Acquity UPLC H-Class Binary Solvent Manager
Acquity UPLC H-Class Binary Solvent Manager
Acquity UPLC H-Class Binary Solvent Manager
ii
Table of Contents
Copyright notice ................................................................................................... ii
Overview ................................................................................................................ 1
Flow rate accuracy ............................................................................................... 4
Delay volume and mixing characteristics........................................................... 5
Intelligent intake valves...................................................................................... 5
Connections INSIGHT® and ACQUITY UPLC Console diagnostics ................ 5
Location of binary solvent manager in an example ACQUITY UPLC system. 6
Major components................................................................................................ 7
Flow path through the solvent management system......................................... 9
iv Table of Contents
Overview
The Waters ACQUITY UPLC Binary Solvent Manager delivers solvent
compositions for isocratic and binary gradient methods at flow rates of 0.01 to
2.0 mL per minute. Its features include in-line filters upstream of a primary
® 2
check valve, the Waters Intelligent Intake Valve (i Valve), automated
priming functions, and daily system-setup routines.
Under high, system backpressure, and by means of a mixing tee, the binary
solvent manager combines the outputs of two, parallel, flow-path channels, A
and B. Four solvents—the chromatographic solvents—are divided into pairs,
one pair for each pump channel. As such, they are designated A1 and A2 (for
channel A) and B1 and B2 (for channel B). Each preselected solvent, A1 or A2,
B1 or B2, passes from the degasser through a selector valve. The solvent then
flows into primary and accumulator pump heads, then the vent valve, the
mixing tee and, finally, the sample manager. The device has an integral
six-channel solvent degasser. Two of the channels are reserved for sample
manager wash solvents.
The binary solvent manager’s two pump channels permit quick solvent
changeovers at a flow rate of 4.0 mL per minute. Low-volume degasser
channels (480 L internal volume), automated solvent-selection valves, and
an automated vent valve make the quick changeovers possible.
The pumps, whose flows are additive, can support fast or ballistic separations
on short columns, with flow rates exceeding 1 mL/min at system
backpressures greater than 82,737 kPa (827 bar, 12,000 psi). Their active,
integral, plunger-seal-wash functions flush away particulates or precipitants
from their plungers and plunger seal interfaces.
The design of the binary solvent manager (and, indeed, the system overall) is
optimized for sub-2 µm particle liquid chromatography. The upper pressure
limit is set at 124,106 kPa (1241 bar, 18,000 psi) per pump. A maximum flow
rate of 1 mL/min is permitted at system backpressures to 124,106 kPa
(1241 bar, 18,000 psi), which falls within the range of optimal linear velocity
for sub-2 µm particle columns of ID 1.0 mm to 3.0 mm.
Overview 1
The following figure depicts the binary solvent manager's flow rate and
pressure envelope.
124,106 kPa
(1241 bar,
18,000 psi)
Pressure
82,737 kPa
(827 bar,
12,000 psi)
1 2
Flow rate (mL/min)
2
The figure below shows the chromatogram of a typical application with a
separation of antidepressants. For such an analysis, you can run the system at
1.6 mL/min, with 98,595 kPa (986 bar, 14,300 psi) system backpressure, using
both pumps in a proportioning mode. The example shows the capability of
running extremely fast separations utilizing the very low injection cycle time
and minimal system delay volume of the ACQUITY UPLC system.
Desipramine - 0.380
Nortriptyline - 0.406
Imipramine - 0.473
Nordoxepin - 0.355
Amitriptyline - 0.494
Doxepin - 0.439
Trimipramine - 0.532
AU
v - 0.066
Minutes
Overview 3
• System volume: 85 L
• Gradient:
– Solvent A: 10 mM ammonium bicarbonate pH 10.0
– Solvent B: acetonitrile
Gradient:
Flow
Time %A1 %B1
(ml/min)
1.60 75.0 25.0
0.02 1.60 75.0 25.0
0.41 1.60 20.0 80.0
0.51 1.60 20.0 80.0
0.52 1.60 75.0 25.0
1.08 1.60 75.0 25.0
4
nature of the mobile phase, the partial pressure of dissolved air in the mobile
phase, and the operating temperature. To verify the linearity and
reproducibility of the flow rate at system pressure, Waters qualification tests
measure the retention time, not volume. Such a test better suits
measurements of UPLC systems at multiple flow rates. It also better suits
those where the measured flow rates exceed 1 mL/min, and system
backpressure can exceed 68,948 kPa (689 bar, 10,000 psi).
®
Connections INSIGHT and ACQUITY UPLC Console diagnostics
The ACQUITY UPLC Console software supports the operation of ACQUITY
UPLC system modules, including the binary solvent manager. The ACQUITY
Console, a graphical, real-time software interface, monitors system status,
directly controls the system, troubleshoots errors, and performs automated
maintenance functions. For example, via the ACQUITY Console, you can
monitor the change in system backpressure as a measure of pump
performance. Monitoring each pump head, you can isolate a failure, thus
identifying a particular failed check valve or seal. You can then select the
appropriate maintenance option for replacing the failed component.
The ACQUITY Console is also the interface for the Waters Connections
INSIGHT remote service and alert capabilities. Through Connections
Overview 5
INSIGHT, Waters Service professionals can instantaneously respond to any
alarms or errors raised by the ACQUITY UPLC system components. The
software also provides the interface for particular functions, like generating a
service profile, that you can send directly to the Waters Global Service and
Support organization for analysis and interpretation.
Bottle tray
Detector
Column heater
Sample manager -
flow through needle
Binary solvent
manager
6
Major components
The following diagram shows the binary solvent manager’s major components.
Accumulator Pressure transducer Primary i2Valve cable
Check valve (2) pump A cable connector (4) pump B connector (2)
Check valve
in-line filter (2)
Accumulator
Primary pump B
pump A
i2Valve
actuator B
2
i Valve
actuator A Six-channel
degasser
Solvent chamber
selection
valve (2) Seal-wash
waste tubing
Pump drip
tray waste
line Leak sensor
Drip tray
Mixer/filter Drain tubing
Seal-wash Vent valve (to waste)
pump Solvent vent
TP03402
tubing
Leak sensor Degasser vent
connector tubing
Component Description
Accumulator pump (A Receives solvent from the primary pump and
and B) delivers it to the system.
Check valve A dual ball check valve that allows flow in only one
direction.
Check valve in-line Removes particulates from the incoming solvent.
filter
Degasser vent tubing Vents exhaust from the degasser pump.
Overview 7
Binary solvent manager’s major components: (Continued)
Component Description
Drain tubing (to Routes waste from the leak management system to
waste) the waste container.
Drip tray Manages fluid leaks.
2
i Valve An electronically controlled ball check valve that
allows flow in only one direction.
Leak sensor Continuously monitors the binary solvent manager
for leaks and stops the system flow when its optical
sensor detects about 1.5 mL of accumulated, leaked
liquid in its surrounding reservoir.
Mixer/filter Mixes and filters the solvent before it reaches the
sample manager. Mixer/filter sizes include 50 L,
100 L, and 380 L.
6-channel degasser Removes dissolved gasses from mobile-phase
chamber solvents and exhausts them, and any condensates,
through waste tubing.
Note: Vacuum degassing can change the
composition of mixed solvents.
Pressure transducer The electrical connections for the pressure
cable connectors transducers located on the front of the actuators.
Primary pump (A and Draws solvent, transferring it to the accumulator
B) pump and system as part of the serial flow design.
Pump drip tray waste Directs pump waste to the drip tray.
line
Seal-wash pump The pump that circulates solvent, to keep the
actuator’s high-pressure seals and plungers free of
contaminants.
Seal-wash waste Directs seal-wash waste to the drip tray.
tubing
Solvent selection Allows the user to select one solvent from A1 or A2
valve and one solvent from B1 or B2.
Solvent vent tubing Tubing that vents solvent to waste during priming.
8
Binary solvent manager’s major components: (Continued)
Component Description
Vent valve A 3-position valve that automatically switches to the
waste/vent position during priming or column
switching, the block position during the binary
solvent manager maintenance leak test, and the
system position during normal operation (or during
the system leak test).
Sample
manager
Vent valve
Solvent selection valves
Seal-wash
pump Degasser
Overview 9
Solvent flow sequence:
1. The solvent is stored in solvent reservoirs A1, A2, B1, and B2.
2. The in-line vacuum degasser degasses the solvent.
3. The solvent selection valve selects solvents for pump A and pump B from
solvent reservoirs A1, A2, B1, and B2.
2
4. The solvents flow through the i Valve, a check valve, which is located on
the inlet of the primary pump’s actuator.
5. The primary piston fills the accumulator pump and delivers solvent to
the system.
6. The accumulator pump’s piston delivers solvent, under pressure, to the
vent valve.
7. From the vent valve, the solvent flows to the in-line mixer, and then to
the sample management system.
10
Installing the waste and degasser vent tubing
Required materials
• Degasser vent tubing
• Gloves: clean, powder-free, chemical-resistant
• Waste tubing
Caution: To avoid distorting the drip tray or causing the drain cup
to leak, restrain the drain cup when attaching or removing the
waste tubing.
1. Wet the barbed drain fitting located at the bottom of the binary solvent
manager with methanol.
2. Hold the back of the drain cup, slide the waste tubing over the barbed
drain fitting, and route the tubing to a suitable waste container.
Drip tray
Solvent tray
Waste tubing
TP03416
12
5. Route the degasser vent tubing to a suitable waste container.
Degasser
vent tubing
Degasser
vent tubing
The solvent tray located on top of the system holds up to 2 L of spilled solvent.
You need a suitable waste container to collect any spill from the waste line at
the rear of the tray.
1. Choose solvent reservoirs that snugly fit the reservoir caps supplied in
the startup kit.
Requirement: Do not pressurize solvent reservoirs.
Recommendation: Use 1-L solvent reservoirs.
2. Remove the solvent filter from the short piece of fluoropolymer tubing.
Solvent filter
Fluoropolymer tubing
Solvent tubing
TP02484
14
5. Insert the filtered end of the solvent tubing into the solvent bottle, in the
tray atop the sample manager or (optional) detector.
Solvent tubing
Solvent bottles
Solvent tray
6. Repeat step 2 through step 5, for the remaining solvent supply lines.
Caution:
• To avoid damaging the solenoid valve seats and seals of the seal-wash
pump in the solvent path, do not use a nonvolatile buffer as the
seal-wash solvent.
• To prevent contaminating the system, do not recycle seal-wash.
Tip: The seal-wash self-primes, but you can use a syringe to hasten the
process.
Required materials
• 30-mL syringe (startup kit)
• Gloves: clean, powder-free, chemical-resistant
• Seal-wash solution
• Tubing adapter (startup kit)
Seal-wash
outlet tube
Drip tray
16
5. In the ACQUITY UPLC Console, select Binary Solvent Manager from
the system tree.
Note: If your system is controlled by UNIFI software, consult the UNIFI
online Help.
6. Click Control > Prime seal-wash, and then click Yes to begin the
seal-wash priming process.
7. Slowly withdraw on the syringe plunger, pulling seal-wash solvent
through the system.
8. When the seal-wash solution begins to flow into the syringe without
bubbles, disconnect the tubing and reinstall it on the fitting on the drip
tray.
9. Click Control > Prime seal-wash, and then click Yes to stop the priming
process.
Ensure the solvent reservoirs contain enough solvent for adequate priming
and the waste container has sufficient capacity for used solvent. The priming
flow rate is 4 mL/min for each pump, or 8 mL/min total. For example, priming
both solvents for 5 minutes requires approximately 20 mL of each solvent.
18
To prime a dry binary solvent manager using a syringe:
1. Open the instrument’s front door.
2. Locate the appropriate solvent vent tube.
• If you are priming solvent A, follow the stainless steel vent tube
labeled “A-VENT” from port 4 on the vent valve, and lift it from the
drip tray.
• If you are priming solvent B, follow the stainless steel vent tube
labeled “B-VENT” from port 1 on the vent valve, and lift it from the
drip tray.
Vent valve
20
2. In the Start up system dialog box, review the settings and select a
different option, if needed.
• Solvent line A only (default)
• Solvent line B only
• Both A and B
3. Click OK.
Result: The system primes the selected solvents, primes the sample
manager with one weak-wash prime (using the wash and sample
syringes), and ignites the lamp in the detector.
The system uses gold-plated compression screws and two-piece ferrules. See
the diagram below for assembly orientation.
Ferrule with
locking ring Compression screw Tubing
Recommendations:
• To prevent bandspreading, ensure the tubing fully bottoms in the fitting
hole before tightening the compression screw.
• For easier accessibility, use long compression screws to attach tubes to
the vent valve.
• Perform the solvent manager leak test whenever you replace or loosen
fittings during maintenance (see the ACQUITY UPLC online Help).
• Whenever you loosen fittings during maintenance, examine for cracks,
stripped threads, and deformations.
• Do not reseat stainless steel ferrules more than six times.
Tip: To avoid problems with the leak management system, do not use long
compression screws on the injector valve.
Required material
Gloves: clean, powder-free, chemical-resistant
22
When tightening system fittings, consult the following table.
(OD) tubing
End of lock ring with smaller
inside diameter (ID)
TP03377
TP03379
3/4-turn
TP03378
1/6-turn
TP02713
3/4-turn
TP03375
24
Installation recommendations for ACQUITY UPLC binary solvent manager
fittings: (Continued)
1/6-turn
TP03374
26
5. Turn the vent tube retainer clockwise, and then lift the A-vent and
B-vent tubes from the drip tray by pulling up on them and moving them
to the left-hand side of the leak sensor.
Leak sensor
TP03421
6. Remove the leak sensor from its reservoir by grasping the sensor by its
serrations and pulling upward.
Tip: If you cannot easily manipulate the leak sensor after removing it
from its reservoir, detach the connector from the front of the instrument
(see page 36).
Serrations
Prism
Lint-free wipe
TP02891
8. Roll up a nonabrasive, lint-free wipe, and use it to absorb the liquid from
the leak sensor reservoir and its surrounding area.
Leak sensor
reservoir
Rolled up
lint-free wipe
28
9. With a cotton swab, absorb any remaining liquid from the corners of the
leak sensor reservoir and its surrounding area.
Leak sensor
reservoir
Cotton swab
10. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
TP02892
TP02889
T-bar
Slot in leak
sensor reservoir
Leak sensor
installed in
reservoir
TP03422
Leak sensor
TP03421
13. If you detached the connector from the front of the instrument, reattach
it.
14. Power-on the binary solvent manager.
15. In the ACQUITY UPLC Console, select Binary Solvent Manager from
the system tree.
Note: If your system is controlled by UNIFI software, consult the UNIFI
online Help.
16. In the binary solvent manager information window, click Control >
Reset BSM, to reset the binary solvent manager.
30
Maintaining the binary solvent manager
Perform the procedures in this section when you discover a problem with a
binary solvent manager component or during routine maintenance. For
information about isolating problems in the binary solvent manager, consult
the ACQUITY UPLC Console online Help.
Maintenance schedule
Perform the following routine maintenance on the binary solvent manager to
ensure reliable operation and accurate results. When using the system
throughout the day (and on nights and weekends), or when using aggressive
solvents such as buffers, perform these maintenance tasks more frequently.
32
Recommended routine maintenance schedule:
34
Replacing the leak sensor
Required materials
• Gloves: clean, powder-free, chemical-resistant
• Leak sensor
Leak sensor
connector
4. Turn the vent tube retainer clockwise, and then lift the A-vent and
B-vent tubes from the drip tray by pulling upward on them and moving
them to the left-hand side of the leak sensor.
Leak sensor
TP03421
36
5. Remove the leak sensor from its reservoir by grasping the sensor by its
serrations and pulling upward.
Serrations
TP02892
TP02889
T-bar
Slot in leak
sensor reservoir
Leak sensor
installed in
reservoir
TP03422
Leak sensor
TP03421
10. Connect the leak sensor connector to the front of the instrument.
11. Power-on the binary solvent manager.
12. In the ACQUITY UPLC Console, select Binary Solvent Manager from
the system tree.
Note: If your system is controlled by UNIFI software, consult the UNIFI
online Help.
13. In the binary solvent manager information window, click Control >
Reset BSM, to reset the binary solvent manager.
38
Replacing the mixer
Required materials
• Gloves: clean, powder-free, chemical-resistant
• Mixer
Required tools
• 1/4-inch open-end wrench
• 5/8-inch open-end wrench
Mixer
Outlet
compression fitting
5/8-inch
open-end wrench
1/4-inch
open-end wrench
4. Using the 5/8-inch open-end wrench to hold the mixer, disconnect the
two inlet compression fittings by using the 1/4-inch wrench.
Mixer
Inlet compression 5/8-inch
fittings open-end wrench
1/4-inch
open-end wrench
40
Replacing the i2Valve actuator
Caution:
2
• To avoid damaging the i Valve actuator, do not attempt to push or
pull liquid or gas through the valve's inlet or outlet ports.
• To prevent contaminating system components, wear clean,
2
chemical-resistant, powder-free gloves when replacing the i Valve
actuator.
Required materials
• Gloves: clean, powder-free, chemical-resistant
2
• i Valve actuator
• i2Valve cartridge (recommended)
Required tools
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
®
• T8 TORX driver
3. Move the solvent bottles to a location below the binary solvent manager.
Connector receptacle
i2Valve connector
Knurled diameter
42
5. Loosen the cap nut on the in-line filter, so that it is removed from the
threads of the ferrule holder fitting.
Cap nut
TP03380
Ferrule holder
fitting
Cap nut
44
Caution:
• To avoid leaks, ensure the PEEK washer, which is normally on
2
the top face of the i Valve cartridge, does not remain in the head
when you remove the valve assembly (see page 59).
• To avoid failure of the fuse on the i2Valve actuator’s PCB board,
never place the actuator assembly or electrical connector in the
drip tray.
7. Remove the i2Valve actuator housing from the bottom of the primary
pump head.
Bottom of primary
pump head
i2Valve actuator
TP03382
TP02960
9. Use the T8 TORX driver to loosen 1/2-turn the 4 screws that secure the
clamping plates.
10. Ensure the shell nut remains free to rotate and that the plates slide
open.
Tips:
• Avoid touching the clamping plate tabs when loosening the screws.
• You can rotate the shell nut to gain access to all 4 screws.
Clamping
plate tab (2)
Shell nut
46
11. When both plates are in the maximum open position, remove the
2
cartridge from the i Valve actuator, and ensure that the low-pressure
gasket is removed with the cartridge.
Tip: If you cannot remove the cartridge from the valve actuator, rotate
the cartridge 1/2-turn, and then remove it.
12. Ensure the PEEK washer is inserted into the cartridge, its chamfered
edge facing away from the cartridge.
Chamfered edge PEEK washer
facing away from
cartridge
i2Valve cartridge
Low-pressure gasket
13. Ensure the plates are loose and in their maximum open position.
Tips:
• Avoid touching the clamping plate tabs when loosening the screws.
• You can rotate the shell nut to gain access to all 4 screws.
Grooved end of
cartridge
i2Valve actuator
2
15. With one hand, squeeze the 2 clamping plate tabs on the i Valve
actuator, to hold the clamping plates against the cartridge.
Requirement: The clamping plates must be fully engaged in the
cartridge groove.
Clamping Clamping
plate tab plate tab
Clamping
plates
48
16. While squeezing the clamping plate tabs, use the T8 TORX driver to
tighten the 4 screws that secure the plates, repeating the torquing
pattern shown below at least 3 times and gradually increasing the
torque until the screws are uniformly tight.
Start here
Tip: You can rotate the shell nut to gain access to all 4 screws.
Clamping
plate tab (2)
Shell nut
18. Orient the i2Valve assembly so that the cable exits from the left-hand
side.
19. Insert the i2Valve assembly into the bottom of the primary pump head,
and route the cable behind the valve actuator.
20. Finger-tighten the shell nut, rotating it approximately 5 full turns, to
secure the valve.
21. Use the 5/16-inch open-end wrench to tighten the nut an additional
1/8-turn.
50
22. Reinsert the in-line filter and tube into the ferrule holder fitting.
Ferrule holder
fitting
In-line filter
23. Place the cap nut over the ferrule holder fitting and finger-tighten the
cap nut to the extent possible.
Ferrule holder
fitting
Cap nut
Connector receptacle
i2Valve connector
Knurled diameter
52
Replacing the i2Valve cartridge
Caution:
2
• To avoid damaging the i Valve actuator, do not attempt to push or
pull liquid or gas through the valve's inlet or outlet ports.
• To prevent contaminating system components, wear clean,
2
chemical-resistant, powder-free gloves when replacing the i Valve
cartridge.
Required materials
• Gloves: clean, powder-free, chemical-resistant
2
• i Valve cartridge
Required tools
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
• T8 TORX driver
2
To replace the i Valve cartridge:
3. Move the solvent bottles to a location below the binary solvent manager.
4. Grasp the i2Valve connector by the knurled diameter, and pull it toward
you, disconnecting it from its receptacle.
Connector receptacle
i2Valve connector
Knurled diameter
54
5. Loosen the cap nut on the in-line filter so that it is removed from the
threads of the ferrule holder fitting.
Cap nut
TP03380
Ferrule holder
fitting
Cap nut
56
Caution:
• To avoid leaks, ensure the PEEK washer, which is normally on
2
the top face of the i Valve cartridge, does not remain in the head
when you remove the valve assembly (see page 59).
• To avoid failure of the fuse on the i2Valve actuator’s PCB board,
never place the actuator assembly or electrical connector in the
drip tray.
7. Remove the i2Valve actuator from the bottom of the primary pump head.
Bottom of primary
pump head
i2Valve actuator
TP03382
TP02960
2
Caution: To avoid damaging the i Valve actuator, do not back the
screws out all the way.
9. Use the T8 TORX driver to loosen 1/2-turn the 4 screws that secure the
clamping plates.
10. Ensure the shell nut remains free to rotate and that the plates slide
open.
Tips:
• Avoid touching the clamping plate tabs when loosening the screws.
• You can rotate the shell nut to gain access to all 4 screws.
Clamping
plate tab (2)
Shell nut
58
11. When both plates are in the maximum open position, remove the
2
cartridge from the i Valve actuator.
Tip: If you cannot remove the cartridge from the valve actuator, rotate
the cartridge 1/2-turn, and then remove it.
12. Unpack the new cartridge.
13. Ensure the PEEK washer is inserted into the cartridge, its chamfered
edge facing away from the cartridge.
Chamfered edge PEEK washer
facing away from
cartridge
2
i Valve cartridge
Grooved end of
cartridge
i2Valve actuator
15. With one hand, squeeze the 2 clamping plate tabs on the i2Valve
actuator, to hold the clamping plates against the cartridge.
Requirement: The clamping plates must be fully engaged in the
cartridge groove.
Clamping Clamping
plate tab plate tab
Clamping
plates
16. While squeezing the clamping plate tabs, use the T8 TORX driver to
tighten the 4 screws that secure the plates, repeating the torquing
60
pattern shown below at least 3 times and gradually increasing the
torque until the screws are uniformly tight.
Start here
Tip: You can rotate the shell nut to gain access to all 4 screws.
Clamping
plate tab (2)
Shell nut
2
17. Insert the ferrule holder fitting into the inlet port on the i Valve
assembly, and tighten it finger-tight.
Ferrule holder
fitting
62
22. Reinsert the in-line filter and tube into the ferrule holder fitting.
Ferrule holder
fitting
In-line filter
23. Place the cap nut over the ferrule holder fitting and finger-tighten the
cap nut to the extent possible.
Ferrule holder
fitting
Cap nut
Connector receptacle
i2Valve connector
Knurled diameter
64
Replacing the in-line filter cartridge on the i2Valve actuator
Required materials
• Gloves: clean, powder-free, chemical-resistant
• In-line filter cartridge
3. Move the solvent bottles to a location below the binary solvent manager.
Cap nut
i2Valve actuator
5. Pull the cap nut off the tube to remove the in-line filter cartridge.
Cap nut
Tube
In-line filter
cartridge
66
In-line filter cartridge and cap nut:
6. Put the cap nut over the end of the tube (see the figure on page 68).
7. Slide the metal locking ring onto the tube, ensuring that the thicker end
of the metal locking ring is facing toward the cap nut.
Cross-sectional view
9. Insert the in-line filter cartridge with tubing into the ferrule holder
fitting.
Ferrule holder
fitting
In-line filter
10. Screw the cap nut onto the ferrule holder fitting, ensuring that the tube
is bottomed out in the in-line filter cartridge. Finger-tighten it to the
extent possible.
11. Return the solvent bottles to their original location.
12. Power-on the binary solvent manager.
13. Prime the binary solvent manager (see page 17).
68
Replacing the accumulator check valve
Required materials
• Accumulator check-valve assembly
• Gloves: clean, powder-free, chemical-resistant
Required tools
• 1/2-inch open-end wrench
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
3. Move the solvent bottles to a location below the binary solvent manager.
Compression fitting
5. Use the 1/2-inch open-end wrench to loosen the check valve, and then
remove the check-valve assembly from the head.
70
6. Unpack the new check valve.
7. Ensure the new PEEK washer is inserted into the new check valve, its
chamfered edge facing away from the check valve.
Chamfered edge PEEK washer
facing away from
check valve
Check valve
8. Insert the check-valve assembly into the head, and use the 1/2-inch
wrench to tighten the check-valve nut 1/8-turn beyond finger-tight.
9. Finger-tighten the compression screw and ferrule, and then, using the
5/16-inch open-end wrench to hold the check valve in place, use the
1/4-inch wrench to tighten the compression screw up to 1/6-turn beyond
finger-tight for an existing stainless steel tubing assembly, or 3/4-turn
beyond finger-tight for a new stainless steel tubing assembly.
10. Return the solvent bottles to their original location.
11. Power-on the binary solvent manager.
12. Prime the binary solvent manager (see page 17).
Required materials
• Gloves: clean, powder-free, chemical-resistant
• New solvent filter
Solvent filter
Fluoropolymer tubing
Solvent tubing
4. Insert the filtered end of the solvent tubing into the solvent bottle.
5. Shake the filter to remove any air from it.
6. Prime the binary solvent manager (see page 17).
72
Cleaning the air filters in the binary solvent manager door
Required material
Mild detergent and water
Required tool
T10 TORX driver
Screw (8)
Air filter
Air filter
Frame
Frame
TP02429
Required material
Binary solvent manager air filters
Required material
T10 TORX driver
Screw (8)
Air filter
Air filter
Frame
Frame
TP02429
2. Remove the old air filters from the air filter frames, and discard them.
3. Align the new air filters with the air filter frames.
4. Attach the air filters and frames to the inside of the binary solvent
manager door, using the T10 TORX driver to secure the 8 screws.
74
Replacing the primary head plunger and seals
See the ACQUITY UPLC online Help to help determine whether you need to
replace the primary plunger seals.
Required materials
• Compressed air
• Gloves: clean, powder-free, chemical-resistant
• Methanol
• Plunger (recommended)
• Plunger seal and plunger seal spacer
• Fluoropolymer O-ring
• Seal-wash seal
Required tools
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
• Pliers
• Plunger removal tool (recommended)
• Seal extraction tool
• Sharp tool
• T27 TORX driver (startup kit)
7. Move the solvent bottles to a location below the binary solvent manager.
76
Caution: To avoid damaging the connector or cable, grasp the
i2Valve connector by the knurled diameter.
8. Grasp the i2Valve connector by the knurled diameter and pull it toward
you, disconnecting it from the receptacle.
Connector receptacle
i2Valve connector
Knurled diameter
Cap nut
TP03380
Ferrule holder
fitting
Cap nut
78
10. Using the 5/16-inch open-end wrench, loosen the shell nut, and then
fully unscrew it.
11. Remove the i2Valve actuator from the bottom of the primary pump head,
ensuring that the PEEK washer remains in the cartridge.
PEEK washer
Cartridge
i2Valve actuator
TP03382
80
12. Remove the seal-wash tubing secured to the seal-wash housing by
barbed fittings by using a tool or by pulling on the tubing as close to the
head as possible.
Seal-wash tubing
Seal-wash tubing
13. Using a pliers, remove the drip wire from the head assembly.
Drip wire
Pliers
Transducer
Outlet tubing
Pressure transducer
cable connector
Tabs
82
16. Using the T27 TORX driver, loosen the 2 head bolts 1/2-turn.
Tip: The bolts are accessible from the front of the pressure transducer.
17. Using the T27 TORX driver, loosen and remove the 2 support plate
bolts, and then gently pull the head and support plate off the actuator
housing, making sure not to tilt the head during the extraction.
Plunger
84
Warning: To avoid hand lacerations, use care when removing the
old plunger. Bending the plunger shaft can cause it to break.
18. Use the recessed side of the plunger removal tool to apply pressure to
both sides of the release collar, and then remove the old plunger.
TP02990
Spring-loaded
release collar
Plunger
19. Remove the plunger removal tool from the release collar.
Pump head
Seal-wash housing
Support plate
86
4. Lift the seal-wash housing from the support plate.
Seal-wash housing
Support plate
Plunger seals:
Head
Fluoropolymer
O-ring
Seal-wash
housing
Seal-wash seal
Plunger seal
5. Using the smooth end of the seal extraction tool, pull the seal-wash seal
from the seal-wash housing.
Seal-wash seal
6. Taking care not to scratch any surfaces, screw the threaded end of the
seal extraction tool into the plunger seal on the reverse side of the seal-
wash housing and carefully withdraw the plunger seal and spacer.
Plunger seal
7. Taking care not to scratch any surfaces, use a sharp tool to remove the
fluoropolymer O-ring from the pump head.
Sharp tool
Fluoropolymer O-ring
88
8. Inspect the seal-wash housing surface and pump head surface, ensuring
both are free from scratches and particulates.
9. Lubricate the new fluoropolymer O-ring with methanol, and press the
O-ring into its seat with your thumbs.
10. Lubricate the new plunger seal spacer with methanol, and use the
smooth end of the seal extraction tool to place it in the seal-wash
housing.
TP
11. Spray the new plunger seal with compressed air to remove any
particulates.
Plunger seal
13. Orient the seal-wash housing so that the holes on its side align with the
holes on the side of the pump head, and then guide it into place.
Seal-wash housing
Pump head
90
14. Spray the seal-wash seal with compressed air to remove any
particulates.
15. Lubricate the new seal-wash seal with methanol, place it in the
seal-wash housing, and press it into place.
Seal-wash seal
16. Place the support plate on top of the pump head, ensuring the round side
of the plate is oriented toward the bottom side of the head.
Round side of
support plate
Bottom side of
pump head
Seals
2. Carefully insert the sapphire plunger shaft into the pump head until the
plunger shaft is no longer visible, ensuring the shaft does not contact the
support plate.
92
Recommendation: Replace the plunger whenever you replace the
plunger seal.
Plunger shaft
Support plate
3. Carefully slide the head assembly and sapphire plunger into the
actuator housing, making sure not to tilt the head.
4. Connect the pressure transducer cable to the bulkhead.
Pressure transducer
cable connector
10. Hold the head assembly securely against the actuator housing, and then
use the T27 TORX driver to tighten the support plate bolts securely.
11. Use the torque driver to tighten each pump head bolt to 40 inch-pounds
of torque.
12. After tightening each pump head bolt to 40 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 50 inch-pounds of
torque.
13. After tightening each pump head bolt to 50 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 60 inch-pounds of
torque.
14. After tightening each pump head bolt to 60 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 70 inch-pounds of
torque.
94
Caution: To avoid pinching the drip wire between the head
assembly and support plate, be sure to install the drip wire after
tightening the head bolts.
15. Reinstall the drip wire around the head assembly, ensuring that the tip
is in the 6 o’clock position.
16. Reattach all fittings and seal-wash tubing.
Tip: When reattaching the outlet tubing to the transducer, tighten the
inlet-tubing fitting finger-tight plus as much as 1/6-turn, for existing
fittings, or 3/4-turn for new fittings.
2
17. Orient the i Valve assembly so that the cable exits from the left-hand
side.
2
18. Insert the i Valve assembly into the bottom of the primary pump head,
and route the cable behind the valve actuator.
19. Finger-tighten the shell nut, rotating it approximately 5 full turns, to
secure the valve.
20. Use the 5/16-inch open-end wrench to tighten the nut an additional
1/8-turn.
Ferrule holder
fitting
In-line filter
22. Place the cap nut over the ferrule holder fitting, and finger-tighten the
cap nut to the extent possible.
Ferrule holder
fitting
Cap nut
96
2
23. Align the red dot on the i Valve connector with the red dot on the
receptacle, in the 12 o’clock position, and insert the connector into the
receptacle.
Red dot on connector
Connector receptacle
i2Valve connector
Knurled diameter
Required materials
• Compressed air
• Gloves: clean, powder-free, chemical-resistant
• Methanol
• Plunger (recommended)
• Plunger seal and plunger seal spacer
• Fluoropolymer O-ring
• Seal-wash seal
Required tools
• 1/4-inch open-end wrench
• 5/16-inch open-end wrench
• Pliers
• Plunger removal tool (recommended)
• Seal extraction tool
• Sharp tool
• T27 TORX driver (startup kit)
98
To remove the accumulator head:
1. Flush the binary solvent manager with nonhazardous solvent.
2. In the ACQUITY UPLC Console, select Binary Solvent Manager from
the system tree.
Note: If your system is controlled by UNIFI software, consult the UNIFI
online Help.
3. In the binary solvent manager information window, click Maintain >
Heads.
4. In the Head Maintenance dialog box, select the accumulator head (A or
B).
5. Click Move Backward, and then wait for the plunger to stop.
7. Move the solvent bottles to a location below the binary solvent manager.
Seal-wash tubing
Seal-wash tubing
9. Using a pliers, remove the drip wire from the head assembly.
Drip wire
Pliers
100
10. Using the 1/4-inch open-end wrench, disconnect the outlet tubing from
the transducer.
Transducer
Outlet tubing
11. Using the 5/16-inch open-end wrench to hold the check-valve cartridge
in place, disconnect the tubing connection from the check valve with the
1/4-inch open-end wrench.
Compression fitting
Pressure transducer
cable connector
Tabs
13. Using the T27 TORX driver, loosen the 2 head bolts 1/2-turn.
Tip: The bolts are accessible from the front of the pressure transducer.
102
Caution: To avoid damaging the plunger, support the pump head
from below as you remove it.
14. Using the T27 TORX driver, loosen and remove the 2 support plate
bolts, and then gently pull the head and support plate off the actuator
housing, making sure not to tilt the head during the extraction.
Plunger
15. Use the recessed side of the plunger removal tool to apply pressure to
both sides of the release collar, and then remove the old plunger.
TP02990
Spring-loaded
release collar
Plunger
16. Remove the plunger removal tool from the release collar.
104
To remove the accumulator plunger seals:
1. Stand the head upright on a clean surface.
2. Using the T27 TORX driver, completely loosen the 2 head bolts to
release the support plate from the pump head.
Requirement: If you remove the transducer and head bolts, be sure to
reuse the head bolt washers when reassembling the pump head.
Pump head
Seal-wash housing
Support plate
Seal-wash housing
Support plate
Plunger seals:
Head
Fluoropolymer
O-ring
Seal-wash
housing
Seal-wash seal
Plunger seal
5. Using the smooth end of the seal extraction tool, remove the plunger seal
spacer from the head.
Seal-wash seal
106
Caution: To avoid scratching any metal surfaces, use care when
screwing the threaded end of the seal extraction tool into the
plunger seal.
6. Taking care not to scratch any surfaces, screw the threaded end of the
seal extraction tool into the plunger seal and carefully withdraw the seal
from the head.
Plunger seal
TP03393
Sharp tool
Fluoropolymer O-ring
10. Spray the new plunger seal with compressed air to remove any
particulates.
11. Lubricate the new plunger seal with methanol, and use the smooth end
of the seal extraction tool to place it in the seal-wash housing, over the
plunger seal spacer.
Plunger seal
108
12. Orient the seal-wash housing so that the holes on its side align with the
holes on the side of the pump head, and then guide it into place.
Seal-wash housing
Pump head
13. Spray the seal-wash seal with compressed air to remove any
particulates.
14. Lubricate the new seal-wash seal with methanol, place it in the
seal-wash housing, and press it into place.
Seal-wash seal
Bottom side of
pump head
16. Holding the assembly together, use the T27 TORX driver to minimally
tighten the 2 head bolts.
110
To reattach the accumulator head:
1. Flip the assembly over, and then lubricate the seals with methanol.
Seals
2. Carefully insert the sapphire plunger shaft into the pump head until the
plunger shaft is no longer visible, ensuring the shaft does not contact the
support plate.
Recommendation: Replace the plunger whenever you replace the
plunger seal.
Plunger shaft
Support plate
3. Carefully slide the head assembly and sapphire plunger into the
actuator housing, making sure not to tilt the head.
4. Connect the pressure transducer cable to the bulkhead.
Pressure transducer
cable connector
112
Caution: To avoid damaging the plunger, alternately tighten the
the support plate screws 1/4-turn so that they are uniformly
torqued.
10. Hold the head assembly securely against the actuator housing, and then
use the T27 TORX driver to tighten the support plate bolts securely.
11. Use the torque driver to tighten each pump head bolt to 40 inch-pounds
of torque.
12. After tightening each pump head bolt to 40 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 50 inch-pounds of
torque.
13. After tightening each pump head bolt to 50 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 60 inch-pounds of
torque.
14. After tightening each pump head bolt to 60 inch-pounds of torque, use
the torque driver to tighten each pump head bolt to 70 inch-pounds of
torque.
15. Reinstall the drip wire around the head assembly, ensuring that the tip
is in the 6 o’clock position.
16. Reattach all fittings and seal-wash tubing.
Required materials
• Gloves: clean, powder-free, chemical-resistant
• Vent valve cartridge
Required tools
• 1/4-inch open-end wrench
• 2-mm Allen wrench (startup kit)
114
To replace the vent valve cartridge:
1. In the ACQUITY UPLC Console, select Binary Solvent Manager from
the system tree.
Note: If your system is controlled by UNIFI software, consult the UNIFI
online Help.
2. In the binary solvent manager information window, click Interactive
Display.
3. In the binary solvent manager interactive display dialog box, click
Control .
4. Ensure the vent valve is set to Vent.
Tip: To change the setting to Vent, click the underlined vent valve
position, and select Vent.
6. Use the 2-mm Allen wrench to remove the hex screw at the 10 o’clock
position on the vent valve cartridge.
7. Remove the vent valve cartridge from the vent valve assembly by pulling
straight forward.
8. Unpack the replacement vent valve cartridge.
116
Caution: To avoid scratching the drive clamp or cartridge housing,
handle it with care.
9. Ensure that the groove in the cartridge housing aligns with the groove
on the drive clamp.
Tip: If the grooves fail to align, turn the drive clamp until they do.
Vent valve cartridge housing
Aligned grooves
Drive clamp
10. Insert the new vent valve cartridge into the vent valve cartridge
chamber.
Requirements:
• Orient the new cartridge exactly as the old one was oriented.
• The vent valve cartridge must slide fully into the vent valve
assembly. If it does not, contact Waters Technical Service.
11. Insert the 2-mm hex screw at the 10 o’clock position on the vent valve
cartridge and then use the 2-mm Allen wrench to tighten the hex screw
to 9 inch-pounds of torque.
12. Use the 1/4-inch wrench to reattach all fittings and tighten them up to
1/6-turn beyond finger-tight for existing fittings, or 3/4-turn beyond
finger-tight for new fittings.
13. Prime the binary solvent manager (see page 17).
118