Scan Abort During Fastcal: Replace Power Module On Djinn Tonix Generator

A scan was aborted during fastcal on a Djinn Tonix generator, showing error code 26011852. Troubleshooting steps in the Djinn Tonix Troubleshooting Guide indicate replacing the power module (part number 5495652-2) on the generator will resolve the issue. The specific service request and work order numbers are provided for reference.

Uploaded by

Luis Batta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
372 views2 pages

Scan Abort During Fastcal: Replace Power Module On Djinn Tonix Generator

A scan was aborted during fastcal on a Djinn Tonix generator, showing error code 26011852. Troubleshooting steps in the Djinn Tonix Troubleshooting Guide indicate replacing the power module (part number 5495652-2) on the generator will resolve the issue. The specific service request and work order numbers are provided for reference.

Uploaded by

Luis Batta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

1/7/22, 9:07 Scan abort during fastcal : Replace Power Module on Djinn Tonix generator

Scan abort during fastcal : Replace Power Module on Djinn Tonix generator
Issue

Scan is aborted during Fastcal. Error 26011852 shown in the error log. 

Troubleshooting

Note: It is the Service person's responsibility to use the appropriate documentation when performing and verifying
service.

Refer to 5487207-8EN (Revolution) ACT/ACTs Service Methods (Adv). The following has been provided for
reference.  Always use the most appropriate documentation for your system.
 

Follow troubleshooting steps as per Service Methods > Troubleshooting > Djinn Tonix Generator > Djinn
Tonix Troubleshooting Guide (DOC1872071)

Resolution

Replace Power Module 5495652-2 on Djinn Tonix generator.

Part Number

5495652-2  Power Module -FRU

Modality / Product

Revolution ACT/ACTs

Error Code

Wed Aug  1 07:52:59 2018 260118522 2


 orp orp_subsys
gen_exec.cxx  3145
Wed Aug  1 07:52:59 2018
Host : Jedi    Ermes # : 260118522
Exception Class : Abort    Severity : Primary
File : gen_exec.cxx    Line# : 3145
Function : No System Function Reported
Scan Type : Axial Scan: 50234/8/24

(587h)Input Stage: Top Capacitor Voltage is too high [Vtopcapa > 400V].
Z6Z7= d8h,1hI (Vtopcapa = Z7Z6)

Key Words

PM

https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=216462&sliceId=1 1/2
1/7/22, 9:07 Scan abort during fastcal : Replace Power Module on Djinn Tonix generator

Reference / Defect / Quality / Tracking

SR#(ServiceMax) CASE 02566690 ,  WO-05212831 


SID: CT4630CL

Last Modified Date: 10/14/2018 ID: 216462

https://psdb2.gehealthcare.com/selfservice/viewContent.do?externalId=216462&sliceId=1 2/2