Scan Abort During Fastcal: Replace Power Module On Djinn Tonix Generator
Scan Abort During Fastcal: Replace Power Module On Djinn Tonix Generator
Scan abort during fastcal : Replace Power Module on Djinn Tonix generator
Issue
Scan is aborted during Fastcal. Error 26011852 shown in the error log.
Troubleshooting
Note: It is the Service person's responsibility to use the appropriate documentation when performing and verifying
service.
Refer to 5487207-8EN (Revolution) ACT/ACTs Service Methods (Adv). The following has been provided for
reference. Always use the most appropriate documentation for your system.
Follow troubleshooting steps as per Service Methods > Troubleshooting > Djinn Tonix Generator > Djinn
Tonix Troubleshooting Guide (DOC1872071)
Resolution
Part Number
Modality / Product
Revolution ACT/ACTs
Error Code
(587h)Input Stage: Top Capacitor Voltage is too high [Vtopcapa > 400V].
Z6Z7= d8h,1hI (Vtopcapa = Z7Z6)
Key Words
PM