Product Sp e ci f i ca t i ons
DFS High Resolution GC/MS
High Performance Magnetic Sector GC/MS
• Simple source design - clean quickly set-up is intuitive and rapid. The DFS mass • Calibration independent of mass range,
spectro-meter enables the reliable, routine scan speed, scan mode (including linked
• Inert composition - clean less frequently
use of AUTOTUNE in all ionization modes scan MS/MS), ion polarity and ionization
Reference Inlet without restrictions, including slit control and technique
An independent reference calibrant system automated resolution settings. • No recalibration required after change of
that is continuously flow adjustable and can ion volume or ionization technique
be individually evacuated. It allows syringe
introduction of liquid or gaseous samples.
By separating the reference inlet from the
GC interface, maintainance is reduced and
uptime maximized.
3
Direct Probe Base Unit (optional) Water Cooled DI Probe (DI / DIP) DCI Probe (optional)
• Port for direct insertion probe (optional) • Direct chemical ionization probe
• Required to attach any of the following • Direct insertion probe • Full data system control
probe options • Data system controlled, liquid cooled • Using reusable filaments
• Contains exchange lock and probe • Using disposable aluminum crucibles • Ultra fast heating rates of > 500 °C/s
electronics • Heating rates: 20 °C to 200 °C in 30 s, • Maximum temperature 1600 °C
200 °C to 300 °C in 30 s • High temperature tip for DCI probe
• Temperature range from 10 °C above (optional)
ambient • Using reusable quartz crucibles
• Maximum temperature 350 °C • Maximum temperature 800 °C
4 Software Specifications Data System
Thermo Scientific Xcalibur Data System Thermo Scientific TargetQuan Personal Computer in the following
Xcalibur™ is the uniform software platform Software Suite (optional) minimum configuration*: Dell Optiplex 990
for system control of the Thermo Scientific The special Xcalibur DFS software package MT Core i3-2100
GC/MS and LC/MS systems. for automated data evaluation on target
• 3.1 GHz Processor
compounds including:
The DFS mass spectrometer uses the • Mini Tower case
Xcalibur instrument control software for • Dioxin method setup
• 4096 MB 800 MHz DDR2 RAM
high and low resolution operation, multiple • Support of instrument and quantitation
ion detection MID, selection of positive or • 1TB SATA II hard disk
• Response file and reporting programs
negative ions, linked scans, peak matching, • DVD-RW drive
and full control of analyzer and inlet systems • Data evaluation for isotope dilution
• DVI Add-in card
supporting the following capabilities: methods as well as relative response
factors • Network card 10/100
• INTENSITY AUTOTUNE: automated • 2 PCI and 1 PCI-Express slot space
• Compliant with the published EPA
optimization of tune peak for maximum
methods for dioxin measurements and • 1 parallel port
senstivity
data evaluation according e.g. EPA 1613, • 2 serial ports
• RESOLUTION AUTOTUNE: computer EPA 8280, EPA 8290, EPA 23, EPA 513,
controlled slit settings for maximum • 12 USB 2.0
EN 1948 and equivalent JIS methods
selectivity • Graphics chip on board, up to 128 MB of
• Compliant with the requirement for TEQ
• Control of standard and optional inlet memory
low-med-upper bound reporting
systems • 10/100/1000 Ethernet chip on board
• Standardized output to LIMS systems
• Instrument diagnostics: full advanced • U.S. keyboard black
suite of instrument diagnostics including Library Options
• Dell USB 6-Button Optical Mouse Black
integrated parameter logging • NIST Library
• Microsoft Windows 7 Professional
• Xcalibur application software, • Wiley Library
Operating system (32Bit OS)
incorporating all mass spectrometry • Pfleger-Maurer-Weber Library
• Microsoft Office 2010 Professional
processing tasks such as chromatogram • Thermo Scientific Pesticide Library
and spectrum display, integrated NIST • Mouse and Keyboard
library search, CMASS for accurate mass Software Licenses • High resolution 22 inch TFT color monitor
conversion and averaging, elemental The DFS Xcalibur software licenses are
• Optional laser printer, BW,
composition and isotopic pattern supplied for instrument control as well as
1200 x 1200 dpi, up to 25 pages/min
calculation. reprocessing.
* Minimum data system specifications
• QuanBrowser: quantification package New instrument software releases are
may change without prior notice in case of
• MS data import and export using the supplied free of charge within 12 month after
technological improvement. Call for latest
ANDI/netCDF formats, conversion from delivery.
configuration.
Finnigan MassLab data file formats, ASCII
text export
• Standardized output to LIMS systems
Gas Chromatography Options IEC (Integrated Electronic Control) Detectors 5
Gas Specification
Thermo Scientific TRACE 1310 • Up to 18 channels of integrated electronic • Detectors Available: Flame Ionization
Gas Chromatograph gas control Detector, Thermal Conductivity Detector,
The TRACE™ 1300 Series GC consists of Electron Capture Detector, Nitrogen
• Pressure Set Points Minimum Increments:
two models designed to meet the specific Phosphorus Detector, full range of Thermo
0.01 kPa-0.001 psi in all ranges
needs of all laboratories. The TRACE Scientific mass spectrometers
Carrier Gas Control Common to all • Available as Instant Connect, user-
1310 GC features a complete icon-driven
Injectors exchangeable modules
touch-screen user interface ideal for direct
• Split Ratio: Up to 12500:1
instrument control in larger routine and • Maximum number Installed: 3 including a
method development laboratories. The • Pressure Range: 0–1000 kPa (0–145 PSI) mass spectrometers on the same oven
TRACE 1300 GC is the budget-conscious • Modes: Constant and programmed • Fast Data Acquisition Rate: up to 300 Hz
investment for the basic routine laboratory pressures and flows for FID, TCD, ECD, and NPD
looking for an intuitive single-button system, • Total Flow Setting:
that provides ease of use with minimal Instant Connect Flame Ionization
– Control of split flow in 1 mL/min from 0 Detector
instrument interaction. Both instruments
to 1250 mL/min • Capillary column optimized compatible
offer the same user-exchangeable, instant
connect injector and detector modules and – Purge flow from 0 to 50 mL/min with 1/8” and 1/16” packed column
fast oven performance with exceptional • Flameout detection and automatic re-
Optional Instant Connect Auxiliary Gas
retention time stability to reach an incredibly ignition
Module
high lab productivity at reduced cost of • Allows for the control of three additional • MDL: <1.8 pg C/s
ownership. gas channels • Sensitivity: >0.03 Coulombs/gC
Performance Specifications Injectors • Linear Dynamic Range: >107 (±10%)
• Typical Retention Time Repeatability: • Maximum Number Injectors Installed: 2 • Maximum Temperature: 450 °C in steps
<0.0008 min
• Available as Instant Connect, user- of 0.1 °C
• Typical Peak Area Repeatability: <0.5 % exchangeable modules • IEC:
RSD
Instant Connect Split/Splitless – Air: 0–500 mL/min in 0.1 steps
Oven Specifications • Suitable for all capillary columns
• Column Oven (H × W × D): 27 × 27 × – H2: 0–100 mL/min in 0.1 steps
(50 μm to530 μm i.d.)
17.7 cm; 12.9 L – Makeup gas (N2 or He) 0–50 mL/min in
• Compatible with 1/8” and 1/16” packed 0.1 steps
• Operating Temperature Range: ambient column using adapters. Supports P&T/TD/
+3 °C to 450 °C Instant Connect Thermal Conductivity
HS by special adapter. Compatible Merlin
• Cryogenic Option Minimum Temperature: Detector
Microseal™ septum.
-100 °C with liquid Nitrogen; -50 with • Capillary column optimized (micro TCD)
• Dedicated split/splitless injector with compatible with 1/8” and 1/16” packed
liquid CO2 integrated backflush capabilities column
• Temperature Set Point Resolution: 0.1 °C • Maximum Temperature: 400 °C • Maximum Temperature: 400 °C in steps
• Number of Ramps/Plateaus: 32/33
Instant Connect Programmable of 0.1 °C
• Maximum Heating Rate: 125 °C/min Temperature Vaporizer • MDL: <400 pg tridecane/mL with He
• Supports hot/cold split and splitless carrier or <20 pg tridecane/s with a total
T Range °C H
eating Rate °C/min modes as well as large volume injections flow through the cell of 3 mL/min
(solvent split) and On Column (TPOC). • Linear Dynamic Range: 105
Model: 220 Volts Model: 110 Volts
Compatible Merlin Microseal septum.
50 to 70 125 90
• Dedicated PTV injector with integrated
70 to 115 100 65 backflush capabilities
115 to 175 80 50 • Temperature Range: Air forced cooling to
175 to 300 50 30 ambient +5 °C up to 450 °C
300 to 450 35 20 • Cryogenic Option Minimum Temperature:
• Oven Cool-Down (22 °C ambient): 450 °C -100 °C with liquid nitrogen; -50 °C with
to 50 °C in <4 minutes CO2
washing cycle during the current GC Syringe Cleaning l Wash stations for up to Incubation Oven Temperature Range
cooling phase. 4 different solvents for a total of 40 mL 40 to 200 °C in 1 °C steps, with agitation
Barcode Reader l Two active laser (standard). Optional large washing stations Incubation Time l 0.1 to 600.0 min in 0.1
scanners for all standard vials using for up to 3 × 100 mL solvent bottles. min increments
1-dimension barcodes in a horizontal Waste: 1 × 10 mL or drain to external
Syringe Flush Capability l With inert gas
orientation. waste bottle. Possibility to install multiple
solvent stations to expand solvent and Solvent Syringe Cleaning l Optional
Vortexer l Intensive mixing of 0.5, 0.7, 2, 5, washing stations (4 × 10 mL or 2 × 100
waste volumes.
10, or 20 mL vials with an agitation speed mL vials)
up to 2000 rpm Options
• Upgrade to headspace option and its own Multiple Headspace Extraction
Vial Trays l Heated and cooled trays expand Yes (optional accessory)
accessories
the range of available applications from
• Upgrade to SPME option (with or without Enrichment Sampling l Yes with optional
sample injection to sample/standard
headspace option) kit for cold trap
preparation
• Peltier-controlled drawer for well plates, Injection Speed l 1 to 100 mL/min, in 1
GC Mounting Kits l Thermo Scientific
300 μL fixed insert vials, 2 and 10 mL mL/min increments
TRACE 1300 GC, TRACE 1310 GC,
TRACE GC Ultra and FOCUS GC vials. Temperature selectable between 0 Typical Performance l ≤0.7 RSD % under
and 40 °C Thermo Scientific standard conditions
Configuration for Liquid Sampling
• Cooled trayholders for well plates, 300 μL Needle Lengths l 65 mm, compatible with
Vial Volumes l 300 μL fixed insert vials,
fixed insert vials, 2, 10 and 20 mL vials. every injector port
0.5, 0.7, 2, 2.5, 10, and 20 mL. 96/384
Temperature selectable between 4 and 70
Microtiter or Deep Well plates with Options
°C. Requires external circulator bath.
Automatic Foil Cutter to pierce alumina or • Upgrade to Liquid option
plastic foils prior the needle penetration • Solvent/reagent reservoir
• Upgrade to SPME option
Bottom-Sensing Vials l Capable of liquid • Large Volume Wash Station
• Peltier-cooled trayholder for 300 μL
injection starting from small-volume Injection Volume l Range from 0.1 to fixed insert vials, 2 and 10 mL vials;
samples. Capability to inject from samples 10,000 μL in 0.1 μL steps to 100 μL and temperature selectable between 0 °C and
as low as 5 µL into a vial. Possibility of 1 μL steps between 100 μL and 10 mL. 40 °C
performing up to three 1 µL injections Needle lengths: 57 mm or 85 mm Liquid
• Cooled trayholders for 300 μL fixed insert
from a 5 µL sample, depending on vial Injection Modes – 8 fully customizable
vials, 2, 10 and 20 mL vials; requires
type. method-specific preset menus available:
external circulator bath; temperature
Height from Vial Bottom l User selectable • Basic enrichment selectable between 4 °C and 70 °C
between 0.1 and 32 mm in 0.1 mm steps • Enrichment needle solvent wash • SPME Fiber Cleaning Station
Injection Speed for Liquid Samples • Internal standard double • Solvent/reagent reservoir
Selectable from 0.1 μL/sec up to 2000
• Internal standard post • Large Volume Wash Station
μL/sec and fully programmable
• Needle solvent wash
Sample Capacity (Depending Solid Phase Micro Extraction Option
• Solvent flush double Vial Volumes l 2, 10 and 20 mL
on autosampler, GC and MS
configurations) • Solvent flush post Sample Capacity l Depending on
• Up to 4608 well plates or 6912 well plates Typical Performance – ≤0.3 RSD % Autosampler, GC and MS configurations:
with extended X-arm obtained under Thermo Scientific standard • Up to 840 0.5/0.7 mL vials or 1260
• Up to 840 0.5/0.7 mL vials or 1260 conditions 0.5/0.7 mL vials with the extended X-arm
0.5/0.7 mL vials with the extended X-arm Configuration for Headspace • Up to 648 2 mL sample vials or 972 2 mL
• Up to 648 2 mL sample vials or 972 2 mL Vial Volumes l 2, 10 and 20 mL vials with the extended X-arm
vials with the extended X-arm Syringe Sizes l 1, 2.5, and 5 mL • Up to 180 10 or 20 mL vials or 300 10 or
• Up to 240 10 mL or 20 mL vials or 360 Sample Capacity (Depending 20 mL vials with the extended X-arm
10 mL or 20 mL vials with the extended on autosampler, GC and MS Incubation Oven Capacity l 6 vials
X-arm configurations) l Up to 180 10 or 20 mL Incubation Oven Temperature Range
Syringes (Capable of handling liquid vials or 300 10 or 20 mL vials with the 40 °C to 200 °C in 1 °C steps, with
volumes in the range 0.1 μL – 10 mL) extended X-arm agitation
Capable of using 0.5 μL, 1.0 μL, 5 μL, 10 Injection Volume Range l 0.1 to 5 mL in Vial Penetration Depth l Standard or
μL (standard), 25 μL, 50 μL, 100 μL, 250 0.1 mL steps depending on syringe custom between 5 mm and 20 mm,
μL, 500 μL, 1000 μL, 10000 μL syringes Syringe Temperature l OFF or 40 °C to suitable for liquid or headspace extraction
for sample injection and/or volume 150 °C in 1 °C steps Fiber Conditioning Station l Optional,
transfer
Incubation Oven Capacity l 6 vials 2-ports and a needle guide, 40-350 °C,
inert gas purged. Fiber: 23 gauge. One
fiber holder to automate all fibers.
Installation Requirements Environment
Product Sp e ci f i ca t i ons
Detailed installation requirements are Compressed Air Room Temperature
provided in the DFS Preinstallation Compressed air with a pressure of 6 bar Laboratory room temperature must be
Requirements Guide PN/1194630 (87 psi) is required to operate the pneumatic maintained between 15 °C and 26 °C
valves of the instrument. A suitable (59 °F and 79 °F). The optimum operation
Recirculating Water Chiller
compressor can be ordered from Thermo temperature is between 18 °C and 21 °C
The DFS mass spectrometer can be
Fisher Scientific (Part No. 026 1850). (65 °F and 70 °F).
configured with the optional Thermo
Scientific NESLAB ThermoFlex TF25 Space Requirement Air Conditioning Load
recirculating chiller. The chiller is compact, The average power dissipation during
quiet and efficient. The ThermoFlex TF25 DFS Mass Spectrometer analysis operation for a basic DFS system,
is available in both air-cooled and water • HxD 140.6 cm (55.4”) x 156.8 cm (62.5”) including gas chromatograph and data
cooled configurations, for a quiet laboratory system, is approximately 4.7 kW (4.5 BTU/s).
• Width with one GC installed
for stable and precise cooling of magnet For dual GC configuration, the average air
and turbo molecular pumps. The cooler is 161.1 cm (63.5”) conditioning load amounts approximately
controlled by a temperature sensor mounted • Width with two GCs installed to 4.9 kW (4.7 BTU/s).
on the DFS magnet. 209.1 cm (82.4”) Humidity
• Precise temperature control and • MS console with analyzer and magnet The relative humidity of the operating
temperature stability of up to ± 0.1 °C 900 kg (2000 lb) environment must be between 30% and
• Adjustable high and low temperature • Data system 130 kg (< 290 lb) 70%, with no condensation.
safeties with audible alarm • GC with console 35 kg (78 lb) Recommended Instrument Clearences
• Environmentally-friendly CFC-free air- and Weight Distribution
TRACE 1310 GC
cooled refrigeration system ThermoFlex
• Dimensions (H × W × D): 45 × 44 × 67
TF25 - 79.6 Kg. 44 x 67 x 73 cm
cm
Supplies • Weight (Kg): 35 Kg main unit plus 0.8 Kg
each module (78 lb)
Power • Column Oven (H × W × D): 27 × 27 ×
The DFS mass spectrometer is designed to 17.7 cm; 12.9 L
operate at a nominal voltage of 230 V AC,
50/60 Hz. The minimum and maximum TriPlus RSH Autosampler
voltage tolerances are in compliance with Overall dimensions on the GC:
IEC 950, Amend 2, 1993. Approximate • Length (X axis) 80 cm
consumption values for regular operation are: The extended extended model length
• 0,9 kW for GC operation is 120 cm
thermoscientific.com/dfs
©2005-2012 Thermo Fisher Scientific Inc. All rights reserved. Optiplex is a trademark of Dell Inc. Intel and Pentium are registered trademarks of Intel
Corporation. NIST is a trademark of The National Institute of Standards and Technology (NIST) an agency of the U.S. Commerce Department. Wiley is a
Trademark of Wiley - John Wiley & Sons, Inc.. Pfleger-Maurer library is a product of Wiley. Swagelok is a trademark of Swagelok Company. Microsoft Office and
Windows 7 are trademarks of Microsoft Coroporation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications,
terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Africa-Other +27 11 570 1840 Finland /Norway/Sweden Middle East +43 1 333 50 34 0
Australia +61 2 8844 9500 +46 8 556 468 00 Netherlands +31 76 579 55 55
Austria +43 1 333 50 34 0 France +33 1 60 92 48 00 South Africa +27 11 570 1840
Belgium +32 53 73 42 41 Germany +49 6103 408 1014 Spain +34 914 845 965
Canada +1 800 530 8447 India +91 22 6742 9434 Switzerland +41 61 716 77 00
China +86 10 8419 3588 Italy +39 02 950 591 UK +44 1442 233555
Denmark +45 70 23 62 60 Japan +81 45 453 9100 USA +1 800 532 4752
Europe-Other +43 1 333 50 34 0 Latin America +1 608 276 5659