Application News: Vividclub
Application News: Vividclub
04 - 2007
VividClub
Application
News
GE Healthcare
VividClub
Vivid 7
EchoPAC (analysis)
Vivid S6 (acquisition)
Vivid i (acquisition)
-2-
GE Healthcare
VividClub
Content
THE ECG..........................................................................................................................................................5
General information ............................................................................................................................................................5
Colour coding .........................................................................................................................................................................5
ECG Lead...................................................................................................................................................................................5
HOW TO WORK WITH THE Q-ANALYSIS PACKAGE FOR TVI DATA ........................................................8
Preparation..............................................................................................................................................................................8
Start the analysis ..................................................................................................................................................................8
Place the ROI...........................................................................................................................................................................9
Move the ROI...........................................................................................................................................................................9
Delete the ROI......................................................................................................................................................................10
Change ROI size..................................................................................................................................................................10
The different modalities ..................................................................................................................................................11
Change the mode ..............................................................................................................................................................11
Tissue Velocity Imaging (TVI) ........................................................................................................................................12
Tissue Tracking (TT)...........................................................................................................................................................13
Strain Imaging (SI)..............................................................................................................................................................14
Strain Rate Imaging (SRI) ................................................................................................................................................15
Tissue Synchronisation Imaging (TSI) .......................................................................................................................16
WHEN TO USE…...........................................................................................................................................17
… TVI, TISSUE TRACKING, STRAIN, STRAIN RATE, TSI............................................................................17
Global systolic function:..................................................................................................................................................17
Regional systolic function: .............................................................................................................................................17
Diastolic function – relaxation abnormalities: ......................................................................................................17
Asynchrony ...........................................................................................................................................................................17
-3-
GE Healthcare
VividClub
NOTE
This hand out is additional training material.
For more information please refer to the user manual and/or reference manual.
-4-
GE Healthcare
VividClub
The ECG
General information
It is crucial to have the ECG applied properly when you want to work with TVI data!
It is needed for the Event Timing measurement and for some triggered modes.
Therefore make sure to have a proper and stable ECG.
Colour coding
Put on the ECG stickers following the colour codes:
Red: right arm
Yellow: left arm
Green: left leg
ECG Lead
Lead I: red / yellow right and left arm
Lead II: red / green right arm and left leg
Lead III: yellow / green left arm and left leg
-5-
GE Healthcare
VividClub
Preparation
Applie a proper and stable ECG.
Acquisition
Do the adjustments for all views or walls and store 3 heart cycles each time.
Do every time the same order when storing single walls, that makes it easier in the analysis later
on.
Here an example:
4 CH
• Septum
• Lateral wall
• RV free wall
2 CH
• Anterior wall
• Inferior wall
PLAX
• Posterior wall
• Anteroseptal wall
-6-
GE Healthcare
VividClub
General information
• The event timing measurements can be done in M-Mode or in Doppler.
• The system will measure the time from beginning of R- wave on the ECG to the
measurement point set by the user.
• This time will be stored in ms in the worksheet.
• Mitral valve opening and –closure (MVO and MVC)
• Aortic valve opening and –closure (AVO and AVC)
Note
The system is detecting the beginning of the R- Wave on the ECG.
Therefore it is crucial to have a proper ECG Signal.
If the patient has arrhythmia this can lead to improper results, because the cycle length of each
heart cycle is completely different. There is no algorithm in the event timing to compensate for
this. The measured time is a fixed value. Be aware of this fact when you do your analysis.
Measurements
M-Mode
1. Acquire a nice M-Mode signal where the valve opening and closure is clearly visible.
2. Press Measure.
3. Open the folder for Event Timing.
4. Select the desired valve.
5. Set the marker at the opening.
6. The system automatically jumps to the measurement for the closure.
7. Set the marker for the closure.
Doppler
1. Acquire a nice doppler signal; most likely including the valve clicks.
2. Press Measure.
3. Open the folder for Event Timing.
4. Select the desired valve.
5. Set the marker at the opening.
6. The system automatically jumps to the closure.
7. Set the marker for the closure.
Now the measurements are stored in the worksheet and will be used for TVI Q-Analysis and AFI.
-7-
GE Healthcare
VividClub
Preparation
It is crucial to have the ECG applied properly when you want to work with TVI data!
Measure the event timing before starting the analysis. This will help to correlate the curves with
the timing in the heart cycle.
Recall a TVI loop from the archive.
On EchoPAC:
Press the Advanced button on the right side menu to get all options.
Then you will get the button for the Q-Analysis to click on.
1 3
-8-
GE Healthcare
VividClub
Click once on the ROI. The ROI can be moved now to another position.
To set the new position you need to press the left mouse key.
-9-
GE Healthcare
VividClub
- 10 -
GE Healthcare
VividClub
On EchoPAC:
In the menu on the right side you will find for each mode a button to click on.
- 11 -
GE Healthcare
VividClub
Colour code:
Red: movement towards transducer
Blue: movement away from the transducer
Aliasing: green or white
Unit: cm/s
Systole
IVCT
IVRT
- 12 -
GE Healthcare
VividClub
Colour code:
See the rainbow colours beside. The numbers at the scale can be adjusted by changing
the scale.
Unit: mm
The parametric image is triggered on systole. You will see how the colour bar will build
up. In diastole the image is a still frame to better visualise the end systolic maximum
distances.
- 13 -
GE Healthcare
VividClub
Colour code:
Red: shortening
Blue: lengthening
Unit: %
- 14 -
GE Healthcare
VividClub
Colour code:
Red: indicates how fast the shortening is happening
Blue: indicates how fast the lengthening is happening
Green:no change
Unit: 1/s
É
Á
IVCT
Systole
IVRT
- 15 -
GE Healthcare
VividClub
Colour code:
Green: the peak velocity is very early
Red: peak velocity is at the end of the
specified cardiac cycle
Unit: ms
For those who are red/green blind there is another colour map where red is changed to blue.
- 16 -
GE Healthcare
VividClub
When to use…
… TVI, Tissue Tracking, Strain, Strain rate, TSI
Asynchrony
• TSI
- 17 -
GE Healthcare
VividClub
Tissue Tracking
Regional wall motion abnormality
Global systolic function
Strain
Evaluation of ischemic heart disease
Analysis of a specific piece of myocardium
Strain rate
Evaluation of ischemic heart disease
Analysis of a specific piece of myocardium
TSI
Assessing asynchrony
- 18 -