Iec 80601-2-59-2008
Iec 80601-2-59-2008
Iec 80601-2-59-2008
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Medical electrical equipment –
Part 2-59: Particular requirements for the basic safety and essential performance
of screening thermographs for human febrile temperature screening
Appareils électromédicaux –
Partie 2-59: Exigences particulières pour la sécurité de base et les performances
essentielles des imageurs thermiques pour le dépistage des humains fébriles
IEC 80601-2-59:2008
THIS PUBLICATION IS COPYRIGHT PROTECTED
Copyright © 2008 IEC, Geneva, Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by
any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or
IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Switzerland
Email: [email protected]
Web: www.iec.ch
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue des publications de la CEI: www.iec.ch/searchpub/cur_fut-f.htm
Le Catalogue en-ligne de la CEI vous permet d’effectuer des recherches en utilisant différents critères (numéro de référence,
texte, comité d’études,…). Il donne aussi des informations sur les projets et les publications retirées ou remplacées.
Just Published CEI: www.iec.ch/online_news/justpub
Restez informé sur les nouvelles publications de la CEI. Just Published détaille deux fois par mois les nouvelles
publications parues. Disponible en-ligne et aussi par email.
Electropedia: www.electropedia.org
Le premier dictionnaire en ligne au monde de termes électroniques et électriques. Il contient plus de 20 000 termes et
définitions en anglais et en français, ainsi que les termes équivalents dans les langues additionnelles. Egalement appelé
Vocabulaire Electrotechnique International en ligne.
Service Clients: www.iec.ch/webstore/custserv/custserv_entry-f.htm
Si vous désirez nous donner des commentaires sur cette publication ou si vous avez des questions, visitez le FAQ du
Service clients ou contactez-nous:
Email: [email protected]
Tél.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 80601-2-59
Edition 1.0 2008-10
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Medical electrical equipment –
Part 2-59: Particular requirements for the basic safety and essential performance
of screening thermographs for human febrile temperature screening
Appareils électromédicaux –
Partie 2-59: Exigences particulières pour la sécurité de base et les performances
essentielles des imageurs thermiques pour le dépistage des humains fébriles
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE PRICE CODE
CODE PRIX V
ICS 11.040.55 ISBN 2-8318-1004-5
–2– 80601-2-59 © IEC:2008
CONTENTS
FOREWORD...........................................................................................................................3
INTRODUCTION.....................................................................................................................5
201.1 Scope, object and related standards .....................................................................6
201.2 Normative references ...........................................................................................7
201.3 Terms and definitions ...........................................................................................8
201.4 General requirements ......................................................................................... 10
201.5 General requirements for testing of ME EQUIPMENT .............................................. 10
201.6 Classification of ME EQUIPMENT and ME SYSTEMS ................................................. 11
201.7 M E EQUIPMENT identification, marking and documents ......................................... 11
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.8 Protection against electrical HAZARDS from ME EQUIPMENT ................................... 11
201.9 Protection against mechanical HAZARDS of ME EQUIPMENT and ME SYSTEMS ......... 12
201.10 Protection against unwanted and excessive radiation HAZARDS ........................... 12
201.11 Protection against excessive temperatures and other HAZARDS ........................... 12
201.12 Accuracy of controls and instruments and protection against hazardous
outputs ............................................................................................................... 12
201.13 H AZARDOUS SITUATIONS and fault conditions ........................................................ 13
201.14 P ROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS ( PEMS ) .......................................... 13
201.15 Construction of ME EQUIPMENT ............................................................................. 13
201.16 M E SYSTEMS ........................................................................................................ 13
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS ......................... 13
201.101 Laboratory accuracy of a screening thermograph................................................ 14
201.102 Screening thermograph alarm conditions ............................................................ 17
Annexes ............................................................................................................................... 18
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS .................................................................................................................... 18
Annex AA (informative) Particular guidance and rationale .................................................... 20
Annex BB (normative) C ALIBRATION SOURCE ......................................................................... 26
Annex CC (informative) Reference to the essential principles .............................................. 27
Bibliography.......................................................................................................................... 29
Index of defined terms used in this particular standard.......................................................... 34
Table AA.1 – Example of relevant uncertainty terms for a SCREENING THERMOGRAPH ............. 22
Table CC.1 – Correspondence between this particular standard and the essential
principles .............................................................................................................................. 27
80601-2-59 © IEC:2008 –3–
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 80601-2-59 has been prepared by a Joint Working Group of IEC
subcommittee 62D: Electromedical equipment, of IEC technical committee 62: Electrical
equipment in medical practice, and of ISO subcommittee SC3: Lung ventilators and related
equipment, of ISO technical committee 121: Anaesthetic and respiratory equipment.
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table. In ISO, the standard has been approved by
12 P-members out of 15 having cast a vote.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
–4– 80601-2-59 © IEC:2008
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
80601-2-59 © IEC:2008 –5–
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of ME EQUIPMENT for human febrile
temperature screening.
This document describes ME EQUIPMENT that uses infrared technology to detect naturally
emitted heat at the skin surface of the FACE . Such ME EQUIPMENT can be useful at ports-of-
entry or ports-of-exit and the entrances to buildings under indoor environmental conditions to
separate febrile from afebrile individuals to help prevent the spread of communicable
diseases. Care can be needed when evaluating individuals under changing environmental
conditions, but the inner canthus of the eye has been demonstrated to be a robust
measurement site and is supplied by the internal carotid artery. [40] 1)
A body core temperature of 38 °C or above was used as the criterion to restrict traveling
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
during the SARS (severe acute respiratory syndrome) epidemic (April 2003). [73] The US
Center for Disease Control advises that SARS typically begins with a temperature above
38 °C, which is 1 °C higher than normal human body core temperature which averages around
37 °C. [29] It is hard to give an accurate assessment of how many people were checked by
infrared temperature measurements in China during the SARS epidemic. There is official
Chinese government data indicating that during a two-month period in the spring of 2003, 30
million travelers were screened in China. From this cohort, 9 292 travelers with elevated
temperature were detected and 38 were suspected of being SARS carriers. SARS was
diagnosed in 21 of these cases. All elevated temperatures were confirmed using traditional
clinical temperature measurements of body temperature. Although it is hard to determine the
human body's core temperature accurately by infrared measurement of SKIN TEMPERATURE , it
is a potential method for screening for elevated temperature values. [36] [73] [75]
This particular standard amends and supplements IEC 60601-1 (third edition, 2005): Medical
electrical equipment – Part 1: General requirements for safety and essential performance,
hereinafter referred to as the General Standard (see 1.4).
A "General guidance and rationale" section giving some explanatory notes, where
appropriate, about the more important requirements is included in annex AA. It is considered
that knowledge of the reasons for these requirements will not only facilitate the proper
application of the standard but will, in due course, expedite any revision necessitated by
changes in clinical practice or as a result of developments in technology. However, this annex
does not form part of the requirements of this standard.
—————————
1) Figures in square brackets refer to the Bibliography.
–6– 80601-2-59 © IEC:2008
201.1.1 * Scope
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
SCREENING THERMOGRAPHS intended to be used for the individual non-invasive febrile
temperature screening of humans under indoor environmental conditions, hereafter referred to
as ME EQUIPMENT . This International Standard sets laboratory characterization test limits for
the SCREENING THERMOGRAPH .
NOTE A SCREENING THERMOGRAPH is intended for screening and detection of human subjects with SKIN
TEMPERATURES elevated above normal. An elevated SKIN TEMPERATURE needs to be followed up by a subsequent
temperature measurement using a clinical thermometer (see IEC 80601-2-56).
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for SCREENING THERMOGRAPHS as defined in 201.3.209.
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 2 of this particular standard.
IEC 60601-1-3 does not apply. All other published collateral standards in the IEC 60601-1
series apply as published.
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
—————————
2) The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance.
80601-2-59 © IEC:2008 –7–
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of sections, clauses and subclauses of this particular standard corresponds to
that of the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the
content of Clause 1 of the general standard) or applicable collateral standard with the prefix
“20x” where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in
this particular standard addresses the content of Clause 4 of the 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3
collateral standard, etc.). The changes to the text of the general standard are specified by the
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses or figures which are additional to those of the general standard are numbered
starting from 201.101. However due to the fact that definitions in the general standard are
numbered 3.1 through 3.139, additional definitions in this standard are numbered beginning
from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items aa), bb),
etc.
Subclauses or figures which are additional to those of a collateral standard are numbered
starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
NOTE Informative references are listed in the bibliography beginning on page 29.
Addition:
–8– 80601-2-59 © IEC:2008
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1-8:2006 apply, except as follows:
Addition:
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.3.201
CALIBRATION
set of operations that establish, under specified conditions, the relationship between values of
quantities indicated by a measuring instrument, or measuring system, or values represented
by a material measure or a reference material and the corresponding values realized by
standards
201.3.202
CALIBRATION SOURCE
infrared radiation blackbody reference source of known and traceable temperature and
EMISSIVITY
201.3.203
DETECTOR
infrared thermal sensor or array of sensors able to detect infrared thermal energy radiating
from the surface of the FACE or other object
NOTE The DETECTOR responds to the net infrared radiation and converts that response into electrical signals.
201.3.204
EMISSIVITY
ratio of the emitted thermal rate of propagation of electromagnetic energy emitted by an
object as a consequence of its temperature propagated in a given direction, per unit solid
angle about that direction and per unit area projected normal to the direction of a surface to
that of a ideal blackbody at the same temperature and under the same spectral conditions
201.3.205
EXTERNAL TEMPERATURE REFERENCE SOURCE
part of the SCREENING THERMOGRAPH that is used to ensure accurate operation between
CALIBRATIONS using an infrared radiation source of known temperature and EMISSIVITY
NOTE The EXTERNAL TEMPERATURE REFERENCE SOURCE is normally imaged in each thermogram or prior to each
thermogram.
—————————
3) To be published.
4) ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA, 19428-2959 USA
80601-2-59 © IEC:2008 –9–
201.3.206
FACE
anterior cranial face of the PATIENT being measured
201.3.207
IMAGE PIXEL
individual infrared thermal detections from the DETECTOR
NOTE The number of IMAGE PIXELS is given in an array format. e.g. number of sensors in horizontal (H) by number
of sensors in vertical (V) e.g. for a 120 (H) x 120 (V) DETECTOR array, the number of IMAGE PIXELS would be 14,400.
201.3.208
LABORATORY ACCURACY
closeness of the agreement between the result of a measurement (with SCREENING
THERMOGRAPH ) and the true value of the measurand
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
NOTE L ABORATORY ACCURACY is a qualitative concept. For a quantitative description, the term 'uncertainty’ should
be used.
201.3.209
SCREENING THERMOGRAPH
ME EQUIPMENT or ME SYSTEM that:
– detects infrared radiation emitted from the FACE from which a thermogram is obtained from
the TARGET ;
– detects infrared radiation emitted from an EXTERNAL TEMPERATURE REFERENCE SOURCE ;
– displays a radiometric thermal image;
– obtains a temperature reading from the TARGET ; and
– compares that temperature reading to the THRESHOLD TEMPERATURE to determine if the
PATIENTis febrile
NOTE 1 A SCREENING THERMOGRAPH is a non-contact, non-invasive temperature screening ME EQUIPMENT used to
measure the FACE temperature and indicate the screened region with a different colour if the temperature is above
the THRESHOLD TEMPERATURE setting.
NOTE 2 A SCREENING THERMOGRAPH has to identify the TARGET from the thermogram to obtain the TARGET
temperature reading.
201.3.210
SELF - CORRECTION
automatic process carried out to compensate for DETECTOR drift
NOTE An uncooled microbolometer DETECTOR can be subject to significant drift in its measurements over time.
EXAMPLE To reduce the error caused by possible drift, the SCREENING THERMOGRAPH carries out a SELF -
CORRECTION .
201.3.211
SKIN TEMPERATURE
skin surface temperature as measured from the WORKABLE TARGET PLANE of a SCREENING
THERMOGRAPH , with an appropriate adjustment for skin EMISSIVITY
201.3.212
TARGET
region of the FACE selected for THRESHOLD TEMPERATURE comparison
201.3.213
TARGET PLANE
in-focus plane perpendicular to the line of sight of a SCREENING THERMOGRAPH
– 10 – 80601-2-59 © IEC:2008
201.3.214
THRESHOLD TEMPERATURE
temperature setting, above which the SCREENING THERMOGRAPH indicates that the TARGET is
potentially febrile
201.3.215
WORKABLE TARGET PLANE
region of TARGET PLANE that meets the specified performance requirements
NOTE The W ORKABLE TARGET PLANE can be the whole or part of the TARGET PLANE .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Clause 4 of the general standard applies, except as follows:
Addition:
When applying this document to a SCREENING THERMOGRAPH , definitions and requirements that
use the term PATIENT shall be considered as applying to the person being screened for a
febrile state.
Addition:
Requirement Subclause
NOTE Failure to indicate that the SCREENING THERMOGRAPH is not capable of performing to
specification is considered a loss of ESSENTIAL PERFORMANCE .
Additional subclauses:
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
a) instruct the OPERATOR to ensure that the FACE is unobstructed by hair, eyeglasses, and
other objects because their presence will interfere with the ability of a SCREENING
THERMOGRAPH to detect a febrile condition.
b) disclose the recommended position of the FACE required to obtain the TARGET relative to
the optical pathway of the SCREENING THERMOGRAPH ;
EXAMPLE Position of the FACE relative to the lens of the infrared camera.
c) recommend secondary screening with a clinical thermometer of any individual that the
SCREENING THERMOGRAPH indicates is possibly febrile .
NOTE 3 The area chosen for screening should have a non-reflective background and minimal reflected infrared
radiation from the surroundings.
d) an explanation of the effects due to airflow, and instructions that this should be minimized.
NOTE 4 Drafts from air conditioning ducts can cause forced cooling or heating of the FACE and should be baffled
or diffused to prevent airflow from blowing directly onto the PATIENT .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.12 Accuracy of controls and instruments and protection against
hazardous outputs
201.12.2 Usability
Additional subclauses:
201.12.2.101 Display
The SCREENING THERMOGRAPH shall be provided with an isothermal colour display for visual
interpretation and rapid identification of the thermogram of the WORKABLE TARGET PLANE . The
indication of temperature range, with colour code/temperature scale, shall be displayed. The
SCREENING THERMOGRAPH shall be provided with at least one colour mapping mode where the
colours follow the order of the visible spectrum (e.g., rainbow scale) such that blue is cooler
and red is hotter.
The SCREENING THERMOGRAPH should be capable of operating in near real-time for rapid and
effective screening of a mass population. The MANUFACTURER shall estimate the throughput
(time between measurements) in NORMAL USE of the SCREENING THERMOGRAPH . The average
time required by a SCREENING THERMOGRAPH to measure, process and display the temperature
of the FACE and throughput of the SCREENING THERMOGRAPH shall be disclosed in the technical
description.
Compliance is checked by functional testing and inspection of the instructions for use.
80601-2-59 © IEC:2008 – 13 –
The minimum display of the WORKABLE TARGET PLANE shall be 240 IMAGE PIXELS by 180 IMAGE
PIXELS . In NORMAL USE , the thermogram of the FACE shall fill at least 180 IMAGE PIXELS by 135
IMAGE PIXELS . The WORKABLE TARGET PLANE should be perpendicular to the FACE to improve
performance. [42] If the SCREENING THERMOGRAPH requires that the OPERATOR frame the
thermogram of the FACE in the WORKABLE TARGET PLANE , a guide or mask shall be provided in
the image of the WORKABLE TARGET PLANE on the display. [42]
NOTE 180 IMAGE PIXELS by 135 IMAGE PIXELS is 75 % of the minimum display of the W ORKABLE TARGET PLANE .
The size or coordinates of the WORKABLE TARGET PLANE , if it is smaller than the SCREENING
THERMOGRAPH ' S TARGET PLANE ,shall be disclosed in the instructions for use.
Compliance is checked by functional testing and inspection of the instructions for use.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.13 H AZARDOUS SITUATIONS and fault conditions
Addition:
The SCREENING THERMOGRAPH shall be equipped with a NETWORK / DATA COUPLING that can
communicate the following:
a) radiometric thermal image of the FACE including an indication of temperature range, with
colour code/temperature scale (see 201.12.2.101);
b) date and time that the image was acquired;
c) THRESHOLD TEMPERATURE ;
d) results of the comparison of the TARGET to this THRESHOLD TEMPERATURE .
The SCREENING THERMOGRAPH should be capable of communicating a visible light image of the
FACE .
201.16 M E SYSTEMS
New clauses:
201.101.1 * T ARGET
The measuring site of the TARGET shall include the inner canthus of each eye.
The SCREENING THERMOGRAPH shall be capable of creating a thermogram of the FACE over at
least the range of 30 °C to 40 °C.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Compliance is checked by functional testing.
where:
The total uncertainty of the LABORATORY ACCURACY (i.e. the measured temperature difference)
shall be calculated using Equation 2.
u 2 = u CS 2 + u ST 2 (2)
where:
The measured radiometric temperature LABORATORY ACCURACY shall meet the criterion of
Equation 3 to fulfill the requirements of this subclause.
e) Ensure the temperature measurement of the CALIBRATION SOURCE uses the algorithm used
to determine the THRESHOLD TEMPERATURE from the IMAGE PIXELS in the field of a TARGET
(see 201.102.3.102).
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
f) Calculate the mean of the five measurements.
g) Repeat d) and e) with CALIBRATION SOURCE temperature approximately equal to the top of
the operating range and three different points within the range.
h) Evaluate the SCREENING THERMOGRAPH using Formula (3).
The THRESHOLD TEMPERATURE shall be adjustable over at least the range 34 °C to 39 °C with
increments not greater than 0,1 °C.
EXAMPLE The difference between measured SKIN TEMPERATURE and oral temperature derived from a population of
representative study samples can guide the selection of the THRESHOLD TEMPERATURE .
201.101.3.1 P ERFORMANCE
The EXTERNAL TEMPERATURE REFERENCE SOURCE shall have a known radiance temperature
over the range of 33 °C to 40 °C with an expanded uncertainty of ± 0,3 °C and a combined
stability and drift of ± 0,1 °C over the interval for measurement drift specified in 201.101.4.
201.101.3.2 * S IZE
The EXTERNAL TEMPERATURE REFERENCE SOURCE shall be capable of being imaged by the
thermal imager of the SCREENING THERMOGRAPH . The size of EXTERNAL TEMPERATURE
REFERENCE SOURCE shall be greater than or equal to 20 IMAGE PIXELS by 20 IMAGE PIXELS . The
EXTERNAL TEMPERATURE REFERENCE SOURCE should be imaged by the same IMAGE PIXELS as
the TARGET to minimize the effect of non-uniformity across the WORKABLE TARGET PLANE .
The permissible drift and stability of a SCREENING THERMOGRAPH shall be less than or equal to
0,2 °C over an interval of 14 days or the CALIBRATION interval specified in the technical
description, whichever is longer. A SCREENING THERMOGRAPH may use SELF - CORRECTIONS to
maintain the drift within acceptable limits. An automated service test mode may be utilized to
perform these measurements to allow for the substitution of the CALIBRATION SOURCE for a
PATIENT .
– 16 – 80601-2-59 © IEC:2008
NOTE A low drift and stability ensures that the LABORATORY ACCURACY of the SCREENING THERMOGRAPH is
maintained. Although the drift can be reduced if a SELF - CORRECTION is performed more regularly, SELF - CORRECTION
performed too often creates interruptions in the ability to screen PATIENTS with the SCREENING THERMOGRAPH .
e) Calculate the mean and standard deviation of the mean of the day's measurements.
f) Evaluate stability by ensuring that 3 times the standard deviation is less than 0,1 °C.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
g) Repeat b) through f) consecutively for either 14 days or the CALIBRATION interval,
whichever is longer.
h) Determine the smallest and largest mean value as values calculated in e).
i) Evaluate drift by ensuring that the difference of the values in h) is less than ± 0,1 °C.
NOTE The minimum resolvable temperature difference reflects the smallest temperature (or equivalent
temperature) difference that the SCREENING THERMOGRAPH can consistently detect within the W ORKABLE TARGET
PLANE .
NOTE Good uniformity within the TARGET PLANE allows flexibility of deployment of the SCREENING THERMOGRAPH
for screening the FACE .
The WORKABLE TARGET PLANE of the SCREENING THERMOGRAPH shall accommodate a FACE that
is positioned 0,75 m to 2,2 m above the floor. This requirement may be met by moving the
infrared camera.
The SCREENING THERMOGRAPH shall inhibit operation during the period of stabilization following
start-up until it is capable of performing to drift specification 201.101.4. The maximum
duration of this start-up period shall be disclosed in the instructions for use. See also
subclause 201.102.1.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Compliance is checked by inspection and the tests of 201.101.4.
The horizontal and vertical spatial resolution of an IMAGE PIXEL when imaging the TARGET in
NORMAL USE shall be ≤ 1 mm and the value shall be disclosed in the technical description.
An EMISSIVITY of 0,98 shall be used in the calculation of SKIN TEMPERATURE by the SCREENING
THERMOGRAPH . [25] [61]
A SCREENING THERMOGRAPH shall be provided with an ALARM SYSTEM that includes a TECHNICAL
ALARM CONDITION of at least LOW PRIORITY during the period of stabilization following start-up
until the SCREENING THERMOGRAPH is capable of performing to specification. See also
subclauses 201.101.4 and 201.101.8.
Annexes
Annex C
(informative)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.C.5 ACCOMPANYING DOCUMENTS , instructions for use
Addition:
Addition:
Reference to IEC ISO/TR 80600, Medical electrical equipment — Deployment, implementation 201.7.9.3.101 a)
and operational guidelines for identifying febrile humans using a screening thermograph
Relative humidity recommendation and effects 201.7.9.3.101 b)
Response time 201.12.2.102
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Spatial resolution of an IMAGE PIXEL 201.101.9
Terms included in calculation of uncertainty 201.101.2.2
THRESHOLD TEMPERATURE algorithm 201.102.2
Throughput 201.12.2.102
– 20 – 80601-2-59 © IEC:2008
Annex AA
(informative)
A core body temperature of 38 °C or above was used as the criterion to restrict traveling
during the SARS (severe acute respiratory syndrome) epidemic (April 2003). [73] The
association between disease and temperature has been recognized since the earliest days of
recorded medicine. Fever, one of the earliest observations, was for centuries a subjective
assessment on the part of the physician. [42] The purpose of a SCREENING THERMOGRAPH is to
quickly and accurately determine the surface SKIN TEMPERATURE of a human being by
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
measuring the TARGET . The temperature of the TARGET is used to indirectly estimate the body
core temperature.
S CREENING THERMOGRAPHS use infrared technology to detect naturally emitted heat at the skin
surface of the FACE . S CREENING THERMOGRAPHS have been used at ports-of-entry, ports-of exit
and the entrances to buildings under indoor environmental conditions with the intention of
separating febrile from afebrile individuals to help prevent the spread of communicable
diseases. A SCREENING THERMOGRAPH screens a FACE and allows the OPERATOR to quantify
and visualize the SKIN TEMPERATURE profile. Since the SKIN TEMPERATURE greatly depends on
both the skin blood perfusion and environmental conditions, SCREENING THERMOGRAPHS need
to be used in a stable environment with minimum environmental interference in order to
satisfy the ESSENTIAL PERFORMANCE requirements.
The distribution of body core temperatures among humans is well known from many studies.
The most significant problem associated with the use of SCREENING THERMOGRAPHS to detect
fever is that the relationship between an elevated SKIN TEMPERATURE and febrile body core
temperature is not well established. The relationship can be further complicated by
perspiration, facial skin flushed from exertion, etc. As a result of this the use of SCREENING
THERMOGRAPHS for the detection of fever from FACES is unsubstantiated. Until the use of
SCREENING THERMOGRAPHS for the detection of fever from FACES is substantiated, all people
suspected of being febrile with SCREENING THERMOGRAPHS should be confirmed with a
secondary screening with a clinical thermometer.
The use of infrared technology in the design of a SCREENING THERMOGRAPH to determine the
skin surface temperature of the TARGET requires correction of two radiation thermometry
sources of error:
a) those associated with the measurement device, such as CALIBRATION uncertainty, internal
thermal stability and size-of-source effects; and
b) those associated with the measurement context, such as EMISSIVITY and transmissivity of
the surface, in-band background radiation and sight path absorption/emission and other
details of the TARGET .
This document does not specify the infrared wavelength that should be utilized. Both the Mid-
IR waveband (3 μm to 5 μm) and Long-IR waveband (8 μm to 14 μm) have been used in
thermal screening of human beings. [25]
The following are rationales for specific clauses and subclauses in this particular standard,
with clause and subclause numbers parallel to those in the body of the document. The
numbering is, therefore, not consecutive.
80601-2-59 © IEC:2008 – 21 –
The consequences of the transmission of infectious diseases can be catastrophic. In the past
certain epidemics and pandemics have been controlled by classic public health measures
using screening and quarantine. The entry to a quarantine process requires several levels of
investigation. The process of determining whether or not an individual is possibly infected is a
medical diagnostic process. [50] [57] As a result, equipment used for that purpose is a
medical device.
Subclause 201.5.3 a)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Subclause 201.12.2.101.1 – Display colour scale
Displaying a 'colourized' thermogram permits the OPERATOR to visually interpret and rapidly
identify FACE temperatures above the THRESHOLD TEMPERATURE thereby allowing identification
of potentially febrile individuals.
Two 'colourizing' methods or scales are commonly used. They are frequently referred to as
the 'Iron temperature' scale and 'rainbow' or 'spectrum' scale.
The colour scales have the advantage of a logical progression of colours. The eye can see
and process more colours than it can shades of grey. The colour temperature scales were
originally created for just that purpose - a logical colour sequence and more perceivable steps
- when grey scale monitors had limited useful dynamic range. [48] The iron temperature scale
was devised for industrial purposes (e.g. looking at molten metals) where the temperature
range of interest is very broad. For the relatively narrow range of human SKIN TEMPERATURES ,
the rainbow scale provides better visual discrimination. For this reason, the rainbow scale was
chosen as the requirement.
This requirement is intended to ensure that there are sufficient IMAGE PIXELS in the
thermogram of the FACE and TARGET to permit an accurate assessment.
R ESPONSIBLE ORGANIZATIONS and public health authorities can use this information to manage
their responses to public health emergencies.
The current evidence indicates that inner canthi are the preferred site for fever screening due
to the stability of that measurement site. [40] [50] [51]
– 22 – 80601-2-59 © IEC:2008
This requirement is intended to ensure that a complete thermogram of the FACE will exist. An
incomplete thermogram of the FACE caused by out-of-range values can affect the ability of the
SCREENING THERMOGRAPH to determine the TARGET .
The committee believed that this is the minimum radiometric LABORATORY ACCURACY that will
permit the separation of febrile and afebrile subjects.
Table AA.1 illustrates some of the elements of uncertainty of the measurement of temperature
using the CALIBRATION SOURCE with the SCREENING THERMOGRAPH . Table AA.1 includes
examples of how to treat them in the estimation of overall uncertainty. Table AA.1 is for
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
illustrative purposes only, not all uncertainty components are necessarily included.
Maximum
permissible
measurement non-
B 0,2 2√3 0,06
uniformity of the
SCREENING
THERMOGRAPH , uU
Maximum
permissible
instability of the B 0,2 2√3 0,06
SCREENING
THERMOGRAPH , uS
MRTD of the
SCREENING
B 0,1 2√3 0,03
THERMOGRAPH ,
u MRTD
Uncertainty of the
CALIBRATION SOURCE , A 0,2 2 0,10
u CS
Combined
standard k=1 0,21
uncertainty
Expanded
k=2 0,42
uncertainty
NOTE The drift of the SCREENING THERMOGRAPH includes that of the EXTERNAL TEMPERATURE
REFERENCE SOURCE .
80601-2-59 © IEC:2008 – 23 –
Although the relationship between body core temperature and facial SKIN TEMPERATURES has
not been fully established, data indicate that the expected temperature range is covered by
this requirement. [40] [50] [51]
These requirements are designed to ensure that the EXTERNAL TEMPERATURE REFERENCE
SOURCE can be used to determine whether or not the LABORATORY ACCURACY is sufficient to
permit reliable detection of the TARGET temperature. To meet this objective, the actual
temperature of the EXTERNAL TEMPERATURE REFERENCE SOURCE needs to be well known. The
size of the EXTERNAL TEMPERATURE REFERENCE SOURCE relative to its distance from the
DETECTOR of the SCREENING THERMOGRAPH needs to be such that there are a sufficient number
of IMAGE PIXELS to permit an accurate measurement.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Subclause 201.101.3.2 – Size
This requirement is intended to ensure that the EXTERNAL TEMPERATURE REFERENCE SOURCE
can be used to determine whether or not the LABORATORY ACCURACY is sufficient to permit
reliable detection of the TARGET temperature and detection of a TARGET temperature that
exceeds the THRESHOLD TEMPERATURE . In operational use, RESPONSIBLE ORGANIZATIONS will
want the time interval between CALIBRATIONS to be as long as practible since the SCREENING
THERMOGRAPH has to be taken out of service for CALIBRATION .
This requirement is intended to ensure that the EXTERNAL TEMPERATURE REFERENCE SOURCE
can be used to determine whether or not the LABORATORY ACCURACY is sufficient to permit
reliable determination of the TARGET temperature and detection of a TARGET temperature that
exceeds the THRESHOLD TEMPERATURE .
For a SCREENING THERMOGRAPH to be effective, a small MRTD is needed. This reflects the
sensitivity of the SCREENING THERMOGRAPH when distinguishing between two close
temperature values. A small MRTD permits fine adjustment of the THRESHOLD TEMPERATURE
setting to optimize the effectiveness of the screening operation.
This requirement is intended to ensure that the EXTERNAL TEMPERATURE REFERENCE SOURCE
can be used to determine whether or not the LABORATORY ACCURACY is sufficient to permit
reliable detection of the TARGET temperature and detection of a TARGET temperature that
exceeds the THRESHOLD TEMPERATURE . The SCREENING THERMOGRAPH should have adequate
uniformity in terms of the response of the DETECTOR across the whole field of view, e.g. all
sensors in the DETECTOR should respond to the same electromagnetic energy emitted by an
object as a consequence of its temperature in a similar way.
S CREENING THERMOGRAPHS are used to screen the general population. The committee chose a
minimum range of TARGET positions that permits the most ambulatory individuals to be
– 24 – 80601-2-59 © IEC:2008
screened, including very tall individuals as well as ambulatory small children and individuals
in wheelchairs.
Some DETECTORS and the EXTERNAL TEMPERATURE REFERENCE SOURCE exhibit significant drift
during an initial period of stabilization. The LABORATORY ACCURACY of measurements during
this period is suspect. Figure AA.1 [42] shows the actual drift of some DETECTORS following
start-up. The change in measured temperature of four different DETECTORS is shown. The
square symbols are from a single element cooled DETECTOR . The other symbols are from
three different focal-plane array DETECTORS over time. The triangular symbols show the
temperature readings from a cooled DETECTOR that does not indicate temperature until
stability is achieved. None of these DETECTORS fulfill the LABORATORY ACCURACY requirements
of this standard without correction. This requirement is intended to ensure that the entire
SCREENING THERMOGRAPH as a whole is ready to operate.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
4
1
T (°C)
0
0 20 40 60 80 100 120
–1
–2
–3
t (min)
IEC 1760/08
This spatial resolution was chosen to ensure that there is sufficient resolution to measure the
temperature of the TARGET . Heated resolution charts can be used to perform this test. [63]
Published sources have reported human skin EMISSIVITY is in the range from 0,96 to 0,98. The
measurement of EMISSIVITY is difficult and the uncertainty of the measurement could typically
be 2 % to 5 % (k = 2). Varying the EMISSIVITY setting in the calculation of temperature changes
the reported temperature and would otherwise cause errors. As a result, this standard
specifies a fixed EMISSIVITY value for use in a SCREENING THERMOGRAPH .
Since the TARGET is required to include the inner canthus, which is shaped like the entrance to
a cavity, its shape can cause an enhancement in EMISSIVITY compared to flat surfaces.
Sweating does cause surface cooling of the skin and changes its temperature. The committee
has not found evidence to indicate that sweating skin has an effect on EMISSIVITY . The
80601-2-59 © IEC:2008 – 25 –
standard requires disclosures to the RESPONSIBLE ORGANIZATION to remind them that controlled
environmental conditions are needed to achieve specified performance.
The principle work on establishing the EMISSIVITY of human skin in the living body was carried
out by the American physiologist Prof. James D Hardy between 1935 and 1939. He
demonstrated that human skin has a high value of 0,98 regardless of skin colour. [25] The
only significant work since that time has been published by Prof. Togowa, whose results using
a reflectance radiometric technique has confirmed the data from Hardy. [61] There are no
credible references to indicate that healthy skin deviates in its EMISSIVITY value. Most authors
cite 0,98 or 0,97 as the value, [58] [59] [71] which at the temperatures of the human body
surface do not show significant differences in temperature whichever value is used.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
specification before it is used.
Annex BB
(normative)
C ALIBRATION SOURCE
The CALIBRATION SOURCE is used as a known temperature reference for characterizing the
SCREENING THERMOGRAPH . As such, it is important that radiance temperature is traceable to
ITS-90 [12] through a national temperature standard. The CALIBRATION of the CALIBRATION
SOURCE should be performed by a laboratory that is accredited to ISO 17025 for that purpose.
The radiance temperature of the CALIBRATION SOURCE shall have an expanded uncertainty not
greater than 0,2 °C and a combined stability and drift better than ± 0,05 °C at the operating
wavelength of the SCREENING THERMOGRAPH .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
The CALIBRATION SOURCE shall have a known EMISSIVITY equal to or greater than 0,998. [8][42]
80601-2-59 © IEC:2008 – 27 –
Annex CC
(informative)
This particular standard has been prepared to support the essential principles of safety and
performance of SCREENING THERMOGRAPHS as medical devices according to ISO/TR 16142.
This particular standard is intended to be acceptable for conformity assessment purposes.
Compliance with this particular standard provides one means of demonstrating conformance
with the specific essential principles of ISO/TR 16142. Other means are possible.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
and the essential principles
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
8.10.4, 8.11
- A.12.7.5 Via IEC 60601-1:2005, 4, 8.11.4, 11.1,
15.4.1, 16.9.1, 16.9.2.1
- A.12.8.1 Via IEC 60601-1:2005, 4, 12.1, 12.4
- A.12.8.2 Via IEC 60601-1:2005, 4, 7.8, 12.3,
12.4
201.12 A.12.8.3 And via IEC 60601-1:2005, 4, 7.4, 7.5,
7.6, 7.8, 12.2
201.7 A.13.1 And via IEC 60601-1:2005, 7, 16.2
- A.14.1 Via IEC 60601-1:2005, 4, 11.1
80601-2-59 © IEC:2008 – 29 –
Bibliography
[1] ISO Guide 99:2007, International vocabulary of metrology – Basic and general concepts
and associated terms (VIM))
[3] ISO 17025:2005, General requirements for the competence of testing and calibration
laboratories
[4] ASTM E1256-95(2007), Standard Test Methods for Radiation Thermometers (Single
Waveband Type), American Society for Testing and Materials
[5] ASTM E1311-89(2004), Standard Test Method for Minimum Detectable Temperature
Difference for Thermal Imaging Systems (MDTD), American Society for Testing and
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Materials
[6] ASTM E1543-00(2006), Standard Test Method for Noise Equivalent Temperature
Difference of Thermal Imaging Systems (NETD), American Society for Testing and
Materials
[7] ASTM E1862-97(2002), Standard Test Methods for Measuring and Compensating for
Reflected Temperature Using Infrared Imaging Radiometers, American Society for
Testing and Materials
[9] ASTM E1933-99, Standard Test Method for Measuring and Compensating for Emissivity
Using Infrared Imaging Radiometers, American Society for Testing and Materials
[11] BSI BS 1041-5:1994 Temperature measurement – Part 5: Guide to selection and use of
radiation pyrometers, British Standard
[13] JIS A 1423:1983 Simplified Test methods for Emissivity by Infrared Radio Meter,
Japanese Industrial Standard
[14] JIS C 1612:2000 Test Methods for Radiation Thermometers, Japanese Industrial
Standard
[15] OIML/CD3, Procedure for calibration and verification of the main characteristics of
thermographic instruments, Organisation Internationale de Métrologie Légale 5
[16] B AE J-S. et al. Prolonged residence of temperate natives in the tropics produces a
suppression of sweating. Pflügers Archiv European Journal of Physiology, (2006) 453/1,
pp.67-72
[17] C HRZANOWSKI , K., et al. Testing of the Thermal Imagers for Non-Destructive Thermal
Testing Applications. ASTM Journal of testing and evaluation, (2000) 28/5, pp.395-402
—————————
5 Available at http://www.oiml.org/download/cds/tc11_sc3_3cd.pdf
– 30 – 80601-2-59 © IEC:2008
[19] D EW ITT , D. P., and N UTTER , G. D. Theory and Practice of Radiation Thermometry. John
Wiley and Sons Inc, (1989)
[21] G ERSHON -C OHEN J., B ORDEN A. and H ERMEL M. Thermography of extremities after
smoking. Br J Radiol (1969) 42, pp.189-191
[22] G ERSHON -C OHEN J., H ABERMAN J. Thermography of smoking. Arch Environ (1968) 16,
pp.637-641
[23] G IBSON , C. E., et al. Radiance Temperature Calibration. NIST Special Publication 250-
43
[24] G OPEL , W., et al. Sensor, A comprehensive Survey. Thermal Sensors (Ed. S CHOLZ , J.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
and R ICOLFI , T.), VCH Publishers Inc, (1990) 4
[25] H ARDY , J. D., et al. The radiation of heat from the human body. IV. The emission,
reflection, and transmission of infra-red radiation by the human skin. J Clin Invest
(1934) 13, pp.817-831
[26] H ASSAN , M. et al. “Infrared Imaging for Tissue Characterization and Function”. in The
Biomedical Engineering Handbook (ED B RONZINO , J.), CRC, (2006) Chpt 30, pp.1-25
[27] Infraspection Institute, Guidelines for Measuring Distance/Target Size Values for
Quantitative Thermal Imaging Cameras. Infraspection Institute document, (2000)
[28] J ONES , T. P. “A NML (Australia) facility for the calibration of radiation thermometers
from 50°C to 2000°C” in Temperature: Its Measurement and Control in Science and
Industry, ed. By S CHOOLEY , J.F., American Institute of Physics, New York (1992) 6,
pp.1105-1110
[29] K AYDOS -D ANIELS , S. C. et al. Body Temperature Monitoring and SARS Fever Hotline,
Taiwan. Emerging Infectious Diseases, (2004) 10/2, pp.373-376
[33] L AI , P.L. Use of thermal imaging system – the seaport experience. National Conference
on T HERMAL I MAGER s for Fever Screening – Selection, Usage and Testing, 30 May 2003,
Singapore
[34] M ACHIN , G., and SERGIENKO , R. A comparative study of size of source effect (SSE)
determination techniques. Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-
Verlag, Berlin, (2002) pp.155-160
[35] M ATVEYEV , M.S. New method for measure of a size-of-source effect in standard
radiation thermometry. Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-
Verlag, Berlin, (2002) pp.167-171
[36] N G , Y.K. and K AW , G. “IR Imagers as Fever Monitoring Devices: Physics, Physiology,
and Clinical Accuracy” in The Biomedical Engineering Handbook (ED B RONZINO , J.),
CRC, (2006) Chpt 24, pp.1-20
80601-2-59 © IEC:2008 – 31 –
[38] N ORTON, P. et al. “Infrared Detectors and Detector Arrays”, in The Biomedical
Engineering Handbook (ED B RONZINO , J.), CRC, (2006) Chpt 37, pp.1-26
[39] P ARK , S.N. and K IM, J.T. “A Monte Carlo Calculation of Lens Abberation Effect on the
Size Of Source Effect” in Radiation Pyrometry, Proc. TEMPMEKO 2001, edited by
F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002) pp.173-177
[40] P ASCOE , D. Comparison of measuring sites for the assessment of body temperature, 6)
[41] P EACOCK , R. Human radiation thermometry and screening for elevated body
temperature in humans. Proc. Of SPIE, (2004) 5405, pp.48-53
[42] P LASSMANN , P. R ING , E., J ONES , C. Quality Assurance of Thermal Imaging Systems in
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Medicine. Therm. Int. (2006) 16/1, pp.9-14
[44] P UŠNICK , I. et al. Bilateral comparison of blackbody cavities for calibration of infrared
ear thermometers between NPL and FE/LMK. Physiol. Meas. (2004) 25, pp.1239-1247
[45] Q UINN , T.J. Temperature. Second Edition, Academic Press limited, (1990)
[46] R EINBERG A. Circadian changes in the temperature of human beings. Bibl Radiol (1975)
6, pp.128-139
[47] R ICOLFI , T. and W ANG , L. Experiments and Remarks on the Size-of-Source Effect in
Precision Radiation Thermometry. Proc. TEMPMEKO 1993, Prague, November. (1993)
9/11, pp. 161-165
[48] R ING , E.F. and A MMER , K. The Technique of Infra red Imaging in Medicine. Therm. Int.
(2000) 10/1, pp.7-14
[49] R ING , E.F. E NGEL J. and P AGE -T HOMAS D. Thermologic Methods in Clinical
Pharmacology – Skin Temperature Measurement in Drug Trials. Int. J Clinical
Pharmacol Therapy & Toxicology (1984) 22, pp.20-24
th
[50] R ING , E.F. et al. Detection of fever in Polish children with infrared imaging. 9
Conference on quantitative infrared thermography, Krakow Poland (2008)
[51] R ING , E.F. et al. Thermography, Radiometry and thermal imaging for fever detection
children. Therm In, (2007) 17/4, p.155-156
[52] S AFFRIN , R.J. Thermal Imaging for Detecting Potential SARS Infection, National
Conference on Thermal Imagers for Fever Screening – Selection, Usage and Testing.
30 May 2003, Singapore
[54] S AKUMA , F., et al. “Distance effect and size-of-source effect of radiation thermometers.”
Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002)
pp.161-166
—————————
6) To be published.
– 32 – 80601-2-59 © IEC:2008
[56] S AUNDERS , P. and R ICOLFI , T. “The Characterisation of a CCD Camera for the Purpose
of Temperature Measurement”. Proc. TEMPMEKO 1996, edited by M ARCARINO , P.,
Torino, (1997) pp.329-334
[57] S MOLINSKI , M. S., H AMBURG , M. A., L EDERBERG , J., Institute of Medicine of the National
Academies, Microbial Threats to Health, (2005) National Academies Press website,
http://www.nap.edu/catalog/11371.html
[58] S TEKETEE , J. Spectral emissivity of the skin and pericardium. Phys Med Biol (1973) 18,
pp.686-694
[59] S TEKETEE , J. The influence of cosmetics and ointments on the spectral emissivity of the
skin. Phys Med Biol (1976) 21, pp.920-930
[60] S TOLWIJK , J.A. and H ARDY , J.D. Handbook of Physiology, Control of body temperature,
Bethesda. Maryland: American Physiological Society, (1977)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
[61] T OGAWA , T. Non-Contact Skin Emissivity: Measurement from Reflectance Using Step
Change in Ambient Radiation Temperature. Clin. Phys. Physiol. Meas., (1989) 10/1,
pp.39-48
[62] T EO , C.W. Development of infrared fever scanning system based on military hardware.
National Conference on T HERMAL I MAGER s for Fever Screening – Selection, Usage and
Testing, 30 May 2003, Singapore
[65] VAN DER H AM , E.W.M., et al. “Intercomparison of Local Temperature Scales with
Transfer Radiation Thermometers between –50 °C and 300 °C.” Proc. TEMPMEKO
2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002) pp.831-837
[66] VAN O OIJEN , A.M.J., et al. Heat transfer in humans: ambient temperature, body
temperature and resting metabolic rate. Maastricht University and Eindhoven University
of Technology Document
[67] W ANG , L. Characterisation of thermal imagers for fever screening. National Conference
on T HERMAL I MAGER s for Fever Screening – Selection, Usage and Testing, 30 May 2003,
Singapore
[68] W ANG , L. and Cheong, T.L.. “Facilities and Techniques for Maintaining a Radiation
Temperature Scale Between 420 °C and 1100 °C.” Proc. TEMPMEKO 1996, edited by
M ARCARINO , P., Torino, (1997) pp.257-260
[69] W ANG , L. “Calibration of Infrared Thermometers at Emissivity Setting Less Than 1”.
Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002)
pp.735-740
[70] W ANG , L. Summary on testing of THERMAL IMAGERs for fever screening. Internal report,
National Metrology Centre, SPRING Singapore, August 2003
[71] W ATMOUGH , D. J. et al. Emissivity of human skin in-vivo between 2.0 μ and 5.4 μ
measured at normal incidence. Nature (1968) 218, p.885
[74] W HO , Frequently asked questions about the international health regulations, World
Health Organization website, http://www.who.int/csr/ihr/howtheywork/faq/en/
[75] W U , M. Stop outbreak of SARS with Infrared Cameras. Proc. Of SPIE, (2004) 5405,
pp.98-105
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
– 34 – 80601-2-59 © IEC:2008
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
FACE ......................................................................................................................... 201.3.206
HAZARD ................................................................................................ IEC 60601-1:2005, 3.39
HAZARDOUS SITUATION ........................................................................... IEC 60601-1:2005, 3.40
IMAGE PIXEL ............................................................................................................... 201.3.207
INTENDED USE ....................................................................................... IEC 60601-1:2005, 3.44
LABORATORY ACCURACY .............................................................................................. 201.3.208
LOW PRIORITY ..................................................................................... IEC 60601-1-8:2006, 3.27
MANUFACTURER ..................................................................................... IEC 60601-1:2005, 3.55
MEDICAL ELECTRICAL EQUIPMENT ( ME EQUIPMENT ).................................... IEC 60601-1:2005, 3.63
MEDICAL ELECTRICAL SYSTEM ( ME SYSTEM ) .............................................. IEC 60601-1:2005, 3.64
MEDIUM PRIORITY ................................................................................ IEC 60601-1-8:2006, 3.28
NETWORK / DATA COUPLING ...................................................................... IEC 60601-1:2005, 3.68
NORMAL USE .......................................................................................... IEC 60601-1:2005, 3.71
OPERATOR ............................................................................................ IEC 60601-1:2005, 3.73
PATIENT ................................................................................................ IEC 60601-1:2005, 3.76
PHYSIOLOGICAL ALARM CONDITION ....................................................... IEC 60601-1-8:2006, 3.31
PROGRAMABLE ELECTRICAL MEDICAL SYSTEM ( PEMS ) ................................ IEC 60601-1:2005, 3.90
RESPONSIBLE ORGANIZATION ................................................................ IEC 60601-1:2005, 3.101
SCREENING THERMOGRAPH .......................................................................................... 201.3.209
SELF - CORRECTION ...................................................................................................... 201.3.210
SKIN TEMPERATURE ..................................................................................................... 201.3.211
TARGET ...................................................................................................................... 201.3.212
TARGET PLANE ............................................................................................................ 201.3.213
TECHNICAL ALARM CONDITION .............................................................. IEC 60601-1-8:2006, 3.36
THRESHOLD TEMPERATURE .......................................................................................... 201.3.214
WORKABLE TARGET PLANE ........................................................................................... 201.3.215
___________
LICENSED TO MECON Limited. - RANCHI/BANGALORE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
– 36 – 80601-2-59 © CEI:2008
SOMMAIRE
AVANT-PROPOS .................................................................................................................. 37
INTRODUCTION................................................................................................................... 39
201.1 Domaine d'application, objet et normes connexes ................................................. 40
201.2 Références normatives .......................................................................................... 41
201.3 Termes et définitions ............................................................................................. 42
201.4 Exigences générales ............................................................................................. 44
201.5 Exigences générales relatives aux essais des APPAREILS EM .................................. 44
201.6 Classification des APPAREILS EM et des SYSTÈMES EM .............................................. 45
201.7 Identification, marquage et documentation des APPAREILS EM ................................ 45
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.8 Protection contre les DANGERS d'origine électrique provenant des APPAREILS EM .... 46
201.9 Protection contre les DANGERS mécaniques des APPAREILS EM et SYSTÈMES EM ...... 46
201.10 Protection contre les DANGERS dus aux rayonnements involontaires ou
excessifs............................................................................................................... 46
201.11 Protection contre les températures excessives et les autres DANGERS ................... 46
201.12 Précision des commandes et des instruments et protection contre les
caractéristiques de sortie présentant des dangers ................................................ 46
201.13 S ITUATIONS DANGEREUSES et conditions de défaut .................................................. 47
201.14 S YSTÈMES ÉLECTROMÉDICAUX PROGRAMMABLES ( SEMP ) ............................................ 47
201.15 Construction de l’ APPAREIL EM ................................................................................ 47
201.16 S YSTÈMES EM ......................................................................................................... 48
201.17 Compatibilité électromagnétique des APPAREILS EM et des SYSTÈMES EM ................ 48
201.101 Précision de laboratoire d’un IMAGEUR THERMIQUE .................................................. 48
201.102 Conditions d’alarme de l’imageur thermique .......................................................... 51
Annexes ............................................................................................................................... 53
Annexe C (informative) Guide pour le marquage et exigences d'étiquetage pour les
APPAREILS EM et les SYSTÈMES EM .......................................................................................... 53
APPAREILS ÉLECTROMÉDICAUX –
AVANT-PROPOS
1) La Commission Electrotechnique Internationale (CEI) est une organisation mondiale de normalisation
composée de l'ensemble des comités électrotechniques nationaux (Comités nationaux de la CEI). La CEI a
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
pour objet de favoriser la coopération internationale pour toutes les questions de normalisation dans les
domaines de l'électricité et de l'électronique. A cet effet, la CEI – entre autres activités – publie des Normes
internationales, des Spécifications techniques, des Rapports techniques, des Spécifications accessibles au
public (PAS) et des Guides (ci-après dénommés "Publication(s) de la CEI"). Leur élaboration est confiée à des
comités d'études, aux travaux desquels tout Comité national intéressé par le sujet traité peut participer. Les
organisations internationales, gouvernementales et non gouvernementales, en liaison avec la CEI, participent
également aux travaux. La CEI collabore étroitement avec l'Organisation Internationale de Normalisation (ISO),
selon des conditions fixées par accord entre les deux organisations.
2) Les décisions ou accords officiels de la CEI concernant les questions techniques représentent, dans la mesure
du possible, un accord international sur les sujets étudiés, étant donné que les Comités nationaux de la CEI
intéressés sont représentés dans chaque comité d’études.
3) Les Publications de la CEI se présentent sous la forme de recommandations internationales et sont agréées
comme telles par les Comités nationaux de la CEI. Tous les efforts raisonnables sont entrepris afin que la CEI
s'assure de l'exactitude du contenu technique de ses publications; la CEI ne peut pas être tenue responsable
de l'éventuelle mauvaise utilisation ou interprétation qui en est faite par un quelconque utilisateur final.
4) Dans le but d'encourager l'uniformité internationale, les Comités nationaux de la CEI s'engagent, dans toute la
mesure possible, à appliquer de façon transparente les Publications de la CEI dans leurs publications
nationales et régionales. Toutes divergences entre toutes Publications de la CEI et toutes publications
nationales ou régionales correspondantes doivent être indiquées en termes clairs dans ces dernières.
5) La CEI n’a prévu aucune procédure de marquage valant indication d’approbation et n'engage pas sa
responsabilité pour les équipements déclarés conformes à une de ses Publications.
6) Tous les utilisateurs doivent s'assurer qu'ils sont en possession de la dernière édition de cette publication.
7) Aucune responsabilité ne doit être imputée à la CEI, à ses administrateurs, employés, auxiliaires ou
mandataires, y compris ses experts particuliers et les membres de ses comités d'études et des Comités
nationaux de la CEI, pour tout préjudice causé en cas de dommages corporels et matériels, ou de tout autre
dommage de quelque nature que ce soit, directe ou indirecte, ou pour supporter les coûts (y compris les frais
de justice) et les dépenses découlant de la publication ou de l'utilisation de cette Publication de la CEI ou de
toute autre Publication de la CEI, ou au crédit qui lui est accordé.
8) L'attention est attirée sur les références normatives citées dans cette publication. L'utilisation de publications
référencées est obligatoire pour une application correcte de la présente publication.
9) L’attention est attirée sur le fait que certains des éléments de la présente Publication de la CEI peuvent faire
l’objet de droits de propriété intellectuelle ou de droits analogues. La CEI ne saurait être tenue pour
responsable de ne pas avoir identifié de tels droits de propriété et de ne pas avoir signalé leur existence.
La Norme internationale CEI 80601-2-59 a été établie par un groupe de travail mixte du sous-
comité 62D: Appareils électromédicaux, du comité d'études 62 de la CEI: Equipements
électriques dans la pratique médicale, et du sous-comité SC3: Ventilateurs pulmonaires et
équipements connexes, du comité technique 121 de l'ISO: Matériel d'anesthésie et de
réanimation respiratoire.
Le rapport de vote indiqué dans le tableau ci-dessus donne toute information sur le vote ayant
abouti à l'approbation de cette norme particulière. A l’ISO, la norme a été approuvée par 12
membres P sur 15 ayant voté.
– 38 – 80601-2-59 © CEI:2008
– “article” désigne l’une des dix-sept sections numérotées dans la table des matières, avec
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
toutes ses subdivisions (par exemple, l’Article 7 inclut les paragraphes 7.1, 7.2, etc);
– “paragraphe” désigne une subdivision numérotée d’un article (par exemple, 7.1, 7.2 et
7.2.1 sont tous des paragraphes appartenant à l’Article 7).
Dans la présente norme, les références à des articles sont précédées du mot “Article” suivi du
numéro de l’article concerné. Dans la présente norme particulière, les références aux
paragraphes utilisent uniquement le numéro du paragraphe concerné.
Dans la présente norme, la conjonction "ou" est utilisée avec la valeur d'un "ou inclusif", ainsi
un énoncé est vrai si une combinaison des conditions quelle qu'elle soit est vraie.
Les formes verbales utilisées dans la présente norme sont conformes à l’usage donné à
l’Annexe H des Directives ISO/CEI, Partie 2. Pour les besoins de la présente norme:
Lorsqu'un astérisque (*) est utilisé comme premier caractère devant un titre, ou au début d’un
titre d'alinéa ou de tableau, il indique l'existence d'une ligne directrice ou d'une justification à
consulter à l’Annexe AA.
Une liste de toutes les parties de la série CEI 60601, publiées sous le titre général: Appareils
électromédicaux, peut être consultée sur le site web de la CEI.
Le comité a décidé que le contenu de cette publication ne sera pas modifié avant la date de
maintenance indiquée sur le site web de la CEI sous "http://webstore.iec.ch" dans les
données relatives à la publication recherchée. A cette date, la publication sera
• reconduite,
• supprimée,
• remplacée par une édition révisée, ou
• amendée.
80601-2-59 © CEI:2008 – 39 –
INTRODUCTION
Les exigences minimales de sécurité spécifiées dans la présente norme particulière sont
considérées comme assurant un degré pratique de sécurité dans le fonctionnement des
APPAREILS EM pour le dépistage des humains fébriles.
Le présent document décrit les APPAREILS EM qui utilisent la technique infrarouge pour
détecter la chaleur émise naturellement sur la peau de la FACE . Des APPAREILS EM de ce type
peuvent être utiles aux ports d’entrée et aux ports de sortie et aux entrées des bâtiments,
dans des conditions d’environnement à l’intérieur, pour séparer les individus fébriles des
individus non fébriles, afin de contribuer à prévenir la propagation des maladies contagieuses.
Un soin particulier peut être nécessaire lors de l’examen des individus dans des conditions
d’environnement variables, mais il a été démontré que l’angle interne de l’œil est un
emplacement de mesure fiable, et il est alimenté par l’artère carotide interne. [40] 1)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Une température interne du corps de 38 °C ou plus a été utilisée comme critère pour réduire
les déplacements au cours de l’épidémie de syndrome respiratoire aigu sévère (SRAS)
(avril 2003). [73] L’US Center for Disease Control informe que le SRAS commence
généralement par une température supérieure à 38 °C, ce qui représente 1 °C au-dessus de
la température interne normale du corps humain, qui se situe autour de 37 °C. [29] Il est
difficile de donner une évaluation précise du nombre de personnes dont la température a été
contrôlée par des mesures à infrarouges en Chine au cours de l’épidémie de SRAS. Il y a des
données officielles du gouvernement chinois indiquant que pendant une période de deux mois
au cours du printemps 2003, 30 millions de voyageurs ont été dépistés en Chine. Dans ce
groupe, 9 292 voyageurs présentant une température élevée ont été détectés et 38 ont été
suspectés d’être porteurs de SRAS. Le SRAS a été diagnostiqué dans 21 de ces cas. Toutes
les températures élevées ont été confirmées à l’aide de mesures cliniques traditionnelles de
la température du corps. Bien qu’il soit difficile de déterminer la température interne du corps
humain de façon précise par des mesures à infrarouges de la TEMPÉRATURE CUTANÉE , il s’agit
d’une méthode possible pour dépister des valeurs de températures élevées. [36] [73] [75]
La présente norme particulière modifie et complète la CEI 60601-1 (troisième édition, 2005):
Appareils électromédicaux – Partie 1: Exigences générales pour la sécurité de base et les
performances essentielles, appelée norme générale dans la suite du texte (voir 1.4).
Une section «Guide général et justifications» contenant, quand cela convient, des notes
explicatives concernant les exigences les plus importantes, figure en Annexe AA. On
considère que la connaissance des raisons qui ont conduit à énoncer ces exigences non
seulement facilitera l’application correcte de la norme, mais accélérera en son temps, toute
révision rendue nécessaire par suite de changements dans la pratique clinique ou en raison
des évolutions technologiques. Cependant, cette annexe ne fait pas partie des exigences de
la présente norme.
—————————
1) Les chiffres entre crochets renvoient à la Bibliographie.
– 40 – 80601-2-59 © CEI:2008
APPAREILS ÉLECTROMÉDICAUX –
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Remplacement:
NOTE Un IMAGEUR THERMIQUE est destiné au dépistage et à la détection des sujets humains avec des
TEMPÉRATURES CUTANÉES élevées supérieures à la normale. Il est nécessaire qu’une TEMPÉRATURE CUTANÉE élevée
soit suivie d’une mesure ultérieure de la température à l’aide d’un thermomètre médical (voir CEI 80601-2-56).
201.1.2 Objet
Remplacement:
L’objet de la présente Norme particulière est d’établir des exigences particulières pour la
SÉCURITÉ DE BASE et les PERFORMANCES ESSENTIELLES des IMAGEURS THERMIQUES tels qu’ils
sont définis en 201.3.209.
Addition:
La présente norme particulière fait référence aux normes collatérales applicables énumérées
à l’Article 2 de la norme générale et à l’Article 2 de la présente norme particulière.
La CEI 60601-1-3 ne s’applique pas. Toutes les autres normes collatérales publiées dans la
série CEI 60601-1 s’appliquent telles qu’elles sont publiées.
Remplacement:
—————————
2) La norme générale est la CEI 60601-1:2005, Appareils électromédicaux – Partie 1: Exigences générales pour la
sécurité de base et les performances essentielles.
80601-2-59 © CEI:2008 – 41 –
Dans la série CEI 60601, des normes particulières peuvent modifier, remplacer ou supprimer
des exigences contenues dans la norme générale et dans les normes collatérales en fonction
de ce qui est approprié à l' APPAREIL EM considéré, et elles peuvent ajouter d’autres exigences
de SECURITE DE BASE et de PERFORMANCES ESSENTIELLES .
Une exigence d’une norme particulière prévaut sur l'exigence correspondante de la norme
générale.
Par souci de concision, la CEI 60601-1 est désignée, dans la présente norme particulière, par
le terme “norme générale”. Les normes collatérales sont désignées par leur numéro de
document.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
collatérale applicable avec le préfixe “20x”, où x est (sont) le (les) dernier(s) chiffre(s) du
numéro de document de la norme collatérale (par exemple 202.4 dans la présente norme
particulière aborde le contenu de l’Article 4 de la norme collatérale 60601-1-2, 203.4 dans la
présente norme particulière aborde le contenu de l’Article 4 de la norme collatérale
60601-1-3, etc). Les modifications apportées au texte de la norme générale sont précisées en
utilisant les termes suivants:
“Addition” signifie que le texte de la présente norme particulière vient s’ajouter aux exigences
de la norme générale ou de la norme collatérale applicable.
Les paragraphes ou les figures qui sont ajoutés à ceux de la norme générale sont numérotés
à partir de 201.101. Toutefois, en raison du fait que les définitions dans la norme générale
sont numérotées 3.1 à 3.139, les définitions complémentaires dans la présente norme sont
numérotées à partir de 201.3.201. Les annexes supplémentaires sont notées AA, BB, etc, et
les alinéas supplémentaires aa), bb), etc.
Les paragraphes ou les figures qui sont ajoutés à une norme collatérale sont numérotés à
partir de 20x, où “x” est le chiffre de la norme collatérale, par exemple 202 pour la
CEI 60601-1-2, 203 pour la CEI 60601-1-3, etc.
L’expression “la présente norme” est utilisée pour se référer à la norme générale, à toutes les
normes collatérales applicables et à la présente norme particulière, considérées ensemble.
NOTE Une liste de références informatives est donnée dans la bibliographie commençant à la page 64.
Addition:
Pour les besoins du présent document, les termes et définitions donnés dans la
CEI 60601-1:2005 et la CEI 60601-1-8:2006 s’appliquent avec les exceptions suivantes:
NOTE Un index des termes définis est donné à partir de la page 69.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Addition:
201.3.201
ÉTALONNAGE
ensemble des opérations établissant, dans des conditions spécifiées, la relation entre des
valeurs de grandeurs indiquées par un appareil de mesure, ou un système de mesure, ou des
valeurs représentées par une mesure matérialisée ou un matériau de référence et les valeurs
correspondantes réalisées par des étalons
201.3.202
SOURCE D ’ ÉTALONNAGE
source de référence corps noir à rayonnement infrarouge de température et d’ ÉMISSIVITÉ
connues et permettant la traçabilité
201.3.203
DÉTECTEUR
capteur thermique à infrarouges ou ensemble de capteurs capable(s) de détecter l’énergie
thermique à infrarouges rayonnant de la surface de la FACE ou de tout autre objet
NOTE Le DÉTECTEUR répond au rayonnement infrarouge net et convertit cette réponse en signaux électriques.
201.3.204
ÉMISSIVITÉ
rapport de la vitesse de propagation thermique émise de l’énergie électromagnétique émise
par un objet à cause de sa température propagée dans une direction donnée, par unité
d’angle solide autour de cette direction et par unité de surface projetée perpendiculairement à
la direction d’une surface sur celle d’un corps noir idéal à la même température et dans les
mêmes conditions spectrales
NOTE 1 L’ ÉMISSIVITÉ de la peau sèche d’un homme est acceptée comme étant de 0,98.
201.3.205
SOURCE DE RÉFÉRENCE DE LA TEMPÉRATURE EXTERNE
partie de l’ IMAGEUR THERMIQUE utilisée pour assurer un fonctionnement précis entre les
ÉTALONNAGES à l’aide d’une source de rayonnement à infrarouges de température et
d’ ÉMISSIVITÉ connues
—————————
3) To be published.
4) ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA, 19428-2959 USA
80601-2-59 © CEI:2008 – 43 –
NOTE La SOURCE DE RÉFÉRENCE DE LA TEMPÉRATURE EXTERNE est en principe visualisée dans chaque
thermogramme ou avant chaque thermogramme.
201.3.206
FACE
face crâniale antérieure du PATIENT mesurée
201.3.207
PIXEL IMAGE
détections thermiques à infrarouges individuelles dans le DÉTECTEUR
NOTE Le nombre de PIXELS IMAGE est donné sous forme de grille, par exemple nombre de capteurs à
l'horizontale (H) par nombre de capteurs à la verticale (V), par exemple pour un ensemble de DÉTECTEURS 120 (H)
x 120 (V), le nombre de PIXELS IMAGE serait de 14,400.
201.3.208
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
PRÉCISION DE LABORATOIRE
degré de concordance entre le résultat d’une mesure (avec un IMAGEUR THERMIQUE ) et la
valeur vraie du mesurande
NOTE La PRÉCISION DE LABORATOIRE est un concept qualitatif. Pour une description quantitative, il convient
d’utiliser le terme « incertitude ».
201.3.209
IMAGEUR THERMIQUE
APPAREIL EM OU SYSTÈME EM qui:
– détecte les rayonnements infrarouges émis par la FACE , à partir duquel un thermogramme
est obtenu à partir de la CIBLE ;
– détecte les rayonnements infrarouges émis par une SOURCE DE RÉFÉRENCE DE LA
TEMPÉRATURE EXTERNE ;
NOTE 1 Un IMAGEUR THERMIQUE est un APPAREIL EM de dépistage de températures sans contacts et non invasif,
utilisé pour mesurer la température de la FACE , et pour indiquer la région dépistée avec une couleur différente si la
température est au-dessus du réglage de la TEMPÉRATURE DE SEUIL .
NOTE 2 Un IMAGEUR THERMIQUE doit identifier la CIBLE à partir du thermogramme, afin d’obtenir la valeur de
température CIBLE .
201.3.210
AUTOCORRECTION
processus automatique réalisé pour compenser la dérive du DÉTECTEUR
NOTE Un DÉTECTEUR microbolomètre non refroidi peut être soumis à une dérive significative de ses mesures
dans le temps.
EXEMPLE Pour réduire l’erreur provoquée par une dérive possible, l' IMAGEUR THERMIQUE réalise une
AUTOCORRECTION .
201.3.211
TEMPÉRATURE CUTANÉE
température de la surface de la peau mesurée depuis la SURFACE CIBLE UTILE d’un IMAGEUR
THERMIQUE , avec un réglage approprié de l’ ÉMISSIVITÉ de la peau
NOTE L’ ÉMISSIVITÉ de la peau sèche d’un homme est acceptée comme étant de 0,98.
201.3.212
CIBLE
région de la FACE choisie pour la comparaison de la TEMPÉRATURE DE SEUIL
– 44 – 80601-2-59 © CEI:2008
201.3.213
SURFACE CIBLE
plan de focalisation perpendiculaire à la ligne droite d’un IMAGEUR THERMIQUE
201.3.214
TEMPÉRATURE DE SEUIL
réglage de température, au-dessus duquel l’ IMAGEUR THERMIQUE indique que la CIBLE est
potentiellement fébrile
201.3.215
SURFACE CIBLE UTILE
région de la SURFACE CIBLE qui satisfait aux exigences de performances spécifiées
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
NOTE La SURFACE CIBLE UTILE peut représenter l’ensemble ou une partie de la SURFACE CIBLE .
Addition:
Addition:
Exigence Paragraphe
NOTE La défaillance pour indiquer que l' IMAGEUR THERMIQUE n’est pas capable de
fonctionner conformément à la spécification est considérée comme une perte de
PERFORMANCE ESSENTIELLE .
Paragraphes complémentaires:
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Les instructions d’utilisation doivent:
a) instruire l’OPÉRATEUR pour s’assurer que la FACE n’est pas obstruée par des cheveux, des
lunettes et d’autres objets parce que leur présence perturbera la capacité d'un IMAGEUR
THERMIQUE à détecter un état fébrile;
b) indiquer la position recommandée de la FACE exigée pour obtenir la CIBLE par rapport au
chemin optique de l’ IMAGEUR THERMIQUE ;
EXEMPLE Position de la FACE par rapport à la lentille de la caméra infrarouge.
c) une explication des effets dus aux sources à infrarouges environnementales telles que la
lumière solaire, les sources électriques proches et l’éclairage, et des instructions selon
lesquelles il convient de les réduire;
NOTE 2 L’ ORGANISME RESPONSABLE doit être informé du type d’éclairage utilisé dans la zone de dépistage. Il
convient d’éviter les éclairages tels que les lampes à incandescence, les lampes à halogène, les lampes
tungstène-halogène-quartz et autres types de lampes qui produisent des interférences (chaleur) significatives.
NOTE 3 Il convient que la surface choisie pour le dépistage ait un fond non réfléchissant et un rayonnement
infrarouge réfléchi minimal provenant de l’environnement.
d) une explication des effets dus à la circulation d’air, et des instructions selon lesquelles il
convient de les réduire.
NOTE 4 Les courants d’air provenant des conduits de climatisation peuvent provoquer un refroidissement ou un
réchauffement forcé de la FACE , et il convient qu’ils soient atténués ou diffus pour empêcher la circulation d’air de
souffler directement sur le PATIENT .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
L’Article 10 de la norme générale s’applique.
Paragraphes complémentaires:
201.12.2.101 Affichage
L’ IMAGEUR THERMIQUE doit être muni d’un affichage couleur isothermique pour l’interprétation
visuelle et l’identification rapide du thermogramme de la SURFACE CIBLE UTILE . L’indication de
la plage de températures, avec code couleurs/échelle de températures, doit être affichée.
L’ IMAGEUR THERMIQUE doit être muni d’au moins un mode de correspondance couleur dans
lequel les couleurs suivent l’ordre du spectre visible (par exemple, échelle arc-en-ciel), de
sorte que le bleu représente le plus froid et le rouge le plus chaud.
Il convient que l’ IMAGEUR THERMIQUE soit capable de fonctionner en temps quasi réel pour le
dépistage rapide et efficace d’une population de masse. Le FABRICANT doit estimer la cadence
de mesure (temps entre les mesures) pour une UTILISATION NORMALE de l’ IMAGEUR THERMIQUE .
Le temps moyen exigé par un IMAGEUR THERMIQUE pour mesurer, traiter et afficher la
80601-2-59 © CEI:2008 – 47 –
température de la FACE et la cadence de mesure de l' IMAGEUR THERMIQUE doit être indiqué
dans la description technique.
La conformité est vérifiée par des essais fonctionnels et par l'examen des instructions
d'utilisation.
L’affichage minimal de la SURFACE CIBLE UTILE doit être de 240 PIXELS IMAGE sur 180 PIXELS
IMAGE . En UTILISATION NORMALE , le thermogramme de la FACE doit remplir au moins 180 PIXELS
IMAGE sur 135 PIXELS IMAGE . Il convient que la SURFACE CIBLE UTILE soit perpendiculaire à la
FACE , afin d’améliorer la performance. [42] Si l’ IMAGEUR THERMIQUE exige que l’ OPÉRATEUR
cadre le thermogramme de la FACE sur la SURFACE CIBLE UTILE , un guide ou un masque doit
être prévu dans l’image de la SURFACE CIBLE UTILE sur l’affichage. [42]
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
NOTE Un affichage de 180 PIXELS IMAGE sur 135 PIXELS IMAGE représente 75 % de l’affichage minimal de la
SURFACE CIBLE UTILE .
La taille des coordonnées de la SURFACE CIBLE UTILE , si elle est inférieure à la SURFACE CIBLE
DE L ’ IMAGEUR THERMIQUE ,
doit être indiquée dans les instructions d’utilisation.
La conformité est vérifiée par des essais fonctionnels et par l'examen des instructions
d'utilisation.
Addition:
L’ IMAGEUR THERMIQUE doit être muni d’un COUPLAGE DE RÉSEAUX/ DE DONNÉES pouvant
communiquer les éléments suivants:
Il convient que l’ IMAGEUR THERMIQUE soit capable de communiquer une image lumineuse
visible de la FACE .
201.16 SYSTÈMES EM
Nouveaux articles:
201.101.1 * C IBLE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
L’emplacement de mesure de la CIBLE doit inclure l’angle interne de chaque œil.
L’ IMAGEUR THERMIQUE doit être capable de créer un thermogramme de la FACE au moins sur la
plage comprise entre 30 °C et 40 °C.
L'incertitude de l’ IMAGEUR THERMIQUE , u ST , doit être calculée. Le calcul doit inclure toutes les
composantes appropriées de l’incertitude, conformément à la Équation 1. Les termes inclus
dans le calcul de l’incertitude de l’ IMAGEUR THERMIQUE doivent être indiqués dans la
description technique.
où:
u 2 = u CS 2 + u ST 2 (2)
80601-2-59 © CEI:2008 – 49 –
où:
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
b) Attendre jusqu’à la disparition de la période de stabilisation de la CONDITION D ’ ALARME
TECHNIQUE .
La TEMPÉRATURE DE SEUIL doit être réglable au moins sur la plage comprise entre 34 °C et
39 °C avec des incréments inférieurs ou égaux à 0,1 °C.
EXEMPLE La différence entre la TEMPÉRATURE CUTANÉE mesurée et la température buccale déduite d’une
population d'échantillons d'étude représentatifs peut guider au choix de la TEMPÉRATURE DE SEUIL .
201.101.3.1 P ERFORMANCE
201.101.3.2 * D IMENSIONS
PIXELS IMAGE comme la CIBLE , afin de réduire l’effet de la non-uniformité à travers la SURFACE
CIBLE UTILE .
NOTE Une dérive et une stabilité faibles assurent que la PRÉCISION DE LABORATOIRE de l’ IMAGEUR THERMIQUE est
maintenue. Bien que la dérive puisse être réduite si une AUTOCORRECTION est effectuée plus régulièrement, une
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
AUTOCORRECTION effectuée trop souvent crée des interruptions dans l'aptitude à dépister des PATIENTS avec
l’ IMAGEUR THERMIQUE .
a) La conformité est vérifiée par l’essai suivant:
b) Monter l’IMAGEUR THERMIQUE conformément au DOCUMENT D ’ACCOMPAGNEMENT .
c) Attendre jusqu’à la disparition de la période de stabilisation de la CONDITION D ’ALARME
TECHNIQUE .
NOTE La résolution minimale de la différence de température reflète la différence de température la plus faible
(ou température équivalente) que l' IMAGEUR THERMIQUE peut détecter constamment dans la SURFACE CIBLE UTILE .
NOTE Une bonne uniformité dans la SURFACE CIBLE permet la flexibilité de déploiement de l’ IMAGEUR THERMIQUE
pour dépister la FACE .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.101.7 * P OSITION DE LA FACE
La SURFACE CIBLE UTILE de l’ IMAGEUR THERMIQUE doit recevoir une FACE située à une hauteur
comprise entre 0,75 m et 2,2 m au-dessus du sol. Cette exigence peut être satisfaite en
déplaçant la caméra infrarouge.
Une ÉMISSIVITÉ de 0,98 doit être utilisée dans le calcul de la TEMPÉRATURE CUTANÉE par
l’ IMAGEUR THERMIQUE . [25] [61]
Un IMAGEUR THERMIQUE doit être muni d’un SYSTÈME D ’ ALARME comprenant une CONDITION
D ’ ALARME TECHNIQUE au moins de PRIORITE FAIBLE au cours de la période de stabilisation
suivant le démarrage, jusqu’à ce que l' IMAGEUR THERMIQUE soit capable de fonctionner
conformément à la spécification. Voir aussi les paragraphes 201.101.4 et 201.101.8.
Un IMAGEUR THERMIQUE doit être muni d’un SYSTEME D ’ ALARME comprenant une CONDITION
D ’ ALARME PHYSIOLOGIQUE au moins de PRIORITÉ MOYENNE indiquant lorsque la TEMPÉRATURE DE
SEUIL a été dépassée à l’intérieur de la CIBLE . La description technique doit inclure une
description de l’algorithme utilisé pour déterminer la TEMPÉRATURE DE SEUIL à partir des PIXELS
IMAGE dans le champ de la CIBLE .
La conformité est vérifiée par examen de la description technique et par des essais
fonctionnels.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
80601-2-59 © CEI:2008 – 53 –
Annexes
Annexe C
(informative)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
201.C.5 DOCUMENTS D ’ ACCOMPAGNEMENT , instructions d’utilisation
Addition:
Addition:
Référence à la CEI ISO/TR 80600, Appareils électromédicaux – Déploiement, mise en œuvre 201.7.9.3.101 a)
et lignes directrices opérationnelles pour identifier les humains fébriles à l’aide d’un imageur
thermique
Recommandation et effets de l’humidité relative 201.7.9.3.101 b)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Temps de réponse 201.12.2.102
Résolution spatiale d’un PIXEL IMAGE 201.101.9
Termes inclus dans le calcul de l’incertitude 201.101.2.2
Algorithme de la TEMPÉRATURE DE SEUIL 201.102.2
Cadence de mesure 201.12.2.102
80601-2-59 © CEI:2008 – 55 –
Annexe AA
(informative)
Une température interne du corps de 38 °C ou plus a été utilisée comme critère pour réduire
les déplacements au cours de l’épidémie de syndrome respiratoire aigu sévère (SRAS)
(avril 2003). [73] L’association entre maladie et température a été reconnue depuis les
premiers traités de médecine. La fièvre, l’une des premières observations, a été évaluée
subjectivement par les médecins pendant des siècles. [42] Le but d’un IMAGEUR THERMIQUE
est de déterminer rapidement et précisément la TEMPÉRATURE CUTANÉE en surface d’un être
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
humain en mesurant la CIBLE . La température de la CIBLE est utilisée pour estimer
indirectement la température interne du corps.
Les IMAGEURS THERMIQUES utilisent la technique infrarouge pour détecter la chaleur émise
naturellement sur la peau de la FACE . Les IMAGEURS THERMIQUES ont été utilisés aux ports
d’entrée, aux ports de sortie et aux entrées des bâtiments, dans des conditions
d’environnement à l’intérieur, dans le but de séparer les individus fébriles des individus non
fébriles, afin de contribuer à prévenir la propagation des maladies contagieuses. Un IMAGEUR
THERMIQUE examine la FACE et permet à l’ OPÉRATEUR de quantifier et de visualiser le profil de
TEMPÉRATURE CUTANÉE . Etant donné que la TEMPÉRATURE CUTANÉE dépend fortement de la
circulation de sang dans la peau et des conditions environnementales, il est nécessaire que
les IMAGEURS THERMIQUES soient utilisés dans un environnement stable avec des influences
environnementales minimales, afin de satisfaire aux exigences de PERFORMANCES
ESSENTIELLES .
La distribution des températures internes du corps parmi les humains est bien connue d’après
de nombreuses études. Le problème le plus significatif associé à l’utilisation des IMAGEURS
THERMIQUES pour détecter la fièvre est le fait que la relation entre une TEMPÉRATURE CUTANÉE
élevée et une température interne d'un corps fébrile n'est pas bien établie. La relation peut
être compliquée davantage par la transpiration, la peau du visage en sueur à cause d’un
effort, etc. A la suite de cela, l’utilisation d' IMAGEURS THERMIQUES pour la détection de la fièvre
à partir de FACES n’est pas fondée. A moins que l’utilisation d’ IMAGEURS THERMIQUES pour la
détection de la fièvre à partir de FACES ne soit fondée, il convient que l’état de toutes les
personnes soupçonnées d’être fébriles avec les IMAGEURS THERMIQUES soit confirmé par un
dépistage secondaire avec un thermomètre médical.
Le présent document ne spécifie pas la longueur d'onde infrarouge qu'il convient d'utiliser. La
bande de longueurs d’ondes à rayonnement infrarouge moyen (3 μm à 5 μm) et la bande de
longueurs d’ondes à rayonnement infrarouge long (8 μm à 14 μm) ont été utilisées dans le
dépistage de températures des êtres humains. [25]
– 56 – 80601-2-59 © CEI:2008
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Paragraphe 201.5.3 a)
Afin d’utiliser efficacement un IMAGEUR THERMIQUE , il est nécessaire que les conditions
environnementales soient contrôlées dans une plage que la plupart des individus pourra
supporter. Il convient que les PATIENTS dépistés n’aient pas trop froid ni trop chaud et en
particulier qu’ils ne transpirent pas.
Les échelles de couleurs ont l’avantage de présenter une progression logique des couleurs.
L’œil peut voir et traiter plus de couleurs que des nuances de gris. Les échelles de
températures en couleurs ont été créées à l’origine uniquement à cet effet – une séquence de
couleurs logique et des étapes plus perceptibles – lorsque les moniteurs de l’échelle des gris
avaient une plage dynamique utile limitée. [48] L’échelle température du fer a été conçue
pour des usages industriels (par exemple, pour les métaux en fusion), lorsque la plage de
températures considérée est très large. Pour la plage relativement étroite de TEMPÉRATURES
CUTANÉES chez l'homme, l'échelle arc-en-ciel fournit une meilleure discrimination visuelle.
Pour cette raison, l’échelle arc-en-ciel a été choisie comme exigence.
Les ORGANISMES RESPONSABLES utilisent ces informations pour concevoir les processus de
dépistage appropriés pour leurs institutions.
Cette exigence est destinée à s’assurer qu’il y a suffisamment de PIXELS IMAGE dans le
thermogramme de la FACE et de la CIBLE pour permettre une évaluation précise.
Les ORGANISMES RESPONSABLES et les autorités sanitaires peuvent utiliser ces informations
pour gérer leurs réponses en cas de crises sanitaires.
80601-2-59 © CEI:2008 – 57 –
Les preuves actuelles indiquent que les angles internes des yeux sont l’endroit préféré pour
dépister la fièvre en raison de la stabilité de cet emplacement de mesure. [40] [50] [51]
Cette exigence est destinée à assurer qu’un thermogramme complet de la FACE existera. Un
thermogramme incomplet de la FACE provoqué par des valeurs en dehors de la plage peut
affecter l’aptitude de l’ IMAGEUR THERMIQUE à déterminer la CIBLE .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Le comité pense que c’est la PRÉCISION DE LABORATOIRE radiométrique minimale qui permettra
la séparation entre les sujets fébriles et non fébriles.
Dérive maximale
admissible de
B 0,2 2√3 0,06
l’ IMAGEUR
THERMIQUE , u D
Mesure maximale
admissible de la
non-uniformité de B 0,2 2√3 0,06
l’ IMAGEUR
THERMIQUE , u U
Instabilité maximale
admissible de
B 0,2 2√3 0,06
l’ IMAGEUR
THERMIQUE , u S
MRTD de l’ IMAGEUR
B 0,1 2√3 0,03
THERMIQUE , u MRTD
Incertitude de la
SOURCE A 0,2 2 0,10
D ’ ÉTALONNAGE , u CS
Incertitude type
k=1 0,21
combinée
Incertitude
k=2 0,42
élargie
NOTE La dérive de l’ IMAGEUR THERMIQUE comprend celle de la SOURCE DE RÉFÉRENCE DE LA
TEMPÉRATURE EXTERNE .
– 58 – 80601-2-59 © CEI:2008
Bien que la relation entre la température interne du corps et les TEMPÉRATURES CUTANÉES du
visage n’ait pas été complètement établie, des données indiquent que la plage de
températures attendue est couverte par cette exigence. [40] [50] [51]
Ces exigences sont conçues pour assurer que la SOURCE DE RÉFÉRENCE DE LA TEMPÉRATURE
EXTERNE peut être utilisée pour déterminer si la PRÉCISION DE LABORATOIRE est suffisante ou
non pour permettre une détection fiable de la température CIBLE . Pour répondre à cet objectif,
il est nécessaire de bien connaître la température réelle de la SOURCE DE RÉFÉRENCE DE LA
TEMPÉRATURE EXTERNE . Il est nécessaire que la taille de la SOURCE DE RÉFÉRENCE DE LA
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
TEMPÉRATURE EXTERNE par rapport à sa distance du DÉTECTEUR de l’ IMAGEUR THERMIQUE soit
telle qu’il y ait un nombre suffisant de PIXELS IMAGE , afin de permettre une mesure précise.
Il est nécessaire que la SOURCE D ’ ÉTALONNAGE soit suffisamment grande, de sorte que la
mesure de l’ IMAGEUR THERMIQUE ne soit pas affectée par sa petite taille et pour permettre une
identification claire de la couleur d’affichage dans la SURFACE CIBLE UTILE .
Pour qu’un IMAGEUR THERMIQUE soit efficace, une faible MRTD est nécessaire. Ceci reflète la
sensibilité de l’ IMAGEUR THERMIQUE lorsque l’on fait la distinction entre deux valeurs de
températures proches. Une faible MRTD permet un ajustement précis du réglage de la
TEMPÉRATURE DE SEUIL , afin d’optimiser l'efficacité de l’opération de dépistage.
Les IMAGEURS THERMIQUES sont utilisés pour dépister l’ensemble de la population. Le comité a
choisi une plage minimale d'emplacements CIBLES permettant aux individus les plus valides
d'être dépistés, y compris les individus de très grande taille et les enfants, ainsi que les
individus en fauteuils roulants.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
trois ensembles de DÉTECTEURS différents à plan focal dans le temps. Les symboles
triangulaires montrent les valeurs de températures provenant d’un DÉTECTEUR refroidi qui
n’indique pas la température jusqu’à obtention de la stabilité. Aucun de ces DÉTECTEURS ne
remplit les exigences de PRÉCISION DE LABORATOIRE de la présente norme sans correction.
Cette exigence est destinée à assurer que l’IMAGEUR THERMIQUE dans son ensemble est prêt à
fonctionner.
1
T (°C)
0
0 20 40 60 80 100 120
–1
–2
–3
t (min)
IEC 1760/08
Cette résolution spatiale a été choisie pour assurer qu’il y a une résolution suffisante pour
mesurer la température de la CIBLE . Des mires de résolution de la chaleur peuvent être
utilisées pour réaliser cet essai. [63]
Les publications ont mentionné que l’ ÉMISSIVITÉ de la peau chez l’homme se situe dans la
plage comprise entre 0,96 et 0,98. La mesure de l’ ÉMISSIVITÉ est difficile et l’incertitude de la
mesure pourrait généralement se situer entre 2 % et 5 % (k = 2). La variation du réglage de
l’ÉMISSIVITÉ dans le calcul de la température modifie la température indiquée et entraînerait
autrement des erreurs. En conséquence, la présente norme spécifie une valeur d’ ÉMISSIVITÉ
fixe destinée à être utilisée dans un IMAGEUR THERMIQUE .
– 60 – 80601-2-59 © CEI:2008
Etant donné qu’il est nécessaire que la CIBLE comprenne l'angle interne, qui a la même forme
que l’entrée d’une cavité, sa forme peut entraîner une augmentation de l' ÉMISSIVITÉ par
rapport aux surfaces planes. La sudation provoque un refroidissement à la surface de la peau
et modifie sa température. Le comité n’a pas pu prouver que la sudation de la peau a un effet
sur l’ ÉMISSIVITÉ . La présente norme exige d’informer l’ ORGANISME RESPONSABLE pour lui
rappeler que des conditions environnementales contrôlées sont nécessaires pour atteindre la
performance spécifiée.
Les principaux travaux liés à l’établissement de l’ ÉMISSIVITÉ de la peau d’un humain en vie ont
été réalisés par le physiologiste américain Prof. James D. Hardy entre 1935 et 1939. Il a
démontré que la peau de l’homme a une valeur élevée de 0,98, quelle que soit la couleur de
la peau. [25] Les seuls travaux significatifs depuis cette période ont été publiés par le Prof.
Togowa, dont les résultats en utilisant une technique radiométrique de réflectance ont
confirmé les données de Hardy. [61] Il n’y a pas de références crédibles pour indiquer que la
peau saine s’écarte de sa valeur d’ ÉMISSIVITÉ . La plupart des auteurs citent 0,98 ou 0,97
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
comme la valeur [58] [59] [71] qui, aux températures de la surface du corps humain, ne
présente pas de différences significatives de températures, quelle que soit la valeur utilisée.
Cette exigence assure que l' IMAGEUR THERMIQUE est capable de fonctionner conformément à
la spécification avant d’être utilisé.
Annexe BB
(normative)
S OURCE D ’ ÉTALONNAGE
La SOURCE D ’ ÉTALONNAGE est utilisée comme une référence de température connue pour
caractériser l' IMAGEUR THERMIQUE . Comme telle, il est important que la température de
luminance soit établie par rapport à l’ITS-90 [12] par une température nationale normalisée. Il
convient que l’ ÉTALONNAGE de la SOURCE D ' ÉTALONNAGE soit réalisé par un laboratoire
accrédité selon l’ISO 17025 à cet effet. La température de luminance de la SOURCE
D ’ ÉTALONNAGE doit avoir une incertitude élargie inférieure ou égale à 0,2 °C et une stabilité et
une dérive combinées supérieures à ± 0,05 °C à la longueur d’onde de fonctionnement de
l’ IMAGEUR THERMIQUE .
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
La SOURCE D ’ ÉTALONNAGE doit avoir une ÉMISSIVITÉ connue supérieure ou égale à 0,998.
[8] [42]
– 62 – 80601-2-59 © CEI:2008
Annexe CC
(informative)
La présente norme particulière a été établie pour supporter les principes essentiels de
sécurité et de performance des IMAGEURS THERMIQUES en tant que dispositifs médicaux,
conformément à l’ISO/TR 16142. La présente norme particulière est destinée à être
acceptable à des fins d’évaluation de la conformité.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
et les principes essentiels
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
CEI 60601-1:2005, 4, 10, 12.4.5, 17
- A.11.2.1 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 10, 12.4
- A.11.2.2 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 12
- A.11.3 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 10, 12.4.5.1
- A.11.4 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 7.9.2.1, 7.9.2.2
- A.11.5.1 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 10.1.2, 12.4
- A.11.5.2 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 10.1.2, 12.4
- A.11.5.3 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 10.1.2, 12.4
201.14 A.12.1 Et par l’intermédiaire de la
CEI 60601-1:2005, 4, 14
- A.12.2
- A.12.3 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 7.8, 12.3
201.12 A.12.4 Et par l’intermédiaire de la
CEI 60601-1:2005, 4, 7.8, 12.3
- A.12.5 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 17
- A.12.6 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 8
- A.12.7.1 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 9, 15.3
- A.12.7.2 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 9.6
- A.12.7.3 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 9.6
- A.12.7.4 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 8.10.3, 8.10.4,
8.11
- A.12.7.5 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 8.11.4, 11.1,
15.4.1, 16.9.1, 16.9.2.1
- A.12.8.1 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 12.1, 12.4
- A.12.8.2 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 7.8, 12.3, 12.4
201.12 A.12.8.3 Et par l’intermédiaire de la
CEI 60601-1:2005, 4, 7.4, 7.5, 7.6, 7.8,
12.2
201.7 A.13.1 Et par l’intermédiaire de la
CEI 60601-1:2005, 7, 16.2
- A.14.1 Par l’intermédiaire de la
CEI 60601-1:2005, 4, 11.1
– 64 – 80601-2-59 © CEI:2008
Bibliographie
[2] ISO/TR 16142:2006, Dispositifs médicaux – Lignes directrices pour le choix des normes
correspondant aux principes essentiels reconnus de sécurité et de performance des
dispositifs médicaux
[4] ASTM E1256-95(2007), Standard Test Methods for Radiation Thermometers (Single
Waveband Type), American Society for Testing and Materials
[5] ASTM E1311-89(2004), Standard Test Method for Minimum Detectable Temperature
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Difference for Thermal Imaging Systems (MDTD), American Society for Testing and
Materials
[6] ASTM E1543-00(2006), Standard Test Method for Noise Equivalent Temperature
Difference of Thermal Imaging Systems (NETD), American Society for Testing and
Materials
[7] ASTM E1862-97(2002), Standard Test Methods for Measuring and Compensating for
Reflected Temperature Using Infrared Imaging Radiometers, American Society for
Testing and Materials
[9] ASTM E1933-99, Standard Test Method for Measuring and Compensating for Emissivity
Using Infrared Imaging Radiometers, American Society for Testing and Materials
[11] BSI BS 1041- 5:1989 Temperature measurement – Part 5: Guide to selection and use of
radiation pyrometers, British Standard
[13] JIS A 1423:1983 Simplified Test methods for Emissivity by Infrared Radio Meter,
Japanese Industrial Standard
[14] JIS C 1612:2000 Test Methods for Radiation Thermometers, Japanese Industrial
Standard
[15] OIML/CD3, Procedure for calibration and verification of the main characteristics of
thermographic instruments, Organisation Internationale de Métrologie Légale 5
[16] B AE J-S. et al. Prolonged residence of temperate natives in the tropics produces a
suppression of sweating. Pflügers Archiv European Journal of Physiology, (2006) 453/1,
pp.67-72
[17] C HRZANOWSKI , K., et al. Testing of the Thermal Imagers for Non-Destructive Thermal
Testing Applications. ASTM Journal of testing and evaluation, (2000) 28/5, pp.395-402
[19] D EW ITT , D. P., and N UTTER , G. D. Theory and Practice of Radiation Thermometry. John
Wiley and Sons Inc, (1989)
[21] G ERSHON -C OHEN J., B ORDEN A. and H ERMEL M. Thermography of extremities after
smoking. Br J Radiol (1969) 42, pp.189-191
[22] G ERSHON -C OHEN J., H ABERMAN J. Thermography of smoking. Arch Environ (1968) 16,
pp.637-641
[23] G IBSON , C. E., et al. Radiance Temperature Calibration. NIST Special Publication 250-
43
[24] G OPEL , W., et al. Sensor, A comprehensive Survey. Thermal Sensors (Ed. S CHOLZ , J.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
and R ICOLFI , T.), VCH Publishers Inc, (1990) 4
[25] H ARDY , J. D., et al. The radiation of heat from the human body. IV. The emission,
reflection, and transmission of infra-red radiation by the human skin. J Clin Invest (1934)
13, pp.817-831
[26] H ASSAN , M. et al. “Infrared Imaging for Tissue Characterization and Function”. in The
Biomedical Engineering Handbook (ED B RONZINO , J.), CRC, (2006) Chpt 30, pp.1-25
[27] Infraspection Institute, Guidelines for Measuring Distance/Target Size Values for
Quantitative Thermal Imaging Cameras. Infraspection Institute document, (2000)
[28] J ONES , T. P. “A NML (Australia) facility for the calibration of radiation thermometers
from 50°C to 2000°C” in Temperature: Its Measurement and Control in Science and
Industry, ed. By S CHOOLEY , J.F., American Institute of Physics, New York (1992) 6,
pp.1105-1110
[29] K AYDOS -D ANIELS , S. C. et al. Body Temperature Monitoring and SARS Fever Hotline,
Taiwan. Emerging Infectious Diseases, (2004) 10/2, pp.373-376
[33] L AI , P.L. Use of thermal imaging system – the seaport experience. National Conference
on T HERMAL I MAGER s for Fever Screening – Selection, Usage and Testing, 30 May 2003,
Singapore
[34] M ACHIN , G., and SERGIENKO , R. A comparative study of size of source effect (SSE)
determination techniques. Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-
Verlag, Berlin, (2002) pp.155-160
[35] M ATVEYEV , M.S. New method for measure of a size-of-source effect in standard
radiation thermometry. Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-
Verlag, Berlin, (2002) pp.167-171
[36] N G , Y.K. and K AW , G. “IR Imagers as Fever Monitoring Devices: Physics, Physiology,
and Clinical Accuracy” in The Biomedical Engineering Handbook (ED B RONZINO , J.),
CRC, (2006) Chpt 24, pp.1-20
– 66 – 80601-2-59 © CEI:2008
[38] N ORTON, P. et al. “Infrared Detectors and Detector Arrays”, in The Biomedical
Engineering Handbook (ED B RONZINO , J.), CRC, (2006) Chpt 37, pp.1-26
[39] P ARK , S.N. and K IM, J.T. “A Monte Carlo Calculation of Lens Abberation Effect on the
Size Of Source Effect” in Radiation Pyrometry, Proc. TEMPMEKO 2001, edited by
F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002) pp.173-177
[40] P ASCOE , D. Comparison of measuring sites for the assessment of body temperature, 6)
[41] P EACOCK , R. Human radiation thermometry and screening for elevated body
temperature in humans. Proc. Of SPIE, (2004) 5405, pp.48-53
[42] P LASSMANN , P. R ING , E., J ONES , C. Quality Assurance of Thermal Imaging Systems in
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
Medicine. Therm. Int. (2006) 16/1, pp.9-14
[44] P UŠNICK , I. et al. Bilateral comparison of blackbody cavities for calibration of infrared
ear thermometers between NPL and FE/LMK. Physiol. Meas. (2004) 25, pp.1239-1247
[45] Q UINN , T.J. Temperature. Second Edition, Academic Press limited, (1990)
[46] R EINBERG A. Circadian changes in the temperature of human beings. Bibl Radiol (1975)
6, pp.128-139
[47] R ICOLFI , T. and W ANG , L. Experiments and Remarks on the Size-of-Source Effect in
Precision Radiation Thermometry. Proc. TEMPMEKO 1993, Prague, November. (1993)
9/11, pp. 161-165
[48] R ING , E.F. and A MMER , K. The Technique of Infra red Imaging in Medicine. Therm. Int.
(2000) 10/1, pp.7-14
[49] R ING , E.F. E NGEL J. and P AGE -T HOMAS D. Thermologic Methods in Clinical
Pharmacology – Skin Temperature Measurement in Drug Trials. Int. J Clinical
Pharmacol Therapy & Toxicology (1984) 22, pp.20-24
th
[50] R ING , E.F. et al. Detection of fever in Polish children with infrared imaging. 9
Conference on quantitative infrared thermography, Krakow Poland (2008)
[51] R ING , E.F. et al. Thermography, Radiometry and thermal imaging for fever detection
children. Therm In, (2007) 17/4, p.155-156
[52] S AFFRIN , R.J. Thermal Imaging for Detecting Potential SARS Infection, National
Conference on Thermal Imagers for Fever Screening – Selection, Usage and Testing.
30 May 2003, Singapore
[54] S AKUMA , F., et al. Distance effect and size-of-source effect of radiation thermometers.
Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002)
pp.161-166
—————————
6) Publication à venir.
80601-2-59 © CEI:2008 – 67 –
[56] S AUNDERS , P. and R ICOLFI , T. “The Characterisation of a CCD Camera for the Purpose
of Temperature Measurement”. Proc. TEMPMEKO 1996, edited by M ARCARINO , P.,
Torino, (1997) pp.329-334
[57] S MOLINSKI , M. S., H AMBURG , M. A., L EDERBERG , J., Institute of Medicine of the National
Academies, Microbial Threats to Health, (2005) National Academies Press website,
http://www.nap.edu/catalog/11371.html
[58] S TEKETEE , J. Spectral emissivity of the skin and pericardium. Phys Med Biol (1973) 18,
pp.686-694
[59] S TEKETEE , J. The influence of cosmetics and ointments on the spectral emissivity of the
skin. Phys Med Biol (1976) 21, pp.920-930
[60] S TOLWIJK , J.A. and H ARDY , J.D. Handbook of Physiology, Control of body temperature,
Bethesda. Maryland: American Physiological Society, (1977)
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
[61] T OGAWA , T. Non-Contact Skin Emissivity: Measurement from Reflectance Using Step
Change in Ambient Radiation Temperature. Clin. Phys. Physiol. Meas., (1989) 10/1,
pp.39-48
[62] T EO , C.W. Development of infrared fever scanning system based on military hardware.
National Conference on T HERMAL I MAGER s for Fever Screening - Selection, Usage and
Testing, 30 May 2003, Singapore
[65] VAN DER H AM , E.W.M., et al. “Intercomparison of Local Temperature Scales with
Transfer Radiation Thermometers between –50 °C and 300 °C.” Proc. TEMPMEKO
2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002) pp.831-837
[66] VAN O OIJEN , A.M.J., et al. Heat transfer in humans: ambient temperature, body
temperature and resting metabolic rate. Maastricht University and Eindhoven University
of Technology Document
[67] W ANG , L. Characterisation of thermal imagers for fever screening. National Conference
on T HERMAL I MAGER s for Fever Screening – Selection, Usage and Testing, 30 May 2003,
Singapore
[68] W ANG , L. and Cheong, T.L.. “Facilities and Techniques for Maintaining a Radiation
Temperature Scale Between 420 °C and 1100 °C.” Proc. TEMPMEKO 1996, edited by
M ARCARINO , P., Torino, (1997) pp.257-260
[69] W ANG , L. “Calibration of Infrared Thermometers at Emissivity Setting Less Than 1”.
Proc. TEMPMEKO 2001, edited by F ELLMUTH , B. et al., VDI-Verlag, Berlin, (2002)
pp.735-740
[70] W ANG , L. Summary on testing of THERMAL IMAGER s for fever screening. Internal report,
National Metrology Centre, SPRING Singapore, August 2003
[71] W ATMOUGH , D. J. et al. Emissivity of human skin in-vivo between 2.0 μ and 5.4 μ
measured at normal incidence. Nature (1968) 218, p.885
[74] W HO , Frequently asked questions about the international health regulations, World
Health Organization website, http://www.who.int/csr/ihr/howtheywork/faq/en/
[75] W U , M. Stop outbreak of SARS with Infrared Cameras. Proc. Of SPIE, (2004) 5405,
pp.98-105
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
80601-2-59 © CEI:2008 – 69 –
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
ÉMISSIVITÉ ................................................................................................................. 201.3.204
ÉTALONNAGE .............................................................................................................. 201.3.201
FABRICANT ............................................................................................ CEI 60601-1:2005, 3.55
FACE ......................................................................................................................... 201.3.206
FAIBLE PRIORITÉ ................................................................................. CEI 60601-1-8:2006, 3.27
IMAGEUR THERMIQUE ................................................................................................... 201.3.209
OPÉRATEUR ........................................................................................... CEI 60601-1:2005, 3.73
ORGANISME RESPONSABLE .................................................................... CEI 60601-1:2005, 3.101
PATIENT ................................................................................................ CEI 60601-1:2005, 3.76
PERFORMANCE ESSENTIELLE ................................................................... CEI 60601-1:2005, 3.27
PIXEL IMAGE ............................................................................................................... 201.3.207
PRÉCISION DE LABORATOIRE ........................................................................................ 201.3.208
PRIORITÉ MOYENNE ............................................................................. CEI 60601-1-8:2006, 3.28
SÉCURITÉ DE BASE ................................................................................. CEI 60601-1:2005, 3.10
SITUATION DANGEREUSE ......................................................................... CEI 60601-1:2005, 3.40
SOURCE D ’ ÉTALONNAGE ............................................................................................... 201.3.202
SOURCE DE RÉFÉRENCE DE LA TEMPÉRATURE EXTERNE .................................................. 201.3.205
SURFACE CIBLE UTILE .................................................................................................. 201.3.215
SURFACE CIBLE ........................................................................................................... 201.3.213
SYSTÈME D ' ALARME ............................................................................. CEI 60601-1-8:2006, 3.11
SYSTÈME ÉLECTROMEDICAL ( SYSTEME EM ) ............................................... CEI 60601-1:2005, 3.64
SYSTÈME ÉLECTROMÉDICAL PROGRAMMABLE ( SEMP ) ................................. CEI 60601-1:2005, 3.90
TEMPÉRATURE CUTANÉE .............................................................................................. 201.3.211
TEMPÉRATURE DE SEUIL .............................................................................................. 201.3.214
UTILISATION NORMALE ............................................................................ CEI 60601-1:2005, 3.71
UTILISATION PRÉVUE .............................................................................. CEI 60601-1:2005, 3.44
___________
LICENSED TO MECON Limited. - RANCHI/BANGALORE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
LICENSED TO MECON Limited. - RANCHI/BANGALORE
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
3, rue de Varembé
PO Box 131
CH-1211 Geneva 20
Switzerland
Tel: + 41 22 919 02 11
Fax: + 41 22 919 03 00
[email protected]
www.iec.ch