CV-5000S Manual
CV-5000S Manual
(Ver.4)
TOPCON CORPORATION
July 2011
Index
1. Basic configuration................................................................................................................................................2
2. Mounting the CV-5000S Phoropter head...........................................................................................................4
3. Connection of head/power supply and controller .............................................................................................5
4. Installation of PC software (setup 3)...................................................................................................................7
5. Connection with chart .........................................................................................................................................14
6. Connection with refraction instruments............................................................................................................28
7. Output subjective data to PC.............................................................................................................................44
8. Network system 1................................................................................................................................................52
9. Network system 2................................................................................................................................................55
10. Network system 3..............................................................................................................................................59
11. Network system 4..............................................................................................................................................64
12. Network system 5..............................................................................................................................................69
13. Network system 6..............................................................................................................................................72
14. Details of setting menu.....................................................................................................................................77
15. Customizing pictures and movies (PC-50S only).........................................................................................78
16. Export and import settings with USB memory ..............................................................................................80
17. Update KB-50S..................................................................................................................................................81
1
1. Basic configuration
SETUP 1
(KB-50S)
SETUP 2
(MONITOR & MOUSE)
SETUP 3
(PC)
2
SETUP 2 Basic components
1. 3.
2.
5.
4.
1.
2. 3. 4.
4
3. Connection of head/power supply and controller
3-1. SETUP 1
Connect power supply box, measuring head, KB-50S and power cord.
KB-50S USB/DC
interface cable.
5
3-2. SETUP 2
Connect power supply box, measuring head, analogue monitor, mouse, and power cord.
VGA cable
3-3. SETUP 3
Connect power supply box, measuring head, PC and power cord.
RS-232C cable
Power cord
6
4. Installation of PC software (setup 3)
4-1. PC requirements
Windows XP Professional SP3
Windows 7 Professional 32bit
Windows 7 Professional 64bit
* In case of Internet Explorer Ver.8, install Microsoft security update program “KB2482017”.
At least one serial port is necessary (to connect power supply box and PC), and make sure that COM port is
recognized.
7
4-2.5. Click [Start Setup]
8
4-2.7. Click [Start Setup]
9
4-2.9. Click [Start Setup]
10
4-3. Initial settings
4-3.1. Power up the power supply box and start [CV-5000PC] software.
* Settings can be changed later, this process is necessary for start the software first time.
11
4-3.4. Select [serial port for CV power supply unit].
Select in the pull down menu, as there are COM numbers which are recognized are listed.
12
4-5. Printer setting
SETUP 3 does not have the built-in printer, thus please arrange a printer for PC if required.
The following is recommended to print setting to have report much clearer to see with highlights on title.
13
5. Connection with chart
General information about cables
Wireless DVI
EZ AIR USB/DVI wireless doc
http://www.ezairwireless.com/
424120005 IR emittor
Chart lineups
14
General layout of CV-5000S system and chart device
SETUP 1 or 2
(1)
SETUP 3
(2)
PC-50S
Exclusively in use with CV-5000S
(1) EZAir wireless DVI No remote controller is available
(2) 424280001 DVI (10m)
SETUP 1, 2, or 3
ACP-8 series
(1)
PC-50
(2)
CC-100 series
15
5-1. PC-50S for SETUP 1 and 2.
Please see Instruction Manual PC-50S for how to install PC-50S.
5-1.1. Cabling
Connect DVI cable to DVI socket of power supply.
DVI cable
16
(2) Select “For far vision”,
Device = PC-50S
Type = Chart A, B, E, F,H
(at your convenience)
(2)’ (Especially in SETUP2) Touch the mark, so that you can see the
monitor for control ( #: plug and play monitor ) and #: PC-50S are
recognized.
(PC-50S displayÆ)
17
5-1.3. Settings in PC-50S.
(1)
(SETUP1) Press [Shift]&[Menu] button, and select
[Settings]_[Chart Settings]
18
(4) Select each item.
*Examination distance(2.5m-6.0m)
*slideshow interval (3-10 secs)
*Size of VA indicator
*Masked chart layout
*mirrored chart
*Language (no need to set)
(5) Touch OK
DVI cable
19
(2) Click [Settings] and select [Chart Settings]
Device = PC-50S
Type = Chart A, B, E, F,H
(at your convenience)
Touch the mark, so that you can see # : Plug and Play Monitor
# : Plug and Play Monitor is recognized.
(PC-50S displayÆ)
20
5-2.3. PC-50S
See 5-1.3
5-3.1. Connect USB transmitter to USB socket on control box (SETUP 1/2), KB-50S or on PC (SETUP3).
*Install USB driver for EZ Air in case of SETUP3. Control box (SETUP 1/2) is pre-installed with the driver for
EZ Air.
DVI cable
21
5-4. PC-50 / CC-100 series /ACP-8 / MC-3 (RS-232C)
Please see instruction manual for how to set up ACP-8/MC-3/PC-50.
5-4.1. Connect the chart to ACP/MC port of power supply with serial cable.
For ACP/MC-3 432140006 (RS232C Din8-Din8)
For PC-50 424290061 (RS232C Din8 Dsub9)
(SETUP 1/2)
(SETUP 3)
22
5-4.2. Settings in CV-5000S
(1)
(SETUP1) press [Shift]&[Menu], and
select [Settings]_[Chart Settings]
23
(3)[1. Settings]
24
1. Language (not necessary to change)
Selectable from “Japanese” and “English”. Default is “English”.
2. Chart type
Select chart type from “A” “B” “F” “H1”. Default is “F”. this must be the same with the selected chart in CV
system..
3. Remote type
Select type of controller from “Non LCD” “LCD” “KB-50”.
If setting is different in PC-50 and controller, it looks working normal but some of function might not work.
25
5-5. PC-50 /CC-100 series /ACP-8 / MC-3 (IR emitter)
Please see instruction manual for how to set up ACP-8/MC-3/PC-50.
5-5.1. Cabling
Connect IR emitter to IR port of power supply.
IR emittor 424120005
(SETUP 1/2)
(SETUP3)
26
5-5.2. Settings in CV-5000S
(1) (SETUP1) press [Shift]&[Menu], and
select [Settings]_[Chart Settings]
27
6. Connection with refraction instruments
CL(Rx)
STD1 format data
Ref (AR)
data
Xml format
9 Data from refraction units are transferred in STD1 format by serial communication and saved in xml format
inside shared folder or default folder of CV-5000S system.
9 KR-1, furthermore, is capable of transferring data directly in xml format by LAN network.
9 Only representative value of Ref / Kerato data are saved.
Control box SETUP1/2 has 3 ports for serial data transfer communication.
COM2 Æ
COM1 Æ
Å COM3
Dsub 9 pins male
Din8 female
SETUP 3 has no port for refraction unit, connect the unit directly to PC.
28
General information about cables
418120002 RS-232C 5m
(DIN 8 pins male L-shape – D SUB 9 pins female)
424290061 RS-232C 10m
(DIN 8 pins male L-shape – D SUB 9 pins female)
432140006 RS-232C 5m
(DIN 8 pins male L – DIN 8 pins male L)
432140010 RS-232C 10m
(DIN 8 pins male L – DIN 8 pins male L)
RS232C-USB converter cable
(D SUB 9 pins male – USB (A))
Please arrange locally if necessary.
29
(3) Select [ON-LINE]_[BAUD-RATE]_[9600].
Select [EXIT] to confirm and close.
30
6-1.3. CV-5000S setting
(1) (Setup 1) Press [Shift] & [Menu] button to access the setting screen.
31
(5) Select [Serial configuration] .
32
6-2. KR-1 (RS-232C)
6-2.1. KR-1 setting.
(1) Power on KR-1.
(2) Select [Menu]_[Comm]_[Output data]_[All]
Select [Menu]_[Comm]_[STD1]
Select [Menu]_[Baud Rate]_[9600].
See 6-1.3
33
6-3. KR-1 (LAN)
6-3.1. Settings in KR-1
(1) Power up KR-1 and select [Menu]_[LAN]
(2) set each item as follows.
LAN connection ON
XML file output ON
Ref / Kerato data output OFF
Shared folder configuration
¥¥IP address of shared CV
system¥RM
e.g. ¥¥192.168.128.152¥rm
• Shared folder
Default “TOPCON”
Or user name which is otherwise assigned to access shared
• User name
folder.
*please do not tell this information to the customer
Default “CV-5000S” (all capital)
Or password which is otherwise assigned to access the shared
• Password
folder.
*please do not tell this information to the customer
IP properties Fix.
Assign KR-1 IP address to access network with CV-5000S
IP address
e.g. 192.168.128.153
34
6-3.2. Cabling
LAN (straight) left LAN port to hub or PC
35
(3) Select [LAN configuration]
(4) Tick in “Enable network connection”
4.’ In case of SETUP 3, IP address is retrieved automatically from the Windows network setting.
If there is not any, please input the IP address, Subnet mask, Default gateway in the following.
(Windows XP)
[Control Panels]_[Network] _ [Local Area Connection] _[properties]
(Windows7)
[Control Panels] – [Network and Internet] – [Networks and Sharing Center]
– [Local Area Connection]
36
IP address should be identical with the one input for step 6-3.1
37
6-4. TRK-1P (RS-232C)
6-4.1. Settings in TRK-1P
(1) Power up TRK-1P with holding 2nd button from the right
to enter [INITIAL MENU].
38
Select [ON-LINE]_[BAUD RATE]_[9600]
39
(3) [EXIT]
40
6-5.3. Settings in CV-5000S
(1) (Setup 1) Press [Shift] & [Menu] button to access the setting screen.
41
(5) Select [Serial configuration] .
*To connect the NIDEK lens meter, select “[In] LensMeter data receive”.
(7) Touch OK
42
(3) Select [MENU]_[RS-232C]_[OUTPUT]
Then select [EXIT] to confirm
43
7. Output subjective data to PC
General information about cables
418120002 RS-232C 5m
(DIN 8 pins male L-shape – D SUB 9 pins female)
424290061 RS-232C 10m
(DIN 8 pins male L-shape – D SUB 9 pins female)
RS-232C cross
(D SUB 9 pins female - D SUB 9 pins female)
Please arrange locally
7-1. IMAGEnetR4
STD1 format
RS232C cable MDR
44
7-1.1. Cabling between CV-5000S and PC
418120002 (CV COM3 to PC serial)
D SUB-DSUB (cross) (CV COM1/2 to PC serial)
45
(4) Select [Serial configuration] .
(6) Click OK
(7) (SETUP1) Go to [Function Settings]_[Print Switch]_Action,
Please select [Data export] or [Printout & data export]
(8) Touch OK
7-1.3. IMAGEnetR4 and MDR
(1)Install Prerequisites
*Prerequisites must be Ver.4.0.
(2)Install IMAGEnet
* IMAGEnet R4 must be later than Ver.4.13.
(3)Install “Multi Data Receiver” (MDR).
46
7-1.4. Settings in MDR
(1)Required file
File Detail
CV5000-STD1-InPlugin.dll Interface file for receiving CV-5000 (RS-232C) data
47
(6) in [Setup] dialog, input [COM PORT] and click [OK]
48
(10) in [Out plugin], tick in [Show Register Window].
(SETUP2) Click [Data Export] icon. (*register this icon on tool bar)
49
(4) See MDR windows received data.
(date & time only)
50
7-2. I-base
7-3.2. Format can be selectable among “.xml file”, “export file Ver.1.00” and “export file ver.2.14”.
51
8. Network system 1
To add a new CV-5000S (SETUP 1/2/3) to existing CV-5000 system and RM/CL
8-1. Data flow
¾ RM/KR/CL data will be sent to CV-5000/CV-5000S in STD1 format.
¾ CV-5000 and CV-5000S received same data list sent via RM/KR/CL.
¾ If RM/KR/CL has patient ID capability, patient ID will be sent to CV-5000 as well.
¾ Subjective data will not be saved.
¾ Shared folder setting is not required. In case of one, non-networked CV-5000S is used, data folder
inside CV-5000S is automatically used.
¾ It is possible to set the period to delete and refresh the RM/CL data saved in shared folder.
Shared
folder (AR)
(3)
(4)
RS-232C
RS-232C
Shared Shared
folder (Rx) folder (AR)
8-2. Cabling
(1) 432140005 DIN8 DIN8 (RM/CL to CV RM )
(2) 418120002 Din8-DSUB9(CV COM to CV COM1/2)
432140005 DIN8 DIN8 (CV COM to CV COM3) * cannot be used for SETUP3
(3) 432140005 DIN8 DIN8 (RM/CL to CV COM3 )
418120002 Din8-DSUB9 (RM/CL to CV COM1/2)
(4) 432140005 DIN8 DIN8 (CV COM3 to CV RM)
418120002 Din8-DSUB9(CV COM1/2 to CV RM)
52
8-3. Settings
8-3.1. RM/KR/CL Æ (STD1).
Please see RM/KR(6-1.1), KR-1(6-2.1),TRK-1P(6-4.1), CL-200(6-5.1), EZ-200(6-6.1).
DIP switch 2 1 ON
DIP switch 2 2 ON
DIP switch 2 5 OFF
DIP switch 2 6 OFF
For the other connection, please see the CV-5000 DIP/SW setting manual.
(3)[Serial config.]
53
(4)In any of blank connection, touch the [▼] button and select each item.
“for” : “[In] AR/Rx data receive”.
Port : Port number which (old) CV-5000 is connected.
BaudRate : 2400bps or 9600 bps.
(same baud rate selected for step 8-3.2)
3. [Serial config.]
54
3. In any of blank connection, touch the [▼] button and select each item.
“for” : “[Out] RM/CL data transfer”.
Port : Port number which (old) CV-5000 is connected.
BaudRate : 2400bps or 9600 bps.
9. Network system 2
To have several CV-5000S (System 1 or 2) and CL. KR-1
9-1. Data flow
RM/KR/CL CV-5000S(1) CV-5000S (2)
(1)
RS-232C
STD1
¾ Shared folder is pre-installed in CV-5000S system. In the above picture, supposing shared
folders are the one preinstalled CV-5000S(2).
¾ As far as they are networked, Ref/CL data saved in shared folder can be retrieved any time.
¾ It is possible to set the period to delete and refresh the RM/CL data saved in shared folder.
¾ Subjective data is not saved, they will be printed out or discarded. (since next version, subjective
data will be export in .xml format.
55
¾ If patient ID is input in KR/RM, patient ID will be output together with data.
9-2. Cabling
(1) 432140005 DIN8 DIN8 (RM/CL to CV COM3 ) or
418120002 DIN8-Dsub9 Din8-DSUB9(CL to CV COM1/2)
(2) LAN
9-3. Settings
9-3.1. RM/KR/CL Æ (RS-232C setting)
They should output data in STD1 format.
Please see RM/KR(6-1.1), KR-1(6-2.1),TRK-1P(6-4.1), CL-200(6-5.1), EZ-200(6-6.1).
3. [Serial config.]
4. In any of blank connection, touch the [▼] button and select each item.
56
6. [LAN Configuration]
7. Tick in [Enable Network connection] and input IP address of CV-5000S(1)
9. [Data Folder]
10. if RM/KR is used,
Tick in [use shared folder in CV system] in AR data folder configuration.
Input IP address of CV-5000(2)
Tick in [use shared folder in CV system] in Rx data folder configuration.
Input IP address of CV-5000(2)
57
3. [LAN Configuration]
4. Tick in [Enable Network connection] and
Input IP address of CV-5000S(2)
6. [Data Folder]
7. if RM/KR is used,
Tick in [use shared folder in CV system] in AR data folder configuration.
Input IP address of CV-5000(2)
Tick in [use shared folder in CV system] in Rx data folder configuration.
Input IP address of CV-5000(2)
8. Click OK.
58
10. Network system 3
To have several CV-5000S (Setup 1/2 and Setup 3)
10-1. Data flow diagram
Shared folder RM/KR/CL
CV-5000S (1) CV-5000S (2) (1)
(AR/Rx)
RS-232C
STD1
(2)
(2)
LAN XML
9 RM/K/CL data sent via RS-232C is once received at CV-5000S(2), then saved in shared folder.
9 KR-1 data if sent via LAN is saved directly in shared folder.
9 RM.KR.CL data are shared by CV-5000S(1) and (2).
9 Subjective data is not saved must be printout. (can be exported in file since next version)
10-2. Cabling
(1) 432140005 DIN8 DIN8 (RM/CL to CV COM3 ) or
418120002 DIN8-Dsub9 Din8-DSUB9 (CL to CV COM1/2)
(2) LAN
10-3. Settings
10-3.1. RM/KR/CL Æ RS-232C setting)
They should output data in STD1 format.
Please see RM/KR(6-1.1), KR-1(6-2.1),TRK-1P(6-4.1), CL-200(6-5.1), EZ-200(6-6.1).
59
10-3.2. KR-1 Æ (LAN )
Please see 6-3.1.
In Menu_LAN_Sharedfolder, input IP address of CV-5000S(1)
¥¥ IP address of shared CV 5000S(1)¥rm
e.g. ¥¥ 10.140.19.112¥rm
10-3.3. CV-5000S(1)
60
3. If it is Windows 7, turn off “password protected sharing”.
[My Computer]_[Control Panels]_[Network and Sharing center]_[Advanced Sharing settings]
_[Password protected sharing].
4.
61
10-3.4. CV-5000S(1) & Shared folder
3. [LAN Configuration]
4. Tick in [Enable Network connection] and check IP address of CV-5000S(1)
IP address is automatically generated from Windows network setting.
6. [Data Folder]
7. if RM/KR is used,
Tick in [use shared folder in CV system] in AR data folder configuration
Input IP address of CV-5000(1)
Tick in [use shared folder in CV system] in Rx data folder configuration.
Input IP address of CV-5000(1)
8. Click OK.
62
10-3.5. RM/KR ÆCV-5000S(2) Å Æ Shared folder
3. [Serial config]
4. In any of blank connection, touch the [▼] button and select each item.
6. [LAN Configuration]
7. Tick in “Enable Network connection” and input IP address of CV-5000S(2)
63
9. [Data Folder]
10. if RM/KR is used,
Tick in [use shared folder in CV system] in AR data folder configuration
Input address of CV-5000(1)
Tick in [use shared folder in CV system] in Rx data folder configuration
Input IP address of CV-5000(1)
11. Click OK
(1)
(2)
RS-232C
LAN
STD1
xml
Practice’s network
9 Shared folder can be on network, or can be the preinstalled folder in either CV-5000S.
9 RM/K/CL data sent via RS-232C is once received at CV-5000S(2), then saved in shared folder.
9 KR-1 data if sent via LAN is saved directly in shared folder.
64
9 RM.KR.CL data are shared by CV-5000S(1) and (2).
9 Subjective data is not saved, thus must be printout. (can be exported in file since next version)
9 CV-5000S(1) must not have user account named “TOPCON”.
9 Any of user/log in ID and password of CV-5000S(1) PC nor CV-5000S(2) PC should not include the
shared folder ID (=TOPCON)
(in September, CV-system will access to shared folder regardless of login user. )
11-2. Cabling
(1) 432140005 DIN8 DIN8 (RM/CL to CV COM3 ) or
418120002 DIN8-Dsub9 Din8-DSUB9(CL to CV COM1/2)
(2) LAN
11-3. Settings
65
[User Rights Assignment]_[Deny access to this computer from the network].
If “guest” exists, right click and select properties. Remove “Guest.
7.
66
4. Start CV-5000S PC software and Click [settings] icon.
6. [LAN Configuration]
7. Tick in [Enable Network connection] and check IP address of CV-5000S(1)
6. [Data Folder]
7. if RM/KR is used,
Tick in [use shared folder in CV system] in [AR data folder configuration]
Input IP address of CV-5000(1)
Tick in [use shared folder in CV system] in Rx data folder configuration.
Input IP address of CV-5000(1)
8. Click OK.
67
11-3.2. RM/KRÆ CV-5000(2)Å Æ Shared folder
3. [LAN Configuration]
4. Tick in [Enable Network connection] and check IP address of CV-5000S(2)
IP address is automatically generated from Windows network setting.
6. [Data Folder]
7. if RM/KR is used,
Tick in [use shared folder in CV system] in AR data folder configuration.
Input IP address of CV-5000(1)
Tick in [use shared folder in CV system] in Rx data folder configuration.
Input IP address of CV-5000(1)
8. [Serial config]
68
9. In any of blank connection, touch the [▼] button and select each item.
5. Click OK
(1) (2)
RS232C RS-232C
STD1 STD1
9 CV-5000S receives representative value from RM/KR/CL and send subjective data to IMAGEnetR4
69
together with RM/KR/CL data.
9 In order to keep KR-1 all data (including peripheral kerato data), it must be linked directly with
IMAGEnetR4, via RS-232C cable and in STD4 format.
12-2. Cabling
(1) 418120002 DIN8-Dsub9 Din8-DSUB9(CL to CV COM1/2)
(2) 418120002 DIN8-Dsub9 Din8-DSUB9(CV COM3 to PC serial )
or DSUB DSUB (cross) locally arrange (CVCOM1/2 to PC serial)
(3) 418120002 DIN8-Dsub9 Din8-DSUB9(CV COM3 to PC serial )
12-3. setting
12-3.1. RM/KR/CL Æ (RS-232C)
They should output data in STD1 format. (KR-1 can be STD4)
Please see RM/KR(6-1.1), KR-1(6-2.1),TRK-1P(6-4.1), CL-200(6-5.1), EZ-200(6-6.1)
3. [Serial config]
4. In any of blank connection, touch the [▼] button and select each item.
“for” : “[In] RM/CL data receive”.
Port : Port number which RM/KR/CL is connected.
BaudRate : 2400bps or 9600 bps.
70
5. Select one of blank connection and set as follows.
*for : [I/O] Database connection
*Port No. : COM1-3 which is connected to IMAGEnetPC
*Baud Rate : 2400 bps
6. [LAN Configuration]
7. Tick off in [Enable network connection]
10.Click OK
71
12.Touch OK
12-3.3. IMAGEnetR4
1.create CV-5000 communication plug-in
See. 7-1.3,7-1.4.
*Out Plugi-in*
Same as CV-5000S
(1) LAN
(1) LAN (1) LAN (1) LAN
RM XML
XML (RM/CL) XML (RM/CL) CV data (to IMAGEnet)
Shared folder
RM/CL
IMAGEnet(2)
Fundus images etc
72
9 KR-1 data will be saved in shared folder and not directly saved in IMAGEnet.
9 CL-200 data will be saved in shared folder and not directly saved in IMAGEnet.
9 CV-5000(1) retrieves any RM/CL data from shared folder, and its subjective data is not saved.
9 CV-5000(2) retrieves any RM/CL saved in shared folder and sends subjective data to IMAGEnet (1).
9 Subject data (STD1 format) contains RM/CL data, and subjective data, and can be reviewed on
IMAGEnetR4.
13-2. Cabling
(1) LAN
(2) 418120002 DIN8-Dsub9 Din8-DSUB9 (CL OUT – CV COM1/2)
432140005 DIN8 DIN8 (RM/CL to CV COM3 )
(3) 418120002 DIN8-Dsub9 Din8-DSUB9 (CV COM3 –PC serial)
DSUB DSUB (cross) locally arrange (CVCOM1/2 to PC serial)
13-3. Settings
13-3.1. Server
1. Check IP address of server.
2. Create shared folder in server for RM/CL data (e.g. “rm” folder, “cl” folder).
73
13-3.4. RM/KR/CL ÆCV-5000S (1) Å Æ shared folder
3. [LAN Configuration]
4. Tick in [Enable Network connection]
and check IP address of CV-5000S(1)
6. [Data Folder]
7. if RM/KR is used,
Tick off [use shared folder in CV system] in AR data folder configuration.
*Input IP address of server¥rm
(must be the identical name and path of created folder in 13-3.1)
*Input user name which is permitted to access this shared folder.
*Input password assigned to the above user name.
74
8. [settings]_[Data I/O settings]_[System configuration]
9. [Serial config]
10. In any of blank connection, touch the [▼] button and select each item.
“for” : “[In] RM/CL data receive”.
Port : Port number which RM/KR/CL is connected.
BaudRate : 2400bps or 9600 bps.
3. [LAN Configuration]
4. Tick in [Enable Network connection] and
check IP address of CV-5000S(1)
75
5. [Data Folder]
6. if RM/KR is used,
Tick off [use shared folder in CV system] in AR data folder configuration.
*Input IP address of server¥rm
(must be the identical name and path of created folder in 13-3.1)
*Input user name which is permitted to access this shared folder.
*Input password assigned to the above user name.
9.Click OK
76
10.(SETUP1) Go to [Function Settings]_[Print Switch]_Action,
Please select [Data export] or [Printout & data export]
(11) Touch OK
77
15. Customizing pictures and movies (PC-50S only)
PC-50S enables to add still images, still images for slideshow and movies that are provided by the
user.
[Capability]
z Maximum 20 (JPG) of both still images and still images for slideshow.
z Maximum 20 (BMP) still images, which are displayed on KB-50S.
z Maximum 2 (WMV) movies.
[How to register]
Carry out registration as mentioned below.
78
1. Make folders under USB memory in the following configuration and save the files in the
folders.
Folder Explanation
PC50S_UC Make this folder under the USB memory. This folder
will contain all the folders.
Movie Make this folder under the “PC50S_UC” folder. Save
the movie file in this folder.
Still Make this folder under the “PC50S_UC” folder. This
folder will contain the still images.
KB50S Make this folder under the “Still” folder. The BMP
images are displayed on KB-50S for the still images
displayed on PC-50S. Save the file of these BMP
images in this folder.
Selective Make this folder under the “Still” folder. Save the file of
the JPEG images to be displayed on slideshow in this
folder.
Slide Make this folder under the “Still” folder. Save the file of
the JPEG still images to be displayed on PC-50S in this
folder.
2. Connect the USB memory, which has been arranged in Step 1, to the CV system. Then,
start the CV system.
Still images and movies are not captured into the system but are read directly from the USB
memory. So it is necessary to connect the USB memory to the CV system when using this
added function.
3. Press the [Shift] and [Menu] switches to access the setting screen. Select “Chart settings”
and then “Chart Page registration”.
79
4. Register the icons to access still image, slideshow and movie into the chart page from the
chart icon list according to the files saved in the USB memory.
Icon Explanation
These icons are used to access the still images saved
in the “Selective” folder. The number on each icon
~ corresponds to the number in the filename.
This icon is used to display the still images in the
“Slide” folder in turn as slideshow.
These icons are used to access the movies saved in
the “Movie” folder. The number on each icon
~ corresponds to the number in the filename.
5. After finishing the above registration, return to the test screen and check if the playback of
the still images, slideshow and movies is performed.
B(F)/(M)
D3
D5
H1
To use these charts, press the [Shift] and [Menu] switches to access the setting screen.
Select “Chart settings” and then “Chart Page registration”. Register the above charts in the chart
list to the chart page.
Output of setting
Perform the following operations on the CV system where the setting should be output.
1. Press the [Shift] and [Menu] switches to access the setting screen.
2. Touch the [Update] button.
3. Touch the [Config. copy] button among the function buttons.
4. Connect the USB memory to the CV system.
The [Export configurations] button will be flickering. Please wait utill “Export configuration” button is
lighting without flickering.
Touch this button.
80
The output is finished.
If you move or delete the output file out of the USB memory, the setting data cannot be re-input our
uploaded through this USB memory. Be careful.
Input of setting
Perform the following operations on the CV system where the setting should be input.
1. Press the [Shift] and [Menu] switches to access the setting screen.
2. Touch the [Update] button.
3. Touch the [Config. copy] button among the function buttons.
4. Connect the USB memory saving the setting data of the CV system, and the [Import configurations]
will be flickering. Please wait until [Import configurations] button is lighting without flickering.
Touch this button.
5. To make the setting valid, restart the CV system.
The input of setting is finished.
The setting items shown below are not copied. Be careful.
• Date setting
• Time setting
• Next data number
• Far point vision – Device
• Far point vision – Type
• Far point vision - Interface/Chart monitor selection
• Near chart – Device
• Near chart – Type
• IP properties
• Serial configuration - Connection 1
• Serial configuration - Connection 2
• Serial configuration - Connection 3
• Serial configuration – Connection 4
1. Using the universal personal computer, prepare the install USB memory.
Copy installers from the install CD to the USB memory.
The install CD has “Main Installer_E” folder and “Main Installer_J” folder. “Main Installer_E”
contains the install set for export and “Main Installer_J” contains the install set for Japan.
Open the proper folder. From it, copy the “CV5000_Updater” folder to the route of USB memory.
2. On the KB-50S where you want to perform updating, press the [Shift] and [Menu] switches to access
the setting screen.
3. Touch the [Update] button.
4. Connect the install USB memory to the CV system, and the [Start] button is valid. Touch this
button.
5. Two or more installers are activated in order. Press the [OK] button for each of them to perform
update.
81
6. When update for all the installers is finished, the message instructing to restart the system is
displayed. Check that this message is displayed and then restart the system.
End
82