Lecture 1423813026
Lecture 1423813026
MEASURING INSTRUMENTS
Instrument
An absolute instrument determines the magnitude of the quantity to be measured in terms of the
instrument parameter. This instrument is really used, because each time the value of the
measuring quantities varies. So we have to calculate the magnitude of the measuring quantity,
analytically which is time consuming. These types of instruments are suitable for laboratory use.
Example: Tangent galvanometer.
This instrument determines the value of the quantity to be measured directly. Generally these
instruments are calibrated by comparing with another standard secondary instrument.
Examples of such instruments are voltmeter, ammeter and wattmeter etc. Practically
secondary instruments are suitable for measurement.
Secondary instruments
2
1.3.1 Indicating instrument
This instrument uses a dial and pointer to determine the value of measuring quantity. The pointer
indication gives the magnitude of measuring quantity.
This type of instruments records the magnitude of the quantity to be measured continuously over
a specified period of time.
This type of instrument gives the total amount of the quantity to be measured over a specified
period of time.
1.3.4 Electromechanical indicating instrument
For satisfactory operation electromechanical indicating instrument, three forces are necessary.
They are
(a) Deflecting force
(c)Damping force
When there is no input signal to the instrument, the pointer will be at its zero position. To deflect
the pointer from its zero position, a force is necessary which is known as deflecting force. A
system which produces the deflecting force is known as a deflecting system. Generally a
deflecting system converts an electrical signal to a mechanical force.
3
1.4.1 Magnitude effect
When a current passes through the coil (Fig.1.2), it produces a imaginary bar magnet. When a
soft-iron piece is brought near this coil it is magnetized. Depending upon the current direction
the poles are produced in such a way that there will be a force of attraction between the coil and
the soft iron piece. This principle is used in moving iron attraction type instrument.
Fig. 1.2
If two soft iron pieces are place near a current carrying coil there will be a force of repulsion
between the two soft iron pieces. This principle is utilized in the moving iron repulsion type
instrument.
When a current carrying coil is placed under the influence of magnetic field produced by a
permanent magnet and a force is produced between them. This principle is utilized in the moving
coil type instrument.
Fig. 1.3
When two current carrying coils are placed closer to each other there will be a force of repulsion
between them. If one coil is movable and other is fixed, the movable coil will move away from
the fixed one. This principle is utilized in electrodynamometer type instrument.
4
Fig. 1.4
To make the measurement indicated by the pointer definite (constant) a force is necessary which
will be acting in the opposite direction to the deflecting force. This force is known as controlling
force. A system which produces this force is known as a controlled system. When the external
signal to be measured by the instrument is removed, the pointer should return back to the zero
position. This is possibly due to the controlling force and the pointer will be indicating a steady
value when the deflecting torque is equal to controlling torque.
Td Tc (1.1)
Two springs are attached on either end of spindle (Fig. 1.5).The spindle is placed in jewelled
bearing, so that the frictional force between the pivot and spindle will be minimum. Two springs
are provided in opposite direction to compensate the temperature error. The spring is made of
phosphorous bronze.
When a current is supply, the pointer deflects due to rotation of the spindle. While spindle is
rotate, the spring attached with the spindle will oppose the movements of the pointer. The torque
produced by the spring is directly proportional to the pointer deflection .
TC (1.2)
The deflecting torque produced Td proportional to ‘I’. When TC Td , the pointer will come to a
steady position. Therefore
I (1.3)
5
Fig. 1.5
Since, and I are directly proportional to the scale of such instrument which uses spring
controlled is uniform.
1.6 Damping force
The deflection torque and controlling torque produced by systems are electro mechanical.
Due to inertia produced by this system, the pointer oscillates about it final steady position before
coming to rest. The time required to take the measurement is more. To damp out the oscillation
is quickly, a damping force is necessary. This force is produced by different systems.
The piston is mechanically connected to a spindle through the connecting rod (Fig. 1.6). The
pointer is fixed to the spindle moves over a calibrated dial. When the pointer oscillates in
clockwise direction, the piston goes inside and the cylinder gets compressed. The air pushes the
piston upwards and the pointer tends to move in anticlockwise direction.
6
Fig. 1.6
If the pointer oscillates in anticlockwise direction the piston moves away and the pressure of the
air inside cylinder gets reduced. The external pressure is more than that of the internal pressure.
Therefore the piston moves down wards. The pointer tends to move in clock wise direction.
An aluminum circular disc is fixed to the spindle (Fig. 1.6). This disc is made to move in the
magnetic field produced by a permanent magnet.
7
When the disc oscillates it cuts the magnetic flux produced by damping magnet. An emf is
induced in the circular disc by faradays law. Eddy currents are established in the disc since it has
several closed paths. By Lenz’s law, the current carrying disc produced a force in a direction
opposite to oscillating force. The damping force can be varied by varying the projection of the
magnet over the circular disc.
8
Fig. 1.7
Principle of operation
When D.C. supply is given to the moving coil, D.C. current flows through it. When the current
carrying coil is kept in the magnetic field, it experiences a force. This force produces a torque
and the former rotates. The pointer is attached with the spindle. When the former rotates, the
pointer moves over the calibrated scale. When the polarity is reversed a torque is produced in the
opposite direction. The mechanical stopper does not allow the deflection in the opposite
direction. Therefore the polarity should be maintained with PMMC instrument.
If A.C. is supplied, a reversing torque is produced. This cannot produce a continuous deflection.
Therefore this instrument cannot be used in A.C.
TC = controlling torque
= angle of deflection
K=spring constant
b=width of the coil
9
l=height of the coil or length of coil
N=No. of turns
I=current
B=Flux density
A=area of the coil
The force produced in the coil is given by
F BIL sin
(1.4)
When 90
For N
F NBIL (1.5)
turns,
Torque produced Td
F r distance (1.6)
Td NBIL b BINA
(1.7)
Td BANI
(1.8)
Td I
(1.9)
Advantages
Torque/weight is high
Power consumption is less
Scale is uniform
Damping is very effective
Since operating field is very strong, the effect of stray field is negligible
Range of instrument can be extended
Disadvantages
Use only for D.C.
Cost is high
Error is produced due to ageing effect of PMMC
Friction and temperature error are present
10
1.7.1 Extension of range of PMMC
instrument Case-I: Shunt
A low shunt resistance connected in parrel with the ammeter to extent the range of current. Large
current can be measured using low current rated ammeter by using a shunt.
Fig. 1.8
Le R =Resistance of meter
m
t
Rsh =Resistance of shunt
I= current to be measure
Vm Vsh (1.10)
ImRm IshRsh
Im R
sh (1.11)
Ish R
m
Eqn (1.12) ÷ by Im
I I
1 sh (1.13)
Im Im
11
I
1
Rm (1.14)
Im Rsh
Rm
(1.15)
I Im 1 R
sh
Rm
1 R is called multiplication factor
sh
Shunt resistance is made of manganin. This has least thermoelectric emf. The change is
resistance, due to change in temperature is negligible.
A large resistance is connected in series with voltmeter is called multiplier (Fig. 1.9). A large
voltage can be measured using a voltmeter of small rating with a multiplier.
Fig. 1.9
Let Rm =resistance of meter
Vm Vse
Rm (1.17)
Rse
V Rm
se Rse
Vm
12
(1.18
)
13
Apply KVL, V Vm
(1.19)
Vse
1 R
m m m
Rse
(1.21)
V Vm 1 R
m
Rse
1 R Multiplication factor
m
Principle of operation
The current to be measured is passed through the fixed coil. As the current is flow through the
fixed coil, a magnetic field is produced. By magnetic induction the moving iron gets magnetized.
The north pole of moving coil is attracted by the south pole of fixed coil. Thus the deflecting
force is produced due to force of attraction. Since the moving iron is attached with the spindle,
the spindle rotates and the pointer moves over the calibrated scale. But the force of attraction
depends on the current flowing through the coil.
Eqn (1.24) gives the energy is used in to two forms. Part of energy is stored in the inductance.
Remaining energy is converted in to mechanical energy which produces deflection.
Fig. 1.11
15
Change in energy stored=Final energy-initial energy stored
1 2 1 2
(L dL)(i di) Li
2 2
1 2 2 2
{(L dL)(i di 2idi) Li }
2
1 2
{(L dL)(i 2idi)
2
Li } 2
1 2 2 2
{Li 2Lidi i dL 2ididL Li }
2
1 2
{2Lidi i dL}
2
1 2
Lidi i dL (1.25)
2
Mechanical work to move the pointer by d
Td d (1.26)
By law of conservation of energy,
Electrical energy supplied=Increase in stored energy+ mechanical work done.
Input energy= Energy stored + Mechanical energy
2 1 2
Lidi i dL Lidi i dL Td d (1.27)
2
1 2
i dL Td d (1.28)
2
1 2 dL
T i (1.29)
d
2 d
At steady state condition
Td TC
1 2 dL (1.30)
i K
2 d
1 2 dL
i (1.31)
2K d
i2 (1.32)
When the instruments measure AC, i2rms
Scale of the instrument is non uniform.
16
Advantages
MI can be used in AC and DC
It is cheap
Supply is given to a fixed coil, not in moving coil.
Simple construction
Less friction error.
Disadvantages
It suffers from eddy current and hysteresis error
Scale is not uniform
It consumed more power
Calibration is different for AC and DC operation
1.8.2 Repulsion type moving iron instrument
Construction:The repulsion type instrument has a hollow fixed iron attached to it (Fig. 1.12).
The moving iron is connected to the spindle. The pointer is also attached to the spindle in
supported with jeweled bearing.
Principle of operation: When the current flows through the coil, a magnetic field is produced by
it. So both fixed iron and moving iron are magnetized with the same polarity, since they are kept
in the same magnetic field. Similar poles of fixed and moving iron get repelled. Thus the
deflecting torque is produced due to magnetic repulsion. Since moving iron is attached to
spindle, the spindle will move. So that pointer moves over the calibrated scale.
Damping: Air friction damping is used to reduce the oscillation.
Control: Spring control is used.
17
Fig. 1.12
Fig. 1.13
18
This instrument can be used for the measurement of voltage, current and power. The difference
between the PMMC and dynamometer type instrument is that the permanent magnet is replaced
by an electromagnet.
Construction:A fixed coil is divided in to two equal half. The moving coil is placed between the
two half of the fixed coil. Both the fixed and moving coils are air cored. So that the hysteresis
effect will be zero. The pointer is attached with the spindle. In a non metallic former the moving
coil is wounded.
Control: Spring control is used.
Damping: Air friction damping is used.
Principle of operation:
When the current flows through the fixed coil, it produced a magnetic field, whose flux density is
proportional to the current through the fixed coil. The moving coil is kept in between the fixed
coil. When the current passes through the moving coil, a magnetic field is produced by this coil.
The magnetic poles are produced in such a way that the torque produced on the moving coil
deflects the pointer over the calibrated scale. This instrument works on AC and DC. When AC
voltage is applied, alternating current flows through the fixed coil and moving coil. When the
current in the fixed coil reverses, the current in the moving coil also reverses. Torque remains in
the same direction. Since the current i1 and i2 reverse simultaneously. This is because the fixed
and moving coils are either connected in series or parallel.
Fig. 1.14
19
Let
L1=Self inductance of fixed coil
L2= Self inductance of moving coil
M=mutual inductance between fixed coil and moving coil
i1=current through fixed coil
i2=current through moving coil
Total inductance of system,
Ltotal L1 L2 2M
(1.33)
But we know that in case of M.I
1 2 d (L)
T i (1.34)
2
d
d
1 d
T i2
d
2 (L L 2M ) (1.35)
d 1 2
TC Td
(1.39)
i1i2 dM
(1.40)
K d
Hence the deflection of pointer is proportional to the current passing through fixed coil and
moving coil.
20
1.9.1 Extension of EMMC instrument
Case-I Ammeter connection
Fixed coil and moving coil are connected in parallel for ammeter connection. The coils are
designed such that the resistance of each branch is same.
Therefore
I1 I2 I
Fig. 1.15
To extend the range of current a shunt may be connected in parallel with the meter. The value
Rsh is designed such that equal current flows through moving coil and fixed coil.
Td dM
I1I2 (1.41)
d
Or Td
2 dM (1.42)
I d
T C K
(1.43)
2
I dM
(1.44)
K d
2
I (Scale is not uniform) (1.45)
Fixed coil and moving coil are connected in series for voltmeter connection. A multiplier may be
connected in series to extent the range of voltmeter.
21
Fig. 1.16
V1
I ,I V (1.46)
1 2 Z2
Z1 2
V V dM
Td 1 2 (1.47)
Z1 Z2 d
K V K V dM
Td 1 2 (1.48)
Z1 Z2 d
KV dM
Td 2 d (1.49)
Z1Z2
2
Td V (1.50)
Case-III As wattmeter
When the two coils are connected to parallel, the instrument can be used as a wattmeter. Fixed
coil is connected in series with the load. Moving coil is connected in parallel with the load. The
moving coil is known as voltage coil or pressure coil and fixed coil is known as current coil.
Fig. 1.17
22
Assume that the supply voltage is sinusoidal. If the impedance of the coil is neglected in
comparison with the resistance ‘R’. The current,
v sin wt
I2 m (1.52)
R
Let the phase difference between the currents I1 and I2 is
I1 Im sin(wt ) (1.53)
dM
Td I1I2 (1.54)
d
Vm sin wt
T
dM I sin(wt )
(1.55)
d m
R d
1 dM
T (I V sin wt sin(wt ))
(1.56)
d m m
R d
1
T I V sin wt.sin(wt )
dM (1.57)
d m m
R d
The average deflecting torque
1 2
(Td )avg Td d wt (1.58)
2 0
1 2
1 dM
(Td )avg
R I V sin wt.sin(wt )
m m d wt (1.59)
2 0
d
{cos cos(2wt )}dwt
V I 1
dM
(Td )avg m m
2 2 d
R (1.60)
d
2 2
d avg V I
4R dM
(T ) (1.61)
cos.dwt cos(2wt ).dwt
m m
0 0
(T )
Vm I m dM
cos wt
2
(1.62)
d avg
4R d 0
Vm I m dM
(T ) cos(2 0) (1.63)
23
d avg
4R d
Vm Im 1 dM
(T ) cos (1.64)
d avg
2 R d
1 dM
(Td )avg Vrms Irms cos
(1.65)
R d
24
(Td )avg KVI cos
(1.66)
TC
(1.67)
KVI cos
(1.68)
VI cos
(1.69)
Advantages
It can be used for voltmeter, ammeter and wattmeter
Hysteresis error is nill
Eddy current error is nill
Damping is effective
It can be measure correctively and accurately the rms value of the voltage
Disadvantages
Scale is not uniform
Power consumption is high(because of high resistance )
Cost is more
Error is produced due to frequency, temperature and stray field.
Torque/weight is low.(Because field strength is very low)
Errors in PMMC
The permanent magnet produced error due to ageing effect. By heat treatment, this error
can be eliminated.
The spring produces error due to ageing effect. By heat treating the spring the error can
be eliminated.
When the temperature changes, the resistance of the coil vary and the spring also
produces error in deflection. This error can be minimized by using a spring whose
temperature co-efficient is very low.
1.10 Difference between attraction and repulsion type instrument
An attraction type instrument will usually have a lower inductance, compare to repulsion type
instrument. But in other hand, repulsion type instruments are more suitable for economical
production in manufacture and nearly uniform scale is more easily obtained. They are therefore
much more common than attraction type.
25
1.11 Characteristics of meter
1.11.1 Full scale deflection current( IFSD )
The current required to bring the pointer to full-scale or extreme right side of the
instrument is called full scale deflection current. It must be as small as possible. Typical value is
between 2 A to 30mA.
1.11.2 Resistance of the coil( Rm )
This is ohmic resistance of the moving coil. It is due to , L and A. For an ammeter this should
be as small as possible.
1.11.3 Sensitivity of the meter(S)
1 Z
S (/ volt), S
IFSD V
It is also called ohms/volt rating of the instrument. Larger the sensitivity of an instrument, more
accurate is the instrument. It is measured in Ω/volt. When the sensitivity is high, the impedance
of meter is high. Hence it draws less current and loading affect is negligible. It is also defend as
one over full scale deflection current.
1.12 Error in M.I instrument
1.12.1 Temperature error
Due to temperature variation, the resistance of the coil varies. This affects the deflection of the
instrument. The coil should be made of manganin, so that the resistance is almost constant.
26
1.12.4 Stray field error
Since the operating field is weak, the effect of stray field is more. Due to this, error is produced
in deflection. This can be eliminated by shielding the parts of the instrument.
Fig. 1.18
Deflection of moving iron voltmeter depends upon the current through the coil. Therefore,
deflection for a given voltage will be less at higher frequency than at low frequency. A capacitor
is connected in parallel with multiplier resistance. The net reactance, ( X L XC ) is very small,
when compared to the series resistance. Thus the circuit impedance is made independent of
frequency. This is because of the circuit is almost resistive.
C 0.41 L
(1.72)
2
(RS )
1.13 Electrostatic instrument
In multi cellular construction several vans and quadrants are provided. The voltage is to be
measured is applied between the vanes and quadrant. The force of attraction between the vanes
27
and quadrant produces a deflecting torque. Controlling torque is produced by spring control. Air
friction damping is used.
The instrument is generally used for measuring medium and high voltage. The voltage is reduced
to low value by using capacitor potential divider. The force of attraction is proportional to the
square of the voltage.
Fig. 1.19
Torque develop by electrostatic instrument
V=Voltage applied between vane and quadrant
C=capacitance between vane and quadrant
1 2
Energy stored= CV
(1.73)
2
Let ‘ ’ be the deflection corresponding to a voltage V.
Let the voltage increases by dv, the corresponding deflection is’ d ’
When the voltage is being increased, a capacitive current flows
dq d (CV ) dC dV (1.74)
i V C
dt dt dt dt
V dt multiply on both side of equation (1.74)
28
Fig. 1.20
dC 2 dV
Vidt V dt CV dt (1.75)
dt dt
2
Vidt V dC CVdV (1.76)
1 2 1
Change in stored energy= (C dC)(V dV ) CV
2 (1.77)
2 2
1 2 dC (1.81)
K V
2 d
1 2 dC
V (1.82)
29
2K d
30
Advantages
It is used in both AC and DC.
There is no frequency error.
There is no hysteresis error.
There is no stray magnetic field error. Because the instrument works on electrostatic
principle.
It is used for high voltage
Power consumption is negligible.
Disadvantages
Scale is not uniform
Large in size
Cost is more
1.14 Multi range Ammeter
When the switch is connected to position (1), the supplied current I1
Fig. 1.21
Ish1Rsh1 Im Rm (1.83)
Rsh1 I m I m Rm (1.84)
Rm I I
1 m
Ish1
31
Rm
R ,R
,I m Multiplying power of shunt
1
Rm
sh I1 sh 1
1 1 1 m1 1 Im
Im
Rm I2
Rsh2 , m2 (1.85)
m2 Im
1 I3
R ,m (1.86)
Rm
sh3 3
m3 1 Im
Rsh4 Rm , I4 (1.87)
m4 m Im
1 4
R4 Rsh4 (1.91)
Fig. 1.22
Ayrton shunt is also called universal shunt. Ayrton shunt has more sections of resistance. Taps
are brought out from various points of the resistor. The variable points in the o/p can be
connected to any position. Various meters require different types of shunts. The Aryton shunt is
used in the lab, so that any value of resistance between minimum and maximum specified can be
32
used. It eliminates the possibility of having the meter in the circuit without a shunt.
33
1.16 Multi range D.C. voltmeter
Fig. 1.23
We can obtain different Voltage ranges by connecting different value of multiplier resistor in
series with the meter. The number of these resistors is equal to the number of ranges required.
34
Fig. 1.24
Consider for voltage V1, (R1 Rm )Im
V1
V1 m
V1 V1
(1.94)
V R
R 1 I Rm V Rm Rm
m ( m) m
Rm
R1 (m1 (1.95)
1)Rm
For V2 ,
(R2 R R )I m V2 (1.96)
I R Rm
1 m V2 R2 m 1
R2 VV2
1)Rm Rm (1.97)
(m
1
m
Rm
R2 m2Rm Rm (m1 1)Rm
R3 (m3 m2 )Rm
36
For V4 R4 R3 R2 R1 Rm Im V4
V
R 4
4 R R R Rm
Im 3 2 1
V4
V Rm (m3 m2 )Rm (m2 m1)Rm (m1 1)Rm Rm
m
R4 Rm m4 m3 m2 m2 m1 m1 1 1
R4 m4 m3 Rm
Example: 1.1
A PMMC ammeter has the following specification
Coil dimension are 1cm 1cm. Spring constant is 0.15 106 N m / rad , Flux density is
3
1.5 10 wb / m 2 .Determine the no. of turns required to produce a deflection of 900 when a
current 2mA flows through the coil.
Solution:
At steady state condition TC
Td
BANI K
K
N BAI
4 2
A=110 m
6 N
K= 0.15 10
m
rad
3 2
B=1.5 10 wb / m
3
I= 2 10 A
90 rad
2
N=785 ans.
Example: 1.2
37
The pointer of a moving coil instrument gives full scale deflection of 20mA. The potential
difference across the meter when carrying 20mA is 400mV.The instrument to be used is 200A
for full scale deflection. Find the shunt resistance required to achieve this, if the instrument to be
used as a voltmeter for full scale reading with 1000V. Find the series resistance to be connected
it?
Solution:
Case-1
Vm =400mV
Im 20mA
I=200A
V 400
Rm m 20
Im 20
Rm
I Im 1 R
sh
3 20
200 20 10
1
R
sh
3
Rsh 2 10
Case-II
V=1000V
Rse
V Vm 1 R
m
3 R
4000 400 10 1 se
20
Rse 49.98k
Example: 1.3
A 150 v moving iron voltmeter is intended for 50HZ, has a resistance of 3kΩ. Find the series
resistance required to extent the range of instrument to 300v. If the 300V instrument is used to
measure a d.c. voltage of 200V. Find the voltage across the meter?
Solution:
Rm 3k,Vm 150V ,V 300V
38
Rse
V Vm 1 R
m
R
300 se R 3k
1501 se
3
Rse
Case-II
V Vm 1 R
m
200 V 1 3
m
3
Vm
Ans
100V
Example: 1.4
What is the value of series resistance to be used to extent ‘0’to 200V range of 20,000Ω/volt
voltmeter to 0 to 2000 volt?
Solution:
Vse V V 1800
1 1
IFSD 20000 Sensitivity
Example: 1.5
A moving coil instrument whose resistance is 25Ω gives a full scale deflection with a current of
1mA. This instrument is to be used with a manganin shunt, to extent its range to 100mA.
Calculate the error caused by a 100C rise in temperature when:
(a) Copper moving coil is connected directly across the manganin shunt.
(b) A 75 ohm manganin resistance is used in series with the instrument moving coil.
The temperature co-efficient of copper is 0.004/0C and that of manganin is 0.000150/C.
Solution:
Case-1
Im 1mA
Rm
39
25
40
I=100mA
Rm
I Im 1 R
sh
25 25
100 11 R R 99
sh sh
25
Rsh 0.2525
99
Rm / t 26
10
R
sh / t10 0.2525(1 0.0001510) 0.2529
Current through the meter for 100mA in the main circuit for 100C rise in temperature
Rm
t 10 C
I Im 1 R
sh
100 I 1 26
mt
0.2529
Im 0.963mA
t 10
41
Resistance of the instrument circuit for 100C rise in temperature
Rm
t 10 25(1 0.004 10) 75(1 0.00015 10) 101.11
Shunt resistance for 100C rise in temperature
Rsh
t 10 1.01(1 0.00015 10) 1.0115
Rm
I Im 1
sh
R
100 I 101.11
1
m
1.0115
Im t 10 0.9905mA
Error =(0.9905-1)*100=-0.95%
Example: 1.6
The coil of a 600V M.I meter has an inductance of 1 henery. It gives correct reading at 50HZ and
requires 100mA. For its full scale deflection, what is % error in the meter when connected to
200V D.C. by comparing with 200V A.C?
Solution:
Vm 600V , Im 100mA
Case-I A.C.
Z 600
Vm 6000
0.1
m
Im
X L 2fL 314
Example: 1.7
A 250V M.I. voltmeter has coil resistance of 500Ω, coil inductance 0f 1.04 H and series
resistance of 2kΩ. The meter reads correctively at 250V D.C. What will be the value of
capacitance to be used for shunting the series resistance to make the meter read correctly at
50HZ? What is the reading of voltmeter on A.C. without capacitance?
Solution: L
C 0.41
2
(RS )
1.04
0.41 0.1F
3 2
(2 10 )
2520Ω 250
2520 248V
2500
Example: 1.8
The relationship between inductance of moving iron ammeter, the current and the position of
pointer is as follows:
Reading (A) 1.2 1.4 1.6 1.8
Deflection (degree) 36.5 49.5 61.5 74.5
Inductance ( H ) 575.2 576.5 577.8 578.8
Calculate the deflecting torque and the spring constant when the current is 1.5A?
Solution:
For current I=1.5A, =55.5 degree=0.96865 rad
43
dL 577.65 576.5
0.11H / deg ree 6.3H / rad
d 60 49.5
1 2 dL 1 2 6 6
Deflecting torque , Td I (1.5) 6.3 10 7.09 10 N m
2 d 2
Td 6 Nm
Spring constant, K 7.09 10 7.319 10 6
0.968 rad
Fig. 1.25
Example: 1.9
For a certain dynamometer ammeter the mutual inductance ‘M’ varies with deflection as
M 6 cos( 30 )mH .Find the deflecting torque produced by a direct current of 50mA
corresponding to a deflection of 600.
Solution:
T I I
dM 2 dM
d 1 2 I
d d
M 6 cos( 30 )
dM
6 sin( 30)mH
d
dM
60 6 sin 90 6mH / deg
d
2
T I 3 2 3 6
dM (50 10 ) 6 10 15 10 N m
d
d
44
Example: 1.10
The inductance of a moving iron ammeter with a full scale deflection of 900 at 1.5A, is given by
the expression L 200 40 4 2 3H , where is deflection in radian from the zero
position. Estimate the angular deflection of the pointer for a current of 1.0A.
Solution:
3 36.64 2 40 0
1.008rad,57.8
Example: 1.11
45
dL H
(5 2 )
d rad
1 2 dL
I
2K d
1 2
(5)
(5 2 106
6
2 12 10 )
1.69rad,96.8
Example: 1.12
The following figure gives the relation between deflection and inductance of a moving iron
instrument.
Deflection (degree) 20 30 40 50 60 70 80 90
Inductance ( H ) 335 345 355.5 366.5 376.5 385 391.2 396.5
Find the current and the torque to give a deflection of (a) 300 (b) 800 . Given that control spring
6
constant is 0.4 10 N m / deg ree
Solution:
1 2 dL
I
2K d
(a) For 30
The curve is linear
dL 355.5 335
1.075H / deg ree 58.7H / rad
d 30 40 20
46
Fig. 1.26
Example: 1.13
In an electrostatic voltmeter the full scale deflection is obtained when the moving plate turns
through 900. The torsional constant is 10 106 N m / rad . The relation between the angle of
deflection and capacitance between the fixed and moving plates is given by
Deflection (degree) 0 10 20 30 40 50 60 70 80 90
Capacitance (PF) 81.4 121 156 189.2 220 246 272 294 316 334
Find the voltage applied to the instrument when the deflection is 900?
Solution:
Fig. 1.27
47
dC bc 370 250
tan 1.82PF / deg ree 104.2PF / rad
d ab 110 44
6 Nm 6
Spring constant K 10 10 0.1745 10 N m / deg
ree rad
1 2 dC
V V 2K
2K d
dC
d
2 0.1745 106 90
V 104.2 1012 549volt
Example: 1.14
Design a multi range d.c. mille ammeter using a basic movement with an internal resistance
Rm
50
and a full scale deflection current Im 1mA . The ranges required are 0-10mA; 0-50mA;
48
500 1
49
A moving coil voltmeter with a resistance of 20Ω gives a full scale deflection of 1200, when a
potential difference of 100mV is applied across it. The moving coil has dimension of
30mm*25mm and is wounded with 100 turns. The control spring constant is
6
0.375 10 N m / deg ree. Find the flux density, in the air gap. Find also the diameter of
copper wire of coil winding if 30% of instrument resistance is due to coil winding. The specific
8
resistance for copper=1.7 10 m .
Solution:
Data given
Vm
100mV Rm
20
120
N=100
6
K 0.375 10 N m / deg ree
RC 30%ofRm
8
1.7 10 m
V
I m 3
m 5 10 A
Rm
6 6
Td BANI ,TC K 0.375 10 120 45 10 Nm
6
T 45 10 2
B d 0.12wb / m
6
ANI 30 25 10 100 5 10 3
RC 0.3 20 6
Length of mean turn path =2(a+b) =2(55)=110mm
l
RC N
A
8 3
N (lt 100 1.7 10 110 10
A
) 6
RC
50
8 2
3.11610 m
3 2
31.16 10 mm
2
A d d 0.2mm
4
Example: 1.16
A moving coil instrument gives a full scale deflection of 10mA, when the potential difference
across its terminal is 100mV. Calculate
(1) The shunt resistance for a full scale deflection corresponding to 100A
(2) The resistance for full scale reading with 1000V.
Calculate the power dissipation in each case?
Solution:
Data given
Im 10mA
Vm
100mV I
100 A
Rm
I Im 1 R
sh
3 10
100 10
10 1 R
sh
3
Rsh 1.00110
Rse ??,V 1000V
V 100
Rm m 10
Im 10
Rse
V V m 1 R
m
3 R
1000 100 10 1 se
10
Rse 99.99K
Example: 1.17
51
Design an Aryton shunt to provide an ammeter with current ranges of 1A,5A,10A and 20A. A
basic meter with an internal resistance of 50w and a full scale deflection current of 1mA is to be
used.
Solution: Data given
m1 I1 1000 A
Im
I1 1A
m2 I 2 5000 A
3 I2 5A Im
Im 110
A
Rm 50 I3 10 A I 10000 A
m 3
3
I4 20 A Im
m4 I4 20000 A
Im
Rsh 50
Rm 0.05
1 m 1000 1
1
1 50
R 0.01
Rm
sh2
m2 1 5000 1
50
Rsh3 Rm 0.005
m3 1 10000
1
Rsh4 50
Rm 0.0025
m 1 20000 1
4
52
53