PresionCalibracion PDF
PresionCalibracion PDF
PresionCalibracion PDF
Adapter kits
-3-
Pneumatic Pressure Pumps
Pressure hose
Pressure relief valve
Pump body
Type P 40.2 P 60
OEM version
Full version
-4-
Hydraulic Pressure Pumps
Full version
-5-
Practical and independent
Practical Mobile and independent
The test pumps and pressure generators are designed to SIKA test pumps and pressure generators are ideal for
enable the direct connection of all pressure systems to be mobile use. Their low weight and compact design make
tested using adapters. The test sample is easily connect- them easy to transport directly to the measurement site.
ed using the rugged industrial hose with integrated quick The instruments can be used immediately and do not
coupling and supplied adapters. The reference is fitted require an additional power supply. There is no need to
directly at the top of the pump using a positioning adapter. take along nitrogen bottles or connect the equipment to
a compressed air network. Wear-free manual pressure
generation is simple and easy, regardless of ambient
temperature and orientation.
Fine adjustment
►
Reference connection
► ►
Test connection Discharge
►
► ►
Possible areas of application
Force applied by pressing the handles together SIKA test pumps and pressure generators can be used
everywhere, including on site in workshops, test and
measurement rooms as well as laboratories. They cover
The required test pressure is initially generated using the a broad spectrum of industries with diverse applications.
handles and then adjusted precisely with the fine adjust-
ment valve. As a result, the pressure on both instruments • Assembly and commissioning
is the same. • Manufacturing and production
The pressure relief valve allows continuous pressure re- • Maintenance and service
duction and ensures accurate and easy testing, even with • Quality assurance and test equipment monitoring
decreasing pressure.
• Repair
In the simplest case, the pressure is indicated by an
analogue pressure gauge. An easy to read digital pres- SIKA test pumps and pressure generators are suitable
sure gauge or hand-held instrument can also be used. The for testing, adjusting and calibrating pressure sensors,
accuracy or adjustment of the pressure measuring device pressure gauges, pressure switches, safety valves and
being tested can be checked by comparing the indicated all types of pressure devices. They are optimised in their
reference value with the measured value for the device function and use and assist in the performance of specific
under test. tests and inspections.
-6-
Digital Pressure Gauges
Tara / Zero
User-defined zero point setting at the push of a button
makes offset adjustment easy and eliminates the need
for tedious mechanical adjustment. Single-point adjust-
ment allows the linear characteristic curve to be shifted
in positive or negative direction over the entire measuring
range.
Linearisation
Multi-point adjustment can be performed if it is necessary
to adjust the indicated values at different test points. Two-
point adjustment is available for setting the zero point and
slope of the measuring cell curve. Some digital pressure
gauges allow up to six offset values to be programmed
in order to shift the characteristic curve to meet the most
stringent customer expectations.
-7-
Digital pressure gauge selection table
E2 D2 R P L
Performance
Accuracy (FS) 0.5 % 0.1 % 0.1 % 0.5 % 0.05 %
0.1 % 0.025 %
0.05 % 0.01 %
Pressure range
-1...2 bar
-1...2.5 bar -1...2.5 bar
-1...3 bar
-1...5 bar -1...5 bar
-2...2.5 mbar
-1...25 mbar -20...25 mbar
0...100 mbar
0...250 mbar
-10...350 mbar -200...350 mbar
0...400 mbar
0...1 bar
0...1.3 bar 0...1.3 bar
-1...1.5 bar
-100...2000 mbar -1...2 bar
0...2.5 bar
-1...3 bar
0...4 bar
0...6 bar
0...7 bar
-1...10 bar
0...16 bar
0...25 bar
0...40 bar
0...60 bar
0...100 bar
0...160 bar
0...250 bar
0...400 bar
0...600 bar
0...1000 bar
-9-
Easy on-site use
Display Electronics / Pressure measuring cell
The large illuminated digital local display shows the The measuring cells and electronics used in the gaug-
measured pressure and indicates the current status of es are temperature compensated, so that the ef-
the digital pressure gauge, even under poor lighting con- fect of temperature on the readings is negligible. Liq-
ditions. This eliminates the difficult task of reading a dial uid entry into the measuring system is not necessary,
gauge and avoiding parallax errors. Needle jitter due to which eliminates the risk of damage from media resi-
vibration or pressure fluctuations is eliminated. Display dues. Another unbeatable feature of the electronic
damping or averaging can be configured directly using measuring cells is their immunity to pressure surges.
display filters. This ensures easy, tireless readout.
Min / Max displays and Peak function
Selectable pressure units Experience shows that excess pressure and pressure
Another feature is the large selection of pressure units. peaks significantly higher than normal operating pressure
Up to 13 different units are possible – far more than any occur at some measuring points. Min/max displays and fast
complicated dual-scale or multi-scale gauge can offer. peak value measurement cycles in digital pressure gaug-
The required display unit is selected directly on the es assist in system analysis and allow peak values to be
digital pressure gauge and is clearly indicated on the determined. This allows incorrect readings and violations
display. No conversion necessary; the desired value of range limits to be detected and helps avoid damage
can be read directly. to pressure systems. Preventive service is often less ex-
pensive than repairing or replacing defective instruments.
Protection class
High IP protection classes are available to minimise dust
and water sensitivity. Rugged, impact-resistant digital
pressure gauges are fitted with rubber caps for protection
during transport and field use.
Area of application
The right measuring system is available for every meas-
uring task. For simple applications with air or non-corro-
sive and non-ionising substances, low-cost unenclosed
pressure sensors are used. In difficult applications with
water or other aggressive media, high-quality stainless Direct mounting, built-in version and
steel versions are used. hand-held instrument
Once the intended use has been determined and the The compact, handy design proves its worth in everyday
pressure range has been specified, a digital pressure use. It puts an end to large-diameter gauges with sizes
gauge with an internal measuring cell is preferably used. up to 250 mm, as is common with precision pressure
For frequently changing application conditions, plug-in gauges. The small size simplifies direct mounting. If nec-
pressure sensors for various pressure ranges and appli- essary, built-in versions are available for switchgear cabi-
cations can be fitted using adapters. Automatic sensor net or control panel mounting. Hand-held digital pressure
recognition using standard DIN connectors offers a sim- gauges are especially suitable for applications where
ple Plug-&-Play solution. short-term pressure measurements are desired instead
of continuous measurement.
- 10 -
Supplementary functions
PC connection and software Analogue output
Many digital pressure gauges have a serial interface An electrical output signal enables remote display on a
port to allow measurement values and stored data to control console or in a control room as well as the con-
be transferred directly to a PC and documented. An in- nection of external recorders and indicating instruments.
expensive measurement data acquisition system can
easily be assembled with suitable software and an in- Relay output / Alarm signalling
terface converter. Processes can be readily monitored Digital pressure gauges allow limit contacts to be closed
and analysed using the recorded and visualised meas- even at low pressures. There is no need for high actuation
urements and all data can be exported using standard forces for magnetic spring or inductive contacts, which
programs such as Microsoft Excel. Remote control is makes it easier to signal critical equipment conditions and
also possible. Various software packages with exten- perform supplementary control tasks. A built-in buzzer gen-
sive recording and display functions, logger and alarm erates an alarm when the pressure exceeds the range of
evaluation as well as for calibration are available. the programmed minimum and maximum pressure levels.
Temperature display
Temperature measurement is often required in addition
to pressure measurement. For this purpose, a tempera-
ture sensor is integrated in the measuring cell to detect
the temperature of the medium. The process temperature
can be displayed at the press of a button. This allows two
quantities to be measured at a single measuring point,
which saves costs.
Explosion protection
Explosion-proof versions are also available for use in po-
tentially explosive locations, e.g. oil refineries, chemical
plants and drilling platforms.
Example applications
Data memory • Continuous or temporary checking of a wide variety of
The logger function for local data storage can be used system pressures
to record pressure curves automatically and perform • Air density measurement in building shells for the de-
leak tests. The integrated data memory in digital pres- tection and elimination of problem areas and avoiding
sure gauges allows a variety of data sets to be recorded structural damage
directly. The time interval between samples is program-
mable and the maximum recording interval is configur- • Monitoring the degree of soiling of filter units in venti-
able. The stored values can be displayed on a PC. Data lation or air conditioning systems
import at the press of a button is also possible. In this • Recording pressure drops for the determination of
case the data is shown directly on the display. In this leakage rates in leak tests
process the values are automatically annotated with the • Measurement of barometric air pressure for the deter-
date and time of day using an integrated real-time clock. mination of weather conditions
• Reference pressure gauge for calibration tasks
- 11 -
Types E2, D2 and R
type E2 / D2 type R
- 12 -
Functions
Type E2 / D2 R
Adjustment options
Linearisation - 6 points
Tare/Zero
Selectable units
Pressure bar, mbar, kPa, MPa, PSI bar
Temperature - °C
Features
Measuring inputs 1 x direct 1 x direct
PC connection RS232 (optional)
Analogue output
Relay output
Alarm buzzer
Built-in version
Explosion protection
Data memory
Number of memories 60.000 values (auto)
Recording interval 1 sec...10 h
Recording duration 1 min...1000 h
Real-time clock/date
Store function
Data sets Pressure / Temperature
Display / Representation
Multi-functional LCD 4 ½ digit 4 digit
Bar graph
Illumination
Display filter
Min/max value
Average
Differential pressure
Measuring rate
Standard 10 ms 100 ms
Peak / Fast 10 ms 8 ms
Process connection
Connection options G¼ G½
Material 1.4404 1.4542
Medium temperature -20...80 °C -10...70 °C
for aggressive media
Housing
Protection class IP67 (front) / IP67 IP65 (front) / IP40
Dimension Ø 80 mm T=30 mm H=100 mm Ø 85 mm T=30 mm H=30 mm
Material Zinc casting Aluminium
Operating temperature 0...50 °C -10...70 °C
Weight 540 g 450 g
Power
Auto-off function
Battery type 2x 1.5 V AA 2x 1.5 V AAA
Ext. power
Battery operation 1500 h 8000 h
- 13 -
Typ P und L
- 14 -
Functions
Type P L
Adjustment options
Linearisation 6 points
Tare/Zero
Selectable units
Pressure bar, mbar, kPa, MPa, PSI bar, mbar, hPa, kPa, MPa, PSI, mmHg, inHg,
Temperature - cmH2O, mH2O, inH2O, ftH2O, Kp/cm²
Features
Measuring inputs 1 x direct 1 x direct
PC connection RS232 (optional) RS485
Analogue output 0(4)...20mA / 0...10 V (optional)
Relay output 2 x 24 VDC/1A (optional)
Alarm buzzer
Built-in version
Explosion protection Ex II 2G Ex ia II C T6
Data memory
Number of memories
Recording interval
Recording duration
Real-time clock/date
Store function
Data sets
Display / Representation
Multi-functional LCD 4 digit (0.5 % / 0.1 %), 5 digit (0.05 %) 5 digit
Bar graph
Illumination
Display filter
Min/max value
Average
Differential pressure
Measuring rate
Standard 100 ms 500 ms
Peak / Fast
Process connection
Connection options G½ G¼
Material 1.4542 1.4435
Medium temperature 0...50 °C 0...50 °C
for aggressive media
Housing
Protection class IP65 (front) / IP40 IP65 (front) / IP54
Dimension 80 x 80 mm T=50 mm H=130 mm Ø 80 mm T=40 mm H=120 mm
Material Aluminium ABS plastic
Operating temperature 0...50 °C 0...50 °C
Weight 900 g 210 g
Power
Auto-off function
Battery type 2x 1.5 V AAA 1x 3 VCR
Ext. power 24 VDC
Battery operation 8000 h 2000 h
- 15 -
Type MH 3161 and MH 3181
Technical data
Accuracy (FS) 0.2 %
Pressure ranges Resolution
-1...25 mbar 0.01 mbar
-10...350 mbar 0.1 mbar
0...1,3 bar (abs.) 1 mbar
-100...2000 mbar 1 mbar
- 16 -
Functions
Type MH 3161 MH 3181
Adjustment options
Linearisation 2 points
Tare/Zero
Selectable units
Pressure bar, mbar, kPa, MPa, Pa, PSI, mmHg, mH2O
Temperature -
Features
Measuring inputs 2 x direct 2 x direct
PC connection TTL (USB optional) TTL (USB optional)
Analogue output 0...1 V
Relay output
Alarm buzzer
Built-in version
Explosion protection Ex II 2G Ex ib II C T4 Ex II 2G Ex ib II C T4
Data memory
Number of memories 10.000 values (auto)
Recording interval 1...3600 sec.
Recording duration
Real-time clock/date
Store function 99 Werte (manuell)
Data sets Pressure / min / max
Display / Representation
Multi-functional LCD 4½ digit 4½ digit
Bar graph
Illumination
Display filter
Min/max value
Average
Differential pressure
Measuring rate
Standard 250 ms 250 ms
Peak/Fast 1 ms
Process connection
Connection options Metal connection plugs Ø 6 mm
Material 2.0380 Nickel-plated
Medium temperature -10...50 °C
for aggressive media
Housing
Protection class IP65 (front) / IP40
Dimension 140 x 70 x 30 mm
Material ABS plastic
Operating temperature 0...50 °C
Weight 200 g
Power
Auto-off function
Battery type 1x 9 V Block
Ext. power 12 VDC
Battery operation 350 h
- 17 -
Type MH 3111, MH 3151 and MH 3156 with external pressure sensor
+
See page 20 for pressure ranges of external pressure sensors.
- 18 -
Functions
Type MH 3111 MH 3151 MH 3156
Adjustment options
Linearisation 2 points
Tare / Zero
Selectable units
Pressure bar, mbar, kPa, MPa, Pa, PSI, mmHg, mH2O
Temperature -
Features
Measuring inputs 1 x interchangeable 1 x interchangeable 2 x interchangeable
PC connection TTL (USB optional) TTL (USB optional) TTL (USB optional)
Analogue output 0...1 V 0...1 V
Relay output
Alarm buzzer
Built-in version
Explosion protection Ex II 2G Ex ib II C T4 Ex II 2G Ex ib II C T4 Ex II 2G Ex ib II C T4
Data memory
Number of memories 10.000 values (auto) 4000 values (auto)
Recording interval 1...3600 sec. 1...3600 sec.
Recording duration
Real-time clock/date
Store function 99 values (manual) 99 values (manual)
Data sets Pressure / min / max Pressure 1 & 2 / Min 1 & 2 / Max 1 & 2
Display / Representation
Multi-functional LCD 4½ digit 4½ digit 4½ digit
Bar graph
Illumination
Display filter
Min/max value
Average
Differential pressure optional optional optional
Measuring rate
Standard 250 ms 250 ms 250 ms
Peak / Fast 1 ms 1 ms
Process connection (depending on chosen external pressure sensors)
Connection options Metal connection plugs Ø 6 G½
Material Nylon 1.4542
Medium temperature 0...50 °C 0...70 °C
for aggressive media
Housing
Protection class IP65 (front) / IP40
Dimension 140 x 70 x 30 mm
Material ABS plastic
Operating temperature 0...50 °C
Weight 200 g
Power
Auto-off function
Battery type 1x 9 V Block
Ext. power 12 VDC
Battery operation 350 h
- 19 -
External Pressure Sensors
Type (Nylon)
Accuracy (FS) 0.2 %
Pressure ranges Resolution
MSD 2,5 MR -2...2.5 mbar (rel.) 0.001 mbar
MSD 25 MR -20...25 mbar (rel.) 0.01 mbar
MSD 350 MR -200...350 mbar (rel.) 0.1 mbar
MSD 1,3 BA 0...1.3 bar (abs.) 1 mbar
MSD 2 BA 0...2 bar (abs.) 1 mbar
MSD 2 BR -1...2 bar (rel.) 1 mbar
MSD 7 BA 0...7 bar (abs.) 10 mbar
MSD 10 BR -1...10 bar (rel.) 10 mbar
Type (stainless steel)
Nylon version
MSD 100 MRE 0...100 mbar (rel.) 0.1 mbar
Piezoresistive pressure sensor for air and non-corrosive/
non-ionising gases and liquids MSD 250 MRE 0...250 mbar (rel.) 0.1 mbar
MSD 400 MRE 0...400 mbar (rel.) 0.1 mbar
Two nylon nipples for 6 x 1 mm pressure tubing
MSD 1 BAE 0...1 bar (abs.) 1 mbar
Electronics Integrated amplifier and sensor data MSD 1 BRE 0...1 bar (rel.) 1 mbar
memory MSD -1/1,5 BRE -1...1.5 bar (rel.) 1 mbar
Electrical PVC cable (1 m)
MSD -1/3 BRE -1...3 bar (rel.) 1 mbar
connection with miniature DIN connector
MSD 2,5 BAE 0...2.5 bar (abs.) 1 mbar
Housing Impact-resistant ABS plastic
Water penetration rating IP65 MSD 2,5 BRE 0...2.5 bar (rel.) 1 mbar
Dimensions 70 x 30 x 15 mm (H x W x D) MSD 4 BAE 0...4 bar (abs.) 1 mbar
Weight approx. 75 g MSD 4 BRE 0...4 bar (rel.) 1 mbar
Stainless steel version MSD 6 BAE 0...6 bar (abs.) 1 mbar
Piezoresistive pressure sensor with separate connecting MSD 6 BRE 0...6 bar (rel.) 1 mbar
cable, for aggressive media as well as water, gases and liquids MSD 10 BAE 0...10 bar (abs.) 10 mbar
Stainless steel connection G ½ male thread MSD 10 BRE 0...10 bar (rel.) 10 mbar
MSD 16 BAE 0...16 bar (abs.) 10 mbar
Electronics Integrated amplifier and
sensor data memory MSD 25 BAE 0...25 bar (abs.) 10 mbar
Electrical PVC cable (1 m) MSD 25 BRE 0...25 bar (rel.) 10 mbar
connection with miniature DIN connector MSD 40 BRE 0...40 bar (rel.) 10 mbar
(optional)
MSD 60 BRE 0...60 bar (rel.) 10 mbar
Housing Stainless steel
Water penetration rating IP65 MSD 100 BRE 0...100 bar (rel.) 100 mbar
Dimensions approx. 30 x 90 mm (D x L) MSD 160 BRE 0...160 bar (rel.) 100 mbar
Weight approx. 220 g MSD 250 BRE 0...250 bar (rel.) 100 mbar
MSD 400 BRE 0...400 bar (rel.) 100 mbar
Nylon and stainless steel options
MSD 600 BRE 0...600 bar (rel.) 100 mbar
• Higher sensor accuracy (from 350 mbar)
MSD 1000 BRE 0...1000 bar (rel.) 1 bar
• Ex II 2G Ex ib IIC T4
- 20 -
Pressure Calibrators
Comparison of digital and analogue pressure gauges Manual test pumps or pressure generators are used for
Pressure calibration is the comparison between the in- simple pressure generation.
dicated values of a pressure measuring device with the SIKA offers a complete range of pressure calibrators for
indicated values of a pressure standard with a known ac- a wide variety of applications to allow specified test and
curacy. calibration tasks to be performed.
In many cases, the device to be tested cannot be re- Routine on-site calibrations can be performed very quick-
moved from the active process. ly and economically with the right combination of pres-
Calibration is performed on site to avoid lengthy down- sure generator and reference. This ensures that the indi-
times. Portable SIKA pressure calibrators are especially cated pressure values are correct and reliable and that all
suitable for this purpose. specified requirements are fulfilled.
In order to perform a specified functional test or accuracy
check, the test sample is often connected to the calibra-
tion device with a pressure hose. Digital pressure gauges
with sufficient precision can be used as compact refer-
ence instruments.
- 21 -
Possible combinations
►
▼ Reference A Reference E2 Reference D2 Reference R Reference P Reference L
P 60 PM 60 A PM 60 E2 PM 60 D2 PM 60 R PM 60 P
All manual pressure pumps and references can be combined as indicated above for various measuring ranges, resolutions and accuracy classes.
- 22 -
Calibrator PC 700+
Technical data
Accuracy (FS) 0.025 %
Pressure ranges (Resolution) 0...60 bar (5 mbar)
0...200 bar (20 mbar)
0...350 bar (30 mbar)
0...700 bar (70 mbar)
Function
Adjustment options
Tara / Zero
Selectable units (Pressure) bar, PSI, kPa, MPa, atm, at (kgf/cm²), mmHg, mHg, inHg, cmH2O, mH2O, inH2O, torr
Features
Measuring inputs 1 x direct
PC connection RS232
Explosion protection EEx ia IIC T4
electronic measuring inputs 0...30 mA / 0...15 V
Data memory
Store function 30 x 10 values (manual)
Display/representation
Multi-functional LCD 5 digit
Illumination
Display filter
Min / max value
Measuring rate
Standard 100 ms
Process connection
Connection options G¼
Material 2.4602
Medium temperature -20...80 °C
for aggressive media
Housing
Protection class IP65 (front) / IP40
Dimension, weight 350 x 160 x 180 mm, 8.1 kg
Material Aluminium
Operating temperature 0...50 °C
Power
Auto-off function
Battery type, battery operation 1x 9 V Block, 35 h
Ext. power 12 VDC
- 23 -
Our Production and Sales Range
®
Flow Sensors without moving Parts Turbine Flow Sensors Flow Switches
Pressure Gauges and Pressure Sensors Industrial Thermometers Electronic Digital Thermometers, Dial Thermometers
...measurement...control ...calibration
Phone: 0700 CALL SIKA
Phone: +49 5605 803-0
Fax: +49 5605 803-54
E-Mail: [email protected]
Internet: http://www.sika.net
Struthweg 7-9, 34260 Kaufungen
P. O. Box 11 13, 34254 Kaufungen
Germany
Druckkalibratoren_e 06/2012