I28 UserManual
I28 UserManual
www.cincinnati-test.com
IMPORTANT INFORMATION
Information in this document is subject to change without notice and does not represent a
commitment on the part of Cincinnati Test Systems, Inc. No part of the manual and/or
software may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying, recording, or information storage and retrieval
systems, for any purpose, other than the purchaser’s personal use, without the express
permission of Cincinnati Test Systems, Inc.
No patent liability is assumed with respect to the use of the information contained herein.
While every precaution has been taken in the preparation of this book, Cincinnati Test
Systems, Inc. assumes no responsibility for errors or omissions.
Because of the variety of uses for this equipment and because of the differences between
this solid-state equipment and electromechanical equipment, the user of and those
responsible for applying this equipment must satisfy themselves as to the acceptability of
each application and use of the equipment. In no event will Cincinnati Test Systems be
responsible or liable for indirect or consequential damages resulting from the use or
application of this equipment.
The illustrations, charts, and layout examples shown in this manual are intended solely to
illustrate the text of this manual. Because of the many parameters and requirements
associated with any particular installation, Cincinnati Test Systems cannot assume
responsibility or liability for actual use based on the illustrative uses and applications.
CAUTION – When you see this warning symbol on the product, refer to the instruction
manual for information before proceeding.
CAUTION – To avoid personal injury due to electric shock, disconnect all power services
before servicing.
If this instrument is not used according to the instructions of this manual, the protection
provided by this equipment design may be compromised.
www.cincinnati-test.com
This page is intentionally blank.
www.cincinnati-test.com
S E N T I N E L I 2 8 I N D E X
Table of Contents
Chapter 1 – Introduction ............................................................................. 1
Unpacking ....................................................................................... 1
Instrument Attachment & Clearance ................................................ 1
Installation ....................................................................................... 1
DC Power Connection ..................................................................... 2
User Interface Navigation ................................................................ 3
Menu Structure ................................................................................ 5
Chapter 2 – Setup....................................................................................... 7
Program Linking & Sequencing ....................................................... 8
Selection of Test Types ................................................................... 9
Test Type Table ...................................................................................... 10
Tooling Control .............................................................................. 12
Programming the Inputs and Outputs ............................................ 12
Channel Configuration ................................................................... 12
Setting the Units of Measure ................................................................... 12
Setting the Leak Standard Values .......................................................... 14
Program Configuration .................................................................. 15
Pressure Select ....................................................................................... 15
Selecting the Test Port ............................................................................ 15
Setting the Pneumatic Regulator ............................................................ 15
Setting the Test Parameters .......................................................... 16
Verifying Setup .............................................................................. 16
Security ......................................................................................... 16
Backup the Instrument Settings ..................................................... 16
Chapter 3 – Pressure Decay-∆P ............................................................... 17
Timer Parameters .......................................................................... 18
Pressure Parameters..................................................................... 19
Test Parameters ............................................................................ 20
Chapter 4 – Pressure Decay-∆P/∆T .......................................................... 21
Timer Parameters .......................................................................... 22
Pressure Parameters..................................................................... 23
Test Parameters ............................................................................ 25
Chapter 5 – Pressure Decay-Leak Std...................................................... 27
Auto Setup Sequence .................................................................... 28
Timer Parameters .......................................................................... 30
Pressure Parameters..................................................................... 31
Test Parameters ............................................................................ 32
Program Calibration ....................................................................... 34
i www.cincinnati-test.com
S E N T I N E L I 2 8 I N D E X
iii www.cincinnati-test.com
S E N T I N E L I 2 8 I N D E X
vii www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
Chapter 1 – Introduction
This chapter explains the external pneumatic and electrical connections, and
introduces the simple menu structure and navigation of the instrument.
Thank you for purchasing the Sentinel I28TM precision leak test instrument. The Sentinel I28 is capable of
testing using a variety of pre-packaged test algorithms designed to get you testing as quickly and easily as
possible. Let’s get started!
Unpacking
Carefully remove the instrument from the shipping carton. Locate the AC power cord or the DC cable
supplied with the instrument. The test port should have a ColderTM quick connect test port with Self-Test
cap. Alternatively, you may install any suitable 1/4” NPTM fitting into the manifold test port. If this is not
a standard instrument, see the print packet that shipped with your instrument for the proper thread
specifications.
Please allow 4 inch (100 mm) clearance under the instrument for installation of test line and for
hand clearance.
Note: These recommendations are also clarified in Appendix F with detailed hole to
hole dimensions in detailed “full scale” reference to the clearance dimension.
Installation
The leak test instrument is a modular, flexible instrument designed to operate in a manufacturing, clean
room, or lab environment. To operate trouble-free, it requires:
• Clean, dry instrument air for test and pilot air (recommend 0.3 and 5.0-micron coalescing filter)
- ISO 8573-1, Class 2 filters are recommended, with class 3 being acceptable.
• Vacuum for vacuum test applications
• Supply power: 24 VDC or 88-264 VAC 50/60 Hz auto sensing/switching electrical (based on
configuration)
• I/O: Switch to select internal or external 24 VDC electrical power for the inputs and outputs
Because this is a precision instrument, it is preferable to locate this instrument at least 15 feet (5 meters)
away from high electromagnetic energy devices (induction heat treat equipment and welders) whenever
possible. In addition, plants having poor quality electrical power or ground systems should consider using
isolation transformers on the power drops.
1 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
Note: When supplying power, always be certain that there is nearly zero voltage
(< 1 V) difference between the neutral and earth ground wires connected to the
instrument.
DC Power Connection
2 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
The instrument is designed with a simple graphical icon-based interface. The Main Menu is
activated by pressing the Main Menu button.
The Arrow/Enter Selection buttons are used to navigate through the icons on the screen and
select the appropriate menus. The center arrow is the Enter button. It is used to select
parameters for editing and saving the edits once they are complete.
The Cancel/Back button is used go back to the previous menu. Multiple presses will
eventually exit you out of the menu structure entirely and place you on the Monitor Screen.
This button is also used to cancel out of any edits before saving.
The Information button is used to get a help screen that is associated with the parameter
corresponding to where the cursor is located.
Note: When a USB memory stick mounted in the USB port, holding this button down
until it beeps will take a screen shot of the screen and save it to the USB memory stick. Do
not remove the USB memory stick until the instrument tells you it is finished writing the
image.
3 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
Hint: Press the “+” button for 2 seconds to prompt a screen to display in where the value
of a program may be selected using the arrow buttons. The up and down arrows
will increment and decrement the ones digit. The right and left arrows will
increment and decrement the tens digit.
Note: The Program may also be changed using Digital I/O (see Chapter 21), using
EtherNet/IPTM (see Chapter 22), or using RS232 or TCP/IP Communication
Methods (see Chapter 24).
4 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
Menu Structure
Below is an overview of the menu structure for the instrument.
Note: Not all menus are available for all hardware configurations and Test Types.
Main
Note 1 Note 2
Menu
Cal Misc
IP Inputs USB
Misc
IP Outputs Misc
Autorun
Graph
HW Inputs
Version
HW Outputs
Security
IP Inputs
Note 3
Notes: Modes IP Outputs
1. See Chapter 30
2. See Chapter 31 Basic
Tooling
3. See Chapter 27
Viewing Security is enabled Advanced
in the Misc menu above Admin
5 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 – I N T R O D U C T I O N
6 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Chapter 2 – Setup
In this chapter, you are asked how you plan to use the instrument to conduct the test or
sequence of tests for the intended application. Based on the answers, you will be directed
to modify certain instrument settings.
Setting up the instrument to meet specific application requirements is most effectively accomplished by
answering a few questions. These questions will guide you through the initial stages of the setup and point
you to the appropriate chapters that detail specific instructions based on the answers.
It is highly recommended that each type of part being tested, whether they differ in volume or construction,
have their own set of parameters defined in the instrument using individual Programs. Both Pressure decay
leak testing and Differential Pressure decay leak testing are a volume dependent function. Flow testing can
be setup to be volume independent; however, to optimize the testing, it is recommended that each type of
part have its own set of parameters. In certain cases, it is possible to group similar parts together to use the
same parameters. In this case we recommend you talk to a CTS applications specialist to assure proper
functionality.
7 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Question:
Yes: It is important that you understand the concepts of Parent Program Linking .
See Chapter 14 and then come back and continue the setup.
No: Proceed.
The instrument is capable of conducting tests on multiple parts in a set sequence (if configured). This may
be done to increase the efficiency of the operator. An operator may load up to four parts (if four test ports
were configured) and press start. The instrument will conduct leak tests on all four parts by multiplexing
through the different test ports, one at a time.
Question:
No: Proceed.
8 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Once you determine which test type is appropriate for your application go to
Program 1 (press the + or – buttons until you see a P01 in the lower right hand
corner of the screen.
Press the Main Menu button to view the icons in the Main Menu screen.
Select the Test type icon using the Arrow/Enter Selection buttons.
Select the appropriate test type using the Enter button to enable editing and then the up
and down arrows to view the options. When you see the desired selection press the
Enter button to select.
To simplify the instructions in this manual, the steps above are reduced to something similar to the
following: The Test Type menu is located in Main Menu > Program Config > Tst type icon.
9 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Pressure Verify Measures the Pressure at the isolated test port (no fill). Determined
from the pressure at the end of the Test segment timer. The result is
Chapter 7 presented in units of pressure.
Measures the Mass Flow needed to maintain the part at a set pressure.
Mass Flow
Typically used in applications where flow is expected. The result is
Chapter 8 presented in units of flow.
Measures the Mass Flow needed to maintain the part at a set pressure.
Mass Flow-Leak Std
Typically used in low leak applications. The result is presented in units of
Chapter 9 flow.
Measures the peak pressure before a Pressure Loss (ΔP) Event.
Ramp to ΔP Event Determines the maximum pressure before the pressure loss limit is
Chapter 10 exceeded as well as records the duration of time before the event. The
results are presented in units of pressure and time.
Measures the pressure at an Input Event. During test pressure ramp,
Ramp to Input Event monitors for trigger event on a digital input, then determines the
Chapter 11 pressure as well as records the duration of time before the event. The
results are presented in units of pressure and time.
Primarily used for testing the cracking point of a check valve. Measures
Ramp to Flow Event the point at which the flow crosses a threshold. Determines the
maximum pressure before the Flow Event limit is exceeded as well as
Chapter 12 records the duration of time before the event. The results are presented
in units of pressure and time.
10 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
This test type is used for conducting a pressure decay test measuring a
Differential Pressure (DP) pressure loss over time utilizing a Differential Pressure (DP) transducer.
Decay-ΔP The result of this test is the pressure loss (or gain for a vacuum test)
Chapter 17 measured over a fixed period of time, presented in units of pressure.
This test type is used for conducting a Diff Decay-Leak rate test
Diff Decay-Leak Rate
measuring leak rate by utilizing a Differential Pressure (DP) transducer.
Chapter 18 The result of this test is the leak presented in units of flow.
Differential Pressure (DP) This test is used for conducting a Differential Pressure (DP) Decay Test
Decay-Leak Std with Leak Standard. Pressure loss is measured utilizing a Differential
Pressure (DP) transducer. The pressure loss is correlated to a leak rate
Chapter 19
using a leak standard. This test requires a two-cycle calibration routine
to correlate the pressure loss to a flow rate. The result of this test is
presented in units of flow.
11 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Before we begin setting the parameters of the program we need to determine some of the capabilities and
features required for the application.
Tooling Control
The instrument is capable of controlling tooling to possibly eliminate the need for a PLC or other computer
that would control cylinders and sealing.
Question:
Are you planning to use the Sentinel I28 to control any sealing operations?
Yes: It is important that you understand the concepts and safety requirements of
Tooling Control. See Chapter 20 and then come back and continue the setup.
No: Proceed.
Question:
Are you planning to use the Sentinel I28 to control any ancillary devices using
discrete I/O or EtherNet/IPTM, or communicate with a PLC?
No: Proceed.
Channel Configuration
Setting the Units of Measure
The instrument has the capability to utilize different units of measure for each configured program. In
order to keep things simple and user friendly if you always use the same units of measure, you can set the
units in one place and have them apply instrument wide. This is done in the Channel Configuration menu.
Question:
Are you planning to use the same units of measure for every program?
Yes: Press the Main Menu button, select the Channel Config icon, then select the
Units icon. Change the Unit/Prec Define parameter to “Channel”. Set the
12 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
desired units of measure on that same screen. These units will apply to
every program.
No: Press the Main Menu button, select the Channel Config icon, then select the
Units icon. Change the Unit/Prec Define parameter to “Program”. The Units
icon will now appear under the Program Config menu. When a program is
created, the units for the program will use the units under the Channel Config
menu as the default settings but the units can now be changed per program.
Note: The units set on the Channel Config screen are also the units used for
Self-Test, Auto Setup and transducer Set/Span routines.
Note: The precision displayed for each unit may also be set on this screen.
13 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Question:
Are you planning to use the same leak standard for every program that requires a
leak standard?
Yes: Press the Main Menu button, select the Channel Config icon, then select the
Leak/Cal icon. Change the Leak Std/Cal Define parameter to “Channel”. Set
the leak standard value and the pressure at which it was calibrated on that
same screen. This leak standard value will apply to every program that
utilizes a leak standard for calibration.
No: Press the Main Menu button. Select the Channel Config icon. From the
Channel Config menu select the Leak/Cal icon. Change the Leak Std/Cal
Define parameter to “Program”. The leak standard parameters will now
appear on the TST:*** screen under the Program Config menu. When a
program is created, the units for the program will use the leak standard value
under the Channel Config menu as the default settings but now can be
changed per program.
Note: Every time the leak standard value changes due to a calibration or when a new leak
standard is used this value needs to be modified to reflect the new value.
Note: The parameter that determines where the physical leak standard is located (either
inside the instrument or connected to a port on the outside) is also located in the Leak/Cal
menu or the TST menu. However, to view this parameter you must be in the Display User
Level. To change the Display User Level, see Chapter 28 – Features.
Remember: Press the information button with the cursor on that parameter if
you want to know the definition of the options.
14 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
Program Configuration
Now that you have created a program, configured the tooling and associated I/O, and set the channel level
parameters, it is time to set the regulators to the correct test pressure.
Pressure Select
To select the proper source for the test, we need to inform the instrument which regulator to use for this
program. This is located in Main Menu > Program Config > Misc icon. The parameter called Pressure
Select will need to be set to match the proper source. Find the regulator with the proper range. Locate the
label that has the Source Number on it and select that source for the Pressure Select parameter. There are
many possible options for this setting, depending on the hardware configuration. Options may include
Pressure, Vacuum, Venturi, Tank, Reference, Flow, Internal Leak Std, Downstream Flow, or Sequential
Test Ports. Instruments with only one pressure source will not allow editing of the Pressure Source
parameter.
Tank Source
If the Pressure Select is configured as a Tank Source, then an additional sub-parameter “Shut Off Delay” is
displayed. The Shut Off Delay is the user configured timer, defining how long to keep the Tank Source
valve active after usage. During typical test operation: after the Fill valve has de-energized the Shut Off
Delay timer will activate, and once expired the Tank Source valve will de-energize.
Question:
Yes: From the same Program Config - Misc menu select the Test Port Number you
want to conduct the test on. The port numbers will count upwards from left
to right.
No: Proceed.
Question:
Does the pressure source you are using for this program have an electronic
regulator?
Yes: The electronic regulator has been setup and calibrated at the factory and is
ready to use. For calibrating the electronic regulator, see Chapter 29 –
Instrument Calibration.
15 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 – S E T U P
will need to be blocked which will allow the instrument to hold pressure.
Standard units are supplied with a ColderTM quick connect fitting with an
attached Self-Test cap. If this instrument is not a standard setup this may
be done with a plug put in the test port. For proper plug thread
specifications, consult the print set shipped with your instrument.
Verifying Setup
Once you have everything programmed, run a repeatability study based on your company’s quality standards
to assure you are getting the results you desire.
Security
Now that the instrument is setup the way you want, make sure to lock down the parameters that you don’t
want changed until a key or password is used. See Chapter 27 – Security.
16 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 – P R E S S U R E D E C A Y - ∆ P
The basic principle of operation of a pressure decay leak test instrument is to fill the test part to a specified
target test pressure, isolate the test part from the pressure or vacuum source, allow the pressure to stabilize,
and then measure the pressure loss due to a leak over a defined time.
The Charts below give an overview of the parameters used to set up a Pressure Decay-∆P Test. The Tables
that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Exhaust
Stabilize
Pre-fill
Test
Fill
Maximum Pressure
Minimum Pressure
Time
17 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 – P R E S S U R E D E C A Y - ∆ P
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also Basic
Fill
be used as time to stabilize part pressure with additional air. Advanced, Admin
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
18 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 – P R E S S U R E D E C A Y - ∆ P
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
19 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 – P R E S S U R E D E C A Y - ∆ P
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:PLO icon.
20 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
The basic principle of operation of a pressure decay leak test instrument is to fill the test part to a specified
target test pressure, isolate the test part from the source air and allow the pressure to stabilize, and then
measure the pressure loss due to a leak over a defined time.
The Charts below give an overview of the parameters used to set up a Pressure Decay-∆P/∆T Test. The
Tables that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Exhaust
Stabilize
Pre-fill
Test
Fill
Maximum Pressure
Minimum Pressure
Time
21 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also Basic
Fill
be used as time to stabilize part pressure with additional air. Advanced, Admin
22 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
23 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
24 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:DPT icon.
25 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 4 – P R E S S U R E D E C A Y - ∆ P / ∆ T
26 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
This test type utilizes a two-point calibration sequence with a non-leaking master part to
record the zero leak pressure loss value and the additional pressure loss value due to the
flow rate of the leak standard on the second calibration cycle.
The basic principle of operation of a pressure decay leak test instrument is to fill the test part to a specified
target test pressure, isolate the test part from the source air and allow the pressure to stabilize, and then
measure the pressure loss due to a leak over a defined time. The leak test instrument translates the pressure
loss value measured over the fixed test time to a leak or flow rate.
The charts in this chapter give an overview of the parameters used to set up a Pressure Decay Test and
correlate the results to a flow rate using a leak standard. The Tables that follow give detailed descriptions
of each parameter and also document the Display User Level associated with each parameter.
= Leak standard
flow rate
Master Part + Leak Standard Pressure Loss
= Negative leak
Master Part Pressure Loss rate (because it
is less than zero)
27 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Note: If you want to set up the program parameters manually or don’t want to have to
reconfigure the inputs and outputs after running Auto Setup, then proceed to Timer
Parameters.
The Auto Setup routine is a unique patented process developed by Cincinnati Test Systems for the Sentinel
I28 instrument that makes the initial setup of the instrument to a new test part quick, easy and efficient.
The process will determine the best time sequence to test a part within the limits of a desired maximum
cycle time. If the desired maximum cycle time is set too short, the final test will not produce repeatable
results. Any leak test needs a certain amount of cycle time to achieve sufficient resolution and repeatability
that are dependent on the part volume, leak rate, test pressure, and part characteristics. The Auto Setup
routine will set the best possible test cycle within the maximum cycle time specified for the test.
28 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Upon successful completion of the Auto Setup routine there is no need to continue the setup
parameters in the rest of this chapter unless you want to adjust parameters.
29 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also Basic
Fill
be used as time to stabilize part pressure with additional air. Advanced, Admin
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
30 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
Pressure Decay
Test
Stabilize
Pre-fill
Exhau
Test
Fill
st
Maximum
Pressure
}
Pressure
Target Press
Window
Minimum
Pressure
Time
31 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:PLR icon.
Basic
Low Limit Rate Lower setpoint value used to evaluate test results.
Advanced, Admin
Basic
High Limit Rate Upper setpoint value used to evaluate test results.
Advanced, Admin
Test results are compared to two setpoints for pass or
fail status. There are three areas for evaluation of results:
Test Evaluation Advanced, Admin
above high limit, between limits, and below low limit.
See Table below for codes.
32 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Quik Test Enable Activates Quik Test. See Appendix B. Advanced, Admin
Options include 10%, 25%, 50%, 75%, and 90%.
Quik Test Timer Advanced, Admin
See Appendix B.
Quik Test LL
See Appendix B. Advanced, Admin
Band
Quik Test HL
See Appendix B. Advanced, Admin
Band
Activates Environmental Drift Correction.
EDC Enable Advanced, Admin
See Appendix C.
Options include 10%, 25%, 50%, 75%, and 90%.
EDC Percentage Advanced, Admin
See Appendix C.
EDC Quantity See Appendix C. Advanced, Admin
Note: If the instrument is configured multiple internal leak standards the user must select
which leak standard is to be used in the Program. See Selecting the Proper Internal Leak
Standard in Chapter 28 – Features.
33 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Program Calibration
In order to convert the pressure loss measured by the instrument to a leak (flow) rate, the instrument uses
a leak standard and needs to run the “Program Cal” procedure. This procedure requires at least one
known non-leaking part referred to as a “master part“. This procedure tests a non-leaking master part
connected to the instrument using the timers and pressures established for the program. The procedure
automatically tests the non-leaking master part twice with the Relax timer delay between tests. Within
each program that uses a leak standard, Program Cal can be configured to use one of four methods. See
Setting the Cal Method and Leak Standard Location in Chapter 28 – Features.
Remember: The Leak/Cal menu has a parameter called Leak Std/Cal Define
that determines the location for the leak standard settings.
When set to “Channel”, leak standard settings are located in the Channel Config menu.
When set to “Program”, leak standard settings are located in the Program menu.
Determine how you plan to use your leak test instrument. Review Setting the Leak Standard
Values section on page 14. It is critical to make sure the leak standard values are set correctly
for proper use. To view the Cal Method parameter, you must be in the Advanced or Admin
Display User Level. To change the Display User Level, see Chapter 28– Features.
Measured
Master+Leak
Δpressure (Converted a Leak Rate )
Loss
Measured
Master Part
Loss
Rate
Corr Leak
High Limit
Std Flow
Rate
Low limit
Zero Leak
34 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Performance Factor
Upon the successful completion of the Program Cal routine, the display will show the calculated
Performance Factor for the calibration. The Performance Factor is an estimate of the quality of the
calibration. It combines the ratio of the Hi Limit Leak result to the non-leaking master part result, the test
pressure, and difference between the Hi Lim result and the non-leaking master part result to scale its
anticipated performance. This value ranges from 0 to 100. It is generally desirable to have a Performance
Factor of 35 to 100. The actual acceptable Performance Factor can vary depending on the desired Gage
R&R performance of the test. Here is the Performance Factor equation:
Performance Factor = Cal Ratio x Pressure Loss Penalty x Time Penalty x 100
The pressure loss penalty is greater for smaller pressure loss values. The time penalty is greater for shorter
Test cycle timer values. Generally, longer stabilize and test cycle timer settings will produce higher
Performance Factors and improved Gage R&R performance.
Note: The instrument will require a Program Cal Sequence if any parameters that affect
the calibration are modified.
If the procedure does not meet these conditions, error messages are displayed at the completion of tests.
35 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
If the Program Cal routine is successfully completed, “PROGRAM CALIBRATION PASSED” will be
displayed. If there are any problems during the Program Cal sequence an error will be displayed. If there
is an error, see Appendix A - Message and Error Codes.
36 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
Calibration Parameters
The Calibration parameters menu is located in the Main Menu > Program Config > CAL:PLR
icon. These calibration parameters are values that are measured or calculated as a result of the
Program Cal routine for a Pressure Decay-Leak Std test.
37 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 5 – P R E S S U R E D E C A Y - L E A K S T D
38 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 6 – O C C L U S I O N
Chapter 6 – Occlusion
This chapter explains the theory and parameters for conducting an Occlusion
Test. The result of this test is the measured back pressure at the expiration of the
Test sequence.
The Occlusion Test is a back pressure test. The part is pressurized throughout the test from a fixed pressure
regulator setting. At the end of the Test timer the actual test pressure is compared to the Low Limit Pressure
and High Limit Pressure settings. Low Limit Pressure indicates high flow or minimum blockage or low
back pressure. High Limit Pressure indicates low flow or maximum blockage or high back pressure.
The Chart below give an overview of the parameters used to set up an Occlusion Test. The Tables that
follow give detailed descriptions of each parameter and also document the Display User Level associated
with each parameter.
Occlusion Test
Maximum Pressure
Hi Limit Pressure
Target Pressure
39 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 6 – O C C L U S I O N
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
40 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 6 – O C C L U S I O N
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:OCC icon.
41 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 6 – O C C L U S I O N
42 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 7 – P R E S S U R E V E R I F Y
The Pressure Verify test is intended to test parts that generate vacuum or pressure, or to verify that
the part was pre-charged to the correct pressure, or to verify that the part was evacuated or filled by
an external source.
The principle of operation is to close the isolation valve on the internal manifold to isolate the part.
The pressure transducer then reads the vacuum or pressure generated by the part or that is already in
the part. At the end of the test timer, the test pressure is compared to the low limit pressure and
high limit pressure settings. The Test Result is recorded with an overall ACCEPT/REJECT
If the Pressure Verify test was performed with the Self-Test cap on the test port, the only pressure
measured will be the small pressure that is trapped by the isolation valve as it actuates to close.
The Tables that follow give detailed descriptions of each parameter and also document the Display User
Level associated with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
43 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 7 – P R E S S U R E V E R I F Y
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
44 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 7 – P R E S S U R E V E R I F Y
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:PVF icon.
45 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 7 – P R E S S U R E V E R I F Y
46 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
This test type utilizes a single calibration cycle with a non-leaking master part to set the
zero flow value. It is important to note that this calibration process creates a linear offset
of all the flow readings based on this single calibration point.
The basic principle of operation of a mass flow test leak test instrument is to fill the test part to a specified
target test pressure and allow the pressure regulator to maintain the target pressure in the part. The
instrument will measure the flow of air required to keep the part at the target pressure.
The Charts below give an overview of the parameters used to set up a Mass Flow Test. The Tables that
follow give detailed descriptions of each parameter and also document the Display User Level associated
with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
47 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also Basic
Fill
be used as time to stabilize part pressure with additional air. Advanced, Admin
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
48 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
49 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:FLW icon.
50 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
Program Calibration
In some applications, particularly low leak applications, it is advantageous to create a flow offset based on
testing a non-leaking master part. To enable flow offset, go to Main Menu > Program Config >
TST:FLW icon. Change the Enable Calibration parameter to “Yes”.
Note: The instrument will require a Program Cal Sequence if any parameters that affect
the calibration are modified.
One condition must be met during the Program Calibration procedure for the instrument to accept and
store the calibration result. The condition is:
The Master (part) Flow must be between the Min Master Flow and Max Master Flow limits.
If the procedure does not meet this condition, an error message is displayed at the completion of test.
If the Program Cal routine is successfully completed, “PROGRAM CALIBRATION PASSED” will be
displayed. If there are any problems during the Program Cal sequence an error will be displayed. See
Appendix A - Messages & Error Codes.
Calibration Parameters
The Calibration parameters menu is located in Main Menu > Program Config > CAL:FLW icon.
These calibration parameters are values that are measured or calculated as a result of the Program
Cal routine for a Mass Flow test.
51 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 8 – M A S S F L O W
If the parameter is set to “Enabled: Prefill-Test” then the transducer is active during the Prefill through Test
timer. If the parameter is set to “Enabled: Fill-Test” then the transducer is active during the Fill to test timer.
If the parameter is set to “Disabled”, the transducer is bypassed during the Prefill and Fill timer and active
only during the Test timer.
Note: It may be advantageous to use direct flow when short test cycles are needed with
parts that are expected to flow a certain amount. With direct flow, the flow transducer is
inserted into the test circuit earlier and gives the flow sensor the entire cycle to stabilize
before making the final measurement at the end of the test segment.
52 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
The basic principle of operation of a mass flow test leak test instrument is to fill the test part to a specified
target test pressure and allow the pressure regulator to maintain the target pressure in the part. The
instrument will measure the flow of air required to keep the part at the target pressure.
This test type utilizes a two-point calibration sequence with a non-leaking master part to
set the zero flow value and the flow value at the leak standard value. It is important to note
that this calibration process creates a linear offset of the flow readings based on these two
points.
The charts in this chapter give an overview of the parameters used to set up a Mass Flow Test. The Tables
that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
53 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also Basic
Fill
be used as time to stabilize part pressure with additional air. Advanced, Admin
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
54 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
55 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:FLC icon.
The minimum flow allowed for the first test during Basic
Min Master Flow
the Program Calibration sequence. Advanced, Admin
The maximum flow allowed for the first test during Basic
Max Master Flow
the Program Calibration sequence. Advanced, Admin
Note: If the instrument is configured with multiple internal leak standards, the user must
select which leak standard is to be used in the Program. See Selecting the Proper Internal
Leak Standard in Chapter 28 – Features.
Program Calibration
In order to scale the flow transducer to the “zero flow” reading measured by testing a non-leaking part
and the “reject” reading measured by testing the non-leaking part with a leak standard incorporated in the
circuit, the instrument needs to run the “Program Cal procedure”. This procedure requires at least one
known non-leaking part referred to as a “master part“. This procedure tests a non-leaking master part in
the seal fixture using the timers and pressures established for the program. The procedure automatically
tests the non-leaking master part twice with the Relax timer delay between tests. Within each program
that uses a leak standard, Program Cal can be configured to use one of four methods. See Setting the
Cal Method and Leak Standard Location in Chapter 28 – Features.
Remember: The Leak/Cal menu has a parameter called Leak Std/Cal Define
that determines the location for the leak standard settings.
When set to “Channel”, leak standard settings are located in the Channel Config menu.
When set to “Program”, leak standard settings are located in the Program menu.
Determine how you plan to use your leak test instrument. Review Setting the Leak Standard
Values section on page 14. It is critical to make sure the leak standard values are set correctly
for proper use. To view the Cal Method parameter, you must be in the Advanced or Admin
Display User Level. To change the Display User Level, see Chapter 28– Features.
57 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
part with a known calibrated leak standard included in the test. The flow value result for this test is saved
as the Master+Leak Flow.
Note: The instrument will require a Program Cal Sequence if any parameters that affect
the calibration are modified.
If the procedure does not meet these conditions, error messages are displayed at the completion of tests.
If the Program Cal routine is successfully completed, “PROGRAM CALIBRATION PASSED” will be
displayed. If there are any problems during the Program Cal sequence an error will be displayed. See
Appendix A – Message and Error Codes.
58 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
Calibration Parameters
The Calibration parameters menu is located in the Main Menu > Program Config > CAL:FLC
icon. These calibration parameters are values that are measured or calculated as a result of the
Program Cal routine for a Mass Flow-Leak Std test.
If the parameter is set to “Enabled: Prefill-Test” then the transducer is active during the Prefill to test timer.
If the parameter is set to “Enabled: Fill-Test” then the transducer is active during the Fill to test timer.
If the parameter is set to “Disabled”, the transducer is bypassed during the Prefill and Fill timer and active
only during the Test timer.
Note: It may be advantageous to use direct flow when short test cycles are needed with
parts that are expected to flow a certain amount. With direct flow, the flow transducer is
inserted into the test circuit earlier and gives the flow sensor the entire cycle to stabilize
before making the final measurement at the end of the test segment.
59 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 9 – M A S S F L O W - L E A K S T D
60 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 0 – R A M P T O Δ P E V E N T
A test type to evaluate parts that ‘crack’ or drastically change state under pressure that can be immediately
detected by monitoring for a sudden significant change in system pressure. The system pressure is typically
increased at a consistent rate (ramp) over a defined period of time. The pressure ramp can be controlled by
the instrument with an electronic regulator, or by a mechanical regulator through a fixed restriction.
The Chart below gives an overview of the parameters used to set up a Ramp to Event Test. The Tables
that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Note: To change the Display User Level, see Chapter 28 – Features.
61 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 0 – R A M P T O Δ P E V E N T
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
62 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 0 – R A M P T O Δ P E V E N T
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:RTE icon.
63 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 0 – R A M P T O Δ P E V E N T
64 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 1 – R A M P T O I N P U T E V E N T
This test type is used to evaluate parts that change state under pressure, where the change can be immediately
detected by monitoring a pressure switch, flow switch, or other device with a digital input. The system
pressure is either increased or decreased (ramp) over a defined period of time. The pressure ramp can be
controlled by the instrument via an electronic regulator, or by a mechanical regulator through a fixed
restriction. The digital input on the instrument which is set to Event Trigger and monitors for a rising or
falling edge or a level change from high to low or low to high. The Event Trigger can be ignored if the
duration is less than the Level Time setting.
The Chart below gives an overview of the parameters used to set up a Ramp to Input Event test. The
Tables that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
65 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 1 – R A M P T O I N P U T E V E N T
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
66 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 1 – R A M P T O I N P U T E V E N T
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:RTI icon.
67 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 1 – R A M P T O I N P U T E V E N T
68 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 2 – R A M P T O F L O W E V E N T
This test type is used to evaluate parts that change flow characteristics under pressure, where the change
can be immediately detected by monitoring the flow on the outlet side of the device. The system pressure
is either increased or decreased at a consistent rate (ramp) over a defined period of time. The pressure ramp
can be controlled by the instrument via an electronic regulator, or by a mechanical regulator through a fixed
restriction. The outlet side of the device is connected to the flow sensor port on the instrument and allows
the instrument to measure downstream flow through the device as the pressure ramps up.
The Chart below gives an overview of the parameters used to set up a Ramp to Flow Event test. The
Tables that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
69 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 2 – R A M P T O F L O W E V E N T
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
70 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 2 – R A M P T O F L O W E V E N T
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:RTF icon.
71 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 2 – R A M P T O F L O W E V E N T
72 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 3 – R A M P T O P R O O F
The system pressure is either increased or decreased at a consistent rate (ramp) over a period of time, and
then held at pressure for an additional period of time. The pressure ramp can be controlled by the
instrument with an electronic regulator or by a mechanical regulator through a fixed restriction. Parts are
evaluated based on whether or not pressure was achieved and maintained for the test time period.
The Charts below give an overview of the parameters used to set up a Proof Test. The Tables that follow
give detailed descriptions of each parameter and also document the Display User Level associated with each
parameter.
Maximum Pressure
Target Pressure
Proof Pressure
73 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 3 – R A M P T O P R O O F
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
74 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 3 – R A M P T O P R O O F
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST:PRF icon.
75 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 3 – R A M P T O P R O O F
76 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
The applications for program linking fall into three basic categories: conducting multiple tests on one part,
one part with multiple chambers, or multiple unique parts.
This test type facilitates the linking of individual programs into a testing sequence and provides the overall
control of the tooling outputs and inputs, start and stop inputs, and program result outputs for the test
sequence. The following functions within each individual test program are ignored and controlled by the
Parent Program.
77 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
LINK 2
Child Part Program Child Part Program
Fine Leak Impreg Level Test
Side 2 P3 Side 1 P4
Accept Either Reject
78 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
To program this test sequence, the user goes to Program Config and defines the test type for the Program
1 to be “Parent Program Linking” The test sequence example above requires two links.
The first link screen identifies the program where the testing in the Parent Program will begin and how
additional tests are initiated or linked to the first program based on the first test result. The link defines
the programs that will run next based on the Evaluation Condition.
The second link required by the example above targets Program 3. If Part 3 is Reject, then Program 5 will
run. If Program 3 is Accepted, the test sequence stops with a Program Accept light and a Program
Accept.
79 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
Note: If you are using Parent Program Linking because you have part with multiple chambers
and the instrument is configured with multiple test ports, remember to select the appropriate test
port. This is described in the section titled Selecting the Test Port in Chapter 2 – Setup.
80 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Sequential Linking
The sequential linking test type is designed to run multiple unique parts conducting the tests by
multiplexing (often with multiple sequential test ports). This allows the operator to load and unload the
parts more efficiently. The test results are reported individually for each program. The functionality and
setup of this test type is very similar to the Parent Program Linking test type. The only differences are
that:
a) Each of the linked program results is stored individually.
b) Each result has no bearing on the accept/reject criteria of past or future linked programs
unless they are conditionally linked.
c) The cycle counters increment with each linked program completed.
d) The tooling control uses the defined motions in each of the linked programs (there is no
"Parental tooling control").
This is because this test type is designed to run multiple individual parts.
81 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 4 – P R O G R A M L I N K I N G
82 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 5 – P E R C E N T V O L U M E
The Percent Volume test type is used to test for gross leakage by measuring the escape of pressure or
vacuum from a reference part filled with a metered volume of pressure or vacuum. Percent Volume test
type is useful for parts with high elasticity, which readily change volume, by using a pre-defined volume to
limit their maximum expansion during test. In addition, parts with internal openings connecting multiple
volumes can be verified to be open or closed.
This test type is used to determine the volume of a test part compared to the volume of a reference part.
The reference volume is initially filled and isolated from the supply. The test part remains isolated. The
reference volume pressure is measured at this point. The reference volume is then released into the test
part and allowed to stabilize. The pressure of the test part is measured at this point. The two pressure
measurements are then used to calculate the ratio of the test part volume to the reference volume.
The Charts below give an overview of the parameters used to set up a Percent Volume Test. The Tables
that follow give detailed descriptions of each parameter and also document the Display User Level
associated with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
Test
Maximum Pressure
Pressure
Time
83 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 5 – P E R C E N T V O L U M E
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
84 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 5 – P E R C E N T V O L U M E
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
Exhaust
Time
85 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 5 – P E R C E N T V O L U M E
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST: %VR icon.
86 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
This test type utilizes a two-point calibration sequence with a non-leaking master part to
record the zero leak pressure loss value and the additional pressure loss value due to the
flow rate of the leak standard on the second calibration cycle.
The Percent Volume, Pressure Decay – Leak Std test type is used to test for both severe and fine leaks by
measuring the escape of pressure or vacuum from a reference part filled with a metered volume of pressure
or vacuum. This test type is useful for parts with high elasticity, which readily change volume, by using a
pre-defined volume to limit their maximum expansion during test. In addition, parts with internal openings
connecting multiple volumes can be verified to be open or closed. The Percent Volume portion of this test
type is used to find severe leaks, while the pressure decay – leak std portion is used to find fine leaks.
The basic principle of a Percent Volume, Pressure Decay Leak Test is to perform a percent volume test
followed by a calibrated pressure decay leak test. Details for the basic operation of a percent volume test
can be found in Chapter 15 – Percent Volume. Details for the basic operation of a pressure decay leak test,
including how to perform the required calibration sequence, can be found in Chapter 5 – Pressure Decay-
Leak std.
The charts in this chapter give an overview of the parameters used to set up a Percent Volume, Pressure
Decay Leak Test and correlate the results of the Pressure Decay Leak Test portion to a flow rate using a
leak standard. The Tables that follow give detailed descriptions of each parameter and also document the
Display User Level associated with each parameter.
Note: To change the Display User Level, see Chapter 28– Features.
87 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
88 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Exhaust
Stabilize
Stab Ref
Test 2
Fill Ref
Test 1
Pressure
Maximum Pressure
Minimum Pressure
Time
89 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
90 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Basic
Low Limit Rate Lower setpoint value used to evaluate test results.
Advanced, Admin
Basic
High Limit Rate Upper setpoint value used to evaluate test results.
Advanced, Admin
Test results are compared to two setpoints for pass or
fail status. There are three areas for evaluation of results:
Test Evaluation Advanced, Admin
above high limit, between limits, and below low limit.
See Table below for codes.
Minimum pressure loss allowed during the first test in
Basic
Min Master Loss the Program Cal routine. Prevents the acceptance of a
Advanced, Admin
calibration of a blocked part or test line.
Maximum pressure loss allowed during the first test in
Basic
Max Master Loss the Program Cal routine. Prevents the acceptance of a
Advanced, Admin
calibration with a leaking part or fixture leaks.
Minimum pressure loss allowed during the second test
Min Mstr+Leak Basic
in the Program Cal routine. Prevents the acceptance of
Loss Advanced, Admin
a calibration with a blocked part or test line.
Maximum pressure loss allowed during the second test
in the Program Cal routine. Prevents the acceptance of a
Max Mstr+Leak Basic
calibration with excessive pressure loss due to part or
Loss Advanced, Admin
fixture leaks. Set slightly higher than max loss during
Program Cal.
Minimum pressure loss allowed that represents the
Basic
Min Leak Loss pressure loss due to the Leak Standard used to calibrate
Advanced, Admin
the instrument.
91 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Note: If the instrument is configured multiple internal leak standards the user must select
which leak standard is to be used in the Program. See Selecting the Proper Internal Leak
Standard in Chapter 28 – Features.
92 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
Measured
Master+Leak
Δpressure (Converted a Leak Rate )
Loss
Measured
Master Part
Loss
Rate
Corr Leak
High Limit
Std Flow
Rate
Low limit
Zero Leak
93 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 6 – P E R C E N T V O L U M E , P R E S S U R E
D E C A Y - L E A K S T D
94 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
How it works
In order to detect leakage in a part, the change in pressure due to temperature and part elasticity must be
allowed to settle before taking any pressure readings.
With differential pressure, a reference part volume (should be identical to the nominal test part volume) is
pressurized simultaneously with the test part.
Leaks are determined by detecting the rate of pressure loss between the two parts.
95 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
Test Setup
The test part and the reference volume are simultaneously pressurized to a preset pressure. The air in the
system is then allowed to stabilize, with the supply valves all closed. After the stabilization time, the
Differential Pressure Transducer is automatically zeroed.
During test, the pressure change in the test piece is compared to the pressure change in the reference
volume, using the Differential Pressure Transducer. If the test piece is leaking, the difference will increase
and be measured; an alarm limit may be set for a pass/fail decision.
The charts in this chapter give an overview of the parameters used to set up a Pressure Differential Test
and correlate the results to a flow rate using a leak standard. The Tables that follow give detailed descriptions
of each parameter and also document the Display User Level associated with each parameter.
96 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
97 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to enable part to reach the Target Pressure. It may also be Basic
Fill
used as time to stabilize part pressure with additional air. Advanced, Admin
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
At the end of this timer, the instrument will read the pressure on
Basic
Test the differential pressure transducer. This pressure is due to the
Advanced, Admin
backpressure created in the pneumatic circuit and part.
Time to relieve or vent part pressure before signaling the end of Basic
Exhaust
test. Need time to prevent blowing out debris or fixture seals. Advanced, Admin
When tooling control is specified, there can be one retract timers
Tooling
for one tooling motions. This is a not to exceed timer when Advanced, Admin
Retract
feedback is being utilized.
98 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
User Display
Pressure Description Mode
Minimum test pressure that must be met within the Pre- Basic
Minimum Pressure fill timer and maintained during the Fill and Stabilization Advanced,
timers. This is an early indication of a major leak Admin
Basic
Target test pressure. Also used as a setpoint for the
Target Pressure Advanced,
Electronic Regulator.
Admin
The value that must not be exceeded at any time to
Basic
complete a successful test. If the pressure goes above the
Maximum Pressure Advanced,
Maximum Test Pressure, the testing cycle (during Fill and
Admin
Stabilization) will end as an Over Pressure Malfunction.
Minimum charge pressure that must be met within the
Charge timer and maintained before the Fill timer. This
is an early indication of a major leak. If the pressure is
Basic
measured below the Charge Min Press at end of Charge
Charge Min Press Advanced,
segment, the testing cycle will end as a Below Charge
Admin
Malfunction.
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
Target Charge pressure. Also used as a setpoint for the
Basic
Electronic Regulator in Charge segment.
Charge Target Press Advanced,
This parameter is only available if the Prefill Method selected is
Admin
“Fixed Charge”
The value that must not be exceeded at end of Charge
segment. If the pressure is measured above the Charge
Basic
Max Press at end of Charge segment, the testing cycle will
Charge Max Press Advanced,
end as an Above Charge Malfunction.
Admin
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
99 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 7 – D I F F E R E N T I A L P R E S S U R E ( D P )
D E C A Y - Δ P
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST: DPD icon.
User Display
TST Parameter Description Mode
Basic
Low Limit Loss Lower set point value used to evaluate test results.
Advanced, Admin
Basic
High Limit Loss Upper set point value used to evaluate test results.
Advanced, Admin
Defines the method for calculating the part pressure
decay during test: Loss – charged part decreasing Basic
Decay Direction
pressure or evacuated part decreasing vacuum, Advanced, Admin
Gain – part pressure increasing from external forces.
Manual compensation value which is added to the
Basic
Loss Offset measured loss when calculating the final pressure loss for
Advanced, Admin
the test.
When enabled, Environmental Drift Correction (EDC)
monitors and continually corrects final Leak Rate
EDC Enabled calculations based on the moving average shift of test Advanced, Admin
results within a band about the Master Part calibration
curve.
Percentage (Plus and Minus) band defined by a
percentage of High Limit about the Master Part Curve.
EDC Percentage Advanced, Admin
EDC continually accumulates and calculates an offset
for the test results.
Number of test results utilized in calculating the EDC
EDC Quantity Advanced, Admin
factor.
100 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
How it works
In order to detect leakage in a part, the change in pressure due to temperature and part elasticity must be
allowed to settle before taking any pressure readings.
With differential pressure, a reference part volume (should be identical to the nominal test part volume) is
pressurized simultaneously with the test part.
The Leak Rate is determined from the measured differential loss between the test port and the
reference volume over time, using the provided known part volume. This test will NOT require
calibration, and a Leak Standard is NOT required to determine the final leak rate.
Where:
Q = Leak Rate
ΔP/Δt = Pressure Loss over time
Pstd = Standard Pressure (14.696 psig). This parameter is configured matching Standard
Pressure definition used for instrument and leak calculation
Vinst = Instrument Volume. Volume of the instrument’s isolated pneumatic circuit up to
the test port. This parameter can be located at Main Menu->Channel Config->Hardware.
Vpart = Part Volume.
101 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
Test Setup
The test part and the reference volume are simultaneously pressurized to a preset pressure. The air in the
system is then allowed to stabilize, with the supply valves all closed. After the stabilization time, the
Differential Pressure Transducer is automatically zeroed. Before the beginning of the test measurement, the
part and the reference volume are isolated from the supply valves for the time described in the Isolation
parameter.
During test, the pressure change in the test piece is compared to the pressure change in the reference
volume, using the Differential Pressure Transducer. If the test piece is leaking, the difference will increase
and be measured; an alarm limit may be set for a pass/fail decision.
The Tables that follow give detailed descriptions of each parameter and also document the Display User
Level associated with each parameter.
102 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
103 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
At the end of this timer, the instrument will read the pressure on
Basic
Test the pressure transducer. This pressure is due to the backpressure
Advanced, Admin
created in the pneumatic circuit and part.
Time to relieve or vent part pressure before signaling the end of Basic
Exhaust
test. Need time to prevent blowing out debris or fixture seals. Advanced, Admin
When tooling control is specified, there can be up to five retract
Tooling
timers for up to five tooling motions. This is a not to exceed timer Advanced, Admin
Retract 1
when feedback is being utilized.
Timer used during Program Cal routine as a delay between
calibration cycles to allow the master part to recover to repeatable Basic
Relax
virgin state. (Too short relax times result in decreasing pressure Advanced, Admin
losses/flow results in successive tests.)
104 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
User Display
Pressure Description Mode
Minimum test pressure that must be met within the Pre- Basic
Minimum Pressure fill timer and maintained during the Fill and Stabilization Advanced,
timers. This is an early indication of a major leak Admin
Basic
Target test pressure. Also used as a setpoint for the
Target Pressure Advanced,
Electronic Regulator.
Admin
The value that must not be exceeded at any time to
Basic
complete a successful test. If the pressure goes above the
Maximum Pressure Advanced,
Maximum Test Pressure, the testing cycle (during Fill and
Admin
Stabilization) will end as an Over Pressure Malfunction.
Minimum charge pressure that must be met within the
Charge timer and maintained before the Fill timers. This
is an early indication of a major leak. If the pressure is
Basic
measured below the Charge Min Press at end of Charge
Charge Min Press Advanced,
segment, the testing cycle will end as a Below Charge
Admin
Malfunction.
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
Target Charge pressure. Also used as a set point for the
Basic
Electronic Regulator in Charge segment.
Charge Target Press Advanced,
This parameter is only available if the Prefill Method selected is
Admin
“Fixed Charge”
The value that must not be exceeded at end of Charge
segment. If the pressure is measured above the Charge
Basic
Max Press at end of Charge segment, the testing cycle will
Charge Max Press Advanced,
end as an Above Charge Malfunction.
Admin
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
105 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 8 – D I F F D E C A Y - L E A K R A T E
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST: DDL icon.
User Display
TST Parameter Description Mode
Lower set point value used to evaluate test results of Basic
Low Limit Leak
parts. Advanced, Admin
Upper set point value used to evaluate test results of Basic
High Limit Leak
parts. Advanced, Admin
Defines the method for calculating the part pressure
decay during test: Loss – charged part decreasing
Decay Direction Admin
pressure, Gain – part increasing pressure from external
forces.
Volume of the part under test. This should include the
volume of all pressurized cavities of the part, seals, Basic
Part Volume
fittings and tubing leading up to the instrument’s test Advanced, Admin
port connection.
Manual compensation value which is added to the final
Leak Rate Offset Admin
calculated leak rate for the test.
When enabled, Environmental Drift Correction (EDC)
monitors and continually corrects final Leak Rate
EDC Enabled calculations based on the moving average shift of test Advanced, Admin
results within a band about the Master Part calibration
curve.
Percentage (Plus and Minus) band defined by a
percentage of High Limit about the Master Part Curve.
EDC Percentage Advanced, Admin
EDC continually accumulates and calculates an offset
for the test results.
Number of test results utilized in calculating the EDC
EDC Quantity Advanced, Admin
factor.
106 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
This test type utilizes a two-point calibration sequence with a non-leaking master part to
record the zero leak pressure loss value and the additional pressure loss value due to the
flow rate of the leak standard on the second calibration cycle.
How it works
In order to detect leakage in a part, the change in pressure due to temperature and part elasticity must be
allowed to settle before taking any pressure readings.
With differential pressure, a reference part volume (should be identical to the nominal test part volume) is
pressurized simultaneously with the test part.
Leaks are determined by detecting the rate of pressure loss between the two parts. The instrument can be
configured optionally with internal or external leak standards. External leak standards can be driven
manually. Internal leak standards are driven by valves (V4) and may be manually forced open during
a test to simulate a leak in the test circuit. Open Leak Std is located in Main Menu > Channel
Config > Leak/Cal icon.
107 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
108 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Test Setup
The test part and the reference volume are simultaneously pressurized to a preset pressure. The air in the
system is then allowed to stabilize, with the supply valves all closed. After the stabilization time, the
Differential Pressure Transducer is automatically zeroed.
During test, the pressure change in the test piece is compared to the pressure change in the reference
volume, using the Differential Pressure Transducer. If the test piece is leaking, the difference will increase
and be measured; an alarm limit may be set for a pass/fail decision
The charts in this chapter give an overview of the parameters used to set up a Pressure Differential Test
and correlate the results to a flow rate using a leak standard. The Tables that follow give detailed descriptions
of each parameter and also document the Display User Level associated with each parameter.
109 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
110 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Timer Parameters
The Timers menu is located in the Main Menu > Program Config > Timers icon.
Time to stabilize part pressure while isolated from the pressure Basic
Stabilize
regulator. This time directly affects the repeatability of the test. Advanced, Admin
At the end of this timer, the instrument will read the pressure on
Basic
Test the pressure transducer. This pressure is due to the backpressure
Advanced, Admin
created in the pneumatic circuit and part.
Time to relieve or vent part pressure before signaling the end of Basic
Exhaust
test. Need time to prevent blowing out debris or fixture seals. Advanced, Admin
When tooling control is specified, there can be up to five retract
Tooling
timers for up to five tooling motions. This is a not to exceed timer Advanced, Admin
Retract 1
when feedback is being utilized.
Timer used during Program Cal routine as a delay between
calibration cycles to allow the master part to recover to repeatable Basic
Relax
virgin state. (Too short relax times result in decreasing pressure Advanced, Admin
losses/flow results in successive tests.)
111 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Pressure Parameters
The Pressure menu is located in the Main Menu > Program Config > Pressure icon.
User Display
Pressure Description Mode
Minimum test pressure that must be met within the Pre- Basic
Minimum Pressure fill timer and maintained during the Fill and Stabilization Advanced,
timers. This is an early indication of a major leak Admin
Basic
Target test pressure. Also used as a setpoint for the
Target Pressure Advanced,
Electronic Regulator.
Admin
The value that must not be exceeded at any time to
Basic
complete a successful test. If the pressure goes above the
Maximum Pressure Advanced,
Maximum Test Pressure, the testing cycle (during Fill and
Admin
Stabilization) will end as an Over Pressure Malfunction.
Minimum charge pressure that must be met within the
Charge timer and maintained before the Fill timers. This
is an early indication of a major leak. If the pressure is
Basic
measured below the Charge Min Press at end of Charge
Charge Min Press Advanced,
segment, the testing cycle will end as a Below Charge
Admin
Malfunction.
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
Target Charge pressure. Also used as a setpoint for the
Basic
Electronic Regulator in Charge segment.
Charge Target Press Advanced,
This parameter is only available if the Prefill Method selected is
Admin
“Fixed Charge”
The value that must not be exceeded at end of Charge
segment. If the pressure is measured above the Charge
Basic
Max Press at end of Charge segment, the testing cycle will
Charge Max Press Advanced,
end as an Above Charge Malfunction.
Admin
This parameter is only available if the Prefill Method selected is
“Fixed Charge”
112 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Test Parameters
The Test parameters menu is located in the Main Menu > Program Config > TST: DPL icon.
User Display
TST Parameter Description Mode
Lower set point value used to evaluate test results of Basic
Low Limit Leak
parts. Advanced, Admin
Upper set point value used to evaluate test results of Basic
High Limit Leak
parts. Advanced, Admin
Defines the method for calculating the part pressure
decay during test: Loss – charged part decreasing Basic
Decay Direction
pressure, Gain – part increasing pressure from external Advanced, Admin
forces.
Minimum pressure loss allowable for the master part
Min Mstr+Leak Basic
with Leak Standard during the calibration cycle. Use to
Loss Advanced, Admin
prevent calibration to leaking master parts or fixtures.
Maximum pressure loss allowable for the master part
Max Mstr+Leak Basic
with Leak Standard during the calibration cycle. Use to
Loss Advanced, Admin
prevent calibration to leaking master parts or fixtures.
Minimum pressure loss allowed that represents the
Basic
Min Leak Loss pressure loss due to the Leak Standard used to calibrate
Advanced, Admin
the instrument.
Maximum pressure loss allowed that represents the
Basic
Max Leak Loss pressure loss due to the Leak Standard used to calibrate
Advanced, Admin
the instrument.
Minimum pressure loss allowable for the Master Part
Min Master Loss Basic
during the calibration cycle. Use to prevent calibration to
Advanced, Admin
the Self-Test cap.
Maximum pressure loss allowable for the Master Part
Basic
Max Master Loss during the calibration cycle. Use to prevent calibration to
Advanced, Admin
the Self-Test cap.
Minimum acceptable value for the Performance Factor
Min Perform Basic
compared after the calibration cycle to prevent improper
Factor Advanced, Admin
calibration.
When enabled, Environmental Drift Correction (EDC)
monitors and continually corrects final Leak Rate
EDC Enabled calculations based on the moving average shift of test Advanced, Admin
results within a band about the Master Part calibration
curve.
Percentage (Plus and Minus) band defined by a
percentage of High Limit about the Master Part Curve.
EDC Percentage Advanced, Admin
EDC continually accumulates and calculates an offset
for the test results.
Number of test results utilized in calculating the EDC
EDC Quantity Advanced, Admin
factor.
113 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Program Calibration
In order to convert the pressure loss measured by the instrument to a leak (flow) rate, the instrument uses
a leak standard and needs to run the “Program Cal” routine. This procedure requires at least one known
non-leaking part referred to as a “master part“. This procedure tests a non-leaking master part connected
to the instrument using the timers and pressures established for the program. The procedure
automatically tests the non-leaking master part twice with the Relax timer delay between tests. Within
each program that uses a leak standard, Program Cal can be configured to use one of four methods. See
Setting the Cal Method and Leak Standard Location in Chapter 28 – Features.
Remember: The Leak/Cal menu has a parameter called Leak Std/Cal Define
that determines the location for the leak standard settings.
When set to “Channel”, leak standard settings are located in the Channel Config menu.
When set to “Program”, leak standard settings are located in the Program menu.
Determine how you plan to use your leak test instrument. Review Setting the Leak Standard
Values section on page 14. It is critical to make sure the leak standard values are set correctly
for proper use. To view the Cal Method parameter, you must be in the Advanced or Admin
Display User Level. To change the Display User Level, see Chapter 28 – Features.
114 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Performance Factor
Upon the successful completion of the Program Cal routine, the display will show the calculated
Performance Factor for the calibration. The Performance Factor is an estimate of the quality of the
calibration. It combines the ratio of the Hi Limit Leak result to the non-leaking master part result, the test
pressure, and difference between the Hi Lim result and the non-leaking master part result to scale its
anticipated performance. This value ranges from 0 to 100. It is generally desirable to have a Performance
Factor of 35 to 100. The actual acceptable Performance Factor can vary depending on the desired Gage
R&R performance of the test. Here is the Performance Factor equation:
Performance Factor = Cal Ratio x Pressure Loss Penalty x Time Penalty x 100
The pressure loss penalty is greater for smaller pressure loss values. The time penalty is greater for shorter
Test cycle timer values. Generally, longer stabilize and test cycle timer settings will produce higher
Performance Factors and improved Gage R&R performance.
Note: The instrument will require a Program Cal routine if any parameters that affect the
calibration are modified.
115 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
If the Program Cal routine is successfully completed, “PROGRAM CALIBRATION PASSED” will be
displayed. If there are any problems during the Program Cal sequence an error will be displayed. If there
is an error, see Appendix A - Message and Error Codes.
116 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
Calibration Parameters
The Calibration parameters menu is located in the Main Menu > Program Config > CAL: DPL icon.
These calibration parameters are values that are measured or calculated as a result of the Program
Cal routine for a Differential Pressure Decay-Leak Std test.
Note: * = These CAL parameters are for those customers who want to do CALS on a series of
parts so they can manually input the ideal settings for these parameters to compare again
117 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 1 9 – D I F F E R E N T I A L P R E S S U R E ( D P ) D E C A Y -
L E A K S T D
118 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
Note: The instrument must be in Advanced or Admin Display User Level to view and
modify the tooling functionality. To change the Display User Level, see Chapter 28 –
Features.
The instrument includes user selectable digital inputs and outputs to increase the functionality and simplify
the application to various test requirements. The tooling functions are defined within each program. This
allows the flexibility to use different tooling functions; permitting the ability to seal unique ports for each
program. The tooling functions are located in Main Menu > Program Config > Tooling icon.
It is important that all precautions be taken when using the tooling control functions of the instrument. If
motions are being controlled, it is important to follow best engineering practices while designing the
circuitry. This may include using safety modules in the circuit.
In order to implement the tooling functions, they must be interfaced with the inputs and outputs. See
Chapter 21 – Inputs and Outputs.
Menus
Parameter Description Display User Level
The options for this parameter are On, Off, and
Disabled. On causes the tooling motion timers to
appear on the Timers menu. When On, the
instrument will perform these timers in the order
they are listed. It will also allow the associated
tooling outputs to appear in the selectable list in the
Tooling Option Advanced, Admin
Outputs menu. When Disabled, the tooling motion
timers still appear but are ignored. It will also allow
the associated tooling outputs to appear in the
selectable list; however, the outputs are not active in
this mode. When Off, the timers will not appear or
be active.
This parameter defines the number of motions
Number of Motions Advanced, Admin
intended to be used. You can have up to 5 motions.
For safety purposes, it may be desired to have 2
Two Inputs to Start independent inputs to activate a test cycle. The 2 Advanced, Admin
inputs are start and cycle.
119 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
120 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
There are applications where it may be required that one or more ports on the test part are blocked or
opened between tests to appropriately evaluate multiple chambers of the part. To accomplish this
additional part manipulation, tooling motions are required between program executions. Thus far we
have covered tooling motions pre-test (Tooling Extend) and post-test (Tooling Retract), this feature adds
a method for the instrument to preempt the pre/post-test tooling motions. Again, this feature is only for
the Parent Program Linking test type.
On the Program Config menu select the Link icon. The Link Motion parameter defines the Tooling
Motion to execute between the Target Program and the next appropriate program number. The options
available for this parameter are based on the tooling Number of Motions parameter defined on the
Tooling menu under the Part Config menu.
The instrument will automatically manage the pre-test and post-test tooling extend and tooling retract
motions. The user only needs to determine:
a) The total number of motions needed (with the Number of Motions parameter on the Tooling
menu within the Part Config menu).
b) How many motions may change between each linked Child program (with the Link Motion
Preempt parameter on the same Tooling menu).
c) Which motions are to occur between linked Child programs (with the numbered Link Motion
parameters on the Link menu under the Program Config menu).
121 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
The following example of Parent Program Linking has two tooling motions, with one link. The
instrument automatically places appropriate Tooling Extend motions pre-test and Tooling Retract
motions post-test. However, the operator configured a Link Motion as Tooling Extend 2, thus the
instrument will not execute the pre-test Tooling Extend 2.
Link 1
Program 4
Target
Link
Tooling Extend 2
Motion
Accept Reject
Program 5 Program 6
122 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
Link 1
Program 1
Target
Link
Tooling Retract 3
Motion
Accept Reject
Program 3 Program 4
It is helpful when designing a Parent Program Linking test to diagram the desired part operation
to appropriately define all the link parameters correctly to achieve the correct test sequence.
123 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 0 – T O O L I N G C O N T R O L
124 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Note: The instrument must be in Advanced or Admin Display User Level to view and
modify the Input and Output functionality. To change the Display User Level, see Chapter
28 – Features.
The instrument includes user selectable digital inputs and outputs to increase the functionality and
simplify the application to various test requirements. The inputs and outputs are divided into two groups
– Universal and Program Specific. When a Universal input or output is assigned in a program, it is
automatically assigned and available within all programs. When a Program Specific input or output is
assigned in a program, that particular input or output is constrained to use only the inputs or outputs within
the same Functional Group. When the input or output is set to “Constrained” in a program, it will not
function within that program.
In order to remove the selection of an input or an output, change the parameter to either "Unassign" or
"Unassign all Prog". Selecting “Unassign all Prog” will remove the selection of the input or output for all
of the programs. Selecting “Unassign” will only remove the selection from the current program.
“Unassign” is not available for Universal inputs or outputs.
125 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Detail A
DIGITAL INPUT/OUTPUT
7-12 (24VDC NOMINAL)
BOTTOM CONNECTOR LEVEL I/O B SWITCH JUMPER POSITION
FOR INTERNAL I/O POWER
The instrument is supplied with the option to elect to use the internal power for the I/O supply or an
external supply. If an external supply is used, the I/O Factory Jumper is already set to “External I/O
Power”. See I/O Jumper in “Detail B” above.
126 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
User selectable or programmable inputs are listed in a table on the next page.
127 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
128 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
The Inputs and Outputs above are listed in their menu selection order. All of the options may not
be available, depending on the test type and the tooling control setting (On or Off and Number of
Motions). Descriptions of these menu options are on the following pages are in related groups.
129 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
130 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Vent/Halt Request
The design of a Vent/Halt request (including initial phase of the Stop/Reset request) is to cease operation
of tooling motions, safely remove all energy from the part, and stop operation.
A Vent/Halt request operation has multiple operations depending on the state of the instrument while
testing. For better understanding, the test sequence can be broken into three basic groups: Tooling Extend,
Part Testing, and Tooling Retract. The Program Evaluation (Accept Program/Reject
Program/Malfunction/Severe Leak) is always after the Tooling Retract.
Other than a user generated Vent/Halt input, a number of test errors will generate a Vent/Halt request.
Possible scenarios include:
Stop/Reset Request
The design of a Stop request is to safely remove all energy from the part, place the tooling into a fully
retracted state, and stop operation. Depending on whether the instrument is actively testing or idle, the stop
request has two modes of operations.
Besides a user generated Stop, a number of test errors will generate a Stop/Reset request that causes a
Vent/Halt routine and continues with the tooling reset. The possible scenarios include: Transducer
Malfunction (zero or over-range), External Switch Fault (not low or high at appropriate time), Severe Leak,
Pressure Low/High, Calculation Faults, Calibration Errors, and General Program Fault (setup errors).
131 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
132 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Input Description
Programs can be remotely selected using the Binary Program Selection inputs
or using one of the communication methods: RS232 or TCP/IP. The number
of required inputs for Binary Program Selection depends on the highest
program number of the program to which access is required.
Binary Program
Selection Program Numbers Binary Program Inputs
1 B1
2–3 B1 and B2
4–7 B1, B2, and B3
8 – 15 B1, B2, B3, and B4
16 – 31 B1, B2, B3, B4 and B5
32 – 63 B1, B2, B3, B4, B5, and B6
64 – 99 B1, B2, B3, B4, B5, B6, and B7
Note: Programs can also be remotely selected using one of the communication methods:
RS232 or TCP/IP, see Chapter 24 - Communication.
133 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Failure to receive this input before the end of the Part Mark Fdbk timer will
cause a Malfunction result and an “ERROR: PART MARK FAULT” display.
If enabled by selecting Tool Ext Fdbk #x as one of the inputs, it is required
that the Tool Extend Feedback input go high before the associated Tool Ext
Tool Ext Fdbk
Timer expires. If the timer expires before the input goes high the instrument
1–5
will issue a malfunction. A Stop/Reset will need to be initiated to reset the
tooling.
If enabled by selecting Tool Ret Fdbk #x as one of the inputs, it is required
that the Tool Retract Feedback input go high before the associated Tool Ret
Tool Ret Fdbk
Timer expires. If the timer expires before the input goes high the instrument
1–5
will issue a malfunction. A Stop/Reset will need to be initiated to reset the
tooling.
134 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
135 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
136 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Output Description
This output goes high during the Tooling Retract 5 segment. This option is
Tooling
only available if Tooling is set to “On” or “Disabled” and the Number of
Retract 5
Motions is equal to 5 or more.
This output goes high during the Tooling Retract 4 segment. This option is
Tooling
only available if Tooling is set to “On” or “Disabled” and the Number of
Retract 4
Motions is equal to 4 or more.
This output goes high during the Tooling Retract 3 segment. This option is
Tooling
only available if Tooling is set to “On” or “Disabled” and the Number of
Retract 3
Motions is equal to 3 or more.
This output goes high during the Tooling Retract 2 segment. This option is
Tooling
only available if Tooling is set to “On” or “Disabled” and the Number of
Retract 2
Motions is equal to 2 or more.
Tooling This output goes high during the Tooling Retract 1 segment. This option is
Retract 1 only available if Tooling is set to “On” or “Disabled”.
137 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Tooling Retract 5
Tooling Retract 4
Tooling Retract 3
Tooling Retract 2
Tooling Retract 1
Tooling Extend 1
Tooling Extend 2
Tooling Extend 3
Tooling Extend 4
Tooling Extend 5
Stabilize
Exhaust
Prefill
Test
Fill
Tooling Extend 1
Tooling Extend 2
Tooling Extend 3
Tooling Extend 4
Tooling Extend 5
Press Select
In Prefill
In Fill
Fill Valve
In Stabilize
In Test
Isolate Valve
In Exhaust
Tooling Retract 1
Tooling Retract 2
Tooling Retract 3
Tooling Retract 4
Tooling Retract 5
Test Passed
Test Failed
Between Limits
Program Accept
Program Reject
Severe Leak
Malfunction
138 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
Tooling Example
The example below shows the power and flexibility of the Input and Output functionality. In this example
the inputs and outputs are being used to seal two different types of parts. The two different parts have
sealing ports in different locations. The part presence sensors select the program in the instrument that
controls the proper hardware based on the current part in the sealing nest.
A
B
Input 4 & 6
Part Presence Seal B Input 3 & 5
Switch B Part Presence
Switch A
Start Stop
Button Button
Input 1 Input 2
IOH IOH
139 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 1 – I N P U T S A N D O U T P U T S
140 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
Chapter 22 – EtherNet/IP
This chapter explains the control capabilities of the I28 using the EtherNet/IPTM feature. The instrument is
able to communicate over Ethernet via TELNET. EtherNet/IP is a registered trademark of ODVA, Inc.
Compatibility
Provide full communication capabilities over EtherNet/IP™ with any Allen-Bradley
ControlLogix® or CompactLogix PLC™.
141 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
142 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
The Inputs and Outputs in the previous table are listed in their menu selection order. All of the
options may not be available, depending on the test type and the tooling control setting (On or
Off and Number of Motions). Descriptions of these menu options can be found in Chapter 21 –
Inputs and Outputs.
143 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
144 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
Additional Menus
Turning on this feature will make 2 additional menu icons visible on the Monitor menu screens
and 2 additional menu icons visible on the Program Config menu screen.
Also see Chapter 30 – Monitor Screen Examples to see how the HW and IP monitor screens
appear.
145 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 2 – E T H E R N E T / I P
146 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
Chapter 23 – PROFINET
This chapter explains the control capabilities of the I28 instrument using the PROFINET feature.
147 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
are the same as the lists of hardware inputs and hardware outputs described in the Programmable
I/O Table in the previous chapter titled Chapter 21 – Inputs and Outputs.
148 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
149 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
CAUTION: When changing the PROFINET setting from “Enabled” to “Disabled”, all of
the user assigned IP Inputs and IP Outputs become Unassigned. This is necessary because
when PROFINET is no longer functional, any I/O that may cause program faults must be
removed. This includes Part Present, External Pressure Switch, etc. This means that all the
user assigned IP Inputs and IP Outputs must be set up again.
150 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
Additional Menus
Turning on this feature will make 2 additional menu icons visible on the Monitor menu screens and
2 additional menu icons visible on the Program Config menu screen.
151 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 3 – P R O F I N E T
152 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Chapter 24 – Communication
The instrument is able to communicate over RS232 and/or Ethernet via TELNET.
This chapter explains how to setup communication with the instrument and how to
decode the test results output.
The communication parameters are located in Main Menu > Global Config icon.
153 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Note: The instrument always uses 8 data bits. The Parity is set to “None”.
The instrument uses 1 stop bit. The flow control is always set to “None”.
Once you have established communication with the desired device you may select whether you want the
instrument to “echo” back each character it receives on the TCP/IP 1 communication port. This setting
is located in the Main Menu > Global Config > RS232 1 icon > RS232 1 Echo or
Main Menu > Global Config > RS232 2 icon > RS232 2 Echo. If the parameter is set to “ON”, the
instrument will output an echo for each character it receives. If this parameter is set to “OFF”, the
instrument will not echo anything.
154 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
CTS DataHub
The CTS DataHub interface is a custom interface designed to work with a QualityWorX CTS
Datahub setup. When CTS DataHub is enabled, the QualityWorX CTS Datahub setup will be able
to communicate with the I28 instrument to capture streaming and result data. This data is stored on
the QualityWorX CTS Datahub setup and allows for engineers and technicians to analyze and report
on a collection of data from your production line.
The CTS DataHub parameter is located in Main Menu > Global Config > Network. When CTS
DataHub is enabled, a new parameter Configuration Key will appear. This parameter is currently
unused.
To setup QualityWorx CTS Datahub, please refer to the QualityWorX CTS Datahub Getting
Started guide.
The TCP/IP parameters are located in Main Menu > Global Config > TCP/IP. The first step in
establishing Ethernet communication with the instrument is to set the Obtain Network Settings parameter
to DHCP or Manual (static IP address). If set to “DHCP”, the Instrument IP Address becomes a read only
parameter. If set to “Manual”, then you will need to set the Instrument IP Address manually. You may
choose to let the instrument initially get its settings with DHCP and then change the setting to Manual to
edit the IP Address of the instrument or lock it so that it will not change.
The Mail Server IP Address parameter should be set to the address of the SMTP mail server.
The Subnet Mask parameter will need to be configured next. The most common configuration for this
parameter is “255.255.255.0” unless there is more than one subnet in which case a common configuration
is “255.255.0.0”.
The MAC Address is the hardware address of the instrument. This number is read-only.
Once you have established communication with the desired device you may select whether you want the
instrument to “echo” back each character it receives on the TCP/IP 1 communication port. This setting
is located in the Main Menu > Global Config > Telnet 1 icon. Select Telnet 1 Echo.
If the parameter is set to “On”, the instrument will output an echo for each character it receives.
If this parameter is set to “Off”, the instrument will not echo anything.
155 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Header Description
XX 8-Bit CRC in HEX. Used for error checking.
Sequence Code in HEX. The value increments
from 01 to FF. This value can be used as a
YY
verification that all data has been received and
nothing was missed by the receiving device.
ZZZ Data length in HEX.
Tab
H Data Type Code. See Data Type Code Table
Tab
Note: For the most up to date Data Type codes, type “TABLE HEADER” into the
instrument using the communication port. The instrument will return a list of about 8
Data Type Codes and descriptions. The table below is an example of the instrument
response.
156 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
In order for the instrument to send the test Result Data automatically once the test is complete the parameter
is set to “ON”. Once this parameter is turned on, the Result Field data parameters show on the screen.
Each parameter may be turned on or off depending on the information that is required for to accompany
each result. The Test Field parameter may be set to “All Result Information” or “First 2 Test Results”. The
“First 2 Test Results” will send the two primary results. The table below shows the format of the Test
Result Data.
Number of
Parameter Characters Format Example Text Description of Example
Channel # 4 C## C01 Channel 1
Port # 3 N# N1 Port 1
Program # 4 P## P01 Program 1
Link
4 R-- No Link
Information
Time 13 HH:MM:SS.XXX 16:15:14.123 16 hrs, 15 mins, 14.123 secs
Date 9 MM/DD/YY 40179 02/01/16
Unique Id 11 ########## 0000098353 Unique test number
Program
3 # A Accept
Evaluation
SPC Flag 2 # * SPC Test Data Result
Barcode 41 1…40 12345 Barcode Data
Test Field First 2 Test Results
Pressure Decay-Leak
Test Type 8 ### PLR
Std
Test
2 # P Pass
Evaluation
Test Data Identifier -
Test Data 1 22 TDI Data Unit LR 0.123456 sccm
Value - Unit
Test Data Identifier -
Test Data 2 22 TDI Data Unit LR 0.123456 sccm
Value - Unit
TAB Tab
TAB Tab
CR Carriage Return
LF Line Feed
157 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Note: For the most up to date Test Data Identifier Codes, type “TABLE VARIABLE” into
the instrument using the communication port. The instrument will return a list of about 550
Test Data Identifier Codes and descriptions. See the full list in Appendix E.
Note: For the most up to date Test Evaluation Codes, type “TABLE EVALUATION” into
the instrument using the communication port. The instrument will return a list of about 120
Test Evaluation Codes and descriptions. See the full list in Appendix E.
Note: For the most up to date Program Evaluation Codes, type “TABLE RESULT” into
the instrument using the communication port. The instrument will return a list of about 12
Data Program Evaluation Codes and descriptions. See the full list in Appendix E.
158 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Example
Parameter Format Text Description of Example
Comma
Channel # C## C01 Channel 1
Delimited
Comma
Program # P## P01 Program 1
Delimited
Comma
Segment XXX PRF Prefill Segment
Delimited
Comma TDI Value LR 0.123456 Test Data Identifier - Value -
Test Data
Delimited Unit sccm Unit
TAB Tab
TAB Tab
CR Carriage Return
LF Line Feed
Note: For the most up to date Segment Codes, type “TABLE SEGMENT” into the
instrument using the communication port. The instrument will return a list of about 120
Segment codes and Descriptions. See the full list in Appendix E.
Note: For the most up to date Test Data Identifier Codes, type “TABLE VARIABLE”
into the instrument using the communication port. The instrument will return a list of
about 550 Test Data Identifier codes and Descriptions. See the full list in Appendix E.
Note: Only ONE interface at a time can be selected for live streaming measurement data.
159 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 4 – C O M M U N I C A T I O N
Reports
The instrument is capable of generating a variety of reports through RS232, Ethernet, Email, or USB
memory port located on the front of the unit. The reports available are in the table below.
Report Description
Reports all of the test results for the current active program. The current
Cur Program Res
program can be seen in the bottom right hand corner of the display.
All Results Reports all of the test results in the instrument memory.
Chan Last 1000 Reports the last 1000 test results.
Chan Last 100 Reports the last 100 test results.
Chan Last 20 Reports the last 20 test results.
Chan Last Res Reports the last test result.
Reports all of the parameters and their settings within the Global Config
Global Config
menu.
Reports all of the parameters and their settings within the Channel Config
Channel Config
menu.
Reports all of the parameters and their settings within the current active
Cur Program Config program. The current program can be seen in the bottom right hand
corner of the display.
Reports all of the parameters and their settings within the Program Config
Program Config
menu.
Reports the Electronic Regulator Calibration data if the instrument is
Regulator Cal
configured with an electronic regulator.
Reports the Transducer Verification data of the last Transducer
Transducer Ver
Verification conducted on the instrument.
Reports the Transducer Calibration data of the last Transducer Calibration
Transducer Cal
conducted on the instrument.
Channel Cntrs Reports all of the counters in the instrument.
160 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Chapter 25 – CTSnet LT
This chapter explores the basic use of CTSnet LT. CTSnet LT is a PC based
graphical user interface that allows you to control, configure and monitor your
instrument, all from a single, easy-to-use application. Please refer to the quick
start guide on the USB drive included with your instrument for instructions on
installing the application.
Compatibility
CTSnet LT application will require Win 10 OS installed on PC.
Initial Startup
After Setting up your I28 instrument(s) on the network, launch CTSnet LT. Your instrument(s) will
appear on the Module Management screen. You can connect to an I28 by pressing the connect icon
on each available instrument. The connected instrument’s information will turn green to indicate
that it is connected.
Note: The CTSnet LT application can only connect to one I28 instrument at
time.
161 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Once an I28 instrument is connected, you will be taken to the home menu. From here you can fully
control and configure the I28 instrument to your specifications. This includes running and
monitoring tests and calibration.
162 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Monitor Screen
Summary Tab
The Summary tab will allow you to start, stop and hold tests. You can select the program being run
by using the drop-down selector. Program calibration, if available for the selected program, can be
performed from this tab as well.
The test can be visually monitored through its stages on the top plot. This plot displays the value of
the primary transducer against elapsed time, as well as marking each phase of the test with color
coordination. The bottom plot will display the value of the Secondary Transducer during the test
phase.
Each plot can be interacted with using the icons to the right of the plot or expanded to fit the screen
with the maximize button in the upper left of the plot.
Plots shown are for demonstration, actual view may differ based upon the selected test type.
163 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
IO Tab
The I/O tab will allow you to monitor each input and output state in real time. This window can be
expanded to fit the full screen by pressing the maximize button on the upper left of the I/O list.
When EtherNet/IP is enabled, the EtherNet/IP IO states will also be displayed here.
164 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Results Screen
Results Tab
The Results tab displays the allowed number of results and allows for filtering by pressing the filter
button. Results can also be cleared by pressing the delete button.
Filtering Results
165 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Counters Tab
The Counters tab displays the number of cycles performed on the instrument including the number
of accept and reject cycles. These can be cleared to reset statistics.
Statistics Tab
The Statistics tab displays statistical analysis on the results from a selected program. The population
of data can be adjusted using the Stat History Length parameter.
166 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Configuration Screen
Program Configuration Tab
Setting Timers
Navigating to the Timers menu under the Program Config tab will allow you to set the various stage
timers to the desired values. These timers may be preset at CTS for a specific set of test parameters.
Please contact CTS to get information on how to set timers for your specific test. Timers are set per
program and may vary depending on the selected test type.
167 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Setting Pressure
Navigating to the Pressure menu under the Program Config tab will allow you to set the target
pressure for the selected test, as well as the minimum and maximum constraints for the test to
continue.
168 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Miscellaneous settings
Navigating to the Misc menu under the Program Config tab will allow you to enter in a custom
name for the selected program, as well as copy programs from one program to another. Various
other settings, such as Electronic Regulator Control Method (if an electronic regulator is available),
Prefill Method, Pressure Correction and Exhaust Method. Refer to Chapter 28 for setting
parameters.
169 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
170 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Tooling
Navigating to the Tooling menu under the Program Config tab will allow you to enable or disable
tooling motions. The part present check input may be executed before or after tooling motions.
Refer to Chapter 20 – Tooling Control
171 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Autorun
Navigating to the Autorun menu under the Program Config tab will allow you to configure
automatic tests, where the selected program will be run a selected number of times.
Refer to Chapter 28 - Autorun for setting parameters.
172 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Hardware Configurations
Navigating to the Hardware menu under the Channel Config tab will allow you to view the factory
settings of your I28 instrument(s).
Self-Test
Navigating to the Self Tst menu under the Channel Config tab will allow you to check your I28
instrument(s) for internal leaks at a set pressure. Refer to Chapter 28 - Self-Test for more
information.
173 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Setting Units
Navigating to the Units menu under the Channel Config tab will allow you to define units for every
program. Refer to Chapter 2- Setting the Units of Measure for more information and for setting
parameters.
174 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Calibration
Navigating to the Set/Span menu under the Channel Config tab will allow you to run verifications
on the transducers available in the I28 instrument as well as an Atmospheric Pressure Check to
determine the current atmospheric pressure. Calibrations are done by CTS service. Refer to Chapter
2 - Setup for more information.
Miscellaneous parameters
Navigating to the Misc menu under the Channel Config tab will allow you to set the maximum
allowable programs available up to 99 total.
175 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Network
Navigating to the Network menu under the Global Config tab will allow you to change the network
parameters for your I28 instrument(s). Please contact your IT department for valid network settings
before making changes and refer to Chapter 24– Communication for more information.
176 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Telnet Communications
Navigating to the Telnet menus under the Global Config tab will allow you set various parameters
regarding telnet communications. The I28 instrument can support 4 separate telnet channels at once:
Telnet 1, 2, 3 and 4. Refer to Chapter 24 - Communication for more information.
Email Settings
Navigating to the Email menu under the Global Config tab will allow you to create various email
reports. Refer to Chapter 24 – Communication for more information.
177 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
USB settings
Navigating to the USB menu under the Global Config tab will allow you to generate reports onto
the USB device by directly connecting the USB device to the instrument.
Miscellaneous settings
Navigating to the Misc menu under the Global Config tab will allow you to change your I28
instrument’s display name, date, time and user level. Refer to Chapter 28 – Features for more
information and to set parameters.
178 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Graph
This menu when selected displays following options as per below screenshot.
Show Segment Markers, Graph During and Y Measurement.
179 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Version
Navigating to the Version menu under the Global Config tab will allow will display version
information about your I28 instrument(s).
180 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Security
Navigating to the Security menu under the Global Config tab will allow you to set a password for
your I28 instrument(s) as well as secure various features and functions of the instrument(s).
Whenever a secured feature or function is accessed, the user will be prompted to enter the security
password.
181 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Utilities Screen
Copy Program Tab
The Copy Program tab will allow you to copy any available program to one or more instruments.
1) Select the program you wish to copy from the list of available programs.
2) Select the instruments you wish to copy them to.
3) Select the program you wish to overwrite with the copied program and press Copy
182 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
183 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Backup/Restore Tab
This tab will allow you to create instrument backup of all settings, restore an instrument back up file, create
a backup file of all programs, restore a backup file of all programs as shown in below screenshots.
184 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
185 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Export Screen
The Export section will allow you to export a various results as .txt files.
You can also export System Debug Log to send to CTS for service needs.
186 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
Users Screen
Pressing the user icon on the bottom left of the screen will display a log of user actions along with a
description of the action. This log can be filtered by pressing the filter icon, exported by pressing the
export icon or deleted by pressing the delete icon.
187 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 5 – C T S N E T L T
188 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
Chapter 26 – Webserver
The instrument provides a rich web server to control and configure the instrument.
The Web Server can be used to display and update configuration, backup/restore
instrument configurations, upgrade instrument firmware, and navigate remotely.
System Requirements
To use the Web Server, you must have the following:
• a TCP/IP-based network
• a web browser installed on your computer, preferably either Google Chrome 60.0X or
Microsoft Internet Explorer 11.0X and above
Note: The web server can be accessed from any operating system that supports the
TCP/IP protocol and a web browser. You cannot view the embedded web server pages
from outside a firewall.
Home Page
The top section will always display the Identity information (Instrument name, Firmware Version,
Serial number, and MAC address) of the instrument. The Home page is shown in Figure 1.
189 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
190 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
The Channel config page allows you to configure Channel Config parameters for the instrument.
191 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
The Global config page allows you to configure Global Config parameters for the instrument.
The Program config page allows user to configure Program Config parameters for the selected
program. To select a program, move the cursor onto “Program Config” menu. This will display
available slots in the instrument i.e. 1-9, 10-19, 20-29, 30-39, 40-49, 50-59, 60-69, 70-79, 80-89, 90-
99. Hold the cursor on 1-9 slot to display list of programs under it. Select the required program.
Information for this program will be displayed. This same technique can be used to select and
configure any other available program.
192 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
The Reports menu allows user to generate a various Reports for the instrument. Click Reports
menu so display Test Results, Report Configuration and Log Reports tab. As per below screenshots,
expand individual tabs to display available report selections. User can choose any report and view
information as required.
193 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
194 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
The Navigation page gives user a similar user interface to the one available on the front display of
the instrument.
195 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
The Backup Restore page allows user to generate a backup file of all the Program Configuration
settings and Restore any previously generated backup file.
The Firmware Update page allows user to update the instrument’s main controller firmware, co-
processor firmware, IO-Board firmware.
196 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
197 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 6 – W E B S E R V E R
198 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 7 – S E C U R I T Y
Chapter 27 – Security
This chapter explains the instrument security parameters. The Security menu allows flexibility in
allowing access or locking functions with a password.
It is located in Main Menu > Global Config menu. The Security icon may be hidden. To make the
Security icon visible, the setting is located in Main Menu > Global Config > Misc icon. Select
Edit/View Security, press Enter, use arrows to select “On”, and press Enter. You will be taken to the
Security menu. The Security icon is now visible at the bottom of the Global Config screen.
When the instrument is secured with any of the settings in the Security menu, a password is required to
unlock that function of the instrument. The default password is “5555”. Anyone with access to this
manual will know the default password. To secure the instrument, the password must be changed.
Remember to write down the new password. If the new password is forgotten, Cincinnati Test Systems
Service department can provide a temporary password to unlock the instrument.
The table below describes all of the settings in the Security menu.
Parameter Description Display User Level
Edit allows setting a new password after entering the Basic
Change Password
old password. Advanced, Admin
Applies security to performing a program Basic
Secure Calibration
calibration. Advanced, Admin
Secure Select Applies security to changing test programs by Basic
Program pressing Change Program buttons. Advanced, Admin
Secure Program Basic
Applies security to changing test part parameters.
Config Advanced, Admin
Secure
Applies security to changing instrument Basic
Global/Chan
configuration parameters. Advanced, Admin
Config
Applies security to clearing test Result Data from Basic
Secure Clear Data
instrument. Advanced, Admin
Applies security to clearing the counter registers Basic
Secure Reset Cntrs
from instrument. Advanced, Admin
Secure Hold Basic
Applies security to hold function.
Function Advanced, Admin
Applies security to protect releasing the part on a
Secure Reject Basic
reject. This security option only works if Retract on
Release Advanced, Admin
Reject is set to “No” in the program Tooling menu.
Applies security to the Monitor screen. If set to
Secure Monitor “Yes”, the user will be unable to view any other Basic
Screen screens other than the last Monitor screen viewed Advanced, Admin
before being secured.
Allows the user to save a backup and/or restore
Basic
Backup/Restore various instrument settings. Note: The manifold
Advanced, Admin
code must match to restore channel settings.
199 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 7 – S E C U R I T Y
Note: The security feature is not applicable to commands over Serial, Telnet, EtherNet/IP
and Digital I/O.
200 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Chapter 28 – Features
This chapter explains essential features of the instrument.
Suggestion: Set the Display User Level to the most basic level for your requirements.
Parameters may be locked from editing using Security, see Chapter 27 – Security.
Note: The instrument must be in Advanced or Admin Display User Level to view and
modify the instrument Language. To change the Display User Level, see Chapter 28–
Features.
Note: The copied program will overwrite all parameters in the target program. Make
sure you want to overwrite this program before executing. If a program exists in the
target program number, the instrument will prompt to confirm that you want to continue.
201 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Note: The NUMBER at the end of the backup filename will increment for each backup
of the instrument saved on the same day.
To execute this function, press the USB button on the user interface of the instrument (or press Main
Menu > Global Config > USB icon). Move the cursor down to the Backup/Restore parameter. Make
sure a USB memory stick is placed in the USB port on the front of the instrument. Press Enter to activate
the edit function and select “Backup Instrument”. When it is complete a pop-up window will tell you it
was successful.
Instrument Cloning
This function should only be executed under the advice of CTS personnel. This feature is used for service
and support of our products. It enables a technician to replicate the exact instrument within our facility to
help with any questions or concerns. To execute this function, press the USB button on the user interface
of the instrument (or press Main Menu > Global Config > USB icon). Move the cursor down to the
“Backup/Restore” parameter. Make sure a USB memory stick is placed in the USB port on the front of
the instrument. Press Enter to activate the edit function and select “Instr Clone”. When it is complete a
pop-up window will tell you it was successful.
The Light Feature Brightness parameter, on the same menu, may be adjusted to 20%, 40%, 60%, 80%, or
100% of maximum brightness.
Note: If the instrument is configured with an internal leak standard(s), the instrument
automatically introduces the leak standard during the appropriate segments of a Program
Cal routine.
202 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Self-Test
The Self-Test diagnostic provides a way to check the integrity of the instrument’s pneumatic circuit. This
is a great way to isolate a potential leak by allowing the user to verify the instrument is working correctly.
By isolating the instrument from the test part and the external tooling and plumbing, this special test will
verify that the instrument is performing properly and leak free.
The Self-Test diagnostic is located in Main Menu > Channel Config > Self-Test icon.
Note: It is important that all test ports are plugged during a Self-Test. The instrument
will multiplex through the test ports to verify all circuits are operating leak free.
203 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Update Firmware
The instrument is able to update the firmware using the USB port on the front of the instrument. Upon a
proper Service request, a firmware update can be sent by an authorized CTS person by email for transfer to
a (FAT32) USB memory stick. The file sent will be saved as a ZIP file. Extract the ZIP file to the root
directory of the USB memory stick. This step creates the proper folder structure on a USB memory stick.
Safely remove the USB memory stick from your computer and insert the stick into the USB port on the
front of the instrument. Press the USB button on the user interface. Change the parameter “Update
Firmware” to “Yes” to start the update process. The instrument will list all of the firmware version files on
the screen. Select which file you want to use. This will start the update process.
Note: If the firmware version is a minor revision all of the parameter settings will be kept
the same. However, if the firmware version is a major revision all of the settings will be
cleared. See the firmware documentation to determine if the firmware version you are
upgrading from will cause the settings to be cleared. To be safe, always perform a Backup
Instrument and Instr Clone before updating firmware.
When the Prefill Method is set to “Percent of the Fill Timer”, which is the default value, the Minimum
Pressure must be met before the value in the Prefill is exhausted. For example, if the Fill Timer is set to 10
seconds and the Prefill Timer is set to 80% then the part must reach Minimum Pressure within 8 seconds.
This allows for fixed fill times to be utilized.
When the Prefill Timer is set to “Not to Exceed Timer”, the instrument will begin the cycle in the Prefill
segment. Once the pressure reaches the Minimum Pressure, the instrument will exit the Prefill segment,
regardless of any time remaining, and enter the Fill segment.
When the Prefill Method is set to “Controlled Fill”, the instrument will begin the cycle in the Prefill Segment
and will ramp up the pressure from 0 to the target pressure within the specified Prefill timer. When
instrument enters the Fill segment the instrument is at target pressure.
Note: “Controlled Fill” method is available only when Electronic regulator is present in
the instrument.
When the Prefill Method is set to “Fixed Charge”, the instrument begins the cycle in the Charge Segment
and charge the pneumatic with desired “Charge Target Press”. At the end of charge timer pressure must
reach “Charge Min Press” and below “Charge Max Press”. Once successfully exited from the charge
segment, the instrument will start to fill the part with the charged pneumatic volume.
Exhaust Method
The Exhaust Method parameter is located in Main Menu > Program Config > Misc icon. It can be
set to “Dump” or “Controlled Exhaust”.
204 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
When the Exhaust method is set to “Dump”, which is the default value, the instrument will open the
Exhaust valve and dump the pressure immediately after entering the Exhaust segment.
When the Exhaust Method is set to “Controlled Exhaust”, when the instrument enters the Exhaust
segment, the instrument will start ramping down the pressure from the last noted instrument pressure to 0
within the specified Exhaust timer. At the end of the Exhaust timer, the instrument will open the Exhaust
valve and dump any remaining pressure.
The instrument has the capability to store barcode data with each test result. The instrument will work with
any RS232 barcode scanner that has the ability to supply a barcode in an ASCII string (no longer than 40
characters) followed by a carriage return. To store the barcode information, it must be received before a
Start command is received by the instrument. There is an option to require the barcode data in order for
the Start command to be acknowledged.
The Barcode function is located in Main Menu > Global Config > RS232 1 icon > RS232 1 Interface
or Main Menu > Global Config > RS232 2 icon > RS232 2 Interface. Set the parameter to “Barcode”.
In the same menu, set the RS232 1 or RS232 2 Baud parameter to match the baud rate of the barcode
scanner communication.
Note: The data bits are always configured to 8. The parity is always set to None. The
number of stop bits is always set to 1. The flow control is always set to None.
Note: The instrument must be in Advanced or Admin Display User Level to view and
modify this functionality. To change the Display User Level, see Chapter 28– Features.
205 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
When the Prefill Timer is set to “Percent of the Fill Timer”, which is the default value, the Minimum
Pressure must be met before the value in the Prefill is exhausted. For example, if the Fill Timer is set to 10
seconds and the Prefill Timer is set to 80% then the part must reach Minimum Pressure within 8 seconds.
This allows for fixed fill times to be utilized.
When the Prefill Timer is set to “Not to Exceed Timer”, the instrument will begin the cycle in the Prefill
segment. Once the pressure reaches the Minimum Pressure, the instrument will exit the Prefill segment,
regardless of any time remaining, and enter the Fill segment.
206 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Autorun
Sometimes it is desired to have the instrument cycle automatically without any need for a person to press
the Start button or send a start signal using the digital inputs. The Autorun feature is designed to
accomplish this. Autorun is located in Main Menu > Program Config > Autorun icon. The following
menu will display on the screen.
Batch Calibration
The Batch Calibration feature allows multiple calibration sequences to be performed. Results are averaged
to calculate the final calibration values. The Batch Calibration parameter is located in the Main Menu >
Program Config > Leak/Cal icon and allows for enabling or disabling of batch program calibration
mode. When enabled, three more parameters “Batch Layout”, “Batch Quantity” & “Average Quantity”
are visible to user.
207 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 8 – F E A T U R E S
Note: This parameter is only available if the Prefill Method selected is “Percent of Fill
Time” or “Pre-Pressure Time”.
Valve Detection
Valve Detection is located in Main Menu > Channel Config > Hardware icon with the options
of Enabled and Disabled. This feature, when enabled, allows the instrument to monitor for the
presence of pneumatic valves via electronic load detection.
Note: This parameter is only available for viewing, and the user cannot change the
settings.
208 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 9 – I N S T R U M E N T C A L I B R A T I O N
Note: Your instrument may not have a flow transducer or an electronic regulator.
These parameters are located in Main Menu > Channel Configuration > Set/Span icon.
Verifying a Transducer
The transducer verification process is very similar to the calibration process. You must enter the external
master gauge value at each point to step through the sequence. You will be prompted to input values so the
Transducer Verification Report has this data for later review. This feature does not affect the calibration
of the instrument.
Calibrating a Transducer
The transducers have been calibrated at the factory. A calibration of the transducer should be done
periodically according to the quality standards at your facility. The instrument is capable of doing a multiple
point calibration. The number of points is determined by the user. You may select up to 32 points for
calibration. You will need a calibrated master pressure gauge to calibrate the transducer on the instrument.
On the Set/Span menu select which transducer you want to calibrate. Press Enter to enable the edit
function. Change the value to “Yes”. The following menu will display on the screen.
209 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 9 – I N S T R U M E N T C A L I B R A T I O N
210 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 9 – I N S T R U M E N T C A L I B R A T I O N
Note: The Low Limit Voltage and High Limit Voltage values are disregarded if both are
set to “0” (zero is the default setting). You should only change these values if the
instrument is not able to automatically analyze the electronic regulator range and perform
a proper calibration.
Note: The electronic regulator can remain active in between instrument test cycles or it
can be set to “Idle”. This parameter is edited by going to Channel Config and selecting
the Hardware icon. The parameter is called “E-Regulator Rest” and is available in Basic
Display User Level. To change the Display User Level, see Chapter 28 – Features.
211 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 2 9 – I N S T R U M E N T C A L I B R A T I O N
212 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 0 – M O N I T O R S C R E E N E X A M P L E S
Screen 2: Monitor 2 highlights the location of the reading from the Min to the Max
allowable pressure:
213 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 0 – M O N I T O R S C R E E N E X A M P L E S
Note: You can zoom/adjust the graph by pressing the Enter button.
A menu will display on the right side of the screen. Additional graph
screen parameters are located in Main Menu > Global Config > Graph
icon.
Screen 4: Hardware Input highlights which of the possible 12 inputs are engaged:
Screen 5: Hardware Output highlights which of the possible 12 outputs are engaged:
214 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 0 – M O N I T O R S C R E E N E X A M P L E S
215 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 0 – M O N I T O R S C R E E N E X A M P L E S
216 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 1 – R E S U L T S S C R E E N E X A M P L E S
Counters Screen
Results screen
Stats Screen
217 www.cincinnati-test.com
S E N T I N E L I 2 8 C H A P T E R 3 1 – R E S U L T S S C R E E N E X A M P L E S
218 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
219 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
220 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
Cal Program
CA The Program Calibration was successful.
Accept
Cal Required -
CF Not Used.
Limit Exceeded
221 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
Invalid I/O
IO Not Used.
Configuration
222 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
223 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
Part
There was an error in the way the program
Configuration PE
was configured.
Error
Check pressure
Test pressure was above the Maximum
Test Pressure regulator setting and
PH Pressure during fill or stabilization cycles
High tooling control
resulting in a Malfunction.
pressures
Check the part
Error: Part Not The part present input is set for the active
PP presence sensor and
Present program and the input was not received.
input wiring
This is an error in the Auto Setup Sequence
Part Not
PS that can occur if the instrument fails to
Stabilized
stabilize the part.
225 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
226 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
227 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X A – M E S S A G E S & E R R O R C O D E S
228 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X B – Q U I K T E S T
– Quik Test
This appendix explains the theory and parameters for reducing the test time for obviously good and
obviously bad parts, without sacrificing the accuracy for marginal parts. This capability is called Quik
Test in the instrument. Quik Test is available in the Pressure Decay-Leak Std test type that conducts a
Program Calibration using a leak standard.
How it works
Most manufacturing processes produce the desired quality of product a high majority of the time. These
parts easily pass the leak test being performed on them. Conversely, most defective parts reject by a wide
margin. Quik Test monitors the pressure or flow signal during the test and ends the test early for parts that
easily pass or easily reject. Any marginal parts that are within the parameters set by the Quik Test HL (High
Limit) Band and Quik Test LL (Low Limit) Band will utilize the entire Test time to ensure the accuracy of
the result.
At the desired time in the Test sequence determined by the Quik Test Percent parameter the instrument
calculates the projected leak rate and compares it with the Quik Test Low Limit and High Limit Bands. If
the projected leak rate is within either of these two bands, the test continues to the end of the test timer. If
the instrument’s projected leak rate is outside of these bands, it will stop the test, output the results according
to the Test Evaluation criteria, and display the projected leak rate.
Exhaust
Stabilize
Pre-fill
Fill
Reject Limit
} Quik Test HL Band
Time
229 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X B – Q U I K T E S T
The prior chart shows the master part curve, the Low Limit with its LL Quik Test band limit, and the High
Limit with its HL Quik Test band limit. If the pressure loss is within either the LL Quik Test band or HL
Quik test band at the QT% of Test Timer, the test will continue and complete the entire test cycle. If the
results are outside the limit bands at the QT% of Test Timer, the test will stop and the test result will reflect
the Test Evaluation criteria. Setting narrow bands about the Low Limit and High Limit is a very aggressive
approach that will greatly limit the number of tests that run to the end of the test timer. Set wider bands as
a more conservative approach to Quik Test.
An initial analysis of Quik Test should be performed before actually implementing this feature by
establishing the “QT Test Timer” at 10, 25, 50, 75, or 90% and setting “Quik Test” to OFF. A Program
Cal must be performed after setting the "Quik Test Timer" to measure the typical leak response curve for
the test. Test production parts over a reasonable time (up to 1000 parts in various production conditions).
The instrument projects a leak rate at the Quik Test Timer and then completes the full test cycle and
calculates the leak rate for the complete test. The instrument stores the test results that include both the
actual complete test result and a projected test result based on the Quik Test measurement. By comparing
the two leak rate results, a percent of error for the Quik Test feature can be calculated. If the results are
satisfactory, the Quik Test feature can be implemented by setting appropriate High Limit and Low Limit
Bands based on the test comparison and enabling Quik Test. This feature will save the time difference
between a full test time and the Quik Test Time for almost all of the accepted parts and most of the rejected
parts. If the results are not consistent to each other, test further by increasing the Quik Test Timer, re-
calibrate the instrument, and compare the results again.
Test Parameters
The Quik Test parameters menu is located in Main Menu > Program Config > TST:*** icon.
230 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X C – E N V I R O N M E N T A L D R I F T C O R R E C T I O N
– Environmental Drift
Correction
This appendix explains the theory and parameters for implementing the
Environmental Drift Correction feature.
How it works
The environmental drift correction routine helps to maintain the calibration of the system by
continuously monitoring and calculating a correction factor for changes in the test conditions. This
routine dynamically compensates for slow changes in the test environment such as room temperature
changes, part temperature changes, test air temperature changes, part elasticity changes, part
absorption characteristics, etc. These factors influence the dynamics of how the test part reacts to the
testing process and the determination of the leak rate. “Environmental Drift Correction” defines how
wide of a band around the Master Part Loss or Master Part Flow value will be considered as normal
variations in flow rate for a non-leaking part. The “Environmental Drift Correction” (+EDC %) is
based on High Limit. The instrument continuously calculates a running average of test results that
fall within the band (EDC %) about the original non-leaking master part curve. It corrects each future
result by the calculated drift determined from previous test results. The drift value is calculated as
shown.
Σn Flow
Environmental Drift Correction = ––––––– - “Master Part Flow”
n
Typical settings for “Environmental Drift Correction” are 10%, 25%, 50%, 75%, and 90%. The sample
size is set as EDC quantity (3 or greater). Flexible parts require a greater “Environmental drift correction”
than rigid parts.
The “Environmental Drift Correction” effectively defines how much shift can occur. This limits the
routine from continuing to compensate for possible growing leaks in the seal or test pneumatics.
231 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X C – E N V I R O N M E N T A L D R I F T C O R R E C T I O N
The EDC parameters menu is located in Main Menu > Program Config > TST:*** icon.
Environmental Drift Correction (EDC) will reset based on any of the following conditions:
1. Instrument power-cycle (due to limitations in instrument non-volatile storage)
2. Change of any variables that affect EDC calculations:
a. EDC Configuration change (modification the variables: EDC Enabled, EDC
Percentage, or EDC Quantity)
b. Test Limit change (modification of the variable: High Limit [Loss/Leak/Rate])
3. Program idle time-out: 1 hour elapsed since last test on matching program number
Note: The act of changing the selected active program does not reset the
Environmental Drift Correction feature. A user can switch between multiple programs
(EDC data is stored per program), as long as the amount of time from the last time
the similar program number was run does not exceed the 1-hour time-out.
232 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
– Electronic Regulator
Compensation
This appendix explains the theory and parameters for implementing the Electronic
Regulator Compensation (ERC) feature.
How it works
The Electronic Regulator Compensation (ERC) routine helps the instrument to compensate for
electronic regulator calibration error (drift from calibrated setpoint). This allows the instrument to
reach the desired target pressure. The ERC method can be set to “Optimize Fill Pressure”, “Optimize
Test Pressure”, or “Optimize Source Pressure”.
The below graph represents a test with ERC disabled. In this example, the electronic regulator is
not calibrated for Target Pressure, so the pressure set during the test has deviated.
Exhaust
Stabilize
Pre-fill
Test
Fill
Target Pressure
Pressure
Ti
me
233 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
In the “Optimize Fill Pressure” method, ERC enables the instrument to adjust the regulator
pressure setting to obtain the optimal measured fill pressure, as determined by the Target
pressure.
In this method, the electronic regulator setpoint is adjusted on subsequent tests to force the
measured fill pressure to match the target pressure, based upon the stored result history of previous
passed tests.
Stabilize
Exhaust
Pre-fill
Test
Fill
Target Pressure
Pressure
Test 1
Test 2
Test 3
Time
234 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
In the “Optimize Test Pressure” method, ERC enables the instrument to adjust the regulator
pressure setting to obtain the optimal measured test pressure, as determined by the Target
pressure.
In this method, the electronic regulator setpoint is adjusted on subsequent tests to force the
measured test pressure to match the target pressure, based upon the stored result history of previous
passed tests.
Exhaust
Stabilize
Pre-fill
Test
Fill
Target Pressure
Pressure
Test 1
Test 2
Test 3
Time
235 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
In the “Optimize Source Pressure” method, the electronic regulator will be continuously adjusted
during the fill segment to force the measured fill pressure to match the desired target pressure for
the test.
Exhaust
Stabilize
Pre-fill
Test
Fill
Target Pressure
Pressure
Time
Note:
1. The performance of “Optimize Test Pressure” or “Optimize Fill Pressure”
methods depends on multiple iterations since it is based on the history of previous
Accept tests. For these ERC methods, only tests that pass with the measured
pressure that is inside the ERC target window will be added to the history.
2. The “Optimize Source Pressure” method does not depend on the history of
previous Accept tests. Every time you run this test, the instrument will regulate
pressure to the desired target pressure for that test only.
236 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
Test Parameters
The ERC Test parameters menu is located in Main Menu > Program Config > MISC icon.
237 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X D – E L E C . R E G . C O M P E N S A T I O N
238 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
– Communication Code
Tables
This appendix lists the codes that are referenced in the notes of Chapter 24 Communication.
Each table can be generated within the communications interface by providing the associated
command denoted.
239 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
240 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
241 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
242 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
Segment Codes
type “TABLE SEGMENT”
Segment Segment
Code Description Code Description
1 %VR Percent of Reference Volume Test 31 DPT Rate of Pressure Loss Test
2 APC Setup - Atmospheric Pressure Check 32 DTV Setup - DP Transducer Verification
3 AR Autorun Relax 33 DTZ Setup - DP Transducer Zero
4 BAL Stabilize DP Xdcr Balance 34 ERA Setup - Electronic Regulator Analyze
5 CBC Chamber - blower control 35 ERS Setup - Electronic Regulator Setpoint
6 CC0 Chamber - circulation off 36 ERZ Setup - Electronic Regulator Zero
7 CC1 Chamber - circulation on 37 ESI Internal - Empty-Seq
8 CCX Chamber - accumulation rest 38 EST Fill Evac
9 CE0 Chamber - evacuate off 39 EXE Exhaust until Empty
10 CE1 Chamber - evacuate on 40 EXH Exhaust
11 CHA Exhaust - Chamber Output Rest 41 EXP Exhaust until Pressure
12 CIF Chamber - inlet blower off 42 EXT Tooling Motion Extend
13 CIO Chamber - inlet blower on 43 FFL Fill until Full
14 CLN Stabilize Chamber Cleanup 44 FGN General Fill
15 CO0 Chamber - outlet blower off 45 FLC Mass Flow - Leak Rate Test
16 CO1 Chamber - outlet blower on 46 FLL Fill (without pressure monitoring)
17 CP0 Chamber - pre-purge 47 FLR Precise Mass Flow Test (Differential Flow)
18 CPC Chamber - clamshell purge rings control 48 FLW Mass Flow Test
19 CPG Chamber - Exh/Purge 49 FRF Fill Reference
20 CPO Chamber - clamshell purge rings option 50 FRP Fill Ramp
21 CPR Chamber - clamshell purge rings rest 51 FST Fill Tracer
22 CST Fill Clean 52 FTS Setup - Flow Transducer Setpoint
23 CV0 Chamber - vent off 53 FTV Setup - Flow Transducer Verification
Differential Pressure Decay - Leak Rate
24 DDL 54 FTZ Setup - Flow Transducer Zero
(volume calculated)
25 DLY Delay 55 HVC Chamber - hardvac control
26 DPD Differential Pressure Decay Test 56 LKC Link Control
Differential Pressure Decay Test (no abs
27 DPD 57 LNK Link Decision
pressure)
Differential Pressure Decay - Leak
28 DPL 58 MVX Setup - Mix Verification
Standard Test (no abs pressure)
Differential Pressure Decay - Leak
29 DPL 59 OCC Occlusion Test (Backpressure)
Standard Test
30 DPS Setup - DP Transducer Setpoint 60 PLO Pressure Loss Test
243 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
244 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
245 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
246 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
247 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
248 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
249 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
250 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
251 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
252 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X E – C O M M U N I C A T I O N C O D E T A B L E S
253 www.cincinnati-test.com
S E N T I N E L I 2 8 A P P E N D I X F – I N S T R U M E N T A T T A C H M E N T & C L E A R A N C E
NOTES:
1) THIS DRAWING IS
NOT TO SCALE.
2) ALL MEASUREMENTS ON
THIS PAGE ARE IN INCHS.
4) RECOMMENDED 4 INCH
CLEARANCE UNDER
INSTRUMENT FOR TEST
PORT, PNEUMATIC
CONNECTIONS AND
HAND CLEARANCE.
254 www.cincinnati-test.com
This page is intentionally blank.
255 www.cincinnati-test.com
Index
Anti-Tie-Down, 120, 130, 219 Hold, 129, 130, 199
Auto Setup, 5, 13, 28, 29, 135, 219, 224, 225 Holding Pressure After Test, 205
Autorun, 207 I/O Wiring, 126
Backup & Restore Instrument, 202 Input Initial State, 67
Backup Instrument Settings, 16 Inputs, 12, 125, 129, 130, 133, 134, 141, 143, 144, 145, 148,
Barcode, 205 151, 152, 153, 214, 215
Binary Program Selection, 133 Inputs and Outputs, 12
Cal Method, 29, 34, 114, 206 IP Inputs, 5, 144, 146, 151, 152, 153, 215
Calibration Parameters, 37, 51, 59, 117 IP Outputs, 5, 144, 146, 151, 215
Cancel/Back button, 3 Leak Flow, 59
Channel Configuration, 12, 14, 141, 148, 209 Leak Flow, 58
Clone Instrument, 202 Leak Loss, 37, 117
Common, 120, 130, 219, 225, 226 Leak Loss, 35
Communication, 4, 12, 141, 144, 148, 151, 154, 156, 157, Leak Loss, 115
159, 209, 210, 211 Leak Standard, 10, 11, 14, 27, 29, 32, 33, 34, 37, 57, 87, 91,
Communication Code Tables, 239 92, 95, 96, 107, 109, 113, 114, 115, 117, 133, 135, 202,
Corr. Leak Std Flow, 37, 59 203, 206, 207
Counters Screen, 217 Leak Standard Values, 14, 34, 57, 114
Crossover Pressure, 63, 67 Leak Std Select, 203
CTSnet LT, 162 Level State, 67
Data Type Codes, 157, 219 Level Time, 67
Delay, 80 Light Bar Control, 202
Differential Pressure Decay-Leak Std, 117 Link, 80, 121, 122, 123, 158, 223, 224
Differential Pressure Transducer, 11, 96, 102, 107, 109 Main Menu, 3, 5, 9, 12, 13, 14
Digital IO Cable Diagram, 128 Manifold, 206
Direct Flow Parameter, 52, 59 Mass Flow, 10, 47, 53
Display User Level, 14, 17, 21, 27, 34, 39, 43, 47, 53, 57, 61, Mass Flow-Leak Std, 10, 53
65, 69, 73, 83, 87, 114, 201, 205, 211 Master Flow Offset, 50, 51
EDC, 20, 25, 33, 50, 56, 57, 92, 100, 106, 113, 231, 232 Master Part, 10, 27, 30, 34, 35, 37, 47, 50, 51, 53, 57, 59, 87,
Edge Type, 67 88, 100, 104, 106, 107, 111, 113, 114, 115, 117, 206, 207,
Electromagnetic Energy Devices, 1 220, 221, 223, 224, 226, 230, 231, 232
Electronic Regulator, 15, 61, 73, 211 Master Part Flow, 58, 59, 231
Electronic Regulator Calibration, 161, 211, 222 Master Part Loss, 34, 35, 37, 93, 115, 117
Electronic Regulator Calibration Report, 211 Master Part Press, 37, 59
Enter Button, 3, 9 Master+Leak Flow, 58, 59
Environmental Drift Correction, 25, 33, 50, 56, 92, 100, 106, Master+Leak Loss, 34, 35, 37, 93, 115, 117
113, 231, 232 Master+Leak Press, 37, 59
E-Regulator Rest, 211 Max Leak Flow, 56
EtherNet/IP, 4, 12, 141, 142, 143, 144, 145, 146, 148, 151 Max Master + Leak Flow, 56
Event Type, 63, 67 Max Master Flow, 56
Event Δ Pressure, 63 Message & Error Codes, 219
Example, 78, 122, 123, 139, 157, 158, 160, 204, 213 Min Leak Flow, 56
Execution Pause, 80 Min Master + Leak Flow, 56
Exhaust, 18, 22, 30, 40, 44, 48, 54, 62, 66, 70, 74, 81, 84, 88, Min Master Flow, 56
98, 104, 111, 129, 133, 135, 136, 205 Monitor Icon, 3
FAT32, 202 Monitor Screens, 5, 145, 152, 213
Features, 18, 22, 30, 48, 54, 141, 148, 201 Multiple Unique Parts, 8, 11, 77, 81
Fill, 18, 22, 30, 48, 52, 54, 59, 98, 104, 111, 129, 133, 135, Non-Leaking Master Part, 10, 34, 35, 37, 51, 57, 107, 114,
136, 204, 206 115, 206, 207
Fill the Test Part, 17, 21, 27, 47, 53 Number of Motions, 119, 121, 137, 140
Filters, 1 Occlusion, 10, 39
Global Configuration, 141, 148 Open Leak Std, 107, 129, 133, 202
Graph, 214 Orifice, 206
Hardware Inputs, 5, 149, 214 Outputs, 12, 125, 129, 135, 136, 137, 141, 143, 144, 145,
Hardware Outputs, 5, 149, 214 148, 151, 214, 215
256 www.cincinnati-test.com
Parent Program Linking, 8, 11, 77, 78, 79, 80, 121 Start Program, 129, 130
Parsing Data, 160 Stats Screen, 217
Part Mark, 18, 120, 134, 136, 224 Stop/Reset, 78, 120, 129, 130, 131, 132, 134, 140, 226
Part Marking Feedback, 134 Streaming, 141, 148, 157, 160
Part Present, 120, 129, 130, 140, 144, 151 Target Pressure, 18, 19, 22, 23, 28, 30, 31, 37, 39, 40, 48, 49,
Part Present Check, 120, 140 54, 55, 59, 74, 75, 85, 90, 98, 99, 104, 105, 111, 112, 117,
Pass and Fail, 20, 25, 33, 41, 45, 50, 57, 63, 68, 71, 92 221, 224
Performance Factor, 35, 37, 113, 115, 117, 221 Target Test Pressure, 17, 21, 27, 40, 47, 53, 99, 105, 112
Pneumatic Regulator, 15 TCP/IP, 156, 160
Prefill, 18, 19, 22, 23, 30, 31, 48, 49, 54, 55, 85, 98, 104, 111, Technical Specifications Summary, 259
129, 135, 136, 160, 204, 206, 226 TELNET, 156
Prefill Timer Function, 204, 205 Test, 17, 18, 21, 22, 24, 30, 31, 40, 43, 44, 48, 53, 54, 66, 70,
Press Select, 129 84, 88, 90, 96, 98, 104, 109, 111, 112, 129, 135
Pressure Correction, 206, 210 Test Data Identifier Codes, 159, 160, 246, 247, 248, 249,
Pressure Decay-Leak Std, 10, 27, 28, 87, 159 250, 251
Pressure Decay-ΔP, 10, 17 Test Evaluation, 20, 25, 27, 29, 32, 33, 41, 45, 50, 56, 57, 63,
Pressure Decay-ΔP/ΔT, 10, 21 67, 68, 71, 91, 92, 99, 112, 141, 148, 158, 159, 229, 230
Pressure Regulator, 18, 22, 30, 39, 47, 48, 53, 54, 88, 98, Test Evaluation Codes, 159, 240, 241
104, 111, 224, 225, 226, 227 Test Failed, 129, 136
Pressure Select, 15, 28 Test Parameters, 16, 20, 25, 32, 41, 45, 50, 56, 63, 75, 86,
Pressure Verify, 10, 43 91, 100, 106, 113, 230, 237
PROFINET, 148 Test Passed, 129, 136
Program Cal, 5, 30, 32, 34, 35, 36, 37, 50, 51, 54, 57, 58, 59, Test Port, 1, 10, 15, 43, 203, 220, 226
81, 88, 91, 92, 104, 111, 114, 115, 116, 117, 129, 133, 135, Test Results, 20, 25, 32, 41, 45, 86, 91, 100, 106, 113, 154,
202, 220, 221, 223, 230 158, 231
Program Calibration, 34, 35, 51, 56, 57, 58, 114, 115, 199, Test Time, 10, 19, 27, 39, 85, 99, 112, 220, 221, 230
219, 221, 224, 229 Test Type, 5, 9, 10, 16, 219
Program Configuration, 15, 141, 148 Test Type Menu, 9
Program Evaluation Codes, 159 Test Type Table, 9
Program Linking, 8 Timer Parameters, 18, 22, 30, 40, 44, 48, 54, 62, 66, 70, 74,
Program Select Buttons, 4 81, 84, 88, 98, 111
Programmable Inputs and Outputs, 148 Tool Ext Fdbk, 134
Programmable Inputs and Outputs Menus, 129, 142 Tool Ret Fdbk, 134
Quik Test, 33, 56, 92, 229, 230 Tooling Control, 12, 18, 22, 30, 40, 54, 84, 88, 98, 104, 111,
Ramp Method, 66 119, 121, 224
Ramp Rate, 66 Tooling Extend, 18, 22, 30, 40, 48, 54, 62, 66, 70, 74, 81, 84,
Ramp to Flow Event, 10, 69 88, 98, 104, 111, 121, 122, 123, 131, 137
Ramp to Input Event, 65 Tooling Feedback, 120
Ramp to Proof Test, 11, 73 Tooling Motion, 18, 22, 28, 30, 40, 54, 84, 88, 98, 104, 111,
Ramp to ΔP Event, 10, 61 121, 122, 123, 132, 140, 219, 226
Relax, 30, 54, 81, 88, 129, 135 Tooling Option, 119
Reports, 161, 211 Tooling Retract, 18, 22, 30, 40, 48, 54, 62, 66, 70, 74, 81, 84,
Results Screen, 217 88, 98, 104, 111, 121, 122, 123, 131, 137
Retract on Reject, 120, 140 Transducer, 11, 13, 40, 43, 52, 57, 59, 95, 96, 98, 101, 102,
RS232, 4, 5, 133, 155, 158, 160, 161, 205, 219 104, 107, 109, 111, 209, 210, 221, 222, 227
Security, 5, 16, 199, 201 Transducer Calibrate, 209
Segment Codes, 160, 243, 244, 245 Transducer Calibration Report, 209, 210
Self-Test, 1, 13, 16, 43, 113, 203, 220, 226, 227 Transducer Verification, 227
Self-Test Failed, 226 Transducer Verification Report, 209
Self-Test Passed, 226 Transducer Zero Bad, 222, 227
Sequential Linking, 8, 11, 80, 81 TST Parameter, 20, 25, 32, 41, 45, 50, 56, 63, 67, 71, 75, 86,
Setup, 5, 7, 13, 16, 96, 109, 201 91, 230, 232, 237
Severe Leak, 19, 23, 31, 49, 55, 85, 90, 129, 131, 136, 225, Two Inputs to Start, 119, 130
226 Units of Measure, 12
SPC, 129, 133, 158 Update Firmware, 204
Stabilization Time, 96, 99, 102, 105, 109, 112 USB Memory Stick, 3, 16, 202, 204
Stabilize, 17, 18, 21, 22, 27, 30, 31, 48, 54, 88, 90, 96, 98, USB Menu, 3
102, 104, 109, 111, 115, 129, 135, 225, 229 USB Port, 202, 204
Stabilize Time, 17, 21, 27, 220, 221, 226 Vent/Halt, 78, 129, 130, 131, 132, 141, 148
Start Channel, 129 Verify Transducer, 209
257 www.cincinnati-test.com
258 www.cincinnati-test.com
Sentinel I28 Technical Specifications
Outside Physical Dimensions
Wall mount model (W x H x D): 305 x 235 x 222 (mm) 12 x 9.25 x 8.75 (inches)
Electrical Specifications
Supply power: 100-240 VAC 50-60 Hz auto sensing/switching
Input power fuse: 3 Amp 250VAC, slow, 5X20MM, glass
I/O driver fuse: 1 Amp 250VAC, fast, 5X20MM, glass
Board fuse: 3.125 Amp 250VAC, fast, radial leads, plastic can, vertical plug in fuse
Output current limit for each output: 0.5 amp max
Output current limit for all outputs combined: 1 amp max
Instrument may operate at ± 10% of nominal voltage
Environmental Factors
Operating temperature Range: 5°C to 40° C (41°F to 104°F)
Maximum humidity 90% relative humidity, non-condensing
Maximum operating altitude 2,500 meters (8,200 feet)
Indoor use only, IP20
Inlet Ports & Connection Type Connection Type & Max Pressure Rating
P1 1/4-18 FNPT or 1/4-19 BSPT fitting Vacuum to 200 psi (1380 kPa) *
P2 1/4-18 FNPT or 1/4-19 BSPT fitting Vacuum to 200 psi (1380 kPa) *
Pilot 1/8 FNPT or 1/8 BSPT fitting 65 to 105 psi (540 kPa to 725 kPa)
Hi Press 1/4-18 FNPT or 1/4-19 BSPT fitting 0 to 500 psi (3450 kPa)
* See label on instrument enclosure to determine whether port is for pressure or vacuum
Air Cleanliness Specifications
Max Particulate Size: 0.3 micron
Max Dew Point: -30°C
Max Oil: 0.1mg/m3
Recommended Filter Sets (available separately)
Small filter (includes 5.0 micron and 0.3 micron filters) for parts less than 300 ml
Medium filter (includes 5.0 micron and 0.3 micron filters) for parts from 300 ml to 13,000 ml
Large filter (includes 5.0 micron and 0.3 micron filters) for parts larger than 13,000 ml
www.cincinnati-test.com
Contact Cincinnati Test Systems Service Department
By phone 513-202-5108 or by email [email protected] 8am-5pm eastern US
For after-hours, 24/7 emergency phone support, call 513-202-5174
www.cincinnati-test.com