown in
Fig.3.11(b). The two end pieces have V-jaws of an angle equal to V form
of the thread. The dimensions 𝐽1 and 𝐽2 are the depths from the face to
the apex points of the V form. Assuming the effective diameter and pitch
of the thread to be known, the distance S is found from the formula,
𝑺 =𝑬 +𝒚−𝒁
Where, 𝐸 = Mean effective diameter.
𝑦 = Depth of the thread from apex to the apex of the
V-form. The value of 𝑦 depends on the included
angle of the thread.
𝑦 = 09605 𝑝 for 55° thread and 𝑦 = 0.866 𝑝 for 60° thread.
𝒁 = 𝑱𝟏 + 𝑱𝟐
The assembled slips are set in a holder with a slip equal to
half the pitch, under one end piece to compensate for the helix angle:
3.15
Fig.3.14 Thread form for NO GO effective diameter gauge
3.16
3.4.2 Ring screw gauges
Ring screw gauges are used for checking bolts and external
thread forms. Major, minor and effective diameters will be checked. In
ordinary production practice, GO effective diameter gauge and full
form NO GO gauge are mainly used.
3.17
The following three gauges are recommended for checking
parallel external screw threads,
(i) GO and NO GO caliper gauges to check the tolerance on the major
diameter.
(ii) GO screw caliper gauge to check the maximum effective diameter.
(iii) NO GO screw caliper gauge to check the minimum effective diameter.
The first gauge controls the form of the thread, pitch and
effective diameter. It ensures that the full and core diameter are not
too large. The second gauge ensures that full and core diameter are
not too small.
3.19
1) The full form GO front anvils of the first gauge are shown at (a)
2) The rear pair of NOT GO anvils for effective diameter are shown at (b).
3) The NOT GO full diameter front anvils of the second gauge are
shown at (c)
4) The NOT GO core diameter rear anvils are shown at (d).
3.20
8) Setting plug for NO GO screw caliper gauge : NO GO screw
caliper gauge is set by this gauge.
9) Solid or adjustable NO GO screw ring gauge : it is used to
check the minimum pitch diameter limit.
10) Screw check plugs for new sold NO GO screw rings gauge :
These check plugs (GO and NO GO) are used to check the limits
of the pitch diameter of the new solid NO GO screw ring gauge.
11) Wear check plug for solid and adjustable NO GO screw ring
gauge : It is used for checking the maximum wear limit of the
solid NO GO screw ring gauge.
12) Setting plug for adjustable NO GO screw ring gauge : It has
a portion with complete flanks and the screw ring gauge is
adjusted for a snug fit on this portion and further checked by the
NO GO screw check plug.
13) Gauges for major diameter : The plain GO and NO GO caliper
gauges or ring gauges are provided for checking the major
diameter of the work piece threads. For checking GO limit, the
ring gauges are preferred.
The carriage has two slides which move on a finely ground ‘V’
guide way or an antifriction guide way. The lower slide moves in a
direction parallel to the axis of the screw thread. The top slide moves
in a direction perpendicular to the axis of the screw thread. The
micrometer can measure with an accuracy of 0.001 mm.
3.22
MEASUREMENT OF GEARS
3.8 Types of gear
1) Spur gear : In this gear, teeth are cut on the periphery of a
cylindrical disc at equal space. These gears are used to transmit
motion between two parallel shafts.
2) Helical gear : In this gear, teeth are cut at an angle to the axis.
These gears are used to transmit motion between two parallel
shafts. Helical gears run smoothly without noise.
3) Rack and pinion gear : They are used to convert rotary motion
into linear motion or linear motion into rotary motion.
4) Worm and worm wheel : These gears are used to transmit the
motion between the two shafts which are perpendicular to each
other. Speed can be very much reduced by using these gears.
5) Bevel gear : This gear is used to transmit motion between two
shafts which are inclined at an angle to each other.
3.24
Spur gear tooth proportions
Spur gear tooth proportion as per BIS are given below.
3.25
5) Single pitch error : Single pitch error is the error in actual
measured pitch value between adjacent teeth.
6) Accumulated pitch error : It is the difference between
theoretical summation over any number of teeth intervals and
summation of actual pitch measurement over the same interval.
7) Runout errors : This type of error refers to the runout of the pitch
circle. Radial runout is measured along a perpendicular to the
axis of rotation. Axial runout is measured parallel to the axis of
rotation. Runout causes vibrations and noise, and reduces the
life of the gears. This error occurs due to inaccuracies in the
cutting arbour and tooling system.
8) Eccentricity : It is half the radial runout.
9) Undulation : It is a periodical deviation of the actual tooth
surface from the design surface.
10) Lead errors : This type of error is caused by the deviation of the
actual advance of the gear tooth profile from the ideal value. This
error results in poor contact between the mating teeth, resulting
in loss of power.
11) Cyclic error : it is an error occurring during each revolution of
the element.
12) Periodic error : It is an error occurring at regular intervals.
13) Assembly errors : An error in centre distance between the two
engaging gears results in backlash error or jamming of gears.
The axes of the two gears must be parallel to each other, failing
which misalignment will be a major problem.
3.26
3.11 Measurement of runout
Runout is the eccentricity in the pitch circle. Gears that are
eccentric give vibration in each rotation. A gear failure may occur with
a high eccentric tooth.
3.28
3.13 Measurement of profile
The profile is the portion of the tooth flank between the
specified form circle and the outside circle. Profile error is the deviation
of the actual tooth form from the theoretical profile. Involute is mostly
used in gears. The methods commonly used for profile measurements
are :
3.29
Fig.3.23 Gear involute tester
3.30
The driving gear is held solidly against rotation. The backlash
is determined by moving the driven gear back and forth. This
movement is registered by a dial indicator. The backlash variation is
measured by locating the points of maximum and minimum backlash
and obtaining the difference. For precision gears the variation should
not exceed 0.02 to 0.03 mm.
3.31
Gear tooth vernier is used to measure the chordal thickness
of a gear tooth. Chordal thickness is the thickness of a gear tooth at
which the pitch circle passes through it. This vernier has a horizontal
slide and a vertical slide. The slides can be independently moved.
Main scale and vernier scale graduations are marked in both slides.
3.32
In gear tooth vernier, both the chordal thickness and chordal
addendum are dependent upon the number of teeth. Hence, separate
calculations are required for measuring a large number of gears with
different number of teeth. The measurement of tooth thickness at
constant chord simplifies the problem for all number of teeth.
𝜋
Mathematically,
Length of constant chord, 𝑐 = 𝑚 𝑐𝑜𝑠2 𝜙
2
𝜋
Depth from tip of the tooth 𝑑 = 𝑚 (1 − 𝑐𝑜𝑠𝜙 𝑠𝑖𝑛𝜙)
4
𝜋𝑚
Height of AB above pitch line, ℎ = 𝑠𝑖𝑛 2𝜙
,
8
Addendum, 𝑚 = ℎ + 𝑑
Where, 𝜙 = Pressure angle
𝑚 = Module of the gear
Using gear tooth vernier, the vertical slide is moved up to the
calculated value of addendum and locked. The vertical slide is made
to rest on the top of the gear tooth. The horizontal slide is adjusted to
touch the side of gear tooth. Now the reading noted from the horizontal
scale is the value of chordal thickness.
3.33
The instrument consists of a fixed anvil and a movable anvil.
There is a micrometer on the moving anvil side and this has a very
limited movement. The distance between two opposed involutes is
known as base tangent length. It can be calculated by using the
following formula.
𝝅 𝝅𝑺
Base tangent length, 𝑾 = 𝒁𝒎 𝒄𝒐𝒔 𝝓 [𝒕𝒂𝒏 𝝓 − 𝝓 − +
𝟐𝒁 𝒁 ]
Where, 𝑍 =Total number of teeth in gear
𝑆 = Number of teeth under consideration
𝜙 = Pressure angle
𝑚 = Module of the gear
3.35
The two gears are mounted on respective mandrels. The
master gear slide is fixed at a convenient position. The two gears are
brought into mesh by moving the sliding carriage. The sliding carriage
base is locked in its position. However, the sliding carriage is free to
slide for a small distance on steel rollers under a light spring force.
A vernier scale is attached in the machine to adjust the centre
distance of gears accurately. The dial indicator is set to zero. The gear
under test is rotated. Radial variations of the gear are indicated by the
dial indicator. This variation is plotted on a chart for one complete
rotation.
3.37
3.22 Elements of surface texture
3.38
Centre line : The line about which roughness is measured.
Sampling length: It is also known as cut-off length. The
roughness value measured over a specified distance in
microns is called sampling length.
6) Waviness: Surface irregularities of large size is called waviness.
Waviness results from vibrations, work deflections, warping, heat
treatment, etc.
Waviness height: It is the distance from the peak of the wave
to its valley. It is measured in millimeter.
Waviness width: It is the spacing between successive
waves. It is measured in millimeter.
3.39
It is defined as the average height from a mean line of all ordinates
of the surface, regardless of sign. The 𝑅𝑎 value is the commonly used
standard for measuring surface roughness. From the Fig. 3.30,
𝑨𝟏 + 𝑨𝟐 + 𝑨𝟑 + ⋯ + 𝑨𝒏
𝑹𝒂 =
𝑳
b) Root Mean Square (RMS) Value
3.45
Long Answer Questions
1) Explain the various screw thread terminology with a sketch.
2) Explain the various pitch errors in screw thread.
3) Explain any two methods of measurements of major diameter of
external thread.
4) Explain the measurement of minor diamter of external thread using
bench micrometer.
5) Explain any two methods used to measure the effective diameter
of external thread.
6) Explain the procedure for measurement of effective diameter of
internal thread with a sketch.
7) Explain any two method used for measuring pitch of external thread.
8) Explain plug screw gagues.
9) Explain the constructional details of adjustable thread gauge with a
neat sketch.
10) Explain the functions of various types of screw gauges.
11) Explain the construction and working of floating carriage
micrometer with a line sketch.
12) Explain the various gear terminology with a sketch.
13) Explain any one method used to measure pitch of gears.
14) Explain the various methods used for measuring the profile of gears.
15) Explain any one method for measuring chordal thickness of gears.
16) Expalin the procedure to measure chordal thickness of gear using
gear tooth vernier caliper.
17) Expalin the method of measuring tooth thickness of gear using
David brown tangent comparator.
18) Explain the construction and working of Parkinson gear tester /
Rolling gear tester with a neat sketch.
19) Explain the various elements of surface texture.
20) Explain the various methods of evaluation of surface roughness.
21) Explain the various comparison methods for measuring surface finish.
22) Explain the construction and working of tracer type profilogram.
3.46
Unit – IV
ADVANCES IN METROLOGY
LASER METROLOGY
4.1 Basic concept of lasers
LASER stands for Light Amplification by Stimulated Emission
of Radiation. The photon emitted during stimulated emission has the
same energy, phase and frequency as the incident photon. If a photon
comes in contact with another atom in the higher energy level E2, then
it will cause the atom to return to ground state energy level E1 by
releasing another photon. The sequence of triggered identical photon
from stimulated atom is known as stimulated emission. This
multiplication of photon through stimulated emission leads to
coherent, powerful, monochromatic, collimated beam of light
emission. This light emission is called laser.
3) Solid State lasers : Solid state lasers use high density solid
media as active laser materials. They are high power lasers with
peak value in kilowatts to megawatts.
Example : Ruby laser, Nd:YAG laser, etc.
4.1
4) Chemical laser : A chemical laser obtains its energy from
a chemical reaction. They can reach continuous wave output with
power reaching to megawatt levels.
Example : Chemical oxygen iodine laser (COIL), all gas-phase
iodine laser (AGIL), etc.
5) Semiconductor lasers : They are made using semiconductor
materials on nanometer scale. It has operation like LED, but the
output beam has characteristics of laser light.
Example : Double heterojunction laser, Quantum well laser,
distributed feedback laser, tunable laser, surface emitting laser, etc.
Advantages of lasers
1) The laser beam has high intensity and high power in the range
of Mega Watts.
2) The laser beam can be focused in to a parallel beam or into a
very small point.
3) The laser beam has no divergence up to a great distance.
4) It has high information carrying capacity
5) It has very minimum signal leakage.
6) It is free from electro-magnetic interference.
7) Laser beam is visible and can be observed easily.
8) Its centre can be easily judged to an accuracy of 1mm over 2m.
9) This is used for very accurate measurements in the order of 0.1µm.
Working principle
The transmitter module produces a collimated parallel
scanning laser beam moving at a high, constant, linear speed. The
scanning beam appears as a red line. This laser beam is deflected at
constant speed by a rotating mirror. The laser beam is projected over
the object being measured.
4.3
Advantages
1) It measures at the rate of 150 scans per second.
2) It is possible to detect changes in dimensions when components
are moving.
3) Objects in hot conditions can be measured.
4) The system can be controlled with closed feedback loops.
5) The output can be taken in digital form.
6) The microprocessor can be programmed to take care of measuring
atmosphere.
4.4
Applications
1) Laser distance measurement is used in construction industry to
measure the distance to a wall, length of a pipe, width of a wire, etc.
2) It is used in automotive industry to measure front, back, and side
safety distances.
3) It is used in the artistic design of theater, photography and video
making to estimate the appropriate placement of equipment.
Working principle
4.5
Scanning laser gauge uses a shadow measurement principle.
The laser diode produces a high intensity beam of light. This laser
beam is deflected at constant speed by a rotating mirror. The
transmitting lens projects a parallel beam of lights in the measuring
field. The parallel beam of light is known as scanning laser. The laser
beam is intercepted by the object being measured. The shadow of the
object is projected onto the receiver.
The receiving lens concentrates the light beams on a photo
diode. The edges of the shadow is accurately measured by the
detector array in the photo diode. The processor calculates the
dimension by measuring the time duration between the detection of
edges of shadow. The position of the object can be detected by
measuring the time duration (length) of light from transmitting lens to
the object. The display unit shows the measured dimension. The
dimensions and positions of one or more objects can be measured at
the same time by detecting the corresponding shadows.
Applications
Scanning laser gauges are used to measure diameters
of wires, rods and other cylinders. It also measures gaps, edge
positions and various geometric dimensions of objects.
4.7
Fig.4.5 Laser triangulation sensors
Applications
Laser triangulation sensors are used for measurement and
monitoring tasks in factory automation, electronics production,
robotics and vehicle construction.
4.9 INTERFEROMETRY
Light is electromagnetic wave and different rays of light can
interfere with each other. If two waves are exactly in phase, they will
reinforce each other. If they are exactly out of phase, they will cancel
each other. Thus, an alternate dark and bright bands will be observed
during interference of two monochromatic light rays. This
phenomenon is applied for measurements of very small linear
dimensions, and the measurement technique is known as
interferometry.
4.8
Essential conditions for interference
1) The two sources of light should continuously give light rays.
2) The wavelength and intensity of light rays should be equal.
3) The phase difference should be maintained constant.
4) Two sources should be very narrow and close to each other.
4.9
This is the oldest type of interferometer. It consists of an
extended monochromatic light source, a beam splitter and two mirrors.
A white light from the source falls on a beam splitter. Beam splitter is
a plain parallel plate having a semi-transparent layer of silver at its
back. It splits the light into two rays of equal intensity at right angles.
One ray is transmitted to mirror M1 and other is reflected through beam
splitter to mirror M2. From both these mirrors, the rays are reflected
back and these recombines at the semi reflecting surface. Then they
are transmitted to the eye or a detector.
The two optical paths must be equal for both beams. The
horizontal beam crosses the beam splitter three times, while the
vertical beam crosses the beam splitter once. A compensator plate is
introduced in the path of mirror M2 to equalize the distance travelled
by both the beams.
4.10
4.11 Twyman - Green interferometer
4.12
Working principle
Laser light first falls on the beam splitter P. It is partially
reflected by 90° and falls on the other reflector S. A portion of light
passes through P and strikes the corner cube. Light is turned through
180° by the corner cube and recombines at the beam splitter S.
Applications
Laser interferometers are used to calibrate machine tables, slides,
and axis movements of coordinate measuring machines, etc.
They are used in calibration of measuring standards like slip
gauges.
Advantages
1) The equipment is portable
2) Easy installation
3) High accuracy and precision in measurements.
4) Long range optical path up to 60m.
5) No change in performance due to wear and tear.
6) Six simultaneous measurements can be made in different axes.
4.13
Versions of AC laser interferometer
1) Standard interferometer
The interferometer explained above is a standard
interferometer. It is least expensive. A corner cube is used as a
measurement retroreflector. Displacement is measured between the
interferometer and the corner cube.
Applications
It is used for measuring linear positioning, displacements,
straightness, flatness and squareness of planes, etc.
COMPUTER IN METROLOGY
4.13 Coordinate Measuring Machines (CMM)
Coordinate Measuring Machine (CMM) is an electromechanical
system designed to determine the location, orientation, dimensions
and geometry of a component or object.
4.14 Types of measuring machines
The commonly used types of measuring machines are :
1) Length bar measuring machine
2) Coordinate measuring machines (CMM)
3) Universal measuring machine
4) Computer controlled coordinate measuring machine
4.15
Fig.4.10 Types of CMM
4.16
(f) Column structure: This structure is similar to the
construction of a machine tool. The X–axis and Y–axis movements
are achieved by moving the work table, while the Z–axis motion is
achieved by moving the probe quill vertically.
4.17
4.18 Performance of CMM
The performance of CMM is measured by conducting various
tests. These includes the following:
1) Repeatability test : The center coordinates of a sphere are
measured ten times quickly. Only four probing points per sphere
are allowed. The range of the coordinates is reported.
2) Linear displacement accuracy test : Measurements are made
parallel to the machine axes. This is the key test for traceability.
A step gage or laser interferometer is used.
3) Volumetric performance test : It involves measuring the length
of the ball bar between its two end spheres. This measurement is
repeated in many positions throughout the machine volume. Any
change in the measured length reflects machine geometry errors.
4) Offset probe performance test : This test also uses the ball bar,
but the test is done to find the error in angular motion of the arm.
5) Bidirectional length test: A small gage block is measured in
multiple positions to detect any size measurement error due to
probe tip calibration.
4.19
Fig.4.11 Operation of computer controlled CMM
After a part has been produced on the CNC machine tool, the
part will be checked on a CMM with the inspection program
downloaded from the computer. After CMM checks the part, data
about the part is sent back to the computer, where the original part
geometry is stored. Thus, the geometry of the designed part is
compared with the geometry of the produced part and deviations are
identified. This will help in identifying the problems in manufacturing
and rectification.
4.22.1 Probes
The probe of the CMM is fastened to a mechanical structure
that allows movement of the probe relative to the work piece. The tip
of the probe is usually a rub ball. Probes may have a single tip or
multiple tips. The X, Y and Z coordinates of the probe are accurately
measured by displacement transducers and recorded by the CMM
controller. Rotary encoders, optical scales and magnetic scales are
the commonly used displacement transducers.
4.20
The following two types probes are commonly used in CMMs.
a) Trigger type probe b) Measuring type probe
4.21
4) Calculation of geometric elements
5) Automatic evaluation and recording of complete measurement
6) Automatic compensation of work piece misalignment
7) Statistical evaluation of measured data
8) Curve measurement and gear measurement
9) Measurement of diameter, center distances, lengths,
geometrical and form errors
10) Interface to CAD software.
4.22
head. Magnetic strips are provided in the devices using magnetic
recording principle. Readout is provided by a range of digital display
units.
4.23
Examples :
Height gauges with built-in integral computer with keyboard are
available.
Slip gauge box with built in computer and printer has been
developed. The computer helps in selecting various combinations
of gauges to build any dimension.
The roundness measuring machine fitted with computer carry out
a large number of tedious calculations quickly.
Computer can easily define the circle if three points on its
circumference are completely defined. Manually this job is time
consuming.
In gear measuring machine, the computer carry out all tedious
calculations and give accurate results instantly.
4.24
Built in diagnostic tools is incorporated for the verification of all
input / output signals.
Advantages of computer aided inspection
1) Reduced Inspection time.
2) Reduced rejection of parts due to in-process inspection.
3) Automatic evaluation and recording of complete measurement
4) Self teaching capability of the system.
5) Built-in programs for repeated inspection tasks.
6) More flexibility to inspect different part configuration.
7) More accurate and precise inspection.
8) Interface to CAD software.
9) Statistical evaluation of measured data.
Review Questions
4.25
14) What are the causes of errors in CMM?
15) List out the applications of CMM.
16) What are the advantages and disadvantages of CMM?
17) List out the features of CMM software.
18) Write short notes on digital devices.
19) List out the advantages of computer aided inspection.
4.26
Unit – V
MEASUREMENT OF
MECHANICAL PARAMETERS
5.1 Measurement of force
Force is the product of mass and acceleration, i.e. 𝐹 = 𝑚𝑎 or
𝑊 = 𝑚𝑔 . Force is a vector quantity. The measurement of force
involves the determination of its magnitude and direction.
5.2
The unequal arm balance uses two arms. One is called the
load arm and other is called power arm. The load arm is used to apply
the load to be measured. The power arm is used to apply counter
weights to set the balance in equilibrium.
5.3
Measurement of unknown force
Let, 𝐹2 be the force to be measured. We know that the power
arm distance 𝑎 is calibrated in terms of force 𝐹2 . Apply the force
through knife edge 𝑃 .
5.4
The system is provided with an adjustable counter poise to
get an initial balance. The poise weight 𝑊𝑃 is set at zero of beam
scale, before the unknown load 𝑊 is applied to the platform. The
counterpoise is adjusted to obtain initial zero balance. In order to
simply the analysis, the weight 𝑊 on the platform can be replaced by
two arbitrary weights 𝑊1 and 𝑊2 . When the weight 𝑊𝑃 is applied it is
entirely balanced by the weigh 𝑊𝑠 in the pan.
At equilibrium,
𝑇 × 𝑏 = 𝑊𝑠 × 𝑎 − − − (1)
𝑓
and, 𝑇 × 𝑐 = 𝑊1 × 𝑒 + 𝑊2 ℎ − − − (2)
𝑑
ℎ 𝑓
=
𝑒 𝑑
If the links are so proportional that,
From the above equation (3), it is clear that the weight 𝑊 may
be placed anywhere on the platform. Its position relative to the two
knife edges of platform is immaterial.
𝑎
From equation (1), 𝑇 = 𝑊𝑠 − − − (4)
𝑏
ℎ
From equation (3), 𝑇 = 𝑊 − − − (5)
𝑐
𝑎 ℎ
Equating (4) & (5), 𝑊𝑠 = 𝑊
𝑏 𝑐
𝑎 𝑐
Unknown weight, 𝑊 = . 𝑊𝑠
𝑏 ℎ
𝑊 = 𝑀 𝑊𝑠
𝑎 𝑐
Where, 𝑀 = . is called the multiplication ratio of the scale.
𝑏 ℎ
5.5
𝑊𝑝 by 1 scale division represent a force of 𝑥 kg, then a poise
If beam scale is so divided that a movement of poise weight
5.6
Pendulum scale works on the principle of multiple leverage
system. The force to be measured is transmitted from a suitable
mechanism and applied to the load rod. As the load is applied, the
sectors rotate about pivot. It moves the counter weights outward. This
movement increases the counter weight effective moment until the
load and balance moments are equalised. Motion of the equalizer bar
is converted to indicator movement by a rack and pinion.
5.8
5.7.2 Pneumatic load cell
5.7.4 Shear type load cell (or) Shear beam load cell
Shear beam load cells are used to measure the applied force
in terms of shear strain induced by the force. There are two types of
shear beam load cells : (a) Double ended and (b) Single ended.
5.10
Fig.5.9 Shear type load cell
5.11
The output signal from load cell is very weak. Signal amplifiers
are used to amplify this signal up to convertible range. The amplified and
filtered signal is further converted to digital form by analogue to digital
converter. The display unit is used to display the measured weight.
5.14
The deflection of the shaft can be read directly when the shaft
is stationery. But, this is not possible when the shaft is rotating. In such
cases, stroboscope is used. The flashing light of the stroboscope is
applied onto the scale, The flashing frequency is adjusted till a
stationery image is obtained. The scale reading can now be taken for
measuring the angular twist. The torque can be calculated from the
measured angular twist.
2𝜋𝑁 𝑇 2𝜋𝑁 𝐹 𝐿
𝑃 = =
60 60
Where, 𝑃 = Power dissipated (watts)
𝑁 = Angular speed (rpm)
𝐿 = Length of load arm (m)
𝐹 = Force (N)
5.16
5.9.2 D.C dynamometer
5.17
5.9.3 Inductor dynamometer or eddy current dynamometer
2 𝜋𝑁 𝑇
𝑃 =
60
5.18
5.9.4 Hydraulic dynamometer
It uses fluid friction rather than dry friction for dissipating the
input energy. It consists of rotating disc and a stationary casing. The
rotating disc is connected to the drive shaft of the prime mover and it
revolves inside the stationary casing.
5.19
Advantages of hydraulic dynamometers
1) Power of very large and high speed engines can be measured.
2) More flexibility in controlling the operation.
3) Less space is required for high absorption capacity.
4) No additional water is required for cooling.
5.20
A single diaphragm in its simplest form is shown in Fig.5.18.
It is a thin, flat, circular plate fixed at the two ends. It will deflect when
pressure is applied. The resulting differential pressure is given by
𝑃1 − 𝑃2 . This can be used only for relatively small movements. The
deflection is limited by linearity constraints or stress requirements.
Some modification is required for practical applications.
5.21
The linear variable differential transducer (LVDT) transforms
a mechanical displacement into an electrical signal. The magnetic
core is connected to an elastic pressure transducer like a diaphragm
gauge. The diaphragm gauge senses the applied pressure and
converts it into displacement. Then it moves the core of the LVDT.
5.12 Rotameter
Rotameter is used to measure the volume flow rate of fluids.
It is a type of variable area flow meter. The meter consists of a float
within a vertical tube. The tube is tapered to an increasing cross-
sectional area at its outlet. Flow entering through the bottom passes
over the float, which is free to move. The equilibrium height of the float
indicates the flow rate.
5.24
As fluid flows upward through
the tube, three forces act on the float :
a downward gravity force, an upward
pressure, and viscous drag forces.
5.25
Advantages of rotameter
1) It requires no external power.
2) Simple and economic device
3) The scale is approximately linear.
4) Condition of flow is visible
5) High resistant to thermal shock and chemical action.
Disadvantages of rotameter
1) The meter must be installed in a vertical position
2) The float may not be visible when opaque fluids are used
3) It cannot be used for liquids with solids suspension.
4) It does not provide electronic readout.
5) The calibration is affected by fluid density.
5.26
It is expressed by the following relation :
𝐸=𝑈 𝐵𝐿
Where, 𝐸 = Induced electromotive force (emf)
𝑈 = Average velocity of fluid
𝐵 = Magnetic flux density
𝐿 = Length of the conductor
5.28
The following two operating modes are available :
(a) Constant current mode
In this mode, a fixed current is passed through the sensor to
heat it. The sensor resistance and its temperature are allowed to vary
with the rate of heat transfer. Bridge-deflection voltage provides a
measure of the fluid velocity.
5.29
Fig.5.26 Ultrasonic flow meter
5.30
to the speed of the scattering particle. Laser beam is used as the
incident wave source in laser Doppler anemometer. LDA measures
the time-dependent velocity at a point in the flow.
5.32
Review Questions
5.34
Q & A.1
Unit – I : BASIC CONCEPTS OF MEASUREMENTS
1. Define metrology.
Metrology is a branch of science which deals with measurements
and measuring instruments
2. Define unit. State its categories.
The standards used for measuring each physical quantity is
known as units. The units are divided into three categories.
a) Fundamental units or base units.
b) Supplementary units.
c) Derived units.
3. List out the base quantities and its base units.
Quantity Unit Symbol
Length Metre m
Mass Kilogram kg
Time Second s
Temperature Kelvin K
Electric current Ampere A
Luminous intensity Candela cd
Amount of substance Mole mol
4. Define metre.
Metre is defined as the length equal to 16,50,763.73 wavelengths
in vacuum of the radiation corresponding to the transition between
energy levels 2P10 and 5d5 of Krypton 86 atom.
5. Define second.
It is the fundamental unit of time. Second is equal to the
duration of 9192631770 periods of the radiation corresponding
to the transition between the hyper fine levels of the ground
state of the Caesium –133 atom.
6. Mention the two supplementary units.
1) Radian (rad ) 2) Steradian (sr )
Q & A.2
7. State the needs for inspection.
1) To meet the interchangeability of manufacture.
2) To produce components with acceptable quality levels.
3) To reduce scrap and wastages.
4) To purchase good quality raw materials and tools.
5) To reduce the rejection percentage of products.
8. Define measurement.
Measurement is the comparison of a given unknown quantity
with one of its predetermined standard values adopted as a unit.
Q & A.3
13. List out the linear measuring instruments.
Steel rule, calipers, dividers, depth gauge, Vernier calipers,
micrometers, Vernier height gauge, slip gauge, comparators, etc.
14. Name the angular measuring instruments.
Angle gauges, sprit level, protractors, combination sets, sine
bars, clino meter, angle dekkor, auto collimator, etc.
15. List out the geometric measuring instruments.
Thread micrometers, gear tooth Vernier calipers, profilometers,
profile projectors, roundtesters, Coordinate Measuring
Machines, etc.
16. Name the instruments used for surface measurements.
Profilometers, Taylor-Hobson Tallysurf, tracer type profilogram,
double microscope, etc.
17. List out the scope of metrology.
1) Complete evaluation of newly developed products.
2) Determination of the process capabilities
3) Determination of the measuring instrument capabilities
4) Minimizing the cost of inspection.
5) Reducing the cost of rejects and rework.
6) To standardize the measuring methods:
18. What is standardisation?
Standardisation is the process of creating, issuing and
implementing standards.
19. Name the international standards organisations.
International Organisation of Weights and Measures
International Organization for Standardization (ISO)
International Laboratory Accreditation Conference(ILAC)
International Electrotechnical Commission (IEC)
National Metrology Institute (NMI)
20. What is meant by legal metrology?
Legal metrology is concerned with the regulations and legal
requirements for the measuring instruments and methods of
Q & A.4
measurements. The main objective is to maintain uniformity of
measurement in a particular country.
21. Define range of measuring instruments.
Range is defined as the upper and lower limits an instrument
can measure a value without exceeding the maximum
permissible error.
22. Differentiate resolution and sensitivity.
Resolution is the smallest value that an instrument can
measure.
Sensitivity of the instrument denotes the smallest change in
input value to which the instrument responds.
23. Define hysteresis.
It is the difference between the indications of a measuring
instrument when the same value of the measured quantity is
reached by increasing or by decreasing that quantity.
24. What is response time?
It is the time taken by an instrument to give an indication after
a sudden change in the measured quantity.
25. Define calibration.
The process of periodic checking against measuring
instruments and standards of high accuracy is called as
calibration.
26. Differentiate between precision and accuracy.
Precision is the closeness of agreement between
independent test results obtained under specified conditions.
It indicates the degree of repeatability in the measuring
process.
Accuracy is the closeness of agreement between the
measured value and the true value.
27. List out the factors affecting accuracy of measuring system.
1) Calibration standards 2) Work piece
3) Measuring Instruments 4) Environmental effects
5) Person or Inspector carrying out the measurement
Q & A.5
28. Define reliability.
Reliability is the degree of consistency of a measurement. A
test will be reliable when it gives the same repeated result
under the same conditions.
29. Define error.
Error is the difference between the measured value and the true
value.
30. Differentiate absolute error and relative error.
Absolute error is the algebraic difference between the results
of measurement to the true value of the quantity measured.
Relative error is the ratio of the absolute error and the value
of comparison used for calculation of that absolute error.
31. Give the classification of errors,
1) Static errors
(a) Reading errors
(b) Environmental errors
(c) Characteristic errors
2) Loading errors
3) Dynamic error
(a) Systematic error
(b) Random error
32. What is loading error.
Instrument loading error is the difference between the value of
measurand before and after the measurement.
33. List out the sources of systematic errors.
1) Calibration errors 2) Atmospheric conditions
3) Stylus pressure 4) Parallax error
34. Mention the factors to be considered for the selection of
measuring instruments.
1) Measuring range 2) Accuracy 3) Precision
4) Resolution 5) Measurement method
6) Reliability 7) Cost
Q & A.6
Unit – II: LINEAR AND ANGULAR MEASUREMENTS
1. List out the important linear measuring instruments.
1) Steel rule 2) Calipers 3) Combination set
4) Feeler gauge 5) Pitch screw gauge 6) Vernier caliper
7) Vernier height gauge 8) Micrometer 9) Thread micrometer
2. Mention the types of calipers.
1) Outside caliper 2) Inside caliper 3) Outside spring caliper
4) Inside spring caliper 5) Jenny or Odd leg caliper
3. What is combination set?
Combination set is the most adaptable and widely used non-
precision instrument in layout and inspection work. The
combination set consists of scale, squaring head, protractor
and centre head.
4. List out the applications of feeler gauges.
1) It is used to measure the clearance between the tool and
work piece in the machines.
2) It is used in automobiles for adjusting the spark plug
clearance correctly.
3) It is used to measure the clearance between piston and cylinder.
4) It is used for measuring the clearance between the guide and
guide ways in machines.
5. List the applications of radius gauges and pitch screw gauges.
Radius gauge is useful for measuring and checking the
inside and outside radii of fillets and other round surfaces.
Pitch screw gauge is used to identify or check the pitch of the
threads on different threaded items.
6. What is least count? How it is calculated in vernier caliper?
Least count is the smallest value that can be measured by
using an instrument.
Least count = Length of one main scale division –
Length of one vernier scale division
7. Give the uses of vernier height gauge.
Vernier height gauge is used to measure the height of the
work piece and to mark the specified dimension exactly on
the work piece.
Q & A.7
8. List out the types of micrometers.
1) Outside micrometer 2) Inside micrometer
3) Depth micrometer 4) Thread micrometer
5) Optical micrometer 6) Light wave micrometer
9. What are the uses of inside micrometer and depth micrometer?
Inside micrometer is used to measure the inside diameter of
a hole.
Depth micrometer is used to measure the depth of blind
holes, slots and grooves.
10. What is the use of thread micrometer? Give the formula
used in three wire method for pitch diameter.
Thread micrometer is used for measuring the pitch diameter of
a thread accurately.
= − + +
Where, = Distance between the wires
= Diameter of wire
= Included angle of thread, = pitch of thread
11. What is the principle involved in optical micrometer?
Optical micrometer uses a shadow measurement principle. The
dimension is calculate by measuring the time duration between
the detection of edges of shadow.
12. What is light wave micrometer?
Light wave micrometer is provided with an optical pressure
indicator which uses light wave interference. This is provided to
apply exactly the same pressure on the object for each
measurement.
13. What is slip gauges? Define wringing.
Slip gauges are precision gauge blocks used for measuring
linear dimensions.
Wringing is the process of combining slip gauges by placing
one over another with the help of adhesion between the two
highly finished surfaces.
Q & A.8
14. What are the requirements of slip gauges?
1) The end faces of the gauges must be flat and parallel to each other.
2) All the adjacent faces must be perfectly square to each other.
3) They must possess high degree of surface finish.
4) The actual size must be within tolerances.
5) The edges must be deburrred.
15. Classify the slip gauges.
Grade II is intended for use in workshops during actual
production of components, tools and gauges.
Grade I is of higher accuracy for use in inspection
departments.
Grade 0 is used in laboratories and standard rooms for
periodic calibration of Grade I and Grade II gauges.
16. List out the uses of slip gauges.
1) Slip gauges are used for the calibration of precision
measuring instruments.
2) Slip gauges are used in setting sine bars for establishing
angles.
3) They are used to check comparators and optical inspection
devices.
4) They are used as auxiliary measuring system on milling
machine.
17. What is the use of universal bevel protractor?
Universal bevel protractor is used for measuring angle between
two adjacent surfaces with an accuracy of 5 minutes.
18. What are the types of sine bar?
1) Sine centre 2) Sine table 3) Compound sine table
19. Mention the uses of sine centre and compound sine table.
Sine centers are used for mounting conical work pieces
which cannot be held on a simple sine bar
Compound sine table is used to measure compound angles
of large work pieces.
20. What are the uses of sine bar?
Measuring unknown angle.
Setting up known angle.
Checking of unknown angles of heavy component.
Q & A.9
21. Write the limitations of sine bar.
1) The accuracy of sine bars is limited by measurement of center
distance between the two precision rollers.
2) It cannot be used as a primary standard for angle
measurements.
3) Sine principle is fairly reliable at angles less than 15o, but
becomes inaccurate as the angle increases.
4) Sine bar becomes impractical and inaccurate as the angle
exceeds 45o.
22. What is clinometer? Give its uses.
A clinometer is a special case of application of spirit level for
measuring the inclinations
Uses :
1) It is used for the measurement of an inclined plane with
respect to a horizontal plane
2) It is used for the measurement of the relative position of two
mutually inclined surfaces :
23. What is vial?
Vial is a closed glass tube of accurate size in a spirit level, which
is used for storing the liquid. It is graduated in linear scale and
the bubble moves inside it.
24. What is the principle involved in autocollimator?
When the reflecting plane is tilted through a small angle
θ°, the reflected rays will be inclined at 2θ° to the optical axis.
These rays will be focused at a point in the focal plane with
a distance = 2 ! from the origin, where ! is the focal
length of the lens.
25. How angles are measured in angle dekkor?
Angle dekkor measures angle by comparing the reading from a
standard, such as combination of angle gauges, and the reading
from the work piece under test. Thus it acts as a comparator.
26. List out the requirements of a good comparator.
1) The scale used in the instrument should be linear.
2) There should not be any backlash and lag between the
movement of the parts.
3) The instrument must be precise and accurate.
Q & A.10
4) The indication method should be clear.
5) The indicator should be constant in its return to zero.
27. Classify the comparators.
1) Mechanical comparator 2) Electrical comparator
3) Pneumatic comparator 4) Optical comparator
28. What are the advantages of mechanical comparator?
It is cheaper in comparison to the other comparators.
It has linear scale, which is easily readable.
It is compact and robust in construction.
It is easily handled.
It is suitable for ordinary workshops.
29. Write down the advantages of electrical comparator.
It has very less number of moving parts.
The magnification is comparatively high.
It can be used for various ranges.
It has a compact construction.
30. What are the limitations of pneumatic comparators?
The scale is generally not uniform.
Very high magnification is required to avoid meniscus errors.
Various gauging heads are needed for different dimensions.
It requires a number of auxiliary equipment.
Handling of equipment is difficult.
31. What are the uses of optical comparator?
Optical comparator is used for checking gear teeth, screw
thread, cutting tools, needles, etc.
Q & A.11
Unit – III : FORM MEASUREMENT
Q & A.12
8. Differentiate between progressive error and periodic error.
When the pitch of the screw is uniform, but is shorter or longer
than its nominal value, the error is called progressive error.
The periodic error repeats itself at equal intervals along the
thread. In this case, the successive portions of the thread length
are either shorter or longer than the mean value.
Q & A.13
16. What are the methods available to measure the effective
diameter of screw threads?
1) Using thread micrometer 2) One wire method
3) Two wires method 4) Three wires method
17. Define best size wire.
It is a wire of such diameter that makes contact with the flanks
of the thread exactly on the effective diameter. The best size
wire should always be used for measuring the effective
diameter.
18. List out the instruments used to measure pitch of threads.
a) Screw pitch gauge
b) Pitch measuring machine
c) Tool maker’s microscope
19. What are thread gauges?
Thread gauges are limit gauges used to assure that the screw
is within the prescribed limits of size. They are used for
checking screw threads quickly.
20. Classify thread gauges based on application.
1) Working gauges 2) Inspection gauges 3) Master gauges
21. Classify thread gauges based on their form.
(a)Plug screw gauges : They are used to inspect internal thread forms.
(b)Ring screw gauges : They are used to inspect external thread forms.
(c) Caliper gauges : They are used to inspect external thread forms.
22. What are caliper gauges?
Caliper gauges are equivalent of gap gauges with thread form
on anvils. They are used to inspect external thread forms.
23. List out any four gauges for checking external threads.
1) GO screw ring gauge 2) GO screw caliper gauge
3) NO GO screw ring gauge 4) NO GO screw caliper gauge
24. Write the applications of floating carriage micrometer?
It is very useful for thread plug gauge manufacturers, gauge
calibration laboratories and standard tool rooms.
Q & A.14
25. List out the various types of gears.
1) Spur gear 2) Helical gear 3) Bevel gear
4) Rack and pinion gear 5) Worm and worm wheel
Q & A.17
Unit – IV : ADVANCES IN METROLOGY
1. What is laser?
LASER stands for Light Amplification by Stimulated Emission
of Radiation.
2. What are the types of lasers.
1) Gas lasers 2) Liquid laser 3) Solid state laser
4) Chemical laser 5) Semiconductor laser
3. Give examples for gas lasers and solid state laser
Gas laser : Argon laser, CO2 laser, Helium-Neon laser,
Krypton laser, Xenon laser, etc.
Solid state laser : Ruby laser, Nd:YAG laser, etc.
4. What are the advantages of lasers?
1) The laser beam has high intensity and high power
2) It has no divergence up to a great distance.
3) It has high information carrying capacity
4) It has very minimum signal leakage.
5) It is free from electro-magnetic interference.
5. List out the applications of laser.
1) Data storage in CDs and DVDs.
2) Barcode scanning
3) High speed transmission of information over fiber optic cable.
4) Manufacturing uses such as metal cutting, drilling, welding, etc.
5) Distance monitoring and measurement
6) Holography, tattoo removal, etc.
6. What is scanning laser gauge?
Scanning laser gauge is a non-contact gauge that measures
with a scanning laser beam. It uses a shadow measurement
principle.
7. Give the applications of scanning laser gauge.
It is used to measure diameters of wires, rods and other
cylinders. It also measures gaps, edge positions and various
geometric dimensions of objects.
Q & A.18
8. What is the use of laser triangulation sensors?
They are used for measurement and monitoring tasks in
factory automation, electronics production, robotics and
vehicle construction.
9. What is interferometry?
Interferometry is a measurement technique based on the
interference of light waves. It is used for the measurements of
very small linear dimensions.
10. What are the essential conditions for interferometry?
1)
The two sources of light should continuously give light rays.
2)
The wavelength and intensity of light rays should be equal.
3)
The phase difference should be maintained constant.
4)
Two sources should be very narrow and close to each
other.
11. Mention the uses of interferometry.
1) Distance monitoring and linear measurements.
2) Angular measurements.
3) Checking of flatness, squareness, parallelism, etc.
4) Calibration of slip gauges
5) Measurement of surface finish
12. What is the difference between the Michelson
interferometer and Twyman-Green interferometer?
Michelson Interferometer Twyman-Green interferometer
Extended monochromatic Monochromatic point light
light source is used source is used
No collimating lens is used Collimating lens is used
13. What is the principle involved in measuring the
displacement using interferometer?
If the path difference between the two reflected beams is an
odd number of half wavelengths, they cancel each other. If the
path difference is an even number of half wavelengths, they
reinforce each other. This creates interference fringes with
alternate dark and bright bands. By counting the number of
bands, the displacement can be measured.
Q & A.19
14. What is beam splitter?
Beam splitter is a plain parallel plate having a semi-transparent
layer of silver at its back. It splits the light into two rays of equal
intensity at right angles.
15. What is the use of corner cubes in interferometry?
The corner cube is a glass disk whose back surface has three
polished faces that are mutually at right angles to each other.
It will reflect light at an angle of 180°.
16. What is retroreflectors?
The corner cubes are called retroreflectors. It will reflect the
incident light at an angle of 180°.
17. State the applications of AC laser interferometer.
Calibration of machine tables, slides, and axis movements
of coordinate measuring machines, etc.
Calibration of measuring standards like slip gauges.
18. List out the advantages of laser interferometers.
1) The equipment is portable
2) Easy installation
3) High accuracy and precision in measurements.
4) Long range optical path up to 60m.
5) No change in performance due to wear and tear.
19. What is single beam interferometer?
Single beam Interferometer has the outgoing and returning
beam superimposed on each other. It gives the appearance of
only one beam travelling between the interferometer and the
retroreflector.
20. What is the principle of measurement in two frequency laser
interferometer?
The displacement is measured by counting the frequency and
the phase shift in frequency due to the Doppler effect.
21. What is CMM?
Coordinate Measuring Machine (CMM) is an electromechanical
system designed to determine the location, orientation,
dimensions and geometry of a component or object.
Q & A.20
22. List out the types of measuring machines.
1) Length bar measuring machine
2) Coordinate measuring machines (CMM)
3) Universal measuring machine
4) Computer controlled coordinate measuring machine
23. What are the types of CMM?
1) Cantilever type 2) Moving bridge type 3) Fixed bridge type
4) Horizontal type 5) Gantry type 6) Column type
24. List out the important features of CMM.
1) Hollow box construction.
2) Program for automatic error compensation
3) Crash protection for protecting sensitive components
4) Separate computer for each machine
5) Software for offline programming.
6) Software for statistical analysis.
25. List out the various causes of errors in CMM.
1) Geometric error in table of CMM
2) Alignment errors between table and probes.
3) Runout errors and perpendicularity errors in probes
4) Errors in optical readout of the digital system.
5) Straightness and perpendicularity errors in guide ways.
6) Errors in probe system calibration and zero setting.
26. What are the various tests conducted for measuring the
performance of CMM?
1) Repeatability test
2) Linear displacement accuracy test
3) Volumetric performance test
4) Offset probe performance test
5) Bidirectional length test
27. Mention the applications of CMM.
1) Checking the various dimensions of machined components.
2) Checking the profiles and dimensions of gears, threads and
cams.
Q & A.21
3) Checking of sheet metal panels, subassemblies and complete
car bodies in the automotive industries.
4) Sorting and pairing of similar components within tolerance
limits.
5) Checking and measuring various gauges, fixtures and other
inspection and production tools to validate their continued use.
28. List out the advantages of CMM.
1) Reduced Inspection time.
2) More flexibility to measure different part configuration.
3) Reduced operator errors in measurements and setup.
4) Reduced operator skill requirements.
5) More accurate and precise than manual inspection.
6) Better repeatability.
7) Reduced maintenance cost.
29. Name the two types of probes used in CMM.
1) Trigger type probe 2) Measuring type probe
30. What are the main features of CMM software?
1) Subroutine programs for repeated measurement tasks
2) Automatic calibration of the probe
3) Determination of the work piece coordinate system
4) Automatic evaluation and recording of complete
measurement
5) Statistical evaluation of measured data
6) Curve measurement and gear measurement
7) Interface to CAD software.
31. List out the factors affecting CMM.
1)Temperature 2) Humidity 3) Trapped dust
4) Contaminated oil 5) Vibration
32. Enumerate the advantages of digital devices.
1) Measuring element is free from errors.
2) Learning the procedure is quick and easy.
3) More accurate measurements.
4) No excessive reading errors .
5) Display can be zeroed quickly.
6) Output can be directly fed to computers
Q & A.22
33. What are the main uses of computers in metrology?
1) Controlling the function of measuring equipment
2) Processing the output of the measurement process.
34. What are the advantages of computer aided inspection?
1) Reduced Inspection time.
2) Reduced rejection of parts due to in-process inspection.
3) Automatic evaluation and recording of complete
measurement
4) Self-teaching capability of the system.
5) More flexibility to inspect different part configuration.
6) More accurate and precise inspection.
Q & A.23
5. What is the basic principle involved in electromagnetic
balance?
In electromagnetic type balances, an electromagnetic force is
applied instead of a placed weight to balance the beam
6. Define load cell.
When an elastic member is combined with a strain gauge and
used for the measurement of force, it is termed as a load cell.
7. List out the types of load cells.
1) Hydraulic load cell 2) Pneumatic load cell
3) Strain gauge load cell 4) Shear type load cell
Q & A.24
13. Classify the dynamometers?
a) Absorption dynameters : Absorbs mechanical energy
b) Driving dynamometers : Provides energy to operate the device
c) Transmission Dynamometers : They do not supply or receive
energy.
Q & A.25
When pressure is applied on piezoelectric crystals an electric
charge is produced. The electric charge is measured in terms of
applied pressure.
21. Define flow rate. What are flow meters?
The rate at which a fluid moves through a pipe known as the
flow rate.
The device used for the measurement of flow rate is known as
flow meters.
22. List out the types of flow meters.
(a) Obstruction meters
1) Venturi meters 2) Flow nozzles
3) Orifices 4) Rotameters
(b) Volume flow meters
1) Electromagnetic flow meters 2) Vortex-shedding meters
3) Ultrasonic flow meters 4) Positive-displacement
meters
Q & A.26
26. List out the advantages of electromagnetic flow meters.
1) Very low pressure loss.
2) No obstruction design.
3) Used for both steady and unsteady flow.
4) No internal moving parts.
5) Corrosive and dirty fluids can be measured.
27. What is hot wire anemometer?
Hot wire anemometer is used to measure the velocity of flow
of gases.
28. Mention the two operating modes in hot wire
anemometers.
1) Constant current mode 2) Constant resistance mode
29. What is ultrasonic flow meter?
Ultrasonic flow meters use sound waves to determine flow rate.
It uses the travel time of ultrasonic waves to estimate average
flow velocity.
30. State the working principle of LDA.
When light waves are scattered by particles in a moving fluid,
they undergo a frequency change called, Doppler shift. The
Doppler shift is proportional to the speed of the scattering
particle.
31. What are the two modes of operation in LDA?
a) Reference beam mode b) interference fringe mode
32. Give the difference between reference beam mode and
interference fringe mode.
In reference beam mode, a separate scattering and
reference beams are used.
In interference fringe mode, a single laser beam is divided
into two coherent beams of equal intensity
33. What is optical heterodyne?
The process of crossing and mixing the incident information
from two beams of light is known as optical heterodyne.
Q & A.27
MEASUREMENTS AND METROLOGY
MODEL QUESTION PAPER
Time: 3 Hrs. Max Marks : 100
[N.B: (1) Answer all TEN questions from PART – A. Each question
carries 3 marks.
(2) Answer division (a) or division (b) of each question in
PART – B. Each question carries 14 marks.]
PART – A
[10 × 3 = 30 Marls]
PART – B
[5 × 14 = 70 Marls]
11.(a) (i) Explain the needs for inspection and measurement. (7)
(ii) Differentiate between systematic error and random (7)
error.
(Or)
(b) (i) Explain the various methods of measurements. (7)
(ii) Explain the factors affecting accuracy of measuring (7)
system.
QP.1
12.(a) (i) Explain optical micrometer with neat sketch. (7)
(ii) What are the types of sine bar? Explain its uses and (7)
limitations.
(Or)
(b) (i) With the help of neat sketch, explain angle dekkor. (7)
(ii) Compare electrical comparators with mechanical (7)
comparators.
14.(a) (i) Explain Twyman green interferometer with neat sketch. (7)
(ii) Explain the computer controlled coordinate measuring (7)
machine.
(Or)
(b) (i) Explain dual frequency interferometer with neat sketch. (7)
(ii) Write short notes on computer aided inspection using (7)
robots.
15.(a) (i) Explain unequal arm balance with neat sketch. (7)
(ii) Explain laser Doppler anemometer with neat sketch. (7)
(Or)
(b) (i) Explain laser optical torque measurement with a neat (7)
sketch.
(ii) Explain the hydraulic dynamometer with neat sketch. (7)
QP.2