Installation Guide
INSTALLING CARDIOPERFECT INSTALLING CARDIOPERFECT
WORKSTATION SOFTWARE & WORKSTATION SOFTWARE
ACCESSORIES ON A SINGLE PC
1 Install the CardioPerfect software.
For software version 1.6.7 or later
Caution Do not plug in any USB devices
For network installation, please refer to the Network Installation
during this installation process.
Manual or contact Welch Allyn technical support.
This guide shows how to install the software and accessories for the
Welch Allyn CardioPerfect Workstation on a single laptop or desktop
PC. This guide does not cover installation for networked CardioPerfect
configurations, for EMR-connected computers, or thin-client settings.
System Requirements
You need the following minimum recommended configuration to work
with the CardioPerfect resting ECG, ambulatory blood pressure, or
spirometry software modules:
Computer Tower or desktop PC preferred;
laptop or tablet.
This screen might appear.
Operating System Windows 7 Home Premium, Professional, or
Ultimate; Windows 10
CPU P4 (or equivalent) or higher
Hard disk 20 GB with 5 GB free space
RAM 512 MB; 1 GB or higher recommended
Ports 2 USB ports
If this screen does not appear, explore the disc, and
Monitor size 15" or larger double-click installer ( ).
Monitor resolution 1024 x 768 true color;
1280 x 1024 resolution recommended
Other DVD-ROM drive (next page)
2
Select OK.
(next page)
Close CardioPerfect.
CardioPerfect disc.
NEXT STEP: Install and verify all devices that you have purchased: ECG recorder, spirometer, and/or ABPM 7100
device. When finished, please see the user information provided with your product.
4
RA
R
V1
C1
V2
V3
C3
E1
E1
E2
E2
E3
E3
E4
E4
V4
C4
V6
C6
LA
L
LL
F
a
b
(next page)
(a) Click Go.
(b) Select your test patient.
(c) Click ECG.
5
b
I NSTALLING THE PRO ECG RECORDER
ON A CARDIOPERFECT
WORKSTATION
a Before you start, install the CardioPerfect Workstation software.
1 A
ssemble the ECG recorder, and then plug it into
any USB port.
The system takes a 10-second recording and 2 Verify the installation of the ECG recorder.
returns to the main screen with your recording
displayed. You have now verified that your ECG
recorder is properly installed.
Welch Allyn
CardioPerfect
(next page)
6
a
b
You will see square waves along with this message: (a) Wait at least 20 seconds for the square waves to
“Bad signal on Left Leg.” level off. When the lines become straight, verify that
there are no lead-off messages.
Example lead-off message:
If you receive a lead-off message, reconnect the
indicated leads, and wait 20 seconds again. Repeat
until there are no lead-off messages.
(b) Click Record.
Straighten a large metal paper clip. The system takes a 10-second recording and
returns to the main screen with your recording
Note Do not using a plastic-coated or painted paper clip. displayed. You have now verified that your ECG
recorder is properly installed.
7
aution
C With the spirometer, always use the USB
extension cable, as shown. This cable helps
to prevent damage to the spirometer.
Welch Allyn The lot code and calibration code appear on the
CardioPerfect flow transducer box:
(next page)
8
Follow the instructions on this screen. (a) Verify that you see a flow loop. Each flow loop looks
slightly different.
(b) Click Done. You have verified that your spirometer is
communicating with the Workstation.
..3.. ..2.. ..1..
Push and then pull the plunger in sync with the side-
to-side motion of the blue bar. Move steadily, all the
way in and then all the way out, without hesitating
during the stroke.
9
WORKSTATION
Before you start, install the CardioPerfect Workstation software.
(next page)
(a)
Right-click on (My) Computer or This PC.
10
g b
c
(f) Double-click Ports (COM & LPT).
(g)
Make a note of the number associated with the selected
COM port. You will need this number in the next steps.
a
Close the Device Manager window.
Welch Allyn
d
CardioPerfect (a) Click this icon.
(b) Click this tab.
(c) Select the COM number appropriate for your device.
Use the number you previously made a note of in Step 2, (g).
(d) Click OK.
You can leave the main screen open.
c
a
b
(a) Click Go.
(b) Select your test patient.
(c) Click ABP.
(next page)
11
a
b
(a) Select device.
(b) Click OK.
Disconnect the cable from the device. The device is ready to take
a BP measurement.
Align the cuff tubing connector to the device. Click the tubing
connector onto the device fitting to secure.
You can give the device any name that you find helpful.
When the time reappears on the screen, Press and hold the On/
Off button for two seconds to turn the device off. Remove the cuff
from your arm.
Press On/Off to turn the device on. “CO” appears on the device
screen indicating the device is connected.
(next page)
12