Ultrasound

The report details the inspection, cleaning, performance testing, and safety testing completed on an Affiniti 70 ultrasound system during a scheduled preventative maintenance, and notes that the system and all transducers passed all diagnostic tests and were in working order. Safety tests confirmed the system met electrical safety standards with no issues found.

Uploaded by

Afiq Husaini
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
54 views3 pages

Ultrasound

The report details the inspection, cleaning, performance testing, and safety testing completed on an Affiniti 70 ultrasound system during a scheduled preventative maintenance, and notes that the system and all transducers passed all diagnostic tests and were in working order. Safety tests confirmed the system met electrical safety standards with no issues found.

Uploaded by

Afiq Husaini
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

 

Page:  1 / 3
.
Ship To:  
AVISENA HOLDING SDN BHD
No 4, Jln Ikhtisas, Seksyen 14
Shah Alam
Malaysia
Test and Verification Report
Case Number: 0120705015
Work Order: WO-07159832
WORK DETAILS
Engineer Name: Afiq Husaini Bin Khairudin Event Type: Predictive / Preventative Maintenance
Device Status: Pass Service Activity: Preventative Maintenance
EQUIPMENT DETAILS
Product Name: Affiniti 70 Ultrasound System Serial Number: USN18F1356
Product Number: 795210 Installed Product: 74285664
UDI: (01)00884838056251(21)USN18F1356
  Min Max UoM Measurement Result Comments

Planned Maintenance
System Information
  Software Version 3.0.3
  Hardware Version B.0
Inspection & Cleaning
  Query Customer Performance Issues Pass
  Inspect Exterior Periph. & Power Cords. Pass
  Inspect & Clean Trackball Pass
  Inspect & Clean Monitor Pass
  Inspect/Clean Control Panels & Keyboard Pass
  Footswitch, Key- & Remote Control Pad Not Applicable
  Inspect Casters & Brakes Operation Pass
  Clean/Check System Filters & Fans Pass
  Check Articulation motion.(monitor, etc) Pass
  Check System Battery Pass
Performance & Peripherals
  Execute Full Set of System Diagnostics Pass
  Check, Download & Clear Logs Pass
  Perform Hard-disk Maintenance Pass
  Backup of System Settings and Configs. Pass
  Check Operation All User Ctrls. & Lights Pass
  Check Operation of Major Applications Pass

 
   

Page:  2 / 3
Work Order: WO-07159832
  Min Max UoM Measurement Result Comments

Planned Maintenance
Performance & Peripherals
  Test Printers Pass
  Test VCR Operation & Clean Not Installed
  Check Operation Dig. Recording Devices Not Installed
  Confirm remote connectivity Not applicable
System Safety General
  Protective Earth resistance (<= 200 mOhm) 200 mOhm 199 Pass
  Earth Leakage Current. NC, Highest val. (<= 500 µA) 500 µA 3 Pass
  Earth Leakage Current. SF. Highest val. (<= 1000 µA) 1000 µA 2.85 Pass
Transducer Body - BF
  Transducer model C5-1
  Transducer serial number B2T40P
  Transducer Cables, Connectors & Lens PASS
  XDR Leakage Current "Normal Condition" (<= 100 µA) 100 0.6 Pass
  XDR Leak Curr. Highest SF or 2SF reading" (<= 500 µA) 500 0.7 Pass
  XDR Isolation, ISO/MAP "Sink" (<= 500 µA) 500 0.6 Pass
  Run Transducer Test PASS
Transducer Heart - CF
  Transducer model L12-5 50
  Transducer serial number B2TN0T
  Transducer Cables, Connectors & Lens PASS
  XDR Leakage Current "Normal Condition" (<= 10 µA) 10 0.5 Pass
  XDR Leak Curr. Highest SF or 2SF reading (<= 50 µA) 50 0.7 Pass
  XDR Isolation, ISO/MAP "Sink" (<= 50 µA) 50 0.5 Pass
  Run Transducer Test PASS
Transducer Heart - CF
  Transducer model C8-5
  Transducer serial number B2WY1B
  Transducer Cables, Connectors & Lens PASS
  XDR Leakage Current "Normal Condition" (<= 10 µA) 10 0.5 Pass
  XDR Leak Curr. Highest SF or 2SF reading (<= 50 µA) 50 0.7 Pass
  XDR Isolation, ISO/MAP "Sink" (<= 50 µA) 50 0.5 Pass
  Run Transducer Test PASS
Transducer Heart - CF
  Transducer model S8-3
  Transducer serial number B2PXNC
 
   

Page:  3 / 3
Work Order: WO-07159832
  Min Max UoM Measurement Result Comments

Planned Maintenance
Transducer Heart - CF
  Transducer Cables, Connectors & Lens PASS
  XDR Leakage Current "Normal Condition" (<= 10 µA) 10 0.6 Pass
  XDR Leak Curr. Highest SF or 2SF reading (<= 50 µA) 50 0.8 Pass
  XDR Isolation, ISO/MAP "Sink" (<= 50 µA) 50 0.6 Pass
  Run Transducer Test PASS
 

Philips Representative Signature:    
 
Engineer Name: Afiq Husaini Bin Khairudin
Date: Feb 8, 2023  