The controller provides complete access to all advanced features of the Ultima® X Gas Monitor through its full-
function keypad. The controller is used for basic calibration functions along with displaying minimum,
maximum and average gas values, displaying date of last calibration, setting span-gas values and alarm
values, setting the real time clock, and changing address. It is completely password-protected.
Information Display
Information Retrieval Display All Information
• Ready • Ready
• Send • Send
• Display • Hold display button until display all
• Use +/- to scroll to desired item • Aim
(min, max, avg, zero, span value, alarm values, etc) • Press ‘1’
• Aim
• Enter
Settings
Setting Span Values Setting Alarm Values
• Ready • Ready
• Send • Send
• Span • Alarm
• Use +/- for gas list (i.e. 500 ppm CO) • 0 – set point • 1 – On/Off
• Enter • Use +/- for gas type • Use +/- for gas type
• Put in number (i.e. 300) • Enter • Enter
• Aim • Point # • Point #
• Enter • Enter • Enter
• Point value • (Alarm prompt) or OR
Set Date or Time • Aim • 1 – Enable 0 – Disable
• Enter • Latched / non-latch
• Press and hold Date or Time
• Prompt appears
• Put in new date or time Reset Latched Alarms
• Enter Send any command and it will reset a latched alarm once it is
below the alarm value.
Send Date or Time
• Send
Send Average Time Interval
• Aim • Ready
• Date or Time • Send
• Average
• Use +/- for 1, 8, 24 hours
• Aim
• Enter
Programs
You can create Programs 1–9 for repetitive items. (See Manual)
www.MSAgasdetection.com
Notes Troubleshooting Tips
Regular Calibration Point Controller at detector, CAL FAULT • You did not pass calibration – Do over
press “CAL” button, A. May have used wrong gas
zero and span is automatic B. Adjustment limits exceeded – Do “I Cal”
Initial Calibration Press and Hold the SPAN I CAL • Overrides factory adjustments
button – 3 seconds, • Hold “SPAN” button down until
press 1 for yes, zero and span prompt asked for “I Cal”
is now automatic, • Aim
with 30 second countdowns. • Press “1”
To abort span cal hit, zero/cal/or address button.
SENSOR RESET • Press “SEND” twice
Set Time Press and hold the TIME button • Prompt asked for reset
Set Date Press and Hold the DATE button • Aim
• Press “0” or “1”
Battery Life “READY” displayed in capital
letters indicates strong batteries CALIBRATION ABORT • You can press “CAL” or “ZERO” button
during countdown and it will
“ready” displayed in lower
abort calibration
case indicates weak batteries
Default Controller Password 9999
• Current Gas Value • Current Date • Ultima X Span Value • Sensor Swap Delay
• Minimum Gas Reading • Future Cal Date • Ultima X Range • Alert Option
• Maximum Gas Reading • Previous Cal Date • Ultima X Gas Table • 4-20
• Average Gas Reading • Sensor Address • Cal Sig Enable
• Zero Value • Cal Interval • Days per Auto Cal
• Span Value • Cal Sig Status • Start Hr of Cal
• Alarm 1 Setpoint • Range Value • Next Cal Date
• Alarm 2 Setpoint • Alert Option
• Alarm 3 Setpoint • Gas Table Value
• Current Time • Swap Delay Option
• Future Cal Time
• Average Time
Note: This Bulletin contains only a general MSA – The Safety Company
description of the products shown. While uses 1000 Cranberry Woods Drive
and performance capabilities are described, Cranberry Township, PA 16066 USA
under no circumstances shall the products be Phone 724-776-8600
used by untrained or unqualified individuals www.MSAsafety.com
and not until the product instructions U.S. Customer Service Center
including any warnings or cautions provided Phone 1-800-MSA-INST
have been thoroughly read and understood. Fax 1-800-967-0398
Only they contain the complete
MSA Canada
and detailed information Phone 1-800-672-2222
concerning proper use and Fax 1-800-967-0398
care of these products.
MSA Mexico
Phone 01 800 672 7222
ID 0720-159-MC / January 2015 Fax 52 - 44 2227 3943
© MSA 2015 Printed in U.S.A.
MSA International
Phone 724-776-8626
Toll Free 1-800-672-7777
Fax
Email
724-741-1553
[email protected]
www.MSAsafety.com