User S Manual: Rxfa11G Magnetic Flowmeter Remote Converter (Hardware Edition/Software Edition)
User S Manual: Rxfa11G Magnetic Flowmeter Remote Converter (Hardware Edition/Software Edition)
User S Manual: Rxfa11G Magnetic Flowmeter Remote Converter (Hardware Edition/Software Edition)
Magnetic
Flow Converter
IM 01R21C01-01E-E
Contents
1. INTRODUCTION ...................................................................................1-1
1.1 Using the Magnetic Flowmeter Safely.............................................................1-2
1.2 Warranty..............................................................................................................1-3
2. HANDLING PRECAUTIONS.................................................................2-1
2.1 Checking Model and Specifications...............................................................2-1
2.2 Accessories.......................................................................................................2-1
3. INSTALLATION......................................................................................3-1
3.1 Installation Location.........................................................................................3-1
3.2 Mounting............................................................................................................3-1
4. WIRING..................................................................................................4-1
4.1 Wiring Precautions............................................................................................4-1
4.2 Cables................................................................................................................4-1
4.4.6 Wiring the Remote Flowtube with the RXFA11 Converter................................. 4-6
All Rights Reserved. Copyright © 2006, Rota Yokogawa i IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
CONTENTS
5.3.2 Setting Example for Numeric-Type Data: Flow rate span.................................. 5-7
6. Parameter Description.........................................................................6-1
6.1 Parameters.........................................................................................................6-1
6.2 Parameter Lists.................................................................................................6-1
6.3 Parameter List Overview..................................................................................6-2
6.4 Parameter Description....................................................................................6-12
6.5 Alarm Functions..............................................................................................6-35
6.5.1 Alarm Levels......................................................................................................... 6-35
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 ii All Rights Reserved. Copyright © 2006, Rota Yokogawa
CONTENTS
8.4 Parameters.........................................................................................................8-3
9. Actual Operation...................................................................................9-1
9.1 Pre-operation Zero Adjustment......................................................................9-1
10. Maintenance......................................................................................10-1
10.1 Maintenance...................................................................................................10-1
10.3 Troubleshooting.............................................................................................10-3
All Rights Reserved. Copyright © 2006, Rota Yokogawa iii IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
CONTENTS
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 iv All Rights Reserved. Copyright © 2006, Rota Yokogawa
1. INTRODUCTION
1. INTRODUCTION
This instrument has been adjusted at the factory Safety and Modification Precautions
before shipment. • The following general safety precautions must
To ensure correct use of the instrument, please be observed during all phases of operation,
read this manual thoroughly and fully understand service, and repair of this instrument. Failure to
how to operate the instrument before operating it. comply with these precautions or with specific
WARNINGS given elsewhere in this manual vio-
lates safety standards of design, manufacture,
Regarding This User's Manual
and intended use of the instrument. Yokogawa
• This manual should be provided to the end assumes no liability for the customer's failure to
user. comply with these requirements. If this instru-
• Before use, read this manual thoroughly to ment is used in a manner not specified in this
comprehend its contents. manual, the protection provided by this instru-
• The contents of this manual may be changed ment may be impaired.
without prior notice. • Yokogawa will not be liable for malfunctions or
• All rights are reserved. No part of this manual damage resulting from any modification made
may be reproduced in any form without to this instrument by the customer.
Yokogawa's written permission. • The following safety symbol marks are used in
• Yokogawa makes no warranty of any kind with this user's manual and instrument.
regard to this material, including, but not limited
to, implied warranties of merchantability and
suitability for a particular purpose. WARNING
• All reasonable effort has been made to ensure A WARNING sign denotes a hazard. It calls
the accuracy of the contents of this manual. attention to procedure, practice, condition or the
However, if any errors or omissions are found, like, which, if not correctly performed or adhered
please inform Yokogawa.
to, could result in injury or death of personnel.
• Yokogawa assumes no responsibilities for this
product except as stated in the warranty.
• Please note that this user's manual may not be
revised for any specification changes, construc- CAUTION
tion changes or operating part changes that A CAUTION sign denotes a hazard. It calls atten-
are not considered to affect function or perfor- tion to procedure, practice, condition or the like,
mance. which, if not correctly performed or adhered to,
• If the customer or any third party is harmed by could result in damage to or destruction of part
the use of this product, Yokogawa assumes no or all of the product.
responsibility for any such harm owing to any
defects in the product which were not predict-
able, or for any indirect damages.
IMPORTANT
NOTE
A NOTE sign denotes information necessary for
essential understanding of operation and fea-
tures.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 1-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
1. INTRODUCTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 1-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
1. INTRODUCTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 1-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
1. INTRODUCTION
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 1-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
2. HANDLING PRECAUTIONS
2. HANDLING PRECAUTIONS
This instrument has been inspected carefully at
the factory before shipment. When the instrument
2.3 Storage Precautions
is delivered, visually check that no damage has If the instrument is to be stored for a long period
occurred during transportation. of time after delivery, observe the following points.
Read this section carefully as it contains important · The instrument should be stored in its original
information on handling this instrument. Refer to packing condition in the storage location.
the relevant sections for information not contained · Select a storage location that fulfils the follow-
in this section. If you have any problems or ques- ing conditions:
tions, please contact Yokogawa sales office. • A place where it will not be exposed to rain or
water
• A place subject to minimal vibrations or shocks
2.1 Checking Model and • Temperature and humidity levels should be as
Specifications follows:
Temperature: -30 to 70°C
The model code and specifications are found on Humidity: 5 to 80% RH (no condensation)
the data plate located on the outside of the case. The preferred ambient temperature and
Check that the model code and specifications humidity levels are 25°C and approxi-
match what you have ordered. mately 65% RH.
· If the RXFA11 is transferred to the installation
Be sure you have your model number and serial
site and stored without being installed, its per-
number available when contacting Yokogawa.
formance may be impaired due to the infiltration
of rainwater and so forth. Be sure to install
and wire the RXFA11 as soon as possible after
transferring it to the installation location.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 2-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
2. HANDLING PRECAUTIONS
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 2-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
3. INSTALLATION
3. INSTALLATION
WARNING
Installation of the magnetic flowmeter must be performed by expert engineers or skilled personnel. No
operator shall be permitted to perform procedures relating to installation.
Install the instrument in a location where it is not exposed to direct sunlight. For ambient temperature,
refer to Chapter 11 “OUTLINE”.
The instrument may be used in an ambient humidity where the RH ranges from 0 to 100 %. However,
avoid long-term continuous operation at relative humidity above 95 %.
3.2 Mounting
This instrument can be mounted using surface mounting, 2-inch pipe mounting, or
panel mounting.
(2.56)
65
194 (7.64)
F0301.EPS
IMPORTANT
Mounting fixture on equipment intended to be mounted on a wall or ceiling shall withstand a force of four
times the weight of the equipment (RXFA11: 3.3 kg (7.3 lb)).
All Rights Reserved. Copyright © 2006, Rota Yokogawa 3-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
3. INSTALLATION
2-inch pipe
Mounting fixture
Washer
Clamp screw
Pass the four clamp screws through the mounting fixture, position it on the 2-inch pipe,
and then fasten the RXFA11 in place.
F0302.EPS
Panel Mounting
Unit: mm
(approx. in)
Panel cutout
203 (8.0)
Mounting fixture
172 (6.8)
R3max
Washer
Screw
Clamp screw
Fit the RXFA11 into the panel. Then attach the mounting fixture to the RXFA11 using the screw and the
washer, and secure the instrument with the two clamp screws.
F0303.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 3-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
4. WIRING
This section describes the wiring on the • Ground the remote flowtube and the converter
converter side only. For information relating separately
to wiring on the flowtube side, refer to the • Cover each shield of the signal cable with vinyl
user’s manual of the RXF Remote Flowtube tube or vinyl tape to avoid contact between two
(IM 01R21D01-01E-E). shields or between a shield and a case.
• When waterproof cable glands or union
WARNING equipped waterproof glands are used, avoid
tightening the glands with an excessive torque.
The wiring of the magnetic flowmeter must be
• Be sure to turn power off before opening the
performed by expert engineer or skilled personnel.
cover.
No operator shall be permitted to perform proce-
• Before turning the power on, tighten the cover
dures relating to wiring.
securely.
CAUTION
IMPORTANT
Once all wiring is complete, double check the
Prepare the signal cable and the excitation cable
connections before applying power to the instru-
almost the same length. It is recommended to lay
ment. Improper arrangements or wiring may
them together closely.
cause a unit malfunction or damage.
CAUTION
Outer jacket 10.5 (0.41in)
• In cases where the ambient temperature ex-
ceeds 50 °C (122 °F), use external heat-resis-
tant wiring with a maximum allowable tempera- Shield (C)
ture of 70 °C (158 °F) or above. Insulation
Insulation
• Do not connect cables outdoors in wet weather F0401.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
4. WIRING
55 (2.17)
25 (0.98)
B
A
50 (1.97)
C
SB
90 (3.54)
90 (3.54)
C
70 (2.76)
Crimp Terminal
150 +5
150 +5
(5.9)
On the On the
(5.9)
converter flowtube
side side
EX1
EX2
EX2
EX1
L (Specified Dimensions)
85 (3.35) 85 (3.35)
f10.5 (0.4)
20 (0.8)
RXFC
F0402.EPS
Figure 4.2.2 Treatment of Dedicated Signal Cables On the converter side On the flowtube side
CAUTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 4-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
Plastic gland
F0406.EPS
Conduit Wiring
When wiring the conduits, pass the conduit
through the wiring connection port, and utilize the
waterproof gland to prevent water from flowing in.
Place the conduit pipe on an angle as shown in
Figure 4.3.5. Install a drain valve at the low end
of the vertical pipe, and open the valve regularly.
Drain valve
F0408.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
4. WIRING
F0409.EPS
F0410.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 4-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
420 (1380)
Functional
Power supply cable 100 ( 330)
grounding cable
0
20.4 22 24 26 28.8
Usable range E (V)
F0411.EPS Cable cross section area: 1.25 mm2
Cable cross section area: 2 mm2
Figure 4.4.3 Electric Cable Wiring F0411-2.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
4. WIRING
(3) Setting Power Supply Frequency 4.4.6 Wiring the Remote Flowtube with
the RXFA11 Converter
IMPORTANT
WARNING
• 4 to 20 mA DC Current Output
AXFA11
l+
Receiver
l- Instrument
Resistive Load Max. 1 kOhm
Protective grounding terminals F0414.EPS
Communication Resistive load
BRAIN 250 to 450 Ohm
HART 250 to 600 Ohm
Figure 4.4.4 Protective Grounding Terminal Location F0417.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 4-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
4. WIRING
• Pulse Output
RXFA11 Protective diode
P+
COM
Protective diode
RXFA11 P+
No-voltage status input
PULSE OUT Mechanical Counter F0420.EPS
P-
Figure 4.4.9 Status Input Connection
RXFA11 P+
• Status Output/ Alarm Output
PULSE OUT Load Electronic Counter
P-
30 V DC, 0.2 A max
IMPORTANT
F0418.EPS
Since this is an insulated transistor output, be
Figure 4.4.7 Pulse Output Connection careful of voltage and polarity when wiring.
Do not apply a voltage larger than 30 V DC or a
current larger than 0.2 A in order to prevent dam-
age to the instrument.
This output cannot switch an AC load. To switch
an AC load, an intermediate relay must be in-
serted as shown in Figure 4.4.11 or 4.4.12.
*The alarm output operates from closed (normal)
to open (alarm occurrence) in the default value
(as setup upon plant shipment). Changes can
be made via the parameter settings.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 4-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
4. WIRING
RXFA11
Protective diode
SO1+
Load
Load
COM
External power supply
30 V DC, 0.2 A max
RXFA11
SO1+ Relay
Protective
diode Electromagnetic
valve
F0421.EPS
Load
AL-
RXFA11 AL+
Relay
Electromagnetic
valve
AL-
External power supply AC power supply
30 V DC, 0.2 A max
F0422.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 4-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
WARNING
Be sure to set parameters as "Protect" on the write protect function after finish of
parameter setting work.
Under extremely rare case, the infra-red switches may respond unexpectedly in
such conditions as sticking ball of water or extraneous substances on the surface
of display panel glass according to the principle of infra-red switch operation. Its
probability rises in such cases as sticking rain water by storm or other similar
situation and washing up work near flowmeter installation place.
Either to illuminate or stop illuminating the infra-red switches by the flashlight may
cause the mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protection Items" and section "10.2.3"
how to use the write protect function in detail.
IMPORTANT
Operate the display unit under the condition where direct sunlight, etc... do not
shine to the setting switches directly when the parameter setting operation is
carried out.
NOTE
(1) Always use the setting switches with the cover of the RXFA11 closed.
(2) Use these switches with them covered by the glass window.
(3) If dirt, dust or other substances surfaces on the display panel glass, wipe
them clean with a soft dry cloth.
(4) The operation with dirty gloves may cause a switch response error.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
FR
m3/h
(2) Setting switch operations
0 : : : 100
3274m3 FTL
(2) Setting switches (Press the switch while holding down the switch)
(Infrared switches)
F0501.EPS : Move the cursor down (for selection-type parameters) or
increase values (for numeric parameters)
Actual instantaneous flow rate FR Displays the actual reading for instantaneous flow rate.
Instantaneous flow rate: mA FR Displays the instantaneous flow rate for the span as a current output value.
Bar graph indicating None Displays the instantaneous flow rate for the span using a percentage
instantaneous flow rate bar graph.
Totalized forward-direction flow rate FTL Displays the totalized value for flow rate in the forward direction.
Totalized reverse-direction flow rate RTL Displays the totalized value for flow rate in the reverse direction.
Totalized differential flow rate DTL Displays the totalized value for flow rate between forward
totalization and reverse totalization.
Diagnosis of electrode ADH Displays the adhesion condition in the form of a bar graph.
adhesion (See the description for parameters K10 through K15 from
Chapter 6: Parameter Description for more details.)
Communication COM Displays the communication type.
T0501.EPS
(4)
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 5-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
Before changing any settings, be sure to check the corresponding setting details
in Chapter 6: Parameter Description.
+ +
Select “No”
Setting modes
* The term "mode" is used to refer to a situation
where display and setting are possible.
+
Press the SET switch while holding down the SHIFT switch. F0502.EPS
Sample Display: Procedure for moving from Display Mode to Setting Mode
Start:
Display Mode 1st line: Actual instantaneous flow rate [FR]
FR
m3/h 2nd line: Bar graph indicating instantaneous flow
: : : 100
0
3274m3 FTL rate
Touch SHIFT + key When other operations are carried out, the
system will automatically return to the Display
Mode.
When no operations in this display are carried
out for 20 seconds, the system will automati-
cally return to the Display Mode.
NOTE (*1)
F0503-1.EPS
The Major Item Parameter which is set just
before will be shown when entrying the Setting
Mode again within 1 minute after returning from
Setting Mode to Display Mode.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
NOTE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 5-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
If no operations are carried out for a period of 10 min in this mode, the system will
automatically return to the Display Mode.
(i) Selection-type B21:Base Flow Unit The desired data item is selected from a predefined list.
m3
m3
kl(Kiloliter)
(ii) Numeric-type B23:Flow Span Data is set using the values in each digit and using
100 l/min the decimal point.
000100 l/min
Rng:0.00001 32000
Regarding the alphanumeric-type format (iii), the following alphanumerics are displayed in the following sequence:
Setting Mode Setting mode Major Item Parameter Search Mode has been
Start:
Major Item Major item
P:Protect accessed in this screen.
B:Easy Setup
Parameter parameter C:Basic Setup
Search Mode Touch the switch to access B: Easy
Setup.
F0504-1.EPS
NOTE
X2
When l (Liter) has been selected, touch the
switch.
NOTE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 5-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
Setting mode
Setting Mode Condition
Start: Setting mode
P:Protect
Major Item Major item B:Easy Setup Touch the switch to access B: Easy Setup.
Parameter parameter* C:Basic Setup
Search Mode NOTE
Sub-item Major item B:Easy Setup Sub-Item Parameter Search Mode has been ac-
50:Auto Zero Exe
Parameter Sub-item 10:Language cessed in this screen.
Search Mode parameter 20:Flow Damping
NOTE
F0505-1.EPS
When the value of 120 has been setup, touch
the switch.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
B23:Flow Span
120 l/min When the switch is touched, the entire
display will flash on and off. Confirm that the
setting has been correctly changed to 120, and
then fix this value by touching the switch
once again.
NOTE
F0506-1.EPS
NOTE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 5-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
5. BASIC OPERATING PROCEDURES
switch.
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 5-9 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
5. BASIC OPERATING PROCEDURES
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 5-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
6. PARAMETER DESCRIPTION
6.1 Parameters NOTE
With the exception of parameters that were In order to ensure that correct flow-rate data can
specified by the customer upon ordering, all of the be acquired, it is crucial that the nominal size,
RXFA11’s internal parameters will initially be set to flow rate span and meter factor of the combined
default values. Actions such as the modification of remote flowtube are setup. In cases where a
display details can then be carried out whenever remote flowtube is ordered at the same time as
necessary. the RXFA11, the nominal size and meter factor
will be setup upon shipment from the manufac-
IMPORTANT turing plant, and these will not require additional
setting. If the RXFA11 is ordered individually, the
Make sure to keep the RXFA11’s power on at default value will be setup for the meter fac-
least for 30 s after you set the parameters. If you tor; accordingly, it will be necessary to change
turn the power off immediately after the param- this setting to the meter factor indicated on your
eters are set, the settings will be canceled. remote flowtube data plate. If a flow rate span is
specified upon ordering, this will be set before
shipment. If this is not the case, however, it will
be necessary for the appropriate value to be set
up by the user.
R: Read only
W: Writable
Parameter number.
T0601.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
G42 Alm Record Time1 R 0D 00:00 to Displays the operation time at the
(ALM TIME 1) 99999D 23:59 occurrence of the most recent alarm.
G43 Alm Record2 R See G41 Displays the content of the second
(ALM RECORD2) (ALM RECORD1) most recent alarm.
G44 Alm Record Time2 R 0D 00:00 to Displays the operation time at occurrence
(ALM TIME 2) 99999D 23:59 of the second most recent alarm.
G45 Alm Record3 R See G41 Displays the content of the third most
(ALM RECORD3) (ALM RECORD1) recent alarm.
G46 Alm Record Time3 R 0D 00:00 to Displays the operation time at the
(ALM TIME 3) 99999D 23:59 occurrence of the third most recent alarm.
G47 Alm Record4 R See G41 Displays the content of the fourth
(ALM RECORD4) (ALM RECORD1) most recent alarm.
G48 Alm Record Time4 R 0D 00:00 to Displays the operation time at the
(ALM TIME 4) 99999D 23:59 occurrence of the fourth most recent alarm.
See “6.5 Alarm Functions”.
G60 — R Good
(SELF CHECK) Error
T0608-2.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
T0609.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-9 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
K00 Diagnosis
(DIAGNOSIS)
K10 Adhesion Check W No No Selects whether or not to perform diagnosis
(ADHESION CHK) Yes of adhesion to the electrode.
K11 Adhesion Level1 W 0.00 to 100.00 M ohm 2 0.10 Sets the resistance value for adhesion
(ADH LEVEL1) Level 1 to the electorode.
K12 Adhesion Level2 W 0.00 to 100.00 M ohm 2 0.50 Sets the Level 2 for the diagnosis of
(ADH LEVEL2) adhesion to the electrode.
K13 Adhesion Level3 W 0.00 to 100.00 M ohm 2 1.00 Sets the Level 3 for the diagnosis of
(ADH LEVEL3) adhesion to the electrode.
K14 Adhesion Level4 W 0.00 to 100.00 M ohm 2 3.00 Sets the Level 4 for the diagnosis of
(ADH LEVEL4) adhesion to the electrode.
K15 Adh Measure Value R — M ohm 2 Displays the diagnosis value for
(ADH MEAS VAL) adhesion to the electrode.
K60 — R Good
(SELF CHECK) Error See “6.5 Alarm Functions”.
T0611.EPS
M00 Adjustment
(ADJUSTMENT)
M10 Auto Zero Exe W No Execution No Execution Selects whether or not automatic
(AUTOZERO EXE) Execution zero adjustment is carried out.
Linked with B50.
M11 Magflow Zero R/W -99.999 to 99.999 3 0.000 Displays the result of the automatic
(MAGFLOW ZERO) zero adjustment, or sets the zero point.
M60 — R Good Displays self-check result.
(SELF CHECK) Error See “6.5 Alarm Functions”.
T0612.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
P00 Protect
(PROTECT)
P10 Key Code W 0 to 9999 0 Parameter of the display
(KEY CODE) restriction
P20 Write Protect R No No Displays whether or not overwriting
(W PROTECT) Yes of parameter data is prohibited.
P21 Enable Wrt Passwd W ASCII 8 characters Sets the correct password so that
(ENABLE WRITE) write protection function will be
released.
P22 New Password W ASCII 8 characters Sets the correct password for write
(NEW PASSWORD) protection function.
P23 Software Seal R Break Keep Displays whether or not a Joker
(SOFT SEAL) Keep password was used (Break).
P60 — R Good See “6.5 Alarm Functions”.
(SELF CHECK) Error
T0613-1.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-11 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
6.4 Parameter Description [B22: Base Time Unit] Selection of time units for
the flow rate span
→ This setting is linked with that of parameter C41.
(1) Menu B: Easy Setup items
This parameter selects the time units of the flow
Those parameters with a high frequency of use
rate span; however, if “meter” or "ft" has been
have been grouped together in Easy Setup. All
selected for the flow rate units, “/s” is automatically
basic functions can be controlled using only the
set for this parameter.
parameters from this block. Parameters from Menu
B share identical names with those from other
[B23: Flow Span] Setting of the flow rate span
menus; however, modification of one such pa-
rameter will result in the other being automatically → This setting is linked with that of parameter C42.
modified. The span can be set for the forward flow rate in the
range 0 to 32,000. The units set using B21/C40:
[B10: Language] Selection of language used for Base Flow Unit and B22/C41: Base Time Unit
the display unit will be displayed at this time.
→ This setting is linked with that of parameter H30.
One of the following languages can be selected NOTE
for use by the display unit.
If the flow rate units, time units, and flow rate
Data Range span are specified upon ordering, these param-
Setting item Description eters will be setup before shipment; however, if
English All parameters, alarm messages, etc. displayed this is not the case, it will be necessary for the
in English. appropriate values to be set up by the user.
Japanese All parameters, alarm messages, etc. displayed
in Japanese katakana.
French All parameters, alarm messages, etc. displayed
NOTE
in French.
German All parameters, alarm messages, etc. displayed
Flow rate span is the value for instantaneous
in German. flow rate that corresponds to a current output
Italian All parameters, alarm messages, etc. displayed of 20 mA. The following factors should be taken
in Italian. into consideration when deciding on the flow rate
Spanish All parameters, alarm messages, etc. displayed span.
in Spanish.
T0614.EPS
• In the case of applications with large variations
in flow rate, the maximum flow rate should be
[B20: Flow Damping] Setting of the damping time set. If a flow rate in excess of the flow rate span
constant was to occur, output would be possible up to
→ This setting is linked with that of parameter C11. an upper limit of 108 %, and beyond this, error
The damping time constant should be modified to would occur. Note that the same applies to
suppress an output fluctuation or to change the pulse output and totalization.
response time. This time constant has an effect • In the case of applications that have a relatively
on analog output and on the flow rate display (i.e., stable flow rate, a flow rate span of 1.5 to 2.0
actual instantaneous flow rate, %, current value), times larger than the normal flow rate may be
and in addition, it also affects pulse output and considered suitable.
totalization. However, when “No Damping” has • The flow rate to be adopted should - upon con-
been set for version to flow velocity - correspond to a value
J24: T/P Damp Select, there will be no effect on within the range of 0.1 to 10 m/s. The flow
pulse output or totalization.
velocity can be confirmed using sizing data or
→ Time constant: The time required for the output
with parameter C44: Velocity Check, and in
to reach 63.2 % from 0 %.
the latter case, the value obtained when span
is converted to flow velocity will be displayed.
[B21: Base Flow Unit] Selection of flow units for
the flow rate span • Regardless of the position of the decimal point,
→ This setting is linked with that of parameter C40. the largest value that can be set on the display
This parameter selects the flow rate span units. (In unit is 32,000. Furthermore, it is not possible
case of mass flow, the setting of density is also to set a number of 4 or greater for the highest
required. Refer to C46: Mass Flow Density for order digit. Similarly, if 3 is set for this highest
more details.) order digit, it will not be possible to set a num-
ber of 2 or greater for the next digit to the right,
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-12 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-14 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-15 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-16 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[C46: Mass Density] Setting of the density for [D11: Total Scale] Setting of the totalization scale
mass flow rate → Refer to the description of parameter B31.
This parameter is necessary in situations where
t, kg, g, klb or lb has been selected as the mass [D12: Total Decimal Pnt] Setting of the decimal
unit in B21/C40: Base Flow Unit. If a mass unit is point position for the totalization display
selected in B21/C40: Base Flow Unit and a value This parameter sets the position of the decimal
of 0 is set for this parameter, the setting alarm point for totalization display in terms of the number
“57: Density Set Err.” will be displayed. In such a of digits. Except in cases where 0 is selected, the
case, ensure that the density is set correctly. totalized units are not displayed.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-17 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-18 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
(4) Menu E: Pulse Setting items [E13: Pulse Low Cut] Setting of the pulse output
Menu E contains items relevant to pulse output. stop range
This parameter allows the settings to be made
[E10: Pulse Unit] Setting of the pulse units which prevent pulse output when the flow rate
→ Refer to the description of parameter B32: is at or below the low-cut setting value. In cases
Pulse Unit where there are multiple ranges or forward/reverse
ranges, low cut is carried out at the setting value
[E11: Pulse Scale] Setting of the pulse scale for the smallest span (i.e., an integer between 0
→ Refer to the description of parameter B33: and 10%).
Pulse Scale Example: When the first range = 10 m3/h, the
second range = 100 m3/h, and the Pulse
Low Cut = 3%, no pulse output is carried
[E12: Pulse Width] Setting of the pulse width
out at flow rates of 0.3 m3/h or lower.
This parameter selects the pulse width (i.e., ms :
millisecond) that is output.
[E20: Pulse Active Mode] Setting of pulse output
Data Range
Pulse Scale (pps) active mode
Setting
Maximum Value Minimum Value This parameter sets whether pulse-output active
mode is to be “On Active” or “Off Active”.
(0) 50% Duty 11000 0.0001
Item Description
(1) 0.05ms 10000 (pps: pulses per second) Closed (On) Act When pulse output is carried out,
(2) 0.1ms 5000 contacts are Closed (On).
(3) 0.5ms 1000 Open (Off) Act When pulse output is carried out,
(4) 1ms 500 contacts are Open (Off).
(5) 20ms 25
T0622.EPS
(6) 33ms 15
(7) 50ms 10
Example: When E12: Pulse Width is 1 ms, output
(8) 100ms 5 takes place as follows in accordance
T0621.EPS
with the setting for E20: Pulse Active
Mode.
NOTE
Closed(On)
*: The pulse widths with the exception of 50 %
Open(Off)
Duty is the "Closed (ON)" time for each pulse 1 ms 1 ms
in case that "Closed (ON) Act" is selected for Closed (On) Active Open (Off) Active
E20: Pulse Active Mode. F0604.EPS
Closed (On)
Open (OFF)
Pulse width F0603.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-19 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-20 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Forward totalization preset via The totalization display value is preset or reset to zero in
Ext Ttl Preset accordance with the external status input. The preset value
external status input
is determined using D21: Ttl Set Val Lower and D22: Ttl
Reverse totalization preset via Set Val Upper.
Ext Rev Ttl Set In case that “Start” is selected for D20: Total Execution , the
external status input count for totalization starts from the preset value.
[F13: SI2 Function] Setting of the function for the F14: SO1/2 Active Mode] Setting of the active
SI2 status input terminal mode for status output
This parameter sets the function for the SI2 (status This parameter sets the active mode for the
input 2) terminal. Selectable functions are the same terminals SO1 and SO2. Active modes cannot be
as for F12: SI1 Function; however, the selection of set individually for these two terminals.
Ext 3 Ranges and Ext 4 Ranges is not possible for
Setting Function
F13.
Status output becomes "Closed (On)" when an
Closed (On) Act
event occurs.
Status output becomes "Open (Off)" when an
Open (Off) Act
event occurs.
T0627.EPS
Open(Off)
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-21 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[F15: SI1/2 Active Mode] Setting of status input Multiple Ranges Setting 1:
This parameter sets the active mode for the ter- Automatic multiple ranges switching
minals SI1 and SI2. Active modes cannot be set · When the instantaneous flow rate exceeds
individually for these two terminals. 100% of the range, transition to the next range
Setting Function (up to four ranges) is carried out automatically.
When the status input is set to "Short", occurrence Furthermore, when the flow is in reverse, the re-
Short Active
of the selected event will be recognized.
When the status input is set to "Open", occurrence
verse range can also be automatically selected.
Open Active · Range switching can be confirmed according
of the selected event will be recognized.
T0628.EPS to the status of the SO1 and SO2 status output
Example: When the “0% signal lock” function is terminals. Refer to Table 6.4.1: Status Output for
selected for SI1 or SI2 Automatic Range Switching for details of status
output conditions for each range.
Short
Open
Status Output for Automatic Multiple Ranges
Signal lock status Signal lock status
Switching
Short Active Open Active
F0606.EPS Operations are performed in accordance with the
following table when the active mode has been set
to “Closed (On) Act” using F14: SO1/2 Active
NOTE Mode. Operating patterns are reversed when the
active mode has been set to “Open (Off) Act”.
Multiple ranges setting
Table 6.4.1 Status Output for Automatic Multiple Ranges
Parameters from F30 to F41 are used with
Switching
the automatic multiple ranges and the multiple
ranges switching via external status input. The Parameter setting
Function
Status output
followings will describe the setting method for F10 F11 SO1 SO2
each range. No FunctionNo Function Forward single range – –
Fwd/Rev Auto forward/reverse Forward Open –
Ranges No Function 1 range (SO1)
The multiple ranges use the following parameters: Reverse Closed –
Fwd/Rev Auto forward/reverse Forward – Open
[B23: Flow Span ] Setting of the flow rate span No Function
Ranges 1 range (SO2) Reverse – Closed
Auto Auto forward Forward 1 range Open –
(Setting of the forward No.1 range) 2 Ranges No Function 2 ranges (SO1) Forward 2 range Closed –
[F30: Forward Span 2] Setting of the forward No. Auto Auto forward Forward 1 range – Open
No Function2 Ranges
2 range 2 ranges (SO2) Forward 2 range – Closed
Forward 1 range Open Open
Auto Auto Auto forward
[F31: Forward Span 3] Setting of the forward No. 3 Ranges 3 Ranges 3 ranges Forward 2 range Closed Open
3 range Forward 3 range Open Closed
Forward 1 range Open Open
[F32: Forward Span 4] Setting of the forward No. Auto Auto
Auto forward Forward 2 range Closed Open
4 range 4 Ranges 4 Ranges 4 ranges
Forward 3 range Open Closed
Forward 4 range Closed Closed
[F33: Reverse Span 1] Setting of the reverse No.
Forward 1 range Open Open
1 range Fwd/Rev Auto Auto forward/reverse Forward 2 range Open Closed
Ranges 2 Ranges 2 ranges Reverse 1 range Closed Open
[F34: Reverse Span 2] Setting of the reverse No.
Reverse 2 range Closed Closed
2 range
Forward 1 range Open Open
[F35: Reverse Span 3] Setting of the reverse No. Auto Fwd/Rev Auto forward/reverse Forward 2 range Closed Open
2 Ranges Ranges 2 ranges Reverse 1 range Open Closed
3 range
Reverse 2 range Closed Closed
[F36: Reverse Span 4] Setting of the reverse No. *: "No Function" is the default value. Only SO1 or SO2 terminals are
4 range used for single or dual ranges; accordingly, the unused terminal
remains at the default value (i.e., No Function) and can therefore
[F40: Auto Range Hys] Setting of the automatic be used for other functions.
T0629.EPS
multiple range hysteresis width
[F41: Bi Direction Hys] Setting of the forward/
reverse flow measurement hysteresis.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-22 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-23 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-24 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Parameter setting sequence (for multiple ranges switching via external status input)
F10: SO1 Function F11: SO2 Function F12: SI1 Function F13: SI2 Function
Select a function Select a function Select a function Select a function
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-25 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
alarm is disabled.
HH LL
SO2
NOTE
80
Setting of -110 % or 110 % results in the cor- High Alarm
Output Example 1 20
Low Low Alarm
The high-high alarm (HH) is set to 90 % or more
0 Time (t)
of the flow rate output; the low-low alarm (LL), to
H
SO1
20 % or less; the high alarm (H), to 80 % or more; L Alarm set to
-110% indicates
and the low alarm (L), to 30 % or less. that the alarm is
LL disabled.
SO2
Settings are:
HH Alarm set to 110% indicates
G10: Low Alarm = 30 % that the alarm is disabled.
AL
G11: High Alarm = 80 % LL
G12: Low Low Alarm = 20 % Select “H/L Alarm” for F10: SO1 Function
Select “HH/LL Alarm” for F11: SO2 Function
G13: High High Alarm = 90 % Select “Closed (On) Act” for F14: SO1/2 Active Mode
Select “Open (Off) Act” for G20: Alm Out Act Mode
Select “Yes” for G33: Alm-HH/LL
F0611.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-26 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Open(Off)
[G14: H/L Alarm Hys] Setting of upper/lower Alarm occurs Alarm occurs
alarm value hysteresis width Closed (On) Active Open (Off) Active
This parameter sets the hysteresis width for upper F0614.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-27 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Setting Function
NOTE
No Not recognized as an alarm
Yes Recognized as an alarm The RXFA11 has three different types of alarm
T0633.EPS (i.e., system alarms, process alarms and setting
alarms). For setting alarms and process alarms,
[G31: Alm-Sig Over] Alarm recognition of “Signal settings are made with G30, G31, G32, G33 and
Overflow Alarm” G34 to specify whether these will be recognized
This parameter specifies whether the signal over- as an alarm.
flow process alarm will be recognized as an alarm. Refer to Section 6.5: Alarm Functions for more
A signal overflow occurs when there is an error in details regarding the content of each alarm and
the input signal. the effect of alarm recognition on output.
Setting Function
No Not recognized as an alarm [G40: Operation Time] Display of operation time
Yes Recognized as an alarm This parameter is used to display the operation
T0634.EPS time. The operation time is the total time that is
counted while the device works actually.
[G32: Alm-Emp Pipe] Alarm recognition of When the power supply is off, the operation time
“Empty Pipe Alarm” is not counted. For example, “1D23:45” indicates
This parameter specifies whether the empty pipe an operation time of 1 day, 23 hours, and 45
(flowtube is not filled with fluid) process alarm will minutes.
be recognized as an alarm.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-28 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Item Description
: 16 space characters No issuing of alarms [G43: Alm Record 2] Alarm record 2
(i.e., no display)
This parameter is used to display the second
10 : uP Fault Microprocessor (CPU) failure most-recent alarm, and the alarms that can be
11 : EEPROM Fault EEPROM failure
displayed are the same as those for G41: Alm
12 : A/D(H) Fault A/D converter failure (at high frequency side)
Record1.
13 : A/D(L) Fault A/D converter failure (at low frequency side)
14 : A/D(Z) Fault A/D converter failure (detection of adhesion)
[G44: Alm Record Time 2] Display time of alarm
15 : Coil Open Flowtube coil is open-circuit
record 2
16 : EEPROM Dflt EEPROM returns to default values
This parameter is used to display the operation
18 : Power off Power supply is off.
time at which the alarm indicated by G43: Alm
Instantaneous power fail for tens of
19 : Inst Pwr Fail milliseconds. Record2 was occurred. For example, “1D23:45”
After this fail is released, outputs reach indicates that an alarm was triggered at the op-
the previous value immediately.
The return from excessive instantaneous
eration time of 1 day, 23 hours, and 45 minutes.
28 : WDT noise.
After the noise is released, output return [G45: Alm Record 3] Alarm record 3
the normal condition.
30 : Sig Overflow Input signal error This parameter is used to display the third most-
31 : Empty Pipe Flowtube is not filled with fluid recent alarm, and the alarms that can be displayed
33 : Adhesion Alm Insulation adhered to electrode are the same as those for G41: Alm Record1.
T0637.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-29 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
4 mA
Input 0.5 % 0.5 %
[H10: Display Select1] Setting of the first line of
Setting range: 0 to 10 % Hysteresis
display unit fixed at 1 %
→ Refer to the description for parameter B40 F0615.EPS
[H30: Language] Selection of language used for If “2.4 mA or less” has been set for G21:4-20mA
the display unit. Alarm Out, 2.4 mA or less will be output upon
Refer to the description for parameter B10 an alarm occurrence, regardless of the low limit
This parameter can be used to select the lan- setting.
guage for use on the display unit.
NOTE
(8) Menu J: Auxiliary Function Setting
items • If the setting value for the low limit for current
Menu J contains setting items such as the flow output is not less than the high limit value (as
direction, rate limits, and current output limits. set using J12: 4-20mA High Lmt), the setting
alarm “4-20 Lmt Err” will be displayed.
[J10: 4-20mA Low Cut] Setting of the low-cut • This parameter has no effect on pulse output or
range for current output the totalization function.
This parameter is used to force current output to
0 % (i.e. 4mA) in the vicinity of 0 % output, and
setting for the 4-20 mA output low cut is made [J12: 4-20mA High Lmt] Setting of the high limit
using a percentage of the smallest flowrate span. for current output
However, the low cut function will be terminated if This parameter is used to restrict high current
this parameter is set to 0 %. portions of 4-20 mA output, and it is initially set to
The indication of the instantaneous flow rates (%, 120 %. Setting should be performed when a lower
actual instantaneous flowrates, mA, bargraph) on value is required for the higher limit.
the display unit are the same action. The indication of the instantaneous flow rates (%,
actual instantaneous flowrates, mA, bargraph) on
Example: Situation where low cut is set to 10 % the display unit are the same action.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-30 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Example: Situation where low cut is set to 90 % [J22: Dead Time] Setting of dead time
Output This parameter sets the time for application of the
rate limit, and if a value of 0 is set, the rate limit
18.4 mA (90 %)
function will be terminated.
4 mA NOTE
Input
F0656.EPS Determining rate limit value and dead time
Rate limit value: Dead time (T0):
Determines the level for output This is to be determined using
fluctuation cutoff. For example, the output fluctuation width. If
NOTE
if this is set to 2 %, noise noise exceeds the dead time as
above 2 % will be eliminated shown in the diagram below, the
If “21.6 mA or more” has been set for as shown in the diagram. dead time should be made
G21:4-20 mA Alarm Out, 21.6 mA or more will longer.
T0
be output upon an alarm occurrence, regardless
2%
of the high limit setting. 2%
[J20: Flow Direction] Setting of the flow direction T0
Upon shipment from the manufacturing plant, the F0616.EPS
for determining whether an input corresponds to 100 225 Number of signal samples
Dead time: 3 s
a flow measurement or noise. Specifically, this F0617.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-31 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
[J30: Power Synch] Setting of power synchroni- [J40: Memo 1] Setting of memo 1
zation
[J41: Memo 2] Setting of memo 2
This parameter indicates whether or not the inter-
nal frequency is to be synchronized with that of [J42: Memo 3] Setting of memo 3
the power supply.
These parameters are used with the memo function,
and up to 16 characters can be set for each.
[J50: Software Rev No] Display of software revision
This parameter is used to display the software’s
revision number.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-32 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
(9) Menu K: Diagnostic Function Set- *: The warning 80: Adhesion Wng is displayed
ting items when the adhension level reaches Level 3.
Menu K contains items that are relevant to self *: If “Warning Output” has been selected for F10:
diagnosis of insulation adhesion to the electrode. SO1 Function or F11: SO2 Function, then sta-
tus output will be performed when the adhesion
level reaches Level 3.
[K10: Adhesion Check] Setting of adhesion diag-
nostic function
This parameter selects whether or not the adhesion [K14: Adhesion Level 4] Setting the resistance
diagnostic function will be carried out. value for adhesion diagnostic level 4
Setting Function This parameter sets the resistance value (in MΩ)
No Halt the adhesion diagnostic function for judgment of Level 4.
Yes Carry out the adhesion diagnostic function *: The process alarm 33: Adhesion Alm is dis-
T0642.EPS played when the adhension level reaches Level
4.
NOTE *: Alarm output will be performed if “Yes” has been
selected for G34: Alm-Adhesion.
Adhesion Diagnostic Function
· This function diagnoses adhesion using elec- [K15: Adh Measure Value] Displays the resistance
trode resistance values. value for adhesion diagnose
· When “Adhesion check” has been set for B41/ This parameter displays the value measured using
H11: Display Select 2 or B42/H12: Display the adhesion diagnostic function (in MΩ).
Select 3, the diagnose adhesion is indicated When "No" is selected for K10: Adhesion check,
on the display unit using four different levels. the parameter displays the indeterminated value.
· If the judgment value for Level 3 is exceeded,
a warning is displayed; and if the value for
Level 4 is exceeded, an alarm is displayed. (10) Menu M: Automatic Zero Adjust-
ment Function Setting items
· Available conductivity for this function is lim-
ited to: Menu M contains items that are relevant to auto-
Nominal size 10 mm and smaller: 30 µS/cm matic adjustment of the zero point.
Nominal size 15 mm and larger: 10 µS/cm
Make sure to use the adhesion diagnostic [M10: Auto Zero Exe] Execution of automatic zero
function with the greater conductivity than the adjustment function
above mentioned value. → Refer to the description of parameter B50.
Alarm
Warning [M11: Magflow Zero] Results of automatic zero
adjustment
This parameter is used to display the results
Level: 1 Level: 2 Level: 3 Level: 4
obtained from B50/M10: Auto Zero Exe. Specifi-
F0619.EPS cally, the correction values displayed, and it is also
possible to directly enter correction values.
[K11: Adhesion Level 1] Setting the resistance
value for adhesion level 1
This parameter sets the resistance value (in MΩ) (11) Menu N: Loop Test (Output Test)
for judgment of Level 1. Setting items
Menu N contains items that are relevant to loop
[K12: Adhesion Level 2] Setting the resistance testing.
value for adhesion diagnostic level 2 [N10: Test Mode] Setting for loop test execution
This parameter sets the resistance value (in MΩ)
Setting Function
for judgment of Level 2.
Normal No execution of loop testing
[K13: Adhesion Level 3] Setting the resistance Test Loop testing is started
T0644.EPS
value for adhesion diagnostic level 3
This parameter sets the resistance value (in MΩ)
for judgment of Level 3.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-33 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
values [N24: Test SI2] Display for SI2 status input termi-
During loop testing, current output (4-20 mA), nal condition during testing
totalization, and pulse will be output in accordance This parameter is used to display the condition of
with this parameter’s setting, and values can be the SI2 status input terminal during loop testing.
Setting Function
set when “Test” has been selected for N10: Test
Mode. With multiple ranges or when performing Open SI2 input terminal in Open condition
forward/reverse flow measurements, setting should Short SI2 input terminal in Short condition
T0646-2.EPS
be done using a percentage of the maximum
range.
(12) Menu P: Parameter Protection
[N20: Test SO1] Setting for SO1 output terminal items
condition during testing Menu P contains items that are relevant to write-
This parameter sets the condition of the SO1 protection and passwords.
status output terminal during loop testing. Setting
is possible when “Test” has been selected for N10:
Test Mode. [P10: Key Code] Parameter display limit
Setting Function This parameter restricts access to the Service
Open (Off) SO1 output terminal in Open (Off) condition Mode.
Closed (On) SO1 output terminal in Closed (On) condition
T0645.EPS NOTE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-34 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
[P20: Write Protect] Status indication for write [P23: Software Seal] Display the software seal
protection When the joker password has been used to
This parameter is used to indicate whether or not release write protection, this parameter displays
write protection is currently on “Break”, and when protection is cancelled using
Default setting (Enable)
the password set using P22: New Password, it
P20:Write Protect returns to “Keep”.
No
NOTE
F0620.EPS
[G20: Alm Out Act Mode] Setting of the active [G34: Alm-Adhesion] Alarm recognition of “Adhe-
mode for the alarm output terminal sion Alarm”
[G21: 4-20mA Alarm Out] Setting of the current [G41: Alm Record 1] Alarm record 1
output during an alarm occurring. [G43: Alm Record 2] Alarm record 2
[G30: Alm-Setting] Alarm recognition of “Setting [G45: Alm Record 3] Alarm record 3
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-36 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-37 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Process Alarms (Device is normal but process-related errors make correct measurement impossible.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Carry out an investigation as follows:
· Check the signal cable for breakage.
Check signal cable · Check for contact between signal
30:Sig Overflow Input signal error cable, power cable, and excitation cable.
and grounding
· Check for stray currents in the fluid.
· Check the grounding.
Fill flow tube with Flowtube is not filled with
31:Empty Pipe Fill the flowtube with fluid.
fluid fluid
Flow rate alarm for greater
32:HH/LL Alm Check the flow rate and than High-High limit value Check the flow rate and setting value
setting value or less than Low-Low limit of High High limit and Low Low limit.
value.
33:Adhesion Alm Clean electrodes Electrode adhesion alarm Clean the electrodes.
T0652.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-38 All Rights Reserved. Copyright © 2006, Rota Yokogawa
6. PARAMETER DESCRIPTION
Setting Alarms (Device is normal but errors have been made in the setting of parameters.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Span flow velocity setting is 11 m/s or Check whether parameters
Check parameter C40, C41, more C40, C41, and C42 are correct.
50:Span > 10m/s In case that multiple range or
and C42
forward and reverse flow
measurement functions is used,
Check parameter C40, C41, Span flow velocity setting is 0.05 m/s or check whether parameters F30
51:Span < 0.1m/s less through F36 are correct.
and C42
Check parameter D10 and Totalization rate is 11000 pps or more
52:TTL>10000p/s
D11 Check whether parameters
Check parameter D10 and Totalization rate is 0.00005 pps or less D10 and D11 are correct.
53:TTL<0.0001p/s
D11
54:4-20 Lmt Err Check parameter J11and J12 The condition [4-20 low limit (J11) < Check whether parameters J11
4-20 high limit (J12)] is not satisfied. and J12 are correct.
The condition [No. 1 range < No. 2 Check whether parameters
55:Multi Rng Err Check parameter F30 to F36 range < No. 3 range < No. 4 range] is F30 through F36 are correct.
not satisfied for multiple ranges.
The condition [High Alarm (G11) –
Low Alarm (G10) > H/L Alarm Hys
Check whether parameters
56:H/L HH/LL Set Check parameter G10 to G14 (G14)] or the condition [High High G10 through G14 are correct.
Alarm (G13) – Low Low Alarm (G12)
> H/L Alarm Hys (G14)] is not satisfied.
Mass units have been selected for Base
Check parameter C40, C45, Flow Unit (C40) but density is set to 0. Check whether parameters
57:Dens Set Err C40, C45, and C46 are correct.
and C46
Pulse rate is 11000 pps or more with
Check parameter E10, E11, 50% duty.
60:PLS > 10000p/s Pulse width is 10000 pps or more with
and E12
0.05 ms selection.
Check parameter E10, E11, Pulse rate is 5000 pps or more with
61:PLS > 5000p/s 0.1 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 1000 pps or more with
62:PLS > 1000p/s 0.5 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 500 pps or more with
63:PLS > 500p/s 1 ms selection.
and E12 Check whether parameters
Check parameter E10, E11, Pulse rate is 25 pps or more with E10, E11, and E12 are correct.
64:PLS > 25p/s 20 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 15 pps or more with
65:PLS > 15p/s 33 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 10 pps or more with
66:PLS > 10p/s 50 ms selection.
and E12
Check parameter E10, E11, Pulse rate is 5 pps or more with
67:PLS > 5p/s 100 ms selection.
and E12
Check parameter E10, E11,
70:PLS<0.0001p/s Pulse rate is 0.00005 pps or less.
and E12
Measure Mode (C20) is set to Enhanced
DF without selecting an optional Check whether parameter C20
71:Meas Mod Set Check parameter C20
code HF1 or HF2. is correct.
A value of 3000.1 mm or more is set for Check whether parameter C32
72: Size Set Err Check parameter C32 Nominal Size (C32). is correct.
73: Adh Set Err Check parameter K11 to K14 The condition in Adhesion detection Check whether parameter K11,
level, Level:1<Level:2<Level:3<Level:4 K12, K13 and K14 are correct.
is not satisfied.
T0653.EPS
Setting Alarms (Device and measurements are normal but a warning is issued.)
Display unit/BRAIN Alarm countermeasure
( 60) content message on display unit Alarm description Countermeasure
Slight adhesion to Clean and check the electrodes.
80:Adhesion Wng – electrodes. Refer to parameter K13.
Carry out adjustment as follows:
· Check if the flowtube is filled with
Results of automatic zero fluid.
82:Auto Zero Wng – adjustment are higher than · Check if the flow velocity is
the rated values. completely zero.
· Check the condition of grounding.
Confirm whether the flow rate is in
excess of the upper limit (108 %) or
83:Fix Cur Wng – The current value is fixed. below the lower limit (-8 %), or
whether upon entry to the Test Mode
or not.
84:Disp Over Wng Overflow in the display digits during Check whether parameter C43 is
–
(only for display unit) instantaneous flow rate display. correct.
Investigate whether the display unit
90:Disp SW Wng Display unit switches are not cover is fitted or whether the cover’s
(only for display unit) – operating. glass surface is dirty.
T0654.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 6-39 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
6. PARAMETER DESCRIPTION
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 6-40 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
Alphanumeric keys
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
Use the function key [F2] CAPS to select between Function Command List
uppercase and lowercase (for letters only). The Command Function
case toggles between uppercase and lowercase ADJ Displays the ADJ menu
each time you press [F2] CAPS. CAPS/caps Selects uppercase or lowercase
The following symbols will appear in sequence, SET Displays the SET menu
one at a time, at the cursor each time you press SLOT Returns to the slot selection panel
[F1] CODE: UTIL Calls the utility panel
*COPY Prints out parameters on display
/ . – , + * ) ( ’ & % $ # ” !
*FEED Paper feed
To enter characters next to these symbols, press [
> ] to move the cursor. *LIST Lists all parameters in the menu
*PON/POFF Automatic printout mode on or off
Entry Key-in Sequence
*PRNT Changes to the print mode
symbol command *GO Starts printing
l/m
*STOP Cancels printing
(I) (/) (m)
T0803.EPS
* Available on BT200-P00 (with printer).
T0804.EPS
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC Function commands
Function keys
F0808.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 7-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
NOTE
In case of BT200, the parameters are displayed
4-20 mA Instrument room
DC signal in English only.
Receiving
Relay transmission Terminal instrument Even if the language with the exception of English
I terminals line board is selected at B10/H30: Language, the param-
I eters are displayed in English upon BT200.
Receiving
resistance: 7.2.2 The data update and upload/
250ot 450Ω download function of BT200
(1) The data update of BT200
BT200 BT200 BT200 BT200 When the following parameters are displayed, the
F0705.EPS
measured data is updated automatically every
Figure 7.2 Connecting the BT200 seven seconds.
Item Name (BRAIN) Item Name (BRAIN)
WARNING
A10 FLOW RATE (%) G43 ALM RECORD2
Be sure to set parameters as write protect func- A20 FLOW RATE G44 ALM TIME 2
tion after finish of parameter setting work. Under A21 FLW RATE (mA) G45 ALM RECORD3
extremely rare case, the infra-red switches may
A30 TOTAL G46 ALM TIME 3
respond unexpectedly in such conditions as sti-
A31 REV TOTAL G47 ALM RECORD4
cking ball of water or extraneous substances on
A32 DIF TOTAL G48 ALM TIME 4
the surface of display panel glass according to the
C44 VELOCITY CHK N23 TEST SI1
principle of infra-red switch operation. Its probabi-
lity rises in such cases as sticking rain water by G22 4-20 BURNOUT N24 TEST SI2
storm or other similar situation and washing up G40 OPERATE TIME P20 W PROTECT
work near flowmeter installation place. Either to il- G41 ALM RECORD1 P23 SOFT SEAL
luminate or stop illuminating the infra-red switches G42 ALM TIME 1
by the flashlight may cause the mis-reaction. T0704.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
(2) Upload/download function of BT200 7.2.3 BT200 Panels & Flow Data Dis-
Upload function is used when the parameters of play
one RXFA11 are copied to the BT200. And down-
load function is used when the parameters copied Use the following procedure to display flow data
to the BT200 are set to another RXFA11. on the BT200.
For details, refer to BT200 User’s Manual • The display of flow data is updated every 5 s.
(IM 01C00A11-01E).
The targeted parameters for upload and download When the BT200 is turned
are following. –––WELCOME–––
BRAIN Terminal
on, the message “Please
wait ...” is displayed for
ID:BT200
several seconds, and then
Item Name (BRAIN) Item Name (BRAIN)
Check cable connection the screen on the left is
C11/B20 FLOW DAMPING D10/B30 TOTAL UNIT and press the Ent key. shown.
UNTIL FEED
Press the key.
C31 SIZE UNIT D11/B31 TOTAL SCALE
C32 NOMINAL SIZE E10/B32 PULSE UNIT
C40/B21 FLOW UNIT E11/B33 PULSE SCALE
C41/B22 TIME UNIT H10/B40 DISP SELECT1 PARAMETER Press the key
01:MODEL to display initial data as
C42/B23 FLOW SPAN H11/B41 DISP SELECT2 AXFA11-A00 shown on the left.
02:TAG NO
C43/B24 FLOW DECIMAL H12/B42 DISP SELECT3 FI-1101 The tag number
03:SELF CHECK
T0705.EPS GOOD specified upon
OK ordering is found
here.
Press the ( OK )or
key.
or
( ESC )
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 7-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
IMPORTANT
NOTE
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
MENU When the cursor has been SETTING When selected, “l (Liter)”
A:DISPLAY moved to B: EASY SETUP, B21:FLOW UNIT will begin to flash on and
B:EASY SETUP m3
C:BASIC SETUP press the key to l(Liter) off: Press the
D:TOTAL SET access Parameter Screen key once again to confirm
E:PULSE SET PRINTER OFF
(Sub-Item Parameter F2:PRINTER ON your selection.
F:STATUS FUNC
Search Mode). FEED POFF NO In case of cancel, press the
HOME SET ADJ ESC
( NO ) key.
X2
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 7-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
7. OPERATION VIA BRAIN TERMINAL (BT200)
MENU When the cursor has been SETTING The data will begin to flash
A:DISPLAY moved to B: EASY SETUP. B23:FLOW SPAN on and off: Press the
B:EASY SETUP 100.000l/min
C:BASIC SETUP Press the key to + 120 key once again
D:TOTAL SET to confirm your entry.
E:PULSE SET access Parameter Screen PRINTER OFF
(Sub-Item Parameter F2:PRINTER ON In case of cancel, press the
F:STATUS FUNC
HOME SET ADJ ESC Search Mode). FEED POFF NO ( NO ) key.
In the parameter list, the The value for flow span has
PARAMETER SETTING
B10:LANGUAGE parameter for setting of flow B23:FLOW SPAN been changed to
English rate span is B23: FLOW 120.000l/min “120.000 l/min.”
B20:FLOW DAMPING
3.0s SPAN. Press the Press ( OK ) key to
B21:FLOW UNIT exit Setting Screen
l(Liter) key four times to move the
DATA DIAG PRNT ESC
cousor to B23. FEED NO OK
(Parameter Replacement
(Note that the key Mode), and then the display
returns to Parameter
can be used to move the
Screen (Sub-Item
cursor to the top of the next
Parameter Search Mode) of
screen.
Menu B items.
In case of redo of setting,
press the ( NO ) key
or
to return the Setting Screen
(Parameter Replacement
Mode).
PARAMETER After moving the cursor to
B22:TIME UNIT
/min B23: FLOW SPAN, press the ( OK )
B23:FLOW SPAN key to access
100.000l/min
B24:FLOW DECIMAL Setting Screen (Parameter
l Replacement Mode). PARAMETER The changed setting for
DATA DIAG PRNT ESC B22:TIME UNIT flow rate span can now be
/min confirmed.
B23:FLOW SPAN
120.000l/min
B24:FLOW DECIMAL Press the ( ESC ) key
l
to return to Menu Screen
DATA DIAG PRNT ESC
(Major Item Parameter
SETTING In this mode, new values can Search Mode).
B23:FLOW SPAN
100.000l/min be entered into the
+ 0 highlighted field using the F0708.EPS
alphanumeric keys.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 7-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
7. OPERATION VIA BRAIN TERMINAL (BT200)
X4
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 7-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
example. The same attention must be paid to 2. Confirmation of the device revision for the
the RXF integral flowmeter. Please note that HART Configuration Tool
commands and comments displayed on HART
(1) Turn on the power of the Configuration
interface products will be refered to as AXF... .
Tool under the standalone condition.
The device description (DD) of AXF is identical
(2) Confirm the installed DD revision in ac-
to RXF. Please select from HART interface model
cordance with the procedure of the Con-
AXF for usage with model RXF.
figuration Tool. Refer to its manual how to
Note: HART is a registered trademark of the HART confirm it in detail.
Communication Foundation. The first 2 digits of the DD file are ex-
pressed the device revision, and its last 2
digits are expressed the DD revision.
8.1 Matching of instru- 01 03.XXX
ment (RXFA11) DD DD revision
and HART Configura- Device revision
F0801.EPS
tion Tool’s DD
Before using the HART Configuration Tool (such
as FieldMate), confirm that the DD (Device De-
scription) of the RXFA11 is installed in the Con-
figuration Tool before using.
About the DD, use the device type, device revision
and DD Revision shown in follows.
Model Name DTM Name Device Type Device
Revision
AXFA11 AXFA11 AXFA11 1
V1.1 (0x0050)
T0801.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
4 to 20 mA DC nal
sig line
Terminal board
Relaying
terminals
USB
FieldMate
Receiving instrument
I+ I- modem
load esistanc
r e:
250 Ω to 600 Ω
USB (including cable
RXFA11 resistance)
PC/F
ieldMate
F0802.EP
WARNING IMPORTANT
Be sure to set parameters as "Protect" on the If the power of flowmeter is turned off within 30
write protect function after finish of parameter seconds after parameters have been set, these
setting work. settings will be canceled. Accordingly, please
Under extremely rare case, the infra-red switches keep the power on for at least 30 seconds after
may respond unexpectedly in such conditions as setting parameters.
sticking ball of water or extraneous substances
on the surface of display panel glass according
to the principle of infra-red switch operation. NOTE
Its probability rises in such cases as sticking
rain water by storm or other similar situation Before updating any setting, remember to always
and washing up work near flowmeter installation check the data content you want to change as
place. described in Chapter 6: Parameter Description.
Either to illuminate or stop illuminating the infra-
red switches by the flashlight may cause the
NOTE
mis-reaction.
Refer to Chapter 6 "Menu P: Parameter Protec- In case of HART Configuration, the parameters
tion Items" and section "10.2.3" how to use the are displayed in English only.
write protect function in detail. This menu in the Even if the language with the exception of Eng-
275 is located at [Hot key] → [2 Wrt Protect Menu]. lish is selected at “Language” setting display, the
parameters are displayed in English upon HART
Configuration Tool.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
• Procedure to call up the Self test display 8.4.4 HART Specific Functions
Device Setup → Diag/Service → Test → Self test
Trim Analog Output
If no error is detected, “Self test OK” is displayed
on the configuration tool. This function is used to adjust the analog output at
If the specific diagnostic item is known for the 4 mA and 20 mA with the D/A trim or Scaled D/A
check, you can directly call up the item by using trim.
the Status command. (1) D/A trim
The status is categorized from 1 to 7. Connect a calibration digital ammeter, and
Show an example below to confirm the status of then enter the read value of the ammeter for
Status group 1. each output of RXFA11.
• Procedure to call up the Status display • Procedure to call up the D/A trim display
Device Setup → Diag/Service → Status → Status group 1 DD Device Setup → Diagnosis/Service →
If no error is detected, “Off” is displayed on the Adjustment → D/A trim
configuration tool. DTM Calibration → D/A trim
T0808.EPS
If there is an error, “On” is displayed on the con-
figuration tool, and a countermeasure for that error (2) Scaled D/A trim
is necessary. The output is adjusted by using a voltmeter or
Example of display: Span > 10 m/s On
other types of meters with 0 to 100% scale.
Span < 0.3 m/s Off Example using a voltmeter:
TTL > 10000 p/s Off 4 mA DC → 1 V
TTL < 0.0001 p/s Off 20 mA DC → 5 V
4-20 Lmt Err Off
Connect this meter, and enter the read value
The HART configuration tool diagnoses at each of the meter for each output of RXFA11.
communication.
When an improper operation is performed, the er- • Procedure to call up the Scaled D/A trim display
DD Device Setup → Diagnosis/Service
ror message is displayed.
→ Calibration
(2) Using DTM → Analog output trim →Scaled D/A trim
The Device Status commands are used for DTM Calibration → Scaled D/A trim
self-diagnostics. When Device Status is run, the T0809.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
(any one of PV, % range/current, or process vars/ (2) Activating the multidrop mode
crnt). The data is sent intermittently as a digital About the procedure to call up the Polling dis-
signal at 3 times a second. play, please refer to the User’s Manual of each
The burst mode is set as follows. configuration tool.
(1) Setting the data to be sent
• Procedure to call up the display NOTE
DD Device Setup → Detailed setup When the same polling address is assigned for
→ HART output → Burst option two or more field devices in multidrop mode,
DTM Configuration → HART → Burst option communication with these field devices is dis-
T0810.EPS
abled.
Select the type of data to be sent from the fol-
lowing options:
- Instantaneous flow rate (PV)
- Output in % and current output (% range/cur-
rent)
- Instantaneous flow rate, totalization value*
and current output (Process vars/crnt)
* “Totl,” “Reverse Totl” or “Dif Totl”
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
1. Device setup
Online 2. PV
3. PV AO
Frequency
4. PV Span
Utility
Parameter
Read/Write of BRAIN Data(*1)
protocol
F0803.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter
Read/Write of BRAIN Data(*1)
protocol
Enhanced DF W C20 –
Low MF (EDF) W C23 –
High MF (EDF) W C24 –
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-8 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter
Read/Write of BRAIN Data(*1)
protocol
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-9 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter
Read/Write of BRAIN Data(*1)
protocol
5 Review 1 Review1 R – –
2 Review2 R – –
3 Review3 R – –
R – –
4 Review4
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-10 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
F0807.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-11 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Root Menu
Process Variables A
Device Status B
Diag and Service C
Easy Setup D
Configuration E
Calibration F
Write Protect G
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
A
Process Variables PV R A20 P
PV % rnge R A10 P
PV AO R A21 P
PV Span R B23 P
PV Damping R C11/B20 P
Totl R A30 P
Reverse Totl R A31 P
Dif Totl R A32 P
B
Device Status • Process Variables PV
PV % rnge
Totl
Reverse Totl
Dif Totl
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-12 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
C
Diag and Service Output Test Loop test W – –
Test Mode W N10 –
Test SO1 W N20 –
Test SO2 W N21 –
Test Alarm Out W N22 –
Test SI1 R N23 P
Test SI2 R N24 P
D
Easy Setup Tag W C10 u/d
Language W B10/H30 u/d
PV Damping W B20/C11 u/d
Base Flow Unit W B21/C40 u/d
Base Time Unit W B22/C41 u/d
PV Span W B23/C42 u/d
Flow Decimal Pnt W B24/C43 u/d
Total Unit W B30/D10 u/d
Total Scale W B31/D11 u/d
Pulse Unit W B32/E10 u/d
Pulse Scale W B33/E11 u/d
Display Select1 W B40/H10 u/d
Display Select2 W B41/H11 u/d
Display Select3 W B42/H12 u/d
Auto Zero Exe – – –
F0809.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-13 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
E
Configuration Characterize Meter Select Flow Tube W C30 –
Nominal Size Unit W C31 u/d
Nominal Size W C32 u/d
Measure Mode W C20 –
Low MF W C21 –
High MF W C22 –
Low MF (EDF) W C23 –
High MF (EDF) W C24 –
MF Set – – –
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-14 All Rights Reserved. Copyright © 2006, Rota Yokogawa
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
All Rights Reserved. Copyright © 2006, Rota Yokogawa 8-15 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
8. OPERATION VIA HART COMMUNICATOR TOOL (HART5)
Parameter Data
Read/Write of BRAIN Renewing
protocol (*1)
F
Calibration Auto Zero Exe – M10/B50 –
Magflow Zero R/W M11 –
D/A trim – – –
Scaled D/A trim – – –
G
Write Protect Write protect R P20 P
Enable Wrt Password W P21 –
New Password W P22 –
Software Seal R P23 –
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 8-16 All Rights Reserved. Copyright © 2006, Rota Yokogawa
9. ACTUAL OPERATION
9. ACTUAL OPERATION
After you have installed the flowtube into the
process piping, wired the input/output terminals,
set up the required parameters, and performed
a pre-operation zero adjustment, the magnetic
flowmeter should output an accurate flow signal
from its terminals as soon as flow of the fluid to
be measured begins. This section describes zero
adjustment and the corresponding procedures.
IMPORTANT
All Rights Reserved. Copyright © 2006, Rota Yokogawa 9-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
9. ACTUAL OPERATION
Parameter M10:Auto Zero Exe Touch the switch to move the cursor to
No Execution
Replacement No Execution
“Execution”.
(execute) Mode Execution
F0901-1.EPS
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 9-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
9. ACTUAL OPERATION
Sub-item Major item M:Adjustment When zero adjustment function has been com-
Parameter Sub- item 10:Auto Zero Exe pleted, the system returns automatically to the
Search Mode selection (D) 11:Magflow Zero sub-item selection screen (D).
F0901-2.EPS
NOTE
The results of M10: Auto Zero Exe can be dis-
played using M11: Magflow Zero. Alternatively,
if the results of the automatic zero adjustment
exceed the rated value, the warning 82: Auto
Zero Wng will be displayed.
CAUTION
In certain cases where the multiple range function is being used with other status
inputs, it may not be possible to perform settings for automatic zero adjustment.
For more details, refer to the description of multiple ranges from Chapter 6: Pa-
rameter Description.
In order to carry out zero adjustment via external status input, it will be neces-
sary to set “Ext Auto Zero” using either F12: SI1 Function or F13: SI2 Function.
The parameter F12: SI1 Function will be used in the following description.
X2
F0902-1.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 9-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
9. ACTUAL OPERATION
F12:SI1 Function
No Function
Ext Auto Zero
Ext Ttl Preset
In order to request confirmation, the entire dis-
play flashes on and off. Touch the switch
F12:SI1 Function
once again at this time to fix selection of the
Ext Auto Zero
automatic zero adjustment function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode. When the operations except
are carried out, the parameter cannot be
set.
Sub-item Major item F:Status Function
11:SO2 Function
Parameter Sub-item 12:SI1 Function
Search Mode selection (E) 13:SI2 Function
F0902-2.EPS The system returns automatically to sub-item
selection screen (E).
Zero adjustment will be started if the DIO terminals are shorted (when the active
mode has been set to “Closed/Short Act” using F15: SI1/2 Active Mode).
This process will end after approximately 30 s.
Auto zero adjustment function is being executed.
Now Auto Zero (about 30 s.)
Executing...
F0902-3.EPS
NOTE
When the DIO terminals continue to be shorted, this zero adjustment is automati-
cally repeated.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 9-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
10. MAINTENANCE
10. MAINTENANCE
10.1 Maintenance 10.2 Setting of Switches
WARNING
IMPORTANT
• Maintenance work must be carried out by
the trained personnel having knowledge • Lifting of display unit is necessary for setting
of safety standard and not by operators. switches. Perform display unit lifting as
• When opening the cover, wait for more than described in section 10.2.1.
10 minutes after turning off the power. Further- • To preserve the safety, do not touch the
more, opening of the cover must also be carried electrical circuit and the cables except the
out by the trained personnel having knowledge setting switches.
of safety standard.
F1008.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 10-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
NOTE
NOTE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 10-2 All Rights Reserved. Copyright © 2006 Rota Yokogawa
10. MAINTENANCE
10.3 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when
the instrument is not operated correctly. This section describes troubleshooting pro-
cedures where the cause of the breakdown is identified through receiver indication.
10.3.1 No Indication
START
Is anything else NO
being displayed?
YES
Does the
NO
arrow on the tube match the
flow direction?
Is the flowtube NO
full of fluid?
YES
Change the piping or the
Check the resistance between mounting condition so that
the flowtube’s A, B, and C the tube fills with the fluid.
terminals using a multimeter.
Is the NO
resistance 100 M or more
in each case?
YES
Is foreign NO
matter adhering to
the walls?
YES
All Rights Reserved. Copyright © 2006, Rota Yokogawa 10-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
START
YES
Are valves leaking? Replace the valve(s).
NO
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 10-4 All Rights Reserved. Copyright © 2006 Rota Yokogawa
10. MAINTENANCE
START
Are parameters NO
Set the parameters correctly.
set correctly?
YES
Is the NO
Completely ground the converter and
flowmeter completely flowtube (Refer to the section on
grounded? grounding in the relevant user’s manual).
YES
Does the NO
conductivity stay within the Use within the specified conductivity range.
specified range?
Signal terminals A-C and B-C: NO Dry the tube fully to restore insulation
100 M or more? resistance.
Excitation terminals EX1-C:
1 M or more?
YES
All Rights Reserved. Copyright © 2006, Rota Yokogawa 10-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
10. MAINTENANCE
Blank Page
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 10-6 All Rights Reserved. Copyright © 2006 Rota Yokogawa
11. OUTLINE
11. OUTLINE
STANDARD SPECIFICATIONS Coating/Paint:
Case and Cover: Polyurethane corrosion-resistant
Excitation Method (combined with RXF Remote Flowtube):
Coating Color: Silver grey (Munsell 3.2PB 7.4/1.2 or its
• Standard pulsed DC excitation:
equivalent)
Size 15 to 1000 mm (0.5 to 40 in)
Cover Mounting Screws: Polyurethane corrosion-resistant
Input Signal: Coating Color: Mint green (Munsell 5.6BG 3.3/2.9 or its
Two Status Inputs: Dry contact equivalent)
Load resistance: 200 Ω or less (ON),
Converter Material:
100 kΩ or more (OFF)
Case and Cover: Aluminum alloy
Output Signals:
Mounting/Shapes:
• One Current Output: 4 to 20 mA DC:
• Mounting: 2-inch pipe, panel or surface mounting
Load resistance: 1 kΩ maximum, including cable resis-
• Electrical Connection: ANSI 1/2 NPT female
tance
ISO M20 x 1.5 female
• One Pulse Output:
• Terminal Connection: M4 size screw terminal
Transistor contact output (open collector):
Grounding:
Contact capacity 30 V DC (OFF), 200 mA (ON)
Grounding resistance 100 Ω or less
Output rate 0.0001 to 10,000 pps (pulse/s)
When optional code A is selected, grounding resistance
• One Alarm Output:
10 Ω or less shall be applied.
Transistor contact output (open collector):
Contact capacity 30 V DC (OFF), 200 mA (ON) Combined Remote Flowtube:
• Two Status Outputs: • RXFA11 Converter can be combined with size 15 to 1000
Transistor contact output (open collector): mm (0.5 to 40 in) of RXF Remote Flowtube.
Contact capacity 30 V DC (OFF), 200 mA (ON) • If a combined converter is changed from RXFA11 to
RXFA14 or vice versa, the meter factor must be adjusted
Communication Signals:
by flow calibrations.
BRAIN or HART communication signal (Superimposed
on the 4 to 20 mA DC signal)
Distance from power line: 15 cm (6 in) or more. Functions:
(Parallel wiring should be avoided.)
How to Set Parameters:
BRAIN:
The indicator’s LCD and three infra-red switches enable
Communication Distance: users to set parameters without opening the case cover.
Up to 1.5 km (0.93 miles), when polyethylene insulated Parameters can also be set with the configuration tool
PVC-sheathed cables (CEV cables) are used. (Such as HHT (handheld terminal) or FieldMateTM, etc.).
Communication distance varies depending on the type The language for the HHT is English only.
of cable and wiring used.
Displayed Languages:
Load resistance: 250 to 450 Ω (incl. cable resistance) Users can choose one of the following languages:
Load Capacitance: 0.22 µF or less English, French, German, Italian, Japanese or Spanish.
Load Inductance: 3.3 mH or less Instantaneous Flow Rate/Totalized Value Display
Input Impedance of Communicating Device: Functions:
10 kΩ or more (at 24 kHz) The full dot-matrix LCD enables user selections of dis-
plays from one line to three lines for:
HART: • Instantaneous flow rate
Load resistance: 230 to 600 Ω (incl. cable resistance) • Instantaneous flow rate (%)
Note: HART is a registered trademark of the HART Com- • Instantaneous flow rate (bar graph)
munication Foundation.
• Current output value (mA)
Data Security During Power Failure: • Totalized value
Data (parameters, totalizer value, etc.) storage by • Tag No.
EEPROM. No back-up battery required. • Results of electrode adhesion diagnostics
Indicator: Totalizer Display Function:
Full dot-matrix LCD (32x132 pixels) The flow rate is counted one pulse at a time according
Lightning Protector: to the setting of totalization pulse weights. For forward
The lightning protector is built into the excitation current and reverse flow measurement functions, the totalized
output, the current output, the signal common, and values of the flow direction (forward or reverse) and the
the pulse/alarm/status input and output terminals as flow rate are displayed on the indicator together with
standard. If optional code /A is selected, an additional the units. The difference of totalized values between the
lightning protector is built into the power terminals. totalized forward and reverse flow can also be displayed.
Protection: Totalization for the reverse flow rate is carried out only
IP66, IP67 when “Forward and reverse flow measurement functions”
is selected.
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-1 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
11. OUTLINE
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 11-2 All Rights Reserved. Copyright © 2006, Rota Yokogawa
11. OUTLINE
CAUTION
Pulse Output Accuracy:
*1: When performing the Insulation Resistance Test or the
Hard rubber Lining:
Withstand Voltage Test, please obey the following cau-
tion.
Flow Velocity • Following the relevant test, wait for more than 10 s
Size mm (in) Accuracy
V m/s (ft/s) after the power supply has been turned off before
15 (0.5) V < 0.25 (0.8) ±2.5 mm/s removing the cover.
to 0.25 ≤ V ≤ 10 • Remove all wires from terminals before testing.
400 (16) ±0.5 % of Rate ±1.25 mm/s • When the power terminal has a lightning protector
(0.8) (33)
(optional code A), remove the short bar at the ground
450 (18) V < 0.25 (0.8) ±3.75 mm/s
terminal.
to 0.25 ≤ V < 10
1000 (40) ±0.5 % of Rate ±2.5 mm/s • After testing, be sure to discharge by using a resistor
(0.8) (33)
T02.EPS
and return all wires and the short bar to its correct
position.
• Screws must be tightened to a torque of 1.18 Nm or
Size 15 mm (0.5 in) to 1000 mm (40 in)
more.
% of Rate
• After closing the cover, the power supply can be
2 restored.
Pulse
High grade
OutputAccuracy
Accuracy
1.5 High grade
15 - 400
Accuracy
Safety Requirement Standards:
High 450
grade
- 1000
Accuracy EN 61010-1
1
EN 61010-2-030
0.6
0.5 • Altitude at installation site: Max. 2000 m above sea level
• Installation category based on IEC 1010:
0 Overvoltage category II (“II” applies to electrical equip-
0 2.5 5 7.5 10 ment which is supplied from a fixed installation-like
V[m/s] F03.EPS distribution board.)
Size 25 mm (1.0 in) to 400 mm (16 in) • Pollution degree based on IEC 1010
Pollution degree 2 (“Pollution degree” describes the
degree to which a solid, liquid or gas which deteriorates
Current Output Accuracy: dielectric strength or surface resistivity is adhering. “2”
Pulse output accuracy ± 0.05 % of Span applies to a normal indoor atmosphere.)
Repeatability:
EMC Conformity Standards:
± 0.1 % of Rate (V ≥ 1 m/s (3.3 ft/s))
EN61326-1 Class A, Table 2 (For use in industrial locations)
± 0.05 % of Rate + 0.5 mm/s (V < 1 m/s (3.3 ft/s))
EN61326-2-3
Maximum Power Consumption: EN61000-3-2 Class A
Combined with RXF Remote Flowtube: 20 W EN61000-3-3
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-3 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
11. OUTLINE
15 2.5 0.1
Allowable cable length m(ft)
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 11-4 All Rights Reserved. Copyright © 2006, Rota Yokogawa
11. OUTLINE
Active Pulse Active pulses are output in order to drive an external electromagnetic or electronic counter directly using the
Output converter’s internal power supply. (Nullifies the standard transistor contact pulse output.)
Output voltage: 24 V DC ±20 % EM
Pulse specifications:
• At the drive current of 30 to 150 mA
Pulse rate: 0.0001 to 2 pps (pulse/s); Pulse width: 20, 33, 50, or 100 ms
Stainless Steel Screwed JIS SUS304 (AISI 304 SS/EN 1.4301 equivalent) stainless steel tag plate. Choose this option when SCT
Tag Plate an SS tag plate is required in addition to the standard nameplate with the tag number inscribed on it.
T05.EPS
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-5 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
11. OUTLINE
EXTERNAL DIMENSIONS
Remote Converter RXFA11G
Pipe mounting Unit: mm
180(7.09) 4-ø6 Hole 120(4.72) (approx. in)
(Surface mounting hole)
(2.56)
210(8.27)
65
194(7.64)
(67)
31 31 31 31 (2.64)
(1.22) (1.22) (1.22) (1.22)
(2.64) (0.94)(1.14) (1.93)
(67) 24 29 49
100(3.94)
1 to 10(0.04 to 0.39)
Panel mounting (Panel thickness) 172(6.77)
Mounting Screw
(M6 3100(3.94))
Panel cutout
203(7.99)
(4.13)
105
182(7.17)
R3 MAX
*1
Surface mounting Mounting Screw(M6) 18(0.71)
194(7.64)
62 (2.44)
T08.EPS
Unless otherwise specified, difference in the dimensions are refer to the following table :
General tolerance in the dimensional outline drawing.
Unit : mm (approx.inch)
Category of basic dimension Category of basic dimension
Tolerance Tolerance
Above Equal or below Above Equal or below
3 (0.12) ±0.7 (60.03) 500 (19.69) 630 (24.80) ±5.5 (62.17)
3 (0.12) 6 (0.24) ±0.9 (60.04) 630 (24.80) 800 (31.50) ±6.25 (60.25)
6 (0.24) 10 (0.39) ±1.1 (60.04) 800 (31.50) 1000 (39.37) ±7.0 (60.28)
10 (0.39) 18 (0.71) ±1.35 (60.05) 1000 (39.37) 1250 (49.21) ±8.25 (60.32)
18 (0.71) 30 (1.18) ±1.65 (60.06) 1250 (49.21) 1600 (62.99) ±9.75 (60.38)
30 (1.18) 50 (1.97) ±1.95 (60.08) 1600 (62.99) 2000 (78.74) ±11.5 (60.45)
50 (1.97) 80 (3.15) ±2.3 (60.09) 2000 (78.74) 2500 (98.43) ±14.0 (60.55)
80 (3.15) 120 (4.72) ±2.7 (60.11) 2500 (98.43) 3150 (124.02) ±16.5 (60.65)
120 (4.72) 180 (7.09) ±3.15 (60.12)
180 (7.09) 250 (9.84) ±3.6 (60.14)
250 (9.84) 315 (12.40) ±4.05 (60.16)
315 (12.40) 400 (15.75) ±4.45 (60.18)
400 (15.75) 500 (19.69) ±4.85 (60.19)
Remarks: The numeric is based on criteria of tolerance class IT18 in JIS B 0401.
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 11-6 All Rights Reserved. Copyright © 2006, Rota Yokogawa
11. OUTLINE
55(2.17)
(approx. in)
B
E
A
B
SA
SB
C
F
90(3.54)
90(3.54)
C
G
(WHITE) (BLACK) (RED)
D
(WHITE) (BLACK) (RED)
8(0.31)Max.
8(0.31)Max.
(5.91)
150+0.20
150+0.20
(5.91)
FLOW TUBE
CONVERTER
ø10.5(0.41)
L(SPECIFIED LENGTH)
20(0.79)
Model D E F G
AXFC AXFC-4 70(2.76) 25(0.98)50(1.97)60(2.36)
F04.EPS
RXFC-0-L
ø10.5(0.41)
CAP
AXFC
All Rights Reserved. Copyright © 2006, Rota Yokogawa 11-7 IM 01R21C01-01E-E 2nd edition October 01, 2013 -00
YOKOGAWA ELECTRIC CORPORATION YOKOGAWA ELECTRIC ASIA Pte. LTD. Yokogawa has an extensive sales and
World Headquarters 5 Bedok South Road
9-32, Nakacho 2-chome, Musashino-shi Singapore 469270 distribution network.
Tokyo 180-8750 Singapore
Japan www.yokogawa.com/sg Please refer to the European website
www.yokogawa.com
(www.yokogawa.com/eu) to contact your
YOKOGAWA CORPORATION OF AMERICA YOKOGAWA CHINA CO. LTD. nearest representative.
2 Dart Road 3F Tower D Cartelo Crocodile Building
Newnan GA 30265 No.568 West Tianshan Road Changing District
USA Shanghai, China
www.yokogawa.com/us www.yokogawa.com/cn
IM 01R21C01-01E-E 2nd edition October 01, 2013 -00 All Rights Reserved. Copyright © 2006, Rota Yokogawa
Subject to change without notice