Illuminance Meters - Requirements and Test Methods: British Standard BS 667:2005
Illuminance Meters - Requirements and Test Methods: British Standard BS 667:2005
Illuminance meters —
Requirements and test
methods
Confirmed
ICS 17.180.20; 29.140.40
December 2011
12&23<,1*:,7+287%6,3(50,66,21(;&(37$63(50,77('%<&23<5,*+7/$:
BS 667:2005
Contents
Page
Committees responsible Inside front cover
Foreword ii
Introduction 1
1 Scope 1
2 Normative references 1
3 Terms and definitions 1
4 Classification 2
5 Digital ranges and readouts 3
6 Analogue scales 3
7 Calibration and testing temperature 3
8 Power supply 3
9 Performance requirements for laboratory and field illuminance meters 3
10 Marking 5
Annex A (informative) Guidance on choice of meter range 6
Annex B (normative) Calibration 6
Annex C (normative) Test methods 8
Bibliography 13
Table 1 — Test and calibration methods 4
Table 2 — Error tolerances for illuminance meters 4
Table C.1 — Example of calculation of correction factor for a high pressure
sodium lamp 12
Foreword
Summary of pages
This document consists of a front cover, an inside front cover, pages i and ii,
pages 1 to 13 and a back cover.
The BSI copyright notice displayed in this document indicates when the
document was last issued.
Introduction
As lighting levels reduce, the response characteristics of the human eye change from photopic during the
day through an intermediate mesopic region at dusk to a scotopic response at night time.
At present, eye response functions have not been defined in the mesopic region; research work is still in
progress on this subject. Although photopic measurements are generally only applicable at lighting levels
above about 10 lux, in practice lighting levels for installations are specified at values significantly below
this. It is therefore necessary to make illuminance measurements with known accuracy at these lower
levels to check performance against such specifications, whilst accepting that these measurements do not
accurately represent the visual response.
Photometric measurements, even under laboratory conditions, are more difficult to determine precisely
than many other kinds of physical measurement, and in the case of illuminance meters, liability to error is
increased by the fact that many of the essential components of such instruments are susceptible to
variation with time and use. These errors can be minimized by care in the design of the instrument as a
whole, and also by the user in not exposing it to deleterious conditions of temperature, illumination or
atmosphere.
It is recommended that illuminance meters should be returned to the manufacturer, or a competent
photometric testing authority at intervals as recommended by the manufacturer for checking and, if
necessary, recalibration.
1 Scope
This British Standard specifies performance requirements for illuminance meters for the measurement of
photopic illuminance for applications other than the measurement of luminaires. It specifies the
performance requirements for two types of illuminance meter, Type L (laboratory instruments) and Type F
(field instruments). This British Standard is applicable to meters for the measurement of planar
illuminance. This standard is intended for use by meter manufacturers and users.
NOTE 1 The requirements specified for a Type L meter in this standard are identical to the requirements specified for an
illuminance meter in BS EN 13032-1. However, BS EN 13032-1 specifies additional requirements which are not specified in this
standard.
NOTE 2 If other types of illuminance e.g. semi-cylindrical, are used, reference should be made to CIE Publication 69:1987 [1].
NOTE 3 Annex A gives recommendations for the choice of meter range.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
BS ISO 10526:1999, CIE standard illuminants for colorimetry.
CIE Publication 18.2:1983, The basis of physical photometry. Second edition.
3.4
selective or coloured filter
medium which changes the spectral distribution of radiation by transmission
3.5
colour temperature
temperature of the full radiator which emits radiation of the same (or nearly the same) chromaticity as the
radiation considered
3.6
photometer head
light sensitive detector the spectral response of which is weighted (usually by means of coloured filters) to
approximate the CIE spectral luminous efficiency function V(2) of the human eye for photopic vision, as
defined in CIE 18.2, together with any cosine correction required
NOTE More information on the spectral response and cosine correction is given in C.2 and in C.5 respectively.
3.7
calibration plane
plane normal to the axis of the photometer head to which the measured illuminance values relate
NOTE This plane should be chosen so that the readings are in accordance with the inverse square law when using a point source.
In the case of a photometer head without a diffuser, this plane is the plane of the limiting aperture; for a photometer head with a
plane diffuser, it is the front of the diffuser.
3.8
effective range
range over which the instrument conforms to the accuracy requirements of this specification
3.9
secondary standard lamp
lamp, the photometric characteristics of which are accurately known and calibrated directly against a
national standard
3.10
working standard lamp
lamp in regular use which is calibrated at regular intervals by reference to a secondary standard lamp
3.11
secondary standard illuminance meter
illuminance meter, the photometric characteristics of which are accurately known and calibrated directly
against a national standard
NOTE Further definitions related to photometry are given in CIE Publication 17.4 [2].
4 Classification
Illuminance meters shall be classified as one of the following two types:
a) laboratory illuminance meters, designated Type L, which are generally retained in a laboratory or
standardizing area, and against which other equipment may be calibrated and the highest precision
readings obtained. Such meters may not be suitable for site or field measurements;
b) field illuminance meters, designated Type F, which are for use in the field, on site and in the working
environment. Some accuracy may have to be sacrificed in the interests of the ease of use, robustness and
versatility of field meters.
6 Analogue scales
For multi-range analogue illuminance meters, the factor between adjacent ranges shall not exceed four.
The scales of the illuminance meters shall be marked in lux or multiples of lux with bold graduations, and
so figured as to minimize the chance of error in reading, even in a poor light.
NOTE 1 A built-in lamp is useful for reading the scale in poor illumination, but care should be taken that this light does not affect
the readings.
NOTE 2 BS 3693 gives recommendations for scale marking and figuring.
NOTE 3 The part of the scale below the effective range of the illuminance meter may be unmarked.
8 Power supply
For battery operated instruments a battery check or other warning shall be provided to show when the
battery should be replaced to ensure accuracy. Mains powered instruments shall conform to Clause 9 for
input voltages within the range 207 V to 253 V.
NOTE 1 For digital displays displaying three significant digits there is a tolerance of ±1 on the least significant digit, which
corresponds to 1 % for a reading of 100 and 0.2 % for a reading of 500.
NOTE 2 A meter which just meets the requirements of this standard would have a best measurement capability of ±4 % (Type L)
or ±6 % (Type F) when used on any of its calibrated ranges. For highly coloured sources, such as light emitting diodes (LEDs), larger
uncertainties would apply in practice.
10 Marking
The following information shall be distinctly and durably marked on the illuminance meter:
a) the mark of origin (this may take the form of a trade mark, the manufacturer’s identification mark or
the name of the vendor);
b) the number of this British Standard, i.e. BS 667:20051);
c) type, i.e. L or F;
d) instrument identification number.
To ensure proper use and maintenance, the following shall be marked on the illuminance meters or
otherwise made available to the purchaser:
1) spectral correction error;
2) the calibration and testing temperature;
3) the date, source used and error of the last calibration and the recommended date of the next
calibration;
4) where the photoelectric cell can be disconnected from the indicating instrument they shall both be
marked to ensure that the correct combination of photoelectric cell and indicating instrument are used
together;
5) correction factors for differences between the ambient temperature and testing temperature;
6) the location of the calibration plane; and
7) if the photometer head is not cosine corrected, this shall be stated.
NOTE Correction factors for specified non-Plankian light sources may also be given, see C.9.
1) Marking BS 667:2005 on or in relation to a product represents a manufacturer’s declaration of conformity, i.e. a claim by or on
behalf of the manufacturer that the product meets the requirements of the standard. The accuracy of the claim is solely the
claimant’s responsibility. Such a declaration is not to be confused with third party certification of conformity.
Annex A (informative)
Guidance on choice of meter range
A.1 General
Most illuminance meters cover more than one range; there is usually a small but sometimes
significant difference between the readings made at the top of the lower range and the bottom of the next
higher range. This applies to both analogue and digital meters, but the effect is generally more obvious
with digital meters because the lower figure is usually read more easily with greater precision, and it is not
seen in comparison with the full scale as with an analogue meter.
A.2 Choice of range
Sometimes use can be made of the overlap between the ranges, or it may be worth working in a
higher range to avoid the problem of making range change corrections. However, where practical, the lower
range should be used, as it is more sensitive, the maximum reading in the range is frequently taken as
the calibration point, and zero and linearity errors will tend to increase towards the minimum reading in
the range.
Annex B (normative)
Calibration
B.1 General
Meters shall be calibrated using one of the methods given in B.2. The uncertainty of the calibration
including the uncertainty of the standard lamp or reference meter shall not exceed the value given in
Table 2. If, following calibration, the meter is adjusted, it shall be recalibrated before use.
B.2 Calibration methods
B.2.1 Using a reference lamp
B.2.1.1 Apparatus
B.2.1.1.1 Reference lamp, comprising a standard tungsten filament lamp calibrated for directional
intensity and having a colour temperature of 2 856 K ± 20 K. The lamp shall be operated under the
conditions for which it was standardized for luminous intensity and colour temperature, using a power
supply which can control the stability of supplied current to within 0.02 %.
NOTE It is good practice to use a second working standard lamp to check the calibration.
B.2.1.1.2 Photometric bench, or other suitable rigid structure on which the reference lamp and photometer
head cell are mounted, and on which distances can be measured accurately.
B.2.1.1.3 Baffles or diaphragms, which exclude all but direct rays from the reference lamp from the
photometer head.
B.2.1.2 Procedure
Adjust the position of the photometer head so that the illumination is normal to its geometric centre.
Adjust the illuminance meter before calibration testing to indicate zero with zero illuminance on the
photometer head.
Adjust the distance between the reference lamp and the calibration plane of the photometer head to achieve
suitable illuminance values; at least one value for each meter range.
The minimum distance between the reference lamp and the photometer head shall be greater than
10 times the maximum lamp filament dimension or 10 times the photometer head acceptance dimension if
greater.
Cover the photometer head between taking measurements at each successive point of calibration and
expose it to the illuminant for sufficient time for the reading to settle before taking a measurement.
Adjust the meter to give the same illuminance reading as that calculated in B.2.1.3.
I
E=
d2
where
E is the illuminance, in lux;
I is the luminous intensity, in candelas, of the reference lamp in the marked direction;
d is the distance, in metres, from the optical centre of the reference lamp filament to the
calibration plane of the photometer head.
Annex C (normative)
Test methods
C.1 Test for linearity
C.1.1 Apparatus
Means of illumination for checking linearity that can be varied by a physical method such as varying the
distance from the light source to the photometer head or using a series of light sources which are exposed
separately and in combination.
C.1.2 Procedure
Expose the photometer head to an illuminance of within ±5 % of the value used to calibrate the meter range
being used. Expose the photometer head to a series of illuminances covering each range on the meter being
tested, and note the values displayed by the meter together with the settings of the apparatus producing
the illuminance.
C.1.3 Calculation of non-linearity
Calculate the percentage non-linearity (N) using the following equation:
AB
N = 1 − × 100
CD
where
A is the meter reading at the test illuminance;
C is the meter reading when exposed to the illuminance within ±5 % of the calibration point;
B/D is the ratio of the illuminance, within ±5 %, of the calibration point to the illuminance at the
check point.
780
∑ s ∗ ( λ )rel − V ( λ )
f1′ = 380 × 100
780
∑ V (λ)
380
where
s*(2)rel is the normalized relative spectral responsivity as given by the following equation:
780
∑ S ( λ )A V ( λ )
s ∗ ( λ )rel = 380
s ( λ )rel
780
∑ S ( λ )A s ( λ )rel
380
S(2)A is the spectral distribution of the illuminant used in the calibration (standard illuminant A
in accordance with BS ISO 10526:1999);
s(2)rel is the relative spectral responsivity normalized at an arbitrary wavelength;
V(2) is the spectral luminous efficiency of the human eye for photopic vision, as defined in
CIE 18.2.
2) For information on the availability of infra-red transmitting, visible radiation absorbing filters contact BSI Customer Services,
C.5.2.3 Repeat the process as in C.5.2.2, but in a plane at right angles to that first chosen.
C.5.3 Expression of results
Calculate the sum (Ö) of the moduli of the errors at each angle from the following equation:
+85°
σ = ∑ f (θ ) − cos(θ )
−85°
Calculate the percentage cosine correction error (T) given by the equation:
σ
T = × 100
22.9 × 2
NOTE 22.9 = {cos 85°+ cos 80° + cos 75° +,...,+ cos (–85°)}
H −L
R= × 100
L
where
L is the reading, in lux, on the lower range; and
H is the reading, in lux, on the higher range.
The correct value of illuminance (E) of the test source is given by:
E = F × Et
where
C.9.2 Calculation of F
Calculate F from the following equation:
780 780
∑ St ( λ )V ( λ ) × ∑ Sr (λ ) s(λ )
F = 380 380
780 780
∑ St ( λ ) s(λ ) × ∑ Sr (λ )V (λ )
380 380
where
s(2) is the relative spectral responsivity of the illuminance meter;
Sr(2) is the spectral power distribution of the reference source used to calibrate the illuminance
meter;
St(2) is the spectral power distribution of the source to be measured;
V(2) is the CIE spectral luminous efficiency function of the meter, as defined in CIE 18.2;
and the summations are carried out over the range 380 nm to 780 nm in maximum steps of 10 nm.
C.9.3 Example of calculation of correction factor for high pressure sodium lamp
Example data for the calculation of the correction factor for a high pressure sodium lamp is given in
Table C.1.
Table C.1 — Example of calculation of correction factor for a high pressure sodium lamp
Wavelength V(2) s(2) Sr(2) St(2) St(2)V(2) Sr(2)s(2) St(2)s(2) Sr(2)V(2)
nm
380 0.000 0 0.000 0 0.098 0.010 7 0.000 0 0.000 0 0.000 0 0.000 0
390 0.000 1 0.000 0 0.121 0.013 9 0.000 0 0.000 0 0.000 0 0.000 0
400 0.000 4 0.000 0 0.147 0.018 6 0.000 0 0.000 0 0.000 0 0.000 1
410 0.001 2 0.000 1 0.177 0.022 7 0.000 0 0.000 0 0.000 0 0.000 2
420 0.004 0 0.000 2 0.210 0.027 5 0.000 1 0.000 0 0.000 0 0.000 8
430 0.011 6 0.001 1 0.247 0.034 4 0.000 4 0.000 3 0.000 0 0.002 9
440 0.023 0 0.003 8 0.287 0.041 8 0.001 0 0.001 1 0.000 2 0.006 6
450 0.038 0 0.005 8 0.331 0.058 3 0.002 2 0.001 9 0.000 3 0.012 6
460 0.060 0 0.025 2 0.378 0.033 8 0.002 0 0.009 5 0.000 9 0.022 7
470 0.091 0 0.066 3 0.429 0.961 0 0.087 5 0.028 4 0.063 7 0.039 0
480 0.139 0 0.136 2 0.482 0.017 8 0.002 5 0.065 6 0.002 4 0.067 0
490 0.208 0 0.214 5 0.539 0.020 1 0.004 2 0.115 6 0.004 3 0.112 1
500 0.323 0 0.382 4 0.599 0.221 0 0.071 4 0.229 1 0.084 5 0.193 5
510 0.503 0 0.544 2 0.661 0.025 8 0.013 0 0.359 7 0.014 0 0.332 5
520 0.710 0 0.732 5 0.725 0.037 1 0.026 3 0.531 1 0.027 2 0.514 8
530 0.862 0 0.896 7 0.791 0.012 3 0.010 6 0.709 3 0.011 0 0.681 8
540 0.954 0 0.969 3 0.859 0.016 6 0.015 8 0.832 6 0.016 1 0.819 5
550 0.995 0 1.002 0 0.929 0.061 7 0.061 4 0.930 9 0.061 8 0.924 4
560 0.995 0 0.964 3 1.000 0.137 1 0.136 4 0.964 3 0.132 2 0.995 0
570 0.952 0 0.905 8 1.072 0.839 0 0.798 7 0.971 0 0.760 0 1.020 5
580 0.870 0 0.834 7 1.144 0.665 9 0.579 3 0.954 9 0.555 8 0.995 3
590 0.757 0 0.721 1 1.217 0.997 6 0.755 2 0.877 6 0.719 4 0.921 3
600 0.631 0 0.583 2 1.290 1.000 0 0.631 0 0.752 3 0.583 2 0.814 0
610 0.503 0 0.464 2 1.363 0.478 5 0.240 7 0.632 7 0.222 1 0.685 6
620 0.381 0 0.339 9 1.436 0.343 4 0.130 8 0.488 1 0.116 7 0.547 1
630 0.265 0 0.235 5 1.508 0.175 1 0.046 4 0.355 1 0.041 2 0.399 6
640 0.175 0 0.158 2 1.580 0.135 4 0.023 7 0.250 0 0.021 4 0.276 5
650 0.107 0 0.099 2 1.650 0.110 7 0.011 8 0.163 7 0.011 0 0.176 6
660 0.061 0 0.058 9 1.720 0.095 9 0.005 8 0.101 3 0.005 6 0.104 9
670 0.032 0 0.033 8 1.788 0.095 9 0.003 1 0.060 4 0.003 2 0.057 2
680 0.017 0 0.018 2 1.854 0.074 9 0.001 3 0.033 7 0.001 4 0.031 5
690 0.008 2 0.009 1 1.919 0.046 8 0.000 4 0.017 5 0.000 4 0.015 7
700 0.004 1 0.004 5 1.983 0.038 6 0.000 2 0.008 9 0.000 2 0.008 1
710 0.002 1 0.002 1 2.044 0.035 9 0.000 1 0.004 3 0.000 1 0.004 3
720 0.001 0 0.001 0 2.104 0.033 8 0.000 0 0.002 1 0.000 0 0.002 1
730 0.000 5 0.000 4 2.161 0.032 5 0.000 0 0.000 9 0.000 0 0.001 1
740 0.000 2 0.000 2 2.217 0.032 0 0.000 0 0.000 4 0.000 0 0.000 4
750 0.000 1 0.000 1 2.270 0.034 4 0.000 0 0.000 2 0.000 0 0.000 2
760 0.000 1 0.000 0 2.321 0.043 1 0.000 0 0.000 0 0.000 0 0.000 2
770 0.000 0 0.000 0 2.370 0.780 0 0.000 0 0.000 0 0.000 0 0.000 0
780 0.000 0 0.000 0 2.417 0.034 9 0.000 0 0.000 0 0.000 0 0.000 0
Sum = 3.633 10.455 3.461 10.788
F = (3.663 × 10.455)/(3.461 × 10.788); i.e. F = 1.026 (2.6 %)
Bibliography
Standards publications
BS 3693:1992, Recommendations for design of scales and indexes on analogue indicating instruments.
Other publications
[1] CIE Publication 69:1987, Methods of characterizing illuminance meters and luminance meters:
Performance, characteristics and specifications.
[2] CIE Publication 17.4:1987, International lighting vocabulary.
Revisions
British Standards are updated by amendment or revision. Users of
British Standards should make sure that they possess the latest amendments or
editions.
It is the constant aim of BSI to improve the quality of our products and services.
We would be grateful if anyone finding an inaccuracy or ambiguity while using
this British Standard would inform the Secretary of the technical committee
responsible, the identity of which can be found on the inside front cover.
Tel: +44 (0)20 8996 9000. Fax: +44 (0)20 8996 7400.
BSI offers members an individual updating service called PLUS which ensures
that subscribers automatically receive the latest editions of standards.
Buying standards
Orders for all BSI, international and foreign standards publications should be
addressed to Customer Services. Tel: +44 (0)20 8996 9001.
Fax: +44 (0)20 8996 7001. Email: [email protected]. Standards are also
available from the BSI website at http://www.bsi-global.com.
In response to orders for international standards, it is BSI policy to supply the
BSI implementation of those that have been published as British Standards,
unless otherwise requested.
Information on standards
BSI provides a wide range of information on national, European and
international standards through its Library and its Technical Help to Exporters
Service. Various BSI electronic information services are also available which give
details on all its products and services. Contact the Information Centre.
Tel: +44 (0)20 8996 7111. Fax: +44 (0)20 8996 7048. Email: [email protected].
Subscribing members of BSI are kept up to date with standards developments
and receive substantial discounts on the purchase price of standards. For details
of these and other benefits contact Membership Administration.
Tel: +44 (0)20 8996 7002. Fax: +44 (0)20 8996 7001.
Email: [email protected].
Information regarding online access to British Standards via British Standards
Online can be found at http://www.bsi-global.com/bsonline.
Further information about BSI is available on the BSI website at
http://www.bsi-global.com.
Copyright
Copyright subsists in all BSI publications. BSI also holds the copyright, in the
UK, of the publications of the international standardization bodies. Except as
permitted under the Copyright, Designs and Patents Act 1988 no extract may be
reproduced, stored in a retrieval system or transmitted in any form or by any
means – electronic, photocopying, recording or otherwise – without prior written
permission from BSI.
This does not preclude the free use, in the course of implementing the standard,
of necessary details such as symbols, and size, type or grade designations. If these
details are to be used for any other purpose than implementation then the prior
BSI written permission of BSI must be obtained.
389 Chiswick High Road Details and advice can be obtained from the Copyright & Licensing Manager.
London Tel: +44 (0)20 8996 7070. Fax: +44 (0)20 8996 7553.
Email: [email protected].
W4 4AL