Manual SP20 Creative Medical
Manual SP20 Creative Medical
Manual SP20 Creative Medical
Instructions to User
Dear Customer,
Thank you for purchasing this quality product. Please read the
manual very carefully before using this device. Failure to follow
these instructions can cause measuring abnormality or damage to
the Oximeter.
No part of this manual may be photocopied, reproduced or
translated into another language without the prior written consent.
We reserve the right to revise and amend it at any time without
prior notice.
Version of the Manual: Ver 1.0
Issued Date: December 20th, 2016
All rights reserved.
Notes:
The contents contained in this manual are subject to change
without prior notice.
Information furnished by Creative is believed to be accurate
and reliable. However, no responsibility is assumed by Creative
for its use, or any infringements of patents or other rights of
third parties that may result from its use.
3502-2290079
I
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Instructions for Safe Operation
Check the device to make sure that there is no visible
damage that may affect user’s safety and measurement
performance. It is recommended that the device should be
inspected minimally before each use. If there is obvious
damage, stop using the device.
Necessary service must be performed only by qualified
technicians. Users are not permitted to service this device.
The oximeter must not be used with the devices and
accessories not specified in User Manual.
Cautions
Explosive hazard—DO NOT use the oximeter in
environment with inflammable gas such as some ignitable
anesthetic agents.
DO NOT use the oximeter while the Patient is under MRI or
CT scanning. This device is NOT MRI Compatible.
Warnings
Discomfort or pain may occur if using the sensor of this
device continuously on the same location for a long time,
especially for the patients with poor microcirculation. It is
recommended that the Oximeter should not be applied to the
same location for longer than 2 hours or less if any
abnormal condition is found. Frequently check and
re-position the Oximeter sensor.
Misapplication of a SpO2 probe with excessive pressure for
prolonged periods can induce pressure injury.
Place the SpO2 probe on the finger tightly will cause venous
pulse and effect blood circulation, and lead to interstitial
edema, hypoxia and inaccurate measurement.
Biocompatibility tests have been performed on all the
applied parts, some exceptional allergic patients may still
II
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
have anaphylaxis. Do not apply to those who have
anaphylaxis.
For the individual patients, there should be a more prudent
inspecting in the placing process. The sensor can not be
placed on the edema and tender tissue.
The local law should be followed when disposing of the
expired device or its accessories.
DO NOT operate in the environment where strong
electro-magnetic interference exists, such as radiogram,
television, radiophone, etc.
Please pay attention to the SpO2 probe cable while using to
avoid strangulating patient.
Attentions
Keep the oximeter away from dust, vibration, corrosive
substances, explosive materials, high temperature and
moisture.
If the Oximeter gets wet, please stop operating it and do not
resume operation until it is dry and checked for correct
operation. When it is carried from a cold environment to a
warm and humid environment, please do not use it
immediately. Allow at least 15 minutes for the Oximeter to
reach ambient temperature.
DO NOT operate the button on the front panel with sharp
materials or sharp point.
DO NOT use high temperature or high pressure steam
disinfection on the oximeter and probes. Refer to related
chapter for instructions regarding cleaning and disinfection.
The intended use of this device is not for therapy purpose.
III
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
equipment is protected against solid foreign objects of
12.5mm and greater, and protected against vertically falling
water drops when enclosure tilted up to 15°.
Declaration of Conformity
The manufacturer hereby declares that this device complies with
the following standards:
IEC 60601-1:2005+A1: 2012, IEC60601-1-2:2014,
IEC60601-1-11:2010, ISO 80601-2-61:2011 and follows the
provisions of the council directive MDD93/42/EEC.
IV
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Table of Contents
1 Overview ..................................................................................... 1
1.1 Appearance .......................................................................... 1
1.2 Product Name and Model..................................................... 3
1.3 Structure ............................................................................... 3
1.4 Features ................................................................................ 4
1.5 Intended Use ........................................................................ 4
1.6 Working Environment .......................................................... 4
2 Preparation ................................................................................ 5
2.1 Charger Base Connection ..................................................... 5
2.2 Power Supply ....................................................................... 5
3 Make Measurement ................................................................... 7
3.1 SpO2 Measurement ............................................................... 7
3.2 Temperature Measurement (optional) .................................. 9
4 Operation.................................................................................. 11
4.1 Power on/off the Oximeter ................................................. 11
4.2 Default Display Screen ...................................................... 11
4.3 Menu .................................................................................. 14
4.4 Record ................................................................................ 27
5 Technical Specifications .......................................................... 32
6 Over-limit Indication............................................................. 34
6.1 Limit settings ..................................................................... 34
6.2 Over-limit indication sound mute setting ........................... 34
7 Packing List .............................................................................. 35
8 Repair and Maintenance ......................................................... 36
8.1 Maintenance ....................................................................... 36
8.2 Cleaning and Disinfecting Instruction ................................ 36
9 Troubleshooting ....................................................................... 37
Appendix ..................................................................................... 38
Quality Inspection Certificate .................................................... 53
V
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
1 Overview
1.1 Appearance
Alert
indicator
Display screen
Record/Back
Menu/Confirm
Rotate-lock /Up
Power indicator
Mute /Right
Power/Left
Setting/Down
1
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
SpO2: TEMP:
SpO2 probe Temperature
connector probe
connector
DC power
input socket
1.3 Structure
It consists of the main unit and SpO2 probe.
(Note: with optional temperature probe, this Oximeter can make
temperature measurement.)
3
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
1.4 Features
It is lightweight, small in size and easy to carry
Color LCD to display plethysmogram and parameters
Measure SpO2, Pulse Rate and Temperature simultaneously
PI (Perfusion Index) display is available
Up to 500 hours data storage for SpO2 and PR and can be
recalled.
16 user IDs for making data and can be added.
A built-on holder for convenient standing on desktop and
display viewing.
Real-time battery status display and low battery voltage
indication.
Auto power off is available
Audible and visual alert function is available
Data uploading to PC for management (Optional)
Power saving mode is available
1.5 Intended Use
This Handheld Pulse Oximeter is intended for measuring and
recording the pulse rate, functional oxygen saturation (SpO 2) and
temperature (optional). It is applicable for detecting SpO2, pulse
rate and temperature of adult and pediatric patients in clinical
institutions and homes.
2 Preparation
Oximeter
Charger base
5
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
with " ", and the other end to the USB power source with
output capacity of 5V DC/1A.
Note: When using the supplied charger base for power supply,
please do not tilt the charger base backwards too much, or the
USB cable and the connector may be damaged.
6
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
3 Make Measurement
3.1 SpO2 Measurement
Operation procedures:
1. Connect the SpO2 probe to the connector on the upper-side of
the device marked with "SpO2". (Note: When disconnecting the
connector, be sure to hold the head of the connector firmly and
pull).
2. The red blinking light inside the clip of the SpO2 probe
indicates a successful connection.
3. Insert one finger (index finger is preferred, the nail should be
not too long) into the clip of the probe according to the finger
mark,as shown in figure 3.1.
4. The device will begin to take the measurement, and the
measured result will be displayed on the screen, as shown in figure
4.2.
7
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
8
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Battery cover
Operation procedure:
1. Connect the infrared temperature probe to the connector on the
upper side of device marked with "TEMP".
2. When the screen shows as the big display screen in figure 3.2
9
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
and the temperature unit "℃" is blinking, the user can begin to
take the measurement.
3. Insert the tip of the temperature probe into the earhole and press
the measuring key to start the measurement. A short beep
means the measurement has finished and the result will be
displayed on the big display screen on temperature probe and
the display screen of the Oximeter.
Note:
If the temperature probe detects a hardware failure, the display
screen on the infrared temperature probe will show "Err" and
will not enter into measurement mode.
The infrared temperature probe will switch to standby state
automatically if there is no operation for 1 minute. If a further
measurement is needed, press the measuring key and repeat
step 2 and step 3.
Normal body temperature varies depending on the position/area
the measurement is taken from. The following table shows the
varying temperature ranges of the different body positions.
Temperature varying range at different body positions:
Arm 34.7 ~ 37.3 ℃
Oral 35.5 ~ 37.5 ℃
Rectal 36.6 ~ 38.0 ℃
Ear 35.8 ~ 38.0 ℃
Safety Instruction for Temperature Measurement
Do NOT take a measurement when the patient is moving.
Patients with tympanitis or otitis problems should NOT use
this device.
When the infrared temperature probe is connected to the
device, the probe will consequently be at power-on status,
10
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
therefore pressing the power on/off key on the temperature
probe will not cause any effect.
4 Operation
4.1 Power on/off the Oximeter
Figure 4.1
4.2 Default Display Screen
Press “ ”power key for 2 seconds to start up the Oximeter, then
the screen will display the default screen, as shown in Figure 4.2.
11
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Pulse rate
mark and unit Pulse symbol
HR high/low limit
setting range Pulse rate value
TEMP value
Perfusion index
Pulse
Plethysmogram
Current time
Description:
During measurement, if the finger is not inserted properly, or
the probe is not connected or the probe is off from the finger,
then "Check Probe" message prompts on the screen, and
"bibibi..." alert sound appears simultaneously. Alert sound is
12
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
" ", then the Rotate-lock icon " " turns to blue " ",
it means the auto rotation function is turned on, if you place
this oximeter horizontally, then the display shows in
horizontal, as shown in figure 4.2B.
" " key. Short time press " " key can turn off (or
resume) the device sound (including pulse beep sound,
audible alert and key click), while the pulse symbol " " still
13
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
4.3 Menu
On the default measuring screen, short time press " "
Menu/Confirm key for entering into main menu screen (as shown
in Figure 4.3).
14
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Note: when the device is connecting the neonate SpO2 probe, then
the User icon " " turns to grey " ", and the patient type on
upper left corner turns to pink " ".
Measurement mode: Select the measurement mode,
"Spot-check mode" and "Continuous mode" for option.
15
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
TEMP Record: Review the temperature record list.
Date: Set the time and date, see Section 4.3.6 for details.
Alert: Set the low alert limit for SpO2 and the high/low alert
limit for HR, see Section 4.3.8 for details.
4.3.1 User ID
On main menu screen, move the cursor on "User ID" and press
Confirm key " ", then the oximeter enters into User ID
Setup screen, as shown in figure 4.4.
16
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Move the cursor on "Edit" and press Confirm key " ", when
the cursor turns to blue, then the user can edit the User ID, and
move the cursor on "OK" to confirm the edit, the edit screen is as
shown in figure 4.4B.
17
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
4.3.3 Measurement Mode
On main menu screen, move the cursor on "Measurement Mode"
and press Confirm key " ", then the oximeter enters into
Measurement Mode Setup screen, as shown in figure 4.6.
19
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Figure 4.10C Language setup Figure 4.10D Auto Power OFF setup
22
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
23
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
key operation for 3 minutes, then the oximeter will power off
automatically.
When the Power Saving Mode is set to "On" option, during
the measurement, if there is no key operation for 1 minute,
the screen display will be dim for power saving. The display
brightness will resume to normal condition by pressing any
key.
4.3.8 Alert
On main menu screen, move the cursor on "Alert" and press
Confirm key " ", then the oximeter enters into alert setting
screen, as shown in figure 4.11.
4.3.9 Help
On main menu screen, move the cursor on "Help" and press
Confirm key " ", then the oximeter help information screen,
which shows SpO2 and temperature measurement tips, as shown in
figure 4.12.
25
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
26
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
4.4 Record
4.4.1 Data Recall
On main default screen, short time press Record/Back key
" " to enter into data recall screen, as shown in figure 4.13.
27
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
The
corresponding
User and User
ID for the
selected record
29
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
The
corresponding
User and User
Recording
ID for the
time
selected record
SpO2
trend graph
PR trend graph
30
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
31
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
5 Technical Specifications
A. Display Panel: 3.5 inch color TFT LCD;
B. Power Supply:
Internal power supply: 2000mAh lithium battery
AC power adapter: 5VDC/1A,
Working current: ≤
180mA
Input power for AC power adapter: <15VA
The typical continuous operation time of the battery: 18 hours
(when screen display is automatically off and bluetooth
function is disabled).
The typical service life of the battery: 5 years.
C. SpO2 Measurement
Transducer: dual-wavelength LED sensor with wavelength:
Red light: 663 nm, Infrared light: 890 nm.
Maximal average optical output power: ≤ 2mW
Display range: 35~100%
Measuring accuracy:
Arms is not greater than 3% for SpO2 range from 70% to 100%
*NOTE: Arms is defined as root-mean-square value of deviation
according to ISO 80601-2-61.
SpO2 low alert limit setting range: 50%~99%
The device is calibrated to display functional oxygen saturation.
The functional tester cannot be used to assess the accuracy of the
SpO2 probe or the device.
D. Pulse Rate Measurement
Display and measuring range: 30bpm~240bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
Over-limit setting range: 25bpm~250bpm
E. Data update period
32
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
The SpO2 value update time is less than 30 seconds. The pulse
rate value update time is less than 10 seconds.
F. Perfusion Index Display
Range: 0.2%~20%
G. Temperature Measurement
Measuring range: 32.0℃~43.0℃
Measuring accuracy: ±0.2℃ for temperature range from
35.0℃ to 42.0℃, and ±0.3℃ for the rest.
Response time: ≤5s
H. Operating Environment
Operating Temperature: 5°C ~40°C
Operating Humidity: 15%~93%
Atmospheric pressure: 70kPa~106kPa
Note: portable and mobile RF communications equipment may
affect the performance of the Oximeter.
I. Low Perfusion Performance
The accuracy of SpO2 and PR measurement still meet the
precision described above when the modulation amplitude is as
low as 0.5%.
J. Resistance to interference of surrounding light:
The difference between the SpO2 value measured in the
condition of indoor natural light and that of darkroom is less
than ±1%.
K. Dimensions: 158 mm (L) × 73 mm (W) × 25 mm (H)
Net Weight: about 230g (including battery)
L. Classification
Type of protection against electric shock:
Internally powered equipment and Class II.
Degree of protection:
Type BF applied parts.
33
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Degree of protection against harmful ingress of liquids:
The equipment is IP22 with protection against harmful solid
foreign objects and ingress of liquid.
Mode of operation: Continuous operation.
Electro-Magnetic Compatibility: Group I, Class B
6 Over-limit Indication
6.1 Limit settings
SpO2 low limit setting range: 50% ~ 99%.
Pulse Rate limits setting range:
High: 100bpm--240bpm Low: 30bpm--99bpm
During the measurement, if the measured value exceeds the preset
value, the alert beeping sound will be activated, the value that is
over-limit will blink at the same time.
6.2 Over-limit indication sound mute setting
During the measurement, if the over-limit indication sound is
set to on, short time press Mute/Right key(" "), then the
over-limit indication sound will mute for 90 seconds, but the
over-limited value still keeps blinking. At this moment, the
speaker volume icon becomes " ". If this alert event
persists over 90 seconds, then the over-limit indication sound
will be activated again.
During the measurement, if the probe is off or disconnected,
the message "Check Probe" shows on the display screen.
The alert sound starts (interval is 5 seconds) and lasts for
about 3 minutes. If the probe is still off, the Oximeter will
power off automatically.
34
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
7 Packing List
1. An Oximeter
2. A SpO2 probe
3. User Manual
4. A oximeter rubber cover
5. A charging base
6. A temperature probe (optional)
7. Charging cable (optional)
8. A USB data cable (optional)
Notes:
1. The accessories are subject to change. See the package in your
hand for detailed items and quantity.
2. All the parts of the device should NOT be replaced at will. If
necessary, please use the components provided by the manufacture
or those that are of the same model and standards as the accessories
along with the device which are provided by the same factory.
Otherwise, negative effects concerning safety and biocompatibility
etc. may be caused.
3. This device can only connect with the manufacture nominated
device.
35
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
37
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
38
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
b) Exposure to excessive illumination, such as surgical lamps,
bilirubin lamps, fluorescent lights, infrared heating lamps, or direct
sunlight;
c) Vascular dyes or external used color-up product such as nail
enamel or color skin care;
d) Excessive patient movement;
e) Placement of a sensor on an extremity with a blood pressure
cuff, arterial catheter, or intravascular line;
f) Exposure to the chamber with High pressure oxygen;
g) There is an arterial occlusion proximal to the sensor;
h) Blood vessel contraction caused by peripheral vessel
hyperkinesias or body temperature decreasing;
i) Low perfusion condition (Perfusion Index is small).
Please contact the local distributor or manufacturer if
necessary.
39
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Appendix
I Key of Symbols
Symbol Description
40
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Symbol Description
Power/Left Key
Mode/Right Key
Rotate-lock/Up Key
Date of manufacture
No alarm
41
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
II Common Knowledge
1 Meaning of SpO2
2 Principle of Measurement
42
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Clinical application of pulse oximeters: SpO2 is an important
physiological parameter to reflect the respiration and
ventilation function, so SpO2 monitoring used in clinical
becomes more popularly, such as monitoring the patient with
serious respiratory disease, the patient under anesthesia
during operation, premature and neonate. The status of SpO2
can be determined in time by measurement and find the
hypoxemia patient earlier, thereby preventing or reducing
accidental death caused by hypoxia effectively.
3 Normal SpO2 Range and Default Low Limit
In campagna area, healthy people’s SpO2 value is greater
than 94%, so the values below 94% are determined as
hypoxia. SpO2<90% is considered as the default threshold for
determining anoxia by most researchers, so SpO2 low limit of
the oximeter is set as 90% generally.
4 Factors affecting SpO2 accuracy (interference reason)
Intravascular dyes such as indocyanine green or
methylene blue
Exposure to excessive illumination, such as surgical
lamps, bilirubin lamps, fluorescent lights, infrared
heating lamps, or direct sunlight.
Vascular dyes or external used color-up product such
as nail enamel or color skin care
Excessive patient movement
Placement of a sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line
Exposure to the chamber with High pressure oxygen
There is an arterial occlusion proximal to the sensor
Blood vessel contraction caused by peripheral vessel
43
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
hyperkinesias or body temperature decreasing
5 Factors causing low SpO2 value (pathology reason)
Hypoxemia disease, functional lack of HbO2
Pigmentation or abnormal oxyhemoglobin level
Abnormal oxyhemoglobin variation
Methemoglobin disease
Sulfhemoglobinemia or arterial occlusion exists near
sensor
Obvious venous pulsations
Peripheral arterial pulsation becomes weak
Peripheral blood supply is not enough
44
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
III EMC
The equipment meets the requirements of IEC 60601-1-2:2014.
Table 1
Guidance and manufacturer’s declaration-electromagnetic
emission
The Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Oximeter should
assure that it is used in such an environment.
Electromagnetic
Emissions test Compliance
environment-guidance
45
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Table 2
Guidance and manufacturer’s declaration-electromagnetic
emission
The Oximeter is intended for use in the electromagnetic environment
specified below. the customer or the user of the Oximeter should assure
that it is used in such an environment.
Electromagnetic
IEC60601 test Complianc
Immunity test environment
level e level
-guidance
Floors should be
wood, concrete or
ceramic tile. if
Electrostatic ±6 kV floors are covered
±6 kV contact
discharge(ESD) contact with synthetic
±8kV air
IEC61000-4-2 ±8kV air material, the
relative humidity
should be at least
30%
46
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
<5% UT
(>95% dip in UT)
for 0.5 cycle
Voltage dips,
<40% UT
short
(60% dip in UT) for
interruptions and
5 cycles N/A
voltage variations N/A
<70% UT
on power supply
(30% dip in UT) for
input lines
25 cycles
IEC61000-4-11
<5% UT
(>95% dip in UT)
for 5 s
Power frequency
magnetic fields
should be at levels
Power frequency(
characteristic of a
50Hz/60Hz)
3A/m 3A/m typical location in
magnetic field
a typical
IEC61000-4-8
commercial or
hospital
environment.
NOTE : UT is the a.c. mains voltage prior to application of the test
level.
Table 3
Guidance and manufacturer’s declaration – electromagnetic
immunity
47
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
48
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
determined by an
electromagnetic site
survey ,a should be less
than the compliance
level in each frequency
range .b
Interference may occur
in the vicinity of
equipment marked with
the following symbol.
49
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
Table 4
50
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
RF Exposure
The equipment complies with FCC radiation exposure limits
set forth for an uncontrolled environment.
52
Downloaded from www.Manualslib.com manuals search engine
User Manual for Handheld Pulse Oximeter
53
Downloaded from www.Manualslib.com manuals search engine