Spatial Positioning: (We Know It's There, But Where Is It?)
Spatial Positioning: (We Know It's There, But Where Is It?)
S1 S2 Take Sj = S (1- Dj / D)
where S = S1+S2
PMT
Distance between Tubes
D = D1 + D2
NaI
S1 = S (1- D1 / D)
D1 D2
=(S1 + S2)D2 / D
Colli- Distance from Tube 2
mator D2 = S1 D / (S1 + S2)
Distance from Tube 1
g-Ray
D1 = S2 D / (S1 + S2)
More generally:
Colli-
mator Dc = 0.5D X+ – X-
X+ + X-
g-Ray
Want to use all PMTs in
array at once …
Spatial Positioning
S1 S2 S3 S4 S5
W+i 0 1 2 3 4
W-i 4 3 2 1 0
X = Wi Si
i
Spatial Positioning
S1 S2 S3 S4 S5
S1 = 0
S2 = 0
S3 = S
S4 = 0
S5 = 0
W+i 0 1 2 3 4
W-i 4 3 2 1 0
X = Wi Si X =k
2S - 2S
=0
i 2S + 2S
X+ -X-
X =k
X+ + X-
Spatial Positioning
S1 S2 S3 S4 S5
S1 = S
S2 = 0
S3 = 0
S4 = 0
S5 = 0
W+i 0 1 2 3 4
W-i 4 3 2 1 0
X = Wi Si 0 - 4S
X =k = -k
i
0S + 4S
X+ - X-
X =k + -
k = radius of array (cm)
X +X
Spatial Positioning
S1 S2 S3 S4 S5
S1 = 0
S2 = S
S3 = S
S4 = 0
S5 = 0
W+i 0 1 2 3 4
W-i 4 3 2 1 0
X = Wi Si X =k
(1S + 2S ) - (3S + 2S ) 1
=- k
i (1S + 2S ) + (3S + 2S ) 4
X+ -X-
X =k + - k = radius of array (cm)
X +X
Anger
Logic
X = k(X+ - X-)
(X+ + X-)
Y = k(Y+ - Y-)
(Y+ + Y-)
k = radius of
array (cm)
Energy (E)= X+ + X- = Y+ + Y- = Z
Simple
Anger
Circuit
X = k(X+ - X-)
(X+ + X-)
Y = k(Y+ - Y-)
(Y+ + Y-)
Ohm’s Law
k = radius of V=IR
array (cm)
Digital centroids
• Given as FWHM
• Typically a % of the photopeak
•10% for NaI at 140keV
•7% at 511 keV
• Goes roughly at 1 / sqrt(E)
• Integration time: better for low-count rate systems
•PET vs SPECT
•Degraded by anything that reduces sensitivity
Ideal Spectrum
Actual
spectrum
(blurred out due to
the 10% energy
resolution of the
camera)
Energy resolution is measured as the
FWHM of the measured photopeak
Septal Penetration
More important for higher-energy photons
Amplifier / Pulse shaper
Dead-time (Count-rate effects)
Time during which system does not count incoming signals
(radiation)
paralyzable → count rate goes to zero
non-paralyzable → count rate goes to constant
Baseline Restoration
Pulse-height Analyzer
l
Collimator Resolution
Collimator Resolution
Collimator Resolution
Collimators: Resolution vs Distance
LEHR
typically has
a resolution
of ∼7mm at
10cm
distance
Collimators: Geometric Efficiency
Hole shape
packing efficiency vs rotational symmetry
hexagonal, circular, square
Septal thickness
thicker septa means less holes
Higher resolution
lower acceptance angle so less of solid angle of photon flux
Efficiency is independent of
distance for parallel hole
collimator
Fig 16-14
Collimators: Geometric Efficiency
q r
2
Focused
Collimators:
Magnification
What collimators are used for what?
Correct
peak
Co-57 source
measured with
Tc-99m energy
window
Uniformity
CFOV UFOV
Typical NEMA Integral 2.5% 3.5%
values for Differential 2.0% 2.5%
Intrinsic Uniformity
Field of View
UFOV = Useful Field of View
Typically the entire FOV, may be 95% of the FOV
CFOV = Central Field of View
75% of the UFOV
UFOV
CFOV
Uniformity
Defective Photomultiplier
tube
Uniformity
Collimator
defect
Pin-cushion Barrel
Uncorrected Corrected
Point source too close to the detector
Uniformity
163
Ring Artifact
Fig 34. Transaxial image from the study shown in Fig 33. The ring
artifact (arrow) is now easily identified and is concentric with
the center of the image (cross-hair).
Spatial Resolution / Linearity
Camera’s ability to produce image detail and sharpness
Method – weekly
Similar to uniformity test with a resolution bar phantom
between camera detector and flood source
7mm 4mm
5mm
6mm
Non-Uniformity Artifacts
Pin-cushion Barrel
Linearity phantom,
1mm slits at 30mm spacing
Non-Uniformity/Linearity Artifacts
Corrections applied
Normal
flood-field
and bar
phantom
Rotate to assess
different directions
CFOV UFOV
Absolute 0.4 mm 0.6 mm
Differential 0.15 mm 0.2 mm
System Spatial Resolution
Annual
Collimator dependent
Quoted typically at 10cm: FWHM + FWTM
With and without scattering media (10cm in
front and 5cm behind)
Collect min 10k cnts in peak of each line
source
Use capillary tubes with <1mm int diameter
Energy Resolution
Point source suspended 5 x UFOV
Intrinsic
Convert spectrum for channels to energy
Scan Tc-99m and Co 57 to calibrate system
Resulting pulse-height spectrum is analyzed to
measure FWHM
Single
Photon Type of radioactive decay
Emission
Computed CT: 3D imaging
Tomography
Clinical Cameras Detector Heads
ECam
Infinia Hawkeye-4
NM 530c (CZT)
SPECT : Basic Process
Patient Camera
Radioactive
Sources
Camera
Collimator
Crystal (NaI)
Photomultiplier Tubes
Computer
Projections
Patient Camera
Sources
Sum all Data in Direction of Camera
Patient Camera
Sources
S
Planar (Static) Heart Imaging
camera
Ant
LAO
Potential Problems:
Cannot differentiate overlapping structures
Image contrast reduced
L.Lat
(LAO-70)
Modes of Acquisition
Static
Dynamic
ECG Gated
SPECT
MAG3 Renal Study
Dynamic
Planar
Images
ECG Gating – What is it?
ECG Signal
Gating – How?
ECG Signal
Gating – Issues
Ant
LAO
L.Lat
(LAO-70)
Phase Analysis
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Activity Near Heart Wall Boundary
Counts in voxel
voxel Time
ERNA: Phase TAC Curve
Pixel
Activity Near Heart Wall Boundary Time-Activity Curve
Counts in voxel
voxel Time
Phase Analysis
f (t ) = A0 + A1 cos(q (t ) + )
Gating – Why?
Phase Analysis
f (t ) = A0 + A1 cos(q (t ) + ) ROI
TAC
Pixel
Time-Activity Curve
92
Measurements of
46 Dissynchrony
SD, Bandwidth,
0 etc
0 90 180 270 36
0
Camera Rotates Around Patient
Patient Camera
Sources
S
Projections
Projections
Projections
Patient Motion
Projections = Set of 2D planar images
r
Sinograms
q
r
Sinograms Equivalent
SPECT
projection
q
r
All LORs
at same q
contribute
to the
same
projection
Small cylinder
Small cylinder projection
Small cylinder sinogram
Small cylinder sinogram
0 degrees rotation
Angular direction
Transaxial direction
Small cylinder sinogram
Angular direction
45 degrees rotation
Transaxial direction
Small cylinder sinogram
Angular direction
90 degrees rotation
Transaxial direction
Small cylinder sinogram
Angular direction
Transaxial direction
Basic Reconstruction: FBP
Ie
- m ( x ) dx
-log = -log
I
= m ( x)dx
SPECT is like CT
“sum of signal”
∴use FBP for reconstruction
Central Section Theorem
p. 396
Central Section Theorem
# spokes / (2p R )
p. 398
Applying the ramp filter
FBP 1
FBP 16
FBP 32
FBP 64
FBP 128
Compare BP and FBP
MCAT model
FBP 1
FBP 2
FBP 3
FBP 8
FBP 64
FBP 128