C70 Modular Monitor Product Specifications
C70 Modular Monitor Product Specifications
C70 Modular Monitor Product Specifications
Product Specifications
1) Monitor Type
Classified by Type
Electric shock protection
Externally powered Class-I equipment;
type
Defibrillation-proof
defibrillation resistant equipment with internal power supply.
applied parts
Electric shock protection Equipment with CF applied part (ECG, C.O. and IBP monitoring part) and
level BF applied parts (all other monitoring parts).
IP grade IPX1
Work mode Continuous operation equipment
2) Environmental Specifications
Item Specification
Ambient
5℃~40℃
temperature
Working
RH ≤93%
conditions
Barometric
700hPa~1060hPa
pressure
1 / 15
C70 Product specifications
Transport Please protect the monitor against violent impact, vibration, rain and snow in
conditions transport.
The monitor should be packed and stored in a well-ventilated room without corrosive gas
Storage conditions (ambient temperature: -20 ℃ ~60 ℃ ; RH: ≤93%; Barometric pressure:
700hPa~1060hPa).
3) Power Supply
Item Specification
AC input voltage 100~240V
AC input frequency 50Hz/60Hz
Power supply Powered either by built-in battery or external AC.
Input power 60VA
11.1V 4400mAh rechargeable lithium-ion battery, supplying power for at least 4
Built-in battery
consecutive hours in normal use once fully charged.
Charge time 11.1V 4400mAh: At least 4 hours from depletion to 90% charge in normal use.
Defibrillation output +5V defibrillation synchronization signal during 100ms
Synchronization Max delay: ≤35ms; Pulse width: 100ms±10; Up/Down time≤1ms
Analog Output Bandwidth: 0.5-40Hz; Max delay: ≤35ms; error: ±5%;
4) General Specifications
Item Specification
Dimension About 300mm×280mm×160mm
Weight About 4kg (not include battery and modules)
LCD specification Pixel: 800×600
5) ECG Specifications
Item Specification
Applicable standards: IEC 60601-2-27.
12-lead (R, L, F, N, C1, C2, C3, C4, C5, C6 or RA, LA, LL, RL, V1, V2, V3, V4,
Lead mode
V5, V6)
Method of leads I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform 2 -channel
Lead mode 5-lead (R, L, F, N, C; or RA, LA, LL, RL, V)
Method of leads I, II, III, aVR, aVL, aVF, V
Waveform 2 -channel
Lead mode 3-lead ( R, L, F; or RA, LA, LL)
Method of leads I, II, III
Waveform 1-channel
Load 1V, power frequency, differential-mode AC voltage for 10s without damage
Overload protection
(p-v)
Resp, lead
AC waveform:
disconnection
Current :<0.1µA;
detection and active
Frequency 64kHz, ±10%
noise control
Automatically identify the type of ECG leads.
C70 Product specifications
stability reset
Drift rate in 10s 10µV/s
Baseline drift in 1h ≤500µV
Baseline drift at
≤50µV/℃
working temperature
Amplitude: ±2mV~±700mV; width: 0.1ms~2.0ms; if overshoot < 0.05ɑ p, settling
Non-overshoot
time < 5µs; start time, end time, rise time and fall time of pulse: ≤ 100µs; start time
pacemaker pulse
of pulse: 40ms or earlier before the start time of QRS wave; there is an identical
inhibition
pulse 150ms~250ms before the above pacemaker pulse.
Inhibition of
pacemaker pulse
Minimum input slew rate: 320mV/s RTI
detector on quick
ECG signals
Amplitude: ±2MV~±700MV; width:
Pacemaker pulse 0.5MS~2MS; maximum rise time:
≥0.2mV
display capability 100µS; the ECG display when the
pacemaker pulse appears at 100/min.
Measurement range -2.0mV-+2.0mV (-20.0 mm~+20.0 mm)
ST segment
-0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater.
measurement Measurement error
Other ranges: not defined.
ST alarm limit range -2.0mV-+2.0mV
ST alarm limit error ±0.1mV
Resolution 0.01mV (0.1mm)
Asystole, ventricular fibrillation (VFIB)/ventricular tachycardia (VTAC),
PVCs/min, R on T, VT>2, couplet, PVC, bigeminy, trigeminy, tachycardia
(TACHY),bradycardia (BRADY), supraventricular tachycardia (SVT), extreme
Arrhythmia types tachycardia, extreme bradycardia, missed beats, multiform PVC (multi. PVCs),
VTAC, nonsustained VT (nonsus. VTAC), ventricular rhythm, heart pause,
pause/min, irregular rhythm (irr. rhythm), ventricular bradycardia, atrial fibrillation,
pacemaker not captured (PNC), pacemaker not paced (PNP).
Leakage current <10uA
Electrosurgical
interference HR change caused by interference: ≤±10%
inhibition
Cut mode: 300W
ESU protection Condense mode: 100W
Recovery time: ≤10s
HR Calculation
Tall T-wave rejection
1.2mV
capability
If all of the last 3 RR intervals are longer than 1200ms, the average of the last 4
HR calculation RR intervals is the HR. In other cases, the average of the last 12 RR intervals (with
the longest interval and shortest interval excluded) is the HR.
Cardiotachometer the HR is displayed as follows after the 20s stable segment:
C70 Product specifications
6) Resp Specifications
Item Specification
Method Thoracic impedance method
Adult 0rpm~120rpm
RR alarm limit range and
Neonate/c
error 0rpm~150rpm
hild
Error ±1rpm
Adult: 10s~60s
No breath alarm limit Range
Child/neonate: 10s~20s
range and error
Error ±5s
No breath alarm delay 10s、15s、20s、25s、30s、35s、40s、45s、50s、55s、1min、Off
The Monitor will display the relevant alarm message when the HR is identical
CVA identification
with the RR.
7) SpO2 Specifications
Item Specification
Display range 0%~100%
Display resolution 1%
Data averaging and other 2s
signal processing time
Data update time 8s
Comen SpO2: measurement range: 0%~100%; measurement accuracy:
±2% (adult/child, in non-motion state) or ±3% (neonate, in non-motion
Measurement accuracy state) within the measurement range of 70%~100%. The measurement
accuracy within the measurement range of 1%~69% is not defined.
Masimo SpO2: measurement range: 1%~100%; measurement accuracy:
C70 Product specifications
8) PR Specifications
Item Specification
Comen SpO2 sensor:
Measurement range: 20bpm~254bpm; resolution: 1bpm; measurement
error: ±2bpm.
Masimo SpO2 sensor:
Measurement range: 25bpm~240bpm; resolution: 1bpm; measurement
error: ±3bpm (in non-motion state) or ±5bpm (in motion state).
Nellcor SpO2 sensor:
Measurement range: 20bpm~300bpm; resolution: 1bpm; measurement
Measurement range and
error: ±3bpm within 20bpm~250bpm range. The measurement accuracy
accuracy
within 251bpm~300bpm range is not defined.
NIBP sensor:
Measurement range: 40bpm~240bpm; resolution: 1bpm; measurement
error: ±3bpm or ±3%, whichever is greater.
IBP sensor:
Measurement range: 20bpm~350bpm; resolution: 1bpm; measurement
error within 20bpm~350bpm range: ±1bpm or ±1%, whichever is greater
(excluding the sensor error).
Comen SpO2 sensor:
PR alarm limit range and
Alarm range: 20bpm~254bpm;
resolution
Upper limit: (lower limit +1)bpm~254bpm ;
C70 Product specifications
9) Temp Specifications
Item Specification
Measurement range and Measurement range 0℃~50℃
accuracy Measurement error ±0.1C
Alarm limit range 0℃~50.0℃
Temp alarm limit range
Alarm limit
and resolution ±0.1℃
resolution
Resolution 0.1C
Number of channels 2
Operating mode Direct mode
Transient response No greater than 40 seconds
range
CO2 alarm
±0.1kPa or ±1mmHg ±0.1kPa or ±1mmHg
resolution
awRR
measurement 0~150rpm 0~150rpm
range
awRR
measurement ±1rpm ±1rpm
accuracy
awRR alarm
0rpm~150rpm 0rpm~150rpm
limit range
awRR alarm
1rmp 1rmp
resolution
Adult:10s、15s、20s、25s、30s、35s、
No breath alarm Range 40s、45s、50s、55s、1min
limit range and
Child/neonate: 20s、25s、30s、35s、40s
error
Error ±5s
No breathalarm
10s、15s、20s、25s、30s、35s、40s、45s、50s、55s、1min、Off
delay
Respironics/Nmed/Palconn EtCO2
Respironics/Nmed/Palconn EtCO2 sensor (mainstream)
sensor (sidestream)
CO2 0~150mmHg 0~150mmHg
measurement 0%~19.7% 0%~19.7%
range (0~20.0kPa) (0~20.0kPa)
0~69mmHg: 0.1mmHg 0~69mmHg: 0.1mmHg
CO2 resolution
70~150mmHg: 0.25mmHg 70~150mmHg: 0.25mmHg
0~40mmHg: ±2mmHg 0~40mmHg: ±2mmHg
41~70mmHg: ±5%×reading 41~70mmHg: ±5%×reading
CO2 accuracy
71~100mmHg: ±8%×reading 71~100mmHg: ±8%×reading
101~150mmHg: ±10%×reading 101~150mmHg: ±10%×reading
CO2 alarm limit
0~150mmHg 0~150mmHg
range
CO2 alarm
±0.1kPa or ±1mmHg ±0.1kPa or ±1mmHg
resolution
awRR
measurement 0~150rpm 0~150rpm
range
awRR
measurement ±1rpm ±1rpm
accuracy
awRR alarm
0~150rpm 0~150rpm
limit range
awRR alarm
1rmp 1rmp
resolution
No breath alarm Range Adult:10s、15s、20s、25s、30s、35s、
C70 Product specifications
14) AG Specifications
Item Specification
The AG module complies with ISO 80601-2-55.
AG measurement method Infrared radiation absorption characteristics
The following accuracy standards are applicable to a dry gas under 22±5°C and
1013±40hPa.
0%~15% ±(0.2kPa+reading×2%)
CO2
15%~25% Not defined.
EtHal/EtEnf/EtIs
0%~25% ±0.1%
o/EtSev/EtDes
FiHal/FiEnf/FiIs
0%~25% ±0.1%
o/FiSev/FiDes
Item Specification
Specifications of ISATM (AG) Sidestream Gas Analyzer
Measurement
Infrared gas measurement
method
For adult: 10s, 15s, 20s, 25s, 30s, 35s, 40s ,45s, 50s, 55s or 1min;
[No Breaths Range For pediatric and neonate: 20s, 25s, 30s, 35s or 40s
Timeout]
Error ±5s
[No Breath
10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 1min or Off
Alm Delay]
General Specifications
Ultra-small low-flow sidestream gas analyzer, with integrated micro-pump, zeroing valve
Description
and flow controller.
Working
ISA AX+: 0~50°C (32~122°F); ISA OR+: 5~50°C (41~122°F)
conditions
Storage
-40~70°C (-40~158°F)
conditions
RH <4kPa H2O (non-condensing) 95% RH, 30°C
Barometric
52.5~120kPa (4572m)
pressure
Water
Sampling tube: patented dehydration tube
treatment
Data Output
Fi/Et value CO2, O2, N2O, five anesthetic gases (Hal, Enf, Iso, Sev, Des)
Waveform Display up to 4 gas concentration waveforms at a time
Diagnostic
Barometric pressure
parameter
Gas Analyzer
ISA sensor 2~9-channel NDIR gas analyzer (measurement range: 4~10µm)
Compensation CO2 broadening effect
No calibration is required. The Monitor will auto perform zeroing when powered on and
Calibration
perform auto zeroing every 24h (ISACO2) or 8h (ISA AX+/OR+) subsequently.
Preheating time ISA CO2: <10s; ISA OR+/AX+: < 20s
Gas
Rise time CO2: ≤250ms; N2O: ≤ 350ms; anesthetic gases: ≤ 350ms; O2: ≤450ms
Overall system <3s (2m sampling tube)
C70 Product specifications
response time
Respiration
Self-adaptive threshold (minimum CO2 concentration change: 1 vol%)
detection
RR 0-150 breaths/min
Anesthetic gas Threshold of main anesthetic gases (ISA OR+/AX+): 0.15 vol%. The concentration of
threshold any identified anesthetic gas will be reported, even if it is lower than 0.15 vol%.
Number of waveforms 3