Identification Instruction of Configuration (Monitor Board of Ventilator)

The document describes five methods for identifying different components of a ventilator monitor board: 1. There are two turbine models identified by labels on their cables. 2. Proportional valves have different voltages and labels that determine interchangeability. 3. Flow sensors have different voltages and wire materials and colors to identify models. 4. Voice coil motors connect differently and have different sockets depending on manufacturer. 5. Displays are identified by software version accessed through the menu or presence of a chip. Monitoring boards also differ in software version and tube connections.

Uploaded by

Jayson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
65 views7 pages

Identification Instruction of Configuration (Monitor Board of Ventilator)

The document describes five methods for identifying different components of a ventilator monitor board: 1. There are two turbine models identified by labels on their cables. 2. Proportional valves have different voltages and labels that determine interchangeability. 3. Flow sensors have different voltages and wire materials and colors to identify models. 4. Voice coil motors connect differently and have different sockets depending on manufacturer. 5. Displays are identified by software version accessed through the menu or presence of a chip. Monitoring boards also differ in software version and tube connections.

Uploaded by

Jayson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

Identification instruction of Configuration

(monitor board of Ventilator)


(一)Turbine identification method:There are two types of turbines,

U65H4 and C65MS1. have label on the turbine connection cable. Please

see the picture below.

(U65H4) (C65MS1)

(二)Identification method of proportional valve:The proportional

valve have special voltage supply , which is 7.2V and 9.6V respectively.

Model 0710,KENTUO, ASCO and Norgren proportional valves use same

voltage 7.2V,so they can be used interchangeably;. the voltage of 0712

proportional valve is 9.6V.please see the below picture.


(ASCO) (0710) (KENTUO) (0712)

Valve have label

Voltage supply of proportional valve test point(socket)

(二)Identification method of flow sensor:The flow sensor is divided

into two different models, one is FS1015CL, the other is SHENGSHIRUIEN,

the flow sensors have different voltage, The voltage of FS1015CL flow

sensor is 0.33v,SHENSHIRUIEN flow sensor is 0.66v.tsi flow sensor is


0.25v. the flow sensors’s connection wire have different material and

color composition, FS1015CL wire is plastic material,consists of black,

green, yellow, blue and red,5 wires, while the wire of shengshiruien and

tsi is coarse cloth material.

Flow sensor test point of voltage


(FS1015CL)(0.33v) (SHENGSHIRUIEN 0.66v)(tsi 0.25v)

(四)Identification method of voice coil motor:Voice coil motor is

divided into two types, one is TM voice coil motor, the other is

switzerland voice coil motor. The switzerland voice coil motor connection

line is FPC, and the TM voice coil motor connection line is composed of

black and red 2 lines. The connection line is very different. And the

socket is different,please see the picture below.

(switzerland voice coil motor) (TM voice coil motor)


(五)Identification method of display screen:The display is divided into

two models, one is A9, the other is 2440.Through the motherboard

software information can be distinguished. After turn on the machine,

enter Menu-System information to see the software information. V1.X.X

is A9 version mainboard, V3. X.X is 2440 version mainboard,please see

the picture below.

(A9) (2440)

The aother method is whether has U20 chip,version A9 monitor

board has U20 chip at the position shown in the figure, while the 2440

version monitor board doesn’t have U20 chip.


(A9) (2440)

Two different types of monitoring board software:


Don’t have small board,the tube connection is different,the software

version is V6.1 JUN 15 2020.

(small board)

Have small board,the software version is V6.2 Dec 22 2020.