Isometer® Isomed427X - (PT) : Insulation Monitoring Device For Medical Applications
Isometer® Isomed427X - (PT) : Insulation Monitoring Device For Medical Applications
isoMED427_D00440_00_D_XXEN/06.2022
ISOMETER® isoMED427x-(PT) Insulation monitoring device with integrated load and
temperature monitoring and locating current injector
for insulation fault location systems for medical IT systems
Product description
ISOMETER®s of the isoMED427x-(PT) series monitor the insulation resistance of a medical
IT system with AC 70…264 V. In addition, the load current and the temperature of the IT
system transformer are monitored. Alarms and measured values are provided to other
bus devices via the BMS interface. For display and signalling purposes, the use of special
alarm indicator and test combinations is recommended.
After detection of an insulation fault, the internal locating current injector of the
isoMED427P-(xx) models enables insulation fault location. The use of special devices of
the EDS series is recommended to locate the insulation fault.
Devices of the isoMED427x-(PT) series do not require an additional supply voltage.
The maximum permissible system leakage capacitance is 5 µF.
ISOMETER® isoMED427P In order to meet the requirements of applicable standards, customised parameter settings
must be made on the equipment in order to adapt it to local equipment and operating
Device features conditions. Please heed the limits of the range of application indicated in the technical
data.
• Insulation monitoring for medical IT systems
• Adjustable response value for insulation Application
monitoring edical IT system in accordance with IEC 60364-7-710, IEC 61557-8, IEC 61557-9 and
M
• Locating current injector for insulation DIN VDE 0100-710.
fault location systems
Function
• Load and temperature monitoring for
IT system transformers During regular operation, the display shows the present insulation resistance. Use the
arrow-up or arrow-down button to display the present load current in %. The isoMED427P-PT
• Adjustable load current response value
also displays the present transformer temperature. If the insulation resistance falls below
• Temperature monitoring with PTC the response value, the AL1 LED signals an insulation fault. AL2 lights up if the load current
thermistor or bimetal switch is too high or the temperature of the monitored IT system transformer is too high. The
• Self monitoring with automatic alarm alarm relay K1 signals all alarm categories. In addition, a bus signal is provided at terminals
• PE connection monitoring A, B for insulation fault locators as well as alarm indicator and test combinations.
• Internal/external test button The detected insulation fault activates the internal locating current injector for insulation
fault location if the EDS function was previously enabled in the menu (factory setting = off).
• LEDs: Power On, Alarm 1, Alarm 2
A positive and a negative locating current pulse is injected into the monitored IT system
• Configurable alarm relay: N/O or N/C alternately for 2 s each. There is a 4 s pause between positive and negative pulse.
operation selectable
The isoMED427P-(PT) models can only be operated as BMS slave. Therefore, the alarm
• Compact two-module enclosure (36 mm) indicator and test combination or the respective insulation fault locator take over the
• BMS interface master function. BMS masters always have BMS address 1.
Approvals Standards
The ISOMETER® of the isoMED427x-(PT) series complies with the requirements of the de-
vice standards:
• DIN EN 61557-8 (VDE 0413-8): 2015-12/Cor1: 2016-12
• DIN EN 50155: 2018-05
2 isoMED427_D00440_00_D_XXEN/06.2022
ISOMETER® isoMED427x-(PT)
Operating elements
1 2 3
1 - Power on LED
2 - Alarm LED 1 is lit: Below response value Ran
3 - Alarm LED 2 is lit: Response value % I or/and °C exceeded
4 - LC-Display
5 - "T": Start a self test (2 s)
: Menu item up / increase value
6 - : Menu item down / decrease value
4 7 - Start menu mode (2 s)
Enter button:
(< 1.5 s): ConArm menu item, submenu item, value.
(> 2 s): Return to the next higher menu level
5 6 7
Wiring diagram
STW2
L E KE L1 L2 7
N Z Z/k l
k
ON AL1 AL2
T1 EDS ... MK ...
5
T2
A B A B
A
R 6
B
6A 6A T R MENU
R J-Y(St)Y 2x2x0,6
on
o
11 12 14
T1 T2 A B
3 4 1 2
isoMED427_D00440_00_D_XXEN/06.2022 3
ISOMETER® isoMED427x-(PT)
4 isoMED427_D00440_00_D_XXEN/06.2022
ISOMETER® isoMED427x-(PT)
Connection Other
Connection type Push-wire terminals Operating mode Continuous operation
Nominal current ≤ 10 A Position of normal use Any
Stripping length 10 mm Degree of protection, built-in components (DIN EN 60529) IP30
Opening force 50 N Degree of protection, built-in components (DIN EN 60529) IP20
Test opening, diameter 2.1 mm Enclosure material Polycarbonate
Connection properties: Flammability class UL94V-0
rigid 0.2…2.5 mm2 (AWG 24…14) DIN rail mounting IEC 60715
flexible without ferrule 0.75…2.5 mm2 (AWG 19…14) Screw mounting 2 x M4
flexible with ferrule 0.2…1.5 mm2 (AWG 24…16) Software versions D643 V1.0x (isoMED427-2)
D355 V1.0x (isoMED427P-2)
Connection type Screw-type terminals
D644 V 1.0x (isoMED427P-PT)
Nominal current ≤ 10 A
Weight ≤ 150 g
Tightening torque 0.5…0.6 Nm (5…7 lb-in)
Cross section AWG 24…12 Factory settings isoMED427x-(PT)
Stripping length 8 mm Response value Ran 50 kΩ (< R)
Connection properties: Response value Ialarm 7 A (> I)
rigid / flexible 0.25…2.5 mm2 Response value °C 4 kΩ (fixed value for isoMED427x)
Flexible with ferrules with/without plastic sleeve 0.25…2.5 mm2 120 °C (configurable for isoMED427P-PT)
Multi-conductor rigid/flexible 0.2…1.5 mm2 Operating principle K1 N/C operation (n.c.)
Multi-conductor flexible with ferrule without plastic sleeve 0.25…1.5 mm2 BMS address 3
Multi-conductor flexible with TWIN ferrule with plastic sleeve 0.25…1.5 mm2 Automatic insulation fault location off, deactivated
Password 0, disabled
CT monitoring on, activated
Termination off, deactivated (120 Ω)
Application example
isoMED427P
BMS
AN450
STW2 IT system
ES710 series
EDS151
MK2430 MK2430
PE = Protective earth
PA = Equipotential bonding
ZPA = Supplementary equipotential bonding isoMED427_D00440_00_D_XXEN/06.2022 5
Ordering information Dimension diagram XM420
Dimensions are given in mm
Supply voltage Us Art. No.
Type
36
AC Screw-type terminal Push-wire terminal
isoMED427-2 B92075306 B72075306
70…264 V, 47…63 Hz isoMED427P-2 1) B92075301 B72075301
isoMED427P-PT B92075307 B72075307
1) Only this device has a Lloyds Register approval
Accessories 93 90 45 70,5
isoMED427_D00440_00_D_XXEN/06.2022 / © Bender GmbH & Co. KG, Germany – Subject to change! The specified standards take into account the edition valid until 06.2022 unless otherwise indicated.
Measuring current transformers STW2 B942709
Temperature sensor (PTC) ES0107 B924186
Mounting frame XM420 B990994
BENDER Group