Isa-Rp37.2-1982 (R1995)
Isa-Rp37.2-1982 (R1995)
ISA–RP37.2–1982 (R1995)
NOTICE OF COPYRIGHT
This is a copyrighted document and may not be copied or distributed in any
form or manner without the permission of ISA. This copy of the document was
made for the sole use of the person to whom ISA provided it and is subject to
the restrictions stated in ISA’s license to that person. It may not be provided to
any other person in print, electronic, or any other form. Violations of ISA’s
copyright will be prosecuted to the fullest extent of the law and may result in
substantial civil and criminal penalties.
ISA–The Instrumentation,
Systems, and
Automation Society
ISA–RP37.2–1982 (R1995)
Guide for Specifications and Tests for Piezoelectric Acceleration Transducers for Aerospace Testing
ISBN 0-87664-377-2
Copyright 1982 by the Instrument Society of America. All rights reserved. Printed in the United
States of America. No part of this publication may be reproduced, stored in a retrieval system, or
transmitted in any form or by any means (electronic, mechanical, photocopying, recording, or
otherwise), without the prior written permission of the publisher.
ISA
67 Alexander Drive
P.O. Box 12277
Research Triangle Park, North Carolina 27709
Preface
This preface, as well as all footnotes and annexes, is included for informational purposes and is
not part of ISA-RP37.2-1982 (R1995).
This recommended practice has been prepared as a part of the service of ISA, the international
society for measurement and control, toward a goal of uniformity in the field of instrumentation.
To be of real value, this document should not be static, but should be subject to periodic review.
Toward this end, the Society welcomes all comments and criticisms, and asks that they be
addressed to the Secretary, Standards and Practices Board; ISA; 67 Alexander Drive; P.O. Box
12277; Research Triangle Park, NC 27709; Telephone (919) 549-8411; Fax: (919) 549-8288;
E-mail: [email protected].
The ISA Standards and Practices Department is aware of the growing need for attention to the
metric system of units in general, and the International System of Units (SI) in particular, in the
preparation of instrumentation standards, recommended practices, and technical reports. The
Department is further aware of the benefits to USA users of ISA standards of incorporating
suitable references to the SI (and the metric system) in their business and professional dealings
with other countries. Toward this end, this Department will endeavor to introduce SI-acceptable
metric units in all new and revised standards to the greatest extent possible. The Metric Practice
Guide, which has been published by the Institute of Electrical and Electronics Engineers as
ANSI/IEEE Std. 268-1992, and future revisions, will be the reference guide for definitions,
symbols, abbreviations,and conversion factors.
CAUTION: The information presented within this ISA recommended practice is believed to be
accurate and reflects the current state of knowledge within the field. The information is an
interpretation and condensation of a larger volume of literature and experience, some of which is
contradictory and speculative. Therefore, application of the information to particular situations
requires the exercise of the independent professional judgment of the user. ISA is not
responsible for any results from such use of the information and shall not be liable for any
damages caused by such use.
It is the policy of ISA to encourage and welcome the participation of all concerned individuals and
interests in the development of ISA standards, recommended practices, and technical reports.
Participation in the ISA standards-making process by an individual in no way constitutes
endorsement by the employer of that individual, of ISA, or of any of the standards, recommended
practices, and technical reports that ISA develops.
The development of this recommended practice was initiated as a result of a survey conducted in
December 1960. A total of 240 questionnaires was sent out to transducer users and
manufacturers in the aerospace field. In their replies, a strong majority indicated a need for
standardization of specifications and tests of several types of commonly used aerospace test
transducers with electrical output.
On the basis of these replies, a Project Sub-Committee 8A-RP37.2, Guide for Piezoelectric
Acceleration Transducers for Aerospace Testing, was formed under the cognizance of Committee
8A-RP37, Transducers for Aerospace Testing, of the Aerospace Standards Group, and this
recommended practice was developed.
ISA-RP37.2-1982 (R1995) 3
The following individuals served as members of Subcommittee 8A-RP37.2:
NAME COMPANY
NAME COMPANY
The following individuals served on the ISA Committee SP37, who reaffirmed ISA-RP37.2
in 1995:
NAME COMPANY
4 ISA-RP37.2-1982 (R1995)
This recommended practice was reaffirmed by the ISA Standards and Practices Board on
September 29, 1995.
NAME COMPANY
ISA-RP37.2-1982 (R1995) 5
Contents
1 Scope ................................................................................................................................ 9
2 Purpose ............................................................................................................................. 9
3 Drawing symbol ............................................................................................................... 9
3.1 General .................................................................................................................. 9
3.2 Self-checking accelerometers ............................................................................. 10
4 Characteristics ............................................................................................................... 11
4.1 Design characteristics ......................................................................................... 12
4.2 Performance characteristics ................................................................................ 13
4.3 Additional terminology ......................................................................................... 14
5 Individual acceptance tests and calibration ................................................................ 16
5.1 Visual inspection ................................................................................................. 16
5.2 Initial functional tests ........................................................................................... 16
5.3 Tests and measurements .................................................................................... 18
5.4 Final functional tests ............................................................................................ 19
6 Qualification tests .......................................................................................................... 20
6.1 Effectiveness of mounting technique ................................................................... 20
6.2 Temperature range .............................................................................................. 21
6.3 Mounted resonant frequency ............................................................................... 21
6.4 Frequency response, maximum range (see 5.2.1 and 5.2.2.) ............................. 22
6.5 Amplitude linearity ............................................................................................... 22
6.6 Strain sensitivity .......................................................................................... 23 & 28
6.7 Temperature sensitivity error ....................................................................... 23 & 28
6.8 Transient temperature error ........................................................................ 23 & 28
6.9 Acoustic sensitivity ...................................................................................... 23 & 28
6.10 Electromagnetic response .......................................................................... 24 & 28
6.11 Transducer seal immersion test (Use on all moisture sealed or hermetically
sealed transducers)...................................................................................... 24 & 29
6.12 Other environmental effects ....................................................................... 24 & 29
6.13 Sensitivity stability ...................................................................................... 24 & 29
Annex A — References ..................................................................................................... 31
ISA-RP37.2-1982 (R1995) 7
1 Scope
1.1 This recommended practice covers piezoelectric acceleration transducers, primarily those
used in aerospace test instrumentation.
1.2 Terminology used in this document follows ISA-S37.1, Nomenclature and Specification
Terminology for Aerospace Test Transducers with Electrical Output, except that additional terms
considered applicable to piezoelectric vibration transducers are defined in 4.3.
2 Purpose
This recommended practice establishes the following for piezoelectric acceleration transducers:
2.1 Uniform minimum general specifications for design and performance characteristics.
2.2 Uniform minimum acceptance and qualification test methods, including calibration.
3 Drawing symbol
3.1 General
ISA-RP37.2-1982 (R1995) 9
3.2 Self-checking accelerometers
3.2.1 Active
3.2.2 Passive
10 ISA-RP37.2-1982 (R1995)
4 Characteristics
The first two paragraphs contain, in alphabetical order, the terms to be used in manufacturers'
and users' specifications for piezoelectric acceleration transducers. Some terms not already
defined in ISA-S37.1 (Nomenclature and Specification Terminology for Aerospace Test
Transducers with Electrical Output) are marked with an asterisk (*) and are defined in 4.3.
In 4.1 and 4.2, each characteristic listed is checked (or numbered with reference to Sections 5
or 6) in one or several columns to indicate the characteristics which are included in
manufacturers' literature or users' specifications as follows:
a) Basic — indicates characteristics which are normally specified for each
transducer.
b) Supplemental — indicates additional characteristics which may be specified if
desired.
The remaining three columns indicate the tests required to determine these characteristics.
a) Individual Acceptance Tests (IAT) — tests performed on each transducer.
1) Within Stated Tolerances - these tests are of the "go - no go" type with respect to
stated tolerances.
2) Measured Value - these tests provide individual measured values of the
characteristic.
b) Qualification tests — tests in addition to the IAT's performed on representative samples
of each transducer design. All IAT's are, of course, performed during a formal
qualification test program.
Some of the basic and supplemental characteristics are determined by the manufacturer's
research and development and require no additional testing.
ISA-RP37.2-1982 (R1995) 11
4.1 Design characteristics
*Defined in 4.3
12 ISA-RP37.2-1982 (R1995)
4.2 Performance characteristics
*Defined in 4.3
ISA-RP37.2-1982 (R1995) 13
4.3 Additional terminology
(For terms not defined here, see ISA-S37.1.)
4.3.1 acceleration limit: the maximum vibration and shock acceleration which the transducer can
accept in either direction along its sensitive axis without permanent damage, usually stated as
±____g's. The acceleration limits are usually much wider than the Acceleration Range and thereby
represent a measure of the overload capability of the transducer.
4.3.2 acceleration range: the range of accelerations over which the transducer has the specified
linearity.
4.3.3 acoustic sensitivity: the output of a transducer (not due to rigid body motions) in response
to a specified acoustical environment. This is sometimes expressed as the acceleration in g rms
sufficient to produce the same output as induced by a specified sound pressure level spectrum
having an overall value of 140 db referred to 0.0002 dyne per sq. cm. rms.
4.3.4 amplification factor at resonant frequency: the ratio of the maximum sensitivity of a
transducer (at its resonant frequency to its Reference Sensitivity).
4.3.5 amplitude linearity, shock: closeness of sensitivity to reference sensitivity over a stated
range of acceleration amplitudes, under shock conditions, usually specified as "within
±____percent for acceleration rise times longer than ____ microseconds."
4.3.6 amplitude linearity, vibration: closeness of sensitivity over a stated range of acceleration
amplitudes, at a stated fixed frequency, usually specified as "within ±____percent."
4.3.8 frequency response: the change with frequency of the sensitivity with respect to the
reference sensitivity, for a sinusoidally varying acceleration applied to a transducer within a stated
range of frequencies, usually specified as "within ±____ percent of the reference sensitivity from
____ to ____ cps." The applicable total capacitance and load resistance should be stated.
4.3.10 markings: information shown on the transducer itself, will normally include Manufacturer,
Model Number and Serial Number.
4.3.12 polarity: the relationship between the transducer output and the direction of the applied
acceleration; taken as "standard" when a positive charge or voltage appears on the "high" side of
the transducer for an acceleration directed from the mounting surface into the body of the
accelerometer.
14 ISA-RP37.2-1982 (R1995)
4.3.13 reference sensitivity (charge or voltage): the ratio of the change in charge or voltage
generated by a transducer to the change in value of the acceleration that is measured under a set
of defined conditions (amplitude, frequency, temperature, total capacitance, amplifier input
resistance, mounting torque). Deviations in sensitivity should be reported as deviations from the
reference sensitivity.
4.3.14 sensing mode of transduction element: the method used to stress the transduction
element such as compression, bending or shear.
4.3.15 sensitivity: the ratio of the change in transducer output to a change in the value of the
acceleration.
NOTES
1. Where one sensitivity under defined conditions is the basis for determining deviations
in performance, use "reference sensitivity."
2. Because the use of piezoelectric acceleration transducers for the measurement of both
shock and vibration, the acceleration is required to be known in either g peak or g rms.
A specified sensitivity in millivolts per g is to be understood as meaning "rms millivolts
per rms g" or its equivalent "peak millivolts per peak g." The use of "mixed units" such
as rms millivolts per peak is to be avoided. Note, however, that an output of 10 millivolts
rms is also approximately 14.1 millivolts peak, and an acceleration of 1 g rms is
approximately 1.41 g peak and, therefore,
mv
10 ------- mv rms mv pk mv rms mv pk
- = 10 ------------------- = 10 ---------------- = 7.07 ------------------- = 14.1 ----------------
g g rms g pk g pk g rms
4.3.16 shock: a substantial disturbance characterized by a rise and decay of acceleration from a
constant value in a short period of time.
4.3.17 strain sensitivity: the sensitivity to strains applied to the base by bending, in the absence
of any rigid body motion of the transducer. It is expressed as 10-6 times the equivalent acceleration
level in g's for a strain in the plane of the base.
4.3.18 temperature range, operating: the interval of temperatures in which the transducer is
intended to be used, specified by the limits of this interval.
4.3.19 temperature sensitivity error: the change in sensitivity of a transducer from its reference
sensitivity as a result of changes in its ambient temperature over a specified operating temperature
range.
NOTE — If changes in voltage sensitivity are specified, the total associated capacitance
must be stated.
4.3.20 transient temperature error: the output of a transducer as a result of a specified transient
temperature change within a specified operating temperature range.
NOTE — The associated capacitance and load resistance, as well as the time, after the
applied transient, at which the amplitude peak occurs must be specified.
ISA-RP37.2-1982 (R1995) 15
5 Individual acceptance tests and calibration
NOTES
1. The standard transducer used should have a calibration traceable (see ISA-S37.1) to
a calibration performed at the National Bureau of Standards within the normal calibration
period of one year. The transducer used as a standard should be reserved for this
purpose only; it should not be exposed to large values of shock, vibration or temperature
extremes; and its calibration should be checked periodically by either of the referenced
methods.
2. The surface on which each transducer is mounted and the part of the transducer base
which touches that surface should be clean and flat, with a surface finish of
64 microinches or less. If oil or grease is used as a gasketing material, it should be
clean, freshly applied to both surfaces just before the test, and completely removed
immediately afterwards.
3. The signal used to excite the motion of the transducers should be as nearly sinusoidal
as possible. The wave shape of the output signal of both transducers should be
observed frequently throughout the test and no perceptible distortion should be allowed.
Preferably, a distortion meter should be used, and the distortion kept below 3%.
16 ISA-RP37.2-1982 (R1995)
4. Screw-attached transducers should be mounted with the torque recommended by the
manufacturer, using a good grade torque wrench. A preliminary observation of the
wave shape and amplitude on an oscilloscope should be made when the transducer is
mounted with a torque about 10% less than recommended. Then the recommended
torque should be applied. A calibration is not valid if a small increase in mounting torque
changes the wave shape or amplitude of the output appreciably. Torque larger than
recommended should never be used unless it is certain that no damage will result. The
torque used should be reported with the results of the test.
5. The plate, on opposite sides of which the transducers are mounted, should be thick
enough so that no appreciable flexure occurs and thin enough so that both pickups
have the same motion.
6. The motion applied to the two transducers should be far enough above the noise level
so that the system noise represents a minor error in the calibration. A signal-to-noise
ratio of 40 db is desirable. Increasing the motion beyond this level is not desirable on
most vibration test equipment because it increases the chance of distortion and
non-axial motion. The acceleration level should be reported with the results of the test.
NOTES
1. The plate on which the two transducers are mounted should be made of a high-modulus
material such as a machinable tungsten alloy.
2. If the signal from one transducer is applied to one set of plates of an oscilloscope, and
the signal from the other transducer is applied to the other set of plates, rotation of the
resulting Lissajous figure as the driving frequency is increased from very low to high
frequencies indicates the onset of difference in the motion of the transducers. Distortion
of the figure indicates bad motion or other cause for investigation. The test should be
repeated with the transducer positions interchanged. A calibration is not valid if
interchanging positions changes results by more than 3% (5% above 5,000 cps).
ISA-RP37.2-1982 (R1995) 17
5.3 Tests and measurements
5.3.1 Transducer seal immersion test
Use water at room temperature in a transparent container such as a Pyrex® beaker. Heat the
water to approximately 200°F. Remove detachable cables and connectors from the transducer.
Immerse the transducer beneath the surface of the heated water. Any stream of air bubbles
released from the transducer indicates leakage and constitutes failure. Dry the transducer
without application of heat and measure the insulation resistance of the element. The minimum
insulation resistance shall be met.
Connect the cable electrically as shown in the schematic. Clamp the cable between pieces of
wood to the two anchor points, allowing a 3-inch sag in the center of the cable. Clamp weight (B)
to cable at center of anchored span. Raise the cable by the weight to maximum height (3 inches
above its neutral position) and drop. Monitor output noise of the cable on the oscilloscope.
Repeat test 3 times. Record the maximum value in mv peak to peak.
18 ISA-RP37.2-1982 (R1995)
5.3.3 Transducer capacitance
Measure capacitance of transducer with cable attached by means of a capacitance bridge at
1000 cps; measure capacitance of transducer cable at 1000 cps. Subtract cable capacitance
from total capacitance to obtain transducer capacitance. Temperature should be 70°F to 85°F.
5.3.4 Polarity (use either the transient method or the comparison method.)
NOTE — Back-to-back mounting results in opposite motions being applied to the two
accelerometers at any instant.
1. The standard which is used to establish the vibration amplitude must be known to be
suitable under all test conditions.
2. Measure capacitances at 1000 cps with a capacitance bridge, without motion applied
to the transducer.
ISA-RP37.2-1982 (R1995) 19
3. Measure resistance with a megohmmeter. Test voltage should be 50 vdc unless
otherwise specified.
4. Allowance may have to be made for changes in cable resistance and/or cable
capacitance throughout the transducer temperature range.
5. Monitor the output waveform of the transducer with an oscilloscope to insure that it
appears sinusoidal.
6. Do not leave the transducer in an open circuit condition while the temperature is being
changed; it should be shorted or connected to a preamplifier.
Repeat the room temperature tests as above at the conclusion of the final functional tests.
If applicable, compute the percentage change in transducer voltage sensitivity at rated
temperature compared with room temperature.
If applicable, compute the percentage change in transducer charge sensitivity at rated
temperature compared with room temperature:
( C tmt + C ext ) E mt – ( C trt + C ext ) E rt
100 ---------------------------------------------------------------------------------------------------
( C trt + C ext ) E rt
Low resistance or excessive sensitivity change with temperature may constitute failure.
Nonrepeatability of room temperature output before and after the test may constitute failure.
6 Qualification tests
Tests which are performed on representative transducers in addition to the Individual Acceptance
Tests.
6.1.2 With electrically insulated mounting studs, apply the rated maximum mounting torque to the
stud several times; reject if there is visible evidence of mechanical damage. Use the tested
mounting stud if possible when conducting the final frequency response test. (See 5.4.3).
20 ISA-RP37.2-1982 (R1995)
6.2 Temperature range
6.2.1 Resistance and Capacitance at Maximum Rated Temperature (delete if rated higher than
350°F as measured in 5.4.1).
Insert transducer with cable attached into an oven with the temperature sensing thermocouple in
contact with or adjacent to the base of the transducer. Short the open end of the cable. Increase
oven temperature and stabilize at the maximum rated temperature for 15 minutes. Remove short
and measure the resistance of the transducer-cable combination using a 50 volt megohmmeter.
Measure the capacitance at 1000 cps with a capacitance bridge. Again short the open end of the
cable and return to room temperature. If the resistance is not acceptable, re-run the test with the
cable alone to determine whether the cause of low resistance is within the transducer or the
cable assembly.
In the same way measure the insulation resistance from both sides of the element to ground for
ungrounded transducers.
NOTE — All connectors should be wiped with clean alcohol and a dry cloth before the test.
ISA-RP37.2-1982 (R1995) 21
6.3.1 Sinusoidal method
Measure the accelerometer output using the comparison method throughout the frequency range
including and above the resonant frequency of the accelerometer. The resonant frequency is the
frequency of maximum sensitivity. The phase angle relative to the standard accelerometer
changes by almost 180 degrees in the range of frequencies near the resonant frequency of the
accelerometer.
The standard accelerometer should be built into the exciter. The resonant frequency of the
exciter is determined by measuring the transfer impedance between the driver coil current and
the standard accelerometer output. Use a dummy mass load to simulate a transducer during the
transfer impedance test. Use the exciter only throughout the range up to 95 percent of its
resonant frequency.
NOTE — The low-frequency rigid-body rise in amplitude and change in phase may be
ignored.
The resonant frequency of the exciter may also be determined with an accelerometer whose
damped natural frequency has been determined by the Shock Method and found to be above the
exciter resonant frequency.
22 ISA-RP37.2-1982 (R1995)
6.5.2 Shock linearity
Use a ballistic impact or comparison technique similar to that outlined in ANSI-S2.2-1959,
Sections 4.6 and 5, respectively. Measure and record the basic sensitivity before and after the
shock linearity tests.
6.6 Strain sensitivity
The technique used to measure strain sensitivity of an accelerometer meets the requirement of
ANSI- Z24.21-1957 reference, paragraph 3.1.3.7. The accelerometer is mounted on a simple
cantilever beam. The radius of curvature at the point where the accelerometer is mounted is
1000 inches when the measurements are taken.
A steel beam is held as a cantilever in a vice bolted to a concrete floor. The beam is 3.0" wide by
0.5" thick and 60" long. (The free length is approximately 57 inches.) The natural frequency is
very close to 5 cps. Four strain gages are bonded to the beam adjacent to the accelerometer
mounting hole (two each, top and bottom, about 1.5" from the edge of the clamp.) A two-channel
recorder is used to record the output of both the strain gage bridge and the accelerometer under
test.
The system is excited by manually deflecting the free end of the beam and allowing it to vibrate
freely. The output of the accelerometer is taken from the oscillograph record at a point where the
strain in the surface of the beam is 250 x 10–6 inch per inch. (This is equivalent to a radius of
curvature of 1000 inches.) The strain sensitivity, in g's, for a strain of 10–6 inch per inch is found
by dividing the above accelerometer output by 250 times the accelerometer sensitivity in millivolts
per g.
ISA-RP37.2-1982 (R1995) 23
the sound pressure reaching the transducer generally causes vibratory accelerations of
the structure to which the accelerometer is attached of sufficient magnitude to make
negligible any output due to pressure changes alone. Tests for acoustic sensitivity, including
the one suggested here, tend to be conservative in that they yield an electrical output which
is too large because it includes signals due to rigid-body motions of the accelerometer.
NOTE — Induced mechanical vibrations and electrical ground loops must be eliminated
from the test set up.
6.11 Transducer seal immersion test (Use on all moisture sealed or hermetically sealed
transducers).
De-aerate several inches of water at 70°F in a closed transparent container by reducing the
absolute pressure to about 2 psi. Return the pressure to normal and immediately immerse the
transducer beneath the surface. Reduce the absolute pressure to about 2 psi. Air leakage from
the transducer constitutes failure. Let the pressure return to normal atmospheric with the
transducer still submerged. Dry the transducer without the application of heat and measure the
insulation resistance of the element. The minimum insulation resistance shall be met.
24 ISA-RP37.2-1982 (R1995)
c) Soak each transducer at minimum rated temperature for one hour. Return to room
temperature by allowing to warm for 24 hours. Measure and record new reference
sensitivity for each.
For each transducer, note the maximum percentage change from the original reference sensitivity
anywhere during the above tests, and calculate the arithmetical average of these for the three
transducers. This is a measure of the average Sensitivity Stability of the transducer.
NOTE — The manufacturer may have data on stability sensitivity of any transducer design
based on long term measurements. These measurements are compiled over years' time by
periodic (monthly) reference sensitivity measurements on transducers not otherwise in use.
ISA-RP37.2-1982 (R1995) 25
Model No. Part No.
Reference
Frequency Amplitude Output T1 T2 T3 T4
Sensitivity Temp.
cps g’s mv or pc* °F °F °F °F
°F
Resistance meg
ohm
Transducer
Capacitance pf
Cable
Capacitance pf
System
Capacitance pf
Output mv
% Change
BY______________________DATE_________________APPROVED_____________________
NOTE—Numbers in parentheses refer to sections in ISA-RP37.2.
APPROVED___________________________________________________________________
STAMP
26 ISA-RP37.2-1982 (R1995)
Model No. Part No.
Case Material
Transducer:
Torque Sensitivity at ____ cps
Maximum
Rated
75% Rated
50% Rated
Stud: Maximum Torque Rating ________________ Evidence of Damage _____________
ISA-RP37.2-1982 (R1995) 27
6.5 Amplitude linearity
Preamplifier Model ________________ Gain _________________
Cable Capacitance __________________ pf
VIBRATION SHOCK
Amplitude Signal Sensitivity Deviation from Amplitude Signal Sensitivity Deviation from
Ref. Sens. Ref. Sens.
28 ISA-RP37.2-1982 (R1995)
6.10 Electromagnetic response
Preamplifier Model ________ Gain ________ Cable Capacitance __________ pf
Mounting Material _________ Weight __________
Magnetic field level __________gauss at _________ cps
Transducer output __________ mv _____________ g/gauss
6.11 Transducer seal
Water Temperature ________ °F
Pressure ________ psia
Air Leakage?
Insulation Resistance across Element _______ meg Ω; to ground _______ meg Ω
6.12 Other environmental effects
Give details of tests and results of measurements.
Serial No. Before Tests After (1) 24 hrs. after (2) After (3)
Serial No. Before Tests After (1) 24 hrs. after (2) After (3)
TRANSDUCER
CAPACITANCE (pf)
CABLE
CAPACITANCE (pf)
BY______________________DATE_________________APPROVED______________________
NOTE—Test numbers and numbers in parentheses refer to procedures in ISA-RP37.2.
APPROVED_____________________________________________________________________
STAMP
ISA-RP37.2-1982 (R1995) 29
Annex A — References
Jones, E., Edelman, S., Sizemore, K. O., "Calibration of Vibration Pickups at Large
Amplitudes, " Journal of the Acoustical Society of America, 33, p. 1462, November
1961.
Nisbet, J. S., Brennan, J. N., and Tarpley, H. I., "High Frequency Strain Gage and
Accelerometer Calibration," Journal of the Acoustical Society of America, 32, p. 71,
Jan. 1960.
Schmidt, V. A., Edelman, S., Smith, E. R., and Pierce, E. T., "Modulated
Photoelectric Measurement of Vibration," Journal of the Acoustical Society of
America, 34, p. 355, April 1962.
ISA
ISA-RP37.2-1982 (R1995) 31
That of Gravity by the 'Chatter' Method," Proceedings of ISA, Vol. 9, Part V, 1954,
(Paper 50-40-1).
MISCELLANEOUS
Clements, E. W., and Stone, M. G., "Techniques for the Rapid Estimation of
Accelerometer Natural Frequencies," Report No. 5681, U. S. Naval Research
Laboratory, October 1961.
Den Hartog, J. P., Mechanical Vibrations , New York, Mc Graw-Hill, 1947, Chapter
II, Section 9.10
Harris, Cyril M. and Crede, Charles E., Shock and Vibration Handbook, 1961,
3 volumes
Harrison, M., Sykes, A. O., and Marcotte, P. G., The Reciprocity Calibration of
Piezoelectric Accelerometers, Washington, D. C., David Taylor Model Basin,
March 1952, (Report R-811).
Perls, T. A., "A Simple Objective Test for Cable Noise Due to Shock, Vibration or
Transient Pressures," PB 121583, Office of Technical Services, U. S. Government
Printing Office, 1955.
32 ISA-RP37.2-1982 (R1995)
Roberts, H. C., Mechanical Measurements by Electrical Methods, Pittsburgh,
Instruments Publishing, p. 357,1946.
ISA-RP37.2-1982 (R1995) 33
Developing and promulgating technically sound consensus standards,
recommended practices, and technical reports is one of ISA's primary
goals. To achieve this goal the Standards and Practices Department
relies on the technical expertise and efforts of volunteer committee
members, chairmen, and reviewers.
ISA is an American National Standards Institute (ANSI) accredited
organization. ISA administers United States Technical Advisory
Groups (USTAGs) and provides secretariat support for International
Electrotechnical Commission (IEC) and International Organization for
Standardization (ISO) committees that develop process measurement
and control standards. To obtain additional information on the
Society's standards program, please write:
ISA
Attn: Standards Department
67 Alexander Drive
P.O. Box 12277
Research Triangle Park, NC 27709
ISBN: 0-87664-377-2