Woodpecker Endomatic User Guide
Woodpecker Endomatic User Guide
Woodpecker Endomatic User Guide
1 Product introduction............................................................................... 1
2 Installation ............................................................................................. 4
3 Function and operation of product ..................................................... 1 2
4 Operation instruction .......................................................................... 1 5
5 Troubleshooting................................................................................... 3 0
6 Cleaning, Disinfection and Sterilization.............................................. 3 1
7 Storage, maintenance and transportation ............................................ 3 4
8 Environmental protection .................................................................... 3 4
9 After service ........................................................................................ 3 5
10 Symbol instruction ............................................................................ 3 5
11 Statement ........................................................................................... 3 5
12 EMC-Declaration of comformity....................................................... 3 6
1 Product introduction
1.1 Preface
Guilin Woodpecker Medical Instrument Co., Ltd is a professional
manufacturer researching, developing, and producing dental products.
Woodpecker owns a sound quality control system. Guilin Woodpecker
Medical Instrument Co., Ltd has two brands, Woodpecker and DTE.
Its main products include Ultrasonic Scaler, Curing light, Apex locator,
Ultrasurgery, Endo Motor, etc.
1
1.5 Scope of application
1.5.1 The device can be used for preparation and enlargement of root
canals, or device for measuring canal length.
1.5.2 The device must be operated in hospital and clinic by the qualified
dentists.
1.6 Contraindication
Patients with implanted pacemakers (or other electrical equipment)
who are warned not to use household appliances such as electric razors,
hair dryers, etc. are not recommended to use this device.
1.7 Warnings
1.7.1 Please carefully read this Instruction Manual before first
operation.
1.7.2 This device should be operated by professional and qualified
2
dentist in qualified hospital or clinic.
1.7.3 Do not directly or indirectly place this device near heat source.
Operate and store this device in reliable environment.
1.7.4 This device requires special precautions regarding
electromagnetic compatibility (EMC) and must be in strict accordance
with the EMC information for installation and use. Do not use this
equipment especially in the vicinity of fluorescent lamps, radio
transmitting devices, remote control devices, handheld and mobile high-
frequency communication devices.
1.7.5 Long time use of Reciprocating Motion Mode may result in
motor handpiece overheat, thus it should be left to cool for use. If the
motor handpiece is overheated frequently, please contact local distributor.
1.7.6 Please use the original contra angle. Otherwise it will not be
used or cause adverse consequences.
1.7.7 Please do not make any changes to the device. Any changes
may violate safety regulations, causing harm to the patient. There will be
no promises of any modification.
1.7.8 Please use original power adapter. Other power adapter will
result in damage to lithium battery and control circuit.
1.7.9 The motor handpiece cannot be autoclaved. Use disinfectant of
neutral pH value or ethyl alcohol to wipe its surface.
1.7.10 Before the contra angle stopping rotating, do not press the
push cover of contra angle. Otherwise the contra angle will be broken.
1.7.11 Before the motor handpiece stopping rotating, do not remove
the contra angle. Otherwise the contra angle and the gear inside motor
handpiece will be broken.
1.7.12 Please confirm whether the file is well installed and locked
before starting the motor handpiece.
1.7.13 Please set torque and speed as per the recommended
specifications of file manufacturer.
1.7.14 Error in replacing lithium batteries can lead to unacceptable
risks, so use the original lithium battery and replace the lithium battery
according to the correct steps in the instructions.
1.7.15 Not to position equipment to make it difficult to operate the
disconnection device.
1.7.16 Please remove the battery if the motor handpiece is not likely
to be used for some time.
3
1.8 Device safety classification
1.8.1 Type of operation mode: Continuous operating device
1.8.2 Type of protection against electric shock: Class II equipment
with internal power supply
1.8.3 Degree of protection against electric shock: B type applied part
1.8.4 Degree of protection against harmful ingress of water: Ordinary
equipment (IPX0)
1.8.5 Degree of safety application in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot
be used in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
1.8.6 Applied part: contra angle, lip hook, file clip, touch probe.
1.8.7 The contact duration of applied part: 1 to 10 minutes.
1.8.8 The temperature of the surface of applied part may reach
46.6℃.
1.9 Primary technical specifications
1.9.1 Battery
Lithium battery in motor handpiece: 3.7V /2000mAh
1.9.2 Power adapter:
Input: ~100V-240V 50Hz/60Hz 400mA Max
Output: DC5V/1A
1.9.3 Torque rang: 0.4Ncm-5.0Ncm(4mNm ~ 50mNm)
1.9.4 Speed rang: 100rpm~1000rpm
1.10 Environment parameters
1.10.1 Environment temperature: +5℃ ~ +40℃
1.10.2 Relative humidity: 30% ~ 75%
1.10.3 Atmospheric pressure: 70kPa ~ 106kPa
2 Installation
2.1 Basic accessories of product
4
Charging base Spray nozzle
2.2.1.2 CW Mode
The motor handpiece rotates forward 360º, clockwise direction. Used
5
rotaty files likes DENTSPLY Protaper or WOODPECKER W3-Pro.
6
rotation.
Trigger torque: 0.4Ncm, 0.6Ncm, 0.8Ncm, 1Ncm
Speed: 150rpm, 300rpm, 500rpm
2.2.2 Torque Display
This appears when the motor is running. Meter shows the torque load
on the file.
The meter numbers 1.0, 2.0, 3.0 and digital numbers 00-16 do not
represent the actual length from the apical foramen. It simply indicates
the file progression towards the apex. The digital numbers -1 and -2
indicate that the file has passed the apex foramen. The digital number “00”
indicate that the file has reached the apex foramen. Subtract 0.5-1mm
from the measured file length as the working length. These numbers are
used to estimate the canal’s working length.
2.3 Instructions for contra angle
2.3.1 The contra angle adopts precision gear transmission, and the
transmission ratio is 1.8: 1.
2.3.2 Before the first use and after treatments, please clean and
disinfect contra angle with disinfectant of neutral PH value. After
disinfection, lubricate it with specific cleaning oil. Finally, sterilize
7
it under high temperature and high pressure (134℃, 2.0bar~2.3bar
(0.20MPa~0.23MPa)).
2.3.3 The contra angle can only be used cooperatively with this
device. Otherwise the contra angle will be damaged.
2.4 Installation and removal of contra angle.
2.4.1 Installation
Line up the notch inside the contra angle with the projection inside
the motor handpiece and slide it in until it clicks securely into place.
The contra angle rotates 300° so that the OLED display can always
be viewed easily.
2.4.2 Removal
Pull out the contra angle horizontally when the motor handpiece does
not run.
8
Warnings:
a) Before plugging in or pulling out contra angle, please first stop the
motor handpiece.
b) After installation, please check and confirm that the contra angle
has been well installed.
c) The contra angle does not rotate freely. Do not try to rotate it past
its stopper.
2.5 Installation and removal of file
2.5.1 Installation of file
Before starting the device, plug the file into the hole of contra angle
head.
Hold down the push button on the contra angle and insert the file.
Turn the file back and forth until it is lined up with interior latch groove
and slips into place. Release the but-ton to lock the file into the contra
angle.
Warnings:
After plugging the file into contra angle, let go the hand on push
cover to assure that the file cannot be taken out.
Be careful when inserting files to avoid injury to fingers.
Inserting and removing files without holding the push button may
damage the chuck of contra angle.
Please use files with shanks meet the ISO standard. (ISO standard:
Ø2.334 – 2.350 mm)
2.5.2 Removal of file
Pressing the push cover, and then directly pull out the file.
9
Warnings:
Before plugging and pulling out the file, the motor handpiece must be
stopped.
Be careful when removing files to avoid injury to fingers.
Removing files without holding the push button will damage the
chuck of contra angle.
2.6 Canal measurement functional connection
This is not required if the canal measurement function will not be
used.
Connect the measuring wire to the motor handpiece. Line up the
measuring wire plug with the notch on the back of the motor and push it
all the way in.
Connect the file clip plug into the socket (black) on the measuring
wire. Connect the lip hook to the socket (white) on the measuring wire.
10
Warnings:
Connect the lip hook to the socket (white) on the measuring wire.
Otherwise, the function of root canal preparation and root canal length
measurement cannot be used together.
Make sure the screw is tight enough. Otherwise, it might come out
and be swallowed. Also, canal measurements might not be accurate.
11
3 Function and operation of product
3.1 Button definition and settings
a. Turn power on
Press Main button to turn on motor handpiece.
b. Turn power off
Hold down the Setting button “S”, then press Main button to turn off
motor handpiece.
c. Customized program change
Press Adjusting button “+”/“-” during standby sate.
d. Parameter setting
Press Setting button “S” till target parameters, press Adjusting button
“+”/“-” to change, then press Main button or wait 5 seconds to confirm.
e. Preset program selection
Long press Setting button “S” to entry preset program during standby
state, press Adjusting button “+”/“-” to select file system , press Setting
button “S” to entry select file number, press Adjusting button “+”/“-” to
select file number, then press Main button to confirm.
f. Handpiece functions setting
With the motor handpiece turned off, hold down the Setting button “S”
and press Main button to entry handpiece functions setting, press Setting
button “S” till target setting, press Adjusting button “+”/“-” to adjust,
then press Main button to confirm.
12
3.2 Screen display
Standby interface
a. Customized program sequence
number 0-9, totally 10 programs.
b. Battery consumption
c. Set speed
d. Set torque
e. Operation mode
Working interface
a. Set speed
b. Set torque
c. Real time torque
d. Torque display scale
13
Apical reference point setting
interface
a. Apical reference point flash bar
b. Apical foramen
c. Digital “02” meter reading, very
near physiological apical foramen.
3.3 Terms and definition
Clockwise rotation, forward ration
CW
Be applied to rotaty file
Counter clockwise rotation, reverse rotation
CCW Be applied to special file, inject calcium
hydroxide and other solutions
Reciprocating motion
Be applied to reciprocating file, path file and
REC
rotary file protection by setting some special
angle.
Adaptive torque reverse
Up to setting torque, the motor will move with
ATR
reciprocating ATR mode ; when torque reduce to
normal value, the motor will clockwise rotate.
Activating in REC and ATR operation mode.
ATR mode: only 180º or 240º are available.
Forward Angle
REC mode: adjustable every 10 degrees,
adjustment range: 30º-340º.
Activating in REC operation mode
Reverse Angle Adjustable every 10 degrees, adjustment range:
30º-340º.
Electronic apex locator
EAL In the mode, the device will work like a stand-
alone apex
AP Apical foramen.
The file action when file tip reaches the flash bar
Apical Action
point.
Shows the point inside the canal where specified
Flash Bar Position
apical action is triggered.
14
The file rotation starts automatically when the file
Auto Start
is inserted in the canal.
The file slows down automatically as it
Apical Slow Down approaches the apex.
Activating in CW and CCW operation mode.
5 operation modes for canal shaping and
Operation Mode measurement.
Such as CW, CCW, REC, ATR and EAL.
Speed File rotation speed.
For CW and CCW modes, the torque value
Torque
(Torque Limit) that triggers reverse rotation. For
(Torque Limit /
ATR mode, the torque value (Trigger Torque)
Trigger Torque)
that triggers ATR action.
4 Operation instruction
4.1 Power on and power off
4.1.1 Starting and stopping of motor handpiece
a) Under the power off state of motor handpiece, press Main button,
and then the motor handpiece will enter Standby interface. The interface
displays are as follow:
Standby interface
b) Under Standby mode, press Main button, and then the motor
handpiece will enter Working interface. The interface displays are as
follow:
Working interface
c) Press the Main button again, and then the motor handpiece backs
to Standby mode.
d) Hold down the Setting button “S”, then press Main button to turn
off motor handpiece. In Standby Interface, the motor handpiece would
automatically shut down after 3 minutes without any button-pressing
15
operation. The motor handpiece will also automatically shut down while
it is put into the charging base.
4.2 Selecting customized program sequence number
The motor handpiece has 10 memory programs(M0-M9) and 5 preset
programs, press Adjusting button “+”/“-” to change customized program
sequence number during standby state.
M0-M9 is a memory program for canal shaping and measurement,
every memory program has its own parameters such as Operation mode,
speed and torque, all these parameters can be changed.
4.3 Parameter setting
Before starting of motor handpiece, please
check the operation mode is correct.
All the parameters must be set according to
files, make sure all the parameters are excepted
before starting of motor handpiece, otherwise
has risk of file separate.
It has 5 operation modes for canal shaping
and measurement: CW, CCW, REC, ATR and
EAL(See chapter 3.3 Terms and definition to get
the explanations of these modes.)
Press Setting button “S” once during standby
state, press Adjusting button “+”/“-” to select
correct Operation mode.
CCW mode is used to inject calcium hydroxide
and other medicant. When this mode is being
used, a double-beep sounds continuously,
used for indicating counter clockwise rotation
happening.
Repeatedly press Setting button “S” to check all the next level
parameters of this operation mode are expected, press Adjusting button
“+”/“-” to select if not.
16
The speed setting can be adjusted from 100 rpm
to1000 rpm.
Press Adjusting button “+”/“-” to increase or
decrease speed. Long press to fast increase or
fast decrease speed.
In ATR mode, speed of 150rpm, 300rpm and
500rpm are available.
In REC mode, the speed is not optional.
The torque setting can be adjusted from 0.4Ncm
to 5Ncm.
Press Adjusting button “+”/“-” to increase or
decrease torque. Long press to fast increase or
fast decrease torque.
In ATR mode, the Trigger Torque of 0.4Ncm,
0.6Ncm, 0.8Ncm and 1.0Ncm are available.
In REC mode, the torque is not optional.
Actions that happen automatically when the file
tip reaches the point inside the canal determined
by the Flash Bar setting.
Benefit from integration of length determination,
when the file reaches the reference point, the
motor will response according to setting, it can
be Reverse , Stop and OFF.
P ress Adjusting button “+”/“-” to change.
OFF: Disable Apical Action function, file
rotating as usual even if reach the reference
point.
Stop: automatically rotation stop when reach
the reference point, upward a little bit and will
rotate again.
Reverse: automatically reverses rotation when
reach or pass the reference point, upward a
little bit, the rotation direction will change back
again.
17
Rotation starts automatically when the file is
inserted into the canal and the canal length
indicator bar lights up more than 2 bars.
P ress Adjusting button “+”/“-” to change.
OFF: Motor does not start when file is inserted
into the canal. The Main button is used to start
and stop the motor handpiece.
ON: Motor starts automatically.
This is the reference point where various apical
actions are triggered.
Press Adjusting button “+”/“-” to select
reference point by change the flash bar.
The meter’s 0.5 reading indicates that the file
tip is located very near the physiological apical
foramen.
The reference point (flash bar) can be set from 2
to AP (Apex) on the meter.
Rotation automatically slows down as the file
tip approaches the reference point.
P ress Adjusting button “+”/“-” to change.
OFF: Disable Apical Slow Down function.
ON: Rotation automatically slows down as the
file tip approaches the reference point.
18
Forward Angle: only activating in REC and
ATR operation mode.
Reverse Angle: only activating in REC operation
mode.
F: Forward Angle
R: Reverse Angle
Press Adjusting button “+”/“-” to change angle,
adjustable every 10 degrees.
It is suggested that the difference between the
forward angle and reverse angle should be
greater than or equal to 120 degrees, otherwise,
root canals cannot be prepared effectively.
Forward Angle<Reverse Angle, such as F: 30º/
R: 150º, effective cutting angle is Reverse Angle,
it is suitable for used the reciprocating files likes
DENTSPLY WAVEONE or WOODPECKER
W3-ONE.
Forward Angle>Reverse Angle, such as F: 180º/
R: 30º, effective cutting angle is Forward Angle,
it is suitable for used the reciprocating files likes
SENDONELINE S1.
Remarks: only 180º or 240º are available in ATR
mode.
4.4 Preset program selection
For convenience, we preset some common file
system.
Press Adjusting button “+”/“-” to switch to
preset program(M0-M9, preset program 1-5 ),
the interface will show as left.
Long press Setting button “S” to entry preset
program during standby state, the interface will
show as left.
Press Adjusting button “+”/“-” to select file
system.
19
After select file system, press Setting button
“S” to entry select file number, press Adjusting
button “+”/“-” to select file number, then press
Main button to confirm.
The parameters of "W3-Pro"can also be changed
make it different from default setting,and there
will be 4 corners around the Parameter.
If want to change back to default setting, long
press Setting button “S” to entry preset program
during standby state, select "W3-Pro" and press
"Main" button to confirm, the default setting
will be reloaded, and the 4 corners around will
disappear.
Changing the preset program default setting
is not recommended, otherwise has risk of file
separate.
4.5 Handpiece functions setting
With the motor handpiece turned off, hold down the Setting button “S”
and press Main button to entry handpiece functions setting, press Setting
button “S” till target setting, press Adjusting button “+”/“-” to adjust,
then press Main button to confirm.
With the motor handpiece turned off, hold down
the Setting button “S” and press Main button to
entry handpiece functions setting, the software
version number will appear on the display
screen.
Press Setting button “S”again, the "Dominant
Hand" can be change, press Adjusting button
“+”/“-” to adjust, then press to "Main" button to
confirm.
The right hand and the left hand can be set.
20
Press Setting button “S”again, the "Dominant
Hand" can be change, press Adjusting button
“+”/“-” to select “ON”, then press to "Main"
button to calibration.
Before calibrating, making sure the original
contra angle is installed, and do not install the
file. The torque will not correct if calibration
without original contra angle or any load on
contra angle chuck, andhas risk of file separate.
After replacement of contra angle, the contra
angle shall be calibrated before use.
Press Setting button “S”again, the "Beeper
Volume" can be change,press Adjusting button
“+”/“-” to adjust, then press to "Main" button to
confirm.
The"Beeper Volume"can be set from 0-3.
Vol.0: Mute.
4.6 Protective function of automatic reverse
During operation, if the load value exceeds the preset torque value,
the file rotation mode will automatically change to Reverse Mode. And
the file would return to normal rotation mode when the load is below the
preset torque value again.
Cautions:
1. Protective function of automatic reverse is ONLY suitable for CW
mode.
21
2. In REC mode, when the load value is higher than preset torque
value, if Forward angle is greater than Reverse angle, the file rotation
automatically change to reverse rotation, and if Forward angle is less
than Reverse angle, the file rotation automatically change to forward
rotation.
3. his function is forbidden under CCW mode, ATR mode.
4. When the motor handpiece battery indicator indicates a low
battery capacity, the low battery capacity is insufficient to support the
motor handpiece to reach the limit torque value, that is, the auto-reverse
function will not work properly. Please charge it in time.
5. If the motor handpiece is under load all the time, the machine may
stop automatically as a result of overheat protection. If it happens, turn
off the motor handpiece for a while until the temperature drops.
4.7 Motor operation
Please set operation mode, torque and speed as per the recommended
specifications of file manufacturer.
Motor alone mode
When using as motor alone mode,
the torque bar will show on the
screen.
(more information about torque
bar, please see chapter 3. 2 Screen
display )
22
Motor combined canal
measurement function mode
When using motor combined
canal measurement function, the
measuring wire must be connecting
with motor handpiece by USB
socket, and white socket connects
with patient’s lip by lip hook, keep
the black socket idle.
The canal length indicator bar
will show on the screen (more
information about canal length
indicator bar, please see chapter 3.
2 Screen display)
Setting parameters of automatic
functions as needed, such as Apical
Action, Auto Start, etc(more
information about automatic
functions, please see chapter 4.3
Parameter setting).
Connection testing
Strongly recommend check the
connection testing every time before
use. Touch the lip hook with the
file in the contra angle and check
that all the bars on the meter on
the screen light up, and the motor
should be reversed continuously,
otherwise, the measuring wire or
contra angle should be replace.
Make sure the screw is tight
enough. Otherwise, canal
measurements might not be
accurate.
23
Replacing Built-in Electrode
If the canal length indicator bars
flicker during use, or if all the bars
in the meter do not light up when
the file touches the lip hook, and
cleaning the rotor axle and built-
in electrode does not solve the
problem, then the built-in electrode
is worn out and must be replaced
with a new one according to the
following procedure.
(1) Loosen the screw and remove
the built-in electrode.
(2) Put a little Ethanol for
Disinfection (Ethanol 70 to 80
vol%) on a brush and clean the
rotor axle with it.
(3) Blow air on the electrode to
remove any remaining moisture.
(4) Slide the built-in electrode into
the contra angle and line up the
screw holes.
(5) Slowly turn the screw and make
sure the built-in electrode goes into
the head properly. Make sure the
screw is tight enough.
24
4.8 Canal measurement operation
When using as alone apex locator mode.
We suggest put the motor handpiece on
the charging base to get better visual
angle.
Press Setting button “S” once during
standby state, press Adjusting button
“+”/“-” to select EAL Operation mode,
then press Main button to confirm. (See
chapter 3.3 Terms and definition to get
the explanations of Operation modes.)
The measuring wire must be connecting
with motor handpiece by USB socket,
white socket connects with patient’s lip
by lip hook, and black socket connect
with file clip.
The canal length indicator bar will show
on the screen(more information about
canal length indicator bar, please see
chapter 3. 2 Screen display).
25
Connection testing
Strongly recommend check the
connection testing every time before
use. Clip the holder onto lip hook and
check that all the bars on the meter
on the screen light up, otherwise, the
measuring wire or file clip should be
replace.
Root Canals not suitable for canal measurement
Accurate measurement cannot be obtained if the root canal conditions
shown below.
Root canal with a large apical foramen
Root canal that has an exceptionally
large apical foramen due to a lesion
or incomplete development cannot be
accurately measured. The results may
show shorter measurement than the
actual length.
26
Root canal with blood overflowing from
the opening
If blood overflows from the opening
of the root canal and contacts the
gums, this will result in electrical
leakage and an accurate measurement
cannot be obtained. Wait for bleeding
to stop completely. Clean the inside
and opening of the canal throughly to
get rid of all blood, and then make a
measurement.
Root canal with a chemical solution
overflowing from the opening
An accurate measurement cannot be
obtained if some chemical solution is
overflowing from the canal opening.
In this case, clean the canal and its
opening.
It is important to get rid of any solution
overflowing the opening.
Broken crown
If the crown is broken and a section
of the gingival tissue intrudes into the
cavity surrounding the canal opening,
contact between the gingival tissue and
the file will result in electrical leakage
and an accurate measurement cannot be
obtained. In this case, build up the tooth
with a suitable material to insulate the
gingival tissue.
Fractured tooth
Leakage through a branch canal
Fractured tooth will cause electrical
leakage and an accurate measurement
cannot be obtained.
A branch canal will also cause electrical
leakage.
27
Re-treatment of a root filled with gutta-
percha
The gutta-percha must be completely
removed to eliminate its insulating
effect. After removing the gutta-percha,
pass a small file all the way through
the apical foramen and then put a little
saline in the canal, but do not let it
overflow the canal opening.
Crown or metal prosthesis touching
gingival tissue
Accurate measurement cannot be
obtained if the file touches a metal
prosthesis that is touching gingival
tissue. In this case, widen the opening at
the top of the crown so that the file will
not touch the metal prosthesis before
taking a measurement.
Extremely dry canal
If the canal is extremely dry, the meter
may not move until it is quite close to
the apex. In this case, try moistening the
canal with saline.
28
The actual apex for the canal is not the
same as that for the anatomical apex.
There are frequently cases where the
apical foramen is located up towards the
crown. In these cases, an X-ray might
indicate that the file has not reached the
apex even though it has actually reached
the apical foramen.
29
grease the contra angle for 2-3s till the oil flow out of contra angle head
part.
3. Vertically place the end part of contra angle more than 30 minutes
to let go the redundant oil under gravity.
Warnings
Motor handpiece cannot be filled with oil.
Cautions
a: To avoid the contra angle from flying out for the pressure, use hand
to safely hold the contra angle while greasing.
b: Do not use a swirling nozzle. Swing nozzle can only be used for
injection of gas, not for oiling.
5 Troubleshooting
Failure Possible cause Solutions
The motor handpiece Chose EAL mode, EAL Changing to CW, CCW,
does not rotate. mode is only for canal REC or ATR mode.
measurement.
There is continuous The continuous beep Stop the motor
beep sounds after sound is indicating that handpiece and change
starting the motor the motor handpiece is the operating mode to
handpiece. under CCW mode. CW Mode.
Contra angle Calibration failure Clean the contra angle,
calibration failure caused by strong and recalibrate after oil
resistance of contra injection.
angle
Motor handpiece Under Reciprocating Stop use. Use after the
heating Motion Mode, the temperature of motor
using time is too long. handpiece drops.
30
The time of endurance Battery capacity Please contact
becomes shorter after becomes smaller. local distributor or
charging. manufacturer.
No sound Beeper Volume set to 0. Set Beeper Volume to
Vol.0: Mute. 1,2,3.
The continuously Incorrect specification Choose CCW Mode,
rotating file is stuck at setting. start the motor
the root canal. Too high load torque of handpiece, and take the
file. file out.
31
Health Canada.
6.2.9 To sterilize the endodontic files, refer to the manufacturer's
instructions for use.
6.2.10 The contra angle needs to be lubricated after cleaning and
disinfection, but before sterilization.
6.3 Step-by-Step Procedure
# Operation Operating Mode Warning
1 Preparation Remove
accessories
(contra angle, lip
hook, file clip,
touch probe) from
handpiece and
base.
2 Automated Put the - Avoid any contact between the
Cleaning accessories contra angle and any instruments,
with washer- (contra angle, lip kits, supports or container.
disinfector hook, file clip, - Follow instructions and observe
touch probe) concentrations given by the
into the washer manufacturer (see also general
disinfector (Ao recommendations).
value >3000 or, - Use only approved washer-
at least 5 min at disinfector according to EN ISO
90°C/194°F) 15883, maintain and calibrate it
regularly.
- Make sure accessories (contra
angle, lip hook, file clip and touch
probe) are dry before moving to the
next step.
32
# Operation Operating Mode Warning
4 Packaging Pack the - Check the validity period of the
accessories pouch given by the manufacturer to
(contra angle, lip determine the shelf life.
hook, file clip, - Use packaging which is resistant
touch probe) in to a temperature up to 141°C
"Sterilization (286°F) and in accordance with EN
pouches". ISO 11607.
33
7 Storage, maintenance and transportation
7.1 Storage
7.1.1 This equipment should be stored in a room where the relative
humidity is 10% ~ 93%, atmospheric pressure is 70kPa to106kPa, and
the temperature is -20°C ~ +55°C.
7.1.2 Avoid the storage in a too hot condition. High temperature will
shorten the life of electronic components, damage battery, reshape or
melt some plastic.
7.1.3 Avoid the storage in a too cold condition. Otherwise, when the
temperature of the equipment increases to a normal level, there will be
dew that will possibly damage PCB board.
7.2 Maintenance
7.2.1 This device do not include accessories for repair usage, the
repair should be carried out by authorized person or authorized after
service center.
7.2.2 Keep the equipment in a dry storage condition.
7.2.3 Do not throw, beat or shock the equipment.
7.2.4 Do not smear the equipment with pigments.
7.2.5 Calibration is recommended when using a new/other contra
angle or after an extend period of operation, as the running properties can
change with usage, cleaning and sterilization.
7.2.6 Replace the battery if it seems to be running out of power
sooner than it should.
7.3 Transportation
7.3.1 Excessive impact and shake should be prevented in
transportation. Lay it carefully and lightly and don’t invert it.
7.3.2 Don’t put it together with dangerous goods during
transportation.
7.3.3 Avoid solarization and getting wet in rain and snow during
transportation.
8 Environmental protection
Please dispose according to the local laws.
34
9 After service
From the date this equipment has been sold, based on the warranty
card, we will repair this equipment free of charge if there are quality
problems. Please refer to the warranty card for the warranty period.
10 Symbol instruction
Refer to instrucion
Serial number
manual/booklet
Date of manufacture Manufacturer
Appliance compliance
Handle with care
WEEE directive
11 Statement
All rights of modifying the product are reserved to the manufacturer
without further notice. The pictures are only for reference. The final
interpretation rights belong to GUILIN WOODPECKER MEDICAL
INSTRUMENT CO., LTD. The industrial design, inner structure, etc,
have claimed for several patents by WOODPECKER, any copy or fake
product must undertake legal responsibilities.
35
12 EMC-Declaration of comformity
The device has been tested and homologated in accordance with EN
60601-1-2 for EMC. This does not guarantee in any way that this device
will not be effected by electromagnetic interference Avoid using the
device in high electromagnetic environment.
Technical Description Concerning Electromagnetic Emission
Table 1: Declaration - electromagnetic emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The model EndoMatic is intended for use in the electromagnetic environment
specified below. The customer or the user of the model EndoMatic should assure
that it is used in such an environment.
Emissions test Compliance Electromagnetic environment -
guidance
RF emissions Group 1 The model EndoMatic uses
CISPR 11 RF energy only for its internal
function. Therefore, its
RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions Class B The model EndoMatic
CISPR11 is suitable for used in all
Harmonic emissions Class A establishments, including
lEC 61000-3-2 domestic establishments and
Voltage fluctuations / Complies those directly connected to
flicker emissions the public low-voltage power
lEC 61000-3-3 supply network that supplies
buildings used for domestic
purposes.
36
Immunity test IEC 60601 Compliance level Electromagnetic
test level environment - guidance
Electrostatic ±8kV contact ±8kV contact Floors should be wood,
discharge (ESD) ±2, ±4, ±8, ±2, ±4, ±8, ±15kV concrete or ceramic tile.
lEC 61000-4-2 ±15kV air air If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast ±2kV for power ±2kV for power Mains power quality
transient/burst supply lines supply lines should be that of a
IEC 61000-4-4 ±1kV for Input/ typical commercial or
output lines hospital environment.
Surge ±0.5, ±1kV line ±0.5, ±1kV line to Mains power quality
lEC 61000-4-5 to line line should be that of a
±0.5, ±1, ±2kV ±0.5, ±1, ±2kV typical commercial or
line to earth line to earth hospital environment.
37
Conducted RF & Radiated RF
Guidance & Declaration - Electromagnetic immunity
The model EndoMatic is intended for use in.the electromagnetic environment
specified below. The customer or the user of the models EndoMatic should
assure that it is used in such an environment.
Immunity test IEC 60601 Compliance Electromagnetic environment -
test level level guidance
Conducted RF 3 Vrms 3V Portable and mobile RF
lEC 61000-4-6 150 kHz to 80 6V communications equipment should
Conducted RF MHz 3V/m be used no closer to any part of
lEC 61000-4-6 6 Vrms the models EndoMatic, including
Radiated RF ISM cables, than the recommended
lEC 61000-4-3 frequency separation distance calculated
band from the equation applicable to the
3 V/m frequency of the transmitter.
80 MHz to 2.7 Recommended separation distance
GHz d=1.2×P1/2
d=2×P1/2
d=1.2×P1/2 80 MHz to 800 MHz
d=2.3×P1/2 800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter
In watts (W) according to the
transmitter manufacturer and d
Is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,a
should be less than the compliance
level in each frequency range.b
Interference may occur In the
vicinity of equipment marked with
the following symbol:
NOTE I At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
38
a Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the model EndoMatic is used exceeds the applicable
RF compliance level above, the model EndoMatic should be observed to verify
normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the model EndoMatic.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3V/m.
39
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) accordable to the transmitter
manufacturer.
NOTE I At 80 MHz and 800 MHz. the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
40
ZMN/WI-09-472 V1.0-20190322