Comparison of Different Number of Beams in Intensity Modulated Radiotherapy in Head and Neck Cancer
Comparison of Different Number of Beams in Intensity Modulated Radiotherapy in Head and Neck Cancer
Comparison of Different Number of Beams in Intensity Modulated Radiotherapy in Head and Neck Cancer
ISSN No:-24562165
Walla Taman
Radiotherapy Department, Ayadi Al Mostakbal Oncology Center, Alexandria, Egypt.
Abstract:- Results
The total results showed that, there was significant
Purpose difference between 5, 7 and 9 beams IMRT in term of
The aim of the work was to determine the best mean values for PTV95% coverage were 96.76, 97.51
beams number and segments in order to improve the and 98.22 respectively with p = 0.005. The conformation
plans conformity and homogeneity that generate low mean values were1.60, 1.49 and 1.34 with p = 0.007. HI
monitor units (MUs) and faster irradiated time for values for the PTV were 0.14 ± 0.05, 0.13 ± 0.05 and 0.12
different types of head and neck cancer (HNC). ± 0.04 with p = 0.001. Right parotid were 21.96, 20.72
and 20.43 with p = 0.003. Left parotid were22.14, 21.04
Methods and 20.70 with p = 0.100. Spinal cord 45.34, 44.51 and
This study includes 30 patients with different HNC. 43.23 with p = 0.003. Brain stem were 49.52, 49.77 and
Intensity modulated radiotherapy (IMRT) treatment 48.74 with p = 0.058. Number of segments were 79.85,
planning techniques were done with step and shoot 106.55 and 131.80 with p = 0.001. MUs were 23879.8,
delivery technique, 5, 7 and 9 beams IMRT were carried 24252.6and 22501.8with p = 0.003and the total irradiated
out for each patient. The treatment plans for all patients time were 79.60, 80.84 and 75.0 respectively with p =
were calculated and optimized using fast superposition 0.003. In fact that, the plan quality improved with an
algorithm. All plans were generated using equal spaced increasing the number of intensity modulated beams.
odd beam number around the target. 6 MV were used in
all beams. Multiple segments were created for each Conclusions
beam. Typically maximum iteration was carried out to From this study we can conclude that, the 9 beams
achieve optimized plans. The beam weight optimized to IMRT is superior to techniques using less number of
generate the plan, then the segment weight optimized for beams (5 and 7) where, the 9 beams IMRT significantly
all plans by using sliding window methods. The final improved the PTV coverage, dose distribution,
optimization maps were converted into a way of step and conformity, homogeneity to the PTV with better sparing
shoot sequence map which delivered by linear OARs and reduce the dose to surrounding normal
accelerator using multi leaf collimator (MLC). IMRT tissues. Moreover, the 9 beams significantly reduced the
plans were compared based on several criteria: Isodose mean MUs and pure irradiated time compared with 5
distributions, the mean and standard deviation with p- and 7 beams IMRT.
values for planning target volume (PTV) 95%,
conformity index (CI), homogeneity index (HI), organs Keyword:- IMRT, Radiotherapy Techniques, Beam Number,
at risk (OARs), number of segments, MUs and total Treatment Planning.
irradiated time were presented and compared in all
patients. Statically analyses were compared for all
patients used ANOVA testes.
IJISRT22DEC1494 www.ijisrt.com 94
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
I. INTRODUCTION the OARs. The main disadvantage is increase the treatment
delivery times and MUs (18). This lead to patient discomfort,
The goal of radiotherapy is to deliver a homogenous reduce the machine output and increase the dose to the
dose of radiation to tumor, while delivering a dose as low as surrounding healthy tissues around the PTV which arise
possible to healthy surrounding tissues (1). Conventional from the MLC transmission and scatter radiation from the
three-dimensional conformal radiotherapy (3DCRT) linear accelerator, these doses proportional to the number of
delivers a homogenous dose to tumor volume with MUs. These scatter radiation can increase the risk of
acceptable low dose to normal structures(2) . However, in secondary malignances (19, 20). Reduction of irradiated time
some tumor sites with concave shape such as in HNC, limits can be achieved by using different numbers of beams or
the ability of conventional radiotherapy to shape the dose to segments or by using high modern delivery techniques such
the target volumes and to spare the OARs (3). as VMAT (8).
Significant advances in imaging technology resulted in The treatment planning optimization system helps to
more precise localization of the tumor and critical organs in determine the distribution of the beam intensity which
three-dimensional (3D).These developments have been across the treatment volumes(21). The optimization explores
mainly driven by the need to reduce the dose to normal these possibilities to find the optimum intensity maps that
tissues. To that end, newer IMRT have been developed (4, 5). are matches the dose and volumes constraints with
IMRT is an advanced form of high precision of 3DCRT, objectives for PTV and OARs using system priorities (3).
which use linear accelerator to deliver precise radiation dose The different plans can be evaluated and compared to select
to tumor (6, 7). IMRT allows to deliver radiation dose to the optimum intensity modulation. The optimum pattern
conform more precisely to 3D concave shape tumor by then converted to a complex sequences of beams segment
(16)
modulating the intensity of radiation beam in multiple .
segments which minimizes the dose to surrounding healthy
tissues (5, 8). Typically, combination radiation beams II. PATIENTS AND METHODS
intensity modulated fields coming from different beam
directions produce precise shape of radiation dose (9). IMRT Thirty HNC patients were enrolled in this study.
techniques for treatment of HNC replaced conventional Seventy percent nasopharynx, 10 % hypopharnx, 6 % neck
3DCRT, which resulted in much better dose conformity, lymph nodes, 10 % tongue and 4 % check cancer. The
sparing of OARs and less radiation toxicity (10). patients included 13 males and 17 females with a mean age
of 39 years (range 16 - 76 years). The patients were at
IMRT includes forward and inverse planning. In the different stages I, II and III with exclusion metastatic IV
forward planning, the planner selects the planning stage. Patients were recruited from Ayadi Al-Mostakbal
parameters, the computer then calculates the dose Oncology Center, Alexandria, Egypt.
distribution and the plans are optimized by the manual
iteration. The inverse planning begins by defining the Immobilization and Computed Tomography
prescription dose to the targets volumes with clinical Simulation
objectives then the planning system algorithm determine the Computed tomography (CT) simulator images were
beams parameters which results dose distribution for the obtained in the supine position with a head and neck
targets and the system undergo thousands of iterations to support. Patients were immobilized using thermoplastic
find the best solution for the treatment plans (11, 12). Inverse mask. CT images were done for all patients with thickness 2
IMRT for HNC is complex due to the large number of mm, using SomAtom Emotion Duo Computed tomography,
OARs locates near to the PTV (13) , so the correct selection of Siemens.
the beam number and direction in HNC IMRT improve the
planning target volume (PTV) coverage as well as sparing IMRT Target Volumes and OARs Delineation
the OARs (14). In the contouring, the CT slices of selected patients
were transferred to focal pro computer system by DICOM
The IMRT can delivered by three delivery techniques: network, where outlining of the target volumes and OARs
step and shoot IMRT, dynamic IMRT and intensity were done according to the RTOG guidelines.
modulated arc therapy (IMAT) with tomotherapy or
volumetric arc therapy (VMAT) (3, 9, 15). The step and shoot Planning Objectives
is most commonly available in cancer treatment centers. In For all HNC patients, the treatment goal was to
this delivery techniques, the beams divided into different delivers the prescribed dose to achieve minimum dose more
segments and the radiation is turn off between the segments. than 95% of the prescribed dose and maximum lower than
The MLC shape the first segments then the radiation turn on 107% for the primary target volumes PTV. In all IMRT
to delivers into the segment, the radiation then turn off to plans, the objectives were applied to achieve minimum
allow the MLC move to create the next segment. In the doses to OARs without compromising the PTV coverage.
dynamic the radiation delivered as the leaves are moving (3). The mean dose to the right and left parotid was aimed to be
below 26 Gy, maximum dose allowed for spinal cord was 45
The IMRT process consist of multiple steps for Gy and for brain stem 54 Gy.
treatment planning until delivery of radiation (17). IMRT is
more conformity for irregular targets and reduce the dose to
IJISRT22DEC1494 www.ijisrt.com 95
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
IMRT Planning Techniques assess the conformation of the dose to the PTV. The ideal
IMRT treatment planning techniques were done on conformation is 1 so, the greater conformity than 1 indicates
Xio computerized treatment planning system. Inverse IMRT greater healthy tissues irradiated around the PTV and the CI
planning were used with step and shoot IMRT delivery defined as CI = VTV/VPTV (25, 26). OARs sparing were
technique using ARTISTE Linear Accelerator, Siemens, compared in term of maximum and mean received doses for
with modulator MLC include 160 leafs with 0.5 cm all plans. Dose volume histograms (DVHs) were used to
thickness to deliver the treatment. The IMRT planning assess and compare the different coverage of the PTV and
parameters were defined and selected manually. For each the doses received to OARs. The number of segments was
patient 5, 7 and 9 beams IMRT plans were carried out. compared to assess the plan efficiency for all IMRT
plans. The MUs and pure irradiation time were calculated
The treatment plans for all patients were calculated and compared as important parameters for all patients. The
and optimized using fast superposition algorithm to generate pure irradiated time defined as MU/D where MU is the
beam modulation as a specific objectives and constraints. monitor units and the D is the dose rate (Mu/min) (27).
All IMRT plans were generated using equal spaced odd Statically analyses were compared for all patients used
beam number around the target to avoided opposing beams. ANOVA testes.
In all IMRT beams 6 MV were used. The gantry angle
started from 180 in all cases as the following: The gantry III. RESULTS
angles in 5 beams were 40, 110, 180, 250, 320, in 7 beams
were 30, 80,130, 180, 230, 280, 320 and in 9 beams were The total results including the mean and standard
20, 60, 100, 140, 180, 220, 260, 300, 340 for all IMRT HNC deviation with p-values for PTV95%, CI, HI, right and left
techniques. All plans were normalized to the PTV to achieve parotids, spinal cord , brain stem, DVHs, number of
coverage of the PTV by at least 95 % of the prescribed dose. segments, MUs and total irradiated time were presented and
Objectives were generated priority as the following: PTV, compared in all HNC patients.
right and left parotids, spinal cord and brain stem. The
IMRT dose prescription and constraints for each PTV and PTV95%:
OARs were adjusted to achieve results as the planning Most IMRT plans for HNC were considered
goals. acceptable in term of PTV95% coverage of the prescribed
dose in this study except some plans when using 5 beams
With delivery method step and shoot inverse IMRT IMRT. In general, the plans were acceptable if the 95% of
treatments planning each optimization were started with the isodose surface covers the 100% of the PTV. In fact, the
generation the flunce map. Multiple segments were created larger number of intensity modulated beams, the better PTV
for each beam. Typically maximum iteration was carried out coverage by 95% of the prescribed dose. As present in
to achieve optimized plans. The segmentation parameters Table (1) and Fig. (1) the means and standard deviations
were specified using sliding window methods with discrete comparison between 5, 7 and 9 beams IMRT according
intensity levels10, minimum segment size 2 cm to reduce to PTV95% coverage were 96.76 ± 2.17, 97.51 ± 2.15 and
the number of segments, in the segment weight optimization 98.22 ± 1.61 respectively. Statically, there were significantly
gird spacing 0.3 cm with minimum segment MU 5 were difference between 5, 7 and 9 beams in mean PTV95%
used. The beam weight optimized to generate the plan, then coverage by the prescribed dose p < 0.05. Results show that,
the segment weight optimized for all IMRT HNC plans by the 9 beams for HNC was superior in PTV95% coverage of
using sliding window methods. In the optimization process, the prescribed dose compared with 5 and 7beam IMRT.
the optimal maps were calculated according to priorities,
constraints and objectives. The final optimization maps were Table 1 Comparison between 5, 7 and 9 Beams IMRT
converted into a way of step and shoot sequence map which According to PTV
delivered by linear accelerator using MLC. The plans were PTV95%
5 Beams 7 Beams 9 Beams p
transferred to the LANTIS system for verification and (%)
approval by medical physicist and radiation oncologist and Min. – 91.80 – 91.60 – 95.15 –
were clinically considered acceptable. Max. 99.93 99.88 99.99
Mean ± 96.76 ± 97.51 ± 98.22 ±
IMRT Plans Comparison 0.005*
SD. 2.17 2.15 1.61
For each patents 5, 7 and 9 beams IMRT plans were Median 96.74 97.75 98.57
carried out and compared. IMRT plans were compared *: Statistically significant at p ≤ 0.05
based on several criteria:
Table 2 Comparison between 5, 7 and 9 Beams IMRT
Isodose distributions were compared visually on According to CI
different images slices with respect the degree of conformity CI (No) 5 Beams 7 Beams 9 Beams p
of the prescribed dose to the PTV. The PTV volumes Min. – 1.18 – 1.17 – 0.84 –
(PTV95%) receiving 95% of the prescribed dose were Max. 2.93 2.89 1.79
compared. The HI was used to assess the dose uniformity Mean ± 1.60 ± 1.49 ± 1.34 ±
with the PTV. The ideal values of HI is 0 so, greater value 0.007*
SD. 0.43 0.38 0.24
indicate heterogeneity inside the PTV (24) and is defined as Median 1.39 1.34 1.36
HI = D2% -D98% / D50%(25). The CI also calculated to
IJISRT22DEC1494 www.ijisrt.com 96
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
DVH:
The DVHs were calculated for all PTV and OARs in
all IMRT HNC patients. Fig. (3) shows the DVHs
comparison between 5, 7 and 9 beams IMRT according to
PTV coverage and OARs received doses include right and
left parotid, spinal cord and brain stem as example HNC
case. The DVHs were not sufficient to evaluate the dose
distribution for PTV, so the conformity and homogeneity
were considered. DVHs showed higher doses received by
the right and left parotids in 5 beams compared with 7 and 9
beams. The 9 beams were superior in PTV coverage of
prescribed dose and OARs sparring compared with 5 and 7
beams IMRT.
Isodose Distributions:
Clinically, the dose distribution in most IMRT plans
for HNC was acceptable except few cases in 5 beams. The
dose distribution comparison between 5, 7 and 9 beams
IMRT as shows in Fig. (2), a typical isodose distribution
was superior in 9 beams compared with 5 and 7 beams. The
isodose lines were showed comparable between the different
beams. Some areas received high doses outside the PTV and
more normal tissues irradiated in 5 beams were observed.
The dose is more conformed to the PTV as well as the dose
reduced to the surrounded healthy tissues in 9 beams
compared with 5 and 7 beams IMRT. In fact, the increasing
in number of beams lead to reduce the doses received to the
normal tissues and the dose were more conformed with
more homogenous doses to the PTV.
IJISRT22DEC1494 www.ijisrt.com 97
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
CI of PTV: Table 3 Comparison between 5, 7 and 9 Beams IMRT
The means and standard deviations comparison According to HI
between 5, 7 and 9 beams IMRT according to CI of the PTV HI (No) 5 Beams 7 Beams 9 Beams p
were showed in Table (2) and Fig. (4). The conformation Min. – 0.06 – 0.07 – 0.06 –
values were1.60 ± 0.43, 1.49 ± 0.38 and 1.34 ± 0.24 Max 0.24 0.26 0.20
respectively. The 5 beams was the worst conformation Mean ± 0.14 ± 0.13 ± 0.12 ±
compared with 7 and 9 beams, in fact that, due to the beam 0.001*
SD. 0.05 0.05 0.04
number not sufficient to conform the radiation dose exactly Median 0.15 0.13 0.13
to the PTV. The conformation improve was observed when
the number of the beams increased from 7 to 9 beams
IMRT. There were significant differences between 7 and 9
beams when compared according to conformity degree, with
p < 0.05. The 9 beams were superior in conformation the
dose to the PTV with reduced the radiation received to the
healthy surrounding tissues for HNC IMRT treatment.
Right Parotid:
The comparison values between 5, 7 and 9 beams
IMRT for right parotid in all HNC patients were listed in
Table (4) and Fig. (6). The means and standard deviations
for mean dose to the right parotid were 21.96 ± 2.30, 20.72
± 3.41and 20.43 ± 2.19 respectively. The mean dose to the
Fig 4 Shows the means and standard deviations comparison right parotid was aimed to be below the 26 Gy to preserve
between 5, 7 and 9 IMRT beams according to CI of PTV for the parotid function and reduce the xerostomia. Most
all HNC patients. techniques were meet the requirements criteria except the 5
beams, due to hot areas were found surrounding the PTV, in
HI of PTV: fact that related to the number of beams not conform the
The dose uniformity inside the PTV was evaluated by dose exactly to the PTV and the radiation received to the
the HI. The means and standard deviations values normal surrounding tissues. There were significant
comparison between 5, 7 and 9 beams IMRT for all HNC differences between 5, 7 and 9 beams were observed. The
patients were presented in Table (3) and Fig. (5). The HI significant lower mean doses to the right parotid were
values for the PTV were 0.14 ± 0.05, 0.13 ± 0.05 and 0.12 ± achieved by using 9 beams compared with 5 and 7 beams
0.04 respectively. There were significant differences in the IMRT with the p < 0.05.
dose homogeneity inside the PTV when compared the
different beams with p < 0.05. The HI was indicated that, Left Parotid:
the dose homogeneity was inferior in 5 beams. The results The effect of beams number also evaluated by
showed that when increased the beam number from 7 to 9 comparison the doses received by the left parotid. The
beams, the dose uniformity within the PTV were improved means and standard deviations comparison between 5, 7 and
compared with 5 beams. The typical dose uniformity inside 9 beams for mean dose to the left parotid were 22.14 ± 2.72,
the PTV was achieved by the 9 beams IMRT. 21.04 ± 3.89 and 20.70 ± 2.77 respectively. As mentioned
before when compared the means and standard deviations of
mean dose received by the right parotid, when compared
different beams IMRT, the results were close to it in term of
sparing. The left parotid sparing was improved when the
number of beams increased. The results showed that, there
were significant differences with p < 0.05 as presented in
Table (4) and Fig. (6).
IJISRT22DEC1494 www.ijisrt.com 98
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
Table 4 Comparison between 5, 7 and 9 Beams IMRT Table 5 Comparison between 5, 7 and 9 Beams IMRT
According to Right and Left Parotids According to Spinal Cord
Parotids Spinal
5 Beams 7 Beams 9 Beams p Cord 5 Beams 7 Beams 9 Beams p
(Gy)
Right (Gy)
Min. – 17.93 – 12.54 – 16.85 – Min. – 38.62 – 38.65 – 36.34 –
Max. 25.32 26.40 25.04 Max. 49.57 50.18 46.22
Mean ± 21.96 ± 20.72 ± 20.43 ± Mean ± 45.34 ± 44.51 ± 43.23 ±
0.003* 0.003*
SD. 2.30 3.41 2.19 SD. 3.15 3.31 2.37
Median 21.67 20.68 20.30 Median 45.46 44.44 43.73
Left
Min. – 16.05 – 10.16 – 12.93 –
Max. 25.76 26.23 25.41
Mean ± 22.14 ± 21.04 ± 20.70 ±
0.100
SD. 2.72 3.89 2.77
Median 22.58 21.62 21.24
Brain Stem:
In brain stem, the maximum dose was aimed to be
below 54 Gy to protect it from radiation. The means and
Fig 6 Shows the means and standard deviations comparison
standard deviations comparison between 5, 7 and 9 beams
between 5, 7 and 9 IMRT beams according to right and left
IMRT were showed in Table (6) and Fig. (8) for all
parotids for all HNC patients.
patients. The values of means and standard deviations were
49.52 ± 7.27, 49.77 ± 5.78 and 48.74 ± 6.41 respectively.
Spinal Cord:
The results showed significant difference with p < 0.05. The
The relationship between the mean maximum dose to
dose reduced to the brain stem in 9 beams compared with 5
spinal cord and the number of the beams for all patients
and 7 beams for all HNC treated by IMRT.
were showed in Table (5) and Fig. (7). The means and
standard deviations comparison between 5, 7 and 9 beams
Table 6 Comparison between 5, 7 and 9 Beams IMRT
IMRT were 45.34 ± 3.15, 44.51 ± 3.31 and 43.23 ± 2.37
According to Brain Stem
respectively. In this study, the maximum dose allowed for
Brain
spinal cord was below 45 Gy. In all patients, most plans 5 Beams 7 Beams 9 Beams p
Stem (Gy)
were meet the constraints criteria except some cases with 5
beams. The means maximum dose to spinal cord was Min. – 30.67 – 38.96 – 31.39 –
showed a significant difference between different IMRT Max. 57.93 57.12 55.48
plans with p < 0.05. In this study, the number of the beams Mean ± 49.52 ± 49.77 ± 48.74 ±
0.058
have significant effect on the maximum dose to the spinal SD. 7.27 5.78 6.41
cord, so when increased the number of the beams, the Median 51.70 52.56 51.46
maximum dose to spinal cord reduced. The spinal cord
received low doses in 9 beams compared with 5 and 7
beams IMRT.
IJISRT22DEC1494 www.ijisrt.com 99
Special Issue-(2nd ICTSA-2022) International Journal of Innovative Science and Research Technology
ISSN No:-24562165
MUs:
The means and standard deviations comparison
between 5, 7 and 9 beams IMRT according to total MUs for
all HNC patients were showed in Table (8) and Fig. (10).
The values of means and the standard deviations were
23879.8±4308.1, 24252.6±4891.9 and 22501.8±3566.97
respectively. In this study, the MUs in all plans were not
restricted by the constraints. In few cases with 5 beams, the
MUs were lowest with fewest numbers of segments but the
PTV coverage was worst and the surrounding normal tissues
received high dose. On the other hand, the longest MUs
were observed with 7 beams IMRT in all cases. Statically,
there were significant differences between the different
Fig 8 Shows the means and standard deviations comparison IMRT beams with p < 0.05. The lowest MUs were achieved
between 5, 7 and 9 IMRT beams according to brain with 9 beams as well as the number of segments were
Stem for all HNC patients. highest with better PTV coverage and the dose received by
the normal surrounding tissues were reduced in most cases
Number of Segments: compared with 5 and 7 beams IMRT.
All IMRT beams with different segments were
compared for all HNC patients. The means and standard Table 8 Comparison between 5, 7 and 9 Beams IMRT
deviations comparison between 5, 7 and 9 beams IMRT According to MUs
showed in Table (7) and Fig. (9). The values of means and
standard deviations were 79.85 ± 13.74, 106.55 ± 16.72 and MUs 5 Beams 7 Beams 9 Beams p
131.80 ± 24.62 respectively. In this study, the quality of the Min.
14759.3– 15972.3– 15209.6–
plans was associated with the number of segments and –
30017.1 32650.1 27365.4
irradiated times. When the number of segments decreases, Max.
the quality of the plans was worse in PTV coverage and the Mean 23879.8±43 24252.6±48 22501.8±356 0.00
doses received by the surrounding normal tissues were high. ± SD. 08.1 91.9 6.97 3*
In most patients the results related to number of segments Medi
24225.36 25270.35 23553.32
were acceptable except some plans with 5 beams. The an
results showed that, there were significant difference
between 5, 7 and 9 beams with p < 0.05. Moreover, when
the number of beams increased the number of segment
increased as well as the dose was improved to the PTV and
the irradiated time was reduced.
Currently, the beam number selection in IMRT is In general, when increase the number of the beams in
depended on the experience of treatment planners, many static step and shoot IMRT the MUs were decreased (37, 38, 40)
researchers have attempted to automate the beam selection . The reduced in number of MUs should lead to less leakage
in IMRT. Moreover, such as in complex concave situations, of radiation from the collimator head. This reductions lead
it is difficult to decide a suitable number of beams without to less peripheral doses (41, 42). The dose to the healthy
trails, errors and compromise in dose distribution and tissues around the PTV arise from the collimator